VEGA SON64 User Manual

Ope
rating Instructions
VEGASON 64
Foundation F
ieldbus
Ultraso
nic

Contents

Conten
1 About this document
2 For your safety
3 Product description
4 Mounting
5 Connecting to power supply
1.1 Function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.2 Target group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.3 Symbolism used . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.1 Authorised personnel . . . . . . . . . . . . . . . . . . . . . . . .
2.2 Appropriate use . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.3 Warning about misuse . . . . . . . . . . . . . . . . . . . . . . .
2.4 General safety instructions . . . . . . . . . . . . . . . . . . . .
2.5 Safety approval markings and safety tips . . . . . . . . . .
2.6 CE conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.7 Fulllment of NAMUR recommendations . . . . . . . . . .
2.8 Safety instructions for Ex areas . . . . . . . . . . . . . . . . .
2.9 Environmental instructions. . . . . . . . . . . . . . . . . . . . .
3.1 Conguration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.2 Principle of operation . . . . . . . . . . . . . . . . . . . . . . . .
3.3 Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.4 Packaging, transport and storage . . . . . . . . . . . . . . .
4.1 General instructions . . . . . . . . . . . . . . . . . . . . . . . . .
4.2 Mounting preparations . . . . . . . . . . . . . . . . . . . . . . .
4.3 Mounting instructions . . . . . . . . . . . . . . . . . . . . . . . .
5.1 Preparing the connection . . . . . . . . . . . . . . . . . . . ..
5.2 Connection procedure. . . . . . . . . . . . . . . . . . . . . . . .
5.3 Wiring plan, double chamber housing . . . . . . . . . . . .
5.4 Switch on phase. . . . . . . . . . . . . . . . . . . . . . . . . . . .
4 4 4
5 5 5 5 6 6 6 6 6
8
9 10 10
11 13 18
23 24 26 27
6 Set up with the indicating and adjustment module P LICSCOM
6.1 Short description . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.2 Insert indicating and adjustment module. . . . . . . . . . .
6.3 Adjustment system . . . . . . . . . . . . . . . . . . . . . . . . . .
6.4 Setup procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.5 Menu schematic . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.6 Saving the parameter adjustment data . . . . . . . . . . . .
7 Setup with PACTware and other adjustment programs
7.1 Connect the PC via VEGACONNECT . . . . . . . . . . . .
7.2 Parameter adjustment with PACTware . . . . . . . . . . ..
7.3 Parameter adjustment with AMS™ . . . . . . . . . . . . . .
7.4 Saving the parameter adjustment data . . . . . . . . . . . .
2 VE
29 29 31 32 37 39
40 41 41 41
GASON 64 Foundation Fieldbus
28793-EN-081127
Contents
8 Maint
enance and fault rectication
8.1 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8.2 Remove interferences . . . . . . . . . . . . . . . . . . . . . . . .
8.3 Exchanging the electronics module . . . . . . . . . . . . . .
8.4 Software update . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8.5 Instrument repair . . . . . . . . . . . . . . . . . . . . . . . . . . .
9 Dismounting
9.1 Dismounting steps . . . . . . . . . . . . . . . . . . . . . . . . . .
9.2 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
10 Supplement
10.1 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
10.2 Foundation Fieldbus . . . . . . . . . . . . . . . . . . . . . . . . .
10.3 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
10.4 Industrial property rights . . . . . . . . . . . . . . . . . . . . . .
10.5 Trademark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
42 42 43 44 44
46 46
48 53 58 61 61
28793-EN-081127
VEGASON 64 Foun
Supplementary documentation
tion:
Informa
Supplementary documents appropriate to the ordered version come with the delivery. You can nd them listed in chapter "Product
description".
Instructions manuals for accessories and replacement parts
p:
Ti
To ensure reliable setup and operation of your VEGASON 64, we oer accessories and replacement parts. The associated documents are:
l 27835 - Indicating and adjustment module PLICSCOM l 32628 - Interface adapter VEGACONNECT l 27720 - External indication VEGADIS 61 l 34296 - Protective cover l 30176 - Electronics module VEGASON series 60 l 30205 - Emitting electronics VEGASON 64, 65, 66
dation Fieldbus 3
1 Abou
t this document
1 Abou
t this document

1.1 Function

This operating instructions manual provides all the information you need for mounting, connection and setup as well as important instructions for maintenance and fault rectication. Please read this information before putting the instrument into operation and keep this manual accessible in the immediate vicinity of the device.

1.2 Target group

This operating instructions manual is directed to trained qualied personnel. The contents of this manual should be made available to these personnel and put into practice by them.

1.3 Symbolism used

Inform
ation, tip, note
This symbol indicates helpful additional information.
Cauti
on: If this warning is ignored, faults or malfunctions can
result. Warning: If this warning is ignored, injury to persons and/or serious damage to the instrument can result. Danger: If this warning is ignored, serious injury to persons and/or destruction of the instrument can result.
applications
Ex
This symbol indicates special instructions for Ex applications.
l List
The dot set in front indicates a list with no implied sequence.
à Action
Th
is arrow indicates a single action.
1 Sequence
Numbers set in front indicate successive steps in a procedure.
4 VE
28793-EN-081127
GASON 64 Foundation Fieldbus
your safety
2 For
or your safety
2 F

2.1 Authorised personnel

All operations described in this operating instructions manual must be carried out only by trained specialist personnel authorised by the plant operator.
During work on and with the device the required personal protective equipment must always be worn.

2.2 Appropriate use

VEGASON 64 is a sensor for continuous level measurement.
You can nd detailed information on the application range in chapter
"Product description".
Operational reliability is ensured only if the instrument is properly used according to the specications in the operating instructions manual as well as possible supplementary instructions.
For safety and warranty reasons, any invasive work on t he device beyond that described in the operating instructions manual may be carried out only by personnel authorised by the manufacturer. Arbitrary conversions or modications are explicitly forbidden.

2.3 Warning about misuse

Inappropriate or incorrect use of the instrument can give rise to application-specic hazards, e.g. vessel overll or damage to system components through incorrect mounting or adjustment.
28793-EN-081127
VEGASON 64 Foun

2.4 General safety instructions

This is a high-tech instrument requiring the strict observance of standard regulations and guidelines. The user must take note of the safety instructions in this operating instructions manual, the country­specic installation standards as well as all prevailing safety regulations and accident prevention rules.
The instrument must only be operated in a technically awless and reliable condition. The operator is responsible for trouble-free operation of the instrument.
During the entire duration of use, the user is obliged to determine the compliance of the required occupational safety measures with the current valid rules and regulations and also take note of new regulations.
dation Fieldbus 5
2 For
your safety
2.5 Safety
The safety approval markings and safety tips on the device must be observed.
approval markings and safety tips

2.6 CE conformity

This device fullls the legal requirements of the applicable EC guidelines. By attaching the CE mark, VEGA provides a conrmation of successful testing. You can nd the CE conformity declaration in the download area of
2.7 Fulllmen
With respect to compatibility, the NAMUR recommendation NE 53 is fullled. This applies also to the corresponding indicating and adjustment components. VEGA instruments are generally upward and downward compatible.
l Sensor software for DTM VEGASON 64 HART, PA or FF l DTM VEGASON 64 for adjustment software PACTware l Indicating and adjustment module for sensor software
The parameter adjustment of the basic sensor functions is independ­ent of the software version. The range of available functions depends on the respective software version of the individual components.
The software version of VEGASON 64 can be determined as follows:
l via PACTware l on the type label of the electronics l via the indicating and adjustment module
You can view all software histories on our website
e use of this advantage and get registered for update information
Mak via e-mail.
www.vega.com.
t of NAMUR recommendations
www.vega.com.

2.8 Safety instructions for Ex areas

Please note the Ex-specic safety information for installation and operation in Ex areas. These safety instructions are part of the operating instructions manual and come with the Ex-approved instruments.

