STERIS Amsco Service Manual

OPERATING INSTRUCTIONS
Amsco® Century
®
Gravity and Prevacuum Sterilizers
(11/13/06) P129367-408
© 2006– STERIS Corporation. All rights reserved. Printed in U.S.A
Table of Contents Operating Instructions 129367-408
A WORD FROM STERIS CORPORATION
This manual contains important information on proper use of this sterilizer. All operators and department supervisors are urged to carefully review and become familiar with the warnings, cautions and instructions contained herein.
This sterilizer is specifically designed to process goods using only the cycles as specified in this manual. If there is any doubt about a specific material or product, contact the manufacturer of the product for the recommended sterilization technique.
STERIS carries a complete line of accessories for this unit to simplify, organize and assure sterility of the sterilization process. Instrument trays, pouches and biological/chemical monitoring systems are all available to fulfill your facility’s processing needs. A STERIS representative will gladly review these with you.
Advisory
Service
Information
Indications for
Use
A summary of the servicing this equipment can be found in Section 1 of this manual. Do not operate or service the equipment until you have become familiar with this information.
Any alteration of the sterilizer not authorized or performed by STERIS which could affect its operation will void the warranty, could adversely affect sterilization efficacy, could violate federal, state and local regulations and jeopardize your insurance coverage.
A thorough preventive maintenance program is essential to safe and proper sterilizer operation. You are encouraged to contact STERIS concerning our Preventive Maintenance Agreement. Under terms of this agreement, preven­tive maintenance, adjustments, and replacement of worn parts are done on a scheduled basis to assure equipment performance at peak capability and to help avoid untimely or costly interruptions. STERIS maintains a nationwide staff of well-equipped, factory-trained technicians to provide this service, as well as expert repair services. Contact STERIS for details.
The Amsco® Century® Steam Sterilizer is designed for efficient, sterilization of non-porous and porous, heat and moisture-stable materials used in healthcare facilities. The Century Steam Sterilizer is available in the following configura­tions:
Safety Precautions
to be observed when operating and
16" x 16" x 26" Single Door Gravity 16" x 16" x 26" Double Door Gravity 20" x 20" x 38" Single Door Gravity 20" x 20" x 38" Double Door Gravity
Table of Contents Operating Instructions 129367-408
16" x 16" x 26" Single Door Prevacuum 16" x 16" x 26" Double Door Prevacuum 20" x 20" x 38" Single Door Prevacuum 20" x 20" x 38" Double Door Prevacuum
i
The Amsco Century Steam Sterilizer is equipped with the following factory­programmed sterilization cycles and cycle values:
PREVACUUM CONFIGURATION
CYCLES RECOMMENDED STERILIZE STERILIZE DRY TIME
LOADS TEMP. TIME
FLASH Unwrapped 270°F 3.0 minutes 1.0 minute
Instrument tray with (132°C) a single instrument.
FLASH Unwrapped instrument 270°F 10 minutes 1.0 minute
tray with non-porous (132°C) multiple instruments, maximum weight 17lbs.
EXPRESS Single wrapped 270°F 4.0 minutes 3.0 minutes
instrument tray with (132°C) a single instrument. Non-porous goods only.
PREVAC Up to two double 270°F 4.0 minutes 20 minutes
1
wrapped instrument (132°C) trays, maximum weight 17 lbs. Up to six fabric packs.
PREVAC Up to two double 275°F 3.0 minutes 16 minutes
wrapped instrument (135°C) trays, maximum weight 17 lbs.
1
Five minute Dry Time can be used for processing a single fabric pack.
GRAVITY CONFIGURATION
CYCLES RECOMMENDED STERILIZE STERILIZE DRY TIME
LOADS TEMP. TIME
FLASH Unwrapped 270°F 3.0 minutes 1.0 minute
Instrument tray with (132°C) a single instrument.
FLASH Unwrapped instrument 270°F 10 minutes 1.0 minute
tray with non-porous (132°C) multiple instruments, maximum weight 17lbs.
GRAVITY Up to two double wrapped 270°F 15 minutes 30 minutes
trays, maximum (132°C) weight 17 lbs.
GRAVITY Up to six Fabric Packs. 250°F 30 minutes215 minutes
(121°C)
2
A 270°F (132°C) cycle adjusted to 25 minute Sterilize Time can be used for
processing fabric packs.
ii
129367-408 Operating Instructions Table of Contents

