Vaisala HMDW6070, HMP140 User Manual

Page 1
USER'S GUIDE
Calibration of Series HMD/W60/70 and
HMP140 Transmitters with
Vaisala HUMICAP® Indicator HMI41
U218EN11
Page 2
PUBLISHED BY Vaisala Oyj Phone (int.): +358 9 8949 1
P.O. Box 26 Fax: +358 9 8949 2227 FIN-00421 Helsinki Finland
Visit our Internet pages at http://www.vaisala.com/
© Vaisala 2006 No part of this manual may be reproduced in any form or by any means, electronic or
mechanical (including photocopying), nor may its contents be communicated to a third party without prior written permission of the copyright holder.
The contents are subject to change without prior notice.
Please observe that this manual does not create any legally binding obligations for Vaisala towards the customer or end user. All legally binding commitments and agreements are included exclusively in the applicable supply contract or Conditions of Sale.
Page 3
HMI41 Calibration Option
19116ZZ-U218en-1.1 Operating Manual

Contents

1. INTRODUCTION...................................................................................................................... 1
1.1 HMI41 with calibration option ......................................................................................1
1.2 Selecting the HMI41 calibrator function ......................................................................1
2. OUTPUT SCALE SELECTIONS...............................................................................................3
2.1 Standard transmitter settings............................................................................... 3
2.2 Getting started ..............................................................................................................3
2.3 Scale selection modes ..................................................................................................4
2.3.1 Voltage output selection mode (U)...................................................................4
2.3.2 Current output selection mode (I) .....................................................................5
2.3.3 Customized voltage output selection mode (cU) ..............................................6
2.3.3.1 Humidity channel...........................................................................................7
2.3.3.2 Temperature channel.....................................................................................9
2.3.4 Customized current output selection m ode (cI) .............................................. 10
2.3.4.1 Humidity channel.........................................................................................11
2.3.4.2 Temperature channel...................................................................................12
3. CALIBRATION.......................................................................................................................13
3.1 Getting started ............................................................................................................13
3.2 Humidity calibration....................................................................................................15
3.3 Temperature calibration ............................................................................................. 16
APPENDIX 1: QUICK REFERENCE TO SCALE SELECTIONS AND CALIBRATION
1996-08-29 i
Page 4
HMI41 Calibration Option Operating Manual 19116ZZ-U218en-1.1
This page intentionally left blank.
ii 1996-08-29
Page 5
HMI41 Calibration Option
19116ZZ-U218en-1.1 Operating Manual

1. INTRODUCTION

1.1 HMI41 with calibration option

The HMI41 can be used as a field calibrator for various Vaisala humidity transmitters. There are three different calibration cables available:
19116ZZ for the calibration of HMD/W60/70 and HMP140
series analogue transmitters
19164ZZ for the calibration of HMP230 series digital
transmitters
19165ZZ for the calibration of the HMD/W20/30 and
HMP130 series analogue transmitters
In this manual, only the calibration of the HMD/W60/70 and HMP140 series analogue transmitters is dealt with. For other calibration options, please refer to corresponding manuals.
Calibration is quick and easy to perform. Yet, the HMI41 includes four output scale selection modes which cover all scalable features of analogue transmitters and provide for a simple and reliable on-site calibration of even customized transmitters. In calibration, the HMI41 measures relative humidity and temperature with a separate probe, and receives the RH and T information from the transmitter. These values are then compared on the HMI41 display and the transmitter is adjusted with calibration potentiometers inside the transmitter.