2.9 Environmental instructions

Protection of the environment is one of our most important duties. That is why we have introduced an environment management system with the goal of continuously improving company environmental protection.
The environment management system is certied according to DIN EN ISO 14001.
Please help us full this obligation by observing the environmental
instructions in this manual:
6 VE
28793-EN-081127
GASON 64 Foundation Fieldbus
2 For
your safety
l Chapte l Chapter "Disposal"
r "Packaging, transport and storage"
28793-EN-081127
VEGASON 64 Foun
dation Fieldbus 7
1
2
3
4
3 Produc
t description
Scope of delivery
Components
3 Produc
t description
3.1 Conguration
The scope of delivery encompasses:
l VEGASON 64 ultrasonic sensor l Unassembled cable gland l Documentation
- this operating instructions manual
- Operating instructions manual 27835 "Indicating and adjust-
ment module PLICSCOM" (optional)
- Ex-specic "Safety instructions" (with Ex-versions)
- if necessary, further certicates
VEGASON 64 consists of the following components:
l transducer with process tting (depending on version ange,
swivelling holder or thread)
l housing with electronics (depending on version, separate) l Housing cover, optionally available with indicating and adjustment
module PLICSCOM
The components are available in dierent versions.
l Version A: compact in ange version l Version B: compact with swivelling holder l Version C: separately with swivelling holder l Version D: separately with thread
with swivelling holder
Type label
Fig. 1: VEGASON 64 compact 1 Housing cover with integrated PLICSCOM (optional) 2 Housing with electronics 3 Swivelling holder with ange 4 Transducer
The type label contains the most important data for identication and use of the instrument:
l Sensor type l Article and serial number device l Article numbers documentation
8 VE
28793-EN-081127
GASON 64 Foundation Fieldbus
3 Produc
hnical data: For example approvals, process temperature,
l Tec
process tting/material, signal output, power supply, protection
With the serial number, you can access the delivery data of the instrument via search". In addition to the type label outside, you can also nd the serial number on the inside of the instrument.
www.vega.com, "VEGA Tools" and "serial number
t description

3.2 Principle of operation

Application area
Functional principle
Power supply and bus
communication
VEGASON 64 is an ultrasonic sensor for continuous level measure- ment. It is particularly suitable for solids, but also good for liquids.
The transducer of the ultrasonic sensor transmits short ultrasonic pulses to the measured product. These pulses are reected by product surface and received again by the transducer as echoes. The running time of the ultrasonic pulses from emission to reception is proportional to the distance and hence the level. The determined level is converted into an appropriate output signal and outputted as measured value.
Four-wire electronics with separate power supply.
Power for the signal output is supplied via the H1 Fieldbus. A two-wire
cable according to Fieldbus specication serves as carrier of both power and digital data transmission for multiple sensors. This cable can be operated in two versions:
l via an H1 interface card in the control system and additional power
supply
l via a Linking device
with HSE (High speed Ethernet) and addition-
al power supply ac-
DD/CFF
28793-EN-081127
VEGASON 64 Foun
cording to IEC 61158-2
The DD (Device Descriptions) and CFF (capability les) necessary for
planning and conguration of your FF (Foundation Fieldbus) commu­nication network are available in the download area of the VEGA homepage Foundation Fieldbus". The appropriate certicates are also available there. A CD with the appropriate les and certicates can be ordered via e-mail under info@de.vega.com or by phone from one of the VEGA agencies under the order number "DRIVER.S".
dation Fieldbus 9
www.vega.com under "Services - Downloads - Software -
3 Produc
t description
Packaging
Transport
Transport inspection
3.3 Operatio
n
VEGASON 64 can be adjusted with dierent adjustment media:
l with indicating and adjustment module l with the suitable VEGA DTM in conjunction with an adjustment
software according to the FDT/DTM standard, e.g. PACTware and PC
l a conguration tool
The entered parameters are generally saved in VEGASON 64, optionally also in the indicating and adjustment module or in PACTware.

3.4 Packaging, transport and storage

Your instrument was protected by packaging during transport. Its capacity to handle normal loads during transport is assured by a test according to DIN EN 24180.
The packaging of standard instruments consists of environment­friendly, recyclable cardboard. For special versions, PE foam or PE foil is also used. Dispose of the packaging material via specialised recycling companies.
Transport must be carried out under consideration of the notes on the transport packaging. Nonobservance of these instructions can cause damage to the device.
The delivery must be checked for completeness and possible transit damage immediately at receipt. Ascertained transit damage or concealed defects must be appropriately dealt with.
Storage
Up to the time of installation, the packages must be left closed and stored according to the orientation and storage markings on the outside.
Unless otherwise indicated, the packages must be stored only under the following conditions:
l Not in the open l Dry and dust free l Not exposed to corrosive media l Protected against solar radiation l Avoiding mechanical shock and vibration
Storage and transport temperature
l Storage and transport temperature see chapter "Supplement -
Technical data - Ambient conditions"
l Relative humidity 20 85 %
10 VE
28793-EN-081127
GASON 64 Foundation Fieldbus
4 M
ounting
Suitability for process conditions
Mounting position
Moisture
4 Moun
ting

4.1 General instructions

Make sure that all parts of the instrument in contact with the measured product, especially the sensor element, process seal and process tting, are suitable for the existing process conditions such as process pressure, process temperature as well as the chemical properties of the medium.
You can nd the specications in chapter "Technical data" in the or on the type label.
Select an installation position you can easily reach for mounting and connecting as well as later retrotting of an indicating and adjustment module. The housing can be rotated by 330° without the use of any tools. You can also install the indicating and adjustment module in four dierent positions (each displaced by 90°).
Use the recommended cables (see chapter "Connecting to power supply") and tighten the cable gland.
You can give your VEGASON 64 additional protection against moisture penetration by leading the connection cable downward in front of the cable entry. Rain and condensation water can thus drain off. This applies mainly to outdoor mounting as well as installation in areas where high humidity is expected (e.g. through cleaning processes) or on cooled or heated vessels.
28793-EN-081127
VEGASON 64 Foun
Fig. 2: Measures
dation Fieldbus 11
against moisture penetration
1 32
100%
0%
4
ounting
4 M
Measuring range
e reference plane for the measuring range (version A) is the lower
Th
edge of the ange. The reference plane for versions B, C and D is the lower edge of the transducer.
Remember that with all versions there is a minimum distance below the reference plane - the so-called dead band - in which measurement is not possible; this distance from reference plane to medium must not be underrun. The exact value of the dead band is stated in chapter
"Technical data".
Informa
tion:
If the medium reaches the transducer, buildup can form on it and cause faulty measurements later on.
Fig. 3: Version A - Measuring distance
1 full (dead band) 2 empty (max. measuring distance) 3 Measuring range 4 Reference plane
range (operating range) and max. measuring
12 VE
28793-EN-081127
GASON 64 Foundation Fieldbus
1 32
100%
0%
4
4 M
ounting
Pressure/Vacuum
Mounting - Version B
Fig. 4: Version B, C, D - Measuring distance
1 full (dead band) 2 empty (max. measuring distance) 3 Measuring range 4 Reference plane
range (operating range) and max. measuring
Gauge pressure in the vessel does not inuence VEGASON 64. Low pressure or vacuum does, however, damp the ultrasonic pulses. This inuences the measuring result, particularly if the level is very low. With pressures under -0.2 bar (-20 kPa) you should use a dierent measuring principle, e.g. radar or guided microwave.

