Vaisala HMP260 User Manual

HMP260 SERIES TRANSMITTERS
Operating Manual
HMP260-U017en-1.5
5 January 1996
 Vaisala 1996
HMP260 SERIES
HMP260-U017en-1.5 Operating Manual
Contents
1. PRODUCT DESCRIPTION ......................................................................................... 1
2. TO BE NOTED WHEN M EASURING HUMIDITY........................................................ 3
3. INSTALLATION.......................................................................................................... 4
3.1. Selecting the place of installation ................................................................ 4
3.2. Mounting........................................................................................................ 5
3.2.1. HMP263 transmitter ......................................................................... 6
3.2.2. HMP264 transmitter ......................................................................... 7
3.2.3. HMP265 transmitter ....................................................................... 10
3.3. Grounding .................................................................................................... 11
3.4. Electrical connections................................................................................. 13
4. COMMISSIONING .................................................................................................... 15
4.1. Security lock jumper ................................................................................... 15
4.2. Selecting the analogue outputs .................................................................. 15
4.3. Connecting the RS 232C serial bus............................................................ 17
4.3.1. Reverting to factory settings of the serial port................................. 19
5. COMMANDS ............................................................................................................ 21
5.1. Commands and security lock jumper ......................................................... 21
5.2. LED commands ........................................................................................... 22
5.3. Display/keypad commands......................................................................... 23
5.3.1. Display mode ................................................................................. 23
5.3.2. Command m ode............................................................................. 23
5.3.3. Entering numbers........................................................................... 23
5.3.4. Analogue output commands ........................................................... 24
5.3.4.1. Selecting the output (mA/V)........................................................ 24
5.3.4.2. Selecting and scaling the analogue output quantities.................. 25
5.3.5. Output via the serial bus ................................................................. 26
5.3.5.1. Turning the serial interface echo ON/OFF................................... 26
5.3.5.2. Serial bus settings...................................................................... 26
5.3.5.3. Setting the transmitter address................................................... 27
5.3.5.4. Selecting the output units ........................................................... 27
5.3.6. Output modes................................................................................. 28
5.3.6.1. Setting the serial interface operation mode................................. 28
5.3.7. Others .......................................................................................... 29
5.3.7.1. Setting the measurement integration tim e................................... 29
5.3.7.2. Setting the pressure for m ixing ratio and wet bulb calc ulations ... 29
5.3.7.3. Setting the date .......................................................................... 30
5.3.7.4. Setting the time .......................................................................... 30
1997-11-25 i
HMP260 SERIES Operating Manual HMP260-U017en-1.5
5.4. Serial commands ......................................................................................... 32
5.4.1. Analogue output commands ............................................................ 32
5.4.1.1. Setting the analogue outputs ....................................................... 32
5.4.1.2. Selecting and scaling the analogue output quantities ..................33
5.4.1.3. Scaling the analogue outputs ......................................................33
5.4.2. Output via the serial bus .................................................................34
5.4.2.1. Starting the measurement output ................................................34
5.4.2.2. Stopping the measurement output ............................................... 34
5.4.2.3. Outputting the reading once........................................................ 34
5.4.2.4. Setting the output interval for the RUN mode ..............................35
5.4.2.5. Serial bus settings.......................................................................35
5.4.2.6. Selecting the output units............................................................36
5.4.2.7. Setting the transmitter address ...................................................36
5.4.2.8. Resetting the transmitter.............................................................36
5.4.4. Operating the transmitter via the serial bus.....................................36
5.4.4.1. Setting the serial interface...........................................................36
5.4.4.2. OPEN & CLOSE .........................................................................37
6. CALIBRATION..........................................................................................................38
6.1. Humidity calibration ....................................................................................38
6.1.1. One point humidity calibration.........................................................39
6.1.1.1 Using serial commands...............................................................39
6.1.1.2 Using display/keypad commands ................................................39
6.1.1.3 Using LED commands................................................................. 40
6.1.2. Two point humidity calibration.........................................................40
6.1.2.1 Using serial commands...............................................................40
6.1.2.2 Using display/keypad commands ................................................41
6.1.2.3 Using LED commands................................................................. 41
6.1.3. Calibration procedure after sensor c hange ...................................... 42
6.1.3.1 Using serial commands...............................................................42
6.1.3.2 Using display/keypad commands ................................................43
6.1.3.3 Using LED commands................................................................. 43
6.1.4. Humidity calibration table................................................................44
6.2. Temperature calibration .............................................................................. 44
6.2.1 One point offs et correction..............................................................45
6.2.2.1 Using serial commands...............................................................45
6.2.2.2 Using display/keypad commands ................................................45
6.2.2.3 Using LED commands................................................................. 46
6.2.2 Two point temperature c alibration...................................................46
