Please read this manual carefully, and also take
note of country-specific installation standards
(e.g. the VDE regulations in Germany) as well
as all prevailing safety regulations and accident prevention rules.
For safety and warranty reasons, any internal
work on the instruments, apart from that involved in normal installation and electrical connection, must be carried out only by qualified
VEGA personnel.
2Pressure transmitter D94 (Profibus PA)
Note Ex area
Please note the attached safety instructions
containing important information on installation
and operation in Ex areas.
These safety instructions are part of the
operating instructions and come with the Ex
approved instruments.
Page 3
Contents
Contents
Safety information ........................................................................ 2
Note Ex area ................................................................................ 2
1 Product description
1.1 Function and configuration .................................................. 4
1.2 Types and versions ............................................................. 5
1.3 Technical data ....................................................................... 6
1.4 Approvals and certificates .................................................. 9
in the processing unit ........................................................ 32
6.2 Exchange of hygienic form seal on D94 .......................... 33
Pressure transmitter D94 (Profibus PA)3
Page 4
1 Product description
Product description
1.1 Function and configuration
D94 pressure transmitters are efficient instruments for hydrostatic level measurement. A
dry ceramic-capacitive CERTEC® measuring
cell is used as the pressure sensor element.
The hydrostatic pressure effects a capacitance change on the ceramic diaphragm.
This capacitance change is detected by an
ASIC (Application Specific Integrated Circuit)
and converted into a pressure-proportional
signal by the integrated electronic module
with microcontroller. Precise, high-resolution
digital processing of measurement data
ensures excellent technical data.
The electronics module is powered by a
Profibus DP - PA segment coupler. The
measured pressure is transmitted via the
bus cable as digital signal to the processing
system and can be displayed on the indicating module.
To improve reliability, the functionality of important electronic components is continuously checked, and internal parameters such
as sensor value, temperature and operating
voltage are closely monitored.
The pressure transmitters with ceramic
CERTEC® measuring cell offer the advantage
of continuous self-monitoring: the measuring
and reference capacitances of the measuring
cell constitute a predefined ratio over the
entire measuring range. Any deviation from
these data is a reliable indicator of a malfunction of the measuring cell. If errors or malfunctions are detected during these routines,
fault signals are triggered according to the
Profibus PA specification.
The pressure transmitter D94 consists of two
parts, the transmitter and the processing
unit. Both parts are connected via a threewire screened cable (depending on the version, with or without breather capillaries). The
processing unit contains the adjustment
elements (address switch, keys and LCD)
and the terminals for the Profibus PA cable.
Transmitter
Processing unit
0 - 20 bar
Profibus PA
cable
Profibus PA version
The instrument profile of the pressure transmitter D94 is according to the profile specification version 3.0. The instrument master file
(GSD) is supplied with the pressure transmitter D94. In addition, the GSD is available as a
download file from the VEGA homepage
(www.vega.com).
Adjustment and indication
The following adjustment versions are available:
- menu-driven adjustment (4 keys and DOTMatrix) in the processing unit
- via PC with the adjustment software VEGA
Visual Operating (VVO)
- optional, processing unit equipped with an
indicating module.
0…0.1 bar / 0…10 kPa15 bar / 1 500 kPa-0.2 bar / -20 kPa
0…0.2 bar / 0…20 kPa20 bar / 2 000 kPa-0.4 bar / -40 kPa
0…0.4 bar / 0…40 kPa30 bar / 3 000 kPa-0.8 bar / -80 kPa
0…1.0 bar / 0…100 kPa35 bar / 3 500 kPa-1.0 bar / -100 kPa
0…2.5 bar / 0…250 kPa50 bar / 5 000 kPa-1.0 bar / -100 kPa
0…5.0 bar / 0…500 kPa65 bar / 6 500 kPa-1.0 bar / -100 kPa
0…10.0 bar / 0…1 000 kPa90 bar / 9 000 kPa-1.0 bar / -100 kPa
0…20.0 bar / 0…2 000 kPa130 bar / 13 000 kPa-1.0 bar / -100 kPa
0…40.0 bar / 0…4 000 kPa200 bar / 20 000 kPa-1.0 bar / -100 kPa
0…60.0 bar / 0…6 000 kPa300 bar / 30 000 kPa-1.0 bar / -100 kPa
-0.05…+0.05 bar / -5…+5 kPa15 bar / 1 500 kPa-0.2 bar / -20 kPa
