The operation and maintenance procedures recommended by STERIS® are described in this manual.
Only these recommended maintenance procedures
should be followed.
A listing of the safety precautions to be observed
when operating and servicing this equipment can be
found in Section 1 of this manual. Do not operate or
service this equipment until you have become familiar with this information.
Any alteration of this equipment not authorized or
performed by STERIS Engineering Service which
could affect its operation will void the warranty,
could adversely affect sterilization efficacy, could
violate federal, state, and local regulations, and could
jeopardize your insurance coverage.
The Harmony
tion System is a variable pattern, variable intensity
surgical lighting fixture designed to provide visible
illumination of the surgical field or the patient for the
operating room staff.
Section 1: Listing of Safety Precautions..........................................................................1-1
1.1 Definition of Symbols ..................................................................................................................................1-4
Section 2: General Information ..........................................................................................2-1
2.1 General Description..................................................................................................................................... 2-1
2.2 Harmony LA 300/500/700 Lighthead Optical Performance ........................................................................... 2-1
2.3 Special Consumables ................................................................................................................................. 2-2
3.1 General Instructions ....................................................................................................................................3-1
3.1.1 To Adjust Lighthead Intensity Levels: ................................................................................................ 3-1
3.1.2 Control Center ...................................................................................................................................3-1
3.1.3 Surgeon’s Control Buttons.................................................................................................................3-2
3.2 Harmony LA 500 Lighthead Lamp Failure Indications ..................................................................................3-2
3.3 Harmony LA 500 Lighthead Positioning ....................................................................................................... 3-3
3.4 Harmony LA 500 Lighthead Pattern Adjustment .......................................................................................... 3-4
3.5 Lighthandle and Lighthandle Cover .............................................................................................................. 3-4
3.5.1 Installing and Using Lighthandle and Lighthandle Cover ................................................................... 3-4
3.5.2 Metal Sterilizable Handle ................................................................................................................. 3-6
3.6 Harmony LA Monitor Arms .......................................................................................................................... 3-7
3.9 Lighthead Positioning with Video Camera .................................................................................................... 3-9
3.10 Video Camera Operation.......................................................................................................................... 3-10
3.10.1 HERMES-Ready Control Center .................................................................................................. 3-10
3.10.2 Optional Foot Control ................................................................................................................... 3-10
3.11 Guidelines for Maximizing Video Image ...................................................................................................3-11
3.12 DeepSite control center status indicators ................................................................................................ 3-12
3.12.1 Intensity Level Status ................................................................................................................... 3-12
3.12.2 Lamp Life Status .......................................................................................................................... 3-12
Section 4: Service Mode ....................................................................................................4-1
4.1 Service mode procedure ............................................................................................................................. 4-1
4.2 Service Menu 1 Screen ............................................................................................................................... 4-1
4.3 Service Menu 2 Screen ............................................................................................................................... 4-2
4.4 Service Menu 3 Screen ............................................................................................................................... 4-3
Section 5: Theory of Operation .........................................................................................5-1
5.1 Suspension System .................................................................................................................................... 5-1
5.2 Major System Components .........................................................................................................................5-1
5.2.1 Harmony LA 500 LightHead ............................................................................................................. 5-1
5.2.2 Harmony LA 300 Lighthead ..............................................................................................................5-1
5.2.3 Harmony LA 700 Lighthead ..............................................................................................................5-1
5.2.4 Flat Panel Monitor Arm .................................................................................................................... 5-1
5.2.6 DeepSite Arm .................................................................................................................................. 5-2
5.3 Control System ...........................................................................................................................................5-2
5.4 Lamp Power Explanation .............................................................................................................................5-2
5.7 Serial Number Explanation .......................................................................................................................... 5-3
5.8 Harmony LA Worldwide Release Control System Theory of Operation ......................................................... 5-3
Section 6: Inspection and Maintenance............................................................................6-1
6.1 General ....................................................................................................................................................... 6-1
6.2 Cleaning the Equipment ..............................................................................................................................6-1
Section 7: Field Test Procedure .........................................................................................7-1
7.1 General ....................................................................................................................................................... 7-1
7.1.1 Special Tools Required ...................................................................................................................... 7-1
7.3 General Test ................................................................................................................................................ 7-2
7.4 Mechanical Test .......................................................................................................................................... 7-2
7.5 Optical Test (Harmony LA 500 & 700).......................................................................................................... 7-2
7.6 Suspension System Field Test Procedure ...................................................................................................7-3
7.7 Wall Control Field Test Procedure ................................................................................................................ 7-4
Section 8: Component Repair and Replacement.............................................................8-1
8.2 Light or monitor Spring Arm Spring Tension Adjustment ..............................................................................8-2
8.2.1 Light Spring Arm.............................................................................................................................. 8-2
8.2.2 Monitor Spring Arm.......................................................................................................................... 8-2
8.4 Knuckle Cover Removal (Any) and Assembly .............................................................................................8-3
