NOTE
Selecting externally connected equipment
The CC-100 Series Computerized chart complies with European Medical Device Directive and bears the CE
marking.
Before connecting a external device to the CC-100 Series product make sure that such external equipment
is in compliance with EN 60950-1 and CE marking.
This manual is referred to device with 3.4.x software version.
Manufacturer
VISIA imaging S.r.l.
Via C. E. Gadda, 15
52027 San Giovanni Valdarno (AR)
Italy
Distributor
Topcon Europe Medical B.V.
Essebaan 11
2908 LJ Capelle a/d IJssel
The Netherlands
www.topcon.eu
medical@topcon.eu
2
INTENDED USE
Utilisation
The CC-100 Series device is a instrument for the analysis of visual acuity.
The CC-100 Series device is able to execute the principal and most important visual tests: acuity, binocular
vision, contrast sensitivity, and more.
The entire display screen is capable to display standard visual acuity tests and dissociated tests too.
The instrument display is commanded through an infrared remote control or through a direct connection with
Topcon automated phoropters.
In comparison to the classical chart projectors, the PC functions allows for greater flexibility and more
advanced functions.
Users
Oculists, ophthalmologists, opticians, optometrists.
The device must be uses by qualified personal.
Type of Environment where it could be use in
Ambulatory, physicians studios, opticians shops.
3
PRECAUTIONS
93-42 EC Directive
Classification
Class I
IEC 601-1 Classification
AC 110-120 V or 220-240V -- 50-60 Hz
220-240 V: 20x5mm 2.5A Antisurge
This instrument is an electronic precision unit, so be sure to use and store it in a place controlled under
normal living temperature, humidity and atmospheric pressure, and avoid direct exposure to sunlight.
To ensure right operation, install the instrument on the wall and check for a precise horizontal and vertical
alignment. Also do not put anything on top of the instrument.
To ensure an optimal coverage of polarized tests, align the center of LCD with the line of sight of the patient.
Wear the polarized or Red/Green lenses and select in CC-100 Series a dissociated test, check the horizontal
and vertical alignment until the best coverage for each eye is reached.
Connect all cables properly before using.
Use the power at a rated voltage.
When not in use, switch off the power source and protect the unit from sunlight and dust.
For accurate functioning, keep the instrument clean and free of spots and dust.
Technical Specifications :
Electrical Specifications:
Electromagnetic Compatibility
This product conforms to the EMC standard (IEC 60601-1-2:2007).
MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC and need to be
installed and put into service according to the EMC information provided in the accompanying
documents.
Portable and mobile RF communications equipment can effect medical electrical equipment
The use of accessories and cables other than those supplied with the instrument, exception for
cables sold by the manufacturer of the equipment as replacement parts, may result in increase
emission or decrease immunity of the equipment or system.
The equipment or system should not be used adjacent to or stacked with other equipment. If
adjacent or stacked use is necessary, the equipment or system should be observed to verify normal
operation in the configuration in which it will be used.
4
EMC Table
The CC-100 SERIE is intended for use in the electromagnetic environment specified below. The costumer or the
user of the CC-100 SERIE should assure that it is used in such an environment.
Electromagnetic enviroment guide
The CC-100 SERIE device uses RF energy only for
its internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
The CC-100 SERIE device is suitable for use in all
establishments, except for domestic
establishments and those directly connected to
the public low-voltage power supply network that
supplies buildings used to domestic purposes.
It is possible to use the device in all
establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic Emission
IEC 61000-3-2
It is possible to use the device in all
establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network that
supplies buildings used for
domestic purposes
Voltage Fluctuation/flicker
Emission
IEC 61000-3-3
5
The CC-100 SERIE device is intended for use in the electromagnetic environment specified below. The
customer or the user of the device should assure that it is used in such an environment.
Electromagnetic
enviroment guide
Electrostatic Discharge (ESD)
EN 61000-4-2
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be at
least 30 %.
Electrical Fast Transient/Burst
EN 601000-4-4
±2 kV high voltage
power line
±2 kV high voltage
power line
Mains power quality should
be that of a typical
commercial or hospital
environment.
Differential mode:
1 kV
Common mode:
2kV
Differential mode:
1 kV
Common mode:
2kV
Mains power quality should
be that of a typical
commercial or hospital
environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
EN 61000-4-11
<5% UT
(>95% dip of UT)
for 0,5 cycle
40% UT
(>60% dip of UT)
for 5 cycle
70% UT
(>30% dip of UT)
for 25 cycle
<5% UT
(>95% dip of UT)
for 5 seconds
<5% UT
(>95% dip of UT)
for 0,5 cycle
40% UT
(>60% dip of UT)
for 5 cycle
70% UT
(>30% dip of UT)
for 25 cycle
<5% UT
(>95% dip of UT)
for 5 seconds
Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of
the device requires continued
operation during power
mains interruptions, it is
recommended that the
device be powered from an
uninterruptible power supply
or a battery.
Magnetic field at power
frequency
EN 61000-4-8
Magnetic power frequency
fields
should be that of a typical
commercial or hospital
environment.
6
Immunity Radio Frequency Guidance
The CC-100 SERIE is manufactured to work in the electromagnetic environment specified below. The
operator or user should make sure it is use within those environmental conditions..
Electromagnetic enviroment
guide
Conducted RF
EN 61000-4-6
3 Veff from 150kHz to
80MHz
3 Veff from 150kHz to
80MHz
Portable and mobile RF
communications equipment should
be used no closer to any part of the
device, including cables, than the
recommended separation distance
calculated from that equation
applicable to the frequency of the
transmitter.
Recommended separation distance
d = 1,2 ⋅√P from 150kHz to 80MHz
d = 1,2 ⋅√P from 80 MHz to 800
MHz
d = 2,3 ⋅√P from 800 MHz to 2,5
GHz
where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is
the recommended separation
distance in meters (m)
3 V/m from 80MHz to
2.5GHz
3 V/m from 80MHz to
2.5GHz
The field intensity of the fixed RF transmitters, as checked in an
electromagnetic investigation of the locus, should be less than the
conformity level for each range of frequency. It is possible to have
interference near devices marked with the symbol:
Recommended separation distances between portable and mobile RF communications
equipment and the CC-100 XP device
CC-100 SERIE device is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the device can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the device as recommended below, according to the maximum output
power of the communications equipment.
Output maximum
nominal power of
the transmitter (W)
Separation distance at the transmitter frequency (m)
from 150kHz to 80MHz
D= 1,2 x P
from 80MHz to 800MHz
D= 1,2 x P
from 800MHz to 2GHz
D= 2,3 x P
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in metres (m) can be determined using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
Notes:
(1) At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
(2) These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
7
WARNING INDICATIONS AND POSITIONS
For safety purposes, this equipment includes clearly labelled warnings.
Use the equipment following this warning instructions.
CAUTION
To avoid electrical shock, do not open the instrument. Refer all servicing to only qualified personnel.
WARNING
Electrical shock may cause burns or possible fire. Turn the main power switch OFF and UNPLUG the power
cord before replacing fuses. Replace only with fuses of the correct rating.
9
CC-100 SERIES STANDARD ACCESSORIES
11
Remote Control
(only for CC-100 Series instruments)
CC-Series Remote Control Batteries
(AAA Type)
CC-SERIES OPTIONAL ACCESSESORIES
REMOTE CONTROL
CC-100 WALL MOUNTING
12
PREPARATIONS
Without WALL MOUNTING
With WALL MOUNTING
WARNING: use appropriate screws and plugs to fix the unit to the wall.
13