VEGA Capacitive electrodes User Manual

Operating Instructions
Capacitive electrodes EL 4 … 20 mA - Compact
Level and Pressure

Safety information

Safety information
2 Capacitive electrodes EL 4 … 20 mA - Compact

Note Ex area

Please note the attached safety instructions con­taining important information on installation and operation in Ex areas. These safety instructions are part of the operat­ing instructions manual and come with the Ex approved instruments.
22652-EN-020809
Contents
Contents
Safety information ......................................................................... 2
Note Ex area ................................................................................. 2
1 Product description
1.1 Function and configuration ................................................... 4
1.2 Types and versions .............................................................. 5
1.3 Technical data ....................................................................... 7
1.4 Approvals ........................................................................... 13
1.5 Dimensions ......................................................................... 14
1.6 Type plate ........................................................................... 16
2 Mounting
2.1 Mounting instructions ......................................................... 17
3 Electrical connection
3.1 Connection instructions ...................................................... 22
3.2 Wiring plan .......................................................................... 22
4 Setup
4.1 General adjustment ............................................................ 26
4.2 Adjustment - Oscillators
CAP E32 Ex and CAP E32 H Ex ........................................ 27
4.3 Adjustment with VVO .......................................................... 30
4.4 Adjustment with HART® handheld ...................................... 31
5 Diagnosis
5.1 Simulation ............................................................................ 38
5.2 Maintenance ........................................................................ 38
5.3 Repair .................................................................................. 38
5.4 Fault rectification ................................................................. 39
22652-EN-020809
Capacitive electrodes EL 4 … 20 mA - Compact 3

1 Product description

Product description

1.1 Function and configuration

Capacitive electrodes series EL can detect the level of practically any kind of product, regardless of whether it is a liquid, powder or paste. This also applies to adhesive products.
The electrode simultaneously measures the capacitance and the ohmic resistance (admit­tance processing). By this means, even diffi­cult products such as adhesive, conductive substances and solids with fluctuating mois­ture content can be measured.
Through the use of screening tubes and screening segments, inactive areas can be created on the probe where pollution, conden­sation or permanent buildup cannot influence the measuring result.
Measuring principle
Electrode, medium and vessel wall form an electrical capacitor.
The capacitance of the capacitor is generally influenced by three factors:
- distance between electrode surfaces (a)
- size of the electrode surfaces (b)
- type of dielectric between the electrodes (c)
a
The electrode and the vessel wall are the capacitor surfaces. The medium is the dielec­tric. Due to the higher dielectric constant (DK value) of the medium with respect to air, the capacitance of the capacitor increases as the level rises around the electrode.
Fig. 1.2 Capacitance change through varying immersion level
The capacitance change is processed by the oscillator and converted into a level-propor­tional measured value. The measured value is outputted as a standardised, floating 4 … 20 mA analogue signal, or optionally as a digital signal with the HART® communication protocol. The sensor can be adjusted through the integrated oscillator. As an option, adjust­ment is also possible with a HART® handheld or PC.
In continuous level measurement, the level is
b
continuously detected and converted into a level-proportional signal which is either directly displayed or further processed.
c
Fig. 1.1 Plate capacitor (schematic drawing)
For this type of measurement, you need a capacitive electrode series EL with oscillator CAP E32 Ex or CAP E32 H Ex.
Continuous measurement requires a constant dielectric value ε unchanging properties.
4 Capacitive electrodes EL 4 … 20 mA - Compact
i.e. the medium should have
r
22652-EN-020809
Product description