8.5 De-install harmony LA 500 Lighthead from the Suspension Arm ................................................................. 8-3
8.5.1 Before Installing Monitor Spring Arm ................................................................................................ 8-4
8.6 Removal of the Control Center Front Panel .................................................................................................. 8-5
8.7 Control Center Precautions ......................................................................................................................... 8-6
8.8 Commutator Replacement (Main Suspension, not spring arm) .................................................................... 8-7
8.9 Voltage Adjustment on Power Supply ......................................................................................................... 8-8
8.15 Discoloration of Lightheads ..................................................................................................................... 8-14
8.16 Using Jumper Wires ................................................................................................................................ 8-14
8.16.1 Using Jumper Wires 755717-334 .................................................................................................. 8-14
8.16.2 Using Jumper Wires 755717-352 and 755717-353 ........................................................................ 8-14
8.16.3 Using Jumper Wire 755717-302 For Troubleshooting Harmony LA Camera System ...................... 8-15
Section 9: Illsutrated Parts Breakdown ............................................................................9-1
v
764330-226
List of Illustrations
DescriptionPage Number
Section 3: Operating Instructions
Figure 3-1. Harmony Control Center Main Menu ................................................................................................ 3-1
Figure 3-2. Lighthead Surgeon’s Control Buttons (Harmony LA 500 Lighthead with Standard Handle Shown) .... 3-2
Figure 3-3. Lamp Status LED ............................................................................................................................ 3-2
Figure 3-9. Optional Camera Installation ...........................................................................................................3-9
Figure 3-10. Install Camera Disposable Sterile Cover ........................................................................................ 3-9
Figure 3-11. Optional Camera Controllers ........................................................................................................ 3-11
Figure 3-12. DeepSite Control Center Indicators .............................................................................................. 3-12
Section 4: Service Mode
Figure 4-1. Service Menu 1 Screen .................................................................................................................... 4-1
Figure 4-2. Service Menu 2 Screen .................................................................................................................... 4-2
Figure 4-3. Service Menu 3 Screen .................................................................................................................... 4-3
Section 5: Theory of Operation
Figure 5-1. View of Global Control Power Status Boards ................................................................................... 5-4
Figure 5-2. Wedge Lock and Vertical Tube ......................................................................................................... 5-6
Figure 5-3. Typical Harmony LA System ........................................................................................................... 5-7
Figure 5-14. Domestic Small Lighthead Controller Board ................................................................................. 5-15
Figure 5-15. Domestic Lighthead Camera Module ............................................................................................ 5-15
Figure 5-16. Domestic Video & Power Kits ...................................................................................................... 5-16
Figure 5-17. Global Medium/Large Lighthead Board Controller ......................................................................... 5-17
Figure 5-18. Global Harmony System Block Diagram ...................................................................................... 5-17
Figure 5-19. Global Deepsite Illuminator Box ................................................................................................... 5-18
Figure 5-20. Global Master Control Board w/DeepSite Interface Board & Analog Video Interface Board .......... 5-18
Figure 5-21. Global LCD Interface Board w/Membrane Switch Panel, Service Port AUX Port,
IR Receiver, & Foot Control Assembly ........................................................................................... 5-19
Figure 5-22. Global Power Supply & Power Status Module .............................................................................. 5-20
Figure 5-23. Global Remote Power Module ...................................................................................................... 5-21
vi
764330-226
List of Illustrations (Continued)
DescriptionPage Number
Section 5: Theory of Operation (Continued)
Figure 5-24. Global Hub Interface Board .......................................................................................................... 5-21
Figure 5-25. Global Hub Power Circuit ............................................................................................................. 5-22
Figure 5-26. Global Schematic ........................................................................................................................ 5-23
Figure 5-26. Global Small Lighthead Controller Board ...................................................................................... 5-23
Figure 5-27. Global Lighthead Camera Module ................................................................................................ 5-23
Figure 5-28. Global Power & Video Kits ........................................................................................................... 5-24
Section 6: Inspection and Maintenance
Figure 6-1. Areas To Be Cleaned ....................................................................................................................... 6-2
Figure 6-2. Lamp Status LED ............................................................................................................................ 6-4
Figure 7-2. Place Light Meter on Table ...............................................................................................................7-2
Figure 8-2. Adjust Support Arm/Spring Arm Brake Tension ................................................................................ 8-1
Figure 8-3. Adjust Spring Arm/Yoke Brake Tension ............................................................................................8-1
Figure 8-4. Adjust Upper Support Arm Brake Tension ........................................................................................ 8-1
Figure 8-5. Remove Cover to Access Spring Arm Tension Adjustment .............................................................. 8-2
Figure 8-6. Adjust Spring Arm Tension ...............................................................................................................8-2
Figure 8-7. Monitor Spring Arm Tension Adjustment Access ............................................................................. 8-2
Figure 8-8. Harmony Wall Control ON/OFF Switch............................................................................................. 8-3
Figure 8-9. Push Spring Arm and Lighthead Toward Floor .................................................................................. 8-3
Figure 8-35. Remove Allen Wrench ................................................................................................................. 8-12
Figure 8-36. Lide in New Lens Clip .................................................................................................................. 8-12
Table 2-1. Lighthead Performance Data ............................................................................................................. 2-1
Table 2-1. Special Consumables........................................................................................................................2-2
Section 4: Service Mode
Table 4-1. Default Control Center Settings ......................................................................................................... 4-3
Section 5: Theory Of Operation
Table 5-1. Lighthead: Large Socket: Primary and Secondary ........................................................................... 5-2
Table 5-2. Lighthead: Medium Socket: Primary ................................................................................................ 5-2
Table 5-3. Lighthead: Medium Socket: Secondary ........................................................................................... 5-3
Table 5-4. Lighthead: Small Socket: Primary ...................................................................................................5-3
Table 6-2. Harmony Lighting and Visualization System Preventive Maintenance Guide ................................... 6-18
ix
764330-226
Section 1: Listing of Safety Precautions
The following Safety Precautionsmust be observed when operating or servicing this surgical lighting system.
