Vaisala HMP230 SERIES Operating Manual

HMP230 SERIES TRANSMITTERS
Operating Manual
U116en-2.2
April 1998
Vaisala 1998
electronic or mechanical (including photocopying), nor may its contents be communicated to a third party without prior written permission of the copyright holde.
The contents of instruction manuals are subject to change without prior notice.
Yleisjäljennös 4/98
HMP230 SERIES
U116en-2.2 Operating Manual
i
Contents
1. PRODUCT DESCRIPTION................................................................................................... 1
2. TO BE NOTED WHEN MEASURING HUMIDITY.................................................................. 3
3. INSTALLATION................................................................................................................... 4
3.1. Selecting the place of installation .......................................................................... 4
3.2. Mounting ................................................................................................................. 5
3.2.1. Mounting the HMP231................................................................................ 5
3.2.2. Installing the HMP233, HMP234 and HMP235 ........................................... 5
3.2.3. HMP233 transmitter................................................................................... 6
3.2.4. HMP234 transmitter................................................................................... 7
3.2.5. HMP235 transmitter................................................................................. 10
3.2.6. HMP238 transmitter................................................................................. 11
3.2.6.1. Mounting; overview.................................................................................. 12
3.2.6.2. Installing the probe through the ball valve assembly................................. 15
3.3. Grounding ............................................................................................................. 19
3.4. Electrical connections.......................................................................................... 21
3.4.1. Connection to a 24 VAC supply................................................................ 22
4. COMMISSIONING.............................................................................................................. 23
4.1. Changing the parameters..................................................................................... 23
4.2. Security lock jumper ............................................................................................. 23
4.3. Selecting the analogue outputs ........................................................................... 24
4.4. Connecting the RS 232C serial bus ..................................................................... 26
4.4.1. Reverting to factory settings of the serial port........................................... 28
5. COMMANDS ......................................................................................................................30
5.1. Commands and security lock jumper .................................................................. 30
5.2. LED commands..................................................................................................... 31
5.3. Display/keypad commands................................................................................... 32
5.3.1. Display mode........................................................................................... 32
5.3.2. Command mode....................................................................................... 32
5.3.3. Entering numbers..................................................................................... 32
5.3.4. Analogue output commands..................................................................... 33
5.3.4.1. Selecting the output (mA/V) ..................................................................... 33
5.3.4.2. Selecting and scaling the analogue output quantities ............................... 34
5.3.5. Output via the serial bus .......................................................................... 35
5.3.5.1. Turning the serial interface echo ON/OFF................................................ 35
5.3.5.2. Serial bus settings ................................................................................... 35
5.3.5.3. Setting the transmitter address ................................................................ 36
5.3.5.4. Selecting the output units ......................................................................... 37
5.3.5.5. Selecting the calculation mode ................................................................. 37
5.3.6. Output modes.......................................................................................... 37
5.3.6.1. Setting the serial interface operation mode .............................................. 38
5.3.7. Others...................................................................................................... 39
5.3.7.1. Setting the averaging time ....................................................................... 39
5.3.7.2. Setting the pressure for pressure compenstion of the HUMICAP
sensor
and for mixing ratio, wet bulb and enthalpy calculations ....................... 39
5.3.7.3. Setting the date........................................................................................ 40
HMP230 SERIES Operating Manual U116en-2.2
ii
5.3.7.4. Setting the time........................................................................................40
5.4. Serial commands ................................................................................................... 42
5.4.1. Analogue output commands .....................................................................42
5.4.1.1. Setting the analogue outputs....................................................................42
5.4.1.2. Selecting and scaling the analogue output quantities................................43
5.4.1.3. Scaling the analogue outputs....................................................................43
5.4.2. Output via the serial bus ...........................................................................44
5.4.2.1. Starting the measurement output.............................................................. 44
5.4.2.2. Stopping the measurement output............................................................44
5.4.2.3. Outputting the reading once .....................................................................44
5.4.2.4. Setting the output interval for the RUN mode ............................................45
5.4.2.5. Serial bus settings....................................................................................45
5.4.2.6. Selecting the output units .........................................................................46
5.4.2.7. Setting the averaging time........................................................................46
5.4.2.8. Setting the transmitter address.................................................................46
5.4.2.9. Setting the calculation mode.....................................................................47
5.4.2.10. Resetting the transmitter ..........................................................................47
5.4.3. Operating the transmitter via the serial bus...............................................47