-0.1…+0.1 bar / -10…+10 kPa20 bar / 2 000 kPa-0.4 bar / -40 kPa
-0.2…+0.2 bar / -20…+20 kPa30 bar / 3 000 kPa-0.8 bar / -80 kPa
-0.5…+0.5 bar / -50…+50 kPa35 bar / 3 500 kPa-1.0 bar / -100 kPa
-1.0…0.0 bar / -100…0 kPa35 bar / 3 500 kPa-1.0 bar / -100 kPa
-1.0…+1.5 bar / -100…+150 kPa50 bar /5 000 kPa-1.0 bar / -100 kPa
-1.0…+4.0 bar / -100…+400 kPa65 bar / 6 500 kPa-1.0 bar / -100 kPa
-1.0…+10.0 bar / -100…+1 000 kPa90 bar / 9 000 kPa-1.0 bar / -100 kPa
-1.0…+20.0 bar / -100…+2 000 kPa130 bar / 13 000 kPa-1.0 bar / -100 kPa
-1.0 …+40.0 bar / -100…+4 000 kPa200 bar / 20 000 kPa-1.0 bar / -100 kPa
-1.0…+60.0 bar / -100…+6 000 kPa300 bar / 30 000 kPa-1.0 bar / -100 kPa
Absolute pressure
0…1.0 bar / 0…100 kPa35 bar / 3 500 kPa
0…2.5 bar / 0…250 kPa50 bar / 5 000 kPa
0…5.0 bar / 0 …500 kPa65 bar / 6 500 kPa
0…10.0 bar / 0…1 000 kPa90 bar / 9 000 kPa
0…20.0 bar / 0…2 000 kPa130 bar / 13 000 kPa
0…40.0 bar / 0…4 000 kPa200 bar / 20 000 kPa
0…60.0 bar / 0…6 000 kPa300 bar / 30 000 kPa
Cable outlet
axial
Cable outlet lateralPlug connection
lateral
Pressure transmitter D94 (Profibus PA)5
Page 6
Product description
1.3 Technical data
Mechanical data
Transmitter materials, wetted parts
Process connectionstainless steel 1.4435 or 1.4571
Diaphragmsapphire-ceramic® (99.9 % Oxide-ceramic)
Seal, measuring cellViton, EPDM, Hifluor, Kalrez
Transmitter materials, non-wetted parts
Transmitter housing- Alu die casting, powder-coated
Materials of the processing unit
Housinghigh resistance plastic PBT (Polyester)
Ground terminalstainless steel 1.4305
Window of the indication moduleLexan
Adjustment and indicating elements (in the processing unit)
Menu-driven adjustment with
additional functions
- adjustment elements4 keys
- indicating elementsDOT-matrix display, 3 lines with 7 digits each
Version without adjustment module
- adjustment element8-pole mini coding switch for adjustment of the
Indication moduleLC-display with
- stainless steel 1.4571
bus address
- bar graph (20 segments)
- digital value (4 digits)
- tendency indicators for rising or falling values
Weight
Transmitterapprox. 1.4 kg
Processing unitapprox. 480 g
6Pressure transmitter D94 (Profibus PA)
Page 7
Product description
Electrical data
Adjustment ranges
Zeroadjustable from -20 … +95 % of nominal range
Spanadjustable from 3.3 … 120 % of nominal range
Supply and signal circuit
Supply voltage9 … 32 V DC
Current consumption10 mA ± 1 mA
Integration time0 … 50 s
Rise time70 ms (ti = 0 sec; 0 – 63 %)
Connection cable, processing unit – Profibus PA
Cable entryM20 x 1.5 (for cable ø 5 … 9 mm)
Screw terminalsfor conductor cross-sections of up to 2.5 mm
Recommended bus cable types- SINEC 6XV 830-5AH10 (Siemens AG)
Connection cable, processing unit – Transmitter
Cable entryM20 x 1.5 (for cable ø 5 … 9 mm)
Screw terminalsfor conductor cross-sections of up to 2.5 mm
Transmitter
- with fixed cable outlet3-wire screened with breather capillaries
The use of a properly fitting cable seal in the cable connection is necessary for maintaining the housing
protection. If the seal used does not fit the cable, it should be replaced by a suitable one.
Pressure transmitter D94 (Profibus PA)7
Page 8
Product description
Accuracy (similar to DIN 16 086, DIN V 19 259 - 1 and IEC 770)
Deviation
Reference conditions (acc. to IEC 770)
- temperature15°C … 35°C
- rel. humidity45 % … 75 %
- air pressure860 mbar … 1060 mbar/86 kPa … 106 kPa
Determination of characteristicslimit point adjustment acc. to DIN 16 086
Characteristicslinear
Deviation in characteristicsincl. hysteresis and repeatability
- Turn Down 1 : 1< 0.25 % with accuracy class 0.25
- Turn Down up to 1 : 5typ. < 0.3 % with accuracy class 0.25
- Turn Down up to 1 : 10typ. < 0.4 % with accuracy class 0.25
Influence of the ambient temperature
Average temperature coefficient
of the zero signal
1)
- Turn Down 1 : 1< 0.15 %/10 K with accuracy class 0.25
- Turn Down up to 1 : 5typ. < 0.225 %/10 K with accuracy class 0.25
- Turn Down up to 1 : 10typ. < 0.3 %/10 K with accuracy class 0.25
< 0.1 % with accuracy class 0.1
typ. < 0.1 % with accuracy class 0.1
typ. < 0.2 % with accuracy class 0.1
< 0.05 %/10 K with accuracy class 0.1
typ. < 0.075 %/10 K with accuracy class 0.1
typ. < 0.1 %/10 K with accuracy class 0.1
Long-term stability
Long-term drift of the zero signal
2)
< 0.1 % per 2 years
Other actuating variables
Vibration resistancemechanical vibrations with 4 g and
5 … 100 Hz, tested acc. to the regulations of
German Lloyd GL-characteristics 2
1)
In compensated temperature range of 0°C … +80°C, reference temperature 20°C.
2)
Acc. to IEC 770, item 6.1.2 relating to the nominal range.
