Please read this manual carefully, and also take
note of country-specific installation standards
(e.g. the VDE regulations in Germany) as well
as all prevailing safety regulations and accident prevention rules.
For safety and warranty reasons, any internal
work on the instruments, apart from that involved in normal installation and electrical connection, must be carried out only by qualified
VEGA personnel.
2VEGASCAN 850
Note Ex area
Please note the approval documents (yellow
binder), and especially the included safety
data sheet.
Continuous level measurement with ultrasonic
sensors is based on the running time measurement of ultrasonic pulses.
VEGASON series 51 … 56 sensors are a
newly developed generation of extremely
compact ultrasonic sensors for level measurement. They were especially developed for
liquids (51 ... 53) and for solids and larger
measuring ranges (54 ... 56).
Due to the small housing dimensions and
process fittings, the compact sensors are a
very reasonable solution for your level measurement applications. With the integrated
display and a special sensor intelligence, in
conjunction with large measuring ranges,
they can be used for applications in which
the advantages of non-contact measurement
could never before be realized.
As output and measuring signal, the sensors
produce a digital output signal which is processed and evaluated in VEGASCAN 850.
VEGASCAN 850 then outputs the processed
measured values as a digital communication
protocol.
Measuring principle
Piezoceramic high-performance transducers
emit focused ultrasonic pulses which are
reflected by the product surface. The measurement electronics prepares a precise image of the environment out of the reflected
ultrasonic pulses. The transducers work both
as transmitter and receiver. As receiver, the
transducers are high-sensitivity piezo microphones.
Product description
Meas. distance
emission - reflection - reception
The measurement electronics precisely calculates the distance between transducer and
medium from the speed of sound and the
measured running time of the emitted sound
impulse. The distance is then converted into
a level-proportional signal, and in conjunction
with the sensor parameter settings, made
available as precise, calibrated level value.
Since the speed of sound is subject to temperature influence, the transducer also continuously detects the ambient temperature,
so that the level is precisely measured even
in case of varying ambient temperature.
4VEGASCAN 850
Product description
Output signal
The level-proportional measuring signal of the
sensors is received, processed and
outputted digitally throughout. VEGASCAN
850 reads in the digital sensor measured
values permanently and processes the individual levels according to the user’s selected
options. Depending on the version of
VEGASCAN, max. 15 sensors can be connected to one screened two-wire cable or 30
sensors on two screened two-wire cables.
The two-wire cable transfers beside the digital level signal and adjustment signals also
the supply voltage of 24 V.
The sensor measured values can be processed and outputted in VEGASCAN as levels
in percent, volume or mass units. In addition,
linked process tasks such as scaling, linearisation, differential generation, addition, tendency processing or limit value processing
can be selected individually. As a result,
VEGASCAN can carry out - decentralised
and without costly programming - the entire
measurement-related arithmetic.
The levels and their processed results are
then outputted in the following field bus protocols and interfaces depending on the ordered version of VEGASCAN 850.
The digital processing of the measured signal ensures an accuracy that an analogue
measuring signal could never reach, as the
digital signal is always transferred without
error right down to the last bit and decimal
point position.
Varying line resistance or small leakage currents do not influence the accuracy (digital
technology). The digital signal is always unambiguous.
Display of measured values
As an option, the series 50 ultrasonic sensors
can be equipped with an indicating instrument for direct, local level survey. The indicating instrument shows the precise level by
means of the analogue bar graph and the
digital number value. In addition to the indication in the sensor, you can have the level
displayed by the VEGADIS 50 external indicating instrument at a distance of up to
25 m from the sensor. The external display of
measured values operates, like the integrated display, independently of the output
signal, can be modified through individual
parameter settings, and is connected to and
powered by the sensor.
ProtocolInterface
Siemens 3964- RS 232
VEGA-ASCII- RS 232
Modbus- RS 232
Profibus DP- RS 485
Profibus FMS- RS 485
VEGASCAN 8505
- RS 422
- RS 485
- TTY
- RS 422
- RS 485
- TTY
- RS 422
- RS 485
- TTY
Product description
1.2 Application features
Applications
• Level measurement of all liquids
• Level measurement of solids (only short
measuring distances) such as e.g. coal,
ore, stones, crushed rocks, cement,
gravel, sand, sugar, salt, cereals, flour,
granules, powder, dusts, sawdust,
sawings.
• Flow measurement on various flumes
• Gauge measurement, distance measure-
ment, object monitoring and conveyor belt
monitoring
Two-wire technology
• Supply and output signal on one two-wire
cable (loop powered).
• Output signal and signal processing completely digital.
Rugged and precise
• Measurement unaffected by substance
properties such as density, conductivity,
dielectric constant…
• Suitable for corrosive substances
• Measuring range 0.25 m … 70 m.
• Precise through digital processing and
transmission of measured values.
Means of adjustment
• With adjustment software VEGA Visual
Operating (VVO) on the PC
• With detachable adjustment module
MINICOM
• With VEGAMET signal conditioning instrument.
1.3 Adjustment
Each measuring situation is unique. For that
reason, every ultrasonic sensor needs some
basic information on the application and the
environment, e.g. an empty vessel profile is
important for a reliable measurement. Beside
this, many other settings and adjustments
are possible on VEGASON ultrasonic sensors.
The adjustment and parameter setting of
VEGASCAN 850 and ultrasonic sensors are
carried out with the PC and the adjustment
program VVO (VEGA Visual Operating).
Only sensor-relevant settings can be carried
out with the adjustment module MINICOM.
Adjustment with PC
The program leads quickly through the adjustment and parameter setting by means of
pictures, graphics and process
visualisations.
The PC is connected with a standard cable
(RS 232) directly to VEGASCAN 850.
The adjustment and parameter data of the
connected sensors and the configuration of
VEGASCAN can be saved with the adjustment software on the PC and can be protected by passwords. On request, the
sensor adjustments can be quickly transferred to other sensors.
With the small (3.2 cm x 6.7 cm) 6-key adjustment module with display, the adjustment
can be carried out in clear text dialogue.
The adjustment program recognises the sensor type
and the location of the connection
Visualised input of a vessel linearisation curve
2
P
...
C
8
5
0
BA
on
1...15
Tank 1
m (d)
12.345
ESC
+
-
OK
Detachable adjustment module. The adjustment
module can be plugged into the ultrasonic sensor or
onto the external indicating instrument VEGADIS 50.
The adjustment module can be plugged into
the ultrasonic sensor or into the optional,
external indicating instrument. Unauthorised
sensor adjustments can be prevented by
removing the adjustment module.
Adjustment with the PC on VEGASCAN 850 with the
standard cable RS 232 (with VEGASCAN up to 15
sensors can be operated on one two-wire cable)
VEGASCAN 8507
2 Types and versions
Types and versions
2.1 T ype overview, VEGASCAN 850
Either 15 or 30 series 50 ultrasonic sensors
can be connected to VEGASCAN 850. The
measured data are outputted in the ordered
field bus protocol by the ordered interface.
The field bus protocol and interface type are
determined by the ordered version of VEGASCAN. The type label of VEGASCAN contains
the bus protocol and the interface type (see
chapter "2.3 Type code of VEGASCAN 850“).
Optional field bus protocols:
- Siemens 3964
- Modbus
- VEGA-ASCII
- Profibus DP
- Profibus FMS
Optional interface types:
- RS 232
- RS 422
- RS 485
- TTY
2.2 T ype overview, sensors
VEGASON series 51 … 56 sensors are a
newly developed generation of extremely
compact ultrasonic sensors for large measuring ranges (VEGASON 54 ... 56) or for
shorter measuring ranges (VEGASON
51 … 53).
Due to the small housing dimensions and
process fittings, the compact sensors do
your level monitoring inconspicuously, and
above all, at reasonable cost.
Because of price, reliability and easy handling, ultrasonic level measurement can be
now used in applications in which non-contact level measurement could never be used
before.
VEGASON 50 ultrasonic sensors utilise twowire technology perfectly. The supply voltage
and the digital output signals of 15 sensors
are transmitted between VEGASCAN 850
and the sensors via one two-wire cable.
Swivelling holders allow quick alignment of
the transducer to the product and solid surface. Mounting is simplified by the option of
separating the sensor electronics from the
transducer. The sensor electronics can be
mounted at a distance of 300 m from the
transducer. It is then possible to mount the
transducer in environments with an ambient
temperature up to 150°C (type 56).
8VEGASCAN 850
Types and versions
Common features
• Application to solids and liquids.
• Measuring range 0.25 m … 70 m.
• Ex approved in Zone 1 (IEC) and Zone 1
(ATEX) classification EEx ia [ia] IIC T6.
• Display module integrated in the sensor or
in the external indicating instrument separated up to 25 m from the sensor.
VEGASON 51
VEGASON 52
VEGASON 54
Version AVersion BVersion CVersion D
VEGASON 55
Version AVersion BVersion CVersion D
VEGASON 56
VEGASON 53
Version AVersion BVersion CVersion D
VEGASCAN 8509
Types and versions
Short overview of sensor features
• Use in solids and liquids.
• Measuring range 0.25 … 70 m.
• Ex approved in Zone 1 (IEC) or Zone 1 (ATEX) classification.
• Integrated display of measured values.
VEGASON51V52V53V54V55V56V
Signal output
digital meas. signal••••••
Power supply
- two-wire technology••••••
Process fitting
- G 1 A, 1“ NPT–––•••
-G 11/2 A, 11/2“ NPT•–––––
- G 2 A, 2“ NPT–•––––
- DN 100 compr. flange––•–––
- DN 50–––•••
- DN 80–––•••
- DN 200–––•••
- DN 250–––•••
Adjustment
- with PC and adjustment software••••••
- with adjustment module in sensor••••••
- with adjustment module in ext. indicating
instrument••••••
A - Siemens S5 (3964 R procedure with RK 512)
B - Modbus (RTU and ASCII)
S - Profibus FMS
P - Profibus DP
N - VEGA-ASCII
B - up to 15 sensors can be connected
C - up to 30 sensors can be connected
1
/2“ NPT
VEGASCAN 85011
2.4 T ype code, sensors
Types and versions
VEGASON 54 V EX.XX X X X X X X X
K - Plastic housing PBT, M20 x 1.5 cable entry
N - Plastic housing PBT,
1
/2“ NPT cable entry
A - Aluminium housing, M20 x 1.5 cable entry
G - Process fitting G 11/2 A
N - Process fitting 11/2“ NPT
X - Process fitting DN 100 PN (without sleeve nut)
A - Process fitting DN 100 PN (PPH compression flange)
B - Process fitting DN 100 PN (1.4571 compression flange)
C - 1.4301 mounting strap
FEP - Version A, flange DN 200 (PP)
FFA - Version A, flange DN 200 (Aluminium)
SAS - Version B, flange swivelling holder DN 50
SBS - Version B, flange swivelling holder DN 80
GAS - Version C, flange swivelling holder DN 50
GBS - Version C, flange swivelling holder DN 80
RGS - Thread G 1 A
YYY - Other process fittings
X - without display
A - with integral display
X - without adjustment module MINICOM
B - with adjustment module MINICOM (integrated)
A - 20 … 72 V DC; 20 … 250 V AC; 4 … 20 mA (four-wire)
B - 20 … 72 V DC; 20 … 250 V AC; 4 … 20 mA, HART
®
(four-wire)
E - Power supply via signal conditioning instrument
G - Segment coupler for Profibus PA
P - 90 … 250 V AC (only in USA)
N - 20 … 36 V DC, 24 V AC (only in USA)
Z - Power supply via signal conditioning instr. (only in USA)
.X - without approval
EX.X - Ex approved CENELEC EEx ia IIC T6
EXS.X - StEx Zone 10
K - Analogue 4 … 20 mA output signal (two-wire or four-wire
technology)
V - Digital output signal (two-wire technology - VBUS)
P - Digital output signal (two-wire technology - Profibus)
Type 51 - Meas. range 0.25 … 4 m
Type 52 - Meas. range 0.4 … 7 m
Type 53 - Meas. range 0.6 … 15 m
Type 54 - Meas. range 1.0 … 25 m
Type 55 - Meas. range 0.8 … 45 m
Type 56 - Meas. range 1.6 … 70 m
Meas. technology (SON for ultrasonic)
12VEGASCAN 850
Types and versions
2.5 Approvals
When using ultrasonic sensors in Ex areas or
in marine applications, the instruments must
be suitable and approved for the explosion
zones and application areas.
