Tektronix TDS 794D, TDS 784D, TDS 754D, TDS 724D, TDS 580D Service Manual

...

Service Manual

TDS 500D, TDS 600B & TDS 700D Digitizing Oscilloscopes
071-0136-00
Warning
The servicing instructions are for use by qualified personnel only. To avoid personal injury, do not perform any servicing unless you are qualified to do so. Refer to the Safety Summary prior to performing service.
Copyright T ektronix, Inc. All rights reserved. T ektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes
that in all previously published material. Specifications and price change privileges reserved. Printed in the U.S.A. T ektronix, Inc., P.O. Box 1000, Wilsonville, OR 97070–1000 TEKTRONIX and TEK are registered trademarks of T ektronix, Inc. FlexFormat and KlipChip are trademarks of T ektronix, Inc.
WARRANTY
T ektronix warrants that this product will be free from defects in materials and workmanship for a period of three (3) years from the date of shipment. If any such product proves defective during this warranty period, T ektronix, at its option, either will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the warranty period and make suitable arrangements for the performance of service. Customer shall be responsible for packaging and shipping the defective product to the service center designated by T ektronix, with shipping charges prepaid. T ektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the T ektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance and care. T ektronix shall not be obligated to furnish service under this warranty a) to repair damage resulting from attempts by personnel other than T ektronix representatives to install, repair or service the product; b) to repair damage resulting from improper use or connection to incompatible equipment; or c) to service a product that has been modified or integrated with other products when the effect of such modification or integration increases the time or difficulty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THIS PRODUCT IN LIEU OF ANY OTHER WARRANTIES, EXPRESSED OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX’ RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO THE CUST OMER FOR BREACH OF THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT , SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.

Table of Contents

Specifications

General Safety Summary xi. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Service Safety Summary xiii. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Preface xv. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Manual Structure xv. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Manual Conventions xvi. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Related Manuals xvii. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Contacting T ektronix xviii. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Introduction xix. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Service Strategy xix. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Service Offerings xix. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Before You Begin xxi. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Specifications 1–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Product Description 1–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
User Interface 1–2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Signal Acquisition System 1–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Horizontal System 1–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Trigger System 1–5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Acquisition Control 1–5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
On-Board User Assistance 1–6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Measurement Assistance 1–6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Storage 1–7. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
I/O 1–7. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Display 1–8. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Nominal Traits 1–9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Warranted Characteristics 1–17. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Typical Characteristics 1–25. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Operating Information

Installation 2–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Supplying Operating Power 2–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Operating Environment 2–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Applying and Interrupting Power 2–4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Repackaging Instructions 2–5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Installed Options 2–5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Operating Instructions 2–7. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Screen Layout 2–7. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Basic Procedures 2–7. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
TDS 500D, TDS 600B and TDS 700D Service Manual
i
Table of Contents

Theory of Operation

Theory of Operation 3–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Logic Conventions 3–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Module Overview 3–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Performance Verification Procedures

Performance Verification Procedures 4–1. . . . . . . . . . . . . . . . . . . . . . . . . .
Input Channels vs. Model 4–2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Conventions 4–2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Brief Procedures 4–5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Self Tests 4–5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Functional T ests 4–7. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Performance Tests 4–15. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Prerequisites 4–15. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Equipment Required 4–16. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
TDS 600B T est Record 4–19. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
TDS 500D/700D T est Record 4–23. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
TDS 794D T est Record 4–26. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Signal Acquisition System Checks 4–29. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Time Base System Checks 4–46. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Trigger System Checks 4–49. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Output Signal Checks 4–62. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Option 05 Video Trigger Checks      4–71. . . . . . . . . . .
Sine Wave Generator Leveling Procedure 4–88. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Optical Filters Checks (TDS 500D/TDS 700D Only) 4–91. . . . . . . . . . . . . . . . . . . . .

Adjustment Procedures

Maintenance

Adjustment Procedures 5–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Requirements for Performance 5–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Usage 5–2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Equipment Required 5–5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Adjustment Instructions 5–6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Probe Adjustment for the P6139A Probe 5–10. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Display Assembly Adjustment 5–21. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Maintenance 6–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Related Maintenance Procedures 6–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Preparation 6–2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Inspection and Cleaning 6–4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Removal and Installation Procedures 6–9. . . . . . . . . . . . . . . . . . . . . . . . . .
Preparation — Please Read 6–9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Access Procedure 6–15. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Procedures for External Modules 6–16. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Procedures for Outer-Chassis Modules 6–28. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Procedures for Inner-Chassis Modules 6–47. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Disassembly for Cleaning 6–54. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
ii
TDS 500D, TDS 600B and TDS 700D Service Manual

Options

Electrical Parts List

Diagrams

Mechanical Parts List

Table of Contents
Troubleshooting 6–57. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Diagnostics 6–57. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Firmware Updates 6–58. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Hard Disk Drive Maintenance 6–59. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Options and Accessories 7–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Options 7–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Standard Accessories 7–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Optional Accessories 7–4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Electrical Parts List 8–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Diagrams 9–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Symbols 9–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Mechanical Parts List 10–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Parts Ordering Information 10–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Using the Replaceable Parts List 10–2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
TDS 500D, TDS 600B and TDS 700D Service Manual
iii
Table of Contents

List of Figures

Figure 2–1: Map of Display Functions 2–8. . . . . . . . . . . . . . . . . . . . . . . . .
Figure 4–1: Map of display functions 4–3. . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 4–2: Verifying adjustments and signal-path compensation 4–6. . Figure 4–3: Universal test hookup for functional tests – TDS 600B
shown 4–8. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 4–4: Measurement of DC offset accuracy at zero setting 4–31. . . .
Figure 4–5: Initial test hookup 4–32. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 4–6: Measurement of DC accuracy at maximum offset and
position 4–35. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 4–7: Initial test hookup 4–37. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 4–8: Optional initial test hookup 4–37. . . . . . . . . . . . . . . . . . . . . . . .
Figure 4–9: Measurement of analog bandwidth 4–40. . . . . . . . . . . . . . . . . .
Figure 4–10: Initial test hookup 4–42. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 4–11: Measurement of channel delay – TDS 684B shown 4–44. . . .
Figure 4–12: Initial test hookup 4–46. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 4–13: Measurement of accuracy – Long-term and delay time 4–48
Figure 4–14: Initial test hookup 4–50. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 4–15: Measurement of time accuracy for pulse and glitch
triggering 4–51. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 4–16: Initial test hookup 4–53. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 4–17: Measurement of trigger-level accuracy 4–55. . . . . . . . . . . . .
Figure 4–18: Initial test hookup 4–58. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 4–19: Measurement of trigger sensitivity – 50 MHz results
shown on a TDS 684B screen 4–59. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 4–20: Initial test hookup 4–62. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 4–21: Measurement of main trigger out limits 4–64. . . . . . . . . . . . .
Figure 4–22: Initial test hookup 4–66. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 4–23: Measurement of probe compensator frequency 4–68. . . . . . .
Figure 4–24: Subsequent test hookup 4–69. . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 4–25: Measurement of probe compensator amplitude 4–70. . . . . .
Figure 4–26: Jitter test hookup 4–72. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 4–27: Jitter test displayed waveform – TDS 684B shown 4–73. . . .
Figure 4–28: Jitter test when completed – TDS 684B shown 4–74. . . . . . .
Figure 4–29: Triggered signal range test – 300 mV 4–75. . . . . . . . . . . . . . .
Figure 4–30: Triggered signal range test – 75 mV 4–76. . . . . . . . . . . . . . . .
iv
TDS 500D, TDS 600B and TDS 700D Service Manual
Table of Contents
Figure 4–31: 60 Hz Rejection test hookup 4–77. . . . . . . . . . . . . . . . . . . . . .
Figure 4–32: 60 Hz Rejection test setup signal 4–77. . . . . . . . . . . . . . . . . . .
Figure 4–33: Subsequent 60 Hz Rejection test hookup 4–78. . . . . . . . . . . .
Figure 4–34: 60 Hz Rejection test result – TDS 684B shown 4–79. . . . . . .
Figure 4–35: Line count accuracy test hookup 4–80. . . . . . . . . . . . . . . . . . .
Figure 4–36: Line count accuracy test setup waveform –
TDS 684B shown 4–81. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 4–37: Line count accuracy correct result waveform 4–82. . . . . . . .
Figure 4–38: Setup for sync duty cycle test 4–84. . . . . . . . . . . . . . . . . . . . . .
Figure 4–39: Sync duty cycle test: one-div neg pulse waveform 4–85. . . . .
Figure 4–40: Sync duty cycle test: critically adjusted pulse 4–86. . . . . . . .
Figure 4–41: Sine wave generator leveling equipment setup 4–88. . . . . . .
Figure 4–42: Equipment setup for maximum amplitude 4–90. . . . . . . . . .
Figure 4–43: Reference-receiver performance-verification set up 4–93. . . Figure 4–44: Optical impulse of Ch1 input from OA5022 Optical Atten–
uator (OIG501 / OIG502 fed into optical attenuator in Step 1) 4–95. . Figure 4–45: Optical impulse response for OC–12 SONET Reference
Receiver 4–98. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 4–46: Optical impulse response for OC–12 SONET Reference
Receiver 4–99. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 5–1: Accessing the protection switch 5–9. . . . . . . . . . . . . . . . . . . . .
Figure 5–2: Hookup for probe compensation 5–11. . . . . . . . . . . . . . . . . . . .
Figure 5–3: Performing probe compensation 5–12. . . . . . . . . . . . . . . . . . . .
Figure 5–4: Proper and improper probe compensation 5–12. . . . . . . . . . .
Figure 5–5: Exposing the Inner Probe Tip 5–13. . . . . . . . . . . . . . . . . . . . . .
Figure 5–6: Initial test hookup 5–14. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 5–7: Exposing the probe body 5–16. . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 5–8: Initial test hookup 5–17. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 5–9: Locations of P6139A Probe Adjustments 5–19. . . . . . . . . . . . .
Figure 5–10: Adjustments versus front-corner response 5–20. . . . . . . . . . .
Figure 5–11: Five and Ten Percent Luminance Patches 5–22. . . . . . . . . . .
Figure 6–1: External modules 6–12. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 6–2: Outer-chassis modules 6–13. . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 6–3: Inner-chassis modules 6–14. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 6–4: Knob removal 6–17. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 6–5: Line fuse and line cord removal 6–18. . . . . . . . . . . . . . . . . . . .
Figure 6–6: Rear cover and cabinet removal 6–21. . . . . . . . . . . . . . . . . . . .
TDS 500D, TDS 600B and TDS 700D Service Manual
v
Table of Contents
Figure 6–7: Front cover, trim ring, menu buttons, and attenuator
panel removal (front cover not shown) 6–22. . . . . . . . . . . . . . . . . . . . . .
Figure 6–8: A12 Front-panel assembly removal 6–24. . . . . . . . . . . . . . . . . .
Figure 6–9: Cabinet modules removal 6–27. . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 6–10: A14 D1 bus and analog-power and digital-power cables
removal 6–29. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 6–11: Remove circuit board assembly from oscilloscope 6–30. . . . .
Figure 6–12: Remove circuit board from assembly 6–31. . . . . . . . . . . . . . .
Figure 6–13: Circuit board installed 6–33. . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 6–14: Circuit board removal 6–34. . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 6–15: A11 Processor/display removal 6–36. . . . . . . . . . . . . . . . . . . .
Figure 6–16: Hard disk removal 6–38. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 6–17: Board bracket removal 6–39. . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 6–18: A10 Acquisition board removal 6–42. . . . . . . . . . . . . . . . . . . .
Figure 6–19: Floppy disk drive removal 6–44. . . . . . . . . . . . . . . . . . . . . . . .
Figure 6–20: Rear chassis removal 6–46. . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 6–21: A16 Low voltage power supply removal 6–48. . . . . . . . . . . . .
Figure 6–22: Display assembly removal 6–50. . . . . . . . . . . . . . . . . . . . . . . .
Figure 6–23: Display driver board removal 6–51. . . . . . . . . . . . . . . . . . . . .
Figure 6–24: Front subpanel removal 6–53. . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 6–25: Accessing the protection switch 6–58. . . . . . . . . . . . . . . . . . . .
Figure 6–26: Primary troubleshooting procedure 6–60. . . . . . . . . . . . . . . .
Figure 6–27: Module isolation troubleshooting procedure 6–62. . . . . . . . .
Figure 6–28: A16 Low voltage power supply module isolation
troubleshooting procedure 6–63. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 6–29: Power supply voltage measurement locations 6–64. . . . . . . .
Figure 6–30: Color display troubleshooting procedure 6–65. . . . . . . . . . . .
Figure 6–31: Horizontal and vertical sync signals – color display 6–66. . . Figure 6–32: A video signal with white, black, and blanking levels –
color display 6–66. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 6–33: Processor/acquisition troubleshooting procedure 6–67. . . . .
Figure 6–34: Processor/front panel troubleshooting procedure 6–68. . . . .
Figure 6–35: Attenuator/acquisition troubleshooting procedure 6–69. . . .
Figure 6–36: A11 DRAM Processor/Display module 6–70. . . . . . . . . . . . . .
Figure 6–37: A10 Acquisition module (test points) 6–71. . . . . . . . . . . . . . .
Figure 9–1: Interconnections 9–2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 9–2: Block Diagram 9–4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
vi
TDS 500D, TDS 600B and TDS 700D Service Manual
Table of Contents
Figure 10–1: External modules 10–7. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 10–2: Outer-chassis modules 10–10. . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 10–3: Inner-chassis modules 10–12. . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 10–4: Cables, bottom view 10–14. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 10–5: Cables, top view 10–16. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 10–6: Floppy disk drive 10–18. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 10–7: Accessories 10–20. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
TDS 500D, TDS 600B and TDS 700D Service Manual
vii
Table of Contents

List of Tables

Table 1–1: Key features of the TDS 500D, 600B and 700D
oscilloscopes 1–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Table 1–2: Record length and divisions per record vs. TDS model 1–4. .
Table 1–3: Nominal traits — Signal acquisition system 1–11. . . . . . . . . . .
Table 1–4: Nominal traits — Time base system 1–12. . . . . . . . . . . . . . . . . .
Table 1–5: Nominal traits — Triggering system 1–13. . . . . . . . . . . . . . . . .
Table 1–6: Nominal traits — Display system 1–15. . . . . . . . . . . . . . . . . . . .
Table 1–7: Nominal traits — GPIB interface, output ports, and
power fuse 1–16. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Table 1–8: Nominal traits — Data handling and reliability 1–16. . . . . . . .
Table 1–9: Nominal traits — Mechanical 1–17. . . . . . . . . . . . . . . . . . . . . . .
Table 1–10: Warranted characteristics — Signal acquisition system 1–20
Table 1–11: Warranted characteristics — Time base system 1–22. . . . . . .
Table 1–12: Warranted characteristics — Triggering system 1–22. . . . . .
Table 1–13: Warranted characteristics — Output ports, probe
compensator, and power requirements 1–23. . . . . . . . . . . . . . . . . . . . .
Table 1–14: Warranted characteristics — Environmental 1–24. . . . . . . . .
Table 1–15: Certifications and compliances 1–26. . . . . . . . . . . . . . . . . . . . .
Table 1–16: Typical characteristics — Signal acquisition system 1–27. . .
Table 1–17: Typical characteristics — Triggering system 1–30. . . . . . . . .
viii
Table 2–1: Power cord conductor identification 2–2. . . . . . . . . . . . . . . . .
Table 2–2: Power cord identification 2–2. . . . . . . . . . . . . . . . . . . . . . . . . .
Table 2–3: Effects of corrupted data 2–4. . . . . . . . . . . . . . . . . . . . . . . . . .
Table 4–1: Test equipment 4–16. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Table 4–2: DC offset accuracy (zero setting) 4–30. . . . . . . . . . . . . . . . . . . .
Table 4–3: DC Voltage measurement accuracy 4–34. . . . . . . . . . . . . . . . . .
Table 4–4: Analog bandwidth 4–39. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Table 4–5: Delay between channels worksheet 4–45. . . . . . . . . . . . . . . . . .
Table 4–6: Trigger level accuracy readout 4–54. . . . . . . . . . . . . . . . . . . . . .
Table 4–7: Delayed trigger level accuracy readout 4–56. . . . . . . . . . . . . . .
Table 4–8: Available Filters 4–91. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Table 4–9: Reference Receiver Filter Options 4–91. . . . . . . . . . . . . . . . . . .
Table 4–10: Option 3C and 4C Specifications
(not available on TDS 794D) 4–92. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
TDS 500D, TDS 600B and TDS 700D Service Manual
Table of Contents
Table 4–11: Available receivers 4–92. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Table 4–12: Bessel Thompson frequency response and reference
receiver limits 4–100. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Table 5–1: Calibration Tests 5–3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Table 5–2: Adjustments Required for Module Replaced 5–4. . . . . . . . . .
Table 5–3: Adjustments and Dependencies 5–4. . . . . . . . . . . . . . . . . . . . .
Table 5–4: Test Equipment, Fixtures, and Supplies 5–5. . . . . . . . . . . . . .
Table 5–5: GPIB Board Configuration 5–7. . . . . . . . . . . . . . . . . . . . . . . . .
Table 6–1: Relative susceptibility to static-discharge damage 6–3. . . . . .
Table 6–2: External inspection check list 6–5. . . . . . . . . . . . . . . . . . . . . . .
Table 6–3: Internal inspection check list 6–6. . . . . . . . . . . . . . . . . . . . . . .
Table 6–4: Tools Required for Module Removal 6–11. . . . . . . . . . . . . . . . .
Table 6–5: Normal supply voltages (measured on J26 and J27 on
the A11 DRAM Processor/Display module) 6–64. . . . . . . . . . . . . . . . . .
Table 6–6: No-load supply voltages (measured on J5 and J6 on
the A17 Main LV Power Supply module) 6–64. . . . . . . . . . . . . . . . . . . .
Table 7–1: Options 7–1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Table 7–2: Standard accessories 7–4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Table 7–3: Optional accessories 7–5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Table 7–4: Recommended Probe Cross Reference 7–6. . . . . . . . . . . . . . .
Table 7–5: Accessory software 7–9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
TDS 500D, TDS 600B and TDS 700D Service Manual
ix
Table of Contents
x
TDS 500D, TDS 600B and TDS 700D Service Manual

