VEGA Capacitive electrodes User Manual

Operating Instructions
Capacitive electrodes EL Profibus PA
Level and Pressure
Safety information, Note Ex area

Safety information

Please read this manual carefully, and also take note of country-specific installation standards (e.g. the VDE regulations in Germany) as well as all prevailing safety regulations and acci­dent prevention rules. For safety and warranty reasons, any internal work on the instruments, apart from that in­volved in normal installation and electrical con­nection, must be carried out only by qualified VEGA personnel.
2 Capacitive electrodes EL - Profibus PA

Note Ex area

Please note the attached safety instructions containing important information on installation and operation in Ex areas. These safety instructions are part of the oper­ating instructions manual and come with the Ex approved instruments.
Contents
Contents
Safety information ........................................................................ 2
Note Ex area ................................................................................ 2
1 Product description
1.1 Function and configuration .................................................. 4
1.2 Types and versions ............................................................. 5
1.3 Technical data ....................................................................... 7
1.4 Approvals ........................................................................... 12
1.5 Dimensions ......................................................................... 13
1.6 Type plate ........................................................................... 15
2 Mounting
2.1 Mounting instructions ......................................................... 16
3 Electrical connection
3.1 Connection instructions ..................................................... 20
3.2 Wiring plan .......................................................................... 20
4 Setup
4.1 Adjustment media .............................................................. 28
4.2 Adjustment with adjustment module MINICOM ............... 29
4.3 Adjustment with VVO ......................................................... 36
5 Diagnosis
5.1 Simulation ............................................................................ 49
5.2 Maintenance ....................................................................... 49
5.3 Repair.................................................................................. 49
5.4 Failure rectification ............................................................. 50
Capacitive electrodes EL - Profibus PA 3

1 Product description

Product description

1.1 Function and configuration

Capacitive electrodes series EL can detect the level of practically any kind of product, regardless of whether it is a liquid, powder or paste. This also applies to adhesive prod­ucts.
The electrode simultaneously measures the capacitance and the ohmic resistance (ad­mittance processing). By this means, even difficult products such as adhesive, conduc­tive substances and solids with fluctuating moisture content can be measured.
Through the use of screening tubes and screening segments, inactive areas can be created on the probe where pollution, con­densation or permanent buildup cannot influ­ence the measuring result.
Measuring principle
Electrode, medium and vessel wall form an electrical capacitor.
The capacitance of the capacitor is generally influenced by three factors:
- distance between electrode surfaces (a)
- size of the electrode surfaces (b)
- type of dielectric between the electrodes (c)
a
The electrode and the vessel wall are the capacitor surfaces. The medium is the dielec­tric. Due to the higher dielectric (DK value) of the medium with respect to air, the capaci­tance of the capacitor increases as the level rises around the electrode.
Capacitance change through varying immersion level
The capacitance change is processed by the oscillator and converted into a level-propor­tional measured value. The measured value is outputted digitally as a Profibus PA signal. The sensor can be adjusted through the integrated oscillator. As an option, the adjust­ment is also possible with a PC running the VVO software.
In continuous level measurement, the level is continuously detected and converted into a level-proportional signal which is either di­rectly displayed or further processed.
For this type of measurement, you need a capacitive electrode with oscillator CAP E34 PA EX.
b
c
Plate capacitor (schematic drawing)
4 Capacitive electrodes EL - Profibus PA
Continuous measurement requires a constant dielectric value ε constant properties.
i.e. the medium should have
r
Product description

1.2 Types and versions

Version 11
Type EL EL EL EL EL EL EL EL EL
1)
1)
21
24
1)
29 31
1)
33 42
1)
52 53
Continuous ••••
Partly insulated
Fully insulated
Oscillators
CAP E34 PA Ex ••••
Approvals
PTB 98 ATEX 2086
PTB no. Ex 98.E.2085
Overfill protection acc. to WHG 2)••••• •
German Lloyd
Lloyds Register of Ship
American Bureau of Ship
Bureau Veritas
RINA
2)
2)
2)
2)
2)
•••••••••
•••••••••
•••••••••
•••••••••
•••••••••
Mechanical connection
G 1 A •• ••••
11/2" NPT •• ••••••
G 11/2 A ••• •••••
Tri-Clamp 11/2"•
Tri-Clamp 2"
Bolting DN 50
Flange ••••
Flange plated
Screwed adapter
PP
PTFE
Electrode material
Steel •• •••••
3)
3)
3)
4)
4)
5)
3)
Stainless steel
3)
Concentric tube
Steel
Stainless steel (1.4435)
1)
also Ex0
2)
applied for
3)
1.4435
4)
1.4571
5)
1.4401
Capacitive electrodes EL - Profibus PA 5
Product description
Version 111)21
Type EL EL EL EL EL EL EL EL EL
Insulating material
1)
1)
24
1)
29 31
1)
33 42
1)
52 53
PTFE
PP
PE/PA 12
PFA
FEP
PE
Screening tube (option)
Steel ••• •••••
Stainless steel
Temperature adapter (option)
Steel
Stainless steel (1.4435)
PA ••• ••••
Housing material
Plastic •••••••••
Aluminium •••••••••
Misc.
2)
Adapter
Overvoltage protection CB-2-36 (integrated)
Overvoltage arrester
Angled electrode
3)
4)
•••
••
••
5)
1)
For electrodes certified for Ex zone 0, PTFE and FEP are approved as insulating material.
2)
For combustible liquids in pressurized vessels, e.g. liquid gas and ammonia.
3)
For high electrostatic discharges, e.g. plastic granules.
4)
Bending max. 90°.
5)
EL 21 only in PTFE with 3.2 mm insulation thickness.
6 Capacitive electrodes EL - Profibus PA
Product description

