Vaisala Oyj 2002
No part of this document may be reproduced in any form or by any means,
electronic or mechanical (including photocopying), nor may its contents be
communicated to a third party without a prior written notice of the copyright
holder.
The instruction manuals may be changed without prior notice.
The HMP240 transmitter is a microprocessor based instrument for the
measurement of dewpoint temperature especially in high humidities and/or
fast changing temperatures. The dewpoint temperature is measured through
relative humidity and temperature. The dewpoint temperature, although a
calculated variable, is the primary reading obtained with the HMP240. As the
probe is equipped with the warming function, the relative humidity reading
obtained is not correct as such whereas the dewpoint temperature is. If the
temperature is below 0 °C, the user can select whether the transmitter
calculates dewpoint or frostpoint reading; as default, the transmitter calculates
frostpoint.
The transmitter can be ordered with one or two sensor heads. If the transmitter
has only a humidity sensor head, it can output the dewpoint temperature or the
mixing ratio. If the transmitter is ordered with an additional temperature head,
the user can choose the output from the following readings: dewpoint, relative
humidity, ambient temperature, dewpoint difference (= ambient temperature dewpoint), mixing ratio, absolute humidity, and wet bulb temperature. The
configuration that the user completes in the order form determines the
available readings. The transmitter has two analogue outputs and can be
connected to a serial bus via the RS 232C interface or through an RS 485/422
serial module or a digital current loop module.
There are various possibilities for the configuration of the transmitter. It can
have either a blank cover, or a cover with a local display and keypad with
which to operate the transmitter. Two analogue output signals are selected
from the measured and calculated quantities; the signals can be scaled and the
measurement ranges changed. The HMP240 can be supplied with two, five or
ten meter sensor head cables.
The dewpoint measurement range is -40...+100 °C. The range depends on the
desired accuracy because the dewpoint is calculated through the RH and the T
readings (see Chapter 8). The analogue temperature output can be scaled quite
freely, for example -20...+60 °C can be set to correspond to 0...10 V. The
relative humidity, absolute humidity, dewpoint difference, mixing ratio and
wet bulb temperature ranges are also scalable.
In some specific applications, the sensor gain may gradually decrease because
of an interference caused by some particular chemicals present in the ambient.
These changes can be recovered with an optional re-gaining function.
The transmitter is equipped with a composite humidity and temperature
sensor; the operation of the HUMICAP®KC humidity sensor is based on
changes in the capacitance of the sensor as its thin polymer film absorbs water
molecules.
5
Page 8
HMP240
User's GuideM210300en
Options
Additional T probeFor the measurement of the ambient temperature and
for obtaining other quantities in addition to
dewpoint temperature and mixing ratio
Calculation variables
dewpoint difference, mixing ratio, absolute
humidity, wet bulb temperature
Serial interfaceRS 232C (standard), RS 485/422, digital current
loop module
Display covercover with local display & keypad
Filterssintered filter, PPS grid with steel netting
Cable lengths2, 5 or 10 meters
Installation aids
Installation kit for temperatures up to 180 °C
HMP243MIKMeteorological installation kit
6
Page 9
HMP240 series
M210300enUser's Guide
2. ADVANTAGES OF A WARMED SENSOR HEAD
In some measurement applications, unwanted dew formation makes humidity
measurement difficult or even impossible. At a weather station, for example,
high humidity combined with rapidly changing outdoor temperature can
condense the water vapour in the air onto the sensor head. Until this dew
evaporates or dries, it is impossible to obtain a true reading. Dew formation is
also a problem in environmental chambers and in processes involving high
humidity, such as meat processing.
In some applications, rapid temperature changes are the source of difficulty.
Normally, relative humidity sensors must be at the same temperature as the
ambient air or the measurement is incorrect. At +20 °C and 90 %RH a difference of +1 °C causes an error of +6 %RH. This makes reliable humidity
measurements difficult in rapidly changing temperatures.
In other applications, the problem is a combination of both the dew formation
and rapid changes of temperature; a rapid rise in temperature can cause a
corresponding increase of water vapour in the ambient. In these conditions, the
temperature of the sensor head rises more slowly, and can remain below the
dewpoint of the ambient for a while. Dew then forms on the sensor head, and
it may take several hours or in the worst case, several days for the sensor to
recover from the condensation.
Vaisala's HMP240 dewpoint transmitter offers a reliable solution for humidity
measurement in all these demanding conditions. The humidity sensor head is
kept dry by warming it. As the measurement is fully temperature compensated,
changes in the process temperature do not delay the measurement. This means
that the response time of dewpoint measurement is proportional only to the
diffusion time of water molecules inside the sensor head.
The HMP240 has an excellent performance also at normal humidities and in
stable environments, but it will mostly be used in applications where very high
humidity can take place. In these kinds of applications, the use of traditional
transmitters has been restricted.
7
Page 10
HMP240
User's GuideM210300en
3. INSTALLATION
3.1 Selecting the place of installation
The transmitter should be installed in a place that gives a true picture of the
environment or process and is as clean as possible. Air should circulate freely
around the sensor. If necessary, the transmitter can be installed in a place
where a hot spot may develop. However, cold spots should be avoided.
When the sensor head is installed in a duct or a process channel where the
temperature is different from the ambient temperature, insulate the point of
entry; this is particularly important if the transmitter is installed with the
sensor head pointing downwards. Installing the sensor head of the HMP240
vertically is not recommended because in high humidities, the humidity may
condence on the sensor head cable and then flow on to the sensor head.
When there is no alternative but to install the sensor head in the process
vertically, the point of entry must be carefully insulated. The cable must also
be allowed to hang loosely in order to prevent any water that possibly
condenses on the cable from running onto the sensor head.
Install the humidity sensor head in the process wherever possible. Avoid
sample flows where the gas temperature can drop below dewpoint
temperature; this might result in erroneous measurement readings. Install the
sensor head transversely against the direction of the process flow. If the
process temperature is much higher than that of the environment, the whole
sensor head and preferably part of the cable must be inside the process.
In duct or channel installations drill a hole ready for a reference meter. Plug
the reference hole tightly (see Figure 3.2).
Install the electronics housing away from possible steams escaping from the
process.
8
Page 11
HMP240 series
M210300enUser's Guide
3.2 Mounting
3.2.1 Mounting the HMP243
133
humidity
sensor head
T-sensor head
(optional)
104
ø6.5
Cable length 2000,
5000 or 10000 mm
Sliding PTFE-sleeve for flange installation
and cable-gland installation
Figure 3.1HMP243 transmitter with a humidity sensor head and an
additional T sensor head
When mounted on the side of a duct or channel, the sensor head must be inserted from the side (see Figure 3.2). If this is not possible and the sensor head
must be inserted from the top, the point of entry must be carefully insulated.
NOTE
The two sensor heads should be installed so that the
humidity sensor head does not warm the T sensor head,
i.e. the T sensor head is installed closer to the process
flow. When the RH reading is required, always install
the T sensor head in the place where you need the
reading from.
The HMP243 can be installed in ducts and channels with the help of the installation kit; the kit consists of a flange, a supporting bar for the sensor head
cable and screws for attaching the flange to the wall of a duct. With the help
of the installation kit the distance between the sensor head and the channel
wall can be easily adjusted. The range of adjustment is 100...320 mm; the
distance is measured from the tip of the sensor head to the flange.
9
Page 12
HMP240-series
h
User's Guide M210300en
duct wall
sealing (silicone)
flange
umidity sensor head
supporting bar
a plugged hole for reference
measurements
T-sensor head
(optional)
PTFE sleeve
Figure 3.2Installing the sensor heads of the HMP243 in a channel
with the help of flanges and supporting bars
The sensor head can also be installed vertically.
PTFE-sleeve
a bushing (cable-gland)
(seal:Viton)
to be sealed
(silicone)
stainor a
humidity sensor
head
T-sensor head
(optional)
when a bushing is used, its
size is selected according to
the diameter of the sensor
head; the diameter of the
cable is increased by using
e.g. silicon tape at the bushing
stainless steel
cable tie or a
similar fastener
m
h
h
10
Figure 3.3Vertical installation of the HMP243 sensor heads
Page 13
HMP240 series
M210300enUser's Guide
3.2.2 Mounting the HMP247
HMP247 has a small size probe made of stainless steel. The sensor head
withstands temperatures -40...180 ºC (-40...356 ºF) and pressure up to 10 bar
(1MPa, 145 psi). The probe is suitable for applications where a mechanically
very durable leak proof probe is needed.
Pressure tight installations
A pressure tight installation can be done by using a probe head installation or
cable installation. Vaisala's cable gland HMP247CG is recommended for the
humidity probe's cable installation.
The different pressure tight installations of the HMP247 are shown in the
following pictures. In cases where the probe shall be positioned horizontally in
the measuring chamber (or in duct) a supportive thread bar set HMP247TBS
can be ordered from Vaisala (order code: HMP247TBS).
A) Probe head installation
1. Make hole with M20x1.5 thread
in the measurement chamber or
process wall. The smoothness of
the thread circle should be
R= 6.3 microm.
2. Install the cable gland
(EN50262, version A, example:
AGRO 1160.20.145) on the thread
of the process wall.
3. Push the probe through the
gland so deep that the backside of
the probe is flust with the cable
gland nut.
4. Tighten the cable gland nut,
the probe will lift up slightly.
B) Cable
installation with
Vaisala's
HMP247CG
∅13.6 mm∅ 5.3 mm
Process pressure <
10 bar, temperature
according to the
cable's sealing
A) Pipe fitting
∅30/M20x1.5x13.5
B) Cable
installation with
Vaisala's
HMP247CG
Figure 3.4 HMP247 Pressure tight installations A) Probe head installation
B) Cable installation (cable gland: HMP247CG) .
11
Page 14
HMP240
fitting and the cable.
User's GuideM210300en
Pressure tight installation with Vaisala's supportive thread bar
set HMP247TBS (cable gland HMP247CG included)
1. Make a hole with M20x1.5 to a wall or weld a pipe fitting
(∅30/M20x1.5x12mm) to a wall.
2. Install the cable gland's fitting body on the hole in the wall.
3. Thread the sensor head through the nut and press the cable
through the cutting line of the sealing.
4. Turn the supporting bars into the sensor head's threaded holes.
5. Turn the sleeve on the opposite end of the supporting bar and
trim to the right length.
6. Cut away the overage supporting bar with saw.
7. Press the sleeves together (with cable inside) and install the
sensor head to a process through the fitting body.
8. Tighten the cable gland nut.
For processes up to 120°C (248 °F). More detailed
instructions are included in the set.
Example of pressure tight temperature probe installation
(cable gland example: AGRO Nr. 1100.12.91.065, EN50262,
version A)
Process pressure < 7 bar,
temperature according to the cable
sealing
∅ 6.5 mm
Recommended support to
keep the probe in
horizontal position
Silicon glue between the PTFE
Cable gland
HMP247CG
12
Figure 3.5 HMP247 pressure tight installations with supporting bars.
Page 15
HMP240 series
M210300enUser's Guide
3.3 Grounding
A single electrical cable with a screen and three to ten wires is recommended
for power and analogue output/serial bus connections. The cable diameter
should be 7...10 mm.
The screen of the electrical cable must be grounded properly to achieve best
possible EMC performance. Recommended cable shield is done in the cable
gland as shown below.
• remove the brass disks, rubber ring and nut from the transmitter
housing
• strip 165 mm of the cable insulation, but leave 25 mm of the braid
visible
• slip the nut and rubber ring over the cable insulation
• slip the brass disk that has the bigger hole in it over the braid so that
it rests against the cable insulation
• slip the other brass disk over the wires to the middle of the braid
2
shielding tube
braid
brass disks
braid
brass
disks
rubber
ring
165
nut
cable
140
25
flexible wires 0.5 mm
(AWG 20), stranded wires
3
recommended
D = Ø 7...10 mm
(If the cable diameter is less
than 7mm, use a shrinking
tube or an adhesive tape)
• push back the braid and press it between the two brass disks to
achieve a full 360° grounding; the fold between the disks should have
the same diameter as the brass disks
• secure the braid with a shielding tube
• insert the wires into the transmitter housing through the gland
13
Page 16
HMP240
User's GuideM210300en
• tighten the nut
• connect the wires into the screw terminals and fasten a cable tie
around the wires
cable tie
transmitter housing
gland
brass disks
rubber ring
nut
NOTE
When the cable is grounded as explained on the
previous page, the metallic parts of the sensor head, the
shield of its cable, the transmitter housing and the
shield of the signal cable to external system are all
connected to each other. After this the whole system
can be grounded from one point only. If the grounding
is made via several points (sensor head, transmitter
housing, signal cable), make sure that the different
groundings are made to the same grounding potential.
