VEGA PULS45 User Manual

Operating Instructions
VEGAPULS 42, 44 and 45 – VBUS
Level and Pressure
Contents
Safety information ........................................................................ 3
Note Ex area ................................................................................ 3
1.1 Function................................................................................. 4
1.2 Application features ............................................................. 6
1.3 Adjustment ............................................................................ 6
2 Types and versions
2.1 Overview .............................................................................. 9
2.2 Type code........................................................................... 11
2.3 Antennas............................................................................. 12
3 Mounting and installation
3.1 General installation instructions ........................................ 14
3.2 Measurement of liquids ..................................................... 16
3.3 Measurement in standpipe (surge or bypass tube) ...... 18
3.4 False echoes ...................................................................... 25
3.5 Installation mistakes ........................................................... 27
Contents
4 Electrical connection
4.1 Connection and connection cable .................................... 30
4.2 Connection of the sensor .................................................. 33
4.3 Connection of the external indicating instrument
VEGADIS 50 ....................................................................... 34
4.4 Configuration of measuring systems ............................... 35
2 VEGAPULS 42, 44 and 45 – VBUS
Contents
5Setup
5.1 Adjustment structure ......................................................... 42
5.2 Adjustment with the PC on the signal
conditioning instrument ...................................................... 43
5.3 Adjustment with VEGAMET or MINICOM ........................ 69
5.4 Adjustment with the PC on VEGALOG ............................ 76
6 Diagnosis
6.1 Simulation ............................................................................ 82
6.2 Error code .......................................................................... 82
7 Technical data
7.1 Data ..................................................................................... 83
7.2 Approvals ........................................................................... 88
7.3 Dimensions ......................................................................... 89

Safety information

Please read this manual carefully, and also take note of country-specific installation standards (e.g. the VDE regulations in Germany) as well as all prevailing safety regulations and acci­dent prevention rules. For safety and warranty reasons, any internal work on the instruments, apart from that in­volved in normal installation and electrical con­nection, must be carried out only by qualified VEGA personnel.
VEGAPULS 42, 44 and 45 – VBUS 3

Note Ex area

Please note the attached safety instructions containing important information on installation and operation in Ex areas. These safety instructions are part of the oper­ating instructions and come with the Ex ap­proved instruments.

1 Product description

Product description
VEGAPULS series 40 sensors are a newly developed generation of very compact, small radar sensors for high resolution and accu­racy. They are characterised by special focusing features for level measurement in narrow space applications. With modest space requirements they were developed for measuring distances of 0 … 4 m/10 m/20 m and for standard applications such as stor­age vessels and buffer tanks as well as for process tanks. With small housing dimen­sions and process fittings, the compact sen­sors monitor your levels at reasonable cost. With the integrated display they enable high precision level measurements and due to their advantages, radar sensors have increasingly become the standard solution for many applications, even for those where it was previously unthinkable due to high costs.
VEGAPULS 40 radar sensors utilise two-wire technology perfectly. The supply voltage and the output signal are transmitted via one two­wire cable. They provide a digital output signal (VBUS or Profibus) or an analogue 4 … 20 mA signal as output or measuring signal.
This operating instruction manual relates to VEGAPULS 42, 44 and 45 sensors with digital VBUS output signal (VEGA-Bus).

1.1 Function

Radio detecting and ranging: Radar.
VEGAPULS radar sensors are used for non­contact, continuous distance measurement. The measured distance corresponds to a filling height and is outputted as level.
Measuring principle:
emission – reflection – reception
Tiny 26 GHz radar signals are emitted from the antenna of the radar sensor as short pulses. The radar impulses reflected by the sensor environment and the product are received by the antenna as radar echoes. The running period of the radar impulses from emission to reception is proportional to the distance and hence to the level.
Meas. distance
emission - reflection - reception
The radar impulses are emitted by the an­tenna system as pulse packages with a pulse duration of 1 ns and pulse intervals of 278 ns; this corresponds to a pulse package frequency of 3.6 MHz. In the impulse inter­vals, the antenna system operates as re­ceiver. Signal running periods of less than one billionth of a second must be processed and the echo image evaluated in a fraction of a second.
1 ns
278 ns
Pulse sequence
4 VEGAPULS 42, 44 and 45 – VBUS
Product description
VEGAPULS radar sensors can accomplish this through a special time transformation procedure which spreads out the more than
3.6 million echo images per second in a slow­motion picture, then freezes and processes them.
tt
Time transformation
Hence, it is possible for the VEGAPULS ra­dar sensors to process the slow-motion pictures of the sensor environment precisely and in detail in cycles of 0.5 to 1 second without using time-consuming frequency analysis (e.g. FMCW, required by other radar techniques).
Virtually all products can be measured
Radar signals are physically similar to visible light. According to the quantum theory they propagate through empty space and are therefore not dependent, like e.g. sound waves, on a conductive medium. And like all electromagnetic waves, they travel at the speed of light. The radar signals react to two primary electrical properties:
- the electrical conductivity of a substance.
- the dielectric constant of a substance.
All products which are electrically conductive or have a dielectric constant greater than ε
1.6 have sufficiently good reflection charac-
r
teristics for reliable radar measurement. (Note: air has a dielectric constant ε The signal reflection increases with the con-
of 1).
r
ductivity or with the dielectric constant of the product. Therefore virtually all products can be measured.
%
50 40 30 20 10
5 %
5
0
2
4 6 8 12 14 16 18
0
25 %
10
40 %
20
ε
r
Reflected radar power dependent on the dielectric constant of the measured product
With standard flanges between DN 50 and DN 150, ANSI 2" to ANSI 6" or G 1½ A and 1½" NPT the sensor antenna systems are adapted to the various products and meas­uring environments.
High quality materials withstand also extreme chemical and physical conditions. The sen­sors deliver reliable, precise and long-term stable, reproducible level signals.
Continuous and reliable
Unaffected by temperature, pressure and individual gas atmospheres, VEGAPULS radar sensors are used for quick and reliable continuous level measurement of various products.
%
0,03 0,02 0,01
0
100 500 1000 1300 ˚C
0
0,018 %
Temperature influence: Temperature error absolutely zero (e.g. at 500°C 0.018 %)
%
10
5
0,29 %
0
10
0
1,44 %
20 30 40 60
50
Pressure influence: Error with pressure increase very low (e.g. at 50 bar 1.44 %)
0,023 %
2,8 %
70 80 90 110 120 130 140
100
3,89 %
bar
VEGAPULS 42, 44 and 45 – VBUS 5
Product description