WARNING indicates the potential for personal injury and CAUTION indicates the potential for damage to
equipment. For emphasis, certain Safety Precautions are repeated throughout the manual. It is important to
review ALL Safety Precautions before operating or servicing the unit.
Strictly following these Safety Precautions enhances your ability to safely and effectively utilize the unit and
helps the customer avoid improper maintenance methods which may damage the unit or render it unsafe. It
is important to understand that these Safety Precautions are not exhaustive; customers are encouraged to
develop their own safety policies and procedures to enhance and compliment these Safety Precautions.
WARNING – ELECTRIC SHOCK AND BURN HAZARD:
Disconnect all utilities to lighting fixture before servicing. Do not install the lighting fixture unless all utilities
have been properly locked out. Always follow OSHA Lockout-Tagout and electrical safety-related work
practice standards. (See 29 CFR 1910.147 and .331 through .335.)
Do not remove covers or perform service other than as described in this equipment manual. Refer servicing
to qualified service personnel. (Maintenance Manual P764330-226.)
WARNING – EXPLOSION HAZARD:
Do not use lighting fixture in the presence of flammable anesthetics.
WARNING – PINCH AND CRUSH HAZARD:
Do not pinch fingers between the structural plate and the vertical tube as the assembly is lifted.
The safety ring must be installed.
Pinch points are created during extreme articulation of the suspension system. Do not place hands on or near
the suspension knuckle during lighthead articulations.
WARNING –IMPACT HAZARD:
Do not remove tension screw from the spring arm joint until the lighthead has been securely installed onto
the spring arm. Note: This warning applies at both installation and de-installation procedures.
WARNING – PERSONAL INJURY HAZARD:
Do not attempt to replace the lamp unless power is turned off and the lighthead has cooled sufficiently.
Do not attempt to clean lighthead unless power is turned off and the lighthead has cooled sufficiently.
Do not remove monitor-arm safety pin until monitor is installed.
Monitor extension arm must have monitor (or equivalent counterweight) installed to keep arm in required
positions while installing sliding guards.
Avoid looking directly at high-intensity light, whether at the lamp or directly at the lighthead. Eye injury may
result.
1-1
764330-226
WARNING – POSSIBLE PATIENT INJURY HAZARD:
Failure to engage the lighthandle cover completely may result in cover falling from lighthead during the
procedure.
WARNING – BIOHAZARD:
Sterile disposables are intended for single use only.
Universal precautions must be observed when disposing of any single use disposable item.
WARNING – DISPOSAL HAZARD:
This product contains materials which may require disposal through appropriately licensed and permitted
hazardous waste management firms.
CAUTION – POSSIBLE EQUIPMENT DAMAGE:
Use of any disinfectant solution OTHER than those listed below may cause discoloration or deformation on
the polycarbonate lens surface: Coverage® Spray HBV, Coverage® HBV Concentrate, T.B.Q.®, or Coverage
Plus®, CoverageTM Spray Disinfectant Cleaner, or Germicidal Cloth. Cleaning solutions other than those listed
have NOT been tested for compatibility or effectiveness. Always follow manufacturer instructions for
concentrations and use of cleaning products.
Use only recommended cleaning/disinfecting and/or anti-static agents on this light. Some degree of staining,
pitting, and/or discoloration could occur if a phenolic-, iodophor-, or glutaraldehyde-based disinfectant is
used on the surfaces of this light. Also, use of alcohol or aerosol spray cleaner/disinfectants (e.g., Lysol®)
containing a substantial amount of alcohol in the formula can damage the polycarbonate lens.
Prevent leakage of fluids into interior of lighthead.
Avoid discoloration of control center keypad and display. Do not clean control center with Betadine® solutions
or allow such solutions to contact keypad and display surfaces.
Do not scratch optical coating on accessible portions of optic assembly when cleaning; always wear rubber
gloves and use only a clean, white, lint-free cloth when wiping internal surfaces.
Appropriate components of this lighting system have been tested and found in compliance with IEC 60601-
1-2 (1st Edition 1993-04), Medical Electrical Equipment – Part 1: General Requirements for Safety;
Electromagnetic Compatibility (EMC). There is, however, a potential for electromagnetic or other interference
between this equipment and other devices. Should you experience interference, relocate this device or
minimize the use of the affected equipment while this device is in operation.
Do not touch the glass portion of the lamp with bare fingers. Skin oils can cause deterioration of material
leading to possible failure of the lamp.
Accurate measurement and calculation in finding the point where the first hole is drilled for the safety ring is
of critical importance in correctly installing the vertical tube. Check all measurements and calculations twice
before drilling the first hole.