6.2.2.1 Using serial commands...............................................................46
6.2.2.2 Using display/keypad commands ................................................47
6.2.2.3 Using LED commands................................................................. 47
6.3. Calibration of analogue outputs .................................................................48
6.3.1 Using serial com m ands ................................................................... 48
6.3.2 Using display/keypad commands ....................................................48
6.3.3 Using LED com m ands ....................................................................50
7. MAINTENANCE........................................................................................................50
7.1. Reference measurements............................................................................50
7.2. Self-diagnostics...........................................................................................50
7.3. Changing the HUMICAP® sensor and the filter.......................................... 51
7.4. Measurement of output currents using test points.................................... 51
7.5. Adjusting the contrast of the display..........................................................52
ii 1997-11-25
HMP260 SERIES
HMP260-U017en-1.5 Operating Manual
8. TECHNICAL DATA................................................................................................... 53
8.1. Relative humidity......................................................................................... 53
8.2. Temperature................................................................................................. 53
8.3. Calculated variables.................................................................................... 53
8.4. Pressure ....................................................................................................... 55
8.5. Analogue outputs........................................................................................ 55
8.6. Electronics ................................................................................................... 55
8.7. Mechanics.................................................................................................... 56
8.8. Environmental conditions ........................................................................... 58
8.8.1 Emissions ....................................................................................... 58
8.8.2 Imm unity ........................................................................................ 58
9. OPTIONS ................................................................................................................. 59
10. SPARE PARTS......................................................................................................... 59
Appendix 1 Serial commands Appendix 2 Safety summary Appendix 3 Installing and using the RS 485/422 serial port module Appendix 4 Installing and using the digital current loop module Appendix 5 Error messages Appendix 6 Calculation formulas: dewpoint, m ixing ratio and abs olute hum idity Appendix 7 Wiring diagr am MK4462 Appendix 8 Installation diagram MK4461 Appendix 9 Approvals and specifications of the protection unit
Certificates: Technical Research Centre of Finland No. Ex-94.C.016X
No. Ex-94.C.018X
DEMKO No. 94C.115330X
This manual is f or progr am m e vers ion HMPS260 1.01
1997-11-25 iii
HMP260 SERIES Operating Manual HMP260-U017en-1.5
iv 1997-11-25
HMP260 SERIES
HMP260-U017en-1.5 Operating Manual
1. PRODUCT DESCRIPTION
NOTE
Before installing and using the HMP260 transmitter study carefully Appendix 2: SAFETY SUMMARY.
The HMP260 series transmitters are associated electrical apparatuses. This means that the housing of the transmitter is mounted into a safe area and the sensor head and its cable can go to a potentially explosive environment (hazardous area). The transmitters incorporate a protective unit which contains the power limiting components. This unit prevents the entry of any hazardous energies to the potentially explosive area.
The transmitters are microprocessor based instruments which measure relative humidity and temperature; from these variables they can calculate dewpoint temperature, absolute humidity, mixing ratio and wet bulb temperature. The transmitters have two analogue outputs and can be connected to a serial bus via the RS 232C interface or through an RS 485/422 serial module or a digital current loop module.
The series consists of three types of transmitter:
HMP263, installation in tight places; temperatures up to +120 °C
HMP264, installation in pressure or vacuum chambers
HMP265, installation in high temperatures up to +180 °C
There are various possibilities for the configuration of the transmitters. They can have either a blank cover or a cover with a local display and keypad with which to operate the transmitter. Two analogue output signals are selected from the measured and calculated quantities; the signals can be scaled and the measurement ranges changed within certain limits. The HMP263, HMP264 and HMP265 can be supplied with two, five, ten or 15 metre probe cable.
The humidity measuring range is 0...100 %RH. The temperature is measured with a Pt 100 sensor. Temperature measurement range depends on the model; the HMP264 and HMP265 have the widest range, -40...+180 °C. The analogue temperature output can be scaled quite freely, for example -20...+60 °C can be set to correspond to 0...10 V. The dewpoint temperature, absolute humidity, mixing ratio and wet bulb temperature ranges are also scalable.
The HMP260 series units incorporate the HUMICAP® sensor, which uses an operating principle based on changes in the capacitance of the sensor as its thin polymer film absorbs water molecules.
1997-11-25 1
HMP260 SERIES Operating Manual HMP260-U017en-1.5
Options Calculation variables dewpoint temperature, mixing ratio, absolute hu-
midity, wet bulb temperature
Serial interface RS 232C (standard), RS 485/422, digital current
loop Display cover cover with or without local display & keypad Filters sintered filter, PPS grid with steel netting, PPS
grid Cable length 2, 5,10 or 15 metres Installation aids HMP263: installation kit
HMP264: NPT conical pipe threaded fitting body
(1/2 - 14 NPT)
HMP265: steel and aluminium flanges
2 1997-11-25
HMP260 SERIES
HMP260-U017en-1.5 Operating Manual
2. TO BE NOTED WHEN MEASURING HUMIDITY
In the measurement of humidity and especially in calibration, it is essential that temperature equilibrium is reached. Even a small difference in temperature between the measured object and the sensor causes an error. If the temperature is +20 °C (+68 °F) and the relative humidity 50 %RH, a difference of ±1 °C between the measured object and the sensor causes an error of ±3 %RH. When the humidity is 90 %RH, the corresponding error is ±6 %RH.
The error is at its greatest when the sensor is colder or warmer than the sur­roundings and the humidity is high. A temperature difference of a few degrees can cause water to condense on the sensor surface. In an unventilated space evaporation may take hours; good ventilation accelerates evaporation. The HUMICAP sensor starts to function normally as soon as the water has evapo­rated. If the condensed water is contaminated, the life span of the sensor may shorten and calibration may change.
10
9 8 7 6 5 4
dRH (%RH)
3 2 1 0
-40-200 20406080100 Temperature (°C)
Fig. 2.1 Measurement error at 100 %RH when the difference between the
ambient and sensor temperature is 1 °C
1997-11-25 3
HMP260 SERIES Operating Manual HMP260-U017en-1.5
3. INSTALLATION