1.3 Technical data

Housing
Housing material plastic PBT (Polyester) or Aluminium
Protection
- plastic housing IP 66
- Aluminium housing IP 66 and 67 (meets requirements of both
Cable entry 1 piece M20 x 1.5 Terminals for max. 1.5 mm
Mechanical connection
Material galvanized steel (St 37), 1.4571 (stst, Aluminium Thread G 1
Flange flange versions (plated)
Electrode
Material
- EL 11 1.4435 (316 L)
- EL 21 steel (St 37)
- EL 31 1.4401 (316 L)
- EL 24, 42 1.4571 (316 L) Length
- rod max. 3 m
- cable max. 20 m Insulation see "Insulation materials" Max. tensile load 1 (cable)
- EL 31 3 KN
- EL 42 3 KN
plastic-coated
protection types)
1
/2 A or 11/2" NPT
G 1 A or 1" NPT
2
wire cross-section
Ambient conditions
Ambient temperature on the housing -40°C … +80°C Product temperature see "Product temperature and operating
pressure" Storage and transport temperature -40°C … +80°C Operating pressure see "Product temperature and operating
pressure"
Adjustment
- adjustment software VEGA Visual Operating (VVO) on Master-Class 2 PC
- adjustment module MINICOM in the sensor or in external indicating instrument (optional)
- process adjustment surface PACTware
TM
(runs in VVO as subprogram)
- SIMATIC PDM in conjunction with Electronic Device Description (EDD)
Accessory
Fixing spring of 1.4571
- length approx. 185 mm (strained)
- tensile force approx. 200 N
Capacitive electrodes EL - Profibus PA 7
Product description
Weights
Basic weight (e.g. EL 24) approx. 1.4 kg Rod weight
- ø 6 mm 0.23 kg/m
- ø 10 mm 0.62 kg/m
Connection cables
Two-wire sensors power supply and signal via one two-wire
cable (the cable resistance depends on the
supply voltage, see diagram) Wire cross-section generally 2.5 mm Ground connection max. 4 mm
2
2
Cable entry 2 x M20 x 1.5 (cable diameter 5 … 9 mm)
Oscillators CAP E34 PA
Protection class II Overvoltage category III Frequency 300 kHz Capacitance ranges 0 … 3000 pF Supply voltage 12 … 36 V DC
in Ex applications, the permissible electrical
connection values stated in the certificate
must be noted. Load dependent on segment coupler, see
technical data of the segment couplers and
Profibus specification Potential separation min. 500 V DC
Load in
1000
800
600
200
12
18 24 30 36
Supply voltage in V
Oscillator in two-wire technology for capacitive electrodes EL
Type CAP E34 PA EX Application: continuous level measurement acc. to the principle of phase-selective admit­tance processing with potential separation. For connection to digital communication of Profibus PA, approved for use in hazardous areas of zone 0. Measuring range 0 - 3000 pF Frequency 300 kHz Signal conditioning instrument segment coupler
8 Capacitive electrodes EL - Profibus PA
Product description
Oscillator
The oscillator CAP E34 PA Ex with patented processing (phase-selective admittance processing) extends the area of application of capacitive level measurement technology. This function can be switched on, see "4 Setup".
In conjunction with a fully insulated rod elec­trode, the oscillator also compensates heavy conductive buildup.
Mounted with any rod or cable electrode type EL, this oscillator also enables exact meas­urement of solids with varying moisture con­tent.
The oscillator processes the measuring cur­rents acc. to their phase position. Measuring currents with a defined phase shift, com­monly occurring when buildup forms or mois­ture content changes, are hence filtered out.
Varying moisture content
A change of moisture content in solids causes a change of the dielectric constant (ε
). A simultaneous change of the ohmic
r
resistance of the product also occurs. In addition, the change causes a phase shift of the measuring currents.
If the moisture content goes above 15 % vol., the fully and partly insulated electrodes react differently (see illustration "Varying moisture content"). Whereas the measured value from fully insulated electrodes increases with un­changing product level, the measured value from partly insulated electrodes falls.
Level Meas. value
+
1)
-
0 %
5 %
10 %
1)
Real value level
Varying moisture content
15 % vol.
fully insulated electrode
Solid moisture (%)
partly insulated electrode
With capacitive measurement, even minor changes of moisture content cause measur­ing errors. Typical products are e.g. sand, aggregate in the cement industry, hops or plastic granules (after drying).
When using oscillator CAP E34 PA Ex, changes of moisture content up to 15 % vol. do not influence the accuracy of the meas­urement. Even layers containing different amounts of moisture do not affect the accu­racy.
Capacitive electrodes EL - Profibus PA 9
Product description
Product temperature 1) and operating pressure
The numbers in the tables all relate to the graphs on this page. The stated pressures are valid for screw connections G
" NPT, G 1 A, 1" NPT. When using flanged
4
versions, make sure to note their nominal pressure. All electrodes are also suitable for applications in a vacuum (-1 bar).
For electrodes certified for Ex zone 0, only PTFE and FEP are approved as insulating materials.
Mechanical connection, 1.4571
Electrode type
EL 11 1 3 - -
EL 21 1 3 - -
EL 21 with flange - 2 - -
EL 24 - - 2 -
EL 29 - - 2 -
EL 31 1 3 - -
EL 42 ----
EL 52 - 3 - -
EL 53 - - - 1
1)
Insulation
PE
PTFE
FEP
3
/4 A, 3/
PE PA12
bar
40
16
0
-30
bar
63
-50
-50
0
bar
63
25
0
60 80
100
EL 21 up to 16 bar
˚C
EL 42: up to 16 bar EL 24: from 100°C 6 bar, max. 150°C for 30 min.
˚C
100
Temperature adapter
1
2
3
200
˚C
1)
For Ex applications, the permissible temperatures and pressures stated in the certificate must be noted. Please observe the table on the following page.
10 Capacitive electrodes EL - Profibus PA
Product description
Electronics temperature
Product and ambient temperatures must be kept within the following indicated ranges, so that the limit temperature on the electronics is not exceeded.
The stated values are mandatory for applications in hazardous areas. For
these applications, also note the re­spective official documents (test reports, test certificates, type approvals and conformity certificates).
Temperature class T6
Product temperature -40°C … +60°C Ambient temperature -40°C … +60°C
Temperature class T5
Product temperature -40°C … +75°C Ambient temperature -40°C … +75°C
Temperature class T4 (or no Ex)
Without temperature adapter
- Product temperature -40°C … +135°C
- Ambient temperature
1)
-40°C … +90°C
With temperature adapter
Plastic housing Metal housing
Product temperature -40°C … 180°C 200°C 150°C 175°C 200°C
Ambient temperature
1)
Ambient temperature on the oscillator
Capacitive electrodes EL - Profibus PA 11
1)
-40°C … 80°C 75°C 80°C 69°C 58°C
Product description

1.4 Approvals

Explosion protection
Only certified capacitive electrodes EL** Ex 0 may be used in hazardous areas with combus­tible gases, vapours or fog. Capacitive elec­trodes EL** Ex 0 are suitable for use in hazardous areas of zone 1 and zone 0. Proof of the explosion protection provided by these instruments is the EC type approval and the conformity certificate, possibly with national zone 0 annex issued by the approval authority. These documents are generally attached to the instruments. When the capacitive electrodes are mounted and operated in hazardous ar­eas, the Ex installation regulations must be noted. The information and regulations in the supplied certificates (EC type approval certifi­cate, conformity certificate) of the capacitive electrodes as well as of the accompanying devices (signal conditioning instrument, sepa­rator, safety barrier) must be noted.
• As a rule, the installation of Ex systems must be carried out only by qualified personnel.
• The capacitive electrodes must be powered from an intrinsically safe circuit; the permissi­ble electrical values are stated in the accom­panying certificate.
• Capacitive electrodes with electrostatically chargeable plastic parts come with a warn­ing label informing of measures to be taken to avoid the dangers of electrostatic dis­charges. Note the information on the warning label.
• The explosion protection of the instrument used is only ensured when the limit tempera­tures stated in the certificate are not ex­ceeded.
• In case of danger due to oscillation or vibra­tion, the appropriate part of the capacitive electrodes must be protected.
• After shortening the electrode cable, make sure the gravity weight is sufficiently secured by means of the pins.
Ship approvals
For use on ships, type approval certificates are available from several ship classification authorities (GL, LRS, ABS, BV, RINA).
CE approval
The capacitive electrodes type EL meet the protective regulations of EMC (89/336/EWG) and NSR (72/23/EWG). Conformity has been judged acc. to the following standards: EMC Emission EN 50 081 - 1
Susceptibility EN 50 082 - 2
NSR EN 61 010 - 1
Zone 2
According to DIN VDE 0165, instruments can be used in hazardous areas of zone 2 with­out approval; they must meet the require­ments in section 6.3 of this VDE. The compliance of the instruments with these requirements is confirmed by VEGA in a manufacturer declaration.
12 Capacitive electrodes EL - Profibus PA
Product description
Ø 12
Ø 14
110
35
L (min. 120/max. 4000 mm)
20
SW 60
G 1½ A
85
32,5
M20x1,5

1.5 Dimensions

Dimensions of capacitive electrodes type EL … Ex 0
Type EL 11 (partly insu­lated)
85
32,5
EL 21 (fully insulated)
85
32,5
Type EL 24 (fully insulated, for adhesive products)
110
20
35
L1
L (min. 100/max. 4000 mm)
Ø 20
Ø 15
M20x1,5
SW 60
G 1½ A
Type EL 29 (fully insulated)
85
32,5
15825L (min. 100/max. 4000 mm)
M20x1,5
110
20
35
L (min. 100/max. 4000 mm)
M20x1,5
SW 60
G 1½ A
A
Type EL 31 (partly insu­lated)
85
32,5
110
20
35
Ø 20
M20x1,5
SW 60
L1
G 1½ A
Type EL 33 (partly insu­lated)
110
85
32,5
M20x1,5
20
35
L1
Ø 15,5
SW 60
G 1½ A
Ø 8
Capacitive electrodes EL - Profibus PA 13
Ø 8
L (min. 400/max. 25.000 mm)
L (min. 400/max. 25.000 mm)
Ø 14
48
200
Ø 40
100
16
70
200
Ø 40
Dimensions of the capacitive electrodes type EL … Ex 0
EL 52 (fully insulated)
85
85
32,5
32,5
Product description
Type EL 53 (fully insulated)Type EL 42 (fully insulated)
85
32,5
110
20
35
Ø 5
Ø 3
L (min. 400/max. 25.000 mm)
200
Ø 40
Housing
90
Housing of plastic/Aluminium
Concentric tube
G 1½ A
Ø 40
SW 41
Ø 3
M20x1,5
SW 60
Ø 20
Ø 36
110
20
35
200
L (min. 400/max. 25.000 mm)
120
ø 40
85
32,5
M20x1,5
of 1.4435
M20x1,5
SW 60
G 1½ A
ø 15,5
ø 8
125
ø 20
ø 36
ø 40
Temperature adapter
133
20
Screening tube
110
20
35
L (min. 400/max. 25.000 mm)
200
ø 40
SW 41
SW 41
G ¾ A,
of 1.4435
G 1 A
ø 11
ø 6
G 1½ A
ø 40
M20x1,5
SW 60
ø 20
ø 36
120
SW 41
L
L7
Ø 21,3
Ø 21,3
of 1.4435 with closing cone of PTFE
14 Capacitive electrodes EL - Profibus PA
Product description

1.6 Type plate

Before mounting and electrical connection, please make sure you are using a suitable instrument. Note the type plate which is lo­cated as shown below.
Type plate
The type plate contains important data re­quired for mounting and connection. The configuration and elements of the type plate are hence explained in the following example.
Contents of the type plate (example)
VEGA® EL 11
1
type EK11EXO.XGBVSTXXVKXX
2
see PTB Nr. EX-98.E.2085 EEx ia IIC T6 0032
2
PTB 98 ATEX 2086 II 1/2G EEx ia IIC T6 techn. data see document. / certificates 1998
3
protection: IP 66/ 67 Insp. length: 400mm VVO: 02
4
Ord. no. 123456/000
Z-65.13.XXX
®
D-77757 Schiltach
7
ser. no. 10612892
5
6
1 Master data of the order number 2 Ex certification number
Explosion protection version - note the
information and regulations of the certificate 3 Data of the electronics/Approvals 4 No. of the order confirmation/Pos. no. 5 Number of the electrode type 6 Serial number 7 Test mark when used as part of an overfill
protection system for vessels storing water
endangering liquids - note the information
and regulations of the general type ap-
proval 8 Year of manufacture 9 Number of the test authority
9
8
Order code
You can find detailed information on the order code in the "VEGA Price list".
Capacitive electrodes EL - Profibus PA 15

