WARNING: Product installation & operation
This equipment must be installed and operated in accordance
with the Raymarine instructions provided. Failure to do so could
result in personal injury, damage to your boat and/or poor product
performance.
WARNING: Electrical safety
Make sure you have switched off the power supply before you
start installing this product.
WARNING:
Although we have designed this product to be accurate and
reliable, many factors can affect its performance. Therefore, it
should serve only as an aid to navigation and should never replace
commonsense and navigational judgement. Always maintain a
permanent watch so you can respond to situations as they
develop.
EMC conformance
All Raymarine equipment and accessories are designed to the best industry
standards for use in the recreational marine environment.
The design and manufacture of Raymarine equipment and accessori es conform to
the appropriate Electromagnetic Compatibility (EMC) standards, but correct
installation is required to ensure that performance is not compromised.
Handbook information
To the best of our knowledge, the information in this handbook was correct when
it went to press. However, Raymarine cannot accept liability for any inaccuracies
or omissions it may contain.
In addition, our policy of continuous product improvement may change
specifications without n otice. Therefore, Raymarine cannot ac cept liability for any
differences between the product and the handbook.
Thank you for purchasing a Raymarine prod uct. We are sure your ST60 instrument
will give you many years of trouble-free operation.
This handbook describes how to install and use the Raymarine ST60 Rudder Angle
Indicator instrument. This instrument gives a continuously updated (real time)
display of your vessel’s rudder angle.
The ST60 Rudder Angle Indicator instrument is constructed in a rugged weather
proofed case. It provides a sensitive and stable analog display, to deliver accurate
information under even the most demanding conditions.
Data inputs
The instrument can be used either as a stand-alone unit, or as part of an
integrated SeaTalk instrumentation system.
SeaTalk
When connected to SeaTalk, the ST60 Rudder Angle Indicator can operate as a
repeater to any Raymarine SeaTalk autopilot, or provide rudder position
information to another rudder angle instrument.
SeaTalk enables a number of compatible instruments to be interconnected and
operate as a single, integrated navigational system.
Power and data in a SeaTalk system are fed via a single cable, so that instruments
be connected by plugging them into the network. SeaTalk is flexible enough to
adapt to any number of compatible instruments without requiring a central
processor. SeaTalk can also communicate with non-SeaTalk equipment, using the
internationally-accepted National Marine Electronics Association (NMEA)
protocol.
In a SeaTalk system, each instrument can be either a master or dedicated repeater
unit. A master instrument is directly connected to a transducer (the device that
provides the raw data) and has control of all other equipment on the SeaTalk
network. A slave instrument is not directly connected to a transducer but repeats
information provided by other equipment in the SeaTalk network.
Using with SeaTalk Autopilot
If the ST60 Rudder Angle Indicator is part of a system which includes a SeaTalk
autopilot, the rudder reference transducer is connected to the autopilot and the
Rudder Angle Indicator operates in repeater mode, i.e. as a slave display to the
autopilot, with the autopilot data displayed using the Rudder Angle Indicator
analog pointer.
Stand alone operation
In Stand alone operation, the ST60 Rudder Angle Indicator is connected only to a
rudder reference transducer and does not display information from, or provide
information to, any other instruments.
Options
Linear transducer
A Linear Rudder Reference Transducer option is available for use with ‘bullhorn’
type outboard installations.
Mounting
If you do not want to surface mount your ST60 instrument, options are available
for:
• Flush mounting. If you have ordered the flush mounting option a low-profile
bezel and four fixing screws are also provided.
• Bracket mounting.
Parts supplied
Unpack your ST60 instrument and check that the following items are present:
• Item 1, ST60 Rudder Angle instrument with standard bezel.
This handbook describes how to operate, maintain and install the Raymarine
ST60 Rudder Angle Indicator Instrument. The instrument provides a real-time
indication of rudder position as determined by the associated Rudder Reference
Transducer.
CAUTION: Calibration requirement
The ST60 Rudder Angle Indicator instrument is calibrated to
factory (default) settings when first supplied. To ensure optimum
performance on your boat, this product must be calibrated before
use. Do NOT use the product until it has been calibrated using the
procedures in
Displayed information
The ST60 Rudder Angle Indicator pointer shows the rudder position in real-time.