1.2 Types and versions

Version 11*)21*)24*
Type EL EL EL EL EL EL EL EL EL
)
29 31*
)
33 42*
)
52 53
Level detection ••••••••
Continuous ••••••••
Partly insulated
Fully insulated
Oscillators
CAP E32 EX ••••••••
CAP E32 H EX ••••••••
Approvals
ATEX II 2G EEx ia IIC T6
ATEX II 1G,1/2G, EEx ia IIC T6
ATEX II 1/2D IP65 T
7)
••
Overfill protection WHG ••••
German Lloyd
Lloyds Register of Ship
American Bureau of Ship
Bureau Veritas
RINA
1)
1)
1)
1)
•••••••••
•••••••••
1)
•••••••••
•••••••••
•••••••••
Process fitting
1½ NPT •• ••••••
G1½A ••• •••••
Tri-Clamp 1½“
Tri-Clamp 2“
Bolting DN 50
Flange
Flange plated
Screwed adapter
PP •• ••
PTFE
Electrode material
Steel
2)
2)
2)
3)
3)
4)
Stainless steel
2)
*) also Ex0
1) applied for
22652-EN-020809
2) 1.4435
3) 1.4571
4) 1.4401
5) For electrodes certified for Ex-Zone 0, only PTFE and FEP are approved as insulation materials.
6) Level detection with signal conditioning instrument possible
7) only with Aluminium housing
Capacitive electrodes EL 4 … 20 mA - Compact 5
2)
Product description
Version 11*)21*)24*)29 31*)33 42*)52 53
Type EL EL EL EL EL EL EL EL EL
Insulation material
5)
PTFE
PP
PE/PA 12
PFA
FEP
PE
Screening tube
Steel
Stainless steel (1.4435)
Screening tube (option)
Steel •• •••••
Stainless steel
Temperature adapter (option)
Steel
Stainless steel (1.4435)
PA
Housing material
Plastic •••••••••
Aluminium •••••••••
Misc.
6)
Adapter
•• ••
Overvoltage protect. option CB-2-36 (integrated)
Overvoltage arrester
Angled electrode
7)
8)
••• •••••
9)
••
*) also Ex 0
1) applied for
2) 1.4435
3) 1.4571
4) 1.4401
5) For electrodes certified for Ex-Zone 0, only PTFE and FEP are certified as insulation materials.
6) For combustible liquids in pressurised vessels, e.g. liquid gas, ammonia
7) For high electrostatic discharge, e.g. plastic granules
8) Bending max. 90°
9) EL 21 only in PTFE with 3.2 mm insulation thickness
6 Capacitive electrodes EL 4 … 20 mA - Compact
22652-EN-020809
Product description