5.4.3.1. Setting the serial interface........................................................................47
5.4.3.2. OPEN & CLOSE .......................................................................................48
6. CALIBRATION...................................................................................................................49
6.1. Humidity calibration..............................................................................................49
6.1.1. One point calibration procedure................................................................50
6.1.1.1. With serial commands ..............................................................................50
6.1.1.2. With display/keypad commands................................................................ 51
6.1.1.3. With LED commands................................................................................51
6.1.2. Two point calibration procedure ................................................................ 51
6.1.2.1. With serial commands ..............................................................................52
6.1.2.2. With display/keypad commands................................................................ 52
6.1.2.3. With LED commands................................................................................53
6.1.3. Calibration procedure after sensor change ...............................................53
6.1.3.1. With serial commands ..............................................................................53
6.1.3.2. With display/keypad commands................................................................ 54
6.1.3.3. With LED commands................................................................................54
6.1.4. Humidity calibration table .........................................................................56
6.2. Temperature calibration ........................................................................................ 56
6.2.1. One point offset correction .......................................................................56
6.2.1.1. With serial commands ..............................................................................56
6.2.1.2. With display/keypad commands................................................................ 57
6.2.1.3. With LED commands................................................................................57
6.2.2. Two point temperature calibration.............................................................58
6.2.2.1. With serial commands ..............................................................................58
6.2.2.2. With display/keypad commands................................................................ 58
6.2.2.3. With LED commands................................................................................59
6.3. Calibration of the analogue outputs.....................................................................59
6.3.1. With serial commands ..............................................................................59
6.3.2. With display/keypad commands................................................................ 60
6.3.3. With LED commands................................................................................60
7. MAINTENANCE .................................................................................................................62
7.1. Reference measurements .....................................................................................62
7.2. Self-diagnostics.....................................................................................................62
7.3. Changing the HUMICAP® sensor and the filter ...................................................63
HMP230 SERIES
U116en-2.2 Operating Manual
iii
7.4. Temperature channel adjustment with Pt 100 simulators................................... 63
7.4.1. Adjustment using serial commands.......................................................... 63
7.4.2. Adjustment using display commands........................................................ 64
7.4.3. Adjustment using LED commands............................................................ 64
7.5. Measurement of output currents using test points............................................. 65
7.6. Adjusting the contrast of the display .................................................................. 65
8. TECHNICAL DATA............................................................................................................ 66
8.1. Relative humidity.................................................................................................. 66
8.2. Temperature.......................................................................................................... 66
8.3. Calculated variables ............................................................................................. 66
8.4. Pressure ................................................................................................................ 68
8.5. Analogue outputs ................................................................................................. 68
8.6. Electronics ............................................................................................................ 69
8.7. Mechanics ............................................................................................................. 70
8.8. Electromagnetic compatibility ............................................................................. 72
8.8.1. Emissions ................................................................................................ 73
8.8.2. Immunity.................................................................................................. 73
9. SPARE PARTS AND OPTIONS......................................................................................... 74
Appendix 1 Serial commands......................................................................................... 75
Appendix 2 Installing the power supply module ........................................................... 99
Appendix 3 Installing and using the RS 485/422 serial port module...........................103
Appendix 4 Installing and using the digital current loop module ...............................115
Appendix 5 Error messages ..........................................................................................125
Appendix 6 Calculation formulas..................................................................................131
Appendix 7 Alarm output unit.......................................................................................133
Appendix 8 Connectors.................................................................................................139
Appendix 9 Re-gaining ..................................................................................................143
Appendix 10 Pressure conversion chart ........................................................................155
Appendix 11 Wiring diagramme MK4456 ........................................................................157
Warranty..................................................................................................................................159
HMP230 SERIES Operating Manual U116en-2.2
iv
This page intentionally left blank.