General Safety Summary

Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. To avoid potential hazards, use this product only as specified.
Only qualified personnel should perform service procedures.
To Avoid Fire or Personal Injury
Use Proper Power Cord. Use only the power cord specified for this product and
certified for the country of use. Ground the Product. This product is grounded through the grounding conductor
of the power cord. To avoid electric shock, the grounding conductor must be connected to earth ground. Before making connections to the input or output terminals of the product, ensure that the product is properly grounded.
Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings and markings on the product. Consult the product manual for further ratings information before making connections to the product.
Connect and Disconnect Properly . Do not connect or disconnect probes or test leads while they are connected to a voltage source.
Do Not Operate Without Covers. Do not operate this product with covers or panels removed.
Use Proper Fuse. Use only the fuse type and rating specified for this product. Avoid Exposed Circuitry. Do not touch exposed connections and components
when power is present. Do Not Operate With Suspected Failures. If you suspect there is damage to this
product, have it inspected by qualified service personnel.
Do Not Operate in Wet/Damp Conditions. Do Not Operate in an Explosive Atmosphere. Provide Proper Ventilation. Refer to the manual’s installation instructions for
details on installing the product so it has proper ventilation.
TDS 500D, TDS 600B and TDS 700D Service Manual
xi
General Safety Summary
Symbols and Terms
T erms in this Manual. These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in damage to this product or other property.
T erms on the Product. These terms may appear on the product: DANGER indicates an injury hazard immediately accessible as you read the
marking. WARNING indicates an injury hazard not immediately accessible as you read the
marking. CAUTION indicates a hazard to property including the product. Symbols on the Product. The following symbols may appear on the product:
WARNING
High Voltage
Protective Ground
(Earth) T erminal
CAUTION
Refer to Manual
Double
Insulated
xii
TDS 500D, TDS 600B and TDS 700D Service Manual

Service Safety Summary

Only qualified personnel should perform service procedures. Read this Service Safety Summary and the General Safety Summary before performing any service
procedures. Do Not Service Alone. Do not perform internal service or adjustments of this
product unless another person capable of rendering first aid and resuscitation is present.
Disconnect Power. To avoid electric shock, disconnect the mains power by means of the power cord or, if provided, the power switch.
Use Caution When Servicing the CRT. To avoid electric shock or injury, use extreme caution when handling the CRT. Only qualified personnel familiar with CRT servicing procedures and precautions should remove or install the CRT.
CRTs retain hazardous voltages for long periods of time after power is turned off. Before attempting any servicing, discharge the CRT by shorting the anode to chassis ground. When discharging the CRT, connect the discharge path to ground and then the anode. Rough handling may cause the CRT to implode. Do not nick or scratch the glass or subject it to undue pressure when removing or installing it. When handling the CRT, wear safety goggles and heavy gloves for protection.
Use Care When Servicing With Power On. Dangerous voltages or currents may exist in this product. Disconnect power, remove battery (if applicable), and disconnect test leads before removing protective panels, soldering, or replacing components.
To avoid electric shock, do not touch exposed connections. X-Radiation. To avoid x-radiation exposure, do not modify or otherwise alter the
high-voltage circuitry or the CRT enclosure. X-ray emissions generated within this product have been sufficiently shielded.
TDS 500D, TDS 600B and TDS 700D Service Manual
xiii
Service Safety Summary
xiv
TDS 500D, TDS 600B and TDS 700D Service Manual

Preface

Manual Structure

This preface contains information needed to properly use this manual to service the TDS 500D, TDS 600B and TDS 700D Oscilloscopes, as well as general information critical to safe and effective servicing of this oscilloscope.
The following models are covered:
TDS 500D: TDS 520D, TDS 540D, and TDS 580D
TDS 600B: TDS 620B, TDS 644B, TDS 680B, and TDS 684B
TDS 700D: TDS 724D, TDS 754D, TDS 784D, and TDS 794D
This manual is divided into sections, such as Specifications and Theory of Operation. Further, it is divided into subsections, such as Product Description and Removal and Installation Procedures.
Sections containing procedures also contain introductions to those procedures. Be sure to read these introductions because they provide information needed to do the service correctly and efficiently. The following is a brief description of each manual section.
H Specifications contains a product description of the TDS 500D, TDS 600B
and TDS 700D Oscilloscopes and tables of the characteristics and descrip­tions that apply to it.
H Operating Information includes general information and operating instruc-
tions at the level needed to safely power on and service this oscilloscope. A statement of the service strategy that this manual supports, and instructions for shipment of the oscilloscopes are found in this section.
H Theory of Operation contains circuit descriptions that support general service
and fault isolation.
H Performance Verification contains a collection of procedures for confirming
that these oscilloscopes function properly and meet warranted limits.
H Adjustment Procedures contains a collection of procedures for adjusting
these oscilloscopes to meet warranted limits.
H Maintenance contains information and procedures for doing preventive and
corrective maintenance of these oscilloscopes. Instructions for cleaning, for
TDS 500D, TDS 600B and TDS 700D Service Manual
xv
Preface

Manual Conventions

module removal and installation, and for fault isolation to a module are
found here. H Options contains information on servicing any of the factory-installed
options that may be present in your oscilloscope.
H Electrical Parts List contains a statement referring you to Mechanical
Replaceable Parts, where both electrical and mechanical modules are listed.
H Diagrams contains a block diagram and an interconnection diagram useful
for isolating failed modules. H Mechanical Parts List includes a table of all replaceable modules, their
descriptions, and their Tektronix part numbers.
This manual uses certain conventions which you should become familiar with before doing service.
Modules
Channels
Safety
Symbols
Throughout this manual, any replaceable component, assembly, or part of these oscilloscopes is referred to generically as a module. In general, a module is an assembly, like a circuit board, rather than a component, like a resistor or an integrated circuit. Sometimes a single component is a module; for example, each chassis part of the oscilloscope is a module.
Some TDS models have two auxiliary channels called AUX 1 and AUX 2, instead of CH 3 and CH 4. References to these channels default to CH 3 and CH 4. If your oscilloscope is one of models TDS 520D, TDS 620B, TDS 680B, or TDS 724D, read AUX 1 and AUX 2 respectively for all references to CH 3 and CH 4 in this manual.
Symbols and terms related to safety appear in the Safety Summary found at the beginning of this manual.
Besides the symbols related to safety, this manual uses the following symbols:
STOP. The stop labels information which must be read in order to correctly do service and to avoid incorrectly using or applying service procedures.
xvi
The clock icon labels procedure steps which require a pause to wait for the oscilloscope to complete some operation before you can continue.
TDS 500D, TDS 600B and TDS 700D Service Manual

Related Manuals

Preface
Various icons such as the example icon at the left are used in procedures to help identify certain readouts and menu functions on screen.
The TDS 500D, TDS 600B and TDS 700D Oscilloscopes come with the following manuals:
TDS 500D, TDS 600B & TDS700D User Manual (Tektronix part number 071-0130-XX) contains a tutorial to quickly show you how to operate the TDS 500D, TDS 600B and TDS 700D Oscilloscopes and an in depth discussion of how to more completely use their features. Applications are also discussed.
TDS 500D, TDS 600B & TDS 700D Reference (Tektronix part number 071-0137-XX) contains a brief overview of oscilloscope operation.
TDS Family (500D, 600B, and 700D) Programmer Diskette (part of the User Manual) contains information for programmed operation via the GPIB interface.
Included are the complete command set, setup information, and programming examples.
TDS 500D, TDS 600B & TDS 700D Technical Reference (Tektronix part number 071-0135-XX) contains performance verification procedures and specifications.
TDS 500D, TDS 600B and TDS 700D Service Manual
xvii
Preface

Contacting Tektronix

Product Support
Service Support
For other information
To write us Tektronix, Inc.
For application-oriented questions about a Tektronix measure­ment product, call toll free in North America: 1-800-TEK-WIDE (1-800-835-9433 ext. 2400) 6:00 a.m. – 5:00 p.m. Pacific time
Or, contact us by e-mail: tm_app_supp@tek.com
For product support outside of North America, contact your local Tektronix distributor or sales office.
Contact your local Tektronix distributor or sales office. Or, visit our web site for a listing of worldwide service locations.
http://www.tek.com In North America:
1-800-TEK-WIDE (1-800-835-9433) An operator will direct your call.
P.O. Box 1000 Wilsonville, OR 97070-1000
xviii
TDS 500D, TDS 600B and TDS 700D Service Manual

Introduction

Service Strategy

STOP. Throughout this manual, any field-replaceable component, assembly, or part of this oscilloscope is referred to generically as a module.
This manual contains all the information needed for periodic maintenance of the TDS 500D, TDS 600B and TDS 700D Oscilloscopes. (Examples of such information are procedures for checking performance and for readjustment.) Further, it contains all information for corrective maintenance down to the module level. This means that the procedures, diagrams, and other troubleshoot­ing aids help isolate failures to a specific module, rather than to components of that module. Once a failure is isolated, replace the module with a fresh unit obtained from the factory.
All modules are listed in Mechanical Parts List. To isolate a failure to a module, use the fault isolation procedures found in Section 6, Maintenance Information. To remove and replace any failed module, follow the instructions in Removal and Installation Procedures, also found in Section 6.

Service Offerings

Warranty Repair Service
Tektronix provides service to cover repair under warranty as well as other services that may provide a cost-effective answer to your service needs.
Whether providing warranty repair service or any of the other services listed below, Tektronix service technicians, trained on Tektronix products, are best equipped to service your TDS 500D, TDS 600B and TDS 700D Oscilloscopes. Tektronix technicians are appraised of the latest information on improvements to the product as well as the latest new options to the product.
Tektronix warrants this product for three years from date of purchase, excluding probes for which the warranty is one year. (The warranty appears after the title page and copyright page in this manual.) Tektronix technicians provide warranty service at most Tektronix service locations worldwide. Your Tektronix product catalog lists all service locations worldwide.
TDS 500D, TDS 600B and TDS 700D Service Manual
xix
Introduction
Repair or Calibration
Service
The following services may be purchased to tailor repair and/or calibration of your TDS 500D, TDS 600B and TDS 700D Oscilloscopes to fit your require­ments.
At-Depot Service. Tektronix offers several standard-priced adjustment (calibra­tion) and repair services:
H A single repair and/or adjustment. H Calibrations using equipment and procedures that meet the traceability
standards specific to the local area. H Annual maintenance agreements that provide for either calibration and repair
or calibration only of the oscilloscope. Of these services, the annual maintenance agreement offers a particularly
cost-effective approach to service for many owners of the TDS 500D, TDS 600B and TDS 700D Oscilloscopes. Such agreements can be purchased to span several years.
On-Site Service. The annual maintenance agreement can be purchased with on-site service, with repair and calibration done at your facility. This service reduces the time your oscilloscope is out of service when calibration or repair is required.
Self Service
Tektronix supports repair to the module level by providing Module Exchange and Module Repair and Return.
Module Exchange. This service reduces down time for repair by allowing you to exchange most modules for remanufactured ones. Tektronix ships you an updated and tested exchange module from the Beaverton, Oregon service center, typically within 24 hours. Each module comes with a 90-day service warranty.
Module Repair and Return. This service returns to you within 10 days the same module that you shipped to Tektronix. The module shipped is repaired, tested, and returned to you from the Beaverton, Oregon service center. It is not updated to match current modules of the same type. Again, each module comes with a 90-day service warranty.
For More Information. Contact your local Tektronix service center or sales engineer for more information on any of the repair or adjustment services just described.
xx
TDS 500D, TDS 600B and TDS 700D Service Manual

Before You Begin

Introduction
This manual is for servicing the TDS 500D, TDS 600B and TDS 700D Oscilloscopes. To prevent injury to yourself or damage to the oscilloscope, do the following tasks before you attempt service:
H Read the Safety Summary found at the beginning of this manual. H Read Service Strategy in this section and Supplying Operating Power in
section 2.
When using this manual for servicing, be sure to read and follow all warnings, cautions, and notes.
TDS 500D, TDS 600B and TDS 700D Service Manual
xxi
Introduction
xxii
TDS 500D, TDS 600B and TDS 700D Service Manual

Specifications

Product Description

This section begins with a general description of the traits of the TDS 500D, TDS 600B and TDS 700D oscilloscopes. Three sections follow, one for each of three classes of traits: nominal traits, warranted characteristics, and typical characteristics.
The TDS 500D, TDS 600B and TDS 700D oscilloscopes are portable, four-chan­nel instruments suitable for use in a variety of test and measurement applications and systems. Table 1–1 lists key features.
T able 1–1: Key features of the TDS 500D, 600B and 700D oscilloscopes
Feature TDS 600B TDS 500D & TDS 700D
Digitizing rate,
maximum
TDS 684B: 5 GS/s on ea. of 4 ch TDS 680B: 5 GS/s on ea. of 2 ch TDS 644B: 2.5 GS/s on ea. of 4 ch TDS 620B: 2.5 GS/s on ea. of 2 ch
TDS 580D, TDS 784D, TDS 794D: 4 GS/s TDS 540D, 754D: 2 GS/s TDS 520D, 724D: 1 GS/s
Analog bandwidth TDS 794D: 2 GHz
TDS 580D, TDS 680B, 684B, and 784D: 1 GHz TDS 520D, 540D, 620B, 644B, 724D and 754D: 500 MHz
No. of Channels TDS 644B & 684B: 4
TDS 620B & 680B: 2 + 2
Record lengths,
maximum
Acquisition modes Sample, envelope, peak detect
Trigger modes Modes include:
Display TDS 520D, 540D, 580D, 620B, 680B: Monochrome
15,000 samples 50,000 samples
and average
Edge, logic, and pulse. Video trigger , with option 05, modes include:
NTSC, SECAM, PAL, HDTV, and FlexFormat. (Not available on TDS 794D)
Communications Trigger with option 2C (available on TDS 500D/700D), modes include: AMI, CMI, and NRZ
TDS 644B, 684B, 724D, 754D, 784D, 794D: Color
1
TDS 540D, 580D, TDS 754D, 784D & 794D: 4 TDS 520D & 724D: 2 + 2
(500,000 with option 1M) (8,000,000 with option 2M)
Sample, envelope, average, high-resolution, and peak-detect
1
TDS 500D, TDS 600B and TDS 700D Service Manual
1–1
Specifications
T able 1–1: Key features of the TDS 500D, 600B and 700D oscilloscopes (cont.)
Feature TDS 500D & TDS 700DTDS 600B
Storage Floppy disk drive:
1.44 Mbyte, 3.5 inch, DOS 3.3-or-later floppy disk drive Internal hard disk drive
(optional on the TDS 500D and 700D) Iomega Zip drive compatible NVRAM storage for saving waveforms, hardcopies, and setups
I/O Full GPIB programmability .
Hardcopy output using GPIB, RS-232, or Centronics ports
1
Two plus Two channel operation allows up to two of the four channels to be displayed simultaneously. Channels not displayed can be used to couple a triggering signal to the oscilloscope.

User Interface

Indicators
Menus
Use a combination of front-panel buttons, knobs, and on-screen menus to control the many functions of the oscilloscope. The front-panel controls are grouped according to function: vertical, horizontal, trigger, and special. Set a function you adjust often, such as vertical positioning or the time base setting, directly by its own front-panel knob. Set a function you change less often, such as vertical coupling or horizontal mode, indirectly using a selected menu.
Pressing one (sometimes two) front-panel button(s), such as vertical menu, displays a main menu of related functions, such as coupling and bandwidth, at the bottom of the screen. Pressing a main-menu button, such as coupling, displays a side menu of settings for that function, such as AC, DC, or GND (ground) coupling, at the right side of the screen. Pressing a side-menu button selects a setting such as DC.
On-screen readouts help you keep track of the settings for various functions, such as vertical and horizontal scale and trigger level. Some readouts use the cursors or the automatic parameter extraction feature (called measure) to display the results of measurements made or the status of the instrument.
General Purpose Knob
1–2
Assign the general purpose knob to adjust a selected parameter function. More quickly change parameters by toggling the SHIFT button. Use the same method as for selecting a function, except the final side-menu selection assigns the general purpose knob to adjust some function, such as the position of measure­ment cursors on screen, or the setting for a channel fine gain.
TDS 500D, TDS 600B and TDS 700D Service Manual
Specifications
GUI
The user interface also makes use of a GUI, or Graphical User Interface, to make setting functions and interpreting the display more intuitive. Some menus and status are displayed using iconic representations of function settings, such as those shown here for full, 250 MHz and 20 MHz bandwidth. Such icons allow you to more readily determine status or the available settings.

Signal Acquisition System

The signal acquisition system provides up to four, full-featured vertical channels with calibrated vertical scale factors from 1 mV to 10 V per division, depending on TDS model. All channels can be acquired simultaneously.
Each of the full-featured channels can be displayed, vertically positioned, and offset, and their vertical coupling specified. Some models can have their bandwidth limited (250 MHz or 20 MHz). Fine gain can also be adjusted.
Besides these channels, up to three math waveforms and four reference wave­forms are available for display. (A math waveform results when you specify dual waveform operations, such as add, on any two channels. A reference waveform results when you save a waveform in a reference memory).

