Sokkia SX series Quick Manual

Distance
Vertical angle
Horizontal angle
Soft key
[2] Target type icon
: Level 3 Full power
: Level 1 Half or less power remains.
: Level 0 Little power remains.
[1] [2] [3] [4] [5]
[6] [7] [8]
Page number
(Next page exists.)
[1] Battery icon
[9] PPM setting icon
[4] Laser-pointer
/guide light icon
[5] Tilt angle compensation icon
[6] Communication status with
external devices icon
[7] Input mode icon
[8] Input panel icon
CInputtingletters/figures
Blue
Green
Press and hold the icons [2] to [8], and the selection screens appear.
(Flashes red)
[3] Motor icon
: Less than 20%
: 20 to 50%
: More than 50%
Disk usage icon
Touch panel icon
[10] Config mode icon
Status bar
Switch to Config mode
Power key [ON]
Tilt screen
Top me nu
[VERSION]
[OBSERVE]
[MENU]
[CONFIG]
[ESC]
[ESC]
Config mode
Menu mode
Observation mode
Mode Structure
Nomenclature
Display Functions
Basic Key Operation
7 Optical plummet reticle cover 8 Optical plummet eyepiece 9 Keyboard 10 Circular level 11 Circular level adjusting screws 12 Serial / External power source
connector
13 Levelling base locking screw 14 Luminance sensor 15 Display unit 16 Objective lens (Includes
"Laser-pointer function") 17 Guide light 18 Handle locking screw 19 Sighting collimator 20Tubular compass slot 21Vertical Jog 22Power key 23 Horizontal Jog 24 Trigger key 25 Battery cover 26 Stylus (pen) 27 Levelling foot screw 28 Base plate 29Tribrach clamp 30 Telescope eyepiece 31 Telescope eyepiece screw 32 Telescope knob 33 Telescope focussing ring
z Power ON
Press the power key on the side of the instrument.
z Power OFF
Press and hold (about 1sec) the power key on the side of t he instrument.
z Lighting up the reticle/keys and selecting screen
backlight brightness
z Switching to star key mode
z Switching to program mode
z Switching target type
• Changes can also be mad e by tapping the icon on status bar or in star key mode.
z Switching the page
Switches the reticle illumination and key light On/Off (When Key light is On, backlight brigthtness goes down)
{}
Switches to star key mode / basic mode
{PRG} Switches to program mode / basic mode
Switches between target types Prism/360°Prism/Sheet/N-prism(reflectorless)
{FUNC} Toggle between Observation mode screen pages
z Switching the Laser-pointer/Guide light ON/OFF
• Changes can also be made by tapping t he icon on status bar or in star key mode.
z Inputting letters/figures.
(Press and hold )
To turn the laser-pointer/guide light ON/OFF, press and hold until a beep sounds.
{α}
Switch between numerals and alphabetic character s
{SHIFT} + {1} to {9}
In alphabetic characters mode, switch between lowercase characters and upper case character s
each time {SHIFT} (Press and hold)
In alphabetic characters mode, switch between
lowercase characters and upper case character s {SHIFT} + {α}
<Input Panel> is displayed
{0} to {9}
Input numeral or symbol printed above the key
(during numeric input mode)
Input alphabetic character in the order they ar e listed
(in alphabetic input mode) {.} Input a decimal point (during numeric input mode)
{±}
Input a plus or minus sign (during numeric input
mode) {ESC} Cancel the input data. {TAB} Shift to the next item
.
z Selecting options
z Selecting tabs
z Others
{B.S.} Delete a character on the left.
{S.P.}
Input a blank space (increments by 1 when setting the date and time)
{K}/{L} Right and left cursor/Select other option. {O}/{G} Up and down cursor {ENT} Select/accept input word/value.
{O}/{G}
Move the cursor/selection item up/down
{K}/{L}
Move the cursor/selection item left/right or select other option
{TAB}
Shift to the next item
{S.P.}
Display other options
{ENT}
Select/accept the option
{O}/{G} Move tab/cursor in tab up/down {K}/{L} Display next tab at left/right
{ESC} Return to the previous screen
Press the star key [ ]. In the star key mode, it is possible to change the setting item that is frequently used during measurement.