1.2 Application features

Applications
• Level measurement of all liquids.
• Measurement also in vacuum.
• All slightly conductive substances and all substances with a dielectric constant > 2.0 can be measured.
• Measuring range 0 … 10 m (type 42). Measuring range 0 … 20 m (type 44). Measuring range 0 … 4 m (type 45).
Two-wire technology
• Power supply and output signal on one two-wire cable (loop powered).
• VBUS output signal (VEGA-Bus).
Rugged and wear-free
• Non-contact.
• High resistance materials.
Exact and reliable
• Accuracy 0.05 %.
• Resolution 1 mm.
• Unaffected by noise, steam, dust, gas compositions and inert gas stratification.
• Unaffected by varying density and tem­perature of the medium.
• Measurements of pressures up to 40 bar and temperatures up to 150 °C.

1.3 Adjustment

Each measuring situation is unique. For that reason, every radar sensor needs some basic information on the application and the environment, e.g. which level means "empty" and which level "full". Beside this "empty and full adjustment", many other settings and adjustments are possible with VEGAPULS radar sensors.
Adjustment and parameter setting of the radar sensors is done with
- the PC and the adjustment program VVO
- the detachable adjustment module MINICOM
- the VEGAMET signal conditioning instru­ment
Adjustment with the PC
Setup and adjustment of the radar sensors is generally done on the PC with the adjustment program VEGA Visual Operating (VVO) under Windows graphics and process visualisation, the pro­gram leads you quickly through adjustment and parameterisation.
®.
By means of pictures,
Communicative
• Integrated display of measured values.
• Optional display separate from the sensor.
• Adjustment with detachable adjustment module which can be plugged into the sensor or into the external display.
• Adjustment with signal conditioning instru­ment.
• Adjustment with PC.
Approvals
The adjustment program recognises the sensor type
• CENELEC, ATEX, PTB, FM, CSA, ABS, LRS, GL, LR, FCC, RINA.
Visualised creation of a vessel linearisation curve
6 VEGAPULS 42, 44 and 45 – VBUS
Product description
Note:
The adjustment program VVO must be avail­able in version 2.70 or higher.
The PC can be connected to any position of the system or the signal cable. It is con­nected with the two-wire PC interface con­verter VEGACONNECT 2 to th e sensor, the signal cable or the signal conditioning instru­ment. The sensor is connected directly to the VEGALOG processing system with the standard cable (RS 232).
2
2
Adjustment with the PC on the digital signal and supply cable between sensors and VEGAMET signal conditioning instrument or directly on the sensor
2
1 ... 15
CPU
VEGALOG
VEGALOG
571 CPU
571 EA
Adjustment on the digital signal and supply cable to the VEGALOG 571 processing system or directly on the sensor
The adjustment and parameter setting data can be saved on the PC with the adjustment software and protected by passwords. The adjustments can be quickly transferred to other sensors if required.
2
......
CPU
2
1...15
VEGALOG
VEGALOG
571 CPU
571 EA
2
Adjustment with the PC on the VEGALOG processing system with standard cable RS 232 (up to 15 sensors can be operated on one two-wire cable with the processing system)
Adjustment with the PC on the VEGAMET signal conditioning instrument, to which one or two sensors can be connected
VEGAPULS 42, 44 and 45 – VBUS 7
Product description
Adjustment with adjustment module MINICOM
With the small (3.2 cm x 6.7 cm) 6-key ad­justment module with display you carry out the adjustment in clear text dialogue. For this purpose, the adjustment module can be plugged into the radar sensor or into the optional external indicating instrument.
Tank 1 m (d)
12.345
Detachable adjustment module MINICOM
Unauthorised sensor adjustments can be prevented by removing the adjustment mod­ule.
ESC
+
-
Tank 1
m (d)
OK
12.345
2
ESC
+
-
Tank 1
m (d)
OK
12.345
4
ESC
+
-
OK
%
100
+
-
OK
ESC
CONNECT
!
on
513
Adjustment with the VEGAMET signal conditioning instrument
The radar sensors with digital output signal can be adjusted, besides with the PC, also with the VEGAMET signal conditioning instru­ment.
%
100
+
-
OK
ESC
CONNECT
12
!
on
515V
6-key adjustment field on the instrument front of the VEGAMET signal conditioning instrument
For adjustment, the digital VEGAMET 514V and 515V signal conditioning instruments are provided with a 6-key adjustment field with display. Here you can carry out the param­eter setting in clear text dialogue. Using the 6-key adjustment field, it is also possible to adjust one or two connected sensors with the signal conditioning instrument. The adjustment scheme corresponds to that of the adjustment module MINICOM.
Adjustment with detachable adjustment module. The adjustment module can be plugged into the radar sensor or into the external indicating instrument VEGADIS 50.
8 VEGAPULS 42, 44 and 45 – VBUS
Types and versions

