VEGA Capacitive electrodes User Manual

Operating Instruction
Capacitive electrodes EK … with signal conditioning instrument
Level and Pressure

Safety information

Safety information
The described module must only be installed and operated as described in this operating instruction. Please note that other action can cause damage for which VEGA does not take responsibility.
2 Capacitive electrodes EK with signal conditioning instrument

Note Ex-area

Please note the attached approval documents (yellow binder) and especially the included safety data sheet.
Contents
Contents
Safety information ........................................................................ 2
Note Ex-area ................................................................................ 2
1 Product description
1.1 Function and configuration .................................................. 4
1.2 Types and versions ............................................................. 6
1.3 Technical data ....................................................................... 8
1.4 Approvals ........................................................................... 14
1.5 Dimensions ......................................................................... 15
1.6 Type plate ........................................................................... 17
2 Mounting
2.1 Mounting instructions ......................................................... 18
3 Electrical connection
3.1 Connection instructions ..................................................... 23
3.2 Wiring plan .......................................................................... 24
4 Set-up
4.1 General adjustment ........................................................... 27
4.2 Level detection ................................................................... 28
4.3 Continuous level measurement ......................................... 29
5 Diagnosis
5.1 Simulation ............................................................................ 31
5.2 Maintenance ....................................................................... 31
5.3 Repair .................................................................................. 31
5.4 Failure removal ................................................................... 32
Capacitive electrodes EK with signal conditioning instrument 3

1 Product description

1.1 Function and configuration

Capacitive electrodes series EK detect levels of virtually any medium, unaffected whether liquids, powders, granules or pastes. This is also valid for adhesive mediums.
The electrode measures also the level ca­pacitance and the ohmic resistance (admit­tance processing). Hence also problematic mediums and solids with fluctuating humidity contents can be detected.
By the use of screening tubes and screen segments, inactive areas can be provided on the probe where pollution, condensation or permanent build-up do not influence the measuring result.
Measuring principle
Electrode, medium and vessel wall form an electrical capacitor.
Product description
Electrode and vessel wall are the capacitor plates. The medium is the dielectricum. Due to the higher dielectric constant figure (DK­value) of the medium against air, the capaci­tance of the capacitor increases with raising covering of the electrode.
Fig. 1.2 Capacitance change with covered electrode
The capacitance of the capacitor is mainly influenced by three factors:
- distance of the electrode plates (a)
- size of the electrode plates (b)
The capacitance change is converted by the oscillator into a level proportional, floating current in the range of 4 … 20 mA or into a switching command.
- kind of dielectricum between the electro­des (c)
Continuous level measurement
With the continuous level measurement, the appropriate level is continuously detected and converted into a level proportional signal which is either directly indicated or further
b
c
Fig. 1.1 Plate capacitor (schematic demon­stration)
4 Capacitive electrodes EK with signal conditioning instrument
processed.
You require a capacitive electrode series EK with oscillator and a signal conditioning in­strument VEGAMET, converting the propor­tional current of the oscillator into standardised current and voltage signals.
The continuous measurement requires a constant dielectric constant figure er, i.e. if possible the medium should have steady features.
Product description
The floating signal of the electrode electronics is in the range 4 … 20 mA and can be there­fore connected without additional potential separation to other processing systems such as e.g. VEGALOG.
In addition to the continuous measurement, also levels can be detected (VEGAMET or VEGAMET + VEGASEL).
Level detection
Level switches should signal when certain levels are reached, e.g. max. or min. levels. These levels are detected at a fixed point and converted into a switching command.
For level detection capacitive electrodes EK with appropriate signal conditioning instru­ments VEGATOR are available. A switching command can be either triggered when the electrode is covered or uncovered.
Capacitive electrodes EK with signal conditioning instrument 5
Product description
1.2 Types and versions
*)
Type
Version 11 21 24 26 31 42
Continuous
Level detection
Partly insulated
Fully insulated
Oscillators
E 14
E 15
E 15 Ex
E 17
E 17 Ex
E 18 ••••••
E 18 Ex
Approvals
CENELEC EEx ia IIC T6
PTB-Zone 0 EEx ia IIC T6
Overfill protection acc. to WHG
German Lloyd
Lloyds Register of Shipping
American Bureau of Shipping
Bureau Veritas
RINA
1)
1)
1)
1)
1)
EK EK EK EK EK EK
• •••••
• •••••
• •••••
• •••••
• •••••
Mechanical connection
G 3/4 A • •••••
G 1 A
3/4" NPT
1“ NPT
Flange plated
Electrode material
Steel
StSt
Isolating material
5)
2)
2)
3)
4)
3)
PTFE
FEP
PE
6 Capacitive electrodes EK with signal conditioning instrument
Product description
*)
Version 11 21 24 26 31 42
Type
Concentric tube
StSt
Screening tube (option)
StSt
Temperature adapter (option)
StSt
Housing material
Plastic (IP 66)
Aluminium - plastic coated (IP 66 and 67)
Others
Bending of electrode
6)
*) All instrument types also Ex0
1) applied
2) 1.4435
3) 1.4571
4) 1.4401
5) For electrodes certified for Ex-Zone 0, only PTFE and FEP are approved as isolating material.
6) Bending max. 90°
7) EK 21 only with PTFE with 3,2 mm isolation thickness
EK EK EK EK EK EK
••
7)
Capacitive electrodes EK with signal conditioning instrument 7
Product description