Horizontal System

There are three horizontal display modes: main only, main intensified, and delayed only. You can select among various horizontal record length settings.
A feature called “Fit to Screen” allows you to view entire waveform records within the 10 division screen area. Waveforms are compressed to fit on the screen. See Table 1–2.
Both the delayed only display and the intensified zone on the main intensified display may be delayed by time with respect to the main trigger. Both can be set to display immediately after the delay (delayed runs after main mode). The delayed display can also be set to display at the first valid trigger after the delay (delayed-triggerable modes).
The delayed display (or the intensified zone) may also be delayed by a selected number of events. In this case, the events source is the delayed-trigger source. The delayed trigger can also be set to occur after a number of events plus an amount of time.
TDS 500D, TDS 600B and TDS 700D Service Manual
1–3
Specifications
Standard Model
Record length
All
All c
els
4
Model
1M
Record length
T able 1–2: Record length and divisions per record vs. TDS model
Divisions per record
s
TDS 500D, TDS600B & TDS 700D
hann
FTS1 Off
500 10 divs 10 divs 1,000 20 divs 10 divs 2,500 50 divs 10 divs 5,000 100 divs 10 divs 15,000 300 divs 10 divs
2
FTS1 On
3
TDS 500D & TDS 700D, All channels
s with Option
TDS 500D & TDS 700D, All channels
TDS 500D & TDS 700D, All channels
TDS 500D & TDS 700D, All channels
TDS 520D One channel only
TDS 540D, TDS 580D, & All TDS 700D One or two channels
TDS 540D, TDS 580D, & All TDS 700D One channel only
50,000 1,000 divs 10 divs
Divisions per record FTS1 Off
2
FTS1 On
75,000 1,500 divs 10 divs
100,000 2,000 divs 10 divs
130,000 2,600 divs 10 divs
250,000 5,000 divs 10 divs
500,000 10,000 divs 10 divs
3
1–4
TDS 500D, TDS 600B and TDS 700D Service Manual
Specifications
Model
2M
Record length
T able 1–2: Record length and divisions per record vs. TDS model (cont.)
Divisions per record
s with Option
TDS 520D & TDS 724D
One or two channels
TDS 540D, TDS 580D, TDS 754D,
TDS 784D & TDS 794D
Three or four channels
FTS1 Off
2,000,000 40,000 divs 10 divs
2
FTS1 On
3

Trigger System

TDS 520D & TDS 724D
One channel only
TDS 540D, TDS 580D, TDS 754D,
TDS 784D & TDS 794D
Two channels
TDS 540D, TDS 754D,
TDS 784D & TDS 794D
One channel only
1
Fit to Screen setting
2
Fit to Screen off preserves 50 samples/division in a 1–2–5 sec/division sequence.
3
Fit to Screen on lets the samples/division and the sec/division sequence vary.
4
All channels means all that may be displayed at one time: four channels for some models, two for others. See Table 1–1 and its footnote.
4,000,000 80,000 divs 10 divs
8,000,000 160,000 divs 10 divs
The triggering system supports a varied set of features for triggering the signal-acquisition system. Trigger signals recognized include:
H Edge (main- and delayed-trigger systems): This familiar type of triggering is
fully configurable for source, slope, coupling, mode (auto or normal), and holdoff.
H Logic (main-trigger system): This type of triggering can be based on pattern
(asynchronous) or state (synchronous). In either case, logic triggering is configurable for sources, for Boolean operators to apply to those sources, for logic pattern or state on which to trigger, for mode (auto or normal), and for holdoff. Time qualification may be selected in pattern mode. Another class of logic trigger, setup/hold, triggers when data in one trigger source changes state within the setup and hold times that you specify relative to a clock in another trigger source.
H Pulse (main-trigger system): Pulse triggering is configurable for triggering
on runt or glitch pulses, or on pulse widths or periods inside or outside limits
TDS 500D, TDS 600B and TDS 700D Service Manual
1–5
Specifications
that you specify. It can also trigger on a pulse edge that has a slew rate faster
or slower than the rate you specify. The timeout trigger will act when events
do not occur in a defined time period. The pulse trigger is also configurable
for source, polarity, mode, and holdoff. H Video (with option 05: Video Trigger): Video triggering is compatible with
standard NTSC, PAL, SECAM, and HDTV formats. An additional feature
called FlexFormat
format on which to trigger. H Comm (with option 2C): is provided for triggering on AMI, CMI, or NRZ
communications signals. You can choose where the trigger point is located within the acquired waveform
record by selecting the amount of pretrigger data displayed. Presets of 10%, 50%, and 90% of pretrigger data can be selected in the horizontal menu, or the general purpose knob can be assigned to set pretrigger data to any value within the 0% to 100% limits.
TM
(flexible format) allows the user to define the video

Acquisition Control

You can specify a mode and manner to acquire and process signals that matches your measurement requirements.
H Select the mode for interpolation (linear or sin (x)/x). This can increase the
apparent sample rate on the waveform when the maximum real-time rate is
exceeded. H Use sample, envelope, average and peak detect modes to acquire signals.
With the Oscilloscopes, also use high-resolution mode. H Set the acquisition to stop after a single acquisition (or sequence of
acquisitions if acquiring in average or envelope modes) or after a limit
condition has been met. H Select channel sources for compliance with limit tests. You can direct the
TDS to signal you or generate hard copy output either to a printer or to a
floppy-disk file based on the results. Also, you can create templates for use
in limit tests.
1–6
TDS 500D, TDS 600B and TDS 700D Service Manual

On-Board User Assistance

Help and autoset can assist you in setting up the oscilloscope to make your measurements.
Specifications
Help
Autoset
Help displays operational information about any front-panel control. When help mode is in effect, manipulating any front-panel control causes the oscilloscope to display information about that control. When help is first invoked, an introduc­tion to help is displayed on screen.
Autoset automatically sets up the oscilloscope for a viewable display based on the input signal.

Measurement Assistance

Once you have set up to make your measurements, the cursor and measure features can help you quickly make those measurements.
Cursor
Three types of cursors are provided for making parametric measurements on the displayed waveforms. Horizontal bar cursors (H Bar) measure vertical parame­ters (typically volts). Vertical bar cursors (V Bar) measure horizontal parameters (typically time or frequency). Paired cursors measure both amplitude and time simultaneously. These are delta measurements; that is, measurements based on the difference between two cursors.
Both H Bar and V Bar cursors can also be used to make absolute measurements. For the H Bars, either cursor can be selected to read out its voltage with respect to any channel’s ground reference level. For the V Bars, the cursors measure time with respect to the trigger point (event) of the acquisition. The cursors can also control the portion of the waveform on which automatic measurements are made.
For time measurements, units can be either seconds or hertz (for 1/time). With the video trigger option installed (Option 05), you can measure the video
line number using the vertical cursors. You can measure IRE amplitude (NTSC) using the horizontal cursors with or without the video trigger option installed.
Measure
TDS 500D, TDS 600B and TDS 700D Service Manual
Measure can automatically extract parameters from the signal input to the oscilloscope. Any four out of the 25 parameters available can be displayed to the screen. The waveform parameters are measured continuously with the results updated on-screen as the oscilloscope continues to acquire waveforms.
1–7
Specifications
Digital Signal Processing
(DSP)

Storage

An important component of the multiprocessor architecture of this oscilloscope is Tektronix’ proprietary digital signal processor, the DSP. This dedicated processor supports advanced analysis of your waveforms when doing such compute-intensive tasks as interpolation, waveform math, and signal averaging. It also teams with a custom display system to deliver specialized display modes (See Display, later in this description).
Acquired waveforms may be saved in any of four nonvolatile REF (reference) memories or on a 3.5 inch, DOS 3.3-or-later compatible disk. Any or all of the saved waveforms may be displayed for comparison with the waveforms being currently acquired.
The Oscilloscopes instrument with option 2M can save waveforms to an internal hard disk drive. Any or all of the saved waveforms may be displayed for comparison with the waveforms being currently acquired.
The source and destination of waveforms to be saved may be chosen. You can save any of the four channels to any REF memory or move a stored reference from one REF memory to another. Reference waveforms may also be written into a REF memory location via the GPIB interface.
I/O
The oscilloscope is fully controllable and capable of sending and receiving waveforms over the GPIB interface (IEEE Std 488.1–1987/IEEE Std 488.2–1987 standard). This feature makes the instrument ideal for making automated measurements in a production or research and development environment that calls for repetitive data taking. Self-compensation and self-diagnostic features built into the oscilloscope to aid in fault detection and servicing are also accessible using commands sent from a GPIB controller.
The oscilloscope can also output copies of its display using the hardcopy feature. This feature allows you to output waveforms and other on-screen information to a variety of graphic printers and plotters from the TDS front panel, providing hard copies without requiring you to put the TDS into a system-controller environment.
You can make hardcopies in a variety of popular output formats, such as PCX, TIFF, BMP, RLE, EPS, Interleaf, and EPS mono or color. You can also save hardcopies in a disk file in any of the formats listed in the I/O section.
1–8
TDS 500D, TDS 600B and TDS 700D Service Manual

Display

Specifications
The hardcopies obtained are based on what is displayed on-screen at the time hardcopy is invoked. The hardcopies can be stamped with date and time and spooled to a queue for printing at a later time. You can output screen information via GPIB, RS-232-C, or Centronics interfaces.
The TDS 500D, TDS 600B and TDS 700D Oscilloscopes offer flexible display options. You can customize the following attributes of your display:
H Color (TDS 644B, TDS 684B, and TDS 700D): Waveforms, readouts,
graticule, and variable persistence with color coding
H Intensity: waveforms, readouts, and graticule H Style of waveform display(s): vectors or dots, intensified or nonintensified
samples, infinite persistence, and variable persistence
H Interpolation method: Sin(x)/x or Linear
Zoom
H Display format: xy or yt with various graticule selections including NTSC
and PAL to be used with video trigger (option 05)
This oscilloscope also provides an easy way to focus in on those waveform features you want to examine up close. By invoking zoom, you can magnify the waveform using the vertical and horizontal controls to expand (or contract) and position it for viewing.
TDS 500D, TDS 600B and TDS 700D Service Manual
1–9
Specifications
1–10
TDS 500D, TDS 600B and TDS 700D Service Manual

Nominal Traits

set
All except
This section contains a collection of tables that list the various nominal traits that describe the TDS 500D, TDS 600B, and TDS 700D oscilloscopes. Electrical and mechanical traits are included.
Nominal traits are described using simple statements of fact such as “Four, all identical” for the trait “Input Channels, Number of,” rather than in terms of limits that are performance requirements.
T able 1–3: Nominal traits — Signal acquisition system
Name Description
Bandwidth Selections 20 MHz, 250 MHz, and FULL (No BW limit on TDS 794D) Samplers, Number of TDS 540D, 580D, 644B, 684B, 754D, 784D, and 794D: Four, simultaneous
TDS 520D, 620B, 680B, and 724D: Two, simultaneous Digitized Bits, Number of 8 bits Input Channels, Number of Four Input Coupling DC, AC, or GND (TDS 794D – DC or GND only) Input Impedance Selections
Ranges, Off (
TDS 794D)
1
1 Mor 50 (TDS 794D – 50 only)
Volts/Div setting Offset range
1 mV/div – 100 mV/div
±1 V
101 mV/div – 1 V/div
1.01 V/div – 10 V/div
Ranges, Offset (TDS 794D only) 10 mV/div – 50 mV/div
50.5 mV/div – 99.5 mV/div
100 mV/div – 500 mV/div
500 mV/div – 1 V/div Range, Position ±5 divisions Range, 1 M Sensitivity Range, 50 Sensitivity 1 mV/div to 1 V/div2 (10 mV/div – 1V/div on TDS 794D)
1
Displayed vertically with 25 digitization levels (DLs) per division and 10.24 divisions dynamic range with zoom off. A DL is the smallest voltage level change of the oscilloscope input that can be resolved by the 8-bit A-D Converter. Expressed as a voltage, a DL is equal to 1/25 of a division times the volts/division setting.
2
The sensitivity ranges from 1 mV/div to 10 V/div (for 1 M) or to 1 V/div (for 50 ) in a 1–2–5 sequence of coarse settings with Fit-to-Screen off. Between coarse settings, the sensitivity can be finely adjusted with a resolution equal to 1% of the more sensitive coarse setting. For example, between 50 mV/div and 100 mV/div, the volts/division can be set with 0.5 mV resolution.
3
1 M not available on TDS 794D
3
1 mV/div to 10 V/div
2
±10 V ±100 V
±0.50 V ±0.25 V ±5 V ±2.5 V
TDS 500D, TDS 600B and TDS 700D Service Manual
1–11
Nominal Traits
T able 1–4: Nominal traits — Time base system
Name Description
Range, Sample-Rate
1,3
TDS 684B: 5 Samples/sec to 5 GSamples/sec on four channels simultaneously TDS 680B: 5 Samples/sec to 5 GSamples/sec on two channels simultaneously TDS 644B: 5 Samples/sec to 2.5 GSamples/sec on four channels simultaneously TDS 620B: 5 Samples/sec to 2.5 GSamples/sec on two channels simultaneously TDS 520D, and 724D: 5 Samples/sec to 2 GSamples/sec when acquiring 1 channel,
to 1 GSample/sec when acquiring 2 channels TDS 540D: 5 Samples/sec to 2 GSamples/sec when acquiring 1 or 2 channels,
to 1 GSample/sec when acquiring 3 or 4 channels TDS 540D Opt. 1G: 5 Samples/sec to 1 GSample/sec when acquiring 1 to 4 channels TDS 754D: 5 Samples/sec to 2 GSamples/sec when acquiring 1 or 2 channels,
to 1 GSample/sec when acquiring 3 or 4 channels TDS 754D Opt 1G: 5 Samples/sec to 1 GSample/sec when acquiring 1 to 4 channels TDS 580D, 784D, and 794D: 5 Samples/sec to 4 GSamples/sec when acquiring 1
channel, to 2 GSamples/sec when acquiring 2 channels, or to 1 GSample/sec when acquiring 3 or 4 channels
Range, Interpolated Waveform Rate
Range, Seconds/Division TDS 600B: 0.2 ns/div to 10 s/div
2,3
TDS 600B: 10 GSamples/sec to 250 GSamples/sec TDS 520D, 540D, 724D, and 754D: 1 GSample/sec to 100 GSamples/sec TDS 580D, 784D, and 794D: 2 GSamples/sec to 250 GSamples/sec
TDS 500D, 724D, and 754D: 0.5 ns/div to 10 s/div TDS 580D, 784D, and 794D: 0.2 ns/div to 10 s/div
1–12
TDS 500D, TDS 600B and TDS 700D Service Manual
T able 1–4: Nominal traits — Time base system (cont.)
Name Description
Record Length Selection TDS 500D & TDS 700D:
500, 1,000, 2,500, 5,000, 15,000 and 50,000 samples
In addition to the record lengths previously listed, the following record lengths are
available with the following options:
TDS 520D & TDS 724D with option 1M:
1-channel: up to 250,000 samples
2-channels: up to 130,000 samples
TDS 520D & TDS 724D with option 2M:
1-channel: up to 4,000,000 samples
2-channels: up to 2,000,000 samples
TDS 540D, TDS 580D, TDS 754D, TDS 784D & TDS 794D with option 1M:
1-channel: up to 500,000 samples
2-channels: up to 250,000 samples
3 or 4-channels up to 130,000 samples
Nominal Traits
TDS 540D, TDS 580D, TDS 754D, TDS 784D & TDS 794D with option 2M:
1-channel: up to 8,000,000 samples
2-channels: up to 4,000,000 samples
3 or 4-channels up to 2,000,000 samples
1
The range of real-time rates, expressed in samples/second, at which a digitizer samples signals at its inputs and stores the samples in memory to produce a record of time-sequential samples.
2
The range of waveform rates for interpolated (or equivalent-time on the TDS 700D) waveform records.
3
The Waveform Rate (WR) is the equivalent sample rate of a waveform record. For a waveform record acquired by real-time sampling of a single acquisition, the waveform rate is the same as the real-time sample rate; for a waveform created by interpolation of real-time samples from a single acquisition or, on applicable products, the equivalent-time sampling of multiple acquisitions, the waveform rate created is faster than the real time sample rate. For all these cases, the waveform rate is 1/(Waveform Interval) for the waveform record, where the waveform interval (WI) is the time between the samples in the waveform record.
T able 1–5: Nominal traits — Triggering system
Name Description
Range, Delayed Trigger Time Delay 16 ns to 250 s Range, Events Delay TDS 600B; 2 to 10,000,000
TDS 500D/700D: 1 to 10,000,000 Range (Time) for Pulse-Glitch,
Pulse-Width, Time-Qualified Runt, Timeout, or Slew Rate Trigger, Delta Time
1 ns to 1 s
TDS 500D, TDS 600B and TDS 700D Service Manual
1–13
Nominal Traits
or
imeSet
iolatio
er
evel or
reshol
T able 1–5: Nominal traits — Triggering system (cont.)
Name Description
Ranges, Setup and Hold f T
Ranges, Trigger L
Video Trigger Modes of Operation (Option 05 Video Trigger)
up/Hold V
n Trigg
Th
Feature Min to max
Setup Time Hold Time Setup + Hold Time
For Setup Time, positive numbers mean a data transition before the clock edge and negative means a transition after the clock edge.
For Hold Time, positive numbers mean a data transition after the clock edge and negative means a transition before the clock edge.
Setup + Hold Time is the algebraic sum of the Setup Time and the Hold Time programmed by the user.
d Source Range
Any Channel Auxiliary Line
Supports the following video standards:
H NTSC (525/60) – 2 field mono or 4 field H PAL (625/50) – 2 field mono or SECAM, 8 field H HDTV –
–100 ns to 100 ns –1 ns to 100 ns 2 ns
±12 divisions from center of screen ±8 V ±400 V
1–14
(787.5/60) (1050/60) (1125/60) (1250/60)
H FlexFormat
User can specify: field rate, number of lines, sync pulse width and polarity, line rate, and vertical interval timing.
TM
(user definable standards)
TDS 500D, TDS 600B and TDS 700D Service Manual
T able 1–5: Nominal traits — Triggering system (cont.)
Name Description
Communication Trigger Modes of Operation (Option 2C Comm Trigger)
Standard Name Code
OC1/STM0 NRZ 51.84 Mb/s
OC3/STM1 NRZ 155.52 Mb/s
OC12/STM4 NRZ 622.08 Mb/s
DS0 Sgl Masks
DS0 Dbl Masks
DS0 Data Contra Masks
DS0 Timing Masks
E1 AMI 2.048 Mb/s
E2 AMI 8.44 Mb/s
E3 AMI 34.368 Mb/s
E4 CMI 139.26 Mb/s
E5 (CEPT) NRZ 565 Mb/s
STM1E CMI 155.52 Mb/s
DS1 AMI 1.544 Mb/s
DS1A AMI 2.048 Mb/s
DS1C AMI 3.152 Mb/s
DS2 AMI 6.312 Mb/s
DS3/DS3 Rate AMI 44.736 Mb/s
DS4NA CMI 139.26 Mb/s
STS-1 AMI 51.84 Mb/s
STS-3 CMI 155.52 Mb/s
FC133 NRZ 132.8 Mb/s
FC266 NRZ 265.6 Mb/s
FC531 NRZ 531.2 Mb/s
FC1063 NRZ 1.0625 Mb/s
D2 NRZ 143.18 Mb/s
D1 NRZ 270 Mb/s
FDDI NRZ 125 Mb/s
4:2:2 SMPTE 259M-D NRZ 360 Mb/s
10 Base-T NRZ 10 Mb/s
100 Base-T NRZ 125 Mb/s
Gigabit Ethernet NRZ 1.25 Gb/s
Nominal Traits
1
2 2 2 2
Bit Rate
64 kb/s 64 kb/s 64 kb/s 64 kb/s
1
AMI = Alternate Mark Inversion. CMI = Code Mark Inversion. NRZ = Non-return to Zero.
2
These Telecom DS0 standards are automatically selected from the Mask Menu. The trigger uses Pulse/W idth trigger.
T able 1–6: Nominal traits — Display system
Name Description
Video Display 7 inch diagonal, with a display area of 5.04 inches horizontally by 3.78 inches vertically
TDS 520D, 540D, 580D, 620B, and 680B: Monochrome display
TDS 644B, 684B, 724D, 754D, 784D, and 794D: Color display Video Display Resolution 640 pixels horizontally by 480 pixels vertically
TDS 500D, TDS 600B and TDS 700D Service Manual
1–15
Nominal Traits
T able 1–6: Nominal traits — Display system (cont.)
Name Description
Waveform Display Graticule Single Graticule: 401 × 501 pixels, 8 × 10 divisions, where divisions are 1 cm by 1 cm Waveform Display Levels/Colors TDS 520D, 540D, 580D, 620B, and 680B:
Sixteen levels in infinite-persistence or variable persistence display TDS 644B, 684B, 724D, 754D, 784D, and 794D:
Sixteen colors in infinite-persistence or variable persistence display
T able 1–7: Nominal traits — GPIB interface, output ports, and power fuse
Name Description
Interface, GPIB GPIB interface complies with IEEE Std 488-1987 Interface, RS-232 RS-232 interface complies with EIA/TIA 574 (talk only)
Optional on the TDS 520D, 540D, 580D, 620B, and 680B
Interface, Centronics Centronics interface complies with Centronics interface standard
C332-44 Feb 1977, REV A
Optional on the TDS 520D, 540D, 580D, 620B, and 680B Interface, Video VGA video output with levels that comply with EIA RS 343A standard. DB-15 connector Logic Polarity for Main- and Delayed-
Trigger Outputs Fuse Rating Either of two fuses1 may be used: a 0.25I × 1.25I (UL 198.6, 3AG): 6 A FAST, 250 V
1
Each fuse type requires its own fuse cap.
Negative TRUE. High to low transition indicates the trigger occurred.
or a 5 mm × 20 mm (IEC 127): 5 A (T), 250V.
T able 1–8: Nominal traits — Data handling and reliability
Name Description
Time, Data-Retention, Nonvolatile Memory
1, 2
Floppy disk drive 3.5 inch, 720 K or 1.44 Mbyte, DOS 3.3-or-later compatible Internal hard disk drive
(option 2M available on the TDS 500D and 700D)
External data storage Iomega Zip drive compatible
1
The times that reference waveforms, stored setups, and calibration constants are retained.
2
Data is maintained by small lithium-thionyl-chloride batteries internal to the memory ICs. At the time of manufacture, no special disposal requirements were in effect for these batteries as the amount of hazardous material contained was below the regulated threshold. Consult your local waste disposal agency for proper disposal.
1–16
Battery life 5 years
175 Mbyte capacity
TDS 500D, TDS 600B and TDS 700D Service Manual
Nominal Traits
T able 1–9: Nominal traits — Mechanical
Name Description
Cooling Method Forced-air circulation with no air filter. Clearance is required. Refer to the TDS 500D,
TDS 600B & TDS 700D User Manual for minimum clearance dimensions.
Construction Material Chassis parts constructed of aluminum alloy; front panel constructed of plastic laminate;
circuit boards constructed of glass laminate. Cabinet is aluminum and is clad in Tektronix Blue vinyl material.
Weight Standard Oscilloscope
14.1 kg (31 lbs), with front cover.
24.0 kg (53 lbs), when packaged for domestic shipment
Rackmount Oscilloscopes
14.1 kg (31 lbs) plus weight of rackmount parts, for the rackmounted Oscilloscopes (Option 1R).
Rackmount conversion kit
2.3 kg (5 lbs), parts only; 3.6 kg (8 lbs), parts plus package for domestic shipping
Overall Dimensions Standard Oscilloscope
Height: 193 mm (7.6 in), with the feet installed Width: 445 mm (17.5 in), with the handle Depth: 434 mm (17.1 in), with the front cover installed
Rackmount Oscilloscope
Height: 178 mm (7.0 in) Width: 483 mm (19.0 in) Depth: 558.8 mm (22.0 in)
TDS 500D, TDS 600B and TDS 700D Service Manual
1–17
Nominal Traits
1–18
TDS 500D, TDS 600B and TDS 700D Service Manual

