VEGA SCAN850 User Manual

Operating Instructions
VEGASCAN 850
Level and Pressure
Contents
Safety information ........................................................................ 2
Note Ex area ................................................................................ 2
1.1 Function ................................................................................. 4
1.2 Application features ............................................................. 6
1.3 Adjustment ............................................................................ 6
2 Types and versions
2.1 Type overview, VEGASCAN 850 ........................................ 8
2.2 Type overview, sensors....................................................... 8
2.3 Type code, VEGASCAN 850............................................. 11
2.4 Type code, sensors ........................................................... 12
2.5 Approvals ........................................................................... 13
2.6 Configuration of measuring systems ............................... 13
3 Technical data
3.1 Data ..................................................................................... 18
3.2 Dimensions ......................................................................... 25
Contents

Safety information

Please read this manual carefully, and also take note of country-specific installation standards (e.g. the VDE regulations in Germany) as well as all prevailing safety regulations and acci­dent prevention rules.
For safety and warranty reasons, any internal work on the instruments, apart from that in­volved in normal installation and electrical con­nection, must be carried out only by qualified VEGA personnel.
2 VEGASCAN 850

Note Ex area

Please note the approval documents (yellow binder), and especially the included safety data sheet.
Contents
4 Mounting and installation
4.1 Mounting VEGASON 54 … 56 .......................................... 34
4.2 General installation instructions ........................................ 36
4.3 Measurement of liquids ..................................................... 39
4.4 Measurement of solids ...................................................... 42
4.5 Socket extensions ............................................................. 44
4.6 Flow measurement ............................................................. 46
4.7 False echoes ...................................................................... 47
4.8 Incorrect mounting ............................................................. 49
5 Electrical connection
5.1 Connection, connection cable and shielding ................... 53
5.2 Connection of the sensor .................................................. 56
5.3 Connection of the external indicating instrument
VEGADIS 50 ....................................................................... 57
5.4 Connection of the sensors to VEGASCAN 850 ............... 58
5.5 Signal output/Interface connection ................................... 59
5.6 Connection of the PC to VEGASCAN 850 ........................ 59
6 Setup
6.1 Adjustment methods .......................................................... 60
6.2 Adjustment with the PC ...................................................... 61
6.3 Sensor adjustment with the adjustment
module MINICOM............................................................... 86
7 Diagnostics
7.1 Simulation ............................................................................ 90
7.2 Error codes ........................................................................ 90
VEGASCAN 850 3

1 Product description

1.1 Function

Continuous level measurement with ultrasonic sensors is based on the running time meas­urement of ultrasonic pulses.
VEGASON series 51 … 56 sensors are a newly developed generation of extremely compact ultrasonic sensors for level meas­urement. They were especially developed for liquids (51 ... 53) and for solids and larger measuring ranges (54 ... 56).
Due to the small housing dimensions and process fittings, the compact sensors are a very reasonable solution for your level meas­urement applications. With the integrated display and a special sensor intelligence, in conjunction with large measuring ranges, they can be used for applications in which the advantages of non-contact measurement could never before be realized.
As output and measuring signal, the sensors produce a digital output signal which is proc­essed and evaluated in VEGASCAN 850. VEGASCAN 850 then outputs the processed measured values as a digital communication protocol.
Measuring principle
Piezoceramic high-performance transducers emit focused ultrasonic pulses which are reflected by the product surface. The meas­urement electronics prepares a precise im­age of the environment out of the reflected ultrasonic pulses. The transducers work both as transmitter and receiver. As receiver, the transducers are high-sensitivity piezo micro­phones.
Product description
Meas. dis­tance
emission - reflection - reception
The measurement electronics precisely cal­culates the distance between transducer and medium from the speed of sound and the measured running time of the emitted sound impulse. The distance is then converted into a level-proportional signal, and in conjunction with the sensor parameter settings, made available as precise, calibrated level value.
Since the speed of sound is subject to tem­perature influence, the transducer also con­tinuously detects the ambient temperature, so that the level is precisely measured even in case of varying ambient temperature.
4 VEGASCAN 850
Product description
Output signal
The level-proportional measuring signal of the sensors is received, processed and outputted digitally throughout. VEGASCAN 850 reads in the digital sensor measured values permanently and processes the indi­vidual levels according to the user’s selected options. Depending on the version of VEGASCAN, max. 15 sensors can be con­nected to one screened two-wire cable or 30 sensors on two screened two-wire cables. The two-wire cable transfers beside the dig­ital level signal and adjustment signals also the supply voltage of 24 V.
The sensor measured values can be proc­essed and outputted in VEGASCAN as levels in percent, volume or mass units. In addition, linked process tasks such as scaling, lineari­sation, differential generation, addition, ten­dency processing or limit value processing can be selected individually. As a result, VEGASCAN can carry out - decentralised and without costly programming - the entire measurement-related arithmetic.
The levels and their processed results are then outputted in the following field bus pro­tocols and interfaces depending on the or­dered version of VEGASCAN 850.
The digital processing of the measured si­gnal ensures an accuracy that an analogue measuring signal could never reach, as the digital signal is always transferred without error right down to the last bit and decimal point position.
Varying line resistance or small leakage cur­rents do not influence the accuracy (digital technology). The digital signal is always un­ambiguous.
Display of measured values
As an option, the series 50 ultrasonic sensors can be equipped with an indicating instru­ment for direct, local level survey. The indi­cating instrument shows the precise level by means of the analogue bar graph and the digital number value. In addition to the indica­tion in the sensor, you can have the level displayed by the VEGADIS 50 external indi­cating instrument at a distance of up to 25 m from the sensor. The external display of measured values operates, like the inte­grated display, independently of the output signal, can be modified through individual parameter settings, and is connected to and powered by the sensor.
Protocol Interface
Siemens 3964 - RS 232
VEGA-ASCII - RS 232
Modbus - RS 232
Profibus DP - RS 485
Profibus FMS - RS 485
VEGASCAN 850 5
- RS 422
- RS 485
- TTY
- RS 422
- RS 485
- TTY
- RS 422
- RS 485
- TTY
Product description

1.2 Application features

Applications
Level measurement of all liquids
Level measurement of solids (only short
measuring distances) such as e.g. coal, ore, stones, crushed rocks, cement, gravel, sand, sugar, salt, cereals, flour, granules, powder, dusts, sawdust, sawings.
Flow measurement on various flumes
Gauge measurement, distance measure-
ment, object monitoring and conveyor belt monitoring
Two-wire technology
Supply and output signal on one two-wire cable (loop powered).
Output signal and signal processing com­pletely digital.
Rugged and precise
Measurement unaffected by substance properties such as density, conductivity, dielectric constant
Suitable for corrosive substances
Measuring range 0.25 m … 70 m.
Precise through digital processing and
transmission of measured values.
Means of adjustment
With adjustment software VEGA Visual Operating (VVO) on the PC
With detachable adjustment module MINICOM
With VEGAMET signal conditioning instru­ment.

1.3 Adjustment

Each measuring situation is unique. For that reason, every ultrasonic sensor needs some basic information on the application and the environment, e.g. an empty vessel profile is important for a reliable measurement. Beside this, many other settings and adjustments are possible on VEGASON ultrasonic sen­sors.
The adjustment and parameter setting of VEGASCAN 850 and ultrasonic sensors are carried out with the PC and the adjustment program VVO (VEGA Visual Operating).
Only sensor-relevant settings can be carried out with the adjustment module MINICOM.
Adjustment with PC
The program leads quickly through the ad­justment and parameter setting by means of pictures, graphics and process visualisations.
The PC is connected with a standard cable (RS 232) directly to VEGASCAN 850.
The adjustment and parameter data of the connected sensors and the configuration of VEGASCAN can be saved with the adjust­ment software on the PC and can be pro­tected by passwords. On request, the sensor adjustments can be quickly trans­ferred to other sensors.
Display of measured value
Display of measured value integrated in sensor.
Optional display module separated from sensor.
Process fittings
G 1 A, DN 50, DN 80, DN 200, DN 250
G 11/2A, 11/2“ NPT
G 2 A, 2 NPT
Compression flange DN 100, ANSI 4
Approvals
CENELEC, ATEX, PTB, FM, CSA, ABS, LRS, GL, LR, FCC
6 VEGASCAN 850
Product description
Adjustment with adjustment module MINICOM
With the small (3.2 cm x 6.7 cm) 6-key ad­justment module with display, the adjustment can be carried out in clear text dialogue.
The adjustment program recognises the sensor type and the location of the connection
Visualised input of a vessel linearisation curve
2
P
...
C
8 5 0
BA
on
1...15
Tank 1 m (d)
12.345
ESC
+
-
OK
Detachable adjustment module. The adjustment module can be plugged into the ultrasonic sensor or onto the external indicating instrument VEGADIS 50.
The adjustment module can be plugged into the ultrasonic sensor or into the optional, external indicating instrument. Unauthorised sensor adjustments can be prevented by removing the adjustment module.
Adjustment with the PC on VEGASCAN 850 with the standard cable RS 232 (with VEGASCAN up to 15 sensors can be operated on one two-wire cable)
VEGASCAN 850 7

2 Types and versions

Types and versions

2.1 T ype overview, VEGASCAN 850

Either 15 or 30 series 50 ultrasonic sensors can be connected to VEGASCAN 850. The measured data are outputted in the ordered field bus protocol by the ordered interface. The field bus protocol and interface type are determined by the ordered version of VEGA­SCAN. The type label of VEGASCAN contains the bus protocol and the interface type (see chapter "2.3 Type code of VEGASCAN 850“).
Optional field bus protocols:
- Siemens 3964
- Modbus
- VEGA-ASCII
- Profibus DP
- Profibus FMS
Optional interface types:
- RS 232
- RS 422
- RS 485
- TTY

2.2 T ype overview, sensors

VEGASON series 51 … 56 sensors are a newly developed generation of extremely compact ultrasonic sensors for large measur­ing ranges (VEGASON 54 ... 56) or for shorter measuring ranges (VEGASON 51 53).
Due to the small housing dimensions and process fittings, the compact sensors do your level monitoring inconspicuously, and above all, at reasonable cost.
Because of price, reliability and easy han­dling, ultrasonic level measurement can be now used in applications in which non-con­tact level measurement could never be used before.
VEGASON 50 ultrasonic sensors utilise two­wire technology perfectly. The supply voltage and the digital output signals of 15 sensors are transmitted between VEGASCAN 850 and the sensors via one two-wire cable.
Swivelling holders allow quick alignment of the transducer to the product and solid sur­face. Mounting is simplified by the option of separating the sensor electronics from the transducer. The sensor electronics can be mounted at a distance of 300 m from the transducer. It is then possible to mount the transducer in environments with an ambient temperature up to 150°C (type 56).
8 VEGASCAN 850
Types and versions
Common features
Application to solids and liquids.
Measuring range 0.25 m … 70 m.
Ex approved in Zone 1 (IEC) and Zone 1
(ATEX) classification EEx ia [ia] IIC T6.
Display module integrated in the sensor or in the external indicating instrument sepa­rated up to 25 m from the sensor.
VEGASON 51
VEGASON 52
VEGASON 54
Version A Version B Version C Version D
VEGASON 55
Version A Version B Version C Version D
VEGASON 56
VEGASON 53
Version A Version B Version C Version D
VEGASCAN 850 9
Types and versions
Short overview of sensor features
Use in solids and liquids.
Measuring range 0.25 … 70 m.
Ex approved in Zone 1 (IEC) or Zone 1 (ATEX) classification.
Integrated display of measured values.
VEGASON 51V 52V 53V 54V 55V 56V
Signal output
digital meas. signal ••••••
Power supply
- two-wire technology ••••••
Process fitting
- G 1 A, 1 NPT –––•••
-G 11/2 A, 11/2“ NPT •–––––
- G 2 A, 2 NPT –•––––
- DN 100 compr. flange ––•–––
- DN 50 –––•••
- DN 80 –––•••
- DN 200 –––•••
- DN 250 –––•••
Adjustment
- with PC and adjust­ment software ••••••
- with adjustment mod­ule in sensor ••••••
- with adjustment mod­ule in ext. indicating instrument ••••••
Meas. range in m
- liquid 0.25 4 0.4 7 0.6 15 1 25 0.8 45 1.4 70
- solid 0.3 2 0.5 3.5 0.75 71 25 0.8 45 1.4 70
10 VEGASCAN 850
Types and versions