3.1. Selecting the place of installation

The transmitters should be installed in a place that gives a true picture of the environment or process and is as clean as possible. Air should circulate freely around the sensor. A rapid air flow is recommended; it ensures that the sensor head and the ambient air are at the same temperat ure.
Install the transmitter in a place where no cold or hot spot can develop. When the sensor head is installed in a duct or a process channel where the temperature is different from the ambient temperature, insulate the point of entry; this is particularly important if the transmitter is installed with the sensor head pointing downwards. Installing the sensor head vertically is not recommended. An uninsulated installation could lead to condensation in the sensor head and even when no condensation occurs, the resultant air flow may change the temperature near the sensor and distort the readings.
Install the sensor head in the process wherever possible; avoid sample flows where the gas temperature can drop below dewpoint temperature. Install the sensor head transversely against the direction of the process flow.
In duct or channel installations drill a hole ready for a reference meter. Plug the reference hole tightly.
The transmitters are associated electrical apparatuses: the electronics housing must be in the safe area and only the sensor head and its cable may go into the hazardous area.
Install the electronics housing away from possible steams escaping from the process.
NOTE
To ensure an IP 65 class protection:
1. Always mount the transmitter housing with the cable bushings pointing downwards.
2. Make sure that the connection cable has the right thickness ( 7...10 mm) and that the cable bushing is carefully tightened.
3. Pay always special attention to closing the transmitter cover carefully and remember to tighten all four screws.
4 1997-11-25
HMP260 SERIES
HMP260-U017en-1.5 Operating Manual