Do not pinch wires that exit the vertical tube as the assembly is lifted. Gently route wires from the vertical tube
out of the way as the suspension system is lifted into place.
1-2
764330-226
CAUTION – POSSIBLE EQUIPMENT DAMAGE (Continued):
Do not pinch wire on the ceiling structure or vertical tube.
During all mounting steps, ensure wires are not pinched between mechanical components of suspension,
spring arms, or lightheads.
Harness connectors must be slipped into expandable sheathe before routing through monitor yoke.
• Failure to use sheath may cause connector damage during routing.
• Slip sheathe only far enough onto the harness to protect connector bundle. If sheathe is slipped too far onto
the harness, removing it after routing may be difficult.
Do not touch glass portion of lamp with bare fingers. Skin oils can cause deterioration of material leading to
possible failure of the lamp.
Sliding guards may be damaged if not installed properly into internal grooves.
On early light systems, the wire clamp must be loosened to allow the wiring within the vertical tube to be free.
1-3
764330-226
1.1 DEFINITION OF
SYMBOLS
SymbolDefinition
ON
STANDBY
Lamp Out
Protective Earth (Ground)
Attention, consult manual for further instructions
Hot
SER. NO.Serial Number of Unit
V~Voltage Rating of Unit, Alternating Current
AAmperage Rating of Unit
HzFrequency Rating of Unit
+Increase Intensity (Surgeon's Control Buttons)
-Decrease Intensity (Surgeon's Control Buttons)
Zoom
Rotate
Manual Focus
1-4
764330-226
Auto Focus
Section 2: General Information
The product literature included in this section contains factual data relating to the principal characteristics of the Harmony™ LA Surgical Lighting and
2.2 HARMONY LA 300/500/700
LIGHTHEAD OPTICAL
PERFORMANCE
Visualization System.
The literature is informative rather than instructional.
It provides and conveys, through text and illustrations, a general concept of the equipment, its purpose,
limitations, and technical applications.
Important: Values are typical for the small pattern
size at highest intensity setting (unless otherwise
noted) at 39-3/8" (1m) from the lighthead. Definitions
and measurements are in accordance with
IEC 60601-2-41.
2.1 GENERAL DESCRIPTION
The Harmony LA Surgical Lighting and Visualization System is a variable pattern, variable intensity
surgical lighting fixture designed to provide visible
illumination of the surgical field or the patient for the
operating room staff.
Table 2-1. Lighthead Performance Data
Harmony LA 300Harmony LA 500Harmony LA 700
Central
Illuminance2,180 – 5,500 fc5,480 – 13,470 fc5,480 – 14,860 fc
The wall-mounted Harmony™ LA Control Center
allows the user to adjust the lighthead intensity level
(brightness) by pressing a membrane switch (or “button”). An identifying number on the control display
corresponds to the same number on the lighthead
suspension arm. Additionally, each lighthead has its
own “onboard” intensity controls, located on the
above the lighthandle adjacent to the lens. These
controls are usually referred to as the surgeon’s
control buttons or as the surgeon’s control. (See
Figure 3-2.)
3.1.1 To Adjust Lighthead Intensity Levels:
WARNING-EXPLOSION HAZARD: Do not use
in the presence of flammable anesthetics.
1. Press ON touch pad.
2. Intensity levels for one or two lightheads can be
controlled from one control center. A lighthead’s
surgeon’s control buttons adjust the intensity level
for that lighthead.
Important: Avoid control center faults. Do not continuously press and hold any control center buttons
for more than 15 seconds.
3.1.2 Control Center
WARNING-EXPLOSION HAZARD: Do not use
in the presence of flammable anesthetics.
See Figure 3-1.
1. Use the “Select Light” touch pads to move the
display indicator up or down until it aligns with
the appropriate lighthead.
2. Once the lighthead is selected, press either the
button to increase light intensity level, or the button to decrease light intensity level.
3. If “All Lights” is selected, intensity for all lightheads in the system is increased or decreased
simultaneously.
NOTE: For longer lamp life, use lowest intensity level
suitable for surgical procedure.
Indicates Selected
Lighthead
Lighthead Intensity
Status
Lamp Status
Figure 3-1. Harmony Control Center Main Menu
Indicators
3-1
764330-226
Surgeon's
Control
Buttons
Figure 3-2. Lighthead Surgeon’s Control Buttons (Harmony LA 500 Lighthead with Standard
Handle Shown)
3.1.3 Surgeon’s Control Buttons
(See Figure 3-2)
1. Grasp the handle of the appropriate lighthead. A
ring of membrane buttons (bumps) are located on
the control bezel adjacent to the lighthead lens.
2. Press any of the surgeon’s control buttons molded
with “+” symbols to increase the lighthead intensity. Press any of the surgeon’s control buttons
molded with “-” symbols to decrease lighthead
intensity.
NOTE: For longer lamp life, use lowest intensity level
suitable for surgical procedure.
3. To turn lighthead OFF.
a. At the control center: press the “Select Light”
button until indicator points to the appropriate
lighthead on the display. Press the “Decrease
Intensity” button until the light goes out.
b. At the lighthead, press and hold any of the “-”
membrane buttons untilthe light goes out.