Otherwise harmful grounding currents may be
generated. If you do the grounding via the transmitter
housing, use one serrated lock washer between a mounting screw and the housing; the lock washer breaks the
paint on the housing.
14
Page 17
HMP240 series
POWER SUPPLY
M210300enUser's Guide
3.4 Electrical connections
CH1- and CH2- are conne cted
CH1+
CH1-
CH2+
together internally
CH2-
X2
X1
+-
V
mA
+-
V
mA
CURRENT/VOLTAGE
OUTPUTS
Do not use power supply
ground (-) as output signal
ground
24 V +
OPENED COVER OF THE
HMP 243
Figure 3.6Electrical connections
Power supply24 VDC
24 VAC (see Chapter 3.4.1)
Output signals0...20 mA
4...20 mA
0...1 V
0...5 V
0...10 V
Power supply ground (-) is connected to the housing with parallel connection
of 15 nF capacitor and 300 kΩ resistor.
15
Page 18
HMP240
User's GuideM210300en
3.4.1 Connection to a 24 VAC supply
The HMP240 transmitter can also be connected to a 24 VAC supply without
an external rectifier. However, when more than one transmitter is connected to
one 24 VAC transformer, a common loop is formed and there is an increased
risk of a short-circuit. To avoid this, always use separate floating supply for
each transmitter (see Figure 3.7 A). However, if several transmitters have to
share one transformer, the phase (∼) must always be connected to + connector
in each transmitter (see Figure Figure 3.7 B).
A) NO COMMON LOOP FORMED - RECOMMENDED
HMP243 transmitter
24 VAC
24 VAC
HMP243 transmitter
B) COMMON LOOP FORMED -
HMP243 transmitter
24 VAC
supply voltage
Controller
supply voltage
supply voltage
signal output
signal output
NOT RECOMMENDED!
Controller
signal output
16
shared
common line
supply voltage
signal output
HMP243 transmitter
Figure 3.7Connecting the transmitters to a 24 VAC supply (valid
also for the HMP247 transmitter).
Page 19
HMP240 series
M210300enUser's Guide
4. COMMISSIONING
When the HMP240 transmitter leaves the factory, its measurement ranges and
output signals have already been scaled according to the order form completed
by the customer. The unit is calibrated at the factory and ready to operate
when the power is turned on. If you take into use active current, voltage or
serial bus outputs, make these connections first; appendix 6 describes them in
detail.
NOTE
Make sure that the power is not turned on until cables
have been connected to screw terminals!
In transmitters with display, the software version appears for a few seconds
when the power is turned on. After this, measurement results appear
automatically. Should an error message appear on the display, consult
Appendix 4.
If your transmitter has a blank cover and the LED indicator inside the housing
lights up, consult Appendix 4 for further information.
4.1 Changing the parameters
If necessary, the user can subsequently change the measurement units between
metric and non-metric and select and scale the output signals with software
functions. This is done through commands, either utilizing the menus on the
local display or giving commands through the serial interface (see
Appendices). Most often the commands are used to change the settings of the
two analogue channels.
A limited range of commands can be given with the three press switches (up,
down, enter) inside the transmitter housing. There are four LEDs to indicate
the commands given with the up and down switches. All HMP240 units
incorporate these switches and LED indicators. LED commands can be used to
calibrate the transmitters (both humidity and temperature) or to calibrate the
analogue outputs.
If you need to change some functions, read the following chapters carefully.
4.2 Security lock jumper
Before the settings can be changed, the security lock jumper in connector X15
must be removed (see Figure 4.1). The security lock jumper makes it
impossible to change the transmitter settings by mistake.
17
Page 20
HMP240
OPENED COVER OF THE HMP 243
User's GuideM210300en
X15
CHANGE OF SETTINGS
DISABLED
Figure 4.1Location of the security lock jumper
When the security lock jumper is connected, some commands are not available
(see Chapter 5).
Should the application require variables that are not included in the
configuration of the transmitter, the user is invited to contact Vaisala or a
Vaisala representative.
4.3 Selecting the analogue outputs
The HMP240 transmitter can be ordered ready with the current or voltage
outputs required. If the outputs need to be changed, move the jumpers in connector X15 into positions as shown in Figure 4.2.
18
Page 21
HMP240 series
OPENED COVER OF THE HMP 243
1
M210300enUser's Guide
CH1
CH1
CH1
CH2
CURRENT OUTPUTS
0 ... 20 / 4 ... 20 mA
CH2
VOLTAGE OUTPUTS
0 ... 5 V / 0 ... 10 V
CH2
VOLTAGE OUTPUTS
0 ... 1 V
CH
CH2
CH1
CH2
CH1
CH2
CH1
CH1
CH2
CH2
CH1 0 ... 1 VOLTAGE OUTPUT
CH2 CURRENT OUTPUT
X15
Figure 4.2Selecting the analogue outputs with jumpers
The software has to be informed which outputs are in use. This is done either
through the serial interface or the menus on a local display. The serial
command is AMODE and the display/keypad command "Mode ð Analogoutputs ð Mode" (see Chapter 5). For the scaling of the outputs, see serial
command ASCL and the display command "Mode ð Analog outputs ðScale".
All the jumpers are used only with the 0...1 V outputs. When other outputs are
in use, the spare jumpers are kept in connector X55.
19
Page 22
HMP240
OPENED COVER OF THE HMP 243
User's GuideM210300en
X55
spare jumpers
OPENED COVER OF THE HMP 243
Figure 4.3Spare jumpers
4.4 Connecting the RS 232C serial bus
RX
GND
TX
NC
X6
Figure 4.4Serial bus connections
To connect a PC to the HMP240 transmitter via the RS 232C serial bus, one of
the following cables is required. The type of the cable depends on the terminal
and the connector type.
20
Page 23
HMP240 series
3
M210300enUser's Guide
RXD
RXD
RXD
TXD
TXD
TXD
TX
GND
RX
TX
GND
RX
TX
GND
RX
HMP 24
PC
TERMINAL
D9S
D25S
D25P
2
5
3
4
6
7
8
3
7
2
5
6
8
20
3
7
2
Figure 4.5 Connection of cables
When the serial bus has been connected between the PC and the transmitter,
the PC is switched on. When using a PC, a terminal emulation programme
(e.g. Procomm Plus, Datastorm or Windows terminal) is started.
The factory settings for data transfer are:
• 4800 baud
• even parity
• 7 data bits
• 1 stop bit
• full duplex
NOTE
When the serial bus settings are changed, the transmitter has to be reset before the new settings become effective.
The processor does not allow the following combinations:
• no parity, 7 data bits, 1 stop bit: if this combination is given the
HMP240 programme will change the number of stop bits to 2
• even or odd parity, 8 data bits, 2 stop bits: if this combination is given
the programme changes the number of stop bits to 1
Refer to the manuals of the PC and the terminal emulation programme when
giving serial settings.
The RS 232C screw terminal cannot be used if an RS 485/422 serial module
or a current loop module is used. See Appendices 2 and 3 on how to install
and operate these modules.
21
Page 24
HMP240
RX GND TX
User's GuideM210300en
In calibrating or changing the settings of the transmitter it can be more convenient to use the connector X17, if connector X6 is already in use. This connector, however, transfers only RS 232C signals. If an RS 485/422 serial port
module or a current loop module has been installed, it has to be removed before communicating through the X17 connector.
X17
Figure 4.6Location and connections of connector X17
NOTE
Some PC computers can generate interferences to the
measured humidity and temperature values if the
transmitter and the PC are connected to different mains
outlets. To minimize the possibility of these interferences, always use the same mains outlet (same phase of
the mains electricity) for the PC and the power supply
of HMP240. This is especially the case when using the
serial line connector X17. The serial line connector X6
is more immune to these interferences than connector
X17.
4.4.1 Reverting to factory settings of the serial port
If the serial port settings are not known, no commands can be given via the
serial interface. The settings can be reverted to the factory settings by inserting
a jumper in connector X16. The jumper must be inserted when the power is
on!
X16
OPENED COVER OF THE HMP 243
Figure 4.7Forcing the serial port settings back to factory settings
22
Page 25
HMP240 series
M210300enUser's Guide
When the jumper is inserted the serial line factory settings become valid, but
only temporarily. The transmitter must be given new settings; otherwisethe transmitter uses the old, unknown settings after power-up. When the
new settings have been given, the transmitter must be reset. The jumper must
be removed before the transmitter is reset; if the jumper is in place when
power is turned on, the transmitter does not work.
After jumper insertion the transmitter is in STOP mode, ready to receive
commands.
The same method is used when the transmitter is in POLL mode and the user
has forgotten its address.
CAUTION
Inserting a jumper in any other place in connector X16
voids the guarantee of the transmitter.
23
Page 26
HMP240
User's GuideM210300en
5. COMMANDS
As the HMP240 transmitter is a microprocessor based device, its configuration
can be set to correspond to the specific needs of the user. This is done through
commands, either by utilizing the menus on the local display or by giving
commands through the serial interface (see Appendix 1). Most often the
commands are used to change the settings of the two analogue channels.
A limited range of commands can be given by using the three press switches up, down and enter - inside the transmitter housing. Four LEDs indicate the
command given with the up and down switches. LED commands can be used
to calibrate the transmitter (both humidity and temperature) or to calibrate the
analogue outputs.
A full range of commands can be given through the display/keypad or through
the RS 232C serial bus. The commands can be used e.g. to select and scale the
outputs, to calibrate the humidity and temperature channels as well as the
analogue outputs and to set the serial interface.
5.1 Commands and security lock jumper
In order to prevent any tampering with the transmitter settings, the transmitter
cannot be calibrated, the analogue outputs set or the analogue output
quantities selected or scaled unless the security lock jumper has been disconnected. The commands involved are:
In the following, the description of these functions is preceded with a reminder of the security lock jumper:
Scale
24
Disconnect the security lock jumper!
Page 27
HMP240 series
OPENED COVER OF THE HMP 243
M210300enUser's Guide
5.2 LED commands
NOTE
If the transmitter has a display/keypad cover, the LED
commands cannot be used.
LED commands can be used to operate the transmitter in the field. These
commands can be used in humidity and temperature calibration and calibration
of the analogue outputs.
Open the housing and press any one of the three press switches. The LEDs
will light up for 2 - 3 seconds.
UP
DOWN
ENT
press switches
LEDs
Figure 5.1Location of press switches and LEDs
Use the up and down switches (marked with arrows on the printed board) to
find the desired command code and acknowledge it with the ENT switch. The
command codes are (l = lit, ¡ = dark):
¡¡¡¡(0)return to normal state
¡¡¡l(1)relative humidity calibration
¡¡l¡(2)temperature calibration of the humidity sensor
head
¡l¡¡(3)temperature calibration of the T sensor head
¡¡ll(4)calibration of analogue outputs
l¡¡¡(8)relative humidity calibration after sensor change
25
Page 28
HMP240
User's GuideM210300en
5.3 Display/keypad commands
5.3.1 Display mode
In the display mode the transmitter outputs measurements on the display; different quantities can be scrolled with arrow keys. The first line is scrolled with
button σ and the second line with button τ; all selections are stored with
ENTER. The selected quantities appear on the display also after power failure.
After reset the transmitter is always in the display mode.
The display also shows error messages and alarms if they occur.
5.3.2 Command mode
Press the CL key to enter the command mode. The first display is the main
menu:
The commands can be scrolled with the arrow keys. The currently active
command flashes; a command is selected with the ENT key. When a menu is
displayed, either the first command or the currently valid setting flashes. The
CL key takes the transmitter back to the display mode.
5.3.3 Entering numbers
When numbers need to be entered into the transmitter programme (e.g. when
scaling or setting the analogue outputs, in calibration or when giving the
transmitter an address), the field is either empty or the currently valid figure is
displayed. Any previously given value is deleted with the CL key.
