This document supports firmware version 1.00
and above.
Warning
The servicing instructions are for use by
qualified personnel only. To avoid personal
injury, do not perform any servicing unless you
are qualified to do so. Refer to all safety
summaries prior to performing service.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes
that in all previously published material. Specifications and price change privileges reserved.
Tektronix, Inc., P.O. Box 500, Beaverton, OR 97077
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
WARRANTY
Tektronix warrants that the products that it manufactures and sells will be free from defects in materials and workmanship for
a period of three (3) years from the date of purchase from an authorized Tektronix distributor. If any such product proves
defective during this warranty period, Tektronix, at its option, either will repair the defective product without charge for parts
and labor, or will provide a replacement in exchange for the defective product. Batteries are excluded from this warranty.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the
warranty period and make suitable arrangements for the performance of service. Customer shall be responsible for packaging
and shipping the defective product to the service center designated by Tektronix, shipping charges prepaid, and with a copy of
customer proof of purchase. Tektronix shall pay for the return of the product to Customer if the shipment is to a location
within the country in which the Tektronix service center is located. Customer shall be responsible for paying all shipping
charges, duties, taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance
and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage resulting from attempts
by personnel other than Tektronix representatives to install, repair or service the product; b) to repair damage resulting from
improper use or connection to incompatible equipment; c) to repair any damage or malfunction caused by the use of
non-Tektronix supplies; or d) to service a product that has been modified or integrated with other products when the effect of
such modification or integration increases the time or difficulty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THE LISTED PRODUCTS IN LIEU OF
ANY OTHER WARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY
IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
TEKTRONIX’ RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND
EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX
AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR
CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS
ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.
Table of Contents
Specifications
Operating Information
General Safety Summaryv...................................
Service Safety Summaryvii....................................
TDS 200 Series Digital Oscilloscope Service Manual
General Safety Summary
Review the following safety precautions to avoid injury and prevent damage to
this product or any products connected to it.
Only qualified personnel should perform service procedures.
Injury Precautions
Use Proper Power Cord. To avoid fire hazard, use only the power cord specified
for this product.
Avoid Electric Overload. To avoid electric shock or fire hazard, do not apply a
voltage to a terminal that is outside the range specified for that terminal.
Avoid Overvoltage. To avoid electric shock or fire hazard, do not apply potential
to any terminal, including the common terminal, that varies from ground by more
than the maximum rating for that terminal.
Avoid Electric Shock. To avoid injury or loss of life, do not connect or disconnect
probes or test leads while they are connected to a voltage source.
Ground the Product. This product is grounded through the grounding conductor
of the power cord. To avoid electric shock, the grounding conductor must be
connected to earth ground. Before making connections to the input or output
terminals of the product, ensure that the product is properly grounded.
Connect the Probe Properly. The probe ground lead is at ground potential. Do not
connect the ground lead to an elevated voltage.
Do Not Operate Without Covers. To avoid electric shock or fire hazard, do not
operate this product with covers or panels removed.
Use Proper Fuse. To avoid fire hazard, use only the fuse type and rating specified
for this product.
Do Not Operate in Wet/Damp Conditions. To avoid electric shock, do not operate
this product in wet or damp conditions.
Do Not Operate in an Explosive Atmosphere. To avoid injury or fire hazard, do not
operate this product in an explosive atmosphere.
Product Damage
Precautions
TDS 200 Series Digital Oscilloscope Service Manual
Use Proper Power Source. Do not operate this product from a power source that
applies more than the voltage specified.
Provide Proper Ventilation. To prevent product overheating, provide proper
ventilation.
Do Not Operate With Suspected Failures. If you suspect there is damage to this
product, have it inspected by qualified service personnel.
v
General Safety Summary
Symbols and Terms
Terms in this Manual. These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result
in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in
damage to this product or other property.
Terms on the Product. These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the
marking.
WARNING indicates an injury hazard not immediately accessible as you read the
marking.
CAUTION indicates a hazard to property including the product.