The suitability is tested by approval authorities and is certified in approval documents.
VEGASON 50 ultrasonic sensors are approved for Ex zone 1, 10, 11, 21 and 22.
Please note the attached approval documents when using a sensor in Ex area.
2.6 Configuration of measuring
systems
A measuring system consists of one to 15
(30) sensors and a VEGASCAN 850. The
processing unit (VEGASCAN 850 signal
conditioning instrument) evaluates the levelproportional, digital measuring signals in a
number of processing routines and outputs
the levels as process bus signal.
Beside the output of the levels in percent,
cubic meters and other physical units, the
levels can be also processed by linked
processing algorithms. Scaling, linearisation,
calculation of linearisation curves, differential
generation, addition or tendency processing
are implemented in VEGASCAN 850 as intrinsic processing routines and are easily accessible via the menu.
On the following pages you will find three
instrument configurations (measuring systems) consisting of sensor(s) and processing unit.
Note:
It is possible to operate up to 15 sensors on
one two-wire cable, see following page (configuration A). However, it would be more
suitable to plan the measuring system in
such a way that max. five sensors are operated on one two-wire cable. For this reason,
VEGASCAN is equipped with three clamping
positions for a VBUS input. The sensor arrangement can either be made linearly (configuration B) or radially (configuration C).
VEGASCAN 85013
Types and versions
Configuration A
Up to 15 sensors via one two-wire cable on VEGASCAN 850
• Two-wire technology, 15 sensors with power supply and digital output signals via one twowire cable on VEGASCAN 850 possible. However, it is better to wire in groups of five as in
configuration B and C.
2
1)
Sensor cables should be screened. Grounding of
4
2
2
the cable screens at both ends is recommended.
However make sure that no earth compensation
currents flow via the screens (see chapter "5.1
Connection, connection cable and screening“).
Earth compensation currents can be avoided by
potential equalisation lines or if the cable screen is
grounded at both ends - by connecting one end via
a capacitor (e.g. 0.1 µF; 250 V) to earth potential.
4
2
2
VEGACONNECT 2
4
2
14VEGASCAN 850
Types and versions
• Display of measured value integrated in sensor.
• Optional external indicating instrument (can be mounted up to 25 m separated from the
sensor, also in Ex area).
• Adjustment with PC or adjustment module MINICOM (can be plugged into the sensor or the
external indicating instrument VEGADIS 50).
• Max. resistance of the signal cable 10 W per wire or 1000 m cable length.
screened sensor cable 1)
2
4
2
2
2)
4
2
2
4
2
2)
Sensor cables leading to the same VBUS input can
be looped together in a screened multiple wire
cable.
Sensor cables leading to another VBUS connection
must be looped in a separate, screened cable.
VEGASCAN 85015
Types and versions
Configuration B
VEGASCAN 850 B; 15 sensors can be connected
• Two-wire technology, 3 x 5 sensors grouped in line on three two-wire cables.
• Display of measured value integrated in sensor.
• Optional external indicating instrument (can be mounted up to 25 m separated from sensor
also in Ex area).
• Adjustment with PC or adjustment module MINICOM (can be plugged into the sensor or the
external indicating instrument VEGADIS 50).
• Max. resistance of the signal cable 15 W per wire or 1000 m cable length.
PC
8
5
0
BA
on
2
2
2
16VEGASCAN 850
Types and versions
Configuration C
VEGASCAN 850 B; 15 sensors can be connected
• Two-wire technology, 3 x 5 sensors grouped radially on three two-wire cables.
• Display of measured values integrated in the sensor.
• Optional external indicating instrument (can be mounted up to 25 m separated from the
sensor also in Ex area).
• Adjustment with PC or adjustment module MINICOM (can be plugged into the sensor or
external indicating instrument VEGADIS 50).
• Max. resistance of the signal cable 15 W per wire or 1000 m cable length.
PC
8
5
0
BA
on
2
4
2
2
2
4
2
4
2
VEGASCAN 85017
Technical data
3 Technical data
3.1 Data
Power supply
Supply voltage
- VEGASCAN 850230 V AC, 50/60 Hz
- sensorssensor power supply is provided via the
Min. sensor voltage17 V
Power consumption
- VEGASCAN 850max. 70 VA
- VEGASON 51 … 5381 mW
- VEGASON 54 … 56max. 1.5 W peak power
Current consumption
- VEGASON 51 … 536 mA
- VEGASON 54 … 5690 mA
Resistance of the signal cable
per measuring data inputdependent on the number of connected sensors
115 V AC, 50/60 Hz
VEGASCAN 850
processing system with max.
1 … 15 sensors per two-wire cable
approx. 1 W continuous power
(power) per connection cable, see following
diagram, however max. 15 W per wire
and max. 1000 m cable length
1000 m
900
840
800
700
630
600
500
400
300
200
Length of the connection cable
100
0
051015
Total power of all sensors on one
2
1,0 mm
2
0,75 mm
two-wire cable
2,5 mm
1,5 mm
2
2
23W max
20
Note:
If longer signal cables are used, it is a good idea to distribute the sensors over two or three
inputs.
18VEGASCAN 850
Technical data
Measuring range
(reference plane is the transducer end. On VEGASON 54 … 56 in version A the lower flange
side is the reference plane.)
VEGASON 51
- liquids0.25 … 4 m
- solids0.3 … 2 m
VEGASON 52
- liquids0.4 … 7 m
- solids0.5 … 3.5 m
VEGASON 53
- liquids0.6 … 15 m
- solids0.75 … 7 m
VEGASON 54 in general1.0 … 25 m
VEGASON 55 in general0.8 … 45 m
VEGASON 56
- version A1.8 … 70 m
- version B … D1.4 … 70 m
Output signal of the sensors
Signal output of the sensorsdigital output signal in two-wire technology
(VBUS): the digital output signal (meas.
signal) is superimposed power supply of
VEGASCAN and further processed in
VEGASCAN
Meas. data inputs on VEGASCAN
Number of inputs3 meas. data inputs on VEGASCAN 850B
6 meas. data inputs on VEGASCAN 850C
Supply voltage from VEGASCANapprox. 24 V
Output current per meas. data inputmax. 1 A
Output power
per meas. data inputmax. 23 W (15 W continuous power)
Resistance of the signal cablemax. 15 W per wire or
max. 1000 m cable length
Integration time0 … 999 seconds (selectable in the sensor)
0 … 600 seconds (selectable in VEGASCAN)
Two-wire technology: The digital output signal (meas. signal) and the power supply are led
through one cable.
VEGASCAN 85019
Technical data
Output signal of VEGASCAN (depending on the ordered instrument version)
Galvanic separationgalvanic separation between sensor current
circuit, power supply, output signal (Siemens
3964, Modbus etc.) and RS 232 PC adjustment
interface
- reference voltageup to 500 V
- isolation resistance4 kV
20VEGASCAN 850
Technical data
Display of measured value (optional)
Liquid-crystal display
- in sensorscalable output of measured values as graph
and digital value
- powered externally by the sensorscalable output of measured values as graph
and digital value. Display of measured values
can be mounted up to 25 m separated
from the sensor.
Adjustment
- PC with adjustment software VEGA Visual Operating via RS 232 interface (max. 15 m
cable length)
- adjustment module MINICOM (only the individual sensor is adjustable)
Characteristics
1)
(typical values under reference conditions, all statements relate to the nominal measuring
range)
Min. span
(between empty and full adjustment)> 20 mm (recommended > 50 mm)
Ultrasonic frequency (at 20°C)
- VEGASON 5170 kHz
- VEGASON 5255 kHz
- VEGASON 5338 kHz
- VEGASON 5430 kHz
- VEGASON 5518 kHz
- VEGASON 5610 kHz
Meas. intervals
- VEGASON 511.0 s
- VEGASON 521.0 s
- VEGASON 530.6 s
- VEGASON 541.0 s
- VEGASON 551.5 s
- VEGASON 562.0 s
Beam angle at -3 dB emitted power
- VEGASON 515.5°
- VEGASON 525.5°
- VEGASON 533°
- VEGASON 544°
- VEGASON 555°
- VEGASON 566°
Influence of the process temperature1.8 %/10 K, however is compensated by a
dynamic temperature detection integrated
in the transducer
Influence of the process pressurenegligible within the permitted sensor pressures
Adjustment time
2)
- VEGASON 51 … 54> 2 s (depending on the parameter setting)
- VEGASON 55, 56> 4 s (depending on the parameter setting)
1)
Similar to DIN 16 086, reference conditions acc. to IEC 770;
temperature 15°C … 35°C; moisture 45 % … 75 %; pressure 860 mbar … 1060 mbar
2)
The adjustment time is the time required by the sensors for the correct output of the level (with max. 10 %
deviation) after a sudden level change.
VEGASCAN 85021
Technical data
Accuracy
1)
(typical values under reference conditions relating to the nominal measuring range)
Characteristicslinear
Deviation in characteristics including
linearity, repeatability and
hysteresis (determined acc. to the
limit point method)< 0.1 %
Linearitybetter than 0.05 %
Average temperature coefficient of the
zero signal0.06 %/10 K
Resolutionmax. 1 mm
Resolution of the output signal
- VEGASON 51 … 540.01 % or 1 mm
- VEGASON 55 and 560.01 % or 10 mm
Ambient conditions
Ambient temperature (housing)-20°C … +60°C
Process temperature (transducer)
Note:
VEGASON 54 … 56 sensors with process fittings G 1 A, 1-11.5 NPT, DN 50 and DN 80
generally require an additional adapter flange, if no other access to the vessel interior is
available.
Connection cables
Power supplysupply and signal via one two-wire cable
Electrical connection
- sensors and VEGASCANspring terminals, terminal cross-section
- swivelling holder or thread G 1 A of
galvanized steel (version B, C and D)
Transducer
- VEGASON 51, 52PVDF
- VEGASON 53UP
- VEGASON 54PA (1.4301 with StEx)
- VEGASON 55, 56UP
Transducer diaphragm
- VEGASON 51, 52PVDF
- VEGASON 531.4571
- VEGASON 541.4571
- VEGASON 55Alu/PE foam
- VEGASON 56Alu/PTFE coating
Transducer cable (cable cover)
- VEGASON 54, 55PUR (1.1082)
- VEGASON 56Silicone (1.1083)
Weight
VEGASON 511.2 kg
VEGASON 521.6 kg
VEGASON 532.3 kg
VEGASON 54
- version A5.6 … 10.7 kg
- version B6.9 … 9.7 kg
- version C7.5 … 10.5 kg
- version D4.7 … 6.9 kg
VEGASON 55
- version A8.0 … 13.3 kg
- version B8.7 … 10.3 kg
- version C9.2 … 11.1 kg
- version D6.5 … 7.5 kg
VEGASON 56
- version A7.3 … 11.3 kg
- version B8.7 … 10.3 kg
- version C9.3 … 11.1 kg
- version D6.5 … 7.5 kg
Technical data
CE conformity
VEGASON series 50 ultrasonic sensors and VEGASCAN 850 meet the protective regulations of EMC (89/336/EWG) and NSR (73/23/EWG). Conformity was judged acc. to the
following standards:
EMCEmissionEN 50 081 - 1: 1992
SusceptibilityEN 50 082 - 2: 1995
NSREN 61 010 - 1: 1993
EN 61 326 - 1: 1997/A1:1998
24VEGASCAN 850
Technical data
3.2 Dimensions
VEGASCAN 850
90
80
260235
246
1390
210
236
VEGASCAN 85025
External indicating instrument VEGADIS 50
Technical data
38
ø5
48
10
Pg 13,5
Mounting on carrier rail 35 x 7 .5 acc. to EN 50 022 or flat
screwed
135
118
108
Adjustment module MINICOM
Tank 1
m (d)
12.345
67,5
ESC
+
-
32,5
OK
Adjustment module for insertion into
VEGASON series 50 sensor or into the external indicating instrument VEGADIS 50
74
82
85
Note:
The diameter of the connection cable must be
5 … 9 mm.