2 Types and versions

2.1 Overview

Series 40 sensors are manufactured in three basic versions, VEGAPULS 42, VEGAPULS 44 and VEGAPULS 45.
VEGAPULS 42 sensors are characterised by
1
G 1
/2 A or 11/2" NPT thread as process fitting. These sensors are equipped as standard versions with a ø 40 mm horn as antenna.
VEGAPULS 44 sensors are characterised by a DIN or ANSI flanges as process fitting. In standard version they are manufactured with DN 50, 80, 100 and 150 as well as with ANSI 2", 3", 4" and 6". The bigger flanges come equipped with respectively larger antenna horns (ø 48, 75 and 95 mm).
Generally: The bigger the antenna horn, the better the focusing characteristics, and the better the antenna gain. This ensures that even a weak product echo can be detected reliably as level echo.
VEGAPULS 45 sensors come completely equipped with a measuring tube of 4 m length. These sensors are used in applications where high accuracy is required or for liquids with very low dielectric constant such as e.g. liquid gas.
VEGAPULS 42
VEGAPULS 45 with thread G 1½ A or 1½" NPT
VEGAPULS 44
VEGAPULS 45 with flange
VEGAPULS 42, 44 and 45 – VBUS 9
Types and versions
Features
General features
• Measurement preferably of liquids in storage tanks or process vessels with increased accuracy requirement.
• Measuring range 0 … 4 m, 0 … 10 m or 0 … 20 m.
• Ex approved in Zone 1 (IEC) or Zone 1 (ATEX) classification mark EEx ia [ia] IIC T6.
• Integrated display of measured values.
Overview
PS42XXE… PS44XXE… PS45XXXE…
Signal output
- VBUS (VEGA-Bus)
Antenna
- horn antenna
- pipe antenna
1) 1)
Process fitting
- G 1½ A; 1½" NPT
- DN 50; ANSI 2"
- DN 80; ANSI 3"
- DN 100; ANSI 4"
- DN 150; ANSI 6"
Adjustment
-PC •••
- adjustment module in the sensor
- adjustment module in external indicating instrument
- with signal conditioning instrument
Measuring range 0 … 4 m
- ø 40 mm horn 0 … 10 m (depending on the
- ø 48 mm horn 0 … 15 m 0 … 15 m tube length)
- ø 75 mm horn 0 … 20 m 0 … 20 m
- ø 95 mm horn 0 … 20 m 0 … 20 m
1)
When mounting on a standpipe or bypass tube, a pipe antenna is created. The tube inner diameter should be between 40 mm and 80 mm.
10 VEGAPULS 42, 44 and 45 – VBUS
Types and versions

2.2 Type code

PS 42 .XX X X X XXX X X
K - Plastic housing PBT, M20 x 1.5 cable entry N - Plastic housing PBT, ½" NPT cable entry A - Aluminium housing, M20 x 1.5 cable entry D - Aluminium housing, ½" NPT cable entry in Exd connection
housing
V - Seal of the antenna system: Viton A - Seal of the antenna system: Kalrez
G - Process fitting G 1½ N - Process fitting 1½" NPT
A
ABC- Process fitting DN 50 PN 40 BBE- Pr ocess fitting DN 80 PN 40 CBG-Process fitting DN 100 PN 16 DBG-Process fitting DN 150 PN 16 ARC- Process fitting ANSI 2" 150 psi BRE- Pr ocess fitting ANSI 3" 150 psi CRG-Process fitting ANSI 4" 150 psi DRG-Process fitting ANSI 6" 150 psi YYY- Process fitting on request
X - without display A - with integrated display
X - without adjustment module MINICOM B - with adjustment module MINICOM (mounted)
B - 20 … 72 V DC; 20 … 250 V AC; 4 … 20 mA, HART
(four-wire)
D - Two-wire (loop powered), 4 … 20 mA, HART
®
®
E - Power supply via signal conditioning instrument G - Segment coupler for Profibus PA
XX - FTZ (standard telecommunication approval Germany) AX - Approval in Ex-Zone 1, EEx ia IIC T6 CX - Approval in Ex-Zone 0, EEx ia IIC T6 BX - Approval in Ex-Zone 1 (Exd connection housing) DX - Approval in Ex-Zone 0 (Exd connection housing)
Type 42: with screwed process fitting Type 44:Instrument series with flange connection Type 45: with measuring tube
PS: Radar sensors series 40
VEGAPULS 42, 44 and 45 – VBUS 11
Types and versions