1.2 Selecting the HMI41 calibrator function

The desired calibrator function is selected in the HMI41 setup menu. To enter the setup mode, first press the ON/OFF button and the following appears:
RH
Td T
P setminmax hysthold LobatHi
% °C °F
°C
°F
Pahg/m
Then release the ON/OFF button and within 1...2 seconds press both ENTER and MODE buttons until the following text appears on the display:
After a few seconds, the text changes automatically to show the following:
1996-08-29 1
Page 6
HMI41 Calibration Option Operating Manual 19116ZZ-U218en-1.1
°C
set
If the basic settings of the HMI41 (display units, automatic power off function, display quantities and pressure) have to be changed, please refer to the HMI41 Operating Manual. Otherwise, press ENTER repeatedly until the following text appears:
set
The number on the first line of the display corresponds to the following HMI41 functions:
1 = HMI41 as a humidity and temperature indicator
2 = HMI41 as a calibrator for HMD/W60/70 and HMP140 series analogue transmitters
3 = HMI41 as a calibrator for HMP230 series digital transmitters 4 = HMI41 as a calibrator for the HMD/W20/30 and HMP130 series analogue transmitters
To calibrate the HMD/W60/70 or the HMP140 series transmitters, select number 2 with buttons (number up) and (number down) and then turn the HMI41 off. When the HMI41 is turned on again, it will automatically wake up as a calibrator for these analogue transmitters. If you wish to calibrate transmitters of some other type or use the HMI41 as an indicator, this selection has to be done accordingly; refer to corresponding manuals for further details.
Note that if the default settings in the HMI41 memory have not been changed (e.g. when it is used as a calibrator for the first time), the HMP140 series transmitters with standard settings and voltage outputs can be directly calibrated (see Chapter 3). In all other cases scale selections are necessary; please, consult Chapter 2. The default settings in the HMI41 memory (Table
1.2) are returned with U, I, cU or cI selections according to which changes have been made.
Humidity range Temperature range Voltage range
0...1V
0...100 %RH
-40...+60 °C
0...5V
0...10V
Table 1.2 Defa ult settin gs in the H MI41 memory
2 1996-08-29
Page 7
HMI41 Calibration Option
19116ZZ-U218en-1.1 Operating Manual

2. OUTPUT SCALE SELECTIONS

2.1 Standard transmitter settings

In Table 2.1 you will find the standard settings of the HMD/W60, HMD/W70 and the HMP140 series transmitters.
Transmitter Relative humidity
scaling range
HMP140
HMD60
HMD70
HMW60
HMW70
0...100 %RH
0...100 %RH
0...100 %RH
0...100 %RH
0...100 %RH
Table 2.1 Standard settings of analogue transmitters
Temperature scaling range
-40...+60 °C
-20...+80 °C
-20...+80 °C
-5...+55 °C
-5...+55 °C
NOTE
Voltage
Current output
output
0...1 V
0...5 V 0...20 mA
0...10 V
4...20 mA
0...1 V
0...5 V 0...20 mA
0...10 V
4...20 mA
0...1 V
0...5 V
0...10 V
When calibrating transmitters with current outputs, make sure that the output loop is terminated.

2.2 Getting started

After having selected the calibrator function 2 as explained in Chapter 1.2, turn the HMI41 on with the ON/OFF button and wait until the following text appears on the display:
RH
T
This message remains on the display for about 5 seconds; if no button is pressed, the HMI41 goes on to calibration. To enter the scale selections, press any button (except ON/OFF) within 5 seconds and the following appears:
1996-08-29 3
Page 8
HMI41 Calibration Option Operating Manual 19116ZZ-U218en-1.1
Scale selection mode is entered when this text is on the display.

2.3 Scale selection modes

The HMI41 has four scale selection modes: voltage output (U) and current output (I) selection modes are used for transmitters with standard settings. Customized voltage output (cU) and customized current output (cI) selection modes are meant for analogue transmitters with customized RH, T and/or output range settings. All these selections are stored in the HMI41 memory; therefore, if the transmitter settings or the selections in the HMI41 memory have not been changed by the next calibration, you can calibrate the transmitter directly. In the following, each selection mode is explained in its own section.