1.3 Technical data

Housing
Housing material plastic PBT (Polyester) or Aluminium
Protection
- plastic housing IP 66
- Aluminium housing IP 66 and 67 (meets both protections) Cable entry 1 pce. M20 x 1,5 Terminals for max. 1,5 mm2 cross-section area of
Mechanical connection
Material 1.4435 (316 L) Thread G 3/4 A or 3/4“ NPT
Flange flange versions, plated
Electrode
Material EK 11 1.4435 (316 L)
Length
- rod max. 3 m
- cable max. 20 m Isolation see "Isolating materials" Max. tensile strength
- EK 31 3 KN
- EK 42 3 KN
plastic coated
conductor
G 1 A or 1“ NPT
EK 21 steel (St 37), 1.4435 (316 L) EK 31 1.4401 (316 L) EK 24, 42 1.4571 (316 L)
Ambient conditions
Ambient temperature on the housing -40°C … +80°C Medium temperature see "Product temperature and operating
Storage and transport temperature -40°C … +80°C Operating pressure see "Product temperature and operating
Oscillators E14, E15, E17, E18
Protection class II Overvoltage category III Meas. frequency see table on the following page Capacitance ranges see table on the following page Supply voltage 12 … 36 V DC (powered by signal conditioning
Potential separation min. 500 V DC (except E14)
8 Capacitive electrodes EK with signal conditioning instrument
pressure"
pressure"
instrument)
Product description
For Ex-applications note the permissible electrical connection values stated in the certificate.
Load in W
Accessory
Straining spring of 1.4571 (EK 42, EK 52, EK 53)
- load approx. 185 mm (stressed)
- tensile load approx. 200 N
Weight
Basic weight (e.g. EK 24) approx. 0,8 kg Rod weight ø 6 mm - 0,23 kg/m
ø 10 mm - 0,62 kg/m
Cable weight (EK 31) approx. 40 g/m
Supply voltage in V
Capacitive electrodes EK with signal conditioning instrument 9
Product description
Oscillators in two-wire technology for capacitive electrodes EK
Typ e Application Meas. range Frequency Signal cond.instr.
E14 Level detection in general I: 0 … 25 pF 400 kHz VEGATOR
E15 Level detection in general I: 0 … 25 pF 400 kHz VEGATOR
with potential separation II: 0 … 100 pF VEGALOG
E15 Ex as E15, however for use in I: 0 … 25 pF 400 kHz VEGATOR Ex
hazardous areas acc. to II: 0 … 100 pF VEGALOG CENELEC, PTB zone 0 as well as III: 0 … 400 pF zone 1 and StEx zone 10 and as part of an overfill protection acc. to WHG, (VbF)
E17 Continuous level measurement in I: 0 … 120 pF 40 kHz VEGAMET
general or level detection with II: 0 … 600 pF VEGALOG potential separation III: 0 … 3000 pF (VEGATOR)
E17 Ex as E17, however for use in I: 0 … 120 pF 40 kHz VEGAMET Ex
hazardous areas acc. to II: 0 … 600 pF VEGALOG CENELEC, PTB zone 0 as well as III: 0 … 3000 pF (VEGATOR Ex) zone 1 and StEx zone 10 and as part of an overfill protection acc. to WHG, (VbF)
2)
E18
Continuous level measurement I: 0 … 120 pF 470 kHz VEGAMET or level detection with potential II: 0 … 600 pF VEGALOG separation acc. to the principle of III: 0 … 3000 pF (VEGATOR) phase selective admittance proces­sing especially for adhesive mediums and for the use in solids with varying humidity
2)
E18 Ex
as E18, however for the use in I: 0 … 120 pF 470 kHz VEGAMET Ex hazardous areas acc. to II: 0 … 600 pF VEGALOG CENELEC, PTB zone 0 as well as III: 0 … 3000 pF (VEGATOR Ex) zone 1 and StEx zone 10 and as part of an overfill protection acc. to WHG, (VbF)
1)
with safety barrier
2)
see following page
II: 0 … 100 pF VEGALOG III: 0 … 400 pF
III: 0 … 400 pF
1)
1)
1)
10 Capacitive electrodes EK with signal conditioning instrument
Product description
Oscillator E18
The oscillator E18 with the patented process­ing (phase selective admittance processing) extends the application range of capacitive level measurement technology.
In conjunction with the fully insulated rod electrode EK 24 the oscillator E18 compen­sates even very conductive build-up.
Mounted in an individual rod or cable elec­trode type EK, E18 ensures also the exact measurement in solids with varying humidity contents.
The oscillator E18 processes the measuring currents according to their phase position. Measuring current with a defined phase shifting as they occur with build-up or humid­ity changes are filtered out.
Humidity change
A humidity change in solids causes a change of the dielectric constant figure (er). In parallel the ohmic value of the medium changes. Due to the change also a phase shifting of the measuring currents is caused.
Level measured value
+
1)
0 %
5 %
10 %
1) Actual value level
Fig. 1.3 Humidity change
15 % vol.
fully insulated electrode
Solid humidity (%)
partly insu­lated elec­trode
With a capacitive measurement already low­est humidity changes cause measuring er­rors. Typical products are e.g. sand, aggregate in the cement industry, hop or plastic granules (after drying).
When using the oscillator E18 humidity changes of 15 % vol. do not influence the accuracy of the measurement. Even layering of product with different humidity does not play a role for the measuring accuracy.
When the humidity contents exceeds 15 % vol., fully and partly insulated electrodes react differently (see also "Fig. 1.3 Humidity change“). Whereby the measured value on fully insulated electrodes raises with steady level, the measured value on partly insulated electrodes drops.
Capacitive electrodes EK with signal conditioning instrument 11
Product description
Product temperature and operating pressure
1)
The figures in the tables relate to the pictures on this page. The statements on pressure are valid for screw connections G 3/4 A,
3
/4“NPT, G 1 A, 1“NPT. With flange versions you have to note their nominal pressure. All electrodes are also suitable for vacuum (-1 bar).
For electrodes certified for Ex-Zone 0 only PTFE and FEP are approved as isolating material acc. to ATEX II 1/2 G EEx ia II C T6.
Mechanical connection, 1.4435 (316 L)
Isolation
Electrode type
EK 11 1 3 -
EK 21 1 3 -
EK 21 with flange - 2 -
EK 24 - - 2
EK 26 - 3 -
EK 26 with flange - 2 -
EK 31 1 3 -
EK 42 - 2 -
PE
PTFE
FEP
bar
40
0–50 100
bar
40
25
0–50 100
EK 21 to 16 bar
EK 42: to 16 bar EK 24: from 100°C 6 bar, max. 150°C for 30 mins.
°C
Temperature adapter
1
2
3
200 °C
1) For Ex-applications the permissible temperatures and pressures stated in the certificate should be noted. Additionally note the table on the following page.
12 Capacitive electrodes EK with signal conditioning instrument
Product description
Electronics temperature
The following product and ambient tempera­tures must be maintained so that the limit temperature on the electronics is not ex­ceeded.
The stated values are obligatory for applications in hazardous areas. Note
for these applications also the appro­priate legal documents (test reports, test certificates, type approvals and conformity certificates).
Temperature class T6
Product temperature -40°C … +60°C Ambient temperature -40°C … +60°C
Temperature class T5
Product temperature -40°C … +75°C Ambient temperature -40°C … +75°C
Temperature classes T4 … T1 (or no Ex)
Without temperature adapter
- product temperature -40°C … +100°C
- ambient temperature -40°C … +80°C
With temperature adapter
Plastic housing Metal housing
Product temperature -40°C … 180°C 200°C 150°C 175°C 200°C
Ambient temperature
1) Ambient temperature on the oscillator
Capacitive electrodes EK with signal conditioning instrument 13
1)
-40°C … 80°C 75°C 80°C 69°C 58°C
Product description