3.2. Mounting

HMP263/264/265 (EEx ia) IIC
Supply voltage 24 VDC Analog outputs (mA/V) RS 232C/ RS 485 /digital current loop
SAFE AREA
HAZARDOUS AREA EEx ia IIC T3/T4
HMP263 sensor head
HMP264 sensor head for pressurized spaces
HMP265 sensor head
.
Fig. 3.1 Examples of sensor head mounting
The lead-throughs must be sealed properly to avoid air flowing in or out from the channel. The sealings must fulfil EEx requirements. When using the support bar with the HMP263, the lead-through piece must be sealed e.g. with silicon. See also Appendix 8.
133
104
Ø 6.5
Fig. 3.1 Mounting holes in the HMP260 transmitter housings
1997-11-25 5
HMP260 SERIES Operating Manual HMP260-U017en-1.5
The transmitters should be mounted with the sensor head horizontally; this way, any water condensing on the tube cannot flow onto the sensors. When there is no alternative but to install the sensor head in the process vertically, the point of entry must be carefully insulated. The cable must also be allowed to hang loosely as in Figure 3.6; this prevents any condensed water from running onto the sensor head along the cable.
If the process temperature is much higher than that of the environment, the whole sensor head and preferably part of the cable must be inside the process.
When mounted on the side of a duct or channel, the sensor head must be in­serted from the side (see Figure 3.7). If this is not possible and the sensor head must be inserted from the top, the point of entry must be carefully insulated.
NOTE
During installation the protective cover of the protective unit must not be opened and the sensor head must not be unsoldered from and then resoldered to the protective unit of the transmitter. This procedure is not allowed for the user. Vaisala is not responsible for any damages caused by incorrect use.

3.2.1. HMP263 transmitter

The HMP263 can be installed in ducts and channels with the help of the in­stallation kit available; the kit consists of a flange, a supporting bar for the sensor head cable and screws for attaching the flange to the wall of a duct. With the help of the installation kit the distance between the sensor head and the channel wall can be easily adjusted. The range of adjustment is 100...320 mm; the distance is measured from the tip of the sensor head to the flange.
duct wall
a plugged hole for reference measurements
flange
sealing
supporting bar
HAZARDOUS AREA
SAFE AREA
Fig. 3.2 Installing the sensor head of the HMP263 in a channel with the
help of flange and supporting bar
6 1997-11-25
HMP260 SERIES
when a bushing is used, its size is selected according to the diameter of the sensor head; the diameter of the cable increased by using e.g. tape at the bushing
to be sealed
(loop type) clamp
mounting the sensor head on a support
HAZARDOUS AREA
SAFE AREA
HMP260-U017en-1.5 Operating Manual
NOTE
When using the support bar the lead-through piece and the space between the duct wall and the flange must be sealed e.g. with silicon. The sealings must fulfil EEx requirements.
The sensor head can also be installed vertically.
Fig. 3.3 Vertical installation of the HMP263 sensor head