2 Mounting

Mounting

2.1 Mounting instructions

General
Different mediums and measurement require­ments make different installation positions necessary. Please note the following points.
Lateral load
Make sure that the electrode is not subjected to strong lateral forces. Mount the electrode in a position in the vessel where no disturbing factors, e.g. stirrers, filling openings etc. are present. This is especially important for very long rod and cable electrodes.
In such cases, use a rod electrode for short meas. distances, since a rod is usually more stable. If, due to the length or the installation position, a cable electrode is necessary, the electrode should not be fixed but simply provided with a gravity weight to allow the cable to more easily follow the product move­ments. Make sure that the electrode cable does not touch the vessel wall.
Pressure
If there is pressure or underpressure in the vessel, the mounting boss on the thread must be sealed. Use the attached seal ring. Check whether the seal ring is resistant against the medium.
Insulating measures, such as e.g., covering the thread with Teflon tape, can interrupt the necessary electrical connection. For this reason, ground the instrument on the vessel.
Shortening the electrode
Fully insulated electrodes have fixed dimen­sions which must not be modified in any way. Any modification will destroy the instrument.
Lateral load
Extracting forces
In case of strong extracting forces, e.g. due to material inflow or material slide, high tensile loads can occur.
16 Capacitive electrodes EL - Profibus PA
Partly insulated cable or rod electrodes can be shortened if necessary. The basic capaci­tance of the electrodes is compensated auto­matically at adjustment time. Because of this, the electrodes can be shortened by any desired amount.
Cable electrode EL 31 can also be short­ened. Loosen the two pins on the gravity weight (hexagon socket) and remove them. Pull the cable out of the gravity weight.
Mounting
To avoid fraying the steel cable (EL 31) dur­ing cutting, you must tin the cable around the cutting position with a soldering iron or blowtorch, or tightly wrap the cable with a wire. Shorten the electrode cable with a metal cutting saw or a cutting-off wheel by the requested length.
Carry out an adjustment. The instruction is under "4.1 Adjustment".
Shortening of the electrode
This is mainly valid for outdoor mounting, in areas where high humidity is expected (e.g. from cleaning processes) or for cooled or heated vessels.
Humidity
Cable entries
When mounting outdoors, on cooled vessels or in humid areas where cleaning is done, e.g. with steam or high pressure, the seal of the cable entry is very important.
Use cable with a round cross-section con­ductor and tighten the cable entry. The cable entry is suitable for cable diameters of 5 mm to 9 mm.
Filling opening
Install the electrode such that it does not protrude directly into a strong filling stream. Should such an installation location be neces­sary, mount a suitable protective sheet metal cover above or in front of the electrode e.g. L 80 x 8 DIN 1028, etc.
Humidity from outside
To avoid moisture ingress, the connection cables to the electrode housing of vertically installed electrodes should be led downward directly outside the cable entry. Rain and condensation water can then drain off.
Capacitive electrodes EL - Profibus PA 17
Metal vessel
Make sure that the mechanical connection of the instrument to the vessel is electrically conductive in order to ensure sufficient grounding.
Use conductive seals such as e.g. copper, lead etc. insulating measures, such as cover­ing the thread with Teflon tape, can interrupt the necessary electrical connection. In such case, use the earth terminal on the housing to connect the electrode to the vessel wall.
Non-conductive vessels
In non-conductive vessels, e.g. plastic tanks, the second pole of the capacitor must be provided separately, e.g. in the form of a concentric tube.
When using a standard instrument, a suitable grounding surface is necessary. Provide therefore the largest possible grounding surface, e.g. wire braiding laminated into the vessel wall or metal foil glued to the vessel. Connect the grounding surface to the ground terminal on the housing.
Mounting
Cable electrode in solids
Rod electrode
Mount the rod electrode such that the elec­trode protrudes into the vessel. When mount­ing in a tube or a socket, material buildup that deters the measurement can form. This is particularly the case with viscous or adhesive mediums.
Cable electrode in solids
Depending on the type of solid, the filling method and its location, the cable electrode may "float" inspite of the gravity weight. The electrode (cable) is pushed by the solid against the vessel wall or to the surface, thus causing incorrect measured values. This should be avoided during continuous meas­urement. If this happens, use a fixing weight or insula­tor to secure the cable electrode.
When securing the cable electrode, avoid high tensile forces. An appropriate fixing spring which prevents cable overload can be found in our price list.
Lateral mounting
With measuring probes that deliver continu­ous measured values, the electrode must be installed vertically. If installation from the top is not possible, the electrodes can also be mounted laterally.
Under accessories in our price list, you can find a screening tube and a closing cone or an angled rod electrode with which the instru­ment can also be laterally mounted. Deter­mine a screening tube length (L) that will not allow product bridges to form between cable and vessel wall, and that will not allow the electrode cable to touch the vessel wall through movements. Use a fixing weight or fixing insulator.
L
Fixing weight
Continuous electrodes
18 Capacitive electrodes EL - Profibus PA
Mounting
d
3
1
2
d
10
Material cone
When installing the meas. probe in a vessel, be aware that conical heaps which alter the switching point can form in solid materials. We recommend selecting an installation loca­tion where the electrode detects the average value of the conical heap.
Filling
d 6
d
Material cone, filling and emptying centered
d 6
The measuring probe must be installed in a way that takes the filling and emptying aper­tures of the vessel into account. To compen­sate measurement errors caused by the conical heap, you should install the electrode at a distance of approx.
d
/6 from the vessel
wall.
Emptying
d
Filling Emptying
1 Emptying
Material cone, filling centered, emptying laterally
Capacitive electrodes EL - Profibus PA 19
2 Filling opening 3 Capacitive
electrode

3 Electrical connection

Electrical connection

3.1 Connection instructions

Note
Switch off the power supply before starting connection work.
Connect the supply voltage according to the following connection diagrams.
Note
If strong electromagnetic interference is ex­pected, we recommend using screened cable. The screening of the cable should only be earthed on one end, i.e. the sensor end (electrode).
Always connect the instrument to vessel ground (PA). For this purpose, there is a terminal on the side of the housing. This con­nection is also used to supply the ground reference potential as well as to drain off electrostatic charges.
The electronics module CAP E34 PA requires a supply voltage of 9 … 32 V DC. The supply voltage and the digital output signal are led via the same two-wire connection cable to the terminals. The external energy is provided by a segment coupler.
nection (cable types see chapter "1.3 Tech­nical data"). The shielding must be con­nected on both ends (on the T-adapter and on the instrument).
- For use in Ex areas, the installation regula­tions must be noted.