1.4 Approvals

Explosion protection
Only certified capacitive electrodes EK**Ex 0 must be used in hazardous areas with com­bustible gases, vapours or fog. Capacitive electrodes EK**Ex 0 are suitable for the use in hazardous areas of zone 1 and zone 0. Proof for the explosion protection of these instruments is the EC-type approval and the conformity certificate possibly with national zone 0 - annex. These documents are generally attached to the instrument. When the capacitive electrodes are mounted or operated in hazardous areas, the Ex­installation regulations must be noted. The information and regulations of the sup­plied certificates (EC-type approval, con­formity certificate) of the capacitive electrodes as well as of the appropriate in­strument (signal conditioning instrument, separator, safety barrier) must be noted.
• The mounting of Ex-systems must be gen­erally carried out by skilled staff.
• The capacitive electrodes must be pow­ered by an intrinsically safe circuit; the permissible electrical values are stated in the appropriate certificate.
• Capacitive electrodes with electrostatically chargeable plastic parts are provided with a warning label informing about measures which must be taken to avoid dangers caused by electrostatic discharges. Note the contents of the warning label.
• The explosion protection of the instrument used is only ensured when the limit tem­peratures stated in the certificate are not exceeded.
ˇ In case of danger due to oscillation or
vibration, the appropriate parts of the ca­pacitive electrodes must be secured.
• After shortening of the electrode cable, it must be noted that the weight is sufficiently secured by means of pins.
WHG
The electrodes EK … Ex0 are also approved as part of an overfill protection for vessels storing water endangering liquids.
Ship approvals
For the use on ships, type approval certifi­cates are available of several ship classifica­tion authorities (GL, LRS, ABS, BV, RINA).
For electrodes certified for Ex-Zone 0, only PTFE and FEP are permitted as isolating materials acc. to ATEX II 1/2 G EEx ia IIC T6.
CE-approval
The capacitive electrodes EK meet the pro­tective regulations of EMVG (89/336/EWG) and NSR (72/23/EWG). The conformity has been judged acc. to the following standards: EMVG Emission EN 50 081 - 1
Susceptibility EN 50 082 - 2
NSR EN 61 010 - 1
Zone 2
According to DIN VDE 0165, instruments can be used in hazardous areas of zone 2 with­out approval; they must meet the require­ments in section 6.3 of this VDE. The compliance of the instruments with these requirements is confirmed by Messrs. VEGA in a manufacturer declaration.
14 Capacitive electrodes EK with signal conditioning instrument
Product description

1.5 Dimensions

Dimensions of the capacitive electrodes type EK
Type EK 11 (partly insu­lated)
32,5
85
EK 11 with concentric tube
85
32,5
Type EK 21 (fully insulated)
85
32,5
M20x1,5
SW 41
G 3/4 A
A
B
L (min. 100 mm, max. 3000 mm)
EK 21 with concentric tube
85
32,5
M20x1,5
SW 41
G 3/4 A
M20x1,5
SW 41
G 3/4 A
ø21,3
Type EK 24 (fully insulated, for adhesive products)
85
32,5
M20x1,5
SW 41
G 3/4 A
M20x1,5
SW 41
G 3/4 A
A
L (min. 100 mm, max. 3000 mm)
Type EK 26
32,5
85
M20x1,5
SW 41
G 3/4 A
ø21,3
L (min. 100 mm, max. 3000 mm)
Isolation A B
PE 2 ,0 mm 14 mm 10 mm PTFE 2,0 mm 10 mm 6 mm PTFE 2,0 mm 1 4 mm 10 mm
outer-ø rod-ø
ø14,4
L (min. 120 mm, max. 3000 mm)
ø14
Capacitive electrodes EK with signal conditioning instrument 15
Type EK 31 (partly insulated)
SW 41
G 3/4 A
L (min. 400 mm, max. 20000 mm)
Housing
Product description
Housing of plastic / Aluminium
Temperature adapter
SW 41
SW 41
of galvanized steel or 1.4571
Concentric tube
of 1.4435
ø21,3
Screening tube
ø21,3
of 1.4435 with closing cone of PP or PTFE
Closing cone
16 Capacitive electrodes EK with signal conditioning instrument
Product description