HMP230 SERIES
U116en-2.2 Operating Manual
1
1. PRODUCT DESCRIPTION
The HMP230 series transmitters are microprocessor based instruments for the measurement of relative humidity and temperature; from these variables they can calculate dewpoint temperature, absolute humidity, mixing ratio, wet bulb temperature and enthalpy. The transmitters have two analogue outputs and they can be connected to a serial bus via the RS 232C interface or through an RS 485/422 serial module or a current loop module. At dewpoint temperatures below 0 °C, the user can select whether the transmitter calculates dewpoint or frostpoint reading; as default, the transmitter calculates dewpoint.
The series consists of four types of transmitters:
HMP231, wall installation
HMP233, installation in tight places; temperatures up to +80 or
+120 °C
HMP234, installation in pressure or vacuum chambers
HMP235, installation in high temperatures of +180 °C
HMP238, installation in pressurized processes
The transmitters can be configured in many ways. They can have either a blank cover or a cover with a local display and keypad with which the user can operate the transmitter. The power supply voltage can be selected from three alternatives. Two analogue output signals are selected from the measured and calculated quantities; the signals can be scaled and the measurement ranges changed within certain limits. The HMP233, HMP234, HMP235 and HMP238 probes can be supplied with two, five or ten metre sensor head cable.
The humidity measuring range is 0...100 %RH. The temperature is measured with a Pt 100 sensor. Temperature measurement range depends on the model; the HMP234, HMP235 and HMP238 have the widest range, -40...+180 °C. The analogue temperature output can be scaled quite freely, for example
-20...+60 °C can be set to correspond to 0...10 V. The dewpoint temperature, absolute humidity, mixing ratio, wet bulb temperature and enthalpy ranges are also scalable.
In some applications the sensor gain may gradually decrease because of an interference caused by some chemical present in the ambient. These changes can be recovered with an optional re-gaining function. Transmitters including this function are equipped with a composite humidity and temperature sensor.
The HMP230 series units incorporate the HUMICAP® sensor, which uses an
operating principle based on changes in the capacitance of the sensor as its thin polymer film absorbs water molecules.
HMP230 SERIES Operating Manual U116en-2.2
2
Options Calculation vari-
ables
dewpoint temperature, mixing ratio, absolute hu­midity, wet bulb temperature, enthalpy
Power supply 24 VDC (standard), (for 24 VAC, see page 22)
115/230 VAC Serial interface RS 232C (standard), RS 485/422, current loop Display cover cover with or without local display & keypad Filters sintered filter, PPS grid with steel netting, PPS grid Cable length 2, 5 or 10 metres Alarm output unit Not with 115/230 VAC power supply and not with
HMP231 Cable connectors For 24 VDC supply, for analogue outputs, for RS
232C line and for RS 485 single loop line; see
Appendix 8 for details Installation aids HMP233: installation kits for +80 °C and +120 °C
cables
HMP234: NPT conical pipe threaded fitting body
(1/2 - 14 NPT)
HMP235: steel and aluminium flanges
HMP238: ball valve set
HMP230 SERIES
U116en-2.2 Operating Manual
3
2. TO BE NOTED WHEN MEASURING HUMIDITY
In the measurement of humidity and especially in calibration it is essential that temperature equilibrium is reached. Even a small difference in tempera­ture between the measured object and the sensor causes an error. If the tem-
perature is +20 °C (+68 °F) and the relative humidity 50 %RH, a difference of ±1 °C between the measured object and t he sensor causes an error of ±3 %RH. When the humidity is 90 %RH, the corresponding error is ±5.4 %RH.
The error is at its greatest when the sensor is colder or warm er than the sur­roundings and the humidity is high. A temperature difference of a few degrees can cause water to condense on the sensor surface. In an unventilated space evaporation may take hours; good ventilation accelerates evaporation. The HUMICAP sensor starts to function normally as soon as the water has evapo­rated. If the condensed water is contaminated, the life span of the sensor may shorten and calibration may change.
Temperature (°C)
dRH (%RH)
0
1
2
3
4
5
6
7
8
9
10
-40-200 20406080100
Figure 2.1 Measurement error at 100 %RH when the differen ce
between the ambient and sensor temperatu re is 1 °C
HMP230 SERIES Operating Manual U116en-2.2
4
3. INSTALLATION
3.1. Selecting the place of installation
Choose a place which gives a true picture of the environment or process, and is as clean as possible. Air should circulate freely around the sensor. A rapid air flow is recommended; it ensures that the sensor head and the ambient air are at the same temperature.