2.3 T ype code, VEGASCAN 850

VEGASCAN 850 X X X X X
K - Cable entry M20 x 1.5 N - Cable entry
A - 230 V AC power supply B - 115 V AC power supply
1 - RS 232 interface 2 - RS 422 interface 3 - RS 485 interface 4 - TTY interface
A - Siemens S5 (3964 R procedure with RK 512) B - Modbus (RTU and ASCII) S - Profibus FMS P - Profibus DP N - VEGA-ASCII
B - up to 15 sensors can be connected C - up to 30 sensors can be connected
1
/2“ NPT
VEGASCAN 850 11

2.4 T ype code, sensors

Types and versions
VEGASON 54 V EX.XX X X X X X X X
K - Plastic housing PBT, M20 x 1.5 cable entry N - Plastic housing PBT,
1
/2“ NPT cable entry
A - Aluminium housing, M20 x 1.5 cable entry
G - Process fitting G 11/2 A N - Process fitting 11/2“ NPT X - Process fitting DN 100 PN (without sleeve nut) A - Process fitting DN 100 PN (PPH compression flange) B - Process fitting DN 100 PN (1.4571 compression flange) C - 1.4301 mounting strap FEP - Version A, flange DN 200 (PP) FFA - Version A, flange DN 200 (Aluminium) SAS - Version B, flange swivelling holder DN 50 SBS - Version B, flange swivelling holder DN 80 GAS - Version C, flange swivelling holder DN 50 GBS - Version C, flange swivelling holder DN 80 RGS - Thread G 1 A YYY - Other process fittings
X - without display A - with integral display
X - without adjustment module MINICOM B - with adjustment module MINICOM (integrated)
A - 20 72 V DC; 20 250 V AC; 4 20 mA (four-wire) B - 20 72 V DC; 20 250 V AC; 4 20 mA, HART
®
(four-wire) E - Power supply via signal conditioning instrument G - Segment coupler for Profibus PA P - 90 250 V AC (only in USA) N - 20 36 V DC, 24 V AC (only in USA) Z - Power supply via signal conditioning instr. (only in USA)
.X - without approval EX.X - Ex approved CENELEC EEx ia IIC T6 EXS.X - StEx Zone 10
K - Analogue 4 … 20 mA output signal (two-wire or four-wire
technology) V - Digital output signal (two-wire technology - VBUS) P - Digital output signal (two-wire technology - Profibus)
Type 51 - Meas. range 0.25 4 m Type 52 - Meas. range 0.4 7 m Type 53 - Meas. range 0.6 15 m Type 54 - Meas. range 1.0 25 m Type 55 - Meas. range 0.8 45 m Type 56 - Meas. range 1.6 70 m
Meas. technology (SON for ultrasonic)
12 VEGASCAN 850
Types and versions

2.5 Approvals

When using ultrasonic sensors in Ex areas or in marine applications, the instruments must be suitable and approved for the explosion zones and application areas. The suitability is tested by approval authori­ties and is certified in approval documents.
VEGASON 50 ultrasonic sensors are ap­proved for Ex zone 1, 10, 11, 21 and 22. Please note the attached approval docu­ments when using a sensor in Ex area.

2.6 Configuration of measuring systems

A measuring system consists of one to 15 (30) sensors and a VEGASCAN 850. The processing unit (VEGASCAN 850 signal conditioning instrument) evaluates the level­proportional, digital measuring signals in a number of processing routines and outputs the levels as process bus signal.
Beside the output of the levels in percent, cubic meters and other physical units, the levels can be also processed by linked processing algorithms. Scaling, linearisation, calculation of linearisation curves, differential generation, addition or tendency processing are implemented in VEGASCAN 850 as intrin­sic processing routines and are easily ac­cessible via the menu.
On the following pages you will find three instrument configurations (measuring sys­tems) consisting of sensor(s) and process­ing unit.
Note:
It is possible to operate up to 15 sensors on one two-wire cable, see following page (con­figuration A). However, it would be more suitable to plan the measuring system in such a way that max. five sensors are oper­ated on one two-wire cable. For this reason, VEGASCAN is equipped with three clamping positions for a VBUS input. The sensor ar­rangement can either be made linearly (con­figuration B) or radially (configuration C).
VEGASCAN 850 13
Types and versions
Configuration A Up to 15 sensors via one two-wire cable on VEGASCAN 850
Two-wire technology, 15 sensors with power supply and digital output signals via one two­wire cable on VEGASCAN 850 possible. However, it is better to wire in groups of five as in configuration B and C.
2
1)
Sensor cables should be screened. Grounding of
4
2
2
the cable screens at both ends is recommended. However make sure that no earth compensation currents flow via the screens (see chapter "5.1 Connection, connection cable and screening). Earth compensation currents can be avoided by potential equalisation lines or if the cable screen is grounded at both ends - by connecting one end via a capacitor (e.g. 0.1 µF; 250 V) to earth potential.
4
2
2
VEGACONNECT 2
4
2
14 VEGASCAN 850
Types and versions
Display of measured value integrated in sensor.
Optional external indicating instrument (can be mounted up to 25 m separated from the
sensor, also in Ex area).
Adjustment with PC or adjustment module MINICOM (can be plugged into the sensor or the external indicating instrument VEGADIS 50).
Max. resistance of the signal cable 10 W per wire or 1000 m cable length.
screened sensor cable 1)
2
4
2
2
2)
4
2
2
4
2
2)
Sensor cables leading to the same VBUS input can be looped together in a screened multiple wire cable. Sensor cables leading to another VBUS connection must be looped in a separate, screened cable.
VEGASCAN 850 15
Types and versions
Configuration B VEGASCAN 850 B; 15 sensors can be connected
Two-wire technology, 3 x 5 sensors grouped in line on three two-wire cables.
Display of measured value integrated in sensor.
Optional external indicating instrument (can be mounted up to 25 m separated from sensor
also in Ex area).
Adjustment with PC or adjustment module MINICOM (can be plugged into the sensor or the external indicating instrument VEGADIS 50).
Max. resistance of the signal cable 15 W per wire or 1000 m cable length.
PC
8 5 0
BA
on
2
2
2
16 VEGASCAN 850
Types and versions
Configuration C VEGASCAN 850 B; 15 sensors can be connected
Two-wire technology, 3 x 5 sensors grouped radially on three two-wire cables.
Display of measured values integrated in the sensor.
Optional external indicating instrument (can be mounted up to 25 m separated from the
sensor also in Ex area).
Adjustment with PC or adjustment module MINICOM (can be plugged into the sensor or external indicating instrument VEGADIS 50).
Max. resistance of the signal cable 15 W per wire or 1000 m cable length.
PC
8 5 0
BA
on
2
4
2
2
2
4
2
4
2
VEGASCAN 850 17
Technical data

3 Technical data

3.1 Data

Power supply
Supply voltage
- VEGASCAN 850 230 V AC, 50/60 Hz
- sensors sensor power supply is provided via the
Min. sensor voltage 17 V Power consumption
- VEGASCAN 850 max. 70 VA
- VEGASON 51 53 81 mW
- VEGASON 54 56 max. 1.5 W peak power
Current consumption
- VEGASON 51 53 6 mA
- VEGASON 54 56 90 mA
Resistance of the signal cable per measuring data input dependent on the number of connected sensors
115 V AC, 50/60 Hz
VEGASCAN 850 processing system with max. 1 15 sensors per two-wire cable
approx. 1 W continuous power
(power) per connection cable, see following diagram, however max. 15 W per wire and max. 1000 m cable length
1000 m
900
840
800 700
630
600 500 400 300 200
Length of the connection cable
100
0
0 5 10 15
Total power of all sensors on one
2
1,0 mm
2
0,75 mm
two-wire cable
2,5 mm
1,5 mm
2
2
23W max
20
Note:
If longer signal cables are used, it is a good idea to distribute the sensors over two or three inputs.
18 VEGASCAN 850
Technical data
Measuring range
(reference plane is the transducer end. On VEGASON 54 … 56 in version A the lower flange side is the reference plane.)
VEGASON 51
- liquids 0.25 4 m
- solids 0.3 2 m
VEGASON 52
- liquids 0.4 7 m
- solids 0.5 3.5 m
VEGASON 53
- liquids 0.6 15 m
- solids 0.75 7 m VEGASON 54 in general 1.0 … 25 m VEGASON 55 in general 0.8 … 45 m VEGASON 56
- version A 1.8 … 70 m
- version B D 1.4 70 m
Output signal of the sensors
Signal output of the sensors digital output signal in two-wire technology
(VBUS): the digital output signal (meas. signal) is superimposed power supply of VEGASCAN and further processed in VEGASCAN
Meas. data inputs on VEGASCAN
Number of inputs 3 meas. data inputs on VEGASCAN 850B
6 meas. data inputs on VEGASCAN 850C Supply voltage from VEGASCAN approx. 24 V Output current per meas. data input max. 1 A Output power
per meas. data input max. 23 W (15 W continuous power)
Resistance of the signal cable max. 15 W per wire or
max. 1000 m cable length Integration time 0 999 seconds (selectable in the sensor)
0 600 seconds (selectable in VEGASCAN)
Two-wire technology: The digital output signal (meas. signal) and the power supply are led through one cable.
VEGASCAN 850 19
Technical data
Output signal of VEGASCAN (depending on the ordered instrument version)
Siemens 3964
- interfaces RS 232, RS 422, RS 485, TTY
- transmission rates 110 19200 baud
- transmission mode serially asynchronous, half-duplex
- coding system 8 bit binary
- number of bits 1 start , 8 data, 1 parity, 1 stop bit
- parity NONE, ODD, EVEN
- backup BCC
VEGA-ASCII
- interfaces RS 232, RS 422, RS 485, TTY
- transmission rates 300 38400 baud
- transmission mode serially asynchronous, half-duplex
- coding system 8 bits ASCII
- number of bits 1 start, 8 (7) data, 1 parity, 1 stop bit
- parity NONE, ODD, EVEN
- backup none
Modbus
- interfaces RS 232, RS 422, RS 485, TTY
- transmission rates 300 38400 baud
- transmission mode serially asynchronous, half-duplex
- coding system RTU-Mode 1 start, 8 data, 1 (0) parity, 1 stop bit ASCII-Mode 1 start, 8 (7) data, 1 (0) parity, 1 stop bit
- parity NONE, ODD, EVEN
- backup RTU-Mode CRC-16 ASCII-Mode LRC
Profibus DP and FMS
- interfaces RS 485
- transmission rates Profibus DP 9.6 12000 kbaud Profibus FMS 9.6 … 500 kbaud
- transmission mode serially asynchronous, half-duplex, slip-free
synchronisation
- coding system NRZ code
- backup Haming distance HD = 4
Galvanic separation galvanic separation between sensor current
circuit, power supply, output signal (Siemens 3964, Modbus etc.) and RS 232 PC adjustment interface
- reference voltage up to 500 V
- isolation resistance 4 kV
20 VEGASCAN 850
Technical data
Display of measured value (optional)
Liquid-crystal display
- in sensor scalable output of measured values as graph and digital value
- powered externally by the sensor scalable output of measured values as graph and digital value. Display of measured values can be mounted up to 25 m separated from the sensor.
Adjustment
- PC with adjustment software VEGA Visual Operating via RS 232 interface (max. 15 m
cable length)
- adjustment module MINICOM (only the individual sensor is adjustable)
Characteristics
1)
(typical values under reference conditions, all statements relate to the nominal measuring range)
Min. span
(between empty and full adjustment) > 20 mm (recommended > 50 mm)
Ultrasonic frequency (at 20°C)
- VEGASON 51 70 kHz
- VEGASON 52 55 kHz
- VEGASON 53 38 kHz
- VEGASON 54 30 kHz
- VEGASON 55 18 kHz
- VEGASON 56 10 kHz
Meas. intervals
- VEGASON 51 1.0 s
- VEGASON 52 1.0 s
- VEGASON 53 0.6 s
- VEGASON 54 1.0 s
- VEGASON 55 1.5 s
- VEGASON 56 2.0 s
Beam angle at -3 dB emitted power
- VEGASON 51 5.5°
- VEGASON 52 5.5°
- VEGASON 53 3°
- VEGASON 54 4°
- VEGASON 55 5°
- VEGASON 56 6°
Influence of the process temperature 1.8 %/10 K, however is compensated by a
dynamic temperature detection integrated in the transducer
Influence of the process pressure negligible within the permitted sensor pressures Adjustment time
2)
- VEGASON 51 54 > 2 s (depending on the parameter setting)
- VEGASON 55, 56 > 4 s (depending on the parameter setting)
1)
Similar to DIN 16 086, reference conditions acc. to IEC 770; temperature 15°C 35°C; moisture 45 % 75 %; pressure 860 mbar 1060 mbar
2)
The adjustment time is the time required by the sensors for the correct output of the level (with max. 10 % deviation) after a sudden level change.
VEGASCAN 850 21
Technical data
Accuracy
1)
(typical values under reference conditions relating to the nominal measuring range)
Characteristics linear Deviation in characteristics including
linearity, repeatability and hysteresis (determined acc. to the
limit point method) < 0.1 % Linearity better than 0.05 % Average temperature coefficient of the
zero signal 0.06 %/10 K Resolution max. 1 mm Resolution of the output signal
- VEGASON 51 54 0.01 % or 1 mm
- VEGASON 55 and 56 0.01 % or 10 mm
Ambient conditions
Ambient temperature (housing) -20°C … +60°C Process temperature (transducer)
- VEGASON 51 55 -40°C … +80°C (StEx: -20°C +75°C)
- VEGASON 56 version A -40°C … +120°C
- VEGASON 56 version B and C -40°C … +150°C
- storage and transport temperature -40°C … +80°C
Vessel pressure max. (gauge pressure)
- VEGASON 51 and 52 2.0 bar
- VEGASON 53 1.5 bar
- VEGASON 54 Version A 1.5 bar (flange version) Version B C 0.5 bar Version D 3 bar
- VEGASON 55 Version A 1.5 bar (flange version) Version B C 0.5 bar Version D 3 bar
- VEGASON 56 Version A 3 bar (flange version) Version B C 0.5 bar Version D 3 bar
Protection
- sensor IP 67
- transducer, process IP 68
Protection class
- two-wire sensor II
- four-wire sensor I
Overvoltage category III Selfheating
at 40°C ambient temperature
- sensor 45°C
- transducer, process 55°C
1)
Similar to DIN 16 086, reference conditions acc. to IEC 770; temperature 15°C 35°C; moisture 45 % 75 %; pressure 860 mbar 1060 mbar
22 VEGASCAN 850
Technical data
Ex technical data
(note approval documents)
Classification m (die casting of the transducer)
supply with increased safety
Temperature class (permissible
ambient temperature on the transducer when used in Ex areas)
-T6 45°C
- T5, T4, T3 60°C
Ex approved in category or zone
- VEGASON 51, 52, 56 ATEX Zone 1 (II 2G) IEC, CENELEC, PTB Zone 1
- VEGASON 51 56 ATEX Zone 21/22 (II 2D/3D) IEC, CENELEC, PTB Zone 10/11
Process fittings
VEGASON 51 G 11/2A, 11/2“ NPT VEGASON 52 G 2 A, 2 NPT VEGASON 53 DN 100 compression flange VEGASON 54 G 1 A, 1-11,5 NPT, DN 50, DN 80, DN 200 VEGASON 55 G 1 A, 1-11,5 NPT, DN 50, DN 80, DN 250 VEGASON 56 G 1 A, 1-11,5 NPT, DN 50, DN 80, DN 200
Note: VEGASON 54 … 56 sensors with process fittings G 1 A, 1-11.5 NPT, DN 50 and DN 80 generally require an additional adapter flange, if no other access to the vessel interior is available.
Connection cables
Power supply supply and signal via one two-wire cable Electrical connection
- sensors and VEGASCAN spring terminals, terminal cross-section
generally 2.5 mm
2
- sensor option screw connection
Cable entry in sensor housing
- standard 2 x M20 x 1.5 (cable diameter 5 9 mm)
- option 2 x 1/2“ NPT (cable diameter 3.1 … 8.7 mm
or 0.12 … 0.34 inch)
Cable entry in
VEGASCAN 850 4 … 5 x M20 x 1.5
Ground connection max. 4 mm
(cable diameter 6 … 12 mm)
2
Transducer cable (only type 54 56
in version C and D)
- VEGASON 54, 55 5 300 m (cable diameter 7.2 7.6 mm)
- VEGASON 56 5 … 300 m (cable diameter 9.5 … 9.9 mm)
VEGASCAN 850 23
Materials
Housing PBT (Valox) or Aluminium Process fitting
- VEGASON 51, 52 PVDF (thread)
- VEGASON 53 PP or 1.4571 (compression flange)
1.43019 (mounting strap)
- VEGASON 54 56 - Alu or PP (version A)
- swivelling holder or thread G 1 A of galvanized steel (version B, C and D)
Transducer
- VEGASON 51, 52 PVDF
- VEGASON 53 UP
- VEGASON 54 PA (1.4301 with StEx)
- VEGASON 55, 56 UP
Transducer diaphragm
- VEGASON 51, 52 PVDF
- VEGASON 53 1.4571
- VEGASON 54 1.4571
- VEGASON 55 Alu/PE foam
- VEGASON 56 Alu/PTFE coating
Transducer cable (cable cover)
- VEGASON 54, 55 PUR (1.1082)
- VEGASON 56 Silicone (1.1083)
Weight
VEGASON 51 1.2 kg VEGASON 52 1.6 kg VEGASON 53 2.3 kg VEGASON 54
- version A 5.6 … 10.7 kg
- version B 6.9 … 9.7 kg
- version C 7.5 … 10.5 kg
- version D 4.7 6.9 kg
VEGASON 55
- version A 8.0 … 13.3 kg
- version B 8.7 … 10.3 kg
- version C 9.2 … 11.1 kg
- version D 6.5 7.5 kg
VEGASON 56
- version A 7.3 … 11.3 kg
- version B 8.7 … 10.3 kg
- version C 9.3 … 11.1 kg
- version D 6.5 7.5 kg
Technical data
CE conformity
VEGASON series 50 ultrasonic sensors and VEGASCAN 850 meet the protective regula­tions of EMC (89/336/EWG) and NSR (73/23/EWG). Conformity was judged acc. to the following standards: EMC Emission EN 50 081 - 1: 1992
Susceptibility EN 50 082 - 2: 1995
NSR EN 61 010 - 1: 1993
EN 61 326 - 1: 1997/A1:1998
24 VEGASCAN 850
Technical data