The instrument scale range gives an expanded indication from -40
about zero.
Operation with other instruments
When the ST60 Rudder Angle Indicator is connected to an operational autopilot
via SeaTalk, the instrument acts as slave to the autopilot, which receives data
from the Rudder Angle Indicator Transducer and passes the information via the
SeaTalk bus.
Chapter 4, Calibration
.
° to +40°
1.2Normal use
As the ST60 Rudder Angle Indicator operates in real time, very little operator
action is necessary during normal use except to observe the instrument as
required to ascertain the rudder angle.
Note:
If rudder angle information is not available on the SeaTalk bus, the analog pointer
oscillates ±10° about the top of the dial.
The key is the only key used during normal operation (see
illustration). The
Chapter 4, Calibration
unless otherwise stated.
Key functions
set, < and > keys are reserved for calibration mode, see
. All key presses called for in this handbook are momentary,
When the instrument is first powered up, the display illumination is set to its
lowest (courtesy) level, to facilitate initial access to the keys.
To adjust the level of display illumination:
1. Hold down the key for approximately 1 second, to enter the illuminationadjust mode.
2. There are four preset illumination levels from a courtesy level to full brightness. Momentarily press the key to cycle through these levels until you
reach the level you want. The selected level is transmitted to all other instruments on the SeaTalk bus.
3. Press any other key to leave the illumination-adjust mode.
Note:
The display will return to normal operation 7 seconds after the last key press.
1.3Remote control
The ST60 Rudder Angle Indicator does not support the SeaTalk remote control
facility. You can only control the instrument with the front panel keys.
Page 13
Chapter 2: Maintenance & Troubleshooting3
Chapter 2:Maintenance & Troubleshooting
2.1Maintenance
Servicing and safety
• Raymarine equipment should be serviced only by authorized Raymarine service technicians. They will ensure that service procedures and replacement
parts used will not affect performance. There are no user serviceable parts in
any Raymarine product.
• Some products generate high voltages, so never handle the cables/connectors when power is being supplied to the equipment.
• When powered up, all electrical equipment produces electromagnetic fields.
These can cause adjacent pieces of electrical equipment to interact with one
another, with a consequent adverse effect on operation. In order to minimize
these effects and enable you to get the best possible performance from your
Raymarine equipment, guidelines are given in the installation instructions, to
enable you to ensure minimum interaction between different items of equipment, i.e. ensure optimum Electromagnetic Compatibility (EMC).
• Always report any EMC-related problem to your nearest Raymarine dealer.
We use such information to improve our quality standards.
• In some installations, it may not be possible to prevent the equipment from
being affected by external influences. In general this will not damage the
equipment but it can lead to spurious resetting action, or momentarily may
result in faulty operation.
Instrument
Certain atmospheric conditions may cause condensation to form on inside of the
instrument window. This will not harm the instrument and can be cleared by
increasing the instrument illumination to the brightest setting.
Periodically clean your ST60 instrument with a soft damp cloth. Do NOT use
chemical and/or abrasive materials to clean the instrument.
Transducer
Refer to the Installation and Maintenance instructions supplied with the
transducer.
Examine all cables for chafing or other damage to the outer shield and, where
necessary, replace and re-secure.
2.2Troubleshooting
Preliminary procedures
Changes in the electronic environment may adversely affect the operation of your
ST60 equipment. Typical examples of such changes are:
• Electrical equipment has rec ently been installed or moved aboard your vessel.
• You are in the vicinity of another vessel or shore station emitting radio signals.
If you appear to have a problem, first ensure that the EMC requirements (see
Chapter 3, Installation
problem.
) are still being met before further investigating the
2.3Fault location
All Raymarine products are subjected to comprehensive test and quality
assurance programmes prior to packing and shipping. If a fault occurs, the
following table may help to identify and rectify the problem.
FaultCauseRemedy
No rudder angle informationNo power supplyCheck power supply.
Check SeaTalk cabling and
connector security.
Check fuse/circuit breaker.
No transfer of information
between instruments or a group
of instruments (e.g. illuminati on
levels).
SeaTalk cabling faultCheck security of SeaTalk
connectors.
Check condition of SeaTalk
cables.
Isolate faulty instrume nt by
disconnecting instruments
one by one.