TABLE OF CONTENTS

Section
Title
A WORD FROM STERIS CORPORATION ...................................... i
1 SUMMARY OF SAFETY PRECAUTIONS ............................ 1-1
2 STERILIZATION TECHNIQUES ......................................... 2-1
2.1 General .............................................................................................................. 2-1
2.2 Special Information Regarding the Express Cycle .............................................. 2-1
2.2.1 Preparing Instruments for Express Cycle Sterilization .................................. 2-2
2.2.2 Unloading ..................................................................................................... 2-2
2.3 Special Information Regarding The Flash Cycle ................................................. 2-3
2.4 Control Measures For Verifying Sterilization Process ......................................... 2-4
2.4.1 Biological Monitors ....................................................................................... 2-4
2.4.2 Testing for Prevacuum Efficiency ................................................................ 2-4
2.5 Dart (Bowie-Dick) Test ....................................................................................... 2-5
2.6 Vacuum Leak Test ............................................................................................. 2-5
2.7 Sterilization Process Recommendations .......................................................... 2-5
3 COMPONENT IDENTIFICATION ........................................ 3-1
3.1 General .............................................................................................................. 3-2
3.2 Main Sterilizer and Cycle Controls ...................................................................... 3-2
3.3 Control Displays ................................................................................................. 3-3
3.4 Alarm Displays ................................................................................................... 3-4
3.5 Operating End Control Panel .............................................................................. 3-5
3.6 Cycle Selection Touch Screen Pads .................................................................. 3-6
3.6.1 Values Touch Screen Pads .......................................................................... 3-6
3.6.2 Abort Touch Screen Pad .............................................................................. 3-7
3.7 Printer ................................................................................................................ 3-7
3.8 Printouts ............................................................................................................. 3-8
3.9 Power Door Operation ........................................................................................ 3-9
3.10 Emergency Door Opening Procedure .............................................................. 3-10
3.11 Optional Electric Steam Generator ................................................................. 3-11
Page
4 STERILIZER OPERATION ................................................. 4-1
4.1 Before Operating the Sterilizer............................................................................ 4-1
4.2 Preparing Loads for Sterilization Cycles ............................................................ 4-2
4.3 Guidelines for Placement of Various Loads ........................................................ 4-3
4.4 Unloading the Sterilizer ....................................................................................... 4-4
4.5 Loading Car Instructions: Loading ...................................................................... 4-5
4.6 Loading Car Instructions: Unloading ................................................................... 4-6
4.7 Loading/Unloading Sterilizer Equipped With Rack and Shelves .......................... 4-6
4.8 Sterilizer (Factory) Cycle Settings ...................................................................... 4-7
4.9 270°F Flash Cycle .............................................................................................. 4-8
4.10 Gravity Cycle ................................................................................................. 4-11
4.11 270°F Express Cycle ...................................................................................... 4-14
4.12 270°F Prevac Cycle........................................................................................ 4-17
4.13 275°F Prevac Cycle........................................................................................ 4-20
4.14 DART (Bowie-Dick) Test (Prevacuum Sterilizer only) .................................... 4-23
4.15 Vacuum Leak Test (Prevacuum Sterilizer only) .............................................. 4-25
Section 4 Contents Continued On Next Page
Table of Contents Operating Instructions 129367-408
iii
TABLE OF CONTENTS (Cont'd)
Section
4.16 Optional Liquid Cycle ...................................................................................... 4-27
4.17 Sterilization Techniques for Optional Liquid Cycle .......................................... 4-30
4.18 Recommendations for Sterilizing Liquids ........................................................ 4-31
4.19 Aborting Cycles .............................................................................................. 4-32
4.20 Cycle Graphs ................................................................................................. 4-33
Title
5 STERILIZER CYCLES AND CYCLE VALUES ...................... 5-1
5.1 Sterilizer (Factory) Cycle Settings ...................................................................... 5-2
5.2 Change Values ................................................................................................... 5-3
5.3 Change Cycle Values ......................................................................................... 5-4
5.3.1 Overview ...................................................................................................... 5-4
5.3.2 Step by Step ................................................................................................ 5-4
5.4 Adjusting Sterilize Time And Dry Time ............................................................... 5-7
5.5 Change Time and Date ....................................................................................... 5-8
5.6 Change Machine Setup .................................................................................... 5-10
5.7 Access Code .................................................................................................... 5-11
5.8 Lockout ............................................................................................................ 5-13
5.9 Utilities Control ................................................................................................. 5-14
5.10 Language ........................................................................................................ 5-17
5.11 Machine Number ............................................................................................ 5-18
5.12 Time Format ................................................................................................... 5-19
5.13 Print Format ................................................................................................... 5-20
5.14 Audible Signals .............................................................................................. 5-21
5.15 Units ............................................................................................................... 5-22
5.16 Date Format ................................................................................................... 5-22
5.17 Duplicate Print ................................................................................................ 5-23
5.18 Leaving Change Values .................................................................................. 5-23
Page
LIST OF TABLES
Table
2-1 Cycle Availability ...................................................................................................................... 2-1
2-2 Item Processing Guidelines For Express Cycle ........................................................................ 2-2
3-1 Required Feed Water Quality for Carbon Steel Steam Generators .......................................... 3-12
4-1 Factory-Set Cycles and Cycle Values ....................................................................................... 4-7
4-2 Liquid Cycle Parameters ........................................................................................................ 4-30
5-1 Cycle Availability ...................................................................................................................... 5-1
5-2 Factory-Set Cycles and Cycle Values ....................................................................................... 5-2
5-3 Change Machine Setup .......................................................................................................... 5-10
iv
129367-408 Operating Instructions Table of Contents
Title
Page
!
!
!
!
!
!
!
!
!
!
!

