Vaisala HM34 User Manual [en, de, fr]

User's Manual Mode d'emploi Bedienungsanleitung
Käyttöohje
HM34
Humidity and Humidité- et
temperature meter températuremétre
Temperaturmeter lämpötilamittari
Ref. HM34-U262en-1.1
CONTENTS Page
1. Product description 1
2. Operation 2
3. Calibration and maintenance 3
4. Spare parts and accessories 4
5. Technical specifications 5
TABLE DES MATIERES Page
1. Description de l'appareil 7
2. Mode d'emploi 8
3. Etalonnage et entretien 9
4. Pieces detachees et accessoires 10
5. Caracteristiques techniques 11
INHALT Seite
1. Beschreibung 13
2. Bedienung 14
3. Eichung und Wartung 15
4. Ersatzteile und Zubehör 16
4. Technicshe Daten 17
SISÄLTÖ Sivu
1. Tuotekuvaus 19
2. Käyttö 20
3. Kalibrointi ja huolto 21
4. Varaosat ja lisätarvikkeet 22
5. Tekniset tiedot 23
WARRANTY - GARANTIE- GARANTIE - TAKUU 25
USERS' MANUAL
HM34 Relative humidity and temperature meter
1. PRODUCT DESCRIPTION
The HM34 is a battery-operated pocket-size meter for the meas­urement of ambient relative humidity and temperature. The humidity and temperature sensors are housed at the tip of an extendable probe, where a semipermeable membrane filter protects the humidity probe against water and dust. For easy use, the meter features a HOLD function for the display and an automatic power-off function.
For humidity measurement, the HM34 utilizes Vaisala's high­accuracy, high-reliability HUMICAP sensor. Temperature measurement is based on a Pt 100 tempera-
ture sensor.
The meter is intended for ambient measurements in such appli­cations as:
- spot checks
- industrial monitoring and inspections
- occupational health and safety
- laboratory and research use
The meter is available with the display calibrated for either degrees Celsius (°C) or degrees Fahrenheit (°F).
®
humidity sensor, the H-
1
2. OPERATION
The meter has four function pushbuttons and a display (Fig. 3).
1. On/off pushbutton (ON/OFF)
To switch the power on, press the ON/OFF pushbutton. This will automatically start the measurement of relative humidity (%RH).
2. Temp measurement pushbutton (TEMP)
For temperature measurement, press the TEMP pushbutton. TEMP measurement mode is indicated by a (+) or a (-) on the display. Display calibration is either in degrees Celsius (°C) or Fahrenheit (°F) as marked on the sticker on the backside of the meter.
3. RH measurement pushbutton (%RH)
Relative humidity is always positive. If a colon (:) is indicated in front of the displayed reading, the meter reads "negative" RH. This condition may occur e.g. when the meter has not been calibrated after sensor change.
4. Reading hold pushbutton (HOLD)
Pressing the HOLD pushbutton locks the reading in the display. A (s) indicates the HOLD mode. The meter remains in HOLD mode until measurement is resumed by pressing any of the pushbuttons except ON/OFF.
5. Low battery indication
When battery voltage is low, a (B) is displayed. To ensure troublefree operation, replace the battery after about 50 h of operation.
Automatic power-off
The meter automatically switches off after about 3 minutes, unless a pushbutton is pressed or the meter is set in HOLD mode.
Operating voltage
The meter uses a standard 9 V transistor battery (type IEC 6F22). The battery is located under a protective lid on the backside of the meter (Fig. 2).
2
3. CALIBRATION AND MAINTENANCE
Calibration and maintenance of the meter should be performed at regular intervals which depend on the conditions of use and the desired accuracy, but at least once a year. A one point calibration may be performed for the HM34 meter by adjusting the trimmer potentiometer marked "DRY" (Fig. 2). This field calibration can be performed against a calibrated Vaisala humid­ity meter. For a higher-accuracy two-point calibration use a Vaisala HMK11 calibrator and saturated salt solutions as de­scribed in the Operating Manual of the HMK11. The calibration is done first for the dry end and then for the wet end by adjusting trimmer potentiometers "DRY" and "WET" respectively. The potentiometers are located under the protective lid of the battery compartment (Fig. 2).