3.2 Dimensions

VEGASCAN 850
90
80
260235
246
13 90
210
236
VEGASCAN 850 25
External indicating instrument VEGADIS 50
Technical data
38
ø5
48
10
Pg 13,5
Mounting on carrier rail 35 x 7 .5 acc. to EN 50 022 or flat screwed
135
118
108
Adjustment module MINICOM
Tank 1 m (d)
12.345
67,5
ESC
+
-
32,5
OK
Adjustment module for insertion into VEGASON series 50 sensor or into the exter­nal indicating instrument VEGADIS 50
74
82
85
Note:
The diameter of the connection cable must be 5 9 mm. Otherwise the seal effect of the cable entry will not be ensured.
26 VEGASCAN 850
Technical data
VEGASON 51
Housing PBT
201
Housing aluminium
370
322
182
215
185
10˚
165
M20x1,5
SW 60
G 1½ A/ 1½" NPT
25
101
90
116
2
ø 40
Min. distance to the medium
228
315
20
Reference plane
0,25 m
205
M20x1,5
SW 60
102
252
337
2
G 1½ A/ 1½" NPT
20
Reference plane
ø 40
Min. distance to the medium
0,25 m
VEGASCAN 850 27
VEGASON 52
Housing PBT
Technical data
201
Housing aluminium
370
322
182
215
185
10˚
165
M20x1,5
SW 60
G 2 A
25
101
90
116
232
2
20
ø 50
Min. distance to the medium
315
Reference plane
0,4 m
205
M20x1,5
SW 60
102
256
337
2
G 2 A
20
Reference plane
ø 50
Min. distance to the medium
0,25 m
28 VEGASCAN 850
Technical data
VEGASON 53
201
322
M8x10 12 tief
182
20
ø 148 ø 158
10˚
165
M20x1,5
195
130
101
90
118
218
Min. distance to the medium
270
0,6 m
VEGASCAN 850 29
VEGASON 54 … 56 in version A
201
165
Plastic housing (PBT)
10˚
101
Aluminium housing (Al)
25
116
Technical data
215
185
VEGASON 54
VEGASON 55
397
257
Min. distance to the medium
Min. distance to the medium
1,0 m
0,8 m
90
12xø22
1)
ø190 (ø196)
ø340
ø244 ø405
75
1)
20
110 (126)
12xø26
20
128
12xø22 (12xø26)
445,8
282
Reference plane
Reference plane
20
Reference plane
VEGASON 56
423
Min. distance to the medium
1,8 m
30 VEGASCAN 850
ø198
2)
ø340 (ø405)
Technical data
VEGASON 54 … 56 in version B
486
4xø19
172
386
245
ø165
ø122,8
201
165
10˚
Plug connection
ø 27
> ø200
Plastic housing (PBT)
11,5
11,5
101
90
503
4xø19
ø165
ø122,8
>ø250
435
270
215
185
65
Plug connection
798
Aluminium housing (Al)
25
116
ø165
ø122,8
4xø19
>ø210
ø190 (ø196)
VEGASON 54
1,0 m
Reference plane
189,5
ø 244
VEGASON 55
484,5
0,8 m
ø 198
1,4 m
VEGASON 56
VEGASCAN 850 31
VEGASON 54 … 56 in V ersion C
Aluminium housing (Al)
Plug
Technical data
215
25
116
185
Plug connection
ø 45
78
68
7
130 150
ø 7
85
65
170
445,8
282
10120
1,0 m
0,8 m
VEGASON 54
VEGASON 55
1,4 m
Reference plane
VEGASON 56
32 VEGASCAN 850
Technical data
VEGASON 54 … 56 in version D
Plug connection
3240
214
149
Plastic housing (PBT)
Plug
68
ø 45
78
233
7
3240
167,5
186
101
90
130 150
201
165
10˚
397,2
257,2
ø 7
85
65
120 10
170
3240
527
462,5
Reference plane
1,0 m
0,8 m
1,4 m
VEGASON 54 VEGASON 55 VEGASON 56
VEGASCAN 850 33