SUMMARY OF SAFETY PRECAUTIONS 1

Following is a list of the indicate the potential for danger to personnel, and CAUTIONS indicate the potential for damage to equipment. These precautions are repeated (in whole or in part), where applicable, throughout the manual. This is a listing of all
Precautions
appearing in the manual. Carefully read them before proceeding to use or service the unit.
Safety Precautions
which must be observed when operating this equipment. WARNINGS
Safety
WARNING - BURN HAZARD:
When sterilizing liquids, to prevent personal injury or property damage resulting from bursting bottles and hot fluid, you must observe the following procedures:
Use Liquid cycle only; no other cycle is safe for processing liquids.
Use only vented closures; do not use screw caps or rubber stoppers with crimped seals.
Use only Type I borosilicate glass bottles; do not use ordinary glass bottles or any container not designed for
sterilization.
Do not allow hot bottles to be jolted; this can cause hot-bottle explosions. Do not move bottles if any boiling or bubbling is present.
It is inappropriate for a healthcare facility to sterilize liquids for direct patient contact.
Sterilizer, rack/shelves, and loading car will be hot after cycle is run. Always wear protective gloves and
apron when removing a processed load. Protective gloves and apron must be worn when reloading sterilizer following the previous operation.
Before daily flushing of the generator, generator must be at zero psig and cooled to room temperature.
Do not attempt to open the sterilizer door if a WATER IN CHAMBER ALARM condition exists. Call a qualified
service technician before attempting to use sterilizer further.
Sterilizer operator may be severely burned by scalding water if the water level control malfunctions. The steam generator level control may malfunction if the supply water exceeds 26,000 ohms/cm (38.5 conductivity min.). Do not connect to treated water (e.g., distilled, reverse osmosis, deionized) unless water resistivity is determined to be acceptable. If water exceeds 26,000 ohms/cm, contact STERIS Engineering Service for information concerning modifications required to the generator control system.
After manual exhaust, steam may remain inside the chamber. Always wear protective gloves, apron and a face shield when following emergency procedure to unload sterilizer. Stay as far back from the chamber opening as possible when opening the door.
Allow sterilizer to cool to room temperature before performing any cleaning or maintenance procedures.
Failure to shut off the steam supply when cleaning or replacing strainers can result in serious injury. Jacket
pressure must be 0 psig before beginning work on the steam trap.
Proper testing of the safety valve requires the valve to be operated under pressure. Exhaust from the safety valve is hot and can cause burns. Proper safety attire (gloves, eye protection, insulated overall) as designated by OSHA, is required. Testing is to be performed by qualified service personnel only.
Steam may be released from the chamber when door is opened. Step back from the sterilizer each time the door is opened to minimize contact with steam vapor.
1-1
Summary of Warnings and Cautions Operating Instructions 129367-408
WARNING – ELECTRIC SHOCK AND BURN HAZARD:
!
!
!
!
!
!
!
!
!
!
!
!
!
Disconnect all utilities to sterilizer before servicing. Do not service the sterilizer unless all utilities have been properly locked out. Always follow OSHA Lockout-Tagout and electrical safety-related work practice standards. (See CFR 1910.147 and .331 through .335.)
WARNING – PERSONAL INJURY HAZARD:
Avoid personal injury from bursting bottles. Liquid sterilization cycle must only be used for liquids in borosilicate (Pyrex) flasks with vented closures.
Door must be locked and the key retained prior to entering chamber for servicing. Always follow appropriate Lockout-Tagout and electrical safety-related work practice standards. Emergency stop switch can be depressed and key retained on sliding door units.
WARNING - EXPLOSION HAZARD:
This sterilizer is not designed to process flammable compounds.
WARNING - SLIPPING HAZARD:
To prevent falls keep floors dry by immediately wiping up any spilled liquids or condensation in sterilizer loading or unloading area.
WARNING – PERSONAL INJURY AND/OR EQUIPMENT DAMAGE HAZARD:
Regularly scheduled preventive maintenance is required for safe and reliable operation of this equipment. Contact your STERIS service representative to schedule preventive maintenance.
When closing the chamber door, keep hands and arms out of the door opening and ensure opening is clear of obstructions.
Repairs and adjustments to this equipment must be made only by fully qualified service personnel. Maintenance performed by inexperienced, unqualified persons or installation of unauthorized parts could cause personal injury or result in costly equipment damage.
WARNING - STERILITY ASSURANCE HAZARD:
Load sterility may be compromised if the biological indicator or air leak test indicates a potential problem. If these indicators show a potential problem, refer the situation to a qualified service technician before using the sterilizer further.
According to AAMI standards, a measured leak rate greater than 1.0 mm Hg/minute indicates a problem with the sterilizer. Refer the situation to a qualified service technician before using the sterilizer further.
The Express cycle is only intended for use with a single instrument in a single wrapped instrument tray.
The Express cycle is not intended for processing porous items (except the tray wrapper).
The Flash cycle is not intended for processing porous items.
1-2
129367-408 Operating Instructions Summary of Warnings and Cautions
CAUTION - POSSIBLE EQUIPMENT DAMAGE:
!
!
!
!
!
!
!
!
!
!
!
!
!
!
!
!
!
Gasket must be fully retracted prior to operating sterilizer door.
Failure to flush generator daily could result in malfunction of the generator. Warranty on the generator will be voided unless flushed daily.
Before flushing generator, ensure generator drain valve is fully open to prevent generator heaters from turning on during flush phase.
If zero dry time is selected, sterilizer automatically initiates a vapor removal phase in place of drying. This phase can still draw a vacuum to 5.0 inHg. Consult device manufacturer's recommendations to ensure devices being processed can withstand this depth of vacuum.
Lifting the chamber float switch when cleaning the chamber may cause the sterilizer control to initiate a “Chamber Flooded” alarm. If this alarm condition occurs, the operator must turn the control power OFF then ON to clear the alarm. The control power switch is located in the mechanical area at the side of the sterilizer. Placing the sterilizer in standby does not clear this alarm.
Never use a wire brush, abrasives, or steel wool on door and chamber assembly. Do not use cleaners containing chloride on stainless-steel surfaces. Chloride-based cleaners will deteriorate stainless steel, eventually leading to failure of the vessel.
Do not use cleaners containing chlorides on loading cars. Chloride-based cleaners will deteriorate the loading car metal.
Sterilization of chloride-containing solutions (e.g., saline) can cause chamber corrosion and is not recommended by the manufacturer. If, however, chloride-containing solutions must be processed, clean the chamber after each use.
Allow thermostatic traps to cool down to room temperature before removing cover. Since there is nothing to limit expansion, the bellows may rupture or fatigue if trap is opened while hot.
Actuation at less than 75% of rated pressure can allow debris to contaminate the seat and cause the safety valve to leak. A leaking safety valve must be replaced.
Insufficient service clearance will make repairs more difficult and time-consuming.
Piping sized too small may cause water hammer, resulting in damage to the sterilizer.
After installation, it is mandatory to brace piping at the drain funnel so that it will not move vertically.
Make sure door opening is clear of any obstruction before closing the door(s).
Do not attempt to open sterilizer door during manual operation unless chamber is at 0 psig.
During manual operation, gasket must be fully retracted prior to operating sterilizer door.
Immediately wipe up saline solution spills on loading car, to prevent damage to stainless steel.
1-3
Summary of Warnings and Cautions Operating Instructions 129367-408

STERILIZATION TECHNIQUES 2

The information in this section is intended as a general guide to steam

2.1 General

WARNING – BURN HAZ­ARD: It is inappropriate for a health care facility to ster­ilize liquids for direct pa­tient contact.
sterilization techniques. Also recommended is reference to the standards of the Association for the Advancement of Medical Instrumentation (AAMI ST-46,
Steam Sterilization and Sterility Assurance, 3rd Edition
).
Prior to sterilization, all materials and articles must be thoroughly cleaned.
After sterilization, goods should be stored in conditions that will not compro-
mise the barrier quality of their wrapping materials.
Always carefully segregate items which have been sterilized using conven­tional cycles from those sterilized using Flash or Express cycles.
IMPORTANT: The sterilization cycles listed in Table 2-1 have been validated using techniques documented in AAMI ST-8 and AAMI ST-37. If different cycle parameters (sterilize time and dry time only) other than those in Table 2-1 are required, it is the responsibility of the health care facility to validate the cycle. Reference AAMI guidelines/standards for a guide to validating sterilization cycles and to ensure that proper sterility assurance level (SAL) as well as moisture retention acceptance criteria are met.
NOTE: Contact your customer service representative for information on a wide range of education/training programs designed to meet the educational needs of health care industries.
The E xpress cycle
only
on prevacuum sterilizers, and is not provided on gravity sterilizers.
is an abbreviated prevacuum cycle that is intended for use
Table 2-1. Cycle Availability
Cycle Sterilize Sterilize Dry Prevac Gravity Type Load Temperature Time Time Default Optional Default Optional
Gravity* Full Load Fabric Packs 270°F 25 min 15 min X X
Gravity* Full Load Fabric Packs 250°F 30 min 15 min X X
Gravity* Full Load Instrument Trays 270°F 15 min 30 min X X
Gravity* Full Load Instrument Trays 250°F 30 min 30 min X X
Liquid* Three 1000ml Bottles 250°F 45 min N/A X X
Prevac* Single Fabric Pack 270°F 4 min 5 min X N/A N/A
Prevac* Full Load Instrument Trays 270°F 4 min 20 min X N/A N/A
Prevac* Full Load Instrument Trays 275°F 3 min 16 min X N/A N/A
Flash** Unwrapped, Non-porous
Express** Single-wrapped
Flash Unwrapped, Non-porous
DART* Bowie-Dick Test Pack 270°F 3 1/2 min 1 min X N/A N/A
Leak* Test None N/A N/A N/A X N/A N/A
Instrument Tray 270°F 3 min 1 min X X
Instrument Tray 270°F 4 min 3 min X N/A N/A
Instrument Tray 270°F 10 min 1 min X X
*Cycles qualified to AAMI ST-8
**Cycles qualified to AAMI ST-37
2-1
Sterilization Techniques Operating Instructions 129367-408
2.2 Special
Information
Regarding the
Express Cycle
Appropriate parameters for sterilization are preset by STERIS. It is designed to permit sterilization using a single instrument in a single or textile, but not a peel pouch) on the instrument tray.
Rationale: The single wrapper serves to confine and contain the sterilized item from environmental contaminants that may be encountered enroute from the sterilizer to the point of use. A single wrapped item sterilized with the Express cycle does not have a shelf life.
wrapper (non-woven
WARNING – STERILITY ASSURANCE HAZARD: The Express cycle is only intended for use with a single instrument in a single wrapped instru­ment tray.
2.2.1 Preparing Instruments for
Express Cycle
Sterilization
WARNING – STERILITY ASSURANCE HAZARD: The Express cycle is not intended for processing porous items (except the tray wrapper).
The Express cycle is useful in providing quick turnaround of an instrument using the wrapped technique for transport from the sterilizer to the point of use.
Instrument trays processed using the Express cycle are intended for immediate use.
Decontaminate and then prepare a surgical instrument (open and/or disas­sembled), and place in a perforated or mesh bottom instrument tray. Only non­porous items (except for the wrapper) should be processed using this cycle.
Items with lumens (e.g., needles for injection and diagnostics or metal suction cannulae) are considered porous items, and must not be processed using this cycle.
The Express cycle has fewer prevacuum pulses so air removal and subsequent steam contact within lumens may be difficult to achieve.
Table 2-2. Item Processing Guidelines for Express Cycle
Examples of non-porous items that can be processed using the Express cycle:
Forceps,
Needle Holders,
Scissors and other routine metal instruments.
Examples of porous items that CANNOT be processed using the Express cycle:
Towels,
Rubber or Plastic Items,
Items with Lumens,
Items with sliding parts that prevent sterilant contact with surfaces.
2-2
129367-408 Operating Instructions Sterilization Techniques

