HMP230 SERIES
Operating Manual U116en-2.2
ii
5.3.7.4. Setting the time........................................................................................40
5.4. Serial commands ................................................................................................... 42
5.4.1. Analogue output commands .....................................................................42
5.4.1.1. Setting the analogue outputs....................................................................42
5.4.1.2. Selecting and scaling the analogue output quantities................................43
5.4.1.3. Scaling the analogue outputs....................................................................43
5.4.2. Output via the serial bus ...........................................................................44
5.4.2.1. Starting the measurement output.............................................................. 44
5.4.2.2. Stopping the measurement output............................................................44
5.4.2.3. Outputting the reading once .....................................................................44
5.4.2.4. Setting the output interval for the RUN mode ............................................45
5.4.2.5. Serial bus settings....................................................................................45
5.4.2.6. Selecting the output units .........................................................................46
5.4.2.7. Setting the averaging time........................................................................46
5.4.2.8. Setting the transmitter address.................................................................46
5.4.2.9. Setting the calculation mode.....................................................................47
5.4.2.10. Resetting the transmitter ..........................................................................47
5.4.3. Operating the transmitter via the serial bus...............................................47
5.4.3.1. Setting the serial interface........................................................................47
5.4.3.2. OPEN & CLOSE .......................................................................................48
6. CALIBRATION...................................................................................................................49
6.1. Humidity calibration..............................................................................................49
6.1.1. One point calibration procedure................................................................50
6.1.1.1. With serial commands ..............................................................................50
6.1.1.2. With display/keypad commands................................................................ 51
6.1.1.3. With LED commands................................................................................51
6.1.2. Two point calibration procedure ................................................................ 51
6.1.2.1. With serial commands ..............................................................................52
6.1.2.2. With display/keypad commands................................................................ 52
6.1.2.3. With LED commands................................................................................53
6.1.3. Calibration procedure after sensor change ...............................................53
6.1.3.1. With serial commands ..............................................................................53
6.1.3.2. With display/keypad commands................................................................ 54
6.1.3.3. With LED commands................................................................................54
6.1.4. Humidity calibration table .........................................................................56
6.2. Temperature calibration ........................................................................................ 56
6.2.1. One point offset correction .......................................................................56
6.2.1.1. With serial commands ..............................................................................56
6.2.1.2. With display/keypad commands................................................................ 57
6.2.1.3. With LED commands................................................................................57
6.2.2. Two point temperature calibration.............................................................58
6.2.2.1. With serial commands ..............................................................................58
6.2.2.2. With display/keypad commands................................................................ 58
6.2.2.3. With LED commands................................................................................59
6.3. Calibration of the analogue outputs.....................................................................59
6.3.1. With serial commands ..............................................................................59
6.3.2. With display/keypad commands................................................................ 60
6.3.3. With LED commands................................................................................60
7. MAINTENANCE .................................................................................................................62
7.1. Reference measurements .....................................................................................62
7.2. Self-diagnostics.....................................................................................................62
7.3. Changing the HUMICAP® sensor and the filter ...................................................63