NOTE: Press the decrease intensity button (-) on
the lighthead surgeon’scontrol for an additional 5
seconds, to turn off all lightheads in the system.
nity.
A Lamp Status LED is located on the lighthead (see
Figure 3-3). If this LED is flashing, it indicates that
the lighthead is using the backup lamp and the
primary lamp should be replaced.
Check the lamp status LED each time the surgical
light is used.
A secondary lamp inside Harmony LA 500 lightheads automatically turns on when the primary
lamp fails.
If the lamp status LED is blinking, replace the failed
lamp(s). (See Lamp Replacement, Section 6.) After
the lamp is replaced and power restored to the
lighthead, the primary lamp lights, the LED stops
blinking and the graphic display changes to “lamp
ready.”
3.2 HARMONY LA 500 LIGHTHEAD
LAMP FAILURE INDICATIONS
The main display screen on the control center indicates lamp status for the system lightheads. The
display indicates the status of both the primary and
secondary lamps for each lighthead. A failed lamp
should always be changed at the earliest opportu-
764330-226
Check Lamp Status
LED on Lighthead
Figure 3-3. Lamp Status LED
3-2
3.3 HARMONY LA 300/500/700
Monitor Arm Positioning
Lighthead Positioning
LIGHTHEAD POSITIONING
CAUTION: Do not bump lightheads into walls
or other equipment.
Lightheads can be positioned by using either the
sterile handle, or by grasping the non-sterile handle
around the lighthead housing. The following two
paragraphs describe the positioning characteristics
of the lighthead from outside or within the sterile
field. To optimize shadow control, position the lighthead as appropriate before starting the intended
surgical procedure.
Lighthead may (1) rotate continuously around vertical suspension tube; (2)rotate continuously around
central hub; (3) rotate continuously about suspen-
sion arm; (4) tilt forward or backward in yoke approximately 310°; and (5) move up or down by
pivoting at suspension elbow 15° up, and 85° down
until vertical tube and suspension arm describe a
straight line.
Monitor arms are capable of the following articulations: (1) rotate 300° at the central or secondary
spindle; (2) rotate 330° at the horizontal extension
arm; (3) rotate 320° a the yoke transition; (4) move
up or down by pivoting at spring arm knuckle 20°
up, and 40° down. Flat panel yoke also allows (5) tilt
forward 15° or backward 75° in yoke. Additionally,
the flat panel yoke can be rotated for either portrait
or landscape display. See Figure 3-4.
15°
Typical
300°
Unlimited
330°
320°
85°
Typical
Unlimited
310°
20°
40°
320°
75° Up
15° Down
180°
Portrait to
Landscape
Figure 3-4. Suspension Positioning
3-3
764330-226
13°
Rotate Handle
to Decrease
Rotate Handle
to Increase
Figure 3-5. Light Pattern Adjustment
3.4 HARMONY LA 500 LIGHTHEAD
PATTERN ADJUSTMENT
The illumination pattern for each lighthead can be
adjusted to any diameter between the minimum and
the maximum.
Adjust the light pattern by rotating the lighthandle
clockwise to decrease pattern size.
Rotate the handle counterclockwise to increase pattern size.
NOTE: Typical lighthead positioning is above and slightly
behind the surgeon’s right or left shoulder.
3.5 LIGHTHANDLE AND
LIGHTHANDLE COVER
3.5.1 Installing and Using Lighthandle and
Lighthandle Cover
WARNING – POSSIBLE PATIENT INJURY
HAZARD: Failure to engage the lighthandle
cover completely may result in cover falling
from lighthead during the procedure.
Lighthandles are used to position Harmony LA 500
lightheads, allow access to intensity control on each
lighthead, and to adjust lighthead pattern size. To
3-4
764330-226
avoid accidental contact with the non-sterile
surgeon’s control buttons, the lighthandle should
always be used with a disposable, sterile lighthandle
cover (available separately from STERIS®).
• If the lighthandle adapter is not already in place,
align the tab on the adapter with the channel in
the mounting ring and thread the adapter in until
fully engaged. (See Figures 3-6a and 3-6b.)
• Prior to starting a procedure, ensure that the
lighthandle is in place. Install the lighthandle by
threading onto the adapter and firmly tightening.
(See Figure 3-6c.)
Surgeon's Control Buttons
• Remove the sterile lighthandle cover from its
packaging and install onto the lighthandle. (See
Figure 3-6d.)
• To ensure the lighthandle cover remains in place
during the procedure, the groove in the cover
must be fully engaged with the groove in the
lighthandle.
• The lighthandle can be removed for cleaning or
sterilization by unscrewing it from the handle
adapter. The lighthandle can be sterilized using
standard hospital cycles. Do not use the lighthead during a sterile procedure unless a disposable cover is installed on the lighthandle.
Lighthandle
Adapter
Align Tab on Handle
Adapter with Channel in
Mounting Ring
Figure 3-6a.
Figure 3-6b.
Lighthandle
Adapter
Lighthandle
Figure 3-6c.
Push Cover
Completely onto
Lighthandle
Disposable
Lighthandle
Cover
Figure 3-6. Standard Lighthhandle
3-5
764330-226
Figure 3-6d.