4 Mounting and installation

4.1 Mounting VEGASON 54 … 56

Mounting and installation
VEGASON 54 … 56, version A
Sensors in version A (flange version) are supplied completely mounted and ready for operation. Immediately after mounting on the vessel and electrical connection, they are ready for operation.
VEGASON 54 … 56, version B
The sensors in version B are supplied in two parts (transducer and sensor electronics). First of all, mount the transducer on the ves­sel or above the medium. There is a four-pole jack at the end of the transducer tube. The respective counterpart to the jack protrudes out of the lower side of the sensor electronics housing. Insert the plug of the sensor elec­tronics (only possible in one position) into the jack of the transducer tube. Continue push­ing the electronics housing onto the trans­ducer tube, on which there is a wide and a narrow groove.
Groove for locking the headless screw
Mounting groove (must no longer be visible after mounting)
The wide groove is used for locking the headless screws. The narrow groove is a marking for mounting. Move the electronics housing farther down over the transducer tube until the mounting groove is no longer visible. Fasten the housing with the headless screws to the transducer tube. Use a 5 mm hexagon screwdriver (or Allen wrench).
34 VEGASCAN 850
Mounting and installation
VEGASON 54 … 56, version C, D
The sensors in version C and D are supplied in three parts (transducer, sensor electronics and transducer cable). First mount the trans­ducer (see version B). There is a four-pole jack at the transducer tube end. A respective counterpart to the jack is provided in the connection cylinder of the transducer cable. Insert the connection cylinder plug into the jack of the transducer tube.
Connection cylinder
Mounting bracket
Connection cylinder
Transducer cable
On the end of the transducer tube you find a wide and a narrow groove. The wide groove is used for locking the cylinder with the head­less screws. The narrow groove is the mounting mark.
Then push the connection cylinder over the transducer tube (with a slight swivelling mo­tion) until the mounting mark is no longer visible.
When the mounting mark is covered by the cylinder, fasten the cylinder with the two headless screws. Use a 5 mm hexagon screwdriver (or Allen wrench).
Now mount the sensor electronics in the requested location. The sensor electronics is fastened to a mounting bracket, so that it can be mounted on a plane surface or on the wall. Make sure that the sensor housing is mounted in such a way that there is enough space above the housing to open the cover.
Now insert the plug at the other end of the transducer cable into the jack on the elec­tronics housing.
Note:
Avoid bending the transducer cable too sharply when laying it out. This is a special cable which could otherwise be damaged.
In addition, make sure that the cable is not damaged during operation. A signal with a voltage of approx. 1 kV is transmitted (which could be a danger in Ex areas if the cable is damaged).
Groove for locking the headless screws
Mounting groove (must no longer be visible after mounting)
VEGASCAN 850 35
Mounting and installation
4.2 General installation instructions
transducer end. For VEGASON 54 ... 56 in version A the lower flange side on the sensor
Measuring range
Beside other criteria, you select your instru­ment according to the required measuring range. The reference planes for the min. and
is the reference plane. Please note the infor­mation on the reference planes in chapter "3.2 Dimensions. The max. filling depends on the required min. distance and the mounting location.
max. distance to the product or solid is the
VEGASON 51
Min. distance
0.25 m 0.4 m
Full
Empty
1m
max. meas. range
4 m (type 51), 7 m (type 52), 15 m (type 53)
Min. distance, max. measuring range and span (VEGASON 51 53)
VEGASON 53 VEGASON 52
Min. distance
0.75 m
Span
Reference plane
Min. distance
VEGASON 56 Version B
Reference plane
min. meas. distance 1.4 m
100 %
0 %
Span
0 %
100 %
Span
VEGASON 54 Version A
min. meas. distance 1.0 m
VEGASON 55 Version B
Reference plane
min. meas. distance 0.8 m
100 %
0 %
max. meas. range
Span
max. meas. distance 25 m (type 54), 45 m (type 55), 70 m (type 56)
Min. distance, max. measuring range, span and reference plane (VEGASON 54 56)
36 VEGASCAN 850
Mounting and installation
0
0 m
15 m
100 %
50 %
1,2
1,2
m
0,4
0,4
3˚
8˚
Beam angle and false echoes
The ultrasonic impulses are focused by the transducer. The impulses leave the trans­ducer in conical form similar to the beam pattern of a spotlight.
Any object inside this emission cone will cause a false echo. Especially within the first few meters of the emission cone, pipes, struts, or other installations can interfere with the measurement. At a distance of 6 m, the false echo of a strut has an amplitude nine times greater than at a distance of 18 m.
At greater distances, the energy of the ultra­sonic impulses distributes over a large area, thus causing weaker echoes from obstruct­ing surfaces. The interfering signals are therefore less critical than those at close range.
If possible, orient the sensor axis perpen­dicularly to the product surface and avoid vessel installations (e.g. pipes and struts) within the 100 % area of the emission cone.
The illustrations showing the ultrasonic emis­sion cones are much simplified and repre­sent only the main beam. However, several weaker beams also exist. The transducer must therefore be aligned - especially under difficult measuring conditions - in such a way that very low false echo values result. Putting emphasis only on a strong useful echo is not sufficient under adverse conditions.
Meas. distance
0 m
Meas. distance
7 m
Meas. distance
0 m
4 m
0,4
0,8
0,4
1,2
VEGASON 51
50 % 100 %
5,5˚
0,4
0
0,8
VEGASON 52
5,5˚
100 %
0,4
1,2
0
VEGASON 53
emitted power emitted power
12˚
m
12˚
emitted power
emitted power
50 %
emitted power
m
emitted power
Under difficult measurement conditions, we recommend looking for a mounting location with the weakest possible false echoes. The useful echo will then often appear automati­cally with sufficient quality. With the adjust­ment software VVO on the PC you can have a look at the echo image and optimise the mounting location (see chapter "6.2 Adjust­ment with the PC – Sensor optimisation – Echo curve“).
VEGASCAN 850 37
0 m
Meas. distance
Mounting and installation
VEGASON 56VEGASON 54
0 m
emitted power
50 %
100 %
4˚
emitted power
25 m
Meas. distance
45 m
0 m
4
0,9
2,0
5˚
10˚
242
0
8˚
0,9
2,0
0
VEGASON 55
50 %
emitted power
100 %
m
m
emitted power
Meas. distance
70 m
50 %
emitted power
6˚
100 %
emitted power
12˚
3,7
7,5
3,7
0
m
7,5
38 VEGASCAN 850
Mounting and installation

4.3 Measurement of liquids

Flat vessel top
On flat vessels, the mounting is usually done on a very short DIN socket piece. Reference plane is the lower edge of the flange. The transducer should protrude out of the flange tube.
Reference plane
£ 60 mm
VEGASON 53 on very short DIN socket piece
< 100 mm
Min. distance
Reference plane
Reference plane
< 400 mm
Min. distance
1.8 m
VEGASON 56 in flange version on short DIN socket piece
A mounting location directly on the vessel top is ideal. A round opening in the vessel top is sufficient to fasten the sensor with the flange, or version B and C (VEGASON 54 … 56) with swivelling holder.
Reference plane
Min. distance Type 54: 1 m Type 55: 0.8 m
VEGASON 55 in flange version on short DIN socket piece
VEGASON 53 (compression flange) on flat vessel top
VEGASCAN 850 39
Reference plane
Swivelling holder
Reference plane
Mounting and installation
< 60 mm
Reference plane
Mounting of the transducer with 1 thread (here belonging to VEGASON 54 version D)
Min. meas. distance
1.8 m
Min. meas. distance
1.4 m
Flange version and swivelling holder on flat vessel top
It is also possible to mount the sensors with 11/2“ or 2“ thread to short socket pieces.
£ 60 mm
Reference plane
Mounting on short 11/2“ or 2“ socket pieces (VE­GASON 51)
VEGASON 54 … 56 sensors in version C are mounted in a 1 thread.
Dished tank ceiling
On dished tank ceilings, please do not mount the instrument in the centre, but approx. 1/ vessel radius from the centre. Dished tank ceilings can act as paraboloidal reflectors. If the transducer is placed at the focal point of the parabolic ceiling, the transducer receives amplified false echoes. The transducer should be mounted outside the focal point. Amplified echoes caused by parabolic sur­faces are thereby avoided.
£ 60 mm
VEGASON 51, 52 on dished tank ceiling
Reference plane
1
/2 vessel radius
2
40 VEGASCAN 850
Mounting and installation
£ 60 mm
1
/2 vessel radius
VEGASON 53 on dished tank ceiling
< 400 mm
1
/2 vessel radius
Reference plane
Reference plane
Reference plane
< 100 mm
1
/2 vessel radius
VEGASON 54 version A on dished tank ceiling; the statements are also valid for VEGASON 55
Open vessels
On open vessels, use of instruments on an extended mounting bracket is recom­mended. Mount the low-weight sensor onto such a bracket and ensure a sufficient dis­tance to the vessel wall.
Reference plane
Min. meas.
Reference plane
distance
VEGASON 56 version A on dished tank ceiling
Min. meas. distance
VEGASON 54 on open vessel
VEGASCAN 850 41
Mounting and installation
Pump shaft
Narrow shafts and shaft-like openings (ves­sel openings) with very rough walls and shoulders make an ultrasonic measurement extremely difficult due to strong false echoes. This problem can be overcome by using an extended socket piece or a complete meas­uring tube (see chapter "4.5 Socket exten­sion).
see 4.5 Socket extension
Socket piece
250 mm
min. distance
Meas. range
4.4 Measurement of solids
Flange mounting
As with applications for liquids, the instrument can be mounted on a short DIN socket con­nection on vessels for solids. The transducer axis, however, should point to the vessel outlet or should be oriented perpendicularly to the product surface. The socket length can be max. 60 … 400 mm, depending on the sensor type.
Reference plane
Min. distance
Shaft pump
Measuring tube
Example of a socket extension or measuring tube in a shaft
Shaft pump
VEGASON 53 on inclined vessel flange
Shaft
Very good measuring results can be attained with a measuring tube in continuous narrow shafts, see figure. The applied measuring tube must have smooth walls inside (e.g. PE sewage pipe) and a diameter of 100 mm. This arrangement works well as long as the inside of the measuring tube collects no dirt or buildup (cleaning necessary). You might want to consider using hydrostatic pressure transmitters or capacitive measuring probes. Either the measuring tube should never be immersed in the medium, or it must always
Swivelling holder
We offer as an accessory a swivelling holder (mounting strap) for mounting of VEGASON
53. This simplifies the alignment of the sensor to the product surface.
Suitable is the use of VEGASON 54 56 version B or C. For solids, the swivelling holder enables an optimum alignment of the transducer, thus minimising false echoes.
be immersed (so that the measurement is carried out exclusively in the tube).
42 VEGASCAN 850
Mounting and installation
Reference plane
Min. distance
Mounting boss
Reference plane
Min. distance
VEGASON 53 on swivelling holder
Reference plane
Min. distance
VEGASON 54C with adapter flange and swivelling holder on a DN 200 vessel flange
VEGASON 51 or 52 on the mounting boss. The socket axis should point directly to the product surface. This is why VEGASON 51 and 52 are less suitable for solids.
Reference plane
Min. distance
VEGASON 56 in 1 mounting boss.
The socket axis should point directly to the product surface. Much better would be the use of a swivelling holder version.
VEGASCAN 850 43
Mounting and installation
Material heaps
Large material heaps are best detected with several instruments, which can be mounted on e.g. traverse cranes. For this type of ap­plication, it is advantageous to orient the sensor directly toward the solid surface.
Transducer on traverse crane above a material heap (illustration: VEGASON 54 in version B)
4.5 Socket extensions
The ultrasonic sensors require a min. dis­tance to the liquid or solid product. Take the min. distance into account in your planning. In some situations, it is possible to reach the required min. distance, and hence the de­sired filling height, with a socket extension. However, the socket extension increases the noise level of the ultrasonic signal at the ex­tension outlet and can interfere with the measurement. Only use a socket extension if all other possibilities have to be excluded. Carry out the extension as shown in the fol­lowing illustration.
Socket extensions in liquids
Chamfer and deburr the socket carefully and make sure it has a smooth inner surface. The socket should not protrude into the meas­ured product, in case buildup can form on the socket through pollution or product resi­dues.
Socket piece should not be immersed into adhesive products (illustration: VEGASON 53)
The socket diameter should be as large and the socket length as small as possible. Make sure that the socket outlet is burr-free to minimise false echoes.
44 VEGASCAN 850
Mounting and installation
15˚ 15˚
Type 51/52
LL
45˚
ø
Socket extensions in liquids
Type 53
45˚
ø
Max. socket length in relation to socket dia­meter
ø in mm L in mm
Type 51 Type 52 Type 53
100 200 300 300 150 300 400 400 200 500 500 250 ––600
Type 54
Type 55
Socket extensions for solids
For solids, use a conical socket extension with a taper of at least 15° … 20°.
Socket extension in solids
Measurement in a tube
For nonadhesive measured products, a socket extension in the form of a measuring tube can be permanently submerged in the product. The ultrasonic measurement is then made exclusively in the measuring tube and works very well without interference from other vessel installations (see "Pump shaft).
L
45˚
ø
Socket extensions that do not protrude into the measured product
L
45˚
ø
Max. socket length in relation to socket dia­meter
ø in mm L in mm
Type 54 Type 55 Type 56
200 400 –– –– 250 500 500 500 300 –– –– 600
VEGASCAN 850 45
Mounting and installation
4.6 Flow measurement
The short examples on this page are only basic information on flow measurement. You can get complete planning information from the flume manufacturers and in special litera­ture.
Rectangular flume
- Installation of the sensor on the upstream side
- Note distance to the overfall edge (3 4 x h
- Installation centered to the flume
- Edge opening ³ 2 x h
- Installation perpendicular to the liquid sur­face
- Keep min. distance in relation to h
- Min. distance from edge opening to down­stream water ³ 50 mm
max
90°
)
from ground
max
max
³ max. distance
h
max
Khafagi-Venturi flume
- Installation of the sensor on the inlet side
- Note distance to the Khafagi-Venturi flume (3 4 x h
- Installation perpendicular to the liquid sur­face
- Keep min. distance in relation to the height of damming h
Khafagi-Venturi flume
max
3 4 x h
90°
Sensor
)
max
max
h
max
B
³ 2 x h
Overfall edge
max
Flow measurement on open flumes
Overfall edge
3 4 x h
max
90°
³ 5 cm
Upstream water
Downstream water
Flow measurement on open flumes
46 VEGASCAN 850
Mounting and installation
4.7 False echoes
The mounting location of the ultrasonic sensor must be selected such that no installations or inflowing material are in the path of the ultra­sonic impulses. The following examples and instructions show the most frequent measur­ing problems and how to avoid them.
Vessel protrusions
Vessel forms with flat protrusions can, due to their strong false echoes, adversely effect the measurement. Shields above these flat protrusions scatter the false echoes and guarantee a reliable measurement.
Correct Wrong
Vessel protrusions (slope)
Intake pipes, e.g. for the mixing of materials ­with a flat surface directed towards the sen­sor - should be covered with a sloping shield. This shield will scatter false echoes.
Vessel installations
Vessel installations such as, for example, a ladder, often cause false echoes. Make sure when planning your measurement loop that the ultrasonic signals have free access to the measured product.
Correct Wrong
Ladder
Vessel installations
Ladder
Struts
Struts, like other vessel installations, can cause strong false echoes that are superim­posed over the useful echo signals. Small shields effectively hinder a direct false echo reflection. These false echoes are scattered and diffused in the area and are then filtered out as "echo noise by the measuring elec­tronics.
Correct Wrong
Correct Wrong
Shields
Struts
Vessel protrusions (intake pipe)
VEGASCAN 850 47
Mounting and installation
Inflowing material
Do not mount the instrument in or above the filling stream. Ensure that you detect the product surface and not the inflowing mate­rial.
Correct
Correct
Wrong
Wrong
Correct
Buildup
Wrong
Overflow basin
The expected max. high water determines the installation height, to ensure the min. distance of the transducer even with the highest water level. The low water level should be covered in the transducer area with a shield to filter out echoes from ex­posed basin surfaces.
Correct Wrong
Inflowing material
Min. distance high water
Buildup
If the sensor is mounted too close to the vessel wall, buildup and adhesions of the measured product to the vessel wall can cause false echoes. Position the sensor at a sufficient distance from the vessel wall. Please also note chapter "4.2 General instal­lation instructions“.
48 VEGASCAN 850
60°
Shield
Filtering out of a level echo
Low water
Mounting and installation
Strong product movements
Heavy turbulences in the vessel, e.g. by strong stirrers or strong chemical reactions, can seriously interfere with the measurement. A surge or bypass tube of sufficient size (DN 200, DN 250) always allows, provided the product causes no buildup in the tube, a reliable measurement even with strong turbu­lences in the vessel.
100 %
60 %
0 %
Strong product movements
4.8 Incorrect mounting
Foam generation
Thick foam on the product can cause incor­rect measurements. Take measures to avoid foam, carry out the measurement in a bypass tube, or use a different measuring technol­ogy, e.g. capacitive measuring probes or hydrostatic pressure transmitters.
Foam generation
Wrong orientation to the product
Weak measuring signals are the result if the sensor is not directly pointed at the product surface. Orient the sensor axis perpendicu­larly to the product surface to achieve opti­mum measuring results.
Orient the sensor perpendicularly to the product surface
VEGASCAN 850 49
Mounting and installation
Strong heat fluctuations
Strong heat fluctuations, e.g. due to the sun, cause measuring errors. Please provide a sun shield.
Shield
Strong heat fluctuations
Min. distance to the medium
If the min. distance to the medium is not main­tained, the instruments show wrong meas­ured values. Mount the instrument at the required min. distance.
Sensor too close to the vessel wall
If the sensor is mounted too close to the vessel wall (dimension A in diagram), strong false echoes can be caused. Buildup, rivets, screws or weld joints on the vessel wall su­perimpose their echoes onto the product or useful echo. Please ensure a sufficient dis­tance from the sensor to the vessel wall, depending on the maximum measuring dis­tance (dimension B in diagram).
In case of good reflection conditions (liquids, no vessel installations), we recommend de­termining the sensor distance according to Diagram curve 1. At a max. meas. distance of e.g. 10 m, the distance of the transducer (according to curve 1) should be approx.
1.5 m.
In case of solids with bad reflection proper­ties, determine the distance to the vessel wall according to Diagram curve 2. Under very bad measuring conditions (rough vessel walls, struts), it might be necessary to in­crease the distance to the vessel wall, or to additionally filter out the false echoes by storing them in memory, thereby adapting the sensor more precisely to the environment.
Correct Wrong
Sensor too close to the vessel wall
50 VEGASCAN 850
Mounting and installation
Distance of the transducer to the vessel wall
A
1 m 2 m 3 m 4 m 5 m
Curve 1 (liquids)
5 m
B
10 m
15 m
max. meas. distance
Curve 2 (solids)
Distance from the sensor to the vessel wall, depend­ing on the meas. distance (type 51 53)
Distance of the transducer to the vessel wall
A
Parabolic effects of rounded or arched vessel tops
Round or parabolic tank tops act on the sig­nals like a parabolic mirror. If the sensor is placed at the focal point of such a parabolic tank top, the sensor receives amplified false echoes. The optimum location is generally in the area of half the vessel radius from the centre.
Correct
< 100 mm
~ 1/
2
vessel radius
Wrong
2 m 4 m 6 m 8 m
Wrong
Curve 1 (liquids)
10 m
B
20 m
Curve 2 (solids)
Mounting on a vessel with parabolic tank top
30 m
max. meas. distance
Distance from the sensor to the vessel wall, depend­ing on the meas. distance (type 54 56)
VEGASCAN 850 51
Socket piece too long
If the sensor is mounted in a socket exten­sion that is too long, strong false echoes are caused, and measurement is hindered. Make sure that the transducer protrudes at least 30 mm out of the socket piece.
Reference plane
< 100 mm
Mounting and installation
Correct and wrong length of socket piece
52 VEGASCAN 850
Electrical connection

