Philips IntelliVue MP90 Service Manual

Page 1
IntelliVue MP90
Service Guide
IntelliVue Patient Monitor
MP90
Patient Monitoring
Page 2
Part Number M8000-9351B
Page 3
1Table of Contents
Who Should Use This Guide 13 How to Use This Guide 13 Abbreviations 13 Responsibility of the Manufacturer 14 Passwords 14 Warnings and Cautions 14
Monitor Theory of Operation 15
System Boundaries 15 Hardware Building Blocks 16
IntelliVue MP90 17 Optional Hardware 18 Compatible Devices 18 Power Supply 19 CPU Boards 19 I/O Boards 20
Data Flow 20
Data Acquisition 20 Data Provider System Service 21 Persistent Data Storage System Service 21 Display and User Interface Service 21 Data Output 21 Monitor Applications 22 Internal LAN (Measurement Server Link) 22 Philips Clinical Network 23
How does the Support Tool Work with the Monitor 23 Monitor Software Block Diagram 24 Block Diagram Legend 26
Concepts 31 Test Reporting 31 Recommended Frequency 32 Tests Recommended When Performing... 33
Installation 33 Repair 33 Preventive Maintenance 33
3
Page 4
Performance Verifications 33 Upgrades 34
Te s t s 34
Visual Test 34 Power On Test 34 NBP Tests 34
NBP Accuracy Test 35 NBP Leakage Test 36 NBP Linearity Test 36 Valve Test 36
Sidestream CO2 Per formance Test 37
Barometric Pressure Check and Calibration 37 Leakage Check 38 Pump Check 39 Flow Rate Check and Calibration 39 Noise Check 39 CO
Gas Measurement Calibration Check 39
2
Calibration Verification 40 Reset Time Counters 40
Temperature Accuracy 41 ECG/Resp Performance Test 42
ECG Performance 42 Respiration Performance 42
Invasive Pressure Performance Test 42 SpO2 Performance Test 42 Cardiac Output Performance Test 43
Service Tool Procedure, Version 1 43 Service Tool Procedure, Version 2 43
BIS Performance Test 43
PIC/DSC Test 43
Nurse Call Relay Performance Test 44
Phone Jack Type Connector Test 44 Multi-Port Nurse Call Connector Test 44
ECG Sync Performance Test 45 VueLink Tests using VueLink Test Module 46
Tes t Pr o c e d u r e 4 6
Safety Testing 47
Warnings, Cautions, and Safety Precautions 47 Safety Test Procedures 47
Touchscreen Calibration 50
Introduction 51 How To Use This Section 51 Who Should Perform Repairs 51 Replacement Level Supported 51
4
Page 5
Software Revision Check 52 Obtaining Replacement Parts 52 Troubleshooting Guide 52
Checks for Obvious Problems 52 Checks Before Opening the Instrument 52
Checks with the Instrument switched Off 52 Checks with the Instrument Switched On, AC connected 53 Initial Instrument Boot Phase 53
Troubleshooting Tables 54
How to use the Troubleshooting tables 54 Boot Phase Failures 55 External Display is blank 58 External Touch Display not functioning 58 General Monitor INOP Messages 59 Remote Alarm Device 60 Remote Extension Device 60 Keyboard/Mouse not functioning 61 Network related problems 62 Wireless Network 63 Multi-Measurement Server 64 MSL-related problems 64 Alarm Issues 66 Alarm Lamps 66 Alarm Tones 67 Alarm Behavior 67 Individual Parameter INOPS 67 Flexible Module Server 68 Printer 69 Recorder 70 MIB / RS232 70 Flexible Nurse Call Relay 71 Troubleshooting the ECG OUT 72 Data Flow Marker In and ECG Wave 72
Status Log 73 Troubleshooting with the Support Tool 74 Troubleshooting the Individual Measurements or Applications 75
Tools Required 77 MP90 CMU Disassembly 77
Removing I/O Boards 77 Removing the Top Cover 79 Removing the Plastic Feet and/or the Locking Cam 81 Removing the optional Fans (Dual CPU Versions only)* 82 Replacing the Second (Independent) Video Board
(Dual CPU Versions only)
83
5
Page 6
Removing the Second CPU/Main Board (Dual CPU Versions only) 84 Accessing the Main CPU or Primary Video Board (Dual CPU Versions) 86 Replacing the Primary Video Board 88 Removing the Main Board 90 Removing the Power Supply 93 Removing the Speaker 95 Removing the Power On/Off Switch 95
Flexible Module Server (FMS) Disassembly 97
Removing the Handle and the Measurement Server Mount 97
Plug-in Modules 102
Plug-In Module Disassembly 102 tcpO2/tcpCO2 Calibration Chamber Kit 103 Recorder Module Paper 104
Disassembly Procedures for the Measurement Server Extension 105
Removing the Front Cover 105 Removing the Extension Bottom Cover 106 Removing the CO2 Scrubber 107 Removing the Pump 108
Refit Procedures for the Measurement Server Extension 108
Refitting the CO2 Scrubber 109 Refitting the Pump 109 Refitting the Extension Bottom Cover 109 Refitting the Front Cover 110 General Reassembly/Refitting Comments 110 Following Reassembly 110
6 Parts 111
MP90 Parts 112
Exchange Parts 112 Replacement Parts 113
Flexible Module Server Parts 114
Exchange and Replacement Parts 114
Multi-Measurement Server Parts 115 Measurement Server Extension Parts (M3015A and M3016A) 117
Exchange Parts List 119
Plug-in Modules Part Numbers 119
Part Number Table 119
Exchange Modules, Table 1 120 Exchange Modules, Table 2 120
Plug-In Modules Replaceable Parts 121
Single-Width Plug-In Module 122 Double-Width Plug-In Module 122 Plug-in Module Replaceable Parts 123 Plug-In Module Language Specific Front Housings, Table 1 123 Plug-In Module Language Specific Front Housings, Table 2 123 Plug-In Module Specific Bezels 124
6
Page 7
BIS Module Replaceable Parts 124
BIS Module Components 125
tcpO2/tcpCO2 Module Accessories 125
External Display Part Numbers 126 Remote Input Devices Part Numbers 127 Remote Alarm Device Part Numbers 128 Remote Extension Device Part Numbers 129
7 Installation Instructions 131
Unpacking the Equipment 131 Initial Inspection 132
Mechanical Inspection 132 Electrical Inspection 132 Claims For Damage and Repackaging 132
Claims for Damage 132 Repackaging for Shipment or Storage 132
Installing the Monitor (M8010A) 132
Mounting Instructions 132
Assembling Mounts 132
Connections 133 Installing Interface Boards 134 Connection of MIB Devices 135
Installing Remote Devices 136
Mounting the Remote Display (M8031A) 136
Connections 136
Mounting the 17” Remote Display (M8033A) 137 Flexible Module Server and/or Multi-Measurement Server 137
Positioning the Measurement Server on a Clamp Mount 137 Mounting the MMS Mount to the FMS (M8048A) 137 Attaching the MMS to a Mount 138 Detaching the Measurement Server from a Mount 138 Mounting the Remote Extension Device to the FMS 139 Mounting the BIS Engine to the FMS 139 Mounting the FMS 140 Connections 141 MSL Cable Termination 141
Remote Alarm Device 143
Mounting 143 Connections 143
Remote Extension Device 144
Mounting 144 Connections 145 Cabling 145
Keyboard/Mouse 146
Philips Clinical Network (wired) 146
7
Page 8
Philips Clinical Network (wireless) 146 Flexible Nurse Call Relay 146
Connections 146
ECG Out Functionality 147
Connections 147
Configuration Tasks 147
Setting Altitude and Line Frequency 148 Configuring the Equipment Label 148 Configuring the printer 148 Setting the Display Resolution (M8033A only) 148
8 Site Preparation 149
Introduction 149
Site Planning 149 Roles & Responsibilities 149
Site Preparation Responsibilities 149 Procedures for Local Staff 150 Procedures for Philips Personnel 151
Monitor M8010A Site Requirements 151
Space Requirements 151 Environmental Requirements 152
Te m p e r at u r e 15 2 Humidity 152 Altitude 152
Electrical and Safety Requirements (Customer or Philips) 152
Safety Requirements 152 Electrical Requirements 153
Remote Device Site Requirements 153
Multi-Measurement Server M3001A or Flexible Module Server M8048A 154
Space Requirements Multi-Measurement Server M3001A 154 Space Requirements Flexible Module Server M8048A 154 Environmental Requirements Multi-Measurement Server M3001A 155 Environmental Requirements Flexible Module Server M8048A 155 Cabling Options and Conduit Size Requirements 155
Remote Displays - M8031A 156
Space Requirements 156 Environmental Requirements 156 Electrical and Safety Requirements 157
Remote Displays - M8033A 157
Space Requirements 157 Environmental Requirements 157 Electrical and Safety Requirements 157 Cabling Options and Conduit Size Requirements 158
Remote Alarm Devices 158
Space Requirements 158
8
Page 9
Cabling Options and Conduit Size Requirements 158
Remote Extension Device 159
Space Requirements 159 Cabling Options and Conduit Size Requirements 159 Input Devices 159
Local Printer 160
Philips Medical LAN 160 MIB Interface 160 Flexible Nurse Call Relay Interface 161 ECG Out Interface 161
9 Anesthetic Gas Module 163
Introduction 163
Description 163 Product Structure 163 Physical Specifications 163 Environmental Specifications 164 Performance Specifications 164
CO2 Measurement 165 AWRR derived from CO2 Waveform 165 N2O Measurement 165 O2 Measurement 165 Anesthetic Agent Measurement 165 Alarm Ranges 166 Alarm Delay 166 Apnea Alarm 166 INOP Alarms 166
General Measurement Principles 167 Theory of Operation 167
Main PC Board 168 Power Supply 169 Pneumatic System 169 Pump 170 Wate rt rap 170 Sample Flow Through the Pneumatic Path 171
Agent Identification Assembly 171
Measurement Principle 172
O2 Sensor 172
Specifications 172 Measurement Principle 172
Infrared Measurement Assembly 173
Installation and Patient Safety 174
Physical Installation 174 Environment 175 Label Sheet 175 Making Connections to the AGM 175
9
Page 10
Sample Gas Connections to the Gas Exhaust 176
Returning the Gas Sample 176 Setting Up the Gas Return 177 Removing the Gas Sample 178
Setup and Configuration Procedures 178
Altitude Configuration 178 Connect Sample Input Tubing 178
Preventive Maintenance (PM) Tasks 178 Post-Installation Checks 179 Safety Requirements Compliance and Considerations 179
Explanation of Symbols Used 179 Power Supply Requirements 180 Grounding the System 180 Equipotential Grounding 181 Combining Equipment 181
Checking and Calibrating the Anesthetic Gas Module 181
Access Service Functions of the M1026A Anesthetic Gas Module 181 When and how to check the Philips M1026A Anesthetic Gas Module 183 Equipment required for checking 183 Checks and adjustments 184
Performance Leakage Check 184 Performance Diagnostic Check 185 Performance Flowrate Check 185 Total Flowrate Check and Adjustment in Purge Mode 185 Measurement Path Flowrate Check and Adjustment 186 Total Flowrate Check in Normal Mode 188
Zero Calibration 188 Barometric Pressure Check and Calibration 189 Span Calibration Check 190
Disposal of Empty Calibration Gas Cylinder 192
Maintaining the Anesthetic Gas Module 193
Preventive Maintenance (PM) Tasks 193 Cleaning 194 Replace PM Parts 194
Internal Nafion Tubing with Bacterial Filters and manifold Seals 194
Room-Air Filter 196 Pump Filter 197 Performance Checks 198
Other factors to maximize uptime or reduce cost of ownership: 198
Troubleshooting the Anesthetic Gas Module 198
Compatibility Criteria for the AGM and the IntelliVue Monitors 198 Flow Charts for Communication and Measurement Type Problems 198 Hardware Related Troubleshooting Strategy 203 INOPs 204 Calibration Checks 206
Calibration Checks Troubleshooting Table 207
10
Page 11
Diagnostic Checks 208
Problem Solving Hierarchy 209 Pneumatic System Diagnostic Checks 210 O2 Assembly Diagnostic Checks 210 Optical Path Disgnostic Checks 213 IR Measurement Assembly Diagnostic Checks 214 Agent ID Assmebly Diagnostic Checks 215 Power Supply Diagnostic Checks 216 Operating Temperature Diagnostic Checks 217
Test Points, Connectors and Jumpers 217
Tes t Poi n t s 21 7 Connectors 218 Jumpers 218
Repairing the Anesthetic Gas Module 220
Introduction 220 The Top Cover 222
Removal 222 Replacement 222
Lifting the IR Measurement Mounting Bracket 224
Removal 225 Replacement 225
Infrared Measurement Assembly Head 227
Transferring NVRAM Data to a Replacement Head 227
Sample Cell 231
Removal 231 Replacement 231
Solenoid Valve #1 235
Removal 235 Replacement 235
Power Supply Unit 237
Removal 237 Replacement 237
Main PC Board 238
Removal 238 Replacement 239
O2 Sensor 240
Removal 240 Replacement 241
Agent Identification Head 243
Removal 243 Replacement 244
Pump 245
Removal 245 Replacement 245
Fan 246
11
Page 12
Removal 246 Replacement 246
Solenoid Valve #2 248
Removal 248 Replacement 248
Top C ov e r PC B o a rd 249
Removal 249 Replacement 249
Watertrap Manifold and Protector 251
Removal 251 Replacement 251
Power Fuses 252
Removal 252 Replacement 252
Test and Inspection Matrix 252 When to Perform Test Blocks 256 Safety Test Appendix 257
Parts List 259
Calibration Equipment 264
12
Page 13
This Service Guide contains technical details for the IntelliVue MP90 Patient Monitor, the Multi­Measurement Server (MMS), the Flexible Module Server (FMS) and the Measurement Server Extensions.
This guide provides a technical foundation to support effective troubleshooting and repair. It is not a comprehensive, in-depth explanation of the product architecture or technical implementation. It offers enough information on the functions and operations of the monitoring systems so that engineers who repair them are better able to understand how they work.
It covers the physiological measurements that the products provide, the Measurement Server that acquires those measurements, and the monitoring system that displays them.

