Tektronix TDS Family Instructions User manual

Instructions
TDS Family Option 2F Advanced DSP Math
070-8582-01
www.tektronix.com
*P070858201*
070858201
Copyright © T ektronix, Inc. All rights reserved. T ektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes
that in all previously published material. Specifications and price change privileges reserved. Printed in the U.S.A. T ektronix, Inc., P.O. Box 1000, Wilsonville, OR 97070–1000 TEKTRONIX, TEK, and TEKPROBE are registered trademarks of T ektronix, Inc. SureFoot is a trademark of T ektronix, Inc.

WARRANTY

Tektronix warrants that this product will be free from defects in materials and workmanship for a period of three (3) years from the date of shipment. If any such product proves defective during this warranty period, Tektronix, at its option, either will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the warranty period and make suitable arrangements for the performance of service. Customer shall be responsible for packaging and shipping the defective product to the service center designated by Tektronix, with shipping charges prepaid. Tektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the Tektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage resulting from attempts by personnel other than Tektronix representatives to install, repair or service the product; b) to repair damage resulting from improper use or connection to incompatible equipment; or c) to service a product that has been modified or integrated with other products when the effect of such modification or integration increases the time or difficulty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THIS PRODUCT IN LIEU OF ANY OTHER WARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX' RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.

Welcome

This instruction manual describes Option 2F, the Advanced DSP Math opĆ tion. Included in this manual are the following subsections:
H Product Description (follows this introduction)
H Fast Fourier Transforms
H Waveform Differentiation
H Waveform Integration

Related Manuals

Conventions

This manual only documents Option 2F. The following documents cover all other aspects related to the use or service of the oscilloscope. If ordering any of these manuals, order by the manual title and the model of your oscilloscope.
H The TDS User Manual guides the user in operation of this oscilloscope
and describes its features. It also contains a tutorial, a specification, and other useful information related to your oscilloscope.
H The TDS Series Programmer Manual describes how to use a computer
to control the oscilloscope through the GPIB interface.
H The TDS Reference gives you a quick overview of how to operate your
oscilloscope.
H The TDS Performance Verification tells how to verify the performance of
the oscilloscope.
H The TDS Service Manual provides information for maintaining and servicĆ
ing your oscilloscope.
In this manual, you will find various procedures that contain steps of instrucĆ tions for you to perform. To keep those instructions clear and consistent, this manual uses the following conventions:
TDS Series Option 2F Instructions
H Names of front panel controls and menu labels appear in boldface print.
H Names also appear in the same case (initial capitals, all uppercase, etc.)
in the manual as is used on the oscilloscope front panel and menus. Front panel names are all upper case letters, for example, VERTICAL MENU, CH 1, etc.
H Instruction steps are numbered. The number is omitted if there is only
one step.
v
Welcome
H When steps require that you make a sequence of selections using front
panel controls and menu buttons, an arrow ( between a front panel button and a menu, or between menus. Also, whether a name is a main menu or side menu item is clearly indicated:
Press VERTICAL MENU tion (main)
Using the convention just described results in instructions that are graphically intuitive and simplifies procedures. For example, the instrucĆ tion just given replaces these five steps:
1. Press the front panel button VERTICAL MENU.
2. Press the main menu button Offset.
3. Press the sideĆmenu button Set to 0 V.
4. Press the main menu button Position
5. Press the side menu button Set to 0 divs
H Sometimes you may have to make a selection from a popup menu:
Press SHIFT UTILITY edly press the main menu button SYSTEM until I/O (for example) is highlighted in the popĆup menu.
Set to 0 divs (side).
Offset (main) Set to 0 V (side) PosiĆ
SYSTEM (popup). In this example, you repeatĆ
) marks each transition
vi
Welcome
Fast Fourier Transforms

Contents

Contents vii. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Safety ix. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Product Description xi. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Description 1Ć1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Operation 1Ć2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Displaying an FFT 1Ć2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Cursor Measurements of an FFT 1Ć5. . . . . . . . . . . . . . . . . . . . . . . .
Automated Measurements of an FFT 1Ć7. . . . . . . . . . . . . . . . . . . .
Considerations for Using FFTs 1Ć7. . . . . . . . . . . . . . . . . . . . . . . . . . . .
The FFT Frequency Domain Record 1Ć7. . . . . . . . . . . . . . . . . . . . .
Offset, Position, and Scale 1Ć9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Record Length 1Ć10. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Acquisition Mode 1Ć10. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Zoom and Interpolation 1Ć11. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Undersampling (Aliasing) 1Ć11. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Considerations for Phase Displays 1Ć12. . . . . . . . . . . . . . . . . . . . . .
FFT Windows 1Ć14. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Waveform Differentiation
Waveform Integration
TDS Series Option 2F Instructions
Description 2Ć1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Operation 2Ć1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Displaying a Differentiated Waveform 2Ć1. . . . . . . . . . . . . . . . . . . .
Automated Measurements of a Derivative Waveform 2Ć2. . . . . . .
Cursor Measurement of a Derivative Waveform 2Ć3. . . . . . . . . . . .
Usage Considerations 2Ć3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Offset, Position, and Scale 2Ć4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Zoom 2Ć4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Description 3Ć1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Operation 3Ć1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Displaying an Integral Waveform 3Ć1. . . . . . . . . . . . . . . . . . . . . . . .
Cursor Measurements of an Integral Waveform 3Ć2. . . . . . . . . . .
Automated Measurements of a Integral Waveform 3Ć4. . . . . . . . .
vii
Contents
Index
Usage Considerations 3Ć4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Offset, Position, and Scale 3Ć4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
DC Offset 3Ć4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Zoom 3Ć5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
viii
Contents

Safety

Please take a moment to review these safety precautions. They are provided for your protection and to prevent damage to the digitizing oscilloscope. This safety information applies to all operators and service personnel.

Symbols and Terms

These two terms appear in manuals:
H
statements identify conditions or practices that could result in
damage to the equipment or other property.
H
statements identify conditions or practices that could result in
personal injury or loss of life.
These two terms appear on equipment:
H CAUTION indicates a personal injury hazard not immediately accessible
as one reads the marking, or a hazard to property including the equipĆ ment itself.
H DANGER indicates a personal injury hazard immediately accessible as
one reads the marking.
This symbol appears in manuals:
TDS Series Option 2F Instructions
StaticĆSensitive Devices
These symbols appear on equipment:
DANGER
High Voltage
Protective
ground (earth)
terminal
ATTENTION
Refer to
manual
ix
Safety