W arranted Characteristics

This section lists the various warranted characteristics that describe the TDS 500D, TDS 600B and TDS 700D oscilloscopes. Electrical and environmen­tal characteristics are included.
Warranted characteristics are described in terms of quantifiable performance limits which are warranted.
NOTE. In these tables, those warranted characteristics that are checked in the procedure Performance Verification appear in boldface type under the column
Name.
As stated above, this section lists only warranted characteristics. A list of typical characteristics starts on page 1–27.

Performance Conditions

The performance limits in this specification are valid with these conditions: H The oscilloscope must have been calibrated/adjusted at an ambient tempera-
ture between +20
H The oscilloscope must be in an environment with temperature, altitude,
humidity, and vibration within the operating limits described in these specifications.
H The oscilloscope must have had a warm-up period of at least 20 minutes. H The oscilloscope must have had its signal-path-compensation routine last
executed after at least a 20 minute warm-up period at an ambient temperature within ±5
_ C of the current ambient temperature.
_ C and +30_ C.
TDS 500D, TDS 600B and TDS 700D Service Manual
1–19
Warranted Characteristics
Accuracy, D
age Measureme
Average
Average
Accuracy, O Accuracy, O
TDS 794D Analog Ba
ple
TDS
B
T able 1–10: Warranted characteristics — Signal acquisition system
Name Description
Accuracy, DC Gain TDS 600B: ±1.5% for all sensitivities from 2 mV/div to 10 V/div
± 2.0% at 1 mV/div sensitivity
TDS 500D, 700D: ±1% for all sensitivities from 1 mV/div to 10 V/div with offset from 0 V
to ±100V
TDS 794D: ±1% for all sensitivities from 10 mV/div to 1 V/div over permitted offset range
C Volt
d (using
mode)
nt,
Measurement type DC Accuracy
Average of 16 waveforms
TDS 600B: ±((1.5% × | reading – Net Offset + Offset Accuracy) + (0.06 div × V/div))
TDS 500D, 700D: ±((1.0% × | reading – Net
1
| ) + Offset Accuracy + (0.06 div x V/div))
Offset
1
| )
ffset
All except TDS 794D
ffset
ndwidth, DC-50 Cou
and Bandwidth selection is FULL,
600
Delta volts between any two averages
of 16 waveforms acquired under the
same setup and ambient conditions
TDS 600B: ±((1.5% × | reading | ) + (0.1 div × V/div) + 0.3 mV)
TDS 500D, 700D: ±((1.0% × | reading | ) (0.1 div x V/div) + 0.3 mV)
TDS 600B
Volts/Div setting
Offset accuracy
1 mV/div – 100 mV/div ±((0.2% × | Net Off-
1
set
| ) + 1.5 mV +
(0.6 div x V/div))
101 mV/div – 1 V/div ±((0.25% × | Net Off-
1
set
| ) + 15 mV +
(0.6 div x V/div))
1.01 V/div – 10 V/div
±((0.25% × | Net Off-
1 | )
set
+ 150 mV +
( 0.6 div x V/div))
Volts/Div setting TDS 794D Offset accuracy
10 mV/div – 100 mV/div ±((0.2% × | Net Off-
1
set
| ) + 1.5 mV +
(0.1 div x V/div))
101 mV/div – 1 V/div ±((0.25% × | Net Off-
1
set
| ) + 15 mV +
(0.1 div x V/div))
d
Volts/Div
TDS 620B & 644B Bandwidth
2
10 mV/div – 1 V/div DC – 500 MHz DC – 1 GHz
5 mV/div – 9.95 mV/div DC – 450 MHz DC – 750 MHz
2 mV/div – 4.98 mV/div
DC – 300 MHz DC – 600 MHz
1 mV/div – 1.99 mV/div DC – 250 MHz DC – 500 MHz
+
TDS 500D/700D Offset accuracy
±((0.2% × | Net Off-
1
set
| ) + 1.5 mV +
(0.1 div x V/div)) ±((0.25% × | Net Off-
1
set
| ) + 15 mV +
(0.1 div x V/div)) ±((0.25% × | Net Off-
1
set
| ) + 150 mV +
(0.1 div x V/div))
TDS 680B & 684B Bandwidth
2
1–20
TDS 500D, TDS 600B and TDS 700D Service Manual
T able 1–10: Warranted characteristics — Signal acquisition system (cont.)
Analog Ba
ple
Name Description
Warranted Characteristics
ndwidth, DC-50 Cou and Bandwidth selection is FULL, TDS 500D/700D
d
Volts/Div
TDS 520D, 540D, 724D & 754D Bandwidth
2
TDS 580D & TDS 784D Bandwidth
2
10 mV/div – 1 V/div DC – 500 MHz DC – 1 GHz 5 mV/div – 9.95 mV/div DC – 500 MHz DC – 750 MHz 2 mV/div – 4.98 mV/div DC – 500 MHz DC – 600 MHz 1 mV/div – 1.99 mV/div DC – 450 MHz DC – 500 MHz
TDS 794D
Volts/Div
Bandwidth
2
10 mV/div – 1 V/div DC – 2 GHz
Crosstalk (Channel Isolation) 100:1 at 100 MHz and 30:1 at the rated bandwidth for the channel’s Volt/Div setting,
for any two channels having equal Volts/Div settings
Delay Between Channels, Full Bandwidth
TDS 600B: 100 ps for any two channels with equal Volts/Div and Coupling settings and both channels’ deskew values set to 0
TDS 500D/700D: 50 ps for any two channels with equal Volts/Div and Coupling settings
and both channel deskew values set to 0.0 s. Input Impedance, DC–1 M Coupled 1 M ±0.5% in parallel with 10 pF ±3 pF (DC–50 Coupled only on TDS 794D) Input Impedance, DC–50 Coupled Input Impedance, DC–50 Coupled
TDS 794D Input Voltage, Maximum,
DC–1 M, AC–1 M, or GND Coupled
50 ±1% with VSWR 1.3:1 from DC – 500 MHz, 1.5:1 from 500 MHz – 1 GHz
50 ±1.25% with VSWR 1.5:1 from 100 MHz – 1 GHz, VSWR 1.7:1 from 1 GHz –
2 GHz
TDS 600B: ±300 V CAT II, 400 V peak;
derate at 20 dB/decade above 1 MHz
TDS 500D/700D: ±300 V CAT II, 400 V peak;
derate at 20 dB/decade above 1 MHz
(DC–50 or GND Coupled TDS 794D only)
Input Voltage, Maximum, DC-50 or
5 V
, with peaks ±30 V
RMS
AC–50 Coupled Input Voltage, Maximum, DC-50
5 V
, with peaks ±20 V
RMS
Coupled TDS 794D Lower Frequency Limit, AC Coupled 10 Hz when AC–1 M Coupled; 200 kHz when AC–50 Coupled
1
Net Offset = Offset – (Position × Volts/Div). Net Offset is the nominal voltage level at the oscilloscope input that
3
corresponds to the center of the A-D converter’s dynamic range. Offset Accuracy is the accuracy of this voltage level.
2
The limits given are for the ambient temperature range of 0_C to +30_C. Reduce the upper bandwidth frequencies by 5 MHz for the TDS 600B, by 2.5 MHz for the 500 MHz TDS 500D/700D models, by 5 MHz for the 1 GHz TDS 500D/700D models, and by 10 MHz for the 2 GHz TDS 500D/700D models for each _C above +30_C.
3
The AC Coupled Lower Frequency Limits are reduced by a factor of 10 when 10X passive probes are used.
TDS 500D, TDS 600B and TDS 700D Service Manual
1–21
Warranted Characteristics
Sensitivity, Edge-Type Trigger
pling
1
Widt
lse
or
Pulse Triggeri
e minimum
lse widths
rm
emiimm lse its rerm
2
Pulse-Type triggeri
T able 1–11: Warranted characteristics — Time base system
Name Description Accuracy, Long Term Sample Rate and
Delay Time
TDS 600B: ±100 ppm over any 1 ms interval TDS 500D/700D: ±25 ppm over any 1 ms interval
T able 1–12: Warranted characteristics — T riggering system
Name Description
, Cou
set to “DC”
Trigger source Sensitivity
Any Channel
TDS 620B & 644B: 0.35 division from DC to 50 MHz, increasing to 1 division at 500 MHz
TDS 680B & 684B: 0.35 division from DC to 50 MHz, increasing to 1 division at 1 GHz
TDS 500D, 724D, & 754D: 0.35 division from DC to 50 MHz, increasing to 1 division at 500 MHz
TDS 580D & 784D: 0.35 division from DC to 50 MHz, increasing to 1 division at 1 GHz
TDS 794D: 0.35 division from DC to 50 MHz, increasing to 1 division at 1.5 GHz
Auxiliary
TDS 600B: 250 mV from DC to 50 MHz, increasing to 500 mV at 100 MHz
h, Minimum Pu
ng
Th
pu
widths and transition times
ng.
and Rearm, f
and rea
required for
TDS 500D, 724D, & 754D: 400 mV from DC to 50 MHz, increasing to 750 mV at 100 MHz
TDS 580D & 784D: 250 mV from DC to 50 MHz, increasing to 500 mV at 100 MHz
TDS 794D: 250 mV from DC to 50 MHz, increasing to 500 mV at 100 MHz
For vertical settings > 10 mV/div. and 1 V/div at the BNC input
Minimum pulse
Pulse class
width
Minimum re-arm width
Glitch 1 ns 2 ns + 5% of Glitch Width Setting Runt Time-Qualified
2 ns 2 ns 2 ns TDS 700D: 8.5 ns + 5% of Width Setting
Runt Width 1 ns 2 ns + 5% of Width Upper Limit Setting Timeout 1 ns 2 ns + 5% of Width Upper Limit Setting Slew Rate 600 ps
2
TDS 700D: 8.5 ns + 5% of Delta Time Setting
1–22
TDS 500D, TDS 600B and TDS 700D Service Manual
T able 1–12: Warranted characteristics — T riggering system (Cont.)
Accuracy (Time
e-Glitch or
Pulse-Width Triggering
gic Level
Main- a
Delayed-Trigger
Outp
Outp
age a
Probe
r
Name Description
Warranted Characteristics
) for Puls
Input Signal Sync Amplitude for Stable Triggering, NTSC and PAL modes (Option 05 Video Trigger)
Jitter (Option 05 Video Trigger) 60 ns
1
The minimum sensitivity for obtaining a stable trigger. A stable trigger results in a uniform, regular display triggered on
Time range Accuracy
1 ns to 500 ns
500 ns to 1 s
±(20% of setting + 0.5 ns) ±(100 ns + 0.01% of Setting)
Field selection “Odd”, “Even”, or “All”: 0.6 division to 4 divisions
Field selection “Numeric”: 1 division to 4 divisions (NTSC mode)
on NTSC or PAL signal
p-p
the selected slope. The trigger point must not switch between opposite slopes on the waveform, and the display must not “roll” across the screen on successive acquisitions. The TRIG’D LED stays constantly lighted when the SEC/DIV setting is 2 ms or faster but may flash when the SEC/DIV setting is 10 ms or slower.
2
For Slew Rate Triggering, this is the minimum transition time, defined to be the time the user’s signal spends between the two trigger threshold settings.
T able 1–13: Warranted characteristics — Output ports, probe compensator, and power requirements
Name Description Lo
uts
s,
nd
Characteristic Limits
Vout (HI)
2.5 V open circuit; 1.0 V into a 50 load to ground
Vout (LO)
0.7 V into a load of ≤4 mA; 0.25 V into a 50 load to ground
ut Volt
Compensato
nd Frequency,
Characteristic Limits
Output Voltage
Frequency
0.5 V (base-top) ±1% into a 50 load 1 kHz ±5%
Output Voltage, Signal Out (CH 31) For TDS 600B: 20 mV/division ±20% into a 1 M load;
10 mV/division ±20% into a 50 load
For TDS 500D/700D:
22 mV/division ±20% into a 1 M load; 11 mV/division ±20% into a 50 load
Source Voltage 100 to 240 VAC
, continuous range, CAT II
RMS
Source Frequency 45 Hz to 440 Hz Power Consumption 350 W (450 VA)
1
CH 3 signal out is present at the rear panel if CH 3 (AUX 1 on the TDS 620B or 680B) is selected as the trigger source for the main and/or delayed trigger systems. It is not available when a channel other than CH3 (AUX 1 on the TDS 620B or 680B) is the source for the Video Trigger when Option 05 is installed.
TDS 500D, TDS 600B and TDS 700D Service Manual
1–23
Warranted Characteristics
T able 1–14: Warranted characteristics — Environmental
Name Description
Atmospherics Temperature (no disk in floppy drive):
TDS 600B: Operating: +4_ C to +45_ C Floppy disk drive: Operating: +10_ C to +45_ C
Nonoperating: –22_ C to +60_ C TDS 500D/700D: Operating: +0_ C to +50_ C Floppy disk drive: Operating: +10_ C to +50_ C
Nonoperating: –22_ C to +60_ C
Relative humidity (no disk in floppy drive):
Operating: 20% to 80%, at or below +32_ C, upper limit derates to 30% relative humidity at +45_ C
Nonoperating: 5% to 90%, at or below +31_ C, upper limit derates to 20% relative humidity at 60_ C
Altitude:
To 4570 m (15,000 ft.), operating (excluding hard disk drive) To 3048 m (10,000 ft.), operating (including hard disk drive) To 12190 m (40,000 ft.), nonoperating
Dynamics Random vibration (floppy disk not installed):
0.31 g rms, from 5 to 500 Hz, 10 minutes each axis, operating
3.07 g rms, from 5 to 500 Hz, 10 minutes each axis, nonoperating
,
1
Emissions (TDS 500D/700D)
2
Meets or exceeds the requirements of the following standards:
FCC Code of Federal Regulations, 47 CFR, Part 15, Subpart B, Class A European Community Requirements
EN 55011 Class A Radiated Emissions EN 55011 Class A Conducted Emissions EN 50081–1 EN60555–2 Power Line Harmonic Emissions
,
1
Emissions (TDS 600B)
2
Meets or exceeds the requirements of the following standards:
FCC Code of Federal Regulations, 47 CFR, Part 15, Subpart B, Class A EN 50081–1 European Community Requirements
1–24
EN 55022 Radiated Emissions Class B EN 55022 Class B Conducted Emissions EN60555–2 Power Line Harmonic Emissions
TDS 500D, TDS 600B and TDS 700D Service Manual
Warranted Characteristics
T able 1–14: W arranted characteristics — Environmental (cont.)
Name Description
,
1
Susceptibility
Approvals Conforms to and is certified where appropriate to:
1
VGA output cable needs to be terminated, if connected at all, for the Instrument to meet these standards. The test will pass with LCOM part # CTL3VGAMM–5.
2
The GPIB cable connected to the instrument for certain of the emissions tests must be “low EMI” having a high-quality outer shield connected through a low impedance to both connector housings. Acceptable cables are Tektronix part numbers 012-0991-00, -01, -02, and -03. In order to maintain the EMI performance conforming to the above regulations, the following cables, or their equivalent, should be used: a shielded Centronics cable, 3 meters in length, part number 012-1214-00, and a shielded RS-232 cable, 2.7 meters in length, CA part number 0294-9.
3
UL 3111, CSA 22.2 no.1010 Safety Certification Compliance: Temperature (operating) 5 to +40 C Altitude (maximum operating): 2000 meters Equipment Type: Test and Measurement Safety Class: Class I (as defined in IEC 1010–1, Annex H) – grounded product Overvoltage Category: Overvoltage Category II (as defined in IEC 1010–1, Annex J) Pollution Degree: Pollution Degree 2 (as defined in IEC 1010–1)
2
Meets or exceeds the EMC requirements of the following standards:
EN 50082–1 European Community Requirements
IEC 801-2 Electrostatic Discharge Performance Criteria B
IEC 801-3 Radiated Susceptibility 3 V/meter from 27 MHz to 500 MHz
unmodulated
IEC 801-4 Fast Transients Performance Criteria B
IEC 801-5 AC Surge Performance Criteria B
UL 3111–1 CAN/CSA C22.2 no. 1010.1
3
– Standard for electrical measuring and test equipment
3
– Safety requirements for electrical equipment for
measurement, control and laboratory use
Note – Rated for indoor use only
TDS 500D, TDS 600B and TDS 700D Service Manual
1–25
Warranted Characteristics
T able 1–15: Certifications and compliances
EC Declaration of Conformity (TDS 500D and TDS 700D)
Australian Declaration of Conformity – EMC (TDS 500D and TDS 700D)
EC Declaration of Conformity (TDS 600B)
EC Declaration of Conformity – Low Voltage
Meets intent of Directive 89/336/EEC for Electromagnetic Compatibility. Compliance was demonstrated to the following specifications as listed in the Official Journal of the European Communities:
EMC Directive 89/336/EEC: EN 55011 Class A Radiated and Conducted Emissions EN 50081-1 Emissions:
EN 60555-2 AC Power Line Harmonic Emissions
EN 50082-1 Immunity:
IEC 801-2 Electrostatic Discharge Immunity IEC 801-3 RF Electromagnetic Field Immunity IEC 801-4 Electrical Fast Transient/Burst Immunity IEC 801-5 Power Line Surge Immunity
Conforms with the following standards in accordance with the Electromagnetic Compatibility Framework:
AS/NZS 2064.1/2 Class A radiated and Conducted Emissions Meets intent of Directive 89/336/EEC for Electromagnetic Compatibility. Compliance was
demonstrated to the following specifications as listed in the Official Journal of the European Communities:
EMC Directive 89/336/EEC: EN 50081-1 Emissions:
EN 55022 Class B Radiated and Conducted Emissions EN 60555-2 AC Power Line Harmonic Emissions
EN 50082-1 Immunity:
IEC 801-2 Electrostatic Discharge Immunity IEC 801-3 RF Electromagnetic Field Immunity IEC 801-4 Electrical Fast Transient/Burst Immunity IEC 801-5 Power Line Surge Immunity
Compliance was demonstrated to the following specification as listed in the Official Journal of the European Communities:
1–26
Low Voltage Directive 73/23/EEC EN 61010-1:1993 Safety requirements for electrical equipment for measurement,
control, and laboratory use
TDS 500D, TDS 600B and TDS 700D Service Manual