Who Should Use This Guide

1

1Introduction

This guide is for biomedical engineers or technicians responsible for troubleshooting, repairing, and maintaining Philips’ patient monitoring systems.

How to Use This Guide

This guide is divided into eight sections. Navigate through the table of contents at the left of the screen to select the desired topic. Links to other relevant sections are also provided within the individual topics. In addition, scrolling through the topics with the page up and page down keys is also possible.

Abbreviations

Abbreviations used throughout this guide are:
Name Abbreviation
IntelliVue MP90 Patient Monitor the monitor
Flexible Module Server FMS
Multi-Measurement Server MMS
Measurement Server Link MSL
Medical Information Bus MIB
Anesthetic Gas Module AGM
13
Page 14
1 Introduction Responsibility of the Manufacturer

Responsibility of the Manufacturer

Philips only considers itself responsible for any effects on safety, reliability and performance of the equipment if:
• assembly operations, extensions, re-adjustments, modifications or repairs are carried out by persons authorized by Philips, and
• the electrical installation of the relevant room complies with national standards, and
• the instrument is used in accordance with the instructions for use.
To ensure safety, use only those Philips parts and accessories specified for use with the monitor. If non­Philips parts are used, Philips is not liable for any damage that these parts may cause to the equipment.
This document contains proprietary information which is protected by copyright. All Rights Reserved. Reproduction, adaptation, or translation without prior written permission is prohibited, except as allowed under the copyright laws.
Philips Medizin Systeme Böblingen GmbH
Hewlett-Packard Str. 2
71034 Böblingen, Germany
The information contained in this document is subject to change without notice.
Philips makes no warranty of any kind with regard to this material, including, but not limited to, the implied warranties or merchantability and fitness for a particular purpose.
Philips shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.

Passwords

In order to access different modes within the monitor a password may be required. The passwords are listed below.
Monitoring Mode: No password required
Configuration Mode: 71034
Demo Mode: 14432
Service Mode: 1345
Consult the configuration guide before making any changes to the monitor configuration.

Warnings and Cautions

In this guide:
•A warning alerts you to a potential serious outcome, adverse event or safety hazard. Failure to observe a warning may result in death or serious injury to the user or patient.
14
•A caution alerts you where special care is necessary for the safe and effective use of the product. Failure to observe a caution may result in minor or moderate personal injury or damage to the product or other property, and possibly in a remote risk of more serious injury.
Page 15

2Theory of Operation

Monitor Theory of Operation

The IntelliVue Patient Monitor:
• displays real-time data
• controls the attached measurement servers
• alarms in the case of patient or equipment problems
• offers limited data storage and retrieval (trending)
• interfaces to the Philips Clinical Network and other equipment
2
A monitor with just a single integrated measurement server can be connected to additional building blocks to form a monitoring system with a large number of measurements, additional interface capabilities and multiple slave displays. These elements cooperate as one single integrated real-time measurement system.

System Boundaries

The following diagram discusses specific boundaries within the overall system with respect to their openness and real-time requirements:
15
Page 16
2 Theory of Operation Monitor Theory of Operation
Philips Clinical Network
Measurement LAN
combines components of one patient monitor; real time requirements across all interconnected elements
Philips Clinical Network (wired LAN)
connects multiple patient monitors, information centers, application servers; closed system, only Philips qualified products (tested and with regulatory approval) are connected, Philips is responsible for guaranteed real-time functionality and performance
Philips Clinical Network (wireless)
like Philips Clinical Network (wired) LAN, however due to current wireless technologies available it has reduced bandwidth, longer latencies, reduced functionality
Hospital LAN, Internet
Standard Network, not under Philips control, no guaranteed service, no real-time requirements

Hardware Building Blocks

The following hardware building blocks make up the monitoring system:
16
Page 17
Monitor Theory of Operation 2 Theory of Operation
IntelliVue MP90
The MP90 monitor:
• can be used with the standalone M8031A 15-inch or the M8033A 17-inch color LCD TFT display with touchscreen operation.
• can also be used with other XGA and SXGA standalone off-the-shelf displays which comply with medical standards such as IEC 60601-1 and IEC 60601-1-2.
• has the central processing unit in a separate module
• uses the Philips SpeedPoint as primary input device whereas the Philips Touchscreen and computer devices such as mice, trackball, and keyboard can be added optionally
• supports the Flexible Module Server (FMS)
Building Blocks:
17
Page 18
2 Theory of Operation Monitor Theory of Operation
|| I/F To Local Printer
Power Supply
I/F
Boards
1st Main Board
Remote Device I/F to
Alarm Device I/F*
Alarm Device and Remote Input
MIB To AGM
1 Main Board
I/F
2nd
Main Board
Boards
FMS CPU
FMS M/B
MSL
OR
MMS CPU
Video I/F
Board
Video I/F
Board
To Ext. Display
I/F
To Ext. Display
Optional Hardware
The M8031A 15-inch color LCD TFT display or the M8033A 17-inch color LCD TFT display (both with touchscreen operation) can be ordered optionally. Additional input devices such as mice, trackball or keyboard can also be added. If the monitor is ordered with the wireless LAN option an external wireless transmitter is required. For further details regarding the wireless network please refer to the M3185A Philips Clinical Network documentation.
Compatible Devices
18
Figure 1 M8048A Flexible Module Server (FMS)
Page 19
Monitor Theory of Operation 2 Theory of Operation
Figure 2 M3001A Multi-Measurement Server (MMS)
Power Supply
The AC/DC converter transforms the AC power coming from the power plug into 48 V/120W DC source and isolates the monitoring system from the AC power mains.The 48V is distributed via power bus and supplies power to all the components of the system: The 56 V DC power needed for the FMS, MMS and measurement server extension is created by an isolating DC/DC converter. The CPU is supplied with 3.3 V and 5 V DC power. The transformation is performed in two steps: The first DC/ DC converter is a power regulator which reduces the variations caused by load changes on the 48V power bus. The second DC/DC converter converts the power to the needed voltage. Interface boards require a power of 10V AC.
CPU Boards
The CPU boards have an MPC860 50 MHz processor that provides a number of on-chip, configurable interfaces. An array of 12 fast UARTs with configurable protocol options are implemented in an ASIC (along with other system functions such as independent watchdogs etc.), providing interfacing capabilities to measurement modules and I/O boards. The serial interfaces can easily be electrically isolated. The main board contains additional video hardware.
Flexible Module Server
CPU
Bank of I/Os
Modules
The CPUs provide two LAN interfaces to interconnect CPUs (via the Internal LAN) or to connect to the Philips Clinical Network.
IntelliVue Patient Monitor Multi-Measurement Server
CPU Video Bank of I/Os
CPU Bank of I/Os
Measurement
Interfaces
Acquisition
The CPU capabilities are identical. Different loading options are coded on serial EEPROMs to support the automatic configuration of the operating system at boot time.
19
Page 20
2 Theory of Operation Monitor Theory of Operation
I/O Boards
Interfaces to the monitor are implemented via I/O boards. The location of these boards is restricted by general rules. The I/O slot designations diagram and the I/O matrix which outline the I/O board placement rules can be found in the Installation Instructions section.
The following is a list of Interface (I/O) boards which may be present in your monitor, depending on your purchased configuration:
•MSL
•Video
• Philips Clinical Network (LAN wired or wireless)
•PS/2
• MIB/RS232
•Flexible Nurse Call
•Parallel printer
• Remote devices (Remote Alarm Device, Remote Extension Device)
The specifications for the above listed interfaces can be found in the technical data sheet for the monitor and in the Specifications chapter of the Instructions for Use.