8Pressure transmitter D94 (Profibus PA)
Page 9
Product description
Operating conditions
Ambient conditions
Ambient temperature-40°C … +85°C
- with indicating module-20°C … +70°C
Storage and transport temperature-40°C … +85°C
Product temperature, depending on
seal material of the measuring cell
- Viton-20°C … +100°C (1 h up to +150°C)
- EPDM-40°C … +100°C (1 h up to +150°C)
- Hifluor-10°C … +100°C (1 h up to +150°C)
- Kalrez0°C … +100°C (1 h up to +150°C)
1.4 Approvals and certificates
Approvals
- Ex Zone 2
- CENELEC EEx ia IIC
- ATEX II 1G EEx ia IIC
If the use of approved instruments is required for certain applications, the appropriate
official documents (test reports, test certificates and conformity certificates) must be observed. These are supplied with the respective instrument.
CE conformity
The pressure transmitter D94 meets the requirements of EMC (89/336/EWG) and NSR
(73/23/EWG). Conformity has been judged acc. to the following standards:
EMCEmissionEN 50 081 - 1: 1992
NSREN 61 010 - 1: 1993
SusceptibilityEN 50 082 - 2: 1995
NAMUR regulations
Full compliance with the NAMUR regulations NE 21, May 1993.
Pressure transmitter D94 (Profibus PA)9
Page 10
1.5 Dimensions
Transmitter housing
ø 76
67
Product description
ø 41,6
93
46
ø 41,6
ø 41,6
46
Processing unit
with display module
3882
∅
M20x1,5
with protective cover
63
5
135
118
108
72100
85
90
139
130
10Pressure transmitter D94 (Profibus PA)
Page 11
Product description
Process connections
Ø 41,6
Ø 41,6
Ø 41,6
76,5 (115*)
121 (160*)
121 (160*)
SW 46
SW 46
121 (160*)
SW 60
24,5
Ø 41,6
Ø 78
LA
Ø 41,6
Ø 78
RA
Ø 41,6
G 1½ A
Ø 55
GG
72,5 (111*)
121 (160*)
121 (160*)
SW 46
SW 46
121 (160*)
Ø 41,6
Ø 90
LB
Ø 41,6
Ø 92
RB
SW 60
25
1½ " NPT
GN
Ø 41,6
SW 46
121 (160*)
SW 46
121 (160*)
Ø 41,6
Ø 64
CA
Ø 41,6
Ø 65,8
Ø 105
AA
121 (160*)
SW 46
21
M44 x 1,25
BA
SW 46
121 (160*)
SW 46
121 (160*)
Ø 41,6
Ø 84
TA
Ø 41,6
Ø 100
PA
120
42,5
160
Ø 48
Ø 115
23
FT4
Ø 38
Ø 115
*) VBUS version
FG4
Pressure transmitter D94 (Profibus PA)11
Page 12
2 Mounting
Mounting
2.1 Mounting instructions
The pressure transmitter can be mounted in
any position. Cable entries must point downward to avoid moisture ingress. For this purpose, the housing can be rotated by 330° in
relation to the mounting part.
An appropriate connection seal must be
used for mounting. This seal is either supplied with the pressure transmitter or must be
provided by the customer.
2.2 Compensation of the atmos-
pheric pressure
On instruments for gauge pressure measurement, the atmospheric pressure is compensated via a breather facility integrated in
the housing.
Pressure transmitter with housing
- via an integrated breather facility (PTFE
use 1)), protection IP 66
- via the capillaries of a special cable connected to the terminals, protection IP 67
Pressure transmitter with direct cable
outlet
- via the capillaries of a special cable connected to the terminals, protection IP 68
We recommend leading a special cable to
the processing unit and carrying out the
pressure compensation via the integrated
breather facility there.
Processing unit
Observe the following instructions:
- as a rule, there must be the same atmospheric pressure on the breather facility as
on the vessel
- the pressure compensation must be made
in dry environment
- with vertical wall mounting, cable entries
must point downward to avoid moisture
ingress and buildup on the breather facility.
1)
air permeable and humidity blocking
12Pressure transmitter D94 (Profibus PA)
Page 13
Electrical connection
3 Electrical connection
3.1 Connection instructions
The electronics in pressure transmitter D94
requires a supply voltage of 9 … 32 V DC.
The supply voltage and the digital output
signal are led via the same two-wire cable to
the terminals. The external energy is provided via a segment coupler.
Note the following instructions for electrical
connection:
- Connection must be made according to
country-specific installation standards (e.g.
in Germany acc. to the VDE regulations).
- The electrical connection must have a protective measure against polarity reversal.
- The connection of field instruments is generally made via T-adapters on stubs..
- Screened cable is recommended for connection (cable types, see chapter "1.3
Technical data"). The shielding must be
grounded on both ends (on the T-adapter
and on the connection housing of D94).
- For use in Ex areas, the installation regulations must be observed.
Safety information for Ex applications
As a rule, do all connecting work in the complete absence of line voltage. Always switch
off the power supply before you carry out
connecting work on the sensors. Protect
yourself and the instruments.
Qualified personnel
Instruments used in Ex areas must be installed only by qualified personnel. The qualified personnel must observe and understand
the mounting regulations and the supplied
type approval certificates and conformity
certificates.
If an instrument is used in hazardous areas,
the respective regulations, conformity certificates and type approvals of the sensors,
separators or safety barriers must be noted
(e.g. DIN VDE 0165).