When the field is empty, a cursor blinks on the right side of the display.
Pressing the arrow keys makes either a blank ' ', a comma ',', a dash '-', a full
stop '.' or a number from '0' to '9' appear on the display. The right character is
selected with ENT; after that the number or numbers move left one step.
Entering numbers is ended with selecting a blank ' ' and pressing ENT. The
last character entered can be deleted with CL. If CL or ENT key is pressed
when the field is empty, the programme returns to the previous display.
With some commands (e.g. calibration) the figures are changed using the arrow keys. When an arrow key is pressed continuously for a while, the numbers
start changing at an increasing rate.
26
Page 29
HMP240 series
M210300enUser's Guide
5.3.4 Analogue output commands
5.3.4.1 Selecting the output (mA/V)
Disconnect the security lock jumper!
• Select Mode in the main menu and Analog outputs in the Mode
menu:
• Select Mode ( mA / V ). The current settings for channel 1 are dis-
played:
• If the settings are correct, press ENT.
• If the settings need to be changed, press CL:
− the quantity (mA/V) starts flashing; it can be changed with the ar-
row keys and acknowledged with the ENT key
− the lower limit starts flashing
− acknowledge the lower limit with ENT or start changing it by
pressing CL; a new lower limit is given one character at a time
with the arrow keys
− the upper limit starts flashing
• acknowledge the upper limit with ENT or start changing it by pressing
CL; a new upper limit is given one character at a time with the arrow
keys
When channel 1 has been set, the programme goes on to channel 2; the procedure is the same as with channel 1.
NOTE
The analogue output jumpers must also be set to right
places (see Figure 4.2).
27
Page 30
HMP240
User's GuideM210300en
5.3.4.2 Selecting and scaling the analogue output quantities
Disconnect the security lock jumper!
• Select Mode in the main menu and Analog outputs in the Mode
menu:
• Select Scale. The quantity and scaling for channel 1 are displayed:
• If the settings are correct, press ENT.
• If the settings need to be changed, press CL:
− the quantity (RH, T, Td, dT, x, a, Tw) starts flashing; it can be
changed with the arrow keys and acknowledged with the ENT key
− the lower limit starts flashing
− acknowledge the lower limit with ENT or start changing it by
pressing CL; a new lower limit is given with the arrow keys
− the upper limit starts flashing
− acknowledge the upper limit with ENT or start changing it by
pressing CL; a new upper limit is given with the arrow keys
• When channel 1 has been set, the programme goes on to channel 2;
the procedure is the same as with channel 1.
Please note that the available selections are affected by the choice of output
parameters. Also make sure that the temperature measuring ranges are not
exceeded.
28
Page 31
HMP240 series
M210300enUser's Guide
5.3.5 Output via the serial bus
5.3.5.1 Turning the serial interface echo ON/OFF
• Select More in the main menu, select More in the More menu and
select Echo in the second More menu.
• Use the arrow keys to select the right alternative and press ENT.
5.3.5.2 Serial bus settings
• Select Seri in the main menu; the currently valid serial interface set-
tings are displayed:
• If the settings are correct, press ENT; the programme returns to the
display mode.
• If the settings need to be changed, press CL:
• Select the parameter to be changed with the arrow keys and ENT key.
Selecting baud rate:
Selecting parity:
Selecting data bits:
29
Page 32
HMP240
User's GuideM210300en
Selecting stop bits:
Full duplex/half duplex:
The processor does not allow the following combinations:
• no parity, 7 data bits, 1 stop bit: if this combination is given the
HMP240 programme will change the number of stop bits to 2
• even or odd parity, 8 data bits, 2 stop bits: if this combination is given
the programme changes the number of stop bits to 1
NOTE
The serial bus settings become effective only after reset.
5.3.5.3 Setting the transmitter address
Address is used when more than one transmitter is connected to one serial bus;
this way, it is possible to communicate with one transmitter at a time.
• Select Addr in the main menu; the following is displayed:
• Pressing ENT returns the programme to the main menu.
• Pressing CL deletes the old address; enter the new address with the
arrow keys.
30
Page 33
HMP240 series
M210300enUser's Guide
5.3.5.4 Selecting the output units
• Select Unit in the main menu:
• Use the arrow keys to select the right alternative and press ENT.
metricnon-metric
RH%RH%RH
T°C°F
Td°C°F
dT
ag/m
°C°F
3
gr/ft
3
xg/kggr/lb
Tw°C°F
5.3.5.5 Selecting the calculation mode
Disconnect the security lock jumper!
• Select More and then again More in the second menu:
• Select Frost and then the desired alternative with the arrow keys;
FROST ON (default) for frostpoint and FROST OFF for dewpoint
calculation at dewpoint temperatures below 0 °C.
5.3.6 Output modes
The output modes only affect output through the serial interface: the transmitter accepts all display and LED commands irrespective of which serial output
mode it is in. The HMP240 transmitter has three serial output modes: RUN,
STOP and POLL.
In the RUN mode the transmitter outputs measurements automatically through
the serial interface to a PC or a peripheral. The only command that can be
given through the serial interface is S (stop), which ends the RUN mode.
In the STOP mode serial commands are given to the transmitter. Measurements are then output only by entering command SEND.
The POLL mode is used when more than one transmitter is connected to the
same serial bus; a single transmitter can be addressed and communicated with.
31
Page 34
HMP240
User's GuideM210300en
When the connection to a given transmitter is opened in the POLL mode, the
transmitter goes into STOP mode and can then receive commands normally.
Closing the connection returns the transmitter to POLL mode. In POLL mode,
the transmitter outputs measurement only when requested (command SEND
aa). If the user has forgotten the address of the transmitter and the transmitter
does not have a display, the transmitter has to be reverted to the factory settings (see Chapter 4.4.1). If the transmitter has a display, the settings can be
checked through it.
5.3.6.1 Setting the serial interface operation mode
• Select Mode in the main menu; the following is displayed:
• Select Serial output:
• The currently valid setting flashes. Select the desired mode with the
arrow keys and press ENT. After this the programme returns to the
Mode Menu.
• When Run mode is selected, the currently valid output interval is dis-
played:
The output interval setting can be changed as follows:
• press CL
• the number starts flashing
• if the interval needs to be changed, press CL again and enter the new
interval; otherwise press ENT
• the unit (seconds or hours) starts flashing
• the unit can be changed with the arrow keys and acknowledged with
ENT
32
• after this the programme returns to Mode menu
Page 35
HMP240 series
M210300enUser's Guide
5.3.7 Others
5.3.7.1 Setting the averaging time
Disconnect the security lock jumper!
With command FILT the transmitter can be given the averaging time during
which the individual measurement samples are integrated to get an averaged
reading. The time can be set in seconds within the range of 0 - 1024 (0 = no
averaging time).
• Select More in the main menu, select More in the More menu and
select Filt in the second More menu:
• Pressing ENT returns the programme to the main menu without
changing the filtration time.
• If the filtration time needs to be changed, press CL; enter the new
filtration time with the arrow keys.
5.3.7.2 Setting the pressure for mixing ratio and wet bulb
calculations
The atmospheric pressure has an effect on mixing ratio and wet bulb. Accurate
calculations can be achieved only when the ambient pressure is taken into
consideration.
• Select More in the main menu:
• Select Pressure:
• Pressing ENT returns the programme to the main menu without
changing the pressure reading.
• If the pressure needs to be changed, press CL; enter the new pressure
with the arrow keys
33
Page 36
HMP240
User's GuideM210300en
5.3.7.3 Setting the date
• Select More in the main menu; select Date in the More menu:
• If the date is correct, acknowledge it by pressing ENT; this takes the
programme back to the More menu.
• If the date needs to be changed, press CL.
− first the centuries (19) start flashing; use the arrow keys to change
them and press ENT
− the years (92) start flashing; use the arrow keys to change them and
press ENT
− the months (06) start flashing; use the arrow keys to change them
and press ENT
− the days (17) start flashing; use the arrow keys to change them and
press ENT
5.3.7.4 Setting the time
• Select More in the main menu; select Time in the More menu:
• If the time is correct, acknowledge it by pressing ENT; this takes the
programme back to the More menu.
• If the time needs to be changed, press CL.
− first the hours (14) start flashing; use the arrow keys to change
them and press ENT
− the minutes (25) start flashing; use the arrow keys to change them
and press ENT
− the seconds (32) start flashing; use the arrow keys to change them
and press ENT
34
NOTE
The transmitter does not have a real-time clock with
backup battery. This means that the date and time settings are not permanent.
Page 37
HMP240 series
M210300enUser's Guide
5.3.7.5 Heat on / heat off command
The status of this command should always be HEAT ON, and it should not be
altered. It is meant for service purposes only.
5.4 Serial commands
All available serial commands are described in more detail in Appendix 1. The
following chapters include only the most commonly used command
sequences. See Chapter 4.4 for connecting the HMP240 transmitter to a serial
bus.
Pressing ESC always interrupts any serial command being given. In the commands <cr> stands for carriage return.
5.4.1 Analogue output commands
5.4.1.1 Setting the analogue outputs
Disconnect the security lock jumper!
AMODE a bb.bbb cc.ccc d ee.eee ff.fff <cr>
a=channel 1:U=voltage output
I=current output
bb.bbb =lower limit of channel 1
cc.ccc=upper limit of channel 1
d=channel 2:U=voltage output
I=current output
ee.eee=lower limit of channel 2
ff.fff=upper limit of channel 2
The bb.bbb, cc.ccc, ee.eee and ff.fff parameters are entered in volts or milliamperes.
Example: lower limit of channel 1 is 0 V and upper limit 1 V(U 0 1)
lower limit of channel 2 is 2 V and upper limit 10 V(U 2 10)
>AMODE U 0 1 U 2 10 <cr>
Ch1 : 0.000 ... 1.000 V
Ch2 : 2.000 ... 10.000 V
35
Page 38
HMP240
User's GuideM210300en
5.4.1.2 Selecting and scaling the analogue output quantities
Starts output of measurements to the peripheral devices (RUN mode); the only
command that can be used is S (stop). The output mode can be changed with
command FORM (see Appendix 1).
5.4.2.2 Stopping the measurement output
S<cr>
Ends the RUN mode; after this command all other commands are available.
Page 39
HMP240 series
M210300enUser's Guide
5.4.2.3 Outputting the reading once
SEND <cr> in STOP mode
or
SEND aa <cr>in POLL mode
aa =address of the transmitter when more than one transmitter
is connected to a serial bus (0...99)
The output format depends on which parameters the transmitter can output.
Output types:
The output mode can be changed with command FORM (see Appendix 1).
5.4.2.4 Setting the output interval for the RUN mode
INTV xxx yyy <cr>
xxx=output interval (0...255)
(0 = no pause between outputs)
yyy=unit (s, min or h)
Example: output interval is changed into 10 minutes
>INTV 10 min <cr>
Output intrv. : 10 min
5.4.2.5 Serial bus settings
SERI b p d s x <cr>
b =bauds (300, 600, 1200, 2400, 4800, 9600)
p =parity (n = none, e = even, o = odd)
d =data bits (7 or 8)
s =stop bits (1 or 2)
x =duplex (H = half, F = full)
The settings can be changed one parameter at a time or all parameters at once:
37
Page 40
HMP240
User's GuideM210300en
>SERI O <cr>changing parity only
4800 O 7 1 HDX
>SERI 600 N 8 1 F <cr>changing all parameters
600 N 8 1 FDX
The processor does not allow the following combinations:
• no parity, 7 data bits, 1 stop bit: if this combination is given the
HMP240 programme will change the number of stop bits to 2
• even or odd parity, 8 data bits, 2 stop bits: if this combination is given
the programme changes the number of stop bits to 1
NOTE
The serial bus settings become effective only after reset.
When the half-duplex mode is set, it will automatically turn the echo off. Even
then the ECHO command can indicate that echo is on.
This command is used to set and inspect the averaging time during which the
individual measurement samples are integrated to get an averaged reading.
Page 41
HMP240 series
M210300enUser's Guide
The time can be set in seconds within the range of 0 - 1024 (0 = no averaging
time).
For example:
>FILT
Filter (S):0 ? 1024
>
>FILT 100 <cr>
Filter (S):100
>
5.4.2.8 Setting the transmitter address
ADDR aa <cr>
aa =address (0...99)
Example: transmitter is given address 99
>ADDR <cr>
Address : 2 ? 99 <cr>
5.4.2.9 Setting the calculation mode
Disconnect the security lock jumper!
