Vaisala HMP155 Technical data

www.vaisala.com
HMP155 Humidity and Temperature Probe
Features/Benefits
- superior long-term stability
Optional warmed humidity
probe
Plug-and-play
Chemical purge
USB connection for service use
Installation kits for DTR13 and
DTR502 radiation shields and also for a Stevenson screen
Weather-proof housing IP66
New, fast temperature probe
Dierent output possibilities:
voltage, RS-485, resistive Pt100
HMP155 with a new, stable HUMICAP®180R sensor and an additional temperature probe.
New probe for reliability
The new Vaisala HUMICAP® Humidity and Temperature Probe HMP155 provides reliable humidity and temperature measurement.
Long-term stability
The HMP155 has a new generation Vaisala HUMICAP has excellent stability and withstands well harsh environments. The probe structure is solid and the sensor is protected with a sintered teon lter, which gives maximum protection against liquid water, dust, and dirt.
®
180R sensor that
Warmed probe and high
This is an environment to which Vaisala has designed a patented, warmed probe for reliable measuring. As the sensor head is warmed continuously, the humidity level inside it stays below the ambient level. Thus, it also reduces the risk of condensation forming on the probe.
Fast temperature measurement
What’s more, with its fast response time, the additional temperature probe for the HMP155 is ideal for measurement in environments with changing temperatures.
humidity environment
Measuring humidity reliably is challenging in environments where humidity is near saturation. Measurements may be corrupted by fog, mist, rain, and heavy dew. A wet probe may not give an accurate measurement in the ambient air.
Applications: meteorological
applications, aviation and road weather, instrumentation
Long lifetime
Protecting the sensor from scattered and direct solar radiation, and precipitation will increase its lifetime. Thus, Vaisala recommends installing the HMP155 in one of the following radiation shields: DTR503, DTR13, or a Stevenson screen.
Easy maintenance
The probe can be calibrated using a pc with a USB cable, with the push buttons, or with the MI70 indicator.
Technical data
RS485 output voltage output
-80 0.4 -0.4 0.45 -0.45 20 0.12 -0.12 0.17 -0.17 60 0.22 -0.22 0.4 -0.4
-0.5
-0.4
-0.3
-0.2
-0.1
0
0.1
0.2
0.3
0.4
0.5
-80 -60 -40 -20 0 20 40 60
RS485 output voltage output
ºC
Δ ºC
Performance
RELATIVE HUMIDITY
Measurement range 0 ... 100 %RH Accuracy (incl. non-linearity, hysteresis and repeatability) at
+15 ... +25 °C (+59 ... +77 °F) ±1 %RH (0 ... 90 %RH)
±1.7 %RH (90 ... 100 %RH)
-20 ... +40 °C (-4 ... 104 °F) ±(1.0 + 0.008 x reading) %RH
-40 ... -20 °C (-40 ... -4 °F) ±(1.2 + 0.012 x reading) %RH +40 ... +60 °C (+104 ... +140 °F) ±(1.2 + 0.012 x reading) %RH
-60 ... -40 °C (-76 ... -40 °F) ±(1.4 + 0.032 x reading) %RH Factory calibration uncertainty (+20 °C /+68 °F) * Defined as ±2 standard deviation limits. Small variations possible, see also calibration certificate. Recommended humidity sensor HUMICAP Response time at +20 °C in still air with a sintered PTFE filter
63 % 20 s 90 % 60 s
TEMPERATURE
Measurement range -80 ... +60 °C (-112 ... +140 °F) Accuracy with voltage output at
-80 ... +20 °C ±(0.226 - 0.0028 x temperature) °C
+20 ... +60 °C ±(0.055 + 0.0057 x temperature) °C passive (resistive) output
according to IEC 751
±(0.1 + 0.00167 x |temperature|)°C
1/3 Class B
RS485 output
-80 ... +20 °C ±(0.176 - 0.0028 x temperature) °C
+20 ... +60 °C ±(0.07 + 0.0025 x temperature) °C Accuracy over temperature range (opposite) Temperature sensor Pt100 RTD 1/3 Class B IEC 751 Response time with additional temperature probe in 3 m/s air flow
63 % <20 s 90 % <35 s
OTHER VARIABLES
dewpoint/frost point temperature, wet bulb temperature, mixing ratio
±0.6 %RH (0 ... 40 %RH)*
±1.0 %RH (40 ... 97 %RH)*
®
180R(C)
General
Operating temperature range -80 ... +60 °C (-112 ... +140 °F) Storage temperature range -80 ... +60 °C (-112 ... +140 °F) Connection 8-pin male M12 connector Connection cables 3.5, 10, and 30 m Cable material PUR Wire size AWG26 Service cables USB connection cable
MI70 connection cable Additional T probe cable length 2 m Housing material PC Housing classification IP66 Sensor protection sintered PTFE Weigth (probe) 86 g Electromagnetic compatibility: Complies with the EMC standard EN61326-1, Electrical equipment for measurement control and laboratory use - EMC requirement for use in industrial locations
Inputs and outputs
Operating voltage 7 ... 28 VDC*
*Note: minimum operating voltage 12 V with 0 ... 5 V output and 16 V with 0 ... 10 V output, probe heating, chemical purge or XHEAT.
Outputs
voltage output 0 ... 1 V, 0 ... 5 V, 0 ... 10 V resistive Pt100 (4-wire connection)
RS485 Average current consumption (+15 VDC, load 100 kOhm)
0 ... 1 V output <3 mA
0 ... 10 V output +0.5 mA
RS485 <4 mA
during chemical purge max. 110 mA
with warmed probe max. 150 mA Settling time at power-up
voltage output 2 s
RS485 3 s
For more information, visit www.vaisala.com or contact us at sales@vaisala.com
Ref. B210752EN-B ©Vaisala 2009
This material is subject to copyright protection, with all copyrights retained by Vaisala and its individual partners. All rights reserved. Any logos and/or product names are trademarks of Vaisala or its individual partners. The reproduction, transfer, distribution or storage of information contained in this brochure in any form without the prior written consent of Vaisala is strictly prohibited. All specifications — technical included — are subject to change without notice.