Sensors used in Ex areas must be connected only to intrinsically safe circuits. The
permissible electrical values are stated in the
conformity certificate or the type approval
certificate.
Intrinsically safe circuits with more than one
active instrument (instrument delivering electrical energy) are not allowed. Please note the
special installation regulations (DIN VDE
0165).
Certain pressure transmitters are provided
with a warning label informing on measures to
be taken to avoid the danger of electrostatic
discharge. Note the content of the warning
label.
Pressure transmitter D94 (Profibus PA)13
Page 14
3.2 Wiring plan
TT
ransmitter with dirransmitter with dir
T
ransmitter with dir
TT
ransmitter with dirransmitter with dir
cable outletcable outlet
cable outlet
cable outletcable outlet
Screen
TT
ransmitter with terransmitter with ter
T
ransmitter with ter
TT
ransmitter with terransmitter with ter
housinghousing
housing
housinghousing
Zum Anschluß an
TransmitterTransmitter
Breather capillaries
For connection to
E21 PA
4321543215
Klemmeinsatz
terminal
ectect
ect
ectect
Suspension cable
br
bl
ge
minalminal
minal
minalminal
Processing unitProcessing unit
Processing unit
Processing unitProcessing unit
Transmitter
E21PA
bl
br
ge
Electrical connection
Display moduleDisplay module
Display module
Display moduleDisplay module
OK
121110231
PA+PA
AM10
-
4
PA +
Profibus PA cable
PA -
14Pressure transmitter D94 (Profibus PA)
Page 15
Setup
4 Setup
4.1 Setting the Profibus P A address
Since each bus participant must have its own
address by which it is identified on the bus,
the address must be adjusted before setup
of the transmitter. Depending on the instrument version, this is done either with the
module "Menu-driven adjustment" or the
address switch.
Address adjustment with the address
switch
Address
12345678
on
off
The transmitter is delivered with the address
set as shown (all switches position "on").
If the mini coding switch is set to address
126 and higher (factory setting), the address
can be modified via software (with VVO).
However, this is only possible if the transmitter is connected as the only participant with
address 126 on the bus. For this reason, we
recommend adjusting the address of the
sensor via the hardware (switch) before
connecting to the bus.
Example:
To select address 37, set switch 1 (value 1),
switch 3 (value 4) and switch 6 (value 32) to
position "on". Set all other switches to position
"off".
Address
12345678
on
off
1 2 4 81632 64 128
1+4+32=37
S
H
S
H
value
Address adjustment with the module
"Menu-driven adjustment with additional functions"
The Profibus PA transmitter is delivered with
the address set to "126". To modify the address via the keyboard, the transmitter must
be connected to the power supply. This does
not have to be over the bus cable. You can
use a standard power supply unit or a battery (9 … 32 V DC, 10 mA).
OK
121110231
PA+PA
Transmitter
With the arrow key and the "OK" key, you
move to the menu item "Instrument address".
There you can adjust the address (0 … 125)
with the "+" and "–" key. Save the adjusted
address with the "OK" key.
Operate
0.123
bar
Ad-
just-
ment
E21PA
U
Calcu-
lation
AM10
-
–+
B
Output
Station
address
126
Simula-
tion
Addi-
tional
funct.
Escape
Pressure transmitter D94 (Profibus PA)15
Page 16
Further adjustment steps (adjust PA output,
adjustment etc.) can also be carried out if the
processing unit is connected to a power
supply unit. However, adjustment is also
possible at a later time (when the processing
unit is connected to the bus cable).
Adjustment can be simplified by temporarily
mounting the cover or the display module on
the processing unit laterally or displaced
downwards.
Processing unit
Display module
OK
121110231
PA+PA
Transmitter
AM10
-
E21PA
0 - 20 bar
Setup
Tighten screws lightly
16Pressure transmitter D94 (Profibus PA)
Page 17
Setup
4.2 Setup with module "Menu-driven adjustment with additional functions"
Adjustment elements
Indication of:
- measured value
- functions in menu
t Jump to the menu below
with the "OK" key
s Return to the menu above
with the "OK" key
Depending on parameter:
change value (increase) or
choose out of a list
+
OPERATE
3.00
mbar
Depending on the parameter: change value (decrease) or choose out of a
list
–>
OK
–
Adjustment with the multifunctional module is
menu-orientated and is made via the four
keys in conjunction with the text display. The
switchover from the measured value indication to the main menu is made with the "OK"
key. Use the "→" key to change within a menu
from one menu item to another.
Branching points
A branching point is shown by the t symbol
and enables a jump to the menu below with
the "OK" key. In this menu you can find parameters which belong together according to
function (possibly in other submenus).
Choose menu item or interrupt
adjustment
Depending on the menu item, save
the adjusted value or change the
menu level
Reset
A reset is indicated by the s symbol and
enables a jump up to the menu above with
the "OK" key.
An automatic jump back to measured value
indication is triggered 60 minutes after the
last key push.
Parameters have no s or t symbol. The
value of the parameters can be modified with
the "+" and "–" keys or chosen from a list.
When these keys are pushed once, the value
flashes first. When pushed again, the value is
modified. The modified value can be saved
with the "OK" key. Push the "→" key to cancel
input (without saving the modification).
Certain parameters can be displayed but not
modified.