Typical Characteristics

Analog Ba
ple
243 or P6245 Probe a
Ba
electi
TDS 2 D 4 D 24D 4D Analog Ba
ple
Ba
electi
TDS 8 D TDS 84D Analog Ba
ple
be
Analog Ba
ple
electi
TDS 520D
24D 4D 84D
This subsection contains tables that list the various typical characteristics which describe the TDS 500D, TDS 600B and TDS 700D oscilloscopes.
Typical characteristics are described in terms of typical or average performance. Typical characteristics are not warranted.
T able 1–16: Typical characteristics — Signal acquisition system
Name Description
ndwidth, DC-50 Cou
with P6
ndwidth s
TDS 520D, 540D, 724D & 754D
with P6245 Probe and
ndwidth s
TDS 580D & TDS 784D
with P6217 Pro TDS 794D
on is FULL
ndwidth, DC-50 Cou
on is FULL
ndwidth, DC-50 Cou
nd
Volts/Div as read out on screen 520D, 540D, 724D & 754D Bandwidth
d
10 V/div – 100 V/div Not Applicable 100 mV/div – 10 V/div 50 mV/div – 99.5 mV/div DC – 500 MHz 20 mV/div – 49.8 mV/div DC – 500 MHz 10 mV/div – 19.9 mV/div DC – 450 MHz (P6243)
Volts/Div as read out on screen TDS 580D & TDS 784D
d
10 V/div –100 V/div (Not Applicable) 100 mV/div – 10 V/div 50 mV/div – 99.5 mV/div DC – 750 MHz 20 mV/div – 49.8 mV/div DC – 600 MHz 10 mV/div – 19.9 mV/div DC – 500 MHz
Volts/Div as read out on screen TDS 794D
d
100 mV/div – 10 V/div
1
DC – 500 MHz
DC – 500 MHz (P6245)
DC – 1 GHz
DC – 2 GHz
ndwidth, DC-1 M Cou with P6139A Probe and Bandwidth s
on is FULL
, 540D, 580D,
724D, 754D & 784D
TDS 500D, TDS 600B and TDS 700D Service Manual
d
Volts/Div as read out on screen
10 V/div – 100 V/div 500 MHz 100 mV/div – 10 V/div 50 mV/div – 99.5 mV/div 500 MHz 20 mV/div – 49.8 mV/div 500 MHz 10 mV/div – 19.9 mV/div 500 MHz
TDS 520D, 540D, 724D, 754D & 784D Bandwidth
500 MHz
1
1–27
Typical Characteristics
Analog Ba
ple
TDS 680B
84B) a
Ba
electi
TDS B
Calculate
ise Time
2
T able 1–16: Typical characteristics — Signal acquisition system (cont.)
Name Description
ndwidth, DC-50 Cou
d with P6139A Probe (TDS 620B & 644B) or P6245 Probe (
ndwidth s
on is FULL
& 6
TDS 600B
TDS 620B & 644B
Volts/Div as read out on screen
nd
10 V/div – 100 V/div (Not Applicable) (Not Applicable) 100 mV/div – 10 V/div
Bandwidth
DC – 500 MHz DC – 1 GHz
1
TDS 680B & 684B Bandwidth
1
50 mV/div – 99.5 mV/div DC – 450 MHz DC – 750 MHz 20 mV/div – 49.8 mV/div DC – 300 MHz DC – 600 MHz 10 mV/div – 19.9 mV/div DC – 250 MHz DC – 500 MHz
Accuracy, Delta Time Measurement The limits are given in the following table for signals having amplitude greater than
5 divisions, reference level = 50%, filter set to (sinX/X), acquired at 5 mV/div or greater. For the TDS 700D, pulse duration < 10 div. Channel skew not included.
For the Single Shot condition, 1.4 T S
is the sample interval and Tr is the displayed rise time.
i
TDS 600B: For the averaged condition, 1.4 T
4, where
r/Si
40, where Wi is the Waveform
r/Wi
Interval, as described elsewhere in these specifications. TDS 600B: Extra error in the measurement will occur for two-channel measurements due
to channel-to-channel skew . This is described elsewhere in these specifications.
Time measurement accuracy
Conditions for accuracy listed at right are: Single Shot or Sample mode (or HiRes mode on the TDS 500D/700D), with Full Bandwidth selected.
TDS 600B:±( (0.20 × sample interval) + (100 ppm × | Reading |) + (0.05 × W TDS 600B example: at 5 GS/s, 5 ns/div, measuring a 40 ns wide pulse,
accuracy = ±(40 ps + 4 ps + 5 ps) = ±49 ps.
i
) )
TDS 500D/700D: ±0.15 sample interval + (25 ppm × | Reading |) + t/div/1000
Conditions for accuracy listed at right are: 100 Averages, with Full Band­width selected, and for TDS 500D/700D, repetitive mode.
d R
, TDS 600B
1–28
TDS 500D/700D example: at 4 Gs/s, accuracy = 37.5 ps TDS 600B: ±(10 ps + (100 ppm × | Reading |) + (0.25 × W
i
) )
TDS 500D/700D: 20 ps + (25 ppm × | Reading |) + t/div/1000
Volts/Div setting
620B & 644B Rise time
680B & 684B Rise time
10 mV/div – 1 V/div 900 ps 450 ps 5 mV/div – 9.95 mV/div 1 ns 600 ps 2 mV/div – 4.98 mV/div 1.5 ns 750 ps 1 mV/div – 1.99 mV/div 1.8 ns 900 ps
TDS 500D, TDS 600B and TDS 700D Service Manual
T able 1–16: Typical characteristics — Signal acquisition system (cont.)
Calculate
ise Time
Calculate
ise Time
2
ective Bits
ective Bits
effective bits for
ste
to
ivisions at
ective Bits
cy
effective bits for
ste
to
ivisions at
ective Bits
cy
e ective its or si e ve ste
Name Description
Typical Characteristics
d R
TDS 500D/700D
d R
TDS 794D Eff
Eff
Eff
— TDS 600B Input frequency Effective bits
The chart on the right gives the typical effective bits for a 9-division p-p sine-wave input, 50 mV/div, 10 ns/div (5 GS/s), with a record length of 1000 points
— TDS 520D & 724D Sample rate
The chart on the right gives the typical
9.2 d
25° C.
— TDS 540D & 754D
,
2
,
a sine wave adju
1 MHz, 50 mV/div @
520D, 540D, 724D,
Volts/Div setting
10 mV/div – 1 V/div 800 ps 400 ps 5 mV/div – 9.95 mV/div 800 ps 530 ps 2 mV/div – 4.98 mV/div 800 ps 600 ps 1 mV/div – 1.99 mV/div 890 ps 800 ps
Volts/Div setting 794D Rise time
10 mV/div – 1 V/div 200 ps
98 MHz 245 MHz 490 MHz 990 MHz
Input frequency
d
1 MHz – 9.2 divs 6.8 bits 9.7 bits 490 MHz – 6.5 divs
& 754D Rise time
6.3 bits
6.0 bits
5.5 bits
5.2 bits (TDS 680B & 684B only)
1 GS/s 10 MS/s & HiRes
6.5 bits N/A
Sample rate
580D & 784D Rise time
The chart on the right gives the typical
a sine wave adju
9.2 d
25° C.
Eff
Frequency Limit, Upper, 250 MHz Bandwidth Limited
Frequency Limit, Upper, 20 MHz Bandwidth Limited
— TDS 580D, 784D & 794D
The chart on the right gives the typical effective bits for a sine wave adjusted to 9.2 divisions at 1 MHz, 50 mV/div @ 25° C.
1 MHz, 50 mV/div @
Input frequen
d
1 MHz – 9.2 divs 6.8 bits 9.7 bits 500 MHz
Input frequen
1 MHz – 9.2 divs 6.6 bits 9.7 bits 1 GHz – 6.5 divs 5.5 bits N/A 2 GHz – 6.5 divs (TDS 794D only) 4.5 bits N/A 250 MHz
20 MHz
2 GS/s 10 MS/s & HiRes
6.8 bits N/A
Sample rate
4 GS/s 10 MS/s & HiRes
TDS 500D, TDS 600B and TDS 700D Service Manual
1–29
Typical Characteristics
ettli
rrors
Div setting
pli
e
ettli
rrors
Accuracy, Trigger Level or Threshol
ple
Positio
rror
eri
T able 1–16: Typical characteristics — Signal acquisition system (cont.)
Name Description
Step Response S
Step Response S TDS 794D
1
The limits given are for the ambient temperature range of 0_C to +30_C. Reduce the upper bandwidth frequencies by 5 MHz for the TDS 600B or by 2.5 MHz for the TDS 500D/700D for each _C above +30_C.
2
The numbers given are valid 0_C to +30_C and will increase as the temperature increases due to the degradation in bandwidth. Rise time is calculated from the bandwidth. It is defined by the following formula:
TDS 600B Rise Time (ns)
Note that if you measure rise time, you must take into account the rise time of the test equipment (signal source, etc.) that you use to provide the test signal. That is, the measured rise time (RT the rise time of the test signal source (RTgen) according to the following formula:
RT
m
3
The values given are the maximum absolute difference between the value at the end of a specified time interval after the midlevel crossing of the step and the value one second after the midlevel crossing of the step, expressed as a percentage of the step amplitude.
2
RT
ng E
ng E
i
2
RT
gen
2
Settling error (%)
Volts/
± Step am
tud
20 ns 100 ns 20 ms
1 mV/div – 100 mV/div 2 V 0.5% 0.2% 0.1% 101 mV/div – 1 V/div 20 V 1.0% 0.5% 0.2%
1.01 V/div – 10 V/div 200 V 1.0% 0.5% 0.2% 10 mV/div – 100 mV/div 1.5 V 0.5% 0.2% 0.1% 101 mV/div – 1 V/div
450
BW (MHz)
TDS 500D700D Rise Time (ns)
3 V ≤1.0% ≤0.5% ≤0.2%
400
BW (MHz)
) is determined by the instrument rise time (RTi) and
m
3
at
T able 1–17: Typical characteristics — Triggering system
Name Description
d, DC
Trigger source Accuracy
Cou
d
(for signals having rise and fall times 20 ns)
Input, Auxiliary Trigger
Trigger Edge Trigg
ng
n E
,
Holdoff, V ariable, Main Trigger
1–30
Any Channel
Auxiliary The input resistance is 1.5 kW; the maximum safe input voltage is
±20 V (DC + peak AC).
Acquisition mode Trigger-position error
Sample, Average Envelope
For all Time/Division ranges, the minimum holdof f is 250 ns and the maximum holdof f is 12 seconds. The minimum resolution is 8 ns for settings 1.2 ms.
±((2% × | Setting – Net Offset |) + (0.3 div × Volts/div Setting ) + Offset Accuracy)
Not calibrated or specified
1,2
±(1 Waveform Interval + 1 ns) ±(2 Waveform Intervals + 1 ns)
TDS 500D, TDS 600B and TDS 700D Service Manual
T able 1–17: Typical characteristics — Triggering system (cont.)
sitivity
Not DC Couple
3
Widt
lse
or
Pulse Triggeri
emiimm lse its rerm
8
Pulse-Type triggeri
Name Description
Typical Characteristics
Lowest Frequency for Successful Operation of “Set Level to 50%” Function
Sen
, Edge Trigger,
d
Sensitivities, Logic Trigger and Events Delay , DC Coupled
Sensitivities, Pulse-T ype Runt Trigger
4
5
Sensitivities, Pulse-T ype Trigger Width and
6
Glitch Width, Minimum Pulse and Rearm, for Logic
Triggering or Events Delay
h, Minimum Pu
and Rearm, f
ng
The minimum pulse widths and rearm widths and transition times
required for
ng.
30 Hz
Typical signal level for stable
Trigger source
AC
triggering
Same as the DC-coupled limits for frequen­cies above 60 Hz. Attenuates signals below
60 Hz. Noise Reject High Frequency Reject
Three times the DC-coupled limits.
One and one-half times the DC-coupled
limits from DC to 30 kHz. Attenuates signals
above 30 kHz. Low Frequency Reject
One and one-half times the DC-coupled
limits for frequencies above 80 kHz.
Attenuates signals below 80 kHz.
1.0 division, from DC to 500 MHz, at vertical settings > 10 mV/div and 1 V/div at the BNC input
1.0 division, from DC to 500 MHz, at vertical settings > 10 mV/div and 1 V/div at the BNC input
1.0 division, at vertical settings > 10 mV/div and 1 V/div at the BNC input
For vertical settings > 10 mV/div and 1 V/div at the BNC input
Triggering type
Minimum pulse width
Minimum re-arm
width
Minimum time between channels
Logic Not Applicable 1 ns 1 ns Events Delay 1 ns (for either + or
Not Applicable 2 ns
– pulse widths)
For vertical settings > 10 mV/div. and 3 1 V/div at the BNC input
Minimum pulse
Pulse class
width
Minimum re-arm width
Glitch 1 ns 2 ns + 5% of Glitch Width Setting Runt
2 ns 2 ns Time-Qualified Runt 2 ns TDS 600B: 7 ns + 5% of Width Setting Width 1 ns 2 ns + 5% of Width Upper Limit Setting Timeout 1 ns 2 ns + 5% of Width Upper Limit Setting Slew Rate 600 ps
8
TDS 600B: 7 ns + 5% of Delta Time Setting
7
TDS 500D, TDS 600B and TDS 700D Service Manual
1–31
Typical Characteristics
T able 1–17: Typical characteristics — Triggering system (cont.)
Name Description
Setup/Hold Time Violation Trigger, Minimum Clock Pulse Widths
Input Signal Sync Amplitude for Stable
For vertical settings > 10 mV/div and 1 V/div at the BNC input, the minimum requirements are:
Minimum Pulse Width, Clock High
User Hold Time + 2.5 ns
9
Minimum Pulse Width, Clock Low
2 ns
All field selections: 0.6 division to 4 divisions
Triggering, HDTV and FLEXFMT modes (Option 05 Video Trigger)
Jitter for HDTV mode
17 ns
p-p
(Option 05 Video Trigger)
Sync Width Flex Format and HDTV
min. 400 ns
modes (Option 05 Video Trigger) Sync Duty Cycle, Flex Format and HDTV
min. 50 to 1
modes (Option 05 Video Trigger) Hum Rejection
(Option 05 Video Trigger)
1
The trigger position errors are typically less than the values given here. These values are for triggering signals having a
NTSC and PAL: –20 dB without any trigger spec deterioration. Triggering will continue down to 0 dB with some performance deterioration.
slew rate at the trigger point of 0.5 division/ns.
2
The waveform interval (WI) is the time between the samples in the waveform record. Also, see the footnote for the characteristics Sample Rate Range or Interpolated Waveform Rates in Table 1–4, on page 1–12.
3
The minimum sensitivity for obtaining a stable trigger. A stable trigger results in a uniform, regular display triggered on the selected slope. The trigger point must not switch between opposite slopes on the waveform, and the display must not “roll” across the screen on successive acquisitions. The TRIG’D LED stays constantly lighted when the SEC/DIV setting is 2 ms or faster but may flash when the SEC/DIV setting is 10 ms or slower.
4
The minimum signal levels required for stable logic or pulse triggering of an acquisition, or for stable counting of a DC-coupled, events-delay signal. Also, see the footnote for Sensitivity, Edge-Type T rigger, DC Coupled in this table. (Stable counting of events is counting that misses no events and produces no extra, phantom events).
5
The minimum signal levels required for stable runt pulse triggering of an acquisition. Also, see the footnote for Sensitivity, Edge-Type Trigger, DC Coupled in this table. (Stable counting of events is counting that misses no events).
6
The minimum signal levels required for stable pulse width or glitch triggering of an acquisition. Also, see the footnote for Sensitivity, Edge-T ype T rigger, DC Coupled in this table. (Stable counting of events is counting that misses no events).
7
For Logic, time between channels refers to the length of time a logic state derived from more than one channel must exist to be recognized. For Events, the time is the minimum time between a main and delayed event that will be recognized if more than one channel is used.
8
For Slew Rate Triggering, this is the minimum transition time, defined to be the time the user’s signal spends between the two trigger threshold settings.
9
User Hold Time is the number selected by the user in the Hold Time Menu.
1–32
TDS 500D, TDS 600B and TDS 700D Service Manual