4.2 Mounting preparations

ning:
War
Mounting preparations must be carried out in idle condition. Non­observance can cause damages on the electronics!
Version B is supplied in two parts:
l Transducer l Electronics housing
Assemble the version as follows:
1 Loosen the screws on the xing ring (4) with a hexagon spanner
(size 4), remove the transducer tube out of the swivelling holder
2 Mount the ange
28793-EN-081127
VEGASON 64 Foun
dation Fieldbus 13
1
2
3
4
5
4 M
ounting
3 Insert
the transducer tube in the requested length from below into
the swivelling holder
4 Fasten with screws (4); torque max. 10 Nm 5 Remove the plug from below out of the instrument housing and
plug it into the socket of the transducer tube
6 Plug the electronics housing to the transducer tube. Do not
squeeze any cables. The housing is in the correct position, if the lower notch (2) of the transducer tube is no longer visible
7 Tighten the screws (5) on the housing moderately with an Allen
wrench (size 4)
8 Loosen terminal screw of the swivelling holder (3) with a fork
spanner (SW 13)
9 Direct the sensor with the swivelling holder to the measured
product
10 Fasten the swivelling holder with screw (3), torque max. 20 Nm 11 Secure screw (3) with lacquer or similar to ensure permanent
tightness of the swivelling holder
12 Rotate the housing in such a way that the cable entry points
downwards (moisture can drain off). Fasten the housing with screws (5); torque Aluminium housing max. 5 Nm, torque plastic housing max. 2 Nm
Fig. 5: Mounting - Version B 1 Groove 2 Notch 3 Screw of swivelling holder (hexagon SW 13) 4 Grub screw, xing ring (hexagon spanner - size 4) 5 Terminal screws, housing (hexagon spanner - size 4)
14 VE
to lock the housing
28793-EN-081127
GASON 64 Foundation Fieldbus
1
2
3
4
5
4 M
ounting
Mounting - Version C
ion C is supplied in three parts:
Vers
l Transducer l Transducer cable l electronics for wall mounting
Assemble the version as follows:
1 Loosen the screws on the xing ring (4) with a hexagon spanner
(size 4), remove the transducer tube out of the swivelling holder
2 Mount the ange 3 Insert the transducer tube in the requested length from below into
the swivelling holder
4 Fasten with screws (4); torque max. 10 Nm 5 Remove the plug from below out of the connection head and plug
it into the socket of the transducer tube
6 Plug the connection head to the transducer tube. Do not squeeze
any cables. The cylinder is in the correct position, if the lower notch (2) of the transducer tube is no longer visible
28793-EN-081127
VEGASON 64 Foun
Fig. 6: Mounting - Version C 1 Groove 2 Notch 3 Screw of swivelling holder (hexagon SW 13) 4 Grub screw, xing ring (hexagon spanner - size 4) 5 Terminal screws, connection head (hexagon spanner - size 4)
for locking the connection head
7 Tighten the screws (5) on the cylinder moderately with an Allen
wrench (size 4)
8 Loosen terminal screw of the swivelling holder (3) with a fork
spanner (SW 13)
9 Direct the sensor with the swivelling holder to the measured
product
10 Fasten the swivelling holder with screw (3), torque max. 20 Nm
dation Fieldbus 15
1
2 3
4 M
ounting
Mounting - Version D
11 Secur
e screw (3) with lacquer or similar to ensure permanent
tightness of the swivelling holder
12 Rotate the connection head in such a way that the cable entry
points downwards (moisture can drain off). Fasten the cylinder with screws (5); torque max. 5 Nm
13 Insert the plug at the end of the transducer cable into the socket on
the remote electronics
Fig. 7: Plug 1 Socket electronics housing 2 Plug connector 3 Connection piece, transducer tube
connector between transducer and socket of electronics housing
Version D is supplied in three parts:
l Transducer l Transducer cable l electronics for wall mounting
Assemble the version as follows:
1 Loosen the hexagon nut (3) on the transducer cable 2 Insert the transducer tube from below into the mounting opening
G1A 3 Fasten (SW 46) the hexagon nut (3) 4 Remove the plug from below out of the connection head and plug
it into the socket of the transducer tube
16 VE
28793-EN-081127
GASON 64 Foundation Fieldbus
1
2
3
4
1
2 3
4 M
ounting
5 Plug
the connection head to the transducer tube. Do not squeeze
any cables. The cylinder is in the correct position, if the lower notch (2) of the transducer tube is no longer visible
Fig. 8: Mounting - Version D
1 Groove
2 Notch
3 Hexagon nut SW 46
4 Terminal screws, connection head (hexagon spanner - size 4)
for locking the connection head
6 Tighten the screws (4) on the cylinder moderately with an Allen
wrench (size 4)
7 Rotate the connection head in such a way that the cable entry
points downwards (moisture can drain off). Fasten the cylinder with screws (4); torque max. 5 Nm
8 Insert the plug at the end of the transducer cable into the socket on
the remote electronics
28793-EN-081127
VEGASON 64 Foun
Fig. 9: Plug
1 Socket electronics housing
2 Plug connector
3 Connection piece, transducer tube
connector between transducer and socket of electronics housing
dation Fieldbus 17
1
> 500 mm
4 M
ounting
Mounting position
4.3 Mount
When mounting VEGASON 64, keep a distance of at least 500 mm (19.69 in) to the vessel wall. If the sensor is installed in the center of
dished or round vessel tops, multiple echoes can arise. These can, however, be suppressed by an appropriate adjustment (see chapter
"Setup").
If you cannot keep this distance you should carry out a false echo storage before setup. This applies mainly if buildup on the vessel wall is expected. In this case, we recommend repeating a false echo storage later with existing buildup.
ing instructions
Fig. 10: Mounting 1 Reference plane
position
18 VE
28793-EN-081127
GASON 64 Foundation Fieldbus
ounting
4 M
vessels with conical bottom it can be advantageous to mount the
In
sensor in the center of the vessel, as measurement is then possible
down to the lowest point of the vessel bottom.
Socket
Sensor orientation
Vessel installations
Fig. 11: Vessel
with conical bottom
Preferably the transducer should be mounted without socket ush to
the vessel top.
If the reective properties of the medium are good, you can mount
VEGASON 64 on a socket piece. The socket end should be smooth
and burr-free, if possible also rounded. Carry out a false echo storage.
Align the sensor in liquids as vertical as possible to the product surface
to achieve optimum measuring results.
Fig. 12: Alignment
in liquids
To optimally align the sensor to solid products, a swivelling holder can
be implemented.
The ultrasonic sensor should be installed at a location where no
installations cross the ultrasonic beam.
28793-EN-081127
VEGASON 64 Foun
dation Fieldbus 19
α
α
α
α
α
α
4 M
ounting
l installations such as for example, ladders, limit switches,
Vesse heating spirals, struts etc. can cause false echoes that interfere with the useful echo. Make sure when planning your measuring site that the ultrasonic signals have a "clear view" to the measured product.
In case of existing vessel installations, a false echo storage should be carried out during setup.
If large vessel installations such as struts or supports cause false echoes, these can be attenuated through supplementary measures. Small, inclined sheet metal or plastic baes above the installations scatter the ultrasonic signals and avoid direct false echoes.
Agitato
Fig. 13: Cover
rs
If there are agitators in the vessel, a false echo storage should be
smooth proles with deectors
carried out with the agitators in motion. This ensures that the interfering reections from the agitators are saved with t he blades in dierent positions.
Fig. 14: Agitators
20 VE
28793-EN-081127
GASON 64 Foundation Fieldbus
Material heaps
ounting
4 M
material heaps are best measured with several instruments,
Large
which can be mounted on e.g. traverse cranes. For this type of
application it is advantageous to orient the sensor perpendicular to the
bulk solid surface.
Inowing medium
Foam
Fig. 15: Transducers
on traverse crane
Do not mount the instruments in or above the lling stream. Make sure
that you detect the product surface, not the inowing product.
Fig. 16: Inowing
medium
Through the action of lling, stirring and other processes in the vessel,
dense foams which considerably damp the emitted signals may form
on the product surface.
If foams are causing measurement errors, the sensor should be used
in a standpipe or, alternatively, the more suitable guided radar sensors
(TDR) should be used.
Guided radar is unaected by foam generation and is particularly
suitable for such applications.
28793-EN-081127
VEGASON 64 Foun
dation Fieldbus 21
ounting
4 M
Air
turbulences
there are strong air currents in the vessel, e.g. due to strong winds in
If outdoor installations or air turbulence, e.g. by cyclone extraction you should mount VEGASON 64 in a standpipe or use a dierent measuring principle, e.g. radar or guided radar (TDR).
Heat uctuations
Strong heat uctuations, e.g. caused by the sun, can cause measurement errors. In this case, you should use a sun shield.
Fig. 17: Heat uctuations
22 VE
28793-EN-081127
GASON 64 Foundation Fieldbus
ecting to power supply
5 Conn
Note safety instructions
Take note of
safety instruc­tions for Ex ap­plications
Select power supply
5 Conn
ecting to power supply

5.1 Preparing the connection

Always keep in mind the following safety instructions:
l Connect only in the complete absence of line voltage
l If overvoltage surges are expected, overvoltage arresters should
be installed according to Foundation Fieldbus specication
Ti
p:
We recommend VEGA overvoltage arrester B63-32.
In
hazardous areas you should take note of the appropriate
regulations, conformity and type approval certicates of the sensors
and power supply u nits.
Supply voltage and current output are carried on separate two-wire
connection cables if reliable separation is required. The supply voltage
range can dier depending on the instrument version.
The data for power supply are specied in chapter "Technical data".
This instrument is designed in protection class I. To maintain this
protection class, it is absolutely necessary that the ground conductor
be connected to the internal ground terminal. Take note of the general
installation regulations.
As a rule, connect VEGASON 64 to vessel ground (potential
euqalisation) or in case of plastic vessels to the next ground potential.
For this purpose, there is a ground terminal on the side of the
instrument housing.
Selecting connection cable
Cable screening and grounding
28793-EN-081127
VEGASON 64 Foun
For power supply, an approved installation cable with PE conductor is
necessary.
The signal output is connected with screened cable according to
Foundation Fieldbus specication.
Use cable with round cross-section. A cable outer diameter of 5 … 9 mm
(0.2 0.35 in) ensures the seal eect of the cable gland. If you are
using cable with a dierent diameter or cross-section, exchange the
seal or use a suitable cable gland.
In systems with p otential equalisation, connect the cable screen
directly to ground potential at the power supply unit, in the connection
box and at the sensor. The screen in the sensor must be connected
directly to the internal ground terminal. The ground terminal outside on
the housing must be connected to the potential equalisation (low
impedance).
dation Fieldbus 23
1
2 3
ecting to power supply
5 Conn
systems without potential equalisation, connect the cable screen
In directly to ground potential at the power supply unit and at the sensor. In the connection box or T-distributor, the screen of the short stub to the sensor must not be connected to ground potential or to another cable screen. The cable screens to the power supply unit and to the next distributor must be connected to each other and also connected to ground potential via a ceramic capacitor (e.g. 1 nF, 1500 V). The low frequency potential equalisation currents are thus suppressed, but the protective eect against high frequency interference signals remains.
Select connec­tion cable for Ex applications
Ta
ke note of the corresponding installation regulations for Ex applications. In particular, make sure that no potential equalisation currents ow over the cable screen. In case of grounding on both sides this can be achieved by the use of a capacitor or a separate potential equalisation.