Pressure transmitter D94 (Profibus PA)17
Page 18
Menu schematic
Operate
0,0
skal
Measured value
indication
Setup
Adjust-
ment
Calcu-
lation
Lin.-
curve
activ
with PA
output "Pressure"
Output
PA-
output
Lin%
pressure
is only displayed if a
linearisation curve was
activated with VVO and
the PA output is set to
Lin%
with PA output
"%" or "Lin%"
Adjust
with
pres.
%
ti
1
s
Scalint
out
0
%=
0
Unit
0,0
skal
mbar
psi
kPa
%
Lin%
inHg
Adjust
without
pres.
Zurück
100
%=
1000
Escape
Decimal
Escape
point
888,8
Unit
0,0
%
mbar
psi
kPa
m
ft
inch
inHg
Simula-
tion
Sim.
x,xx
bar
Escape
Escape
Zero
0,0
%
Offset
correc-
tion
Offset
corr.
OK?
Sensor
un-
pres.?
18Pressure transmitter D94 (Profibus PA)
Escape
Only on pressure transmitter
for relative pressure measurement
Span
100
%
Escape
Unit
0,0
bar
psi
kPa
inHg
on sensors with a measuring range <1 bar
the unit
mbar
Offset
correc-
tion
Offset
corr.
OK?
Sensor
un-
pres.?
is displayed instead of
Zero
0,0
bar
Span
1,000
bar
bar
Escape
Page 19
Setup
Addi-
tional
funct.
Station
address
126
Escape
Sensor
data
Fabr.
date
49.98
Lan-
guage
English
Reset
Escape
Franc.
Ital.
Español
Deutsch
P min
-0,3
bar
1)1)1)1)
1)
The min. and max. values can be set to the actual value
Reset
OK ?
Reset
Now !
OK ?
P max
15,0
bar
T emp.
30,7
°C
T min
23,5
°C
T max
36,2
°C
by pushing the "+" and "–" keys simultaneously.
Diagno-
se no.
- - -
Escape
White letters indicate the parameters
which can be modified with the "+"or "–
" key and saved with the "OK" key.
bar
Sensors or measured value information which cannot be modified in these
positions are in italics..
List
These options can be selected with
the "+" or "–" key and saved with the
"OK" key.
Adjust
with
pres.
Escape
Use the arrow key to move to
the right in the menu schematic.
In menu items with these symbols, you can move to the top
or bottom with the "OK" key.
Span
100,0
Lan-
guage
English
Franc.
Ital.
Español
Deutsch
Pressure transmitter D94 (Profibus PA)19
Page 20
Setup
Adjust P A output
Operate
0,0
skal
Adjust-
ment
The PA output is factory-set to "Lin%". With
this adjustment it is possible to carry out an
adjustment and a volume measurement ( to
measure the volume of a non-linear vessel,
the linearisation curve of the vessel also has
to be programmed into the D94 with the software VVO). The adjustment "%" enables a
level measurement (also with adjustment). If
only a pressure value should be outputted,
the PA output must be set to "Pressure". An
adjustment is then not possible. Move with
the arrow and "OK" key to the menu item "PA
output", choose there the requested option
with the "+" or "–" key and save with the "OK"
key.
Under "Scaling out." you can modify the
number values (min/max), the position of the
decimal point and the unit (with the "+" or "–"
key the values can be modified, with the "OK"
key the adjustments are saved).
%=
Calcu-
lation
0
0
100
%=
1000
Output
PA-
output
Lin%
pressure
%
Decimal
point
888,8
Scaling
out.
Unit
0,0
%
mbar
psi
kPa
m
ft
inch
inHg
Escape
Escape
Adjustment
(only possible, if "%" or "Lin%" has been chosen for the PA output)
Adjustment taking the current pressure into account (live adjustment)
Live adjustment comprises two steps:
1 Adjustment of zero
2 Adjustment of span
1 Adjustment of zero
(e.g. process pressure zero or vessel
empty)
Set the percentage value to 0.0 %
Zero
0,0
with the "+" or "–" key. Then push the
%
"OK" key.
2 Adjustment of span
(e.g. process pressure zero or vessel level
max.)
Set the percentage value to 100 %
Span
100
with the "+" or "–" key. Then push the
%
"OK" key.
Note:
- A modification of zero does not influence
the adjusted span, i.e. the final value of the
measuring range is shifted.
- Push the "+" and "-" key simultaneously for
standard adjustment of zero/span. The
value jumps directly to 0 %/100 %.
- When pushing the "+" or "–" keys individually, the output signal remains at the last
value; it takes on the adjusted value only
after saving with the "OK" key.
- It is also possible to set values for partial
fillings or partial pressures, e.g. 25 % and
75 %. The pressure transmitter then automatically calculates the values for 0 % or
100 % (only possible with a level difference
>3.3 %).
20Pressure transmitter D94 (Profibus PA)
Page 21
Setup
Adjustment without taking the current
pressure into account (dry adjustment)
Adjustment without pressure comprises four
steps:
1 Selection of the unit in which the adjustment
is to be carried out
2 Offset correction
3 Adjustment of zero
4 Adjustment of span
The offset correction (only in gauge pressure
mode) defines the reference positions for the
measurement. It can be carried out:
- before or after adjustment of zero and
span
- before or after installation of transmitters.
The transmitter must be unpressurised for
the offset correction!
Adjustment without pressure can be carried
out with the instrument mounted or dismounted (e.g. workroom). Any ambient pressure has no effect on the adjustment.