3.2.2. HMP264 transmitter

The HMP264 is supplied with a nut, a fitting body and a sealing washer. During handling the fitting body and the nut should remain in place on the body of the sensor head to prevent damage to the highly polished surface.
To achieve a leak-tight assembly:
1. Remove the fitting body from the nut and the sensor head.
2. Fasten the fitting body to the chamber wall. Tighten the fitting body
into the threaded sleeve with a torque spanner. The tightening torque is 150 ±10 Nm.
HAZARDOUS AREA
sealing washer
fitting body
nut
tightening cone
SAFE AREA
1997-11-25 7
HMP260 SERIES Operating Manual HMP260-U017en-1.5
3. Insert the body of the sensor head into the fitting body and screw the nut manually to the fitting body until the connection feels tight.
4. Mark both the fitting body and the nut hex.
A pen
nut
fitting body
5. Tighten the nut a further 30° (1/12 turn) or if you have a torque span­ner tighten it with a torque of 80 ±10 Nm.
NOTE
After detachment the nut must be tightened without increased effort.
30°
6. The tightening cone of the fitting body has to be cleaned and greased after each tenth detachment. Use high-vacuum grease (Down Corning, Europe) or a similar grease.
clean cotton stick
tightening cone
sealing washer
fitting body
The sealing washer has to be changed every time the fitting body is detached.
8 1997-11-25
HMP260 SERIES
HMP260-U017en-1.5 Operating Manual
Fasten by threaded sleeve Sealing by Metal sealing washer DIN 7603
M22x1.5
threaded sleeve M22x1.5/Ø40 x15
sheet metal
Fasten by Nut DIN 80705 Check the thickness of the sheet metal
according to the pressure of the chamber
nut DIN 80705-M22x1.5 (AISI 316) tightening torque=150 Nm ±10 Nm
NPT Conical pipe threaded connection
ANSI/ASME B1.20.1-1983
sealing by anaerobic pipe thread seal (SWAK, Cajon Company) or PTFE (teflon) tape
1/2-14 NPT
(boring)
Ø22+0.3
s=3...6mm
sealing by Metal sealing washer DIN 7603
fitting body. VAISALA code 17225 (AISI 316Ti). Body hex = 27 mm tightening torque=150 Nm ±10 Nm
Fig. 3.4 Some examples on the installation of the HMP264 sensor head
1997-11-25 9
HMP260 SERIES Operating Manual HMP260-U017en-1.5

3.2.3. HMP265 transmitter

SAFE AREA
to be sealed
to be insulated
HAZARDOUS AREA
Fig. 3.5 Installing a transmitter in a process with the sensor head hori-
zontally
SAFE AREA
to be sealed
to be insulated
HAZARDOUS AREA
Fig. 3.6 Installing a transmitter in a process with the sensor head down-
wards (not recommended)
10 1997-11-25
HMP260 SERIES
HMP260-U017en-1.5 Operating Manual
a plugged hole
HAZARDOUS AREA
for an Exi-approve d reference meter
SAFE AREA
Fig. 3.7 Mounting the sensor head on a duct or channel
When the sensor head is installed in a duct or a channel, the temperature difference of the air inside and outside the duct must be small as the sensor head conducts heat.

3.3. Grounding

A single electrical cable with a screen and three to ten wires is recommended for power and analogue output/serial bus connections. The cable diameter should be 7...10 mm.
The screen of the electrical cable must be grounded properly to achieve best possible EMC performance. Recommended cable shield is done in the cable gland as shown below.
remove the brass disks, rubber ring and nut from the transmitter
housing
strip 165 mm of the cable insulation, but leave 25 mm of the braid
visible
slip the nut and rubber ring over the cable insulation
slip the brass disk that has the bigger hole in it over the braid so that
it rests against the cable insulation
slip the other brass disk over the wires to the middle of the braid
1997-11-25 11
HMP260 SERIES Operating Manual HMP260-U017en-1.5
braid
rubber ring
165
brass disks
nut
cable
140
25
3
flexible wires 0.5 mm (AWG 20), stranded wires recommended
D = Ø 7...10 (If the cable diameter is less
than 7mm, use a shrinking tube or an adhesive tape)
2
shielding tube
braid
brass disks
push back the braid and press it between the two brass disks to
achieve a full 360° grounding; the fold between the disks should have the same diameter as the brass disks
secure the braid with a shielding tube
insert the wires into the transmitter housing through the gland.
tighten the nut
connect the wires into the screw terminals and fasten a cable tie
around the wires
cable tie
transmitter housing
gland
brass disks
rubber ring
nut
12 1997-11-25
HMP260 SERIES
HMP260-U017en-1.5 Operating Manual
NOTE
When the cable is grounded as above, the metallic parts of the sensor head, the shield of its cable, the transmitter housing and the shield of the signal cable to external system are all connected to each other. The
transmitter housing must be grounded via the grounding terminal located at the lower right hand corner of the housing. In addition to this the negative terminal of the power supply must be grounded. If these groundings have not been made correctly, the transmitter does not work.
HMP26x
Hazardous area
Power supply
The electrical safety of the grounding must comply with the EN 50014/13 standard. When compliance with Factory Mutual Standards is required, the grounding must comply with ANSI/ISA RP 12.6 and ANSI/NFPA 70.
Always use the grounding terminal to connect the transmitter to safety ground.

