mechanical (including photocopying), nor may its contents be communicated to a third
party without prior written permission of the copyright holder.
The contents are subject to change without prior notice.
Please observe that this manual does not create any legally binding obligations for
Vaisala towards the customer or end user. All legally binding commitments and
agreements are included exclusively in the applicable supply contract or Conditions of
Sale.
Chapter 1 _________________________________________________________ General Information
CHAPTER 1
GENERAL INFORMATION
Safety
Throughout the manual, important safety considerations are
highlighted as follows:
WARNING
CAUTION
NOTE
Warning alerts you to a serious hazard. If you do not read and follow
instructions very carefully at this point, there is a risk of injury or
even death.
Caution warns you of a potential hazard. If you do not read and
follow instructions carefully at this point, the product could be
damaged or important data could be lost.
Note highlights important information on using the product.
Vaisala hereby represents and warrants all Products
manufactured by Vaisala and sold hereunder to be
free from defects in workmanship or material during
a period of twelve (12) months from the date of
delivery save for products for which a special
warranty is gi ven. If any P roduct p roves ho wever to
be defective in workmanship or material within the
period herein provided Vaisala undertakes to the
exclusion of any other remedy to repair or at its own
option replace the defective Product or part thereof
free of charge and otherwise on the same conditions
as for the original Product or part without extension
to original warranty time. Defective parts replaced in
accordance with this clause shall be placed at the
disposal of Vaisala.
Vaisala also warrants the quality of all repair and
service works performed by its employees to
products sold by it. In case the repair or service
works should appear inadequate or faulty and should
this cause malfunction or nonfunction of the product
to which the service was performed Vaisala shall at
its free option either repair or have repaired or
replace the product in question. The working hours
used by employees of Vaisala for such repair or
replacement shall be free of charge to the client.
This service warranty shall be valid for a period of
six (6) months from the date the service measures
were completed.
This warranty is however subject to following
conditions:
a) A substantiated written claim as to any alleged
defects shall have been received by Vaisala
within thirty (30) days after the defect or fault
became known or occurred, and
b) The allegedly defective Product or part shall,
should Vaisala so require, be sent to the works of
Vaisala or to such other place as Vaisala may
indicate in writing, freight and insurance prepaid
and properly packed and labelled, unless Vaisala
agrees to inspect and repair the Product or
replace it on site.
a) normal wear and tear or accident;
b) misuse or other unsuitable or unauthorized use of
the Product or negligence or error in storing,
maintaining or in handling the Product or any
equipment thereof;
c) wrong installation or assembly or failure to
service the Product or otherwise follow Vaisala
service instructions including any repairs or
installation or assembly or service made by
unauthoriz ed personnel not a pproved by V aisala
or replacements with parts not manufactured or
supplied by Vaisala;
d) modifications or changes of the Product as well
as any adding to it without Vaisala's prior
authorization;
e) other factors depending on the Customer or a
third party.
Notwithstanding the aforesaid Vaisala's liability
under this clause shall not apply to any defects
arising out of materials, designs or instructions
provided by the Customer.
This warranty is expressly in lieu of and excludes all
other conditions, warranties and liabilities, express
or implied, whether under law, statute or otherwise,
including without limitation ANY IMPLIED
WARRANTIES OF MERCHANTABILITY OR OF
FITNESS FOR A PARTICULAR PURP OSE and all
other obligations and liabilities of Vaisala or its
representatives with respect to any defect or
deficiency applicable to or resulting directly or
indirectly from the Products supplied hereunder,
which obligations and liabilities are hereby
expressly cancelled and waived. Vaisala's liability
shall under no circumstances exceed the invoice
price of any Product for which a warranty claim is
made, nor shall Vaisala in any circumstances be
liable for lost profits or other consequential loss
whether direct or indirect or for special damages.
This warranty does not however apply when the
defect has been caused through
Temperature of the probe and the concrete shall be the same!
In humidity measurement and especially in calibration it is essential that temperature of the
probe and measured object is the same. Even a small difference in temperature between the
measured object and the probe causes an error. As the curve below shows, if the temperature
is +20 °C and the relative humidity 100 %RH, a difference of ±1 °C between the measured
object and the probe causes an error of ±6 %RH. When the humidity is 90 %RH, the
corresponding error is ±5.4 %RH.
A temperature difference of a few degrees can also cause water to condense on the sensor
surface. HUMICAP sensor starts to function normally as soon as the water has evaporated. If
the condensed water is contaminated, the life span of the probe may shorten and calibration
may change.
10
9
8
7
6
5
4
dRH (%RH)
3
2
1
0
-40-200 20406080100
Temperature (°C)
Figure 3.1Measurement error at 1 00 %RH w hen the dif feren ce
between the ambien t and sens or te mpera ture is 1 °C
Chapter 4 _____________________________________________________________ Taken into use
CHAPTER 4
TAKEN INTO USE
Before taking into use the HM44 do the following:
- place the batteries into HMI indicator
- connect the probe
- feed the individual probe adjustment coefficients
- if needed, check and change the settings
The basic display appears when turning the device on:
RH
T
Currently selected probe
%
°C
Relative
humidity
Temperature
Feeding the individual adjustment coefficients
Each probe has an individual adjustment coefficients which are written in the calibration
certificate (see Appendix 1). These coefficients should be fed to the HMI41 memory to reach
the measurement accuracy of ±2 %RH.
- When using one probe, numbering of the probe is not needed.