Data Flow

The following diagram shows how data is passed through the monitoring system. The individual stages of data flow are explained below.
Display
and User
Interface
Data
Acquisition
Data
Provider
Applications
Service
Persistent
Data
Storage
Data
Output
Data Acquisition
Monitoring data (for example patient measurement data in the form of waves, numerics and alerts) is acquired from a variety of sources:
20
Page 21
Monitor Theory of Operation 2 Theory of Operation
•Measurement Servers
The Measurement Servers connected to the internal LAN convert patient signals to digital data and apply measurement algorithms to analyze the signals.
• External measurement devices
Data can be also acquired from devices connected to interface boards of the monitor. Software modules dedicated to such specific devices convert the data received from an external device to the format used internally. This applies to parameter modules and the Anesthetic Gas Module.
• Server systems on the Philips Clinical Network
To enable networked applications such as the other bed overview, data can be acquired from server systems attached to the Philips Clinical Network, for example a Philips Information Center.
Data Provider System Service
All data that is acquired from measurement servers or external measurement devices is temporarily stored by a dedicated data provider system service. All monitor applications use this central service to access the data in a consistent and synchronized way rather than talking to the interfaces directly.
This service makes the applications independent of the actual type of data acquisition device.
The amount of data stored in the data provider system service varies for the different data types. for example several seconds of wave forms and the full set of current numerical values are temorarily stored in RAM.
Persistent Data Storage System Service
Some applications require storage of data over longer periods of time. They can use the persistent data storage system service. Dependent on the application requirements, this service can store data either in battery backed-up (buffered) memory or in flash memory. The buffered memory will lose its contents if the monitor is without power (not connected to mains) for an extended period of time. The flash memory does not lose its contents.
The trend application for example stores vital signs data in a combination of flash memory and buffered memory, while the system configuration information (profiles) is kept purely in flash memory.
Display and User Interface Service
Applications can use high level commands to display monitoring data or status and command windows on the LCD panel. These commands are interpreted by the display manager application. This application controls the dedicated video hardware which includes video memory and a special ASIC.
User input is acquired from a variety of input devices, for example the SpeedPoint, the touchscreen or other standard input devices (keyboard, mouse) which may be attached to I/O boards. The system software makes sure that the user input is directed to the application which has the operating focus.
Data Output
The monitoring system is very flexible and customizable regarding its data output devices. Built-in devices (for example LAN, speaker, video) provide the basic output capabilities.
These capabilities can be enhanced by adding additional I/O boards, as required in the specific end­user setup. The additional I/O boards typically provide data to externally attached devices, for example to printers, RS232 based data collection devices, nurse call systems etc.
21
Page 22
2 Theory of Operation Monitor Theory of Operation
The monitor can identify I/O boards by means of a serial EEPROM device that stores type and version information. The operating system detects the I/O boards and automatically connects them with the associated (interface driver) application. For some multi-purpose cards it is necessary to configure the card for a particular purpose first (for example the dual MIB/RS232 card can support external touch display, data import, data export).
Monitor Applications
The monitor applications provide additional system functionality over the basic measurement and monitoring capabilities. This includes for example trending, report generating, event storage or derived measurements.
In general, the monitor applications use the data provider system service to access the measurement data. Application interfaces to the other system services allow the application to visualize data, to store data over extended periods of time or to output data to other devices.
Internal LAN (Measurement Server Link)
All components of the monitoring system (including measurement servers and CPUs in the monitor) communicate using an IEEE802.3/ Ethernet LAN in the Measurement Server Link (MSL). This network is used to distribute data between the components, for example:
• Digitized patient signals including wave data, numerical data and status information (typically from the measurement server to a display unit)
• Control data representing user interactions (typically from the display unit to a measurement server)
• Shared data structures, for example representing patient demographical data and global configuration items
The internal LAN allows plug and play configuration of the monitoring system. The system automatically detects plugging or unplugging of measurement servers and configures the system accordingly.
The components on the internal LAN are time-synchronized to keep signal data consistent in the system. Dedicated hardware support for synchronization eliminates any latency of the network driver software.
The integrated LAN provides deterministic bandwidth allocation/reservation mechanisms so that the real-time characteristic of signal data and control data exchange is guaranteed. This applies to the data flow from the measurement server to the monitor (for example measurement signal data) and the data flow from the monitor to a measurement server (for example to feed data to a recorder module).
Integrated communication hubs in the monitor and the FMS allow flexible cabling options (star topology, daisy chaining of servers).
22
Page 23
Monitor Theory of Operation 2 Theory of Operation
MDSE Internal LAN
MDSE
Internal
LAN
Philips Clinical Network
The monitoring system may be connected to the Philips Clinical Network, for example to provide central monitoring capabilities or other network services. This connection may be through a normal wired connection or through a wireless connection.
The monitor supports the connection of an external off-the-shelf wireless adapter. This allows a simple field upgrade as well as a technology upgrade in the future. Switching between wired and wireless networks is automatically triggered by the plugging or unplugging of the network cable.
The Philips Clinical Network protocols function very similarly to the protocols used on the internal LAN.
MDSE
Internal
LAN
After configuration, the monitoring system sends the digitized patient signals including wave data, numerical data and status information onto the network. Control data representing user interactions can be exchanged between the monitoring system and a central station bi-directionally.
Additional protocols are supported for networked applications, for example for the other bed overview function, which allows viewing of monitoring data from other patients on the network.
For plug and play operation, the monitoring system uses the standard BootP protocol to automatically acquire a network address.

How does the Support Tool Work with the Monitor

The support tool is a Windows application typically installed on the laptop of a customer engineer or a biomedical engineer working in the customer’s own service department.
The purpose of the support tool is to upgrade, configure and diagnose all monitoring components (modules, measurement servers, and monitors) in the system over the network.
23
Page 24
2 Theory of Operation Monitor Theory of Operation
The service protocol developed for this purpose uses a raw access to the devices without the need for IP addresses etc. over a standard customer network installation, so that even defective devices can be upgraded as long as the few kBytes of initial boot code are working. The boot code itself can also be upgraded using the same protocol.
The tool allows access to internal service information and to serial numbers. It can be remote­controlled, for example via a dial-up connection from a response center, provided the proper infrastructure is in place.
For details see the Instructions for Use for the Support Tool.

Monitor Software Block Diagram

Figure 3 shows the functional block diagram for the monitoring system. A legend explaining terms and diagram elements follows. The information below varies depending on the purchased monitor options.
24
Page 25
Monitor Theory of Operation 2 Theory of Operation
Philips Clinical
Network
LAN
MDSE
Input Devices
(including PS/2)
Remote Extension
Device M8026A
SpeedPoint
(LAN, Centronics, Display Controller, HIF Control,RS-422, PS/2, Remote Device I/F, Nurse Call, MIB)
Applications
System Services
Real Time Operating System
Visual
Indicators
Remote Alarm Device
LEDs
Indicators
M8025A
Loudspeaker
Audio
Video Out
Centronics
Input Devices
(PS/2)
Nurse
MIB
Call
Printer Connection
Interfaces
Interface Managers
Record Alarm Trend HiRes ADT
Reports
Calc Param
Events
ECG-Out Marker-In
ECG-Out
Marker-In
Loudspeaker
MDSE
LAN
LAN
RS-422
M3001A Multi­Measurement Server
12-lead ECG/Resp, NBP, SpO2,
Press/Temp
M3015/16A Measurement Server
Extension
CO2, Press/Temp
M1006B Press
Plug-In Modules
M1012A C.O.
LAN RS-422
M1018A tcPO2/CO2
M1029A Temp
RS-422
Flexible Module Server
M1032A VueLink
M1116B Recorder
Figure 3 IntelliVue Patient Monitoring System Functional Block Diagram
25
Page 26
2 Theory of Operation Monitor Theory of Operation