FROST ON/OFF <cr>
This command is used to select whether the transmitter calculates the
frostpoint (default) or the dewpoint reading at dewpoint temperatures below
0 °C. Select FROST ON for frostpoint and FROST OFF for dewpoint
calculations. For example:
>Frost <cr>
Frost :ON
>Frost off
Frost : OFF
>Frost on
Frost : on
>
5.4.2.10 Resetting the transmitter
RESET <cr>
39
Page 42
HMP240
User's GuideM210300en
5.4.3 Operating the transmitter via the serial bus
5.4.3.1 Setting the serial interface
SMODE xxxx<cr>
xxxx=STOP, RUN or POLL
In STOP mode: measurements output only by command, all commands can be
used
In RUN mode:outputting automatically, only command S can be used
In POLL mode: measurements output only with command SEND. When in
POLL mode, the output mode is changed as follows:
OPEN aa <cr>
SMODE xxxx<cr>
aa=address of the transmitter
xxxx=STOP, RUN or POLL
The OPEN command sets the bus temporarily in STOP mode so that the
SMODE command can be given.
Example:
>SMODE STOP <cr>setting STOP mode
Serial mode : STOP
40
Page 43
HMP240 series
M210300enUser's Guide
5.4.3.2 OPEN & CLOSE
OPEN nn <cr>
nn=address of the transmitter (0...99)
CLOSE <cr>
In STOP mode: command OPEN has no effect, CLOSE sets the transmitter in
POLL mode
In POLL mode: command OPEN sets the transmitter temporarily in STOP
mode, command CLOSE returns the instrument to POLL
mode
Example: relative humidity calibration is performed at transmitter 2 which is
in POLL mode
>OPEN 2 <cr>opens the line to transmitter 2
>CRH <cr>calibration started
...
>CLOSE <cr>line closed
41
Page 44
HMP240
User's GuideM210300en
6. CALIBRATION
The HMP240 transmitter has been fully calibrated at the factory so there
should be no immediate need for recalibration. The transmitter should be
calibrated only if there is reason to believe that the adjustments of the
transmitter have changed. The adjustments of the temperature measurement
channel and the analogue outputs are particularly stable; in normal
circumstances there is no need to recalibrate them. Humidity calibration
should be performed at least once a year.
NOTE
During the calibration procedure, the HMP240 is like
any standard RH transmitter and is therefore calibrated
against known RH values. The warming function is not
active and corrective calculations are not made. It is
therefore essential that the sensor head is allowed
enough time to stabilize to the ambient conditions in
order to ensure the highest possible accuracy in
calibration.
Either a one-point or a two-point calibration is possible. The security lock
jumper has to be disconnected at the moment of taking the transmitter to the
calibration room in order to make sure that the temperature of the sensor head
stabilizes correctly and the warming function does not interfere with the
stabilization. The stabilization of the humidity sensor head can be monitored
by the RH reading or by the T reading: the T reading is actually that of the
humidity sensor head stabilizing to the ambient temperature.
When the security lock jumper is disconnected, the serial port and analogue
outputs show the RH reading and T reading measured directly from the
humidity sensor head (instead of the calculated values). This means that the
readings (RH, T, a, x, dT or Tw) are erroneous until the humidity sensor head
has stabilized to the ambient temperature; the Td reading is correct also during
stabilization. It is recommended that the transmitter is disconnected from the
process during the calibration.
6.1 Humidity calibration
The calibration of the HMP240 transmitter can be checked with a calibrated
Vaisala humidity meter, e.g. the HM34. The HMI38 humidity data processor
with an appropriate reference connection cable can be used as a field calibrator. A two-point calibration can be done with the HMK15 or the HMK13B
calibrator, or the instrument can be sent to Vaisala. We recommend
recalibration at least once a year. The instruments must be recalibrated every
time the HUMICAPHC sensor is changed.
42
Page 45
HMP240 series
M210300enUser's Guide
A ∅ 13.5 mm adapter must be used when calibrating with the HMK13B
calibrator. The adapter (part no. 16611) can be ordered from Vaisala or
Vaisala representatives.
Calibration can be performed by giving the commands with the press switches
inside the housing (see Chapter 5.2), through the serial bus (see Chapter 5.4)
or through the menus on the local display (see Chapter 5.3).
When LED commands are used and when the two analogue channels do not
output either relative humidity and/or temperature, relative humidity is calibrated on channel 1 and temperature is calibrated on channel 2. The calibration ranges are 0...100 %RH and -20...+80 °C. When the transmitters are calibrated at two points, the points must be either 50 %RH or 50 °C apart from
each other.
NOTE
If the transmitter includes the re-gaining option, the
sensor re-gaining must always be done before humidity
calibration. Before starting the calibration make sure
that the temperature of the composite sensor has come
down to ambient temperature (see Appendix 7).
6.1.1 One point humidity calibration
The HMI38 humidity data processor can be used as a one-point field calibrator. A two-point calibration with the HMI38 can also be done, provided that
two separate humidity points with a difference of more than 50%RH between
them are available. The HMI38 is connected to an HMP240 transmitter using
an appropriate reference connection cable. The necessary correction factors
are automatically stored in the HMP240 memory. Detailed calibration
instructions are given in the HMI38 operating manual.
A manual one-point correction can also be done in the field against an accurate reference.
NOTE
If the sensor has been changed, the calibration has to be
done according to the instructions in Chapter 6.1.3.
43
Page 46
HMP240
User's GuideM210300en
6.1.1.1 With serial commands
•Make sure that the sensors of the transmitter and the
reference instrument are close to each other. Allow enough
time for the sensor heads to stabilize to the measurement
conditions
Disconnect the security lock jumper!
•Give command CRH <cr>, enter the humidity value and
press <cr>.
>CRH <cr>
RH : xx.x Ref1 ? yy.y <cr>
Press any key when ready...
•If you want to see how the sensor stabilizes to the refer-
ence humidity, enter c <cr> instead of the first reference:
RH : 11.9 Ref1 ? c <cr>
RH : 11.5 Ref1 ? c <cr>
RH : 11.5 Ref1 ? 11.3 <cr>
Press any key when ready...
•Press any key and <cr> when the transmitter requests the
second point value.
RH : yy.y Ref2 ? <cr>
6.1.1.2 With display / keypad commands
•Make sure that the sensors of the transmitter and the
reference instrument are close to each other. Allow enough
time for the sensor heads to stabilize to the measurement
conditions
Disconnect the security lock jumper!
•Select Cali in the main menu and then RH cal; select Not
changed and then one-point offset correction RH 1 point
cal. Change the humidity reading with the arrow keys to
correspond to the reference value and acknowledge it with
ENT; pressing an arrow once changes the reading by 0.05
%RH.
44
Page 47
HMP240 series
M210300enUser's Guide
6.1.1.3 With LED commands
•Make sure that the sensors of the transmitter and the
reference instrument are close to each other. Allow enough
time for the sensor heads to stabilize to the measurement
conditions
Disconnect the security lock jumper!
•Connect an ammeter/voltmeter to the analogue outputs
(connector X2); if the outputs are already connected e.g. to
a process computer and you do not want to disconnect
them, the current output can be measured at separate test
points located next to connector X15 (see Chapter 7.1).
Give command ¡¡¡l (see Chapter 5.2). At the first
calibration point the LED on the left flashes; adjust the humidity point (offset) with the arrow switches to the reference value. One push of a switch changes the output by
0.05 %RH; the change of the output voltage or current depends on the output scaling. Press ENT switch. The second
LED from left starts flashing; press ENT again.
NOTE
If neither channel normally outputs the RH reading,
during the calibration channel 1 outputs the RH reading
0...100%RH, corresponding to the current/voltage scale
selected for this channel.
6.1.2 Two point humidity calibration
A two-point humidity calibration should be performed in stable conditions
using saturated salt solutions as a reference. Read also the calibrator manual.
NOTE
If the sensor has been changed, the calibration has to be
done according to the instructions in Chapter 6.1.3.
6.1.2.1 With serial commands
•Leave the calibrator and the transmitter for at least 4 hours
in the same space so that their temperatures have time to
equalize. Remove the filter cap on the transmitter.
Disconnect the security lock jumper!
45
Page 48
HMP240
User's GuideM210300en
•Insert the sensor head into the measurement hole of the
LiCl chamber.
•Wait at least 10 minutes for the reading to stabilize.
•Give command CRH <cr>, enter the first point value and
press <cr>.
>CRH <cr>
RH : xx.x Ref1 ? yy.y <cr>
Press any key when ready...
•If you want to see how the sensor stabilizes to the humid-
ity in the calibrator, enter c <cr>:
RH : 11.9 Ref1 ? c <cr>
RH : 11.5 Ref1 ? c <cr>
RH : 11.5 Ref1 ? 11.3 <cr>
Press any key when ready...
•Insert the sensor head into the measurement hole of the
NaCl chamber.
•Wait at least 20 minutes for the reading to stabilize.
• Press any key, enter the second point value and press
<cr>.
RH : xx.x Ref2 ? yy.y <cr>
• The stabilization of the sensor can be monitored by
entering c<cr>.
6.1.2.2 With display / keypad commands
•Leave the calibrator and the transmitter for at least 4 hours
in the same space so that their temperatures have time to
equalize. Remove the filter cap on the transmitter.
Disconnect the security lock jumper!
•Insert the sensor head into the measurement hole of the
LiCl chamber.
•Wait at least 20 minutes for the reading to stabilize.
46
•Select Cali in the main menu and then RH cal; select Not
changed and then two-point calibration RH 2 point cal.
Change the first point reading with the arrow keys to correspond to the reference humidity and press ENT; pressing
an arrow once changes the reading by 0.05 %RH.
Page 49
HMP240 series
M210300enUser's Guide
•Insert the sensor head into the measurement hole of the
NaCl chamber.
•Wait at least 20 minutes for the reading to stabilize.
•If necessary, change the second point reading with arrow
keys and press ENT.
6.1.2.3 With LED commands
•Leave the calibrator and the transmitter for at least 4 hours
in the same space so that their temperatures have time to
equalize. Remove the filter cap on the transmitter.
Disconnect the security lock jumper!
•Insert the sensor head into the measurement hole of the
LiCl chamber.
•Wait at least 10 minutes for the reading to stabilize.
•Connect an ammeter/voltmeter to the analogue outputs
(connector X2). Give command ¡¡¡l. At the first calibration point the LED on the left flashes; adjust the first
point (offset) with the arrow switches to the value given in
the calibration table (Chapter 6.1.4) and press ENT switch.
•Insert the sensor head into the measurement hole of the
NaCl chamber.
•Wait at least 20 minutes for the reading to stabilize.
•Check that the reading corresponds within the desired ac-
curacy to the reading in the calibration table (Chapter
6.1.4). If not, adjust the second point with the arrow
switches to the correct value and press ENT. At the second
calibration point the second LED from the left flashes.
NOTE
If neither channel normally outputs the RH reading,
during the calibration channel 1 outputs the RH reading
0...100 RH%, corresponding to the current/voltage scale
selected.
47
Page 50
HMP240
User's GuideM210300en
6.1.3 Humidity calibration procedure after sensor change
Humidity calibration should be performed in stable conditions using saturated
salt solutions as a reference. If the transmitter has the re-gaining option, the
re-gaining has to be done manually before starting the calibration.
6.1.3.1 With serial commands
•Leave the calibrator and the transmitter for at least 4 hours
in the same space so that their temperatures have time to
equalize. Remove the filter cap on the sensor head.
Disconnect the security lock jumper!
•Insert the sensor head into the measurement hole of the
LiCl chamber.
•Wait at least 20 minutes for the reading to stabilize.
•Give command FCRH <cr>, enter the first point value and
press <cr>:
>FCRH <cr>
RH : xx.x Ref1 ? yy.y <cr>
Press any key when ready...
•The stabilization of the sensor to the reference humidity
can be monitored by entering c <cr>:
RH : 11.9 Ref1 ? c <cr>
RH : 11.5 Ref1 ? c <cr>
RH : 11.5 Ref1 ? 11.3 <cr>
Press any key when ready...
•Insert the sensor head into the measurement hole of the
NaCl chamber.