3.4. Electrical connections

All the component boards are grounded via the housing frame. Make sure that the fixing screws of the boards are firmly tightened before connecting the cables.
WARNING
CAUTION
1997-11-25 13
HMP260 SERIES
P
Operating Manual HMP260-U017en-1.5
CH1- and CH2- are connected
CH1+
CH1-
CH2+
together internally
CH2-
X2
X1
+-
V
mA
+-
V
mA
24 V +
CURRENT/VOLTAGE OUTPUTS
EXTERNAL POWER SUP
Do not use power supply ground (-) as output signal ground
OPENED COVER OF THE HMP260
Fig. 3.8 Electrical connections
Power supply 24 VDC Output signals 0...20 mA
4...20 mA
0...1 V
0...5 V
0...10 V
IF THE CONNECTION INSTRUCTIONS ARE NOT CAREFULLY FOLLOWED OR THE COMPONENTS IN THE PROTECTION UNIT ARE MODIFIED OR TAMPERED WITH, THE DEVICE DOES NOT MEET THE REQUIREMENTS OF EXi CLASSIFICATION. VAISALA IS NOT RESPONSIBLE FOR ANY DAM­AGES CAUSED BY INCORRECT USE.
14 1997-11-25
HMP260 SERIES
HMP260-U017en-1.5 Operating Manual

4. COMMISSIONING

When the HMP260 transmitters leave the factory, their measurement ranges and output signals have already been selected. The user can subsequently change the measurement units between metric and non-metric and select and scale the output signals with software functions, see Chapter 5 Commands and Appendix 1.

4.1. Security lock jumper

Before the settings can be changed, the security lock jumper in connector X15 must be removed (see Fig. 4.1). The security lock jumper makes it impossible to change the transmitter settings by mistake.
CHANGE OF SETTINGS DISABLED
Fig. 4.1 Location of the security lock jumper
When the security lock jumper is connected, some commands are not avail able (see Chapter 5 Commands).
Should the application require variables that are not included in the configuration of the transmitter, the user is invited to contact Vaisala or a Vaisala representative.

4.2 Selecting the analogue outputs

The HMP260 transmitters can be ordered ready with the current or voltage outputs required. If the outputs need to be changed, move the jumpers in con­nector X15 into positions as shown in the Figure 4.2.
X15
1997-11-25 15
HMP260 SERIES Operating Manual HMP260-U017en-1.5
CH1
CH1
CH1
CH1
CH1
CH2
CH1
CH2
CH1
CH2
CH1
CH2
CH2
CURRENT OUTPUTS 0 ... 20 / 4 ... 20 mA
CH2
VOLTAGE OUTPUTS 0 ... 5 V / 0 ... 10 V
CH2
VOLTAGE OUTPUTS 0 ... 1 V
CH2
CH1 0 ... 1 VOLTAGE OUTPUT CH2 CURRENT OUTPUT
X15
OPENED COVER OF THE HMP260
Fig. 4.2 Selecting the analogue outputs with jumpers
The software has to be informed which outputs are in use. This is done either through the serial interface or the menus on local display when one is in use. The serial command is AMODE and the display/keypad command "Mode Analog outputs Mode" (see Chapter 5 Commands). If the outputs need to be scaled, see serial command ASCL and the display command "Mode Analog outputs Scale".
All the jumpers are used only with the 0...1 V outputs. When other outputs are in use, the spare jumpers are kept in connector X55.
16 1997-11-25
HMP260 SERIES
HMP260-U017en-1.5 Operating Manual
X55 spare jumpers
Fig. 4.3 Spare j umpers