Page 15
Chapter 2: Maintenance & Troubleshooting5
Technical support
Raymarine provides a comprehensive customer support service, on the world
wide web and by telephone help line. Please use either of these facilities if you are
unable to rectify a problem.
World wide web
Please visit the Customer Support area of our web site at:
• www.raymarine.com
As well as providing a comprehensive Frequently Asked Questions section and
servicing information, the web site gives e-mail access to the Raymarine Technical
Support Department and a details of the locations of Raymarine agents,
worldwide.
Telephone help line
If you do not have access to the world wide web, please call our help line.
In the USA, call:
• +1 800 539 5539, extension 2444 or
• +1 603 881 5200 extension 2444
In the UK, Europe the Middle East or the Far East, call:
• +44 (0) 23 9271 4713 (voice)
• +44 (0) 23 9266 1228 (fax)
Help us to help you
When requesting service, please quote the following product information:
• Equipment type.
• Model number.
• Serial number.
• Software issue number.
The Software issue number can be ascertained by means of the Intermediate
Calibration facility, see
This chapter describes how to install the ST60 Rudder Angle Indicator instrument
and associated transducer.
Rotary and linear Raymarine Rudder Angle Transducers are available for use with
the ST60 Rudder Angle Indicator instrument.
If the instrument is being installed as stand-alone, the transducer cable is
connected to the rear of the instrument. If being installed as part of a system
incorporating a SeaTalk Autopilot, the transducer is connected to the Autopilot.
For advice, or further information regarding the installation of this equipment,
please contact the Raymarine Product Support Department or your own National
Distributor.
3.1Planning your installation
Before starting the installation, spend some time considering the best positions
for the instrument, such that the
guidelines
are satisfied.
Site requirements
Site requirements
and the
EMC installation
Transducer
The siting of the Rudder Reference Transducer depends on the type of vessel.
Fitting details and dimensions are given in the
chapter.
Fitting transducers
section of this
Instrument
CAUTION:
The presence of moisture at the rear of the instrument could cause
damage either by entering the instrument through the breathing
hole or by coming into contact with the electrical connectors.
ST60 instruments can be fitted either above or below deck, provided that the rear
of the instrument is sited where it is protected from contact with water.
Each instrument must also be positioned where it is:
• Easily read by the helmsman.
• Protected against physical damage.
• At least 9 in (230 mm) from a compass.
• At least 20 in (500 mm) from radio receiving equipment.
• Reasonably accessible from the rear for installation and servicing.
All Raymarine equipment and accessories are designed to the best industry
standards for use in the recreational marine environment.
Their design and manufacture conforms to the appropriate Electromagnetic
Compatibility (EMC) standards, but correct installation is required to ensure that
performance is not compromised. Although every effort has been taken to ensure
that they will perform under all conditions, it is important to understand what
factors could affect the operation of the product.
The guidelines given here describe the conditions for optimum EMC performance,
but it is recognized that it may not be possible to meet all of these conditions in all
situations. To ensure the best possible conditions for EMC performance within the
constraints imposed by any location, always ensure the maximum separation
possible between different items of electrical equipment.
For optimum EMC performance, it is recommended that wherever possible:
• Raymarine equipment and cables connected to it are:
• At least 3 ft (1 m) from any equipment transmitting or cables carrying
radio signals e.g. VHF radios, cables and antennas. In the case of SSB
radios, the distance should be increased to 7 ft (2 m).
Page 19
Chapter 3: Installation9
• More than 7 ft (2 m) from the path of a radar beam. A radar beam can nor-
mally be assumed to spread 20 degrees above and below the radiating
element.
• The equipment is supplied from a separate battery from that used for engine
start. Voltage drops below 10 V in the power supply to our products, and
starter motor transients, can cause the equipment to reset. This will not damage the equipment, but may cause the loss of some information and may
change the operating mode.
• Raymarine specified cables are used. Cutting and rejoining these cables can
compromise EMC performance and must be avoided unless doing so is
detailed in the installation manual.
• If a suppression ferrite is attached to a cable, this ferrite should not be
removed. If the ferrite needs to be removed during installation it must be reassembled in the same position.
Suppression ferrites
The following illustration shows typical cable suppression ferrites used with
Raymarine equipment. Always use the ferrites supplied by Raymarine.