Install the transmitter in a place where no cold or hot spot can develop. When the sensor head is installed in a duct or process channel where the temperat ure is different from the ambient temperature, insulate the point of entry; this is particularly important if the transmitter is installed with the sensor head pointing downwards. Installing the sensor head of the HMP233, HMP234 and HMP235 vertically is not recommended. An uninsulated installation could lead to condensation in the sensor head and even when no condensation oc­curs, the resultant air flow may change the temperature near the sensor and distort the readings.
Install the sensor head in the process wherever possible; avoid sample flows where the gas temperature can drop below dewpoint temperature. Install the sensor head transversely against the direction of the process flow.
In duct or channel installations drill a hole ready for a reference meter. Plug the reference hole tightly.
Install the electronics housing away from possible steams escaping from the process.
NOTE
To ensure an IP 65 class protection:
1. Always mount the transmitter housing with the cable bushings pointing downwards.
2. Make sure that the connection cable has the right thickness ( 7...10 mm) and that the cable bushing is carefully tightened.
3. Pay always special attention to closing the transmitter cover carefully and remember to tighten all four screws.
HMP230 SERIES
U116en-2.2 Operating Manual
5
3.2. Mounting
3.2.1. Mounting the HMP231
The best position for mounting the HMP231 is with the probe head pointing downwards. Due to internal heat transfer, the transmitter should not be mounted with the probe pointing upwards.
133
104
Ø 6.5
Figure 3.1 Mounting holes in the HMP230 transmitter housing
3.2.2. Installing the HMP233, HMP234 and HMP235
It is recommended that the cable models HMP233, HMP234 and HMP235 be mounted with the sensor head horizontally; this way, any water condensing on the tube cannot flow onto the sensors. When there is no alternative but to install the sensor head in the process vertically, the point of entry must be carefully insulated. The cable must also be allowed to hang loosely as in Figure 3.6; this prevents any condensed water from running onto the sensor head along the cable.
If the process temperature is much higher than that of the environment, the whole sensor head and preferably a piece of the cable must be insi de the proc­ess.
When mounted on the side of a duct or channel, the sensor head must be in­serted from the side (see Figure 3.7). If this is not possible and the sensor head must be inserted from the top, the point of entry must be carefully insulated.
HMP230 SERIES Operating Manual U116en-2.2
6
NOTE
During installation the sensor head must not be unsol­dered from and again resoldered to the main printed board of the transmitter. This procedure may damage the humidity calibration of the transmitter.
3.2.3. HMP233 transmitter
The HMP233 can be installed in ducts and channels with the help of the in­stallation kit available; the kit consists of a flange, a supporting bar for the sensor head cable and screws for attaching the flange to the wall of a duct. With the help of the installation kit the distance between the sensor head and the channel wall can be easily adjusted. The range of adjustment is 100...320 mm; the distance is measured from the tip of the sensor head to the flange.
supporting bar
sealing
flange
a plugged hole for reference
measurements
duct wall
Figure 3.2 Installing the sensor head of the HMP233 in a channel
with the help of flange and supporting bar
HMP230 SERIES
U116en-2.2 Operating Manual
7
The sensor head can also be installed vertically.
to be sealed
(loop type) clamp
Figure 3.3Vertical installation of the HMP233 sensor head
3.2.4. HMP234 transmitter
The atmospheric pressure has an effect on mixing ratio, wet bulb temperature and enthalpy. Therefore, accurate calculations can be achieved only when the ambient pressure is taken into consideration. The pressure is used for pressure compensation of the HUMICAP sensor in order to ensure the best possible
measurement accuracy. If the process pressure differs from normal
ambient pressure, the value has to be en tered in the transmitter memory when using the transmitters HMP234 or HMP238. The pressure to be
entered is the absolute pressure in hPa or mbar (for pressure unit conversion, see Appendix 10).
The HMP234 is supplied with a nut, a fitting body and a sealing washer. Dur­ing handling the fitting body and the nut should remain in place on the body of the sensor head to prevent damage to the highly polished surface.