2.3 Antennas

The antenna is the eye of the radar sensor. A casual observer would never guess how carefully the antenna geometry must be adapted to the physical properties of electro­magnetic fields. The geometrical form deter­mines focal properties and sensitivity - the same way it determines the sensitivity of a unidirectional microphone.
For various application purposes, measure­ment conditions and process requirements, series 40 sensors are available with three antenna forms as VEGAPULS 42, VEGAPULS 44 and VEGAPULS 45.
VEGAPULS 42 sensors are characterised by
1
G 1
/2 A or 11/2" NPT thread as process fitting. These sensors are equipped as standard versions with a ø 40 mm horn as antenna.
VEGAPULS 44 sensors are characterised by a DIN or ANSI flanges as process fitting. In standard version they are manufactured with DN 50, 80, 100 and 150 as well as with ANSI 2", 3", 4" and 6". The bigger flanges come equipped with respectively larger antenna horns (ø 48, 75 and 95 mm).
Generally: The bigger the antenna horn, the better the focusing characteristics, and the better the antenna gain. This ensures that even a weak product echo can be detected reliably as level echo.
VEGAPULS 45 sensors are characterised by integrated measuring tube. The measuring tube serves the radar signals as a waveguide in which no emitted energy is lost. The complete emitted energy is returned as reflection energy. Due to this, also products with very weak reflection characteristics,
such as light petrol, liquid gas with of 1.4 … 2.0 can be reliably detected.
ε
values
r
Viscous or adhesive products cannot be measured with VEGAPULS 45.
Horn antennas
Horn antennas focus the radar signals very well. Manufactured from 1.4435 (StSt) or Hastelloy C22, they are very rugged, and both physically and chemically resistant. They are suitable for pressures up to 40 bar and for medium tempera­tures up to 150°C. The horn
VEGAPULS 42
diameters determine the focusing of the radar sig­nals. The antenna gain in­creases with increasing horn diameter (40, 48, 75, 95 mm). The antenna gain represents the relation between transmitted energy and received echo energy.
VEGAPULS 44
12 VEGAPULS 42, 44 and 45 – VBUS
Types and versions
Pipe antennas
Pipe antennas composed of horn antenna and standpipe or bypass tube
Horn antennas on a stand­pipe or bypass tube form a complete antenna system in conjunction with a measuring pipe (which must not be perfectly straight, but can also have bends). For the radar signals, the measuring pipe acts as a conductor. The running time of the radar signal changes in the pipe and depends on the pipe diameter. The sensor must be therefore informed about the tube inner diameter so that the running time can be taken into account and precise level signals are delivered. Pipe antennas are best suited for turbulent applications and products with very low dielectric con­stant.
The antennas are characterised by a very high antenna gain. As a result, high reliability is achieved even with products having very poor reflection proper­ties.
VEGAPULS 42 pipe antenna composed of horn antenna and bypass tube
VEGAPULS 44 pipe antenna composed of horn antenna and bypass tube
VEGAPULS 45 with thread, pipe antenna inte­grated in the sensor
VEGAPULS 45 with flange, pipe antenna inte­grated in the sensor
Pipe antenna integrated in the sensor
VEGAPULS 45 series originated as a further development of VEGAPULS 42 and 44 sensors combined with a surge pipe or bypass tube. This sensor is equipped with an optimised measuring tube up to 4 m long and enables high precision level measurement also in products with very small dielectric figures of ε liquid gas).
VEGAPULS 42, 44 and 45 – VBUS 13
= 1,4 … 1,8 (e.g.
r