5 Electrical connection

5.1 Connection, connection cable and shielding

Qualified personnel
Instruments which are not operated with protective low voltage or DC voltage must only be connected by qualified personnel. This is especially valid for the connection of the power supply on VEGASCAN.
Safety information
As a rule, do the work in the complete ab­sence of voltage. Always switch off the power supply before you carry out connecting work. Protect yourself and the instruments.
The shielded transducer cables transmit a signal with a voltage of approx. 1 kV. In Ex areas, cable damage can be dangerous. Do not carry out any mounting or connection work on the transducer when VEGASCAN is switched on.
Connection cable
Please note that the connection cables are specified for the expected operating tempera­tures in your systems. The power supply and the sensor cables must have an outer diam­eter of 6 12 mm, to ensure the seal effect of the cable entry.
Transducer cable
When wiring the transducer cable (version C and D), strong bending of the cable should be avoided. This is a special cable which can be damaged.
Make sure when wiring the transducer cable that no operating influences can damage it.
Earth conductor terminal
The electronics housing of the sensors has a protective insulation. The earth conductor terminal and the earth terminal in the electron­ics housing are galvanically connected with the metallic transducer diaphragm. On sensors in version B, the earth conductor terminal is galvanically connected to the transducer diaphragm via the transducer tube when the sensor is completely mounted. On version C and D, the connection is made via the cable screen of the transducer cable and the transducer tube.
Shielding of the sensor cables
The "Electromagnetic pollution from elec­tronic actuators, power lines and transmitting stations is often so considerable that the two­wire cable of VEGASCAN to the sensors must be shielded.
Power supply
For power supply, standard three-wire cable up to max. 2.5 mm2 can be used. The electri­cal connection is made via spring terminals. You open the terminal opening with a small screwdriver by inserting it into the opening slot above the terminal position and lever it upward. This opens the terminal and the copper core of the connection wire can be inserted.
VEGASCAN 850 53
Electrical connection
We recommend a screening on both ends, see the following sketch. Screening is a good preventative measure against future sources of interference. However, you must make sure that no ground potential currents flow through the sensor cable shields. Ground potential currents can be avoided by poten­tial equalisation cables. When grounding at both ends, it is possible to connect the cable
Note Ex protection!
In Ex applications, grounding on both ends is not allowed due to potential losses. If an instrument is used in hazardous areas, the respective regulations, conformity certificates and type approvals for systems in Ex areas must be noted (e.g. DIN 0165). Please note the approval documents with the
safety data sheet attached to the Ex sensors. shield on one side (e.g. in the switching cabi­net) via a capacitor (e.g. 0.1 µF; 250 V) to the ground potential. Use a low-resistance ground connection (foundation, plate or mains earth).
Linear (serial) arrangement of sensors
Grounding of the cable screen on both ends, at the end of each sensor line via a ground ca­pacitor.
PC
850
BA
on
2
2
54 VEGASCAN 850
Electrical connection
Radial arrangement of sensors
Grounding on at least two ends, on VEGASCAN and once on the sensor star, i.e. on the long­est sensor line. If the individual sensor lines are longer than approx. 15 m, a grounding of each longer line should be made via a ground capacitor.
PC
850
BA
on
Longest stub of the sensor star
2
4
2
2
4
2
VEGASCAN 850 55
Electrical connection
5.2 Connection of the sensor
After mounting the sensor at the measure­ment location according to the instructions in chapter "4 Mounting and installation, loosen the closing screw on top of the sensor. The sensor lid with the optional indication display can then be opened. Unscrew the sleeve nut and slip it over the connection cable (after removing about 10 cm of insulation). The sleeve nut of the cable entry has a self-lock­ing ratchet that prevents it from opening on its own.
Version with aluminium housing
Voltage supply and digital meas. signal
M20 x 1.5 (diameter of the connection cable 69 mm)
+
-
To the indicating instru­ment in the cover or to the external indicating instrument VEGADIS 50
M20 x 1.5
Now insert the cable through the cable entry
into the sensor. Screw the sleeve nut back
onto the cable entry and clamp the stripped
wires of the cable into the proper terminal
positions.
The spring terminals hold the wire without a
screw. Press the white opening tabs with a
small screwdriver and insert the copper core
of the connection cable into the terminal
opening. Check the hold of the individual
wires in the terminals by lightly pulling on
them.
Version with plastic housing
Voltage supply and digital meas. signal
-
+
To the indicating instrument in
Cable entry M20 x 1.5
the cover or to the external indicating instrument
Sockets for connection of VEGACON­NECT 2 (communica­tion sockets)
12 C 567843
12 C 5 6 7 8
(+) (-)
Commu­nication
Display
ESC
-
+
VBUS
Terminals (max. 2.5 mm wire cross-section)
OK
Sockets for connec­tion of the HART handheld or VEGA­CONNECT
pluggable adjustment module MINI­COM
2
12 C 5678
12 C 5678
Commu-
(+) (-)
nication
VBUS
®
Tank 1 m (d)
12.345
Display
+
ESC
-
OK
Opening tabs
56 VEGASCAN 850
Electrical connection
5.3 Connection of the external indicating instrument VEGADIS 50
Loosen the four screws of the housing cover on VEGADIS 50. The connection procedure can be facilitated by fixing the housing cover during connec­tion work with one or two screws on the right of the housing (figure).
OUTPUT (to the sensor)
3
2
1
4
5
8
6
7
Adjustment module
VEGADIS 50
+
-
Tank 1 m (d)
12.345
ESC
OK
Voltage supply and digital meas. signal
-
+
12 C 5678
2
1
(+) (-)
VBUS
Tank 1 m (d)
12.345
C 5678
Commu nication
+
-
Display
ESC
OK
DISPLAY
(in the cover of the indicating instrument)
Screws
VEGASCAN 850 57
Electrical connection
5.4 Connection of the sensors to
VEGASCAN 850
off
on
---
POWER
GNDTxD
N
RxDDTR
FIELDBUS VBUS1 VBUS2
PC2
PE L
VBUS1
Use the terminal VBUS1, when using a VEGASCAN 850 for 15 sensors. Sensors 16 30 are connected to the terminals VBUS2.
+++
--­+++
VBUS2
The electrical connection is made by spring
terminals. The clamping opening can be
opened with a small screwdriver by inserting
it into the opening slot above the clamping
position and pushing upwards. Then the
terminal and the copper core of the connec-
tion wire can be inserted.
Up to 15 sensors can be connected to each VBUS branch on one two-wire cable.
---
+++
Sensor 1 15
VBUS1
It is a good idea to optimise the current distri­bution by dividing the BUS into three branches with five sensors each.
Sensor
Sensor 6 10
Sensor 11 15
1 5
---
+++
VBUS1
58 VEGASCAN 850
Electrical connection
5.5 Signal output/Interface connec­tion
The connection terminal is labelled for the field bus cable according to the selected interface and instrument version.
Modbus, VEGA-ASCII, Siemens 3964
TXRX
DATA
(B/B´)
DATA
(B/B´)
Interface RS 232
/TX/RX
Interface RS 422
Interface RS 485
DATA
(A/A´)
Interface RS TTY
Interface RS 485
DATA
(A/A´)
RxD
GND TxD
FIELDBUS
GND
FIELDBUS
GND
(C/C´)
FIELDBUS
T-
R+ R+
GND T+
FIELDBUS
Profibus DP , Profibus FMS
GND
(C/C´)
FIELDBUS

5.6 Connection of the PC to VEGASCAN 850

The PC with the adjustment software VVO can be connected in the front panel or in the terminal compartment of VEGASCAN.
For connection of the PC in the front panel (9­pole socket) you require a standard RS 232 DTE-DTE (Data Terminal Equipment) inter­face cable.
DCD
RxD
TxD
DTR
GND
...
6
...
7
...
8
...
9
For connection of the PC in the terminal com­partment, connect the opened PC cable according to the following sketches to termi­nal block PC2. For communication, only three wires of the 9-pole cable are used.
DCD
1
RxD
2
TxD
3
DTR
4
GND
5
...
6
...
7
...
8
...
9
1
DCD1
2
RxD2
3
TxD3
4
DTR4
5
GND5
...
6
...
7
...
8
...
9
GNDTxD
RxD DTR
PC2
Terminal in VEGASCAN
VEGASCAN 850 59

