Parts marked with "" are important for maintaining the safety of the set. Be sure to replace these parts with specified ones for maintaining the
safety and performance of the set.
This document has been published to be used for
after sales service only.
The contents are subject to change without notice.
XG-C330X/C430X
XG-C330X
SAFETY PRECAUTION
IMPORTANT SERVICE SAFETY NOTES
IMPORTANT SERVICE SAFETY NOTES (for USA)
Service work should be performed only by qualified service technicians who are
thoroughly familiar with all safety checks and servicing guidelines as follows:
WARNING
1. For continued safety, no modification of any circuit
should be attempted.
2. Disconnect AC power before servicing.
BEFORE RETURNING THE PROJECTOR:
(Fire & Shock Hazard)
Before returning the projector to the user, perform
the following safety checks:
1. Inspect lead wires are not pinched between the chassis
and other metal parts of the projector.
2. Inspect all protective devices such as non-metallic
control knobs, insulating materials, cabinet backs,
adjustment and compartment covers or shields,
isolation resistor-capacity networks, mechanical
insulators, etc.
3. To be sure that no shock hazard exists, check for
current leakage in the following manner:
Plug the AC cord directly into a 120-volt AC outlet, (Do
not use an isolation transformer for this test).
Using two clip leads, connect a 1.5k ohm, 10 watt
resistor paralleled by a 0.15μF capacitor in parallel
between all exposed metal cabinet parts and earth
ground.
Service Manual
Use an AC voltmeter with sensitivity of 5000 ohm per
volt., or higher, sensitivity to measure the AC voltage
drop across the resistor.
All checks must be repeated with the AC plug
connection reversed. (If necessary, a non-polarized
adapter plug must be used only for the purpose of
completing these checks.)
Any reading of 0.3 volts RMS (this corresponds to 0.2
milliamp. AC.) or more is excessive and indicates a
potential shock hazard which must be corrected before
returning the unit to the owner.
Many electrical and mechanical parts in DMD™
Projector have special safety-related characteristics.
These characteristics are often not evident from visual
inspection, nor can protection afforded by them be
necessarily increased by using replacement components
rated for higher voltage, wattage, etc.
Replacement parts which have these special safety
characteristics are identified in this manual; electrical
components having such features are identified by “”
and shaded areas in the Replacement Parts Lists and
Schematic Diagrams. For continued protection,
replacement parts must be identical to those used in
the original circuit. The use of a substitute replacement
parts which do not have the same safety characteristics
as the factory recommended replacement parts shown
in this service manual, may create shock, fire or other
hazards.
WARNING: The bimetallic component has the primary
conductive side exposed. Be very careful in
handling this component when the power is on.
AVIS POUR LA SECURITE
De nombreuses pièces, électriques et mécaniques, dans
les projecteur à DMD™ présentent des caractéristiques
spéciales relatives à la sécurité, qui ne sont souvent pas
évidentes à vue.
Le degré de protection ne peut pas être nécessairement
augmentée en utilisant des pièces de remplacement
étalonnées pour haute tension, puissance, etc.
Les pièces de remplacement qui présentent ces
caractéristiques sont identifiées dans ce manuel;
les pièces électriques qui présentent ces particularités
sont identifiées par la marque “” et hachurées dans la
liste des pièces de remplacement et les diagrammes
schématiques. Pour assurer la protection, ces pièces
doivent être identiques à celles utilisées dans le circuit
d’origine. L’utilisation de pièces qui n’ont pas les mêmes
caractéristiques que les pièces recommandées par l’usine,
indiquées dans ce manuel, peut provoquer des
électrocutions, incendies ou autres accidents.
AVERTISSEMENT: La composante bimétallique dispose du
conducteur primaire dénudé. Faire attention
lors de la manipulation de cette
composante sous tension.
Ne peut effectuer la réparation qu' un technicien spécialisé qui s'est parfaitement
accoutumé à toute vérification de sécurité et aux conseils suivants.
AVERTISSEMENT
1. N'entreprendre aucune modification de tout circuit.
C'est dangereux.
2. Débrancher le récepteur avant toute réparation.
VERIFICATIONS CONTRE L'INCEN-DIE ET
LE CHOC ELECTRIQUE
Avant de rendre le récepteur à l'utilisateur, effectuer
les vérifications suivantes.
1. Inspecter tous les faisceaux de câbles pour s'assurer
que les fils ne soient pas pincés ou qu'un outil ne soit
pas placé entre le châssis et les autres pièces
métalliques du récepteur.
2. Inspecter tous les dispositifs de protection comme les
boutons de commande non-métalliques, les isolants, le
dos du coffret, les couvercles ou blindages de réglage
et de compartiment, les réseaux de résistance-capacité,
les isolateurs mécaniques, etc.
3. S'assurer qu'il n'y ait pas de danger d'électrocution en
vérifiant la fuite de courant, de la facon suivante:
Brancher le cordon d'alimentation directem-ent à une
prise de courant de 120 V. (Ne pas utiliser de transformateur d'isolation pour cet essai).
A l'aide de deux fils à pinces, brancher une résistance
de 1.5 kΩ 10 watts en parallèle avec un condensateur
de 0.15μF en série avec toutes les pièces métalliques
exposées du cof fret et une terre connue comme une
conduite électrique ou une prise de terre branchée à la
terre.
Utiliser un voltmètre CA d'une sensibilité d'au moins
5000Ω/V pour mesurer la chute de tension en travers
de la résistance.
Toucher avec la sonde d'essai les pièces métalliques
exposées qui présentent une voie de retour au châssis
(antenne, coffret métallique, tête des vis, arbres de
commande et des boutons, écusson, etc.) et mesurer la
chute de tension CA en-travers de la résistance. Toutes
les vérifications doivent être refaites après avoir inversé
la fiche du cordon d'alimentation. (Si nécessaire, une
prise d'adpatation non polarisée peut être utilisée dans
le but de terminer ces vérifications.)
Tous les courants mesurés ne doivent pas dépasser 0.2
mA.