5.2 Connection procedure

War
ning:
Before connecting to voltage supply, the plug connection between transducer and remote electronics must be provided in idle condition (see illustration below). Non-observance will damage the electronics!
Fig. 18: Plug 1 Socket electronics housing 2 Plug connector 3 Connection piece, transducer tube
connector between transducer and socket of electronics housing
Only then, the sensor may be connected to voltage supply.
Proceed as follows:
1 Unscrew the housing cover 2 Loosen compression nut of the cable entry
24 VE
28793-EN-081127
GASON 64 Foundation Fieldbus
ecting to power supply
5 Conn
3 Remo
ve approx. 10 cm (4 in) of the cable mantle (current output),
strip approx. 1 cm (0.4 in) insulation from the ends of the individual wires
4 Insert the cable through the cable gland into the sensor 5 Lift the opening levers of the terminals with a screwdriver 6 Insert the wire ends into the open terminals according to the wiring
plan
7 Press down the opening levers of the terminals, you will hear the
terminal spring closing
8 Check the hold of the wires in the terminals by lightly pulling on
them
9 Connect the screen to the internal ground terminal, connect the
outer ground terminal with potential equalisation
10 Tighten the compression nut of the cable entry. The seal ring must
completely encircle the cable
11 Connect the lead cable for power supply in the same way
according to the wiring plan, in addition connect the ground conductor to the inner ground terminal.
12 Screw the housing cover on
The electrical connection is nished.
28793-EN-081127
VEGASON 64 Foun
Fig. 19: Connection
steps 5 and 6
dation Fieldbus 25
1 2 3
4
5
1
3
2
Display
1 2 5 6 7 8
I²C
ecting to power supply
5 Conn
Housing overview
Electronics compart-
ment
5.3 Wir
Fig. 20: Double 1 Housing cover, connection compartment 2 Blind stopper or plug M12 x 1 for VEGADIS 61 (optional) 3 Housing cover, electronics compartment 4 Filter element for air pressure compensation 5 Cable gland
ing plan, double chamber housing
chamber housing
compartment, double chamber housing
Fig. 21: Electronics 1 Plug connector for VEGACONNECT (I²C interface) 2 Internal connection cable to the connection compartment 3 Terminals for VEGADIS 61
26 VE
28793-EN-081127
GASON 64 Foundation Fieldbus
3
1
2
3 41 2
L1 N
BUS
PE
/ L / N
L1 N
1 2
1 2
3
4
BUS
Connection compart­ment
ecting to power supply
5 Conn
Wiring plan
Switch on phase
Fig. 22: Connection 1 Spring-loaded terminals for signal output 2 Ground terminal for connection of the ground conductor and screen 3 Spring-loaded terminals for voltage supply
Fig. 23: Wiring 1 Power supply 2 Signal output
compartment, double chamber housing
plan, double chamber housing

5.4 Switch on phase

After VEGASON 64 is connected to voltage supply or after voltage recurrence, the instrument carries out a self-check for approx. 30 seconds. The following steps are carried out:
l Internal check of the electronics l Indication of the instrument type, the rmware as well as the
sensor TAGs (sensor designation)
l Status byte goes briey to fault value
28793-EN-081127
VEGASON 64 Foun
dation Fieldbus 27
ecting to power supply
5 Conn
en the current measured value will be displayed and the
Th corresponding digital output signal will be outputted to the cable.
1)
1)
The values correspond to the actual measured level as well as to the set­tings already carried out, e.g. default setting.
28 VE
28793-EN-081127
GASON 64 Foundation Fieldbus
up with the indicating and adjustment module PLICSCOM
6 Set
Function/Conguration
Mount/Dismount indicat-
ing and adjustment module
6 Set
up with the indicating and adjustment
module PLICSCOM

6.1 Short description

The indicating and adjustment module PLICSCOM is used for measured value display, adjustment and diagnosis. It can be mounted in the following housing versions and instruments:
®
l All sensors of the plics
in the double chamber housing (optionally in the electronics or connection compartment)
l External indicating and adjustment unit VEGADIS 61
Note:
You
can nd detailed information on the adjustment in the operating
instructions manual of the indicating and adjustment module
PLICSCOM.

6.2 Insert indicating and adjustment module

The indicating and adjustment module can be inserted into the sensor and removed again at any time. It is not necessary to interrupt the power supply.
Proceed as follows:
1 Unscrew the housing cover 2 Place the indicating and adjustment module in the desired position
on the electronics (you can choose any one of four dierent positions - each displaced by 90°)
3 Press the indicating and adjustment module onto the electronics
and turn it to the right until it snaps in.
4 Screw housing cover with inspection window tightly back on
Removal is carried out in reverse order.
The indicating and adjustment module is powered by the sensor, an
additional connection is not necessary.
instrument family, in the single as well as
28793-EN-081127
VEGASON 64 Foun
dation Fieldbus 29
up with the indicating and adjustment module PLICSCOM
6 Set
Fig. 24: Mounting
the indicating and adjustment module
Note:
If
you intend to retrot the instrument with an indicating and adjustment
module for continuous measured value indication, a higher cover with an inspection glass is required.
28793-EN-081127
30 VE
GASON 64 Foundation Fieldbus
1.1
2
3
1
up with the indicating and adjustment module PLICSCOM
6 Set
Key functions

