Operating Instructions
Disabling the Auto Power Off Function
To disable the auto power off function for long-term measurements,
turn the meter on while holding down the “HOLD” push-button. Auto
power-off will be disabled until the meter is manually turned off.
“CMP” Mode
Set the function and range switch to the desired location and turn
on the meter. Depress the “CMP” push-button. You are now in the
compare mode.
Now press the “EDIT” push-button. The “1” at the right side of the
LCD will begin to blink and the arrow at the right of the bar graph will
come on. This indicates that the value you put in will be the highest
acceptable reading. Using the arrow push-buttons found on the “CMP,
REL, % and REC” buttons, enter the HI value that you want your
readings to be compared to.
Once the value is entered, press the “HI/LO” push-button. Now the
arrow at the left of the bar graph will come on. This indicates that the
value you put in will be the lowest acceptable reading. Using the
arrow push-buttons, enter the LO value that you want your readings
compared to.
Press the “EDIT” push-button to exit the mode. From this point on, all
readings will be compared to the stored values and displayed on the
LCD in one of three ways:
• “LO” - The reading is lower than the LO stored value
• “PASS” - The reading is between the stored values
• “HI” - The reading is higher than the HI stored value
“REL” Mode
Set the function and range switch to the desired location. In the “REL”
mode, the meter displays the difference between the stored value and
the measured value.
Turn on the meter and press the “REL” push-button. “REL” will appear
on the LCD and the bar graph pointer will be positioned at the center of
the scale.
Press the “EDIT” push-button to exit the mode.
All measurements will be displayed as the difference between the stored
value and the measured value. Negative readings are lower than the
stored value. Positive readings are higher than the stored value. Pushing
the “REL” button again exits the mode.
“%” Mode
Set the function and range switch to the desired location. In the “%”
mode, the meter displays the difference in percentage between the
stored value and the measured value.
Turn the meter on and press the “%” push-button. The “%” sign will
appear on the LCD and the bar graph pointer will be positioned at zero
on the scale.
Press the “EDIT” push-button. Using the arrow buttons, enter the value
you want the readings displayed as a percentage of. Once the value is
entered, press the “EDIT” push-button to exit the mode.
C o n t rols and Indicators
1. Digital Display: Readings are displayed on a 4-4/5 digit,
50,000 count (primary display) and 5,000 count
(secondary display).
2. “CMP”, “REL” and “%” Push-buttons: These three functions
allow you to evaluate a stored reference value or values in the
meter by comparing “CMP” or in relation to “REL” or as a
percentage “%”.
3. REC Push-button: Function allows you to record minimum,
maximum and average values for a series of measurements on
the same function and range.
4. RANGE Push-button: Puts the meter in manual range mode.
holding it down for two seconds exits this mode.
5. Hold Push-buttons: Freezes the reading on the LCD for all
functions and ranges.
6. ON/OFF Push-buttons: Turns the meter on and off. If the rotary
switch or a push-button is not activated within 30 minutes, the
meter automatically turns itself off.
7. A Input Jack: The red test lead is plugged into this jack for
measuring current on the 10 AC or DC amp functions.
8. mAµA Input Jack: The red test lead is plugged into this jack for
measuring current on the 4m and 400m AC or DC amp functions.
9. VΩHz Input Jack: The red test lead is plugged into this jack for
all AC V, DC V, Ohm, Continuity Buzzer, and Diode test functions.
10. COM Input Jack: The black test lead is plugged into this jack for
all measurements except temperature.
6
DM397-MAN P. 2
7
2
3
4
5
1
8
9
10