Otherwise the seal effect of the cable entry will
not be ensured.
26VEGASCAN 850
Technical data
VEGASON 51
Housing PBT
201
Housing aluminium
370
322
182
215
185
10˚
165
M20x1,5
SW 60
G 1½ A/
1½" NPT
25
101
90
116
2
ø 40
Min. distance
to the medium
228
315
20
Reference plane
0,25 m
205
M20x1,5
SW 60
102
252
337
2
G 1½ A/
1½" NPT
20
Reference plane
ø 40
Min. distance to
the medium
0,25 m
VEGASCAN 85027
VEGASON 52
Housing PBT
Technical data
201
Housing aluminium
370
322
182
215
185
10˚
165
M20x1,5
SW 60
G 2 A
25
101
90
116
232
2
20
ø 50
Min. distance to
the medium
315
Reference plane
0,4 m
205
M20x1,5
SW 60
102
256
337
2
G 2 A
20
Reference plane
ø 50
Min. distance to
the medium
0,25 m
28VEGASCAN 850
Technical data
VEGASON 53
201
322
M8x10
12 tief
182
20
ø 148
ø 158
10˚
165
M20x1,5
195
130
101
90
118
218
Min. distance to
the medium
270
0,6 m
VEGASCAN 85029
VEGASON 54 … 56 in version A
201
165
Plastic housing
(PBT)
10˚
101
Aluminium
housing (Al)
25
116
Technical data
215
185
VEGASON 54
VEGASON 55
397
257
Min. distance to
the medium
Min. distance to
the medium
1,0 m
0,8 m
90
12xø22
1)
ø190 (ø196)
ø340
ø244
ø405
75
1)
20
110 (126)
12xø26
20
128
12xø22 (12xø26)
445,8
282
Reference plane
Reference plane
20
Reference plane
VEGASON 56
423
Min. distance to
the medium
1,8 m
30VEGASCAN 850
ø198
2)
ø340 (ø405)
Technical data
VEGASON 54 … 56 in version B
486
4xø19
172
386
245
ø165
ø122,8
201
165
10˚
Plug connection
ø 27
> ø200
Plastic housing
(PBT)
11,5
11,5
101
90
503
4xø19
ø165
ø122,8
>ø250
435
270
215
185
65
Plug connection
798
Aluminium
housing (Al)
25
116
ø165
ø122,8
4xø19
>ø210
ø190 (ø196)
VEGASON 54
1,0 m
Reference plane
189,5
ø 244
VEGASON 55
484,5
0,8 m
ø 198
1,4 m
VEGASON 56
VEGASCAN 85031
VEGASON 54 … 56 in V ersion C
Aluminium housing (Al)
Plug
Technical data
215
25
116
185
Plug connection
ø 45
78
68
7
130
150
ø 7
85
65
170
445,8
282
10120
1,0 m
0,8 m
VEGASON 54
VEGASON 55
1,4 m
Reference plane
VEGASON 56
32VEGASCAN 850
Technical data
VEGASON 54 … 56 in version D
Plug connection
3240
214
149
Plastic housing (PBT)
Plug
68
ø 45
78
233
7
3240
167,5
186
101
90
130
150
201
165
10˚
397,2
257,2
ø 7
85
65
12010
170
3240
527
462,5
Reference plane
1,0 m
0,8 m
1,4 m
VEGASON 54VEGASON 55VEGASON 56
VEGASCAN 85033
4 Mounting and installation
4.1 Mounting VEGASON 54 … 56
Mounting and installation
VEGASON 54 … 56, version A
Sensors in version A (flange version) are
supplied completely mounted and ready for
operation. Immediately after mounting on the
vessel and electrical connection, they are
ready for operation.
VEGASON 54 … 56, version B
The sensors in version B are supplied in two
parts (transducer and sensor electronics).
First of all, mount the transducer on the vessel or above the medium. There is a four-pole
jack at the end of the transducer tube. The
respective counterpart to the jack protrudes
out of the lower side of the sensor electronics
housing. Insert the plug of the sensor electronics (only possible in one position) into the
jack of the transducer tube. Continue pushing the electronics housing onto the transducer tube, on which there is a wide and a
narrow groove.
Groove for locking the
headless screw
Mounting groove (must
no longer be visible after
mounting)
The wide groove is used for locking the
headless screws. The narrow groove is a
marking for mounting. Move the electronics
housing farther down over the transducer
tube until the mounting groove is no longer
visible. Fasten the housing with the headless
screws to the transducer tube. Use a 5 mm
hexagon screwdriver (or Allen wrench).
34VEGASCAN 850
Mounting and installation
VEGASON 54 … 56, version C, D
The sensors in version C and D are supplied
in three parts (transducer, sensor electronics
and transducer cable). First mount the transducer (see version B). There is a four-pole
jack at the transducer tube end. A respective
counterpart to the jack is provided in the
connection cylinder of the transducer cable.
Insert the connection cylinder plug into the
jack of the transducer tube.
Connection
cylinder
Mounting
bracket
Connection
cylinder
Transducer
cable
On the end of the transducer tube you find a
wide and a narrow groove. The wide groove
is used for locking the cylinder with the headless screws. The narrow groove is the
mounting mark.
Then push the connection cylinder over the
transducer tube (with a slight swivelling motion) until the mounting mark is no longer
visible.
When the mounting mark is covered by the
cylinder, fasten the cylinder with the two
headless screws. Use a 5 mm hexagon
screwdriver (or Allen wrench).
Now mount the sensor electronics in the
requested location. The sensor electronics is
fastened to a mounting bracket, so that it can
be mounted on a plane surface or on the wall.
Make sure that the sensor housing is
mounted in such a way that there is enough
space above the housing to open the cover.
Now insert the plug at the other end of the
transducer cable into the jack on the electronics housing.
Note:
Avoid bending the transducer cable too
sharply when laying it out. This is a special
cable which could otherwise be damaged.
In addition, make sure that the cable is not
damaged during operation. A signal with a
voltage of approx. 1 kV is transmitted (which
could be a danger in Ex areas if the cable is
damaged).
Groove for locking the
headless screws
Mounting groove (must
no longer be visible after
mounting)
VEGASCAN 85035
Mounting and installation
4.2 General installation instructions
transducer end. For VEGASON 54 ... 56 in
version A the lower flange side on the sensor
Measuring range
Beside other criteria, you select your instrument according to the required measuring
range. The reference planes for the min. and
is the reference plane. Please note the information on the reference planes in chapter
"3.2 Dimensions“. The max. filling depends on
the required min. distance and the mounting
location.
max. distance to the product or solid is the
VEGASON 51
Min.
distance
0.25 m0.4 m
Full
Empty
1m
max. meas. range
4 m (type 51), 7 m (type 52), 15 m (type 53)
Min. distance, max. measuring range and span (VEGASON 51 … 53)
VEGASON 53VEGASON 52
Min.
distance
0.75 m
Span
Reference plane
Min.
distance
VEGASON 56
Version B
Reference plane
min. meas. distance 1.4 m
100 %
0 %
Span
0 %
100 %
Span
VEGASON 54
Version A
min. meas.
distance 1.0 m
VEGASON 55
Version B
Reference plane
min. meas. distance 0.8 m
100 %
0 %
max. meas. range
Span
max. meas. distance 25 m (type 54), 45 m (type 55), 70 m (type 56)
Min. distance, max. measuring range, span and reference plane (VEGASON 54 … 56)
36VEGASCAN 850
Mounting and installation
0
0 m
15 m
100 %
50 %
1,2
1,2
m
0,4
0,4
3˚
8˚
Beam angle and false echoes
The ultrasonic impulses are focused by the
transducer. The impulses leave the transducer in conical form similar to the beam
pattern of a spotlight.
Any object inside this emission cone will
cause a false echo. Especially within the first
few meters of the emission cone, pipes,
struts, or other installations can interfere with
the measurement. At a distance of 6 m, the
false echo of a strut has an amplitude nine
times greater than at a distance of 18 m.
At greater distances, the energy of the ultrasonic impulses distributes over a large area,
thus causing weaker echoes from obstructing surfaces. The interfering signals are
therefore less critical than those at close
range.
If possible, orient the sensor axis perpendicularly to the product surface and avoid
vessel installations (e.g. pipes and struts)
within the 100 % area of the emission cone.
The illustrations showing the ultrasonic emission cones are much simplified and represent only the main beam. However, several
weaker beams also exist. The transducer
must therefore be aligned - especially under
difficult measuring conditions - in such a way
that very low false echo values result. Putting
emphasis only on a strong useful echo is not
sufficient under adverse conditions.
Meas.
distance
0 m
Meas.
distance
7 m
Meas.
distance
0 m
4 m
0,4
0,8
0,4
1,2
VEGASON 51
50 %
100 %
5,5˚
0,4
0
0,8
VEGASON 52
5,5˚
100 %
0,4
1,2
0
VEGASON 53
emitted power
emitted power
12˚
m
12˚
emitted power
emitted power
50 %
emitted power
m
emitted power
Under difficult measurement conditions, we
recommend looking for a mounting location
with the weakest possible false echoes. The
useful echo will then often appear automatically with sufficient quality. With the adjustment software VVO on the PC you can have
a look at the echo image and optimise the
mounting location (see chapter "6.2 Adjustment with the PC – Sensor optimisation –
Echo curve“).
VEGASCAN 85037
0 m
Meas.
distance
Mounting and installation
VEGASON 56VEGASON 54
0 m
emitted power
50 %
100 %
4˚
emitted power
25 m
Meas.
distance
45 m
0 m
4
0,9
2,0
5˚
10˚
242
0
8˚
0,9
2,0
0
VEGASON 55
50 %
emitted power
100 %
m
m
emitted power
Meas.
distance
70 m
50 %
emitted power
6˚
100 %
emitted power
12˚
3,7
7,5
3,7
0
m
7,5
38VEGASCAN 850
Mounting and installation
4.3 Measurement of liquids
Flat vessel top
On flat vessels, the mounting is usually done
on a very short DIN socket piece. Reference
plane is the lower edge of the flange. The
transducer should protrude out of the flange
tube.
Reference plane
£ 60 mm
VEGASON 53 on very short DIN socket piece
< 100 mm
Min.
distance
Reference plane
Reference plane
< 400 mm
Min. distance
1.8 m
VEGASON 56 in flange version on short
DIN socket piece
A mounting location directly on the vessel top
is ideal. A round opening in the vessel top is
sufficient to fasten the sensor with the flange,
or version B and C (VEGASON 54 … 56) with
swivelling holder.
Reference plane
Min. distance
Type 54: 1 m
Type 55: 0.8 m
VEGASON 55 in flange version on short
DIN socket piece
VEGASON 53 (compression flange) on flat vessel top
VEGASCAN 85039
Reference
plane
Swivelling
holder
Reference
plane
Mounting and installation
< 60 mm
Reference
plane
Mounting of the transducer with 1“ thread (here
belonging to VEGASON 54 version D)
Min. meas. distance
1.8 m
Min. meas. distance
1.4 m
Flange version and swivelling holder on flat vessel
top
It is also possible to mount the sensors with
11/2“ or 2“ thread to short socket pieces.
£ 60 mm
Reference plane
Mounting on short 11/2“ or 2“ socket pieces (VEGASON 51)
VEGASON 54 … 56 sensors in version C are
mounted in a 1“ thread.