Installation

Supplying Operating Power

NOTE. Read all information and heed all warnings in this subsection before connecting the oscilloscope to a power source.
WARNING. AC POWER SOURCE AND CONNECTION. The oscilloscope operates from a single-phase power source. It has a three-wire power cord and two-pole, three-terminal grounding type plug. The voltage to ground (earth) from either pole of the power source must not exceed the maximum rated operating voltage, 250 volts.
Before making connection to the power source, be sure the digitizing oscilloscope has a suitable two-pole, three-terminal grounding-type plug.
GROUNDING. This instrument is safety Class 1 equipment (IEC designation). All accessible conductive parts are directly connected through the grounding conductor of the power cord to the grounded (earthing) contact of the power plug.
WARNING. The power input plug must be inserted only in a mating receptacle with a grounding contact where earth ground has been verified by a qualified service person. Do not defeat the grounding connection. Any interruption of the grounding connection can create an electric shock hazard.
For electric shock protection, the grounding connection must be made before making connection to the instrument’s input or output terminals.
Power Cord Information
TDS 500D, TDS 600B and TDS 700D Service Manual
A power cord with the appropriate plug configuration is supplied with each oscilloscope. Table 2–1 gives the color-coding of the conductors in the power cord. If you require a power cord other than the one supplied, refer to Table 2–2, Power Cord Identification.
2–1
Installation
T able 2–1: Power cord conductor identification
Conductor Color Alternate Color
Ungrounded (Line) Brown Black Grounded (Neutral) Light Blue White Grounded (Earthing) Green/Yellow Green
T able 2–2: Power cord identification
Plug configuration Normal usage Option number
North America 125 V
Europe 230 V
United Kingdom 230 V
Australia 230 V
North America 230 V
Switzerland 230 V
Standard
A1
A2
A3
A4
A5
2–2
Operating Voltage
This oscilloscope operates with any line voltage from 90–250 VAC
RMS
with any line frequency from 45–440 Hz. There are two fuses, either of which may be used throughout the line voltage and frequency ranges. (The two fuses are not totally interchangeable as each requires a different fuse cap.)
TDS 500D, TDS 600B and TDS 700D Service Manual
Installation
Memory Backup Power

Operating Environment

Operating Temperature
Memory modules with on-board batteries allow the TDS 500D, TDS 600B and TDS 700D Oscilloscopes to retain some types of data upon loss of the AC power source. The stored adjustment constants, saved front-panel settings, current front-panel settings (instrument status), and waveforms saved in memory are retained.
The on-board batteries of the memory modules have a shelf life of about five years. Partial or total loss of stored settings upon power on may indicate that the memory modules need to be replaced.
The following environmental requirements are provided to ensure proper operation and long instrument life.
Operate oscilloscopes where the ambient air temperature is from 4_ C to +45_ C (+50_ C for the TDS 500D and 700D) with no diskette in the floppy drive. Store the oscilloscopes in ambient temperatures from –22_ C to +60_ C with no diskette in the floppy drive. After storage at temperatures outside the operating limits, allow the chassis to stabilize at a safe operating temperature before applying power.
Ventilation Requirements
The oscilloscopes are cooled by air drawn in and exhausted through their cabinet side panels by an internal fan. To ensure proper cooling of the oscilloscope, allow at least two inches clearance on both sides and 3/4 inch on the bottom of the oscilloscope. (The feet on the bottom of the oscilloscope provide the required clearance when set on flat surfaces.) The top of the oscilloscope does not require ventilation clearance.
CAUTION. If air flow is restricted, the digitizing oscilloscope’s power supply may temporarily shut down.
TDS 500D, TDS 600B and TDS 700D Service Manual
2–3
Installation

Applying and Interrupting Power

Consider the following information when you power on or power off the instrument, or when power is interrupted due to an external power failure.
Power On
Power Off
Upon power on, the oscilloscope runs a power-on self check. If it passes, the oscilloscope displays a “passed” status message and a prompt to press CLEAR MENU to continue. If it fails, the oscilloscope displays a diagnostic log that identifies the area(s) that failed and a prompt to press CLEAR MENU to continue. See Section 6, Maintenance, for information on diagnostics and fault isolation.
CAUTION. DO NOT power off the oscilloscope when either running a signal path compensation or when doing any of the adjustments described in Section 5, Adjustment Procedures. To do so might result in the loss of internally stored adjustment constants.
In general, do not power off the instrument when doing operations that affect the data types listed in Table 1–1. Wait for the instrument to finish the operation when doing adjustments, saving waveforms, or saving setups.
Improper power off or unexpected loss of power to the oscilloscope can result in the following corruptions of nonvolatile RAM (NVRAM). The following table describes the messages displayed when power is restored after an abnormal power off.
2–4
T able 2–3: Effects of corrupted data
Corrupted Data Type Results
Adjustment Constants:
Signal Path Compensation A signal path compensation is required Voltage Reference A voltage reference adjustment is required
(Section 5)
Frequency Response A frequency response adjustment is required
(Section 5) Error Log Errors logged are lost Reference Waveforms W aveform Lost Saved Setups Setup Lost
TDS 500D, TDS 600B and TDS 700D Service Manual

Repackaging Instructions

Use a corrugated cardboard shipping carton having a test strength of at least 275 pounds and with an inside dimension at least six inches greater than the instrument dimensions.
If the instrument is being shipped to a Tektronix Service Center, enclose the following information: the owner’s address, name and phone number of a contact person, type and serial number of the instrument, reason for returning, and a complete description of the service required.
Seal the shipping carton with an industrial stapler or strapping tape. Mark the address of the Tektronix Service Center and also your own return
address on the shipping carton in two prominent locations.

Installed Options

Your instrument may be equipped with one or more instrument options. Except for the line-cord options described by Table 2–2 (on page 2–2 of this section), all options and optional accessories are listed and described in Section 7, Options. For further information and prices of instrument options, see your Tektronix Products catalog or contact your Tektronix Field Office.
Installation
TDS 500D, TDS 600B and TDS 700D Service Manual
2–5
Installation
2–6
TDS 500D, TDS 600B and TDS 700D Service Manual

Operating Instructions

Before doing service, read the following operating instructions. These instruc­tions are at the level appropriate for servicing these oscilloscopes. The User Manual contains more complete operating instructions.
Additional instructions are integrated into the service procedures found in later sections of this manual. For instance, the procedures found in the section Performance Verification contain instructions for making the front-panel settings required to check each instrument characteristic included there. Also, the general instructions for operating these oscilloscopes’ internal diagnostic routines are found in Section 6, Maintenance. You may also find the Product Description in Section 1 useful for understanding how the oscilloscope functions.

Screen Layout

Figure 2–1 on page 2–8 shows the screen layout. This figure illustrates a full graticule. You may also select a grid, crosshair, or frame graticule from the display menu.

Basic Procedures

How to Power On
Push the principal power switch found on the rear panel of the oscilloscope, then push the ON/STBY (standby) switch to toggle the oscilloscope into operation. The switch at the rear panel is the true power disconnect switch. The ON/ STBY(standby) switch simply toggles operation on and off.
WARNING. The principal power switch at the rear panel is the true power disconnect switch. The ON/STBY (standby) switch simply toggles operation on and off. When connected to a power source and when the principal power switch is on, the internal power supplies and much of the other circuitry of these oscilloscopes are energized regardless of the setting of the ON/STBY switch.
TDS 500D, TDS 600B and TDS 700D Service Manual
2–7
Operating Instructions
Brief status information
Graticule and waveforms
Waveform reference
symbols shows ground
levels and waveform
sources
Vertical scale, horizontal scale,
and trigger level readouts
Pop-up menu
Position of waveform
record relative to
the screen and display
General purpose
knob readout
Side menu area. Readouts for measurements move here when CLEAR MENU is pressed.
How to Use Help
How to Use the Status
Menu
Main menu display area. Readouts in lower graticule
area move here when CLEAR MENU is pressed.
Figure 2–1: Map of Display Functions
Push the HELP front-panel button to enter help mode. Front-panel knobs and buttons now display information about their function when turned or pushed. Push HELP again to exit help mode.
To get help information on a menu item, display the menu desired (if you are in help mode, exit help first). Push HELP. Now the menu buttons display informa- tion about their function when pushed.
Push the SHIFT, then the STATUS front-panel buttons to display the status menu. You will find messages reflecting the state of the acquisition system, whether it is running or stopped (and if it is stopped, why), as well as setup-related information.
2–8
TDS 500D, TDS 600B and TDS 700D Service Manual
Operating Instructions
How to Set Functions
Special function controls.
Access all functions except
vertical, horizontal, and
trigger functions from this
front-panel control block.
Most functions can be set either by using one (or two) front-panel button(s) or knob(s), or by pushing a front-panel button to use a main menu, and then a side-menu button to set the function. The following steps illustrate both procedures.
1. Locate the block that contains the function to be set.
Vertical function controls. Access all
vertical functions and inputs from
this front-panel control block.
Horizontal function controls.
Access all horizontal functions
from this front-panel block.
Trigger function controls. Access all trigger functions from this front-panel block.
TDS 500D, TDS 600B and TDS 700D Service Manual
2–9
Operating Instructions
2. Select the waveform source(s). Position, scale, and set trigger level for
waveform source(s) directly from the front panel. The steps that follow provide examples of possible control selections.
8
7
123
4
2–10
65
a. Input waveforms into these channels (7). Example: CH 1. b. Push any channel’s button (8) to display its waveform. The last channel
turned on determines which waveform is positioned and scaled. The indicator above the channel last selected is lighted. Example: Push CH 1; then CH 2.
c. Vertically (1) and horizontally (2) scale and position the waveform(s)
selected. Example: Set the scale to 100 mV/div and center the waveform on screen.
d. Stop and start acquiring waveforms (3). Example: Push RUN/STOP if
not acquiring.
TDS 500D, TDS 600B and TDS 700D Service Manual
Operating Instructions
e. Adjust trigger level (6) to trigger the waveform(s) selected or use these
buttons (5) to either set a trigger level at the mid-amplitude level of the selected waveform or to force a single trigger. Example: Push SET
LEVEL TO 50%.
3. Set all other functions using menus. a. Choose the waveform source (8) first if setting a vertical function; else
skip to step b. Example: Push CH 2.
b. Push SHIFT (4) if the function to be set is highlighted in blue; else skip
to step c.
9
14
13
10 11
12
c. Push the front-panel button that corresponds to the menu containing the
function. A main menu (14) for selecting among related functions appears. Example: Push VERTICAL MENU.
Note the two labels: the top label is a function to choose from; the bottom label tells you the current setting for that function. Offset is currently set to 0 V.
d. Select a function from the main menu using the main-menu buttons (12).
A side menu for selecting among that the available settings for that function appears. Example: Push Coupling (13).
TDS 500D, TDS 600B and TDS 700D Service Manual
2–11
Operating Instructions
e. Select the setting for the function from the side menu (9) using the
side-menu buttons (11). Example: Push AC (10).
How to Set Complex
Functions
18
A few functions require more than just two levels (main and side) of menus to completely specify their settings. In such cases, either the main menu, the side menu, or both are modified to supply additional choices. The procedures that follow show both schemes.
1. Set up a function using pop-up menus:
15
17
16
a. For some selections, pushing a main-menu button pops up a menu (18)
of subfunctions. Example: Push Type (17). Note the pop-up menu for Type is set to Edge. All the main-menu
buttons to the right of the pop-up menu are labeled with subfunctions of
Edge.
b. Pushing the button that popped up the menu (17) toggles through the
pop-up menu choices. Example: Repeatedly push Type to move through the pop-up menu. Notice the other main-menu button labels change accordingly. Push Type to move back to Edge.
c. Complete the setting of the desired mode by selecting from the main
menu and the side menu that results. Example: Push Mode & Holdoff (16), and then push Normal (15).
2–12
TDS 500D, TDS 600B and TDS 700D Service Manual
Operating Instructions
2. Set up a function using the general purpose knob (20). (The examples of possible menu selections in the substeps that follow assume you have pushed TRIGGER MENU.)
19 20
21
a. Pushing some main-menu buttons displays a side menu with labels
containing readouts that can be varied. Example: Push Level (21).
b. Pushing the side-menu button Level assigns the general purpose knob to
control the readout appearing in the button label. It also copies the readout to the general purpose knob readout area in the right corner of the screen. Example: Push Level (19).
c. Use the general purpose knob (20) to adjust the trigger level to the
setting desired. Example: Turn the knob to –20 mV.
More About the General Purpose Knob. As you’ve just seen, the general purpose knob can be used to extend the number of choices available to a side menu button. The general purpose knob can also be assigned to control the following functions:
H Cursor positioning H Display intensities H Delay time
TDS 500D, TDS 600B and TDS 700D Service Manual
2–13
Operating Instructions
H Gated measurements H Number of events H Template generation H Trigger position H Holdoff H Offset H Variable persistence
In all cases, the menus are used to select the function to which the general purpose knob is assigned. The following attributes apply to this knob:
H Depending on the function it is assigned to control, the general purpose knob
may vary numerical readouts, position objects (cursors) on screen, or select between icon-label settings that show up in side-menu labels.
H The general purpose knob has a readout area at the upper right corner of the
screen. (See Figure 2–1.) This readout always reflects the name and value of the function that the general purpose knob is currently controlling.
H Whenever the general purpose knob assignment is changed, a knob icon
appears immediately to the left of the general purpose knob readout to notify you of the assignment change. The icon is removed as soon as you use the general purpose knob to change the value of the function it is assigned to.
H To assign the general purpose knob to control a function, display the menu
containing the function; then select the function. (Note that not all functions can be controlled by the general purpose knob.)
H Whenever the menu is removed, the general purpose knob is not assigned
and doesn’t control any function. (An exception is the cursor function. If cursors are turned on, removing the menu leaves the knob assigned to control the cursors until reassigned by selecting another menu and function that uses the knob.)
H When the SHIFT button is lighted, the general purpose knob becomes a
coarse adjustment. When the SHIFT button is not lighted, the general purpose knob becomes a fine adjustment.
H The general purpose knob also has a SELECT button. This button is used to
toggle the knob between the control of either of the two cursors displayed when H-bar or V-bar cursors are turned on in the cursor menu.
2–14
TDS 500D, TDS 600B and TDS 700D Service Manual
Operating Instructions
Display and Utility Menus. Using the techniques described for using menus, you can access and change functions in the display menu and utilities menu. In the Display menu, you can set the following functions:
H Color: waveforms, readouts, graticule H Intensity: waveforms, readouts, graticule, etc. H Style of waveform display(s): vectors or dots, intensified or nonintensified
samples, and infinite or variable persistence
H Display format: XY or YT H Graticule format: type H Waveform interpolation filter and readout options
From the Utility menu, you can configure the GPIB port (talk/listen, address, etc.) and access internal routines for self diagnostics and self compensation. Instructions for setting up communication over the GPIB are found in Section 5, Adjustment Procedures.
TDS 500D, TDS 600B and TDS 700D Service Manual
2–15
Operating Instructions
2–16
TDS 500D, TDS 600B and TDS 700D Service Manual

Theory of Operation

This section describes the electrical operation of the TDS 500D, TDS 600B and TDS 700D Oscilloscopes using the major circuit blocks or modules.
This section has two main parts: H Logic Conventions describes how logic functions are discussed and
represented in this manual.
H Module Overview describes circuit operation from a functional-circuit block
perspective.