Symbols on the Product. The following symbols may appear on the product:
Certifications and
Compliances
DANGER
High Voltage
Protective Ground
(Earth) Terminal
ATTENTION
Refer to Manual
Double
Insulated
Refer to the specifications section for a listing of certifications and compliances
that apply to this product.
vi
TDS 200 Series Digital Oscilloscope Service Manual
Service Safety Summary
Only qualified personnel should perform service procedures. Read this Service
Safety Summary and the General Safety Summary before performing any service
procedures.
Do Not Service Alone. Do not perform internal service or adjustments of this
product unless another person capable of rendering first aid and resuscitation is
present.
Disconnect Power. To avoid electric shock, disconnect the main power by means
of the power cord or, if provided, the power switch.
Use Care When Servicing With Power On. Dangerous voltages or currents may
exist in this product. Disconnect power, remove battery (if applicable), and
disconnect test leads before removing protective panels, soldering, or replacing
components.
To avoid electric shock, do not touch exposed connections.
TDS 200 Series Digital Oscilloscope Service Manual
vii
Service Safety Summary
viii
TDS 200 Series Digital Oscilloscope Service Manual
Preface
Related Manuals
The service manual for the TDS 200-Series Digital Real-Time Oscilloscope
provides information to troubleshoot and repair the instrument to the module
level.
Some TDS models have two input channels and an external trigger while other
models have four input channels. Most illustrations in this manual show a two
channel model. Some parts for the four channel model are slightly different
because of the additional channels.
Additional documentation for the instrument is contained in the related manuals
listed below.
Extension module
User manual
Language
English071-0398-XX071-0409-XX071-0493-XX
French071-0400-XX*071-0483-XX
German071-0402-XX*071-0485-XX
Italian071-0401-XX*071-0484-XX
Spanish071-0399-XX*071-0482-XX
Portuguese071-0403-XX*071-0486-XX
Japanese071-0405-XX*071-0488-XX
Korean071-0408-XX*071-0491-XX
Simplified Chinese071-0406-XX*071-0489-XX
Traditional Chinese071-0407-XX*071-0490-XX
Russian071-0404-XX071-0487-XX
*These manuals contain a language overlay for the front-panel controls.
part number
instructions part
number
Programmer manual
part number
TDS 200 Series Digital Oscilloscope Service Manual
ix
Preface
Contacting Tektronix
Phone1-800-833-9200*
AddressTektronix, Inc.
Department or name (if known)
14200 SW Karl Braun Drive
P.O. Box 500
Beaverton, OR 97077
USA
Web sitewww.tektronix.com
Sales support1-800-833-9200, select option 1*
Service support1-800-833-9200, select option 2*
Technical supportEmail: techsupport@tektronix.com
*This phone number is toll free in North America. After office hours, please leave a
voice mail message.
Outside North America, contact a Tektronix sales office or distributor; see the
Tektronix web site for a list of offices.
Product End-of-Life Handling
Components that Contain Mercury. The cold cathode fluorescent tube located in
the liquid crystal display backlight contains trace elements of mercury. When you
are ready to reclaim the instrument, you must properly transfer it according to
local regulations concerning mercury-containing equipment or ship the instrument to the Tektronix Recycling Operations (RAMS). You can contact
Tektronix for the RAMS shipping address and instructions.
1-800-833-9200, select option 3*
1-503-627-2400
6:00 a.m. -- 5:00 p.m. Pacific time
x
TDS 200 Series Digital Oscilloscope Service Manual
Specifications
g
q
p
,
g
All specifications apply to the TDS 200-Series Digital Real-Time Oscilloscope
with a P2100 probe with the Attenuation switch set to 10X unless noted
otherwise. To meet specifications, two conditions must first be met:
HThe instrument must have been operating continuously for twenty minutes
within the specified operating temperature.
HYou must perform the Self Cal operation, accessible through the utility menu,
if the operating temperature changes by more than 5° C.
All specifications are guaranteed unless noted “typical.” Specifications that are
marked with the n symbol are checked in the chapter Performance Verification.