•Wait at least 20 minutes for the reading to stabilize.
•Press any key, enter the second point value and press <cr>.
RH : xx.x Ref2 ? yy.y <cr>
•The stabilization of the sensor can be monitored by
entering c <cr>.
48
6.1.3.2 With display / keypad commands
•Leave the calibrator and the transmitter for at least 4 hours
in the same space so that their temperatures have time to
equalize. Remove the filter cap on the sensor head.
Disconnect the security lock jumper!
Page 51
HMP240 series
M210300enUser's Guide
•Insert the sensor head into the measurement hole of the
LiCl chamber.
•Wait at least 20 minutes for the reading to stabilize.
•Select Cali in the main menu and then RH cal; select Sen-
sor changed. Change the first point reading with the arrow keys and press ENT.
•Insert the sensor head into the measurement hole of the
NaCl chamber.
•Wait at least 20 minutes for the reading to stabilize.
•If necessary, change the second point reading with arrow
keys and press ENT.
6.1.3.3 With LED commands
•Leave the calibrator and the transmitter for at least 4 hours
in the same space so that their temperatures have time to
equalize. Remove the filter cap on the sensor head.
Disconnect the security lock jumper!
•Insert the sensor head into the measurement hole of the
LiCl chamber.
•Wait at least 20 minutes for the reading to stabilize.
•Connect an ammeter/voltmeter to the analogue outputs
(connector X2). Give command l¡¡¡. At the first calibration point the LED on the left flashes; adjust the first
point with the arrow switches to the value given in the
calibration table (Chapter 6.1.4) and press ENT switch.
•Insert the sensor head into the measurement hole of the
NaCl chamber.
•Wait at least 20 minutes for the reading to stabilize.
•Check that the reading corresponds within the desired ac-
curacy to the reading of the calibration table (Chapter
6.1.4). If not, adjust the second point with the arrow
switches to the correct value and press ENT. At the second
calibration point the second LED from the left flashes.
NOTE
If neither channel normally outputs the RH reading,
during the calibration channel 1 outputs the RH reading
0...100%RH, corresponding to the current/voltage scale
selected for this channel.
49
Page 52
HMP240
User's GuideM210300en
The basic capacitance of the new sensor may differ considerably from that of
the previous one. Therefore the corresponding humidity reading of the
transmitter may be below 0 %RH at the low or above 100 %RH at the high
calibration point. However, the current/voltage reading of the analogue output
shows only the minimum or maximum value of the selected current/voltage
scale and the output value may not change even though the arrow switches are
pressed several times. If this happens, press the up or down arrow switch
continuously to bring the output back into the selected scale; this may take as
long as half a minute.
6.1.4 Humidity calibration table
Temperature°C1520253035
°F5968778695
LiCl%RH*)11.311.311.311.3
4...20 mA5.815.815.815.81
0...20 mA2.262.262.262.26
0...1 V0.1130.1130.1130.113
0...5 V0.5650.5650.5650.565
0...10 V1.131.131.131.13
NaCl%RH75.675.575.375.174.9
4...20 mA16.1016.0816.0516.0215.98
0...20 mA15.1215.1015.0615.0214.98
0...1 V0.7560.7550.7530.7510.749
0...5 V3.7803.7753.7653.7553.745
0...10 V7.567.557.537.517.49
Table 1Greenspan's calibration table
*) If the LiCl solution is used or stored at temperatures below +18 °C
(+64 °F), its equilibrium humidity changes permanently.
6.2 Temperature calibration
The temperature channel has been calibrated at the factory and since it is very
stable, calibration should be performed only when there is strong reason to
believe that the adjustments have changed.
Temperature calibration should be done against some accurate temperature
reference. It can be done either by using the press switches inside the housing,
through the serial bus or the menus on the local display. Either a one-point
offset correction or a two-point calibration are possible.
50
NOTE
If neither channel normally outputs the T reading,
during the calibration channel 2 outputs the T reading
-20...+80°C, corresponding to the current/voltage scale
selected for this channel.
Page 53
HMP240 series
M210300enUser's Guide
6.2.1 One point offset calibration
6.2.1.1 With serial commands
•Leave the reference instrument and the transmitter for at
least 4 hours in the same space so that their temperatures
have time to equalize. Remove the filter cap prior to
calibration.
Disconnect the security lock jumper!
•Check the transmitter against the reference.
•After giving the correct temperature value (Ref1) and
pressing <cr>, press any key and then <cr>.
6.2.1.2 With display / keypad commands
•Leave the reference instrument and the transmitter for at
least 4 hours in the same space so that their temperatures
have time to equalize. Remove the filter cap prior to
calibration.
Disconnect the security lock jumper!
•Check the transmitter against the reference.
•Select T 1 point cal (or Ta 1 point cal), change the
reading to correspond to the reference and press ENT.
6.2.1.3 With LED commands
•Leave the reference instrument and the transmitter for at
least 4 hours in the same space so that their temperatures
have time to equalize. Remove the filter cap prior to
calibration.
Disconnect the security lock jumper!
•Check the transmitter against the reference.
• After adjusting the offset point and pressing ENT, the sec-
ond LED from left flashes. Press ENT without changing
the output value.
51
Page 54
HMP240
User's GuideM210300en
6.2.2 Two point temperature calibration
6.2.2.1 With serial commands
•Leave the reference instrument and the transmitter for at
least 4 hours in the same space so that their temperatures
have time to equalize. Remove the filter cap prior to
calibration.
Disconnect the security lock jumper!
•Check the transmitter against the reference.
•For the humidity sensor head: give command CT <cr>,
enter the first point value and press <cr>:
>CT <cr>
T : xx.x Ref1 ? yy.y <cr>
Press any key when ready
• If you want to see how the sensor stabilizes to the refer-
ence temperature, enter c <cr> instead of the first reference:
T : 0.90 Ref1 ? c <cr>
T : 0.55 Ref1 ? c <cr>
T : 0.55 Ref1 ? 0.0 <cr>
Press any key when ready...
•For the additional temperature sensor head: give command
CTA <cr>, enter the first point value and press <cr>:
>CTA<cr>
Ta: -20.07 Ref1 ? -20.08
Press any key when ready...
•Change the temperature and check the transmitter again
against the reference.
•Check that the reading corresponds to the reading of the
reference instrument. If not, adjust the second point.
•Press any key, enter the second point value and press <cr>.
T (or Ta) : xx.x Ref2 ? yy.y <cr>
•The stabilization of the sensor can be monitored by
entering c <cr> instead of the reference value.
52
Page 55
HMP240 series
M210300enUser's Guide
6.2.2.2 With display / keypad commands
•Leave the reference instrument and the transmitter for at
least 4 hours in the same space so that their temperatures
have time to equalize. Remove the filter cap prior to
calibration.
Disconnect the security lock jumper!
•Check the transmitter against the reference.
• Select Cali in the main menu and then T (Ta for the
additional T sensor head); select two-point calibration T
(or Ta) 2 point cal. Change the first point reading with the
arrow keys and press ENT.
•Change the temperature and check the transmitter again
against the reference.
•Check that the reading corresponds to the reading of the
reference instrument. If not, adjust the second point.
•If necessary, change the second point reading with the ar-
row keys and press ENT.
6.2.2.3 With LED commands
•Leave the reference instrument and the transmitter for at
least 4 hours in the same space so that their temperatures
have time to equalize. Remove the filter cap prior to
calibration.
Disconnect the security lock jumper!
•Check the transmitter against the reference.
• Connect an ammeter/voltmeter to the analogue outputs
(connector X2). Give command ¡¡l¡ (or ¡l¡¡ for
the T sensor head). At the first calibration point the LED
on the left flashes; adjust the first point (offset) with the
arrow switches to the same reading with the reference and
press ENT switch.
•Change the temperature and check the transmitter again
against the reference.
•Check that the reading corresponds to the reading of the
reference instrument. If not, adjust the second point.
•If necessary, adjust with the arrow switches to the correct
value and press ENT. At the second calibration point the
second LED from the left flashes.
53
Page 56
HMP240
User's GuideM210300en
6.3 Calibration of analogue outputs
The analogue outputs have been calibrated at the factory and since they are
very stable, calibration should be performed only when there is reason to
believe that their adjustments have changed.
6.3.1 With serial commands
Disconnect the security lock jumper!
ACAL <cr>
The outputs on channels 1 and 2 are measured and the measured values (mA
or V) entered as calibration coefficients.
Example: both channels have 0...10 V outputs (set with AMODE command);
enter the voltages measured at the analogue outputs:
>ACAL <cr>
Ch1 U1 ( V ) ? 0.123 <cr>
Ch1 U2 ( V ) ? 9.98 <cr>
Ch2 U1 ( V ) ? 0.120 <cr>
Ch2 U2 ( V ) ? 9.98 <cr>
6.3.2 With display / keypad commands
Disconnect the security lock jumper!
• Connect an ammeter/voltmeter to the output of channel 1, select Cali
in the main menu and Analog outputs in the Cali menu. The following is displayed (the quantity can be either mA or V):
• Enter the measured lower end current/voltage on channel 1:
54
• Enter the measured upper end current/voltage on channel 1:
• Connect the meter to the output of channel 2 and enter the measured
lower end current/voltage on channel 2:
Page 57
HMP240 series
M210300enUser's Guide
• Enter the measured upper end current/voltage on channel 2.
6.3.3 With LED commands
If both the analogue outputs and humidity/temperature channels are calibrated,
the analogue outputs should be calibrated first. This applies only when the
calibrations are done with LED commands!
• connect an ammeter/voltmeter to the analogue outputs (connector X2)
Disconnect the security lock jumper!
• Give command ¡¡ll.
• the LED on the left flashes; set the low end of channel 1 with the ar-
row keys and press ENT
• the second LED from the left flashes; set the high end of channel 1
with the arrow keys and press ENT
• the LED on the left flashes; set the low end of channel 2 with the ar-
row keys and press ENT
• the second LED from the left flashes; set the high end of channel 2
with the arrow keys and press ENT
The analogue outputs are calibrated to ensure the correct scaling: for example,
when the output is scaled to 4...20 mA, the low end of the scale is 4 mA and
high end 20 mA exactly. However, when 0... 20 mA output is used, the output
can not be adjusted to exactly 0 mA, but to 50 µA. When 0...1 V, 0...5 V or
0...10 V output is in use, the output is adjusted to 50 mV. The following table
summarizes the correct output values.
Summary of the correct output values:
Output scale:
0...20 mA4...20 mA0...1 V0...5 V0...10 V
low end:50 µA4 mA50 mV50 mV50 mV
high end:20 mA20 mA1 V5 V10 V
55
Page 58
HMP240
User's GuideM210300en
7. MAINTENANCE
7.1 Reference measurements
Reference measurements are necessary for the verification of whether the
transmitter readings are within specifications. This way the user can check if
the transmitter needs calibration or service. If the transmitter can be taken out
of the process or the control system, perform the reference measurement in a
laboratory with stable conditions.
The reference meter should preferably use the same technology as the instrument checked, i.e. transmitters as HMP240 which use a capacitive RH sensor,
should be checked with instruments that also use capacitive RH sensors. This
minimizes the risk of errors caused by different measurement techniques.
Whatever technique is used, make sure that the reference instrument is at the
same ambient conditions as the checked instrument in order to avoid errors
caused by temperature differences. The reference measurement should be
made as close to the checked sensor as possible and the readings should be
read at the same time, when possible.
When an RH reference meter is used, disconnect the security lock jumper and
wait until the temperature of the probe has stabilized to the same temperature
with the reference meter. When the security lock jumper is disconnected, the
HMP240 is like any other RH transmitter outputting the RH and T values (for
further details, see Chapter 6).
7.2 Self-diagnostics
The HMP240 transmitter goes through a self-diagnostics procedure when the
power is switched on. If the procedure does not reveal any errors or faults, the
transmitter starts operating normally. If errors or faults are found, first check
whether the humidity and temperature sensors are damaged. If they are intact,
send the transmitter to Vaisala or a Vaisala representative for repairs. The
error messages the transmitter outputs are listed in Appendix 4.
