mechanical (including photocopying), nor may its contents be communicated to a third
party without prior written permission of the copyright holder.
The contents are subject to change without prior notice.
Please observe that this manual does not create any legally binding obligations for
Vaisala towards the customer or end user. All legally binding commitments and
agreements are included exclusively in the applicable supply contract or Conditions of
Sale.
7.SPARE PARTS AND ACCESSORIES ................................................................................14
180 sensor and the filter ............................................11
APPENDIX 1: CURRENT OUTPUT MODULE
APPENDIX 2: SETTINGS OF THE LOCAL DISPLAY
APPENDIX 3: HMP143A PROBE CABLE EXTENSION
TECHNICAL DATA
1999-12-08i
Page 4
HMP140A SERIES
Operating ManualU175en-1.5
This page intentionally left blank.
ii1999-12-08
Page 5
HMP140A SERIES
U175en-1.5Operating Manual
1.PRODUCT DESCRIPTION
The HMP140A series transmitters are reliable and easy to use instruments for
the measurement of relative humidity and temperature. The series consists of
three different types of transmitters to be chosen according to the requirements
of the application:
• HMP141A for wall installations
• HMP142A for duct installations
• HMP143A for installation in tight places
The configuration can be chosen from various possibilities to meet the specific
requirements of the customer. The transmitters can be ordered with a blank
cover or a cover with a local display which outputs relative humidity and
temperature readings. The reading to be displayed is chosen with a pushbutton
on the cover and the desired temperature unit (Celsius or Fahrenheit) can be
chosen with a jumper (see Appendix 2).
All HMP140A transmitters measure relative humidity in the range of
0...100 %RH and temperature from -40 to +60 °C; the measurement is
temperature compensated. The HMP140A transmitters have four different
analogue outputs: 0...20 mA (4...20 mA), 0...1 V, 0...5 V and 0...10 V. The
outputs are also scalable if necessary, e.g. 4...20 mA from 0...20 mA.
The durable plastic cover provides IP65 protection from dust and sprayed
water. The HMP140A transmitters are therefore suitable for most indoor and
outdoor applications, including those with high humidities like indoor
swimming pools etc. These versatile transmitters are also easy to install and to
use. When necessary, they can be recalibrated on site with Vaisala’s HMI41
indicator equipped with an appropriate probe and optional calibration cable
(19116ZZ).
The HMP140A series transmitters incorporate the HUMICAP180 sensor, the
operation of which is based on the changes in its capacitance as the thin
polymer film absorbs water molecules. The sensor is immune to most
chemicals and has an excellent long-term stability. The temperature is
measured with a Pt 1000 sensor.
1999-12-081
Page 6
HMP140A SERIES
Operating ManualU175en-1.5
2.TO BE NOTED WHEN MEASURING HUMIDITY
In the measurement of humidity and especially in calibration, it is essential
that the temperature equilibrium is reached. Even a slight difference in the
temperature between the measured object and the sensor causes an error. For
example, at +20 °C (+ 68 °F) and 50 %RH, a temperature difference of +1 °C
between the measured object and the sensor causes an error of +3 %RH. If
relative humidity is 90 %RH, the error is about +5.4 %RH.
The error is at its greatest when the temperature of the sensor differs from that
of the surroundings and the humidity is high. A few degrees’ difference in
temperature may cause water to condense on sensor surface. Efficient
ventilation accelerates evaporation whereas in an unventilated space, it may
take hours. The HUMICAP180 sensor returns to its normal functioning as
soon as water has evaporated. Any contaminated water condensing on the
sensor may shorten its life span and change the calibration.
10
9
8
7
6
5
4
dRH (%RH)
3
2
1
0
-40-200 20406080100
Temperature (°C)
Figure 2.1Measurement error at 100 %RH when the temperature
difference between th e ambient and the sensor i s 1 °C
21999-12-08
Page 7
HMP140A SERIES
U175en-1.5Operating Manual
3.INSTALLATION
3.1Selecting the place of installation
Select a place that gives a true picture of the environment or process and is as
clean as possible. Air should circulate freely around the sensor. A rapid air
flow is recommended as it ensures the same temperature for the ambient air
and the sensor head.
Install the transmitter in a place where no cold or hot spot can develop. If the
sensor head is installed in a duct or channel where the temperature is different
from the ambient temperature, insulate the point of entry; this is particularly
important if the transmitter is installed with the sensor head pointing
downwards. Vertical installation of the HMP142A and HMP143A is not
recommended; they should be mounted with the sensor head horizontally
whenever possible. An uninsulated installation might lead to condensation on
the sensor head and even if no condensation occurs, the resultant air flow may
change the temperature near the sensor head and distort the readings.