1.6 Type plate

Before mounting and electrical connection please check if you use the suitable instru­ment. Therefore note the type plate which is located as follows:
Type plate
The type plate contains important data re­quired for mounting and connection. The configuration and components of the type plate are hence explained in the following example.
Configuration of the type plate (example)
VEGA® EK 11
1
type EK11EXO.XGBVSTXXVKXX
see PTB no. EX-98.E.2085 EEx ia IIC T6 0032
2
PTB 98 ATEX 2086 II 1/2G EEx ia IIC T6
3
techn. data see document. / certificates 1998
protection: IP 66/67 Insp.
length: 400mm VVO: 02
4
Ord. no. 123456/000
®
D-77757 Schiltach
5
Z-65.13.XXX
7
1 Master data of the order no. 2 Ex-certification number
Explosion protection version - note the
information and regulations of the certificate 3 Data of the electronics / Approvals 4 No. of the order confirmation/Pos.-no. 5 Number of the electrode type 6 Serial number 7 Test mark when used as part of an overfill
protection for vessels storing water endan-
gering liquids - note the information and
regulations of the general type approval 8 Manufacturing year 9 Number of the test authority
9
8
6
Order code
Detailed information on the order code you will find in the "Product Information Capaciti­ve“ or in the "VEGA-Pricelist“.
Capacitive electrodes EK with signal conditioning instrument 17

2 Mounting

Mounting

2.1 Mounting instructions

General
Different mediums and requirements to the measurement require various installations. Hence the following instructions should be noted.
Length of the level electrode
Note when ordering the electrode, that the electrode must be sufficiently covered ac­cording to the electrical features of the me­dium (DK-value).
E.g. an electrode for level detection in oil
er ~ 2) requires a considerably higher cover­ing than in water (er ~ 81).
Lateral load
Note that the electrode is not subjected to strong lateral forces. Mount the electrode in a position in the vessel where no interfering influences such as e.g. stirrers, filling open­ing etc. occur. This is mainly valid for very long rod and cable electrodes.
Extraction forces
In case of strong extraction forces e.g. dur­ing filling or settling of solids, high tensile loads can be caused.
In these cases use for short measuring dis­tances a rod electrode, as a rod is generally more stable. If due to the length or the mount­ing position a cable electrode should be necessary, the electrode should not be strained, but only equipped with a gravity weight as then the cable can more easily follow the product movements. Note that the electrode cable does not touch the vessel wall.
Pressure
In case of gauge or low pressure in the ves­sel, the mounting boss must be sealed at the thread. Use the attached seal ring. Check if the seal ring is resistant against the medium.
Isolating measures such as e.g. the covering of the thread with Teflon tape can interrupt the electrical connection in case of metal vessels. Hence earth the electrode on the vessel.
Fig. 2.1 Lateral load
18 Capacitive electrodes EK with signal conditioning instrument
Mounting
Shortening of the electrode
The dimensions of fully insulated electrodes are fixed and must not be modified. Each modification will destroy the instrument.
Partly insulated cable or rod electrodes can be shortened afterwards. Note that due to the change of the basic capacitance also the switch point can change.
When the cable should be considerably shortened, it can happen that an adjustment of the electrode is not possible. A new com­pensation of the electrode will be necessary in this case. Note the serial number of the electrode and call one of our sales engineers.
The electrode is compensated to the appro­priate electrode length. For this reason you should already state in the order if you want to shorten the electrode.
Cable electrode EK 31 can be shortened afterwards (see fig. 2.2). Loosen the two pins on the gravity weight (hexagon) and remove the two pins. Pull the cable out of the gravity weight.
To avoid splicing of the steel cable (EK 31) during cutting, you have to tin the cable around the cutting position with a copper bit or strongly tighten the cable with a wire. Shorten the electrode cable with a metal cutting saw or a cutting-off wheel by the requested length.
Carry out the adjustment. The instruction is under "4.1 Adjustment“.
Fig. 2.2 Shortening of the electrode
Filling opening
Install the electrode such that it does not protrude directly into the filling stream. Should such an installation place be neces­sary, mount a suitable sheet above or in front of the electrode e.g. L 80 x 8 DIN 1028, etc. Fig. 2.3
Horizontal installation
The electrodes can be mounted horizontally to reach a very exact switch point with level detection. We recommend to mount the elec­trode approx. 20° inclined to the bottom, so that build-up can be avoided. Fig. 2.3