Dans le cas contraire, ilyaunepossibilité de choc
électrique qui doit être supprimée avant de rendre le
récepteur au client.
DVM
ECHELLE CA
1.5k ohm
10W
AUX PIECES
METALLIQUES
EXPOSEES
ii
0.15 μF
SONDE D'ESSAI
BRANCHER A UNE
TERRE CONNUE
XG-C330X/C430X
NOTE POUR LE PERSONNEL
D’ENTRETIEN
PRECAUTION POUR LES RADIATIONS UV
La source de lumière, la lampe UHP, dans le projecteur
LCD émet de petites quantités de radiation UV.
EVITEZ TOUTE EXPOSITION DIRECTE DES
YEUX ET DE LA PEAU.
Pour votre sécurité, nous vous prions de respecter
les points suivants:
Précautions pour les radiations UV
et la lampe moyenne pression
NOTE TO SERVICE
PERSONNEL
UV-RADIATION PRECAUTION
The light so urce, UHP lamp, in the LCD projector
emits small amounts of UV-Radiation.
1. Toujours porter des lunettes de soleil lors d’un entretien
du projecteur
avec la lampe allumée
et le haut du coffret retiré.
2. Ne pas faire fonctionner la lampe à l’extérieur du boîtier
de lampe.
3. Do not operate for more than 2 hours with the enclosure
removed.
1. Be sure to disconnect the AC plug when replacing the
lamp.
2. Allow one hour for the unit to cool down before
servicing.
3. Replace only with same type lamp. Type AN-C430LP
rated 275 W AC.
4. The lamp emits small amounts of UV-Radiation, avoid
direct-eye contact.
5. The medium pressure lamp involves a risk of explosion.
Be sure to follow installation instructions described
below and handle the lamp with care.
3. Ne pas faire fonctionner plus de 2 heures avec le coffret
retiré.
1. Toujours débrancher la fiche AC lors du remplacement
de la lampe.
2. Laisser l’unité refroidir pendant une heure avant de
procéder à l’entretien.
3. Ne remplacer qu’avec une lampe du même type. Type
AN-C430LP, caractéristique 275 W-AC.
4. La lampe émet de petites quantités de radiation UVéviter tout contact direct avec les yeux.
5. La lampe moyenne pression implique un risque
d’explosion. Toujours suivre les instructions
d’installation décrites ci-dessous et manipuler la lampe
avec soin.
Since the lamp reaches a very high temperature during
units operation replacement of the lamp should be
done at least one hour after the power has been turned
off. (to allow the lamp to cool off.)
Installing the new lamp, make sure not to touch the
lamp (bulb) replace the lamp by holding its reflector
2
.
[Use original replacement only.]
Lamp
1
Reflector
2
–– Never turn the power on without the
lamp to avoid electric-shock or damage of the devices
since the stabilizer generates high voltages at its start.
Comme la lampe devient très chaude pendant le
fonctionnement de l’unité, son remplacement ne doit
être effectué au moins une heure après avoir coupé
l’alimentation (pour permettre à la lampe de refroidir).
En installant la nouvelle lampe, s’assurer de ne pas
toucher la lampe (ampoule). Remplacer la lampe en
tenant son réflecteur.
[N’utiliser qu’un remplacement d’origine.]
–– Ne jamais mettre sous tension sans la
lampe pour éviter un choc électrique ou des
dommages des appareils car le stabilisateur génère
de hautes tensions à sa mise en route.
2
TIONS UV (Suite)
1
Lampe
2
Reflecteur
Since small amounts of UV-radiation are emitted
from an opening between the exhaust fans, it is recommended to place the cap of the optional lens on
the opening during servicing to avoid eye and skin
exposure.
Comme de petites quantités de radiation UV sont
émises par une ouverture entre les ventilateurs aspirants, il est recommandé de placer le capuchon de
l’optique optionnelle sur l’ouverture pendant l’entretien
pour éviter une exposition des yeux et la peau.
iv
XG-C330X/C430X
WARNING:High brightness light source, do not stare into the beam of light, or view directly. Be especially
careful that
children do not stare directly in to the beam of light.
WARNING:TO REDUCE THE RISK OF FIRE OR ELECT RIC SHOCK, DO NOT EXPOSE
THIS UNIT TO
MOISTURE OR WET LOCATIONS.
CAUTION
AVERTISSEMENT: Source lumineuse de grande intensité. Ne pas fixer le faisceau lumineux ou le regarder
directement. Ve
iller particulièrement à éviter que les enfants ne fixent directement le
faisceau lumineux.
AVERTISSEMENT: AFIN D’EV
ITER TOUT RISQUE D’INCENDIE OU D’ELECTROCUTION, NE PAS PLACER
CET APPAREIL DANS UN ENDROIT HUMIDE OU MOUILLE.
ATTENTION
CAUTION
(INLET Unit)
PRECAUTION
(Unité d'admission)
CAUTION: TO REDUCE THE RISK OF ELECTRIC SHOCK,
6.3A 250V
NO USER-SERVICEABLE PARTS EXCEPT LAMP UNIT.
REFER SERVICING TO QUALIFIED SERVICE
RISK OF ELECTRIC SHOCK.
DO NOT REMOVE SCREWS
EXCEPT SPECIFIED USER
SERVICE SCREW.
DO NOT REMOVE CABINET.
PERSONNEL.
For continued
protection against a
risk of fire, replace
only with same type
fuse. 6.3A, AC250V (F801)
The lighting flash with arrowhead within a
triangle is intended to tell the user that
parts inside the product are risk of electric
shock to persons.
The exclamation point within a triangle is
intended to tell the user that important
operating and servicing instructions are in
the manual with the projector.
L’éclair terminé d’une flèche à l’intérieur
d’un triangle indique à l’utilisateur que les
pi‘eces se trouvant dans l’appareil sont
susceptibles de provoquer une décharge
électrique.
SPECIFIEES
Le point d’exclamation à l’intérieur d’un
D’ELECTROCUTION, NE PAS RETIRER LE CAPOT.