2.2.2 Unloading Use instrument processed in an Express cycle immediately. Do not store for

later use.
When opening and unloading the sterilizer at the conclusion of the Express cycle always use aseptic techniques. Sterile towels and sterile gloves are used to remove the tray from the sterilizer. The still-wrapped tray is transported to the point of use, being careful not to contact any unsterile surfaces. The tray is placed on a previously draped sterile surface field and the wrapper opened by the circulating nurse. The draping material used must be impervious to liquids and able to withstand contact with hot surfaces.
To shorten the sterilization process, drying time has been reduced in the Express cycle. Thus, the single wrapper will appear dry at the conclusion of the cycle, but the content of the package will, in all probability, still be wet with condensation formed during sterilization. Water will penetrate the wrapper, especially on the under side of the tray because of gravity. The wrapper only serves the function of protecting the tray and its content from particulate matter encountered on the way from the sterilizer to the point of use.
Any moisture present can strike through the wrapper, so procedures should be developed to avoid contamination by contact with unsterile surfaces. For example:
Wear sterile gloves and use sterile towels as “pot holders” when removing the goods from the sterilizer.
Never place the wrapped tray on an unsterile surface.
2.3 Special
Information
Regarding The
Flash Cycle
WARNING – STERILITY ASSURANCE HAZARD: The Flash cycle is not in­tended for processing po­rous items.
The Flash Cycle Amsco® Century® prevacuum and gravity configurations. The Flash cycle has been designed to sterilize an unwrapped item at sterilization parameters preset by STERIS. There is no storage or shelf life of flash sterilized items.
Rationale: The Flash Cycle is intended for sterilizing an unwrapped item intended for immediate use (e.g. a dropped instrument). In any method of sterilization, it is important to adhere to good processing practices. This is particularly important in flash sterilization.
Preparing Instruments For Flash Cycle Sterilization
As prescribed in AAMI ST-37, prior to flash sterilization of a dropped instrument, ensure the item is free of soil by the appropriate decontamination procedure. The flash sterilized item then must be transferred immediately, using aseptic technique, from the sterilizer to the actual point of use, usually the sterile field in an ongoing sterile procedure.
Use items processed in a Flash Cycle immediately. Do not store the processed items for later use.
is an abbreviated gravity cycle
and is provided on both the
2-3
Sterilization Techniques Operating Instructions 129367-408
2.4 Control
Measures For
Verifying
Sterilization
Process

2.4.1 Biological Monitors

As part of the operator's verification of the sterilization process, biological indicators may be used to demonstrate that sterilization conditions have been met.
NOTE: Contact STERIS for information on specific biological indicators recommended for use with this sterilizer.
A live spore test utilizing form of biological monitoring. This type of product utilizes controlled populations of a controlled resistance, so that survival time and kill time can be demonstrated.
To verify the process, insert the biological indicator in a test pack and place pack on the bottom shelf. Run test pack through a typical cycle. On completion, forward test pack and monitor to appropriate personnel for evaluation. Refer to AAMI guidelines to conduct routine biological monitoring.
G. stearothermophilus
is the most reliable
2.4.2 Testing for
Prevacuum Efficiency
WARNING – STERILITY ASSURANCE HAZARD: Load sterility may be com­promised if the biological indicator or air leak test indicates a potential prob­lem. If these indicators show a potential problem, refer the situation to a qualified service techni­cian before using the ster­ilizer further.
Run a Dart® (Bowie-Dick test) cycle daily before processing any loads. The first prevacuum cycle of each day should be used to test the adequacy of air removal from the chamber and load, so that steam can penetrate the load. It is not a test for adequate exposure to heat in terms of time-at­temperature.
Tests such as the Bowie-Dick or the Dart® (Daily Air Removal Test)* are designed to document the removal of residual air from a sample challenge load.
In the case of these tests, following exposure in a prevacuum sterilizing cycle, the pack is opened, the indicator examined and conclusions are drawn as to the pattern of residual air, if any, that remained in the pack during the sterilizing cycle. Any indication of a malfunction must be reported to the supervisor. Sterilizer should not be used during this time.
* Available from STERIS.
.
2-4
129367-408 Operating Instructions Sterilization Techniques
2.5 Dart (Bowie­Dick) Test