Block Diagram Legend

Functional Block Description
Services
Operating System The Operating System (OS) provides a layer of isolation between
the specific hardware implementation and the application software. The OS performs system checks and allocates resources to ensure safe operation when the system is first started. This includes internal self-tests on several hardware modules and configuration checks for validity of configuration with the operating software. During normal operation, the OS continues to run checks on system integrity. If error conditions are detected the OS will halt monitoring operations and inform the operator about the error condition.
System Services The System Services provide generic common system services.
In particular: It uses a real-time clock component to track time. It synchronizes to network time sources and verifies the accuracy of the system time information. It is also responsible for managing persistent user configuration data for all Measurement Servers, Flexible Module Servers and IntelliVue Patient Monitoring System software modules. User configuration data is stored in a non­volatile read/write storage device
Applications
Reports The Reports Service retrieves current and stored physiological
data and status data to format reports for printing paper documentation. The following reports are supported:
•Vital Signs Report
• Graphical Trend Report
• Event Review Report
•Event Episode Report
• ECG Report (12 Lead/Multi-Lead)
• Cardiac Output Report
• Calculations Report (Hemodynamic/Oxygenation/ Ventilation)
• Calculations Review Report
•Wedge Report
•Test Report
• Other reports (e.g. Loops, Review Applications, Drug report)
The Reports service generates report data which can be printed on a local or a central printer.
26
Page 27
Monitor Theory of Operation 2 Theory of Operation
Functional Block Description
Record The Record Service retrieves current and stored physiological
data and status data to format a continuous strip recording. A recording can be triggered manually by the operator or automatically by an alarm condition. The Record Service uses the services of the Recorder Interface to control an M1116B Recorder in the FMS. The Record Service can also send data to a central recorder.
Alarm The Alarm Service contains logic that prioritizes alarm conditions
that are generated either by the Measurement Servers, Flexible Module Server, or by IntelliVue Patient Monitoring System software modules. Visual alarm messages are displayed at the top of the IntelliVue Patient Monitoring System display and alarm sounds are generated by a loudspeaker. Alarm conditions may be generated when a physiological parameter exceeds preselected alarm limits or when a physiological parameter or any other software module reports an inoperative status (technical alarm, for example, the ECG leads may have fallen off the patient). The Alarm service manages the alarm inactivation states, for example suspension of alarms, silencing of alarms, and alarm reminder. Alarm signals may also be configured as latching (alarm signals are issued until they are acknowledged by the operator, even when the alarm condition is no longer true). The Alarm service controls the visual alarm signals (alarm lamps).
Trend The Trend service stores the sample values of physiological data
and status data with a resolution of 12 seconds, 1 minute or 5 minutes for a period of up to 48 hours. The data is kept in battery buffered read/write storage and flash memory devices to be preserved across power failures. The stored data is protected via consistency checks and checksums. When a new patient is admitted, the trend database erases all data of the previous patient.
HiRes The OxyCRG (Oxygen CardioRespiroGram) service derives a
high-resolution trend graph from the Beat-to-Beat Heart Rate, SpO
or tcpO2, and Respiration physiological data. The
2
OxyCRG is specialized for neonatal applications, allowing the opeartor to identify sudden drops in Heart Rate (Bradycardia) and SpO
or tcpO2 (Desaturations), and supporting the operator
2
in visualizing Apnea situations.
ADT The ADT (Admit/Discharge/Transmit) service maintains the
patient demographics information. The operator may admit a new patient, discharge the old patient and enter or modify the patient demographics. The ADT service also supports the transport of a patient (trend database) with the M3001A Multi­Measurement Server. The ADT service controls the deletion of old patient data, the upload of trend data from the M3001A and the switching back of all settings to user defaults. It also synchronizes patient information with a central station on the network.
27
Page 28
2 Theory of Operation Monitor Theory of Operation
Functional Block Description
Calc Param The Calc Param (Calculated Parameters) service accesses current,
stored and manually entered physiological data as input to calculation formulas. With these formulas, derived hemodynamic, oxygenation and ventilation variables are computed. The calculation results, including the input parameters, are stored for later review using the Trend service.
Interface Managers
MDSE The MDSE (Medical Data Service Element) Interface Manager is
responsible for the exchange of real-time data between the IntelliVue Patient Monitoring System display unit and the Measurement Servers and Flexible Module Server as well as between the IntelliVue Patient Monitoring System display unit and other devices attached to the network. MDSE establishes and maintains a data communication link between the devices. It provides configuration information about the remote device to applications in the local device and it allows the exchange of measurement data and status information between the devices.
Printer The Printer Interface Manager provides a high level interface to a
printer. It provides means to:
• establish a connection to the printer
• transfer data to the printer
• get status of the printer
• close connection to the printer
The Printer Interface Manager also supervises the connection to the printer and whether the printer accepts data (for example paper out). The Printer Interface Manager notifies the operator in such cases.
28
Page 29
Monitor Theory of Operation 2 Theory of Operation
Functional Block Description
Display & Operator Interface The Display and Operator Interface Manager performs the
following tasks:
• Screen presentation of real-time and stored physiological measurement data, alarm condition data and status information received from the MDSE interface manager, the Alarm service or other IntelliVue Patient Monitoring System modules
• Screen presentation of operating controls (control windows)
• Processing of operating control commands received from HIF Control interface. The module verifies and interprets the received commands and forwards them to other software modules of the IntelliVue Patient Monitoring System display unit, Measurement Servers or Flexible Module Server
• Sound generation (issues audible alarm signals and generates audible information signals, for example QRS and SpO
Interfaces
LAN The LAN interface implements the physical layer of IEEE 802.3.
The LAN interface performs Manchester encoding/decoding, receive clock recovery, transmit pulse shaping, jabber, link integrity testing, reverse polarity detection/correction, electrical isolation, and ESD protection. Electronically separated interfaces are used for communication to the Measurement Servers or Flexible Module Server and to the network.
Centronics The Centronics interface implements the standard signaling
method for bi-directional parallel peripheral devices according to IEEE 1284-I. The interface is used as a parallel interface to a standard printer with electrical isolation and ESD protection.
Display Controller The Display Controller Interface consists of a video controller
chip, video RAM and the controlling software. The Display Controller interface processes the high level display commands (character and graphic generation, wave drawing) and translates them into pixels, which are written into the video RAM where the video controller chip generates the video synchronization signals and the pixel stream for the external display.
HIF Control The HIF (Human Interface Control) interface scans the Human
Interface devices for operator controls (Touch Screen, Trim Knob, and PS/2 devices), formats the collected data and sends it to the display and Operating Interface.
tones, operator audible feedback)
2
29
Page 30
2 Theory of Operation Monitor Theory of Operation
Functional Block Description
ECG-Out Marker-In The ECG Out/Marker In interface receives the ECG waveform
directly from the ECG/Resp Arrhythmia ST-Segment physiological algorithm via an RS-422 serial interface and converts the digital ECG signal to an analog ECG signal. In addition, the ECG Out controller receives from a connected device the marker information and forwards this data to the ECG/Resp Arrhythmia ST-Segment physiological algorithm. The converted analog signal is used to synchronize a connected device to the patient’s ECG
RS-422 The serial link RS-422 interface communicates the ECG signal
to the ECG Output/Marker In of the IntelliVue Patient Monitoring System display unit. The interface is a serial, differential, full-duplex link. The interface is ESD protected.
PS/2 The PS/2 interface supports the serial protocol of standard PS/2
devices (mouse). The PS/2 serial protocol is interpreted by the HIF Control interface.
Remote Device Interface The remote device interface is an interface to the remote alarm
device and the remote extension device. The interface communicates with the remote devices via a differential serial link.
Nurse Call The Nurse Call board contains 2 connectors. A phone jack type
connector and a multi-port connector. The phone jack type connector has a single close-on-alarm relay. The multi-port connector has three alarm relays which are configurable to be open or closed on alarm. In addition, this interface has an audible alert capability for loss of AC power.
MIB The MIB interface allows full-duplex, short-haul asynchronous
binary communication between the monitor and an arbitrary (medical/non-medical) device using an eight-pin RJ45 modular connector. Switching between MIB and RS232 protocol is possible.
30
Page 31

3Testing and Maintenance

Concepts

This chapter provides a checklist of the testing and maintenance procedures for the monitor, the MMS the Measurement Server Extensions and the FMS associated modules.
Preventive Maintenance refers specifically to the series of tests required to make sure the Instrument measurement results are accurate. The measurements requiring these reported tests are NBP and sidestream CO readings are in question or as specified.
. The accuracy and performance procedures are designed to be completed when
2

Test Reporting

Authorized Philips personnel report test results back to Philips to add to the product development database. Hospital personnel, however, do not need to report results. This table shows you what to record on the service record after completing the tests in this chapter.
3
Test What to record
Visual V:P or V:F
Power On PO:P or PO:F
P NIBP PN:P/X1/X2/X3/X4 or
PN:F/X1/X2/X3/X4
P CO
2
Safety S(1):P/x1/x2 or
Where P = Pass, F = Fail and X/x are the measured values as defined in the tests described in this chapter.
PCO2:P/X1/X2/X3/X4/X5/X6/X7/X8 or
PCO2:F/X1/X2/X3/X4/X5/X6/X7/X8
S(1):F/x1/x2
S(2): P/x1 or
S(2): F/x1
S(3): P/x1 or
S(3): F/x1
31
Page 32
3 Testing and Maintenance Recommended Frequency

Recommended Frequency

The testing checklist appears in the next section of this chapter. Perform the procedures as indicated in the suggested testing timetable. These timetable recommendations do not supersede local requirements.
Suggested Testing Timetable Frequency
Preventive Maintenance Tests
•NBP Calibration
• Sidestream CO
•CO
pump / CO2 scrubber replacement
2
Calibration
2
Performance and Safety Tests
• Temperature Accuracy
• ECG/Resp Performance
• Invasive Pressure Performance
•SpO
•Mainstream CO
Per formance
2
Per formance
2
•EEG Performance
• C.O. Performance
•BIS Performance
•SvO2 Performance
Required
• Once a year, or as specified by local laws.
• Once a year or after 4,000 hours continuous use and following any instrument repairs or the replacement of any instrument parts.
• Once every three years or after 15 000 operating hours
Once every two years, or if you suspect the measurement is incorrect
32
•tcGas Performance
• VueLink Performance
• Nurse Call Relay Performance*
• ECG Sync Performance*
*Only when in use as part of hospital protocols
Safety Checks (in accordance with IEC 60601-
1)
• System Enclosure Leakage Current
•Protective Earth
• Patient Leakage Current
Once every two years and after repairs where the power supply is replaced or the monitor has been damaged by impact.
Page 33
Tests Recommended When Performing... 3 Testing and Maintenance

Tests Recommended When Performing...

Installation

Service Event
(When performing...
Installation of monitor with a non-medical device connected.

Repair

Service Event
(When performing...
Repairs of M3015A Perform Power On and M3015A tests
Repairs where the monitor has been damaged by impact
Repairs where the power supply is replaced Perform Safety (2) Test Block
All other IntelliVue Monitoring System repairs Perform Power On Test Block

Preventive Maintenance

Perform preventive maintenance tests:
Tes t Bl o c k s R e q u i r e d
...Complete these tests)
Perform Visual, Power On and Safety (1) Test Blocks
Tes t Bl o c k s R e q u i r e d
...Complete these tests)
Perform Power On and Safety (2) and (3) Test Blocks
•NBP calibration
• Sidestream CO
• Pump and scrubber replacement.
calibration
2

Performance Verifications

Perform all safety, accuracy and performance test procedures listed in the following sections. If a particular measurement is in question, perform the measurement performance test only.
33
Page 34
3 Testing and Maintenance Tests

Upgrades

Service Event
(When performing...
Hardware and software upgrades Perform Power On Test Block unless otherwise

Tests

Some of the following testprocedures must be performed in service mode. To enter service mode select Operating Modes in the main menu. Then select Service Mode and enter the password.
If required, open the screen menu in the monitor info line at the top of the screen and select Service to access the service screen. This is required particularly for Anesthetic Gas Module testing procedures.

Visual Test

Inspect the system for obvious signs of damage. Also check all external leads and accessories.
The expected test result is pass: the system has no obvious signs of damage.

Power On Test

Tes t Bl o c k s R e q u i r e d
...Complete these tests)
specified in the Upgrade Installation Notes shipped with the upgrade.
NOTE An external display must be connected to the MP90 computer module unit.
1 Switch on the monitor and connect the MMS.
2 Observe whether the system boots up successfully and if an ECG wave appears on the screen.
The expected test result is pass: the monitor boots up and displays an ECG wave. The wave might be a flat line if no simulator is attached.