Connections to other equipment
If your Raymarine equipment is to be connected to other equipment using a cable
not supplied by Raymarine, a suppression ferrite MUST always be attached to the
cable near the Raymarine unit.
3.2Procedures
As it is not possible to describe procedures for all possible installation scenarios,
the procedures given here describe the broad requirements for installing an ST60
Rudder Angle Indicator instrument and its associated transducer. Adapt these
procedures as appropriate, to suit your individual requirement.
CAUTION:
Where it is necessary to cut holes (e.g. for cable routing and
instrument mounting), ensure that these will not cause a hazard
by weakening critical parts of the vessel’s structure.
Unpacking
Unpack your ST60 instrument and check that the items detailed in the
Introduction to this handbook are present.
Each ST60 instrument is supplied with a standard bezel for surface mounting.
Optional mounting kits are available for flush mounting and bracket mounting
the instrument. If you have ordered the flush mounting option, a low-profile bezel
and four fixing screws are also provided.
Fitting the instrument
The ST60 Rudder Angle Indicator instrument can be installed using one of a
number of different mounting options:
• Surface Mounting. Gives a profile of approximately 0.95 in (24 mm).
• Flush Mounting. Gives a profile of approximately 0.25 in (6 mm)
• Bracket Mounting.
The ST60 Rudder Angle Indicator instrument can also be mounted behind a panel
with just the instrument dial and keys visible.
Surface mounting
To surface mount your ST60 instrument (see the
1. Ensure that:
• The selected location is clean, smooth and flat.
• There is sufficient space behind the location to accommodate the rear of
the instrument and connectors.
2. Apply the surface mount template (supplied at the rear of this handbook) to
the selected location and mark the centers for the fixing studs (1) and the
aperture (3) that will take the rear casing of the instrument.
3. Drill out the two 0.2 in (5 mm) fixing stud clearance holes (2).
4. Cut out the clearance hole (3) then remove the template.
5. Peel off the protective sheet from the self-adhesive gasket (4) then stick the
gasket into position on the rear of the instrument.
6. Screw the two fixing studs into the threaded sockets on the rear of the instrument.
Surface mounting
illustration):
Page 21
Chapter 3: Installation11
7. Mount the assembled instrument, studs, bezel and gasket into the panel.
Secure from behind with the thumb nuts (5).
412135 2 5
Surface mounting
D4331-2
Flush mounting
The Flush Mounting Kit uses a low-profile bezel to reduce the fitted profile of the
instrument to approximately 0.25 in (6 mm) above the panel fascia.
Fitting the low-profile bezel
In order to flush-mount your ST60 instrument, replace the standard bezel with the
low-profile bezel as follows:
1. Hold the instrument in both hands with the display towards you.
2. Using both thumbs, gently press an upper corner of the instrument from the
bezel, then remove the bezel from the instrument. Retain the rubber keypad
which is released when the bezel is removed.
3. Place the instrument face upwards on a flat surface and place the rubber keypad (7) in position around the display window (i.e. so that each key outline is
located over its associated key on the instrument).
4. Snap the low-profile bezel (8) in position over the instrument, so that the rubber keys are correctly located in the holes on the bezel.
9
D4358-2
Fitting the low profile bezel
8
7
CAUTION:
It is essential that only screws of the correct size are used to
secure the instrument to the bezel. Failure to observe this caution
could result in damage to both the instrument and the bezel.
5. Using the four, self-tapping screws (9) provided, secure the instrument and
bezel together. Fit the screws from the rear of the instrument and tighten
them sufficiently to secure the instrument and bezel together. DO NOT OVERTIGHTEN.
Page 23
Chapter 3: Installation13
Flush mounting procedure
Flush mount your instrument (see the
1. Assemble the ST60 instrument and low-profile bezel as described under
ting the low-profile bezel
.
Flush mounting
illustration) as follows:
Fit-
2. Ensure that:
• The panel on which you intend to mount the instrument is between
0.12 in (3 mm) and 0.78 in (20 mm) thickness.
• The selected location is clean, smooth and flat.
• There is sufficient space behind the location to accommodate the rear of
the instrument and connectors.
1315 65
Flush mounting
4
D4332-3
3. Apply the flush mount template (supplied at the rear of this handbook) to the
selected location and mark out the aperture into which the assembled instrument and bezel will sit.