6 Setup

Setup

6.1 Adjustment methods

VEGASCAN 850 and the series 50 ultrasonic sensors are adjusted with the PC (adjust­ment program VVO). With the detachable adjustment module MINICOM, it is possible to adjust the sensors individually. The adjust­ment can be carried out with only one adjust­ment instrument at a time.
With the PC
The PC with the adjustment program VVO (VEGA Visual Operating) can be connected to:
- VEGASCAN 850
- the sensor
- the signal cable
With the adjustment program VVO (VEGA Visual Operating) on the PC you adjust the level measuring system consisting of VEGASCAN and ultrasonic sensors. The PC communicates via the standard RS 232 inter­face cable with VEGASCAN and all con­nected sensors. For this purpose, a digital adjustment signal is superimposed on the signal and supply cable.
In the following chapter "6.2 Adjustment with the PC, the adjustment is described when connecting the PC via a RS 232 interface cable to VEGASCAN. When connecting the PC via VEGACONNECT to the sensor signal cable (see "2.6 Configuration of measuring systems configuration A), all sensors con­nected to the affected signal cable are avail­able for adjustment. However, the configuration and measured data processing functions of VEGASCAN are not available.
When inserting VEGACONNECT into the communication socket of an individual sen­sor, only sensor-relevant adjustment options such as e.g. sensor optimisation on the af­fected sensor are adjustable. Other sensors and VEGASCAN are then not adjustable. This is due to an adjustment hierarchy beginning with VEGASCAN and ending with the sensor.
With the adjustment module MINICOM
With the adjustment module MINICOM you adjust the individual sensor directly in the sensor or in the external indicating instrument VEGADIS 50. The adjustment module MINI­COM with the 6 key adjustment field with text display enables the parameter adjustment of the sensor with an array of functions compa­rable to the adjustment program VVO when connected via VEGACONNECT to the sen­sor, but not the configuration of the measur­ing system.
Beside the connection of the PC to VEGA­SCAN, it is possible to connect the PC with the interface converter VEGACONNECT to the signal cable between sensor and VEGA­SCAN or also directly to the individual sensor.
60 VEGASCAN 850
Setup