1.3 Technical data

Housing
Housing material plastic PBT (Polyester) or Aluminium
Protection
- plastic housing IP 66
- Aluminium housing IP 66/67 (meets requirements of both protection
Cable entry 1 piece M20 x 1.5 Terminals for max. 1.5 mm2 wire cross-section
Mechanical connection
Material galvanized steel (St 37), 1.4571 (stst), Aluminium Thread G 1 Flange flange versions
Electrode
Material
- EL 11 1.4435 (316 L)
- EL 21 steel (St 37)
- EL 31 1.4401 (316 L)
- EL 24, 42 1.4571 (316 L) Length
- r od max. 3 m
- cable max. 20 m Insulation see "Insulation materials" Max. tensile load (cable)
- EL 31 20 KN
- EL 33, 34, 52, 53 15 KN
plastic-coated
types)
1/
A or 1
2
1/
" NPT
2
Ambient conditions
Ambient temperature on the housing -40°C … +80°C Product temperature see "Product temperature and operating
pressure" Storage and transport temperature -40°C … +80°C Operating pressure see "Product temperature and operating
pressure"
22652-EN-020809
Capacitive electrodes EL 4 … 20 mA - Compact 7
Oscillators CAP E32 Ex, CAP E32 H Ex
Protection class II Overvoltage category III Frequency 300 KHz Capacitance ranges 0 … 3000 pF Supply voltage 12 … 36 V DC
in Ex applications note the permissible electrical
connection values stated in the certificate Permissible residual ripple USS less or equal 1 V Potential separation min. 500 V DC
Load in
Ohm
1000
800
600
Product description
200
12
18 24 30 36
Supply voltage in V
Accessory
Fixing spring of 1.4571
- length approx. 185 mm (stressed)
- tensile force approx. 200 N
Weight
Basic weight (e.g. EL 24) approx. 0.8 kg Rod weight ø 6 mm - 0.23 kg/m
ø 10 mm - 0.62 kg/m
Oscillators in two-wire technology for capacitive electrodes EL
T y pe App licat ion Meas. range Freq uency Signal condit.
CAP E32 Compact electronics 4 … 20 mA for 0 - 3000 pF 300 KHz not necessary Ex continuous level measurement,
acc. to the principle of phase­selective admittance processing. Parameter adjustment via key pushing on the oscillator
CAP E32 Compact electronics 4 … 20 mA for 0 - 3000 pF 300 KHz not necessary HEx continuous level measurement,
acc. to the principle of phase­selective admittance processing. Parameter adjustment via key pushing on the oscillator, VVO from V. 2.30 or HART® handheld
instrument
8 Capacitive electrodes EL 4 … 20 mA - Compact
22652-EN-020809
Product description
Oscillator
The oscillator CAP E32 (H) Ex with patented processing (phase-selective admittance processing) extends the area of application of capacitive level measurement technology. This function can be switched on, see 4 Setup.
In conjunction with a fully insulated rod elec­trode, the oscillator also compensates heavy conductive buildup.
Mounted with any rod or cable electrode type EL, this oscillator also enables exact measure­ment of solids with varying moisture content.
The oscillator processes the measuring cur­rents acc. to their phase position. Measuring currents with a defined phase shift, commonly occurring when buildup forms or moisture content changes, are hence filtered out.
Varying moisture content
A change of moisture content in solids causes a change of the dielectric constant (er). A simultaneous change of the ohmic resistance of the product also occurs. In addition, the change causes a phase shift of the measuring currents.
If the moisture content goes above 15 % vol., the fully and partly insulated electrodes react differently (see also „Fig. 1.3 Varying moisture content“). Whereas the measured value from fully insulated electrodes increases with un­changing product level, the measured value from partly insulated electrodes falls.
Level Meas. value
+
1)
0 %
1) Real value level
5 %
15 % vol.
10 %
fully insulated electrode
Solid moisture (%)
partly insu­lated electrode
Fig. 1.3 Varying moisture content
With capacitive measurement, even minor changes of moisture content cause measure­ment errors. Typical products are e.g. sand, aggregate in the cement industry, hops or plastic granules (after drying).
When using oscillator CAP E32 (H) Ex, changes of moisture content up to 15 % vol. do not influence the accuracy of the measure­ment. Even layers containing different amounts of moisture do not affect the accu­racy.
22652-EN-020809
Capacitive electrodes EL 4 … 20 mA - Compact 9
Product description
Product temperature1) and operating pressure
1)
The numbers in the tables all relate to the graphs on this page. The stated pressures are valid for screw connections G 1½ A, 1½ NPT and R 1½. Boltings DN 50 acc. to DIN 11 851 only up to max. 25 bar. When using flanged versions, make sure to note their nominal pressure. All electrodes are also suitable for applications in a vacuum (-1 bar). For electrodes certified for Ex-Zone 0, only PTFE and FEP are approved as insulating materials.
Process fitting, 1.4571
Insulation
Electrode type
PE
PP
PTFE
PE/PA12
PFA
FEP
EL 11 - 1 3 - - -
EL 21 1-3-3-
EL 21 (flange) - - 2 - - -
EL 24 - - - - - 2
EL 29 - - 2 - - -
EL 31 - 1 3 - - -
EL 33 - - 3 - - -
EL 42 - - 2 - - -
EL 52 - - 3 - - -
EL 53 - - - 1 - -
Process fitting, steel (St 37)
Insulation
Electrode type
EL 11 -46---
EL 21 4-6-6-
EL 21 (flange) - - 5 - - -
EL 24 -----5
EL 31 -46---
EL 33 --6---
EL 42 --5---
EL 52 --6---
EL 53 ---4--
PE
PP
PTFE
PE/PA 12
PFA
FEP
bar
40
16
0
-30
bar
63
40
16
-50 100
0
bar 63
40 25
0-50 100
bar
40
16
0
-10
bar
63
40
16
0-10 100
bar 63
25
0
-10
60 80
60 80
˚C
˚C
100
˚C
Temperature adapter
˚C
Temperature adapter
EL 53: PE and PE / PA 12: up to 16 bar
11
1
EL 53: PE and PE / PA 12: up to 16 bar EL 34: unpressurised
EL 42: up to 16 bar EL 24: up to 40 bar max. 150°C for 30 min.
EL 42: up to 16 bar EL 24: up to 40 bar max. 150°C for 30 min.
2)
11
22
2
22
33
3
33
200 ˚C
1)
44
4
44
55
5
55
66
6
66
˚C
200
1)
Temperature adapter of PA up to 150°C, from 100°C unpressurised
10 Capacitive electrodes EL 4 … 20 mA - Compact
22652-EN-020809
Product description
Process fitting, Aluminium
Insulation
Electrode type
EL 11 - 7 8 - -
EL 21 7 - 8 - 8
EL 31 - 7 8 - -
EL 33 - - 8 - -
EL 42 - - 8 - -
EL 52 - - 8 - -
EL 53 - - - 7 -
PE
PP
PTFE
PE/PA 12
Process fitting, plastic (screw adapter)
Insulation
PVC
PP
Electrode type
EL 29 9 9 9
PTFE
PFA
77
bar
16
0
-30
bar
40
0
-20
60 80
bar
16
0-50 100
60
80
˚C
˚C
PVC, PP: unpressurised up to 80°C PTFE: 3 bar, -20 … 100°C
˚C
100
7
77
88
8
88
99
9
99
22652-EN-020809
1)
For Ex applications, the permissible temperatures and pressures stated in the certificate must be noted. Please observe the table on the following page.
2)
Temperature adapter of PA up to 150°C, from 100°C unpressurised
Capacitive electrodes EL 4 … 20 mA - Compact 11
Electronics temperature
Product and ambient temperatures must be kept at levels that do not exceed the limit temperature of the electronics.
Temperature classes T4 (or no Ex)
Without temperature adapter
- product temperature -40°C … +135°C
- ambient temperature
1)
-40°C … +90°C
Temperature class T3
Product description
With temperature adapter
Plastic housing Metal housing
Product temperature -40°C … 180°C 200°C 150°C 175°C 200°C
Ambient temperature
1)
-40°C … 80°C 75°C 80°C 69°C 58°C
1) Ambient temperature on the oscillator
12 Capacitive electrodes EL 4 … 20 mA - Compact
22652-EN-020809
Product description