3 Mounting and installation

3.1 General installation instructions

Mounting and installation
Measuring range
The reference plane for the measuring range of the sensors is the flange face or the seal shoulder of the thread. For measurements in
Keep in mind that in measuring environments where the medium can reach the sensor flange, buildup can occur on the antenna and
can cause measurement errors. surge or bypass pipes with VEGAPULS 45 the max. measuring distance depends on the tube length.
Reference plane
max. filling
Measuring range (operating range) and max. measuring distance Note: Use of the sensors for applications with solids is limited.
full
empty
Meas. range
max. measuring distance 20 m (type 45: 4 m)
Note: The use of series 40 sensors for solids
is restricted.
max.
min.
Interfering reflections
Flat obstructions and struts cause large interfering reflections. They reflect the radar signal with high energy density. Rounded profile interfering surfaces scatter the radar signals in all directions and thus cause interfering reflections of lower energy density. Hence, they ar e less critical than reflections from a flat surface.
If flat obstructions in the range of the radar
signals cannot be avoided, we recommend
diverting the interfering signals with a deflec-
tor. Through scattering, the interfering signals
will be low in amplitude and so diffuse that
they can be filtered out by the sensor.
max.
max.
min.
Rounded profiles diffuse radar signals
Profiles with flat surfaces cause stronger interfering signals
A deflector causes signal scattering
14 VEGAPULS 42, 44 and 45 – VBUS
Mounting and installation
Emission cone and interfering reflec­tions
The radar signals are focused by the an­tenna system. The signals leave the antenna in a conical path similar to the beam pattern of a spotlight. The form and intensity of the emission cone depend on the antenna used.
Any object in this beam cone causes a reflec­tion of the radar signals. Within the first few meters of the beam cone, tubes, struts or other installations can interfere with the meas­urement. At a distance of 6 m, the false echo of a strut has an amplitude nine times greater than at a distance of 18 m.
At greater distances, the energy of the radar signal distributes itself over a larger area, thus causing weaker echoes from obstruct­ing surfaces. The interfering signals are therefore less critical than those at close range.
If possible, orient the sensor axis perpen­dicularly to the product surface and avoid vessel installations within the 100 % area of the emission cone, e.g. tubes or struts.
The illustrations of the emission cones are simplified and represent only the main beam
- a number of weaker beams also exist. Un­der difficult measuring conditions, the align­ment of the antenna must be such that the lowest possible false echo values appear. Only giving attention to the size of the useful echo is not always adequate when measur­ing conditions are unfavourable.
Optimum measuring conditions exist when the emission cone reaches the measured product perpendicularly and when the emis­sion cone is free of obstructions.
0 m
Meas. distance
10 m
Emission cone of a VEGAPULS 42 with screw-on antenna and with ø 40 mm horn
Meas. distance
3,50 1,90 3,501,900
0 m
VEGAPULS 42
22˚
30˚
50%
18˚
25˚
25%
VEGAPULS 44 with ø 48 mm horn
m
In a difficult measurement environment, searching for a mounting location with the lowest possible false echo intensity will bring the best results. In most cases, the useful echo will then be present with sufficient strength. With the adjustment software VVO
25%
on the PC, you can have a look at the echo image and optimise the mounting location (see chapter "5.2 Adjustment with the PC – Sensor optimisation – Echo curve").
If possible, provide a "clear view" to the prod-
15 m
4,0 2,3 4,02,30
Emission cone of a DN 50 flange antenna
50%
m
uct inside the emission cone and avoid ves­sel installations in the first third of the emission cone.
VEGAPULS 42, 44 and 45 – VBUS 15
Mounting and installation
0 m
Meas. distance
20 m
3,0 1,7 3,01,70m
VEGAPULS 44 with ø 75 mm horn
10˚
20˚
50%
Emission cone of a DN 80 flange antenna
0 m
VEGAPULS 44 with ø 95 mm horn
25%