4. Cut out the aperture (3) for the assembled instrument and bezel and remove
the template.
5. Peel off the protective sheet from the self-adhesive gasket (4) then stick the
gasket into position on the rear of the bezel.
6. Screw the two fixing studs (1) into the threaded sockets on the rear of the
instrument.
7. Mount the assembled instrument, studs, bezel and gasket into the panel.
8. Locate the flush mount bracket (6) onto the fixing studs and secure the assembly to the panel with the thumb-nuts (5).
Bracket mounting
A Control Unit Mounting Bracket (Part No. E25009) enables you to mount your
ST60 instrument in locations where other forms of mounting are impractical.
Although bracket mounting provides a useful alternative method for securing
your instrument, it is only suitable for use in positions where the instrument will
not be exposed to water.
To bracket mount your ST60 instrument, do so in accordance with the Control Unit
Mounting Bracket Instruction Sheet.
Fitting transducers
The Rudder Reference Transducer duplicates movement of the tiller arm to
provide the instrument or the autopilot as appropriate, with the exact position of
the vessel’s rudder.
Rotary and Linear Rudder Reference Transducers are available. The Rotary type is
suitable for most installations, but the Linear type is designed for use with
‘bullhorn’ type outboard installations. Refer to the fitting instructions for either
the
Rotary Rudder Reference Transducer
Transduc er
, as appropriate.
or the
Linear Rudder Reference
If you are fitting an ST60 Rudder Angle Indicator instrument as part of a system
which includes a SeaTalk autopilot, the transducer must be connected directly to
the autopilot. The ST60 Rudder Angle Indicator will then act as a slave to the
autopilot.
Rotary Rudder Reference Transducer
152mm (6in)
139.7mm (5.5in)
61mm (2.4in)
69.5mm (2.7in)
Rotary Rudder Reference Transducer - dimensions
D4426-1
Page 25
Chapter 3: Installation15
Mounting
1. Ensure that the base height of the Rotary Rudder Reference Transducer is such
that it is able to maintain the correct vertical alignment between the transducer and tiller arm (as shown).
Tiller arm
Vertical alignment of transducer and tiller arm
Transducer
Mounting base
D4427-1
2. Using Rotary Rudder Reference Transducer template (supplied at the rear of
this handbook), mark the centers for the three securing screws at a suitable
location adjacent to the rudder stock.
3. Drill out the three 3 mm holes then, using the three self-tapping screws provided, secure the Rotary Rudder Transducer in position.
Note:
To ensure precise rudder position indication, the Rotary Rudder Reference Trans-
ducer incorporates a spring to remove any mechanical backlash in the tiller linkage.
Connecting to the tiller
CAUTION:
The Rotary Rudder Reference Transducer arm movement is
limited to ±60°. Care must be taken during installation to make
sure that the arm is opposite the point of cable entry when the
rudder is amidships. Failure to do this could result in damage if the
arm is driven onto its end stops by the steering system.
When connecting the Rotary Rudder Reference Transducer to the tiller, ensure this
is done in accordance with the limits shown in the
Transducer - as fitted
illustration (below), and that the tiller and arm of the Rotary
Rotary Rudder Reference
Rudder Reference Transducer are parallel to each other.
1. With the rudder amidships, ensure that the Rotary Rudder Reference Transducer arm is opposite the point of cable entry and at 90° to the connecting
bar. If any minor adjustment is necessary, loosen the three securing screws,
rotate the Rotary Rudder Reference Transducer body to achieve the required
adjustment, then re-tighten the screws.
± 60
maximum
travel
permitted
Min
75mm
(3in)
Max
310mm
(12.2in)
Parallel
A
Cable
entry
40
max
90
Rudder amidships
Rotary Rudder Reference Transducer - as fitted
40 max
Min 101mm (4in)
'A' 140mm (5.5in)
Max 190mm (7.5in)
D4428-1
2. Position the tiller pin within the limits shown. Ideally, dimension ‘A’ should be
5.5 in (140 mm). Secure the tiller ball-pin to the tiller arm using the two selftapping screws provided.
Note:
Variation within the given limits will not degrade autopilot performance but will
slightly alter the scaling of the ST60 Rudder Angle Indicator display.