1.4 Approvals

Explosion protection
Only certified capacitive electrodes EL** Ex 0 may be used in hazardous areas with com­bustible gases, vapours or fog. Capacitive electrodes EL** Ex 0 are suitable for use in hazardous areas of zone 1 and zone 0. The explosion protection provided by these instruments is certified by the EC type ap­proval and the conformity certificate, possibly with national zone 0 annex, issued by the approval authority. These documents are generally attached to the instruments. When the capacitive electrodes are mounted and operated in hazardous areas, the Ex installation regulations must be noted. The information and regulations in the sup­plied certificates (EC type approval certificate, conformity certificate) of the capacitive elec­trodes as well as of the accompanying de­vices (signal conditioning instrument, separator, safety barrier) must be noted.
• As a rule, the installation of Ex systems must be carried out only by qualified per­sonnel.
• The capacitive electrodes must be powered from an intrinsically safe circuit; the permis­sible electrical values are stated in the accompanying certificate.
• Capacitive electrodes with electrostatically chargeable plastic parts come with a warn­ing label informing of measures to be taken to avoid the dangers of electrostatic dis­charges. Note the information on the warn­ing label.
• The explosion protection of the instrument used is only ensured when the limit tem­peratures stated in the certificate are not exceeded.
• In case of danger due to oscillation or vibra­tion, the appropriate parts of the capacitive electrodes must be protected.
• After shortening the electrode cable, make sure the gravity weight is sufficiently se-
22652-EN-020809
cured by means of the pins.
Overfill protection acc. to WHG
The statements and regulations of the at­tached type approvals must be noted if a capacitive electrode is used as part of an overfill protection system for vessels storing water-endangering liquids. The general type approval must be available at the application site. Electrodes must only be used in liquids against which the materials of the wetted parts are sufficiently chemically resistant.
Ship approvals
For use on ships, type approval certificates are available from several ship classification authorities (GL, LRS, ABS, BV, RINA).
CE approval
The capacitive electrodes EL meet the protec­tive regulations of EMC (89/336/EWG) and NSR (72/23/EWG). Conformity has been judged acc. to the following standards: EMC Emission EN 50 081 - 1
Susceptibility EN 50 082 - 2
NSR EN 61 010 - 1
Zone 2
According to DIN VDE 0165, instruments can be used in hazardous areas of zone 2 without approval; they must meet the requirements in section 6.3 of this VDE. The compliance of the instruments with these requirements is con­firmed by VEGA in a manufacturer declaration.
Capacitive electrodes EL 4 … 20 mA - Compact 13

1.5 Dimensions

Dimensions of capacitive electrode type EL … Ex 0
Type EL 11 (partly insu­lated) (Ex 0)
85
32,5
Type EL 21 (fully insulated) (Ex 0)
85
32,5
Product description
Type EL 24 (fully insulated, for adhesive products) (Ex 0)
85
32,5
110
20
35
L1
L (min. 100/max. 4000 mm)
M20x1,5
SW 60
G 1½ A
Ø 20
Ø 15
Insulation length L1: PP: 100 mm PTFE: 50 mm StEx: max. 100 mm
L (min. 100 mm, max. 4000 mm)
Type EL 29 (fully insulated)
85
32,5
15825L (min. 100/max. 4000 mm)
M20x1,5
110
20
35
L (min. 100/max. 4000 mm)
M20x1,5
SW 60
G 1½ A
A
L (min. 100 mm, max. 4000 mm)
Type EL 31 (partly insu­lated)
85
(Ex 0)
110
20
35
Ø 20
32,5
M20x1,5
SW 60
L1
G 1½ A
Ø 8
110
20
35
L (min. 120/max. 4000 mm)
M20x1,5
SW 60
G 1½ A
Ø 12
Ø 14
L (min. 120 mm, max. 4000 mm)
Insulation A
PE 2.0 mm 20 mm PTFE 2.0 mm 2 0 mm PTFE 3.2 mm 1 6 mm PFA 2.0 mm 20 mm
outer-ø
Type EL 42 (fully insulated) (Ex 0)
85
32,5
110
20
35
Ø 5
M20x1,5
SW 60
G 1½ A
Ø 14
48
L (min. 120, max. 3000 mm)
L (min. 400/max. 25.000 mm)
200
Ø 40
16
70
L (min. 400 mm, max. 25000 mm) Insulation length L1: PP: 100 mm PTFE: 50 mm StEx: max. 100 mm
100
L (min. 400/max. 25.000 mm)
200
Ø 40
L (min. 400 mm, max. 25000 mm)
Ø 3
Ø 3
120
Ø 20
Ø 36
Ø 40
14 Capacitive electrodes EL 4 … 20 mA - Compact
22652-EN-020809
Loading...