AUCUNE DES PIECES INTERIEURES N’EST REPARABLE
PAR L’UTILISATEUR, A L’EXCEPTION DE L’UNITE DE
LAMPE. POUR TOUTE REPARATION, S’ADRESSER A UN
6.3A 250V
ATTENTION: POUR EVITER TOUT RISQUE
TECHNICIEN D’ENTRETIEN QUALIFIE.
Pour une protection
continue contre un
risques d’incendie, ne
remplacer qu’avec un
fusible du même type fuse.
6.3A, AC250V (F801)
triangle indique à l’utilisateur que les
instructions de fonctionnement et
d’entretien sont détaillées dans les
documents fournis avec le projecteur.
v
XG-C330X/C430X
Precautions for using lead-free solder
Employing lead-free solder
• "PWBs" of this model employs lead-free solder. The LF symbol indicates lead-free solder, and is attached on the PWBs and service manuals. The
alphabetical character following LF shows the type of lead-free solder.
Example:
L Fa
Indicates lead-free solder of tin, silver and copper.
Using lead-free wire solder
• When fixing the PWB soldered with the lead-free solder, apply lead-free wire solder. Repairing with conventional lead wire solder may cause damage or accident due to cracks.
As the melting point of lead-free solder (Sn-Ag-Cu) is higher than the lead wire solder by 40 °C, we recommend you to use a dedicated soldering
bit, if you are not familiar with how to obtain lead-free wire solder or soldering bit, contact our service station or service branch in your area.
Soldering
• As the melting point of lead-free solder (Sn-Ag-Cu) is about 220 °C which is higher than the conventional lead solder by 40 °C, and as it has poor
solder wettability, you may be apt to keep the soldering bit in contact with the PWB for extended period of time. However, Since the land may be
peeled off or the maximum heat-resistance temperature of parts may be exceeded, remove the bit from the PWB as soon as you confirm the
steady soldering condition.
Lead-free solder contains more tin, and the end of the soldering bit may be easily corroded. Make sure to turn on and off the power of the bit as
required.
If a different type of solder stays on the tip of the soldering bit, it is alloyed with lead-free solder. Clean the bit after every use of it.
When the tip of the soldering bit is blackened during use, file it with steel wool or fine sandpaper.
• Be careful when replacing parts with polarity indication on the PWB silk.
• Depending on the controlling device used, it may be necessary to connect Pin 4 and Pin 6 on the controlling
456
456
1TX+ 5
2TX- 6 RX-
3RX+ 7
48
Pin No. Signal Pin No.Signal
LAN Terminal : 8-pin RJ-45 modular connector
8... 1
1 – 5
[3] Dimensions
XG-C330X/C430X
Units: inches (mm)
Side ViewSide View
Rear View
Top View
1227/32 (326)
31
3
/64 (88.5)
17
(6)
4
/
1
(279.5)
64
/
1
11
64
/
(6.5)
Front View
Bottom View
32
/
31
(50)
1
4 (101.5)
64
/
63
(25)
15
/16
2
(76.5)
64
/
1
3
(74.5)
53
/64 (148)553/64 (148)
5
29
/32
15/64
(27.25)
2
(73.5)
(146.5)
32
/
25
5
(235.5)
32
/
9
9
(268.5)
64
/
37
10
261/64
(75)
1 – 6
XG-C330X/C430X
[4] Resetting the total lamp timer
When replacing the lamp, reset the total lamp timer in the procedure
below.
Resetting the Lamp Timer
Reset the lamp timer after replacing the lamp.
Info
Make sure to reset the lamp timer only
when replacing the lamp. If you reset the
lamp timer and continue to use the same
lamp, this may cause the lamp to become
damaged or explode.
You can also reset the lamp timer via the
LAN.
(Refer to the SETUP MANUAL on the supplied CD-ROM for details.)
1 Connect the power cord.
Plug the power cord into the AC socket
of the projector.
2 Reset the lamp timer.
While simultaneously holding down ,
ENTER and MENU on the projector,
STANDBY/ON on the projector.
press
"LAMP 0000H" is displayed, indicating
that the lamp timer is reset.
AC socket
STANDBY/ON button
ENTER button
button
MENU button
Lamp
• It is recommended that the lamp unit (optional: AN-C430LP) be
replaced when the remaining lamp life becomes 5% or less, or
when you notice a significant deterioration in the picture and color
quality. The lamp life (percentage) can be checked with the onscreen display.
• Purchase a replacement lamp unit of type AN-C430LP from your
place of purchase, nearest Sharp Authorized Projector Dealer or
Service Center.
Maintenance Indicators
Power indicator
Lamp indicator
Temperature warning indicator
About the temperature warning indicator
If the temperature inside the projector increases, due to blockage of the air vents, or the setting location,
"" will illuminate in the lower left corner of the picture. If the temperature keeps on rising, the lamp will
turn off and the temperature war ning indicator will blink, the cooling fan will run for a further 90 seconds, and
then the projector will enter the standby mode. After "
• The warning lights on the projector indicate problems inside the
projector.
• If a problem occurs, either the temperature indicator or the lamp
indicator will illuminate red, and the projector will enter the standby
mode. After the projector has entered the standby mode, follow the
procedures given below.
" appears, ensure you perform the measures.
About the lamp indicator
When the remaining lamp life becomes 5% or less, " " (yellow) and "Change The Lamp." will be
displayed on the screen. When the percentage becomes 0%, it will change to "
automatically turn off and then the projector will automatically enter the standby mode. At this time,
the lamp indicator will illuminate in red.
If you try to turn on the projector a fourth time without replacing the lamp, the projector will not turn on.
" (red), the lamp will
1 – 7
XG-C330X/C430X
Maintenance indicator
Temperature
warning
indicator
Lamp
indicator
Power
indicator
Normal
Off
Green on
Green blinks
when the lamp
is warming up
Green on/
Red on
Green blinks
(Cooling)
Abnormal
Abnormal
Red on
(Standby)
Red on
Red on
(Standby)
Red blinks
Problem
The internal
temperature is
abnormally high.