2.3.1 Voltage output selection mode (U)

If the transmitter has voltage outputs, its temperature range is given to the HMI41 with the voltage output selection mode (U). The humidity range is always 0...100 %RH. Select the U selection mode with buttons or and acknowledge the selection with ENTER. A text similar to the following is displayed:
C T
set
°C
Lo
Numbers on the second line are blinking. They indicate the lower limit of the transmitter’s temperature range currently stored in the HMI41 memory. Text in the lower righthand corner is Lo. You can now set the lower limit with buttons (number up) or (number down) to correspond to that of the transmitter. Acknowledge the setting with ENTER:
4 1996-08-29
Page 9
HMI41 Calibration Option
19116ZZ-U218en-1.1 Operating Manual
C T
°C
set
Hi
Numbers on the first line start to blink. They indicate the upper limit of the transmitter’s temperature range currently stored in the HMI41 memory. Text in the lower righthand corner is Hi. You can now set the upper limit with buttons (number up) or (number down) to correspond to that of the transmitter. Acknowledge the setting with ENTER and turn the HMI41 off with the ON/OFF button.You can now calibrate the transmitter (see Chapter
3).

2.3.2 Current output selection mode (I)

NOTE
When calibrating the transmitters with current outputs, make sure that the output loop is terminated.
If the transmitter has current outputs, its temperature range is given to the HMI41 with the current output selection mode (I). The RH range is always
0...100 %RH. Select the I selection mode with buttons or and acknowledge the selection with ENTER. A text similar to the following is displayed:
set
Numbers on the display are blinking, and you can now select the output (0...20 or 4...20 mA) with buttons or according to the transmitter to be calibrated. Acknowledge the selection with ENTER and a text similar to the following is displayed:
1996-08-29 5
Page 10
HMI41 Calibration Option Operating Manual 19116ZZ-U218en-1.1
C T
°C
set
Lo
Numbers on the second line are blinking. They indicate the lower limit of the transmitter’s temperature range currently stored in the HMI41 memory. Text in the lower righthand corner is Lo. You can now set the lower limit with buttons (number up) or (number down) to correspond to that of the transmitter. Acknowledge the setting with ENTER:
C T
set
°C
Hi
Numbers on the first line start to blink. They indicate the upper limit of the transmitter’s temperature range currently stored in the HMI41 memory. Text in the lower righthand corner is Hi. You can now set the upper limit with buttons (number up) or (number down) to correspond to that of the transmitter. Acknowledge the setting with ENTER and turn the HMI41 off with the ON/OFF button. You can now calibrate the transmitter (see Chapter
3).