NBP Tests

This section describes NBP test procedures.The monitor must be in service mode to perform these tests.
34
Page 35
Tests 3 Testing and Maintenance
NBP Accuracy Test
This test checks the performance of the non-invasive blood pressure measurement. Connect the equipment as shown:
Expansion Chamber
Tubing
Manometer
To NBP Input
Tools required:
• Reference manometer (includes hand pump and valve), accuracy 0.2% of reading.
• Expansion chamber (volume 250 ml +/- 10%)
• Appropriate tubing.
In service mode, the systolic and diastolic readings indicate the noise of NBP channels 1 and 2 respectively. When static pressure is applied, the reading in NBP channel 1 should be below 50. The value in parentheses indicates the actual pressure applied to the system.
1 Connect the manometer and the pump with tubing to the NBP connector on the MMS and to the
expansion chamber.
2 In service mode, select the Setup NBP menu. 3 Select Close Valves: On
4 Raise the pressure to 280 mmHg with the manometer pump.
5 Wait 10 seconds for the measurement to stabilize.
6 Compare the manometer values with the displayed values.
7 Document the value displayed by the monitor (x1).
8 If the difference between the manometer and displayed values is greater than 3 mmHg, calibrate
the MMS. If not, proceed to the leakage test.
9 To calibrate the MMS, select Close Valves off then Cali brate NBP and wait for the
instrument to pump up the expansion chamber.Wait a few seconds after pumping stops until EnterPrVal is highlighted and then move the cursor to the value shown on the manometer. If one of the following prompt messages appears during this step, check whether there is leakage in the setup:
– NBP unable to calibrate – cannot adjust pressure
– NBP unable to calibrate – unstable signal
10 Press Confirm.
If the INOP NBP Equipment Malfunction message occurs in monitoring mode, go back to service mode and repeat the calibration procedure.
35
Page 36
3 Testing and Maintenance Tests
NBP Leakage Test
The NBP leakage test checks the integrity of the system and of the valve. It is required once per year and when you repair the monitor or replace parts.
1 If you have calibrated, repeat steps 2 to 6 from the accuracy test procedure so that you have 280
mmHg pressure on the expansion chamber.
2 Watch the pressure value for 60 seconds.
3 Calculate and document the leakage test value (x2).
x2 = P1 - P2 where P1 is the pressure at the beginning of the leakage test and P2 is the pressure displayed after 60 seconds. The leakage test value should be less than 6 mmHg.
NBP Linearity Test
1 Reduce the manometer pressure to 150 mmHg.
2 Wait 10 seconds for the measurement to stabilize.
3 After these 10 seconds, compare the manometer value with the displayed value.
4 Document the value displayed by the monitor (x3)
Valve Test
5 If the difference is greater than 3 mmHg, calibrate the MMS (see steps 9 to 10 in the accuracy test
procedure).
1 Raise the pressure again to 280 mmHg. 2 Select Close valves: Off.
3 Wait five seconds and then document the value displayed. The value should be less than 10
mmHg.
4 Document the value displayed by the monitor (x4).
Test Expected test results
Accuracy test x1 = value displayed by monitor
Difference 3mmHg
Leakage test x2 = leakage test value
x2 < 6 mmHg
Linearity test x3 = value displayed by monitor
Difference 3mmHg
36
Valve Test x4 = value < 10 mmHg
Page 37
Tests 3 Testing and Maintenance

Sidestream CO2 Performance Test

Allow five seconds between individual service procedures to ensure stable equipment conditions. When certain monitor procedures are running, service procedures are not possible and trying to start them will result in a message Service Operation Failed in the monitor’s status line. Wait until the monitor completes the current operation, then restart the service procedure.
This test checks the performance of the CO
measurement for the sidestream extension. The CO2
2
performance test is required once per year and when the instrument is repaired or when parts are replaced.
This test uses calibration equipment that you can order (see the Parts section for the part number). The procedure is summarized in the following steps. Refer to the documentation accompanying the equipment for detailed instructions.
Tools Required:
• Standard tools, such as screwdriver, tweezers
• Electronic flowmeter, M1026-60144.
• Gas calibration equipment:
• Cal 1 gas 15210-64010 (5% CO
• Cal 2 gas 15210-64020 (10% CO
)
2
)
2
• Cal gas flow regulator M2267A
• Cal tube 13907A
You also need a local barometric pressure rating received from a reliable local source (airport, regional weather station or hospital weather station) which is located at the same altitude as the hospital.
The CO
calibration for the sidestream extension consists of the following steps:
2
• Barometric pressure check and calibration, if required.
• Leakage check
•Pump check
• Flow check and calibration, if required.
•Noise check
•CO
•CO
Cal check and calibration, if required.
2
Cal verification using 2nd cal gas
2
Perform all checks in the same session.
Barometric Pressure Check and Calibration
Check the barometric pressure value in the sidestream CO2 extension as follows:
1 Go into service mode and select Setup CO
2 Connect a FilterLine to the sidestream CO
Extension.
3 The status line at the bottom of the screen displays “CO2 pressure reading (ambient/cell) xxx/yyy”
where xxx is the ambient pressure and yyy is the measured cell pressure. Check whether the ambient pressure value (x1) matches (within the acceptable tolerance of ±12mm Hg) the reference
menu.
2
input. This activates the pump in the sidestream CO2
2
37
Page 38
3 Testing and Maintenance Tests
value you have received. If so, proceed to the leakage check. If the value is not correct, calibrate as follows.
Leakage Check
The leakage check consists of checking the tubing between:
• the pump outlet and the measurement server extension outlet and
• the pump inlet and FilterLine inlet.
Check the user’s guide of the flowmeter for details on how to make a correct flow reading.
Part 1
1 Go into service mode and select Setup CO
2 Connect a FilterLine to the sidestream CO
3 Check the ambient pressure and the cell pressure shown in the monitor’s status line. The cell
a. Select CO
then select Barom.Press to activate a table of values.
2
b. Select the value in the table which matches the reference value received from a reliable local
source (airport, regional weather station or hospital weather station). (The values are displayed with a resolution of 2 mmHg up to 500 mmHg and a resolution of 1 mmHg from 500 mmHg to 825 mmHg.) Note: the selected value must be within ±10% of the current measured ambient pressure, otherwise an error message will occur at restarting the monitor.
c. Confirm the barometric pressure setting.
d. Check that the ambient pressure displayed in the status line at the bottom of the screen is the
same as the value which you selected from the list in step b.
menu.
2
input to start the pump running.
2
pressure should be approximately 20 mmHg lower than ambient pressure.
4 Connect the flowmeter outlet to the FilterLine inlet using a flexible connecting tube.
5 Block the measurement server extension outlet using your fingertip and observe the flowmeter
display. The value on the flowmeter (x2) should decrease to between 0 and 4 ml/min, accompanied by an audible increase in pump noise. If the value is within the tolerance limits, continue with part 2 of the leakage check.
6 If the value is outside the tolerance limits, there is a leakage between the pump outlet and the
measurement server extension gas outlet.
7 Open the measurement server extension and check the tubing connections at the pump outlet and
the extension gas outlet. If the connections are good, then there is a leakage in the tubing and you must exchange the measurement server extension.
Part 2
1 Disconnect the flowmeter from the Part 1 setup and connect the flowmeter inlet to the M3015A
gas outlet.
2 Leave the Filterline connected to the M3015A inlet.
3 Block the inlet of the FilterLine using your fingertip and observe the flowmeter display. The value
on the flowmeter (x3) should decrease to between 0 and 4 ml/min, accompanied by an audible increase in pump noise. Do not block the inlet for longer than 25 seconds as this will lead to an “Occlusion” INOP. If the value is within the tolerance limits, there are no leakages and the leakage check is completed; proceed to the pump check.
38
Page 39
Tests 3 Testing and Maintenance
If the value is not within the tolerance limits, there is a leakage between the FilterLine inlet and the
4
pump inlet.
5 Check the FilterLine connections and open the M3015A to check the tubing connections at the
pump inlet and the M3015A gas inlet. If the connections are good, try replacing the FilterLine and repeating the leakage check. If the situation remains, there is a leakage in the tubing and the M3015A must be exchanged.
Pump Check
1 Connect the flowmeter inlet to the M3015A gas outlet.
2 Connect the FilterLine to the M3015A inlet.
3 Block the inlet of the FilterLine using your fingertip and observe the cell pressure on the M3046A
display. The cell pressure (x4) should be more than 120 mmHg below the ambient pressure shown. If the pressure difference is less than 120 mmHg, the pump is not strong enough and you should replace it, irrespective of the Pump OpTime.
Flow Rate Check and Calibration
Check the flow rate in the sidestream CO2 extension as follows:
Noise Check
1 Connect the flowmeter to the CO
2 Check on the flowmeter the flow that the sidestream CO
50 ml/min ± 7.5 ml/min. If the value is within tolerance, proceed to the CO
FilterLine.
2
extension pump draws (x5). It should be
2
Gas calibration
2
check. If the value is not within tolerance, calibrate as follows.
3 Adjust the flow in the instrument by selecting Increase Flow or Decrease Flow until it
is as close as possible to 50 ml per minute as indicated on the flowmeter gauge.
4 When you are satisfied that the flow is set as close as possible to 50 ml per minute, select Store
Flow and confirm the setting. If you do not store the adjusted flow within 60 seconds of the
adjustment, the old flow setting is restored.
5 If you cannot adjust the flow to within tolerance, replace the pump. If you still cannot make the
flow adjustment, this indicates a fault in the measurement extension, which must be replaced.
1 With the monitor in service mode, select Setup CO
2 Disconnect the flowmeter and connect the 5% calibration gas and flow regulator in its place.
3 Open the valve to apply the 5% calibration gas and wait until the value is stable.
4 Check the noise index (x6) displayed next to the CO
of noise on the CO
wave). If the value exceeds 3 mmHg, replace the measurement extension.
2
menu.
2
value on the display (this indicates the level
2
CO2 Gas Measurement Calibration Check
After switching the measurement extension on, wait at least 20 minutes before checking the calibration. Check the calibration of the CO
1 Check that the 5% calibration gas and flow regulator are connected.
gas measurement as follows:
2
39
Page 40
3 Testing and Maintenance Tests
Calculate the expected measurement value in mmHg as follows:
2
0.05 x (ambient pressure) = value mmHg for example 0.05 x 736 = 36.8 mmHg (with an ambient pressure of 736 mmHg)
3 Open the valve on the flow regulator to allow 5% CO
gas to flow into the extension. Allow the
2
value to stabilize.
4 Check that the value on the instrument (measurement value on the main screen, x7)) matches the
calculated mmHg value ± 2.6 mmHg. If the value is outside the tolerance, calibrate as described in step 9 in this procedure onwards.
5 Disconnect the 5% calibration gas and connect the 10% calibration gas.
6 Calculate the expected measurement value and tolerance in mmHg as follows:
0.1 x (ambient pressure) = value mmHg ±0.07 x (value mmHg) = tolerance
for example 0.1 x 737 mmHg = 73.7 mmHg (with an ambient pressure of 737 mmHg) ±0.07 x 73.7 mmHg = ±5.16 mmHg tolerance
7 Open the valve on the flow regulator to allow 10% CO
gas to flow into the extension. Allow the
2
value to stabilize.
8 Check that the value on the instrument (x8) matches the calculated mmHg value within the
calculated tolerance. If so, the measurement extension is correctly calibrated. If the value is outside the tolerance, calibrate as follows.
9 If not already connected, connect the 5% calibration gas. 10 Select Cal. CO
11 Select the value for the calibration gas. (The default value is 5.0%.)
.
2
12 Open the valve on the calibration gas to allow CO
to stabilize before the start of the calibration. Leave the valve open until the instrument gives a prompt that gas can be removed.
13 The extension calibrates and prompts when calibration is successful.
Calibration Verification
1 Reopen the 5% gas valve and allow the value to stabilize.
2 Check that the value displayed on the monitor is correct within the tolerance (see step 2 above).
3 Disconnect the 5% calibration gas and connect the 10% calibration gas.
4 Open the valve on the flow regulator to allow 10% CO
value to stabilize.
5 Check that the value displayed on the monitor is correct within the tolerance (see step 6 above).
If one or both values are not within tolerances, you must exchange the measurement server extension.
Reset Time Counters
You must check the time counters on the sidestream CO2 extension before calibrating the instrument. As well, when parts are replaced, the appropriate counters must be reset to zero.
gas to flow into the extension. Allow the value
2
gas to flow into the extension. Allow the
2
40
Page 41
Tests 3 Testing and Maintenance
The counters for CO2 pump, IR Src and Last Cal are displayed in the status line. The values are updated when entering the Setup CO
menu.
2
Observe the following guidelines:
• When calibrating the CO
extension, if no parts have been replaced, check the displayed values of
2
Reset PumpOpTi me and Reset IRSourceTime selections to make sure that they are within suggested guidelines for use (15, 000 hours of continuous use). If the counter time is greater than 15, 000 hours, replace the appropriate part. See Repair and Disassembly for details.
• When calibrating the CO
extension, if parts have been replaced, reset the appropriate values using
2
the Reset PumpOpTime and Reset IRSourceTime selections. See Repair and Disassembly for details.
Resetting the PumpOpTime generates the INOP: “CO
OCCLUSION”. To clear this INOP you
2
must perform a flow check and store the flow in service mode (select Store Flow).
Table 1 Documenting CO
Test Results
2
Test Expected Test Results
Barometric Pressure Check
x1 = difference between the reference pressure and the measured ambient pressure displayed on the monitor
(x1<12 mmHg)
Leakage Check parts 1 and 2
x2 = value of part 1 leakage check on flowmeter (x2< 4.0 ml/min)
x3 = value of part 2 leakage check on flowmeter (x3< 4.0 ml/min)
Pump Check x4 = difference in pressure between cell pressure and ambient
Flow Check x5 = difference between measured value and 50.0 ml/min
Noise Check x6 = noise index displayed on monitor (x6<3.0)
CO
Gas
2
Calibration Check
CO
2 Cal Verification