2.6 Vacuum Leak Test

Conduct a residual air test (e.g., Bowie-Dick test) at the beginning of each day according to the AAMI standard ST-46. STERIS can provide a product called Dart® (Daily Air Removal Test), designed to be as sensitive as the standard AAMI Bowie-Dick test pack in detecting air leaks. Refer to instructions for running Dart test given in Section 4 of this manual. If a Dart is not available, construct Bowie-Dick test package in accordance with instructions given in AAMI standard ST-8.
Run the Vacuum Leak test cycle daily or weekly. This test measures the integrity of the sealed pressure vessel and associated piping to assure air is not being admitted to the sterilizer during the vacuum draw downs. Refer to appropriate cycle description in Section 4 of this manual.
WARNING – STERILITY ASSURANCE HAZARD: According to AAMI stan­dards, a measured leak rate greater than 1.0 mm Hg/minute indicates a problem with the sterilizer. Refer the situation to a qualified service techni­cian before using the ster­ilizer further.
2.7 Sterilization Process
Recommendations
After running a vacuum leak test, a value or leak rate will be printed on the printer tape. This value will help define a trend over a period of time if the integrity of the system begins to deteriorate (i.e., allowing air to enter the system). By running a vacuum leak test cycle daily or weekly, the operator or maintenance personnel can always monitor the air tightness of the system and make repairs or adjustments when necessary.
NOTE: A leak rate of greater than 1.0 mmHg per minute indicates a problem with the sterilizer that must be addressed.
Saturated steam is a well controlled, reliable method for processing items which can withstand the temperatures and pressures associated with steam steriliza­tion. The requirements for achieving reproducible results are well known by many users, but are not always understood by all users.
The condition most likely to result in sterilization problems is a failure to remove all of the air from the items being processed. For example, placing an empty beaker or bowl in an upright position in a gravity displacement sterilizer may result in the object not being sterilized, or may require exceptionally long sterilization times. This problem is because air has almost twice the density as does saturated steam under the same conditions. Thus, the air sits in the bottom of the container, and the steam forms a stable layer over the air. This effect is similar to oil forming a stable layer over water. As long as there is no mechanism for actively mixing the two, the bottom of the container will only see dry heat, which is not an effective sterilization method at the temperatures typically used in steam processes.
There are two traditional methods for enhancing the sterilization of solid bottom containers in gravity displacement cycles. These are:
• Place 1.0 mL of water for each liter of volume in the bottom of each container. The expansion of the water into steam as the product is heated will force most of the air out of the object, thus allowing steam to reach all surfaces and effect sterilization.
• The better, more reliable method is to orient all objects in a manner which would allow water to flow out. When the steam enters the chamber, it will tend to layer over the air. However, the object is now oriented so the air can flow out. As the air flows out of the container, it will be replaced by the steam. The steam can now reach all surfaces and effect sterilization.
2-5
Sterilization Techniques Operating Instructions 129367-408

COMPONENT IDENTIFICATION 3

Steam Supply
Valve
Touch Screen
Access Panel
(Front panel of
unit is hinged
for service
access)
Century Sterilizer
Generator Water Supply Valve
Foot Pedal
Century Control Panel
Century Control Panel w/Door Open
Control Access
Door
Printer
Touch Screen
Sight Glass
JACKET
CHAMBER
Optional Electric Steam Generator
Generator Drain Valve
Main Power Disconnect Switch. This should be left on at all times.
Figure 3-1. Amsco® Century® Sterilizer Components
Component Identification Operating Instructions 129367-408
3-1

3.1 General

Use this manual to become familiar with control locations and functions before operating the sterilizer (refer to Fig. 3-1). The controls for this sterilizer are contained within the control touch screen. Control touch pads appear on the screen as needed during each operation. Available controls change as the sterilizer steps through different operations.
3.2 Main Sterilizer and Cycle
Controls
Open
Closed
JACKET
CHAMB ER
Steam Supply Valve
••
Main Power Disconnect Switch
••
(refer to Fig. 3-1) – Located behind the access door, on the main control box, this switch disconnects power to the control. Under normal operation, this switch is left in the open position at all times.
••
Steam Supply Valve – This is located behind the front access door, above
••
the chamber door. Refer to Figure 3-2. Ensure this is in the open position before trying to operate the sterilizer.
••
Water Supply Valve – This is located behind the front access door, below
••
the chamber door. Refer to Figure 3-2. Ensure this is in the open position before trying to operate the sterilizer.
••
Steam Generator Control – The optional integral electric steam generator
••
(if present) is located in the space below the sterilizer chamber. The generator is automatically turned on by the sterilizer’s control when the sterilizer is turned on. Refer to the instructions later in this section before operating a sterilizer equipped with an integral steam generator.
Sterilizer Status
STATUS ............ STANDBY 0
TIME .................. 00:00 AM
Current Time & Date
DATE ................. 00-0 0-00
Open
Touch Screen Pad—Press to Turn the Sterilizer ON
ON
Standby Screen
STATUS ............ DOOR OPEN 1
Closed
Water Supply Valve
TEMP ................. 000F
PRESS ............... 00 PSIG
12 34
FLASH FLASH EXPRESS PREVAC
270F 270F 270F 270F S=03M S=10M S=04M S=04M D=01M D=01M D=03M D=20M
00:00 AM 00-00-00
PAPER MORE
FEED CYCLES
MENU STANDBY
ON Screen
Figure 3-2. Main Sterilizer and Cycle Controls
3-2
129367-408 Operating Instructions Component Identification

3.3 Control Displays

» Generator Water Supply Valve (refer to Fig. 3-1) – This valve should
always be open. Ensure it is open before attempting to operate the sterilizer.
» Generator Drain Valve (refer to Fig. 3-1) – This should be opened once
a day, while generator is at room temperature, to flush the generator of residual solids (present in most water) that may have accumulated in the generator’s boiler. After the boiler has been flushed ensure the drain valve is then closed. Refer to instructions later in this section.
••
Sterilizer Control Touch Pad
••
––
– This is visible on the control touch screen
––
whenever the sterilizer is in Standby mode. Refer to Figure 3-2.
NOTE: Screen touch pads respond to very slight pressure, and only need to be pressed lightly.
The sterilizer enters operating mode when the ON touch pad is pressed. This touch pad switches the sterilizer control between Standby and Ready conditions (Standby mode is usually used at night when the sterilizer is not being operated—steam is turned off and machine cools, saving energy).
A screen reference number appears in the upper right corner of each display. Numbers are used for reference only, and do not relate to the operating sequence of the screen.
Control displays can be divided into two categories, those occurring when the sterilizer is “out-of-cycle” and those occurring when the sterilizer is "in-cycle".
Typical out-of-cycle and in-cycle displays are shown in Figure 3-3.
Out-of-cycle displays are used to start cycles, or set up and adjust sterilizer operation. With the exception of the cycle starting displays, most out-of-cycle displays will only be used occasionally. Detailed instructions for adjusting the sterilizer operating parameters are in Section 5 of this manual.
Generally, when the sterilizer is in-cycle, displays appear automatically and, unless an abnormal conditions occurs, require no special attention or instructions. In-cycle displays tell the operator at what temperature and pressure the sterilizer chamber is operating, show the current cycle phase and indicate when the processing cycle is complete. For more details about operating cycles, refer to Section 4 of this manual.
STATUS ............ DOOR OPEN 1
TEMP ................. 000F
PRESS ............... 00 PSIG
12 34
FLASH FLASH EXPRESS PREVAC
270F 270F 270F 270F S=03M S=10M S=04M S=04M D=01M D=01M D=03M D=20M
00:00 AM 00-00-00
PAPER MORE
FEED CYCLES
MENU STANDBY
Out-of-Cycle In-Cycle
STATUS ............ STERILIZE 00:00 4
TEMP ................ 000 F
PRESS ............. 00 PSIG
CYCLE .............. 1, FLASH, 270F, S=03M, D=01M
PROJECTED CYCLE COMPLETION TIME:
0:00
MINUTES SECONDS
PAPER STATUS
FEED PRINT
ABORT
Figure 3-3. Typical In and Out-Of-Cycle Display
3-3
Component Identification Operating Instructions 129367-408