The lamp does
not illuminate.
Time to change
the lamp.
The lamp does
not illuminate.
The power
indicator blinks
in red when the
projector is on.
CausePossible solution
• Relocate the projector to an area
• Blocked air intake
• Cooling fan breakdown
• Internal circuit failure
• Clogged air intake
• The lamp is shut
down abnormally.
• Remaining lamp life
becomes 5% or less.
• Burnt-out lamp
• Lamp circuit failure
• The filter cover, lamp
unit cover or lens
housing cover is
open.
• Cooling fan breakdown
with proper ventilation
• Take the projector to your nearest
Sharp Authorized Projector Dealer
or Service Center
for repair.
• Clean the exhaust and intake
vents.
• Disconnect the power cord from
the AC outlet, and then connect it
again.
• Carefully replace the lamp.
• Take the projector to your nearest
Sharp Authorized Projector Dealer
or Service Center
for repair.
• Please exercise care when
replacing the lamp.
• Securely install the lamp unit cover.
• If the power indicator blinks even
when the covers are securely
installed, or if the cooling fan does
not run normally, then contact your
nearest Sharp Authorized
Projector Dealer or Service Center
for advice.
Info
• If the temperature warning indicator illuminates and the projector enters the standby mode, check whether
any of the ventilation holes are blocked and then try turning the power back on. Wait until the projector
has cooled down completely before plugging in the power cord and turning the power back on.
(At least 10 minutes.)
• If the power is turned off for a brief moment due to power outage or some other cause while using the
projector, and the power supply recovers immediately after that, the lamp indicator will illuminate in red
and the lamp may not be lit. In this case, unplug the power cord from the AC outlet, replace the power
cord in the AC outlet and then tur n the power on again.
• The cooling fan keeps the internal temperature of the projector constant and this function is controlled
automatically. The sound of the cooling fan may change dur ing operation because the fan speed may
change and this is not a malfunction.
1 – 8
XG-C330X/C430X
XG-C330X
CHAPTER 2. REMOVING OF MAJOR PARTS
Service Manual
[1] Removing the Top Body and the Intake Cover.
1. Remove the 2 screws from the rear side.
2. Remove the 8 screws from the Top Body.
3. Lift and hold the Top Body . Remove [LF] connector.
4. Detact the Top Body.
5. Detach the Intake Cover.
1-4
Top Body
1-1
1-2
1-3
LF
2 – 1
1-5
Intake Cover
[2] Removing the Connecting Cables and Main PWB.
1. Remove the 3 screws from the rear side.
2. Detach all the Connecting Cables from PWBs.
3. Detach the FFCs (RP, GP, BP) from Main PWB.
4. Remove the 3 screws and detach the Main PWB.
5. Detach the 1 screw and detach the Lamp Door.
2-4
2-4
CP
SO
LL
FE
FD
FC
T0
CP
2-3
RP
BP
GP
T1
FF
FB
EA
XG-C330X/C430X
2-1
T2
FA
Lamp Door
2-5
2 – 2
XG-C330X/C430X
[3] Removing the Ballast Ass'y and the Speaker.
1. Remove the 1 screw and detach the Trigger Box.
2. Remove the 3 screws.
3. Remove the 2 cables and detach the Ballast Ass’y.
4. Detach the Speaker.
5. Remove the 1 screw and detach the Bimetal Holder.
3-1
3-2
Trigger
3-3-1
Box
3-2
Ballast Shield
3-5
3-3-2
3-3-2
Bimetal
Holder
3-4
2 – 3
[4] Removing the Optical Mechanism Unit.
1. Remove the 6 screws.
2. Remove the 2 screws.
3. Remove 1 screw and detach the earth lead.
4. Detach the Optical Mechanism Unit.
XG-C330X/C430X
4-3
4-1
4-2
Optical
Mechanism
unit
2 – 4
XG-C330X/C430X
[5] Removing the Power Unit, Inlet Unit, RC Unit and CONDENSER Unit.
1. Remove the 3 screws and Detach the Power Unit.
2. Detach the Inlet Unit.
3. Detach the R/C unit.
4. Remove the 1screw and detach the Condenser Unit.
5-1
5-4
Inlet
Unit
POWER
Unit
Condenser
Unit
5-3
R/C Unit
2 – 5
XG-C330X
CHAPTER 3. THE OPTICAL UNIT OUTLINE
Service Manual
[1] THE OPTICAL UNIT OUTLINE
1. Layout for proper setup of the optical components and parts (top view)
XG-C330X/C430X
Output polarizer R
Output Pre-polarizer R
LCD (R)
Input polarizer R
Dichroic filter R
Condencer lens R
Condencer lens 5
Input polarizer G
Input Pre-polarizer G
Condencer lens G
CL5
M3
M3
Output polarizer B
Input pre-polarizer B
B reflect
M4
o
r
B
r
e
fl
ec
M1
Output pre-polarizer B
LCD (B)
Input polarizer B
Condencer lens B
Output polarizer G
Output pre-polarizer G
Output pre-pre-polarizer G
CL2
tor
PBS
FLYEYE LENS (OUTPUT)
FLYEYE LENS (INPUT)
CL1
Condencer lens 1
Condencer lens 2
Light source
(Lamp)
Projection Lens
Cross dichroic
r
o
t
ec
fl
e
r
R
CL4
R
r
e
f
lect
CL4
o
r
prism
CL3-G
G
ref
M2
Condencer lens 4
LCD (G)
l
e
c
t
o
r
CL3-B
3 – 1
XG-C330X/C430X
2. Adjusting the mirrors
This adjustment is needed when any of the optical parts of the optical mechanism has been replaced.
1. Disconnect the flat cables from all the LCD panels.
2. Light up the lamp.
3. Project a white-light image and check to see if there is any color tint in any direction. If any, use the M2, M4 and M5
adjusting levers.