Table 1--1: Specifications
Acquisition
Acquisition ModesSample, Peak detect, and Average
Acquisition Rate,
typical
Single SequenceAcquisition ModeAcquisition Stops After
Up to 180 waveforms per second, per channel (Sample acquisition mode, no measurements)
Sample, Peak DetectSingle acquisition, all channels simultaneously
AverageN acquisitions, all channels simultaneously, N is
selectable from 4, 16, 64, and 128
Inputs
Input CouplingDC, AC, or GND
Input Impedance,
DC Coupled, all channela
Input Impedance,
DC Coupled,
EXT TRIG only
P2100 Probe Attenuation1X, 10X
Probe Attenuation Factors1X, 10X, 100X, 1000X
Maximum Voltage Between Overvoltage CategoryMaximum Voltage
Signal and Common at
input BNC
*Bandwidth is not valid for the P2100 probe when the switch is set to 1X.
1MΩ ±2% in parallel with 20 pF ±3pF
TDS 210 (B099188 to B119999 & C021679 to C029999)
TDS 220 (B065810 to B079999 & C021127 to C029999)
1.2 MΩ ±5% in parallel with 20 pF ±5pF
CAT I and CAT II
CAT III150 V
For steady-state sinusoidal waveforms, derate at 20 dB/decade above 100 kHz to 13 Vpkat 3 MHz* and
above. Also, refer to Overvoltage Category description on page 1--7.
All other TDS 210, TDS 220, and TDS 224
1MΩ ±5% in parallel with 20 pF ±5pF
300 V
width < 100 msec.)
(420 V peak, duty factor < 50%, pulse
RMS
RMS
TDS 200 Series Digital Oscilloscope Service Manual
1--1
Specifications
g
inputBN
C
g
tBN
C
ithP210
0
g
5s/div),t
ypical
Table 1--1: Specifications (cont.)
Inputs
Maximum Voltage Between Overvoltage CategoryMaximum Voltage
Probe Tip and ground
using P2100 connected to
Time delay between
channels, typical
Channel-to-ChannelTDS 210TDS 220 and TDS 224
Common Mode Rejection,
typical
Channel-to-Channel Cross- TDS 210TDS 220 and TDS 224
talk
CAT I and CAT II300 V
(500 V peak, duty factor < 35%, pulse
RMS
width < 100 msec.)
CAT III100 V
Derate at 20 dB/decade above 900 kHz to 13 V
at 27 MHz* and above. Also, refer to Overvoltage
RMS
RMS
Category description on page 1--7.
150 ps
100:1 at 60 Hz
20:1at30MHz*
100:1 at 60 Hz
20:1at50MHz*
Measured on MATH Ch1 -- Ch2 waveform, with test signal applied between signal and common of both
channels, and with the same VOLTS/DIV and coupling settings on each channel. Also measured on MATH
Ch3 -- Ch4 waveform for the TDS 224.
≥100:1 at 30 MHz*
≥100:1 at 50 MHz*
Measured on one channel, with 8 division test signal applied between signal and common of the other
channel, the same VOLTS/DIV and coupling settings on each channel, and 50 Ω terminators on each
channel.
Vertical
Digitizers8 bit resolution (except when set to 2 mV/div), each channel sampled simultaneously
VOLTS/DIV Range2 mV/div to 5 V/div at input BNC
(Full bandwidth at >5 mV/div to 5 V/div, 20 MHz at 2 mV/dif to 5 mV/div, except in Peak Detect mode full
bandwidth at >10 mV/div to 5 V/div [20 MHz at 2 mV/div to 10 mV/div])
Position Range2 mV/div to 200 mV/div, ±2V
> 200 mV/div to 5 V/div, ±50 V
n Analog Bandwidth in
Sample and Average modes
a
or w
DC Coupled
Analog Bandwidth in PeakTDS 210TDS 220 and TDS 224
Detect mode (5 s/div to
Selectable Analog Bandwidth Limit, typical
probe,
Lower Frequency Limit,
AC Coupled
TDS 210TDS 220 and TDS 224
60 MHz* (when vertical scale set to > 5 mV/div)100 MHz* (when vertical scale set to > 5 mV/div)
* (when vertical scale set to ≤ 5mV/div)
20 MHz
50 MHz* (when vertical scale set to > 10 mV/div)75 MHz* (when vertical scale set to > 10 mV/div)
20 MHz* (when vertical scale set to ≤ 10 mV/div)
20 MHz*
≤10 Hz at BNC
≤1 Hz when using a 10X passive probe
*Bandwidth is not valid for the P2100 probe when the switch is set to 1X.