To achieve a leak-tight assembly:
1. Remove the fitting body from the nut and the sensor head.
2. Fasten the fitting body to the chamber wall. Tighten the fitting body
into the threaded sleeve with a torque spanner. The tightening torque
is 150 ±10 Nm.
3. Insert the body of the sensor head into the fitting body and screw the
nut manually to the fitting body until the connection feels tight.
When a bushing is used, its size is selected according to the diameter of the sensor head; the diameter of the cable is increased by using e.g. tape at the bushing
HMP230 SERIES Operating Manual U116en-2.2
8
tightening cone
sealing washer
fitting body
nut
4. Mark both the fitting body and the nut hex.
A pen
fitting body
nut
5. Tighten the nut a further 30° (1/12 turn) or if you have a torque span­ner tighten it with a torque of 80 ±10 Nm. NOTE: after detachment the nut must be tightened without increased effort.
30°
6. The tightening cone of the fitting body has to be cleaned and greased after every tenth detachment. Use high-vacuum grease (Down Corning, Europe) or a similar grease.
clean cotton stick
tightening cone
sealing
washer
fitting body
The sealing washer has to be changed every time the fitting body is detached.
HMP230 SERIES
U116en-2.2 Operating Manual
9
(boring)
threaded sleeve M22x1.5/Ø40 x15
sheet metal
nut DIN 80705-M22x1.5 (AISI 316)
Sealing by Metal sealing washer DIN 7603
Check the thickness of the sheet metal according to the pressure of the chamber
sealing by anaerobic pip e thread seal
(SWAK, Cajon Company) or PTFE
(teflon) tape
fitting body. VAISALA code 17225 (AISI 316Ti). Body hex = 27 mm
tightening torqu e=15 0 Nm ±10 Nm
ANSI/ASME B1.20.1-1983
Fasten by threaded sleeve
Fasten by Nut DIN 80705
tightening torque=150 Nm ±10 Nm
NPT Conical pipe threaded connection
sealing by Metal sealing washer
DIN 7603
Ø22+0.3
s=3...6mm
M22x1.5
1/2 -14 NPT
Figure 3.4 Some examples on the installation of the HMP234 sensor
head
HMP230 SERIES Operating Manual U116en-2.2
10
3.2.5. HMP235 transmitter
to be sealed
to be insulated
Figure 3.5 Installing a transmitter in a process with the sensor head
horizontally
to be sealed
to be insulated
Figure 3.6 Installing a transmitter in a process with the sensor head
downwards (not recommended)
HMP230 SERIES
U116en-2.2 Operating Manual
11
Figure 3.7 Mounting the sensor head on a duct or channel
When the sensor head is installed in a duct or channel, the temperature differ­ence between the air in the duct and outside it must be small as the sensor head conducts heat.
3.2.6. HMP238 transmitter
The atmospheric pressure has an effect on mixing ratio, wet bulb temperature and enthalpy. Therefore, accurate calculations can be achieved only when the ambient pressure is taken into consideration. The pressure is used for pressure compensation of the HUMICAP sensor in order to ensure the best possible
measurement accuracy. If the process pressure differs from normal
ambient pressure, the value has to be en tered in the transmitter memory when using the transmitters HMP234 or HMP238. The pressure to be
entered is the absolute pressure in hPa or mbar (for pressure unit conversion, see Appendix 10).
It is recommended that the sensor head is installed directly in the process through the ball valve assembly. When the ball valve set is used, the chamber or the duct does not have to be emptied or shut down for installation or removal of the probe. Install the sensor head transversely against the direction of the process flow.
However, if direct installation is not possible for some reason, the probe can be installed in a "leak-through" position provided that there is a slight overpressure in the process. In this installation, the probe is mounted behind the ball valve assembly. The flow passes through the sensor head and leaks out through a vent hole in the fitting body. However, make sure that the temperature at the measurement point is equal to t hat of the process.
HMP230 SERIES Operating Manual U116en-2.2
12
PROBE UP
PROBE PUSHED DOWN
cable length 2, 5 or 10 m
ø5.5
178
31
ø13.5
adjustment
range120 mm
29
R1/2 ISO 7/1
non leaking screw (A) (factory setting) or leak screw (B) (included in the package)
149
clasp nut fitting body
Figure 3.8 Probe dimensions (in mm)
NOTE
Take care not to damage the pipe of the probe. If the pipe is damaged, the probe head is less tight and it will not go through the clasp nut.