4. Loosen the adjusting lever lock screws, make adjustments, and tighten up the lock screws.
Black
Black
Black
3. Correcting color irregularities on white-only screen when replacing the lamp
If color irregularities are found at the right and left on a white-only screen after replacing the lamp, it is necessary to
readjust the optical axis of the new lamp. Take the following steps.
1. Open the lamp cover and loosen the screws A and B at the top of the lamp.
2. Using a screwdriver or the like, move the lever D in the arrow direction.
3. Temporarily fix the screws A and B, close the lamp cover, and check the white-only screen again for color
irregularities.
4. Repeat the above steps 1 and 2 until there will be no color irregularities. Now tighten up the screws A and B.
5. Finally secure the lamp cover back in position.
3 – 2
4. Replacing the prism holder unit
1. Remove the two lock screws, and take the prism holder unit out of the optical unit.
2. Replace the prism holder unit with new one. Take the above step 1 in reverse order.
NOTE: Even if just one of the LCD panels is defective, it is necessary to replace the entire prism unit. Do not replace
just the defective LCD panel only. Under the present circumstances, please do not touch the display part of
LCD.
XG-C330X/C430X
Optical unit
Prism holder unit lock screws
R-LCD panel
G-LCD panel
B-LCD panel
3 – 3
XG-C330X/C430X
XG-C330X
CHAPTER 4. ELECTRICAL ADJUSTMENT
Service Manual
[1] PW software rewriting procedure
1. Prepare before rewriting.
1. Please prepare one RS232C cross cable and conversion cable.
Connect the conversion cable to the Projector, and it connects with a RS232C cross cable.
Connect opposite side to the serial port of personal computer.
2. Open the filter cover of the bottom cabinet and insert the LEAF SW JIG:PSPAZB245WJKZ.
Projector will be rewriting mode.
LEAF SW JIG:PSPAZB245WJKZ
If you don’t have LEAF SW JIG, open the top cabinet and select S2701 of main PWB to “WRITE” side.
Projector will be rewriting mode.
4 – 1
2. How to rewrite the software.
1. Projector software rewriting will be done with the Flash Up grader.
2. Unpack the software file.
3. Double-click on the "FlashUpgrader.exe ", so below window will be displayed on PC.
XG-C330X/C430X
4. Click on the “Choose” button.
Then, below window will be displayed on PC.
4 – 2
XG-C330X/C430X
5. Clock on the above position and select the latest software
If you mistook the folder, please select the unpacked folder.
6. There is file named " pwSDK.inf " in this folder.
4 – 3
7. Double-click on the "pwSDK.inf".
XG-C330X/C430X
Note:
Options: Continuous Flash Mode; Do not put check point.
Notify On Completion; Do not put check point.
Reset Target After download; Put the check point.
Modes: Put the check point in “Flash”.
8. Click on the “Flash” button.
Then, below image will be appeared.
Now, preparation is completed.
4 – 4
XG-C330X/C430X
9. Next, connect the AC cord of the projector and software rewriting will start automatically.
If the gauge is not started, please check the “LEAF SW JIG” or “Write SW”.
Rewriting times are about 3 or 4 minutes.
Do not disconnect the AC cord by at this time.
Below window is under writing.
And below window will appear after software rewriting is completed.
Please click on “Close “ and software rewriting is finished.
10.Remove the LEAF SW JIG : PSPAZB245WJKZ and attach the filter cover.
Or select S2701 of main PWB to “NORMAL” side and attach the top cabinet.
Projector will be Normal mode(Non-writing mode)
4 – 5
XG-C330X/C430X
3. Rewriting of the model name
1. Rewriting of the Model name will be done with the Tera Term.
2. Send the following command.
“MF1_0000” for XG-C430X
“MF1_0001” for XG-C330X
3. Remove the AC cord
4. Confirmation of version.
1. Enter the “factory mode”,
With the menu not displayed, press the “ENTER”, “ENTER”, “VOL+”, “VOL-”, “ENTER”, “ENTER” and “MENU” keys on the projector
2. Below image is appeared and select “VERSION”.
Latest version will be appered.
1EEPROM initialization1) Turn on the power (the lamp turns on)
and perform aging for 15 minutes.
2R-bright check
Contrast adjustment
(manual or automatic)
1) Input the white/black wind pattern signal of amplitude level 96% (0.67 Vpp)/
0% of XGA60Hz to COMPUTER1.
2) Select the following.
Group: AD
Adjustment item: R-Bright
R-Contrast
(Process gamma linkage)
1) Perform the following setting.
Press S6501 to enter the process mode, and execute SS2 of the
SS menu.
1) Check the fixed value.
R-Bright: 63
2) Observe the chromaticity of 96% white wind pattern portion using
CA100.
3) On the screen where bit dropouts occur, raise the R-Contrast
value. Adjust to the point where black to bright red bit dropouts
occur and the x value changes by 30/1000 or more.
2) Check that the white pattern for process adjustment is displayed.
3) Connect a synchroscope to 2P of
(P1401).
4) For red and blue, connect to (P14011P) and (P1401-3P) respectively.
1) Check the fixed value.
G-Bright: 63
2) Observe the chromaticity of 96% white wind pattern portion using
CA100.
3) On the screen where bit dropouts occur, raise the G-Contrast
value. Adjust to the point where black to bright green bit dropouts
occur and the y value changes by 30/1000 or more.
1) Check the fixed value.
B-Bright: 63
2) Observe the chromaticity of 96% white wind pattern portion using
CA100.
3) On the screen where bit dropouts occur, raise the B-Contrast
value. Adjust to the point where black to bright blue bit dropouts
occur and the y value changes by 30/1000 or more.
1) Check the fixed value.
R-Bright2: 63
2) Observe the chromaticity of 100% white wind pattern portion using
CA100.
3) On the screen where bit dropouts occur, raise the R-Contrast2
value. Adjust to the point where black to bright red bit dropouts
occur and the x value changes by 30/1000 or more.
1) Check the fixed value.
G-Bright2: 63
2) Observe the chromaticity of 100% white wind pattern portion using
CA100.