3.2 Wiring plan

Note
The oscillator is dependent on the electrode and can be exchanged on site.
12 5678
+-
PROFIBUS
R
EEx ia IIC see certificate
on
87654321
Note the following instructions for electrical connection:
- Connection must be made according to the country-specific installation standards (e.g. in Germany acc. to the VDE regulations).
- The electrical connection must have a pro­tective measure against polarity reversal.
- The connection of field instruments is gen­erally made via T-adapters on stubs.
Segment coupler
- Screened cable is recommended for con-
20 Capacitive electrodes EL - Profibus PA
Electrical connection
Connection of the indicating instrument VEGADIS 50
Loosen the 4 screws of the housing lid on VEGADIS 50. To make wiring easier, we rec­ommend fastening (with a screw) the cover to the side of the housing during connection work.
Adjustment module
SENSOR
DISPLAY1234 5678
Tank 1 m (d)
12.345
ESC
+
-
OK
Screws
12 5678
+-
PROFIBUS
R
EEx ia IIC see certificate
on
87654321
Capacitive electrodes EL - Profibus PA 21
Electrical connection
Qualified personnel
Instruments which are operated in Ex areas must only be connected by qualified person­nel. They have to note the installation regula­tions and attached EC type approval certificates and conformity certificates.
If the capacitive electrodes are to be mounted on vessels 0.8 which must be pro­tected against ignition by lightning strokes (acc. to TRbF 100 no. 8, para. 1), they have to be equipped with the external overvoltage protection unit, type B62-30W.
Connection cable and cable length
Connection cables must be in conformity with the Profibus specification and the FISCO model. The sensor cable must be in conform­ity with the values of the reference cable acc. to IEC 1158-2:
2
0.8 mm
Z C
The max. cable length first of all depends on the transmission speed: up to 32 kbit/s: 1900 m Prup to 32 kbit/s: 1900 m Pr
up to 32 kbit/s: 1900 m Pr
up to 32 kbit/s: 1900 m Prup to 32 kbit/s: 1900 m Pr up to 94 kbit/s: 1200 m Profibus DP up to 188 kbit/s: 1000 m Profibus DP up to 500 kbit/s: 500 m Profibus DP up to 1500 kbit/s: 200 m Profibus DP up to 12000 kbit/s: 100 m Profibus DP
The distributed resistance of the cable, in conjunction with the output voltage of the segment coupler and the current requirement or the voltage requirement of the sensors, determines the greatest possible length of the cable.
In the practical application of a PA bus branch, the max. length of the cable is also determined (beside the required supply voltage and max. current consumption of all participants on the PA bus branch) by the bus structure and the type of segment cou­pler used.
; R
= 44 Ω/km;
DCmax.
= 80 … 120 ; damping = 3 dB/km;
31.25kHz
= 2 nF/km.
asymmetric
ofibus Pofibus P
ofibus P
ofibus Pofibus P
AA
A
AA
The cable length results from the sum of all cable sections and the length of all stubs. The length of the individual stubs must not exceed the following lengths: 1 … 12 stubs 120 m (Ex: 30 m) 13 … 18 stubs 60 m (Ex: 30 m) 19 … 24 stubs 30 m (Ex: 30 m)
No more than 24 stubs are permitted, whereby each branch longer than 1.2 m is counted as a stub. The total length of the cable must not exceed 1900 m (in Ex version 1000 m).
Ground terminal
The electronics housings of the sensors have a protective insulation. The ground terminal in the electronics housing is galvanically con­nected with the metallic process connection. For sensors with a plastic thread as process fitting, the sensor grounding must be made by a ground connection to the external ground terminal.
Screening
According to the Profibus specification, screening should be done on both ends. To avoid potential equalisation currents, a po­tential equalisation system must be provided in addition to the screening.
According to the specification, we recom­mend the use of twisted and screened two­wire cable, e.g.: SINEC 6XV1 830-5AH10 (Siemens), SINEC L26XV1 830-35H10 (Sie­mens), 3079A (Belden).
Alternatively, when grounding at both ends in non-Ex areas, the cable shielding can be connected on one ground side (in the switch­ing cabinet) via a capacitor (e.g. 0.1 µF; 250 V) to the ground potential. Make sure that the ground connection has the lowest possible resistance (foundation, plate or mains earth).
22 Capacitive electrodes EL - Profibus PA
Electrical connection
Profibus PA in Ex environment
For application in Ex areas, the PA bus and all connected instruments must be certified as intrinsically safe protective class "i". Four­wire instruments requiring a separate power supply must at least have an intrinsically safe PA connection. VEGA sensors for PA-Ex environment are always "ia two-wire instru­ments".
In the so-called Fieldbus Intrinsically Safe Concept (FISCO), the general conditions for an Ex safe bus configuration have been laid down. Therein, the participants and the bus cables with their respective electrical data have been determined, so that the linking of these components always meets Ex require­ments. The more time-consuming Ex calcula­tion normally required is therefore not necessary. Build your Ex bus according to the IEC standard 1158-2.
The Ex segment coupler delivers a controlled power supply to the PA bus. All other compo­nents (field instruments and bus terminators) are only consumers. A field instrument must consume at least 10 mA. Ideally, an individual sensor should not consume more than 10 mA, so that the number of instruments can be as large as possible.
Watch out for potential losses
Due to potential losses, earthing on both sides without a potential equalisation system is prohibited in Ex applications. If an instru­ment is used in hazardous areas, the re­quired regulations, conformity and type approval certificates for systems in Ex areas must be noted (e.g. DIN 0165). Please also note the approval documents with the safety data sheet attached to the Ex sensors.
Ex protection
If an instrument is used in dust Ex areas, the required regulations as well as the StEx cer­tificates for systems in StEx areas must be observed.
Capacitive electrodes EL - Profibus PA 23
Electrical data of the cables
Electrical connection
R
DC
No. of A in Z cores mm
2
31.25kHz
C in Damping Screen nF/km
SINEC 6XV1 44 Ω/km 2 0.75 100 Ω < 90 < 3 dB/km Cu braiding 830-5AH10 +/- 20 39 kHz (Siemens)
SINEC L26XV1 44 Ω/km 2 0.75 100 Ω < 90 < 3 dB/km Cu braiding 830-35H10 +/- 20 39 kHz (Siemens)
3079A 105 /km 2 0.32 150 29.5 < 3 dB/km Foil (Belden) 39 kHz
24 Capacitive electrodes EL - Profibus PA
Electrical connection
Overvoltage arresters for Ex systems
B62-30 W can be used as overvoltage arrester in Ex systems. In Ex areas, overvoltage ar­resters must only be mounted in zone 1 or zone 2 (not in zone 0). You must differentiate between overvoltage protection of intrinsically and non-intrinsically safe circuits.
For Ex systems, please note the installation conditions as well as the special prerequisites cited in the respective conformity certificates.
B62-30 W: 9 … 36 V DC, max. 1 A
CAP E34 PA Ex
+
Profibus PA
e.g.
12 5678
+-
PROFIBUS
R
EEx ia IIC see certificate
87654321
on
B62-30 W
IN E1­E2+
OUT A1+ A2-
Carrier rail
- Terminal row (IN) unprunpr
generally
otectedotected
unpr
otected side of the overvoltage arrester.
unprunpr
otectedotected
- Terminal row (OUT) prpr
generally
Capacitive electrodes EL - Profibus PA 25
otected otected
pr
otected side of the overvoltage arrester.
prpr
otected otected
Electrical connection
Connection example for the installation of an Ex system on vessels without cathodic corrosion protection
The following overvoltage arrester can be used:
--
B62-30 W: B62-30 W:
-
B62-30 W: 9 … 36 V DC, max. 1 A
--
B62-30 W: B62-30 W:
Ex area, Zone 1 or Zone 2
Overvoltage arrester
Oscillator
Screen
Transmitter
External ground terminal
Sensor
Vessel Zone 0
A
A cable with either metal cover, shielding or metallic protection tube accor ding to VDE must be used
2 1
to the potential equalisation cable
A
min. 4 mm
in metal or plastic housing
A2 A1
B62-30 W
2
Cu
B
E2 E1
Screen
Non-Ex area
Switching cabinet/Control room
optionally overvoltage arrester
E2
A2 A1
E1
B62-30 W
Screen -if necessary- connect only on one cable end to ground or PA
between overvoltage arrester and transmitter (metal cover, shielding or protection tube must be con­nected to the potential equalisation). Test voltage of the cable "A" ³ 1500 V AC.
B
A cable acc. to VDE, if necessary with metal cover or shielding, must be used between control room and overvoltage arrester (connect metal cover or shielding, if necessary, only on one end of the cable to earth or PA). Test voltage of the cable "B" ≥ 500 V AC.
Signal condition­ing instrument or safety barrier
26 Capacitive electrodes EL - Profibus PA
Electrical connection
Connection example for installation of an Ex system on vessels with cathodic cor­rosion protection
The following overvoltage arrester can be used:
--
B62-30 W: B62-30 W:
-
B62-30 W: 9 … 36 V DC, max. 1 A
--
B62-30 W: B62-30 W:
Ex area, Zone 1 or Zone 2
Overvoltage-
Oscillator
Screen
Transmitter
External ground terminal
Sensor
Vessel Zone 0
Cathodic tank protection 2 … 24 V (object voltage)
A
A cable with either metal cover, screening or metallic protection tube according to VDE must be used between overvoltage arrester and transmitter (metal cover, screening or protective tube must not be
2 1
Insulated cable min. 4 mm
A
2
Cu
arrester in metal or plastic housing
A2 A1
B62-30 W
B
E2 E1
Screen
Non-Ex area
Switching cabinet/Control room
optionally overvoltage arrester
E2
A2 A1
E1
B62-30 W
grounded). Test voltage of the cable "A" 1500 V AC.
B
A cable acc. to VDE, if necessary with metal cover or screen, must be used between control room and overvoltage arrester. Metal cover or screen - if necessary - must only be connected on the transmitter side of the overvoltage arrester. The cable must have an outer insulation. Test voltage of the cable "B" ³ 500 V AC.
Signal condition­ing instrument or safety barrier
Capacitive electrodes EL - Profibus PA 27