Config mode
Measuring screen
Input screen
Observation mode screen
Operation panel
Star key mode
[1] Battery icon [2] Target type icon [3] Motor icon [4] Laser-pointer/guide light icon [5] Tilt angle compensation icon [6] Communication status with
external devices icon [7] Input mode icon [8] Input panel icon [9] PPM setting icon [10]Config mode icon
Motor configuration
Version information
Laser is emitted
21305 90080
Instrument name
Serial number
Switch to <Motor configuration>
18
16
15
13
11 12
10
8
7
5
2
4
1
9
6
3
17
19
14
21
23
24
25
33
32
29
26
27
28
20
22
31
30
[1] [2] [3] [4]
[9] [6] [7] [10]
[5]
[8]
1 Wireless antenna 2 Handle 3 Beam d etector for Remote
Controler 4 Instrument height mark 5 External interface hatch 6 Optical plummet focussing ring
{ }
Star key mode
{ }
Display unit
{ }
Illumination
{ }
{ }
{ }
{ }
Luminance sensor / Microphone
SHIFT
{ } { }
FUNC
Switching input mode
Switching target type
Program mode
Switching the page
PRG
{ }
ENT
{ }
ESC
{ }
TAB
B.S.
z Disk usage icon
zTouch panel icon
SURVEYING INSTRUMENTS
SX series
Quick Manual
SX_Quick_E.fm 1 ページ 2012年5月14日 月曜日 午後 2時39分
Input over 3 letters !
The input password consists of less than 3 characters. Input a password at least 3 characters in length.
Motor error EXXX
A problem has occurred with the motor drive and operati on stops. Power the instrument OFF then ON to correct the problem. If this error message appears frequently, contact your local dealer.
Need base pt. obs
During REM measurement, the observation of the target was not completed normally. Reset and sight the prism and perform measurement again.
Need 1st obs
During missing line measurement, the observation of the starting position was not completed normally. Sight the starting position accurately and press [MEAS] to perform the measurement again.
New password Diff.
During new password setting, the passwords input twice are different. Input new password twice correctly.
No solution
The calculation of the instrument station coordinate s during resection does not converge. Access the results and if necessary, perform the observations again.
Out of range
During gradient % display, the display range (less than ± 1000%) has been exceeded. During REM measurement, either the vertical angle has exceeded horizontal ±89° or the measured distance is greater than 9999.999m. Install the instrument station far from the target.
Please input an address consisting of 12 hexadecimal characters (0 to 9, A to F)
The Bluetooth address is not input correctly for Bluetooth communication. Input the Bluetooth address with 12 characters of 0 to 9 or A to F.
Reflectorless not supported !!
Auto Tracking cannot be performed in reflectorless mode. Use the prism to carry out automatic sighting.
Remote Control communication err !!
Communication between the Remote controller for Remote Cont rol System and the instrument failed. Check the status (communic ations setup, power supply, cable connections etc.) of the Remote controller, wireless modem and cables.
Sheet not supported !!
Auto Tracking cannot be performed with the sheet. Use the prism to carry out automatic sighting.
Signal off
The reflected light is not observed when distance me asurement begins. Or, during measurement, the reflected light has weakened or is block ed. Either sight the target again or, when using a reflective prism, increase the number of reflective prisms.
Auto Tracking Settings in <Motor configuration>
1.Set search conditions.
• Auto AIM : Fine/Rapid
• MEAS Acc. : Advanced/Standard
• Track Setting : Track
• Srch method : G.S./R.C. •
TURN : 3"/5"/10"/20"/30"/60"
2.Set the search area.
• Search pattern : 1/2/3
• Vertical : 0 to 90°
• Horizontal : 0 to 180° (when "Search pattern" is set to “3” : 0 to 90°)
3.Set the Jog turning speed.
• Lo (1 to 4) / Hi (1 to 7) / Shift (1 to 6)
4.Set "Forecast time" and "Target lost".
• Forecast time : 0.5 sec. /1 sec. / 2 sec. / 3 sec. / 4 sec. / 5 sec. / Auto
• Wait time (sec.) : 0 to 3600
5.Press [OK].
Measurement with Auto Tracking
1.Aim the objective lens in the general direction of the prism.
2.Press [TRK On].
When the target is found, the target is aligned in the field-of-view and Auto Tracking starts. To stop Auto Tracking, press [STOP].
Auto Tracking (Only for Auto Tracking model)
1. Collimate the first target "A".
2. Set the horizontal angle of target "A" to 0°00'00".
Press [0 SET], and it flashes. Press [0 SET] again at once, and the horizontal angle of target "A" is set to 0°.
3. Collimate the second target "B".
The horizontal angle and vertical angle of target "B", which you want to know, are displayed.
1. Collimate the target center.
Make sure that setting is applicable to the measurement conditions.