If any errors occur during operation, the error messages are output on the local
display if the transmitter displays measurements; if the menus are used, error
messages are not output. The LEDs indicate errors at all times. During operation, however, the error messages are not output automatically through the serial interface. If there is any reason to doubt that there is something wrong
with the transmitter, use command ERRS:
56
ERRS <cr>
If there are no error messages, only a prompt is displayed:
>ERRS <cr>
Page 59
HMP240 series
M210300enUser's Guide
>
When errors have occurred, the transmitter outputs an error code (see Appendix 4 for all error messages):
>ERRS <cr>
E40 f ( all ) out of range
>
7.3 Replacing the composite sensor
The humidity and temperature sensors (HUMICAP H-sensor and Pt100)
have been glued together. If one or both sensors are damaged, the whole
composite sensor must be changed.
• Cut the pins of the Pt100 sensor at the top.
• Lift the humidity sensor gently.
• Loosen the pins of the Pt100 sensor by soldering.
• Remove the old tin from the connectors.
• Put a new humidity sensor in place.
• Bend the pins of the Pt100 sensor carefully into the connectors.
• Fasten the pins by soldering them with Multicore 96S-Flux 381.
• Make sure the solder joints are good and check that the pins are not
short-circuited.
• Remove possible splashes from the solder joint with distilled water,
but take care not to damage the humidity sensor.
After sensor change the humidity calibration must be done according to the
instructions in Chapter 6.1.3.
Replace a dirty membrane or sintered filter to ensure a maximum lifetime for
the sensor. Do not try to clean the filter. The sensor can be cleaned with distilled water; if this does not work, replace the sensor.
7.4 Temperature channel (additional) adjustment with Pt 100 simulators
Switch the power off and disconnect the wires to the Pt 100 sensor from solder
lugs TP1, TP2 and TP3.
57
Page 60
HMP240
User's GuideM210300en
TP3
TP2
8
X8
X77
TP1
Figure 7.1Location of solder lugs TP1, TP2 and TP3 and connector
X77
Connect a Pt 100 simulator to connector X77 and set it at the lowest temperature to be calibrated.
Pt 100
X77
Figure 7.2Connecting the Pt 100 simulator to connector X77
Switch the power on.
7.4.1 With serial commands
Give command CTA, enter the first point value and press <cr>:
>CTA <cr>
Ta : xx.x Ref1 ? yy.y <cr>
Press any key when ready
Set the Pt 100 simulator at the highest temperature to be calibrated and press
any key. Enter the second point (gain) reference reading.
7.4.2 With display commands
Select Cali in the main menu and then Ta; select two-point calibration Ta 2
point cal. Change the first point reading with the arrow keys and press ENT.
58
Set the Pt 100 simulator at the highest temperature to be calibrated and adjust
the second point (gain) to the reference reading.
7.4.3 With LED commands
Connect an ammeter/voltmeter to the analogue outputs (connector X2). Give
command ¡l¡¡ and adjust the first point (offset) with the arrow switches to
the same reading with the reference; press ENT switch.
Set the Pt 100 simulator at the highest temperature to be calibrated and adjust
the second point (gain) to the reference reading.
Page 61
HMP240 series
M210300enUser's Guide
Disconnect the Pt 100 simulator and reconnect the Pt 100 wires to solder lugs
TP1, TP2 and TP3.
The correct connections according to wire colours are:
TP1TP2TP3
bluewhitered
If a Pt 100 simulator is not available, the adjustment can be made with two
resistors of 84 Ω and 154 Ω with precisely known resistance. Measure
resistors with a resistance meter. Look up the corresponding temperature value
from a Pt 100 conversion table or calculate it according to the following
equation:
7.5 Temperature channel adjustment with Pt 100 simulators
(composite sensor)
Switch the power off and disconnect the wires to the Pt 100 sensor from solder
lugs TP5, TP6 and TP7.
TP7
TP6
8
X77
X8
TP5
Figure 7.3Location of solder lugs TP5, TP6 and TP7 and connector
X88
Connect a Pt 100 simulator to connector X88 and set it at the lowest temperature to be calibrated.
59
Page 62
HMP240
User's GuideM210300en
Pt 100
X88
Figure 7.4Connecting the Pt 100 simulator to connector X88
Switch the power on.
7.5.1.1 With serial commands
Give command CT, enter the first point value and press <cr>:
>CT <cr>
C : xx.x Ref1 ? yy.y <cr>
Press any key when ready
Set the Pt 100 simulator at the highest temperature to be calibrated and press
any key. Enter the second point (gain) reference reading.
7.5.1.2 With display commands
Select Cali in the main menu and then T; select two-point calibration T 2 point
cal. Change the first point reading with arrow keys and press ENT.
Set the Pt 100 simulator at the highest temperature to be calibrated and adjust
the second point (gain) to the reference reading.
7.5.2 With LED commands
Connect an ammeter/voltmeter to the analogue outputs (connector X2). Give
command ¡¡l¡ and adjust the first point (offset) with arrow switches to the
same reading with the reference and press ENT switch.
Set the Pt 100 simulator at the highest temperature to be calibrated and adjust
the second point (gain) to the reference reading.
Disconnect the Pt 100 simulator and reconnect the Pt 100 wires to solder lugs
TP5, TP6 and TP7.
The correct connections according to wire colours are:
TP5TP6TP7
HMP240bluegreenyellow
7.6 Measurement of output currents using test points
If a current output has been connected e.g. to a process computer, the output
current cannot be measured at the output connector X2 without disconnecting
60
Page 63
HMP240 series
M210300enUser's Guide
the external load. The output current can, however, be measured at test points
CH1+/CH1- and CH2+/CH2- without disconnecting the output wires. These
test points can therefore be used in one point offset correction against an accurate reference or in checking the current output without disconnecting the
analogue output from the process.
CH1+
CH1-
CH2+
CH2-
X2
CH1 +
X15
+
CH2
OPENED COVER OF THE HMP 243
Figure 7.5Location of the CH1 and CH2 test points
mA
X2
CH1+
TEST POINTS
CH1-
Figure 7.6Circuit diagram of the analogue output current test points
7.7 Adjusting the contrast of the display
CH1+
CH1-
R
L
The contrast of the display can be adjusted using the trimmer "LCD display
contrast" located next to press switches.
61
Page 64
HMP240
Difference between ambient temperature and dewpoint temperature
Accuracy in dewpoint
User's GuideM210300en
8. TECHNICAL DATA
SensorRH & T composite sensor
HUMICAP®KC(part no.
18258HM)
T sensor in the humidity sensor headPt 100 IEC 751 class 1/3 B
Voltage proof, AC: DC supply (+ or -) to housing 250 VAC, 1 minute
(300 kΩ and 15 nF parallel)
8.1 Dewpoint temperature
typical range of dewpoint temperature -40...+100 °C
Response time (90 %) at +20 °C in15 s
still air (with sintered filter)
Dewpoint temperature (°C)
6
5
4
0
-20
20
40
60-40
80
3
2
1
measurement (°C)
100
0
020406080100
Dewpoint difference (°C)
Figure 8.1Accuracy in dewpoint measurement as a function of
depression and dewpoint (NOTE: dewpoint < 0 °C = calculation of
frostpoint)
62
Page 65
HMP240 series
M210300enUser's Guide
8.2 Temperature (with additional T sensor head)
Measurement range-40...+180 °C
Typical accuracy at +20 °C (+68 °F) ±0.1 °C
Typical temperature dependence of
electronics0.005 °C/°C
Temperature sensor Pt 100 RTD DIN IEC 751,
class 1/4 B
8.3 Calculated variables
Without the additional temperature head only dewpoint temperature and
mixing ratio are available.
Typical ranges
relative humidity0...100 %RH
dewpoint difference 0...+50 °C
mixing ratio0...500 g/kg dry air
absolute humidity0...600 g/m
3
wet bulb temperature0...+100 °C
Response time (90 %) at +20 °C in
still air (with sintered filter)15 s
Accuracies of the calculated quantities depend on the accuracies of the measured quantities: relative humidity and temperature. In the dewpoint accuracy
graph and accuracy tables for x, a and Tw the accuracies of relative humidity
and temperature are assumed to be ±(0.5 + 2.5 % of the reading) %RH for
humidity and 0.1°C for temperature (at 25°C). Note that the user can select
whether the HMP240 calculates the frostpoint (default) or the dewpoint when
the dewpoint is < 0 °C. In addition, the HMP240 also calculates the dewpoint
difference, i.e. the difference between the ambient temperature and the
dewpoint ( = dT).
8.3.1 Relative humidity (with additional T sensor head)
Accuracy (including nonlinearity and repeatability)
maximum achievable accuracy when calibrated against high quality,
certified humidity standards:+(0.5 + 2.5 % of the reading) % RH
for example, accuracy at:
0 % RH+0.5 % RH
60 % RH+2.0 % RH
Response time (90 %) at +20 °C in
still air (with sintered filter)15 s
63
Page 66
HMP240
User's GuideM210300en
8.3.2 Accuracy of other calculated variables
Accuracy of mixing ratio (g/kg) when ambient pressure is 1013 mbar
Weight with
HMP240 (duct mounting, max
180°C, without T probe)
T probe for HMP240 80 g150 g250 g
HMP243FA (duct installation
kit for HMP240; 180°C)
2 x HMP243FA 280 g
Weight of display cover420 g
8.7 Electromagnetic compatibility
Complies with the following standard:
EN 61326-1:1997 + Am 1:1998, Electrical equipment for measurement,
control and laboratory use - EMC requirements, Industrial environment.
Filters of the sensor: PPS grid & stainless steel netting filter (16720) and
sintered steel filter (16452).
Test methods
2 m cable5 m cable10 m cable
1100 g1300 g1500 g
140 g
Emission
Radiated emissionsCISPR16 class B (CISPR22 Class B)
Harmonic currents to AC mains EN/IEC 61000-3-2
Voltage fluctuationsEN/IEC 61000-3-3
HMP243FAMounting kit for HMP240, +180 °C cable
16452Sintered filter, stainless steel
16720PPS grid with stainless steel netting
HUMICAP®KCComposite humidity and temperature sensor
16611Calibration adapter for the HMK13B Calibrator
HMP240RSRS 485/422 serial module
HMP247CGCable gland for HMP243 and HMP247
HMP247TBSThread bar set for HMP247
HMP240CLCurrent loop module
The commands function as described when the serial interface is in fullduplex mode and echo is on. All commands except FORM can be given in
either capital or small letters.
In the commands <cr> means carriage return, <lf> line feed and <ht> horizontal tabulation.
69
Page 72
HMP240
Appendix 1: Serial commandsM210300en
1. ANALOGUE OUTPUT COMMANDS
AMODESetting the analogue outputs
Disconnect the security lock jumper!
AMODE a bb.bbb cc.ccc d ee.eee ff.fff <cr>
a=channel 1:U=voltage output
I=current output
bb.bbb =lower limit of channel 1
cc.ccc=upper limit of channel 1
d=channel 2:U=voltage output
I=current output
ee.eee=lower limit of channel 2
ff.fff=upper limit of channel 2
The bb.bbb, cc.ccc, ee.eee and ff.fff parameters are entered in volts or milliamperes.
Sets the analogue outputs on channels 1 and 2. An example of this is when the
voltage output on channel 1 is set to be 0...1 V and channel 2 set to 2...10 V:
>AMODE U 0 1 U 2 10 <cr>
Ch1 : 0.000 ... 1.000 V
Ch2 : 2.000 ... 10.000 V
The current settings can be checked by giving the command without any parameters:
>AMODE <cr>
Ch1 : 0.000 ... 20.000 mA
Ch2 : 0.000 ... 20.000 mA
ASELSelecting the scaling the analogue output quantities
With this command the transmitter can be calibrated against a reference. A
one-point calibration can be done against an accurate transfer standard in the
field and a two-point calibration using saturated salt solutions in controlled
conditions. A two-point calibration is performed as follows:
If the stabilization of the sensor to the humidity in the calibrator needs to be
monitored, the measurement output can be repeated by giving command c<cr>
at Ref1 and Ref2:
>CRH <cr>
RH : 12.00 Ref1 ? c <cr>
RH : 11.70 Ref1 ? c <cr>
RH : 11.50 Ref1 ? 11.3 <cr>
Press any key when ready ...
RH : 76.00 Ref2 ? 75.5 <cr>
FCRHRelative humidity calibration after sensor change
Disconnect the security lock jumper!