3.5.2METAL STERILIZABLE
HANDLE
WARNING-STERILITY ASSURANCE HAZARD: Do not use the surgeon’s control buttons unless a disposable sterile cover is
installed. If the sterilizable lighthandle
(metal) is used without a disposable cover,
the surgeon’s control buttons are not protected by a sterile covering.
The metal sterilizable handle is used to position the
Harmony LA lighthead and to adjust the light pattern size. The sterilizable handle may be used in
place of a disposable sterile lighthandle cover.
The plastic lighthandle should be removed prior to
installing the metal sterilizable handle. To do so,
simply unscrew it from the threaded handle adapter.
Prior to starting a procedure, install a clean, steril-
ized metal handle by screwing it onto the threaded
handle adapter. Ensure handle is firmly tightened
prior to use. The gap between the flange on the metal
steriliizable handle and the surgeon’s control buttons prevent accidental contact with the non-sterile
surface of the buttons.
The metal sterilizable handle can be removed for
sterilization by unscrewing it from the adapter.
The metal sterilizable handle can be sterilized using
standard hospital cycles steam cycles intended for
lumens. Always sterilize handle between surgical
procedures.
3-6
764330-226
CRT Monitor Support Arm
Flat Panel Montior Support Arm with
Monitor in Landscape Position
Figure 3-7. Harmony Monitor Configurations
Flat Panel Montior Support Arm with
Monitor in Portrait Position
Adjust CRT Monitor
Shelf by Loosening/
Tightening Knob
3.6 HARMONY LA MONITOR ARMS
The Harmony LA Surgical Lighting and Visualization System can be equipped with one or two monitor supportarms. Monitor arms can be included in
the Harmony system either as the uppermostarm on
the central spindle, or as the secondary spindle arm.
Refer to 3.3 Harmony LA Lighthead Positioning for
positioning range details. See Figure 3-4.
Refer to separate Operating Instructions supplied
with the monitor.
Signal and monitor power wiring is provided at the
installation with each monitor support arm.
Input signals to monitors can be routed through the
suspension wiring froman external video source
(not by STERIS).
WARNING – BIOHAZARD: Sterile disposables are intended for single use only. Universal precautions must be observed when
disposing of any single use disposable item.
Important: Use new covers for each handle for each
procedure where the monitor will be used. Covers
are sterile and shipped in a protective wrapper.
Wear sterile gloves when removing covers from
packaging. When installing covers, follow
sterilityassurance procedures as outlined by facility
protocols.
764330-226
Refer to Figure 3-8.
1. Remove cover from packaging.
2. Pull the cover until it fully covers the handle.
3. Pull closure tie until tight enough to prevent
cover from slipping down the handle.
4. Repeat procedure for second handle.
3-8
3.8 VIDEO CAMERA INSTALLATION
OR REMOVAL
3.9 LIGHTHEAD POSITIONING WITH
VIDEO CAMERA
3.8.1 Installation
1. Rotate lighthead until lens faces the room ceiling.
2. Remove the standard handle by unthreading it
from the support mounting ring.
3. The camera is secured to the lighthead by threading it into the same support used by the standard
handle.
a. Align the post with recess. This should align
system connector properly. Install camera as-
sembly to lighthead.
b. Select Camera Menu 1 on the control center.
c. Verify that all functions of the camera operate
correctly using the control center. Verify func-
tions using the optional wireless hand-held
control or optional foot control, if applicable.
4. Once camera functions are verified, press the OFF
touch pad on the face of the control center.
3.8.2 Removal
1. Rotate lighthead until perpendicular to floor.
2. Unthread camera housing from mounting ring
and remove.
When the lighthead is used with a camera, a dispos-
able sterile cover is placed over the camera housing. (Sterile cover is available from STERIS.)
Figure 3-10. Insta ll Camera Disposable
Sterile Cover
3. Install standard handle.
Threaded
Mounting Ring
Optional Video
Camera
Alignment Recess
Alignment Post
Figure 3-9. Optional Camera Installation
3-9
764330-226
3.10VIDEO CAMERA OPERATION
The optional video camera is integrated into a removable handle that can be fitted to any Harmony
LA 500 or 700 lighthead. Whenever the video option
is used, a separated sterile disposable cover must be
fitted over the camera before each procedure. This
cover allows the camera to be grasped and used in
the same way as a standard handle, including positioning the lighthead for optimal illumination, adjusting lighthead intensity and pattern size, as well
as positioning the video image field for the best view
of the procedure. The sterile cover must be removed
from the camera and discarded after each procedure.
Video and control signals are turned on or off at the
control center; any of the optional remote control
units (see below) provide similar control functions.
The control center also provides additional control
for video brightness and contrast, color gain, stillimage capture and time and date display.
Three optional remote devices can control camera
functions: a hand-held wireless control, Hermesvoice-activated control (verbal commands and handheld pendant), or a foot control.
The control center functions by transmitting signals
via a control cable to the camera located in the
lighthead.
The optional wireless control functions by transmitting infrared signals to the control center. To initiate
camera functions with either control, press the appropriate touch pad on the control face panel. The
hand-held control’s IR (infrared) transmitter must
be within approximately 15 feet maximum and 50° of
the control center receiver. The control touch pads
are used to control the following camera functions:
Zoom–for determining the level of detail visible in
the image field. The Zoom function adjusts the image field continuously between two extremes:
+ (Telephoto). At extreme telephoto, the camera
captures an image showing great detail in a
small area.