3.2Mounting
The electronics of the HMP140A series can be disconnected for ease of
installation and service. The housing can be conveniently installed and the
electronics then consequently mounted. For service and maintenance purposes
the electronics can be disconnected without disconnecting the cabling, and
taken to an appropriate environment for necessary adjustments.
60
6.5
60
83.5
4.2
Figure 3Drilling dimensions (in mm)
1999-12-083
Page 8
HMP140A SERIES
Operating ManualU175en-1.5
3.2.1Mounting the HMP141A
The optimal installation of the HMP141A is with the sensor head pointing
downwards. Upward mounting should be avoided due to internal heat transfer.
1. Mounting and wiring:
cable feedthrough
on the surface or
in the connection box.
2. Mount the protection
lid to the conn ection
box.
3. Push the s e nsor h ea d
through the protection
lid a s in fig ure a n d
connect the electronics
housing to the connection
box. Make sure that the
pins at the bottom of
the electronics housing
are in co nt ac t with the
sc r ew termin a l .
4. Fasten the four
conn ec ting s cre w s of
the electronics housing.
5. Place the sensor head
into the p ro t e ct io n lid .
6. Close the cover and fasten
the two co n ne c tion s c rews.
1
3
4
4
6
4
4
5
2
Figure 3.2.1Mounting of the HMP141A
3.2.2Mounting the HMP142A
The HMP142A transmitter should be mounted with the sensor head
horizontally whenever possible. This way, any water that might condense on
the tube cannot flow on to the sensors. When there is no alternative but to
install the sensor head in the process vertically, the point of entry must be
carefully insulated.
41999-12-08
Page 9
HMP140A SERIES
U175en-1.5Operating Manual
225
Ø 15
Ø 12
Figure 3.2.2Dimensions of the HMP142A (in mm)
3.2.3Mounting the HMP143A
The cable model HMP143A should also be mounted with the sensor head
horizontally whenever possible. This model is provided with two plastic nuts
and a gasket. Mount the nuts on the sensor head: disconnect the cable from the
transmitter, insert the nuts and fasten them on the sensor head. Then reconnect
the cable. A mounting flange is also available (see Fig. 3.2.3). The cable can
also be extended up to 100 metres; see Appendix 3 for details.
M12x1
Ø 12
69
duct wall
sealing
supporting bar
a plugged hole for
reference measurements
flange
Figure 3.2.3Dimensions of the HMP143A probe (in mm) and
installation in a channel with the help of flange
(part no. HMP233FA) and supporting bar
1999-12-085
Page 10
HMP140A SERIES
Operating ManualU175en-1.5
3.3Electrical connections
The wiring can be done in either of the two ways shown in Figure 3.3.
mounting on a juction boxsurface mount
GND
Tout
SHIELD
RHout
Vs
(OHFWULFDOFRQQHFWLRQVP$RXWSXW
9V
*1'
$
$
7
RXW
6+,(/'
5+
RXW
9V
(OHFWULFDOFRQQHFWLRQV9RXWSXW
9V
9
9
Figure 3.3Electrical connections
Outputs
0...20 mA (4...20 mA)20...35 VDC17...24 VAC
min.
L
+ VDC
)(12ohmR
50
0...1 V 9...35 VDC 9...24 VAC
0...5 V15...35 VDC12...24 VAC
0...10 V20...35 VDC16...24 VAC
Current consumption6...10 mA + output currents
*1'
7
RXW
6+,(/'
5+
RXW
9V
Electrical connectionsscrew terminals for 0.5 - 1.5 mm2 wires (AWG
15...20), stranded wires recommended
61999-12-08
Page 11
HMP140A SERIES
U175en-1.5Operating Manual
3.4Connectors and potentiometers on the main PCB
DISPLAY CONNECTOR
PROBE CONNECTOR
VIO
X15
X16
SHIELD
GRN
YEL
BRN
Toffset
RHoffset
RHoutput
selection
RHgain
Tgain
Spare
jumpers
Toutput
selection
Test connector
for factory use;
leave the jumper
as indicated here
Test connector for
one point calibration
Connectors
for the
current module
Figure 3.4Connectors and potentiometers on the main PCB
3.5Selecting the outputs
The HMP140A series transmitters can be ordered with the desired output
signals already selected. If the outputs are changed, the jumpers have to be
moved to corresponding places (see Figure 3.5). For current outputs, see
Appendix 1. Jumper changes may change the output < ±0.15 %FS within the
chosen range; this usually does not give cause to recalibration.