3) On the screen where bit dropouts occur, raise the G-Contrast2
value. Adjust to the point where black to bright green bit dropouts
occur and the y value changes by 30/1000 or more.
1) Check the fixed value.
B-Bright2: 63
2) Observe the chromaticity of 100% white wind pattern portion using
CA100.
3) On the screen where bit dropouts occur, raise the B-Contrast2
value. Adjust to the point where black to bright blue bit dropouts
occur and the y value changes by 30/1000 or more.
1) Select G1-GAIN, and adjust the signal amplitude as follows using
the operation SW or button on the remote control (see figure (1)
below).
4.6 Vpp
2) Select G1-BLK, and adjust the white-white level as follows using
the operation SW or button on the remote control (see figure (2)
below).
0.5 Vpp
3) Adjust red and blue in the same way.
(G waveform diagram of process pattern)
± 0.05 V
± 0.1 V
* The part defining the amplitude to the waveform is the same for both
R and B.
2) Select the following.
Group: OUTPUT1
Adjustment item: S-G1-BLK
S-B1-BLK
1) Select the following.
Group: LINE
Adjustment item: Calibration
1) Check the blur and ghost of characters of OUTPUT3.
2) Raise the ENBR-DLY value until pre-ghost appears on the left of
characters of OSD.
3) When the value is reduced by 1 to 2 points, pre-ghost disappears.
Set to the value reduced by 3 from the value where pre-ghost disappears.
(Set to 16 in the following case: When the ENBR-DLY value is 20,
pre-ghost appears; when it is 19, pre-ghost disappears.)
4) Adjust green and blue in the same way.
NOTE: EPSON panels may vary by 1 to 2 points from lot to lot. To
improve the ghost quality, perform this adjustment.
1) Adjust flicker using the operation SW or button on the remote control so that it becomes minimum.
2) When flicker appears on the screen unevenly, adjust the center
area considering the whole balance.
1) 1. Adjust R1-BLK and B1-BLK so that the following chromaticity
values are obtained in the middle of the screen using CL200.
x = 294 ± 5, y = 327 ± 5
1) Adjust S-G1-BLK and S-B1-BLK so that the following chromaticity
values are obtained in the middle of the screen using CL200.
x = 322 ± 5, y = 341 ± 5
1) With the set placed on a level surface, press the ENTER key and
check that K-Sensor value is 0 ± 3 (125 - 3).
Check that keystone is caused by tilting the unit.
XG-C330X/C430X
14Model name writing1) Write the model name with RS232C
command.
15Tracking check1) Input the RGB 10-step gray scale sig-
nal (inversion on top and bottom).
2) Select the following.
Group: OUTPUT3
Adjustment item: GAMMA
16Auto color shading
correction
1) Perform the auto color shading correction using the color shading correction
system (ccdc).
1) C430: MF_10000 (default)
C330: MF_10001
1) Check the lower gradation tracking.
When it is improper, optimize the GAMMA value.
1: Default (normal)
2: Red and blue down
3: Red and blue up
NOTE: When changing the GAMMA value, readjust the white bal-
ance.
1) Check that remarkable color shading is not found on the screen.
2. Cautions for 3D color shading correction/adjustment
The following measurement and adjustment must be performed under the same conditions using the following:
• Dark room (Block out stray light.)
• Diffuse reflection (mat type) screen
The setting of the projector must be performed under the following conditions:
• Power save disabled (to keep the lamp stable during measurement)
• Auto search canceled (to fix to INPUT1)
• RGB mode of the input terminal COMPUTER1 (component mode canceled - just to be sure)
1. System setting (“Setting -” System setting””)
Condition setting for C330/C430
System option
Simultaneous writing of 3D correction value:No (not checked)
Char transmission interval:0 ms
Binary interval:2 ms
Creation of correction value by linear correction factor:No (not checked)
Number of image captures (during adjustment or measurement): Twice
4 – 8
XG-C330X/C430X
System option
Number of image captures (during teaching):4 times
2. Teaching
Basically, teaching should be measured only once from model to model. However, if the optical characteristics of the set are changed significantly,
it is necessary to reset the conditions according to changes and to perform remeasurement.
Measurement method of correction factor
1) Determine the standard set of C330/C430. Each one unit (n units)
2) Switch the set to the RGB mode of the input terminal COMPUTER1.
3) Execute teaching of the standard set under the following conditions.
Condition setting for C330/C430
Model name:Enter the model name newly.
Resolution:1024 x S768
Binary interval:2 ms
Correction value, X direction:33
Correction value, Y direction:25
Number of correction bits:8
Adjustment value, minimum:-32
Adjustment value, center:0
Adjustment value, maximum:32
Number of teachings:5
Teaching interval:32
Brightness variation direction by adjustment value: Brightness UP by increasing the adjustment value (checked)
Adjustment point:3D (checked)
Adjustment point (Test pattern display level)
Tone 0:44%
Tone 1:64%
Adjustment color:Adjustment with RGB (checked)
3. Color shading correction (adjustment)
Condition setting for C330/C430
TG system (selected)
Tone 0 (checked)
Tone 1 (checked)
Other tones are not checked.
Correction setting
Tone 7:Default
Tone 6:Default
Tone 5:Default
Tone 4:Default
Tone 3:Default
Tone 2:Default
Tone 1:Correction by correction factor
Tone 0:Correction by correction factor
Correction method
Select the model name created in the section (2) and perform color shading correction/adjustment.
If the set has significantly different optical characteristics from the standard or shows color shading or line dropouts, execute teaching and adjust it
using the correction factor.
32Monitor out check1) Input the video and audio signals to the
33RS232C operation
check
34USB operation check1) Connect to a PC with an USB cable.1) Check that forward/reverse operation is enabled on the PC screen
35LAN operation check1) Connect to a PC with a LAN cable.1) Check that communication is possible.
36Remote control check1) Check the wired remote operation.
1) Receive the RGB gray scale signal.1) It must be better than the vertical stripe boundary sample.
1) Heat the thermistor with a hair dryer.1) Make sure the temperature is displayed.
1) Receive the phase checking pattern
signal.