1 Selection of the unit
Choose the unit with the "+" or "–"
Adjust
without
key. Save the selected unit with
pres.
the "OK" key.
Unit
0,0
mbar
psi
kPa
inHg
2 Adjustment of zero
Adjust zero with the "+" or "–" key
Zero
0,0
and save with the "OK" key.
mbar
4 Adjustment of span
Span
Adjust span with the "+" or "–" key
100,0
and save with the "OK" key.
mbar
Note:
- A modification of zero does not influence
the span.
- By pushing the "+" and "–" keys, the output
remains on the last value - only after saving
with the "OK" key will the output take on the
adjusted value.
Evaluation
Linearisation curve
Is only displayed if a linearisation
Lin.-
curve
curve has been activated with VVO.
activ
Adjustment of the integration time
An integration time ti of 0 s … 50 s can be
the "OK" key.
Selection of the indicated unit
adjusted with the "+" or "–" key for
ti
1
damping of pressure shocks. The
s
adjusted value can be saved with
Unit
The measurement unit for the dis-
0,0
play of measured values as well as
skal
for the display module can be
psi
selected here (with the "+" or "–"
kPa
key and the "OK" key). If you select
%
m
"scal.", the scaled PA output value
ft
will be displayed (as adjusted
inch
under "Scaling outp.").
mbar
Pressure transmitter D94 (Profibus PA)21
Page 22
Setup
Simulation
Simula-
You can adjust with the "+" or "–"
tion
keys an individual pressure or %
value to check the outputs of the
pressure transmitter and the con-
Sim.
nected components. The adjusted
xx,x
mbar
value flashes during activated simulation. The simulation can be
stopped with the "OK" key.
Additional functions
Sensor data
Important sensor data can be displayed via
the DOT-matrix for information and diagnostic
purposes:
- manufacturing date
- pointer function (p
- pointer function (p
- actual temperature value (Temp)
- pointer function (T
- pointer function (T
- diagnostic number
Additional
funct.
min
max
min
max
)
)
)
)
Language
Lan-
guage
English
Franc.
Ital.
Español
Deutsch
Reset
Reset
Reset
Now?
Reset
Now !
OK ?
The pressure transmitter comes
adjusted to the language in which
it is ordered. The languages German, English, French, Italian or
Spanish can be chosen with the
"+" or "–" key and the "OK" key.
The reset function deletes the adjusted values and resets the parameters to the default values. The
adjustment data again correspond
to the nominal measuring range. The
following data are not reset:
- instrument address
- language
- diagnostics number
- offset correction
- pointer
Station
address
Fabr.
date
49.98
1)
The min. and max. values can be set to the actual
value by pushing the "+" and "–" keys simultaneously.
22Pressure transmitter D94 (Profibus PA)
xxx
Sensor
data
P min
P max
-0,3
150
mbar
mbar
1)
1)1)1)
Temp.
30,7
°C
T min
23,5
°C
T max
36,2
°C
Diagno-
se no.
- - -
Page 23
Setup
4.3 Setup with adjustment software
VEGA Visual Operating (VV O)
VVO connection to the b us cable
To operate the sensor with the adjustment
software VVO, you require a Profibus DP
interface card (from Messrs. Softing) in your
PC. This card must be used to provide the
connection between computer and Profibus
DP cable (a connection directly to the sensor
or to the Profibus PA cable is not possible).
The computer with VVO then has the status
"Master, class 2".
Segment
coupler
Profibus PA
cable
VVO
DP cable
Profibus DP
interface card
(Messrs. Softing)
Field distributor
SensorSensorSensorSensor
If the computer is connected to the Profibus
DP cable, you can start VVO. If the following
presentation is shown on the monitor, you
have to modify the communication settings in
VVO. To do this, click to Configuration, point
to Program and click to Communication.
First click in the window "Communication" to
Profibus DP for PA sensors. Then select the
Profibus card installed in your PC. It is also
possible to modify the settings for baud rate,
further bus parameters and master address.
Note:
Make sure that the master address you are
using is not assigned to another bus participant!
Click to OK, if you have finished the communication setting. Then you have to restart
VVO.
Pressure transmitter D94 (Profibus PA)23
Page 24
Setup
Sensor configuration
After starting VVO as previously described
or having made connection to the Profibus,
the following presentation will be displayed
on the monitor. Click to Configuration, then
to Measuring system.
If you have already connected several VEGA
sensors to the bus, you can select the sensor to be configured, in this case pressure
transmitter D94. The serial number and the
bus address of the sensors are displayed to
enable clear identification.
The menu language of the sensor can be
selected (only relevant with a sensor with
menu-driven adjustment) and the units for the
output of sensor values can be set. If address 126 is adjusted on the sensor, you can
assign a bus address to the sensor in the
field "Sensor address". Then click to Save.
Information about the sensor identification number:
You do not have to modify the adjustment
"VEGA specific No.". Only if you want to make
sure that the pressure transmitter can be
replaced on the Profibus by a sensor of
another manufacturer do you have to choose
the option "Profile specific No.".
Configuration measurement loop
Click to Configuration, point to Measurement loop, then click to Modify.
Now all instruments which are connected to
the bus are shown (identifiable by the bus
address and the serial number). Notice that
in this example the same serial number appears twice. This is a sensor with a measurement loop for pressure and a second measurement loop for temperature. Choose the
sensor (the measurement loop) with which
you want to create a measurement loop and
click to OK.