NOTE: At extreme telephoto, any motion of the lighthead/camera will be exaggerated (jerky). The field of
focus has little depth at this extreme, forcing the Auto
Focus function (if enabled) to refocus the camera when
any object (such as a hand) enters the image field, or if
camera position is adjusted. The camera is also sensitive to light level changes at extreme telephoto.
– (Wide Angle). At extreme wide angle, the cam-
era captures a large, image with less detail than
telephoto.
NOTE: At extreme wide angle, the image field possesses a greater depth of focus and less sensitivity to
light level changes.
Rotate–Use this function to change orientation of
the video field. Image field can be rotated in either
clockwise or counterclockwise directions.
Bright–Use this function to affect the overall bright-
ness (or darkness ) of the video image. The Bright
control can be set to adjust itself automatically.
Manual Focus–Use this function to manually set the
focus for close-up shots or other special applications. The Auto Focus function must be toggled off
to enable manual focusing. Adjust the clarity of
focus by pressing the + or – touch pads.
NOTE:
subject;
+
touch pad moves the lens slightly closer to
–
touch pad moves the lens slightly away from
subject.
Auto Focus – Use this function to toggle Auto Focus
on or off. When Auto Focus is ON, the camera
automatically focuses on the object in the image
field closest to the camera lens. When Auto Focus is
turned OFF, the camera maintains focus on the last
object upon which it was focused, until Auto Focus
is turned back on.
NOTE: It may be necessary to toggle Auto Focus OFF
when using the camera at extreme telephoto (close up),
to prevent the camera from refocusing on hands and
other objects introduced into the image field while
videoing procedures.
Red Gain/Blue Gain – These controls are used to
adjust image color settings, and are intended to
allow the user to compensate for changes in color
temperature provided by differing light sources.
The gain control can be set to adjust automatically.
3.10.1HERMES-Ready Control Center
The Hermes Control Center functions by transmitting signals through an interface cable connected to
the wall control. When used, the interface cable
should be routed to avoid surgical personnel foot
traffic. The control center is intended to allow personnel within the sterile field to operate all camera
functions using verbal commands or a hand-held
pendant. Refer to the Hermes operating manual for
more information.
3.10.2Optional Foot Control
The foot control functions by transmitting signals
through a cable connected to the control center.
When the foot control is used, the cable should be
routed to avoid surgical personnel foot traffic. The
foot control is intended to allow personnel within
3-10
764330-226
the sterile field to operate the zoom and rotation
Amsco® SQ240 SurgiVision
TM
Surgical Lighting and Video System
ZOOMRO TATE
functions of the camera.
NOTE: Zooming completely out causes image distortion within the illuminated area.
Foot control is used to control zoom and rotation
functions. Press on appropriate sides of foot pedals
to activate foot control functions. Functions controlled through the foot control unit are identical to
those controlled through the optional hand-held
wireless control or control center.
3.11 GUIDELINES FOR MAXIMIZING
VIDEO IMAGE
The following steps will aid in maximizing video
imaging effectiveness.
1. Energize the light at medium setting on the control center, and position the lighthead approximately 39" (1000 mm) from the surgical site.
2. Using the hand-held remote, or the control center,
zoom in or out (+ or -) until the desired image fills
the viewing screen of the monitor.
3. If the light pattern (white circle) does not fill the
monitor’s screen (shaded square), the image inside the pattern will be distorted.
4. Reposition the camera to orient the focal point
(center of the desired image) at the center of the
viewing screen.
Adjust zoom and rotation as needed:
5. The camera may be zoomed to the full 10x zoom
and rotation orientation adjusted as needed.
6. Clockwise and/or counterclockwise orientation
is adjusted by using the curved arrow(s) on the
hand-held control, the control center or the optional foot control.
3.11.1Engage Manual Focus
As instruments are introduced to and removed from
the surgical field, the auto focus will attempt to
focus on the nearest object, possibly causing the
image to blur intermittently. To prevent this effect,
once the camera has been positioned and focused
onto the surgical site, the manual focus mode may be
engaged to maintain image clarity during the procedure.
For deep cavity illumination, it may be necessary to
engage the manual focus to focus beyond the nearest
object (i.e., the surface area surrounding the incision) so that the desired image can be viewed clearly.
Optional Hand-held Wireless
Remote Control
Figure 3-11. Optional Camera Controllers
Optional Foot Control
3-11
764330-226
3.12 DeepSite CONTROL CENTER
STATUS INDICATORS
NOTE: See DeepSite Maintenance Manual P764330210 for more information on the DeepSite Light.
3.12.1Intensity Level Status
This shows the current intensity level of the DeepSite
light.
3.12.2Lamp Life Status
The DeepSite lighthead uses a high-intensity
Xenonlamp module as a light source. The Harmony
LA Surgical Lighting and Visualization System can
track the approximate life remaining for this lamp
module and display it at the Control Center.
NOTE: Refer to Lamp Replacement procedure in Section 6 of this manual when it becomes necessary to
replace the lamp.