Replacing the sensors
When you replace a defective H-sensor package with a new one, make sure you handle the package with care. After installation, calibrate the humidity sensor. If the "WET" trimmer potentiometer has not been adjusted from its factory setting, a one point calibration is sufficient. Otherwise, perform a two point calibra­tion with an HMK11 calibrator. Replacement of the Pt 100
temperature sensor requires no recalibration.
3
Chemical tolerance of the humidity sensor (HUMICAP® )
Long-term exposure of the sensor to certain chemicals and gases may shorten its life. The following list gives suggested maximum
ambient concentrations for some typical chemicals:
1. Organic solvents: 1000...10 000 ppm (typ.)
2. Aggressive chemicals:
(e.g. strong acids such as
SO
3. Weak acids: 100...1000 ppm (typ.)
4. Bases: 10 000...100 000 ppm (typ.)
Detailed information on allowed concentrations are available on request from your Vaisala representative.
, H2S, HCl, Cl2 etc.) 1...10 ppm (typ.)
2,H2SO4
4. SPARE PARTS AND ACCESSORIES
Part no.
HUMICAP Temperature sensor Pt 100 (1/3 DIN 43760B) 10429 Sintered filter 133 Plastic grid 6221 HMK11 humidity meter calibrator Battery IEC 6F22 0332
®
H-sensor 15202
µm 0195
4
5. TECHNICAL SPECIFICATIONS
Measurement range humidity 0...100 %RH
temperature -20...+60
Accuracy (at +20 (including nonlinearity and hysteresis)
relative humidity against factory references against field references
temperature NOTE: When the HM34 is used in an electromagnetic field of 3 V/m, with the frequency of 200 - 220 Mhz, the temperature deviation is max. ±3 °C.
Temperature dependence:
Resolution: 0.1 %RH
Response time for humidity (at +20 °C): 15 s Display: 3 HOLD function: pushbutton hold
Supply voltage: 9 V battery (IEC 6F22) Battery operating time: 50 h (typ.) Automatic power-off: after 3 minutes, unless
°C):
+2 %RH (0...90 %RH)
+3 %RH (90...100 %RH)
+0.3 °C (+0.5 °F)
for displayed value
HOLD is activated
°C or
-4...+140
+1 %RH
+0.04 %RH/°C
+0.02 °C/°C
0.1 °C (0.1 °F)
1
/2-digit LCD
°F
5
Sensors Humidity: HUMICAP
Temperature: Pt 100 (1/3 DIN 43760B) Sensor protection:
standard plastic grid optional 133 µm sintered filter
Weight: 250 g Ordering information: HM34C
(with °C temp. display)
(with °F temp. display)
Information and specifications in this manual subject to change without notice.
Copyright Vaisala Oy, 1996.
HUMICAP is a registered trademark of the Vaisala Company.
®
H-sensor
(part no. 15202)
(part no. 6221)
(part no. 0195)
HM34F
6
MANUEL D'UTILISATION
HM34 Appareil de mesure de l'umidité et de la temperature
1. DESCRIPTION DE L'APPAREIL
Le HM34 est un appareil de mesure destiné aux mesures de température et d'humidité relative. Les capteurs de température et d'humidité sont situés à la pointe de la sonde rétractable et le capteur de température est protégé par un filtre à membrane. L'appareil possède une fonction de maintien de la valeur lue et un arrêt automatique.
Le HM34 est un appareil de grande précision et très fiable, équipé du capteur-H performant de Vaisala, HUMICAP de tempétature Pt 100.
Cet appareil de mesure fonctionnant sur pile convient aux applications suivantes:
- surveillance des locaux
- industrie
- protection du travail
- laboratoires et recherche
Le HM34 est disponible avec des échelles de mesure Celsius (°C) ou Fahrenheit (°F).
®