Logic Conventions

The oscilloscopes contain many digital logic circuits. This manual refers to these circuits with standard logic symbols and terms. Unless otherwise stated, all logic functions are described using the positive-logic convention: the more positive of the two logic levels is the high (1) state, and the more negative level is the low (0) state. Signal states may also be described as “true” meaning their active state or “false” meaning their nonactive state. The specific voltages that constitute a high or low state vary among the electronic devices.

Module Overview

General
Input Signal Path
Active-low signals are indicated by a tilde prefixed to the signal name (~RESET). Signal names are considered to be either active-high, active-low, or to have both active-high and active-low states.
This module overview describes the basic operation of each functional circuit block as shown in Figure 9-2.
The TDS 500D, TDS 600B and TDS 700D Oscilloscopes have four channels. Each channel provides a calibrated vertical scale factor.
A signal enters the oscilloscope through a probe connected to a BNC on the A10 Attenuator/Acquisition board.
Attenuators. Circuitry in the attenuator selects the input coupling, termination, and the attenuation factor. The processor system, by way of the acquisition system, controls the attenuators. For example, if 50 input termination is
TDS 500D, TDS 600B and TDS 700D Service Manual
3–1
Theory of Operation
selected and the input is overloaded, the processor system switches the input to the 1 MW position (on TDS models with 1 MW termination).
Probe Coding Interface. Probe coding interface signals pass through the attenuator portion of the A10 Attenuator/Acquisition to the acquisition system, where they are sensed and controlled.
Acquisition System. The acquisition system amplifies the input signals, samples them, converts them to digital signals, and controls the acquisition process under direction of the processor system. The acquisition system includes the trigger, acquisition timing, and acquisition mode generation and control circuitry.
V oltage Controlled Oscillator (VCO). Master clocks for the acquisition system are generated by the circuitry on the A10 Acquisition board. The circuitry makes up a phase-locked loop. The master clock is divided down by the A10 Acquisition circuitry under control of the processor system.
D1 Bus. The acquisition system passes the digital values representing the acquired waveform through the A14 D1 Bus to the A11 DRAM Processor/Dis­play board. This happens after a waveform acquisition is complete if the digital signal processor in the processor system requests the waveform.
Processor System. The processor system contains a microprocessor that controls the entire instrument. This system also includes the firmware and a GPIB interface. You can reprogram the firmware from a remote controller using the GPIB and an external software package.
The processor also includes a digital signal processor. This signal processor processes each waveform as directed by the system processor. Waveforms and any text to be displayed are passed on to the display system. The A11 DRAM Processor/Display board contains both the processor and display systems.
Display System. Text and waveforms are processed by different parts of the display circuitry. The display system sends the text and waveform information to the tube assembly as a video signal. The display system also generates and sends vertical (VSYNC) and horizontal (HSYNC) sync signals to the tube assembly. A VGA-compatible video output is at the rear of the TDS 500D, TDS 600B and TDS 700D.
3–2
Tube Assembly
All information (waveforms, text, graticules, and pictographs) is displayed by the A30/31/32 Display system. The A30 generates the high voltages necessary to drive the display tube. It also contains the video amplifier, horizontal oscillator, and the vertical and horizontal yoke driver circuitry.
TDS 500D, TDS 600B and TDS 700D Service Manual
Theory of Operation
Front Panel
Rear Panel
Low Voltage Power
Supply
The processor system sends instructions to and receives information from the Front Panel Processor on the A12 Front Panel board. The Front Panel Processor reads the front-panel switches and potentiometers. Any changes in their settings are reported to the processor system. The Front Panel Processor also turns the LEDs on and off and generates the bell signal.
Front-panel menu switches are also read by the Front Panel Processor. The processor sends any changes in menu selections to the processor system. The ON/STBY switch is one of the menu switches. However, it is not read by the Front Panel Processor, but passes through the A12 Front Panel board and the A11 DRAM Processor/Display board to the low voltage power supply.
The front panel also generates the probe compensation signals SIGNAL and GND.
The GPIB connector provides access to stored waveforms and allows external control of the oscilloscope.
You can make hardcopies on the GPIB, RS-232 and Centronics ports.
The low voltage power supply is a switching power converter with active power factor control. It supplies power to all of the circuitry in the oscilloscope.
Fan
The principal POWER switch, located on the rear panel, controls all power to the oscilloscope including the Low Voltage Power Supply. The ON/STBY switch, located on the front panel, also controls all of the power to the oscillo­scope except for part of the circuitry in the Low Voltage Power Supply.
The power supply sends a power fail (~PF) warning to the processor system if the power is going down.
The fan provides forced air cooling for the oscilloscope. It connects to +25 V from the Low Voltage Power Supply by way of the A11 DRAM Processor/Dis­play module.
TDS 500D, TDS 600B and TDS 700D Service Manual
3–3
Theory of Operation
3–4
TDS 500D, TDS 600B and TDS 700D Service Manual

Performance Verification Procedures

Two types of Performance Verification procedures can be performed on this product: Brief Procedures and Performance Tests. You may not need to perform all of these procedures, depending on what you want to accomplish.
H To rapidly confirm that the oscilloscope functions and was adjusted properly,
just do the brief procedures under Self Tests, which begin on page 4–5. Advantages: These procedures are quick to do, require no external
equipment or signal sources, and perform extensive functional and accuracy testing to provide high confidence that the oscilloscope will perform properly. They can be used as a quick check before making a series of important measurements.
H To further check functionality, first do the Self Tests just mentioned; then do
the brief procedures under Functional Tests that begin on page 4–7. Advantages: These procedures require minimal additional time to perform,
require no additional equipment other than a standard-accessory probe, and more completely test the internal hardware of the oscilloscope. They can be used to quickly determine if the oscilloscope is suitable for putting into service, such as when it is first received.
H If more extensive confirmation of performance is desired, do the Perform-
ance Tests, beginning on page 4–15, after doing the Functional and Self Tests
just referenced. Advantages: These procedures add direct checking of warranted specifica-
tions. They require more time to perform and suitable test equipment is required. (See Equipment Required beginning on page 4–16).
If you are not familiar with operating this oscilloscope, read the TDS 500D,
TDS 600B & TDS 700D Reference (071-0137-XX) or the TDS 500D, TDS 600B & TDS 700D User Manual (071-0130-XX). These contain instructions that will
acquaint you with the use of the front-panel controls and the menu system.
TDS 500D, TDS 600B and TDS 700D Service Manual
4–1
Performance Verification Procedures

Input Channels vs. Model

When performing the procedures in this section, be aware that some TDS models refer to input channels Ch 3 and Ch 4 as Aux 1 and Aux 2 respectively. Where appropriate, both names will appear in the procedure, for example, Ch 3 (Aux 1). The channel names for the various TDS models are shown below.
TDS Model Channel Names
TDS 540D, 580D, 644B, 684B, 754D, 784D, and 794D TDS 520D, 620B, 680B, and 724D Ch 1, Ch 2, Aux 1, and Aux 2

Conventions

Throughout these procedures the following conventions apply:
Ch 1, Ch 2, Ch 3, and Ch 4
H Each test procedure uses the following general format:
Title of Test Equipment Required Prerequisites Procedure
H Each procedure consists of as many steps, substeps, and subparts as required
to do the test. Steps, substeps, and subparts are sequenced as follows:
1. First Step a. First Substep
H First Subpart H Second Subpart
b. Second Substep
2. Second Step
H In steps and substeps, the lead-in statement in italics instructs you what to
do, while the instructions that follow tell you how to do it, as in the example step below, “Initialize the oscilloscope” by doing “Press save/recall SETUP. Now, press the main-menu button...”.
4–2
TDS 500D, TDS 600B and TDS 700D Service Manual
Performance Verification Procedures
Initialize the oscilloscope: Press save/recall SETUP. Now, press the main-menu button Recall Factory Setup; then the side-menu button OK Confirm Factory Init.
H Where instructed to use a front-panel button or knob, or select from a main
or side menu, or verify a readout or status message, the name of the button or knob appears in boldface type: “press SHIFT; then UTILITY, press the main-menu button System until Cal is highlighted in the pop-up menu. Verify that the status message is Pass in the main menu under the Voltage Reference label.”
STOP. The STOP notation at the left is accompanied by information you must read to do the procedure properly.
H Refer to Figure 4–1: “Main menu” refers to the menu that labels the seven
menu buttons under the display; “side menu” refers to the menu that labels the five buttons to the right of the display. “Pop-up menu” refers to a menu that pops up when a main-menu button is pressed.
Brief status
information
Graticule and waveforms
Waveform reference
symbols show ground levels
and waveform sources
Vertical scale, horizontal scale,
and trigger level readouts
Pop-up menu
Position of waveform
record relative to
the screen and display
General purpose
knob readout
Side menu area. Readouts for measurements move here when CLEAR MENU is pressed.
Main menu display area. Readouts in lower graticule
area move here when CLEAR MENU is pressed.
Figure 4–1: Map of display functions
TDS 500D, TDS 600B and TDS 700D Service Manual
4–3
Performance Verification Procedures
4–4
TDS 500D, TDS 600B and TDS 700D Service Manual

Brief Procedures

Self Tests

The Self Tests use internal routines to confirm basic functionality and proper adjustment. No test equipment is required to do these test procedures.
The Functional Tests utilize the probe-compensation output at the front panel as a test-signal source for further verifying that the oscilloscope functions properly. A probe is required to do these test procedures.
This procedure uses internal routines to verify that the oscilloscope functions and was adjusted properly. No test equipment or hookups are required.
Verify Internal Adjustment,
Self Compensation, and
Diagnostics
Equipment required
Prerequisites Power on the oscilloscope and allow a 20 minute warm-up before doing
1. Verify that internal diagnostics pass: Do the following substeps to verify
passing of internal diagnostics.
a. Display the System diagnostics menu:
H Press SHIFT; then press UTILITY. H Repeatedly press the main-menu button System until Diag/Err is
highlighted in the pop-up menu.
b. Run the System Diagnostics:
H First disconnect any input signals from all four channels. H Press the main-menu button Execute; then press the side-menu
button OK Confirm Run Test.
c. Wait: The internal diagnostics do an exhaustive verification of proper
oscilloscope function. This verification will take up to three and a half minutes on some models. When the verification is finished, the resulting status will appear on the screen.
None
this procedure.
d. Verify that no failures are found and reported on-screen. If any failures
occur do step 1a, then press the main menu button Error Log for details.
e. Confirm the three adjustment sections have passed status:
TDS 500D, TDS 600B and TDS 700D Service Manual
4–5
Brief Procedures
H Press SHIFT; then press UTILITY. H Highlight Cal in the pop-up menu by repeatedly pressing the
main-menu button System. See Figure 4–2.
H Verify that the word Pass appears in the main menu under the
following menu labels: Voltage Reference, Frequency Response, and Pulse Trigger. See Figure 4–2.
f. Run the signal-path compensation: Press the main-menu button Signal
Path; then press the side-menu button OK Compensate Signal Paths.
g. Wait: Signal-path compensation may take five minutes on the
TDS 500D/700D and fifteen minutes on the 600B to run. While it progresses, a “clock” icon (shown at left) is displayed on-screen. When compensation completes, the status message will be updated to Pass or
Fail in the main menu. See step h.
h. Confirm signal-path compensation returns passed status: Verify that the
word Pass appears under Signal Path in the main menu. See Figure 4–2.
4–6
Highlight Cal
1
3
Verify Pass
2
Verify Pass
Figure 4–2: Verifying adjustments and signal-path compensation
2. Return to regular service: Press CLEAR MENU to exit the system menus.
TDS 500D, TDS 600B and TDS 700D Service Manual

