![](/html/89/8954/89542fdef92889e24db3d4626b7d2b57ad6302b2677205d07fab08248921dd94/bg1.png)
Digital Clamp-On
Multimeter
INSTRUCTION MANUAL
ENGLISH
600V CAT III 400A
DL49B
Frequency (Hz) / Duty Cycle
600V CAT III 400A
DL49B
True RMS
Rel / Zero
Hz / Duty
x1 = Hz
x2 = Duty
Hz / Duty
CATII 1000V
CATII 750V
CATIII 600V
FEATURE LEGEND
• Press to hold the reading on the display.
• Press again to return to live reading.
• Press to enter Max / Min mode; the largest and smallest
values will be saved while in this mode.
• Press repeatedly to alternate between the maximum and
minimum readings.
• Press for 2 seconds to return to live reading and clear the
stored maximum and minimum values.
Note: Select range prior to selecting Min/Max to capture
large values
• Press repeatedly to cycle through manual ranges.
• Press for 2 seconds to return to auto ranging mode.
• AT is displayed on LCD only during auto ranging mode.
Note: Select range prior to Min/Max for best results.
Rel / Zero
Hz / Duty
• Press to enable/disable mode and set reference value.
• Display will show the difference between the set reference
• Select or setting.
• Select between AC and DC voltage.
• Select between AC and DC current.
DL49B
True RMS
Rel / Zero
CATII 1000V
CATII 750V
CATIII 600V
Hz / Duty
1-800-547-5740 • Fax: (503) 643-6322
www.ueitest.com • email: info@ueitest.com
DL49B
Instruction Manual
GENERAL SPECIFICATIONS
• Altitude: Operating - up to 2000m
Storage - 10,000m
• Humidity: 80% max
• Operating Temperature: 32°F to 122°F (0°C to 50°C) at < 75% R.H
• Storage Temperature: -4°F to 140°F (-20°C to 60°C) at < 80% R.H
• Relative humidity: 0% to 80% at 32°F to 95°F (0°C to 35°C),
0% to 70% at 32°F to 131°F (0°C to 55°C)
• Temperature Coefcient: Nominal 0.1 x (Specified accuracy) / °C
(<18°C or >28°C ; <64°F or >82°F)
• Pollution degree: 2
• Display: 3-3/4 digits 4000 counts single LCD display(s) with 20
segments bargraph
• Refresh Rate: 3 times/sec
• Overrange: "OL" is displayed
• Polarity: Automatic(no indication for positive polarity) ; Minus(-) sign
for negative polarity
• Dimensions: 8.625" x 3.25" x 1.5"
• Weight: 11.9oz.
• Calibration: Accurate for one year
• CAT Rating: CAT III 600V, CAT II 1000V
• Certications: UL & cUL Listed IEC61010-2-032
• Battery type: 2 x 1.5V AAA or LR03
• IEC61010-2-031 Silicon Test Lead
• Accuracy: ± (% of reading + # of least signicant digits)
WARNINGS
To ensure safe operation and service of the tester, follow these instructions.
Failure to observe these warnings can result in severe injury or death.
• Before each use, verify meter operation by measuring a known voltage
or current.
• Never use the meter on a circuit with voltages that exceed the category
based rating of this meter.
• Do not use the meter during electrical storms, or in wet weather.
• Do not use the meter or test leads if they appear to be damaged.
• Ensure meter leads are fully seated, and keep ngers away from the
metal probe contacts when making measurements.
• Do not open the meter to replace batteries while the probes are
connected.
• Use caution when working with voltages above 60V DC, or 25V AC RMS.
Such voltages pose a shock hazard.
• To avoid false readings that can lead to electrical shock, replace
batteries if a low battery indicator appears.
• Unless measuring voltage or current, shut off and lock out power before
measuring resistance or capacitance.
• Always adhere to local and national safety codes. Use Personal
Protective Equipment (PPE) to prevent shock and arc blast injury.
Features:
Frequency Measurement
Range Resolution Accuracy Overload Protection
99.99Hz 0.01Hz
999.9Hz 0.1Hz
9.999kHz 1Hz
99.99kHz 10Hz
Sensitivity: 1.8V RMS
Duty Cycle Measurement
Range Accuracy Overload Protection
1.0 ˜ 99.0% ±(0.2% per kHz + 0.1% + 5 digits) 600V RMS
± (0.1% + 4 digits) 600V RMS
FUNCTIONS
• AC/DC Voltage
• AC/DC Current
• Resistance ..................................................