4.3 Connecting the RS 232C serial bus

RX GND
TX NC
X6
Fig. 4.4 Serial bus connections
To connect a PC to the HMP260 transmitters via the RS 232C serial bus, one of the following cables is required. The type of cable depends on the terminal and the connector type.
1997-11-25 17
HMP260 SERIES
3
Operating Manual HMP260-U017en-1.5
RXD
RXD
RXD
TXD
TXD
TXD
TX GND RX
TX GND RX
TX GND RX
HMP 2
PC
TERMINAL
D9S
D25S
D25P
2 5 3
4 6 7 8
3 7 2
5 6 8
20
3 7 2
Fig. 4.5 Connection of cables
When the serial bus has been connected between the PC and the transmitter, the PC is switched on. When using a PC, a terminal emulation programme (e.g. Procomm Plus, Datastorm or Windows terminal) is started.
The factory settings for data transfer are:
4800 baud
even parity
7 data bits
1 stop bit
full duplex
NOTE
When the serial bus settings are changed, the transmit­ter has to be reset before the new settings become ef­fective.
The processor does not allow the following combinations:
no parity, 7 data bits, 1 stop bit: if this combination is given the
HMP260 programme will change the number of stop bits to 2
even or odd parity, 8 data bits, 2 stop bits: if this combination is given
the programme changes the number of stop bits to 1
Refer to the manuals of the PC and the terminal emulation programme when giving serial settings.
18 1997-11-25
HMP260 SERIES
X
HMP260-U017en-1.5 Operating Manual
The RS 232C screw terminal cannot be used if an RS 485/422 serial module or a digital current loop module is used. See appendices 3 and 4 on how to install and operate these modules.
In calibrating or changing the settings of the transmitter it can be more con­venient to use the connector X17, if connector X6 is already in use. This con­nector, however, transfers only RS 232 signals. If an RS 485/422 serial port module or a digital current loop module has been installed, it has to be removed before communicating through the X17 connector.
RX GND T
X17
Fig. 4.6 Location and connections of connector X17

4.3.1 Reverting to factory settings of the serial port

If the serial port settings are not known, no commands can be given via the serial interface. The settings can be reverted to the factory settings by inserting a jumper in connector X16. The jumper must be i nserted wh en the pow er is
on!
cover of the protection unit
X16
Fig. 4.7 Forcing the serial port settings back to factory settings
When the jumper is inserted the serial line factory settings become valid, but only temporarily. The transmitter must be given new settings; otherwise the transmitter uses the old, unknown settings after power-up. When the new settings have been given, the transmitter must be reset. Note that the
1997-11-25 19
HMP260 SERIES Operating Manual HMP260-U017en-1.5
jumper must be removed before the transmitter is reset. If the jumper is in place when power i s turned on, the tran smitter does not w ork.
After jumper insertion the transmitter is in STOP mode, ready to receive commands.
The same method is used when the transmitter is in POLL mode and the user has forgotten its address.
NOTE
Inserting a jumper in any other place in connector X16 voids the guarantee of the transmitter.
20 1997-11-25
HMP260 SERIES
HMP260-U017en-1.5 Operating Manual

5. COMMANDS

As the HMP260 transmitters are microprocessor based devices, their configuration can be set to correspond to the specific needs of the user. This is done through commands, either utilizing the menus on the local display or by giving commands through the serial interface (see Appendix 1). Most often the commands are used to change the settings of the two analogue channels.
A limited range of commands can be given by using the three press switches ­up, down and enter - inside the transmitter housing. Four LEDs indicate the command given with the up and down switches. The switches and LEDs are in all HMP260 transmitters. LED commands can be used to calibrate the trans­mitters (both humidity and temperature) or to calibrate the analogue outputs.
A full range of commands can be given through the display/keypad or through the RS 232C serial bus. The commands can be used e.g. to select and scale the outputs, to calibrate the humidity and temperature channels as well as the analogue outputs and to set the serial interface.

