Tektronix tds210 schematic

Service Manual
TDS 200-Series Digital Real-Time Oscilloscope
071-0492-03
This document supports firmware version 1.00 and above.
Warning
The servicing instructions are for use by qualified personnel only. To avoid personal injury, do not perform any servicing unless you are qualified to do so. Refer to all safety summaries prior to performing service.
www.tektronix.com
Copyright © Tektronix, Inc. All rights reserved.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes that in all previously published material. Specifications and price change privileges reserved.
Tektronix, Inc., P.O. Box 500, Beaverton, OR 97077
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.

WARRANTY

Tektronix warrants that the products that it manufactures and sells will be free from defects in materials and workmanship for a period of three (3) years from the date of purchase from an authorized Tektronix distributor. If any such product proves defective during this warranty period, Tektronix, at its option, either will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product. Batteries are excluded from this warranty.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the warranty period and make suitable arrangements for the performance of service. Customer shall be responsible for packaging and shipping the defective product to the service center designated by Tektronix, shipping charges prepaid, and with a copy of customer proof of purchase. Tektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the Tektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage resulting from attempts by personnel other than Tektronix representatives to install, repair or service the product; b) to repair damage resulting from improper use or connection to incompatible equipment; c) to repair any damage or malfunction caused by the use of non-Tektronix supplies; or d) to service a product that has been modified or integrated with other products when the effect of such modification or integration increases the time or difficulty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THE LISTED PRODUCTS IN LIEU OF ANY OTHER WARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX’ RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.

Table of Contents

Specifications
Operating Information
General Safety Summary v...................................
Service Safety Summary vii....................................
Preface ix...................................................
Related Manuals ix.................................................
Contacting Tektronix x..............................................
Product End-of-Life Handling x......................................
Specifications 1--1.............................................
Operating Information 2--1.....................................
General Features 2--2.................................................
Installation 2--3......................................................
Extension Modules 2--4...............................................
Functional Check 2--5................................................
Probe Compensation 2--6..............................................
Self Calibration 2--7..................................................
Probe Safety 2--7....................................................
Probe Attenuation Setting 2--7..........................................
Factory Setup 2--8...................................................
Theory of Operation
Theory of Operation 3--1.......................................
Main Board 3--1.....................................................
Power Supply 3--2...................................................
Display Module 3--2..................................................
Front Panel 3--2.....................................................
Extension Modules 3--2...............................................
Performance Verification
Performance Verification 4--1...................................
Test Record 4--2.....................................................
Performance Verification Procedures 4--4.................................
Adjustment Procedures
Adjustment Procedures 5--1.....................................
Required Equipment 5--1..............................................
Adjustment Procedure 5--2.............................................
TDS 200 Series Digital Oscilloscope Service Manual
i
Table of Contents
Maintenance
Options
Electrical Parts List
Diagrams
Maintenance 6--1..............................................
Preparation 6--1.....................................................
Preventing ESD 6--1..................................................
Inspection and Cleaning 6--2...........................................
Removal and Installation Procedures 6--5.................................
Troubleshooting 6--38.................................................
Repackaging Instructions 6--48..........................................
Options 7--1..................................................
Electrical Parts List 8--1........................................
Diagrams 9--1.................................................
Replaceable Parts
Replaceable Parts 10--1..........................................
Parts Ordering Information 10--1.........................................
Using the Replaceable Parts List 10--2.....................................
ii
TDS 200 Series Digital Oscilloscope Service Manual

List of Figures

Table of Contents
Figure 2--1: Routing the power cord and security cable 2--3..........
Figure 2--2: Installing an extension module 2--4....................
Figure 3--1: Module-level block diagram (two channel) 3--3..........
Figure 3--2: Module-level block diagram (four channel) 3--4..........
Figure 5--1: Service menu enable button 5--2.......................
Figure 5--2: Adjustment setups 5--3...............................
Figure 6--1: Removing the rear feet 6--8...........................
Figure 6--2: Installing the rear feet 6--9............................
Figure 6--3: Installing a front-case label 6--10.......................
Figure 6--4: Installing a new rear-case label 6--11....................
Figure 6--5: Removing the handle 6--12............................
Figure 6--6: Installing the handle 6--13.............................
Figure 6--7: Removing and installing the rear cover 6--15.............
Figure 6--8: Removing and installing the front feet 6--16..............
Figure 6--9: Removing and installing the EMI clips 6--17..............
Figure 6--10: Removing the flip stand 6--18.........................
Figure 6--11: Installing the flip stand 6--19..........................
Figure 6--12: Line fuse location 6--20..............................
Figure 6--13: Removing the power supply module 6--21...............
Figure 6--14: Installing the power supply module 6--22...............
Figure 6--15: Removing and installing the internal assembly 6--24......
Figure 6--16: Installing the copper mesh grounding tube 6--25.........
Figure 6--17: Main board removal 6--26............................
Figure 6--18: Main board installation 6--27.........................
Figure 6--19: Removing the display module 6--28....................
Figure 6--20: Installing the display module 6--30.....................
Figure 6--21: Removing the front panel module 6--31.................
Figure 6--22: Installing the front panel module 6--32.................
Figure 6--23: Removing and installing the keypad 6--33...............
Figure 6--24: Removing the display shield 6--35......................
Figure 6--25: Installing the display shield 6--36......................
Figure 6--26: Measuring the backlight voltage 6--41..................
Figure 10--1: Exploded diagram 10--7..............................
TDS 200 Series Digital Oscilloscope Service Manual
iii
Table of Contents

List of Tables

Table 1--1: Specifications 1--1...................................
Table 1--2: General specifications 1--5............................
Table 2--1: Factory setup settings 2--8............................
Table 5--1: Required equipment 5--1.............................
Table 5--2: TDS 210 and TDS 220 Adjustment steps 5--5.............
Table 5--3: TDS 224 Adjustment steps 5--7........................
Table 6--1: Internal inspection check list 6--3......................
Table 6--2: List of procedures 6--7................................
Table 6--3: List of error codes 6--46...............................
Table 10--1: Parts list column descriptions 10--2.....................
Table 10--2: Manufacturers cross index 10--3.......................
Table 10--3: Replaceable parts list 10--3............................
Table 10--4: Replaceable standard accessories 10--8..................
Table 10--5: Replaceable optional accessories 10--8..................
iv
TDS 200 Series Digital Oscilloscope Service Manual

General Safety Summary

Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it.
Only qualified personnel should perform service procedures.
Injury Precautions
Use Proper Power Cord. To avoid fire hazard, use only the power cord specified for this product.
Avoid Electric Overload. To avoid electric shock or fire hazard, do not apply a voltage to a terminal that is outside the range specified for that terminal.
Avoid Overvoltage. To avoid electric shock or fire hazard, do not apply potential to any terminal, including the common terminal, that varies from ground by more than the maximum rating for that terminal.
Avoid Electric Shock. To avoid injury or loss of life, do not connect or disconnect probes or test leads while they are connected to a voltage source.
Ground the Product. This product is grounded through the grounding conductor of the power cord. To avoid electric shock, the grounding conductor must be connected to earth ground. Before making connections to the input or output terminals of the product, ensure that the product is properly grounded.
Connect the Probe Properly. The probe ground lead is at ground potential. Do not connect the ground lead to an elevated voltage.
Do Not Operate Without Covers. To avoid electric shock or fire hazard, do not operate this product with covers or panels removed.
Use Proper Fuse. To avoid fire hazard, use only the fuse type and rating specified for this product.
Do Not Operate in Wet/Damp Conditions. To avoid electric shock, do not operate this product in wet or damp conditions.
Do Not Operate in an Explosive Atmosphere. To avoid injury or fire hazard, do not operate this product in an explosive atmosphere.
Product Damage
Precautions
TDS 200 Series Digital Oscilloscope Service Manual
Use Proper Power Source. Do not operate this product from a power source that applies more than the voltage specified.
Provide Proper Ventilation. To prevent product overheating, provide proper ventilation.
Do Not Operate With Suspected Failures. If you suspect there is damage to this product, have it inspected by qualified service personnel.
v
General Safety Summary
Symbols and Terms
Terms in this Manual. These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in damage to this product or other property.
Terms on the Product. These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the marking.
WARNING indicates an injury hazard not immediately accessible as you read the marking.
CAUTION indicates a hazard to property including the product.
Symbols on the Product. The following symbols may appear on the product:
Certifications and
Compliances
DANGER
High Voltage
Protective Ground
(Earth) Terminal
ATTENTION
Refer to Manual
Double
Insulated
Refer to the specifications section for a listing of certifications and compliances that apply to this product.
vi
TDS 200 Series Digital Oscilloscope Service Manual

Service Safety Summary

Only qualified personnel should perform service procedures. Read this Service Safety Summary and the General Safety Summary before performing any service
procedures.
Do Not Service Alone. Do not perform internal service or adjustments of this product unless another person capable of rendering first aid and resuscitation is present.
Disconnect Power. To avoid electric shock, disconnect the main power by means of the power cord or, if provided, the power switch.
Use Care When Servicing With Power On. Dangerous voltages or currents may exist in this product. Disconnect power, remove battery (if applicable), and disconnect test leads before removing protective panels, soldering, or replacing components.
To avoid electric shock, do not touch exposed connections.
TDS 200 Series Digital Oscilloscope Service Manual
vii
Service Safety Summary
viii
TDS 200 Series Digital Oscilloscope Service Manual

Preface

Related Manuals

The service manual for the TDS 200-Series Digital Real-Time Oscilloscope provides information to troubleshoot and repair the instrument to the module level.
Some TDS models have two input channels and an external trigger while other models have four input channels. Most illustrations in this manual show a two channel model. Some parts for the four channel model are slightly different because of the additional channels.
Additional documentation for the instrument is contained in the related manuals listed below.
Extension module
User manual
Language
English 071-0398-XX 071-0409-XX 071-0493-XX
French 071-0400-XX* 071-0483-XX
German 071-0402-XX* 071-0485-XX
Italian 071-0401-XX* 071-0484-XX
Spanish 071-0399-XX* 071-0482-XX
Portuguese 071-0403-XX* 071-0486-XX
Japanese 071-0405-XX* 071-0488-XX
Korean 071-0408-XX* 071-0491-XX
Simplified Chinese 071-0406-XX* 071-0489-XX
Traditional Chinese 071-0407-XX* 071-0490-XX
Russian 071-0404-XX 071-0487-XX
*These manuals contain a language overlay for the front-panel controls.
part number
instructions part number
Programmer manual part number
TDS 200 Series Digital Oscilloscope Service Manual
ix
Preface

Contacting Tektronix

Phone 1-800-833-9200*
Address Tektronix, Inc.
Department or name (if known) 14200 SW Karl Braun Drive P.O. Box 500 Beaverton, OR 97077 USA
Web site www.tektronix.com
Sales support 1-800-833-9200, select option 1*
Service support 1-800-833-9200, select option 2*
Technical support Email: techsupport@tektronix.com
* This phone number is toll free in North America. After office hours, please leave a
voice mail message. Outside North America, contact a Tektronix sales office or distributor; see the Tektronix web site for a list of offices.