Approximate Lamp Life
Used/Remaining
New Lamp
Spent Lamp
DeepSite Intensity Status
Indicator
Figure 3-12. DeepSite Control Center Indicators
DeepSite Lamp Life
Status Indicator
3-12
764330-226
Section 4: Service Mode
4.1 SERVICE MODE PROCEDURE
4.2 SERVICE MENU 1 SCREEN
Service screens shall be available to service personnel
for diagnostics and firmware updates of the wall
control and lighthead units. The service screens shall
be accessible by performing the following sequence:
1. Turn power ON by pressing systems button.
2. Bring up the System Menu by pressing systems
button.
3. Push and hold the ON button used to place the Wall
Control Unit into the Active Mode.
4. Depress the Select Function Up button.
After performing the sequence described above, the
first of a series of Service Menus shall be displayed.
Service Menu 1, shown in Figure 4-1, is used to read
the firmware part number and revision for the wall
control unit and all installed lightheads.
1. The context-sensitive switch under the "Lights"
description field shall change the LCD display to
the Light Control screen.
2. The context-sensitive switch under the "Menu 2"
description field shall change the LCD display to
the Service Menu 2 screen.
3. The pointer (triangle on left side of lighthead) shall
not be affected by the depression of the Up/Down
button.
4. The Left/Right buttons under the lighthead field
shall select the lighthead whose firmware part
numbers are to be displayed. If a lighthead is not
installed, the message "Not Installed" shall be displayed in the area where the firmware part numbers are displayed.
5. The lighthead numbers shall correspond to the
serial communication port/head numbers at the
Hub Interface PWB.
6. The three TYPES of lightheads shall be Small, Medium, or Large.
Figure 4-1. Service Menu 1 Screen
4-1
764330-226
4.3 SERVICE MENU 2 SCREEN
Service Menu 2, shown in Figure 4-2, shall be used to
display the programmed power and operating performance for all of the lighthead units installed.
1. The context-sensitive switch under the "Lights"
description field shall change the LCD display to
the Light Control screen.
2. The context-sensitive switch under the "Menu 3"
description field shall change the LCD display to
the Service Menu 3 screen.
3. The context switch Reset Temp shall send a command to all lightheads to clear the stored maximum
temperature reading.
4. The Lamp numbers shall correspond to the serial
communication port/Head numbers at the Hub
Interface PWB.
5. The Left/Right buttons under Lamp shall be used
to select the lighthead lamp whose intensity is to be
selected with the Up/Down buttons.
6. The Up/Down buttons under Power shall be used
to adjust the intensity (from off to a maximum
intensity of 7) of the selected lighthead lamp. Depressing and holding the Down button for more
that two seconds shall turn the lamp off. If a lamp
is off, depressing the Up button shall turn the lamp
on at its last intensity setting.
7. The duty cycle listed is in units of percent and is
rounded off to the nearest integer (less than 0.5% is
rounded down, 0.5% or greater is rounded up).
Intensity Setting of Lighthead 1-7
Power in Watts
Lamp Voltage in 22-24 V
Current to Lamp
Duty Cycle of Lamp Control Board
Temperature on Control Board, °C
Maximum Temperature seen on Control Board, °C
Figure 4-2. Service Menu 2 Screen
4-2
764330-226
4.4 SERVICE MENU 3 SCREEN
Service Menu 3, shown in Figure 4-3, shall be used to
force the Wall Control Unit to undergo a soft reset,
place the unit into a mode for reprogramming Flash
memory with updated firmware, or reinitialize user
settings, such as lamp intensities, in non-volatile
RAM.
5. Depressing the Right buttons under Reset shall
cause the Wall Control Unit to reset and perform
the selected function.
RESET WCU CPU – The WCU performs the same
startup, testing, and display of messages as if the
power was applied to the system.
RESET/REPROGRAM FLASH – After changing
screen choice to "Yes," the WCU will startup and
wait for a new application to be downloaded from
a PC running STERIS AppLoad. As an alternative,
the power can be turned off (at the incoming ac
voltage rocker switch in the rough-in box) to the
wall control, the AppLoad program started, then
upon turning the wall control back on, the software
will automatically download.
RESET/CLEAR USER SETTING – The WCU performs the same startup, testing, and display of
messages as if the power was applied to the system.
The only difference is that the user settings for the
following will be set to default values.
Table 4-1. Control Center Default Values
LANGUAGE:ENGLISH
Figure 4-3. Service Menu 3 Screen
1. The context-sensitive switch under the "Lights"
description field shall change the LCD display to
the Light Control screen described in SECTION 4.2.
2. The context-sensitive switch under the "Menu 1"
description field shall change the LCD display to
the Service Menu 1 screen described in SECTION 4.2.
3. Control of the Ambient Light is described in SECTION
5.2.7.
4. The Up/Down buttons under the Select Function
shall select the type of Wall Control Reset.
AMBIENT LIGHTNOT INSTALLED
AMBIENT LIGHT MODEMANUAL
AMBIENT LIGHTOFF
CAMERA ZOOM SPEED3
CAMERA FOCUS SPEED1
LCD CONTRASTNominal
4-3
764330-226
Loading...
+ 120 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.