6.2 Adjustment with the PC

In the following setup and adjustment instruc­tions you will find information on the following topics and adjustment points:
Configuration
- configuration info
- create new measurement loop
Parameter adjustment 1
- adjustment
- scaling
Sensor optimisation
- meas. environment
- sound velocity
- echo curve
- false echo storage
Parameter adjustment 2 (optional)
- linearisation
- parameter adjustment sensor display
- meas. loop data
Display of measured value
Simulation
Select available linearisation curve
Print configuration and adjustments
Backup
Note:
Keep in mind that you first have to log in a sensor in the configuration menu before you can continue with the parameter setting, and if necessary, carry out adjustments in the menu Sensor optimisation”.
For connection of the PC to VEGASCAN, a standard RS 232- DTE-DTE (Data Terminal equipment) interface cable is required. With the cable, the PC is connected to VEGASCAN.
DTE DTE
Beside the measured values, the adjust­ments signals are also transmitted digitally via the signal supply cable between sensor and VEGASCAN. The adjustment program VVO can then communicate with VEGASCAN and all connected sensors. In chapter "2.6 Configuration of measuring systems the connection of the PC to the different sensor arrangements is shown.
Before starting the setup:
Do not be confused by the many pictures, adjustment steps and menus on the following pages. Just carry out the setup with the PC step-by-step and you will soon no longer need the following pages. Whenever you should adjust or choose something, this is indicated in the following by a large black dot, like this:
Choose
Start
Click to
Connect the standard output of your PC by
the standard RS 232 interlink cable (9­pole) to VEGASCAN.
Now switch on the power supply of the
processing system.
By this means, the actions to be carried out are clearly separated from supplementary information in the following adjustment in­structions.
Approx. 1 … 2 minutes after switching on the supply voltage, the sensors generally show measured values on the sensor display.
After the adjustment software on the PC has attained connection to VEGASCAN, you start with the "Configuration of your measuring system.
DCD 1 1 DCD RxD 2 2 RxD TxD 3 3 TxD DTR 4 4 DTR GND 5 5 GND
--- 6 6 ---
--- 7 7 ---
--- 8 8 ---
--- 9 9 ---
VEGASCAN 850 61
First start the adjustment software VVO on your PC.
Choose with the arrow keys or the mouse the item "
Planning
on the entrance screen
and click to "OK“.
In the next window you are asked for the user identification.
Setup - Adjustment with the PC
If the VVO (adjustment software) gets in contact with VEGASCAN for the first time, you are asked if you want to transfer the data from VEGASCAN to the PC.
Enter under name "VEGA.
Also enter "VEGA under password.
The adjustment program VEGA Visual Ope­rating (VVO), called in the following VVO, gets into contact with VEGASCAN and the sensors
and asks in which mode you want to use the adjustment software. After a few seconds the software indicates if and with which sen­sor a connection exists.
Click to "
YES
.
If you want to connect the VVO to a VEGASCAN, to which in the meantime an­other PC was connected, you will be asked, if you want to transfer the saved data from the actual PC to VEGASCAN or the data of VEGASCAN to the PC.
When clicking to "
Quit
, you are immediately in the main menu window of the adjustment software VVO.
The preset user identification can be modified later in the menu "
User access
Configuration/Program/
.
62 VEGASCAN 850
Setup - Adjustment with the PC
Note:
If you get no sensor connection, check the following:
- Is the sensor being supplied with voltage from VEGASCAN (min. 17 V)?
- Are you using the correct interface cable?
- Is the RS 232 cable connected to the cor­rect COM-Port?
Change COM-Port
In the menu "
nication
eters of your PC or change the activated COM-Port.
Configuration/Program/Commu-
you can adjust the interface param-
Configuration
During the setup of VEGASCAN you are confronted with two terms: "Configuration and "Parameter adjustment. The meas. sys­tem VEGASCAN is first set up with a configu­ration and then with a parameter adjustment.
Configuration
The term "Configuration means the basic adjustments of VEGASCAN. You inform VEGASCAN about the application (level measurement, gauge, distance ), the measurement loop name and the DCS output address of the sensors. The configuration corresponds to an electronic wiring and labelling of your VEGASCAN or just which sensor for what application and where.
Parameter adjustment
After the configuration, you carry out the parameter adjustment for each individual sensor. This means, to adjust the sensors to the respective operating range and to adjust to the actual application. You inform the sen­sor which product distance (which level) is "empty and which "full. This is called adjust­ment. Here you choose in which physical unit (volume, mass) and unit of measurement (m3, gal, liters ) the adjusted measured value should be outputted. In the submenu "Sensor optimisation you inform the sensor electron­ics about the actual environment, such as e.g. quick changes of the measured value, foam generation, gas stratification, solid or liquid.
VEGASCAN 850 63
Setup - Adjustment with the PC
Configuration info
Choose the menu "Configuration/Measur-
ing system.
You reach the menu window "
measuring system
. Here you get all informa-
Configuration
tion on the connected VEGASCAN.
User authorisation
Save your password with "
Save
and you are again in the initial menu of the configu­ration.
Communication
Click in the menu window "
measuring system
Configuration
to "
Communication
.
Unauthorised access to VEGASCAN can be prevented by a password. For each connec­tion location of the PC, on the D-SUB plug in the front of VEGASCAN (PC) or fixed con­nection in the terminal compartment (PC2), a respective password can be allocated.
Click to "
64 VEGASCAN 850
Access protection activated“.
Here you adjust the interface parameters for your signal output (here Modbus).
Click to " with "
Click to "
Quit
Save
or quit the menu window
“.
VBUS1
.
Setup - Adjustment with the PC
You get an overview of the sensors con­nected to the VBUS1 terminal position and their configuration status.
If a sensor was logged in as described in the following menu " status "
Configure
overview.
Create new measurement loop
Choose the menu "Configuration/Measure-
ment loop/New
In the menu window "
ment loop tion
with "OK.
Measurement loop/New
is written in the sensor
“.
, you choose "
Create new measure-
a new applica-
, the
You are in the menu window "
measurement loop - Application
Choose the parameter "
ment
and the sensor type "
Click to "
Choose in the next menu window "
ard level measurement
Click to "
After a few seconds, the menu window "
Continue
Continue
“.
“.
Create new
.
Level measure-
Ultrasonic
and "
Stand-
no options
“.
“.
Cre­ate new measurement loop - Sensor configu­ration
opens.
Click to "
The small menu window "
tion
VEGASCAN 850 65
Sensor coordination
opens.
“.
Sensor coordina-
Setup - Adjustment with the PC
Choose the serial number of the sensor to be assigned and confirm with "OK“.
In the menu window "
ment loop - Sensor configuration
number and the serial number of the sensor are displayed instead of "not configured“.
Now click in the menu window "
Create new measure-
, the VBUS
-Create new
measurement loop - Sensor configuration
to "
Continue
Click in the menu window "
“.
Create new measurement loop - Measurement loop designation
to "
Level
.
The menu window "
Create new measurement
loop - Measurement loop designation
Enter here the measurement loop number and a measurement loop description.
Assign here also one or several PC/DCS addresses to the actual sensor by clicking to "
PC/DCS address
“.
opens.
66 VEGASCAN 850
Setup - Adjustment with the PC
Confirm the adjustment with "OK“.
Also here confirm with "OK and you are
again in the initial window "
Create new measurement loop - Measurement loop designation
, in which the measurement loop number and the previously entered measurement loop description are now displayed.
In the menu "
Configuration/Measuring sys-
tem“ you now see the configured sensors.
Now, you only have to carry out the param­eter adjustment for the sensor. The param­eter adjustment is described in the following under Parameter adjustment 1 and Param-
Click to "OK and you are again in the main menu window.
You have just carried out the configuration
eter adjustment 2. Parameter adjustment 1 is obligatory, whereas parameter adjustment 2 (after sensor optimisation) must only be car­ried out if necessary.
settings for one sensor on VEGASCAN.
VEGASCAN 850 67
Parameter adjustment 1
Setup - Adjustment with the PC
In the menu "
justment
adjustments.
Adjustment
Choose the menu "
eter adjustment
want to adjust.
In the opening menu window, you now see the previously entered measurement loop names and the measurement loop descrip­tions. If you have only configured one sensor, naturally you will have a choice of only one sensor.
Instrument data/Parameter ad-
you carry out all important sensor
Instrument data/Param-
and then the sensor you
Then click to "OK, the menu window "
strument data parameter adjustment
opens.
Click to "
Adjustment
.
In-
Click in the menu window "
"
Min/Max-adjustment
Click to the sensor or the measurement loop on which you want to carry out the parameter adjustment.
68 VEGASCAN 850
.
Adjustment
to
Setup - Adjustment with the PC
You can carry out the min./max. adjustment
"with medium
actual level) or taking the actual level into consideration, i.e. with empty vessel).
Generally, you will carry out the adjustment without medium. Then you are not concerned about the actual vessel filling during adjust­ment. When you want to carry out the adjust­ment with medium, you have to carry out the min. adjustment with emptied (also partly emptied) vessel and the max. adjustment with filled (also partly filled) vessel. It is con­venient and quick to carry out the adjustment without medium, as shown in the example.
Choose "
Choose if you want the carry out the ad-
justment in
Enter a distance for the upper and lower level and the volume in % corresponding to each distance.
In the example, the 0 % filling is at a product distance of 3.400 m and the 100 % filling at a product distance of 0.500 m.
Note:
The instrument is supplied with the operating range identical to the sensor measuring range. After sensor adjustment, the operating range corresponds to the adjustment. The sensor can only detect levels within the de­fined operating range. For the detection of levels outside the operating range, e.g. new adjustment, it is necessary to correct the operating range in the menu "
sation/Meas. environment
(adjustment by means of the
"without medium
no (adjustment without medium)
meters
(m) or in
(without
feet
(ft).
Sensor optimi-
Confirm your adjustments with "OK and
you are again in the menu window "
ment
“.
Click in the menu window "
"
Quit
.
You are again in the menu window "
.
ment data parameter adjustment
Now the sensor electronics has two charac­teristics points, one for min. and one for max., out of which a linear correlation between product distance and the percentage filling of a vessel is generated. Naturally the characteristics point must not be at 0 % and 100 %, however the distance should be as big as possible (e.g. at 20 % and at 80 %). The difference between the characteristics points for the min./max. ad­justment should be at least 20 mm product distance. This is a theoretical value, because if the characteristics points are too close together, the possible measuring error in­creases. Ideal would be, to carry out the adjustment as shown in the example, at 0 % and at 100 %.
Adjustment
.
Adjust-
to
Instru-
VEGASCAN 850 69
Setup - Adjustment with the PC
In the menu "
justment/Conditioning/Linearisation
enter later, if necessary, a correlation be­tween product distance and percent value of volume other than linear (see later subitem linearisation).
Scaling
Click in the menu window "
parameter adjustment
The menu window "
Instrument data/Parameter ad-
you can
Instrument data
to "
Conditioning
Conditioning
opens.
“.
As parameter you can choose
less
(plain numbers),
and distance
measuring unit (e.g. l, hl). The sensor display then shows the measured value in the se­lected parameter and unit.
Save the adjustments in the menu " by clicking "OK.
volume, mass, height
and assign an appropriate
"dimension-
Scaling
The adjustments are now transferred to the sensor and you are again in the menu win­dow "
Conditioning
Click in the menu window " "
Quit
“.
Click to "
In the menu " 0 % and 100 % values of the parameter and their unit. You thereby inform the sensor, e.g. that at 0 % filling there are still 45 liters and at 100 % filling 1200 liters in the vessel. The sensor display then shows with empty vessel (0 %) 45 liters and with full vessel (100 %) 1200 liters.
70 VEGASCAN 850
Scaling
Scaling
“.
you enter the actual
Click in the menu window "
parameter adjustment
.
Conditioning
to
Instrument data
to "
Quit
.
Setup - Adjustment with the PC
Sensor optimisation
In the menu "Sensor optimisation you can adapt the sensor to the environment. Here you carry out special optimising adjustments and optimise, e.g. by means of the echo curve, the mounting location of the sensor.
Meas. environment/Operating range
Choose the menu "
eter adjustment
Choose in the menu window "
data parameter adjustment
"
Sensor optimisation and click to "
A
“.
First click to "Meas. environment.
Instrument data param-
Instrument
the menu item
Sensor
The window "
With the menu item " can define the operating range of the sensor deviating from the measuring range (de­pending on the sensor type) and the "
Max adjustment
range corresponds otherwise to the min./ max. adjustment (span) or the measuring range.
Generally, it is better to choose the operating range approx. 5 % bigger than the adjusted measuring range (span) determined by the min./max. adjustment. In the example:
- Min. adjustment to 1.270 m,
- Max. adjustment to 5.85 m. In the example you would have adjusted the operating range from 1 m to 6 m.
Meas. environment
opens.
Operating range
. By default, the operating
you
Min/
Save the adjustments and quit the menu window " with "
VEGASCAN 850 71
Limitation of the operating range
Quit
“.
Meas. environment/Meas. conditions
Setup - Adjustment with the PC
Click in the window "
"
Measuring conditions.
In the menu window " you click on the options corresponding to your application.
Confirm with "OK.
After a few seconds of saving during which the adjustments are permanently saved in the sensor, you are again in the window "
Meas. environment
Sonic velocity
When measuring in a gas composition devi­ating from air, e.g. pure nitrogen, the sonic velocity will change. Without correction, this would distort the measuring result.
Meas. environment
Measuring conditions
“.
to
Confirm with "OK“.
Click in the menu window "
ment
to "Quit.
You are again in the menu window "
optimisation
Echo curve
If, due to vessel installations, you expect strong false echoes, a correction (if possible) of the mounting location and orientation (dur­ing simultaneous monitoring of the echo curve) can help localise the false echoes and reduce their intensity. With the menu item "
Echo curve
optimisation
intensity of the detected ultrasonic echoes.
“.
in the menu window " you can see the form and the
Meas. environ-
Sensor
Sensor
In the menu " can be adapted to the correct value in case of another gas composition.
72 VEGASCAN 850
Sonic velocity
the sonic velocity
Setup - Adjustment with the PC
In the illustration, you see the echo curve (before correction of the sensor orientation) with a false echo nearly as large as the prod­uct echo.
In the next illustration you then see the echo curve after optimum directing of the sensor to the product surface and away from a large false echo source (sensor axis perpendicular to the product surface). The false echo, caused e.g. by a strut, is now reduced by more than 30 dB and will no longer influence the measurement.
Quit the menu "
Echo curve
with "Quit.
False echo storage
With the menu item " the menu "
Sensor optimisation
False echo storage
you can au-
in
thorise the sensor to save false echoes. The sensor electronics then saves the false ech­oes in an internal database and assigns them a lower level of importance than the useful echo. The false echo storage should be car­ried out with empty vessel.
Click in the menu window "
tion
to the menu item "False echo storage.
Sensor optimisa-
VEGASCAN 850 73
Setup - Adjustment with the PC
Now click in the opening menu window "
False echo storage
oes. A small window opens.
Enter here the verified product distance and click to "Create new
You hereby authorise the sensor to mark all echoes before the product echo as false echoes. This prevents the sensor from erro­neously detecting a false echo as level echo.
Click to "Show echo curve.
to "Learn false ech-
.
Quit the menu with "Quit.
You are again in the menu window "
optimisation
reset all options out of the menu "
optimisation
Quit the menu window "
tion
"
Sensor optimisation sensor selection
with "
You are then in the initial menu window "
. With the menu item "
to default.
Sensor optimisa-
with "Quit and the menu window
Quit
.
strument data parameter adjustment
Sensor
Reset
Sensor
“.
you
also
In-
The curve of the false echo marking and the real echo curve (top) are shown.
74 VEGASCAN 850
Setup - Adjustment with the PC
Parameter adjustment 2
Linearisation
The relation between level and filling volume is described with so-called linearisation curves. If, in your vessel, there is a correla­tion other than linear between level ("
age value
(value of the filling volume), choose in the menu window " "
Instrument data/Parameter adjustment/Con-
ditioning
of the level) and the filling volume
Conditioning
“.
the menu item
Percent-
Beside the two programmed linearisation curves "
Horizontal cylindrical tank
"
Spherical tank
programmable curves
is a linear correlation between level and vol­ume.
User programmable linearisation curves
Click to "User programmable curveto enter your own vessel geometry or a user programmable filling curve.
Click to "Edit.
you can also enter "
. Linear means there
and
user
Click in the menu window " the menu item "Linearisation.
The menu window " correlation between percentage value of the level and the percentage value of the filling volume is preadjusted.
VEGASCAN 850 75
Linearisation
Conditioning
, in which a
to
Setup - Adjustment with the PC
The user programmable linearisation curve is generated by index markers. Each index marker consists of one value pair. A value pair is generated out of a value " and a value "
value
the level. "
Percentage value
represents the percentage value of
Linearised
represents the per-
Linearised
. "
Percentage
centage vessel volume at a certain percent­age value of the level.
In the field "
Transfer measured value
the current level as percent of the adjusted span is displayed. The measuring span has al­ready been adjusted with the min./max. ad­justment. In our example, the measuring span is 4.58 m and is in the range of 5.85 m (empty) up to 1.27 m (full), see the following illustration.
0 m
100 % or 1.27 m
The percentage value of 95.79 % means that now 4.387 m of the adjusted measuring span (4.58 m) are reached:
4.58 0.9579 = 4.387 m.
The distance (product distance) outputted by the sensor, in case you have chosen "
distance
as output parameter, is then:
5.85 – (4.58 0.9579) = 1.463 m.
If the index markers or value pairs of your vessel are not known, you have to define the linearisation curve by incremental filling or calculate it with the vessel calculation pro­gram of VVO (see next page).
Defining the linearisation curve by incre­mental filling
In the characteristics of the example, you see four index markers or value pairs. There is always a linear interpolation between the index markers. The example vessel consists of three cylindrical segments of different height and diameter. The middle segment has a considerably smaller diameter.
95.79 % or 1.463 m
Span
4.58 m 100 %
4.387 m
(95.79 %)
5.85 m or 0 %
Click the check box "Show scaled values, to have the selected measuring unit dis­played on the y-axis (bottom left in the
5.85 m meas. distance correspond to 0 %
menu window).
level. 1.27 m meas. distance correspond to 100 % level. The measuring span is therefore
4.58 m (5.85 m – 1.27 m = 4.58 m).
76 VEGASCAN 850
Setup - Adjustment with the PC
Index marker 1 is at 0 % filling (
value [%]
), corresponding in the example to
percentage
an actual distance to the product surface of
5.850 m (empty vessel). The volume quantity is 45 liters (fluid remaining in the vessel). Index marker 2 is at a filling level of 30 % (30 % of the meas. distance of
1.270 m 5.850 m). At a filling level of 30 %, there are 576 liters in the vessel (in our exam­ple). Index marker 3 is at a filling level of 60 %. At this filling level there are 646 liters in the ves­sel. Index marker 4 is at a filling level of 100 % (product distance 1.270 m), where 1200 liters are in the vessel.
Max. 32 index markers can be entered per linearisation curve (value pairs).
Max.
Min.
100 % (1.270 m) correspond to 1200 liters
Span (4.58 m)
0 % (5.850 m) correspond to 45 liters
Calculating the linearisation curve
(using previous tank example)
In the menu window "
programmable curve --
vessel calculation program. With the vessel calculation program you can calculate (using dimensions from the technical drawings of the vessel) the correlation of filling height to filling volume. If the curve is defined this way, gauging by incremental filling is not neces­sary - your sensor will output volume as a function of level.
Linearisation -- user
you can start the
Click to "Calculate.
The tank calculation program starts. In the top left corner you choose the vessel type (upright tank, cylindrical tank, spherical tank, individual tank form or matrix). When choos­ing matrix, you can enter a user-programma­ble linearisation curve by means of index markers. This corresponds to the input of value pairs (linearisation points), as previ­ously described. In the following example, the tank calculation program calculates the linearisation curve of a vessel, corresponding to the vessel in the previous gauging example.
Click to individual tank form and choose three round tank segments with the dimen­sions 0.9 m 0.9 m (height by diameter),
0.68 m 0.37 m and 0.68 m 1.02 m (this tank form corresponds to the tank form of the gauging example).
.
VEGASCAN 850 77
Setup - Adjustment with the PC
Click to
After a short calculation time, the levels in percentage of span and the respective vol­ume percentage will be shown. The outputted curves show the correlation in a diagram.
Quit the linearisation table with "OK.
"Calculate
.
You are again in the menu window "
calculation
Click to "OKto save the tank calculation.
You are again in the menu window "
.
tion -- user programmable curve --
volume percent values to the percentage values of the level are shown. When clicking the bottom left menu window "
values
, liters will be displayed depending on
the adjustment in the menu "
Tank
Linearisa-
. The
Show in scaled
Instrument data/
Parameter adjustment/Conditioning/Scaling
Calculate cylindrical tank
Click in the menu window "
user programmable curve --
"
Calculate and in the menu window "
calculation
tanks.
to the symbol for cylindrical
Linearisation --
to
Tank
.
78 VEGASCAN 850
Setup - Adjustment with the PC
The menu window for the adjustment of the cylindrical tank opens.
Choose the measuring unit which should be used for the input of the vessel dimen­sions, e.g. mm.
The following example shows a cylindrical tank, which inclined by 3°, has a cylinder length of 10000 mm, at a diameter of 5000 mm. The cylindrical tank has a 1500 mm wide, spherical form at the right end and a dished tank end at the left.
Above the information "
internal dimensions
All dimensions are
, you will find two fields with the percentage values 0 % and 100 %. Here you can shift the 100 % line or the 0 % line. In the example, the 100 % filling line was defined at a distance of 650 mm from the upper vessel edge (inner).
Click to "Calculate.
You will get the calculated linearisation table after a short calculation time. By means of 32 linearisation points, a function correlating vessel volume to filling height is outputted. The example vessel has a filling of 216561 liters at the 100 % line or 216.6 m3. It is possi­ble to output the volume in barrels, gallons, cubic yards or cubic feet.
Note:
In the bottom left corner in the menu window "
Tank calculation
dimensions are internal dimensions
you find the information "
. The
All
calculation programs calculates by means of the vessel inner dimensions the vessel vol­ume. The adjustment of the wall thickness is not necessary for the calculation of the dished boiler end as its mathematical calcula­tion is based on the outer dimension.
VEGASCAN 850 79
There is a linear interpolation between the linearisation points.
Click to "OK and you are again in the menu window "
Again click in the menu window "
culation
sation menu.
Here the calculated linearisation curve is again outputted. The volume information under "
Linearised
longer to the calculated volume of the tank calculation program. Why?
In the menu " tioning/Scaling) you entered that at 0 % filling there are 45 liters in the tank and at 100 % filling 1200 liters. The geometry of the calcu­lated cylindrical tank is reduced to a size that has a volume of only 1200 liters. The lineari­sation curve of the calculated vessel is there­fore modified to fit the volume data you entered in the menu "
If the true content of the calculated vessel should be outputted, the volume that was determined in the tank calculation program must be entered in the menu "Scaling“.
Tank calculation
.
Tank cal-
to "OK and you are in the lineari-
now corresponds no
Scaling
(Instrument data/Condi-
Scaling
.
Setup - Adjustment with the PC
The sensor then outputs the actual filling volume by means of the adjusted vessel dimensions.
Quit the menu with "OK.
Confirm with "OK and your individual lin-
earisation curve is saved in the sensor.
Again in the menu window " can enter with the menu item "
time
a measured value integration. This is recommended for agitated product surfaces, to prevent the measured value indication and output from changing constantly. As a standard feature, an integration time of 0 seconds is preset.
Quit the menu with "OK., you are again in
the menu window "
eter adjustment
Quit the menu window with "OK.
“.
Conditioning
, you
Integration
Instrument data param-
80 VEGASCAN 850
Setup - Adjustment with the PC
Parameter adjustment sensor display
In the menu item "Outputs you choose the scale and the unit in which your level should be displayed.
Choose in the main menu window "
ment data/Parameter adjustment
the menu item "
Outputs
“.
Instru-
and then
Meas. loop data
Click in the window "
rameter adjustment
Then click to " and "
VEGASCAN
tion on your measurement loop in the infor­mation windows.
Instrument data pa-
to "
Meas. loop data
Application
“, "
, to get detailed informa-
Input no.A
“.
In the menu window "
"
Display of measured value
Choose the sensor and carry out the ad­justments.
If you want to leave the adjustments un­changed, click to "
Click in the menu window " "
Quit
and you are in the menu window
"
Instrument data parameter adjustment
Click in the menu window "
parameter adjustment
VEGASCAN 850 81
Outputs
Quit
“.
again to "
you click to
“.
Outputs
to
“.
Instrument data
Quit
“.
Close the information window.
Quit the menu "
Click in the menu window "Instrument data
parameter adjustment to "
You are again in the main menu window.
Meas. loop data
Quit
“.
“.
Setup - Adjustment with the PC
Show measured value
Click in the main menu window to the menu "
Display/Display of measured value
choose the measurement loop or the sen­sor to be displayed.
Choose in the line " output "
Distance
distance will be displayed. When choosing "
Scaled
, e.g. the measured value will be displayed in liters and volume percent as well as the actual signal current in the 0/4 20 mA signal cable.
Indication value
and the sensor product
and
the
Simulation
Click to the menu " and choose the measurement loop.
The menu window " similar to the previous menu window, opens. In this menu window however, you can set the filling of the vessel or the signal current and the indication to any value (simulate measured value).
First of all, the actual measured value and the signal current are displayed.
Click to "Startin the turquoise window segment.
Diagnostics/Simulation
Simulation of outputs
,
82 VEGASCAN 850
Setup - Adjustment with the PC
The grey scrollbar becomes active. With this scrollbar your can modify the measured value in the range of -10 % 110 % and thereby simulate the filling or emptying of the vessel. In the field of the turquoise window cutout you can enter any % value for the filling percentage.
Choose available linearisation curve
On the previous pages, the entering or the calculation of a linearisation curve was shown as an example. In this menu item, you can choose from the available predefined or user­entered linearisation curves.
Note:
The simulated measured value is outputted during adjustment with the PC until you termi­nate the simulation mode.
A selected curve can be modified later (edit) or calculated with the calculation program.
VEGASCAN 850 83
Print configuration and adjustments
Setup - Adjustment with the PC
Click to "
Before printing the complete configuration of VEGASCAN and all sensors, you can have a look at the individual pages and
Services/Print
.
With the menu " justments displayed in detail.
View
you can have the ad-
then print everything or just certain pages.
84 VEGASCAN 850
Setup - Adjustment with the PC
Backup
With the menu items "
conditioning instruments Backup/Sensors
and parameter settings of VEGASCAN and of each sensor.
The settings can later be quickly transferred to new sensors, e.g. in case of a sensor exchange, or certain product settings can be quickly carried out. This is done in the menu "
Services/Restore configuration
For further information take the manual "VEGA Visual Operating (VVO).
Services/Backup/Signal
and "
you save the configuration
Services/
.
VEGASCAN 850 85
Setup - Sensor adjustment with the adjustment module MINICOM
6.3 Sensor adjustment with the ad­justment module MINICOM
Tank 1 m (d)
12.345
Beside the PC, the ultrasonic sensors VE­GASON 51V … 56V can also be adjusted with the small, detachable adjustment module MINICOM in the sensor.
Only the sensor-relevant adjustments such as e.g. scaling of the sensor display, operat­ing range, meas. conditions, sensor display scaling or false echo storage are possible with the adjustment module MINICOM. Not possible are all adjustment steps relating to the configuration, the conditioning and the signal processing (configuration of the inputs and outputs, linearisation curves, simulation ). This is only possible with the PC directly on VEGASCAN 850.
You carry out all adjustment steps with the 6 keys of the adjustment module. A small dis­play shows you, apart from the measured value, a short message on the menu item or on the value of a menu adjustment.
The information volume of the small display, however, cannot be compared with that of the adjustment program VVO, but you will soon get used to it and will be able to carry out your adjustments quickly and directly with the small MINICOM (see menu plan on the following pages).
ESC
+
-
OK
1. Measurement in gases
2. Operating range
3. Meas. conditions
4. False echo storage (only necessary, if failures occur during operation).
5. Indication of the useful and noise level
6. Processing/Scaling of the sensor display
In the following you will find the sensor-spe­cific setup items 1 … 6.
1. Measurement in gases
Adjustment is only necessary when the measurement is made in gases (Co2, He, etc.) deviating from air. In case of measure­ment in gases, sound the distance of the sensor to the product surface and enter it in the menu item "Measurement in gases. The sensor can then take the modified sonic velocity in gases (as opposed to air) and output correct levels.
2. Operating range
Without special adjustment, the operating range corresponds to the measuring range. The measuring range has already been ad­justed with the min./max. adjustment. Gener­ally it is useful to choose a slightly bigger (approx. 5 %) operating range than the measuring range.
Example: Min./max. adjustment: 0.500 … 3.500 m; adjust operating range to approx.
0.400 3.600 m.
86 VEGASCAN 850
Setup - Sensor adjustment with the adjustment module MINICOM
3. Meas. conditions
Here you inform each sensor under which application conditions the measurement is carried out, e.g. measurement in gas compo­sitions deviating from air, which will influence the running time of the ultrasonic signals or if a liquid (hard echo) or a solid (weak echo) should be measured.
4. False echo storage
A false echo storage is always useful when unavoidable false echo sources (e.g. struts) must be minimised. By creating a false echo storage, you authorise the sensor electronics to note the false echoes and save them in an internal database. The sensor electronics treats these (false) echoes differently from the useful echoes and filters them out.
5. Useful and noise level
In the menu
you get important information on the signal quality of the product echo. The higher the "S-N value, the more reliable the measure­ment (menu plan MINICOM).
Ampl.:
XX dB
S-N:
XX
dB
Example:
Ampl. = 68 dB S-N = 53 dB
68 dB – 53 dB = 15 dB noise level
53 dB signal level difference indicates very high measurement reliability.
6. Processing/Scaling of the sensor display
(see menu plan)
Ampl.: means amplitude of the level echo in
dB (useful level)
S-N: means Signal-Noise, i.e. the useful
level minus the level of the back­ground noise
The bigger the "S-N value (difference be­tween the amplitude useful level and the noise level), the better the measurement: > 50 dB Measurement excellent 40 50 dB Measurement very good 20 40 dB Measurement good 10 20 dB Measurement satisfactory 5 10 dB Measurement sufficient < 5 dB Measurement poor
VEGASCAN 850 87
Setup - Sensor adjustment with the adjustment module MINICOM
Menu schematic of the adjustment module MINICOM
Sensor
m(d)
4.700
SON52 V
1.00
When switching on, the sensor type and the software version are displayed for a few seconds.
Para­meter
2.
Sensor opti­mise
Meas. enviro nment
Opera­ting rage
Begin
m (d)
0.50
End
m (d)
6.00
3.
Meas. condit ions
Condit ion
Liquid
Fast change
No
Measur ing in gas
Meas. dist.
m (d)
4.700
Correc tion Now !
OK ?
Agitat ed sur face
No
1.
Foam­ing prod.
No
Measur ing in gas
No
4.
False echo memory
Create new
Meas. dist.
m (d)
X.XX
Echo learn Now!
OK ?
learn­ing
Mul­tiple echoes
No
Update
Meas. dist.
m (d)
X.XX
Update Now!
OK ?
learn­ing
Confi­gura­tion
Sensor Tag
Sensor
‘Delete
Delete Now!
OK ?
delet­ing
Meas. unit
m (d)
High
Condit ion
Solid
Fast change
No
dust level
No
Large angle repose
No
Measur ing in gas
No
Mul­tiple echoes
No
6.
Displ. adjust ment
Adjust ment in
m(d)
0.0%
at
m (d)
XX.XXX
Signal condit ioning
Sca­ling
0 % corres ponds
XXXX
100 % corres ponds
XXXX
Deci­mal­point
8888
Prop. to
Volume
Unit
l
88 VEGASCAN 850
Out­puts
Sensor dis­play
Prop. to
Dis­tance
Setup - Sensor adjustment with the adjustment module MINICOM
With these keys you move in the menu field to the left, right, top and bottom
ESC
Basic Reset
Reset Now!
OK ?
reset­ing
Dis­tance
m (d)
4.700
5.
Ampl.:
XXdB
S-N:
XX
max. range
m (d)
OK
7.000
Add´l func­tions
Basic
Info
reset
Reset
dB
Sensor Tag
Sensor
Now !
Reset­ing
Sensor typ
SON 52 V
OK ?
Serial nubmer
1094 8218
Lan­guage
Eng­lish
Softw. Vers.
3.00
Softw. Date
11.09. 1997
Ampl.:
Meas. dist.
X,XX
Fast change
No
Foam­ing prod.
No
Menu items in bolt print provide sensor and measured value information and cannot be modified in this position.
Light grey menu fields are only displayed if required (dependent on the adjustments in other menus).
White menu items can be modified with the "+“ or "–“ key and saved with the "OK“ key.
Dis­tance
m (d)
3.700
S-N:
71dB
48
dB
Tempe­rature
Actual temp.
Max.
Min.
temp.
temp.
°C
18
°C
°C
59
4
VEGASCAN 850 89