Product End-of-Life Handling

Components that Contain Mercury. The cold cathode fluorescent tube located in the liquid crystal display backlight contains trace elements of mercury. When you are ready to reclaim the instrument, you must properly transfer it according to local regulations concerning mercury-containing equipment or ship the instru­ment to the Tektronix Recycling Operations (RAMS). You can contact Tektronix for the RAMS shipping address and instructions.
1-800-833-9200, select option 3* 1-503-627-2400
6:00 a.m. -- 5:00 p.m. Pacific time
x
TDS 200 Series Digital Oscilloscope Service Manual

Specifications

g
q
p
,
g
All specifications apply to the TDS 200-Series Digital Real-Time Oscilloscope with a P2100 probe with the Attenuation switch set to 10X unless noted otherwise. To meet specifications, two conditions must first be met:
H The instrument must have been operating continuously for twenty minutes
within the specified operating temperature.
H You must perform the Self Cal operation, accessible through the utility menu,
if the operating temperature changes by more than 5° C.
All specifications are guaranteed unless noted “typical.” Specifications that are marked with the n symbol are checked in the chapter Performance Verification.
Table 1--1: Specifications
Acquisition
Acquisition Modes Sample, Peak detect, and Average
Acquisition Rate, typical
Single Sequence Acquisition Mode Acquisition Stops After
Up to 180 waveforms per second, per channel (Sample acquisition mode, no measurements)
Sample, Peak Detect Single acquisition, all channels simultaneously
Average N acquisitions, all channels simultaneously, N is
selectable from 4, 16, 64, and 128
Inputs
Input Coupling DC, AC, or GND
Input Impedance, DC Coupled, all channela
Input Impedance, DC Coupled, EXT TRIG only
P2100 Probe Attenuation 1X, 10X
Probe Attenuation Factors 1X, 10X, 100X, 1000X
Maximum Voltage Between Overvoltage Category Maximum Voltage Signal and Common at input BNC
* Bandwidth is not valid for the P2100 probe when the switch is set to 1X.
1M±2% in parallel with 20 pF ±3pF
TDS 210 (B099188 to B119999 & C021679 to C029999) TDS 220 (B065810 to B079999 & C021127 to C029999)
1.2 M±5% in parallel with 20 pF ±5pF
CAT I and CAT II
CAT III 150 V
For steady-state sinusoidal waveforms, derate at 20 dB/decade above 100 kHz to 13 Vpkat 3 MHz* and above. Also, refer to Overvoltage Category description on page 1--7.
All other TDS 210, TDS 220, and TDS 224
1M±5% in parallel with 20 pF ±5pF
300 V width < 100 msec.)
(420 V peak, duty factor < 50%, pulse
RMS
RMS
TDS 200 Series Digital Oscilloscope Service Manual
1--1
Specifications
g
inputBN
C
g
tBN
C
ithP210
0
g
5s/div),t
ypical
Table 1--1: Specifications (cont.)
Inputs
Maximum Voltage Between Overvoltage Category Maximum Voltage Probe Tip and ground using P2100 connected to
Time delay between channels, typical
Channel-to-Channel TDS 210 TDS 220 and TDS 224 Common Mode Rejection, typical
Channel-to-Channel Cross- TDS 210 TDS 220 and TDS 224 talk
CAT I and CAT II 300 V
(500 V peak, duty factor < 35%, pulse
RMS
width < 100 msec.)
CAT III 100 V
Derate at 20 dB/decade above 900 kHz to 13 V
at 27 MHz* and above. Also, refer to Overvoltage
RMS
RMS
Category description on page 1--7.
150 ps
100:1 at 60 Hz 20:1at30MHz*
100:1 at 60 Hz 20:1at50MHz*
Measured on MATH Ch1 -- Ch2 waveform, with test signal applied between signal and common of both channels, and with the same VOLTS/DIV and coupling settings on each channel. Also measured on MATH Ch3 -- Ch4 waveform for the TDS 224.
100:1 at 30 MHz*
100:1 at 50 MHz*
Measured on one channel, with 8 division test signal applied between signal and common of the other channel, the same VOLTS/DIV and coupling settings on each channel, and 50 terminators on each channel.
Vertical
Digitizers 8 bit resolution (except when set to 2 mV/div), each channel sampled simultaneously
VOLTS/DIV Range 2 mV/div to 5 V/div at input BNC
(Full bandwidth at >5 mV/div to 5 V/div, 20 MHz at 2 mV/dif to 5 mV/div, except in Peak Detect mode full bandwidth at >10 mV/div to 5 V/div [20 MHz at 2 mV/div to 10 mV/div])
Position Range 2 mV/div to 200 mV/div, ±2V
> 200 mV/div to 5 V/div, ±50 V
n Analog Bandwidth in
Sample and Average modes a
or w
DC Coupled
Analog Bandwidth in Peak TDS 210 TDS 220 and TDS 224 Detect mode (5 s/div to
Selectable Analog Band­width Limit, typical
probe,
Lower Frequency Limit, AC Coupled
TDS 210 TDS 220 and TDS 224
60 MHz* (when vertical scale set to > 5 mV/div) 100 MHz* (when vertical scale set to > 5 mV/div)
* (when vertical scale set to 5mV/div)
20 MHz
50 MHz* (when vertical scale set to > 10 mV/div) 75 MHz* (when vertical scale set to > 10 mV/div)
20 MHz* (when vertical scale set to 10 mV/div)
20 MHz*
10 Hz at BNC
1 Hz when using a 10X passive probe
* Bandwidth is not valid for the P2100 probe when the switch is set to 1X.
1--2
TDS 200 Series Digital Oscilloscope Service Manual
Table 1--1: Specifications (cont.)
,yp
y
Vertical
Rise Time at BNC, typical TDS 210 TDS220 and TDS 224
<5.8 ns <3.5 ns
Peak Detect Response
n DC Gain Accuracy
DC Measurement Accu- Measurement Type Accuracy racy, Average Acquisition Mode
Captures 50% or greater amplitude of pulses 10 ns wide (5 s/div to 5 s/div) in the center 6 vertical divisions. The oscilloscope reverts to Sample mode when the sec/div (Horizontal scale) setting is from
2.5 s/div to 5 ns/div. The Sample mode can still capture 10 ns glitches.
TDS210 and TDS220 TDS224
2 mV/div and larger, ±3% for Sample or Average acquisition mode
2 mV/div to 5 mV/div, ±4% for Sample or Average acquisition mode
10 mV/div and larger, ±3% for Sample or Average acquisition mode
Average of 16 waveforms with vertical position at zero
TDS224:
±(4% × reading + 0.1 div + 1 mV) and either 2 mV/div or 5 mV/div is selected.
Specifications
±(3% × reading + 0.1 div + 1 mV) and 10 mV/div or greater is selected
TDS210 and TDS220:
±(3% × reading + 0.1 div + 1 mV) for all ranges
Average of 16 waveforms with vertical position not at zero
TDS210, TDS220 and TDS224:
±[3% × (reading + vertical position) + 1% of vertical position + 0.2 div]
Add 2 mV for settings from 2 mV/div to 200 mV/ div. Add 50 mV for settings from >200 mV/div to 5V/div.
Delta Volts Measurement Accuracy, Average Acquisi­tion Mode
P2100 passive probe
Bandwidth DC to 100 MHz DC to 7 MHz
Attenuation ratio 10:1 1:1
Compensation Range 18 pf-35 pf
Input resistance
Input capacitance 14.5 pf-17.5 pf 80 pf-110 pf
Maximum input voltage
Delta volts between any two averages of 16
3% of reading + 0.05 div waveforms acquired under same setup and ambient conditions
10X position 1X position
All oscilloscopes with 1 M input
10 M ± 2% 1M± 2%
300 V 300 V 100 V
CAT I or 300 V DC CAT I
RMS
CATIIor300VDCCATII
RMS
CAT III or 100 V DC CAT III
RMS
150 V
150 V
100 V
RMS RMS RMS
Horizontal
Sample Rate Range 50 S/s to 1 GS/s
Waveform Interpolation (sin x)/x
CAT I or 150 V DC CAT I CATIIor150VDCCATII CAT III or 100 V DC CAT III
TDS 200 Series Digital Oscilloscope Service Manual
1--3
Specifications
ggy,g
ggg
ggy
,
Table 1--1: Specifications (cont.)
Horizontal
Record Length 2500 samples for each channel
SEC/DIV Range 5 ns/div to 5 s/div, in a 1, 2.5, 5 sequence
n Sample Rate and Delay Time Accuracy
Delta Time Measurement Conditions Accuracy Accuracy (Full Bandwidth)
Position Range 5ns/divto10ns/div
Trigger
±100 ppm over any 1 ms time interval
Single-shot, Sample mode
>16 averages ±(1 sample interval + 100 ppm × reading +
Sample interval = s/div ÷ 250
25 ns/div to 100 s/div 250 s/divto5s/div
(--4 div × s/div) to 20 ms (--4 div × s/div) to 50 ms (--4 div × s/div) to 50 s
±(1 sample interval + 100 ppm × reading +
0.6 ns)
0.4 ns)
n Trigger Sensitivity, Edge Trigger Type
Trigger Sensitivity, Edge Trigger Type, typical
Trigger Level Range Source Range
Trigger Level Accuracy, Accuracies are for signals having rise and fall times 20 ns typical
Coupling Sensitivity
DC
Coupling Sensitivity
AC Same as DC at 50 MHz and above
NOISE REJ Reduces DC-coupled trigger sensitivity by 2 times for > 10 mV/div to 5 V/div
HF REJ Same as DC-coupled limit from DC to 7 kHz, attenuates signals above
LF REJ Same as the DC-coupled limits for frequencies above 300 kHz, attenuates
Internal ±8 divisions from center of screen
TDS 210 and TDS 220
EXT ±1.6 V
EXT/5 ±8V
Source
Internal
CH 1, CH 2, CH 3 & CH 4
TDS 210 and TDS 220
EXT 100 mV from DC to 10 MHz*,
EXT/5 500 mV from DC to 10 MHz*,
80 kHz
signals below 300 kHz
Accuracy
±0.2 div × volts/div within ±4 divisions from center screen
1 div from DC to 10 MHz*,
1.5 div from 10 MHz* to Full
150 mV from 10 MHz* to Full
750 mV from 10 MHz* to Full
1--4
TDS 200 Series Digital Oscilloscope Service Manual
Table 1--1: Specifications (cont.)
y
,
Trigger
TDS 210 and TDS 220
EXT ±(6% of setting + 40 mV)
EXT/5 ±(6% of setting + 200 mV)
* Bandwidth is not valid for the P2100 probe when the switch is set to 1X.
SET LEVEL TO 50%, typical
Default Settings, Video Trigger
Sensitivity, Video Composite video signal Trigger Type, typical
Signal Formats and Field Rates, Video Trigger Type
Holdoff Range 500 ns to 10 s
Operates with input signals 50 Hz
Trigger mode is Auto and Coupling is AC
Source
Internal Pk-pk amplitude of 2 divisions
TDS 210 and TDS 220
EXT 400 mV
EXT/5 2V
Supports NTSC, PAL, and SECAM broadcast systems for any field or any line
Range
Specifications
Measurements
Cursors
Automated Measurements Cycle RMS, Mean, Pk -- Pk, Period, Frequency
Voltage difference between cursors (V) Time difference between cursors (T) Reciprocal of T in Hertz (1/T)
Table 1--2: General specifications
Display
Display Type 5.9 in (152 mm) diagonal liquid crystal
Display Resolution 320 horizontal by 240 vertical pixels
Display Contrast Adjustable, temperature compensated
Backlight Intensity, typical 60 cd/m
Probe compensator output
Output Voltage, typical
Frequency, typical 1 kHz
2
5Vinto1MΩ load
TDS 200 Series Digital Oscilloscope Service Manual
1--5
Specifications
p
y
Table 1--2: General specifications (cont.)
Power Source
Source Voltage
100 -- 120 VAC
10% from 45 Hz through 440 Hz, CAT II
RMS
120 -- 250 VAC
10% from 45 Hz through 66 Hz, CAT II
RMS
Power Consumption TDS 210 and TDS 220 TDS 224
Less than 20 W Less than 25 W
Fuse 1 A, T rating, 250 V
Environmental
Temperature Operating 0° Cto+50° C
Nonoperating -- 2 0 ° Cto+60° C
Cooling Method Convection
Humidity +40° C or below 90% relative humidity
+41° Cto+50° C ≤60% relative humidity
Altitude Operating 3,000 m
Nonoperating 15,000 m
Random Vibration Operating 0.31 g
from 5 Hz to 500 Hz, 10 minutes on each
RMS
axis
Nonoperating 2.46 g
from 5 Hz to 500 Hz, 10 minutes on each
RMS
axis
Mechanical Shock Operating 50 g, 11 ms, half sine
Mechanical
Size Height 151.4 mm (5.96 in.)
Width 304.8 mm (12 in.)
Depth 120.7 mm (4.75 in.)
Weight (approximate) When packaged for domestic shipment 3.6 kg (8.0 lbs)
1--6
TDS 200 Series Digital Oscilloscope Service Manual
Table 1--2: General specifications (cont.)
Certifications and compliances
Specifications
EC Declaration of Conformity
(TDS 210, TDS 220 & TDS 224)
(TDS 210, TDS 220, TDS 224, P2100)
(P2100)
Certifications
(TDS 210, TDS 220, TDS 224)
(P2100)
CSA Certified Power Cords CSA Certification includes the products and power cords appropriate for use in the North America power
Pollution Degree 2 Do not operate in environments where conductive pollutants may be present.
Overvoltage Category
Meets intent of Directive 89/336/EEC for Electromagnetic Compatibility*; meets the requirements of FCC Code of Federal Regulations, 47 CFR, Part 15, Subpart B, Class A:
CISPR 11 (Class A): Electromagnetic radiated and conducted emissions EN 50082-1 European Community Requirements:
IEC 61000-4-2 electrostatic discharge IEC 61000-4-3 radiated susceptibility** IEC 61000-4-4 power line fast transients IEC 61000-4-5 power line surge immunity IEC 61000-4-6 conducted immunity to RF fields IEC 61000-4-11 voltage dips, interrupts and variations
AS/NZS 2064, Australian emissions standard for Industrial, Scientific, and Medical Equipment
Low Voltage Directive 73/23/EEC as amended by 93/68/EED:
EN61010-1/A2:1995 Safety requirements for electrical equipment for measurement control, and
laboratory use
EN 61010-2-031:1994 Particular requirements for hand-held probe assemblies for electrical
measurement and test
CAN/CSA C22.2 No. 1010.1-92 UL3111-1, First Edition
IEC61010-1/A2 IEC61010-2-031
network. All other power cords supplied are approved for the country of use.
Category: Examples of Products in this Category:
CAT III Distribution-level mains, fixed installation
CAT II Local-level mains, appliances, portable equipment
CAT I Signal levels in special equipment or parts of equipment, telecommunications,
electronics
* Instrument tested with shielded cables.
** Meets Performance Criterion A with allowed degradation of up to 5 major divisions of trace noise increase from 80 MHz to
200 MHz and up to 2 divisions of trace noise increase over the range of 200 MHz to 1000 MHz under a 3 V/m RF field. Ambient RF fields may induce triggering when trigger threshold is offset less than 2.5 major divisions from ground reference.
Adjustment interval
The recommended adjustment interval is one year
TDS 200 Series Digital Oscilloscope Service Manual
1--7
Specifications
1--8
TDS 200 Series Digital Oscilloscope Service Manual