3.4 Alarm Displays

Alarm displays tell operators and technicians when the sterilizer is experienc­ing an abnormal condition. Alarm conditions can be caused by failure of utility supplies or sterilizer components. Section 5 of ROUTINE MAINTENANCE
MANUAL (P129367-410), details the steps an operator can take to solve most
alarm conditions. Typical alarm displays are shown in Figure 3-4.
When an alarm occurs during cycle operation, a display appears on the screen, accompanied by an audible tone. This display indicates the problem as determined by control sensors, and lists a brief troubleshooting list. The operator should follow the instructions on the screen, if possible. If these instructions fail to clear the alarm, consult your departmental supervisor or a trained service technician before using the sterilizer further.
STATUS ............ ALARM! 205
CHAMBER: 70.0 F 00.0 psig
STERILIZER WILL:
AUTOMATICALLY TRY TO COMPLETE CYCLE
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CHECK STEAM SUPPLY VALVE
3. IF ALARM RECURS, CALL SERVICE
SILENCE
STATUS ............ SERVICE INFORMATION: 206
→→
CHAMBER DID NOT REACH STERILIZE
→→
CAUSES AND CORRECTION:
1. STEAM PRESSURE LESS THAN 50 PSIG
2. STEAM REGULATOR MALFUNCTION
2. SOLENOID VALVE MALFUNCTION
3. CONTROL OUT OF CALIBRATION
TOO LONG IN CHARGE
→→
IF CLOSED, OPEN VALVE
→→
ABORT
ALARM PRINT FEED HELP
TEMPERATURE WITHIN ALLOTTED TIME
→→
CHECK STEAM SUPPLY PIPING
→→
→→
REPAIR
→→
→→
REPAIR S09
→→
→→
REPAIR S02
→→
→→
RECALIBRATE (CONTACT QUALIFIED
→→
SERVICE PERSON)
ABORT
STATUS PAPER
TOO LONG IN CHARGE
PAPER
FEED
SERVICE
EXIT
Figure 3-4. Typical Alarm Displays
3-4
129367-408 Operating Instructions Component Identification

3.5 Operating End Control Panel

STATUS ............ STERILIZE 00:00 4
TEMP ................ 000 F
PRESS ............. 00 PSIG
CYCLE .............. 1, FLASH, 270F, S=03M, D=01M
A sterilizer equipped with two doors, will also be equipped with two control panels. The control panel at the loading door of the sterilizer is referred to as the "operating end control," the control panel located at the unloading door is referred to as the "non-operating end control."
A single-door sterilizer is equipped with an "operating end control," only.
NOTE: Except for the presence of the printer (which is only present at the operating end of the unit), control panels at both ends of the unit are similar and each can be used to start or abort the sterilizer.
The operating end control panel (see Fig. 3-5) is used to:
••
Select and start cycles.
••
••
Abort cycles.
••
••
Set cycles and cycle values.
••
••
Obtain status printouts (see “Printer” paragraph later in this section).
••
The operating end control includes a printer for cycle documentation.
Cycle status and control messages are shown on a 30 column x 40 line graphics display. Cycles can be started or aborted using the touch screen pads. Cycles and cycle values can be set using the Change Values procedure (accessible from the sterilizer MENU screen). If changing cycle values becomes neces­sary, refer to Section 5 of this manual.
PROJECTED CYCLE COMPLETION TIME:
PAPER STATUS
FEED PRINT
When in-cycle, the display shows a status of the sterilizer, status of the cycle, and large numbers indicate the approximate time remaining in the current cycle.
STATUS ............ MENU 2
TEMP ................. 128F
PRESS ............... 0 PSIG
CHANGE
TIME
DATE
PAPER
FEED
CHANGE CHANGE DUPLICATE
&
CYCLE MACHINE PRINT
VALUES SETUP
EXIT
Menu Screen
0:00
MINUTES SECONDS
ABORT
Typical Menu Screen
AMSCO Time 00:00:00
AMSCO Time 00:00:00 AMSCO Time 00:00:00
AMSCO Time 00:00:00 AMSCO Time 00:00:00 AMSCO Time 00:00:00
Cycle
Cycle
Printout
Printout
Touch
Touch
Screen
Screen
Figure 3-5. Operating End Control Panel
3-5
Component Identification Operating Instructions 129367-408
3.6 Cycle
Selection Touch
Screen Pads
Press to StartPress to Start
Press to Start
Press to StartPress to Start (Select) Cycles(Select) Cycles
(Select) Cycles
(Select) Cycles(Select) Cycles
Four cycle selection touch pads are shown on the screen in Figure 3-6. These pads display the basic parameters of the cycle (cycle name, sterilization exposure temperature, sterilization exposure time and dry time), additional cycles may be selected by pressing MORE CYCLES. Details on individual cycles are in Section 4.8.
Up to seven additional cycles can be programmed and displayed. It is the responsibility of the healthcare facility to validate the cycle. Reference AAMI for guidelines and standards for a guide to validating sterilization cycles and to ensure that proper sterility assurance level (SAL) as well as moisture retention acceptance criteria are met.
STATUS ............ DOOR OPEN 1
TEMP ................. 000F
PRESS ............... 00 PSIG
12 34
FLASH FLASH EXPRESS PREVAC
270F 270F 270F 270F S=03M S=10M S=04M S=04M D=01M D=01M D=03M D=20M
00:00 AM 00-00-00
PAPER MORE
FEED CYCLES
MENU STANDBY
Press to StartPress to Start
Press to Start
Press to StartPress to Start
(Select) Cycles(Select) Cycles
(Select) Cycles
(Select) Cycles(Select) Cycles
Press for More CyclesPress for More Cycles
Press for More Cycles
Press for More CyclesPress for More Cycles