3.2 Measurement of liquids

Flange antennas
In most cases, the mounting of radar sensors
is done on short DIN socket pieces. The
lower side of the instrument flange is the
reference plane for the measuring range. The
antenna must always protrude out of the
flange tube.
When the DIN socket piece is longer, please
make sure that the horn antenna is not cov-
ered completely by the socket. It is better if
the antenna protrudes slightly out of the
socket.
Reference plane
< 135 mm (DN 50) < 210 mm (DN 80) < 310 mm (DN 100, DN 150)
Mounting on DIN socket piece
When mounting on dished vessel tops, the
antenna length should at least correspond to
the length of the longer sockets.
Vessel center or symmetry axis
Meas. distance
14˚
25%
<135…310mm (250…425mm with antenna extension)
Mounting on a dished vessel top; max. socket length
20 m
2,5 1,3 2,51,30
50%
m
depending on flange size and, if applicable, on the length of the antenna extension (see "7.3 Dimen­sions").
Emission cone of a DN 100 and DN 150 flange an­tenna
16 VEGAPULS 42, 44 and 45 – VBUS
Mounting and installation
On dished tank ends, please do not mount the instrument in the centre or close to the vessel wall, but approx.
1
/2 vessel radius from
the centre or from the vessel wall.
Dished tank ends can act as paraboloidal reflectors. If the radar sensor is placed in the focal point of the parabolic tank, the radar sensor receives amplified false echoes. The radar sensor should be mounted outside the focal point. Parabolically amplified echoes can be thereby avoided.
Vessel center or symmetry axis
Reference plane
½ vessel radius
Mounting on dished tank ends
If the stability of the vessel will allow it (sensor weight), flat mounting directly on the vessel top is a good and economical solution. The top side of the vessel is the reference plane.
Screw-on antenna
The screw-on antenna is mainly used on small vessels. The antenna fits on small ves­sel openings down to 1 must not be longer than 135 mm (when using the longer antenna, not longer than 250 mm).
Screw-on antenna on socket piece 11/2"
Screw-on antenna with antenna extension on socket piece 1½"
1
/2" socket. The socket
Reference plane
135 mm
250 mm
As an alternative to socket mounting, the screw-on antenna can be screwed into the
Reference plane
Mounting directly on the flat vessel top
VEGAPULS 42, 44 and 45 – VBUS 17
hole in the vessel.
Rod antenna directly on vessel opening
Mounting and installation
3.3 Measurement in standpipe
(surge or bypass tube)
General instructions
Pipe antennas are preferred in vessels which contain many installations, e.g. heating tubes, heat exchangers or fast-running stirrers. Measurement is then possible where the product surface is very turbulent, and vessel installations cannot cause false echoes.
Through the focusing of the radar signals within the measuring tube, even products with low dielectric constant (ε be measured reliably in the surge or bypass tube. Note the following instructions.
Surge pipe welded to the tank
Type label
= 1.6 to 3) can
r
Surge pipe in the socket piece
maxmax
Make sure the required upper vent hole in
the surge pipe is aligned with the sensor
type label.
As an alternative to a surge pipe in the ves-
sel, a pipe antenna system outside the ves-
sel in a bypass tube is also possible.
The surge and bypass tubes must generally
be made of metal. For plastic tubes, a
closed, conductive jacket is always required.
For metal tubes with plastic inner coating,
make sure that the thickness of the coating is
minimal (approx. 2 … 4 mm).
Align the sensor such that the type label lies
on one axis with the tube holes or the tube
connection openings. The polarisation of the
radar signal enables a considerably more
stable measurement with this alignment.
Type label
> 300 mm
100 %
Vent hole ø 5 … 10 mm
0 %
Tube flange system as bypass tube
When mounting a VEGAPULS 42 or 44 sen-
sor on a bypass tube (e.g. on a previous
min
without deflector
Pipe antenna systems in the tank
with deflector
min
floating or displacer unit), the radar sensor
should be mounted at a distance of approx.
300 mm or more from the max. level.
The surge or bypass tubes open at the bot­tom should reach to the requested min. level, since a reliable measurement is only possible within the pipe. The tube inner diameter should be max. 100 mm and correspond with the size of the antenna horn.
18 VEGAPULS 42, 44 and 45 – VBUS
Mounting and installation
> 300 mm
100 %
75 %
0 %
Extended bypass tube on a vessel with intense product movements
For products with small dielectric constants (< 4), a bypass tube longer than that re­quired by the lower tube connection should be used. Products with small dielectric con­stants are partly penetrated by the radar signals, so that the tube bottom delivers a stronger echo than the product (when the bypass tube is nearly empty). As a result of the extension of the lower tube end, sufficient liquid will remain even when the vessel is emptied.
With a liquid quantity of 300 … 800 mm in the blind lower end of the tube, the portion of the signal that penetrates the liquid and reflects from the tube bottom is sufficiently damped ­the sensor can then easily distinguish it from the echo of the liquid surface. If not enough liquid remains, a deflection plate located at the bottom of a vertical pipe can provide the same function. It deflects the signal reflected from the tube end sideways into the standard tube opening.
Connections to the bypass tube
The connections to the bypass tubes must be fashioned in such a way that only minimal reflections are caused by the walls of the connecting tubes. This is especially important for the breather connection in the upper part of the tube. Observe the following points:
• Use small openings for the connection.
• The diameter of the connecting tubes should not exceed 1/3 of the bypass diam­eter.
• The tube connections must not protrude into the bypass.
• Large welding beads in the tubes should be avoided.
• Additional connections to the bypass tube are more suitable if they lie on the same plane as the upper and lower vessel con­nection (superimposed or displaced by 180°).
Type label
> 300 mm
100 %
Optimum connection to the bypass tube
0 %
Bypass tube with tube stub
VEGAPULS 42, 44 and 45 – VBUS 19
300 ... 800 mm
Welding beads too large
Tube connection protrudes
Mounting and installation
Conducting tube
Conducting tube in existing surge or bypass tube
To increase the min. distance, the conducting tube can protrude out of the surge or bypass tube. For this purpose, a plain flange can be welded at the required position on the out­side of the extended conducting tube. In both cases, an adequate vent hole must be provided.
Additional connection to the bypass tube in one plane
Use of conducting tubes
(VEGAPULS 44)