Adjustment system

6.3 Adj
Fig. 25: Indicating 1 LC display 2 Indication of the menu item number 3 Adjustment keys
l [OK] key:
l [->
ustment system
and adjustment elements
- Move to the menu overview
- Conrm selected menu
- Edit parameter
- Save
- menu change
- list entry
value
] key to select:
- Select editing position
l [+
] key:
- Change value of the parameter
l [ESC] key:
- interrupt input
- jump to the next higher menu
The
sensor is adjusted via the four keys of the indicating and
adjustment module. The LC display indicates the individual menu items. The functions of the individual keys are shown in the above illustration. Approx. 10 minutes after the last pressing of a key, an automatic reset to measured value indication is triggered. Any values not conrmed with [OK] will not be saved.
28793-EN-081127
VEGASON 64 Foun
dation Fieldbus 31
up with the indicating and adjustment module PLICSCOM
6 Set
Parameter adjustment
Carrying out min. ad-
justment
6.4 Setup
procedure
As VEGASON 64 is a distance measuring instrument, the distance from the sensor to the product surface is measured. To have the real product level displayed, an allocation of the measured distance to the percentage height must be made. To carry out this adjustment, the distance is entered with full and empty vessel. If these values are not known, an adjustment with the distance values, e.g. 10 % and 90 % is also possible. Starting point for these distance specications is always the lower side of the ange, with all other versions the lower side of the transducer.
The actual level is then calculated on the basis of these entered values. At the same time, the operating range of the sensor is limited from maximum range to the requested range.
The real product level during this adjustment is not important, because the min./max. adjustment is always carried out without changing the product level. These settings can be made ahead of time without the instrument having to be installed.
In the main menu item "Basic adjustment", the individual submenu items should be selected one after the other and provided with the correct parameter values.
Start your parameter adjustment with the following menu items of the basic adjustment:
Proceed as follows:
1 Move from the measured value display to the main menu by
pushing [OK].
Basic adjustment Display Diagnostics Service Info
2 Select the menu item "Basic adjustment" with [->] and conrm with
[OK]. Now the menu item "Min. adjustment" is displayed.
Min. adjustment
0.00 %
=
5.000 m(d)
4.000 m(d)
3 Prepare the % value for editing with [OK] and set the cursor to the
requested position with [->]. Set the requested percentage value with [+] and save with [OK]. The cursor jumps now to the distance value.
4 Enter the suitable distance value in m for the empty vessel (e.g.
distance from the sensor to the vessel bottom) corresponding to the percentage value.
32 VE
28793-EN-081127
GASON 64 Foundation Fieldbus
up with the indicating and adjustment module PLICSCOM
6 Set
Carrying out max. ad­justment
Medium selection
5 Save
the settings with [OK] and move to "Max. adjustment" with [-
>].
Proceed as follows:
Max. adjustment
100.00 %
=
1.000 m(d)
2.000 m(d)
1 Prepare the % value for editing with [OK] and set the cursor to the
requested position with [->]. Set the requested percentage value with [+] and save with [OK]. The cursor jumps now to the distance value.
2 Enter the appropriate distance value in m (corresponding to the
percentage value) for the full vessel. Keep in mind that the max. level must lie below the dead band.
3 Save the settings with [OK] and move to "Medium selection" with
[->].
Each product has dierent reective properties. In addition, there are various interfering factors which have to be taken into account: agitated product surfaces and foam generation (with liquids); dust generation, material cones and echoes from the vessel wall (with solids). To adapt the sensor to these dierent conditions, you should rst select "Liquid" or "Solid".
Medium
Liquid
Vessel form
28793-EN-081127
VEGASON 64 Foun
With solids, you can also choose between "Powder/Dust", "Granular/
Pellets" or "Ballast/Pebbels".
Through this additional selection, the sensor is adapted perfectly to the product and measurement reliability, particularly in products with bad reective properties, is considerably increased.
Enter the requested parameter via the appropriate keys, save your settings and jump to the next menu item with the [->] key.
Apart from the medium, the vessel shape can also inuence the measurement. To adapt the sensor to these measuring conditions, this menu item oers dierent options depending on whether liquid or solid is selected. With "Liquids" these are "Storage tank", "Stilling tube", "Open vessel" or "Stirred vessel", with "Solid", "Silo" or "Bunker".
Vessel form
Storage tank
dation Fieldbus 33
up with the indicating and adjustment module PLICSCOM
6 Set
Enter the requested parameter via the appropriate keys, save your
settings and jump to the next menu item with the [->] key.
Damping
Linearisation curve
To suppress uctuations in the measured value display, e. g. caused by an agitated product surface, a damping can be set. This time can be between 0 and 999 seconds. Keep in mind that the reaction time of the entire measurement will then be longer and the sensor will react to measured value changes with a delay. In general, a period of a few seconds is sucient to smooth the measured value display.
Damping
0 s
Enter the requested parameter via the appropriate keys, save your settings and jump to the next menu item with the [->] key.
A linearization is necessary for all vessels in which the vessel volume does not increase linearly with the level - e. g. with a cylindrical or spherical tank - and the indication or output of the volume is required. Corresponding linearization curves are preprogrammed for these vessels. They represent the correlation between the level percentage and vessel volume. By activating the appropriate curve, the volume percentage of the vessel is displayed correctly. If the volume should not be displayed in percent but e.g. in l or kg, a scaling can be also set in the menu item "Display".
Linearisation curve
linear
Enter the requested parameter via the appropriate keys, save your settings and jump to the next menu item with the [->] key.
Gating out of false sig­nals
High sockets or vessel installations, such as e. g. struts or agitators as well as buildup and weld joints on the vessel walls cause interfering reections which can impair the measurement. A false echo storage detects and marks these false echoes, so that they are no longer taken into account for the level measurement. A false echo memory should be created with low level so that all potential interfering reections will be detected.
Gating out of false signals
Change now?
34 VE
28793-EN-081127
GASON 64 Foundation Fieldbus
up with the indicating and adjustment module PLICSCOM
6 Set
eed as follows:
Proc
1 Move from the measured value display to the main menu by
pushing [OK].
2 Select the menu item "Service" with [->] and conrm with [OK].
Now the menu item "False signal suppression" is d isplayed.
3 Conrm "False signal suppression - Change now" with [OK] and
select in the below menu "Create new". Enter the actual distance from the sensor to the product surface. All false signals in this area are detected by the sensor and saved after conrming with [OK].
Note:
Check
the distance to the product surface, because if an incorrect (too large) value is entered, the existing level will be saved as false signal. The lling level would then no longer be detectable in this area.
Extended setting/Quick level change
Copy sensor data
The menu item "Extended setting" oers the possibility to optimise VEGASON 64 for applications in which the level changes very quickly. For this reason, select the function "Quick level change > 1 m/min.".
Extended setting
quick level change > 1 m/min.
Note:
Since
with the function "Quick level change > 1 m/min." the generation
of an average value of the signal processing is considerably reduced, false reections by agitators or vessel installations can cause measured value uctuations. A false echo memory is thus recom­mended.
This function enables reading out parameter adjustment data as well as writing parameter adjustment data into the sensor via the indicating and adjustment module. A description of the function is available in the operating instructions manual "Indicating and adjustment module".
The following data are read out or written with this function:
l Measured value presentation l Adjustment l Medium l Vessel form l Damping l Linearisation curve l Sensor-TAG l Displayed value l Unit of measurement l Language
The following safety-relevant data are not read out or written:
l PIN
28793-EN-081127
VEGASON 64 Foun
dation Fieldbus 35
up with the indicating and adjustment module PLICSCOM
6 Set
Copy sensor data
Copy sensor data?
Reset
Basic adjustment
If the "Reset" is carried out, the sensor resets the values of the following menu items to the reset values (see chart):
Function Reset value
Max. adjustment 0 m(d)
Min. adjustment Meas. range end in m(d)
Medium Liquid
Vessel form not known
Damping 0 s
Linearisation linear
Sensor-TAG Sensor
Displayed value AI-Out
Unit of measurement m(d)
2)
3)
The values of the following menu items are not reset to the reset values (see chart) with "Reset":
Function Reset value
Language no reset
Factory setting
Like basic adjustment, furthermore special parameters are reset to default values.
4)
Pointer
The min. and max. distance and temperature values are reset to the actual value.
Optional settings
Additional adjustment and diagnosis options such as e.g. scaling, simulation or trend curve presentation are shown in the following menu schematic. You will nd a detailed description of these menu items in the operating instructions manual "Indicating and adjustment module".
2)
Sensor-specic basic adjustment.
3)
Depending on the sensor type, see chapter "Technical data".
4)
Special parameters are parameters which are set customer-specically on the service level with the adjustment software PACTware.
36 VE
28793-EN-081127
GASON 64 Foundation Fieldbus
up with the indicating and adjustment module PLICSCOM
6 Set
Basic adjustment
Basic adjustment Display Diagnostics Service Info
Min. adjustment
000.0 %
=
10.000 m(d)
1.245 m(d)
Damping
0 s
Display
Basic adjustment
Display Diagnostics Service Info
Displayed value
AI-Out
1.1
1.5
2.1
1
2
6.5 Menu
Max. adjustment
100.0 %
=
0.000 m(d)
6.789 m(d)
Linearisation curve
linear
schematic
1.2
1.6
Medium
Bulk solid
1.3
Vessel form
1.4
Silo
Diagnostics
Basic adjustment Display
Diagnostics Service Info
Pointer Distance min.: 0.580 m(d) Dist. max.: 16.785 m(d) T-max.: 30.6 °C T-min.: 4.5 °C
28793-EN-081127
VEGASON 64 Foun
3
3.1
Meas. reliability
36 dB
Sensor status
OK
3.2
Curve selection
Echo curve
3.3
Echo curve
Presentation of the echo
curve
dation Fieldbus 37
3.4
up with the indicating and adjustment module PLICSCOM
6 Set
Service
Basic adjustment Display Diagnostics
Service Info
4
Gating out of false signals
Change now?
Unit of measurement
m(d)
Info
Basic adjustment Display Diagnostics Service
Info
Device ID
< max. 32 characters >
Sensor-TAG (PD_TAG)
< max. 32 characters >
Sensor characteristics
Display now?
4.1
4.5
5.1
5.5
5
Extended setting
Fast
level change
(> 1 m/min.)
Language
Deutsch
Sensor type
VEGASON 6x
Serial number
12345678
4.2
4.6
5.2
Simulation
Start simulation?
Copy sensor data
Copy sensor data?
Date of manufacture
10. April 2007 Software version
3.50
4.3
4.7
5.3
Reset
Reset now?
PIN
Enable?
Date of last change
10. April 2007
4.4
4.8
5.4
38 VE
28793-EN-081127
GASON 64 Foundation Fieldbus
up with the indicating and adjustment module PLICSCOM
6 Set
6.6 Saving
It is recommended noting the adjusted data, e.g. in this operating instructions manual and archive them afterwards. They are hence available for multiple use or service purposes.
If VEGASON 64 is equipped with an indicating and adjustment module, the most important data can be read out of the sensor into indicating and adjustment module. The procedure is described in the operating instructions manual "Indicating and adjustment module" in the menu item "Copy sensor data". The data remain there permanently even if the sensor power supply fails.
If it is necessary to exchange the sensor, the indicating and adjustment module is inserted into the replacement instrument and the data are written into the sensor under the menu item "Copy sensor data".
the parameter adjustment data
28793-EN-081127
VEGASON 64 Foun
dation Fieldbus 39
3
1
2
1
2
3
4
OPEN
TWIST
USB
LOCK
7 Setup
with PACTware and other adjustment programs
Internal connection via I²C interface
External connection via I²C interface
7 Setup
with PACTware and other adjustment
programs

7.1 Connect the PC via VEGACONNECT

Fig. 26: Connection 1 USB cable to the PC 2 VEGACONNECT 3 Sensor
of the PC via VEGACONNECT directly to the sensor
40 VE
Fig. 27: Connection 1 I²C bus (com.) interface on the sensor 2 I²C connection cable of VEGACONNECT 3 VEGACONNECT 4 USB cable to the PC
via C connection cable
28793-EN-081127
GASON 64 Foundation Fieldbus
7 Setup
ary components:
Necess
l VEGASON 64 l PC with PACTware and suitable VEGA DTM l VEGACONNECT l Power supply unit or processing system
with PACTware and other adjustment programs

7.2 Parameter adjustment with PACTware

Further setup steps are described in the operating instructions manual
"DTM Collection/PACTware" attached to each CD and which can also be downloaded from our homepage. A detailed description is available in the online help of PACTware and the VEGA DTMs.
Note:
Keep
in mind that for parameter adjustment of a VEGASON 64, DTM
Collection 04/2004 or a newer version must be used.
All currently available VEGA DTMs are provided in the DTM Collection
on CD and can be obtained from the responsible VEGA agency for a token fee. This CD includes also the up-to-date PACTware version. The basic version of this DTM Collection incl. PACTware is also available as a free-of-charge download from the Internet.
www.vega.co
Go via
m and "Downloads" to the item "Software".