3. Cut the threaded rod to length and screw one lock-nut and ball socket onto
each end of the rod.
4. Offer up the rod (with ball sockets) to the Rotary Rudder Reference Arm and
tiller, making adjustments to the ball sockets as required then, supporting the
ball sockets (by means of a suitable wrench) tighten the locking nuts.
5. Press each socket onto its corresponding ball-pin.
6. Move the rudder from side to side and ensure that the linkage is free from any
obstruction at all rudder angles.
Note:
If the Rotary Rudder Reference Transducer has to be mounted in an inverted position, the connections for the red and green wires at the ST60 Rudder Angle Indicator instrument (or course computer) must be reversed.
Page 27
Chapter 3: Installation17
Linear Rudder Reference Transducer
The Linear Rudder Reference Transducer is designed for use with ‘bullhorn’ type
outboard installations.
2
78 39
1110
Fitting the Linear Rudder Reference Transducer
564121
D4429-1
Mounting
1. Operate the steering system so that the ‘bullhorn’ ram (1) is positioned amidships.
2. Release the hydraulic pressure from the vessel’s hydraulic steering system (if
required). Refer to the manufacturer’s instructions for correct procedures.
3. Loosen the starboard bolt that secures the ‘bullhorn’ ram (1) shaft to the end
bracket (2).
4. Assemble the U-bracket (3) over the end bracket (2) and the shaft of the ’bull-
horn’ ram (1).
5. Hand tighten the starboard ‘bullhorn’ bolt to hold the U-bracket (3) in position.
6. Fully open the hose clamps (6) using a flat-blade screwdriver.
7. Hang the hose clamps (6) over the ‘bullhorn’ ram (1).
8. Site the spacers (4) on the ‘bullhorn’ ram (1) and hold, temporarily, with adhesive tape.
9. Pull the shaft (9) out of the linear feedback transducer (5) until the alignment
mark (10) is level with the end of the body (11).
10. Position the linear feedback transducer (5) on top of the spacers (4) so that the
threaded end of the shaft passes through the U-bracket (3).
The linear feedback transducer shou ld, under normal circumstances, be assembled
with the shaft (9) pointing to starboard. However, if it is not possible to orientate the unit in
this way, port installation is possible provided that the red and green wires are reversed at
the ST60 Rudder Angle Indicator instrument (or course computer).
11. With the adjustment screw and barrel aligned with the spacers, close the hose
clamps (6) around the linear feedback transducer (5) and the ‘bullhorn’ ram
(1).
12. Tighten the ‘bullhorn’ bolt to retain the U-bracket (3).
13. Fit and tighten the nut (7) and washer (8) to the shaft of the linear feedback
transducer (5).
Transducer cabling
General
Each transducer type is supplied with sufficient cable to run from the mounted
position to either the ST60 Rudder Angle Indicator instrument (for operation as a
master) or to an autopilot (for operation as a repeater).
If you are connecting for repeater operation, use the instructions in your autopilot
handbook to connect the transducer to the autopilot.
Running the cable
The manner in which you run the cable will depend on the locations of the
transducer and instrument (or autopilot). The following guidelines are provided:
• When running cable from a Linear Rudder Reference Transducer, leave a loop
of cable at the end of the Linear Rudder Reference Transducer, sufficient to
allow for movement of the ‘bullhorn’.
• If the cable has to be fed through the deck, always use a proprietary deck
gland.
• Where cables are fed through holes, always use grommets to prevent chafing.
• Secure long cable runs so they do not present a hazard.
• Although the transducer cable is fitted with spade connectors for direct con-
nection to the rear of the instrument, it may be necessary to remove these to
facilitate installation e.g. if you want incorporate a junction box in the cable
run or if the cable has to be routed through narrow apertures. Extra spade
connectors are provided, to replace any that are removed when running the
cable. When fitting spade connectors, prepare the cable as at (a) in the following illustration, then fold back the wire strands and insert into the spade connector as at (b). Ensure the wire strands do not extend beyond the rear of the
spade connector insulation, then crimp the connector to the wire.
Page 29
Chapter 3: Installation19
50 mm
6 mm
(a)
3 mm
(b)
D4467-6
Connecting the instrument
Introduction
The ST60 Rudder Angle Indicator instrument can be connected:
• To the transducer as a stand-alone instrument.