3.6.1 Values Touch Screen Pads

Sterilization ExposureSterilization Exposure
Sterilization Exposure
Sterilization ExposureSterilization Exposure
TemperatureTemperature
Temperature
TemperatureTemperature
Cycle NameCycle Name
Cycle Name
Cycle NameCycle Name
Figure 3-6. Cycle Selection Touch Pads
These touch screen pads are accessed through the MENU screen by pressing CHANGE CYCLE VALUES. The values touch pads are used for changing the operating values used in cycles, changing the cycles displayed on the cycle selection menus, and for changing the operating settings of the sterilizer. Instructions for changing sterilizer cycle parameters are in Section 5 of this manual.
Sterilization ExposureSterilization Exposure
Sterilization Exposure
Sterilization ExposureSterilization Exposure
TimeTime
Time
TimeTime
Dry TimeDry Time
Dry Time
Dry TimeDry Time
SELECT ITEM TO CHANGE
1 TEMP STER DRY
FLASH 270 F 03 M 01 M
EXIT
11
Figure 3-7. Values Touch Screen Pads
3-6
129367-408 Operating Instructions Component Identification

3.6.2 Abort Touch Screen Pad

The ABORT touch screen pad is used to end a cycle before it finishes normally. A cycle only needs to be aborted if an abnormal condition or a control problem develops during the cycle. Pressing ABORT causes the sterilizer chamber to depressurize (if pressurized), or Air Break (if in vacuum), the door seal deactivates, the control prompts the operator to open the door, and the sterilizer returns to its normal out-of-cycle state. If an abnormal condition persists after fully aborting the cycle, contact your supervisor or a qualified service technician before trying to operate the sterilizer further.
STATUS ............ STERILIZE 00:00 4
TEMP ................ 000 F
PRESS ............. 00 PSIG
CYCLE .............. 1, FLASH, 270F, S=03M, D=01M
PROJECTED CYCLE COMPLETION TIME:
Figure 3-8. Cycle AbortTouch Screen Pad

3.7 Printer

PAPER PRINT
FEED STATUS
0:00
MINUTES SECONDS
ABORT
Abort Touch Screen Pad
Refer to Figure 3-1.
Printer records all cycle data on 2-1/4 inch wide single-ply paper. See ROUTINE MAINTENANCE MANUAL
(P129367-410)
for paper changing procedure.
Printer functions controlled by touch screen pads are as follows:
Paper Feed — Press to feed out paper from the roll stored inside the control. Accessible during all phases of operation, including alarm conditions. Press and hold for continuous feed.
Duplicate Print
— Press to obtain a complete duplicate printout of the last previously run cycle (when unit is not in cycle). This touch pad is only visible on the screen during Complete and Change Values menu. The Duplicate Print touch screen pad is not visible upon first power-up of the day.
Status Print — Press to obtain a printout of current cycle phase and conditions (when unit is in cycle). This touch pad is only visible during cycle operation.
Print Values — Press to obtain a printout of all currently set cycles and cycle values. Only accessible when the unit is not in cycle. This touch screen pad appears on Change Values menu only.
3-7
Component Identification Operating Instructions 129367-408

3.8 Printouts

Cycle Type
Cycle Start Time & Date
Total Cycle Count Operator I.D. Machine Number
Sterilize Temp. Control Overdrive Temp. Sterilize Time Dry Time
Refer to Figure 3-9.
The printout reports useful information about each cycle the sterilizer runs. This includes the load number, which is a unique identifying code. Each load number is printed in the following format: a two digit month (e.g., January = 01), a two digit day (e.g., second day of any given month = 02) and a two digit daily cycle count (e.g., third cycle of the day = 03). In our examples the complete load number would then be 010203.
During the cycle, status lines on the printouts show the time the line was printed, chamber temperature and the level of vacuum or pressure in the chamber. Each status line also begins with a letter code. This code indicates during which cycle phase the print line occurred, or what kind of event caused the print line to occur.
Refer below to see other features of the printout.
Additional Status Print Codes:
F = Alarm (Failure) L = Leak Test (Vacuum or Hold) D = Demand Print (Print Status)
Load Number
Sterilize Temp. Min./Max.
Time in Phase
Total Cycle Time
Review and verify critical cycle parameters were achieved during processing, then sign printout to indicate verification.
Conditioning
• Charge
• Vacuum Pulses
Sterilize
Exhaust & Dry
Complete
3-8
129367-408 Operating Instructions Component Identification
Figure 3-9. Typical Printout