Operating Information

TDS 200-Series Digital Oscilloscopes are two or four channel oscilloscopes in small, lightweight, benchtop packages that you can use to take ground-referenced measurements.
In addition to the list of general features, this chapter covers the following topics:
H How to install your product
H How to add extended functions
H How to perform a brief functional check
H How to compensate probes
H How to use the self calibration routine
H How to match your probe attenuation factor
For more detailed information about instrument operation, refer to your user manual.
TDS 200 Series Digital Oscilloscope Service Manual
2--1
Operating Information

General Features

H 100 MHz (TDS 220 and TDS 224) or 60 MHz (TDS 210) bandwidth with
selectable 20 MHz bandwidth limit
H 1 GS/s sample rate and 2,500 point record length for each channel
H Cursors with readout
H Five automated measurements
H High-resolution, high-contrast LCD display with temperature compensation
and replaceable backlight
H Setup and waveform storage
H Autoset for quick setup
H Waveform averaging and peak detection
H Digital real-time oscilloscope (at least ten-times over sampling)
H Dual time base
H Video trigger capability
H RS-232, GPIB, and Centronics communication ports easily added with
optional extension modules
H Variable persistence display
H User interface available in ten user selectable languages
2--2
TDS 200 Series Digital Oscilloscope Service Manual

Installation

Operating Information
Power Cord
Use only power cords designed for your oscilloscope. Use a power source that delivers 85 to 275 VAC
,47to63Hz.Refertopage10--9foralistof
RMS
available power cords.
Use the power cord notch to help route the cord to the rear of the instrument and avoid inadvertently disconnecting the power source.
Power cord
notch
Securing cable
Figure 2-1: Routing the power cord and security cable
Security Loop
Use the built-in cable channels to secure both your instrument and extension module to your location.
TDS 200 Series Digital Oscilloscope Service Manual
2--3
Operating Information

Extension Modules

You can increase the feature set of your oscilloscope by inserting an extension module. Refer to page 10--9 for information about the available modules. Figure 2-2 shows the proper way to install a module without bending the connecting pins.
CAUTION. Electrostatic discharge (ESD) can damage components in the extension module and the oscilloscope. Do not operate your instrument with the extension module connector exposed.
1
2
Figure 2-2: Installing an extension module
2--4
TDS 200 Series Digital Oscilloscope Service Manual

Functional Check

Operating Information
Perform this quick functional check to verify that your instrument is operating correctly.
ON/OFF
button
PASSED
PROBE COMP
1. Turn on the instrument.
Wait until the display shows that all self tests passed. Push the SAVE/RECALL button, select Setups in the top menu box and push the Recall Factory menu box. The default Probe menu attenuation setting is 10X.
2. Set the switch to 10X on the P2100 probe and connect the oscilloscope probe to
channel 1. To do this, align the slot in the probe connector with the key on the CH 1 BNC, push to connect, and twist to the right to lock the probe in place.
Attach the probe tip and reference lead to the PROBE COMP connectors.
CH 1
3. Push the AUTOSET button. Within a few seconds, you should see a square
wave in the display (approximately 5 V at 1 kHz peak-to-peak).
Push the CH 1 MENU button twice to turn off channel 1, push the CH 2 MENU button to turn on channel 2, repeat steps 2 and 3. For TDS 224, repeat for CH 3 and CH 4.
TDS 200 Series Digital Oscilloscope Service Manual
2--5
Operating Information
A

Probe Compensation

Perform this adjustment to match your probe to the input channel. This should be done whenever you attach a probe for the first time to any input channel.
PROBE
COMP
Overcompensated
Undercompensated
Compensated correctly
UTOSE
T button
CH 1
1. Set the Probe menu attenuation to 10X. Set the switch to 10X on the P2100
probe and connect the oscilloscope probe to channel 1. Attach the probe tip and reference lead to the PROBE COMP connectors and then press AUTOSET.
If using the probe hook-tip, ensure a proper connection by firmly twisting the tip onto the probe.
2. Check the shape of the displayed waveform.
3. If necessary, adjust your probe.
Repeat as necessary.
2--6
TDS 200 Series Digital Oscilloscope Service Manual

Self Calibration

Probe Safety

Operating Information
The self calibration routine lets you quickly optimize the oscilloscope signal path for maximum measurement accuracy. You can run the routine at anytime but you should always run the routine if the ambient temperature changes by 5_ Cor more.
To compensate the signal path, disconnect any probes or cables from the channel input connectors. Then, press the UTILITY button and select Do Self Cal to confirm that you are ready to proceed.
A guard around the probe body provides a finger barrier for protection from electric shock.
WARNING. To avoid electric shock when using the probe, keep fingers behind the guard on the probe body.
To avoid electric shock while using the probe, do not touch metallic portions of the probe head while it is connected to a voltage source.
Connect the probe to the instrument and connect the ground terminal to ground before you take any measurements.

Probe Attenuation Setting

Probes are available with various attenuation factors which affect the vertical scale readout of the oscilloscope.
To change (or check) the probe attenuation setting, press the VERTICAL MENU button (of the channel youre using) and then press the menu selection next to Probe until the correct setting is displayed.
This setting remains in effect until changed again.
Finger guard
TDS 200 Series Digital Oscilloscope Service Manual
2--7
Operating Information
NOTE. The default Probe menu attenuation setting is 10X when the oscilloscope is shipped.
Be sure that the Attenuation switch on the P2100 probe is set to match the Probe menu selection in the oscilloscope. The probe switch settings are 1X and 10X.
Attenuation switch
NOTE. When the Attenuation switch is set to 1X, the P2100 probe limits the bandwidth of the oscilloscope to 7 MHz. To use the full bandwidth of the oscilloscope, be sure to set the switch to 10X.

Factory Setup

The table below lists the state of the instrument after you select Recall Factory from the SAVE/RECALL menu.
Tab le 2 - -1: Factory setup settings
Control Setting
Acquire mode Sample
Acquire # of averages 16
Channel selection Channel 1 on
Cursor H Bar position --3.2 divisions from the center
Cursor V Bar position --2 divisions from the center
Cursor function Off
Cursor time units Seconds
Delayed time base time/div
Delay time, delayed runs after main
Display format YT
Display graticule type Full
Display contrast 50%
Display style Vectors
Display accumulate time 500 ms
Edge trigger coupling DC
50 s/div
50 s
2--8
TDS 200 Series Digital Oscilloscope Service Manual
Tab le 2 - -1: Factory setup settings (cont.)
Control Setting
Edge trigger level 0.0 V
Edge trigger slope Rising
Edge trigger source Channel 1
Horizontal -- main trigger position 50%
Horizontal -- time base Main only
Main time base time/div
Math waveform function CH1 + CH2
Saved setups No change
Saved waveforms No change
Trigger holdoff Minimum (500 ns)
Trigger mode Auto
Trigger type Edge
Vertical bandwidth (all channels) Full
Vertical coupling (all channels) DC
Vertical position (all channels) 0div
Vertical volts/div. (all channels) 100 mV/div
500 s/div
Operating Information
TDS 200 Series Digital Oscilloscope Service Manual
2--9
Operating Information
2--10
TDS 200 Series Digital Oscilloscope Service Manual

Theory of Operation

This chapter describes the electrical operation of the TDS 200-Series Digital Real-Time Oscilloscope to the module level. It describes the basic operation of each functional circuit block shown in figures 3-1 and 3-2.

Main Board

The Main board contains a two or four channel acquisition system, trigger system, acquisition/display controller, and system processor. Interfaces are provided for two signal inputs, external trigger input, probe compensation, front panel, LCD display, extension module, and power supply. Acquisition and display functions are implemented in two application specific integrated circuits (ASICs).
Acquisition System
Display System
Processor System
Input Signal Interface
Signals from the CH 1, CH 2, and EXT TRIG, or CH 1, CH 2, CH 3, and CH 4 input connectors pass through attenuators to the acquisition ASIC. The acquisi­tion ASIC contains amplifiers and samplers for each input channel, A/D converter, and trigger logic. The digitized waveform samples are transferred to the display ASIC. Low voltage regulators supply isolated power to the acquisi­tion ASIC.
The display system consists of the display ASIC, DRAM memory, and system oscillator. Digitized acquisition samples are processed by the display ASIC and stored into DRAM. The Display ASIC formats the waveform data and generates drive signals for the LCD display. Front panel scanning is also performed by the display ASIC.
The processor system consists of the microprocessor and memory. The processor system interprets the front panel controls, programs acquisition and display parameters, computes waveform measurements, and manages the extension module interface. Saved setups, waveforms, and calibration constants are stored in nonvolatile memory. The processor system shares DRAM with the display system.
BNC connectors are mounted on the main board for the CH 1, CH 2, and trigger or CH 1, CH 2, CH 3, and CH 4 inputs. The signal inputs are compatible with the P2100 probes supplied.
Probe Compensation
TDS 200 Series Digital Oscilloscope Service Manual
The PROBE COMP and ground terminals are provided for probe adjustment.
3--1
Theory of Operation

Power Supply

Display Module

Front Panel

The power supply module generates low voltage DC, a line trigger signal for the main board, and high voltage AC to power the LCD backlight. The power supply operates over a wide power line voltage and frequency range.
The display module consists of a liquid crystal display (LCD), a display driver, and a florescent backlight.
The front panel board holds the encoders and push button contacts that the operator uses to control the instrument. An elastomer mat holds the conductive pads for the push buttons. Circuits on the front panel convert the matrix into serial data which is read by the display ASIC on the main Board.

Extension Modules

Each extension module also contains a GPIB, RS--232, and Centronics parallel printer interface. A ROM on the extension module contains processor instruc­tions to support the communications interfaces. Extension modules interface with the processor system through the extension module port connector.
3--2
TDS 200 Series Digital Oscilloscope Service Manual
AC Line
Theory of Operation
Power Supply
3
AC Line In
GND
P3
Backlight Power
DC Out
&Line
Trigger
2
Front Panel
Probe Comp
Probe Comp
CH1
CH2
EXT
TRIG
J602
J603
P53
5
J604
Signal
J605
GND
J51
A1 Main
GND
DC In
LCD
J131
J202
P131
12
7
Display
J90 1
Backlight
J902
Display Control
Extension
Module
J52
J53
J601
J1 5 0
9
J2
24
J4
25
J3
RS232
GPIB
CENTRONICS
Figure 3-1: Module-level block diagram (two channel)
TDS 200 Series Digital Oscilloscope Service Manual
3--3
Theory of Operation
AC Line
Power Supply
3
AC Line In
GND
P3
Backlight Power
DC Out
&Line
Trigger
2
Front Panel
5
Probe Comp
Probe Comp
CH1
CH2
CH3
Signal
J200
J250
J300
J1
GND
J101
J103
J104
J104
A1 Main
GND
DC In
LCD
J131
J102
J690
P131
5 0
7
12
Extension
Module
J1
Display
J90 1
Backlight
J902
Display Control
9
J2
24
J4
25
J3
RS232
GPIB
CENTRONICS
J350
CH4
Figure 3-2: Module-level block diagram (four channel)
3--4
TDS 200 Series Digital Oscilloscope Service Manual