2.3.3 Customized voltage output selection mode (cU)

The customized voltage output selection mode (cU) is selected when the scaling of the RH, T and/or voltage range is not a standard one.
Select the cU selection mode with buttons or and acknowledge the selection with ENTER. A text similar to the following is displayed:
RH
set
6 1996-08-29
Page 11
HMI41 Calibration Option
19116ZZ-U218en-1.1 Operating Manual
2.3.3.1 Humidity channel
RH
set
Numbers on the display are blinking. They indicate the voltage range of the transmitter’s RH channel (jumper selectable) currently stored in the HMI41 memory. Please, consult Figures 2.3.1 - 2.3.2 when checking the transmitter’s jumper positions.
RH output selections
0...1V
0...5 V
0...10 V
Figure 2.3.1 Jumper selections of the HMD/W70
RH output selections
0...1V
0...5 V
0...10 V
Figure 2.3.2 Jumper selections of the HMP140
Numbers on the second line are blinking, and you can now select the range with buttons or according to the transmitter’s jumper position. Acknowledge the selection with ENTER. A text similar to the following is displayed:
RH
set
Lo
Numbers on the second line are blinking. They indicate the lower limit of the transmitter’s voltage range currently stored in the HMI41 memory. Text in the lower righthand corner is Lo. You can now set the lower limit of the voltage range to correspond to that of the transmitter. Set the limit with buttons or and acknowledge the setting with ENTER:
1996-08-29 7
Page 12
HMI41 Calibration Option Operating Manual 19116ZZ-U218en-1.1
RH
set
Hi
Numbers on the first line start to blink. They indicate the upper limit of the transmitter’s voltage range currently stored in the HMI41 memory and the text in the lower righthand corner is Hi. You can now set the upper limit of the voltage range to correspond to that of the transmitter. Set the limit with buttons or and acknowledge the setting with ENTER. A tex t similar to the following is displayed:
RH
set
%
Lo
Numbers on the second line are blinking. They indicate the lower limit of the transmitter’s RH range currently stored in the HMI41 memory. Text in the lower righthand corner is Lo. You can now set the lower limit of the RH range to correspond to that of the transmitter. Set the limit with buttons or and acknowledge the setting with ENTER:
RH
set
%
Hi
Numbers on the first line start to blink. They indicate the upper limit of the transmitter’s RH range currently stored in the HMI41 memory. Text in the lower righthand corner is Hi. You can now set the upper limit of the RH range to correspond to that of the transmitter. Set the limit with buttons or and acknowledge the setting with ENTER. The HMI41 goes on to T channel settings, and a text similar to the following is displayed:
T
set
8 1996-08-29
Page 13
HMI41 Calibration Option
19116ZZ-U218en-1.1 Operating Manual
2.3.3.2 Temperature channel
T
set
Numbers on the second line are blinking. They indicate the voltage range of the transmitter’s T channel (jumper selectable) currently stored in the HMI41 memory. Please, consult Figures 2.3.3 - 2.3.4 when checking the transmitter’s jumper positions.
T output selections
0...1V
0...5 V
0...10 V
Figure 2.3.3 Jumper selections for the HMD/W70
T output selections
0...1V
0...5 V
0...10 V
Figure 2.3.4 Jumper selections for the HMP140
You can now set the range with buttons or to correspond to the jumper position. Acknowledge the setting with ENTER. A text similar to the following is displayed:
T
set
Lo
Numbers on the second line are blinking. They indicate the lower limit of the transmitter’s voltage range currently stored in the HMI41 memory. Text in the lower righthand corner is Lo. You can now set the lower limit with buttons or to correspond to that of the transmitter. Acknowledge the selection with ENTER. A text similar to the following is displayed:
1996-08-29 9
Page 14
HMI41 Calibration Option Operating Manual 19116ZZ-U218en-1.1
T
set
Hi
Numbers on the first line start to blink. They indicate the upper limit of the transmitter’s voltage range currently stored in the HMI41 memory. Text in the lower righthand corner is Hi. You can now set the upper limit with buttons or to correspond to that of the transmitter. Acknowledge the setting with ENTER. A text similar to the following is displayed:
C T
set
°C
Lo
Numbers on the second line are blinking. They indicate the lower limit of the transmitter’s temperature range currently stored in the HMI41 memory. Text in the lower righthand corner is Lo. You can now set the lower limit to correspond to that of the transmitter. Set the limit with buttons or and acknowledge the setting with ENTER:
C T
set
°C
Hi
Numbers on the first line start to blink. They indicate the upper limit of the transmitter’s temperature range currently stored in the HMI41 memory. Text in the lower righthand corner is Hi. You can now set the upper limit to correspond to that of the transmitter. Set the limit with buttons or and acknowledge the setting with ENTER. Turn the HMI41 off with the ON/OFF button. You can now calibrate the transmitter (see Chapter 3).