Temperature Accuracy

This test checks the performance of the temperature measurement.
Tools required: Patient simulator (with 0.1
1 Connect the patient simulator to the temperature connector on the MMS or measurement server
extension.
2 Configure the patient simulator to 40
pressure displayed on the monitor during occlusion (x4 >120 mmHg)
(x5<7.5 ml/min)
x7 = difference between measured CO2 value and calculated value, based on 5% CO
cal. gas. (x7 < 2.6 mmHg)
2
x8 = difference between measured CO2 value and calculated value, based on 10% CO
cal. gas.
2
(x8 < ± {0.07 x value calculated})
o
C or 0.2oF).
o
C or 100 oF.
41
Page 42
3 Testing and Maintenance Tests
The value should be 40 oC ± 0.2 oC or 100 oF ± 0.4 oF.
3

ECG/Resp Performance Test

This test checks the performance of the ECG and respiration measurements.
Tools required: Patient simulator.
ECG Performance
1 Connect the patient simulator to the ECG/Resp connector on the measurement server.
2 Configure the patient simulator as follows:
– ECG sinus rhythm.
– HR = 100 bpm.
3 Check the displayed ECG wave and HR value against the simulator configuration.
4 The value should be 100bpm +/- 2bpm.
Respiration Performance
1 Change the Patient Simulator configuration to:
– Base impedance line 1500 Ohm.
– Delta impedance 0.5 Ohm.
– Respiration rate 40 rpm.
2 The value should be 40 rpm +/- 2 rpm.

Invasive Pressure Performance Test

This test checks the performance of the invasive pressure measurement.
Tools required: Patient simulator.
1 Connect the patient simulator to the pressure connector on the MMS or the measurement server
extension.
2 Set the patient simulator to 0 pressure.
3 Make a zero calibration.
4 Configure the patient simulator as P(static) = 200 mmHg.
5 Wait for t he d isp lay.
6 The value should be 200 mmHg ± 5 mmHg. If the value is outside these tolerances, calibrate the
MMS or measurement server extension. If the MMS was calibrated with a dedicated reusable catheter, check the calibration together with this catheter.

SpO2 Performance Test

This test checks the performance of the SpO2 measurement.
Tools required: none
42
1 Connect an adult SpO
2 Measure the SpO
value on your finger (this assumes that you are healthy).
2
transducer to the SpO2 connector on the MMS.
2
Page 43
Tests 3 Testing and Maintenance
The value should be between 95% and 100%.
3

Cardiac Output Performance Test

These tests check the performance of the cardiac output measurement.
1 Connect the patient simulator to the C.O. module using the patient cable.
2 Configure the patient simulator as follows:
Injection temperature: 2 °C Computation Const: 0.542 (Edward's Catheter) Flow: 5 l/min
3 Check displayed value against the simulator configuration.
4 Expected test result: C.O. = 5 +/– 1 l/min.
Service Tool Procedure, Version 1
This procedure applies for Service Tool M1012-61601 and/or C.O. modules without option C10.
1 In monitoring mode, connect the C.O. interface cable to the module.
2 Connect one side of the service tool to the injectate receptacle of C.O. interface cable and the other
side to catheter cable receptacle.
3 Enter the C.O. Procedure window and check the results. The expected test result is:
o
– Tblood = 37.0
C +/- 0.1oC
Service Tool Procedure, Version 2
This procedure applies only for Service Tool M1012-61601 in combination with C.O. modules with option C10.
1 In monitoring mode, connect the C.O. interface cable to the module.
2 Connect one side of the service tool to the injectate receptacle of the C.O. interface cable and the
other side to the catheter cable receptacle.
3 Enter Setup C.O. menu and check results for:
– Method of measurement
– Catheter constant
–Tblood
4 Enter C.O. Procedure window and check results. The expected results are
–Transpulmonary 341
o
– Tblood = 37.0
C +/- 0.1oC

BIS Performance Test

These tests check the performance of the BIS measurement.
PIC/DSC Test
1 In monitoring mode connect the sensor simulator (for maximum usage please refer to the
documentation delivered with the sensor simulator) to the patient interface cable.
43
Page 44
3 Testing and Maintenance Tests
Enter the BIS menu and select Open Window.
2 3 Start impedance check by pressing StartCyclicCheck. Check the displayed results. Expected
results are:
– Electrode 1 (+): 4-6 k – Electrode 2 (Ref): 8-12 k – Electrode 3 (1-): 1-3 k – Electrode 4 (2-): 1-3 k

Nurse Call Relay Performance Test

The nurse call relay performance test can be performed either at the phone jack type connector (this only tests one relay) or at the multi-port nurse call connector (to test all three relays).
Phone Jack Type Connector Test
This test checks the operation of the Nurse Call Relay. The Nurse Call Relay test is recommended for customer sites where the nurse call is in use. The Nurse Call relay functions as follows:
• Standard Operation—Relay open.
• Alarm Condition—Relay closed.
Tools required: Ohmmeter.
1 Plug a phono connector into the Nurse Call Relay connector.
2 Connect the ohmmeter.
3 If no alarm occurs, the relay contacts are open. When an alarm occurs, the relay contacts close.
Multi-Port Nurse Call Connector Test
This test checks the operation of the Flexible Nurse Call Relay. The Nurse Call Relay test is recommended for customer sites where the nurse call is in use. The following diagram and table show the pins and relay identifiers of the connector:
44
Page 45
Tests 3 Testing and Maintenance
Pin Cable Color Coding Relay
1black R2-closure
2brown R2-middle
3red R2-opener
4orange R3-closure
5yellow R3-middle
6 green R3-opener
7blue n/a
8purple n/a
9gray n/a
10 white n/a
11 pink R1-closure
12 light green R1-middle
13 black/white R1-opener
14 brown/white n/a
15 red/white n/a
16 orange/white n/a
17 blue/white R_failure_closure
18 purple/white R_failure_middle
19 green/white R_failure_opener
20 red/black n/a
The Nurse Call relay functions as follows:
• During standard operation R1,R2,R3 _opener are closed; R1,R2,R3_closure are open.
• During alarm condition R1,R2,R3_opener are open; R1,R2,R3_closure are closed.
Tools required: Ohmmeter.
1 Plug an M8087-61001 cable into the Nurse Call Relay connector.
2 Connect the ohmmeter and measure the pins as indicated in the diagram and table.
3 The relay contacts should behave as described above. The behavior may vary depending on
configuration choices. See the Configuration Guide for details on Alarm Relay settings.

ECG Sync Performance Test

This test checks the performance of ECG synchronization between the monitor and a defibrillator. It only needs to be performed when this feature is in use as a protocol at the customer site.
Tools required:
• Defibrillator with ECG Sync and Marker Output.
•Patient simulator.
1 Connect the patient simulator to the ECG connector on the Measurement server and the
defibrillator to the ECG Sync Output on the monitoring.
45
Page 46
3 Testing and Maintenance Tests
Set the patient simulator to the following configuration:
2
– HR = 100 bpm.
– ECG sinus rhythm.
3 Switch the defibrillator to simulation mode.
4 Check that the marker pulse is displayed before the T-wave begins.

