Declaration of Conformity
Manufacturer
Type
Corresponds to the requirements of the following EC directives
Electro Magnetic Compatibility Directive
Low Voltage Directive
The application harmonised standards are: EN60950-1:2006+A1:2009
EN61000-6-1:2007
EN61000-6-3:2007
A copy of the original signed declaration for this instrument is available
from the below UK address:
Avery Weigh-Tronix
Brecknell MS20
2004/108/EC
2006/95/EC
Electromagnetic Compatibility Statement for North America
This equipment has been tested and found to comply with the limits for a Class B
digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to
provide reasonable protection against harmful interference when the equipment is
operated in a residential install
radiate radio frequency energy and if not installed and used in accordance with the
instruction manual, may cause harmful interference to radio communications.
However, there is no guarantee that interference will not occur in a particular
installation. If this equipment does cause harmful interference to radio or television
reception which can be determined by turning the equipment off and on, the
encouraged to try to correct the interference by one of the following measures:
ation. This equipment generates, uses and can
user is
Re-orient or relocate the receiving antenna.
Increase the separation between the equipment and the receiver.
Connect the equipment to an outlet different from that to which the receiver is
connected.
Consult the dealer or an experienced radio / TV technician for help
Brecknell USA
1000 Armstrong Drive
Fairmont, MN 56031
Tel: 507-238-8702
Fax: 507-238-8271
Email: sales@brecknellscales.com
http://www.brecknellscales.com
IN-0787 3014H
Brecknell UK
Foundry Lane,
Smethwick, West Midlands
England B66 2LP
Tel: +44 (0) 845 246 6717
Fax: +44 (0) 845 246 6717
Email: sales@brecknellscales.co.uk
http://www.brecknellscales.co.uk
.
'S
USER
MANUAL
Model MS-20

SETTING UP YOUR MS-20 SCALE
To insure that your scale operates correctly, position the scale on a level sturdy
worktop.
The display head can be either attached to the wall or positioned in front of the
scale.
Pull down feet on back of display head enable the operator to change the
viewing angle of the display head.
Any excess cable can be stored in the cavity on the underside of the display
head.
POWER SUPPLY
The MS-20 uses 9V alkaline battery or
9V 100mA AC adapter.
CALIBRATION PROCEDURE
1) With the scale switched off, press and hold the “ON/OFF/TARE key, without
releasing press the kg/lb key three times and then release.
2) Dis pl a y will show zero (500 ~ 4000).
3) Press “ON/OFF/TARE” key once.
4) Display will show “CAL” and then “0”.
5) Load 20kg weight onto the scale.
6) Display will show span count (1250 ~ 3750)
7) Press “ON/OFF/TARE” key again.
8) Display will show “CAL” and then “20.00kg”.
9) Remove the weight from the scale and the calibration is finished
ERROR MESSAGES
CAUTION: always connect the AC
adapter to the MS-20 before
connecting to the mains power supply.
Disconnect from the mains power
supply before disconnecting the
adapter from the MS-20
KEY FUNCTION
ON/TARE (1) Press once to switch scale on.
OFF -
(2) Press while scale is switched on to
re-zero the scale.
(3) Press and hold this key for 2 seconds
to switch scale off. Scale will auto shut-off
if not used after 120 seconds
UNIT - Selects between kg mode and lb mode.
HOLD - When pressed the displayed weight will be
held on the display once the weight becomes
stable.
By repressing the Hold key again will release
the held display and allow the scale to return
back to normal weighing mode.
1. Low Battery
2. High zero counts
3. Low zero count
4. Overload
SPECIFICATIONS
Model # MS-20
Maximum Capacity 44 lb / 20 kg
Readability ½ oz / 10 g
Weight units lb / kg
Power supply 9V alkaline battery or 9V 100mA AC adapter
Dimensions 20” x 12” x 2 ½” / 505mm x 300mm x 65 mm
Display 1.0” LCD display with 4 active digits
Tray Plastic top cover
Operating Temperature
Function keys ON/TARE/OFF, UNIT, HOLD
5°C ~ 35°C