Dished tank ceiling
On dished tank ceilings, please do not mount
the instrument in the centre, but approx. 1/
vessel radius from the centre. Dished tank
ceilings can act as paraboloidal reflectors. If
the transducer is placed at the focal point of
the parabolic ceiling, the transducer receives
amplified false echoes. The transducer
should be mounted outside the focal point.
Amplified echoes caused by parabolic surfaces are thereby avoided.
£ 60 mm
VEGASON 51, 52 on dished tank ceiling
Reference plane
1
/2 vessel radius
2
40VEGASCAN 850
Mounting and installation
£ 60 mm
1
/2 vessel radius
VEGASON 53 on dished tank ceiling
< 400 mm
1
/2 vessel radius
Reference
plane
Reference plane
Reference
plane
< 100 mm
1
/2 vessel radius
VEGASON 54 version A on dished tank ceiling; the
statements are also valid for VEGASON 55
Open vessels
On open vessels, use of instruments on an
extended mounting bracket is recommended. Mount the low-weight sensor onto
such a bracket and ensure a sufficient distance to the vessel wall.
Reference plane
Min. meas.
Reference plane
distance
VEGASON 56 version A on dished tank ceiling
Min. meas.
distance
VEGASON 54 on open vessel
VEGASCAN 85041
Mounting and installation
Pump shaft
Narrow shafts and shaft-like openings (vessel openings) with very rough walls and
shoulders make an ultrasonic measurement
extremely difficult due to strong false echoes.
This problem can be overcome by using an
extended socket piece or a complete measuring tube (see chapter "4.5 Socket extension“).
see “4.5 Socket extension“
Socket piece
≥ 250 mm
min.
distance
Meas. range
4.4 Measurement of solids
Flange mounting
As with applications for liquids, the instrument
can be mounted on a short DIN socket connection on vessels for solids. The transducer
axis, however, should point to the vessel
outlet or should be oriented perpendicularly
to the product surface. The socket length can
be max. 60 … 400 mm, depending on the
sensor type.
Reference plane
Min. distance
Shaft pump
Measuring tube
Example of a socket extension or measuring tube in a
shaft
Shaft pump
VEGASON 53 on inclined vessel flange
Shaft
Very good measuring results can be attained
with a measuring tube in continuous narrow
shafts, see figure. The applied measuring
tube must have smooth walls inside (e.g. PE
sewage pipe) and a diameter of 100 mm. This
arrangement works well as long as the inside
of the measuring tube collects no dirt or
buildup (cleaning necessary). You might
want to consider using hydrostatic pressure
transmitters or capacitive measuring probes.
Either the measuring tube should never be
immersed in the medium, or it must always
Swivelling holder
We offer as an accessory a swivelling holder
(mounting strap) for mounting of VEGASON
53. This simplifies the alignment of the sensor
to the product surface.
Suitable is the use of VEGASON 54 … 56
version B or C. For solids, the swivelling
holder enables an optimum alignment of the
transducer, thus minimising false echoes.
be immersed (so that the measurement is
carried out exclusively in the tube).
42VEGASCAN 850
Mounting and installation
Reference plane
Min. distance
Mounting boss
Reference plane
Min. distance
VEGASON 53 on swivelling holder
Reference plane
Min. distance
VEGASON 54C with adapter flange and swivelling
holder on a DN 200 vessel flange
VEGASON 51 or 52 on the mounting boss. The socket
axis should point directly to the product surface. This
is why VEGASON 51 and 52 are less suitable for
solids.
Reference plane
Min. distance
VEGASON 56 in 1“ mounting boss.
The socket axis should point directly to the
product surface. Much better would be the
use of a swivelling holder version.
VEGASCAN 85043
Mounting and installation
Material heaps
Large material heaps are best detected with
several instruments, which can be mounted
on e.g. traverse cranes. For this type of application, it is advantageous to orient the
sensor directly toward the solid surface.
Transducer on traverse crane above a material heap
(illustration: VEGASON 54 in version B)
4.5 Socket extensions
The ultrasonic sensors require a min. distance to the liquid or solid product. Take the
min. distance into account in your planning. In
some situations, it is possible to reach the
required min. distance, and hence the desired filling height, with a socket extension.
However, the socket extension increases the
noise level of the ultrasonic signal at the extension outlet and can interfere with the
measurement. Only use a socket extension if
all other possibilities have to be excluded.
Carry out the extension as shown in the following illustration.
Socket extensions in liquids
Chamfer and deburr the socket carefully and
make sure it has a smooth inner surface. The
socket should not protrude into the measured product, in case buildup can form on
the socket through pollution or product residues.
Socket piece should not be immersed into adhesive
products (illustration: VEGASON 53)
The socket diameter should be as large and
the socket length as small as possible. Make
sure that the socket outlet is burr-free to
minimise false echoes.
44VEGASCAN 850
Mounting and installation
15˚ 15˚
Type 51/52
LL
45˚
ø
Socket extensions in liquids
Type 53
45˚
ø
Max. socket length in relation to socket diameter
ø in mmL in mm
Type 51Type 52Type 53
100200300300
150300400400
200–500500
250––600
Type 54
Type 55
Socket extensions for solids
For solids, use a conical socket extension
with a taper of at least 15° … 20°.
Socket extension in solids
Measurement in a tube
For nonadhesive measured products, a
socket extension in the form of a measuring
tube can be permanently submerged in the
product. The ultrasonic measurement is then
made exclusively in the measuring tube and
works very well without interference from
other vessel installations (see "Pump shaft“).
L
45˚
ø
Socket extensions that do not protrude into the
measured product
L
45˚
ø
Max. socket length in relation to socket diameter
ø in mmL in mm
Type 54Type 55Type 56
200400––––
250500500500
300––––600
VEGASCAN 85045
Mounting and installation
4.6 Flow measurement
The short examples on this page are only
basic information on flow measurement. You
can get complete planning information from
the flume manufacturers and in special literature.
Rectangular flume
- Installation of the sensor on the upstream
side
- Note distance to the overfall edge
(3 … 4 x h
- Installation centered to the flume
- Edge opening ³ 2 x h
- Installation perpendicular to the liquid surface
- Keep min. distance in relation to h
- Min. distance from edge opening to downstream water ³ 50 mm
max
90°
)
from ground
max
max
³ max.
distance
h
max
Khafagi-Venturi flume
- Installation of the sensor on the inlet side
- Note distance to the Khafagi-Venturi flume
(3 … 4 x h
- Installation perpendicular to the liquid surface
- Keep min. distance in relation to the height
of damming h
Khafagi-Venturi flume
max
3 … 4 x h
90°
Sensor
)
max
max
h
max
B
³ 2 x h
Overfall edge
max
Flow measurement on open flumes
Overfall edge
3 … 4 x h
max
90°
³ 5 cm
Upstream water
Downstream water
Flow measurement on open flumes
46VEGASCAN 850
Mounting and installation
4.7 False echoes
The mounting location of the ultrasonic sensor
must be selected such that no installations or
inflowing material are in the path of the ultrasonic impulses. The following examples and
instructions show the most frequent measuring problems and how to avoid them.
Vessel protrusions
Vessel forms with flat protrusions can, due to
their strong false echoes, adversely effect
the measurement. Shields above these flat
protrusions scatter the false echoes and
guarantee a reliable measurement.
CorrectWrong
Vessel protrusions (slope)
Intake pipes, e.g. for the mixing of materials with a flat surface directed towards the sensor - should be covered with a sloping
shield. This shield will scatter false echoes.
Vessel installations
Vessel installations such as, for example, a
ladder, often cause false echoes. Make sure
when planning your measurement loop that
the ultrasonic signals have free access to the
measured product.
CorrectWrong
Ladder
Vessel installations
Ladder
Struts
Struts, like other vessel installations, can
cause strong false echoes that are superimposed over the useful echo signals. Small
shields effectively hinder a direct false echo
reflection. These false echoes are scattered
and diffused in the area and are then filtered
out as "echo noise“ by the measuring electronics.
CorrectWrong
CorrectWrong
Shields
Struts
Vessel protrusions (intake pipe)
VEGASCAN 85047
Mounting and installation
Inflowing material
Do not mount the instrument in or above the
filling stream. Ensure that you detect the
product surface and not the inflowing material.
Correct
Correct
Wrong
Wrong
Correct
Buildup
Wrong
Overflow basin
The expected max. high water determines
the installation height, to ensure the min.
distance of the transducer even with the
highest water level. The low water level
should be covered in the transducer area
with a shield to filter out echoes from exposed basin surfaces.
CorrectWrong
Inflowing material
Min. distance
high water
Buildup
If the sensor is mounted too close to the
vessel wall, buildup and adhesions of the
measured product to the vessel wall can
cause false echoes. Position the sensor at a
sufficient distance from the vessel wall.
Please also note chapter "4.2 General installation instructions“.
48VEGASCAN 850
60°
Shield
Filtering out of a level echo
Low water
Mounting and installation
Strong product movements
Heavy turbulences in the vessel, e.g. by
strong stirrers or strong chemical reactions,
can seriously interfere with the measurement.
A surge or bypass tube of sufficient size
(DN 200, DN 250) always allows, provided
the product causes no buildup in the tube, a
reliable measurement even with strong turbulences in the vessel.
100 %
60 %
0 %
Strong product movements
4.8 Incorrect mounting
Foam generation
Thick foam on the product can cause incorrect measurements. Take measures to avoid
foam, carry out the measurement in a bypass
tube, or use a different measuring technology, e.g. capacitive measuring probes or
hydrostatic pressure transmitters.
Foam generation
Wrong orientation to the product
Weak measuring signals are the result if the
sensor is not directly pointed at the product
surface. Orient the sensor axis perpendicularly to the product surface to achieve optimum measuring results.
Orient the sensor perpendicularly to the product
surface
VEGASCAN 85049
Mounting and installation
Strong heat fluctuations
Strong heat fluctuations, e.g. due to the sun,
cause measuring errors. Please provide a
sun shield.
Shield
Strong heat fluctuations
Min. distance to the medium
If the min. distance to the medium is not maintained, the instruments show wrong measured values. Mount the instrument at the
required min. distance.
Sensor too close to the vessel wall
If the sensor is mounted too close to the
vessel wall (dimension A in diagram), strong
false echoes can be caused. Buildup, rivets,
screws or weld joints on the vessel wall superimpose their echoes onto the product or
useful echo. Please ensure a sufficient distance from the sensor to the vessel wall,
depending on the maximum measuring distance (dimension B in diagram).
In case of good reflection conditions (liquids,
no vessel installations), we recommend determining the sensor distance according to
Diagram curve 1. At a max. meas. distance
of e.g. 10 m, the distance of the transducer
(according to curve 1) should be approx.
1.5 m.
In case of solids with bad reflection properties, determine the distance to the vessel wall
according to Diagram curve 2. Under very
bad measuring conditions (rough vessel
walls, struts), it might be necessary to increase the distance to the vessel wall, or to
additionally filter out the false echoes by
storing them in memory, thereby adapting the
sensor more precisely to the environment.
CorrectWrong
Sensor too close to the vessel wall
50VEGASCAN 850
Mounting and installation
Distance of the
transducer to the
vessel wall
A
1 m2 m3 m4 m5 m
Curve 1 (liquids)
5 m
B
10 m
15 m
max. meas.
distance
Curve 2 (solids)
Distance from the sensor to the vessel wall, depending on the meas. distance (type 51 … 53)
Distance of the
transducer to the
vessel wall
A
Parabolic effects of rounded or arched
vessel tops
Round or parabolic tank tops act on the signals like a parabolic mirror. If the sensor is
placed at the focal point of such a parabolic
tank top, the sensor receives amplified false
echoes. The optimum location is generally in
the area of half the vessel radius from the
centre.
