![](/html/f2/f26e/f26e06ea515e6108adcd09111cbfc0b4350bb433b28aab8c70906d4ba726e6dd/bg1.png)
CAVIWAVE ULTRASONIC CLEANING SYSTEM
APPLICATION
The Caviwave Sonic Cleaner is designed to
thoroughly wash (remove tissue, blood and other
contaminants) instruments prior to final
disinfection and sterilization. The Caviwave
Ultrasonic System’s compact design makes it
suitable for use in the Operating Room, Central
Supply or other high volume area to process precleaned instruments.
DESCRIPTION
The Caviwave Ultrasonic Cleaning System Sonic
Cleaner is a freestanding instrument washer
supplied in a compact stainless-steel cabinet. The
Sonic Cleaner contains all needed components to
effectively clean (utilizing ultrasonic cavitation
passing through a detergent-balanced wash
solution) surgical instruments prior to disinfection
and sterilization.
The Sonic Cleaner is equipped with an automatic
opening lid and load elevator for easy and safe
instrument loading.
STANDARDS
SONIC CLEANER
(Typical only - some details may vary.)
FEATURES
The Caviwave Ultrasonic Cleaning System meets
the applicable requirements of the following
standards:
• EMC Directive 89/336/EEC, 92/31/EEC,
93/68/EEC.
• Low Voltage Directive 73/23/EEC, 93/68/EE.
• ETL Listed.
• Underwriters Laboratories (UL) Standard
61010-1.
• Underwriters Laboratories (UL) Standard
61010A-2-010.
• Canadian Standards Association (CSA)
Standard C22.2 No. 61010-1.
• Canadian Standards Association (CSA)
Standard C22.2 No. 61010-2-010.
The Selections Checked Below Apply To This Equipment
The Caviwave Ultrasonic Cleaning System Sonic Cleaner is wired
for operation on either 208 V (60 Hz, three phase or single phase),
or 240 V, 60 Hz, single phase voltage. Unit can be configured for
single phase. STERIS supplies all components necessary to obtain
a complete working unit ready for (but not including) installation and
connection to facility service lines. The Sonic Cleaner includes the
following features:
Optimum Ultrasonic Cleaning Power is ensured by constant
power output generators operating at 132 kHz and complemented
with complex resonance frequencies to ensure lock-box cleaning
efficiency. The generators are mounted with individual slide out
modules for ease of service. Also, patented ceramically-enhanced
transducer modules are bonded to the wash tank providing
maximum energy transmission for instrument cleaning.
Automatic Lid and E-Z Load Tray Elevator improve safety and
increase instrument loading ease. E-Z Load Tray Elevator is
positioned at a convenient working height and is oriented laterally
for ease of loading heavy instrument trays.
Port Flushing System processes up to six rigid lumened surgical
devices with both ultrasonic energy and internal flushing.
VOLTAGES
❑ 208 V, 60 Hz, 3 ph (Standard)
❑ 208 V, 60 Hz, 1 ph
❑ 240 V, 60 Hz, 1 ph
ADDITIONAL TANK SIZES
❑ 11-gal (42 L) Tank
❑ 15-gal (57 L) Tank
❑ 20-gal (76 L) Tank
ACCESSORIES
❑ Seismic Design
❑ Additional Port Flushing Kit
❑ Additional Port Flushing Tray
❑ 11-gal (42 L) Tray
❑ 15-gal (57 L) Tray
❑ 20-gal (76 L) Tray
Item ________________________
Location(s)___________________
____________________________
SD847 (08/01/08)
![](/html/f2/f26e/f26e06ea515e6108adcd09111cbfc0b4350bb433b28aab8c70906d4ba726e6dd/bg2.png)
Illuminated Main Controls include power ON/OFF, automatic
fill, drain alert with cycle counter, adjustable wash cycle timer,
drain actuator and lid actuator.
Locking Caster Wheels are supplied for ease of Caviwave
Ultrasonic Cleaning System Sonic Cleaner movement. The four
caster wheels add 2" (51 mm) to total Sonic Cleaner height.