a.
b.
20°
Sheet
Fig. 2.3 Horizontal installation
Capacitive electrodes EK with signal conditioning instrument 19
Mounting
Humidity from outside
After installation turn the cable entries of horizontally mounted instruments to the bot­tom to avoid humidity ingress. The instrument housing is rotational by approx. 330°. In case of vertically installed electrodes, loop the connection line to the electrode housing to the bottom so that rain or condensation water can drain off.
This is mainly valid for mounting outside, in areas where humidity must be expected (e.g. by cleaning processes) or on cooled or heated vessels (see fig. 2.4).
Fig. 2.4 Humidity
Cable entries
When mounting outside, on cooled vessels or in humid areas where cleaning is made e.g. with steam or high pressure, the sealing of the cable entry is very important.
Use cable with a round cross-section area of conductor and tighten the cable entry. The cable entry is suitable for cable diameters of 5 mm to 9 mm.
Aluminium vessels
Use for Aluminium vessels an electrode with steel thread. The combination Aluminium on Aluminium should be avoided, as the thread "seizes“ when being screwed and cannot be loosened after some time without being dam­aged.
Metal vessels
Note that the mechanical connection of the electrode is electrically conductive connected with the vessel to ensure sufficient earth.
Use conductive seals such as e.g. copper, lead etc. Isolating measures such as cover­ing the thread with Teflon tape can interrupt the necessary electrical connection. In this case use the earth terminal on the housing to connect the electrode to the vessel wall.
Non-conductive vessels
In non-conductive vessels, e.g. plastic tanks, the second pole of the capacitor must be provided separately, e.g. by a concentric tube or the use of a double rod electrode.
When using a standard electrode, a suitable earth plate is necessary. Hence provide a possibly large earth plate, e.g. wire braiding laminated into the vessel wall or metal foil which is glued to the vessel. Connect the earth plate with the earth terminal on the housing.
Rod electrode
Mount the electrode such that the electrode protrudes into the vessel. When mounting in a tube or a socket, build-up can be caused which can influence the measurement. This is particularly the case with viscous or adhesive products. (see fig. 2.5)
20 Capacitive electrodes EK with signal conditioning instrument
Mounting
max. 80 mm
Lateral installation
With electrodes, delivering continuous meas­ured values, the electrode must only be in­stalled vertically. Should the installation from top not be possible, the electrodes can also be mounted laterally (see fig. 2.7)
If there are struts or a roof at the installation place of the electrode, you should check if a rod electrode of the requested length can be mounted. If a mounting of the rod electrode is not possible, use a cable electrode.
Fig. 2.5 Rod electrodes
Cable electrodes in solids
Dependent on the kind of solid and position or kind of filling, the cable electrode can "float“ despite of the gravity weight. The elec­trode (cable) is pushed by the solid to the vessel wall or to the top and wrong meas­ured values are caused. This should be avoided with the continuous level measure­ment.
In this case use a fixing weight to fasten the electrode.
When fixing the cable electrode avoid high tensile strengths. An appropriate fixing spring avoiding overloading of the cable is listed in our pricelist as accessory (see fig.
2.6)
Under the accessory in our pricelist you find a screening tube and a closing cone or a bent rod electrode by which the electrode can be also mounted laterally. Choose the length (l) of the screening tube such that no product bridges can be caused between cable and vessel wall and that the electrode cable cannot touch the vessel wall due to product movements. Use a fixing weight or a fixing insulator.
Fixing weight
Fig. 2.7 Continuous electrodes
Fig. 2.6 Cable electrode in solids
Capacitive electrodes EK with signal conditioning instrument 21
Mounting
Material cone
Note when installing the electrodes into the vessel, that material cones can be caused with solids which can change the switch point. We recommend to choose an installa­tion place where the electrode detects an
According to the position of the filling and emptying opening in the vessel, the electrode must be installed appropriately. To compen­sate the measuring errors caused by the material cone, you should install the electrode at a distance of d/6 from the vessel wall.
average value of the material cone.
Filling
Fig. 2.8 Material cone, filling and emptying centered
Filling Emptying
Emptying
3
2
1 Emptying 2 Filling opening
Fig. 2.9 Material cone, filling centered, emptying laterally
22 Capacitive electrodes EK with signal conditioning instrument
3 Capacitive
electrode
1
Electrical connection