Correct
< 100 mm
~ 1/
2
vessel
radius
Wrong
2 m 4 m 6 m 8 m
Wrong
Curve 1 (liquids)
10 m
B
20 m
Curve 2 (solids)
Mounting on a vessel with parabolic tank top
30 m
max. meas.
distance
Distance from the sensor to the vessel wall, depending on the meas. distance (type 54 … 56)
VEGASCAN 85051
Socket piece too long
If the sensor is mounted in a socket extension that is too long, strong false echoes are
caused, and measurement is hindered. Make
sure that the transducer protrudes at least
30 mm out of the socket piece.
Reference plane
< 100 mm
Mounting and installation
Correct and wrong length of socket piece
52VEGASCAN 850
Electrical connection
5 Electrical connection
5.1 Connection, connection cable
and shielding
Qualified personnel
Instruments which are not operated with
protective low voltage or DC voltage must
only be connected by qualified personnel.
This is especially valid for the connection of
the power supply on VEGASCAN.
Safety information
As a rule, do the work in the complete absence of voltage. Always switch off the power
supply before you carry out connecting
work. Protect yourself and the instruments.
The shielded transducer cables transmit a
signal with a voltage of approx. 1 kV. In Ex
areas, cable damage can be dangerous. Do
not carry out any mounting or connection
work on the transducer when VEGASCAN is
switched on.
Connection cable
Please note that the connection cables are
specified for the expected operating temperatures in your systems. The power supply and
the sensor cables must have an outer diameter of 6 … 12 mm, to ensure the seal effect of
the cable entry.
Transducer cable
When wiring the transducer cable (version C
and D), strong bending of the cable should
be avoided. This is a special cable which can
be damaged.
Make sure when wiring the transducer cable
that no operating influences can damage it.
Earth conductor terminal
The electronics housing of the sensors has a
protective insulation. The earth conductor
terminal and the earth terminal in the electronics housing are galvanically connected with
the metallic transducer diaphragm.
On sensors in version B, the earth conductor
terminal is galvanically connected to the
transducer diaphragm via the transducer
tube when the sensor is completely mounted.
On version C and D, the connection is made
via the cable screen of the transducer cable
and the transducer tube.
Shielding of the sensor cables
The "Electromagnetic pollution“ from electronic actuators, power lines and transmitting
stations is often so considerable that the twowire cable of VEGASCAN to the sensors
must be shielded.
Power supply
For power supply, standard three-wire cable
up to max. 2.5 mm2 can be used. The electrical connection is made via spring terminals.
You open the terminal opening with a small
screwdriver by inserting it into the opening
slot above the terminal position and lever it
upward. This opens the terminal and the
copper core of the connection wire can be
inserted.
VEGASCAN 85053
Electrical connection
We recommend a screening on both ends,
see the following sketch. Screening is a good
preventative measure against future sources
of interference. However, you must make
sure that no ground potential currents flow
through the sensor cable shields. Ground
potential currents can be avoided by potential equalisation cables. When grounding at
both ends, it is possible to connect the cable
Note Ex protection!
In Ex applications, grounding on both ends is
not allowed due to potential losses. If an
instrument is used in hazardous areas, the
respective regulations, conformity certificates
and type approvals for systems in Ex areas
must be noted (e.g. DIN 0165).
Please note the approval documents with the
safety data sheet attached to the Ex sensors.
shield on one side (e.g. in the switching cabinet) via a capacitor (e.g. 0.1 µF; 250 V) to the
ground potential. Use a low-resistance
ground connection (foundation, plate or
mains earth).
Linear (serial) arrangement of sensors
Grounding of the cable screen on both ends, at the end of each sensor line via a ground capacitor.
PC
850
BA
on
2
2
54VEGASCAN 850
Electrical connection
Radial arrangement of sensors
Grounding on at least two ends, on VEGASCAN and once on the sensor star, i.e. on the longest sensor line. If the individual sensor lines are longer than approx. 15 m, a grounding of
each longer line should be made via a ground capacitor.
PC
850
BA
on
Longest stub of the
sensor star
2
4
2
2
4
2
VEGASCAN 85055
Electrical connection
5.2 Connection of the sensor
After mounting the sensor at the measurement location according to the instructions in
chapter "4 Mounting and installation“, loosen
the closing screw on top of the sensor. The
sensor lid with the optional indication display
can then be opened. Unscrew the sleeve nut
and slip it over the connection cable (after
removing about 10 cm of insulation). The
sleeve nut of the cable entry has a self-locking ratchet that prevents it from opening on
its own.
Version with aluminium housing
Voltage supply and digital
meas. signal
M20 x 1.5
(diameter of the
connection
cable 6…9 mm)
+
-
To the indicating instrument in the cover or to the
external indicating
instrument VEGADIS 50
M20 x 1.5
Now insert the cable through the cable entry
into the sensor. Screw the sleeve nut back
onto the cable entry and clamp the stripped
wires of the cable into the proper terminal
positions.
The spring terminals hold the wire without a
screw. Press the white opening tabs with a
small screwdriver and insert the copper core
of the connection cable into the terminal
opening. Check the hold of the individual
wires in the terminals by lightly pulling on
them.
Version with plastic housing
Voltage supply and digital
meas. signal
-
+
To the indicating instrument in
Cable entry
M20 x 1.5
the cover or to the external
indicating instrument
Sockets for
connection of
VEGACONNECT 2
(communication sockets)
12C567843
12C5 6 7 8
(+) (-)
Communication
Display
ESC
-
+
VBUS
Terminals
(max. 2.5 mm
wire cross-section)
OK
Sockets for connection of the HART
handheld or VEGACONNECT
pluggable
adjustment
module MINICOM
2
12C5678
12C5678
Commu-
(+) (-)
nication
VBUS
®
Tank 1
m (d)
12.345
Display
+
ESC
-
OK
Opening
tabs
56VEGASCAN 850
Electrical connection
5.3 Connection of the external indicating instrument VEGADIS 50
Loosen the four screws of the housing cover
on VEGADIS 50.
The connection procedure can be facilitated
by fixing the housing cover during connection work with one or two screws on the right
of the housing (figure).
OUTPUT
(to the sensor)
3
2
1
4
5
8
6
7
Adjustment
module
VEGADIS 50
+
-
Tank 1
m (d)
12.345
ESC
OK
Voltage supply and
digital meas. signal
-
+
12C5678
2
1
(+) (-)
VBUS
Tank 1
m (d)
12.345
C5678
Commu
nication
+
-
Display
ESC
OK
DISPLAY
(in the cover of
the indicating
instrument)
Screws
VEGASCAN 85057
Electrical connection
5.4 Connection of the sensors to
VEGASCAN 850
off
on
---
POWER
GNDTxD
N
RxDDTR
FIELDBUSVBUS1 VBUS2
PC2
PE L
VBUS1
Use the terminal VBUS1, when using a
VEGASCAN 850 for 15 sensors. Sensors
16 … 30 are connected to the terminals
VBUS2.
+++
--+++
VBUS2
The electrical connection is made by spring
terminals. The clamping opening can be
opened with a small screwdriver by inserting
it into the opening slot above the clamping
position and pushing upwards. Then the
terminal and the copper core of the connec-
tion wire can be inserted.
Up to 15 sensors can be connected to each
VBUS branch on one two-wire cable.
---
+++
Sensor 1 … 15
VBUS1
It is a good idea to optimise the current distribution by dividing the BUS into three
branches with five sensors each.
Sensor
Sensor
6 … 10
Sensor
11 … 15
1 … 5
---
+++
VBUS1
58VEGASCAN 850
Electrical connection
5.5 Signal output/Interface connection
The connection terminal is labelled for the
field bus cable according to the selected
interface and instrument version.
Modbus, VEGA-ASCII, Siemens 3964
TXRX
DATA
(B/B´)
DATA
(B/B´)
Interface RS 232
/TX/RX
Interface RS 422
Interface RS 485
DATA
(A/A´)
Interface RS TTY
Interface RS 485
DATA
(A/A´)
RxD
GND TxD
FIELDBUS
GND
FIELDBUS
GND
(C/C´)
FIELDBUS
T-
R+ R+
GND T+
FIELDBUS
Profibus DP , Profibus FMS
GND
(C/C´)
FIELDBUS
5.6 Connection of the PC to
VEGASCAN 850
The PC with the adjustment software VVO
can be connected in the front panel or in the
terminal compartment of VEGASCAN.
For connection of the PC in the front panel (9pole socket) you require a standard RS 232
DTE-DTE (Data Terminal Equipment) interface cable.
DCD
RxD
TxD
DTR
GND
...
6
...
7
...
8
...
9
For connection of the PC in the terminal compartment, connect the opened PC cable
according to the following sketches to terminal block PC2. For communication, only three
wires of the 9-pole cable are used.
DCD
1
RxD
2
TxD
3
DTR
4
GND
5
...
6
...
7
...
8
...
9
1
DCD1
2
RxD2
3
TxD3
4
DTR4
5
GND5
...
6
...
7
...
8
...
9
GNDTxD
RxD DTR
PC2
Terminal in VEGASCAN
VEGASCAN 85059
6 Setup
Setup
6.1 Adjustment methods
VEGASCAN 850 and the series 50 ultrasonic
sensors are adjusted with the PC (adjustment program VVO). With the detachable
adjustment module MINICOM, it is possible
to adjust the sensors individually. The adjustment can be carried out with only one adjustment instrument at a time.
With the PC
The PC with the adjustment program VVO
(VEGA Visual Operating) can be connected
to:
- VEGASCAN 850
- the sensor
- the signal cable
With the adjustment program VVO(VEGA
Visual Operating) on the PC you adjust the
level measuring system consisting of
VEGASCAN and ultrasonic sensors. The PC
communicates via the standard RS 232 interface cable with VEGASCAN and all connected sensors. For this purpose, a digital
adjustment signal is superimposed on the
signal and supply cable.
In the following chapter "6.2 Adjustment with
the PC“, the adjustment is described when
connecting the PC via a RS 232 interface
cable to VEGASCAN. When connecting the
PC via VEGACONNECT to the sensor signal
cable (see "2.6 Configuration of measuring
systems“ configuration A), all sensors connected to the affected signal cable are available for adjustment. However, the
configuration and measured data processing
functions of VEGASCAN are not available.
When inserting VEGACONNECT into the
communication socket of an individual sensor, only sensor-relevant adjustment options
such as e.g. sensor optimisation on the affected sensor are adjustable. Other sensors
and VEGASCAN are then not adjustable. This
is due to an adjustment hierarchy beginning
with VEGASCAN and ending with the sensor.
With the adjustment module MINICOM
With the adjustment module MINICOM you
adjust the individual sensor directly in the
sensor or in the external indicating instrument
VEGADIS 50. The adjustment module MINICOM with the 6 key adjustment field with text
display enables the parameter adjustment of
the sensor with an array of functions comparable to the adjustment program VVO when
connected via VEGACONNECT to the sensor, but not the configuration of the measuring system.
Beside the connection of the PC to VEGASCAN, it is possible to connect the PC with
the interface converter VEGACONNECT to
the signal cable between sensor and VEGASCAN or also directly to the individual sensor.
60VEGASCAN 850
Setup
6.2 Adjustment with the PC
In the following setup and adjustment instructions you will find information on the following
topics and adjustment points:
• Configuration
- configuration info
- create new measurement loop
• Parameter adjustment 1
- adjustment
- scaling
• Sensor optimisation
- meas. environment
- sound velocity
- echo curve
- false echo storage
• Parameter adjustment 2 (optional)
- linearisation
- parameter adjustment sensor display
- meas. loop data
• Display of measured value
• Simulation
• Select available linearisation curve
• Print configuration and adjustments
• Backup
Note:
Keep in mind that you first have to log in a
sensor in the configuration menu before you
can continue with the parameter setting, and
if necessary, carry out adjustments in the
menu “Sensor optimisation”.
For connection of the PC to VEGASCAN, a
standard RS 232- DTE-DTE (Data Terminal
equipment) interface cable is required. With
the cable, the PC is connected to
VEGASCAN.