4 Setup

Setup

4.1 Adjustment media

All VEGA Profibus sensors operate in profile 3 and can be adjusted with:
- the adjustment program VVO on a PC with Profibus card
- VVO via VEGACONNECT 3
- the adjustment program PACTware which VVO runs as a subprogram
- the Siemens software PDM in conjunction with an EDD (Electronic Device Descrip­tion)
- the adjustment module MINICOM in the sensor.
Adjustment with VVO on the PC
The adjustment program VVO enables user­friendly adjustment of VEGA Profibus PA sensors. All functions and options of sensor adjustment are accessible. The program runs under Windows Profibus-Master-Class 2 interface card on Profibus DP level as Master-Class 2 tool. The VVO program accesses the VEGA PA sen­sors via the DP bus, the segment coupler and the PA bus. You can also directly access the sensors via a VEGACONNECT 3.
The adjustment program VVO from VEGA makes the full range of functions of VEGA sensors available and can update the entire sensor software, if necessary. The adjust­ment program must be installed on a PC which is equipped with a Profibus-Master­Class 2 interface card (Softing) (see illustra­tion on the next page).
®
on a PC with a
TM
, under
In practice, the adjustment program VVO is often installed as a tool on an engineering station or an adjustment station. VVO then accesses over the bus all VEGA sensors via the Profibus interface card (e.g. from Softing) as Master-Class 2, from the DP level, through the segment coupler on the PA level right down to the individual sensor.
Setting the P rofibus PA address
Since each bus participant must have its own address by which it is identified on the bus, the address must be adjusted before setup of the meas. instrument. The address setting can be done in different ways:
- mini coding switch
- Minicom
- VVO
Address adjustment with the mini cod­ing switch
The transmitter is delivered with the address set as shown (all switches position "on").
12345678 on
If the mini coding switch is set to address 126 or higher (factory setting), the address can be modified via software (with VVO) or with the adjustment module MINICOM. We recommend adjusting the address of the sensor via the hardware (switch) before connecting to the bus.
The PC with the Profibus interface card can be connected directly at any location on the DP bus by means of the standard RS 485 Profibus cable.
The adjustment and parameter data can be saved with the adjustment software on the PC and protected by passwords. On request, the adjustment can be also transferred to other sensors.
28 Capacitive electrodes EL - Profibus PA
Setup
If all sensors are already connected to the bus cable, the address must be set via VEGACONNECT 3, the adjustment module MINICOM or via the mini coding switch.
Example: To select address 37, set switch 1 (value 1), switch 3 (value 4) and switch 6 (value 32) to position "on". Set all other switches to position "off".
LV 8
12345678 on
1 2 4 81632 64128
1+4+32=37
6 I
value
Device description (DD)
Beside the instrument master file (GSD), by which a sensor is logged into the Profibus system, and specific adjustment software, the majority of all Profibus sensors also re­quires a so-called EDD (Electronic Device Description) to enable access and adjust­ment of each sensor from the bus levels. This is not the case when using VVO. The adjust­ment software VVO can communicate at any time with all VEGA sensors without the help of a special database. Of course, all other VEGA sensors can be adjusted with the adjustment software as well (4 … 20 mA sensors or VBUS sensors). With VEGA sen­sors, it is not necessary to go looking for the latest EDD. Expected by many users, this feature is an essential requirement for a manufacturer-independent adjustment pro­gram.
Adjustment with PACTware
TM
The adjustment with PACTwareTM corre­sponds to VVO adjustment, in which case VVO runs as a subprogram of PACTware The adjustment instructions can be found in the documentation of PACTware
TM
TM
.
.
Adjustment with PDM
The sensors can be adjusted completely with PDM. However, some convenient functions and many special features, like, e.g. display of an echo curve, are not available. In addi­tion to the PDM software, an EDD (available upon request from VEGA) is required for each sensor type. The adjustment instruc­tions for PDM are described in the PDM documentation.

Adjustment with adjustment module MINICOM

With the adjustment module MINICOM, you adjust the individual sensor directly in the sensor or in the external indicating instrument VEGADIS 50. The adjustment module MINICOM enables (with the 6 key adjustment field with text display) all essential functions of parameter setting and adjustment.
Capacitive electrodes EL - Profibus PA 29
Setup
4.2 Adjustment with adjustment
module MINICOM
As with the PC, the sensor can also be ad­justed with the small detachable adjustment module MINICOM. The adjustment module is inserted into the sensor or into the external indicating instrument (optional).
ESC
+
-
Tank 1 m (d)
12.345
OK
2
ESC
+
-
Tank 1 m (d)
12.345
OK
4
The adjustment module also provides all the adjustment options available on the PC with the adjustment program VVO. All adjustment steps can be carried out with the 6 keys of the adjustment module. A small display gives you, beside the measured value, a short information on the menu item or the value of a menu adjustment.
Not possible, however, are parts of the ad­justment relating to the configuration and signal processing (e.g. linearisation curve). This is only possible with the PC.
Adjustment elements
The adjustment module MINICOM is menu­driven. The clear text indications on the dis­play lead through the menu. The functions of the keys are described in the following.
2
1
Tank 1
-
m (d)
12.345
5
OK key (4)
With the OK key, adjustments can be con­firmed.
When the symbol move with the OK key to the next lower menu level.
When the symbol appears, there is no branching below the menu item, but only another menu item belonging to the same function.
ESC key (3)
With the Escape key (ESC) you can, de­pending on the menu item, interrupt an ad­justment or a function or change to the next higher menu level.
E.g. to reach the top menu level, push the ESC key several times.
3
ESC
+
OK
4
or is displayed, you can
▼ ▼
+ and - key (2)
With the keys + and - you can modify the values of the parameters or choose from several options.
After pushing the first time, the values to be adjusted flash. The value will be modified with each additional push of the key.
30 Capacitive electrodes EL - Profibus PA
Setup
Arrow keys (5)
With the keys > and <, you can move within the menu level from one menu item to the other.
Digital indication (1)
On the digital indication, the actual measured value is displayed during operation. When you adjust the instrument, the clear text indi­cation displays the appropriate function.
Branching point from which you can change to a lower menu.
This symbol informs you about an ensuing
safety enquiry.
Adjustment steps
On the following pages, you will find the menu schematic of the adjustment module MINICOM.
Set up the sensor in the following sequence. The numbers correspond to the setup se­quence. You will find the numbers near the appropriate menu items in the menu sche­matic on the following pages.
1. Sensor optimisation
2. Adjustment
3. Conditioning
4.Outputs
1. Sensor optimisation
With this function, you can set the sensor optimisation suitable for your application. With it, the phase angle of the phase selective admittance processing (PSA) is modified. "Mode 1" is preset.
Mode 1 = Standard (90°)
The standard setting is recommended for unproblematic applications such as e.g. liquids with small DK values or with elec­trodes in conjunction with a concentric tube.
The standard setting is also recommended for non-conductive vessels with an exterior grounding surface.
Mode 1 is a pure capacitance measurement, the ohmic conductance is not taken into ac­count in the measuring result.
Application:
- standard setting
- non-conductive liquids up to approx. 50 µS
- compensation of conductance changes in liquids
- generally with partly insulated electrodes in liquids
- non-conductive solids without moisture content
- with insufficient product grounding
- with electrodes in conjunction with a con­centric tube
- in non-conductive vessels with exterior grounding surface
Mode 2 = PSA (45°)
For buildup, varying product moisture con­tent With this mode, the patented phase selective admittance processing (PSA) is set. This setting is recommended for problematic applications in solids, mainly if fluctuations of the moisture content occur, and in conductive liquids that stick to the electrode.
The capacitance and the ohmic conductance are measured separately, the capacitance value is corrected by calculating it with the ohmic conductance, so that measurement errors caused by conductive buildup or varying product moisture content are com­pensated.
Capacitive electrodes EL - Profibus PA 31
Setup
Application:
- highly conductive products
- adhesive, conductive products
- solids with fluctuating moisture content
For use in conductive, viscous liquids, a suitable electrode of type EL 24 or EL 24 should be used.
2. Adjustment
Under the menu item " the sensor of the meas. span it should oper­ate in.
Max.
Min.
You can carry out the adjustment with me­dium or without. Generally, you will carry out the adjustment with medium, because the electrode can then adapt to the conditions in the vessel.
Adjustment
100 % corresponds to 1200 l
0 % corresponds to 456 l
", you inform
Span
The following criteria apply to the conductiv­ity:
Mode 1 (90°) Mode 2 (PSA 45°)
with conc. tube > 50 µS/cm > 150 µS/cm w.o conc. tube > 100 µS/cm > 300 µS/cm
Adjustment in
First of all, choose the unit you want to carry out the adjustment in. Possible are pF, m or ft.
Offset correction
Carry out the offset correction with unimmersed electrode.
This is necessary in order to save the initial capacitance in the electronics. The electrode must be mounted in the vessel and it must be completely uncovered by the medium.
For electrodes with concentric tube, this function is already carried out at VEGA, as the vessel does not influence the measure­ment. The offset correction must only be repeated if the oscillator is exchanged.
Conductivity
If you have selected adjustment in m or ft, you have to enter whether the medium is conductive or non-conductive. If the medium is non-conductive, you also have to enter the dielectric constant in the next menu item.
The above table shows you the criteria that
Adjustment without medium
apply to conductivity in capacitive measure-
ment. If you want to carry out the adjustment with­out medium, your measurement loop must meet certain requirements.
In conductive medium
- electrode is fully insulated
Adjustment with medium
Fill the vessel e.g. to 10 %, enter in the menu
"
Min. adjustment
" with the "+" and "–" keys 10
% and confirm with the "OK" key. Then fill the In non-conductive medium
- concentric tube electrode
- you know the DK value of the medium
vessel e.g. to 80 % or 100 %, enter in the
menu "
Max. adjustment
" with the "+" and "–" keys 80 % or 100 % and confirm with the "OK" key.
32 Capacitive electrodes EL - Profibus PA
Setup
Key Display
Sensor
m(d)
4.700
Para-
OK
OK
OK
OK
meter adjust­ment
Adjust­ment
with medium
Min. adjust at %
-------
(Min. adjustment)
3. Conditioning
Under the menu item " can scale the display, set an integration time and choose a linearisation curve.
Scaling
Here, you can scale the digital display of the electrode.
Signal conditioning
" you
If you need for the max. possible value a number with more than four digits, then you should choose the next higher unit, e.g. 14000 l = 140 hl.
Decimal point
If necessary, set the position of the decimal point. Keep in mind that max. 4 digits can be displayed.
Unit
In the menu " unit (%, t, ft yd, m
Unit
", you choose the physical
3
, in3, gal, in, lb, pF, m, cm, mm, ft,
3
, kg, l, hl).
Linearisation curve
In the menu "
Lin. curve
", you can choose between three standard linearisation curves. Preadjusted is a linear depedency between percentage value of the level and the per­centage value of the filling volume.
You can choose between linear, cylindrical tank, spherical tank and user-programmable linearisation curve. Programming your own linearisation curve is only possible with the PC and the adjustment program VVO. A user-programmable curve, however, can be transferred from the PC to the electronics of the instrument and called up under this menu item.
0 % correspond
Here, you enter the value which should be displayed when 0 % level is reached, e.g. 20 l.
100 % correspond
Here, you enter the value which should be displayed when 100 % level is reached, e.g. 1200 l.
Integration time
In the menu item "
Integration time
", you can
set a delay (in seconds) for the signal output.
4. Outputs
Under the menu " display and determine which parameter should be outputted by the sensor display
Outputs
", you can scale the
and the PA output.
Capacitive electrodes EL - Profibus PA 33
Menu schematic of the adjustment module MINICOM
1.
Para­meter
Sensor optimi­ze Mode 1
Confi­gura­tion
Add’l func­tions
Info
Sensor Tag
Sensor type
CAP E34
Serial number
Softw. vers.
Sensor addr.
max. range
Setup
1)
Offset
Capac-
pF
itance
pF
pF/m
Adjust­ment
with medium
Min. adjust at %
XXX.X
2.
Max. adjust at %
XXX.X
Sensor Tag
w.out medium
Adjust­ment in
Elec-
Sensor
trode
addr.
type
No. 01
3.
Signal condit­ioning
Lin.
Sca­ling
0 % corres ponds
XXXX
1)
2)
Offset
Me­dium conduc­tive
DK value (Er)
correc­tion
pF
Elec­trode uncov­ered?
Offset
pF Now! OK?
0 %
100 %
at
at
pF
curve
Linear
100 % corres ponds
XXXX
pF
Integr ation time
Deci­mal point
888.8
0 s
Unit
%
Menu items only available with:
1)
Adjustment in m or ft
2)
Non-conductive medium
3)
Concentric tube or double rod electrode
34 Capacitive electrodes EL - Profibus PA
Setup
Outputs
4.
Simu­lation
Reset
Reset to de­fault
Reset to de­fault OK
Reset Now!
OK ?
Total reset
Total reset
OK?
Reset Now!
OK ?
Lan­guage
Eng­lish
DK measure­ment
Immers­ion
DK meas. Now! OK?
DK value (Er)
3)
3)
3)
3)
PA output
Failure
prop.
mode
to
Error
Vol u -
value
me
With these keys you move in the menu field to the left, to the right, top and bottom.
Sensor dis­play
prop. to
Per­cent
ESC
OK
Simu­lation Now!
Simu­lation
in V%
XXX.X
OK?
act. dist. m
X,XX
Fast change
Yes
Bolt print figures are sensor or measured value information and cannot be modified in this posi­tion.
The menu items in white letters stand for figures which can be modified with the "+" or "–" key and saved with the "OK" key.
Capacitive electrodes EL - Profibus PA 35
Setup

