RKI Instruments GX-8000 User Manual

4 (2)

H4E-0050

Portable Gas Monitor

GX-8000

User Maintenance Manual

(H4-0050)

Need of Maintenance and Servicing

This gas monitor must be maintained in a normal state at all times to prevent accidents due to gas leaks. Daily and regular maintenance is required to keep the gas monitor in a normal state. Neglecting maintenance will result in failures and false alarms.

The specified maintenance procedures must be performed for the sake of maintenance of a normal state. Also, maintenance by the manufacturer (comprehensive maintenance) should be performed in addition to these maintenance specified in this manual.

This manual complements the Operating Manual (main manual). Be sure to read and understand thoroughly both this manual and the Operating Manual including precautions to ensure proper use of this gas monitor.

1 Maintenance Intervals and Items

1

Maintenance Intervals and

Items

Daily maintenance: Perform maintenance before beginning to work.

Monthly maintenance: Perform alarm test once a month.

Regular maintenance: Perform a maintenance once or more for every six months to maintain the performance as a safety unit.

Maintenance

Maintenance content

Daily

Monthly

Regular

item

 

maintenance

maintenance

maintenance

Battery level

Check that the battery level is sufficient.

{

{

{

check

 

 

 

 

Concentration

Make the gas monitor draw in fresh air. Check

 

 

 

display check

that the concentration display value is zero (or

 

 

 

 

20.9 vol% on the oxygen deficiency meter).

{

{

{

 

When the reading is incorrect, perform the zero

 

 

 

 

 

adjustment (fresh air adjustment) after ensuring

 

 

 

 

that no other gases exist around it.

 

 

 

Flow rate check

See the flow rate indicator to check for

{

{

{

 

abnormalities.

 

 

 

 

Filter check

Check the dust filter for dust or clogging.

{

{

{

Alarm test

Check the alarm lamp and buzzer for normal

 

{

{

 

operation by using the alarm test function.

 

 

 

 

 

Span adjustment

Perform the span adjustment by using the

 

 

{

 

calibration gas.

 

 

 

 

 

 

Gas alarm check

Check the gas alarm by using the calibration gas.

 

 

{

GX-8000

- 2 -

2 Regular Maintenance Mode

2

Regular Maintenance Mode

Regular maintenance mode is an operation (adjustment) mode in which span adjustment and setup such as change of alarm setpoints can be performed.

WARNING

After the adjustment is completed, do not forget to return to the detection mode.

(If the gas monitor remains in the regular maintenance mode, it does not automatically return to the detection mode.)

The regular maintenance mode comes in two types: Calibration and maintenance modes. To enter the calibration mode, press the DISPLAY switch with the T switch pressed in the detection mode and release them when the buzzer sounds. To enter the maintenance mode, press the POWER switch with the S and T switches pressed at power-on and release them when the buzzer sounds. The maintenance mode is password-protected. The input of a password is needed in the beginning.

- 3 -

GX-8000

2 Regular Maintenance Mode

The regular maintenance mode has the menus shown below.

Mode

Item

LCD display

Details

 

Calibration

Fresh air adjustment

AIR CAL

Perform the fresh air adjustment.

*

mode

 

 

 

 

Zero adjustment for

VOLZ.CAL

Perform the zero adjustment for

 

 

high-concentration

 

high-concentration combustible

 

 

combustible gases

 

gases.

 

 

Simultaneous span

AUTO CAL

Perform the span adjustment

**

 

adjustment for all channels

 

simultaneously for all channels.

 

 

Span adjustment for each

ONE CAL

Perform the span adjustment for

**

 

channel

 

each channel.

 

 

Bump test

BUMP

Perform the bump test.

**

 

Exit calibration mode

NORMAL

Exit the calibration mode and

**

 

 

 

return to the detection mode.

 

Maintenance

Date/time setting

DATE

Set the date and time.

 

mode

 

 

 

*

Fresh air adjustment

AIR CAL

Perform the fresh air adjustment.

 

Zero adjustment for

VOL Z.CAL

Perform the zero adjustment for

 

 

high-concentration

 

high-concentration combustible

 

 

combustible gases

 

gases.

 

 

Simultaneous span

AUTO CAL

Perform the span adjustment

 

 

adjustment for all channels

 

simultaneously for all channels.