VueLink Tests using VueLink Test Module

Use the VueLink plug-in test module ( M1186-60510) to test M1032A VueLink modules.
Test Procedure
You must preselect the test module to ON in Configuration Mode. Therefore, the test module must be one of the devices made available for selection during configuration of the VueLink module.
Carry out the test itself in monitoring mode. For information concerning the configuration of VueLink modules see the M1032A VueLink Module Handbook.
1 Plug the VueLink module into the FMS.
2 Press the Setup key on the front of the VueLink module. 3 In the Setup Vu eL in k menu select Device, then select Test Module. 4 Select Confirm to store the selection and wait for the message “Switched to new device”.
5 Plug in the test module.
6 Connect the modules by plugging one end of the cable (part number M1032-61661) into the
connector on the front of the VueLink Module, and the other end into the connector on the front of the Test Module.
7 Select the wave segment on the screen, where you want the waves to appear. In the wave menu,
select Change Wave, then select WAVE.
8 Select the VueLink SmartKey, then select the TEST Plug-In pop-up key
The test module acts in the same way as an external device would, and sends signals to the VueLink module in both analog and digital form. The computer module checks these signals for validity, and then displays “passed” or “failed” on the screen.
The wave segment displays two waveforms, a triangular one and a rectangular one. These are displayed alternately and for a period of ten seconds each. The expected curve type is indicated below the wave.
There are two pairs of gridlines that indicate the permitted range for the max/min values of these waves. If all the data received by the test module is correct, the waves will lie within the specified ranges. If either limit of either wave falls outside the respective gridlines, then the module being tested is faulty regardless of the passed/failed messages.
When the test is complete:
46
1 Disconnect the cable that joins the test module to the VueLink module.
2 In configuration mode, ensure that the test module is not selected, and the preselected devices are
the same ones as before the test. Also, verify the settings for these devices.
3 Return to monitoring mode.
4 Press the Setup key on the front of the VueLink module and select the required device by selecting
Device in the Setup VueLink menu.
Page 47
Tests 3 Testing and Maintenance
NOTE
It is important to ensure that the preselected device drivers are configured exactly the same as they were before the test, including their default settings.

Safety Testing

You are recommended to file the results of safety tests. This may help to identify a problem early particularly if the test results deteriorate over a period of time.
Warnings, Cautions, and Safety Precautions
• These tests are a proven means of detecting abnormalities that, if undetected, could prove dangerous to either the patient or the operator.
• You can perform all tests using commercially available Safety Analyzer test equipment. You can perform basic measurements with widely available multifunction instruments such as the HP 3469A multimeter or equivalent.
• The consistent use of a Safety Analyzer as a routine step in closing a repair or upgrade is emphasized as a mandatory step to maintain approval agency status. You can also use the Safety Analyzer as a troubleshooting tool to detect abnormalities of line voltage and grounding plus total current loads.
• For Europe and Asia/Pacific according to: IEC60601-1:1988 + A1:1991 + A2:1995 = EN60601-1:1990 +A1:1991 + A2:1995 For USA according to: UL2601-1
• Additional tests may be required according to local regulations.
•Normally, a Safety Analyzer is used to perform these procedures. Popular testers include the DEMPSEY 232D, or for use in Europe, testers like the Rigel, Metron or Gerb. Follow the instructions of the Instrument manufacturer. If the Dempsey is used for an extended length of time, it could be damaged by the high amp current draw of the system.
• Any device with mains connection that is connected to the medical device must comply with IEC60601-1 if within patient vicinity and be separately tested at the same intervals as the monitor
Safety Test Procedures
Use the test procedures outlined here only for verifying safe installation or service of the product. The setups used for these tests and the acceptable ranges of values are derived from local and international standards but may not be equivalent. These tests are not a substitute for local safety testing where it is required for an installation or a service event. If using the Metron Safety tester, perform the tests in accordance with your local regulations, for example in Europe use IEC60601-1/IEC60601-1-1 and in the US use UL2601-1. The Metron Report should print results with the names listed below, together with other data.
NOTE For any system with external displays: Disconnect the display from the medical device and perform
S(1) and S(2) on each device with a mains cable. If both pass the tests reconnect the display and procede with normal use.
47
Page 48
3 Testing and Maintenance Tests
S(1) Part 1: System Enclosure Leakage Current - NC (normal condition)
Medical electrical system
Instrument under test
L (N)
Signal parts
in- and/or
output
Signal parts
(**)
in- and/or
output
Other
Instrument
N (L)
S1
S4
(*)
S2
PE
MD
(*) Not present in Class 2. (**) Can be multiple different connections to different equipment at same time.
Applied part
Expected test results:
© Normal condition maximum leakage current x1 100µA
This measures leakage current of exposed metal parts of Instrument under Test (IUT) and between parts of the system within the patient environment; normal and reversed polarity using S2.
Safety test according IEC 60601-1 / UL2601-1
S(1) Part 2: System Enclosure Leakage current - Single Fault (open earth)
48
Medical electrical system
Instrument under test
L (N)
S2
PE
(**) Can be multiple different connections to different equipment at same time.
Applied part
N (L)
S1
S4
MD
Expected test results:
© Single Fault maximum leakage current x2 500µA (IEC 60601-1)
Signal parts
in- and/or
output
(**)
Signal parts
in- and/or
output
Other
Ins tru m en t
Page 49
Tests 3 Testing and Maintenance
300µA (UL2601-1)
This measures leakage current of exposed metal parts of Instrument under Test (IUT) with Protective Earth (PE) open circuit (S4 = open) and between parts of the system within the patient environment; normal and reversed polarity using S2.
S(2) Protective Earth Continuity
Instrument under test
L (N)
Ri
Ohm
6V
50 Hz 25 A or 1.5 Ir
N (L)
PE
(*)
Insulating pad
(*) If equipotential connection present : measure also with yellow/green E.P. conductor connected.
Applied part
Expected test results:
© With mains cable, maximum impedance x = 100 mOhms (IEC 60601-1 and UL2601-1)
This measures impedance of Protective Earth (PE) terminal to all exposed metal parts of Instrument under Test (IUT), which are for safety reasons connected to the Protective Earth (PE). Test current 25 Amp applied for 5 to 10 seconds.
49
Page 50
3 Testing and Maintenance Touchscreen Calibration
S(3) Patient Leakage current - Single Fault Condition (S.F.C.) mains on applied part
(*) Not present in Class 2
L (N)
S2
(*)
S6
N (L)
S1
S4
L (N)
N (L)
Expected test results:
© Maximum leakage current, x = 50µA @ 250V (IEC60601-1 and UL2601-1)
Measures patient leakage current from applied Part to earth caused by external main voltage on applied Part with switch S5 open and closed. Each polarity combination possible is tested using S2 and S6. This test is applicable for every measurement input .

Touchscreen Calibration

Please refer to the documentation shipped with your selected display for details on touchscreen calibration procedures.
Instrument under test
PE
Insulating pad
S5
Applied part
Signal part in- and/or output
R
MD
50
Page 51

4Troubleshooting

Introduction

This section explains how to troubleshoot the monitor if problems arise. Links to tables that list possible monitor difficulties are supplied, along with probable causes, and recommended actions to correct the difficulty.

How To Use This Section

Use this section in conjunction with the sections Testing and Maintenance and Parts. To remove and replace a part you suspect is defective, follow the instructions in the section Repair and Disassembly. The Theory of Operation section offers information on how the monitor functions.
4

Who Should Perform Repairs

Only qualified service personnel should open the monitor housing, remove and replace components, or make adjustments. If your medical facility does not have qualified service personnel, contact Philips’ Response Center or your local Philips representative.
WARNING High Voltage - Voltages dangerous to life are present in the instrument when it is connected to the
mains power supply. Do not perform any disassembly procedures (other than server and extension removal) with power applied to the instrument. Failure to adhere to this warning could cause serious injury or death.

Replacement Level Supported

The replacement level supported for this product is to the printed circuit board (PCB) and major subassembly level. Once you isolate a suspected PCB, follow the procedures in the Repair and Disassembly section, to replace the PCB with a known good PCB. Check to see if the symptom disappears and that the monitor passes all performance tests. If the symptom persists, swap back the replacement PCB with the suspected malfunctioning PCB (the original PCB that was installed when you started troubleshooting) and continue troubleshooting as directed in this section.
51
Page 52
4 Troubleshooting Software Revision Check

Software Revision Check

Some troubleshooting tasks may require that you identify the Software Revision of your monitor. You can find the software revision along with other information, such as the system serial number, in the monitor revision screen.To access the monitor revision screen:
1 Enter the Main Setup menu and select Revision 2 Select the pop-up key Software
NOTE The system serial number can also be found on the right side and on the bottom of the MP90 CMU.

Obtaining Replacement Parts

See Parts section for details on part replacements.

Troubleshooting Guide

Problems with the monitor are separated into the categories indicated in the following sections and tables. Check for obvious problems first. If further troubleshooting instructions are required refer to the Troubleshooting Tables.
Taking the recommended actions discussed in this section will correct the majority of problems you may encounter. However, problems not covered here can be resolved by calling Philips Response Center or your local representative.

Checks for Obvious Problems

When first troubleshooting the instrument, check for obvious problems by answering basic questions such as the following:
1 Is the power switch turned on?
2 Is the AC power cord connected to the instrument and plugged into an AC outlet?
3 Are the MMS and, if present, the measurement server extension inserted correctly?
4 Are the cables connected properly to the FMS?
5 Are the parameter modules plugged into the FMS correctly?