5.1 Commands and security lock jumper

In order to prevent any tampering with the transmitter settings, the transmitters can not be calibrated, the analogue outputs set or the analogue output quanti­ties selected or scaled unless the security lock jumper has been disconnected. The commands involved are:
all LED commands
display/keypad commands:
Cali RH cal T cal
Analog outputs
Mode Analog outputs Mode
serial commands:
CRH, CT, FCRH, ACAL; AMODE, ASEL, ASCL
In the following, the description of these functions is preceded with a re­minder of the security lock jumper:
Disconnect the security lock jumper!
Scale
1997-11-25 21
HMP260 SERIES Operating Manual HMP260-U017en-1.5

5.2 LED commands

NOTE
If the transmitter has a display/keypad cover, the LED commands cannot be used.
LED commands can be used to operate the transmitters in the field. These commands can be used in humidity and temperature calibration and calibration of the analogue outputs.
Open the housing and press any one of the three press switches. The LEDs will light up for 2...3 seconds.
UP
DOWN
ENT
press switches
LEDs
Fig. 5.1 Location of press switches and LEDs
Use the up and down switches (marked with arrows on the printed board) to find the desired command code and acknowledge it with the ENT switch. The command codes are (l = lit, ¡ = dark):
¡¡¡¡ (0) return to normal state ¡¡¡l (1) relative humidity calibration ¡¡l¡ (2) tem perature cali bration ¡¡ll (3) calibration of analogue outputs l¡¡¡ (8) relative humidity calibration after sensor change
22 1997-11-25
HMP260 SERIES
HMP260-U017en-1.5 Operating Manual

5.3 Display/keypad commands

5.3.1 Display mode

In the display mode the transmitters output measurements on the display; dif­ferent quantities can be scrolled with the arrow keys according to the variables selected when ordering the transmitter. After reset the transmitters are always in the display mode.
The display also shows error messages and alarms if they occur.

5.3.2 Command mode

Press the CL key to enter the command mode. The first display is the main menu:
The commands can be scrolled with the arrow keys. The currently active commands flashes; a command is selected with the ENT key. When a menu is displayed, either the first command or the currently valid setting flashes. The CL key takes the transmitter back to the display mode.

5.3.3 Entering numbers

When the transmitter needs numbers to be entered into the programme (e.g. when scaling or setting the analogue outputs, in calibration or when giving the transmitter an address), the field is either empty or the currently valid figure is displayed. Any previously given value is deleted with the CL key.
When the field is empty, a cursor blinks at the right side of the display. Pressing the arrow keys brings either a blank ' ', a comma ',', a dash '-', a full stop '.' or a number from '0' to '9' on the display. The right character is selected with ENT; after that the number or numbers move left one step. Entering
1997-11-25 23
HMP260 SERIES Operating Manual HMP260-U017en-1.5
numbers is ended with selecting a blank ' ' and pressing ENT. The last charac­ter entered can be deleted with CL. If CL or ENT key is pressed when the field is empty, the programme returns to the previous display.
With some commands (e.g. calibration) the figures are changed using the ar­row keys. When an arrow key is pressed continuously for a while, the numbers start changing at an increasing rate.

5.3.4 Analogue output commands

5.3.4.1 Selecting the output (mA/V)
Disconnect the security lock jumper!
Select Mode in the main menu and Analog outputs in the Mode
menu:
Select Mode ( mA / V ). The current settings for channel 1 are dis-
played:
If the settings are correct, press ENT.
If the settings need to be changed, press CL:
the quantity (mA/V) starts flashing; it can be changed with the ar-
row keys and acknowledged with the ENT key
the lower limit starts flashing
acknowledge the lower limit with ENT or start changing it by
pressing CL; a new lower limit is given one character at a time with the arrow keys
the upper limit starts flashing
acknowledge the upper limit with ENT or start changing it by
pressing CL; a new upper limit is given one character at a time with the arrow keys
When channel 1 has been set, the programme goes on to channel 2; the procedure is the same as with channel 1.
24 1997-11-25
Loading...
+ 88 hidden pages