DTEDTE
Beside the measured values, the adjustments signals are also transmitted digitally
via the signal supply cable between sensor
and VEGASCAN. The adjustment program
VVO can then communicate with VEGASCAN
and all connected sensors. In chapter "2.6
Configuration of measuring systems“ the
connection of the PC to the different sensor
arrangements is shown.
Before starting the setup:
Do not be confused by the many pictures,
adjustment steps and menus on the following
pages. Just carry out the setup with the PC
step-by-step and you will soon no longer
need the following pages. Whenever you
should adjust or choose something, this is
indicated in the following by a large black
dot, like this:
• Choose …
• Start …
• Click to …
• Connect the standard output of your PC by
the standard RS 232 interlink cable (9pole) to VEGASCAN.
• Now switch on the power supply of the
processing system.
By this means, the actions to be carried out
are clearly separated from supplementary
information in the following adjustment instructions.
Approx. 1 … 2 minutes after switching on the
supply voltage, the sensors generally show
measured values on the sensor display.
After the adjustment software on the PC has
attained connection to VEGASCAN, you start
with the "Configuration“ of your measuring
system.
DCD11DCD
RxD22RxD
TxD33 TxD
DTR44DTR
GND55GND
---66---
---77---
---88---
---99---
VEGASCAN 85061
• First start the adjustment software VVO on
your PC.
• Choose with the arrow keys or the mouse
the item "
Planning
“ on the entrance screen
and click to "OK“.
In the next window you are asked for the
user identification.
Setup - Adjustment with the PC
If the VVO (adjustment software) gets in
contact with VEGASCAN for the first time, you
are asked if you want to transfer the data
from VEGASCAN to the PC.
• Enter under name "VEGA.
• Also enter "VEGA under password.
The adjustment program VEGA Visual Operating (VVO), called in the following VVO,
gets into contact with VEGASCAN and the
sensors …
… and asks in which mode you want to use
the adjustment software. After a few seconds
the software indicates if and with which sensor a connection exists.
• Click to "
YES
“.
If you want to connect the VVO to a
VEGASCAN, to which in the meantime another PC was connected, you will be asked, if
you want to transfer the saved data from the
actual PC to VEGASCAN or the data of
VEGASCAN to the PC.
When clicking to "
Quit
“, you are immediately
in the main menu window of the adjustment
software VVO.
The preset user identification can be modified
later in the menu "
User access
Configuration/Program/
“.
62VEGASCAN 850
Setup - Adjustment with the PC
Note:
If you get no sensor connection, check the
following:
- Is the sensor being supplied with voltage
from VEGASCAN (min. 17 V)?
- Are you using the correct interface cable?
- Is the RS 232 cable connected to the correct COM-Port?
Change COM-Port
In the menu "
nication
eters of your PC or change the activated
COM-Port.
Configuration/Program/Commu-
“ you can adjust the interface param-
Configuration
During the setup of VEGASCAN you are
confronted with two terms: "Configuration“
and "Parameter adjustment“. The meas. system VEGASCAN is first set up with a configuration and then with a parameter adjustment.
Configuration
The term "Configuration“ means the basic
adjustments of VEGASCAN. You inform
VEGASCAN about the application (level
measurement, gauge, distance …), the
measurement loop name and the DCS output
address of the sensors. The configuration
corresponds to an electronic wiring and
labelling of your VEGASCAN or just which
sensor for what application and where.
Parameter adjustment
After the configuration, you carry out the
parameter adjustment for each individual
sensor. This means, to adjust the sensors to
the respective operating range and to adjust
to the actual application. You inform the sensor which product distance (which level) is
"empty“ and which "full“. This is called adjustment. Here you choose in which physical unit
(volume, mass) and unit of measurement (m3,
gal, liters …) the adjusted measured value
should be outputted. In the submenu "Sensor
optimisation“ you inform the sensor electronics about the actual environment, such as
e.g. quick changes of the measured value,
foam generation, gas stratification, solid or
liquid.
VEGASCAN 85063
Setup - Adjustment with the PC
Configuration info
• Choose the menu "Configuration/Measur-
ing system.
You reach the menu window "
measuring system
“. Here you get all informa-
Configuration
tion on the connected VEGASCAN.
User authorisation
• Save your password with "
Save
“ and you
are again in the initial menu of the configuration.
Communication
• Click in the menu window "
measuring system
Configuration
“ to "
Communication
“.
Unauthorised access to VEGASCAN can be
prevented by a password. For each connection location of the PC, on the D-SUB plug in
the front of VEGASCAN (PC) or fixed connection in the terminal compartment (PC2), a
respective password can be allocated.
• Click to "
64VEGASCAN 850
Access protection activated“.
Here you adjust the interface parameters for
your signal output (here Modbus).
• Click to "
with "
• Click to "
Quit
Save
“ or quit the menu window
“.
VBUS1
“.
Setup - Adjustment with the PC
You get an overview of the sensors connected to the VBUS1 terminal position and
their configuration status.
If a sensor was logged in as described in the
following menu "
status "
Configure
overview.
Create new measurement loop
• Choose the menu "Configuration/Measure-
ment loop/New
• In the menu window "
ment looption
“ with "OK“.
Measurement loop/New
“ is written in the sensor
“.
“, you choose "
Create new measure-
a new applica-
“, the
You are in the menu window "
measurement loop - Application
• Choose the parameter "
ment
“ and the sensor type "
• Click to "
• Choose in the next menu window "
ard level measurement
• Click to "
After a few seconds, the menu window "
Continue
Continue
“.
“.
Create new
“.
Level measure-
Ultrasonic
“ and "
Stand-
no options
“.
“.
Create new measurement loop - Sensor configuration
“ opens.
• Click to "
The small menu window "
tion
VEGASCAN 85065
Sensor coordination
“ opens.
“.
Sensor coordina-
Setup - Adjustment with the PC
• Choose the serial number of the sensor to
be assigned and confirm with "OK“.
In the menu window "
ment loop - Sensor configuration
number and the serial number of the sensor
are displayed instead of "not configured“.
• Now click in the menu window "
Create new measure-
“, the VBUS
-Create new
measurement loop - Sensor configuration
to "
Continue
• Click in the menu window "
“.
Create new
measurement loop - Measurement loop
designation
“ to "
Level
“.
The menu window "
Create new measurement
loop - Measurement loop designation
Enter here the measurement loop number
and a measurement loop description.
• Assign here also one or several PC/DCS
addresses to the actual sensor by clicking
to "
PC/DCS address
“
“.
“ opens.
66VEGASCAN 850
Setup - Adjustment with the PC
• Confirm the adjustment with "OK“.
• Also here confirm with "OK“ and you are
again in the initial window "
Create new
measurement loop - Measurement loop
designation
“, in which the measurement
loop number and the previously entered
measurement loop description are now
displayed.
In the menu "
Configuration/Measuring sys-
tem“ you now see the configured sensors.
Now, you only have to carry out the parameter adjustment for the sensor. The parameter adjustment is described in the following
under Parameter adjustment 1 and Param-
• Click to "OK“ and you are again in the main
menu window.
You have just carried out the configuration
eter adjustment 2. Parameter adjustment 1 is
obligatory, whereas parameter adjustment 2
(after sensor optimisation) must only be carried out if necessary.
settings for one sensor on VEGASCAN.
VEGASCAN 85067
Parameter adjustment 1
Setup - Adjustment with the PC
In the menu "
justment
adjustments.
Adjustment
• Choose the menu "
eter adjustment
want to adjust.
In the opening menu window, you now see
the previously entered measurement loop
names and the measurement loop descriptions. If you have only configured one sensor,
naturally you will have a choice of only one
sensor.
Instrument data/Parameter ad-
“ you carry out all important sensor
Instrument data/Param-
“ and then the sensor you
• Then click to "OK“, the menu window "
strument data parameter adjustment
opens.
• Click to "
Adjustment
“.
In-
“
• Click in the menu window "
"
Min/Max-adjustment
• Click to the sensor or the measurement
loop on which you want to carry out the
parameter adjustment.
68VEGASCAN 850
“.
Adjustment
“ to
Setup - Adjustment with the PC
You can carry out the min./max. adjustment
"with medium“
actual level) or
taking the actual level into consideration, i.e.
with empty vessel).
Generally, you will carry out the adjustment
without medium. Then you are not concerned
about the actual vessel filling during adjustment. When you want to carry out the adjustment with medium, you have to carry out the
min. adjustment with emptied (also partly
emptied) vessel and the max. adjustment
with filled (also partly filled) vessel. It is convenient and quick to carry out the adjustment
without medium, as shown in the example.
• Choose "
• Choose if you want the carry out the ad-
justment in
• Enter a distance for the upper and lower
level and the volume in % corresponding to
each distance.
In the example, the 0 % filling is at a product
distance of 3.400 m and the 100 % filling at a
product distance of 0.500 m.
Note:
The instrument is supplied with the operating
range identical to the sensor measuring
range. After sensor adjustment, the operating
range corresponds to the adjustment. The
sensor can only detect levels within the defined operating range. For the detection of
levels outside the operating range, e.g. new
adjustment, it is necessary to correct the
operating range in the menu "
sation/Meas. environment
(adjustment by means of the
"without medium“
no (adjustment without medium)
meters
(m) or in
(without
feet
(ft).
Sensor optimi-
“
• Confirm your adjustments with "OK and
you are again in the menu window "
ment
“.
• Click in the menu window "
"
Quit
“.
You are again in the menu window "
“.
ment data parameter adjustment
Now the sensor electronics has two characteristics points, one for min. and one for max.,
out of which a linear correlation between
product distance and the percentage filling of
a vessel is generated.
Naturally the characteristics point must not
be at 0 % and 100 %, however the distance
should be as big as possible (e.g. at 20 %
and at 80 %). The difference between the
characteristics points for the min./max. adjustment should be at least 20 mm product
distance. This is a theoretical value, because
if the characteristics points are too close
together, the possible measuring error increases. Ideal would be, to carry out the
adjustment as shown in the example, at 0 %
and at 100 %.
Adjustment
“.
Adjust-
“ to
Instru-
VEGASCAN 85069
Setup - Adjustment with the PC
In the menu "
justment/Conditioning/Linearisation
enter later, if necessary, a correlation between product distance and percent value of
volume other than linear (see later subitem
linearisation).
Scaling
• Click in the menu window "
parameter adjustment
The menu window "
Instrument data/Parameter ad-
“ you can
Instrument data
“ to "
Conditioning
Conditioning
“ opens.
“.
As parameter you can choose
less
(plain numbers),
and distance“
measuring unit (e.g. l, hl). The sensor display
then shows the measured value in the selected parameter and unit.
• Save the adjustments in the menu "
by clicking "OK.
volume, mass, height
and assign an appropriate
"dimension-
Scaling
“
The adjustments are now transferred to the
sensor and you are again in the menu window "
Conditioning
• Click in the menu window "
"
Quit
“.
• Click to "
In the menu "
0 % and 100 % values of the parameter and
their unit. You thereby inform the sensor, e.g.
that at 0 % filling there are still 45 liters and at
100 % filling 1200 liters in the vessel. The
sensor display then shows with empty vessel
(0 %) 45 liters and with full vessel (100 %)
1200 liters.
70VEGASCAN 850
Scaling
Scaling
“.
“ you enter the actual
• Click in the menu window "
parameter adjustment
“.
Conditioning
“ to
Instrument data
“ to "
Quit
“.
Setup - Adjustment with the PC
Sensor optimisation
In the menu "Sensor optimisation“ you can
adapt the sensor to the environment. Here
you carry out special optimising adjustments
and optimise, e.g. by means of the echo
curve, the mounting location of the sensor.
Meas. environment/Operating range
• Choose the menu "
eter adjustment
• Choose in the menu window "
data parameter adjustment
"
Sensor optimisation and click to "
A
“.
• First click to "Meas. environment.