2.3.4 Customized current output selection mode (cI)

The customized current output selection mode (cI) is selected when the scaling of the RH, T and/or current range is not a standard one.
NOTE
When calibrating transmitters with current outputs, make sure that the output loop is terminated.
10 1996-08-29
Page 15
HMI41 Calibration Option
19116ZZ-U218en-1.1 Operating Manual
Select the cI selection mode with buttons or and acknowledge the selection with ENTER. A text similar to the following is displayed:
RH
set
2.3.4.1 Humidity channel
RH
set
Numbers on the second line are blinking. They indicate the current range of the the transmitter’s RH channel currently stored in the HMI41 memory. Select the range with buttons ▲ or ▼ to correspond to that of the transmitter. Acknowledge the selection with ENTER. A text similar to the following is displayed:
RH
set
%
Lo
Numbers on the second line are blinking. They indicate the lower limit of the transmitter’s RH range currently stored in the HMI41 memory. Text in the lower righthand corner is Lo. You can now set the lower limit to correspond to that of the transmitter. Set the limit with buttons or and acknowledge the setting with ENTER:
RH
set
%
Hi
Numbers on the first line start to blink. They indicate the upper limit of the transmitter’s RH range currently stored in the HMI41 memory. Text in the
1996-08-29 11
Page 16
HMI41 Calibration Option Operating Manual 19116ZZ-U218en-1.1
lower righthand corner is Hi. You can now set the upper limit to correspond to that of the transmitter. Set the limit with buttons or and acknowledge the setting with ENTER. The HMI41 goes on to T channel settings and a text similar to the following is displayed:
T
set
2.3.4.2 Temperature channel
T
set
Numbers on the second line are blinking. They indicate the current range of the transmitter’s T channel currently stored in the HMI41 memory. Select the range with buttons or to correspond to that of the transmitter. Acknowledge the selection with ENTER. A text similar to the following is displayed:
C T
set
Lo
°C
Numbers on the second line are blinking. They indicate the lower limit of the transmitter’s temperature range currently stored in the HMI41 memory. Text in the lower righthand corner is Lo. You can now set the lower limit to correspond to that of the transmitter. Set the limit with buttons or and acknowledge the setting with ENTER:
C T
°C
set
Hi
Numbers on the first line start to blink. They indicate the upper limit of the transmitter’s temperature range currently stored in the HMI41 memory. Text in the lower righthand corner is Hi. You can now set the upper limit to correspond to that of the transmitter. Set the limit with buttons or and acknowledge the setting with ENTER. Turn the HMI41 off with the ON/OFF button. You can now calibrate the transmitter (Chapter 3).
12 1996-08-29
Page 17
HMI41 Calibration Option
ENTE
R
MOD
E HOL
D ON/
OFF
EXT
Connector for the calibration cable
19116ZZ-U218en-1.1 Operating Manual

3. CALIBRATION

For a successful calibration, it is essential that the probe of the HMI41 and that of the transmitter are at the same temperature, and that the reference probe has been previously calibrated. Always allow enough time for the readings to stabilize. Note that the stabilization time depends on the ambient conditions and may vary from 10 minutes to a couple of hours.

3.1 Getting started

After having selected the desired function (Chapter 1.2) and necessary scale selections (Chapter 2), turn the HMI41 off and connect the calibration cable to the EXT connector at the bottom of the HMI41 (see Figure 3.1) and to the appropriate connector in the transmitter (see Figures 3.2 - 3.4).
Figure 3.1 Location of the HMI41 calibration connector
TEST CONNECTOR
HMD60U/Y
RH TEST
T
RH OFFSET
RH GAIN
T GAIN
T OFFSET
TEST
Figure 3.2 Calibration connectors and p otentiometers in
HMD70U/Y
CONNECTOR
TEST FOR VOLTAGE OUTPUTS
RH GAIN
T GAIN
FOR CURRENT OUTPUTS
RH OFFSET
T OFFSET
1
U
UU
I-module 0-20mA
1
U
the HMD60 and HMD70 transmitters
1996-08-29 13
Page 18
HMI41 Calibration Option Operating Manual 19116ZZ-U218en-1.1
Figure 3.3 Calibration connectors and potentiometers in
RH TEST
RH
RH
OFFSET
GAIN
HMW60U/Y
the HMW60 and HMW70 transmitters
T TEST
X3 X4
T GAIN
T OFFSET
RH OFFSET
RH GAIN
HMD70U/Y
TEST CONNECTOR
T GAIN
T OFFSET
Tgain
Toffset
Test connector for voltage outputs
Test connector for current outputs
Current Module
RHoffset
RHgain
Figure 3.4 Calibration connectors and p otentiometers in the
HMP140 series transmitters
14 1996-08-29
Page 19
HMI41 Calibration Option
19116ZZ-U218en-1.1 Operating Manual