• Frequency & Duty cycle ................................
• Diode test ..................................................
• Continuity ..................................................
..........................................
..........................................
Hz / Duty
FEATURES
• True RMS
• Data hold mode
• MIN/MAX (All ranges except Frequency & Capacitance)
• DC A Zero mode
• Test lead storage
• Autoranging measurements with manual ranging capability
• Bar Graph
a. The bar graph shows an approximate analog representation of a
measurement.
b. The bar graph responds much faster than the digital display.
c. The scale of the bar graph is zero to the maximum reading of
the selected range.
• Auto-Power-Off: After 30 minutes of non-use
• Low battery: is displayed if battery voltage drops below
operating voltage.
......................................
..
Min/Max
....................................
Rel/Zero
HOLD
SYMBOLS USED ON LCD
AC Measurement DC Measurement
~
Negative DC Value
Overload: Range Exceeded
O
.L.
Low Battery
Minimum Reading
MIN
Duty Cycle Mode Frequency Mode
%
Voltage Measurement
V
Resistance in Ohms Diode Test
Ω
Relative / Zero Mode
m
Milli ( x 10
-3
)
Auto Range Active
AT
Auto Power-Off Active
Apo
Hold Active
HOLD
Maximum Reading
M
AX
Current in Amps
A
Kilo ( x 103)
k
Mega ( x 106 )
M
INTERNATIONAL SYMBOLS
AC Alternating Current Warning or Caution
~
DC Direct Current Dangerous levels
DC/AC Voltage or Current Double Insulated Class II
Ground
AC Source
Safe for disconnect
from live conductors
Auto / Manual Range
600V CAT III 400A
DL49B
True RMS
Rel / Zero
Hz / Duty
CATII 1000V
CATII 750V
CATIII 600V
x1x1
WARRANTY
The DL49B is warranted to be free from defects in materials and workmanship
for a period of three year from the date of purchase. If within the warranty
period your instrument should become inoperative from such defects, the unit
will be repaired or replaced at UEi’s option. This warranty covers normal use
and does not cover damage which occurs in shipment or failure which results
from alteration, tampering, accident, misuse, abuse, neglect or improper
maintenance. Batteries and consequential damage resulting from failed
batteries are not covered by warranty.
Any implied warranties, including but not limited to implied warranties of
merchantability and fitness for a particular purpose, are limited to the express
warranty. UEi shall not be liable for loss of use of the instrument or other
incidental or consequential damages, expenses, or economic loss, or for any
claim or claims for such damage, expenses or economic loss.
A purchase receipt or other proof of original purchase date will be required
before warranty repairs will be rendered. Instruments out of warranty will be
repaired (when repairable) for a service charge.
For more information on warranty and service:
www.ueitest.com • Email: info@ueitest.com
1-800-547-5740 • FAX: (503) 643-6322
This warranty gives you specific legal rights. You may also have other rights
which vary from state to state.
BATTERY REPLACEMENT
• When indicator is displayed on the LCD, batteries must be replaced.
•Remove the back screw and replace 2 x AAA batteries.
CLEANING
Turn instrument off and disconnect test leads. Clean the instrument by using
a damp cloth. Do not use abrasive cleaners or solvents.
STORAGE
Remove the batteries when instrument is not in use for a prolonged period
of time. Do not expose to high temperatures or humidity. After a period
of storage in extreme conditions exceeding the limits mentioned in the
Specifications section, allow the instrument to return to normal operating
conditions before using it.
DISPOSAL / RECYCLE
Caution: This symbol indicates that equipment and its
accessories shall be subject to a separate collection and correct
disposal.
PLEASE
RECYCLE
Copyright © 2014 UEi. All Rights Reserved
17131 05/14
![](/html/89/8954/89542fdef92889e24db3d4626b7d2b57ad6302b2677205d07fab08248921dd94/bg2.png)
AC Current (large): < 400A
600V CAT III 400A
Single
Conductor
Only
AC / DC Voltage: < 750V AC or 1000V DC
600V CAT III 400A
Diode
GOOD DIODE
600V CAT III 400A
DL49B
True RMS
Rel / Zero
Hz / Duty
CATII 1000V
CATII 750V
CATIII 600V
• Center wire in guides for best accuracy.