- When using more than one probe, firstly select a number for the probe and then feed the
probe individual coefficients to the HMI41 memory. Remember to write the probe number
(1-10) in the product label to be able to identify the probes afterwards.
The first number of the second line of display indicates the currently selected probe (1-10).
Connect a probe and change the number on the display by pressing ENTER and
simultaneously buttons ▲ or ▼. Remember to write the probe number (1-10) in the product
label to be able to identify the probes.
Feeding the offset correction
1. Press the button ENTER continuously for about 5 seconds until a text similar
to the following appears on the display (numbers on the second line are
blinking):
RH
set
2. Feed the individual offset correction (see the calibration certificate) in the
HMI41 memory with buttons ▲ and ▼. Each time the button is pressed, the
numbers change 0.1 %RH. If the button is pressed continuously, the numbers
change in a faster rate.
3. When the reading is correct, press ENTER and the display changes to show
the gain correction.
Feeding the gain correction
RH
set
4. Numbers on the second line are blinking; you can now enter the individual
gain correction in the HMI41 memory. The numbers are changed with
buttons ▲and ▼. Each time the button is pressed, numbers change 0.001. If
the button is pressed continuously, the numbers change in a faster rate.
NOTE
If the correction given in the calibration certificate has only two
decimals (e.g. 1.03), enter a zero in place of the third decimal (e.g.
Chapter 4 _____________________________________________________________ Taken into use
5. When the reading is correct, press ENTER and the display changes to show
the humidity and temperature readings. However, an arrow is now displayed
as an indication of the individual corrections in the HMI41 memory:
RH
T
%
°C
Changing the settings
Change the settings ONLY if there is a need to change the following factory settings:
Factory settings
• display units:0(metric)
• automatic power-off:5 (minutes)
• display quantities:0(RH, T and Td)
• pressure:1013.25hPa (1 hPa = 1 mbar)
• probe t ypeAUT
• start5
To change the settings, enter the setup menu.
Enter the setup menu
Turn the device ON (by using ON/OFF button) and immediately (within 1..2 seconds) after
turning on, press together both the buttons ENTER and MODE for about 3 seconds, until the
following appears:
In a few seconds, the text changes automatically to the following:
•Select the display units with buttons ▲ or ▼. Choose 0 for metric units or 1 for non-
metric units (see Table 1).
• Press ENTER and the display changes to show the setting of the automatic power-off
function.
Table 1. Metric and non-metric units
quantitymetricnon-metric
relative humidity (RH)%RH%RH
temperature (T)
dewpoint temperature (Td)
absolute humidity (a)g/m
mixing ratio (x)g/kggr/lb
wet bulb temperature (Tw)
°C°F
°C°F
3
gr/ft
°C°F
3
Setting the automatic power-off function
set
The number (or text NO) on the first line of the display indicates in minutes (1...60) the time
that the HMI41 stays on before an automatic power-off if no buttons are pressed.
• changed with buttons ▲ and ▼. If NO is chosen, the automatic power-off
function is not activated.
• Press ENTER and the display changes to show the selection of the display
quantities. NOTE: during data collecting this setting is not valid.
Chapter 4 _____________________________________________________________ Taken into use
Selecting the display quantities
set
The HMI41 displays relative humidity (RH), temperature (T) and dewpoint temperature (Td)
readings. Additionally, one of the following quantities can be chosen: absolute humidity (a),
wet bulb temperature (Tw) and mixing ratio (x). The number on the display indicates the
following quantities:
Chapter 5 ____________________________________________ Humidity measurements in concrete
CHAPTER 5
HUMIDITY MEASUREMENTS IN
CONCRETE
General about humidity measurements in a
drilled hole
The HMP44 probe measures humidity of concrete in a drilled hole. In equilibrium stage, the
humidity of a concrete and air humidity in a drilled hole become same.
Drilling changes the stage of the concrete. This is due to the fact that heat is formed during
drilling.(Meantime the humidity of the concrete changes, because of the temperature
changes). To reach the equilibrium stage after drilling takes about three days. When the
equilibrium stage has been reached, the air humidity in a drilled hole and the concrete
humidity become equal. Then it is possible to make the humidity measurement.
For getting more reliable results, it is recommended to have several measurement points.
Measurement depth
Depth of a drilled hole
When drying of a concrete slab takes place from both sides, the minimum depth of
measurement (d) is made 20 % of the depth of the slab. When drying occurs from one side
only, the measurement depth (d ) is 40 % of the depth of a slab. See Figure 5.1.
Humidity profile (Figure 5.1)
The humidity profile of the slab depends on the drying of the concrete, i.e. whether the
concrete dries in one or in two directions. In concrete slabs drying in two directions, the
humidity is highest in the middle. After coating, the humidity spreads evenly throughout the
slab until it corresponds to readings measured at the depth of approximately 20% of the slab.
In concrete slabs drying in one direction only, the humidity is highest at the bottom. After
coating, the humidity spreads evenly throughout the slab until it corresponds to readings
measured at the depth of approximately 40% of the slab.
DRYING IN TWO DIRECTIONS
coating
hole for the
plastic tube
humidity profile
hole for the
plastic tube
humidity profile
concrete slab
BEFORE COATING
hole for the
plastic tube
concrete slab
0.2xd
d
concrete slab
0.2xd
AFTER COATING
DRYING IN ONE DIRECTION
coating
humidity profile
0.4xd
BEFORE COATING
concrete slab
d
hole for the
plastic tube
AFTER COATING
humidity profile
0.4xd
Figure 5.1 Humidity profile before and after the coating