3 Electrical connection

3.1 Connection instructions

Note
Switch off the power supply before starting connection work.
The electrical connection must be made de­pendent on the installed oscillator. The in­stalled electronics type is stated on the type plate of the oscillator. Connect the supply voltage according to the following connection diagrams.
Note
If strong electromagnetic interferences have to be expected, we recommend to use screened cable. The screening of the cable should only be earthed at one sensor end (electrode).
Generally connect the electrode with vessel ground (PA). For this purpose there is a terminal laterally on the housing. This connec­tion is additionally for the mass reference potential as well as to drain off electrostatic charges.
Skilled staff
Instruments operated in Ex-areas must only be mounted by skilled staff. They must note the mounting regulations and the supplied EC-type approvals and conformity certifi­cates.
When capacitive electrodes are mounted on vessels which must be protected according to TRbF 100 no. 8, para.1 against inflamma­tion due to lightning, they have to be equipped with the external overvoltage ar­rester type B 62-36 G or the internal overvolt­age protection unit type CB 2-36.
Capacitive electrodes EK with signal conditioning instrument 23

3.2 Wiring plan

The electrical connection of the sensor to the signal conditioning instrument is stated in the operating instruction of the appropriate signal conditioning instrument.
Note
The oscillator is independent of the electrode and can be exchanged locally.
As the oscillators have different characteris­tics (approx. 5 %), it can be necessary to readjust the signal conditioning instrument after electronics exchange.
Electrical connection
0/4 … 20 mA
DISBUS
24 Capacitive electrodes EK with signal conditioning instrument
Electrical connection
Capacitive electrode with external overvoltage protection unit
Vessel without cathodic corrosion protection
Ex-area
Electrode EK
optionally overvoltage arrester B62 - 36 G Ex 0 (metal housing) or B62 - 36 G Ex (plastic housing) (Li = 0,15 mH, Ci = 1,5 nF)
Outer isolation
1 2
B
E1
A1
Typ
E2
A2
B62-36G
Screen
Outer isolation
Screen
Not-Ex-area (control room)
optionally overvoltage arrester
A
E1
Typ
B62-36G
Signal conditioning instrument or safety barrier
A1 A2E2
min. 4 mm2 Cu
Vessel Zone 0
External earth terminal
to potential euqaliza­tion line
Earth/ PA-connection terminal
Note:
[A] Between control room and overvoltage protection system a suitable cable, if necessary a metal cover or
screen, should be used. Screen or metal cover - if necessary - must only be connected to the electrode side of the overvoltage arrester.
[B] Between overvoltage protection system and capacitive electrode a suitable cable with metal cover, screen
or a suitable cable with metal protection tube should be used (screen, metal cover or protection tube must be connected to the potential equalisation).
Test voltage of the cable A and B: £ 500 V AC
Capacitive electrodes EK with signal conditioning instrument 25
Vessel with cathodic corrosion protection
Ex-area
Electrode EK
External earth terminal
optionally overvoltage arrester B62 - 36 G Ex 0 (metal housing) or B62 - 36 G Ex (plastic housing) (Li = 0,15 mH, Ci = 1,5 nF)
Outer
B A
isolation
A1 A1
2
A2
B62-36G
Typ
Metal housing mounted earth-free
E1 E2
Screen
min. 4 mm2 Cu
Vessel Zone 0
Earth / PA-connection terminal
Note:
[A] and [B] see "Vessel without cathodic corrosion protection“ on
Cathodic corrosion protection 2 … 24 V (object voltage)
the previous page.
Outer isolation
Screen
Not-Ex-area (control room)
optionally overvoltage arrester
E11
Typ
B62-36G
Capacitive electrode with integral overvoltage protection module
Vessel with/without cathodic corrosion protection
Electrical connection
Signal conditioning instrument or safety barrier
A2E2
Ex-area
Not-Ex-area (control room)
optionally overvoltage arrester
B62-36G
A1
Typ
A2E2
Signal conditioning instrument or safety barrier
Note:
Electrode EK
[1]
1 2
Vessel Zone 0
Outer
A
isolation
E1
Screen
[A] Between control room and overvoltage
protection system a suitable cable, if
External earth terminal
Without cathodic corrosion protection: to potential equalisation line
Cathodic corrosion protection 2 … 24 V (object voltage)
necessary with screen or metal cover, should be used. Connect screen or metal cover - if necessary - only to the electrode side of the overvoltage protection system. With cathodic corrosion protection the cable must be provided with an outer isolation.
26 Capacitive electrodes EK with signal conditioning instrument
Set-up

4 Set-up

4.1 General adjustment

When you state with your order already the medium to be measured, the electrode will be adjusted by VEGA. The DK-value and the conductivity of the medium as well as the length of the electrode will be considered.
For the set-up the electrode must be ad­justed with the original medium.
For adjustment of the electrode, the housing cover must be opened. With the changeover switch on the oscillator you can choose the sensitivity range of the electrode.
E14, E15 E17, E18
Stage I 0…25 pF 0…120 pF (sensitive)
Stage II (standard) 0…100 pF 0…600 pF
Stage III (less sensitive) 0…400 pF 0…3000 pF
The instruction for adjustment or switch point adjustment is stated in the operating instruc­tion of the appropriate signal conditioning instrument.
Capacitive electrodes EK with signal conditioning instrument 27
Set-up

4.2 Level detection

Vertically installed electrodes
• Set the changeover switch on the oscillator of the electrode to stage I.
• Choose the requested mode (A - overfill protection, B - dry run protection) on the signal conditioning instrument.
• Fill the vessel to the requested level.
• Carry out the adjustment. Turn the potenti­ometer on the signal conditioning instru­ment (VEGATOR) very slowly until the signal lamp changes condition. Should the adjusted measuring range not be sufficient to find the switch point, you have to set the changeover switch to the next higher stage (II or III). Note also the switch point adjustment of the signal condi­tioning instrument.
Horizontally installed electrodes
• Set the changeover switch on the oscillator of the electrode to the appropriate stage. Note the following conditions: EK with E14, E15
- non-conductive medium range 1
- conductive medium to 0,25 m range 2
to 0,25 m range 3
- EK 26 with E18 range 1 If there is no other information, the oscilla­tors E14, E15 and E18 are preadjusted to range 1.
• Choose the requested mode (A - overfill protection, B - dry run protection) on the signal conditioning instrument.
• Make sure that the integration time is switched off (see operating instruction of the signal conditioning instrument).
• Empty the vessel or lower the level at least up to 100 mm below the electrode.
• Carry out the empty adjustment. Therefore turn the potentiometer on the signal condi­tioning instrument (VEGATOR) very slowly clockwise until the relay control lamp lights (mode A) or extinguishes (mode B).
- when the potentiometer on the signal
conditioning instrument is above the value 7, you should set the changeover switch to range 2.
• Note the position of the potentiometer.
• Fill the vessel until the electrode is com­pletely covered.
• Carry out the full adjustment. Turn the po­tentiometer on the signal conditioning in­strument (VEGATOR) very slowly clockwise until the relay control lamp lights (mode A) or extinguishes (mode B).
• Note the position of the potentiometer.
• Set the potentiometer to the average value of the two noted values.
28 Capacitive electrodes EK with signal conditioning instrument
Set-up