4.3 Adjustment with VVO

If the oscillator CAP E34 PA Ex is installed, the electrode can also be adjusted via a PC with the adjustment software VVO (from ver­sion 2.60).
Before the sensors can be operated and adjusted with the adjustment program VVO (VEGA Visual Operating) via Profibus PA, they must be integrated into the Profibus system. Address the sensors first (chapter
4.1 Adjustment media) and then connect them to the PA segment. With the attached GSD file (instrument master file), you then integrate the sensors into your system. The GSD file can be integrated by means of an appropriate tool, e.g. Siematic Manager, into your system.
To connect the PC to the DP bus, a standard RS 485-DTE interface cable (Data Terminal equipment) is necessary. With the cable, you connect the DP interface card with the bus or with the segment coupler.
PCPC
PC
PCPC
Screen 1 1 Screen – – 2 2 M24 RxD/TxD-PRxD/TxD-P
RxD/TxD-P
RxD/TxD-PRxD/TxD-P
33
3
33 – – 4 4 CNTR-P GNDGND
GND
GNDGND
55
5
55 – – 6 6 VP – – 7 7 P24 RxD/TxD-NRxD/TxD-N
RxD/TxD-N
RxD/TxD-NRxD/TxD-N
88
8
88 – – 9 9 CNTR-P
Profibus-DP DP-Bus (in brackets interface card the PIN number of the P+F
BUSBUS
BUS
BUSBUS
33
RxD/TxD-PRxD/TxD-P
3
RxD/TxD-P (40)
33
RxD/TxD-PRxD/TxD-P
55
GNDGND
5
GND (55)
55
GNDGND
88
RxD/TxD-NRxD/TxD-N
8
RxD/TxD-N (41)
88
RxD/TxD-NRxD/TxD-N
segment couplers)
VVO connection to the b us cable
You need a Profibus DP interface card (from Softing) in your PC in order to adjust the sensor with the adjustment software VVO. This card must be used to make the connec­tion between computer and Profibus DP cable (a direct connection or a connection to the Profibus PA cable is not possible). With it, the computer with VVO has the status "Mas­ter class 2".
Segment coupler
DP cable
VVO
Profibus DP interface card (Softing)
Sensor Sensor Sensor Sensor
VEGACONNECT 3
If the computer is connected to the Profibus DP cable or via VEGACONNECT 3, you can start VVO. When connecting via VEGACON­NECT 3, you need VVO version 2.80 or higher.
Profibus PA cable
36 Capacitive electrodes EL - Profibus PA
Setup
• Switch on the power supply of the con­nected sensor.
• Start the adjustment software VVO (VEGA Visual Operating) on your PC.
• In the entrance screen, you choose with the arrow keys or the mouse the item
ning
and click to OK. You should only
choose
Planning
if you are authorised to
Plan-
modify the instrument parameters. Other­wise choose
Operator
or
Maintenance
. In the window "User identification", you are asked for the name and the password.
• For setup ( name:
VEGA
Planning
VEGA
) you enter under
and under password also:
. It does not matter if you use capital
or small letters.
VVO recognises automatically the type of the connected sensor and immediately after­wards indicates with which sensor a connec­tion exists.
The following adjustment steps are de­scribed in their sequence and should be carried out in the same order during the initial setup.
For further information, see the operating instruction manual of the adjustment program VEGA Visual Operating (VVO).
First of all, click to "Profibus DP (for PA sen­sors". Then choose the Profibus card which is installed in your PC and click to OK. Then restart VVO. If you are connected via a VEGACONNECT 3, choose the option Standard. For a connection via VEGACONNECT 3, you need VVO version 2.80 or higher.
Sensor configuration
If you start VVO as described above, i.e. if you have connection to the Profibus, you will see the following presentation on your screen. Click to Configuration, then to Measuring system.
If you do not get a connection to the instru­ment, click to Configuration, choose Pro- gram and click to Communication .
If you have already connected several VEGA sensors to the bus, you select the sensor you want to configure - in this case the ca­pacitive electrode. The serial number and the bus address of the sensor will be displayed to ensure a clear identification. Through VEGACONNECT 3 you can access only one sensor.
Configuration
Display Diagnostics Instrument data Configuration Services Quit Help
Capacitive electrodes EL - Profibus PA 37
system
Measuring Measurement loop > Program >
Language User access Communication
Setup
Under the menu item "Configuration", you can choose the following functions:
• Measuring system
• Measurement loop
•Program
The electrode is already preset at VEGA. The measuring system must only be configured if the oscillator is exchanged.
Measuring system
In this window, you can choose the respec­tive electrode. You can find the electrode type on the type plate of the instrument.
Choose the correct electrode from the list, e.g. EL 21: 14 mm PTFE. In front of the listed electrodes you see the electrode type number. The number belonging to your ca­pacitive electrode is stated on the type plate (see also chapter "1.6 Type plate"). If your electrode is not included in the list, choose
´not configured´
confirm the setting.
. Push the Save button to
Profibus PA address
In this window, you can assign a bus ad­dress to the sensor in the field "Sensor ad­dress". This function is only available if the mini coding switch on the oscillator is set to >126.
Reset
In this menu window, you can carry out two different resets.
Standard reset (reset to default)
Adjustment 0 % - 0 pF
100 % - 3000 pF Adjustment unit pF Integration time 0 s Linearisation linear Sensor display scaled Scaling 0.0 % 100 % PA output volume Failure mode last value Sensor optimisationMode 1 (90°)
Total reset
With a total reset, all adjustments which can be carried out via VVO or MINICOM are reset. This includes, e.g., the following data:
- electrode type
- Tag name
- offset correction A total reset is necessary, e.g. if the oscilla­tors are installed on another electrode.
Information on the sensor ident number:
You do not have to modify the setting "VEGA specific No.". Only if you want to make sure the sensor can be replaced on the Profibus by a sensor from another manufacturer do you have to choose the option "Profile spe­cific No.".
38 Capacitive electrodes EL - Profibus PA
Setup
Configuration of measurement loops
Click to Configuration, point to Measure­ment loop, then click to Modify.
Now all instruments connected on the bus are listed (identifiable by means of bus ad­dress and serial number). Choose the sen­sor (the measurement loop) you want to use to create a measurement loop and click to
OK.
Measurement loop description
In this field, you can give your measurement loop a descriptive label, e.g. level - cleaning detergent. Up to 80 characters can be en­tered. The measurement loop description is not saved in the sensor, but in the PC.
Application
For capacitive electrodes,
ment
is permanently set and cannot be modi-
fied.
Program
In this menu item, you can modify settings of the program.
Level measure-
Language
Here you can change the program language. With OK you reach again the window "VEGA
Modify measurement loop configuration
In this window, you can more clearly specify your measurement loop.
Measurement loop no. (Sensor-TAG)
In this field, you can enter a measurement loop number, e.g. tank 15/3. Max. 16 charac­ters are available.
Capacitive electrodes EL - Profibus PA 39
Visual Operating".
User access
With this function, you can modify the user name and the password or deactivate the password enquiry.
Communication
Here you can determine the settings for data transmission.
Options
In this function, program settings such as backup, etc., are available.
Instrument data
Display Diagnostics Instrument data Configuration Services Quit Help
Under the menu item "Instrument data", you can choose the function:
• Parameter adjustment.
Parameter adjustment
Setup
From the window "Instrument data parameter adjustment", you can reach all submenus of the sensor. Just test the various buttons one after the other. You will always return to this window. Click to the button Adjustment to carry out the adjustment.
There you find the following functions:
- Adjustment
- Conditioning
- Outputs
- Sensor optimisation
- Additional functions
Sensor parameter adjustment
Click to Instrument data, then to Parameter adjustment.
Adjustment - Min/Max. adjustment
With this function, you can adjust the capaci­tive electrode (Min/Max. adjustment). You can choose if you want to carry out the adjust­ment with or without medium.
In the windows "Min. adjustment" and "Max. adjustment" (with medium) as well as "Min/ Max. adjustment" (without medium) it is pos­sible to enter an offset in percent. By doing this, the initial value is shifted by the entered
Choose the measurement loop on which you want to carry out the parameter adjustment and click to OK.
40 Capacitive electrodes EL - Profibus PA
percentage value. A modification of this offset is normally not necessary.
Setup
Note! Do not confuse this offset with the offset correction (Instrument data/Parameter adjustment/Additional functions).
Adjustment with medium
Min. adjustment
There must be min. level in the vessel. If you push the button entered percentage value.
For example, if you know that your vessel is filled to 10 %, this can be entered under min. adjustment.
Max. adjustment
There must be max. level in the vessel. If you push the button entered percentage value. You can also assign a certain percentage value to a known level.
For example, if you know that your vessel is filled to 90 %, this can be entered under max. adjustment.
Adjustment without medium (dry adjust­ment)
Save
, the output is set to the
Save
, the output is set to the
Under certain circumstances, an adjustment can also be carried out without medium with the help of the processing software VVO.
The requirements for an adjustment in m are:
In conductive medium:
- electrode is fully insulated
The following criteria apply to the conductiv­ity:
Mode 1 (90°) Mode 2 (45°) w. conc. tube > 50 µS/cm >150 µS/cm
w.o.conc.tube >100 µS/cm >300 µS/cm
In non-conductive medium:
- concentric tube electrode
- you know the DK value of the medium