24Pressure transmitter D94 (Profibus PA)
Page 25
Setup
In the window "Modify meas. loop configuration" you can assign a name and a description to the measurement loop. State under
"Application" if you want to create a process
pressure or level measurement. With OK you
return again to the window "VEGA Visual
Operating". In the following explanations, we
assume that you have chosen the application
"Level measurement". The adjustment "Process pressure measurement" differs only
slightly from the following presentation.
Sensor parameter setting – Adjustment
Click to Instrument data, then to Parameter
adjustment.
From the window "Instrument data parameter
adjustment" you reach all submenus of the
sensor. To get familiar with the available options, try out the buttons one after the other.
You will automatically return to this window. To
carry out the adjustment, click the button
Adjustment.
Click in the window "Adjustment" to Min/Maxadjustment.
Choose the measurement loop in which you
want to make the parameter adjustment and
click to OK.
Pressure transmitter D94 (Profibus PA)25
Then in the window "Min/Max adjustment" you
have to choose the type of adjustment. If you
want to carry out the "Adjustment without
taking the actual product into account" (dry
adjustment), click to no, then to OK.
Page 26
Setup
Here you can choose in which unit the adjustment should be made (mbar or bar, psi, kPa,
inHg). Then you enter the pressure values
corresponding to 0 % and 100 %. Then click
to OK. The adjustment is now finished.
If you want to carry out the "Adjustment taking the actual product into account" (live
adjustment), click in this window to yes, then
to OK.
Click to Min-Adjustment or to Max-Adjust-ment to carry out the respective adjustment.
In the window "Min-Adjustment" you choose in
which unit the adjustment should be carried
out (bar, mbar, psi, kPa, inHg). Then click to
Save. The min. adjustment is finished.
In the window "Max-Adjustment" you choose
in which unit the adjustment should be carried out (bar, mbar, psi, kPa, inHg). Then
click to Save. The max. adjustment is finished.
Sensor parameter setting – Linearisation curves
Click in the window "Instrument data parameter adjustment" to Conditioning.
Note:
For min. adjustment and max. adjustment,
defined levels (e.g. 0 % and 100 % or e.g.
20 % and 80 %) have to be present in the
vessel.
26Pressure transmitter D94 (Profibus PA)
Page 27
Setup
Click in the window "Conditioning" to Lineari-sation.
In the window "Linearisation" you can choose
the linearisation curves of a cylindrical or a
spherical tank. If you choose a user-programmable curve, you can then click to Edit
and open the program "Tank Calculation".
With this program you can calculate the
curves of different tank shapes (see the
manual "VEGA Visual Operating"). After selection of the curve, click to OK.
Click in the window "Outputs" to Profibusoutput.
In the window "Profibus output" you can determine the options for the Profibus output
(acc. to Profibus PA instrument profile). Confirm your adjustments with OK.
Sensor parameter setting – Outputs
Click in the window "Instrument data parameter adjustment" to Outputs.
Pressure transmitter D94 (Profibus PA)27
Page 28
Setup
Sensor parameter setting – Integration
time
Click in the window "Instrument data parameter adjustment" to Conditioning.
Click in the window "Conditioning" to Integra-tion time. In the window "Integration time"
you can enter a time interval of max. 50 seconds. Then click to OK.
Sensor parameter setting – Offset correction
Click in the window "Instrument data parameter adjustment" to Additional functions,
then in the window "Additional functions" to
Offset correction.
Confirm the question with OK, in case the
conditions are met.
Click in the window "Offset correction" to
Correct.
28Pressure transmitter D94 (Profibus PA)
Page 29
Setup
Pointer functions
Click in the window "Instrument data parameter adjustment" to Meas. loop data.
In the window "Meas. loop data" all available,
current sensor values and the peak values
(pointer function) are displayed. With the
button reset you can reset all displayed peak
values to the actual value all at once.
Choose the requested measurement loop
and click to OK.
In the window "Simulation of outputs", you
click to Start to begin the simulation. With the
buttons "<–" and "–>" (or the thumb of the
scroll bar) you can adjust values between 10 % and 110 %. Click to Stop to terminate
the simulation.
Simulation
For checking outputs of the pressure transmitter and connected devices and components, a pressure can be simulated. Click to
Diagnostics, then to Simulation.
Note:
The simulation mode is not terminated automatically, but remains active until you switch it
off!
Pressure transmitter D94 (Profibus PA)29
Page 30
Display of measured values
The current values of the measurement loops
can be displayed from the main menu. Click
to Display and then to "Display of measured
value".
Choose the requested measurement loop
and click to OK.
Setup
In the window "Display of measured value"
you can choose the unit in which the measured value should be displayed. In addition,
you will see the current value. By clicking to
Quit you return to the window "VEGA Visual
Operating".
30Pressure transmitter D94 (Profibus PA)
Page 31
Diagnostics
5 Diagnostics
5.1 Maintenance
Pressure transmitters D94 are maintenancefree.
5.2 Failure rectification
Fault signals
The pressure transmitter D94 provides maximum reliability through its self-test and continuous self-monitoring. If, however, malfunctions occur, the pressure transmitter D94
diagnosis distinguishes between atypical
process conditions and faults in the pressure
transmitter D94.