4.3 Continuous level measurement

When you state the medium with your order, the oscillator of the electrode is already pread­justed. In this case the medium is stated on the order confirmation.
• Choose with the changeover switch on the oscillator the stage according to the following schedule. Take the column corresponding to your medium and choose by means of the length of your electrode the suitable range. The stated lengths partly do not correspond to the actually available electrodes. When the medium has a DK-value which is between the values stated in the table, the max. permissible electrode length per measuring range must be averaged appropriately. In case of larger lengths or when there is nothing else stated in the table, choose range 3. If you are not sure, set the changeover switch always to the next higher stage.
Medium Non-conductive Non-conductive Conductive
EK-type and DK = 2 and DK = 10 or DK > 50
EK 11 0 - 5,5 m range 1 0 - 0,8 m range 1 ––
EK 11 with concentric tube 0 - 1,5 m range 1 0 - 0,15 m range 1 ––
EK 18 0 - 6 m range 1 0 - 1 m range 1 ––
EK 21 (2 mm-isolation) 0 - 5 m range 1 0 - 1,1 m range 1 0 - 0,25 m range 1
EK 21 (2 mm-isolation) 0 - 2 m range 1 0 - 0,5 m range 1 0 - 0,3 m range 1 with concentric tube – – 0,5 - 2,5 m range 2 0,3 - 1,7 m range 2
EK 21 (3,2 mm-isolation) 0 - 6 m range 1 0 - 1,5 m range 1 0 - 0,6 m range 1
EK 21 (3,2 mm-isolation) 0 - 3,5 m range 1 0 - 1 m range 1 0 - 0, m range 1 with concentric tube – – 1 - 5,5 m range 2 0,7 - 4 m range 2
EK 24 0 - 6 m range 1 0 - 1 m range 1 0 - 0,15 m range 1
EK 24 with concentric tube 0 - 2 m range 1 0 - 0,4 m range 1 0 - 0,15 m range 1
EK 26 0 - 6 m range 1 0 - 1,5 m range 1 0 - 0,5 m range 1
EK 31 0 - 6 m range 1 0 - 1 m range 1 ––
EK 42 0 - 9 m range 1 0 - 2 m range 1 0 - 0,8 m range 1
1)
Oscillator E17 from a conductivity > 10 µS, oscillator E18 from a conductivity > 100 µS.
–– 0,8 - 4,5 m range 2 ––
–– 0,15 - 0,9 m range 2 ––
–– 1 - 5,5 m range 2 – –
–– 1,1 - 6 m range 2 0,25 - 1,4 m range 2
–– 1,5 - 6 m range 2 0,6 - 3 m range 2
–– 1 - 5,5 m range 2 0,15 - 0,8 m range 2
–– 0,4 - 2 m range 2 0,15 - 0,9 m range 2
–– 1,5 - 6 m range 2 0,5 - 2,8 m range 2
6 - 30 m range 2 1 - 5 m range 2 – –
9 - 30 m range 2 2 - 10 m range 2 0,8 - 4,5 m range 2
1)
Capacitive electrodes EK with signal conditioning instrument 29
Set-up
Signal conditioning instrument VEGAMET series 300
When the electrode is not already pread­justed by VEGA, you have to choose the measuring range.
• Set the changeover switch on the capaci­tive electrode EK according to the table on the previous page.
• Turn the potentiometer for the full adjust­ment approx. 22 turns clockwise.
• Empty the vessel to the requested min. level.
• Turn the potentiometer for the empty ad­justment anticlockwise until the pointer of the indicating instrument is at 0.
• Fill the vessel to the requested max. level. The indication shows more than 100 %.
• Turn the potentiometer for the full adjust­ment anticlockwise until the pointer is at 100 %. When the indication cannot be set to 100 %, proceed as follows:
- when the indication does not reach
100 %, you have to set the changeover switch on the electrode one stage lower.
- when the indication exceeds 100 % and
cannot be reset, you have to set the changeover switch to the next higher
stage. In both cases you have to repeat the ad­justment.
Signal conditioning instrument VEGAMET 614, VEGAMET series 500
When the electrode is not already pread­justed by VEGA, you have to choose the measuring range.
• Set the changeover switch on the capaci­tive electrode EK according to the table on the previous page.
• Carry out the adjustment on the signal conditioning instrument (see operating instruction of the signal conditioning instru­ment: "Adjustment with medium“).
- when the signal conditioning instrument
displays the fault signal E014, this mes­sage is explained in the operating in­struction of the signal conditioning instrument under "Error code“. When sensor and connection line work without problems, probably a wrong range of the capacitive electrode was chosen. Set the changeover switch to the next higher stage.
- When the signal conditioning instrument
displays the fault signal E017, set the changeover switch on the electrode to the next lower stage.
Signal conditioning instrument VEGAMET series 600
(except VEGAMET 614)
When the electrode is not already pread­justed by VEGA, you have to choose the measuring range.
• Set the changeover switch on the capaci­tive electrode EK according to the table on the previous page.
• Carry out the adjustment according to the operating instruction of the signal condi­tioning instrument (see operating instruc­tion of the signal conditioning instrument "Adjustment“). When the indication cannot be set to 100 % (10,0 V), proceed as follows:
- when the indication does not reach
100 % (10,0 V), you have to set the changeover switch on the electrode to the next lower stage.
- When the indication exceeds 100 %
(10,0 V) and cannot be reset, you have to set the changeover switch of the elec­trode to the next higher stage. In both cases you have to repeat the adjustment.
30 Capacitive electrodes EK with signal conditioning instrument
Diagnosis

5 Diagnosis

5.1 Simulation

Test switch
A test switch can be optionally integrated in the housing to simulate a switching condition. By pushing the test switch, an additional capacitance is connected. Just the function of the oscillator and the connected instru­ments are tested.
Note
When the changeover switch is set to posi­tion 3, it can be possible that the connected capacitance is not sufficient to carry out a test.
The test switch can be used to simulate a max. level (overfill protection).
Test
Set the test switch to position 1. The changeover of the test switch to position 1 increases the capacitance of the uncovered electrode, so that the oscillator responds and triggers a full signal. The connected instru­ments will be activated.
Note
It is absolutely necessary to reset the test switch to basic position (position 0) after the test.