Specific Precautions

Observe all of these precautions to ensure your personal safety and to prevent damage to either the digitizing oscilloscope or equipment conĆ nected to it.
Power Source
The digitizing oscilloscope is intended to operate from a power source that will not apply more than 250 V tween either supply conductor and ground. A protective ground connection, through the grounding conductor in the power cord, is essential for safe system operation.
between the supply conductors or beĆ
RMS
Grounding the Digitizing Oscilloscope
The digitizing oscilloscope is grounded through the power cord. To avoid electric shock, plug the power cord into a properly wired receptacle where earth ground has been verified by a qualified service person. Do this before making connections to the input or output terminals of the digitizing oscilloĆ scope.
Without the protective ground connection, all parts of the digitizing oscilloĆ scope are potential shock hazards. This includes knobs and controls that may appear to be insulators.
Use the Proper Power Cord
Use only the power cord and connector specified for your product. Use only a power cord that is in good condition.
Use the Proper Fuse
To avoid fire hazard, use only the fuse specified in the parts list for your product, matched by type, voltage rating, and current rating.
Do Not Remove Covers or Panels
To avoid personal injury, do not operate the digitizing oscilloscope without the panels or covers.
Electric Overload
Never apply to a connector on the digitizing oscilloscope a voltage that is outside the range specified for that connector.
Do Not Operate in Explosive Atmospheres
The digitizing oscilloscope provides no explosion protection from static discharges or arcing components. Do not operate the digitizing oscilloscope in an atmosphere of explosive gases.
x
Safety

Product Description

Option 2F, Advanced DSP Math, is an option to the TDS Family Digitizing Oscilloscopes. It can be ordered only at the time the oscilloscope is purĆ chased.
TDS oscilloscopes use a proprietary Digital Signal Processor (DSP) to convert normally acquired waveforms into simple math waveforms (inverts a waveform, adds, subtracts, multiplies two waveforms). Option 2F adds three new complex math waveforms to the oscilloscope:
FFT (Fast Fourier Transform) MathĊtransforms a displayed waveform from the time domain to the frequency domain by applying the Fast Fourier Transform. FFT Math features the following capabilities:
H Displays the magnitude of the various frequencies the source waveform
contains or, optionally, the phase angle of those frequencies
H Measures magnitude in linear V
sures phase in degrees or radians
H Provides phase suppression to reduce the phase angle to zero for
frequencies with magnitudes below a userĆspecified threshold
H Transforms source waveforms with record lengths of 500, 1,000, 5,000,
15,000, 30,000, 50,000, and 60,000 points. (Your model oscilloscope will not have all of these these record lengths; consult your User manual to determine which are available with your TDS model.)
H Operates on source waveforms displayed in any channel or reference
memory (the source cannot be another math waveform)
H Provides the following FFT windows to optimize frequency resolution
and magnitude measurement accuracy: rectangular, Hamming, HanĆ ning, and BlackmanĆHarris
H Executes an FFT in as little as 64.2 milliseconds (1,000 point FFT)
H Updates the display up to ten times per second (1,000 point FFT)
H Corrects FFT DC error automatically
H Allows cursor and automatic measurements of FFT math waveforms
Derivative MathĊdifferentiates, with respect to time, a waveform displayed from a channel or from a reference memory. Derivative math waveforms
measure the instantaneous rate of change of a waveform.
or dB with respect to 1 V
RMS
RMS
; meaĆ
TDS Series Option 2F Instructions
Integral MathĊintegrates over time a waveform displayed from a channel or from a reference memory. Integral math waveforms display the total area under a waveform with respect to ground.
xi
Product Description
xii
Product Description
Fast Fourier Transforms

Fast Fourier Transforms

Using the Fast Fourier Transform (FFT), you can transform a waveform from a display of its amplitude against time to one that plots the amplitudes of the various discrete frequencies the waveform contains. Further, you can also display the phase shifts of those frequencies. Use FFT math waveforms in the following applications:
H Testing impulse response of filters and systems
H Measuring harmonic content and distortion in systems
H Characterizing the frequency content of DC power supplies
H Analyzing vibration
H Analyzing harmonics in 50 and 60 cycle lines
H Identifying noise sources in digital logic circuits

Description

The FFT computes and displays the frequency content of a waveform you acquire as an FFT math waveform. This frequency domain waveform is based on the following equation:
N
* 1
2
X(k) +
Where: x(n) is a point in the time domain record data array
The resulting waveform is a display of the magnitude or phase angle of the various frequencies the waveform contains with respect to those frequenĆ cies. For example, Figure 1Ć1 shows the nonĆtransformed impulse response of a system in channel 2 at the top of the screen. The FFTĆtransformed magĆ nitude and phase appear in the two math waveforms below the impulse. The horizontal scale for FFT math waveforms is always expressed in frequenĆ cy/division with the beginning (leftĆmost point) of the waveform representing zero frequency (DC).
1
N
Ă SĂ
n +
X(k) is a point in the frequency domain record data array
n is the index to the time domain data array
k is the index to the frequency domain data array
N is the FFT length
j is the square root of −1
x(n)e
* N
2
*
j2pnk
N
Ă
ĂĂĂĂĂ
Ă
for : k + 0Ă toĂ N * 1
TDS Series Option 2F Instructions
1Ć1
Fast Fourier Transforms
Normal Waveform of an
Impulse Response
FFT Waveform of the
Magnitude Response
The FFT waveform is based on digital signal processing (DSP) of data, which allows more versatility in measuring the frequency content of waveĆ forms. For example, DSP allows the oscilloscope to compute FFTs of source waveforms that must be acquired based on a single trigger, making it useful for measuring the frequency content of single events. (The TDS 800 model oscilloscopes must have repetitive triggers; therefore, they cannot compute FFTs of single events.) DSP also allows the phase as well as the magnitude to be displayed.