COMPUTER1 and COMPUTER2 terminals.
2) Connect another monitor to the monitor
out terminal.
3) Connect a speaker with a built-in amplifier to the audio out terminal.
1) Connect the unit and a PC with the
RS232C cable.
2) Check the remote control operation.
1) Check that there is no disturbance in the white balance using a
monitor.
1) Select OFF in the process mode.
Make sure that the off-timer starts with 5-minute display, counts
down 1minute for 1second, and turns off when 0minute is displayed.
1) Make sure that Clock, Phase, H-POS and V-POS can be automatically adjusted in the VGA/S-VGA/XGA mode.
1) Check that image selected with the projector is displayed on
another monitor connected to monitor out terminal and sound is
heard from the speaker connected to the audio out terminal.
1) Send a command from the PC, and check it functions correctly.
using the remote control.
1) Check that the wired remote control terminal is activated.
History 1
History 2
History 3
History 4
TOTAL TIME
Process modeLINEOFF
LED CHECK
Calibration
K-Sensor
V-S ize
Fan-mode
Nfan-Normal
4 – 13
XG-C330X
CHAPTER 5. TROUBLE SHOOTING TABLE
Service Manual
[1] TROUBLE SHOOTING TABLE
Checking the basic operation
XG-C330X/C430X
Does the power LED light up or
flash in red or green?
YES
Does the set function with its keys
or the remote controller?
YES
Lamp does not turn on or it turns
off in a short time.
NO
Colors are not display correctly.YESHas P1101, P1201 or P1301
NO
Are RGB and component input
displayed normally?
NOHas P1704 come off or is it
loose?
NOGo to "Main PWB check".
YESGo to "Lamp fails to light-up".
come off or is it loose?
NOGo to "RGB and component sig-
nal check".
YESFully insert the connector.
YESFully insert the connector.
YES
Is VIDEO input displayed normally?
YES
Is S-VIDEO input displayed nor-
mally?
YES
Does the LAN function?NOGo to “checking the LAN”.
YES
END
NOGo to "VIDEO input check".
NOGo to "S-VIDEO input check".
5 – 1
XG-C330X/C430X
Main PWB check
Is 6 V applied to TL1711, 13 V to
TL1712 and 17V to TL1727?
YES
Is approx. 6 V applied to TL1709?NOCheck IC1710 and peripheral circuits.
YES
Is 1.5 V applied to TL1705?NOCheck IC1706 and peripheral circuits.
YES
Is 3.3 V applied to TL1706?NOCheck IC1704 and peripheral circuits.
YES
Is 2.5 V applied to TL1704?NOCheck IC1702 and peripheral circuits.
YES
Is 5 V applied to TL1708?NOCheck IC1707 and peripheral circuits.
YES
Is 3.3 V applied to TL1703?NOCheck IC1708 and peripheral circuits.
NOGo to "Power supply ballast PWB check".
YES
Is 15.5 V applied to TL1707?NOCheck IC1705 and peripheral circuits.
YES
Is5 V applied to TL1701?NOCheck IC1701 and peripheral circuits.
YES
Is 3.3V applied to TL1702?NOCheck IC1703 and peripheral circuits.
YES
Is 5.4V applied to TL2075?NOCheck IC2001 and peripheral circuits.
YES
Is 3.3 V applied to pin(2) of S2701?NOCheck that it is set to the NORMAL side.
Or remove S2701.
YES
Is rectangular wave of 14.318 MHz
sent to X2003 pin(3)?
YES
Is rectangular wave of 48 MHz sent
to X2201 pin(3)?
NOCheck X2003 and peripheral circuits.
NOCheck X2201 and peripheral circuits.
YES
Check IC2001, IC2201, IC2202 and
peripheral circuits.
5 – 2
Lamp fails to light-up
XG-C330X/C430X
Turn on the power switch.Is
dischargeing sound heard
from lamp?
NO
Is the ballast cooling fan
runnning?
YES
Is the DC voltage of 380V
applied to pin(1) of P702 connector?
YES
Is LOW LEVEL voltage
applied between pins(3) and
(4) of Main's P1702 connec-
tor?
YESYES
Replace the ballast.
YES
NO
NO
NO
Is the lamp out of socket?
NO
Repalce the lamp.
Check the power supply circuits.
Check the power supply circuits.
Is the P1702 connector disconnect?
Reconnect it into socket correctly.
YES
NO
Reconnect the lamp into
socket.
Check the microprocessor
and its peripheral circuits.
5 – 3
XG-C330X/C430X
Checking the power unit
There is no voltage output at P703?
YES
Is AC voltage (90-264V) applied
acress the P801 connector?
YES
Is AC voltage (90-264V) applied
acress the P701 connector?
YES
Is the fuse F801 open?
NO
Is the Bimetal SW open?
NO
Replace the Inlet unit.
NO
NO
NO
YES
YES
Check the Main PWB
unit.
Check the P801 or AC
Cord.
Is the resistor R729
open?
NO
Replace power unit?
Replace the F801.
Push the red button of
the bimetal.
YES
Replace R729.
5 – 4
RGB and component signal
check
Send component signal or
RGB signal of 1080i (or 720P)
from INPUT 1 or INPUT 2.
Select INPUT 1 or INPUT 2
using keys on the main unit or
the remote control.
XG-C330X/C430X
Are picture signals sent to
TL2502, TL2503, and
TL2504?
YES
Is 3.3V applied to TL1702?NOCheck IC1703 and peripheral cir-
YES
Is 3.3V applied to TL1703?NOCheck IC1708 and peripheral cir-
YES
Check IC2505 and peripheral
circuits.
NOCheck IC2403 and peripheral cir-
cuits .
cuits.
cuits.
5 – 5
XG-C330X/C430X
S-VIDEO (INPUT 3)
check
Send S-Video signal
(color signal) from
INPUT 3. Select INPUT
3 using keys on the
main unit or the remote
control.
Is image displayed in
black and white?
NOYES
Is S-VIDEO signal (Y)
sent to C3139?