7 Diagnostics

Diagnostics

7.1 Simulation

For simulation of a certain filling, you can call up the function "Simulation on the adjustment module MINICOM, in the software program VVO or on the signal conditioning instrument.
You simulate a vessel filling and thereby a certain sensor current. Please note that con­nected instruments, such as e.g. a PLC react according to their adjustments and will prob­ably activate alarms or system functions.
Simulation with VVO
If you start the simulation mode with the ad­justment program VVO on the PC, the simu­lated level is outputted until you quit the simulation mode.
Simulation with MINICOM
If you start the simulation mode on the adjust­ment module MINICOM, the sensor returns to standard operating mode after one hour.

7.2 Error codes

Error codes Removal
E013 No valid measured value Message is displayed during the warm-up phase
- Sensor in the warm-up phase
- Loss of the useful echo If the message remains, a false echo storage (with the adjustment software on the PC - see Echo curve under Sensor optimisation) must be carried out together with a modification of mounting location and orientation to achieve the lowest possible false echo background.
E017 Adjustment span too small Carry out a readjustment.
E036 Sensor software does not run Sensor requires new software (service).
E040 Hardware failure/Electronics defec- Check all connection cables.
tive Transducer defective.
90 VEGASCAN 850
Make sure that the difference between min. and max. adjustment is at least 10 mm.
Message appears during a software update.
Contact our service department.
Notes
VEGASCAN 850 91
VEGA Grieshaber KG Am Hohenstein 113 D-77761 Schiltach Phone (0 78 36) 50 - 0 Fax (0 78 36) 50 - 201 e-mail info@vega-g.de internet www.vega-g.de
ISO 9001
The statements on types, application, use and operating conditions of the sensors and processing systems correspond to the latest infor­mation at the time of printing.
Technical data subject to alterations
2.24 631 / March 2000
Loading...