Performance Verification

p
OscilloscopeCalibration
This chapter contains performance verification procedures for the specifications marked with the n symbol. The following equipment, or a suitable equivalent, is required to complete these procedures.
Description
DC Voltage Source 17.5mVto7V,±0.5%
Leveled Sine Wave Generator 50 kHz and 100 MHz, ±3%
Time Mark Generator 10 ms period, ±25 ppm
Minimum requirements
accuracy
amplitude accuracy
accuracy
Examples
Wavetek 9100 Universal Calibration System with Oscillosco Module (Option 250)
Fluke 5500A Multi-product Calibrator with Oscilloscope Calibration Option (Option 5500A-SC)
e Calibration
50 Ω BNC Cable
50 Ω BNC Cable
50 Feedthrough Termination
Dual Banana to BNC Adapter Banana plugs to BNC female Tektronix part number
BNC T Adapter BNC male to dual BNC female
BNCmaletoBNCmale, 36 in (1 m) long
BNCmaletoBNCmale, 10 in (25 cm) long
BNC male and female connectors
connectors
Tektronix part number 012-0482-00
Tektronix part number 012-0208-00
Tektronix part number 011-0049-01
103-0090-00
Tektronix part number 103-0030-00
TDS 200 Series Digital Oscilloscope Service Manual
4--1
Performance Verification

Test Record

Serial number
Test Passed Failed
Self Test
Oscilloscope tests Low limit Test result High limit
DC Gain Accuracy, TDS210 & TDS220
Channel 1 5 mV/div 34.0 mV 36.0 mV
Procedure performed by Date
200 mV/div 1.358 V 1.442 V
2V/div 13.58 V 14.42 V
Channel 2 5 mV/div 34.0 mV 36.0 mV
200 mV/div 1.358 V 1.442 V
2V/div 13.58 V 14.42 V
DC Gain Accuracy, TDS224
Channel 1 5 mV/div 33.6 mV 36.4 mV
200 mV/div 1.358 V 1.442 V
2V/div 13.58 V 14.42 V
Channel 2 5 mV/div 33.6 mV 36.4 mV
200 mV/div 1.358 V 1.442 V
2V/div 13.58 V 14.42 V
Channel 3 5 mV/div 33.6 mV 36.4 mV
200 mV/div 1.358 V 1.442 V
2V/div 13.58 V 14.42 V
Channel 4 5 mV/div 33.6 mV 36.4 mV
200 mV/div 1.358 V 1.442 V
4--2
2V/div 13.58 V 14.42 V
TDS 200 Series Digital Oscilloscope Service Manual
Performance Verification
Oscilloscope tests High limitTest resultLow limit
Channel 1 Bandwidth 2.12 V
Channel 2 Bandwidth 2.12 V
Channel 3 Bandwidth
Channel 4 Bandwidth
Sample Rate and Delay Time Accuracy -- 2 d i v s +2 divs
Channel 1 Edge Trigger Sensitivity Stable trigger
Channel 2 Edge Trigger Sensitivity Stable trigger
Channel 3 Edge Trigger Sensitivity
Channel 4 Edge Trigger Sensitivity
External Edge Trigger Sensitivity
1
Channels 3 and 4 are only on four channel instruments, and External Edge trigger is only on two channel instruments.
1
1
1
1
1
2.12 V
2.12 V
Stable trigger
Stable trigger
Stable trigger
TDS 200 Series Digital Oscilloscope Service Manual
4--3
Performance Verification

Performance Verification Procedures

Before beginning these procedures, two conditions must first be met:
H The instrument must have been operating continuously for twenty minutes
within the operating temperature range specified.
H You must perform the Self Calibration operation described below. If the
ambient temperature changes by more than 5° C, you must perform the Self Calibration operation again.
The time required to complete the entire procedure is approximately one hour.
WARNING. Some procedures use hazardous voltages. To prevent electrical shock, always set voltage source outputs to 0 V before making or changing any interconnections.
Self Test
Self Calibration
This internal procedure is automatically performed every time the instrument is powered on. No test equipment or hookups are required. Verify that no error messages are displayed before continuing with this procedure.
The self calibration routine lets you quickly optimize the oscilloscope signal path for maximum measurement accuracy. You can run the routine at anytime but you should always run the routine if the ambient temperature changes by 5_ Cor more.
1. Disconnect any probes or cables from the channel 1 and channel 2 input
connectors.
2. Press the UTILITY button and select Do Self Cal to start the routine. The
routine takes approximately one minute to complete.
3. Verify that Self calibration passed.
4--4
TDS 200 Series Digital Oscilloscope Service Manual
Performance Verification
Check DC Gain Accuracy
This test checks the DC gain accuracy of all input channels.
1. Set the DC voltage source output level to 0V.
2. Set up the instrument using the following steps:
Press menu button Select menu item Select setting
SAVE/ RECALL
CH 1 Probe 1X
ACQUIRE Average 16
MEASURE
Recall Factory
Source An unchecked channel
Type Mean
3. As shown below, connect the instrument channel selected in the table to the
DC voltage source.
DC voltage
source
Digitizing oscilloscope
+--
Dual banana to BNC adapter
BNC cable
4. For each VOLTS/DIV setting listed in the table at the top of page 4--6,
perform the following steps:
a. Set the DC voltage source output level to the positive voltage listed and
then record the mean measurement as V
pos
.
b. Reverse the polarity of the DC voltage source and then record the mean
measurement as V
c. Calculate V
diff=Vpos
neg
.
-- V
and then compare V
neg
to the accuracy
diff
limits in the table.
TDS 200 Series Digital Oscilloscope Service Manual
4--5
Performance Verification
Check Bandwidth
VOLTS/DIV setting DC voltage source output levels Accuracy limits for V
TDS210 & TDS220
5 mV/div +17.5 mV, --17.5 mV 34.0 mV to 36.0 mV
200 mV/div +700 mV, --700 mV 1.358 V to 1.442 V
2V/div +7.00 V, --7.00 V 13.58 V to 14.42 V
TDS224
5 mV/div +17.5 mV, --17.5 mV 33.6 mV to 36.4 mV
200 mV/div +700 mV, --700 mV 1.358 V to 1.442 V
2V/div +7.00 V, --7.00 V 13.58 V to 14.42 V
5. Set DC voltage source output level to 0V.
6. Disconnect the test setup.
7. Repeat steps 1 through 6 until all input channels have been checked.
This test checks the bandwidth of all input channels.
1. Set up the instrument using the following steps:
diff
Press menu button Select menu item Select setting
SAVE/ RECALL
CH 1 Probe 1X
ACQUIRE Average 16
TRIGGER Coupling Noise Reject
MEASURE
Recall Factory
Source An unchecked channel
Type Pk-Pk
4--6
TDS 200 Series Digital Oscilloscope Service Manual
Performance Verification
2. As shown below, connect the instrument channel selected in the table to the
leveled sine wave generator.
Leveled
sine wave
generator
Output
BNC cable
Digitizing oscilloscope
50 feedthrough terminator
3. Set the instrument VOLTS/DIV to 500 mV/div.
4. Set the instrument SEC/DIV to 10 s/div.
5. Set the leveled sine wave generator frequency to 50 kHz.
6. Set the leveled sine wave generator output level so the peak-to-peak measurement is between 2.98 V and 3.02 V.
7. Set the leveled sine wave generator frequency to 60 MHz if you are checking a TDS 210 or to 100 MHz if you are checking a TDS 220 or TDS 224.
8. Set the instrument SEC/DIV to 10 ns/div.
Check Sample Rate and
Delay Time Accuracy
9. Check that the peak-to-peak measurement is 2.12 V.
10. Disconnect the test setup.
11. Repeat steps 1 through 10 until all input channels have been checked.
This test checks the time base accuracy.
1. Set up the instrument using the following steps:
Press menu button Select menu item Select setting
SAVE/ RECALL
CH 1 Probe 1X
Recall Factory
TDS 200 Series Digital Oscilloscope Service Manual
4--7
Performance Verification
2. Connect the instrument to the time mark generator as shown below.
Time mark
generator
Output
BNC cable
Digitizing oscilloscope
50 feedthrough terminator
3. Set the time mark generator period to 10 ms.
4. Set the instrument VOLTS/DIV to 500 mV/div.
5. Set the instrument Main SEC/DIV to 1ms/div.
6. Press SET LEVEL TO 50%.
7. Use the vertical POSITION control to center the test signal on screen.
8. Change the instrument setup using the following steps:
Press menu button Select menu item Select setting
HORIZONTAL Window
4--8
9. Set the instrument Window SEC/DIV to 500 s/div.
NOTE. Setting the Window SEC/DIV to a setting slower than the Main SEC/DIV setting resets the main timebase. If this happens, reset the Main SEC/DIV and then the Window SEC/DIV settings.
10. Use the horizontal POSITION control to set the window position to
10.00 ms.
11. Set the instrument Window SEC/DIV to 500 ns/div.
12. Check that the rising edge of the marker crosses the center horizontal
graticule line within ±2 divisions of center graticule.
TDS 200 Series Digital Oscilloscope Service Manual
Performance Verification
NOTE. One division of displacement from graticule center corresponds to a 50 ppm time base error.
13. Disconnect the test setup.
Check Edge Trigger
Sensitivity
This test checks the edge trigger sensitivity for all input channels.
1. Set up the instrument using the following steps:
Press menu button Select menu item Select setting
SAVE/ RECALL
CH 1 Probe 1X
TRIGGER Mode Normal
ACQUIRE Sample
MEASURE
Recall Factory
Source An unchecked channel
Type Pk-Pk
2. As shown below, connect the instrument channel selected in the table to the leveled sine wave generator.
Leveled
sine wave
generator
Digitizing oscilloscope
Output
3. Set the leveled sine wave generator frequency to 60 MHz if you are checking a TDS 210 or to 100 MHz if you are checking a TDS 220 or TDS 224.
4. Set the instrument VOLTS/DIV to 500 mV/div.
5. Set the instrument SEC/DIV to 10 ns/div.
6. Set the leveled sine wave generator output level to approximately 750 mV
so that the measured amplitude is approximately 750 mV. (The measured amplitude can fluctuate around 750 mV.)
TDS 200 Series Digital Oscilloscope Service Manual
BNC cable
50 feedthrough terminator
p-p
4--9
Performance Verification
7. Press SET LEVEL TO 50%. Adjust TRIGGER LEVEL as necessary and
then check that triggering is stable.
8. Change the instrument setup using the following steps:
Press menu button Select menu item Select setting
TRIGGER Slope Falling
9. Press SET LEVEL TO 50%. Adjust TRIGGER LEVEL as necessary and
then check that triggering is stable.
10. Change the instrument setup using the following steps:
Press menu button Select menu item Select setting
TRIGGER Slope Rising
Check External Edge
Trigger Sensitivity
11. Disconnect the test setup.
12. Repeat steps 1 through 11 until all input channels have been checked.
TDS 210 and TDS 220 only: this test checks the edge trigger sensitivity for the external trigger.
1. Set up the instrument using the following steps:
Press menu button Select menu item Select setting
SAVE/ RECALL
CH 1 Probe 1X
TRIGGER
ACQUIRE Sample
MEASURE
Recall Factory
Source Ext
Mode Normal
Source CH1
Type Pk-Pk
4--10
TDS 200 Series Digital Oscilloscope Service Manual
Performance Verification
2. Connect the instrument to the leveled sine wave generator as shown below.
Leveled sine wave generator
Output
Digitizing oscilloscope
50 feedthrough terminator
BNC cable
BNC T connector
BNC
cable
3. Set the leveled sine wave generator frequency to 60 MHz if you are checking a TDS 210 or to 100 MHz if you are checking a TDS 220.
4. Set the instrument VOLTS/DIV to 500 mV/div.
5. Set the instrument SEC/DIV to 10 ns/div.
6. Set the leveled sine wave generator output level to approximately 750 mV
p-p
so that the measured amplitude is approximately 750 mV. (The measured amplitude can fluctuate around 750 mV.)
7. Press SET LEVEL TO 50%. Adjust TRIGGER LEVEL as necessary and then check that triggering is stable.
8. Change the instrument setup using the following steps:
Press menu button Select menu item Select setting
TRIGGER Slope Falling
9. Press SET LEVEL TO 50%. Adjust TRIGGER LEVEL as necessary and then check that triggering is stable.
10. Change the instrument setup using the following steps:
Press menu button Select menu item Select setting
TRIGGER Slope Rising
11. Disconnect the test setup.
TDS 200 Series Digital Oscilloscope Service Manual
4--11
Performance Verification
4--12
TDS 200 Series Digital Oscilloscope Service Manual

Adjustment Procedures

This chapter contains adjustment procedures for the TDS 200-Series Digital Real-Time Oscilloscope.
Only qualified personnel should perform service procedures. Read the Service Safety Summary and the General Safety Summary before performing any service procedures. Also refer to the chapter Operating Information for information about using the TDS 200s.
NOTE. The voltage references inside the TDS 200s are very stable over time and should not require routine updates. Before performing any procedure in this chapter, first verify that the instrument does not meet specifications. Refer to the chapter Performance Verification to verify the specifications.