YESYES
Is 5 V applied to
TL1701?
YESCheck S-VIDEO ter-
minal and pin(43) of
IC3103.
NOCheck S-VIDEO ter-
minal and peripheral
circuits of IC3103.
NOCheck IC1701 and
peripheral circuits.
VIDEO (INPUT 4)
check
Send VIDEO signal from
INPUT 4. Use buttons
on the main unit or the
remote control to select
INPUT 4.
Is VIDEO signal sent to
C3137?
Is 5 V applied to
TL1701?
Check IC3103 and
peripheral circuits.
NOCheck VIDEO termi-
nal and peripheral circuits of Q3104 and
Q3107.
NOCheck IC1701 and
peripheral circuits.
YES
Check IC3103 and
peripheral circuits.
5 – 6
Checking the LAN
LAN communication is disabled even
though connecting the control PC with a
LAN cable. The LINK LED (green) does not
light up.
Are the connection of connecting cable and
the settings of IP address, sub-net address
and default gateway correct?
YES
XG-C330X/C430X
NOCheck the connecting cable and LAN settings. If the LAN
settings are unknown, contact the network administrator.
If the network settings cannot be saved, replace IC6708
(EEP ROM) as it may be defective.
Note: When IC6708 (EEP ROM) is replaced, the MAC
address returns to default. Refer to "Resetting after replacing IC6708 (EEP ROM)" to reset.
Are the network device (e.g. Hub) between
the control PC and XG-C430X and cable
normal?
YES
When data is sent from the control PC, do
pins (30) and (31) of IC8201 (PHY receiver)
receive signals?
YES
Check the PHY receiver IC8201 and the
peripheral circuits of the microprocessor
IC8001.
NORepair defective parts in the network.
NOCheck the peripheral circuit of J8201 (RJ-45 jack with built-
1. Voltages at test points are measured at the supply voltage of AC
220V. Signals are fed by a color bar signal generator for servicing
purpose and the above voltages are measured with a 20k ohm/V
tester.
2. WAVEFORM MEASUREMENT CONDITION:
1. Waveforms at test points are observed at the supply voltage of AC
220V. Signals are fed by a color bar signal generator for servicing
purpose.
3. INDICATION OF RESISTOR & CAPACITOR:
RESISTOR
1. he unit of resistance “Ω” is omitted.
(K=kΩ=1000 Ω, M=MΩ).
2. All resistors are ± 5%, unless otherwise noted.
(J=± 5%, F= ± 1%, D= ± 0.5%)
3. All resistors are 1/10W, unless otherwise noted.
4. All resistors are Carbon type, unless otherwise noted.
XG-C330X/C430X
Service Manual
: Solid : Cement
C
S
: Oxide Film : Special
N
: Metal Coating
CAPACITOR
1. All capacitors are μF, unless otherwise noted.
(P=pF=μμF).
2. All capacitors are 50V, unless otherwise noted.
3. All capacitors are Ceramic type, unless otherwise noted.
(ML): Mylar (TA): Tantalum
(PF): Polypro Film (ST): Styrol
CAUTION:
This circuit diagram is original one, therefore there may be a
slight difference from yours.
W
T
SAFETY NOTES:
1. DISCONNECT THE AC PLUG FROM THE AC
OUTLET BEFORE REPLACING PARTS.
2. SEMICONDUCTOR HEAT SINKS SHOULD BE
REGARDED AS POTENTIAL SHOCK HAZARDS
WHEN THE CHASSIS IS OPERATING.
IMPORTANT SAFETY NOTICE:
PARTS MARKED WITH “” () ARE
IMPORTANT FOR MAINTAINING THE SAFETY OF THE
SET. BE SURE TO REPLACE THESE PARTS WITH
SPECIFIED ONES FOR MAINTAINING THE SAFETY AND
PERFORMANCE OF THE SET.
8 – 1
XG-C330X/C430X
[2] SCHEMATIC DIAGRAM
عMAIN Unit-1/11
,
+
*
)
(
'
&
%
$
#
8 – 2
XG-C330X/C430X
8 – 3
XG-C330X/C430X
,
#
$
%
&
'
(
)
*
+
عMAIN Unit-2/11
8 – 4
XG-C330X/C430X
8 – 5
XG-C330X/C430X
عMAIN Unit-3/11
,
+
*
)
(
'
&
%
$
#
8 – 6
XG-C330X/C430X
8 – 7
XG-C330X/C430X
عMAIN Unit-4/11
,
+
*
)
(
'
&
%
$
#
8 – 8
XG-C330X/C430X
8 – 9
XG-C330X/C430X
عMAIN Unit-5/11
,
+
*
)
(
'
&
%
$
#
8 – 10
XG-C330X/C430X
8 – 11
XG-C330X/C430X
عNAIN Unit-6/11
,
+
*
)
(
'
&
%
$
#
8 – 12
XG-C330X/C430X
8 – 13
XG-C330X/C430X
عNAIN Unit-7/11
,
+
*
)
(
'
&
%
$
#
8 – 14
XG-C330X/C430X
8 – 15
XG-C330X/C430X
عMAIN Unit-8/11
,
+
*
)
(
'
&
%
$
#
8 – 16
XG-C330X/C430X
8 – 17
XG-C330X/C430X
عMAIN Unit-9/11
,
+
*
)
(
'
&
%
$
#
8 – 18
XG-C330X/C430X
8 – 19
XG-C330X/C430X
عMAIN Unit-10/11
,
+
*
)
(
'
&
%
$
#
8 – 20
XG-C330X/C430X
8 – 21
XG-C330X/C430X
عMAIN Unit-11/11
,
+
*
)
(
'
&
%
$
#
8 – 22
XG-C330X/C430X
8 – 23
XG-C330X/C430X
عTERMINAL Unit-1/4
,
+
*
)
(
'
&
%
$
#
8 – 24
XG-C330X/C430X
8 – 25
XG-C330X/C430X
عTERMINAL Unit-2/4
,
+
*
)
(
'
&
%
$
#
8 – 26
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.