3.2 Humidity calibration

When performing the humidity calibration of the HMD/W60 transmitters, make sure that the calibration cable is connected to the RH test connector (Figures 3.2 and 3.3).
After having connected the cable, turn the HMI41 on with the ON/OFF button and wait until the following text appears on the display:
RH
T
This message remains on the display for about 5 seconds. Press no button when this text is displayed and the HMI41 will automatically go on to calibration; a text similar to the following appears:
RH %
Numbers on the first line indicate the relative humidity measured by the transmitter, and numbers on the second line indicate the RH measured by the HMI41 reference probe. If you wish, you can now adjust the transmitter:
Wait until the readings have stabilized and adjust the calibration
potentiometers (Figures 3.2 - 3.4) until the readings on the first and the second line are the same. It is recommended to use the offset potentiometer if the RH is <65 %RH and the gain potentiometer if the RH is >65 %RH.
After having completed the calibration, press MODE if you wish to
perform a temperature calibration (see next page), or ON/OFF (disconnect the cable).
If you press the HOLD button in the first RH calibration mode, you will go on to the second RH calibration mode which is an alternative for the first one. A text similar to the following is displayed:
RH %
1996-08-29 15
Page 20
HMI41 Calibration Option Operating Manual 19116ZZ-U218en-1.1
Numbers on the first line indicate how much the RH measured by the transmitter differs from that measured by the HMI41 reference probe. Numbers on the second line indicate the relative humidity measured by the reference probe. If you wish, you can now adjust the transmitter (by pressing HOLD again you will return to the previous RH calibration mode):
Wait until the readings have stabilized and adjust the calibration
potentiometers (Figures 3.2 - 3.4) until the numbers on the first line are at zero. It is recommended to use the offset potentiometer if the RH is <65 %RH and the gain potentiometer if the RH is >65 %RH.
After having completed the calibration, press MODE if you wish to
perform a temperature calibration (see below), or ON/OFF (disconnect the cable).

3.3 Temperature calibration

As the temperature sensor is extremely stable, temperature calibration should be used for check-ups only. Before starting the temperature calibration, leave the transmitter and the HMI41 to the same room at least for an hour so that their temperatures have enough time to equalize. If the difference between the readings is less than 0.5 °C, make no adjustments.
When performing the temperature calibration of the HMD/W60 transmitters, make sure that the calibration cable is connected to the T test connector (Figures 3.2 and 3.3).
Temperature calibration is entered through the RH calibration modes (see Chapter 3.2). By pressing MODE in the first RH calibration mode, a text similar to the following is displayed:
C T
°C
Numbers on the first line indicate the temperature measured by the transmitter, and numbers on the second line indicate the temperature measured by the HMI41. If you wish, you can now adjust the transmitter (by pressing the MODE button you will return to the RH calibration mode):
Wait until the readings have stabilized and then adjust the calibration
potentiometer (Figures 3.2 - 3.4) until the readings on the first and the second line are the same. It is recommended to use the offset potentiometer.
After having completed the calibration, press ON/OFF and disconnect
the cable.
16 1996-08-29
Page 21
HMI41 Calibration Option
19116ZZ-U218en-1.1 Operating Manual
If you press the HOLD button in the first T calibration mode, you will go on to the second T calibration mode which is an alternative for the first one. A text similar to the following is displayed:
C T
°C
Numbers on the first line indicate how much the t emperature measured by the transmitter differs from that measured by the HMI41 reference probe. Numbers on the second line indicate the temperature measured by the reference probe. If you wish, you can now adjust the transmitter (by pressing the HOLD button again you will return to the previous T calibration mode and by pressing the MODE button you will return to the second RH calibration mode):
Wait until the readings have stabilized and then adjust the calibration
potentiometer (Figures 3.2 - 3.4) until the numbers on the first line are at zero. It is recommended to use the offset potentiometer.
After having completed the calibration, press ON/OFF and disconnect
the cable.
GUARANTEE
Vaisala issues a guarantee for the material and workmanship of this product under normal operating conditions for one (1) year from the date of delivery. Exceptional operating conditions, damage due to careless handling and mis­application will void the guarantee.
1996-08-29 17
Page 22
HMI41 Calibration Option
19116ZZ-U218-1.1 Appendix 1
APPENDIX 1: QUICK REFERENCE TO SCALE SELECTIONS AND
CALIBRATION
It is recommended that this quick reference guide is used only as a checking list for those who already know how to operate the HMI41 as a calibrator. For those who take it into use for the first time, the Operating Manual gives useful information that is not included in this quick reference guide.