Large Capacity Process Tank accepts a large variety of
instrument trays. One instrument tray is included with each unit.
Tank weight capacity is as follows:
• 11-gal (42 L) tank holds 22 lb (10 kg) of instruments.
• 15-gal (57 L) tank holds 22 lb (10 kg) of instruments.
• 20-gal (76 L) tank holds 35 lb (16 kg) of instruments.
Automatic Pushbutton Ultrasonic Drain with drain indicator
light and built-in counter to provide wash cycle number
tracking, alerts operator when ultrasonic tank needs drained.
Adjustable cycle drain count preprogrammed to 12 cycles.
Toe Touch Control lowers the instruments into the cleaning
solution and starts the processing cycle. When cycle is
complete, the lid automatically opens and instrument tray is
raised to unload position.
Safety Interlocks and Controls are provided to ensure
complete operator safety.
Front and Side Service Access Panels are supplied on the
unit for ease of any maintenance procedures. Access doors
are scratch and dent resistant and sound dampened. Front
doors have automatic power cut-off feature to protect
components when doors open.
PREVENTIVE MAINTENANCE
well-equipped, factory-trained technicians to provide these
services, as well as on-site installation, training and expert
repair services. Contact STERIS for details.
NOTES
1. Pipe sizes shown indicate terminal outlets only. Building
service lines (not provided by STERIS) must supply the
specified pressures and flow rates.
2. STERIS recommends that a dedicated, grounded electrical
circuit be provided for each unit. Extension cord and plug
use is not recommended.
3. Approximate net weight:
» Model CR101: 425 lb (193 kg).
» Model CR115: 425 lb (193 kg).
» Model CR12 : 500 lb (227 kg).
4. Sonic Cleaner must be mounted on a hard, level surface.
5. Caviwave Sonic Cleaner is manufactured for STERIS by
Crest Ultrasonics.
UTILITY REQUIREMENTS
Refer to equipment drawing for installation details and
specifications.
Electrical
208 or 240 V, 60 Hz, 1-Phase (4 Wire) or 3-Phase (5 Wires
including Neutral and Ground)
Drain
1-1/4" NPT CPVC Male Slip (trap and waste lines are not
provided by STERIS).
Hot Water
1/2" NPT CPVC Male Slip (15-50 psi); 100-120°F (38-49°C).
Customers are encouraged to contact STERIS concerning our
annual maintenance program. Under the terms of the program,
preventive maintenance, adjustments and replacement of worn
parts are provided on a scheduled basis to help ensure optimal
equipment performance and help minimize untimely or costly
schedule interruptions. STERIS maintains a worldwide staff of
TECHNICAL DATA
POWER
REQUIREMENTS
(Amps)† SONIC
MODEL
POWER
(Watts)
1 PH 3 PH 1 PH
CR101 13.2 9.9 12.9 750 571 lb
CR115 13.2 9.9 12.9 1000 699 lb
CR12 19.6 16.3 19.8 1000 815 lb
WEIGHT
The base language of this document is ENGLISH. Any
translations must be made from the base language
document.
CUSTOMER IS RESPONSIBLE FOR COMPLIANCE WITH
APPLICABLE LOCAL AND NATIONAL CODES AND
SHIP
(259 kg)
(317 kg)
(370 kg)
TAN K
FLUID
CAPACITY
11 gal
(42 L)
15 gal
(57 L)
20 gal
(76 L)
HEAT
LOSS
(BTU)
210
225
275
L x W x D*208 V 240 V
20 x 10-1/2 x 3-1/2"
(508 x 267 x 89 mm)
20 x 13-1/2 x 3-1/2"
(508 x 343 x 89 mm)
24 x 13-1/2 x 3-1/2"
(610 x 343 x 89 mm)
TRAY
SIZE
*Maximum tray size
†Amp value is actual current drain. Disconnect should be at least 25% more.
See Technical Data.
2