In case of very rough inner surfaces in exist­ing bypass tubes (e.g. due to corrosion) or in case of big connection openings as well as bypass tubes with an inner diameter of more than 100 mm, the use of a conducting tube in the existing bypass tube is recommended. This reduces the noise level and increases reliability considerably. The flange of the conducting tube can be easily mounted as a sandwich flange between vessel and sensor flange.
20 VEGAPULS 42, 44 and 45 – VBUS
Extended conducting tube
Mounting and installation
Seals on tube connections and tube ex­tensions
Microwaves are very sensitive to gaps in flange connections. If connections are made without proper care, distinct false echoes as well as increased signal noise can result. Observe the following points:
• The seal used should correspond to the tube inner diameter.
• If possible, conductive seals such as con­ductive PTFE or graphite should be used, especially for thicker seals.
• Make sure that there are only a few seal positions on a conducting tube.
Flange connections on bypass tubes
Adhesive products
For slightly adhesive products, choose a surge pipe with e.g. a nominal width of 50 mm. VEGAPULS 42 and 44 radar sensors with 26 GHz technology are relatively insensi­tive to buildup in the tube. Nevertheless, buildup must not be allowed to plug up the tube completely.
For adhesive products, the use of a DN 80 to max. DN 100 stand/surge pipe can enable measurement in spite of buildup. Products that cause excessive buildup cannot be measured in a standpipe.
VEGAPULS 45 with integrated measur­ing tube
The sensor version VEGAPULS 45 is espe­cially developed for measuring tube applica­tions and is supplied complete with a 27 mm measuring tube. With a measuring tube length up to 4 m, this sensor version is suit­able e.g. for the use in existing standpipes.
max
min
VEGAPULS 45 with integrated measuring tube in a free vessel or in an existing bypass pipe
Note:
VEGAPULS 45 sensors with integrated measuring tube are equipped with a threaded process fitting or a flange connec­tion. For sensors with threaded process fitting, make sure that the complete sensor, i.e. also the sensor housing, turns when screwing in. The sensor housing is fastened with a clipped connection. This sensor hous­ing must not be turned with respect to the measuring tube or even loosened. Otherwise the radar signal coupling would be destroyed and the sensor would not function proberly.
VEGAPULS 42, 44 and 45 – VBUS 21
Mounting and installation
If you want to measure inhomogeneous prod­ucts or stratified products in a surge pipe, it must have holes, elongated holes or slots. These openings ensure that the liquid is mixed and corresponds to the liquid in the vessel.
The more inhomogeneous the measured product, the closer the openings should be spaced.
Due to radar signal polarisation, the holes or slots must be positioned in two rows offset by 180°. The radar sensor must then be mounted so that the type label of the sensor is aligned with the rows of holes.
VEGAPULS 45 with integrated measuring tube (surge pipe) measures between the heating spirals
Standpipe measurement of inhomoge­neous products
ø 5...15
homogeneous liquids
slightly inhomogeneous liquids
ø 5...15
Every wider slot causes a false echo. The slots should therefore not exceed a width of 10 mm, to keep the signal-to-noise ratio at a minimum. Round slot ends are better than rectangular ones.
Type label
ø 5...15
VEGAPULS 44: Rows of holes on one axis with the type label
inhomogeneous liquids
Openings in a surge pipe for mixing of inhomogene­ous products
22 VEGAPULS 42, 44 and 45 – VBUS
Mounting and installation
Surge pipe with ball valve
If a ball valve is mounted in the surge pipe, maintenance and servicing can be carried out without opening the vessel (e.g. if it con­tains liquid gas or toxic products).
Ball valve
ø50
Tube antenna system with ball valve cutoff in measur­ing tube
A prerequisite for trouble-free operation is a ball valve throat that corresponds to the pipe diameter and provides a flush surface with the pipe inner wall. The valve must not have any rough edges or constrictions in its chan­nel and should have a min. distance of 300 mm from the sensor flange.
> 300 mm
Vent hole
Deflector
Guidelines for standpipe construction
Radar sensors for measurement on surge or bypass pipes are used with G 1½ A screwed-on antenna or flange sizes DN 50, DN 80, DN 100 and DN 150. The radar sensor with a DN 50 flange is only a functional measuring system in conjunction with a measuring tube.
On the left you see the constructional fea­tures of a measuring pipe (surge or bypass tube) as exemplified by a sensor with DN 50 flange.
The measuring pipe must be smooth inside (average surface quality Rz ≤ 30). Use stain- less steel tubing (drawn or welded length­wise) for construction of the measuring pipe. Extend the measuring pipe to the required length with welding neck flanges or with con­necting sleeves. Make sure that no shoulders or projections are created during welding. Before welding, join pipe and flange with their inner surfaces flush and exactly fitting.
Avoid welding through the pipe wall. The pipe must remain smooth inside. Roughness or welding beads on the inner surfaces must be carefully removed and burnished, as they cause false echoes and encourage product adhesion.
On the following page you will see the constructional features of a measuring pipe as exemplified by a sensor with DN 100 flange.
If the vessel contains agitated products, fasten the measuring pipe to the vessel bot­tom. Provide additional fastenings for longer measuring pipes.
VEGAPULS 42, 44 and 45 – VBUS 23
Mounting and installation
0 %
Flange DN 100
Deburr the holes
150…500
Connecting sleeve
Welding neck flanges
VEGAPULS 44
When measuring products with lower dielec­tric values (< 4), a part of the radar signal penetrates the medium. If the vessel is nearly empty, an echo is generated by the medium and the vessel bottom. In some cases, the vessel bottom generates a stronger signal echo than the product surface. With a deflec­tor on the measuring pipe end, the radar signals are scattered. In nearly empty ves­sels and products with low dielectric value, the medium then generates a stronger echo than the vessel bottom.
Thanks to the deflector, only the useful signal is received in a nearly empty vessel - the
Welding of the plain welded flange
100 %
ø 95
2
Welding of the connect­ing sleeves
5…10
correct measured value is thus transmitted and the 0 % level reliably detected.
Instead of a deflector, the standpipe or surge pipe can be equipped with a quadrant pipe at the end. This reflects the radar signals that penetrate the medium diffusely to the side and reduces strong echoes from the tube
0,0…0,4
3,6
Welding of the welding neck flange
3,6
1,5…2
0,0…0,4
end or vessel bottom.
0 %
Quadrant pipe on the bypass tube end
Deflector
0 %
ø 100,8
~45˚
Meas. pipe fastening
Vessel bottom
Quadrant pipe on the standpipe end
24 VEGAPULS 42, 44 and 45 – VBUS
Mounting and installation