Operation

FFT Waveform of the
Phase Response
Figure 1Ć1:ăSystem Response to an Impulse
To obtain an FFT of your waveform, do these basic tasks:
H Acquire and display it normally (that is, in the time domain) in your
choice of input channels.
H Transform it to the frequency domain using the math waveform menu.
H Use cursors or automated measurements to measure its parameters.
Use the following procedure to perform these tasks.
1Ć2
Displaying an FFT
1. Connect the waveform to the desired channel input and select that channel.
Fast Fourier Transforms
Fast Fourier Transforms
2. Adjust the vertical and horizontal scales and trigger the display (or press AUTOSET).
The topic Offset, Position, and Scale, on page 1Ć9, provides in depth information about optimizing your setup for FFT displays.
3. Press MORE to access the menu for turning on math waveforms.
4. Select a math waveform. Your choices are Math1, Math2, and Math3 (main).
5. Press Change Math Definition (side) See Figure 1Ć2.
FFT (main).
TDS Series Option 2F Instructions
Figure 1Ć2:ăDefine FFT Waveform Menu
6. Press Set FFT Source to (side) repeatedly (or use the general purpose knob) until the channel source selected in step 1 appears in the menu label.
7. Press Set FFT Vert Scale to (side) repeatedly to choose from the followĆ ing vertical scale types:
H dBV RMSĊMagnitude is displayed using log scale, expressed in dB
relative to 1 V
where 0 dB =1 V
RMS
RMS
.
H Linear RMSĊMagnitude is displayed using voltage as the scale.
H Phase (deg)ĊPhase is displayed using degrees as the scale,
where degrees wrap from -180_ to +180_.
1Ć3
Fast Fourier Transforms
H Phase (rad)ĊPhase is displayed using radians as the scale, where
radians wrap from -p to +p.
The topic Considerations for Phase Displays, on page 1Ć12, provides in depth information on setup for phase displays.
8. Press Set FFT Window to (side) repeatedly to choose from the following window types:
H RectangularĊBest type for resolving frequencies that are very close
to the same value but worst for accurately measuring the amplitude of those frequencies. Best type for measuring the frequency specĆ trum of nonĆrepetitive signals and measuring frequency components near DC.
H HammingĊVery good window for resolving frequencies that are
very close to the same value with somewhat improved amplitude accuracy over the rectangular window.
H Hanning ĊVery good window for measuring amplitude accuracy
but degraded for resolving frequencies.
H BlackmanĆHarrisĊBest window for measuring the amplitude of
frequencies but worst at resolving frequencies.
The topic Selecting the Window, on page 1Ć14, provides in depth inĆ formation on choosing the right window for your application.
9. If you did not select Phase (deg) or Phase (rad) in step 7, skip to step
12. Phase suppression is only used to reduce noise in phase FFTs.
10. If you need to reduce the effect of noise in your phase FFT, press SupĆ press phase at amplitudes (side).
11. Use the general purpose knob (or the keypad if your oscilloscope is so equipped) to adjust the phase suppression level. FFT magnitudes below this level will have their phase set to zero.
The topic Adjust Phase Suppression, on page 1Ć13, provides additional information on phase suppression.
12. Press OK Create Math Wfm (side) to display the FFT of the waveform you input in step 1.
1Ć4
Fast Fourier Transforms
Fast Fourier Transforms
Figure 1Ć3:ăFFT Math Waveform in Math1
Cursor Measurements of an FFT
Once you have displayed an FFT math waveform, use cursors to measure its frequency amplitude or phase angle.
1. Be sure MORE is selected in the channel selection buttons and that the FFT math waveform is selected in the More main menu.
2. Press CURSOR tion (main)
3. Use the general purpose knob to align the selected cursor (solid line) to the top (or to any amplitude on the waveform you choose).
4. Press TOGGLE to select the other cursor. Use the general purpose knob to align the selected cursor to the bottom (or to any amplitude on the waveform you choose).
5. Read the amplitude between the two cursors from the D: readout. Read the amplitude of the selected cursor relative to either 1 V ground (0 volts), or the zero phase level the @: readout. (The waveform reference indicator at the left side of the graticule indicates the level where phase is zero for phase FFTs.)
Mode (main) Independent (side) FuncĆ
H Bars (side).
(0 degrees or 0 radians) from
RMS
(0 dB),
TDS Series Option 2F Instructions
Figure 1Ć4 shows the cursor measurement of a frequency magnitude on an FFT. The @: readout reads 0 dB because it is aligned with the 1 V
RMS
level. The D: readout reads 24.4 dB indicating the magnitude of the frequency it is measuring is -24.4 dB relative to 1 V
. The source
RMS
waveform is turned off in the display.
1Ć5
Fast Fourier Transforms
The cursor units will be in dB or volts for FFTs measuring magnitude and in degrees or radians for those FFTs measuring phase. The cursor unit
depends on the selection made for Set FFT Vert Scale to (side).
See
step 7 on page 1Ć3 for more information.
6. Press V Bars (side). Use the general purpose knob to align one of the two vertical cursors to a point of interest along the horizontal axis of the waveform.
7. Press TOGGLE to select the alternate cursor.
8. Align the alternate cursor to another point of interest on the math waveĆ form.
9. Read the frequency difference between the cursors from the D: readout. Read the frequency of the selected cursor relative to the zero frequency point from the @: readout.
The cursor units will always be in Hz, regardless of the setting in the Time Units side menu. The first point of the FFT record is the zero frequency point for the @: readout.
1Ć6
Figure 1Ć4:ăCursor Measurement of an FFT Waveform
10. Press Function (main)
Paired (side).
11. Use the technique just outlined to place the vertical bar of each paired cursor to the points along the horizontal axis you are interested in.
Fast Fourier Transforms
Fast Fourier Transforms
12. Read the amplitude between the short horizontal bar of the two paired cursors from the topĆmost D: readout. Read the amplitude of the short horizontal bar of the selected (solid) cursor relative to either 1 V (0 dB), ground (0 volts), or zero phase level (0 degrees or 0 radians) from the @: readout. Read the frequency between the long horizontal bars of both paired cursors from the bottom D: readout.
RMS
Automated Measurements of an FFT
You can also use automated measurements to measure FFT math waveĆ forms. Use the same procedure as is found under Waveform Differentiation on page 2Ć2.