Instrument data param-
“
Instrument
“ the menu item
Sensor
The window "
With the menu item "
can define the operating range of the sensor
deviating from the measuring range (depending on the sensor type) and the "
Max adjustment
range corresponds otherwise to the min./
max. adjustment (span) or the measuring
range.
Generally, it is better to choose the operating
range approx. 5 % bigger than the adjusted
measuring range (span) determined by the
min./max. adjustment.
In the example:
- Min. adjustment to 1.270 m,
- Max. adjustment to 5.85 m.
In the example you would have adjusted the
operating range from 1 m to 6 m.
Meas. environment
“ opens.
Operating range
“. By default, the operating
“ you
Min/
• Save the adjustments and quit the menu
window "
with "
VEGASCAN 85071
Limitation of the operating range
Quit
“.
“
Meas. environment/Meas. conditions
Setup - Adjustment with the PC
• Click in the window "
"
Measuring conditions.
• In the menu window "
you click on the options corresponding to
your application.
• Confirm with "OK.
After a few seconds of saving during which
the adjustments are permanently saved in
the sensor, you are again in the window
"
Meas. environment
Sonic velocity
When measuring in a gas composition deviating from air, e.g. pure nitrogen, the sonic
velocity will change. Without correction, this
would distort the measuring result.
Meas. environment
Measuring conditions
“.
“ to
• Confirm with "OK“.
• Click in the menu window "
ment
“ to "Quit.
You are again in the menu window "
optimisation
Echo curve
“
If, due to vessel installations, you expect
strong false echoes, a correction (if possible)
of the mounting location and orientation (during simultaneous monitoring of the echo
curve) can help localise the false echoes and
reduce their intensity. With the menu item
"
Echo curve
optimisation
intensity of the detected ultrasonic echoes.
“.
“ in the menu window "
“ you can see the form and the
Meas. environ-
Sensor
Sensor
In the menu "
can be adapted to the correct value in case
of another gas composition.
72VEGASCAN 850
Sonic velocity
“ the sonic velocity
Setup - Adjustment with the PC
In the illustration, you see the echo curve
(before correction of the sensor orientation)
with a false echo nearly as large as the product echo.
In the next illustration you then see the echo
curve after optimum directing of the sensor to
the product surface and away from a large
false echo source (sensor axis perpendicular
to the product surface). The false echo,
caused e.g. by a strut, is now reduced by
more than 30 dB and will no longer influence
the measurement.
• Quit the menu "
Echo curve
“ with "Quit.
False echo storage
With the menu item "
the menu "
Sensor optimisation
False echo storage
“ you can au-
“ in
thorise the sensor to save false echoes. The
sensor electronics then saves the false echoes in an internal database and assigns them
a lower level of importance than the useful
echo. The false echo storage should be carried out with empty vessel.
• Click in the menu window "
tion
“ to the menu item "False echo storage.
Sensor optimisa-
VEGASCAN 85073
Setup - Adjustment with the PC
• Now click in the opening menu window
"
False echo storage
oes. A small window opens.
• Enter here the verified product distance
and click to "Create new
You hereby authorise the sensor to mark all
echoes before the product echo as false
echoes. This prevents the sensor from erroneously detecting a false echo as level echo.
• Click to "Show echo curve.
“ to "Learn false ech-
.
• Quit the menu with "Quit.
You are again in the menu window "
optimisation
reset all options out of the menu "
optimisation
• Quit the menu window "
tion
"
Sensor optimisation sensor selection
with "
You are then in the initial menu window "
“. With the menu item "
“ to default.
Sensor optimisa-
“ with "Quit and the menu window
Quit
“.
strument data parameter adjustment
Sensor
Reset
Sensor
“.
“ you
“ also
In-
The curve of the false echo marking and the
real echo curve (top) are shown.
74VEGASCAN 850
Setup - Adjustment with the PC
Parameter adjustment 2
Linearisation
The relation between level and filling volume
is described with so-called linearisation
curves. If, in your vessel, there is a correlation other than linear between level ("
age value
(value of the filling volume), choose in the
menu window "
"
Instrument data/Parameter adjustment/Con-
ditioning
“ of the level) and the filling volume
Conditioning
“.
“ the menu item
Percent-
Beside the two programmed linearisation
curves "
Horizontal cylindrical tank
"
Spherical tank
programmable curves
is a linear correlation between level and volume.
User programmable linearisation curves
• Click to "User programmable curve to
enter your own vessel geometry or a user
programmable filling curve.
• Click to "Edit.
“ you can also enter "
“. Linear means there
“ and
user
• Click in the menu window "
the menu item "Linearisation.
The menu window "
correlation between percentage value of the
level and the percentage value of the filling
volume is preadjusted.
VEGASCAN 85075
Linearisation
Conditioning
“, in which a
“ to
Setup - Adjustment with the PC
The user programmable linearisation curve is
generated by index markers. Each index
marker consists of one value pair. A value
pair is generated out of a value "
and a value "
value
the level. "
Percentage value“
“ represents the percentage value of
Linearised
“ represents the per-
Linearised
. "
Percentage
“
centage vessel volume at a certain percentage value of the level.
In the field "
Transfer measured value
“ the
current level as percent of the adjusted span
is displayed. The measuring span has already been adjusted with the min./max. adjustment. In our example, the measuring
span is 4.58 m and is in the range of 5.85 m
(empty) up to 1.27 m (full), see the following
illustration.
0 m
100 % or 1.27 m
The percentage value of 95.79 % means that
now 4.387 m of the adjusted measuring
span (4.58 m) are reached:
4.58 • 0.9579 = 4.387 m.
The distance (product distance) outputted
by the sensor, in case you have chosen
"
distance
“ as output parameter, is then:
5.85 – (4.58 • 0.9579) = 1.463 m.
If the index markers or value pairs of your
vessel are not known, you have to define the
linearisation curve by incremental filling or
calculate it with the vessel calculation program of VVO (see next page).
Defining the linearisation curve by incremental filling
In the characteristics of the example, you see
four index markers or value pairs. There is
always a linear interpolation between the
index markers. The example vessel consists
of three cylindrical segments of different
height and diameter. The middle segment
has a considerably smaller diameter.
95.79 % or 1.463 m
Span
4.58 m
100 %
4.387 m
(95.79 %)
5.85 m or 0 %
• Click the check box "Show scaled values,
to have the selected measuring unit displayed on the y-axis (bottom left in the
5.85 m meas. distance correspond to 0 %
menu window).
level. 1.27 m meas. distance correspond to
100 % level. The measuring span is therefore
4.58 m (5.85 m – 1.27 m = 4.58 m).
76VEGASCAN 850
Setup - Adjustment with the PC
Index marker 1 is at 0 % filling (
value [%]
), corresponding in the example to
percentage
an actual distance to the product surface of
5.850 m (empty vessel). The volume quantity
is 45 liters (fluid remaining in the vessel).
Index marker 2 is at a filling level of 30 %
(30 % of the meas. distance of
1.270 m … 5.850 m). At a filling level of 30 %,
there are 576 liters in the vessel (in our example).
Index marker 3 is at a filling level of 60 %. At
this filling level there are 646 liters in the vessel.
Index marker 4 is at a filling level of 100 %
(product distance 1.270 m), where 1200 liters
are in the vessel.
Max. 32 index markers can be entered per
linearisation curve (value pairs).
Max.
Min.
100 % (1.270 m) correspond
to 1200 liters
Span (4.58 m)
0 % (5.850 m) correspond to
45 liters
Calculating the linearisation curve
(using previous tank example)
In the menu window "
programmable curve --
vessel calculation program. With the vessel
calculation program you can calculate (using
dimensions from the technical drawings of
the vessel) the correlation of filling height to
filling volume. If the curve is defined this way,
gauging by incremental filling is not necessary - your sensor will output volume as a
function of level.
Linearisation -- user
“ you can start the
• Click to "Calculate.
The tank calculation program starts. In the
top left corner you choose the vessel type
(upright tank, cylindrical tank, spherical tank,
individual tank form or matrix). When choosing matrix, you can enter a user-programmable linearisation curve by means of index
markers. This corresponds to the input of
value pairs (linearisation points), as previously described.
In the following example, the tank calculation
program calculates the linearisation curve of
a vessel, corresponding to the vessel in the
previous gauging example.
• Click to individual tank form and choose
three round tank segments with the dimensions 0.9 m • 0.9 m (height by diameter),
0.68 m • 0.37 m and 0.68 m • 1.02 m (this
tank form corresponds to the tank form of
the gauging example).
.
VEGASCAN 85077
Setup - Adjustment with the PC
• Click to
After a short calculation time, the levels in
percentage of span and the respective volume percentage will be shown. The outputted
curves show the correlation in a diagram.
• Quit the linearisation table with "OK.
"Calculate
.
You are again in the menu window "
calculation
• Click to "OK to save the tank calculation.
You are again in the menu window "
“.
tion -- user programmable curve --
volume percent values to the percentage
values of the level are shown. When clicking
the bottom left menu window "
values
“, liters will be displayed depending on
the adjustment in the menu "
Tank
Linearisa-
“. The
Show in scaled
Instrument data/
Parameter adjustment/Conditioning/Scaling
Calculate cylindrical tank
• Click in the menu window "
user programmable curve --
"
Calculate and in the menu window "
calculation
tanks.
“ to the symbol for cylindrical
Linearisation --
“ to
Tank
“.
78VEGASCAN 850
Setup - Adjustment with the PC
The menu window for the adjustment of the
cylindrical tank opens.
• Choose the measuring unit which should
be used for the input of the vessel dimensions, e.g. mm.
The following example shows a cylindrical
tank, which inclined by 3°, has a cylinder
length of 10000 mm, at a diameter of
5000 mm. The cylindrical tank has a 1500 mm
wide, spherical form at the right end and a
dished tank end at the left.
Above the information "
internal dimensions
All dimensions are
“, you will find two fields
with the percentage values 0 % and 100 %.
Here you can shift the 100 % line or the 0 %
line. In the example, the 100 % filling line was
defined at a distance of 650 mm from the
upper vessel edge (inner).
• Click to "Calculate.
You will get the calculated linearisation table
after a short calculation time. By means of 32
linearisation points, a function correlating
vessel volume to filling height is outputted.
The example vessel has a filling of 216561
liters at the 100 % line or 216.6 m3. It is possible to output the volume in barrels, gallons,
cubic yards or cubic feet.
Note:
In the bottom left corner in the menu window
"
Tank calculation
dimensions are internal dimensions
“ you find the information "
“. The
All
calculation programs calculates by means of
the vessel inner dimensions the vessel volume. The adjustment of the wall thickness is
not necessary for the calculation of the
dished boiler end as its mathematical calculation is based on the outer dimension.
VEGASCAN 85079
There is a linear interpolation between the
linearisation points.
• Click to "OK and you are again in the
menu window "
• Again click in the menu window "
culation
sation menu.
Here the calculated linearisation curve is
again outputted. The volume information
under "
Linearised
longer to the calculated volume of the tank
calculation program. Why?
In the menu "
tioning/Scaling) you entered that at 0 % filling
there are 45 liters in the tank and at 100 %
filling 1200 liters. The geometry of the calculated cylindrical tank is reduced to a size that
has a volume of only 1200 liters. The linearisation curve of the calculated vessel is therefore modified to fit the volume data you
entered in the menu "
If the true content of the calculated vessel
should be outputted, the volume that was
determined in the tank calculation program
must be entered in the menu "Scaling“.
Tank calculation
“.
Tank cal-
“ to "OK and you are in the lineari-
“ now corresponds no
Scaling
“ (Instrument data/Condi-
Scaling
“.
Setup - Adjustment with the PC
The sensor then outputs the actual filling
volume by means of the adjusted vessel
dimensions.
• Quit the menu with "OK.
• Confirm with "OK and your individual lin-
earisation curve is saved in the sensor.
Again in the menu window "
can enter with the menu item "
time
“ a measured value integration. This is
recommended for agitated product surfaces,
to prevent the measured value indication and
output from changing constantly.
As a standard feature, an integration time of
0 seconds is preset.