7.3 Parameter adjustment with AMS

For VEGA sensors, instrument descriptions for the adjustment program AMSare available as DD. The instrument descriptions are already implemented in the current version of AMS™. For older versions of AMS™, a free-of-charge download is available via Internet.
Go via
www.vega.co
m and "Downloads" to the item "Software".
28793-EN-081127
VEGASON 64 Foun

7.4 Saving the parameter adjustment data

It is recommended to document or save the parameter adjustment data. They are hence available for multiple use or service purposes.
The VEGA DTM Collection and PACTware in the licensed, profes­sional version provide suitable tools for systematic project documen­tation and storage.
dation Fieldbus 41
aintenance and fault rectication
8 M
Reaction when malfunc­tions occur
Causes of malfunction
Fault rectication
24 hour service hotline
8 Maint
enance and fault rectication

8.1 Maintenance

When used in the correct way, no special maintenance is required in normal operation.

8.2 Remove interferences

The operator of the system is responsible for taken suitable measures to remove interferences.
A maximum of reliability is ensured. Nevertheless, faults can occur during operation. These may be caused by the following, e.g.:
l Sensor l Process l Power supply l Signal processing
The rst measures to be taken are to check the output signals as well as to evaluate the error messages via the indicating and adjustment module. The procedure is described below. Further comprehensive diagnostics can be carried out on a PC with the software PACTware and the suitable DTM. In many cases, the causes can be determined in this way and faults can be rectied.
However, should these measures not be successful, call the VEGA service hotline in urgent cases under the phone no. +49 1805 858550.
The hotline is available to you 7 days a week round-the-clock. Since we oer this service world-wide, the support is only available in the English language. The service is free of charge, only the standard telephone costs will be charged.
Checking Foundation Fieldbus
? When an additional instrument is connected, the H1 segment fails.
l Max. supply current of the segment coupler exceeded
à Measure the current consumption, reduce size of segment
? Measured value on the indicating and adjustment module does not
correspond to the value in the PLC
l The menu item "Display - Display value" is not set to "AI-Out"
à Check values and correct, if necessary
? Instrument does not appear during connection setup
l Incorrect termination
à Check termination at the beginning and end points of the bus
and terminate, if necessary, according to the specication
l Instrument not connected to the segment
à Check and correct
42 VE
28793-EN-081127
GASON 64 Foundation Fieldbus
aintenance and fault rectication
8 M
applications, the regulations for the wiring of intrinsically safe
In Ex circuits must be observed.
Fault messages via the indicating/adjustment module
Reaction after fault rec­tication
? E013
l no measured value available
à sensor in boot phase
à Sensor does not nd an echo, e.g. due to faulty installation or
wrong parameter adjustment
? E017
l Adjustment span too small
à Carry out a fresh adjustment and increase the distance
between min. and max. adjustment
? E036
l no operable sensor software
à Carry out a software update or send the instrument for repair
? E041
l Hardware error, electronics defective
à Exchange instrument or return instrument for repair
? E113
l Communication conict
à Exchange instrument or return instrument for repair
Depending on the failure reason and measures taken, the steps
described in chapter "Set up" must be carried out again, if necessary.

8.3 Exchanging the electronics module

If the electronics module is defective, it can be replaced by the user.
Sensor serial number
28793-EN-081127
VEGASON 64 Foun
applications only one instrument and one electronics module with
In Ex respective Ex approval may be used.
If there is no electronics module available on site, one can be ordered from the VEGA agency serving you.
The settings of the sensor must be downloaded into the new electronics module. This can be done:
l At the factory by VEGA l Or on site by the user
In both cases, the sensor serial number is needed. The serial numbers are stated on the type label of the instrument, inside the housing or on the delivery note.
dation Fieldbus 43
aintenance and fault rectication
8 M
tion:
Informa
When loading on site, rst of all the order data must be downloaded from the Internet (see operating instructions manual "Oscillator").
Assignment
Load sensor software to PC
Prepare update
Load software into sen-
sor
The electronics modules are adapted to the respective sensor and distinguish also in the signal output or power supply.

8.4 Software update

The following components are required to update the sensor software:
l Sensor l Power supply l VEGACONNECT l PC with PACTware l Current sensor software as le
At "
www.vega.co instruments and sensors" the suitable instrument series. Load the zip le via the right mouse key with "Save target as" e.g. on the desktop of your PC. Extract all les available in the zip le, e.g. to the desktop.
Connect the sensor to power supply and provide connection from PC to the instrument via VEGACONNECT. Start PACTware and provide connection to the sensor, e.g. via the VEGA project assistant. Close the parameter window of the sensor, as far as open.
Go in the PACTware menu bar to "Instrument data", "Additional functions" and "Update instrument software".
PACTware checks now the actual hardware and software version of the sensor and displays the data. This procedure lasts approx. 60 s.
Push the button " Update software" and select the previously extracted hex le. Then the software update can be started. Further les are installed automatically. Depending on the sensor, this procedure lasts approximately 1 h.
m/downloads" go to "Software". Select under "plics

8.5 Instrument repair

If a repair is necessary, please proceed as follows:
You can download a return form (23 KB) from our Internet homepage
www.vega.com
form".
By doing this you help us carry out the repair quickly and without having to call back for needed information.
l Print and ll out one form per instrument l Clean the instrument and pack it damage-proof l Attach the completed form and, if need be, also a safety data
sheet outside on the packaging
44 VE
under: "Downloads - Forms and certicates - Repair
28793-EN-081127
GASON 64 Foundation Fieldbus
aintenance and fault rectication
8 M
ase ask the agency serving you for the address of your return
l Ple
shipment. You can nd the respective agency on our website www.vega.com under: "Company - VEGA worldwide"
28793-EN-081127
VEGASON 64 Foun
dation Fieldbus 45
1
2 3
9 Dismoun
ting
9 Dismou
nting

9.1 Dismounting steps

Danger:
The individual wires of the transducer cable carry a voltage of approx.
70 V during operation (see chapter "Mounting preparations"). There- fore dismount VEGASON 64 only in complete absence of line voltage.
War
ning:
Before separating the transducer cable (see chapter "Mounting preparations", example see illustration below), idle condition must be
provided. Non-observance can cause damage to the electronics!
Fig. 28: Plug 1 Socket electronics housing2
Take note of chapters "Mounting" and "Connecting to power supply" and carry out the listed steps in reverse order.
connector between transducer and socket of electronics housing
Plug connector on the transducer cable3 Connection piece, transducer tube

9.2 Disposal

The instrument consists of materials which can be recycled by specialised recycling companies. We use recyclable materials and have designed the electronics to be easily separable.
WEEE directive 2002/96/EG
46 VE
This instrument is not subject to the WEEE directive 2002/96/EG and
the respective national laws. Pass the instrument directly on to a specialised recycling company and do not use the municipal collecting points. These may be used only for privately used products according to the WEEE directive.
28793-EN-081127
GASON 64 Foundation Fieldbus
9 Dismoun
disposal avoids negative eects to persons and environment
Correct and ensures recycling of useful raw materials.
Materials: see chapter "Technical data"
If you have no possibility to dispose of the old instrument
professionally, please contact us concerning return and disposal.
ting
28793-EN-081127
VEGASON 64 Foun
dation Fieldbus 47
10 Suppl
ement
10 Supplem
ent