2. Press [MEAS].
Distance measurement starts.
$
Press the trigger key when [MEAS] and [STOP] are indicated on the display unit. You can start/stop measurement.
The results of the measurement are displayed. Hereafter, the results are updated whenever measurement is performed.
$
Press “SHVdist” tab, and the displayed contents are changed. (Slope distance / Height difference / Horizontal distance / Vertical angle / Horizontal angle)
Trigge r key
Target not found !!
The prism cannot be found within the Search area range. Reset and sight the prism and perform measurement again.
Tem p Rng e OUT
The instrument is outside useable temperature range and accu rate measurement cannot be performed. Repeat measurement within the appropriate temperatur e range.
Tilt over range !!
The tilt angle exceeds the tilt angle compensation ra nge of the sensor. Level the instrument again.
Time out !!
Measurement is not carried out in the allotted time. Reset and sight the prism and perform measurement again. When designating the angle of rotation or automatically s ighting the prism, there is a problem with the positioning o f the prism or the operation of the instrument and measurement is not obtai ned within the fixed time. Check the positioning of the instrument and prism and perfor m measurement again. If observation is still not possible, sight the target manually.
If you are experiencing problems with the instrument and sus­pect a fault in the program, you should try a warm boot.
1. Power OFF the instrument.
2. Press the power key on the side of the instrument while pressing {ENT}.
The instrument is reset and powers ON as normal.
If the problem is not resolved with a warm boot the next step is to perform a cold boot. For cold boot, refer to “Resolving Software Issues” of the manual (PDF).
Listen for the click to ensure that the battery cover and external interface hatch are properly closed.
Resolving Software Issues
General Handling Precautions
To remove the handle, hold both sides of the handle and lift it straight above. If you hold the handle by one hand or incline it, the terminal attached on the handle may be damaged.
Auto Pointing Settings in <Motor configuration>
1.Set search conditions.
• Auto AIM : Fine/Rapid
• MEAS Acc. : Advanced/Standard
• Track Setting : Search
• Srch method : G.S./R.C.
• TURN : 3"/5"/10"/20"/30"/60"
2.Set the search area.
• Search pattern : 1/2/3
• Vertical : 0 to 90°
• Horizontal : 0 to 180° (when "Search pattern" is set to “3” : 0 to 90°)
3.Set the Jog turning speed.
• Lo (1 to 4) / Hi (1 to 7) / Shift (1 to 6)
4.Press [OK].
Auto-Pointing Function for Target Sighting
1.Aim the objective lens in the general direction of the target (Prism or Reflective sheet).
2.Press [SRCH] in any Observation mode screen.
The telescope and top half of the instrument rotate and target auto-search begins. When the target is found, the instrument sights the center of the prism and stops.
The following is a list of the error messages displayed by the instrument and the meaning of each message. If the same error message is repeated or if any message not shown below appears, the instrument has malfunctioned. Contact your local dealer.
Backup battery dead. Clock display may no longer be correct.
The voltage supplied by the lithium battery either declines or is completely discharged. Ask your local dealer to replace the battery for yo u.
Bad condition
The air is shimmering a lot, etc., measuring conditions ar e poor.
The center of the target cannot be sighted. Resight the target.
Unsuitable distance measurement conditions when reflectorless measurement is set. When reflectorless measurement is set, distance cannot be measured because the laser beam is striking at least two surfaces at the same time. Choose a single surface target for distance measurement.
Bad letter !!
A file name containing special characters s uch as "*" and "?" is invalid.
Calculation error
During resection measurement the same point is regist ered multiple times. Set another known point so that the known point coor dinates do not coincide.
Coordinates identical to the known point coordina tes observed during resection exist.
During surface area calculation, conditions nec essary for calculations are not met. Check conditions and try again.
An error occurred during calculation.
Device list is full !!
No more Bluetooth devices can currently be registered. Delete unnecessary devices from the list and try again.
Error: Read Build Info.
Error: Read sysflg
Error: Self check
Error: Read OS Parameter
Error: Write sysflg
Press [OK] to cancel the message. If this error message appe ars frequently, contact your local dealer.
Incorrect password.
Input password does not match set password. Input correct password.
Input device name !!
Bluetooth device name not input. Input device name and complete device registration.
Message List
Target A
0°00'00"
107°09'26"
Horizontal angle
Target B
Distance Measurement
Angular Measurement
Target A
0°00'00"
Auto Pointing
SX_Quick_E.fm 2 ページ 2012年5月14日 月曜日 午後 2時39分