3.9 Power Door Operation

The sterilizer door is operated at the foot pedal (refer to Fig. 3-10).
Pressing the foot pedal while the door is in the closed (up) position causes the door to open (lower).
Pressing the foot pedal while the door is in the open (lowered) position, causes the door to close (by raising).
NOTE: The foot pedal only provides control of the visible door. For double-door units, the foot pedal cannot be used to control the door at the opposite end of the sterilizer.
Figure 3-10. Foot Pedal
WARNING - PERSONAL IN­JURY HAZARD: When clos-
ing the chamber door keep hands and arms out of the door opening and ensure opening is clear of any obstructions.
Important:
Operation of Power Door if Motor Fails
If the power door cannot be operated using the foot pedal (such as during a power failure or if the drive mechanism is damaged), it is still possible to operate the door manually.
Keep the door closed when the unit is not in use.
Using hand pressure, pull up or push down on the handle to operate the door. Increased effort is required on the part of the operator to slide the door. Do not try to raise or lower door rapidly as fast operation may damage the door drive mechanism.
NOTE: If the power fails, and pressure remains in the sterilizer chamber, refer to Section 3.10, Emergency Door Opening Procedure.
3-9
Component Identification Operating Instructions 129367-408
3.10 Emergency Door Opening
Procedure
WARNING – BURN HAZ­ARD: Do not attempt to open sterilizer door if a water in chamber alarm condition exists.
This procedure should only be used when pressure remains in the sterilizer chamber, and the door cannot be opened normally because the sterilizer has lost either electrical or water utilities. This emergency door opening procedure can be used to retrieve a load in the chamber. This procedure requires pushing on the door cover to retract the door seal into the groove, then pushing the door downward manually.
Procedure:
1. Swing open the access panel of the sterilizer. Open the emergency exhaust hand valve until chamber exhausts to 0.0 psig. See Figure 3-11.
2. Using the door release tool provided, press on the upper left hand and right hand corners of the chamber door (see Fig. 3-11). The door should give inward slightly, indicating that the seal has been pushed into the groove. The door must “bottom out” against the metal end-frame of the sterilizer chamber.
3. Close the access panel of the sterilizer.
4. Press down on the door handle to lower the door. The door will slide stiffly, so be prepared to press down hard.
5. Once the door is open, do not use the sterilizer until the unit has been examined by a qualified service technician. Further use without attention may damage the sterilizer.
6. Close the emergency exhaust valve.
Figure 3-11. Emergency Door Opening Procedure
3-10
129367-408 Operating Instructions Component Identification
3.11 Optional
Electric Steam
Generator
CAUTION – POSSIBLE EQUIPMENT DAMAGE: Failure to flush generator on a daily basis could re­sult in generator malfunc­tion.
If a building steam source is not available, the sterilizer may be equipped with an electric steam generator. The generator automatically converts water to steam using electric heat. The steam created is then used to power the sterilizer.
Steam generators are highly susceptible to mineral scaling if the supplied water has any level of hardness. Refer to Table 3-1 for water quality requirements.
IMPORTANT: Regardless of the hardness level of supplied water, the genera­tor must be flushed every day before use to prevent mineral scaling or carryover of debris into the chamber.
ATTENTION: Warranty on this steam generator will be voided unless daily flushing procedures are performed.
Daily Generator Start Up Procedure
1. Press the ON button on the sterilizer touch screen (screen #0). Display advances to screen #72. Instructions on how to flush the generator are listed on screen #72.
Pressing STOP TIMER stops flush phase and resets timer to three minutes.
Pressing CANCEL bypasses flush phase and advances display to screen #1. Do not press CANCEL unless generator has already been flushed that day.
WARNING – BURN HAZ­ARD: Before daily flushing of the generator, genera­tor must be at zero psig and cooled to room tem­perature.
CAUTION – POSSIBLE EQUIPMENT DAMAGE: Before flushing generator, ensure generator drain valve is fully open.
FLUSH GENERATOR
THE GENERATOR SHOULD BE FLUSHED EVERY DAY.
TO FLUSH GENERATOR:
1. VERIFY GENERATOR PRESSURE IS 0 psig.
2. OPEN GENERATOR MANUAL DRAIN VALVE.
3. PRESS START TIMER TO BEGIN FLUSH PHASE.
4. ALLOW GENERATOR TO FLUSH FOR 5 MINUTES.
FLUSH TIMER: 5:00
START
TIMER
STOP
TIMER
CANCEL
72
2. Check generator pressure gauge (see Fig. 3-12). Generator must be at
0.0 psig and room temperature before flushing.
Water Supply Valve to the Generator
Water-level Sight Glass
Generator Pressure Gauge
Generator Drain Valve
Generator Electric Box
Figure 3-12. Optional Steam Generator
3-11
Component Identification Operating Instructions 129367-408
NOTE: If generator is not at 0.0 psi, the Flush can be bypassed by pressing CANCEL, however the flush should not be bypassed on a continuous basis or else damage to the generator may occur. Failure to flush the generator on a daily basis will void the generator warranty.
To ensure generator is at 0.0 psig the sterilizer can be shut off at end of the day and by next morning the unit will be able to be flushed. Approximately seven hours are required for generator to cool down to less than 140°F.
3. Open drain valve on the side of the generator electric box (see Fig. 3-12).
4. Verify that the water supply valve to the sterilizer is open.
5. Ensure the water supply valve to the generator is open (see Fig. 3-12).
6. Press START TIMER on screen #72. Water automatically flushes through the generator and out the drain for 3.0 minutes. Flush timer on screen #72 counts down time remaining in the flush phase.
7. After 5.0 minutes, alarm buzzer sounds and display advances to screen #73. Instructions on how to restart the generator are listed on screen #73.
73
WARNING – BURN HAZARD:
FLUSHING OF GENERATOR COMPLETE
TO RESTART GENERATOR:
1. CLOSE GENERATOR MANUAL DRAIN VALVE.
2. PRESS CONTINUE TO START THE GENERATOR AND ADVANCE TO THE MAIN MENU.
CONTINUE
Sterilizer operator may be se­verely burned by scalding wa­ter if the water level control malfunctions. The steam gen­erator level control may mal­function if the supply water exceeds 26,00 ohms/cm (38.5 micromhos conductiv­ity min.). Do not connect to
8. Close the generator drain valve.
9. Press CONTINUE on screen #73. Generator automatically fills to the proper level and starts to heat. Display screen advances to the Main Menu screen (#1). Allow ten minutes warm-up time once generator starts to fill.
10. Close front cabinet panel.
NOTE: The generator must be flushed every day before use.
treated water (e.g., distilled, reverse osmosis, deionized) unless water resistivity is de­termined to be acceptable. If water exceeds 26,000 ohms/ cm, contact STERIS for infor­mation concerning modifica­tions required to the genera­tor control system.
Total Hardness as CaCO3* 0-17 mg/L 130 mg/L
Total Dissolved Solids 50-150 mg/L 250 mg/L
Total Alkalinity as CaCO
Resistivity - ohms/cm 2000-6000 26000
*17.1 mg/L = 1 grain hardness
Table 3-1. Required Feed Water Quality for
Carbon Steel Steam Generators
Nominal Maximum
Condition Conditions Conditions
Temperature as supplied 140° F (60° C)
3
pH 6.8-7.5 6.5-8.5
Total Silica 0.1-1.0 mg/L 2.5 mg/L
50-100 mg/L 180 mg/L
3-12
129367-408 Operating Instructions Component Identification

STERILIZER OPERATION 4

Operate sterilizer by referring to the appropriate cycle description in this
4.1 Before
Operating the
Sterilizer
WARNING – BURN HAZ­ARD: Sterilizer, rack/ shelves, and loading car will be hot after cycle is run. Always wear protec­tive gloves and apron when removing a pro­cessed load. Protective gloves and apron must be worn when reloading ster­ilizer following the previ­ous operation.
WARNING – SLIPPING HAZARD: To prevent falls keep floors dry by imme­diately wiping up any spilled liquids or conden­sation in sterilizer loading or unloading area.
section. The steps described on the next three pages are general instructions that apply to all cycle operations.
1. Press ON touch screen pad on the sterilizer control display.
The printer records the time and date that the power is turned ON.
2. Open chamber door
a. Check that drain strainer is clean and in place and that chamber
interior is clean. See ROUTINE MAINTENANCE MANUAL
(P129367-410)
if cleaning is necessary.
b. Close chamber door.
3. Open access door on the operating end of the sterilizer — refer to Figure 4-1. Turn on steam (Fig. 4-1a) and water (Fig. 4-1b) supplies. Close access door.
Steam enters jacket and begins to warm chamber.
4. Open control access door.
Control Access Door
Closed
Open
JACKET
Access Panel
Steam Supply Valve (4-1a)
CHAMB ER
4-1a Steam Supply Valve
Open
Water Supply
Closed
Valve (4-1b)
4-1b Water Supply Valve
Figure 4-1. Steam Valves
Sterilizer Operation Operating Instructions 129367-408
4-1
Loading...
+ 58 hidden pages