Considerations for Using FFTs

There are several characteristics of FFTs that affect how they are displayed and should be interpreted. Read the following topics to learn how to optiĆ mize the oscilloscope setup for good display of your FFT waveforms.
The FFT Frequency Domain Record
The following topics discuss the relation of the source waveform to the record length, frequency resolution, and frequency range of the FFT freĆ quency domain record. (The FFT frequency domain waveform is the FFT math waveform that you display.)
FFTs May Not Use All of the Waveform RecordĊThe FFT math waveĆ
form is a display of the magnitude or phase data from the FFT frequency domain record. This frequency domain record is derived from the FFT time
domain record, which is derived from the waveform record. All three records are described below.
Waveform RecordĊthe complete waveform record acquired from an input channel and displayed from the same channel or a reference memory. The length of this time domain record is userĆspecified from the Horizontal menu. The waveform record is not a DSP Math waveform.
FFT Time Domain RecordĊthat part of the waveform record that is input to the FFT. This time domain record waveform becomes the FFT math waveĆ form after it is transformed. Its record length depends on the length of the waveform record defined above.
TDS Series Option 2F Instructions
FFT Frequency Domain RecordĊthe FFT math waveform after digital signal processing converts data from the FFT time domain record into a frequency domain record.
Figure 1Ć5 compares the waveform record to the FFT time domain record. Note the following relationships:
H For waveform records ≤10 K points in length, the FFT uses all of the
waveform record as input.
1Ć7
Fast Fourier Transforms
H For waveform records >10 K points, the first 10 K points of the waveform
record becomes the FFT time domain record.
H Each FFT time domain record starts at the beginning of the acquired
waveform record.
H The zero phase reference point for a phase FFT math waveform is in the
middle of the FFT time domain record regardless of the waveform record length.
FFT Time Domain Record =
Waveform Record
Waveform Record 10 K
Zero Phase
Reference
FFT Time Domain Record = 10k
Waveform Record > 10 K
Zero Phase
Reference
Figure 1Ć5:ăWaveform Record vs. FFT Time Domain Record
FFTs Transform Time Records to Frequency RecordsĊThe FFT time
domain record just described is input for the FFT. Figure 1Ć6 shows the transformation of that time domain data record into an FFT frequency doĆ main record. The resulting frequency domain record is one half the length of the FFT input because the FFT computes both positive and negative freĆ quencies. Since the negative values mirror the positive values, only the positive values are displayed.
FFT Time Domain Record
FFT
1Ć8
FFT Frequency Domain Record
Figure 1Ć6:ăFFT Time Domain Record vs. FFT Frequency Domain
Record
Fast Fourier Transforms
Fast Fourier Transforms
FFT Frequency Range and ResolutionĊWhen you turn on an FFT
waveform, the oscilloscope displays either the magnitude or phase angle of the FFT frequency domain record. The resolution between the discrete frequencies displayed in this waveform is determined by the following equaĆ tion:
SampleĂ Rate
DF +
Where: DF is the frequency resolution.
Sample Rate is the sample rate of the source waveform.
FFT Length is the length of the FFT Time Domain waveform record.
The sample rate also determines the range these frequencies span; they span from 0 to ½ the sample rate of the waveform record. (The value of ½ the sample rate is often referred to as the Nyquist frequency or point.) For example, a sample rate of 20 Megasamples/second would yield an FFT with a range of 0 to 10 MHz. The sample rates available for acquiring data reĆ cords vary over a range the limits of which depend on your oscilloscope model. TDS oscilloscopes (except for TDS 800 models) display the sample rate in the acquisition readout at the top of the oscilloscope screen.
FFTĂ Length
Offset, Position, and Scale
The following topics contain information to help you display your FFT propĆ erly.
Adjust for a NonĆClipped DisplayĊTo properly display your FFT waveform, scale the source waveform so it is not clipped.
H You should scale and position the source waveform so it is contained on
screen. (Off screen waveforms may be clipped," which will result in errors in the FFT waveform).
Alternately, to get maximum vertical resolution, you can display source waveforms with amplitudes up to two divisions greater than that of the screen. If you do, turn on PkĆPk in the measurement menu and monitor the source waveform for clipping.
H Use vertical position and vertical offset to position your source waveĆ
form. As long as the source waveform is not clipped, its vertical position and vertical offset will not affect your FFT waveform except at DC. (DC correction is discussed below.)
Adjust Offset and Position to Zero for DC CorrectionĊNormally, the output of a standard FFT computation yields a DC value that is twice as large as it should be with respect to the other frequencies. Also, the selecĆ
tion of window type introduces errors in the DC value of an FFT.
TDS Series Option 2F Instructions
1Ć9
Fast Fourier Transforms
The displayed output of the FFT on TDS oscilloscopes is corrected for these errors to show the true value for the DC component of the input signal. The Position and Offset must be set to zero for the source waveform in the Vertical menu. When measuring the amplitude at DC, remember that 1 VDC equals 1 V
and the display is in dB.
RMS
Record Length
Most often, you will want to use a short record length because more of the FFT waveform can be seen on screen and long record lengths can slow oscilloscope response. However, long record lengths lower the noise relative to the signal and increase the frequency resolution for the FFT. More imporĆ tant, they might be needed to capture the waveform feature you want to include in the FFT.
To speed up oscilloscope response when using long record lengths, you can save your source waveform in a reference memory and perform an FFT on the saved waveform. That way the DSP will compute the FFT based on saved, static data and will only update if you save a new waveform.
Acquisition Mode
Selecting the right acquisition mode can produce less noisy FFTs.
Set up in Sample or Normal ModeĊUse sample mode (all models except TDS 800 Oscilloscopes) until you have set up and turned on your FFT. Sample mode can acquire repetitive and nonrepetitive waveforms and does not affect the frequency response of the source waveform.
Use normal mode if your oscilloscope is a TDS 800 Oscilloscope (sample mode is not available). Like sample mode for other models, normal mode does not affect the frequency response; however, unlike sample mode, it can only acquire repetitive waveforms.
Hi Res and Average Reduce NoiseĊAfter the FFT is set up and displayed, it might be useful to turn on Hi Res mode, on TDS models so equipped, to reduce the effect of noise in the signal. Hi Res operates on both repetitive and nonrepetitive waveforms; however, it does affect the frequency reĆ sponse of the source waveform.
If the pulse is repetitive, then Average mode may be used to reduce noise in the signal at a cost of slower display response. Average operates on repetiĆ tive waveforms only, and averaging does effect the frequency response of the source waveform.
1Ć10
Peak Detect (on TDS models so equipped) and Envelope mode can add significant distortion to the FFT results and are not recommended for use with FFTs.
Fast Fourier Transforms
Fast Fourier Transforms
Zoom and Interpolation
Once you have your waveform displayed optimally, you may magnify (or reduce) it vertically and horizontally to inspect any feature you desire. Just be sure the FFT waveform is the selected waveform. (Press MORE, then select the FFT waveform in the More main menu. Then use the Vertical and Horizontal SCALE knobs to adjust the math waveform size.)
If you wish to see the zoom factor (2X, 5X, etc.) you need to turn Zoom on: press ZOOM on screen.
Whether Zoom is on or off, you can press Reset Zoom Factors (side) to return the zoomed FFT waveform to no magnification.
Zoom always uses either sin(x)/x or linear interpolation when expanding displayed waveforms. To select the interpolation method: press DISPLAY Filter (main) Sin(x)/x or Linear (side).
If the source waveform record length is 500 points, the FFT will use 2X Zoom to increase the 250 point FFT frequency domain record to 500 points. ThereĆ fore, FFT math waveforms of 500 point waveforms are always zoomed 2X or more with interpolation. Waveforms with other record lengths can be zoomed or not and can have minimum Zooms of 1X or less.
On (side). The vertical and horizontal zoom factors appear
Sin(x)/x interpolation may distort the magnitude and phase displays of the FFT depending on which window was used. You can easily check the efĆ fects of the interpolation by switching between sin(x)/x and linear interpolaĆ tion and observing the difference in measurement results on the display. If significant differences occur, use linear interpolation.
Undersampling (Aliasing)
Aliasing occurs when the oscilloscope acquires a source waveform with frequency components outside of the frequency range for the current samĆ ple rate. In the FFT waveform, the actual higher frequency components are undersampled, and therefore, they appear as lower frequency aliases that fold back" around the Nyquist point (see Figure 1Ć7).
The greatest frequency that can be input into any sampler without aliasing is ½ the sample frequency. Since source waveforms often have a fundamental frequency that does not alias but have harmonic frequencies that do, you should have methods for recognizing and dealing with aliases:
H Be aware that a source waveform with fast edge transition times creates
many high frequency harmonics. These harmonics typically decrease in amplitude as their frequency increases.
H Sample the source signal at rates that are at least 2X that of the highest
frequency component having significant amplitude.
TDS Series Option 2F Instructions