Atypical process conditions
Exceeding or falling short of measuring
range limits (fault signal extinguishes when
the measured value is again in the measuring
range).
Failure in the D94 pressure transmitter
Failure in the electronics, interference or
damage in the measuring cell.
The following table assists in the analysis of
fault signals.
Cause ofFault signal via
failureDOT-matrixBar graph
Clearly"OPERATEBar graph 0 %
outside????or 100 %
measuringbar"digital value
rangeflashes
Digital indication
On instruments with menu-driven adjustment
with additional functions, the possible causes
are displayed under the menu item "Diagnostics no." in case of failure.
DiagnosticsMeaning
no.
1No frequency from C/D-converter
2Frequency signal of the capacitor not
within the limit values
3Frequency signal of the reference
capacitor not within the limit values
4Frequency signal temperature not
within the limit values
7Communication with EEPROM dis-
rupted
9Error in EEPROM CRC-checksum
11Process connection or electronics
module has been exchanged (appears for approx. 20 s after switching
on for the first time after exchange)
Troubleshooting
If the displayed value does not correspond
to the level in the vessel or to the process
pressure, the following measures must be
taken:
- check the pressure compensation (only
with gauge pressure measuring ranges)
- check the electrical connections.
OverloadBar graph 0 %
rangeor 100 %,
of thedigital value
meas. cellfour segments
Failure inall segments
D94flash
flash "- - - -"
Checking pressure compensation
Open the processing unit of the D94 pressure transmitter. The measured value must
not change. However, if the displayed value
changes, compensation of the atmospheric
pressure is not ensured, which can result
distortion of the measured value. Check the
breather on the processing unit.
Pressure transmitter D94 (Profibus PA)31
Page 32
6 Instrument modification
6.1 Exchange of adjustment modules in the processing unit
Display
module
Module for
menu-driven
adjustment
Module without
adjustment
Instrument modification
Exchange of the adjustment module
Removal of the adjustment module
• Separate processing unit from the power
supply.
• Remove the screws on the upper side of
the housing and remove the cover or the
display module.
• Disconnect the cables from the terminals,
and if applicable, pull out the plug connection of the display module.
• Remove the two screws of the adjustment
module.
• Remove adjustment module and pull out
plug connection.
Insertion of an adjustment module
• Plug the plug connection of the new adjustment module into the plug-in socket of
the electronics module.
• Fasten the new adjustment module.
• Reconnect the cables, and if applicable,
also the cable from the display module.
• Replace the cover or the display module
of processing unit.
• Reconnect processing unit to power
supply.
Processing unit
The modular construction of the processing
unit allows retrofitting, interchange and removal of the adjustment modules and the
display module. Data previously saved (e.g.
adjustment values) are not stored in the adjustment module but in an EEPROM of the
electronics module and readjustment is
therefore not necessary. The connection of
the modules is made via a 4-pole plug connection.
32Pressure transmitter D94 (Profibus PA)
Page 33
Notes
6.2 Exchange of hygienic form seal
on D94
On the hygienic connections AA, CA, LA, LB,
RA, RB, TA of pressure transmitter D94, the
ceramic measuring cell is radially, gaplessly
sealed with a form seal. This form seal (material EPDM-FDA) can be exchanged by the
user without the need of a fresh adjustment.
The criteria and the time intervals for such an
exchange of the seal are generally defined
by the process and hygienic requirements.
Important:
Use only original replacement seals from
VEGA!
Please note the following exchange procedure.
Removal procedure:
• Pressure transmitter should be unpres-
surized (switch off process pressure or
empty vessel).
• Loosen the hexagon pressure screw (4)
(width across flats, 46 mm) by turning to
the left.
Note!
The process connection (6) (e.g. compression nut) must not be loosened!
• Turn the adapter (3) below the pressure
transmitter housing with a screwdriver
(with across flats, 36 mm) briefly to the
left (to loosen the seal).
• Pull the complete pressure transmitter on
the housing (1) out of the process connection by carefully turning to the left.
• Slightly lift the form seal (5) and loosen
from the ceramic measuring cell.
• If the form seal (5) does not completely
encircle the measuring cell, it must be
carefully removed from the process connection.
Installation procedure:
• Put the new form seal (5) over the measuring cell (the conical end must point to
the process connection!).
• Carefully insert pressure transmitter into
the process connection by turning clockwise.
• Turn the housing (1) of the pressure
transmitter into the original position.
The exchange of the form seal is now finished; the removed seal must be properly
disposed of.
1
2
3
4
5
6
1 Pressure transmitter
2 Headless screw (avoids overwinding and must not be
removed)
3 Adapter
4 Hexagon screw
5 Form seal
6 Process connection
Pressure transmitter D94 (Profibus PA)33
Page 34
Notes
34Pressure transmitter D94 (Profibus PA)
Page 35
Notes
Pressure transmitter D94 (Profibus PA)35
Page 36
VEGA Grieshaber KG
Am Hohenstein 113
D-77761 Schiltach
Germany
Phone+49 (0) 7836 50-0
Fax+49 (0) 7836 50-201
E-Mailinfo@de.vega.com
www.vega.com
ISO 9001
All statements concerning scope of delivery, application, practical
use and operating conditions of the sensors and processing systems correspond to the latest information at the time of printing.
Technical data subject to alteration
2.24 843 / June 2001
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.