• To SeaTalk .
Instruments connected to SeaTalk derive their power directly from SeaTalk and no
separate power connection is necessary. Where a SeaTalk system includes an
autopilot, the power for the system is provided by the autopilot.
A range of Raymarine SeaTalk extension cables is available to connect separated
instruments. These cables are supplied with a SeaTalk connector fitted to each
end. A junction box can be used to join cables.
Signal connections
Make the necessary connections to your ST60 instrument (see the
ST60 Rudder Angle Indicator instrument
illustration).
Connections to
SeaTalk cable
Note:
If the ST60 Rudder Angle
Indicator pointer moves in
the wrong direction with
respect to the rudder
movement, swap the Red
and Green connections.
This may be necessary if
the Rudder Reference
Transducer is fitted in a
non-standard manner.
Cable from transducer
Connections to ST60 Rudder Angle Indicator instrument
When instruments are connected to SeaTalk, ensure that the
power supply for the SeaTalk 12 V line is protected by a 5 A fuse.
Systems with a large number of instruments on the SeaTalk bus may require
connections to the power supply from each end of the system (‘ring-main’ style),
to maintain sufficient voltage throughout the system.
Red
5 A fused,
12 V dc supply
(typically provided
by autopilot)
Screen
Red
Screen
1234
Instruments
5 to 16
20
19
18
17
SeaTalk power connections
D4311-1
This requirement depends on the total length of the cable run and the total
number of instruments in the system, as follows:
Cable runNo. of instrumentsPower connections
Up to 10 m13 maximum
26 maximum
Up to 20 m7 max imu m
13 maximum
1
2
1
2
Stand alone instruments
Stand-alone instruments are not connected to SeaTalk and therefore need to be
connected to an alternative 12 V power source. Power cables are available in 2 m
and 9 m lengths.
Page 31
Chapter 3: Installation21
To fit a power cable:
1. Ensure the intended power source is switched off.
2. Run the power cable from the instrument to a suitable 12 V dc power source.
3 A over-current
circuit breaker
12 V dc
supply
Red
Screen
Power connections for stand-alone instrument
D4310-6
3. If the cable has not already been trimmed at the power supply end:
i. Cut the cable to length and trim back an appropriate amount of the outer
sheath.
ii. Cut back and insulate the yellow wire.
4. Connect the screen to the power supply 0 V terminal.
5. Connect the red wire via a 3 A circuit breaker to the power supply +12 V terminal.
6. Insert the power cable connector into one of the SeaTalk connectors at the
rear of the instrument.
Before using the Rudder Angle Indicator instrument for navigational purposes, it
must be aligned with the Rudder Reference Transducer to counter any offset and
ensure that the rudder position is accurately monitored.
EMC conformance
• Always check the installation before going to sea to make sure that it is not
affected by radio transmissions, engine starting etc.
• In some installations, it may not be possible to prevent the equipment from
being affected by external influences. Although this will not damage the
equipment, it can lead to sp urious resetting action, or momentarily may result
in faulty operation.
4.2Rudder angle offset adjustment
Rudder angle offset adjustment removes any deviation between the reading on
the ST60 Rudder Angle Indicator display and the Rudder Reference Transducer. To
adjust for any offset, proceed as follows:
1. Center the rudder and check that the display analog pointer is within ±7° of
zero. If the analog pointer reading exceeds the limits ±7°, the Rudder Reference Transducer should be realigned in order to avoid scale inaccuracies, see
Chapter 3, Installation
2. With the rudder centered, press and hold the
ment will emit an audible beep to indicate that you are in calibration mode.
Note:
You can enter calibration mode only when the instrument is in master mode. If the
instrument is in repeater mode and you attempt to enter calibration mode, the instrument
will beep twice when the
3. Use the < (decrement) and/or > (increment) keys to adjust the Rudder Angle
Indicator display pointer to the zero position in 1° steps, or press and hold
either
< or > for rapid change.
4. Press the
mode.
Note:
is exited without change.
set key for 2 seconds to save the new setting and exit calibration
If, during calibration, no key is pressed for a period of 60 seconds, calibration mode
.
set key f or 2 seconds. The instru -
set
key is released, to indicate an invalid key press.