FCRH <cr>
After humidity sensor change the transmitter must be calibrated using this
command and the calibration must be done at two reference points. The calibration is performed as follows:
>FCRH <cr>
RH : 1.90Ref1 ?11.3 <cr>
Press any key when ready ...
RH : 76.30 Ref2 ?74.9 <cr>
73
Page 76
HMP240
Appendix 1: Serial commandsM210300en
The command can also be divided into two commands, so the computer can be
used for other purposes while waiting for the sensor to stabilize to the higher
humidity.
>FCRH 1 <cr>
RH : 1.90Ref1 ?11.3 <cr>
FCRH 2 <cr> ...
RH : 76.30 Ref2 ?74.9 <cr>
If the stabilization of the sensor to the humidity in the calibrator needs to be
monitored, the measurement output can be repeated by giving command c<cr>
at Ref1 and Ref2:
>FCRH <cr>
RH : 12.00 Ref1 ? c <cr>
RH : 11.70 Ref1 ? c <cr>
RH : 11.50 Ref1 ? 11.3 <cr>
Press any key when ready ...
RH : 76.00 Ref2 ? 75.5 <cr>
CTTemperature calibration of the humidity sensor head
Disconnect the security lock jumper!
CT <cr>
With this command the humidity sensor head can be calibrated against an
accurate reference, such as a Pt 100 simulator. A two-point calibration is performed as follows:
>CT <cr>
T : 0.80Ref1 ?0.0 <cr>
Press any key when ready ...
T : 56.20Ref2 ?55.0 <cr>
In one-point offset correction, the Ref2 prompt is acknowledged with <cr>:
>CT <cr>
T : 0.80Ref1 ?0.0 <cr>
Press any key when ready ...
T : 75.50Ref2 ?<cr>
If the stabilization of the sensor to the temperature of the calibrator or the reference needs to be monitored, the measurement output can be repeated by giving command c<cr> at Ref1 and Ref2:
74
>CT <cr>
T : 0.80Ref1 ?c <cr>
T : 0.40Ref1 ?0.00 <cr>
Press any key when ready ...
T : 56.20Ref2 ?55.0 <cr>
Page 77
HMP240
M210300enAppendix 1: Serial commands
CTATemperature calibration of the optional T sensor head
Disconnect the security lock jumper!
CTA <cr>
With this command the optional temperature sensor head can be calibrated
against an accurate reference, such as a Pt 100 simulator. A two-point
calibration is performed as follows:
>CTA <cr>
Ta : 0.80Ref1 ?0.0 <cr>
Press any key when ready ...
Ta : 56.20 Ref2 ?55.0 <cr>
In one-point offset correction, the Ref2 prompt is acknowledged with <cr>:
>CTA <cr>
Ta : 0.80Ref1 ?0.0 <cr>
Press any key when ready ...
Ta : 75.50 Ref2 ?<cr>
If the stabilization of the sensor to the temperature of the calibrator or the reference needs to be monitored, the measurement output can be repeated by giving command c<cr> at Ref1 and Ref2:
>CTA <cr>
Ta : 0.80Ref1 ?c <cr>
Ta : 0.40Ref1 ?0.00 <cr>
Press any key when ready ...
Ta : 56.20 Ref2 ?55.0 <cr>
ACALCalibrating the analogue outputs
Disconnect the security lock jumper!
ACAL <cr>
Calibrates the outputs selected on channels 1 and 2. The output is measured
and the measured values (mA or V) entered as calibration coefficients.
For example, calibrating the outputs when 0...10 V signal has been selected on
both channels (set with AMODE command)
where: Ts = temperature of the humidity sensor head
Ta = temperature of the temperature sensor head
The factory settings are offset 0 and gain 1. The transmitter can be returned to
its factory calibration by giving these values .
NOTE
76
The temperature unit in offset correction is always degrees Centigrade, even if the transmitter is using nonmetric units (Fahrenheit) in its measurement output.
Page 79
HMP240
M210300enAppendix 1: Serial commands
3. OUTPUT VIA THE SERIAL BUS
RStarting the measurement output
R <cr>
Starts output of measurements to the peripheral devices (PC display or
printer); output interval is set with command INTV. The output format
depends on the transmitter configuration and variables used. The sequence,
however, is always the same: relative humidity, temperature, dewpoint
temperature, absolute humidity, mixing ratio, wet bulb temperature and
dewpoint difference. An example:
When the transmitter sends out the readings, the serial interface does not echo
any commands; the only command that can be used is S (stop).
The output mode can be changed with command FORM.
SStopping the measurement output
S<cr>
Ends the RUN state; after this command all other commands are available.
SENDOutputting a reading once
SEND <cr> in STOP mode
or
SEND aa <cr>in POLL mode
aa=address of the transmitter when more than one transmitter is
connected to a serial bus (0...99; set with command ADDR)
Outputs the current measurement readings via the serial line. The output format depends on which parameters the transmitter can output. Output types are:
DSENDOutputting readings of all connected transmitters once
DSEND <cr>
All transmitters connected to the serial bus send their addresses and current
measurement readings in a sequence defined by their addresses. After
receiving DSEND command a transmitter sets a delay time according to its
address value and sends the data after this delay. DSEND works also in POLL
mode. With this command the user can, for example, easily find out the
addresses of the transmitters.
The output when four transmitters with addresses 4, 5, 10, 33 have been connected to the serial bus:
During operation, error messages are not automatically output through the serial interface. If there is any reason to doubt that there is something wrong
with the transmitter, possible error messages can be output with command
ERRS.
If there are no error messages, only a prompt is displayed:
>ERRS <cr>
>
If errors have occurred, the transmitter outputs the error code (see Appendix 4
for error messages):
>ERRS <cr>
E40 f ( all ) out of range
>
78
Page 81
HMP240
M210300enAppendix 1: Serial commands
ECHOTurning the serial interface echo ON/OFF
ECHO xxx <cr>
xxx=ON or OFF
When the echo is off, the commands given through the serial interface or the
prompt > cannot be seen on the display.
When the serial interface is in half-duplex mode, the echo is always off. Even
then the ECHO command can indicate that echo is on.
INTVSetting the output interval for the RUN mode
INTV xxx yyy <cr>
xxx=output interval (0...255)
0: no pause between outputs
yyy=unit (s, min or h)
Sets the output interval when the transmitter outputs measurement readings to
a peripheral device.
For example, the currently valid settings are output with:
>INTV <cr>
Output intrv. : 0 min
When this is changed into 10 minutes, the command is:
>INTV 10 <cr>
Output intrv. : 10 min
The unit is changed into seconds with:
>INTV S <cr>
Output intrv. : 10 s
The change can also be done with one command:
>INTV 10 S <cr>
Output intrv. : 10 s
79
Page 82
HMP240
Appendix 1: Serial commandsM210300en
FORMSetting the output format
FORM <cr>
"xxx...xxx"
? zzz...zzz <cr>
xxx...xxx=old format
zzz...zzz=new format
The FORM command sets the format of the outputs generated in RUN state
and by SEND command. Please note that capital and small letters have dif-ferent meanings.
\UU..UU\relative humidity
\TT..TT\temperature
\DD..DD\dewpoint temperature
\AA..AA\absolute humidity
\XX..XX\mixing ratio
\WW..WW\ wet bulb temperature
\dd..d\dewpoint difference
\uu..uu\unit according to the preceding variable
\nline feed <lf>
\rcarriage return <cr>
\thorizontal tabulation <ht> or <tab>
\\\
When FTIME is activated, the current time is output at the beginning of the
output line. The time is set with command TIME. After RESET or power on
the current time is 00:00:00.
When FDATE is activated, the current date is output at the beginning of the
output line. The time is set with command DATE. After RESET or power on
the current date is 1991-01-01.
b=bauds (300, 600, 1200, 2400, 4800, 9600)
p=parity (n = none, e = even, o = odd)
d=data bits (7 or 8)
s=stop bits (1 or 2)
x=duplex (H = half, F = full)
Giving the command on its own outputs the current settings:
>SERI <cr>
4800 E 7 1 FDX
The settings can be changed one parameter at a time or all parameters at once:
>SERI O H <cr>changing parity and duplex
4800 O 7 1 HDX
>SERI 600 N 8 1 F <cr>changing all parameters
600 N 8 1 FDX
82
NOTE
The serial bus settings become effective only after reset.
Page 85
HMP240
M210300enAppendix 1: Serial commands
The processor does not allow the following combinations:
• no parity, 7 data bits, 1 stop bit: if this combination is given the
HMP240 programme will change the number of stop bits to 2
• even or odd parity, 8 data bits, 2 stop bits: if this combination is given
the programme changes the number of stop bits to 1
When the half-duplex mode is set, it will automatically turn the echo off. Even
then the ECHO command can indicate that echo is on.
UNITSelecting the output units
UNIT x <cr>
x=m(etric units) n(on-metric units)
metric unitsnon-metric
units
RH%RH%RH
T
Td
dT
ag/m
°C°F
°C°F
°C°F
3
gr/ft
3
xg/kggr/lb
Tw
°C°F
For example, the command for setting the non-metric units is:
>UNIT N <cr>
Output units : non metric
When the command is given with no parameters, the transmitter outputs the
currently valid setting.
ADDRSetting the transmitter address
ADDR aa <cr>
aa=address (0...99)
The address is used when more than one transmitter is connected to one serial
bus. The ADDR command makes it possible to communicate with one
transmitter at a time in POLL state.
For example, transmitter is given address 99
>ADDR <cr>
Address : 2 ? 99 <cr>
83
Page 86
HMP240
Appendix 1: Serial commandsM210300en
When asking the current address, no address number is given:
>ADDR <cr>
Address : 2 ? <cr>
FROSTSetting the calculation mode
Disconnect the security lock jumper!
FROST ON/OFF <cr>
This command is used to select whether the transmitter calculates the
frostpoint (default) or the dewpoint reading at dewpoint temperatures below
0 °C. Select FROST ON for frostpoint and FROST OFF for dewpoint
calculations. For example:
>Frost <cr>
Frost :ON
>Frost off
Frost : OFF
>Frost on
Frost : on
>
RESETResetting the transmitter
Resets the transmitter. All settings that have been changed remain in the memory even after reset or power failure.
3.1 Operation modes
SMODESetting the serial interface
xxxx=STOP, RUN or POLL
In STOP mode: measurements output only by command, all commands can be
used
In RUN mode:outputting automatically, only command S can be used
RESET <cr>
SMODE xxxx<cr>
84
In POLL mode: measurements output only with command SEND. When in
POLL mode, the output state is changed as follows:
OPEN aa <cr>
Page 87
HMP240
M210300enAppendix 1: Serial commands
SMODE xxxx<cr>
aa=address of the transmitter
xxxx=STOP, RUN or POLL
The OPEN command sets the bus temporarily in STOP MODE so that the
SMODE command can be given. For example:
>SMODE <cr>which mode is in use at the moment
Serial mode : STOP
>SMODE STOP <cr>setting STOP mode
Serial mode : STOP
OPEN & CLOSE
OPEN nn <cr>
nn=address of the transmitter (0...99)
CLOSE <cr>
In STOP mode: command OPEN has no effect, CLOSE sets the transmitter in
POLL mode
In POLL mode: command OPEN sets the transmitter temporarily in STOP
mode, command CLOSE returns the instrument to POLL
mode
When more than one transmitter is connected to the same serial bus, the POLL
mode makes it possible to communicate with the transmitters. For example, a
relative humidity calibration is performed at transmitter 2 (<bel> = ASCII 7):
>OPEN 2 <cr>
<cr><lf> 'HMP nn line opened for operator commands'
<cr><lf><lf><bel>
>CRH <cr>
...
>CLOSE <cr>
<cr><lf> 'line closed' <cr><lf>
85
Page 88
HMP240
Appendix 1: Serial commandsM210300en
4. OTHERS
ITESTTesting the analogue outputs
ITEST <cr>
or
ITEST a b <cr>
a=current/voltage of channel 1
b=current/voltage of channel 2
The operation of the analogue outputs can be tested by forcing the outputs to
given values which can then be measured with a current/voltage meter from
the analogue outputs. The response to ITEST command gives six outputs/parameters. Only the first two are relevant; they show the channel current
or voltage in mA or V. The other four figures contain information for service
purposes only.