 

 

Span adjustment for each

ONE CAL

Perform the span adjustment for

 

 

channel

 

each channel.

 

 

Bump test

BUMP

Perform the bump test.

***

 

Alarm setting

ALARM-P

Change the alarm setpoint value.

 

 

Bump test setting

BUMP-SET

Set the bump test.

***

 

Beep setting

BEEP SET

Set the confirmation beep

***

 

 

 

operation.

 

 

Exit maintenance mode

START

Exit the maintenance mode and

 

 

 

 

enter the detection mode.

 

*mark indicates items that can be operated (adjusted) in other modes such as [Detection mode] than the regular maintenance mode. Note that operation can be performed in any mode.

**mark indicates the menus in the calibration modes available in the overseas specification: AUTO CAL, ONE CAL, BUMP, and NORMAL.

***mark indicates the items not available in the Japanese specifications.

NOTE

If any mode that is not described in this manual or Operating Manual (main manual) (any mode but the regular maintenance mode) is entered by mistake, turn off the power once and try again.

GX-8000

- 4 -

RKI Instruments GX-8000 User Manual

2 Regular Maintenance Mode

<<Calibration Mode>>

In the detection mode, keep T pressed and press the DISPLAY switch. When the buzzer beeps, release the switches.

Calibration mode

AIR CAL

Perform the fresh air adjustment.

VOLZ.CAL

Perform the zero adjustment for high-concentration combustible gases.

AUTO CAL

Perform the span adjustment simultaneously for all channels.

(Set the span gas concentration in advance and make adjustment at once.)

ONE CAL

Perform the span adjustment for each channel. (While gas is supplied, perform span adjustment using UP/DOWN.)

BUMP

Perform the bump test.

NORMAL

Exit the calibration mode.

Fresh air adjustment => P16

Zero adjustment for highconcentration combustible gases

=> P20

Simultaneous span adjustment for all channels

=> P18

Span adjustment for each channel

=> P19

Bump test => P6

Detection mode

- 5 -

GX-8000

2 Regular Maintenance Mode

<Bump Test "BUMP">

Required equipment/material

• Bump test gas (collected in a gas sampling bag)

Connection

Connect the equipment as shown below to perform the bump test.

 

Gas sampling probe GAS IN

 

(with a tube)

Gas

 

sampling

GAS OUT

(Open)

bag

GX-8000 main unit

WARNING

About the bump test gas

The bump test gas is a hazardous gas (toxic, oxygen deficient, etc.). Handle the gas and related jigs and tools with due care (e.g., the gas must not be inhaled or the gas sampling bag must not have any hole).

About the place for span adjustment

Perform span adjustment where no silicon, organic solvent, spray can gases, etc. are used.

Perform span adjustment indoors at normal temperature without remarkable fluctuation (within ±5°C).

Perform span adjustment in an exhaust booth.

CAUTION

The GAS OUT side of the pipe must be left open without any pipe connected. A supplied gas must be discharged to a safe place.

GX-8000

- 6 -

2 Regular Maintenance Mode

BUMP

Press the ENTER switch.

The bump test gas concentration is displayed. The concentration of the prepared test gas must be consistent with the displayed test gas concentration.

To change the test gas concentration value, change the value in span adjustment menu. The test gas concentration is set to the same value as the span gas concentration. Simultaneous span adjustment for all channels

=> P18

Supply the test gas.

Supply the test gas and press ENTER. BUMP and APPLY are displayed alternately, and the countdown is started.

Go to (1) if all the channels are Pass in about 30 seconds. Go to (2) if any of the channels is Fail.

NOTE

Only on TYPE-A and TYPE-E, every time the S or T switch is pressed, the concentration of the span gas for high-concentration combustible gases is displayed alternately.

<If all the channels are Pass>

The result is displayed in about 30 seconds. Stop supplying the gas.

The results and values are displayed alternately every time the S or T switch is pressed.

P: Pass, F: Fail

After checking the result, press ENTER. The gas monitor returns to the calibration mode menu.

(1)

Return to

 

Stop supplying

BUMP.

the test gas.

- 7 -

GX-8000

Loading...
+ 16 hidden pages