3.2.6.1. Mounting; overview
>10.5mm
fitting bo dy hex = 24mm tapered thread R1 /2 I S O 7/1
ø19mm drilling
sealing with:
1. LOCT ITE® N o 542 + activ. No 7649 (t=-55...+150 °C)
2. M E G A -P IP E E X T R A N o 71 88 (t= -55 ...+1 70 °C )
3. PTFE tape (t=-60...+210 °C) NOTE: the tape does not lock the parts together. Therefore, use two fork spanners (hex 24 and 27 mm ) for tightening and opening the clasp nut of the probe
parallel thread G1/2 ISO 228/1 (BS 2779, JIS B0202)
>40mm
Process or pipe wall
Figure 3.9 Sealing and thread cutting for the fitting body
The fitting body can be installed e.g. on standard pipe fittings (G 1/2 ISO 228/1) or on a thread in the process wall. If the wall thickness is more than
10.5 mm, it is recommended to use a welded sleeve (see Figure 3.11). Note
HMP230 SERIES
U116en-2.2 Operating Manual
13
that the minimum recommended distance of the fitting body and probe head is 40 mm (see Figure 3.9).
Adjust the probe to a suitable distance according to the type of installation and tighten the clasp nut first manually; mark the fitting body and the clasp nut and tighten the nut a further 50...60° with a fork spanner (see Figure 3.10).
60°
probe
fitting body
clasp nut
max.
a pen
Figure 3.10 Tightening the clasp nut
NOTE
Be careful not to tighten the clasp nut more than 60° as this may result in difficulties when trying to open it.
When the probe is installed directly on the process wall or pipe, note that a closing valve may be needed on both sides of the installed probe so that the sensor head can be removed from the process for calibration and maintenance.
If the sensor head is installed in a pressurized chamber, always make sure that the pressure of the chamber is equalized with the ambient pressure prior to removing the probe.
HMP230 SERIES Operating Manual U116en-2.2
14
process pipe
welded sleeve
(G1/2)
Non leaking screw (scre w A )
closing valve (ball valve)
sealing
capped nut DIN 917-M22x1.5
when the probe is pulled out for maintenance, cap the hole with a capped nut; this way, th e pro c e s s c a n b e open although the probe is not in place
Figure 3.11 Installing the sensor head directly on the process wall
HMP230 SERIES
U116en-2.2 Operating Manual
15
3.2.6.2. Installing the probe through the ball valve assembly
The best way to install the sensor head is through the ball valve assembly. Use
a 1/2” ball valve assembly with a ball hole of ∅14 mm or more. In this kind of installation, it is not necessary to empty or shut down the process for installing or removing the sensor head. If the sensor head is installed in a process pipe, please note that the nominal size of the pipe must be at least 1 inch. See pages 15 - 18 for detailed instructions.
handle
probe
screw A
process pipe / chamber
ball of the ball valve (hole diameter at least 14 mm)
>30 mm
>31 m m
Figure 3.12 Installing the sensor head through the DMP248BVS
ball valve assembly
NOTE
The probe can be installed in the process through the ball valve assembly provided that the process pressure is less than 10 bars. This way, the process does not have to be shut down when installing or removing the probe. However, if the process is shut down before removing the probe, the process pressure can be max. 40 bars.
HMP230 SERIES Operating Manual U116en-2.2
16
See Figure 3.13 through Figure 3.15 for detailed description of installation through the ball valve assembly. This installation is possible provided that the process pressure is less than 10 bars. Note also that if the sensor head is installed in a process pipe, the nominal size of the pipe must be at least 1 inch.