DISPLAY CONNECTOR
Test connector for
one point calibration
PROBE CONNECTOR
VIO
SHIELD
GRN
YEL
BRN
RHoutput
selection
Toutput
selection
Voltage
Toutput selection
0-1V
0-5V
0-10V
Voltage
RHoutput selection
0-1V
0-10V
0-5V
Figure 3.5Jumper positions for voltage outputs
1999-12-087
Page 12
HMP140A SERIES
Operating ManualU175en-1.5
3.6Connection to an AC supply
The HMP140A transmitters can also be connected to an AC supply without an
external rectifier. However, when two or more transmitters are connected to
one 24 VAC transformer, a common loop is formed and there is an increased
risk of a short-circuit. To avoid this, always use separate floating supply for
each transmitter (see Figure 3.6 A). However, if several transmitters have to
share one transformer, always connect the phase (∼) to + connector in each
transmitter (see Figure 3.6 B).
A) NO COMMON LOOP FORMED - RECOMMENDED
HMP140 transmitter
HMP140 transmitter
B) COMMON LOOP FORMED -
HMP140 transmitter
supply voltage
Controller
supply volt a g e
supply voltage
signal output
signal output
NOT RECOMMENDED!
Controller
signal output
shared
common line
supply voltage
signal output
HMP140 transmitter
Figure 3.6Connecting the transmitters to a 24 VAC supply
81999-12-08
Page 13
HMP140A SERIES
U175en-1.5Operating Manual
4.CALIBRATION
4.1One-point humidity calibration
The HMP140A series transmitters should be recalibrated approximately once a
year. The interval depends on the operating conditions and the required
accuracy. One-point calibration can be done with Vaisala’s portable HMI41
indicator equipped with an appropriate probe and optional calibration cable
(19116ZZ). The cable is connected to a test connector either on the main
component board of the HMP140A or, if a current output module is installed,
on the module (see Figure 4.1). For more detailed instructions, see the
operating manual of the calibration cable (U218en).
DISPLAY CONNECTOR
PROBE CONNECTOR
VIO
SHIELD
GRN
YEL
BRN
Test connector for
one point calibration
on the main PCB
X15
X16
Test connector for
one point calibration
on the current
module
Current Module
Figure 4.1Location of the test connectors
4.2Two-point humidity calibration
The calibration can also be done with the HMK11 or the HMK13B Calibrator,
or the instrument can be sent to Vaisala or a Vaisala represent ative.
4.2.1Two-point humidity calibration procedure
• Leave the calibrator and the transmitter for at least 30 minutes in the
same space so that their temperatures have time to equalize.
• Place the probe into the calibration hole of the LiCl bottle in the
humidity calibrator.
• Wait for 10 minutes.
1999-12-089
Page 14
HMP140A SERIES
Operating ManualU175en-1.5
• Use the RH offset potentiometer (see Figure 4.2) to adjust the output
signal to the value given in the calibration table (Chapter 4.4).
• Place the probe into the calibration hole of the NaCl bottle in the
calibrator.
• Wait for 10 minutes.
• Check that the reading corresponds within the desired accuracy to the
reading given in the calibration table. If not, adjust the reading with
the RH gain potentiometer (see Figure 4.2).
• Check again the reading at the first point and adjust if necessary.
DISPLAY CONNECTOR
PROBE CONNECTOR
Figure 4.2Calibration potentiometers
4.3Temperature calibration
One point temperature calibration can be performed with the HMI41 equipped
with an appropriate probe and optional calibration cable if the accuracy of
+0.5 °C is sufficient. For further details, see the operating manual of the
HMI41 and calibration option.