Required Equipment

The following equipment, or a suitable equivalent, is required to complete these procedures.
Table 5--1: Required equipment
Description Minimum requirements Examples
DC Voltage Source -- 2 0 V t o 2 0 V, ±0.1% accuracy Wavetek 9100 Universal Calibration
System with Oscilloscope Calibration Module (Option 250)
Leveled Sinewave Generator
50 Ω BNC Cable
50 BNC Cable (two)
50 Feedthrough Termination
Dual Banana to BNC Adapter
BNC T One male and two female
50 kHz and 100 MHz, ±0.6% amplitude accuracy
BNCmaletoBNCmale, 36 in (1 m) long
BNCmaletoBNCmale, 10 in (25 cm) long
BNC male and female connectors
Banana plugs to BNC female Tektronix part number 103-0090-00
BNC connectors
Fluke 5500A Multi-product Calibrator with Oscilloscope Calibration Option (Option 5500A-SC)
Tektronix part number 012-0482-00
Tektronix part number 012-0208-00
Tektronix part number 011-0049-01
Tektronix part number 103-0030-00
TDS 200 Series Digital Oscilloscope Service Manual
5--1
Adjustment Procedures

Adjustment Procedure

The adjustment procedure consists of 55 steps. Each step requires an external voltage source for new calibration constants.
Before performing the adjustment procedure, you must warm up the instrument for at least ten minutes in an ambient temperature between 20° C and 30° C. Adjustments performed prior to warm-up or outside this temperature range may result in poor performance.
If all steps in the procedure are completed successfully, a Passmessage is displayed and the new calibration constants take affect. If any step fails, the procedure is aborted and the current calibration is not affected. You can choose to abort the procedure at any step by selecting FCAL ABORT without affecting the current calibration.
The equipment setups, shown in Figure 5-2, are required to complete the adjustment procedure. Tables 5--2 and 5--3 list the steps in the procedure and the signal requirement for each step.
Enable the Service Menu
You must enable the service menu to perform the adjustment procedure. With the instrument on, insert a narrow object such as a plastic alignment tool into the access hole, shown in Figure 5-1, until you feel the click of the momentary button. Hold this button in and press the front panel UTILITY button.
Access to
service menu
enable button
Figure 5-1: Service menu enable button
5--2
A Service menu selection replaces the Language menu selection, giving you access to the service routines.
After adjustment is complete, disable the service menu by again holding in the internal momentary button and pressing the front panel UTILITY button.
TDS 200 Series Digital Oscilloscope Service Manual
Adjustment Procedures
DC voltage setup 1
DC voltage
source
+--
DC voltage setup 2
DC voltage
source
+--
Dual banana to BNC adapter
BNC cable
BNC cable
Two Channel
Digitizing oscilloscope
10 inch
BNC cables
Digitizing oscilloscope
BNC ‘T’ connector
Four Channel
Digitizing oscilloscope
10 inch
BNC cables
10 inch
BNC cable
BNC ‘T’ connector
10 inch BNC cable
10 inch
BNC cables
BNC ‘T’
connector
s
Sinewave generator setup
Leveled
sine wave
generator
Output
(Table 5--2 and 5--3 list
BNC cable
Dual banana to BNC adapter
Digitizing oscilloscope
50 feedthrough
terminator
proper channel)
Figure 5-2: Adjustment setups
BNC cable
Edge setup
Square wave
generator
Four Channel
Digitizing oscilloscope
Output
50 feedthrough
terminator
(Table 5--2 and 5--3 list
proper channel)
BNC cable
TDS 200 Series Digital Oscilloscope Service Manual
5--3
Adjustment Procedures
Adjustment Procedure
Use this procedure to load new calibration constants.
1. Turn on the instrument and let it warm up for at least ten minutes.
2. Enable the Service menu as described on Page 5--2.
3. Press the UTILITY button and select Service from the menus.
4. Select FCAL INIT to initiate the factory calibration routine. You are now in
step 1 of the procedure.
5. As indicated by Step 1 in Table 5--2 on page 5--5 (TDS 210 and TDS 220)
or Table 5--3 on page 5--7 (TDS 224), connect a 0.0 VDC source to the channel inputs.
6. Press FCAL STEP to load the calibration constant for step 1.
7. Follow the screen prompts along with Table 5--2 on page 5--5 (TDS 210 and
TDS 220) or Table 5--3 on page 5--7 (TDS 224) for the remainder of the routine, applying the proper signal for each step.
NOTE. During some steps, the instrument may appear to be idle for several minutes while it is processing information internally. Please have patience.
If any step in the procedure fails, the procedure terminates. Failure will result if an incorrect signal source is connected. If this happens, you must start the procedure over from the beginning.
If the procedure completes successfully, a Passmessage is displayed and the new adjustment takes effect.
8. Compensate the signal path by performing the self calibration routine.
Disconnect any probes or cables from the channel input connectors. Then, press the UTILITY button and select Do Self Cal to confirm that you are ready to proceed.
9. Disable the Service menu option.
5--4
TDS 200 Series Digital Oscilloscope Service Manual
Adjustment Procedures
Table 5--2:
Step: Firmware Below V 2.00
1 1 DC Voltage 1 0.0 V
2 2 DC Voltage 1 --20.0 V
3 3 DC Voltage 1 -- 2 . 0 V
4 4 DC Voltage 1 -- 1 . 6 V
5 5 DC Voltage 1 -- 1 . 2 V
6 6 DC Voltage 1 -- 0 . 8 V
7 7 DC Voltage 1 -- 0 . 4 V
8 8 DC Voltage 1 --0.32 V
9 9 DC Voltage 1 -- 0 . 2 V
10 10 DC Voltage 1 --0.16 V
11 11 DC Voltage 1 --0.08 V
12 12 DC Voltage 1 --0.04 V
13 13 DC Voltage 1 --0.03 V
14 14 DC Voltage 1 --0.02 V
15 15 DC Voltage 1 --0.015 V
16 16 DC Voltage 1 0.015 V
17 17 DC Voltage 1 0.02 V
18 18 DC Voltage 1 0.03 V
19 19 DC Voltage 1 0.04 V
20 20 DC Voltage 1 0.08 V
21 21 DC Voltage 1 0.16 V
22 22 DC Voltage 1 0.2 V
23 23 DC Voltage 1 0.32 V
24 24 DC Voltage 1 0.4 V
25 25 DC Voltage 1 0.8 V
26 26 DC Voltage 1 1.2 V
27 27 DC Voltage 1 1.6 V
28 28 DC Voltage 1 2.0 V
29 29 DC Voltage 1 20.0 V
30 30 DC Voltage 1 0.0 V
31 31 DC Voltage 2 -- 5 . 0 V
32 32 DC Voltage 2 -- 1 . 0 V
33 33 DC Voltage 2 1.0 V
34 34 DC Voltage 2 5.0 V
TDS 210 and TDS 220 Adjustment steps
Step Firmware V2.00andUp
Equipment setup Signal source setting Input
CH1 & CH2
EXT TRIG
TDS 200 Series Digital Oscilloscope Service Manual
5--5
Adjustment Procedures
,TD
S
M
H
V
,TD
S
M
H
V
,TD
S
M
H
V
,
,TD
S
M
H
V
Table 5--2:
Step: Firmware Below V 2.00
35 35 Sinewave Generator, 50 Ohm 5 kHz, 1 V
36 36 Sinewave Generator, 50 Ohm 1MHz,1V
37 37 Sinewave Generator, 50 Ohm 50 kHz, 1 V
38 38 Sinewave Generator, 50 Ohm 20 MHz, 1 V
39 39 Sinewave Generator, 50 Ohm
TDS 210 and TDS 220 Adjustment steps (cont.)
Step Firmware V2.00andUp
p-p
p-p
p-p
p-p
TDS 210: 60 MHz, 1 V
220: 100
40 40 Sinewave Generator, 50 Ohm 5 kHz, 2.5 V
41 41 Sinewave Generator, 50 Ohm 1MHz,2.5V
42 42 Sinewave Generator, 50 Ohm 50 kHz, 2.5 V
43 43 Sinewave Generator, 50 Ohm 20 MHz, 2.5 V
44 44 Sinewave Generator, 50 Ohm
TDS 210: 60 MHz, 2.5 V
p-p
p-p
p-p
p-p
220: 100
45 45 Sinewave Generator, 50 Ohm
TDS 210: 60 MHz, 2.5 V
220: 100
None 46 Edge, 50 Ohm 1 kHz, 0 to --800 mV
46 47 Sinewave Generator, 50 Ohm 5 kHz, 1 V
47 48 Sinewave Generator, 50 Ohm 1MHz,1V
48 49 Sinewave Generator, 50 Ohm 50 kHz, 1 V
49 50 Sinewave Generator, 50 Ohm 20 MHz, 1 V
50 51 Sinewave Generator, 50 Ohm
TDS 210: 60 MHz, 1 V
p-p
p-p
p-p
p-p
TDS 220: 100 MHz, 1 V
51 52 Sinewave Generator, 50 Ohm
5 kHz, 2.5 V
52 53 Sinewave Generator, 50 Ohm 1MHz,2.5V
53 54 Sinewave Generator, 50 Ohm 50 kHz, 2.5 V
54 55 Sinewave Generator, 50 Ohm 20 MHz, 2.5 V
55 56 Sinewave Generator, 50 Ohm
TDS 210: 60 MHz, 2.5 V
p-p
p-p
p-p
p-p
220: 100
z, 1
z, 2.5
z, 2.5
z, 2.5
InputSignal source settingEquipment setup
CH1
p-p
p-p
p-p
p-p
p-p
p-p
CH2
p-p
p-p
p-p
p-p
5--6
TDS 200 Series Digital Oscilloscope Service Manual
Adjustment Procedures
,
,
Table 5--3:
Step Equipment setup Signal source setting Input
1 DC Voltage 1 0.0 V
2 DC Voltage 1 --20.0 V
3 DC Voltage 1 -- 2 . 0 V
4 DC Voltage 1 -- 1 . 6 V
5 DC Voltage 1 -- 1 . 2 V
6 DC Voltage 1 -- 0 . 8 V
7 DC Voltage 1 -- 0 . 4 V
8 DC Voltage 1 --0.32 V
9 DC Voltage 1 -- 0 . 2 V
10 DC Voltage 1 --0.16 V
11 DC Voltage 1 --0.08 V
12 DC Voltage 1 --0.04 V
13 DC Voltage 1 --0.03 V
14 DC Voltage 1 --0.02 V
15 DC Voltage 1 --0.015 V
16 DC Voltage 1 0.015 V
17 DC Voltage 1 0.02 V
18 DC Voltage 1 0.03 V
19 DC Voltage 1 0.04 V
20 DC Voltage 1 0.08 V
21 DC Voltage 1 0.16 V
22 DC Voltage 1 0.2 V
23 DC Voltage 1 0.32 V
24 DC Voltage 1 0.4 V
25 DC Voltage 1 0.8 V
26 DC Voltage 1 1.2 V
27 DC Voltage 1 1.6 V
28 DC Voltage 1 2.0 V
29 DC Voltage 1 20.0 V
30 DC Voltage 1 0.0 V
TDS 224 Adjustment steps
CH1, CH2, CH3 & CH4
TDS 200 Series Digital Oscilloscope Service Manual
5--7
Adjustment Procedures
Table 5--3:
TDS 224 Adjustment steps (cont.)
Step Equipment setup Signal source setting Input
31 Sinewave Generator, 50 Ohm 5 kHz, 1 V
32 Sinewave Generator, 50 Ohm 1MHz,1V
33 Sinewave Generator, 50 Ohm 50 kHz, 1 V
34 Sinewave Generator, 50 Ohm 20 MHz, 1 V
35 Sinewave Generator, 50 Ohm 100 MHz, 1 V
36 Sinewave Generator, 50 Ohm 5 kHz, 2.5 V
37 Sinewave Generator, 50 Ohm 1MHz,2.5V
38 Sinewave Generator, 50 Ohm 50 kHz, 2.5 V
39 Sinewave Generator, 50 Ohm 20 MHz, 2.5 V
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
40 Sinewave Generator, 50 Ohm 100 MHz, 2.5 V
41 Sinewave Generator, 50 Ohm 77.133 MHz, 1 V
p-p
p-p
CH1
42 Edge, 50 Ohm 1 kHz, 0 to --800 mV
43 Sinewave Generator, 50 Ohm 5 kHz, 1 V
44 Sinewave Generator, 50 Ohm 1MHz,1V
45 Sinewave Generator, 50 Ohm 50 kHz, 1 V
46 Sinewave Generator, 50 Ohm 20 MHz, 1 V
47 Sinewave Generator, 50 Ohm 100 MHz, 1 V
48 Sinewave Generator, 50 Ohm 5 kHz, 2.5 V
49 Sinewave Generator, 50 Ohm 1MHz,2.5V
50 Sinewave Generator, 50 Ohm 50 kHz, 2.5 V
51 Sinewave Generator, 50 Ohm 20 MHz, 2.5 V
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
52 Sinewave Generator, 50 Ohm 100 MHz, 2.5 V
53 Sinewave Generator, 50 Ohm 5 kHz, 1 V
54 Sinewave Generator, 50 Ohm 1MHz,1V
55 Sinewave Generator, 50 Ohm 50 kHz, 1 V
56 Sinewave Generator, 50 Ohm 20 MHz, 1 V
57 Sinewave Generator, 50 Ohm 100 MHz, 1 V
58 Sinewave Generator, 50 Ohm 5 kHz, 2.5 V
59 Sinewave Generator, 50 Ohm 1MHz,2.5V
60 Sinewave Generator, 50 Ohm 50 kHz, 2.5 V
61 Sinewave Generator, 50 Ohm 20 MHz, 2.5 V
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
62 Sinewave Generator, 50 Ohm 100 MHz, 2.5 V
63 Sinewave Generator, 50 Ohm 77.133 MHz, 1 V
p-p
p-p
p-p
CH2
CH3
64 Edge, 50 Ohm 1 kHz, 0 to --800 mV
5--8
TDS 200 Series Digital Oscilloscope Service Manual
Adjustment Procedures
Table 5--3:
TDS 224 Adjustment steps (cont.)
Step Equipment setup Signal source setting Input
65 Sinewave Generator, 50 Ohm 5 kHz, 1 V
66 Sinewave Generator, 50 Ohm 1MHz,1V
67 Sinewave Generator, 50 Ohm 50 kHz, 1 V
68 Sinewave Generator, 50 Ohm 20 MHz, 1 V
p-p
p-p
p-p
p-p
69 Sinewave Generator, 50 Ohm 100 MHz, 1 V
70 Sinewave Generator, 50 Ohm 5 kHz, 2.5 V
71 Sinewave Generator, 50 Ohm 1MHz,2.5V
p-p
p-p
72 Sinewave Generator, 50 Ohm 50 kHz, 2.5 V
73 Sinewave Generator, 50 Ohm 20 MHz, 2.5 V
74 Sinewave Generator, 50 Ohm 100 MHz, 2.5 V
p-p
p-p
p-p
p-p
CH4
TDS 200 Series Digital Oscilloscope Service Manual
5--9
Adjustment Procedures
5--10
TDS 200 Series Digital Oscilloscope Service Manual