• Opposing currents cancel (use line-splitter when necessary).
Features:
Rel / Zero
AC Current Measurement
Range Resolution Accuracy Overload Protection
40A 0.01A ± (2.9% + 15 digits)
400A 0.1A ± (1.9% + 8 digits)
600V RMS
True RMS: 45Hz to 400Hz
* Minimum Current for Clamp Measurement: 0.2A
DC Current (large): < 400A
600V CAT III 400A
DL49B
True RMS
Rel / Zero
Rel / Zero
Hz / Duty
CATII 1000V
CATII 750V
x1 =
CATIII 600V
x2 =
Before
DC Test
• Center wire in guides for best accuracy.
• Opposing currents cancel (use line-splitter when necessary).
Features:
Rel / Zero
DC Current Measurement
Range Resolution Accuracy Overload Protection
40A 0.01A ± (2.5% + 15 digits)
400A 0.1A ± (1.5% + 8 digits)
600V RMS
True RMS: 45Hz to 400Hz
* Minimum Current for Clamp Measurement: 0.2A
Resistance: < 40MΩ
DL49B
True RMS
Rel / Zero
Hz / Duty
CATII 1000V
CATII 750V
CATIII 600V
x1 =
x2 =
Use CAT III rated leads or higher.
Do not attempt to measure more than 1000V DC or 750V AC.
Keep hands below line when measuring high current levels.
• Select AC or DC voltage source.
Features:
DC Voltage Measurement
Range Resolution Accuracy Overload Protection
400mV 0.1mV
4V 1mV
40V 10mV
400V 100mV
1000V 1V ± (0.8% + 10 digits)
Overload Protection: 1000V
AC Voltage Measurement
Range Resolution Accuracy Overload Protection
400mV 0.1mV
4V 1mV
40V 10mV
400V 100mV
750V 1V
True RMS: 45Hz to 400Hz
• Display shows resistance.
• Buzzer sounds if less than 40Ω.
Features:
Rel / Zero
± (0.5% + 4 digits)
1000V
± (2.0% + 5 digits) 750V RMS
Continuity
600V CAT III 400A
DL49B
True RMS
Rel / Zero
Hz / Duty
CATII 1000V
CATII 750V
CATIII 600V
Forward Bias
Displays approx.
voltage drop
Reverse Bias
Displays "OL"
BAD DIODE
Open Diode
Displays "OL"
Rel / Zero
Rel / Zero
Rel / Zero
DL49B
True RMS
CATII 1000V
CATII 750V
CATIII 600V
600V CAT III 400A
DL49B
True RMS
CATII 1000V
CATII 750V
CATIII 600V
600V CAT III 400A
DL49B
True RMS
CATII 1000V
CATII 750V
CATIII 600V
600V CAT III 400A
DL49B
True RMS
Hz / Duty
Hz / Duty
Hz / Duty
600V CAT III 400A
DL49B
True RMS
Rel / Zero
Hz / Duty
CATII 1000V
CATII 750V
CATIII 600V
• Do not measure resistance on a live circuit.
Features:
Rel / Zero
Range Resolution Accuracy Overload Protection
400Ω 0.1Ω
4kΩ 1Ω
40kΩ 10Ω
400kΩ 100Ω
± (1.0% + 4 digits)
600V RMS
4MΩ 1kΩ
40MΩ 10kΩ
± (2.0% + 4 digits)
Continuity Test
Overload Protection Open Circuit Voltage
600V RMS
<
0.44V
Threshold Approx : < 40Ω
Relative Reading / DC A Zero
600V CAT III 400A
DL49B
Rel / Zero
True RMS
Rel / Zero
x1 = ON
x2 = OFF
CATII 1000V
CATII 750V
CATIII 600V
Hz / Duty
Rel / Zero
Shorted Diode
Displays "0000"
CATII 1000V
CATII 750V
CATIII 600V
Hz / Duty
or
• Forward voltage drop if forward biased.
• “O.L.” if reverse biased.
Features:
Diode Test
Overload Protection Range Test Current Open Circuit Voltage
600V RMS 2.0V Appx. 0.25mA< 1.6V DC