1. Selecting the calibrator function

For selecting the calibrator function, press ON/OFF until you can see some text on the display. Then release the ON/OFF button and press within 1...2 seconds both ENTER and MODE buttons until the text "setup" appears on the display. Then follow the instructions of the table below.
DISPLAY WHAT TO DO PRESS:
Wait for a few seconds.
If the basic settings have been given,
press ENTER repeatedly:
press ENTER.
°C
set
If they have not been given, refer to the HMI41 Operating Manual.
select the HMI41 function: 1 = indicator
2 = calibrator for HMD/W60/70 and
set
HMP140 series analogue transmitters
(number up) or (number down)
ON/OFF 3 = calibrator for HMP230 transmitters 4 = calibrator for HMD/W20/30 and HMP130 series transmitters
1996-08-26 1
Page 23
HMI41 Calibration Option Appendix 1 19116ZZ-U218en-1.1

2. Scale selection modes

The HMI41 automatically assumes the function selected in the setup menu when it is turned on again with the ON/OFF button. In the following, you will find a summary of the scale selection modes.
DISPLAY DESCRIPTION
RH Td °C°F T
Pabs Pa setminmax hysthold Lobat Hi
RH T
bat
%
°C°F
g/m
Hi
The HMI41 has been turned on.
Indication of the software version (if the version is 1.02 or more)
Indication of the battery voltage.
Press any button except ON/OFF within 5 seconds to enter the scale selection modes.
Select the desired scale selection mode with buttons or and acknowledge the selection with ENTER. Then follow the instructions of tables 2.1 - 2.5 accordingly.

2.1 Voltage output scale selection mode (U)

DISPLAY WHAT TO DO PRESS:
(number up) or
Select the voltage output scale selection mode (U)
(number down) ENTER to acknowledge the selection
(number up) or
C T
set
°C
Lo
Set the lower limit of the temperature range
(number down) ENTER to acknowledge the selection
(number up) or
C T
set
°C
Hi
Set the upper limit of the temperature range
(number down) ENTER to acknowledge the selection ON/OFF
2 1996-08-26
Page 24
HMI41 Calibration Option
19116ZZ-U218-1.1 Appendix 1

2.2 Current output scale selection mode (I)

DISPLAY WHAT TO DO PRESS:
(number up) or Select the current output scale selection mode (I)
(number down)
ENTER to acknowledge the
selection
(number up) or Select the output (0...20 mA or 4...20 mA)
set
(number down)
ENTER to acknowledge the
selection
(number up) or
C T
set
°C
Lo
Set the lower limit of the temperature range
(number down)
ENTER to acknowledge the
selection
C T
set
°C
Hi
Set the upper limit of the temperature range
(number up) or
(number down)
ENTER to acknowledge the
selection
ON/OFF
1996-08-26 3
Page 25
HMI41 Calibration Option Appendix 1 19116ZZ-U218en-1.1