Toffset
RHoffset
RHgain
Tgain
101999-12-08
Page 15
HMP140A SERIES
U175en-1.5Operating Manual
4.4Calibration table
Temperature°C1520253035
°F5968778695
LiCl%RH*)11.311.311.311.3
4...20 mAmA5.815.815.815.81
0...20 mAmA2.262.262.262.26
0...1 VV0.1130.1130.1130.113
0...5 VV0.5650.5650.5650.565
0...10 VV1.131.131.131.13
NaCl%RH75.675.575.375.174.9
4...20 mAmA16.1016.0816.0516.0215.98
0...20 mAmA15.1215.1015.0615.0214.98
0...1 VV0.7560.7550.7530.7510.749
0...5 VV3.7803.7753.7653.7553.745
0...10 VV7.567.557.537.517.49
*)LiCl solution must not be used or stored in temperatures below +18 °C (+64 °F) ;
otherwise the equilibrium humidity of the salt solution changes permanently
Table 1Greenspan’s calibration table with output values accordin g to
the chosen scale
5.MAINTENANCE
5.1Replacing the HUMICAP180 sensor and the filter
Remove the damaged sensor and insert a new one. Handle the sensor by the
plastic socket. DO NOT TOUCH THE SENSOR PLATE. Check the output
after sensor change. If it deviates too much from the reference value,
recalibrate the transmitter. Note that even after sensor change, the accuracy is
still +7 %RH.
Replace a dirty filter to ensure a maximum lifetime for the sensor. Do not try
to clean the filter. The sensor can be cleaned with distilled water; if this does
not help, replace the sensor. Do not use any mechanical methods.
1999-12-0811
Page 16
HMP140A SERIES
0
Operating ManualU175en-1.5
6.TECHNICAL DATA
6.1Relative humidity
Measurement range0...100 %RH
Accuracy (including non-linearity and repeatibility)
when calibrated against salt solutions (ASTM E104-85):
0...90 %RH+2 %RH
90...100 %RH+3 %RH
Temperature dependence:
dRH (%RH)
2
1.5
1
0.5
0
-0.5
-1
-1.5
-2
-20
-10
0102030
40
50
6
T (°C)
Response time (90%)15 s at +20 °C in still air (with sintered filter)
Maximum error between
the display and the outputs 0.2 %RH
Humidity sensorHUMICAP180
6.2Temperature
Measurement range-20...+60 °C
Typical accuracy at:
+20 °C+0.2 °C
-10...+40 °C+0.3 °C
-20...+60 °C+0.4 °C
Temperature sensorPt 1000 IEC 751 Class B
121999-12-08
Page 17
HMP140A SERIES
U175en-1.5Operating Manual
6.3General
Power supply
current outputs: 20...35 VDC 17...24 VAC
min.12 +
RL(ohm)
50
0...1 V 9...35 VDC 9...24 VAC
0...5 V15...35 VDC12...24 VAC
0...10 V20...35 VDC16...24 VAC
Current consumption6...10 mA + output currents
Electrical connectionsscrew terminals for 0.5 - 1.5 mm2 wires (AWG
15...20), stranded wires recommended
Alternative analogue
output signals (scalable)0...20 mA (4...20 mA)
0...1 V, 0...5 V, 0...10 V
NOTE! Output scales correspond to 0…100 %RH and -40…+60°C.
External loads for
current outputs< 500 Ω
0...1 V> 1 kΩ (to ground)
0...5 V> 5 kΩ (to ground)
0...10 V>10 kΩ (to ground)
Operating temperature range
for electronics-20...+60 °C
with display option 0...+50 °C
Storage temperature range -40...+80 °C
Operating humidity range0...100 %RH
Bushingfor ∅ 7...10 mm cables
Sensor protection
standardmembrane filter, part no. 17039HM
optionplastic grid, part no. 17038
Cable length of HMP143A
standard2.5 m
maximum100 m (user extendable, see Appendix 3)
Housing materialABS plastic
Housing classificationIP65 (NEMA 4)
1999-12-0813
Page 18
HMP140A SERIES
Operating ManualU175en-1.5
6.4Electromagnetic compatibility
6.4.1Emissions
Test:Setup according to:
Radiated interferenceEN55022(class B)
6.4.2Interference
Test:Setup according to: Performance:
Electrostatic dischargeIEC 801-2 criteria B
RF-radiated fieldsIEC 801-3criteria A
(allowed temporary effect
±1 °C in temperature)
Electrical fast transientsIEC 801-4criteria B
7.SPARE PARTS AND ACCESSORIES
Order codeDescription
HMP233FAmounting flange for the HMP143A
19326junction box for the HMP143A cable ex tension
(seeApp.3)
1535extension cable for the HMP143A (see Appendix 3)
17039HMmembrane filter
17038plastic grid
HUMICAP®180humidity sensor
GUARANTEE
Vaisala issues a guarantee for the material and workmanship of this product
under normal operating conditions for one (1) year from the date of delivery.
141999-12-08
Page 19
HMP140A SERIES
U175en-1.5Operating Manual
Exceptional operating conditions, damage due to careless handling and misapplication will void the guarantee.
