GelDoc-It™ TS Imaging System
Installation and User Instructions
UVP, LLC Ultra-Violet Products Ltd.
2066 W. 11th Street Unit 1, Trinity Hall Farm Estate
Upland, CA 91786 Nuffield Road, Cambridge CB4 1TG UK
Phone: (800) 452-6788 Phone: +44(0)1223-420022
Fax: (909) 946-3597 Fax: +44(0)1223-420561
Web Site: www.uvp.com
81-0259-01 Rev. K
GelDoc-It TS Imaging System 2
Gel Camera with Lens Kit
(hidden behind the LCD
touch screen)
Filter Tray (located under
the camera) with EtBr
Filter
Transilluminator on
roll-out tray
Introduction
The GelDoc-It TS System is designed as a self-contained imaging system with a built-in computer, no
external computer required. The system enables capture and documentation of fluorescent gels with the
ability to save images to the internal system or to an external USB storage device for later quantitative
analysis or enhancement for publication.
System Components
Components
Refer to the packing slip and pictured components for parts included with the system.
Computer
Operating System: Windows XP
Processor: 1.5GHz (minimum)
USB storage device capacity: 1G (minimum)
Internal storage (hard disk): 60GB (minimum)
USB 2.0 ports: 1 front, 4 in the back
Software Options: TS software (supplied only with the manual lens)
VisionWorks®LS software (supplied with the
motorized lens)
GelDoc-It TS Imaging System 3
Camera
GelCam 310 Camera, CCD monochrome model with a resolution of 1600 x 1200 with a USB 2.0 PC
interface. All camera settings are factory pre-set for optimum performance when viewing gels, films, or
membranes under low light level conditions. UVP Technical Support should be contacted before making
any adjustments to the camera settings.
Lens Kit (Motorized or Manual)
The zoom lens comes in two configurations.
The motorized version is controlled via the computer.
The manual version needs manual adjustment from the user.
Both lenses are fitted with a close-up diopter. The diopter is used for focusing on objects at the focus
length of the GelDoc-It TS System.
Ethidium Bromide Filter (EtBr)
The Ethidium Bromide (EtBr) Filter (50mm sq.) UV blocking band pass interference filter blocks UV and
IR radiation emitted from the transilluminator. The filter is placed in the filter tray below the camera
assembly. The filter allows visualization of fluorophores from 580-630nm, targeting the EtBr emission
which peaks at 605nm. The EtBr filter can be substituted for other specific fluorophore filters or removed
when imaging non-fluorescent media (protein gels, colony plates, etc.) in order to produce brighter
images.
Darkroom
The darkroom is light tight for optimum imaging conditions. The darkroom features:
Overhead white light
Roll out tray for the transilluminator
Gel viewer window
Side pocket for storage of materials
Brackets for installing optional UV lamps for overhead UV illumination
Built in power cord for an optional LED white light tray
Transilluminator
A transilluminator may be included with the system; refer to the packing list. The GelDoc-It TS System
can accommodate UVP’s Benchtop and FirstLight transilluminator models. UVP offers a variety of
configurations, from the benchtop models with 8-watt, single wavelength and single intensity to multiple
wavelengths and high/low intensities, or the high-end FirstLight transilluminator.
Note: For UV protection, there is a 10-minute timer installed in the system. The timer shuts off the
transilluminator after 10 minutes of operation. A switch on the front of the darkroom allows the user to
reset the timer. When the darkroom door is opened, this automatically resets the timer.
LCD Touch Screen
The display is a 15-inch VGA color touch screen connected to the darkroom cabinet with tilt adjustment.
The touch screen allows the user to preview, snap, save and print images, as well as select certain
preference options, without the need of an external mouse or keyboard.
A stylus pen is included for increased pointer control in selecting software options on the touch screen.
Focus Target
The focus target fluoresces when placed on a transilluminator or when exposed to overhead UV. The
target provides sharp, fluorescent images to aid in adjusting the focus of the camera.
GelDoc-It TS Imaging System 4
Optional Equipment
Refer to the ordering information section of this manual for
optional equipment part numbers.
Thermal Printer
The thermal printer provides archive quality, 256 grayscale
prints, and five optional cost-effective print sizes.
UV Handheld Lamps for Overhead lighting
A set of two 4-watt ultraviolet lamps can be connected inside
the darkroom for UV epi-illumination. The lamps can be
switched On/Off using the switch on the front of the
darkroom.
Converter Plates
The White Light Tray plugs into the darkroom for uniform
white light transillumination.
The UV/White Converter Plate allows imaging of nonfluorescent stained media with an ultraviolet transilluminator.
The converter plate is specially coated to convert the 302nm
UV output to white light, rather than using a separate white
light box.
The Visi-Blue Converter Plate converts UV to 460-470nm
designed for use with blue excitation samples and SYBR
green, SYPRO orange and GFP stains.
Transilluminators
If the system does not include a transilluminator (check the
packing slip), contact UVP for ordering information or go to
UVP’s web site at www.uvp.com. A large selection of
transilluminators is available from UVP. These units include
compact 8-watt models to our high performance FirstLight
Transilluminator.
VisionWorksLS Analysis Software
The GelDoc-It TS System can be configured with VisionWorks®LS software for analysis of gels, plates,
and membranes. The software features image enhancement, complete analysis tools and reporting
capabilities.
Mouse and Keyboard
An external keyboard and mouse can be connected to the GelDoc-It system. These components
connect to any available USB port in the back of the darkroom.
GelDoc-It TS Imaging System 5
1. Carefully remove the filter from the protective plastic
case, holding the filter at the edges to prevent
fingerprints.
2. The filter tray is located under the camera bracket on
the right side top of the darkroom. Slide the tray open.
3. Place the filter inside the filter tray so that the text on
the edge of the filter is positioned so the letters are
right side up when facing you.
Set-up Instructions
Components
Camera with zoom lens, bracket, filter, cables
GelDoc-It TS darkroom
Windows XP operating system and software to control the GelDoc-It TS
USB storage device, 64MB (minimum)
Transilluminator (may be ordered separately)
WARNING: DO NOT ATTEMPT TO CONNECT ANY WIRING WHILE THE
EQUIPMENT IS CONNECTED TO ANY POWER SUPPLY
CAUTION
Do not install the system in places with high moisture, dust, or high temperature. Do not use any oil or
petroleum based cleaner for the cabinet. Use only mild soap or detergent solution for cleaning. Ensure that
the system is turned OFF during cleaning. Keep the equipment away from motors or other large magnetic
equipment apparatus.
Installing the Filter
To install the 50mm sq Ethidium Bromide (EtBr) filter:
Additional filters are available. Refer to Accessories for ordering information.