H Filter the input to bandwidth limit it to frequencies below that of the
Nyquist frequency.
1Ć11
Fast Fourier Transforms
H Recognize and ignore the aliased frequencies.
If you think you have aliased frequencies in your FFT, select the source channel and adjust the horizontal scale to increase the sample rate. Since you increase the Nyquist frequency as you increase the sample rate, the alias signals should unfold" and appear at their proper frequency.
Nyquist Frequency
Point
Amplitude
Frequency
Aliased Frequencies Actual Frequencies
Figure 1Ć7:ăHow Aliased Frequencies Appear in an FFT
Considerations for Phase Displays
When you setup an FFT math waveform to display the phase angle of the frequencies contained in a waveform, you should take into account the reference point the phase is measured against. You may also need to use phase suppression to reduce noise in your FFTs.
Establish a Zero Phase Reference PointĊThe phase of each frequenĆ
cy is measured with respect to the zero phase reference point. The zero reference point is the point at the center of the FFT math waveform but corresponds to various points on the source (time domain) record. (See Figure 1Ć5 on page 1Ć8.)
To measure the phase relative to most source waveforms, you need only to center the positive peak around the zero phase point. (For instance, center the positive half cycle for a sine or square wave around the zero phase point.) Use the following method:
H First be sure the FFT math waveform is selected in the More menu, then
set horizontal position to 50% in the Horizontal menu. This positions the zero phase reference point to the horizontal center of the screen.
1Ć12
Fast Fourier Transforms
Fast Fourier Transforms
H In the Horizontal menu, vary the trigger position (time base position for
TDS 800 model oscilloscopes) to center the positive peak of the source waveform at the horizontal center of screen. Alternately, you can adjust the trigger level (knob) to bring the positive peak to center screen if the phase reference waveform has slow enough edges.
When impulse testing and measuring phase, align the impulse input into the system to the zero reference point of the FFT time domain waveform:
H Set the trigger position (time base position for TDS 800 model oscilloĆ
scopes) to 50% and horizontal position to 50% for all record lengths less than 15 K. (Your model oscilloscope may not have record lengths of 15 K or longerĊconsult your User manual.)
H For records with a 15 K length, set the trigger position to 33% for all
models except TDS 800 oscilloscopes. Use the horizontal position knob to move the trigger T on screen to the center horizontal graticule line.
With TDS 800 oscilloscopes only, first set the record length to 5000 points in 1000 divisions. Then turn the horizontal position knob full counter clockwise, set the record length to 15000 points in 300 diviĆ sions, and adjust the time base position to move the impulse to the center horizontal graticule line.
H For records with 30 K, 50 K, or 60 K lengths (not all lengths are available
for all TDS modelsĊconsult your User manual), set the trigger position to 16.6%,10%, or 8.3%, respectively. Use the horizontal position knob to move the trigger T on screen and to the center horizontal graticule line.
H Trigger on the input impulse.
Adjust Phase SuppressionĊYour source waveform record may have a
noise component with phase angles that randomly vary from pi to pi. This noise could make the phase display unusable. In such a case, use phase suppression to control the noise.
You specify the phase suppression level in dB with respect to 1 V magnitude of the frequency is greater than this threshold, then its phase angle will be displayed. However, if it is less than this threshold, then the phase angle will be set to zero and be displayed as zero degrees or radians. (The waveform reference indicator at the left side of the graticule indicates the level where phase is zero for phase FFTs.)
It is easier to determine the level of phase suppression you need if you first create a frequency FFT math waveform of the source and then create a phase FFT waveform of the same source. Do the following steps to use a cursor measurement to determine the suppression level:
1. Do steps 1 through 7 of Display an FFT that begins on page 1Ć2. Select
dBV
RMS (side) for the Set FFT Vert Scale to (side).
RMS
. If the
TDS Series Option 2F Instructions
1Ć13
Fast Fourier Transforms
2. Press CURSOR Mode (main) Independent (side) FuncĆ tion (main)
selected cursor to a level that places the tops of the magnitudes of frequencies of interest above the cursor but places other magnitudes completely below the cursor.
3. Read the level in dB from the @: readout. Note the level for use in step
5.
H Bars (side). Use the general purpose knob to align the
4. Press MORE (main)
Press Set FFT Vert Scale to (side) repeatedly to choose either Phase (rad) or Phase (deg).
5. Press Suppress Phase at Amplitudes (side). Use the general purpose knob (or keypad if your oscilloscope is so equipped) to set phase supĆ pression to the value obtained using the H Bar cursor. Do not change the window selection or you will invalidate the results obtained using the cursor.
Change Waveform Definition menu (side).
FFT Windows
To learn how to optimize your display of FFT data, read about how the FFT windows data before computing the FFT math waveform. Understanding FFT windowing can help you get more useful displays.
Windowing ProcessĊThe oscilloscope multiplies the FFT time domain
record by one of four FFT windows it provides before it inputs the record to the FFT function. Figure 1Ć8 shows how the time domain record is proĆ cessed.
The FFT windowing acts like a bandpass filter between the FFT time domain record and the FFT frequency domain record. The shape of the window controls the ability of the FFT to resolve (separate) the frequencies and to accurately measure the amplitude of those frequencies.
1Ć14
Selecting a WindowĊYou can select your window to provide better
frequency resolution at the expense of better amplitude measurement accuracy in your FFT, better amplitude accuracy over frequency resolution, or to provide a compromise between both. You can choose from these four
windows: Rectangular, Hamming, Hanning, and BlackmanĆHarris.
In step 8 (page 1Ć4) in Displaying an FFT, the four windows are listed in order according to their ability to resolve frequencies versus their ability to accurately measure the amplitude of those frequencies. The list indicates that the ability of a given window to resolve a frequency is inversely proporĆ tional to its ability to accurately measure the amplitude of that frequency. In general, then, choose a window that can just resolve between the frequenĆ cies you want to measure. That way, you will have the best amplitude accuĆ racy and leakage elimination while still separating the frequencies.
Fast Fourier Transforms
FFT Time Domain Record
FFT Window
Fast Fourier Transforms
X's
FFT Time Domain Record
FFT Frequency Domain Record
After Windowing
FFT
Figure 1Ć8:ăWindowing the FFT Time Domain Record
You can often determine the best window empirically by first using the window with the most frequency resolution (rectangular), then proceeding toward that window with the least (BlackmanĆHarris) until the frequencies merge. Use the window just before the window that lets the frequencies merge for best compromise between resolution and amplitude accuracy.
NOTE
TDS Series Option 2F Instructions
If the Hanning window merges the frequencies, try the Hamming window before settling on the rectangular window. Depending on the distance of the frequencies you are trying to measure from the fundamental, the Hamming window sometimes resolves frequenĆ cies better than the Hanning.
1Ć15
Fast Fourier Transforms
Window CharacteristicsĊWhen evaluating a window for use, you may
want to examine how it modifies the FFT time domain data. Figure 1Ć9 shows each window, its bandpass characteristic, bandwidth, and highest side lobe. Consider the following characteristics:
H The narrower the central lobe for a given window, the better it can
resolve a frequency.
H The lower the lobes on the side of each central lobe are, the better the
amplitude accuracy of the frequency measured in the FFT using that window.
H Narrow lobes increase frequency resolution because they are more
selective. Lower side lobe amplitudes increases accuracy because they reduce leakage.
Leakage results when the FFT time domain waveform delivered to the FFT function contains a nonĆinteger number of waveform cycles. Since there are fractions of cycles in such records, there are discontinuities at the ends of the record. These discontinuities cause energy from each discrete frequency to leak" over on to adjacent frequencies. The result is amplitude error when measuring those frequencies.
The rectangular window does not modify the waveform record points; it generally gives the best frequency resolution because it results in the most narrow lobe width in the FFT output record. If the time domain records you measured always had an integer number of cycles, you would only need this window.
Hamming, Hanning, and BlackmanĆHarris are all somewhat bellĆshaped widows that taper the waveform record at the record ends. The Hanning and Blackman/Harris windows taper the data at the end of the record to zero; therefore, they are generally better choices to eliminate leakage.
1Ć16
Fast Fourier Transforms
Fast Fourier Transforms
FFT Window Type Bandpass Filter
Rectangular Window
Hamming Window
Hanning Window
BlackmanĆHarris
Window
0 dB
-20
-40
-50
0 dB
-20
-40
-60
0 dB
-20
-40
-60
-80
0 dB
-20
-40
-60
-80
-100
-101
Ć3 dB Bandwidth Highest Side Lobe
0.89
1.28
1.28
1.28
Ć13 dB
-43 dB
-32 dB
-94 dB
Figure 1Ć9:ăFFT Windows and Bandpass Characteristics
Care should be taken when using bell shaped widows to be sure that the most interesting parts of the signal in the time domain record are positioned in the center region of the window so that the tapering does not cause severe errors.
TDS Series Option 2F Instructions
1Ć17
Fast Fourier Transforms
1Ć18
Fast Fourier Transforms
Waveform Differentiation