If you already know the electrode capaci­tance of another similar measurement loop (same electrode, same installation conditions, same medium), you can also carry out a dry adjustment in pF.
With an unimmersed electrode, first carry out an offset correction (only for adjustment in m:
Instrument data - Parameter adjustment ­Additional functions Adjustment without medium
You can choose if you want to carry out the adjustment in pF or in m. If you have chosen adjustment in m, you have to enter whether your medium is conductive or non-conduc­tive. If the medium is non-conductive, you also have to enter the DK value.
The above table shows you the conductivity criteria applicable to capacitive measurement and whether you have to choose or
Non-conductive
). Go again to the function
.
Conductive
.
If you have chosen m, you can assign a corresponding value in m to the percentage value. Example 0 % = 0.2 m 100 % = 3.2 m
Capacitive electrodes EL - Profibus PA 41
If you have chosen pF, you can assign a corresponding value in Picofarad to the per­centage value. Example 0 % = 97.2 pF 100 % = 1428.0 pF
The values for the adjustment will be ac­cepted by pushing the button
OK
.
Conditioning
The following functions are available in this window:
- Scaling
- Linearisation
- Integration time
Setup
Click in the window "Conditioning" to
Linearisation.
Scaling
With this button, a message appears telling you that for Profibus sensors, this function can be found in the menu window "
data parameter adjustment
item "
Output
". Carry out the scaling there.
" under the menu
Instrument
With oscillator CAP E34 PA Ex you can also enter your own linearisation curve for special vessels, e.g. cylindrical vessel with large funnel-shaped outlet, under "user-program­mable curve".
Push the button
Edit.
You can enter the value
pairs (percentage value - volume value) for
Linearisation
32 index markers.
In this window, you can linearise non-linear vessels, such as e.g. a cylindrical tank. The following linearisation curves are available:
- linear
- cylindrical tank
- spherical tank
- user-programmable curve
Click in the window "Instrument data param­eter adjustment" to Conditioning.
42 Capacitive electrodes EL - Profibus PA
Setup
Push the button linearisation curves.
In the window "Linearisation", you can choose the linearisation curves of a horizontal cylin­drical tank or a spherical tank. If you choose a user-programmable curve, you can then click to Edit and open the program "Tank Calculation". With this program, you can cal­culate the curves of different tank forms (see the manual "VEGA Visual Operating"). After the selection of the curve, click to OK.
Transfer
, to call up further
Sensor parameter adjustment – Inte­gration time
Choose this function if you want to set an integration time (damping).
The range for the integration time is between 0 and 600 seconds. As a standard feature, an integration time of zero seconds is preset.
After the set integration time has expired, 63 % of the measured value change remain on the output.
In the window "Instrument data parameter adjustment", you click to Conditioning.
Click in the window "Conditioning" to Integra- tion time.
In the window "Integration time" you can enter a time of max. 600 seconds. Then click to
OK.
Sensor parameter adjustment – Out­puts
In this window, you can change the preset values for the PA output and the measured value display. Push the button fer the modified values.
Save
to trans-
Click in the window "Instrument data param­eter adjustment" to Outputs.
Capacitive electrodes EL - Profibus PA 43
Setup
Click in the window "Outputs" to Profibus output.
In the window "Profibus output" you can de­termine the options for Profibus output (acc. to Profibus PA instrument profile). Confirm the settings with OK. The values under change the Profibus output and the value on any display that may be connected. To use the values for a display, "scaled" must be chosen under "Display of measured value".
"Scaling"
Sensor parameter adjustment – Sensor optimisation
In this window, the mode of the sensor can be set. With this, you modify the phase angle of the phase selective admittance processing (PSA). "Mode 1" is preset.
Mode 1 = Phase angle 90°
Mode 1 is a pure capacitance measurement, the ohmic resistance is not taken into account in the measuring result. Application:
- standard adjustment
- non-conductive liquids up to approx. 50 µS
- compensation of conductance changes in liquids
- generally with partly insulated electrodes in liquids
- non-conductive solids without moisture content
- with insufficient product grounding
- for electrodes in conjunction with a concen­tric tube
- in non-conductive vessels with exterior grounding plate outside
Mode 2 = Phase angle 45°
The capacitance and the ohmic conductance are measured separately, the capacitance value is corrected by calculating it with the ohmic conductance, so that measuring errors caused by conductive buildup or changes of moisture content can be compensated. Application:
- highly conductive products
- adhesive, conductive products
- solids with fluctuating moisture content
When clicking "Display of measured value", you can choose twhat should be displayed on the sensor display, e.g. level, capacitance or scaled value.
44 Capacitive electrodes EL - Profibus PA
Setup
Sensor parameter adjustment – Offset correction
This function is necessary if an adjustment should be carried out in m. With this function, the initial capacitance is saved in the elec­tronics. A prerequisite is that the electrode is installed in the vessel and completely unimmersed.
An offset correction is only necessary if ad­justment without medium is to be carried out. For electrodes with concentric tube, this function is already carried out by VEGA, since the vessel does not influence the meas­urement. The offset correction must only be repeated of the oscillator is exchanged.
Click in the window "Instrument data param­eter adjustment" to Additional functions, then in the window "Additional functions" to Offset correction.
Confirm the question with OK, if the condi­tions are met.
Click in the window "Offset correction" to
Correct.
Display
Display Diagnostics Instrument data Configuration Services Quit Help
Display of measured value
Under the menu item "Display", you can choose the function:
• Display of measured value.
Display of measured value
In this window, the actual measured value is displayed digitally and as bar graph. The upper bar shows the current measured value in percent, pF or scaled in the set unit.
The actual values of the measurement loops can be displayed from the main menu. Click to "Display" and then to "Display of measured value".
Capacitive electrodes EL - Profibus PA 45
Choose the requested measurement loop and click to OK.
In the window "Display of measured value" you can choose the unit in which the meas­ured value should be displayed. By clicking to Quit you return to the window "VEGA Visual Operating".
Setup
Diagnostics
Display Diagnostics Instrument data Configuration Services Quit Help
Status Si
mulation
Under the menu item "Diagnostics" you can choose the following functions.
• Status
• Simulation
Status
This function is not available.
Simulation
With this function, you can simulate a certain level. Push the start button to activate the simulation. Shift the scrollbar to the request PA output value or enter a specific PA output value. The scaled display bar will be modified accordingly.
To interrupt the simulation, push the button
Stop
.
The simulation will be stopped automatically if you close the window
To check the outputs of the electrodes and connected instruments or components, you have to click to Diagnostics, then to Simula- tion.
Simulation
.
46 Capacitive electrodes EL - Profibus PA
Setup
Select the requested measurement loop and click to OK.
In the window "Simulation of outputs", you click to Start, to start the simulation. With the buttons "<–" and "–>" (or with the slide switch) you can set values between -10 % and 110 %. To quit the simulation, click to Stop.
Under the menu item "Services", you can select the following functions:
• Print
•View
• Backup
• Restore configuration
• Edit database
• User level
• Connection
• Modem
Print
With this function, you can print a list of all sensor settings.
View
With this function, you can display on screen the list of sensor settings.
Backup - Sensors
In this window, you can save the settings of the sensor in a database.
Restore configuration - Sensors
In this window, you can call up the sensor settings from the database.
Note:
The simulation mode will not be stopped automatically, it remains active until you
Edit database
In this window, you can edit the saved sensor data in the database.
switch it off!
User level
Services
Display Diagnostics Instrument data Configuration Services Quit Help
Capacitive electrodes EL - Profibus PA 47
Print View
Backup > Restore configuration > Edit database
User level Connection F8 Modem F7
Signal conditioning Sensors
instruments
In this window, you can select the user levels. The following user levels are available.
- Operator
- Maintenance
- Planning
Operator
This user level is for the operator. In this level, you can have the measured value displayed and the sensor data printed. There is no password required for this level.
Maintenance
If you select the level Maintenance, all func­tions are available except configuration. The password for this level is: VEGA.
Planning
If you select the level Planning, all functions are accessible. The passwords for the planning level are: VEGA - VEGA.
Connection
With this function you can restart connection set-up. This function can also be activated with the key F8.
Modem
This window enables adjustments for data transfer via modem. This function can also be activated with the key F7. The modem does not have to meet special requirements.
Setup
Bits & Bytes
In this window, you can call up the current measured values of all connected sensors with the function "Display of measured value". The function Live-List shows all data of the connected sensors.
48 Capacitive electrodes EL - Profibus PA