STEP I: mount the probe with the ball valve assembly closed; tighten the
clasp nut manually
fitting body R1/2 cone, seale
d
bushing R 1/2 c one sealed
bu s h ing R1/2 c o n e /G1/2 (4 0 b a r) e.g . Camoz z i 2 5 2 0-1 /2 - 1 /2
ball valve 1/2" (40 bar) e.g. Atlas Copco:BAL-1A 15 (G1/2)
nipple R 1 /2 co ne sealed
(the bushing serves for moving the probe (sinter) to such a distance from the ball valve that the valve can be closed)
>30 mm
clasp nut
Figure 3.13 Installing the probe through the ball valve
assembly; step 1
HMP230 SERIES
U116en-2.2 Operating Manual
17
STEP 2: open the ball valve assembly
R 1/2 ISO 7/1
120mm
ø5.5
61
(40)
148 mm
manual press tool
probe pipe
fitting ferrule
marking groove
leak scre w (B ) (hex. 1.5 mm)
ø14
29
15
ø13.5
DRYCAP® sensor
filter
fitting b o d y (hex. 24 mm )
handle
adjustment range
ball of the ball valve
clasp nut (hex 27 mm)
> 14
=
O-ring
Figure 3.14 Installing the probe through the ball valve
assembly; step 2 (measures in mm)
HMP230 SERIES Operating Manual U116en-2.2
18
STEP 3: push the probe head through the ball valve assembly into the
process. If the pressure is high, use a manual press tool. Note that the sensor head must be pushed so deep that the filter is completely inside the process flow.
VALVE OPEN
MANUAL PRESS TOOL
VALVE CLOSED
FILTER
Figure 3.15 Installing the probe through the ball valve
assembly: step 3
For by-pass measurements, the probe is mounted behind the ball valve assembly and non-leaking screw A on the fitting body is replaced with leaking screw B and O-ring is placed on the groove of the sintered filter. Screw B has a small (0.08 mm) laser-made hole in the middle; the gas or air to be measured passes through the sintered filter and by the sensor, and leaks out through the screw. The process pressure reduces in the hole of the screw B. This installation is recommended if the process flow rate is >20 m/s and there is an over-pressure in the process.
HMP230 SERIES
U116en-2.2 Operating Manual
19
NOTE keep the marking groove in sight when using leak screw (B)
gas escape channel: use this position for by-pass measurements
leak screw (B)
DRYCAP® sensor
sintered filter
process pipe or chamber
O-ring
Figure 3.16 Installing the sensor head for by-pass measurements
When pushing the probe head through the ball valve assembly, be careful not to break the sintered filter. Open and close the ball valve assembly with the marking groove always in sight. In by-pass measurements, the clasp nut is tightened manually prior to pressing the probe through the valve. When the probe has been pressed through and the valve is open, the nut is tightened
50...60° with a fork spanner (hexagon 27 mm).
3.3. Grounding
A single electrical cable with a screen and three to ten wires is recommended for power and analogue output/serial bus connections. The cable diameter should be 7...10 mm.
The screen of the electrical cable must be grounded properly to achieve best possible EMC performance. Recommended cable shield is done in the cable gland as shown below.
remove the brass disks, rubber ring and nut from the transmitter
housing
strip 165 mm of the cable insulation, but leave 25 mm of the braid
visible
slip the nut and rubber ring over the cable insulation
HMP230 SERIES Operating Manual U116en-2.2
20
slip the brass disk that has the bigger hole in it over the braid so that
it rests against the cable insulation
slip the other brass disk over the wires to the middle of the braid
(If the cable diameter is less than 7mm, use a shrinking
tube or an adhesive tape)
cable
braid
flexible wires 0.5 mm² (AWG 20), stranded wires
recommended
D = Ø 7...10 mm
3
brass disks
25
rubber ring
nut
brass disks
140
165
shielding tube
braid
push back the braid and press it between the two brass disks to
achieve a full 360° grounding; the fold between the disks should have the same diameter as the brass disks
secure the braid with a shielding tube
insert the wires into the transmitter housing through the gland.
tighten the nut
connect the wires into the screw terminals and fasten a cable tie
around the wires
transmitter housing
cable tie
gland
brass disks
rubber ring nut
HMP230 SERIES
U116en-2.2 Operating Manual
21
NOTE
When the cable is grounded as above, the metallic parts of the sensor head, the shield of its cable, the transmit­ter housing and the shield of the signal cable to external system are all connected to each other. After this the whole system can be grounded from one point only. If the grounding is made via several points (sensor head, transmitter housing, signal cable), make sure that the different groundings are made to the same grounding potential. Otherwise harmful grounding currents may be generated. If you do the grounding via the transmitter housing, use one serrated lock washer between a mount­ing screw and the housing; the lock washer breaks the paint on the housing.