Waveform Differentiation

The Advanced DSP Math option provides waveform differentiation that allows you to display a derivative math waveform that indicates the instantaĆ neous rate of change of the waveform acquired. Such waveforms are used in the measurement of slew rate of amplifiers and in educational applicaĆ tions. You can store and display a derivative math waveform in a reference memory, then use it as a source for another derivative waveform. The result is the second derivative of the waveform that was first differentiated.

Description

Operation

The math waveform, derived from the sampled waveform, is computed based on the following equation:
Y
+ (X
n
(n)1)
Where: X is the source waveform
Y is the derivative math waveform
T is the time between samples
Since the resultant math waveform is a derivative waveform, its vertical scale is in volts/second (its horizontal scale is in seconds). The source signal is differentiated over its entire record length; therefore, the math waveform record length equals that of the source waveform.
To obtain a derivative math waveform, do the following tasks:
H Acquire and display the waveform in your choice of input channels.
H Differentiate it using the math waveform menu.
* Xn)
1 T
TDS Series Option 2F Instructions
H Use cursors or automated measurements to measure its parameters.
Use the following procedure to perform these tasks.
Displaying a Differentiated Waveform
1. Connect the waveform to the desired channel input and select that channel.
2. Adjust the vertical and horizontal scales and trigger the display (or press AUTOSET).
2Ć1
Waveform Differentation
Derivative Math
Waveform
Source
Waveform
3. Press MORE Math1, Math2, or Math3 (main) Change Math Definition (side)
Single Wfm Math (main). See Figure 1Ć2.
Figure 2Ć1:ăDerivative Math Waveform
4. Press Set Single Source to (side). Repeatedly press the same button (or use the general purpose knob) until the channel source selected in step 1 appears in the menu label.
5. Press Set Function to (side). Repeatedly press the same button (or use the general purpose knob) until diff appears in the menu label.
6. Press OK Create Math Wfm (side) to display the derivative of the waveĆ form you input in step 1.
You should now have your derivative math waveform on screen. Use the Vertical SCALE and POSITION knobs to size and position your waveĆ form as you require.
Automated Measurements of a Derivative Waveform
Once you have displayed your derivative math waveform, you can use automated measurements to make various parameter measurements. Do the following steps to display automated measurements of the waveform:
1. Be sure MORE is selected in the channel selection buttons and that the differentiated math waveform is selected in the More main menu.
2Ć2
2. Press MEASURE
Select Measrmnt (main).
Waveform Differentation
Waveform Differentation
3. Select up to four measurements in the side menu. (Push More (side) to see more measurement choices.) See your User Manual for descriptions of each measurement.
You can press Remove Measrmnts (main) to display a side menu for removing measurements from the display.

Usage Considerations

Figure 2Ć2:ăPeakĆPeak Amplitude Measurement of a Derivative
Waveform
You can also get a snapshot of all single waveform measurements at once:
4. Press Snapshot (main). Most of the measurements found in the MeaĆ sure side menus are now in a single display.
Cursor Measurement of a Derivative Waveform
You can also use cursors to measure derivative waveforms. Use the same procedure as is found under Waveform Integration on page 3Ć2. When using that procedure, note that the amplitude measurements on a derivative waveform will be in volts/second rather than in voltĆseconds as is indicated for the integral waveform measured in the procedure.
When creating differentiated math waveforms from live channel waveforms, consider the following topics.
TDS Series Option 2F Instructions
2Ć3
Waveform Differentation
Offset, Position, and Scale
Note the following tips for obtaining a good display:
H You should scale and position the source waveform so it is contained on
screen. (Off screen waveforms may be clipped," which will result in errors in the derivative waveform).
H You can use vertical position and vertical offset to position your source
waveform. The vertical position and vertical offset will not affect your derivative waveform unless you position the source waveform off screen so it is clipped.
H When using the vertical scale knob to scale the source waveform, note
that it also scales your derivative waveform.
Because of the method the oscilloscope uses to scale the source waveform before differentiating that waveform, the derivative math waveform may be too large vertically to fit on screenĊeven if the source waveforms is only a few divisions on screen. You can use Zoom to reduce the size of the waveĆ form on screen (see Zoom that follows), but If your waveform is clipped before zooming, it will still be clipped after it is zoomed.
If your math waveform is a narrow differentiated pulse, it may not appear to be clipped when viewed on screen. You can detect if your derivative math waveform is clipped by expanding it horizontally using Zoom so you can see the clipped portion. Also, the automated measurement PkĆPk will display a clipping error message if turned on (see Automated Measurements of a Derivative Waveform on page 2Ć2).
If your derivative waveform is clipped, try either of the following methods to eliminate clipping:
H Reduce the size of the source waveform on screen. (Select the source
channel and use the vertical SCALE knob.)
H Expand the waveform horizontally on screen. (Select the source channel
and increase the horizontal scale using the horizontal SCALE knob.) For instance, if you display the source waveform illustrated in Figure 2Ć1 on page 2Ć2 so its rising and falling edges are displayed over more horiĆ zontal divisions, the amplitude of the corresponding derivative pulse will decrease.
Whichever method you use, be sure Zoom is off and the zoom factors are reset (see Zoom below).
Zoom
2Ć4
Once you have your waveform optimally displayed, you can also magnify (or contract) it vertically and horizontally to inspect any feature. Just be sure the differentiated waveform is the selected waveform. (Press MORE, then select the differentiated waveform in the More main menu. Then use the Vertical and Horizontal SCALE knob to adjust the math waveform size.)
Waveform Differentation
Waveform Differentation
If you wish to see the zoom factor (2X, 5X, etc.), you need to turn zoom on: press ZOOM on screen.
Whether zoom is on or off, you can press Reset Zoom Factors (side) to return the zoomed derivative waveform to no magnification.
ON (side). The vertical and horizontal zoom factors appear
TDS Series Option 2F Instructions
2Ć5
Waveform Differentation
2Ć6
Waveform Differentation
Waveform Integration