1--2
TDS 200 Series Digital Oscilloscope Service Manual
Table 1--1: Specifications (cont.)
,yp
y
Vertical
Rise Time at BNC, typicalTDS 210TDS220 and TDS 224
<5.8 ns<3.5 ns
Peak Detect Response
n DC Gain Accuracy
DC Measurement Accu-Measurement TypeAccuracy
racy, Average Acquisition
Mode
Captures 50% or greater amplitude of pulses ≥10 ns wide (5 s/div to 5 s/div) in the center 6 vertical
divisions. The oscilloscope reverts to Sample mode when the sec/div (Horizontal scale) setting is from
2.5 s/div to 5 ns/div. The Sample mode can still capture 10 ns glitches.
TDS210 and TDS220TDS224
2 mV/div and larger, ±3% for Sample or Average
acquisition mode
2 mV/div to 5 mV/div, ±4% for Sample or
Average acquisition mode
10 mV/div and larger, ±3% for Sample or
Average acquisition mode
Average of ≥16 waveforms with vertical position at zero
TDS224:
±(4% × reading + 0.1 div + 1 mV) and either
2 mV/div or 5 mV/div is selected.
Specifications
±(3% × reading + 0.1 div + 1 mV) and 10 mV/div
or greater is selected
TDS210 and TDS220:
±(3% × reading + 0.1 div + 1 mV) for all ranges
Average of ≥16 waveforms with vertical position not at
zero
TDS210, TDS220 and TDS224:
±[3% × (reading + vertical position) + 1% of
vertical position + 0.2 div]
Add 2 mV for settings from 2 mV/div to 200 mV/
div. Add 50 mV for settings from >200 mV/div to
5V/div.
Delta Volts Measurement
Accuracy, Average Acquisition Mode
P2100 passive probe
BandwidthDC to 100 MHzDC to 7 MHz
Attenuation ratio10:11:1
Compensation Range18 pf-35 pf
Input resistance
Input capacitance14.5 pf-17.5 pf80 pf-110 pf
Maximum input voltage
Delta volts between any two averages of ≥16
3% of reading + 0.05 div
waveforms acquired under same setup and ambient
conditions
10X position1X position
All oscilloscopes with 1 M Ω input
10 M Ω ± 2%1MΩ ± 2%
300 V
300 V
100 V
CAT I or 300 V DC CAT I
RMS
CATIIor300VDCCATII
RMS
CAT III or 100 V DC CAT III
RMS
150 V
150 V
100 V
RMS
RMS
RMS
Horizontal
Sample Rate Range50 S/s to 1 GS/s
Waveform Interpolation(sin x)/x
CAT I or 150 V DC CAT I
CATIIor150VDCCATII
CAT III or 100 V DC CAT III
TDS 200 Series Digital Oscilloscope Service Manual
1--3
Specifications
ggy,g
ggg
ggy
,
Table 1--1: Specifications (cont.)
Horizontal
Record Length2500 samples for each channel
SEC/DIV Range5 ns/div to 5 s/div, in a 1, 2.5, 5 sequence
n Sample Rate and Delay
Time Accuracy
Delta Time MeasurementConditionsAccuracy
Accuracy (Full Bandwidth)
(--4 div × s/div) to 20 ms(--4 div × s/div) to 50 ms (--4 div × s/div) to 50 s
±(1 sample interval + 100 ppm × reading +
0.6 ns)
0.4 ns)
n Trigger Sensitivity, Edge
Trigger Type
Trigger Sensitivity, Edge
Trigger Type, typical
Trigger Level RangeSourceRange
Trigger Level Accuracy,Accuracies are for signals having rise and fall times ≥20 ns
typical
CouplingSensitivity
DC
CouplingSensitivity
ACSame as DC at 50 MHz and above
NOISE REJReduces DC-coupled trigger sensitivity by 2 times for > 10 mV/div to 5 V/div
HF REJSame as DC-coupled limit from DC to 7 kHz, attenuates signals above
LF REJSame as the DC-coupled limits for frequencies above 300 kHz, attenuates
Internal±8 divisions from center of screen
TDS 210 and TDS 220
EXT±1.6 V
EXT/5±8V
Source
Internal
CH 1, CH 2,
CH 3 & CH 4
TDS 210 and TDS 220
EXT100 mV from DC to 10 MHz*,
EXT/5500 mV from DC to 10 MHz*,
80 kHz
signals below 300 kHz
Accuracy
±0.2 div × volts/div within ±4 divisions from center screen
1 div from DC to 10 MHz*,
1.5 div from 10 MHz* to Full
150 mV from 10 MHz* to Full
750 mV from 10 MHz* to Full
1--4
TDS 200 Series Digital Oscilloscope Service Manual
Table 1--1: Specifications (cont.)