3.4 False echoes

The mounting location of the radar sensor must be selected such that no installations or inflowing material cross the radar impulses. The following examples and instructions show the most frequent measuring problems and how to avoid them.
Vessel protrusions
Vessel forms with flat protrusions can, due to their strong false echoes, greatly effect the measurement. Shields above these flat pro­trusions scatter the false echoes and guaran­tee a reliable measurement.
Correct Wrong
Vessel protrusions (slope)
Intake pipes, i.e. for the mixing of materials ­with a flat surface directed towards the sen­sor - should be covered with a sloping shield that will scatter false echoes.
Vessel installations
Vessel installations such as, for example, a ladder, often cause false echoes. Make sure when planning your measuring location that the radar signals have free access to the measured product.
Correct Wrong
Ladder
Vessel installations
Ladder
Struts
Struts, like other vessel installations, can cause strong false echoes superimposed on the useful echoes. Small shields effectively hinder a direct false echo reflection. These false echoes are scattered and diffused in the area and are then filtered out as "echo noise" by the measuring electronics.
Correct Wrong
Correct Wrong
Struts
Vessel protrusions (intake pipe)
VEGAPULS 42, 44 and 45 – VBUS 25
Shields
Mounting and installation
Inflowing material
Do not mount the instrument in or above the filling stream. Ensure that you detect the product surface and not the inflowing mate­rial.
Correct
Inflowing material
Wrong
Buildup
If the sensor is mounted too close to the vessel wall, buildup and adhesions of the measured product to the vessel wall will cause false echoes. Position the sensor at a sufficient distance from the vessel wall. Please also note chapter "3.1 General instal­lation instructions".
Correct Wrong
Strong product movements
Strong turbulences in the vessel, e.g. caused by stirrers or strong chemical reactions, can seriously interfere with the measurement. A surge or bypass tube (figure) allows, pro­vided the product causes no buildup in the tube, a reliable measurement even with turbulences in the vessel.
Correct Wrong
100 %
75 %
0 %
Strong product movements
Buildup
26 VEGAPULS 42, 44 and 45 – VBUS
Mounting and installation

3.5 Installation mistakes

Socket piece too long
If the sensor is mounted in a socket exten­sion that is too long, false reflections are caused and measurement is hindered. Make sure that the horn antenna protrudes out of the socket piece.
Correct Unfavourable
Reference plane
Flange antenna: Correct and unfavourable socket length
Unfavour­able
Correct
Parabolic effects on dished or arched vessel tops
Round or parabolic tank tops act like a para­bolic mirror on the radar signals. If the radar sensor is placed at the focal point of such a parabolic tank top, the sensor receives am­plified false echoes. The optimum mounting position is generally in the range of half the vessel radius from the centre.
Correct
~ ½ vessel radius
Unfavourable
Unfavourable
< 135 mm (250 mm)
Screw-on antenna: Correct and unfavourable socket length
VEGAPULS 42, 44 and 45 – VBUS 27
Mounting on a vessel with parabolic tank top
Mounting and installation
Wrong orientation to the product
Weak measuring signals are caused if the sensor is not directly pointed at the product surface. Orient the sensor axis perpendicu­larly to the product surface to achieve reliable measuring results.
Correct Wrong
Ladder
Direct sensor vertically to the product surface
Ladder
Sensor too close to the vessel wall
If the radar sensor is mounted too close to the vessel wall, strong false echoes can be caused. Buildup, rivets, screws or weld joints superimpose their echoes onto the product or useful echo. Please ensure sufficient dis­tance from the sensor to the vessel wall.
Foam generation
Conductive foam is penetrated by the radar signals to different depths and generates a number of single (bubble) echoes. The sig­nals in the foam are also damped, like heat radiation that tries to penetrate styrofoam. Thick, dense, creamy foam, and especially conductive foam, on the product surface can cause incorrect measurements.
conductive foam
Liquid
In case of good reflection conditions (liquids
Foam generation
without vessel installations), we recommend selecting the sensor distance so that there is no vessel wall within the inner emission cone. For products in less favourable reflection environments, it is a good idea to also keep the outer emission cone free of interfering installations. Note chapter "3.1 General instal­lation instructions".
Take measures to avoid foam, measure in a bypass tube or use a different measuring technology, e.g. capacitive electrodes or hydrostatic pressure transmitters.
In many cases, VEGAPULS 54 radar sensors with 5.8 GHz operating frequency achieve considerably better and more reliable meas­uring results in foam applications than type 40 sensors with 26 GHz technology.
28 VEGAPULS 42, 44 and 45 – VBUS
Loading...
+ 64 hidden pages