2.3 Customized voltage output selection mode (cU)

DISPLAY WHAT TO DO PRESS:
(number up) or
Select the customized voltage output scale selection mode (cU)
(number down) ENTER to acknowledge the selection
RH
Select the jumper selected RH channel’s
set
voltage output range (0...1V, 0...5V, 0...10V)
RH
Set the lower limit of the RH channel’s
set
Lo
voltage range
(number up) or (number down)
ENTER to acknowledge the selection
(number up) or (number down)
ENTER to acknowledge the selection
RH
Set the upper limit of the RH channel’s
set
Hi
voltage range
(number up) or (number down)
ENTER to acknowledge the selection
RH
set
%
Set the lower limit of the humidity range
Lo
(number up) or (number down)
ENTER to acknowledge the selection
RH
set
%
Set the upper limit of the humidity range
Hi
(number up) or (number down)
ENTER to acknowledge the selection
T
Select the jumper selected T channel’s
set
voltage output range (0...1V, 0...5V, 0...10V)
(number up) or (number down)
ENTER to acknowledge the selection (number up) or
T
set
Lo
Set the lower limit of the T channel’s voltage range
(number down) ENTER to acknowledge the selection (number up) or
T
set
Hi
Set the upper limit of the T channel’s voltage range
(number down) ENTER to acknowledge the selection (number up) or
C T
set
°C
Lo
Set the lower limit of the temperature range
(number down) ENTER to acknowledge the selection (number up) or
C T
set
°C
Hi
Set the upper limit of the temperature range
(number down) ENTER to acknowledge the selection press ON/OFF
4 1996-08-26
Page 26
HMI41 Calibration Option
19116ZZ-U218-1.1 Appendix 1

2.4 Customized current output selection mode (cI)

DISPLAY WHAT TO DO PRESS:
(number up) or Select the customized current output scale selection mode (cI)
(number down)
ENTER to acknowledge the
selection
(number up) or
RH
set
Select the RH channel’s current output range (4...20 mA, 0...20 mA)
(number down)
ENTER to acknowledge the
selection
(number up) or
RH
%
Set the lower limit of the humidity range
(number down)
ENTER to acknowledge the
set
Lo
selection
(number up) or
RH
%
Set the upper limit of the humidity range
(number down)
ENTER to acknowledge the
set
Hi
selection
(number up) or
T
set
range (4...20 mA, 0...20 mA)
Select the T channel’s current output
(number down)
ENTER to acknowledge the
selection
C T
set
C T
set
°C
Lo
°C
Hi
Set the lower limit of the temperature range
Set the upper limit of the temperature range
(number up) or
(number down)
ENTER to acknowledge the
selection
(number up) or
(number down)
ENTER to acknowledge the
selection
press ON/OFF
1996-08-26 5
Page 27
HMI41 Calibration Option Appendix 1 19116ZZ-U218en-1.1

2.5 Calibration

DISPLAY WHAT TO DO PRESS:
RH T
Wait for about 5 seconds.
Let the readings stabilize and then adjust
RH %
the transmitter potentiometers until the readings are the same. Alternatively, you can calibrate the transmitter in the second RH calibration mode (see below).
Let the readings stabilize and then adjust
RH %
the transmitter potentiometers until the reading on the first line is at zero
Let the readings stabilize and then adjust
C T
°C
the transmitter offset potentiometer until the readings are the same. Alternatively, you can calibrate the transmitter in the second T calibration mode (see below).
Let the readings stabilize and then adjust
C T
°C
the transmitter offset potentiometer until the reading on the first line is at zero.
HOLD to enter the second RH calibration mode MODE to enter the first T calibration mode ON/OFF if you have completed the calibration HOLD to return to the first RH calibration mode MODE to enter the second T calibration mode ON/OFF if you have completed the calibration HOLD to enter the second T calibration mode MODE to return to the first RH calibration mode ON/OFF if you have completed the calibration
HOLD to return to the first T calibration mode MODE to return to the second RH calibration mode ON/OFF if you have completed the calibration
6 1996-08-26
Page 28
Page 29
Page 30
*U218
EN
*
www.vaisala.com
Loading...