Maintenance

This chapter contains the information needed to do periodic and corrective maintenance on the oscilloscope. The following sections are included:
H Preparation refers you to important information contained elsewhere in this
manual.
H Preventing ESD provides general information on preventing damage to
internal modules when doing maintenance.
H Inspection and Cleaning provides procedures for inspecting the instrument
and cleaning its external and internal modules.
H Removal and Installation Procedures provide procedures for the removal and
installation of modules.
H Troubleshooting provides information for isolating failed modules.
H Repackaging Instructions provides packaging information for shipment or
storage.

Preparation

Preventing ESD

Only qualified personnel should perform service procedures. Read the Service Safety Summary and the General Safety Summary before performing any service
procedures. Also refer to the chapter Operating Information for information about using the oscilloscope.
When performing any service which requires internal access to the instrument, adhere to the following precautions to avoid damaging internal modules and their components due to electrostatic discharge (ESD).
CAUTION. Static discharge can damage any semiconductor component in this
instrument.
1. Minimize handling of static-sensitive modules.
2. Transport and store static-sensitive modules in their static protected
containers or on a metal rail. Label any package that contains static-sensitive modules.
TDS 200 Series Digital Oscilloscope Service Manual
6--1
Maintenance

Inspection and Cleaning

3. Discharge the static voltage from your body by wearing a grounded antistatic
wrist strap while handling these modules. Do service of static-sensitive modules only at a static-free work station.
4. Nothing capable of generating or holding a static charge should be allowed
on the work station surface.
5. Handle circuit boards by the edges when possible.
6. Do not slide the modules over any surface.
7. Avoid handling modules in areas that have a floor or work-surface covering
capable of generating a static charge.
This section describes how to inspect for dirt and damage, and how to clean the instrument. Inspection and cleaning are done as preventive maintenance. Preventive maintenance, when done regularly, may prevent malfunctions and enhance reliability.
General Care
Preventive maintenance consists of visually inspecting and cleaning the instrument, and using general care when operating it. How often to do mainte­nance depends on the severity of the environment in which the instrument is used. A proper time to perform preventive maintenance is just before adjustment of the instrument.
Protect the instrument from adverse weather conditions. The instrument is not waterproof.
Do not store or leave the instrument where the liquid crystal display (LCD) will be exposed to direct sunlight for long periods of time.
The front and back cases help keep dust out of the instrument and must be in place during normal operation.
CAUTION. To avoid damage to the instrument or probes, do not expose it to any sprays, liquids, or solvents.
6--2
TDS 200 Series Digital Oscilloscope Service Manual
Maintenance
Inspection and Cleaning
Procedures
Inspect and clean the instrument and probes as often as operating conditions require. The collection of dirt on internal components can cause them to overheat and breakdown. Dirt also provides an electrical conduction path that could cause an instrument failure, especially under high-humidity conditions.
Inspection — Exterior. Inspect the outside of the instrument for damage, wear, and missing parts. Instruments that appear to have been dropped or otherwise abused should be checked thoroughly to verify correct operation and performance. Immediately repair defects that could cause personal injury or lead to further damage to the instrument.
Cleaning Procedure — Exterior. To clean the instrument exterior and probes, perform the following steps:
1. Remove loose dust on the outside of the instrument and probes with a lint free cloth. Use care to avoid scratching the clear plastic display shield.
2. Use a soft cloth dampened with water to clean the instrument. You can use a 75% isopropyl alcohol solution for more efficient cleaning.
CAUTION. To avoid damage to the surface of the instrument or probes, do not use any abrasive or chemical cleaning agents.
Inspection — Interior. During normal use, the instrument case guards the internal components against common dust and dirt. However, if it appears that the instrument may have been damaged or has been exposed to an unusually dirty environment, you should inspect and clean the interior.
To inspect and clean the interior, disassemble the instrument by completing each of the disassembly procedures in the order given starting on page 6--12.
Inspect the internal portions of the instrument for damage and wear using Table 6--1 as a guide. Defects found should be repaired immediately.
Table 6--1: Internal inspection check list
Item Inspect for Repair action
Front and Back Case
Gasket and Foam Pads
Cracks or deformations. Scratched lettering or display filter. Loose connectors or labels.
Misplaced or missing pieces. Relocate or replace defective pieces.
Repair or replace defective hardware.
TDS 200 Series Digital Oscilloscope Service Manual
6--3
Maintenance
Table 6--1: Internal inspection check list (cont.)
Item Repair actionInspect for
Circuit boards Loose, broken, or corroded
solder connections. Burned circuit boards. Burned, broken, or cracked circuit-run plating.
Solder connections Cold solder or rosin joints. Resolder joint and clean with isopropyl
Remove failed module and replace with a new module.
alcohol.
Wiring and cables Loose plugs or connectors.
Burned, broken, or frayed wiring.
Firmly seat connectors. Repair or replace modules with defective wires or cables.
Cleaning Procedure Interior. To clean the instrument interior, perform the following steps:
1. Blow off dust with dry, low-pressure, deionized air (approximately 9 psi).
2. Remove any remaining dust with a lint-free cloth dampened in isopropyl
alcohol (75% solution) and rinse with warm deionized water. (A cotton­tipped applicator is useful for cleaning in narrow spaces and on circuit boards.)
STOP. If, after performing steps 1 and 2, a module is clean upon inspection, skip steps 3 and 4. Proceed with step 5.
If steps 1 and 2 do not remove all the dust or dirt, all modules in the instrument except the Display Module may be spray washed using a solution of 75% isopropyl alcohol. Proceed with step 3.
3. Spray wash dirty parts with the 75% isopropyl alcohol solution and wait
60 seconds for the majority of the alcohol to evaporate.
6--4
4. Use hot (49_ Cto60_ C / 120_ F to 140_ F) deionized water to thoroughly
rinse the parts.
5. Dry all parts with low-pressure, deionized air.
6. Dry all components and assemblies in an oven or drying compartment using
low-temperature (52_ Cto66_ C / 125_ F to 150_ F) circulating air.
Lubrication. There is no periodic lubrication required for the oscilloscope.
TDS 200 Series Digital Oscilloscope Service Manual