y
,
Trigger
TDS 210 and TDS 220
EXT±(6% of setting + 40 mV)
EXT/5±(6% of setting + 200 mV)
*Bandwidth is not valid for the P2100 probe when the switch is set to 1X.
SET LEVEL TO 50%,
typical
Default Settings, Video
Trigger
Sensitivity, VideoComposite video signal
Trigger Type, typical
Signal Formats and Field
Rates, Video Trigger Type
Holdoff Range500 ns to 10 s
Operates with input signals ≥50 Hz
Trigger mode is Auto and Coupling is AC
Source
InternalPk-pk amplitude of 2 divisions
TDS 210 and TDS 220
EXT400 mV
EXT/52V
Supports NTSC, PAL, and SECAM broadcast systems for any field or any line
Range
Specifications
Measurements
Cursors
Automated MeasurementsCycle RMS, Mean, Pk -- Pk, Period, Frequency
Voltage difference between cursors (∆V)
Time difference between cursors (∆T)
Reciprocal of ∆T in Hertz (1/∆T)
Table 1--2: General specifications
Display
Display Type5.9 in (152 mm) diagonal liquid crystal
Display Resolution320 horizontal by 240 vertical pixels
Display ContrastAdjustable, temperature compensated
Backlight Intensity, typical60 cd/m
Probe compensator output
Output Voltage, typical
Frequency, typical1 kHz
2
5Vinto≥1MΩ load
TDS 200 Series Digital Oscilloscope Service Manual
1--5
Specifications
p
y
Table 1--2: General specifications (cont.)
Power Source
Source Voltage
100 -- 120 VAC
10% from 45 Hz through 440 Hz, CAT II
RMS
120 -- 250 VAC
10% from 45 Hz through 66 Hz, CAT II
RMS
Power ConsumptionTDS 210 and TDS 220TDS 224
Less than 20 WLess than 25 W
Fuse1 A, T rating, 250 V
Environmental
TemperatureOperating0° Cto+50° C
Nonoperating-- 2 0 ° Cto+60° C
Cooling MethodConvection
Humidity+40° C or below≤90% relative humidity
+41° Cto+50° C≤60% relative humidity
AltitudeOperating3,000 m
Nonoperating15,000 m
Random VibrationOperating0.31 g
from 5 Hz to 500 Hz, 10 minutes on each
RMS
axis
Nonoperating2.46 g
from 5 Hz to 500 Hz, 10 minutes on each
RMS
axis
Mechanical ShockOperating50 g, 11 ms, half sine
Mechanical
SizeHeight151.4 mm (5.96 in.)
Width304.8 mm (12 in.)
Depth120.7 mm (4.75 in.)
Weight (approximate)When packaged for domestic shipment3.6 kg (8.0 lbs)
1--6
TDS 200 Series Digital Oscilloscope Service Manual
Table 1--2: General specifications (cont.)
Certifications and compliances
Specifications
EC Declaration of
Conformity
(TDS 210, TDS 220 &
TDS 224)
(TDS 210, TDS 220,
TDS 224, P2100)
(P2100)
Certifications
(TDS 210, TDS 220,
TDS 224)
(P2100)
CSA Certified Power Cords CSA Certification includes the products and power cords appropriate for use in the North America power
Pollution Degree 2Do not operate in environments where conductive pollutants may be present.