When mains power supply is in use, the housing must be grounded by protec­tive ground wire using a grounding screw at the right-hand side of the power supply module (see Appendix 2).
3.4. Electrical connections
V
mA
+
-
CH1+
CH1-
CH2
+
CH2-
24 V +
X2
X1
V
mA
+
-
CH1- and CH2- are connected together internally
CURRENT/VOLTAGE OUTPUTS
POWER SUPPLY
(INTERNAL OR EXTERNAL)
Do not use power supply ground (-) as output signal ground
OPENED COVER OF THE HMP230
Figure 3.17 Electrical connections
HMP230 SERIES Operating Manual U116en-2.2
22
Power supply 24 VDC
24 VAC (see page 22)
with power supply module 115/230 VAC
Output signals 0...20 mA, 4...20 mA
0...1 V, 0...5 V, 0...10 V
Power supply ground (-) is connected to the housing with parallel connection of 15 nF capacitor and 300 kΩ resistor.
See Appendix 2 on how to connect the power supply module to the transmitter.
3.4.1. Connection to a 24 VAC supply
The HMP230 transmitters can also be connected to a 24 VAC supply without an external rectifier. However, when more than one transmitter is connected to one 24 VAC transformer, a common loop is formed and there is an increased risk of a short-circuit. To avoid this, always use separate floating supply for each transmitter (see Figure 3.18 A). However, if several transmitters have to share one transformer, the phase () must al ways be connected to + connector in each transmitter (see Figure 3.18 B).
24 VAC
HMP230 transmitter
HMP230 transmitter
Controller
shared common
24 VAC
24 VAC
HMP230 transmitter
HMP230 transmitter
Controller
B) COMMON LOOP FORMED -
NOT RECOMMENDED!
A) NO COMMON LOOP FORMED - RECOMMENDED
supply
voltage
signal
output
supply
voltage
signal
output
supply
voltage
supply
voltage
signal
output
signal
output
line
Figure 3.18 Connecting the transmitters to a 24 VAC supply
HMP230 SERIES
U116en-2.2 Operating Manual
23
4. COMMISSIONING
When HMP230 transmitters leave the factory, their measurement ranges and output signals have already been scaled according to the order form completed by the customer. Units are calibrated at the factory and ready to operate when the power is turned on. If you take into use active current, voltage or serial bus outputs, make these connections first; Appendix 9 describes them in detail.
NOTE
Make sure that the power is not turned on until cables have been connected to screw terminals!
In transmitters with display, the software version appears for a few seconds when the power is turned on. After this, measurement results appear automatically. Should an error message appear on the display, consult Appendix 5.
If your transmitter has a blank cover and the LED indicator inside the housing lights up, consult Appendix 5 for further information.
Appendix 7 contains information on how to determine the ranges for alarm outputs and alarm controls when an alarm output unit is used.
4.1. Changing the parameters
If necessary, the user can subsequently change the measurement units between metric and non-metric and select and scale the output signals with software functions. This is done through commands, either utilizing the menus on the local display or giving commands through the serial interface (see Appendices). Most often the commands are used to change the settings of the two analogue channels.
4.2. Security lock jumper
Before the settings can be changed, the user must first remove the security lock jumper in connector X15 (see Figure 4.1). The security lock jumper makes it impossible to change the transmitter settings by mistake.
HMP230 SERIES Operating Manual U116en-2.2
24
OPENED COVER OF THE HMP 230
X15
CHANGE OF SETTINGS DISABLED
Figure 4.1 Location of the security lo ck jumper
When the security lock jumper is connected, some commands cannot be used, see Chapter 5.
If you wish to take into use variables that are not included in the configuration of your transmitter, contact Vaisala or a Vaisala representative.
4.3. Selecting the analogue outputs
The HMP230 transmitters can be ordered with the required current or voltage outputs already selected. If the outputs need to be changed, move the jumpers in connector X15 into positions as shown in Figure 4.2.
Loading...
+ 134 hidden pages