Removal and Installation Procedures

This section describes how to remove and install the major mechanical and electrical modules. It provides the following information:
H Tools Required on page 6--6 describes the tools needed to perform the
procedures.
H List of Modules on page 6--6 describes where to locate a list of replaceable
modules.
H Summary of Procedures on page 6--7 lists the procedures for removal and
installation of modules.
H Rear Feet, on page 6--8, begins a series of procedures that describe the
removal and installation of modules.
Maintenance
Preparation
Please read the following warning statement. Then read the following general instructions before removing a module.
WARNING. Only qualified personnel should perform service procedures. Before doing this or any other procedure in this manual, read the General Safety Summary and Service Safety Summary found at the beginning of this manual.
Always disconnect the power cord before disassembling any portion of the instrument.
Also, to prevent damage to electrical components, read Preventing ESD on page 6--1.
1. Use Figure 10-1 (the exploded view of the instrument) on page 10--8 to give you an overview of how the entire instrument is assembled.
2. Read the Summary of Procedures on page 6--7 to understand which procedure to use to access a given module. Then read Tools Required on page 6--6 for a list of tools needed to remove and install modules in the oscilloscope.
3. If you are disassembling the oscilloscope for cleaning, refer to the Inspection and Cleaning procedure on page 6--3 for cleaning instructions.
TDS 200 Series Digital Oscilloscope Service Manual
6--5
Maintenance
Tools Required
List of Modules
You will need the following tools to remove and replace all the modules in the oscilloscope.
Tools
Torque-limiting screwdriver, 12 in lb (1.3 N
Screwdriver with !@4 inch flat blade
Pliers
Sharp knife (X-ACTOknife or equivalent)
m) range with TorxT-15 tip
The Mechanical Parts List chapter provides a list of all replaceable modules. Any electrical or mechanical module, assembly, or part listed in the parts list is referred to as a module.
6--6
TDS 200 Series Digital Oscilloscope Service Manual
Maintenance
Summary of Procedures
Table 6--2 lists all procedures in order of increasing complexity. Complete disassembly is best achieved by following the procedures in the order given.
Table 6--2: List of procedures
Procedure Begins on Page
Rear Feet 6--8
Front-Case Label 6--10
Rear-Case Label 6--11
Handle 6--12
Power Button 6--13
Rear Case 6--14
Front Feet 6--16
EMI Clips 6--17
Flip Stand 6--18
PowerLineFuse 6--20
Power Supply Module 6--21
Front Panel Knobs 6--23
Internal Assembly 6--24
Main Board Module 6--26
Display Module 6--28
Front Panel Module 6--30
Keypad 6--33
Display Shield 6--35
Front Case 6--37
TDS 200 Series Digital Oscilloscope Service Manual
6--7
Maintenance
Rear Feet
No tools are required to remove the feet.
Removal. Use this procedure to remove the rear feet.
1. Place the instrument face down on a soft surface (such as an anti-static mat).
2. Slide the feet up and out of the rear case as shown in Figure 6-1.
NOTE. You may want to use a flat blade screwdriver to pry up on the foot. Use care not to mar the surface of the rear case.
Rear foot (2)
6--8
Figure 6-1: Removing the rear feet
TDS 200 Series Digital Oscilloscope Service Manual
Maintenance
Installation. Use this procedure to install the feet.
1. Place the rear case face down on a soft surface (such as an anti-static mat).
2. Align the slots of the rear foot with the guides in the rear case as shown in
Figure 6-2 and press the foot in.
Rear foot (2)
Figure 6-2: Installing the rear feet
TDS 200 Series Digital Oscilloscope Service Manual
6--9
Maintenance
Front-Case Label
You will need a sharp knife and needle-nose pliers to remove the front-case label.
CAUTION. Removal of a front-case label will destroy it. Do not remove it unless it needs to be replaced.
Removal. Use the sharp knife to pry up a corner of the label and use the pliers to peel the label off.
Installation. Use this procedure to install a new front-panel label in either a new or the existing front case.
1. Peel the paper off the back side of the new label.
2. Handling the label by its edges only, align it with the recess in the front case
as shown in Figure 6-3.
3. When properly aligned, press the new label into place.
6--10
Figure 6-3: Installing a front-case label
TDS 200 Series Digital Oscilloscope Service Manual
Maintenance
Rear-Case Label
You will need a sharp knife and needle-nose pliers to remove the rear-case label.
CAUTION. Removal of a rear-case label will destroy it. Do not remove it unless it needs to be replaced.
Removal. Use the sharp knife to pry up a corner of the label and use the pliers to peel the label off.
Installation. Use this procedure to install a new rear-case label in either a new or the existing rear case.
1. Peel the paper off the back side of the new label.
2. Handling the label by its edges only, align it with the recess in the rear case
as shown in Figure 6-4.
3. When properly aligned, press the new label into place.
Rear label
Figure 6-4: Installing a new rear-case label
TDS 200 Series Digital Oscilloscope Service Manual
6--11
Maintenance
Handle
You will need a flat-blade screw driver to remove the handle.
Removal. Use this procedure to remove the handle.
1. Place the instrument on its feet.
2. With the handle laying down, pry one end of the handle out of the case as
shown in Figure 6-5.
NOTE. Use care not to mar the surface with the screwdriver.
3. Slide the handle towards the back of the rear case.
6--12
Figure 6-5: Removing the handle
TDS 200 Series Digital Oscilloscope Service Manual
Maintenance
Installation. Use this procedure to install the handle.
1. Place the instrument on its feet.
2. From behind the rear case with the handle laying down, slide the handle
along the groves in the case until the handle snaps into place as shown in Figure 6-6.
Figure 6-6: Installing the handle
Power Button
TDS 200 Series Digital Oscilloscope Service Manual
Remove the power button by firmly grasping the button (with pliers if necessary) and pull it straight out of the cabinet. Reinstall the button by pressing it onto the power button shaft until it snaps into place.
NOTE. Wrap the plier jaws with tape (such as electrical) to avoid marring the button surface.
6--13
Maintenance
Rear Case
You will need a torque-limiting Torx T-15 screwdriver and a flat-blade screwdriver for this procedure.
Removal. First remove the handle and power button using the procedures on pages 6--12 and 6--13, then use this procedure to remove the rear case.
1. If installed, remove the extension module from the rear of the instrument.
CAUTION. Do not remove the rear case with an extension module installed.
2. Place the instrument face down on a soft surface (such as an anti-static mat).
3. Remove the two screws located in the handle recesses. Refer to Figure 6-7.
4. Slip the flat-blade screwdriver between the rear case and front case at the
bottom of the instrument to disengage the tabs and lift the rear case off.
Installation. Use this procedure to install the rear case.
1. Place the instrument face down on a soft surface (such as an anti-static mat).
2. Slide the rear case over the assembly until the bottom tabs of the front case
lock into the rear case.
CAUTION. Be careful to align the guide ribs and to not pinch any wires or other parts between the front and rear cases. If used, make sure that the copper mesh grounding tube is in the proper place as shown in Figure 6-7.
3. Install the two screws located in the handle recesses.
CAUTION. Avoid cross threading the screws that secure the rear case to the front case. An easy way to avoid this is to slip the screw into the hole and, using slight pressure, slowly rotate the screw counterclockwise until you feel the screw slip into the screw threads. Then tighten the screw clockwise. A properly started screw will have very little resistance going in.
4. Install the handle and the power button.
6--14
TDS 200 Series Digital Oscilloscope Service Manual
Retaining screw
Maintenance
Retaining screw
Rear case
Front case
Figure 6-7: Removing and installing the rear cover
Retaining tab (7)
Copper mesh tubing
TDS 200 Series Digital Oscilloscope Service Manual
6--15
Maintenance
Front Feet
You will need a torque-limiting Torx T-15 screwdriver and a flat-blade screwdriver for this procedure.
Removal. First remove the handle, power button, and rear case using the procedures on pages 6--12, 6--13, and 6--14. Then, slide the front feet out of the front case as shown in Figure 6-8.
Installation. Slide the feet into the slots provided in the front case as shown in Figure 6-8. Use the installation procedures for each module removed to reassemble the instrument.
Front foot (2)
Figure 6-8: Removing and installing the front feet
Front case
6--16
TDS 200 Series Digital Oscilloscope Service Manual
Maintenance
EMI Clips
You will need a torque-limiting Torx T-15 screwdriver and a flat-blade screwdriver for this procedure.
Removal. First remove the handle, power button, and rear case using the procedures on pages 6--12, 6--13, and 6--14. Then, slide the four EMI clips out of the rear case as shown in Figure 6-9.
NOTE. The EMI clips can be pushed out from the rear of the case using a narrow flat-blade screwdriver.
Installation. Slide the four EMI clips into the slots provided in the rear case as shown in Figure 6-9. Use the installation procedures for each module removed to reassemble the instrument.
Figure 6-9: Removing and installing the EMI clips
TDS 200 Series Digital Oscilloscope Service Manual
6--17
Maintenance
Flip Stand
You will need a torque-limiting Torx T-15 screwdriver and a flat-blade screwdriver for this procedure.
Removal. First remove the handle, power button, and rear case using the procedures on pages 6--12, 6--13, and 6--14. Then, use the following procedure to remove the flip stand.
1. Place the rear case face down on a soft surface (such as an anti-static mat).
2. With the stand folded against the rear case (in the up position), push the stand
out of the rear case channels as shown in Figure 6-10.
Rear case
6--18
Flip stand
Figure 6-10: Removing the flip stand
TDS 200 Series Digital Oscilloscope Service Manual
Maintenance
Installation. Use this procedure to install the flip stand.
1. Orient the flip stand to the rear case, aligning the flip stand slots to the rear case tabs as shown in Figure 6-11.
2. With the stand in the folded up position, snap the stand into place.
3. Use the installation procedures for each module removed to reassemble the
instrument.
Rear case
Figure 6-11: Installing the flip stand
TDS 200 Series Digital Oscilloscope Service Manual
Flip stand
6--19
Maintenance
Power Line Fuse
You will need a torque-limiting Torx T-15 screwdriver and a flat-blade screwdriver for this procedure.
First remove the handle, power button, and rear case using the procedures on pages 6--12, 6--13, and 6--14 to access the line fuse.
The line fuse is located on the power supply module as shown in Figure 6-12. If the fuse tests bad, replace it with a fuse of the same type and rating.
Use the installation procedures for each module removed to reassemble the instrument.
Line fuse
Power supply module
Figure 6-12: Line fuse location
Front case
6--20
TDS 200 Series Digital Oscilloscope Service Manual
Maintenance
Power Supply Module
You will need a torque-limiting Torx T-15 screwdriver and a flat-blade screwdriver for this procedure.
Removal. First remove the handle, power button, and rear case using the procedures on pages 6--12, 6--13, and 6--14. Then, use the following procedure to remove the power supply module. Refer to Figure 6-13.
1. Disconnect the following wires.
a. The line cord ground wire on the main board at J602 or J101.
b. The seven-conductor ribbon cable on the main board at J131.
c. The two-conductor backlight cable on the power supply module.
2. Slightly bend back the two securing tabs at the bottom of the power supply
module and lift the module out of the inner chassis.
Two-conductor backlight cable
Seven-conductor
ribbon cable
Power supply module
Line cord
ground wire
Figure 6-13: Removing the power supply module
Front case
TDS 200 Series Digital Oscilloscope Service Manual
6--21
Maintenance
Installation. Use this procedure to install the power supply module. Refer to Figure 6-14.
1. Place the power supply module into the inner chassis, placing the board
under the three tabs at the top of the inner chassis.
2. Snap the bottom of the module into place.
3. Reconnect the following wires.
a. The line cord ground wire on the main board at J602 or J101.
b. The seven-conductor ribbon cable on the main board at J131.
c. The two-conductor backlight cable on the power supply module.
4. Use the installation procedures for each module removed to reassemble the
instrument.
Two-conductor backlight cable
Seven-conductor
ribbon cable
Power supply module
Line cord
ground wire
Figure 6-14: Installing the power supply module
Front case
6--22
TDS 200 Series Digital Oscilloscope Service Manual
Maintenance
Front-Panel Knobs
Remove the front panel knobs by firmly grasping the knob (with pliers if necessary) and pull it away from the front panel. Install front panel knobs by aligning the keyed knob with the shaft and pressing the knob onto the shaft.
NOTE. Wrap the plier jaws with tape (such as electrical) to avoid marring the knob surface.
TDS 200 Series Digital Oscilloscope Service Manual
6--23
Maintenance
Internal Assembly
You will need a torque-limiting Torx T-15 screwdriver and a flat-blade screwdriver for this procedure.
Removal. First remove the handle, power button, and rear case using the procedures on pages 6--12, 6--13, and 6--14. Then, use the following procedure to remove the internal assembly.
1. Remove all the front panel knobs.
2. Lift the entire internal assembly (housing all of the boards and the display
screen) out of the front case as shown in Figure 6-15.
NOTE. The switch keypad will most likely remain inside the front case. It does not need to be removed with the internal assembly.
Internal assembly
6--24
Keypad
Conductive foam pad
Front case
Figure 6-15: Removing and installing the internal assembly
TDS 200 Series Digital Oscilloscope Service Manual
Maintenance
Installation. Use this procedure to install the internal assembly.
1. Place the internal assembly into the front case as shown in Figure 6-15.
2. If used, place the copper mesh grounding tube between the main board
module shield and the front case as shown in Figure 6-16. This part is not used on later instruments; check the Mechanical Parts List for your serial number (see page 10--8).
3. Use the installation procedures for each module removed to reassemble the instrument.
Copper mesh tubing
Figure 6-16: Installing the copper mesh grounding tube
Front case
TDS 200 Series Digital Oscilloscope Service Manual
6--25
Maintenance
Main Board Module
You will need a torque-limiting Torx T-15 screwdriver and a flat-blade screwdriver for this procedure.
Removal. First remove the handle, power button, and rear case using the procedures on pages 6--12, 6--13, and 6--14. Then, use the following procedure to remove the main board module.
1. Remove the entire internal assembly from the front case using the procedure
on page 6--24.
2. Disconnect the following wires on the main board. Refer to Figure 6-17.
a. The line cord ground wire at J602 or J101 and the seven-conductor
ribbon cable at J131 from the power supply module.
b. The ribbon cable at J603 or J103 from the front panel module.
c. The ribbon cable at J202 or J102 from the display module.
3. From the bottom of the internal chassis, slightly bend the two securing tabs
and tilt the board down until it clears the tabs. Refer to Figure 6-17.
4. Pull the board out of the front slots of the internal chassis (near the BNC
connectors).
6--26
TDS 200 Series Digital Oscilloscope Service Manual
Internal chassis
Display
module ribbon
cable
Front panel cable
Maintenance
Line cord ground wire
Power supply
ribbon cable
Main board
Figure 6-17: Main board removal
Installation. Use this procedure to install the main board module. Refer to
Figure 6-18.
1. Working from the bottom of the internal chassis, place the front of the main board module into the slots on the internal chassis (near the BNC connec­tors).
2. Keeping the module in place, press the board into place so that the securing tabs lock onto the bottom of the board. Note that there are two board guides on the chassis that must match the cutouts on the main board.
3. Reconnect the following wires on the main board. Refer to Figure 6-17.
a. The line cord ground wire at J602 or J101 and the seven-conductor
ribbon cable at J131 from the power supply module.
b. The ribbon cable at J603 or J103 from the front panel module.
c. The ribbon cable at J202 or J102 from the display module.
4. Place the entire internal assembly (housing all of the boards and the display
screen) into the front case as shown in Figure 6-16.
TDS 200 Series Digital Oscilloscope Service Manual
6--27
Maintenance
5. Use the installation procedures for each module removed to reassemble the
instrument.
Line cord ground wire
Power supply
ribbon cable
Internal chassis
Display
module ribbon
cable
Front panel cable
Display Module
Board guides
Cutout
Main board
Figure 6-18: Main board installation
You will need a torque-limiting Torx T-15 screwdriver and a flat-blade screwdriver for this procedure.
Removal. First remove the handle, power button, and rear case using the procedures on pages 6--12, 6--13, and 6--14. Then, use the following procedure to remove the display module.
1. Remove the entire internal assembly from the front case using the procedure
on page 6--24.
2. Disconnect the display module ribbon cable at J202 or J102 on the main
board by pulling straight up from the connector. Refer to Figure 6-17.
3. Disconnect the backlight cable from the power supply module. Refer to
Figure 6-14.
6--28
4. Slightly bend the two securing tabs on the left side of the display module and
lift the left side up until it clears the tabs. Then slide the display to the left until the right-side tabs are disengaged. Refer to Figure 6-19.
TDS 200 Series Digital Oscilloscope Service Manual
Maintenance
CAUTION. The display shield is attached to the display module. Removing the shield will destroy the attaching gasket.
Internal chassis
Backlight
cable
Display module
Ribbon cable
Figure 6-19: Removing the display module
Installation. Use this procedure to install the display module.
1. Route the display ribbon cable through the opening in the internal chassis as shown in Figure 6-20. If installing a new module, fold the ribbon cable as shown.
2. Place the tabs on the right side of the display module into the slots of the internal chassis located under the front panel board. It is not necessary to remove the front panel board. Refer to Figure 6-20.
3. Route the backlight cable through the opening in the internal chassis as shown in Figure 6-20.
4. Lower the left side of the display module into the internal chassis until the two securing tabs lock on to the display module. Note that the backlight cable protruding from each end of the module must align with the cutouts of the inner chassis. Refer to Figure 6-20.
5. Reconnect the display module ribbon cable at J202 or J102 on the main board by pushing the cable straight down into the connector. Refer to Figure 6-17.
TDS 200 Series Digital Oscilloscope Service Manual
6--29
Maintenance
6. Reconnect the backlight cable to the power supply module. Refer to
Figure 6-14.
7. Place the entire internal assembly into the front cover. Refer to Figure 6-16.
8. Use the installation procedures for each module removed to reassemble the
instrument.
Backlight cable to
power supply
module
Display module
Do not pinch backlight cable when installing the display module.
Left side view of
internal chassis
Figure 6-20: Installing the display module
Internal chassis
Ribbon cable to
main board
6--30
Front Panel Module
You will need a torque-limiting Torx T-15 screwdriver and a flat-blade screwdriver for this procedure.
Removal. First remove the handle, power button, and rear case using the procedures on pages 6--12, 6--13, and 6--14. Then, use the following procedure to remove the front panel module.
1. Remove the entire internal assembly from the front case using the procedure
on page 6--24.
2. Disconnect the front panel ribbon cable at J200 on the main board. Refer to
Figure 6-17.
TDS 200 Series Digital Oscilloscope Service Manual
Maintenance
3. Disengage the two securing tabs at the bottom of the board and slide the board down and out of the internal chassis. Refer to Figure 6-21
Internal chassis
Front panel module
Figure 6-21: Removing the front panel module
TDS 200 Series Digital Oscilloscope Service Manual
6--31
Maintenance
Installation. Use this procedure to install the front panel module.
1. Route the front panel ribbon cable through the opening in the chassis as
shown in Figure 6-22.
2. Insert the top of the front panel module under the tabs at the top of the
chassis and press the board down until the securing tabs at the bottom of the chassis lock into place. Refer to Figure 6-21.
3. Reconnect the front panel ribbon cable at J603 or J103 on the main board.
Refer to Figure 6-17.
4. Place the entire internal assembly into the front case as shown in Figure 6-15.
5. Use the installation procedures for each module removed to reassemble the
instrument.
Internal chassis
6--32
Ribbon cable to J202 on main board
Front panel module
Figure 6-22: Installing the front panel module
TDS 200 Series Digital Oscilloscope Service Manual
Maintenance
Keypad
You will need a torque-limiting Torx T-15 screwdriver and a flat-blade screwdriver for this procedure.
Removal. First remove the handle, power button, and rear case using the procedures on pages 6--12, 6--13, and 6--14. Then, use the following procedure to remove the switch keypad.
1. Remove the entire internal assembly from the front case using the procedure on page 6--24.
2. Lift the keypad out of the front case. Refer to Figure 6-23.
CAUTION. Keep the electrical contacts on the back of the keypad clean. Skin oils and dust will inhibit good electrical contact.
Keypad
Guide post
Figure 6-23: Removing and installing the keypad
TDS 200 Series Digital Oscilloscope Service Manual
Conductive foam pad
Front case
6--33
Maintenance
Installation. Use this procedure to install the switch keypad.
1. Align the switch keypad into the front case using the five guide posts as
shown in Figure 6-23.
CAUTION. Keep the electrical contacts on the back of the switch pad clean. Skin oils and dust will inhibit good electrical contact.
2. Check that all buttons are properly seated into place before proceeding.
3. Place the entire internal assembly into the front case as shown in Figure 6-16.
4. Use the installation procedures for each module removed to reassemble the
instrument.
6--34
TDS 200 Series Digital Oscilloscope Service Manual
Maintenance
Display Shield
You will need a torque-limiting Torx T-15 screwdriver and a flat-blade screwdriver for this procedure.
STOP. Removing the display shield from the display module will destroy the display shield gasket. Only remove the shield if you intend to replace it with a new shield and gasket assembly.
Removal. First remove the handle, power button, and rear case using the procedures on pages 6--12, 6--13, and 6--14. Then, use the following procedure to remove the display shield.
1. Remove the entire internal assembly from the front case using the procedure on page 6--24.
2. Remove the shield from the display module. Refer to Figure 6-24.
NOTE. You may have to cut through the display shield gasket to remove the shield. Carefully remove any remaining gasket and adhesive from the LCD display module.
Display shield
Figure 6-24: Removing the display shield
Internal chassis
Front case
TDS 200 Series Digital Oscilloscope Service Manual
6--35
Maintenance
Installation. Use this procedure to install the display shield.
1. Align the display shield into the front case, using the two guide posts as
shown in Figure 6-25.
2. Remove the paper backing from the adhesive lining of the gasket.
3. Carefully place the entire internal assembly into the front case as shown in
Figure 6-25 making sure that the display module properly contacts the display shield foam.
4. Use the installation procedures for each module removed to reassemble the
instrument.
Internal assembly
Align display shield with
guide
posts on the inside of
the front case.
Figure 6-25: Installing the display shield
Display shield foam with adhesive
Displa y shield
Guide post
Front case
Conductive foam pad (TDS 210 and TDS 220)
6--36
TDS 200 Series Digital Oscilloscope Service Manual
Maintenance
Front Case
You will need a torque-limiting Torx T-15 screwdriver and a flat-blade screwdriver for this procedure.
Removal. First remove the handle, power button, and rear case using the procedures on pages 6--12, 6--13, and 6--14. Then, use the following procedure to remove the display shield.
1. Remove the entire internal assembly from the front case using the procedure on page 6--24.
2. Remove the switch keypad using the procedure on page 6--33.
3. Remove (TDS 210 and TDS 220 only) the conductive foam pad from the
front case (see Figure 6-15 on page 6--24).
Installation. Use this procedure to install the front case. Refer to Figure 6-15 on page 6--24.
1. Place (TDS 210 and TDS 220 only) the conductive foam pad into the front case.
2. Place the switch keypad into the front case.
3. Place the entire internal assembly into the front case.
4. Use the installation procedures for each module removed to reassemble the
instrument.
5. If installing a new front case, place a new front-panel label onto the case as described on page 6--10.
TDS 200 Series Digital Oscilloscope Service Manual
6--37
Maintenance