Diagnosis

5 Diagnostics

5.1 Simulation

To simulate a certain filling level, you can call up the function Simulation in the oscillator or in the adjustment program VVO. You hereby simulate a certain PA output value. Remem­ber that connected instruments, such as e.g. a PLC, react according to their settings and that alarms may be triggered or system functions activated.

5.2 Maintenance

The instrument is maintenance-free.

5.3 Repair

Repair work requires opening the instrument to correct defects. For safety and warranty reasons, all internal work on the instruments, apart from that involved in normal installation and connection, must be carried out only by qualified VEGA personnel.
In case of a defect, please return the affected instrument with a short description of the error to our repair department.
Failures are short-term malfunctions of the instrument which are caused by wrong ad­justment or defects in the sensor or the con­nection cables.
Failures, their possible causes and their rectification are stated under "5.4 Failure rectification".
Capacitive electrodes EL - Profibus PA 49
Diagnosis

5.4 Failure rectification

Failure Measure, failure rectification
Sensor defective, Test the internal connections: measurement does • Loosen the 4 screws of the housing cover with a Phillips screwdriver
not react to level changes • Loosen the two small screws with a Phillips screwdriver and pull the no display of oscillator out of the housing. measured values
and remove the housing cover.
• Measure with an ohmmeter (range MOhm) the resistance values between the following contacts:
6
CM
1 M
CM - Capacitor
Screen
3
4
Contact 4 and middle pin (1)
The resistance must be 1 MOhm. If the resistance is less, it indicates moisture in the housing or a defect in the electrode insulation. A possible reason could be a non-insulated electrode used in a conductive (moist) medium. If the resistance is higher or if the connection is interrupted, the reason is usually a bonding failure in the adapter plate or a resistor defective due to strong electrostatic discharge. In both cases, return the sensor to VEGA for repair.
50 Capacitive electrodes EL - Profibus PA
Diagnosis
Fail ure Measure, failure rectification
Contact 4 and vessel
The connection between contact 4 and the metal vessel (not instrument hexagon or electrode flange) should be as good as possible. Measure with an ohmmeter (range very low) the resistance value between con­tact 4 and the vessel.
• Short-circuit (0 … 3 Ohm, optimum connection)
• Resistance > 3 Ohm
- corrosion on the mounting thread or flange
- possibly the mounting thread was covered with Teflon tape or a
similar material Check the connection to the vessel. If there is no connection, you can connect a cable from the external grounding terminal to the vessel. Make sure that coated flanges are always connected via the grounding terminal to the vessel.
Contact 4 and 6
Values > 3 Ohms indicate a defect.
If you cannot find a failure in the sensor, then replace the oscillator with a similar type (if available) or send the sensor for repair to VEGA. After replacement of the electronics, a fresh adjustment must be carried out (see "4 Setup").
Error message Invalid measured value
Invalid measured The current measured value is outside the valid measuring range. value E030 Causes can be e.g. considerable changes of the DK value or PA status 81, 82 incorrect settings.
Change the current measuring conditions.
Error message Instrument error Instrument error An error in the sensor was found.
E068 PA status 16, 68, 72 Possible causes are:
- supply voltage not within tolerance
- defective oscillator
- damage to the electrode or to the insulation
- electrode short-circuited against vessel wall
• Separate the sensor from power supply and then reconnect (cold
start).
• If the error message does not extinguish, check the sensor cable
and the correct power supply.
• If the error message does not extinguish, check the sensor for vis-
ible damage.
If you cannot find any faults, please call our service department.
Capacitive electrodes EL - Profibus PA 51
Fail ure Measure, failure rectification
Error messageError message
Error message Increase the span between empty and full adjustment.
Error messageError message Adjustment span too small E017
Error messageError message
Error message Instrument error
Error messageError message General An error in the sensor was found. instrument error E034 Possible causes are : PA status 12 - supply voltage not within tolerance
- defective oscillator
- damage to the electrode or to the insulation
- electrode short-circuited against vessel wall
• Separate the sensor from power supply and then reconnect (cold start).
• If the error message does not extinguish, check the sensor cable and the power supply (correct voltage).
• If the error message does not extinguish, check the sensor for vis­ible damage. If you cannot find any faults, please call our service department.
Notes
52 Capacitive electrodes EL - Profibus PA
Notes
Capacitive electrodes EL - Profibus PA 53
Notes
54 Capacitive electrodes EL - Profibus PA
Notes
Capacitive electrodes EL - Profibus PA 55
VEGA Grieshaber KG Am Hohenstein 113 D-77761 Schiltach Phone (07836) 50-0 Fax (07836) 50-201 E-Mail info@de.vega.com
www.vega.com
ISO 9001
All statements relating to scope of delivery, application, practical use and operating conditions of the sensors and processing systems correspond to the latest information at the time of printing.
Technical data subject to alterations
2.25 712 / August 2001
Loading...