Functional Tests

Brief Procedures
The purpose of these procedures is to confirm that the oscilloscope functions properly. The only equipment required is one of the standard-accessory probes and, to check the file system, a 3.5 inch, 720 K or 1.44 Mbyte floppy disk.
CAUTION. The P6217, P6243, and P6245 probes that can be used with this oscilloscope provide an extremely low loading capacitance (<1 pF) to ensure the best possible signal reproduction. These probes should not be used to measure signals exceeding ±8 volts, or errors in signal measurement will be observed. Above 40 volts, damage to the probe may result. To make measurements beyond ±10 volts, use either the P6139A probe (good to 500 volts peak), the P6339A probe (for the TDS 794D), or refer to the catalog for a recommended probe.
STOP. These procedures verify functions; that is, they verify that the oscilloscope features operate. They do not verify that they operate within limits.
Verify All Input Channels
Therefore, when the instructions in the functional tests that follow call for you to verify that a signal appears on-screen “that is about five divisions in amplitude” or “has a period of about six horizontal divisions,” etc., do NOT interpret the quantities given as limits. Operation within limits is checked in Performance Tests, which begin on page 4–15.
STOP. DO NOT make changes to the front-panel settings that are not called out in the procedures. Each verification procedure will require you to set the oscilloscope to certain default settings before verifying functions. If you make changes to these settings, other than those called out in the procedure, you may obtain invalid results. In this case, just redo the procedure from step 1.
When you are instructed to press a menu button, the button may already be selected (its label will be highlighted). If this is the case, it is not necessary to press the button.
Equipment required
Prerequisites None
One probe such as the P6243, P6245, P6139A, or P6339A
1. Install the test hookup and preset the oscilloscope controls: a. Hook up the signal source: Install the probe on CH 1. Connect the probe
tip to PROBE COMPENSATION SIGNAL on the front panel;
TDS 500D, TDS 600B and TDS 700D Service Manual
4–7
Brief Procedures
connect the probe ground (typically black) to PROBE COMPENSA- TION GND. If using a P6243 or P6245 probe, you may want to attach a Y-lead connector and two SMD KlipChips as shown in Figure 4–3.
Signal Gnd
Figure 4–3: Universal test hookup for functional tests – TDS 600B shown
b. Initialize the oscilloscope:
H Press save/recall SETUP. H Press the main-menu button Recall Factory Setup. H Press the side-menu button OK Confirm Factory Init.
2. Verify that all channels operate: Do the following substeps test CH 1 first, skipping substep a and b since CH 1 is already set up for verification
and as the trigger source from step 1. a. Select an unverified channel:
H Press WAVEFORM OFF to remove the channel just verified from
display.
4–8
H Press the front-panel button that corresponds to the channel you are
to verify.
TDS 500D, TDS 600B and TDS 700D Service Manual
Brief Procedures
H Move the probe to the channel you selected.
b. Match the trigger source to the channel selected:
H Press TRIGGER MENU. H Press the main-menu button Source. H Press the side-menu button that corresponds to the channel selected,
Ch2, Ch3, or Ch4. (Some TDS models use Ax1 and Ax2 instead of Ch3 and Ch4).
c. Set up the selected channel:
H Set the vertical SCALE to 200 mV. H Set the horizontal SCALE to 200 s. Press CLEAR MENU to
remove any menu that may be on the screen.
H Press SET LEVEL TO 50%.
d. Verify that the channel is operational: Confirm that the following
statements are true. H The vertical scale readout for the channel under test shows a setting
of 200 mV, and a square-wave probe-compensation signal about
2.5 divisions in amplitude is on-screen. See Figure 4–1 on page 4–3 to locate the readout.
H The vertical POSITION knob moves the signal up and down the
screen when rotated.
H Turning the vertical SCALE knob counterclockwise decreases the
amplitude of the waveform on-screen, turning the knob clockwise increases the amplitude, and returning the knob to 200 mV returns the amplitude to about 2.5 divisions.
e. Verify that the channel acquires in all acquisition modes: Press SHIFT;
then press ACQUIRE MENU. Use the side menu to select, in turn, each of the three hardware acquire modes and confirm that the following statements are true. Refer to the icons at the left of each statement as you confirm those statements.
H Sample mode displays an actively acquiring waveform on-screen.
(Note that there is noise present on the peaks of the square wave).
H Peak Detect mode displays an actively acquiring waveform
on-screen with the noise present in Sample mode “peak detected.”
H Hi Res mode (TDS 500D and 700D only) displays an actively
acquiring waveform on-screen with the noise that was present in Sample mode reduced.
TDS 500D, TDS 600B and TDS 700D Service Manual
4–9
Brief Procedures
H Envelope mode displays an actively acquiring waveform on-screen
with the noise displayed.
H Average mode displays an actively acquiring waveform on-screen
with the noise reduced.
f. Test all channels: Repeat substeps a through e until all four input
channels are verified.
3. Remove the test hookup: Disconnect the probe from the channel input and the probe-compensation terminals.
Verify the Time Base
Equipment required
Prerequisites None
One probe such as the P6243, P6245, P6139A, or P6339A
1. Install the test hookup and preset the oscilloscope controls: a. Hook up the signal source: Install the probe on CH 1. Connect the probe
tip to PROBE COMPENSATION SIGNAL on the front panel; connect the probe ground to PROBE COMPENSATION GND. See Figure 4–3 on page 4–8.
b. Initialize the oscilloscope:
H Press save/recall SETUP. H Press the main-menu button Recall Factory Setup; then press the
side-menu button OK Confirm Factory Init.
c. Modify default settings:
H Set the vertical SCALE to 200 mV. H Set the horizontal SCALE to 200 s. H Press SET LEVEL TO 50%.
4–10
H Press CLEAR MENU to remove the menus from the screen.
2. Verify that the time base operates: Confirm the following statements. a. One period of the square-wave probe-compensation signal is about five
horizontal divisions on-screen for the 200 s horizontal scale setting (set in step 1c).
b. Rotating the horizontal SCALE knob clockwise expands the waveform
on-screen (more horizontal divisions per waveform period), counter­clockwise rotation contracts it, and returning the horizontal scale to 200 s returns the period to about five divisions.
TDS 500D, TDS 600B and TDS 700D Service Manual
Brief Procedures
c. The horizontal POSITION knob positions the signal left and right
on-screen when rotated.
3. Remove the test hookup: Disconnect the probe from the channel input and
the probe-compensation terminals.
Verify the Main and
Delayed Trigger Systems
Equipment required
Prerequisites None
One probe such as the P6243, P6245, P6139A, or P6339A
1. Install the test hookup and preset the oscilloscope controls: a. Hook up the signal source: Install the probe on CH 1. Connect the probe
tip to PROBE COMPENSATION SIGNAL on the front panel; connect the probe ground to PROBE COMPENSATION GND. See Figure 4–3 on page 4–8.
b. Initialize the oscilloscope:
H Press save/recall SETUP. H Press the main-menu button Recall Factory Setup. H Press the side-menu button OK Confirm Factory Init.
c. Modify default settings:
H Set the vertical SCALE to 200 mV. H Set the horizontal SCALE for the M (main) time base to 200 s. H Press SET LEVEL TO 50%. H Press TRIGGER MENU. H Press the main-menu button Mode & Holdoff. H Press the side-menu button Normal. H Press CLEAR MENU to remove the menus from the screen.
2. Verify that the main trigger system operates: Confirm that the following statements are true.
H The trigger level readout for the main trigger system changes with the
trigger-LEVEL knob.
H The trigger-LEVEL knob can trigger and untrigger the square-wave
signal as you rotate it. (Leave the signal untriggered, which is indicated by the display not updating).
TDS 500D, TDS 600B and TDS 700D Service Manual
4–11
Brief Procedures
H Pressing SET LEVEL TO 50% triggers the signal that you just left
untriggered. (Leave the signal triggered).
3. Verify that the delayed trigger system operates: a. Select the delayed time base:
H Press HORIZONTAL MENU. H Press the main-menu button Time Base. H Press the side-menu button Delayed Triggerable; then press the
side-menu button Delayed Only.
H Set the horizontal SCALE for the D (delayed) time base to 200 s.
b. Select the delayed trigger level menu:
H Press SHIFT; then press DELAYED TRIG. H Press the main-menu button Level; then press the side-menu button
Level.
c. Confirm that the following statements are true:
H The trigger-level readout for the delayed trigger system changes as
you turn the general purpose knob.
H As you rotate the general purpose knob, the square-wave probe-com-
pensation signal can become triggered and untriggered. (Leave the signal untriggered, which is indicated by the display not updating).
H Pressing the side-menu button Set to 50% triggers the probe-com-
pensation signal that you just left untriggered. (Leave the signal triggered).
d. Verify the delayed trigger counter:
H Press the main-menu button Delay by Time. H Use the keypad to enter a delay time of 1 second. Press 1, then press
ENTER.
H Verify that the trigger READY indicator on the front panel flashes
about once every second as the waveform is updated on-screen.
4. Remove the test hookup: Disconnect the probe from the channel input and the probe-compensation terminals.
4–12
TDS 500D, TDS 600B and TDS 700D Service Manual
Brief Procedures
Verify the File System
Equipment required
Prerequisites None
One probe such as the P6243, P6245, P6139A, or P6339A One 720 K or 1.44 Mbyte, 3.5 inch DOS-compatible disk.
You can use a disk of your own or you can use the TDS Family Programmer Disk (Tektronix part number 063-3002-XX) included in the TDS 500D, TDS 600B & TDS 700D User Manual (Tektronix part number 071–0130–XX).
1. Install the test hookup and preset the oscilloscope controls: a. Hook up the signal source: Install the probe on CH 1. Connect the probe
tip to PROBE COMPENSATION SIGNAL on the front panel; connect the probe ground to PROBE COMPENSATION GND. See Figure 4–3 on page 4–8.
b. Insert the test disk: Insert the floppy disk in the floppy disk drive to the
left of the monitor.
H Position the disk so the metal shutter faces the drive. H Position the disk so the stamped arrow is on the top right side. In
other words, place the angled corner in the front bottom location.
H Push the disk into the drive until it goes all the way in and clicks
into place.
c. Initialize the oscilloscope:
H Press save/recall SETUP. H Press the main-menu button Recall Factory Setup. H Press the side-menu button OK Confirm Factory Init.
d. Modify default settings:
H Set the vertical SCALE to 200 mV. H Set the horizontal SCALE for the M (main) time base to 200 s.
Notice the waveform on the display now shows two cycles instead of five.
H Press SET LEVEL TO 50%. H Press CLEAR MENU to remove the menus from the screen.
e. Save the settings:
H Press SETUP. H Press the main-menu button Save Current Setup; then press the
side-menu button To File.
TDS 500D, TDS 600B and TDS 700D Service Manual
4–13
Brief Procedures
H Turn the general purpose knob to select the file to save. Choose
 (or fd0:, the floppy disk drive, hd0:, the optional hard disk drive, or Zip:, if an external Zip drive is connected). With this choice, you will save a file starting with , then containing 5-numbers, and a  extension. For example, the first time you run this on a blank, formatted disk or on the Example Programs Disk, the oscilloscope will assign the name  to your file. If you ran the procedure again, the oscilloscope would increment the name and call the file .
NOTE. If testing an external Zip drive, it must first be connected to the oscillo­scope, then be powered on simultaneously with the oscilloscope, or immediately after the oscilloscope is powered-on. This ensures proper communications are set up between the oscilloscope and the Zip drive.
f. To test the optional hard disk drive or external Zip drive, choose either
the hard disk drive (hd0:), or the external Zip drive (Zip:). Then use the general purpose knob to select the file to save. Save the file as in step e.
H Press the side-menu button Save To Selected File.
2. Verify the file system works:
H Press the main-menu button Recall Factory Setup and the side-menu
button OK Confirm Factory Init to restore the 500 s time base and the five cycle waveform.
H Press the main-menu button Recall Saved Setup; then press the
side-menu button From File.
H Turn the general purpose knob to select the file to recall. For example, if
you followed the instructions previously and saved the file to either the floppy disk drive (fd0:), or the optional hard disk drive (hd0:), you had the oscilloscope assign the name TEK00000.SET to your file.
H Press the side-menu button Recall From Selected File. H Verify that the oscilloscope retrieved the saved setup from the disk. Do
this by noticing the horizontal SCALE for the M (main) time base is again 200 s and the waveform shows only two cycles just as it was when you saved the setup.
3. Remove the test hookup:
4–14
H Disconnect the probe from the channel input and the probe-compensa-
tion terminals.
H Remove the floppy disk from the floppy disk drive.
TDS 500D, TDS 600B and TDS 700D Service Manual

Performance Tests

This section contains a collection of procedures for checking that the TDS 500D, TDS 600B, and TDS 700D Oscilloscopes perform as warranted.
The procedures are arranged in four logical groupings: Signal Acquisition System
Checks, Time Base System Checks, Triggering System Checks, and Output Ports Checks. They check all the characteristics that are designated as checked in Specifications. (The characteristics that are checked appear in boldface type under Warranted Characteristics in Specifications).
STOP. These procedures extend the confidence level provided by the basic procedures described on page 4–5. The basic procedures should be done first, then these procedures performed if desired.

Prerequisites

The tests in this section comprise an extensive, valid confirmation of perform­ance and functionality when the following requirements are met:
H The cabinet must be installed on the oscilloscope. H You must have performed and passed the procedures under Self Tests, found
on page 4–5, and those under Functional Tests, found on page 4–7.
H A signal-path compensation must have been done within the recommended
calibration interval and at a temperature within ±5_ C of the present operating temperature. (If at the time you did the prerequisite Self Tests, the temperature was within the limits just stated, consider this prerequisite met).
H The oscilloscope must have been last adjusted at an ambient temperature
between +20_ C and +30_ C, must have been operating for a warm-up period of at least 20 minutes, and must be operating at an ambient tempera­ture between +4_ C and either +45_ C for the TDS 600B or +50_ C for the TDS 500D and 700D. (The warm-up requirement is usually met in the course of meeting the Self Tests and Functional Tests prerequisites listed above).
TDS 500D, TDS 600B and TDS 700D Service Manual
4–15
Performance Tests

Equipment Required

T able 4–1: Test equipment
Item number and description
These procedures use external, traceable signal sources to directly check warranted characteristics. The required equipment list follows this introduction.
Minimum requirements Example Purpose
1. Attenuator ,10X (two required)
2. Attenuator , 5X Ratio: 5X; impedance 50 ; connec-
3. Adapter, BNC female-to­Clip Leads
4. Terminator, 50 Impedance 50 ; connectors: female
5. Cable, Precision 50 Coaxial (two required)
6. Connector , Dual-Banana (two required)
7. Connector , BNC “T” Male BNC-to-dual female BNC Tektronix part number
8. Coupler, Dual-Input Female BNC-to-dual male BNC Tektronix part number
9. Generator , DC Calibra­tion
10. Generator, Calibration 500 mV square wave calibrator
11. Generator, Time Mark (optional)
12. Probe, 10X A P6139A, P6243, P6245, or P6339A
13. 3.5 inch, 720 K or
1.44 Mbyte, DOS-com­patible floppy disk
14. Generator, Sine Wave 250 kHz to at least 500 MHz (higher
Ratio: 10X; impedance 50 ; connec­tors: female BNC input, male BNC output
tors: female BNC input, male BNC output
BNC female-to-Clip Leads Tektronix part number
BNC input, male BNC output 50 , 36 in, male-to-male BNC
connectors Female BNC-to-dual banana Tektronix part number
Variable amplitude to ±104 V; accura­cy to 0.1%
amplitude; accuracy to 0.25% Variable marker frequency from 10 ms
to 10 ns; accuracy within 2 ppm
3
probe
for higher-bandwidth scopes). Vari­able amplitude from 60 mV to 2 V into 50 . Frequency accuracy >2.0%
Tektronix part number 01 1-0059-02
Tektronix part number 01 1-0060-02
013-0076-00 Tektronix part number
01 1-0049-01 Tektronix part number
012-0482-00
103-0090-00
103-0030-00
067-0525-02 Data Precision 8200
Tegam/Tektronix PG 506A Calibration Generator
Tegam/Tektronix TG 501A Time Mark Generator
Tektronix part number P6139A or P6245
TDS Family Programmer Disk (Tektronix part number 063–3002–00), included with User Manual (Tektronix part number 071–0130-XX)
Rohde & Schwarz SMT or
1,4
SMY
p-p
Signal Attenuation
Signal Attenuation
Signal Coupling for Probe Compensator Output Check
Signal Termination for Channel Delay Test
Signal Interconnection
Various Accuracy Tests
Checking Trigger Sensitivity
Checking Delay Between Channels
1
1,2
1,2
Checking DC Offset, Gain, and Measurement Accuracy
To check accuracy of CH 3 Signal Out
Checking Sample-Rate and Delay-time Accuracy
Signal Interconnection
Checking File System Basic Functionality
Checking Analog Bandwidth, Trigger Sensitivity , Sample­rate, External Clock, and Delay-Time Accuracy
4–16
TDS 500D, TDS 600B and TDS 700D Service Manual
Performance Tests
T able 4–1: Test equipment (cont.)
Item number and description
15. Meter, Level and Power Sensor
16. Splitter, Power Frequency range: DC to 1 GHz.
Frequency range:10 MHz to scope bandwidth (.5, 1, or 2 GHz) Amplitude range: 6 mV
p-p
to 2 V
p-p
Rohde & Schwarz URV 35 or NRVS 55, with NRV -Z24 power sensor
Rohde & Schwarz RVZ
1,4
1,4
Tracking: >2.0%
17. Adapter (four required) Male N-to-female BNC Tektronix part number
103-0045-00
18. Adapter Female N-to-male BNC Tektronix part number
103-0058-00
19. Generator, Video Signal Provides PAL compatible outputs Tektronix TSG 121 Used to Test Video Option 05
20. Oscillator, Leveled Sine wave Generator
60 Hz Sine wave Tegam/Tektronix SG 502 (if
available)
1,2
21. Pulse Generator T ektronix CFG280
(or Tegam/Tektronix PG 502)
22. Cable, Coaxial (two required)
23. Terminator, 75 (two required)
24. Generator, Optical Im­pulse
25. Generator, Optical Im­pulse
75 , 36 in, male to male BNC connectors
Impedance 75 ; connectors: female BNC input, male BNC output
Tektronix part number 012-1338-00
Tektronix part number 01 1-0102-01
850 nm optical impulse T ektronix OIG501 Optical
Impulse Generator
2
1300 nm optical impulse T ektronix OIG502 Optical
Impulse Generator
2
26. Attenuator, Optical 62.5 micron optical attenuator Tektronix OA5022 Optical
Attenuator
2
27. Cable, Fiber Optic FC-FC fiber cable Tektronix part number
174-2322-00
28. Optical-to-Electrical Converter
P6701B (used with Option 3C) or P6703B (used with Option 4C)
Tektronix part number P6701B or P6703B
optical-to-electrical converter
1
You may replace items 9, 10, 11, 14, 15, 16, and 20 with a Wavetek 9100 (with options 100 and 600) for 500 MHz oscillo­scopes, or a Wavetek 9500 (with option 100) and output head appropriate for the bandwidth of the oscilloscope(s) being tested.
2
Requires a TM 500 or TM 5000 Series Power Module Mainframe.
3
Warning: The P6243 and P6245 probes that may be used with this oscilloscope provide an extremely low loading capacitance (<1 pF) to ensure the best possible signal reproduction. These probes should not be used to measure signals exceeding ±8 V , or errors in signal measurement will be observed. Above 40 V, damage to the probe may result. To make measurements beyond ±8 V, use either the P6139A probe (good to 500 V), or refer to the catalog for a recom­mended probe.
4
If available, you may replace items 14, 15, and 16 with a Tegam/Tektronix SG5030 and its leveling head for bandwidths up to 500 MHz, or a Tegam/Tektronix SG503 and SG504 with SG504 leveling head for bandwidths up to 1 GHz.
PurposeExampleMinimum requirements
Checking Analog Bandwidth and Trigger Sensitivity
Checking Analog Bandwidth
Checking Analog Bandwidth
Checking Analog Bandwidth
Equipped Instruments Only Used to Test Video Option 05
Equipped Instruments Only Used to Test Video Option 05
2
Equipped Instruments Only Used to Test Video Option 05
Equipped Instruments Only Used to Test Video Option 05
Equipped Instruments Only Checking Option 3C
Checking Option 4C
Checking Option 3C and 4C
Checking Option 3C and 4C
Checking Option 3C and 4C
TDS 500D, TDS 600B and TDS 700D Service Manual
4–17
Performance Tests

TDS 600B Test Record

Photocopy this and the next three pages and use them to record the performance test results for your TDS 600B.
TDS 600B T est Record
Instrument Serial Number: Certificate Number: Temperature: RH %: Date of Calibration: Technician:
TDS 600B Performance Test Minimum Incoming Outgoing Maximum
Offset Accuracy CH1 Offset +1 mV
+101 mV +1.01 V
CH2 Offset +1 mV
+101 mV +1.01 V
CH3 or AX1 +1 mV Offset +101 mV
+1.01 V
CH4 or AX2 +1 mV Offset +101 mV
+1.01 V DC Voltage Measurement Accuracy (Averaged) CH1 5 mV Vert scale setting,
–5 Div position setting, +1 V offset
CH1 5 mV Vert scale setting,
+5 Div position setting, –1 V offset
CH1 200 mV Vert scale setting,
–5 Div position setting, +10 V offset
CH1 200 mV Vert scale setting,
+5 Div position setting, –10 V offset
CH1 1.01 V Vert scale setting,
–5 Div position setting, +10 V offset
CH1 1.01 V Vert scale setting,
+5 Div position setting, –10 V offset
CH2 5 mV Vert scale setting,
–5 Div position setting, +1 V offset
CH2 5 mV Vert scale setting,
+5 Div position setting, –1 V offset
CH2 200 mV Vert scale setting,
–5 Div position setting, +10 V offset
– 2.1 mV – 75.6 mV – 756 mV
– 2.1 mV – 75.6 mV – 756 mV
– 2.1 mV – 75.6 mV – 756 mV
– 2.1 mV – 75.6 mV – 756 mV
+ 1.0329 V __________ __________ + 1.0471 V
– 1.0471 V __________ __________ – 1.0329 V
+ 11.4165 V __________ __________ + 11.7835 V
– 11.7835 V __________ __________ – 11.4165 V
+ 17.102 V __________ __________ + 18.899 V
– 18.899 V __________ __________ – 17.102 V
+ 1.0329 V __________ __________ + 1.0471 V
– 1.0471 V __________ __________ – 1.0329 V
+ 11.4165 V __________ __________ + 11.7835 V
__________ __________ __________
__________ __________ __________
__________ __________ __________
__________ __________ __________
__________ __________ __________
__________ __________ __________
__________ __________ __________
__________ __________ __________
+ 2.1 mV + 75.6 mV + 756 mV
+ 2.1 mV + 75.6 mV + 756 mV
+ 2.1 mV + 75.6 mV + 756 mV
+ 2.1 mV + 75.6 mV + 756 mV
4–18
TDS 500D, TDS 600B and TDS 700D Service Manual