Waveform Integration

The Advanced DSP Math option provides waveform integration that allows you to display an integral math waveform that is an integrated version of the acquired waveform. Such waveforms find use in the following applications:
H Measuring of power and energy, such as in switching power supplies
H Characterizing mechanical transducers, as when integrating the output
of an accelerometer to obtain velocity

Description

Operation
The integral math waveform, derived from the sampled waveform, is comĆ puted based on the following equation:
n
y(n) + scale
Where: x(i) is the source waveform
y(n) is a point in the integral math waveform
scale is the output scale factor
T is the time between samples
Since the resultant math waveform is an integral waveform, its vertical scale is in voltĆseconds (its horizontal scale is in seconds). The source signal is integrated over its entire record length; therefore, the math waveform record length equals that of the source waveform.
To obtain an integral math waveform, you first acquire it in your choice of input channels. Then you integrate it by creating an integral math waveform from the math waveform menu. Once the math waveform is displayed, you can use cursors or automated measurements to measure its parameters.
x(i) ) x(i * 1)
S
i + 1
2
T
TDS Series Option 2F Instructions
Displaying an Integral Waveform
Do the following steps to integrate and display your waveform.
1. Connect the waveform to the desired channel input and select that channel.
2. Adjust the vertical and horizontal scales and trigger the display (or press AUTOSET).
3Ć1
Waveform Integration
3. Press MORE Math1, Math2, or Math3 (main) Change Math waveform definition (side)
Single Wfm Math (main).
4. Press Set Single Source to (side). Repeatedly press the same button (or use the general purpose knob) until the channel source selected in step 1 appears in the menu label.
5. Press Set Function to (side). Repeatedly press the same button (or use the general purpose knob) until intg appears in the menu label.
6. Press OK Create Math Waveform (side) to turn on the integral math waveform.
You should now have your integral math waveform on screen. See Figure 3Ć1. Use the Vertical SCALE and POSITION knobs to size and position your waveform as you require.
Integral Math
Waveform
Source
Waveform
Figure 3Ć1:ăIntegral Math Waveform
Cursor Measurements of an Integral Waveform
Once you have displayed your integrated math waveform, use cursors to measure its voltage over time.
3Ć2
1. Be sure MORE is selected (lit) in the channel selection buttons and that the integrated math waveform is selected in the More main menu.
2. Press CURSOR tion (main)
Mode (main) Independent (side) FuncĆ
H Bars (side).
Waveform Integration
Integral Math
Waveform
Waveform Integration
3. Use the general purpose knob to align the selected cursor (solid) to the top (or to any amplitude level you choose).
4. Press TOGGLE to select the other cursor.
5. Use the general purpose knob to align the selected cursor (to the botĆ tom (or to any amplitude level you choose).
6. Read the integrated voltage over time between the cursors in voltĆ seconds from the D: readout. Read the integrated voltage over time between the selected cursor and the reference indicator of the math waveform from the @: readout. See Figure 3Ć2.
Source
Waveform
Figure 3Ć2:ăH Bars Cursors Measure an Integral Math Waveform
7. Press Function (main)
V Bars (side).Use the general purpose knob
to align one of the two vertical cursors to a point of interest along the horizontal axis of the waveform.
8. Press TOGGLE to select the alternate cursor.
9. Align the alternate cursor to another point of interest on the math waveĆ form.
TDS Series Option 2F Instructions
10. Read the time difference between the cursors from the D: readout. Read the time difference between the selected cursor and the trigger point for the source waveform from the @: readout.
11. Press Function (main)
Paired (side).
3Ć3
Waveform Integration
12. Use the technique just outlined to place the long vertical bar of each paired cursor to the points along the horizontal axis you are interested in.
13. Read the following values from the cursor readouts:
H Read the integrated voltage over time between the short horizontal
bars of both paired cursors in voltĆseconds from the D: readout.
H Read the integrated voltage over time between the short horizontal
bar of the selected cursor and the reference indicator of the math waveform from the @: readout.
H Read the time difference between the long vertical bars of the paired
cursors from the D: readout.
Automated Measurements of a Integral Waveform
You can also use automated measurements to measure integral math waveĆ forms. Use the same procedure as is found under Waveform Differentiation on page 2Ć2. When using that procedure, note that your measurements on an integral waveform will be in voltĆseconds rather than in volts/second as is indicated for the differential waveform measured in the procedure.

Usage Considerations

When creating integrated math waveforms from live channel waveforms, consider the following topics.
Offset, Position, and Scale
Note the following requirements for obtaining a good display:
H You should scale and position the source waveform so it is contained on
screen. (Off screen waveforms may be clipped," which will result in errors in the integral waveform).
H You can use vertical position and vertical offset to position your source
waveform. The vertical position and vertical offset will not affect your integral waveform unless you position the source waveform off screen so it is clipped.
H When using the vertical scale knob to scale the source waveform, note
that it also scales your integral waveform.
DC Offset
The source waveforms that you connect to the oscilloscope often have a DC offset component. The oscilloscope integrates this offset along with the time varying portions of your waveform. Even a few divisions of offset in the source waveform may be enough to ensure that the integral waveform saturates (clips), especially with long record lengths.
3Ć4
Waveform Integration
Waveform Integration
You may be able to avoid saturating your integral waveform if you choose a shorter record length. (Press HORIZONTAL MENU Length (main).) Reducing the sample rate (use the HORIZONTAL SCALE
knob) with the source channel selected might also prevent clipping. You can also select AC coupling (on TDS models so equipped) in the vertical menu of the source waveform or otherwise DC filter it before applying it to the oscilloscope input.
Record
Zoom
Once you have your waveform optimally displayed, you may magnify (or reduce) it vertically and horizontally to inspect any feature you desire. Just be sure the integrated waveform is the selected waveform. (Press MORE, then select the integrated waveform in the More main menu. Then use the Vertical and Horizontal SCALE knobs to adjust the math waveform size.)
If you wish to see the zoom factor (2X, 5X, etc.) you need to turn Zoom on: press ZOOM on screen.
Whether Zoom is on or off, you can press Reset Zoom Factors (side) to return the zoomed integral waveform to no magnification.
On (side). The vertical and horizontal zoom factors appear
TDS Series Option 2F Instructions
3Ć5
Waveform Integration
3Ć6
Waveform Integration
Index