Examples:
• reading the channel outputs and parameters
>itest <cr>
1.9438 2.3483 1.00694 10.64634 1.97374 2.17665
>
• forcing outputs 0.5 V and 4 V to channels 1 and 2
>itest 0.5 4 <cr>
0.5000 4.0000 1.00694 10.62970 1.23336 3.01722
>
• releasing the forced control and reading the outputs
>itest <cr>
1.9427 2.3392 1.00731 10.62428 1.97157 2.16978
>
FILTSetting the averaging time
Disconnect the security lock jumper!
FILT nnnn <cr>
86
nnnn=averaging time (0 - 1024 seconds)
This command is used to set and inspect the averaging time during which the
individual measurement samples are integrated to get an averaged reading.
Page 89
HMP240
M210300enAppendix 1: Serial commands
The time can be set in seconds within the range of 0 - 1024 (0 = no averaging
time).
For example:
>FILT <cr>
Filter (S):0 ? 1024
>
>FILT 100 <cr>
Filter (S):100
>
PRESSetting the pressure for mixing ratio and wet bulb
temperature calculations
PRES pppp.pp <cr>
pppp.pp=pressure (hPa)
The atmospheric pressure has an effect on mixing ratio and wet bulb
temperature. Therefore, accurate calculations can be achieved only when the
ambient pressure is taken into consideration. The pressure is used for pressure
compensation of the HUMICAP sensor in order to ensure the best possible
measurement accuracy.
When the command is given, the transmitter first gives the currently used
pressure; after this a new value can be entered or the old one acknowledged.
>PRES <cr>
Pressure : 1013.25 ? 1000.00 <cr>
When the currently used pressure is known, a new pressure can also be entered
directly:
>PRES 1010 <cr>
Pressure : 1010
NOTE
If the pressure setting is frequently adjusted, e.g. by
using an external barometer as a pressure input source,
the command XPRES is recommended.
87
Page 90
HMP240
Appendix 1: Serial commandsM210300en
XPRESSetting the pressure for mixing ratio and wet bulb
temperature calculations temporarily
XPRES pppp.pp <cr>
pppp.pp=pressure (hPa)
The function and format of XPRES are the same as those of the PRES command except that by using XPRES the setting is valid only until a reset is
given or power is turned off or pressure is set to zero using XPRES. After this
the pressure stored using command PRES is valid again.
CDATEEntering calibration date
CDATE xxxxxx <cr>
xxxxxx =calibration date (000101...991231)
When the latest calibration date has to kept in memory, it is entered as follows:
>CDATE 940506 <cr>
If the command is given without the date, the transmitter outputs the latest
calibration already in memory.
>CDATE <cr>
940420
The date can be given in any format; however, the maximum number of digits
is six.
DATESetting the date
DATE <cr>
For example, to enter a new date:
88
>DATE <cr>
Current date is 1993-01-30
Enter new date (yyyy-mm-dd) : 1993-06-12 <cr>
When the current date is asked, the new date is passed with <cr>.
TIMESetting the time
Page 91
HMP240
M210300enAppendix 1: Serial commands
TIME <cr>
For example, to enter a new time:
>TIME <cr>
Current time is 01:35:54
Enter new time (hh:mm:ss) : 13:25:56 <cr>
When the current time is asked, the new time is passed with <cr>.
VERSName and version of the programme
VERS <cr>
For example:
>VERS <cr>
HMP 240 / x.yy
where x.yy is the programme version.
?Outputting the transmitter settings
? <cr>
For example:
>? <cr>
HMP 240 / 1.02
CPU serial nr : 0
Keyboard type : 0
Address: 7
Output units : metric
Baud P D S: 4800 E 7 1 FDX
Serial mode: STOP
Output intrv. : 0 min
Mtim: 32
Pressure : 1013.25
Analog outputs
Ch1 0.00 ... 10.00 V
Ch2 0.00 ... 10.00 V
Ch1 ( RH ) lo0.000 %RH
Ch1 ( RH ) hi100.000 %RH
Ch2 ( T ) lo-20.000 'C
Ch2 ( T ) hi180.000 'C
Transducer:
PRB serial nr : 0
Calibr. date : 0
??Outputting the transmitter settings also in POLL mode
89
Page 92
HMP240
Appendix 1: Serial commandsM210300en
?? <cr>
Command ?? outputs the same information as command ? but it works also
when the transmitter has been set to POLL mode. However, if there are more
than one addressed transmitters connected to the serial bus, they all will respond at the same time and the output on the screen will be chaotic.
90
Page 93
HMP240
M210300en Appendix 2: RS 485/422 serial port module
APPENDIX 2: INSTALLING AND USING THE RS 485/422 SERIAL
PORT MODULE
HMP240
Appendix 2: RS 485/422 serial port moduleM210300en
1. INSTALLATION
Switch the transmitter off.
Resistors R2, R3 and R4 between connectors X4 and X5 in the component
board in the cover of the transmitter are removed with side-cutting pliers. The
module is plugged in connectors X4 and X5 on the main board of the HMP240
transmitter; connector X1 on the module board to connector X4 and connector
X2 to connector X5.
RS485/422
serial bus
module
X5
X1X2
RX GND TX
H I
TX
LO
HI
RX
LO
R2
R3
R4
X4
+
Ch1
TX HI
TX LO
RX HI
RX LO
Ch2
24
+
X6
Jumper ( ) selections for the
RS 485/422 serial bus module
Single pair
X1
-- +
Dual pair
V
-
X1
X2
X2
Cut off the resistors R2, R3 and R4 on the main board.
New signal names for X6 screw terminal are on the module.
Follow the instructions on the module:
X1 to X4 and
X2 to X5 on the mother board
NOTE! If the transmitter is NOT at the end
of the bus OR the line has a dynamic line
termination, the resistors R3 and R6 have
to be removed!
R6
R3
92
Connect the data wires to screw terminal X6 on the main board. Switch the
power on.
Page 95
HMP240
HMP243 transmitters
HMP243 transmitters
M210300en Appendix 2: RS 485/422 serial port module
2. OPERATION
The HMP240 transmitter can either be given an address or operated without an
address. Both single and dual loop wiring with half duplex connection can be
used. No address is needed when only one HMP240 transmitter is used; when
several transmitters are connected to the same line, each transmitter must be
given an address in the initial configuration.
A single transmitter can get its operating voltage from the master or it can
have its own (floating) power supply.
The serial line structure is a parallel interfaced chain (daisy chain). At the end
of the serial line there must be an HMP240 transmitter, dynamic line adapter
(120 ohm resistor in series with a 33 nF capacitor) or line master. If a branch
line is made with a junction box, the branch should be shorter than 3 meters.
When connecting the device, follow the instructions given in the figure in
Chapter 1.
HOST COMPUTER
TX HI
TX LO
RX HI
RX LO
-
NOTE: DYNAMIC LINE TERMINATION
REQUIRED IF NO DEVICE
AT THE END OF THE BUS.
HOST COMPUTER
TX HI
TX LO
RX HI
RX LO
SINGLE LOOP WIRING
TWISTED PAIR WIRING
120R
DUAL LOOP WIRING
TWISTED PAIR WIRING
RX HI
RX LO
TX HI
TX LO
1 to n pieces
RX HI
RX LO
TX HI
TX LO
33 nF
RX HI
RX LO
TX HI
TX LO
1 to n pieces
NOTE: DYNAMIC LINE TERMINATION
REQUIRED IF NO DEVICE
AT THE END OF THE BUS.
120R
33 nF
120R
RX HI
RX LO
TX HI
TX LO
33 nF
93
Page 96
HMP240
Appendix 2: RS 485/422 serial port moduleM210300en
The RS 485/422 module has separate lines for transmitting and receiving, but
they can be connected together with jumpers. Dual loop connection is the
factory setting; when a single loop connection is used, the positions of jumpers in connector X4 on the module must be changed.
The HI of the receiving line is approx. 0.6 V and its LO is approx. 0 V in order to reduce noise on the lines when no data is transferred (idling). Both lines
are terminated with a 120 ohm resistor in series with a 33 nF capacitor. When
operating the transmitter through a single pair, only one line terminal
impedance is in use. The line must not be terminated with a resistor alone, as
the power consumption increases too much.
The data lines can withstand short circuit to ground and to each other. They do
not survive connection of supply voltage to the data lines.
The module must be mounted on the main board in the right direction. If
mounted in the wrong direction or to the wrong pins, the module will not
break but it does not work.
94
Page 97
HMP240
RX GND TX
M210300en Appendix 2: RS 485/422 serial port module
3. NETWORK CONFIGURATION
3.1 Single loop operation
Bi-directional data on one pair is one of the great advantages of the RS 485
line. Set jumpers in connector X4 on the module board as shown in the figure
below.
X1X2
X4
X4
This jumper setting connects RX HI to TX HI and RX LO to TX LO and selects only one common line termination. The HI and LO terminals of the RX
pair can now be used for operation.
Supplying power from the same end to the whole network prevents common
mode voltages from rising too high (over 7 V).
• Connect wires to the transmitter's serial connector.
• Check the wiring.
The following procedure must be repeated with all transmitters.
• Open the transmitter cover.
• Pull out the RS 485/422 serial port module, if it is already mounted.
• Set the serial port of the terminal to 4800 baud, even parity, seven
data bits and one stop bit, full duplex (4800 E 7 1 FDX).
X17
• The serial settings of the transmitter must also be 4800 E 7 1 FDX
and the transmitter must be in STOP mode. If these factory settings
have been changed, they must be returned. Connect the RS 232C port
of the terminal to connector X17 on the top of the main board and
switch the power on.
95
Page 98
HMP240
Appendix 2: RS 485/422 serial port moduleM210300en
• Set the address of the transmitter; you can choose any number
between 1 and 99. In this example the address is 22:
>addr 22
Address : 22
• Set the serial bus settings according to your network specifications.
This setting will become valid after next RESET or power off:
>seri 2400 e 7 1 h
2400 E 7 1 HDX
• Set the transmitter in POLL mode:
>smode poll
Serial mode : POLL
NOTE 1
The SMODE command must be given last.
NOTE 2
The transmitter outputs no prompt (>) after the SMODE
POLL command and it only reacts to commands which
include its address.
• Check that the transmitter responds to its address:
>send 22
RH= 24.4 %RH T= 29.1 'C
• Disconnect the terminal.
• Check that the jumpers in connector X4 are in the right places.
X1X2
X4
X4
96
• Remount the RS 485/422 serial module.
• Close the cover.
• When all transmitters on the network have been configured, switch
them off.
Page 99
HMP240
M210300en Appendix 2: RS 485/422 serial port module
3.2 Dual loop operation
The following procedure must be repeated with all transmitters.
• Open the transmitter cover.
• Pull out the RS 485/422 serial port module, if it is already mounted.
• Set the serial port of the terminal to 4800 baud, even parity, seven
data bits and one stop bit, full duplex (4800 E 7 1 FDX).
• The serial settings of the transmitter must also be 4800 E 7 1 FDX
and the transmitter must be in STOP mode. If these factory settings
have been changed, they must be returned. Connect the RS 232C port
of the terminal to connector X17 on the top of the main board and
switch the power on.
When dual loop is used, the jumpers in connector X4 on the module board
must be as shown below.
X1X2
X4
X4
• Set the address of the transmitter; you can choose any number
between 1 and 99. In this example the address is 22:
>addr 22
Address : 22
• Set the serial bus settings according to your system. This setting will
become valid after next RESET or power off:
>seri 2400 e 7 1 f
2400 E 7 1 FDX
• Switch echo on:
>echo on
ECHO : ON
>
• Change the serial output mode into POLL:
>smode poll
Serial mode : POLL
NOTE 1
The SMODE command must be given last.
97
Page 100
HMP240
Appendix 2: RS 485/422 serial port moduleM210300en
NOTE 2
The transmitter outputs no prompt (>) after the SMODE
POLL command and it only reacts to commands which
include its address.
• Check that the transmitter responds to its address:
>send 22
RH= 24.4 %RH T= 29.1 'C
• Disconnect the terminal.
• Check that the jumpers in connector X4 are in the right place
X1X2
X4
X4
• Remount the RS 485/422 serial module.
• Close the cover.
• Repeat this setting procedure with each transmitter.
• When all transmitters on the network have been configured, switch
them off.
98
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.