Troubleshooting

This section contains information and procedures to help you isolate a defective module in the oscilloscope.
WARNING. Before doing this or any other procedure in this manual, read the
General Safety Summary and Service Safety Summary found at the beginning of this manual. Also, to prevent possible injury to service personnel or damage to electrical components, read Preventing ESD on page 6--1.
Adjustment After Repair
Required Tools and
Equipment
If you replace the main board or power supply, you must adjust the instrument after repair. Refer to the chapter Adjustment Procedures.
You may need the the following tools and equipment to troubleshoot the instrument.
Tools and equipment Example
Oscilloscope with 1X/10X and 100X probes Tektronix TDS 200- or TDS 300-series with
standard accessory 1X/10X probe and optional 100X probe
DMM Tektronix TX-series digital multimeter
6--38
TDS 200 Series Digital Oscilloscope Service Manual
Maintenance
Begin Here
Begin with this troubleshooting tree to isolate a problem to a specific trouble­shooting procedure.
Plug the instrument into a proper power source and
press the power button.
Does the
instrument
turn on?
Yes
Is the display functioning
properly?
Yes
Check diagnostic screen
displayed at power-on.
No
No
Go to
Troubleshooting
Power Supply
(page 6--40).
Is there
a display?
Yes
Is display
color abnormal
(dark blue, green, or
orange)?
No
Go to Checking
Probe Comp
(page 6--40).
Yes
Go to Troubleshooting
Display (page 6-- 42).
Any
failed
messages?
No
Yes
Go to Checking
Probe Comp
(page 6--40).
No
Does
backlight
come on?
Yes
Instrument should be functioning.
Try the Other Troubleshooting
procedures (page 6-- 46) and the
Front-Panel Troubleshooting
(page 6--43) if other problems exist.
No
Go to Troubleshooting
Backlight (page 6-- 41).
TDS 200 Series Digital Oscilloscope Service Manual
6--39
Maintenance
PROBE COMP Output
Power Supply
Troubleshooting
When the instrument is on and operating properly, the PROBE COMP output should generate a square wave, approximately 5 V in amplitude, at a 1 kHz frequency. Use the oscilloscope and set the Attenuation switch to 10X on the P2100 probe to probe this output.
Signal at PROBE COMP Possible problem
1 kHz signal, no display Go to Display Troubleshooting (Page 6--42)
No signal Go to Power Supply Troubleshooting (Page 6--40)
> 1.2 kHz signal Go to Power Supply Troubleshooting (Page 6--40)
Follow these steps to troubleshoot the power supply.
1. Remove the rear case using the procedure Rear Case on page 6--14.
2. Use the test oscilloscope to measure the voltages from the power supply
module at J131 on the main board module. The table below lists the voltages you should expect to see.
Voltage at J131 Description
Pin 1 (marked with A ) +4.85 VDC to +5.15 VDC
Pin 2 +4.85 VDC to +5.15 VDC
Pin 3 GND
Pin 4 --4.35 VDC to --5.15 VDC (slightly higher if disconnected)
Pin 5 -- 2 3 V D C t o -- 2 5 V D C
Pin 6 GND
Pin 7 1.4 V
pk-pk
to5V
, 47 to 63 Hz square wave
pk-pk
3. If all of the voltages are present, the main board is probably defective.
Replace it.
4. If all or some of the voltages are missing, continue with step 5.
5. Turn off the instrument and disconnect the cable at connector J131 on the
main board module.
6. Turn on the instrument and check the voltages at the loose end of the cable
disconnected from J131, checking for the same voltages as in step 2.
6--40
TDS 200 Series Digital Oscilloscope Service Manual
Maintenance
7. If all of the voltages are missing, check the line fuse on the power supply
module. If the fuse is defective, replace the fuse. If the fuse is ok, continue with step 8.
8. If all or some of the voltages are missing, the power supply module is
probably defective. Replace it.
Backlight Troubleshooting
Follow these steps to troubleshoot a nonfunctional backlight.
1. Remove the rear case using the procedure Rear Case on page 6--14.
2. Turn the instrument on.
WARNING. To avoid electrical shock, do not touch the two-wire backlight connector on the power supply module. The output connector is the one closest to the power line cord.
3. Set the test oscilloscope (with a 100X probe) to measure a 1270 V
(450 V
) signal at approximately 60 kHz. Attach the probe ground lead to
RMS
pk-pk
the ground lug on the AC line connector and the probe tip to the backlight connector as shown in Figure 6-26.
AC ground lug
Figure 6-26: Measuring the backlight voltage
4. If the 1270 V
pk-pk
Replace the display. If the signal is not present proceed with step 5.
TDS 200 Series Digital Oscilloscope Service Manual
Backlight voltage
test point
signal is present, the backlight is probably defective.
6--41
Maintenance
Display Troubleshooting
5. If the 1270 V
signal is missing, the power supply module is probably
pk-pk
defective. Replace it.
Follow these steps to troubleshoot a nonfunctional display.
1. Remove the rear case using the procedure Rear Case on page 6--14.
2. Turn the instrument on.
NOTE. Without a functional display module, the simplest way to verify that the instrument is on and functioning is to check for a signal at the PROBE COMP output. The PROBE COMP output generates approximately a 5 V, 1 kHz squarewave.
3. Use the test oscilloscope and set the Attenuation switch to 10X on the P2100
probe to probe the display cable connector at J202 (two channel instruments) or J102 (four channel instruments) on the main board. The table below describes the signals you should expect to see.
Signal at J202/J102 Description
Pin 1 (marked with A ) -- 1 0 V DC t o -- 2 2 V DC
Pin 2 -- 2 3 V D C t o -- 2 5 V D C
Pin 3 Digital data, high = 5 V, low = 0 V
Pin 4 Digital data, high = 5 V, low = 0 V
Pin 5 Digital data, high = 5 V, low = 0 V
Pin 6 Digital data, high = 5 V, low = 0 V
Pin 7 0 V (ground)
Pin 8 0 V (ground)
Pin 9 +5 V
Pin 10 1.89 MHz clock (528 ns period), 50% duty cycle, appears
Pin 11
Pin 12 (closest to BNC end of instrument)
DC
in bursts, high = 5 V, low = 0 V
19.3 kHz clock (51.7 s period), 16% duty cycle, high=5V,low=0V
75.5 Hz clock (13.2 ms period), 0.4% duty cycle, high=5V,low=0V
4. If all the signals are present, the display module is probably defective.
Replace it. If all or some of the signals are missing, continue with step 5.
6--42
TDS 200 Series Digital Oscilloscope Service Manual
Loading...