1999-12-0815
Page 20
Page 21
HMP140A SERIES
U175en-1.5Appendix 1
APPENDIX 1: CURRENT OUTPUT MODULE
Spare jumpersCurrent module
X15
T output
selection
X16
RH output
selection
Figure 1
The current module is available only for HMP142A and HMP143A
transmitters. To install the current module to connectors X15 and X16, first
disconnect the jumpers (4 pcs). Insert them to the connector for spare jumpers
(see Figure 1) and install the current module. Note that the module must be
installed exactly as indicated in Figure 1. The jumpers for RH and T output
selections must be connected in two middle pins of the connectors (Figure 1).
Recalibrate the transmitter after having installed the current module. The
maximum error in output without recalibration is ± 0.5 %FS.
1999-12-081
Page 22
Page 23
HMP140A SERIES
U175en-1.5Appendix 2
APPENDIX 2: SETTINGS OF THE LOCAL DISPLAY
°F
RH/T
°C
Figure 1Location of the RH/T button and the °C/°F jumpers
The reading on the local display (relative humidity or temperature) can be
chosen by pressing the RH/T button beside the display. At power-up, the
display always shows the relative humidity reading. This can be changed to
temperature reading by pressing the RH/T button once. Further pressings will
change the reading between RH and T.
The temperature unit (°Celsius or °Fahrenheit) is chosen with the jumper on
back of the local display unit. For °C, the jumper is connected to the lower
position and for °F, it is connected to the upper position (see Figure 1).
NOTE!
The HMP140A series transmitters with the local
display and current output cannot be re-scaled.
1999-12-081
Page 24
HMI41
Appendix 2 HMI41-Uxxxen-1.1
21999-12-08
Page 25
HMP140A SERIES
S
U175en-1.5Appendix 3
APPENDIX 3: HMP143A PROBE CABLE EXTENSION
The cable of the HMP143A probe can be extended up to 100 metres with the
help of the extension set including a junction box (part no. 19326) and a
shielded six wire extension cable (part no. 1535).
The junction box has a 5-pole terminal block and two PG9 lead-throughs. The
extension cable is a shielded 6 x 0.22 mm2 copper wire cable. This cable
enables the extension without the need for transmitter adjustments. The user
can also use some other cable for this purpose provided that it is a shielded
copper wire cable. A shielded cable is recommended as it ensures the best
possible EMC protection. The diameter of the cable should not exceed 7 mm
as larger cables do not fit to the lead-through of the transmitter.
The HMP143A probe head uses a combined signal ground and supply voltage
ground. Ground wire resistance Rc causes an offset type error to the
transmitter output. The error is -0.11 x Rc both in %RH and in °C. This means
that if maximum error allowed is e.g. -0.3 %RH or -0.3 °C, the ground wire
resistance should not exceed 2.7Ω. The resistance of the cable 1535 is
minimized by connecting three wires as a ground conductor (see Figure 1).
Possible error can also be leveled out by readjusting the transmitter offset
(RH offset and T offset).
PROBE CONNECTOR ON THE
HMP143 PCB; THE COLOURS
SHOW THE DIRECT CONNECTION
OF THE HMP143 PROBE HEAD
PROBE CONNECTOR
VIO
SHIELD
GRN
YEL
BRN
Figure 1Schematic of the extension set connections
HMP143 PROBE
CONNECTOR
THREE WIRES CONNECTED
TOGETHER (WHT, GRN, RED)
EXTENSION
CABLE
CABLE SHIELD
BLU
GREY
YEL
BLK
JUNCTION
BOX
HMP143 CABLE
VIO
SHIELD
GRN (ground)
YEL
BRN
1999-12-081
Page 26
HMP140A
Appendix 3 U175en-1.5
TECHNICAL DATA
Junction box (part no. 19326)
Housing:
materialABS plastic
dimensions65 x 50 x 35 mm (l x w x h)
classificationIP 65
temperature range-40...+80 °C
Lead-throughs:
sizePG 9
cable diameter4...7 mm
housing classificationIP 68
Terminal block:
max. wire area1.5 mm
Extension cable (part no. 1535)
TypePFSK 6 x 0.22 mm
Wire resistance80 Ω / km
Coveringbraided, tinned copper wire
Cross-sectional area
of the wires0.22 mm
CoverPVC, light grey
Temperature range-30...+70 °C
Wire coloursred, blue, green, yellow, white, black
2
2
2
(DEF-STAN 61-12)
21999-12-08
Page 27
Page 28
www.vaisala.com
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.