Index

A
Aliasing, 1Ć11
Applications
derivative math waveforms, 2Ć1 FFT math waveforms, 1Ć1 integral math waveforms, 3Ć1
Automated measurements
of derivative math waveforms, 2Ć2
(procedure), 2Ć2 of FFT math waveforms, 1Ć7 of integral math waveforms, 3Ć4
B
BlackmanĆHarris window, 1Ć4
C
CAUTION
statement in manuals, ix statement on equipment, ix
Clipping
derivative math waveforms, 2Ć4 FFT math waveforms, 1Ć9 how to avoid, 1Ć9, 2Ć4, 3Ć4 integral math waveforms, 3Ć4
Conventions, v
Cursor menu, 1Ć5, 3Ć2
Cursor readout
HĆBars, 1Ć5, 2Ć3, 3Ć3 Paired, 2Ć3 Paired cursors, 1Ć7, 3Ć4 VĆBars, 1Ć6, 2Ć3, 3Ć3
Cursors
with derivative waveforms, 2Ć3 with FFT waveforms, 1Ć5 with integral waveforms, 3Ć2
D
DANGER, statement on equipment, ix
DC offset, 1Ć9
for DC correction of FFTs, 1Ć9 with math waveforms, 1Ć9, 3Ć4
Derivative math, description, xi
Derivative math waveform, 2Ć1
applications, 2Ć1 derivation of, 2Ć1 procedure for displaying, 2Ć1 procedure for measuring, 2Ć2, 2Ć3 record length of, 2Ć1
Description
derivative math, xi FFT math, xi integral math, xi product, xi
Differentiation
of a derivative, 2Ć1 waveform, 2Ć1
F
Fast Fourier Transforms, description, 1Ć1
Fast Fourier Transforms (FFTs), applicaĆ
tions, 1Ć1
FFT frequency domain record, 1Ć7
defined, 1Ć7ć1Ć9 length of, 1Ć8
FFT math, description, xi
FFT math waveform, 1Ć1
acquisition mode, 1Ć10 aliasing, 1Ć11 automated measurements of, 1Ć7 DC correction, 1Ć9 derivation of, 1Ć1 displaying phase, 1Ć3 frequency range, 1Ć9 frequency resolution, 1Ć9
interpolation mode, 1Ć11 list of features, xi magnifying, 1Ć11 phase display, setup considerations,
1Ć12ć1Ć14
phase suppression, 1Ć4, 1Ć13 procedure for displaying, 1Ć2 procedure for measuring, 1Ć5 record length, 1Ć8 reducing noise, 1Ć10 undersampling, 1Ć11 zero phase reference, 1Ć12
FFT time domain record, defined, 1Ć7
H
Hamming window, 1Ć4
Hanning Window, 1Ć4
I
Integral math, description, xi
Integral math waveform, 3Ć1
applications, 3Ć1 automated measurements of, 3Ć4 derivation of, 3Ć1 magnifying, 2Ć4, 3Ć5 procedure for displaying, 3Ć1 procedure for measuring, 3Ć2 record length of, 3Ć1
Integration, Waveform, 3Ć1
Interpolation
FFT distortion, 1Ć11 linear versus sin(x)/x, 1Ć11
TDS Series Option 2F Instructions
I-1
Index
M
Math waveform
derivative. See Derivative math waveĆ
form
FFT. See FFT math waveform integral. See Integral math waveform
Measurements, snapshot of, 2Ć3
Menu, More, 1Ć3, 2Ć2
See also More menu
More menu, 2Ć2
BlackmanĆHarris, 1Ć4 Change Math waveform definition,
1Ć3, 2Ć2, 3Ć2
dBV RMS, 1Ć3 diff, 2Ć2 FFT, 1Ć3 Hamming, 1Ć4 Hanning, 1Ć4 intg, 3Ć2 Linear RMS, 1Ć3 Math1, Math2, Math3, 1Ć3, 2Ć2, 3Ć2 OK Create Math Waveform, 3Ć2 Phase (deg), 1Ć3 Phase (rad), 1Ć4 Rectangular, 1Ć4 Set FFT Source to:, 1Ć3 Set FFT Vert Scale to:, 1Ć3 Set FFT Window to:, 1Ć4 Set Function to:, 2Ć2, 3Ć2 Set Single Source to:, 2Ć2, 3Ć2 Single Wfm Math, 2Ć2, 3Ć2
P
Phase suppression, 1Ć13
Position, vertical, 1Ć9, 2Ć4, 3Ć4
Product description, xi
R
Readout, Cursor, Paired, 2Ć3
Readout, cursor
HĆBars, 1Ć5, 2Ć3, 3Ć3 Paired cursors, 1Ć7, 3Ć4 VĆBars, 1Ć6, 2Ć3, 3Ć3
Record length
derivative math waveforms, 2Ć1 integral math waveforms, 3Ć1
Rectangular window, 1Ć4
S
Safety, ix
symbols, ix
Scale, vertical, 1Ć9, 2Ć4, 3Ć4
Snapshot, 2Ć3
W
WARNING, statement in manual, ix
Waveform clipping. See Clipping
Waveform differentiation, 2Ć1
Waveform FFTs, 1Ć1
Waveform integration, 3Ć1
Waveform record
FFT, 1Ć7 FFT frequency domain, 1Ć7
length of, 1Ć7
FFT source, 1Ć7
acquisition mode, 1Ć10 defined, 1Ć7 long versus short, 1Ć10
FFT time domain, 1Ć7ć1Ć9
Window, 1Ć14
BlackmanĆHarris, 1Ć4, 1Ć14, 1Ć17 characteristics of, 1Ć16 Hamming, 1Ć4, 1Ć14, 1Ć17 Hanning, 1Ć4, 1Ć14, 1Ć17 rectangular, 1Ć4, 1Ć14, 1Ć17 rectangular vs. bellĆshaped, 1Ć16 selecting, 1Ć14
Windowing, process, 1Ć14
Windows, descriptions of, 1Ć4
Z
N
Noise
reducing in FFTs, 1Ć10 reducing in phase FFTs, 1Ć4, 1Ć13
Nyquist frequency, 1Ć11
O
Offset
DC. See DC Offset vertical, 1Ć9, 2Ć4, 3Ć4
I-2
V
Vertical position, for DC correction of
FFTs, 1Ć9
Zero phase reference point, 1Ć8, 1Ć12
establishing for impulse testing, 1Ć12,
1Ć13ć1Ć14
Zoom
derivative math waveforms, 2Ć4 on FFT math waveforms, 1Ć11 on integral math waveforms, 3Ć5
Index
Loading...