• Quit the menu with "OK., you are again in
the menu window "
eter adjustment
• Quit the menu window with "OK.
“.
Conditioning
“, you
Integration
Instrument data param-
80VEGASCAN 850
Setup - Adjustment with the PC
Parameter adjustment sensor display
In the menu item "Outputs“ you choose the
scale and the unit in which your level should
be displayed.
• Choose in the main menu window "
ment data/Parameter adjustment
the menu item "
Outputs
“.
Instru-
“ and then
Meas. loop data
• Click in the window "
rameter adjustment
• Then click to "
and "
VEGASCAN
tion on your measurement loop in the information windows.
Instrument data pa-
“ to "
Meas. loop data
Application
“, "
“, to get detailed informa-
Input no.A
“.
“
• In the menu window "
"
Display of measured value
• Choose the sensor and carry out the adjustments.
• If you want to leave the adjustments unchanged, click to "
• Click in the menu window "
"
Quit
“ and you are in the menu window
"
Instrument data parameter adjustment
• Click in the menu window "
parameter adjustment
VEGASCAN 85081
Outputs
Quit
“.
“ again to "
“ you click to
“.
Outputs
“ to
“.
Instrument data
Quit
“.
• Close the information window.
• Quit the menu "
• Click in the menu window "Instrument data
parameter adjustment“ to "
You are again in the main menu window.
Meas. loop data
Quit
“.
“.
Setup - Adjustment with the PC
Show measured value
• Click in the main menu window to the menu
"
Display/Display of measured value
choose the measurement loop or the sensor to be displayed.
• Choose in the line "
output "
Distance
distance will be displayed. When choosing
"
Scaled
“, e.g. the measured value will be
displayed in liters and volume percent as
well as the actual signal current in the
0/4 … 20 mA signal cable.
Indication value
“ and the sensor product
“ and
“ the
Simulation
• Click to the menu "
and choose the measurement loop.
The menu window "
similar to the previous menu window, opens.
In this menu window however, you can set
the filling of the vessel or the signal current
and the indication to any value (simulate
measured value).
First of all, the actual measured value and the
signal current are displayed.
• Click to "Start in the turquoise window
segment.
Diagnostics/Simulation
Simulation of outputs
“,
82VEGASCAN 850
Setup - Adjustment with the PC
The grey scrollbar becomes active. With this
scrollbar your can modify the measured
value in the range of -10 % … 110 % and
thereby simulate the filling or emptying of the
vessel. In the field of the turquoise window
cutout you can enter any % value for the
filling percentage.
Choose available linearisation curve
On the previous pages, the entering or the
calculation of a linearisation curve was shown
as an example. In this menu item, you can
choose from the available predefined or userentered linearisation curves.
Note:
The simulated measured value is outputted
during adjustment with the PC until you terminate the simulation mode.
A selected curve can be modified later (edit)
or calculated with the calculation program.
VEGASCAN 85083
Print configuration and adjustments
Setup - Adjustment with the PC
• Click to "
Before printing the complete configuration of
VEGASCAN and all sensors, you can have a
look at the individual pages and …
Services/Print
“.
With the menu "
justments displayed in detail.
View
“ you can have the ad-
… then print everything or just certain pages.
84VEGASCAN 850
Setup - Adjustment with the PC
Backup
With the menu items "
conditioning instruments
Backup/Sensors
and parameter settings of VEGASCAN and of
each sensor.
The settings can later be quickly transferred
to new sensors, e.g. in case of a sensor
exchange, or certain product settings can be
quickly carried out. This is done in the menu
"
Services/Restore configuration
For further information take the manual "VEGA
Visual Operating“ (VVO).
Services/Backup/Signal
“ and "
“ you save the configuration
Services/
“.
VEGASCAN 85085
Setup - Sensor adjustment with the adjustment module MINICOM
6.3 Sensor adjustment with the adjustment module MINICOM
Tank 1
m (d)
12.345
Beside the PC, the ultrasonic sensors VEGASON 51V … 56V can also be adjusted
with the small, detachable adjustment module
MINICOM in the sensor.
Only the sensor-relevant adjustments such
as e.g. scaling of the sensor display, operating range, meas. conditions, sensor display
scaling or false echo storage are possible
with the adjustment module MINICOM. Not
possible are all adjustment steps relating to
the configuration, the conditioning and the
signal processing (configuration of the inputs
and outputs, linearisation curves, simulation
…). This is only possible with the PC directly
on VEGASCAN 850.
You carry out all adjustment steps with the 6
keys of the adjustment module. A small display shows you, apart from the measured
value, a short message on the menu item or
on the value of a menu adjustment.
The information volume of the small display,
however, cannot be compared with that of the
adjustment program VVO, but you will soon
get used to it and will be able to carry out
your adjustments quickly and directly with
the small MINICOM (see menu plan on the
following pages).
ESC
+
-
OK
1. Measurement in gases
2. Operating range
3. Meas. conditions
4. False echo storage (only necessary, if
failures occur during operation).
5. Indication of the useful and noise level
6. Processing/Scaling of the sensor display
In the following you will find the sensor-specific setup items 1 … 6.
1. Measurement in gases
Adjustment is only necessary when the
measurement is made in gases (Co2, He,
etc.) deviating from air. In case of measurement in gases, sound the distance of the
sensor to the product surface and enter it in
the menu item "Measurement in gases“. The
sensor can then take the modified sonic
velocity in gases (as opposed to air) and
output correct levels.
2. Operating range
Without special adjustment, the operating
range corresponds to the measuring range.
The measuring range has already been adjusted with the min./max. adjustment. Generally it is useful to choose a slightly bigger
(approx. 5 %) operating range than the
measuring range.
Example:
Min./max. adjustment: 0.500 … 3.500 m;
adjust operating range to approx.
0.400 … 3.600 m.
86VEGASCAN 850
Setup - Sensor adjustment with the adjustment module MINICOM
3. Meas. conditions
Here you inform each sensor under which
application conditions the measurement is
carried out, e.g. measurement in gas compositions deviating from air, which will influence
the running time of the ultrasonic signals or if
a liquid (hard echo) or a solid (weak echo)
should be measured.
4. False echo storage
A false echo storage is always useful when
unavoidable false echo sources (e.g. struts)
must be minimised. By creating a false echo
storage, you authorise the sensor electronics
to note the false echoes and save them in an
internal database. The sensor electronics
treats these (false) echoes differently from
the useful echoes and filters them out.
5. Useful and noise level
In the menu
you get important information on the signal
quality of the product echo. The higher the
"S-N“ value, the more reliable the measurement (menu plan MINICOM).
Ampl.:
XX dB
S-N:
XX
dB
Example:
Ampl. = 68 dB
S-N = 53 dB
68 dB – 53 dB = 15 dB noise level
53 dB signal level difference indicates very
high measurement reliability.
6. Processing/Scaling of the sensor
display
(see menu plan)
Ampl.: means amplitude of the level echo in
dB (useful level)
S-N:means Signal-Noise, i.e. the useful
level minus the level of the background noise
The bigger the "S-N“ value (difference between the amplitude useful level and the
noise level), the better the measurement:
> 50 dBMeasurement excellent
40 … 50 dBMeasurement very good
20 … 40 dBMeasurement good
10 … 20 dBMeasurement satisfactory
5 … 10 dBMeasurement sufficient
< 5 dBMeasurement poor
VEGASCAN 85087
Setup - Sensor adjustment with the adjustment module MINICOM
Menu schematic of the adjustment module MINICOM
Sensor
m(d)
4.700
SON52
V
1.00
When switching on, the sensor type and the
software version are displayed for a few
seconds.
Parameter
2.
Sensor
optimise
Meas.
enviro
nment
Operating
rage
Begin
m (d)
0.50
End
m (d)
6.00
3.
Meas.
condit
ions
Condit
ion
Liquid
Fast
change
No
Measur
ing in
gas
Meas.
dist.
m (d)
4.700
Correc
tion
Now !
OK ?
Agitat
ed sur
face
No
1.
Foaming
prod.
No
Measur
ing in
gas
No
4.
False
echo
memory
Create
new
Meas.
dist.
m (d)
X.XX
Echo
learn
Now!
OK ?
learning
Multiple
echoes
No
Update
Meas.
dist.
m (d)
X.XX
Update
Now!
OK ?
learning
Configuration
Sensor
Tag
Sensor
‘Delete
Delete
Now!
OK ?
deleting
Meas.
unit
m (d)
High
Condit
ion
Solid
Fast
change
No
dust
level
No
Large
angle
repose
No
Measur
ing in
gas
No
Multiple
echoes
No
6.
Displ.
adjust
ment
Adjust
ment
in
m(d)
0.0%
at
m (d)
XX.XXX
Signal
condit
ioning
Scaling
0 %
corres
ponds
XXXX
100 %
corres
ponds
XXXX
Decimalpoint
8888
Prop.
to
Volume
Unit
l
88VEGASCAN 850
Outputs
Sensor
display
Prop.
to
Distance
Setup - Sensor adjustment with the adjustment module MINICOM
With these keys you move in
the menu field to the left, right,
top and bottom
ESC
Basic
Reset
Reset
Now!
OK ?
reseting
Distance
m (d)
4.700
5.
Ampl.:
XXdB
S-N:
XX
max.
range
m (d)
OK
7.000
Add´l
functions
Basic
Info
reset
Reset
dB
Sensor
Tag
Sensor
Now !
Reseting
Sensor
typ
SON
52 V
OK ?
Serial
nubmer
1094
8218
Language
English
Softw.
Vers.
3.00
Softw.
Date
11.09.
1997
Ampl.:
Meas.
dist.
X,XX
Fast
change
No
Foaming
prod.
No
Menu items in bolt print provide
sensor and measured value
information and cannot be modified
in this position.
Light grey menu fields are only
displayed if required (dependent on
the adjustments in other menus).
White menu items can be modified
with the "+“ or "–“ key and saved
with the "OK“ key.
Distance
m (d)
3.700
S-N:
71dB
48
dB
Temperature
Actual
temp.
Max.
Min.
temp.
temp.
°C
18
°C
°C
59
4
VEGASCAN 85089
7 Diagnostics
Diagnostics
7.1 Simulation
For simulation of a certain filling, you can call
up the function "Simulation“ on the adjustment
module MINICOM, in the software program
VVO or on the signal conditioning instrument.
You simulate a vessel filling and thereby a
certain sensor current. Please note that connected instruments, such as e.g. a PLC react
according to their adjustments and will probably activate alarms or system functions.
Simulation with VVO
If you start the simulation mode with the adjustment program VVO on the PC, the simulated level is outputted until you quit the
simulation mode.
Simulation with MINICOM
If you start the simulation mode on the adjustment module MINICOM, the sensor returns to
standard operating mode after one hour.
7.2 Error codes
Error codesRemoval
E013 No valid measured valueMessage is displayed during the warm-up phase
- Sensor in the warm-up phase
- Loss of the useful echo
If the message remains, a false echo storage
(with the adjustment software on the PC - see
“Echo curve” under “Sensor optimisation”) must
be carried out together with a modification of
mounting location and orientation to achieve the
lowest possible false echo background.
E017 Adjustment span too smallCarry out a readjustment.
E036 Sensor software does not runSensor requires new software (service).
E040 Hardware failure/Electronics defec- Check all connection cables.
tiveTransducer defective.
90VEGASCAN 850
Make sure that the difference between
min. and max. adjustment is at least 10 mm.
Message appears during a software update.
Contact our service department.
Notes
VEGASCAN 85091
VEGA Grieshaber KG
Am Hohenstein 113
D-77761 Schiltach
Phone (0 78 36) 50 - 0
Fax (0 78 36) 50 - 201
e-mail info@vega-g.de
internet www.vega-g.de
ISO 9001
The statements on types, application, use and operating conditions of
the sensors and processing systems correspond to the latest information at the time of printing.
Technical data subject to alterations
2.24 631 / March 2000
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.