Overvoltage Category
Meets intent of Directive 89/336/EEC for Electromagnetic Compatibility*; meets the requirements of FCC
Code of Federal Regulations, 47 CFR, Part 15, Subpart B, Class A:
CISPR 11 (Class A): Electromagnetic radiated and conducted emissions
EN 50082-1 European Community Requirements:
IEC 61000-4-2 electrostatic discharge
IEC 61000-4-3 radiated susceptibility**
IEC 61000-4-4 power line fast transients
IEC 61000-4-5 power line surge immunity
IEC 61000-4-6 conducted immunity to RF fields
IEC 61000-4-11 voltage dips, interrupts and variations
AS/NZS 2064, Australian emissions standard for Industrial, Scientific, and Medical Equipment
Low Voltage Directive 73/23/EEC as amended by 93/68/EED:
EN61010-1/A2:1995Safety requirements for electrical equipment for measurement control, and
laboratory use
EN 61010-2-031:1994Particular requirements for hand-held probe assemblies for electrical
measurement and test
CAN/CSA C22.2 No. 1010.1-92
UL3111-1, First Edition
IEC61010-1/A2
IEC61010-2-031
network. All other power cords supplied are approved for the country of use.
CAT ISignal levels in special equipment or parts of equipment, telecommunications,
electronics
*Instrument tested with shielded cables.
** Meets Performance Criterion A with allowed degradation of up to 5 major divisions of trace noise increase from 80 MHz to
200 MHz and up to 2 divisions of trace noise increase over the range of 200 MHz to 1000 MHz under a 3 V/m RF field. Ambient RF
fields may induce triggering when trigger threshold is offset less than 2.5 major divisions from ground reference.
Adjustment interval
The recommended adjustment interval is one year
TDS 200 Series Digital Oscilloscope Service Manual
1--7
Specifications
1--8
TDS 200 Series Digital Oscilloscope Service Manual
Operating Information
TDS 200-Series Digital Oscilloscopes are two or four channel oscilloscopes in
small, lightweight, benchtop packages that you can use to take ground-referenced
measurements.
In addition to the list of general features, this chapter covers the following topics:
HHow to install your product
HHow to add extended functions
HHow to perform a brief functional check
HHow to compensate probes
HHow to use the self calibration routine
HHow to match your probe attenuation factor
For more detailed information about instrument operation, refer to your user
manual.
TDS 200 Series Digital Oscilloscope Service Manual
2--1
Operating Information
General Features
H100 MHz (TDS 220 and TDS 224) or 60 MHz (TDS 210) bandwidth with
selectable 20 MHz bandwidth limit
H1 GS/s sample rate and 2,500 point record length for each channel
HCursors with readout
HFive automated measurements
HHigh-resolution, high-contrast LCD display with temperature compensation
and replaceable backlight
HSetup and waveform storage
HAutoset for quick setup
HWaveform averaging and peak detection
HDigital real-time oscilloscope (at least ten-times over sampling)
HDual time base
HVideo trigger capability
HRS-232, GPIB, and Centronics communication ports easily added with
optional extension modules
HVariable persistence display
HUser interface available in ten user selectable languages
2--2
TDS 200 Series Digital Oscilloscope Service Manual
Installation
Operating Information
Power Cord
Use only power cords designed for your oscilloscope. Use a power source that
delivers 85 to 275 VAC
,47to63Hz.Refertopage10--9foralistof
RMS
available power cords.
Use the power cord notch to help route the cord to the rear of the instrument and
avoid inadvertently disconnecting the power source.
Power cord
notch
Securing cable
Figure 2-1: Routing the power cord and security cable
Security Loop
Use the built-in cable channels to secure both your instrument and extension
module to your location.
TDS 200 Series Digital Oscilloscope Service Manual
2--3
Operating Information
Extension Modules
You can increase the feature set of your oscilloscope by inserting an extension
module. Refer to page 10--9 for information about the available modules.
Figure 2-2 shows the proper way to install a module without bending the
connecting pins.
CAUTION. Electrostatic discharge (ESD) can damage components in the
extension module and the oscilloscope. Do not operate your instrument with the
extension module connector exposed.
1
2
Figure 2-2: Installing an extension module
2--4
TDS 200 Series Digital Oscilloscope Service Manual
Functional Check
Operating Information
Perform this quick functional check to verify that your instrument is operating
correctly.