5.3 Repair

Due to safety and guarantee reasons repair work beside the wiring must only be made by VEGA-staff.
In case of a defect, please return the appro­priate instrument with a short description of the error to our repair department.
Failures are short-term malfunctions of the instrument which are caused by wrong ad­justment or defect on the sensor or the con­nection lines.
Failures, possible reasons and their removal are stated under "5.4 Failure removal“.

5.2 Maintenance

The instrument is maintenance free.
Capacitive electrodes EK with signal conditioning instrument 31

5.4 Failure removal

Failure Measure, failure removal
The red failure- Check the sensor inputs on the following failures: LED of the signal - short-circuit on the input conditioning instru- - sensor not correctly connected ment lights - sensor line interrupted
- supply voltage too low
Check if the sensor is connected correctly.
- failures on the sensor, effecting a current change below 2 mA or above 23 mA, trigger a fault signal on the signal conditioning instru­ments.
Measure the current on the connection line to the sensor.
The terminal voltage of the sensor in standard condition is at least 12 V.
4 … 20 mA 12 … 24 V
mA
+
Sensor
V
28
+
30
32
Signal conditioning instrument e.g. VEGATOR
Diagnosis
Note with Ex-systems that the Ex-protection is not influenced by the measuring instruments.
a. Current value < 2 mA
- Check the supply voltage on the connection line to the sensor. The voltage should be approx. 17 … 24 V.
Should you measure a value below 17 V, the signal conditioning in­strument is defect. In this case send the instrument to VEGA for repair.
- Should the red failure lamp continue to light, separate the connection
line from the signal conditioning instrument and connect a resistor of 2,2 kOhm to the sensor inputs of the signal conditioning instrument. Should the failure lamp continue to light, the signal conditioning instru­ment is defect. In this case return the instrument to VEGA for repair.
- Should the failure lamp extinguish, connect the signal conditioning
instrument again. Separate the sensor from the connection line and connect instead a resistor of 2,2 kOhm.
- Should the failure lamp continue to light, perhaps the connection line
is interrupted. Check the connection line to the sensor.
- When the failure lamp extinguishes, the sensor is defect. Check the
connected sensor.
32 Capacitive electrodes EK with signal conditioning instrument
Diagnosis
b.Current value > 22 mA
- Check all connections and the connection line to the sensor.
- Should the red failure lamp continue to light, separate the sensor from the connection line and connect instead a resistor of 2,2 kOhm. When the failure lamp extinguishes, the sensor is defect. Check the connected sensor.
- Should the failure lamp continue to light, connect the sensor again. Separate the signal conditioning instrument from the connection line and connect a resistor of 2,2 kOhm to the sensor input.
- Should the failure lamp continue to light, the signal conditioning instru­ment is defect. In this case send the instrument to VEGA for repair.
- When the failure lamp extinguishes, this is perhaps a short-circuit in the connection line. Check the connection line to the sensor.
Failure Measure, failure removal
Sensor defect, Test of the internal connections: measurement does • Remove the housing cover. not react to level • Loosen the three screws with a screwdriver and pull changes the oscillator out of the housing. Use flat pliers to pull the
oscillator easily out of the housing. Hold the oscillator on the extension of the separating wall between the sensor connections.
• Measure with an ohmmeter (range MW) the resistance values be­tween the following contacts:
Screen
CM - Capacitor
Capacitive electrodes EK with signal conditioning instrument 33
Diagnosis
Contact 4 against middle pin (1)
The resistor must be 1 MW. If the resistor is less, this means humidity in the housing or a failure in the electrode isolation. A possible reason could be a not isolated elec­trode used in a conductive (humid) medium. If the resistor is higher or if the connection is interrupted, the reason is mainly a bonding failure in the adapter plate or a defect resistor due to strong electrostatic discharge. In both cases return the electrode for repair to VEGA.
Contact 4 against vessel
The connection between contact 4 and the metal vessel (not instrument hexagon or electrode flange) should be as good as possible. Measure with an ohmmeter (range very low) the resistance value between con­tact 4 and the vessel.
• Short-circuit (0 … 3 W), optimum connection
• Resistance > 3 W
- corrosion on the mounting boss or flange
- probably the mounting boss was covered with Teflon tape or similar
Check the connection to the vessel. If there is no connection, you can connect a line from the earth terminal outside to the vessel. Note that covered flanges must be in any case connected via the earth terminal to the vessel.
Contact 7 against middle pin (1)
The resistor must be infinite (>10 MW). If the resistor is less, humidity penetrated or the compensation capaci­tor is defect.
Contact 3 against 4
In case of electrodes without screen, the resistor is infinite (>10 MW). With EK 26 the resistor should be 1 MW. In case of lower values, the electrode isolation is defect or humidity penetrated into the instrument housing. In case of higher values there is a bonding failure in the adapter plate or the resistor is defect.
Contact 4 against 6
With older sensors the resistor is infinite, with newer electrodes <3 W. In case of values between 3 W and 10 MW there is a defect.
When you cannot find a failure in the electrode, then replace the oscilla­tor by a similar type (if available) or send the electrode for repair to VEGA. If the failure disappears after insertion of the new replacement electron­ics, the adjustment on the signal conditioning instrument should be repeated as the oscillators have different characteristics.
34 Capacitive electrodes EK with signal conditioning instrument
Notes
Capacitive electrodes EK with signal conditioning instrument 35
VEGA Grieshaber KG Am Hohenstein 113 D-77761 Schiltach Phone (0 78 36) 50 - 0 Fax (0 78 36) 50 - 201 e-mail info@vega-g.de
The statements on types, application, use and operating conditions of the sensors and processing systems correspond to the actual knowledge at the date of printing.
Technical data subject to alteration
2.22 647 / Nov. 98
Loading...