10.1 Technical data

General data
Materials, wetted parts
- Flange PP or Aluminium
- Swivelling holder, ange galvanized steel
- Transducer PA (1.4301 with StEx)
- Transducer diaphragm 316Ti
Materials, non-wetted parts
- Housing Aluminium die-casting powder coated
- Seal between housing and housing
cover
- Inspection window in housing cover Polycarbonate (UL-746-C listed)
- Ground terminal 316Ti
- transducer cable with separate version C
and D
5)
Weight
- Compact with ange (version A) 5.6 10.7 kg (12.3 23.6 lbs)
- Compact with swivelling holder
(version B)
- Separate with swivelling holder
(version C)
- Separate with threaded tting (version D) 4.7 6.9 kg (10.4 15.2 lbs)
Silicone
PUR (1.1082)
6.9 … 9.7 kg (15.2 … 21.4 lbs)
7.5 … 10.5 kg (16.5 … 23.1 lbs)
Output variable
Output
- Signal digital output signal, Foundation Fieldbus protocol
- Physical layer according to IEC 61158-2
Channel Numbers
- Channel 1 Primary value
- Channel 2 Secondary value 1
- Channel 3 Secondary value 2
Transmission rate 31.25 Kbit/s Current value 10 mA, ±0.5 mA Damping (63 % of the input variable) 0 999 s, adjustable
5)
Depending on size and process tting material.
48 VE
GASON 64 Foundation Fieldbus
28793-EN-081127
plement
10 Sup
variable
Input
Measured value
- Version A distance between lower edge of the ange and
product surface
- version B, C and D distance between lower edge of the transducer and
product surface
Measuring range
- Liquids up to 25 m (82.02 ft)
- Bulk solids up to 15 m (49.21 ft)
Dead band 1 m (3.3 ft)
Reference conditions to measuring accuracy (similar to DIN EN 60770-1)
Reference conditions according to DIN EN 61298-1
- Temperature +18 … +30 °C (+64 +86 °F)
- Relative humidity 45 75 %
- Air pressure 860 … 1060 mbar/86 106 kPa (12.5 15.4 psig)
Other reference conditions
- Reector ideal reector, e.g. metal plate 2 x 2 m (6.56 x 6.56 ft)
- False reections Biggest false echo, 20 dB smaller than the useful
echo
Measuring characteristics
Ultrasonic frequency 30 kHz Interval > 2 s (dependent on the parameter adjustment) Beam angle at 3 dB 4°
Adjustment time
6)
> 3 s (dependent on the parameter adjustment)
Measuring accuracy
Resolution, general > 1 mm (0.039 in)
Deviation
7)
6)
Time to output the correct level (with max. 10 % deviation) after a sudden level change.
7)
Incl. non-linearity, hysteresis and non-repeatability.
see diagram
28793-EN-081127
VEGASON 64 Foun
dation Fieldbus 49
25 m
(82.021 ft)
10 m
(32.8 ft)
3 m
(9.48 ft)
25 mm (0.984 in)
50 mm (1.196 in)
6 mm (0.236 in)
-6 mm (-0.236 in)
-25 mm (-0.984 in)
-50 mm (-1.196 in)
5 m (16.4 ft)
15 m
(49.213 ft)
20 m
(65.617 ft)
25 m
(82.021 ft)
10 m
(32.8 ft)
3 m
(9.48 ft)
25 mm (0.984 in)
50 mm (1.196 in)
6 mm (0.236 in)
-6 mm (-0.236 in)
-25 mm (-0.984 in)
-50 mm (-1.196 in)
5 m (16.4 ft)
15 m
(49.213 ft)
20 m
(65.617 ft)
10 Suppl
ement
Fig. 29: Deviation VEGASON 64 version A
Fig. 30: Deviation VEGASON 64 versions B, C and D
uence of the ambient temperature to the sensor electronics
In
Average temperature coecient of the zero signal (temperature error)
50 VE
8)
Relating to the nominal measuring range.
0.06 %/10 K
8)
GASON 64 Foundation Fieldbus
28793-EN-081127
plement
10 Sup
ent conditions
Ambi
Ambient, storage and transport temperature -40 +80 °C (-40 +176 °F)
Process conditions
Process pressure
- Version A with PP ange 0 kPa
- Other versions -20 50 kPa/-0.2 0.5 bar (-2.9 7.3 psi)
Process temperature (transducer temper-
-40 +80 °C (-40 +176 °F)
ature)
Vibration resistance mechanical vibration with 1 g in the frequency
range 5 200 Hz
9)
Electromechanical data
Cable entry
- Double chamber housing l 1 x cable gland M20 x 1.5 (cable: ø 5 9 mm), 1 x blind stopper M20 x 1.5; plug M12 x 1 for
VEGADIS 61 (optional)
or:
l 1 x closing cap ½ NPT, 1 x blind stopper
½ NPT, plug M12 x 1 for VEGADIS 61
(optional)
or:
l 1 x plug (depending on the version), 1 x blind
stopper M20 x 1.5; plug M12 x 1 for VEGADIS 61 (optional)
Spring-loaded terminals for wire cross-sec­tion up to
Transducer cable
10)
- Length 5 … 300 m (16.4 984.3 ft)
- Diameter 7.2 … 7.6 mm (0.283 0.299 in)
2.5 mm² (AWG 14)
Indicating and adjustment module
Voltage supply and data transmission through the sensor
Indication LC display in dot matrix Adjustment elements 4 keys
Protection
- unassembled IP 20
- mounted into the sensor without cover IP 40
9)
Tested according to the regulations of German Lloyd, GL directive 1.
10)
With separate version C and D.
28793-EN-081127
VEGASON 64 Foun
dation Fieldbus 51
als
ement
www.vega.com
via "VEGA Tools" and
10 Suppl
Materi
- Housing ABS
- Inspection window Polyester foil
Power supply
Operating voltage
- Non-Ex and Ex-d instrument 20 … 72 V DC, 20 253 V AC, 50/60 Hz Power consumption max. 4 VA; 2.1 W
Electrical protective measures
Protection IP 66/IP 67
Overvoltage category III
Protection class I
Approvals
Depending on the version, instruments with approvals can have dierent technical data.
For these instruments, the corresponding approval documents have to be taken into account. These are part of the delivery or can be downloaded under
"serial number search" as well as via "Downloads" a nd "Approvals".
52 VE
28793-EN-081127
GASON 64 Foundation Fieldbus
t
i
Simulate
ALARM
S
OutPV
FIELD-VALUE
M
ode &
St
a
t
us
PV
F-
T
I
M
E
LOW-CUTOFF
L-
T
YPE
OUT-SC
ALE
PV-SCALE
DIRECT
INDIRECT
INDIRECT
SQRT
m(d)
% Lin%
min/max
adjustment
Linearization
Primary
Value
Secondary
Value 2
Secondary
Value 1
Sensor_Value
AI 2
AI 1
TB
AI 3
Temperature
Channel
Channel = 1: Primary Value Channel = 2: Secondary Value 1 Channel = 3: Secondary Value 2 Channel = 4: Temperature
10 Sup
plement
10.2 F
oundation Fieldbus
Block diagram, measured value processing
The following illustration shows the Transducer Block (TB) and Function block (FB) in simplied form.
Fig. 31: VEGASON 64 measured
Diagram, adjustment
The following illustration shows the function of the adjustment:
28793-EN-081127
VEGASON 64 Foun
value processing
dation Fieldbus 53
Secondary_value_1
Calibration
Lowest Point
Calibration
Highest Point
Sensor_value
cal_level_lo
cal_level_hi
Calibration Lowest Point
Calibration Highest Point
Sensor reference plane
cal_level_lo [%]
cal_level_hi [%]
10 Suppl
Fig. 32: Adjustment VEGASON 64
ement
Param
eter list for Device revision 3.0
The following list contains the most important parameters and their meaning:
l primary_value
- This is the process value after adjustment and Line
block
l primary_value_unit
- Unit code of 'Prima
l secondary_value_1
ry_value'
arization with the status of the transducer
- Value after min./max.-adjustment (level + level oset). Selected as input to AIFB by setting
'Channel' = 2. Unit derives from 'Secondary_value_1_unit'
l secondary_value_1_unit
- Unit code of 'Secon
l secondary_value_2
- Sensor
l secondary_value_2_unit
- Unit
54 VE
value + senso
from 'Secondary_value_2_unit'
code of 'Secondary_value_2'
dary_value_1'
r oset. Selected as input to AIFB by setting 'Channel' = 3. Unit derives
GASON 64 Foundation Fieldbus
28793-EN-081127
10 Sup
plement
l sensor_valu
- Raw sensor value, i.e. the
'Sensor_range.unit'
l sensor_range
e
uncalibrated measurement value from the sensor. Unit derives from
- Sensor_range.unit' refers to 'Sensor_value', 'Max/Min_peak_sensor_value', 'Cal_point_hi/lo'
l simulate_primary_value l simulate_secondary_value_1 l simulate_secondary_value_2 l Device Status l Linearization Type
- Possi
l CURVE_POINTS_1_10
- X and Y values for
l CURVE_POINTS_11_20
- X and Y values
l CURVE_POINTS_21_30
ble types of linearization are: linear, user
container
the user dened linearization curve
for
the user dened linearization curve
dened, cylindrical lying container, spherical
- X and Y values for the user dened linearization curve
l CURVE_POINTS_31_33
- X and Y values for
l CURVE_STATUS
the user dened linearization curve
- Result of table plausibility check
l SUB_DEVICE_NUMBER l SENSOR_ELEMENT_TYPE l display_source_selector
- Selects the type of value, whic
l max_peak_sensor_value
h is displayed on the indicating and adjustment module
- Holds the maximum sensor value. Write access resets to current value. Unit derives from
'Sensor_range.unit'
l min_peak_sensor_value
- Holds the minimum sensor value. Write
'Sensor_range.unit'
l CAL_POINT_HI
- Min./max.-adjustme
dened in 'Sensor_range.unit' hi
l CAL_POINT_LO
nt: Upper calibrated point of the sensor. It refers to 'Cal_level_hi'. The unit is
access resets to current value. Unit derives from
- Min./max.-adjustment: Lower calibrated point of the sensor. It refers to 'Cal_level_lo'. The unit is
dened in 'Sensor_range.unit'
l CAL_LEVEL_HI
- Min./max.-adjustme
(Online) the 'Cal_point_hi' is automatically set to the current sensor value. The unit is dened in
'Level_unit'
l CAL_LEVEL_LO
nt: Level at 'Cal_point_hi'. When writing 'Cal_level_hi' and 'Cal_type' = 1
28793-EN-081127
VEGASON 64 Foun
dation Fieldbus 55
10 Suppl
ement
- Min./max.-adjustme
(Online), the 'Cal_point_lo' is automatically set to the current sensor value. The unit is dened in
'Level_unit'
l CAL_TYPE
- Min./max.-adjust
current sensor value determines 'Cal_point_hi/lo'
l level
- Value after min./max.-ad
l level_unit
nt: Level at 'Cal_point_lo'. When writing 'Cal_level_lo' and 'Cal_type' = 1
me
nt: Denes type of calibration: Dry: no inuence of sensor value. Online:
justment
- Unit code of 'Level', 'Level_oset', 'Cal_level_hi', 'Cal_level_lo'
l level_oset
- Oset that is added to the 'Level' valu
l SENSOR_OFFSET
e. Unit derives from 'Level_unit'
- Offset that is added to the 'Sensor_value'. Unit derives from 'Sensor_range.unit'
l end_of_operation_range
- Set up to suit the process conditions
l begi
n_of_operation_range
- Set up to suit the process conditions
l product_type
- Set up to suit the process conditions. If Spec
parameter cannot be written
l liquids_medium_type
ial-Parameter adjustment has been utilized this
- Set up to suit the process conditions. If Special-Parameter adjustment has been utilized this
parameter cannot be written
l solids_medium_type
- Set up to suit the process conditions. If Spec
parameter cannot be written
l liquids_vessel_type
- Set up to suit the process conditions. If Spec
parameter cannot be written
l solids_vessel_type
ial-Parameter adjustment has been utilized this
ial-Parameter adjustment has been utilized this
- Set up to suit the process conditions. If Special-Parameter adjustment has been utilized this
parameter cannot be written
l fast_level_change
up to suit the process conditions. If Spec
- Set
parameter cannot be written
l first_echo_factor
ial-Parameter adjustment has been utilized this
- Set up to suit the process conditions
l pulse_velo
city_correction
- Set up to suit the process conditions
l echo_quality
rat
- Signal/Noise
l empty_vessel_curve_corr_dist
io
28793-EN-081127
56 VE
GASON 64 Foundation Fieldbus
10 Sup
plement
- Distance from the sensor to the product surface. Unit
l empty_vessel_curve_corr_op_code
- Update, create
l sound_velocity
new or delete the empty vessel curve
derives from 'Sensor_range.unit'
- Set up to suit the process conditions
l sound_velocity_unit
- Unit code of 'Soun
l Temperature
d_velocity'
- Process temperature. Selected as input to AIFB by setting 'Channel' = 4. Unit derives from
'Temperature.unit'
l temperature_unit
- Unit code of 'T
l max_peak_temperature_value
emperature', 'Max./Min._peak_temperature_value'
- Holds the maximum process temperature. Write access resets to current value. Unit derives from
'Temperature.unit'
l min_peak_temperature_value
- Holds
the minimum process temperature. Write
'Temperature.unit'
access resets to current value. Unit derives from
28793-EN-081127
VEGASON 64 Foun
dation Fieldbus 57
ø
84mm
~
27
/
")
/
32
* 129mm (5
5
/
64
")
5
/
16
")
M20x1,5/
½
NPT
M12x
10 Suppl
ement
10.3 Dim
ensions
Housing
Fig. 33: Aluminium * Dimension with integrated indicating and adjustment module
double chamber housing
58 VE
28793-EN-081127
GASON 64 Foundation Fieldbus
VEGASON 64
3 4
3 4
110 mm (4.33")
ø 189 mm /
(7.44")
ø 196 mm
(7.72")
b
130 mm (5.12")
k
D
240 mm (9.45")
1
2
>ø 200 mm
(7.87")
250 mm (9.84")
440 mm (17.32")
max. 15°
(0.59")
k
D
Dmm
DIN
ASME
b k
d DN 50 165 11,5 125 4x ø 18 DN 80 200 11,5 160 4x ø 18
Dinch b k
d
DN 50 / 2" DN 80 / 3"
Dmm b k
d
2" 152,4 11,5 120,7 4x ø 19,1 3" 190,5 11,5 152,4 4x ø 19,1
Dinch b k
d
2" 3"
6.50" 4.92" 4x ø 0.71" 4x ø 0.71"6.3"0.45"
0.45"
7.87"
6" 4.75" 4x ø 0.75"
4x ø 0.75"6"0.45"
0.45"
7.5"
plement
10 Sup
Fig. 34: VEGASON 64, the 1 Version A 2 Version B 3 Dead zone: 1 m (3.281 ft) 4 Measuring range: with liquids up to 25 m (82.02 ft), with solids up to 15 m (49.21 ft)
28793-EN-081127
VEGASON 64 Foun
specication to the transducer diameter apply to standard/StEx
dation Fieldbus 59
3 4
1 2
250 mm (9.84") 312 mm (12.28")
150 mm (5.91")
>ø 200 mm
(7.87")
220 mm (8.66")
149 mm (5.87")
18 mm
(0.71")
143 mm (5.63")
G1A
k
D
Dmm
DIN
ASME
b k
d DN 50 165 11,5 125 4x ø 18 DN 80 200 11,5 160 4x ø 18
Dinch b k
d DN 50 / 2" DN 80 / 3"
Dmm b k
d 2" 152,4 11,5 120,7 4x ø 19,1 3" 190,5 11,5 152,4 4x ø 19,1
Dinch b k
d 2" 3"
6.50" 4.92" 4x ø 0.71" 4x ø 0.71"6.3"0.45"
0.45"
7.87"
6" 4.75" 4x ø 0.75"
4x ø 0.75"6"0.45"
0.45"
7.5"
10 Suppl
ement
VEGASON 64
Fig. 35: VEGASON 64 1 Version C 2 Version D 3 Dead 4 Measuring range: with liquids up to 25 m (82.02 ft), with solids up to 15 m (49.21 ft)
zone: 1 m (3.281 ft)
60 VE
GASON 64 Foundation Fieldbus
28793-EN-081127
10 Sup
plement
10.4 In
dustrial property rights