ON/OFF
button
PASSED
PROBE COMP
1. Turn on the instrument.
Wait until the display shows that all self tests passed. Push the SAVE/RECALL
button, select Setups in the top menu box and push the Recall Factory menu
box. The default Probe menu attenuation setting is 10X.
2. Set the switch to 10X on the P2100 probe and connect the oscilloscope probe to
channel 1. To do this, align the slot in the probe connector with the key on the
CH 1 BNC, push to connect, and twist to the right to lock the probe in place.
Attach the probe tip and reference lead to the PROBE COMP connectors.
CH 1
3. Push the AUTOSET button. Within a few seconds, you should see a square
wave in the display (approximately 5 V at 1 kHz peak-to-peak).
Push the CH 1 MENU button twice to turn off channel 1, push the CH 2 MENU
button to turn on channel 2, repeat steps 2 and 3. For TDS 224, repeat for CH 3
and CH 4.
TDS 200 Series Digital Oscilloscope Service Manual
2--5
Operating Information
A
Probe Compensation
Perform this adjustment to match your probe to the input channel. This should be
done whenever you attach a probe for the first time to any input channel.
PROBE
COMP
Overcompensated
Undercompensated
Compensated correctly
UTOSE
T button
CH 1
1. Set the Probe menu attenuation to 10X. Set the switch to 10X on the P2100
probe and connect the oscilloscope probe to channel 1. Attach the probe tip and
reference lead to the PROBE COMP connectors and then press AUTOSET.
If using the probe hook-tip, ensure a proper connection by firmly twisting the
tip onto the probe.
2. Check the shape of the displayed waveform.
3. If necessary, adjust your probe.
Repeat as necessary.
2--6
TDS 200 Series Digital Oscilloscope Service Manual
Self Calibration
Probe Safety
Operating Information
The self calibration routine lets you quickly optimize the oscilloscope signal path
for maximum measurement accuracy. You can run the routine at anytime but you
should always run the routine if the ambient temperature changes by 5_ Cor
more.
To compensate the signal path, disconnect any probes or cables from the channel
input connectors. Then, press the UTILITY button and select Do Self Cal to
confirm that you are ready to proceed.
A guard around the probe body provides a finger barrier for protection from
electric shock.
WARNING. To avoid electric shock when using the probe, keep fingers behind the
guard on the probe body.
To avoid electric shock while using the probe, do not touch metallic portions of
the probe head while it is connected to a voltage source.
Connect the probe to the instrument and connect the ground terminal to ground
before you take any measurements.
Probe Attenuation Setting
Probes are available with various attenuation factors which affect the vertical
scale readout of the oscilloscope.
To change (or check) the probe attenuation setting, press the VERTICAL MENU
button (of the channel you’re using) and then press the menu selection next to
Probe until the correct setting is displayed.
This setting remains in effect until changed again.
Finger guard
TDS 200 Series Digital Oscilloscope Service Manual
2--7
Operating Information
NOTE. The default Probe menu attenuation setting is 10X when the oscilloscope
is shipped.
Be sure that the Attenuation switch on the P2100 probe is set to match the Probe
menu selection in the oscilloscope. The probe switch settings are 1X and 10X.
Attenuation switch
NOTE. When the Attenuation switch is set to 1X, the P2100 probe limits the
bandwidth of the oscilloscope to 7 MHz. To use the full bandwidth of the
oscilloscope, be sure to set the switch to 10X.
Factory Setup
The table below lists the state of the instrument after you select Recall Factory
from the SAVE/RECALL menu.
Tab le 2 - -1: Factory setup settings
ControlSetting
Acquire modeSample
Acquire # of averages16
Channel selectionChannel 1 on
Cursor H Bar position--3.2 divisions from the center
Cursor V Bar position--2 divisions from the center
Cursor functionOff
Cursor time unitsSeconds
Delayed time base time/div
Delay time, delayed runs after main
Display formatYT
Display graticule typeFull
Display contrast50%
Display styleVectors
Display accumulate time500 ms
Edge trigger couplingDC
50 s/div
50 s
2--8
TDS 200 Series Digital Oscilloscope Service Manual
Loading...
+ 92 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.