10.5 Trademark

VEGA product lines are global protected by industrial property rights. Further information see http://www.vega.com.
Only in U.S.A.: Further information see patent label at the sensor housing.
VEGA Produktfamilien sind weltweit geschützt durch gewerbliche Schutzrechte.
Nähere Informationen unter http://www.vega.com.
Les lignes de produits VEGA sont globalement protégées par des droits de propriété intellectuelle. Pour plus d'informations, on pourra se référer au site http://www.vega.com.
VEGA lineas de productos están protegidas por los derechos en el campo de la propiedad industrial. Para mayor información revise la pagina web http://www.vega.com.
Линии продукции фирмы ВЕГА защищаются по всему миру правами на интеллектуальную собственность. Дальнейшую информацию смотрите на сайте http://www.vega.com.
VEGA 站<http://www.vega.com>
All brands used as well as trade and company names are property of their lawful proprietor/originator.
28793-EN-081127
VEGASON 64 Foun
dation Fieldbus 61
10 Suppl
ement
62 VE
28793-EN-081127
GASON 64 Foundation Fieldbus
10 Sup
plement
28793-EN-081127
VEGASON 64 Foun
dation Fieldbus 63
VEGA Grieshaber KG
ISO 9001
Am Hohenstein 113 77761 Schiltach
Germany Phone +49 7836 50-0 Fax +49 7836 50-201 E-mail: info@de.vega.com
www.vega.com
Printing date:
statements concerning scope of delivery, application,
All
practical use and operating conditions of the sensors and
processing systems correspond to the information avail-
able at the time of printing.
© VEGA Grieshaber KG, Schiltach/Germany 2008
Subject to change without prior notice 28793-EN-081127
Loading...