Checks Before Opening the Instrument

You can isolate many problems by observing indicators on the instrument before it is necessary to open the instrument.
NOTE It takes several seconds for the AC Power LED to switch on / off after the mains power cord has been
connected / disconnected.
Checks with the Instrument switched Off
• AC connected:
– AC Power LED is on (green).
• No AC connected:
52
Page 53
Troubleshooting Guide 4 Troubleshooting
–All LEDs are off.
Checks with the Instrument Switched On, AC connected
When the monitor is first switched on, all the LEDs on the power supply and the remote alarm device light up momentarily. The location of the power supply LEDs is shown in the following photograph:
AC Power LED
(Green)
Error LED
(Red)
Power On LED
(Green)
Initial Instrument Boot Phase
The following tables describe the regular initial boot phase of the monitor and its components. If the boot phase does not proceed as described below go to Boot Phase Failures for Troubleshooting information.
Monitor Boot Phase:
For these steps it is assumed that the Monitor is powered correctly and the +5V System Board supply voltage is okay. This is indicated by the green Power On LED.
Time (sec.) after Power On
0 When the Power On/Off button is pressed, the green Power On LED and the red error LED
1 The alarm LEDs on the remote alarm device are switched on with low intensity. Colors: Left
3 Red Error LED is switched off.
4 Boot Screen with the Philips Logo appears on the display. Test Sound is issued.
5 All Alarm LEDs are switched off.
8 Fixed screen elements (for example smart keys, alarm fields) appear on the screen.
15-30 First measurement information appears on the screen,user input devices (for example
Event
switch on immediately.
LED:cyan; Middle LED:red; Alarm Suspend LED (right): red
Boot Screen with the Philips Logo disappears
Mouse, Touch, Speed Point) are functional
53
Page 54
4 Troubleshooting Troubleshooting Guide
Flexible Module Server Boot Phase
For these steps it is assumed that the Flexible Module Server is connected via MSL-cable to the monitor
Time (sec. ) after Monitor Power On
0 Red Error LED switches on immediately
1 Green “Ready” LED switches on
3 Red Error LED is switched off
5 Module Power is switched on
5-8 Module Status LEDs blink once or twice (Module dependent)
Event

Troubleshooting Tables

The following tables list troubleshooting activities sorted according to symptoms. Click on the links below to view a particular table.
How to use the Troubleshooting tables
The possible causes of failure and the remedies listed in the troubleshooting tables should be checked and performed in the order they appear in the tables. Always move on to the next symptom until the problem is solved.
Boot Phase Failures
External Display is blank
External Touch Display not functioning
Remote Alarm Device
Remote Extension Device
Keyboard/Mouse not functioning
Network related problems
Wireless Network
Multi-Measurement Server
MSL-related problems
Alarm Lamps
Alarm Tones
Individual Parameter INOPS
Flexible Module Server
Printer
MIB / RS232
Flexible Nurse Call Relay
54
Page 55
Troubleshooting Guide 4 Troubleshooting
Boot Phase Failures
Symptoms Possible Causes of Failure Failure Isolation and Remedy
AC LED does not light up
Green Power On LED and Red Error LED remain off after pressing power on button on power supply
AC Connection not ok Check that the AC-Mains are powered and
the power cord is ok and connected
LED defective
Power Switch board not connected to the main board
Power supply defective
Secondary main board defective
Primary Video Board defective
Backplane Bottom defective
Primary Main Board defective
Remote Devices
Power Switch Micro Controller hung
Power switch board not connected to the main board
Power Switch Board defective
I/O Board defective
Primary Video Board defective
Primary Main Board defective
Try to switch on the monitor. If it operates normally , the LED is defective => exchange Power Switch board.
Check if power switch board is connected correctly to the Main Board
Remove power supply and check if output voltage is within the specifications (47V ­49V). Measure on multi-colored wired connection between red and black wires Exchange power supply if defective
Disconnect cable from the backplane top to backplane bottom.
If failure persists continue with ‘Primary Video Board defective”
Otherwise: Reconnect cable, remove secondary video board and check again.
Exchange main board and check again.
Replace backplane top.
Remove primary Video Board and check again
Remove Backplane Bottom and check again
Exchange primary Main Board
Disconnect all connections to the remote devices and try to switch on the monitor again
Unplug AC Mains and replug after 10 seconds. Try to switch on the monitor again.
Check if power switch board is connected correctly to the primary main board.
Exchange Power Switch BOard and try to switch the monitor on again.
Remove all I/O boards and try to switch the monitor on again
Remove primary video board and try again
Exchange primary main board. Add boards in reverse order and try again with each board.
55
Page 56
4 Troubleshooting Troubleshooting Guide
Symptoms Possible Causes of Failure Failure Isolation and Remedy
Green Power On LED and Red Error LED remain off after pressing power on button on remote extension device or remote alarm device.
Green On/ Standby LED or Red Error LED remain off after pressing Power on button:
Red Error LED stays on continuously
Red Error LED blinks (indicating cyclic reboots)
Cable to remote device
Remote device
Power switch board not connected to the main board
Remote device interface board
Power Switch Board not connected to the main board
Power Switch board defective
Main board defective
External connected device defective disconnect all external cables (except AC)
I/O Board defective
Backplane top defective or secondary main board defective (opt. #E30)
Primary Video Board defective
Backplane Bottom defective
Primary Main board defective
Hardware Failure
Software Fault
Hardware Failure
Check if cable to remote device is connected correctly to the remote device interface board and the remote device.
Use another remote device on this port of the remote device interface board and try to switch on the monitor again.
Check if power switch board cable is connected to the primary and secondary main board (opt.#E30)
Check if monitor can be switched on by the power switch on the power supply. If this is possible replace the remote device interface board. Otherwise proceed as described in “Green Power On LED and Red Error LED remain off after pressing power on button on power supply” above.
check if power switch board is connected correctly to the main board
exchange Power switch board
exchange main board
and switch the monitor on again
Remove all I/O boards and switch the monitor on again.
Disconnect cable from the backplane top to the backplane bottom.
Remove primary video board and switch on again
Remove Backplane Bottom and check again.
Exchange primary Main board
connect Support Tool directly to monitor with crossover cable and start “search for defective devices”
If no device is detected, proceed as described above in section “Red error LED stays on continuously”
If the Support Tool can detect the device and it indicates the Operating Mode is ‘Boot’, download and store the status log. Reload software and re-clone the monitor. If this fixes the problem e-mail the status log to your local response center
If this does not rectify the problem follow instructions under “Red Error LED stays on continuously”.
56
Page 57
Troubleshooting Guide 4 Troubleshooting
Symptoms Possible Causes of Failure Failure Isolation and Remedy
No Test Sound issued
Speaker defective
Main board defective
check for INOPs and follow instructions
exchange speaker
exchange main board
57
Page 58
4 Troubleshooting Troubleshooting Guide
External Display is blank
Symptoms Possible Causes of Failure Failure Isolation and Remedy
External Display is blank Video cable to external display
not connected
External display has no power
External display is defective
Video board defective
Main board defective
External Touch Display not functioning
Symptoms Possible Causes of Failure Failure Isolation and Remedy
Touch Screen not functioning
Touch position invalid Touch not calibrated Perform touch calibration:
External Touch cable not connected Check cable connection from
External Touch driver configuration
MIB Board defective
External touch defective
Main board defective
Check video cable connection to external display
Check electricity supply of external display
Check external display and video cable on another monitor or PC
Replace video baord
Replace main board
external touch to MIB board
Check RS232/MIB configuration:
1. Enter Main Setup menu
2. Select Monitor
3. Select Hardware
4. Reconfigure RS232/MIB drivers
5. if problem persists, proceed to the next step
Replace MIB board
Replace external touch
Replace Main board
1. Enter Main Setup menu
2. Select Monitor
3. Select Hardware
4. Select Touch Driver
5. Select Calibrate
58
Page 59
Troubleshooting Guide 4 Troubleshooting
General Monitor INOP Messages
INOP Message Possible Causes of Failure Failure Isolation and Remedy
CHECKINTERNVOLTAGE Problem with the voltages (5V, 12V) in
the monitor
CHECK MENU LABELS
This message indicates that the monitor has performed a cold start and attempted to reset the menu labels to defaults. If the defaults are present, the recovery was successful. Default settings should be recorder as part of your documented configuration setup. If the menu labels do not match the configuration file settings or the INOP persists, perform the troubleshooting tasks listed in this table.
CHECK MONITOR TEMP The temperature inside the monitor is
CHECK SETTINGS
This message indicates that the monitor has performed a cold start and attempted to reset the menu labels to defaults. If the defaults are present, the recovery was successful. Default settings should be recorded as part of your documented configuration setup. If the menu labels do not match the configuration file settings or the INOP persists, perform the troubleshooting tasks listed in this table.
Memory space in which the menu labels are stored has been corrupted
Main Board defective
too high
Monitor ventilation obstructed
Main Board defective
Memory space in which the settings are stored has been corrupted
Main board defective
Remove all I/O boards and put them back in one at a time to isolate any defective board. If this does not resolve the problem, replace the main board
Reclone configuration file. This will reload the memory space
Replace Main Board
Check the environment for possible causes
Clean the monitor ventilation internally and then cool monitor down for 8 hours
replace Main Board
Reclone configuration file. This will reload the memory space.
Replace Main board
59
Page 60
4 Troubleshooting Troubleshooting Guide
INOP Message Possible Causes of Failure Failure Isolation and Remedy
SETTINGS MALFUNCTION
INTERNAL COMM.MALF. Problem with the I2C Bus
Problem during cloning process. Reclone configuration file
Memory space in which the settings are stored has been corrupted
Main board defective
communication in the monitor
Video board defective
Main board defective
Remote Alarm Device
Symptoms Possible Causes of Failure Failure Isolation and Remedy
Alarm LEDs illuminate, but no alarm sound is issued
Alarm occurs on screen, but no LED or alarm sound on the alarm device
Alarm sound is isued, but no LEDs light up
wrong I/O slot check I/O matrix in the Theory
speaker defective
remote device I/F defective
cabling not connected check cabling
cabling defetive
I/O board defective
Remote Alarm Device defective
LED failure Replace Alarm Device
Reclone configuration file. This will reload the memory space.
Replace Main board
Disconnect the external display and try another one
Replace Video board
Replace Main board
of Operation section of this manual
replace remote alarm device
replace I/O board
replace cable
replace I/O board
replace Remote Alarm Device
Remote Extension Device
Symptoms Possible Causes of Failure Failure Isolation and Remedy
Remote input device (for example mouse/keyboard) attached to the Remote Extension Device does not function
Buttons on the Remote Extension Device do not function but input device attached is functioning
60
See table Keyboard/Mouse not functioning
Remote Extension Device defective
See tables Keyboard/Mouse not
functioning
replace Remote Extension Device
Page 61
Troubleshooting Guide 4 Troubleshooting
Symptoms Possible Causes of Failure Failure Isolation and Remedy
Speed Point attached to Remote Extension Device not functioning
Speed Point Knob Rotation, Joystick Control or Selection control not functioning
INOP Message CHECK INPUT DEVICES is issued
Keyboard/Mouse not functioning
Remote Extension Device is not connected to the monitor.
SpeedPoint not connected properly
SpeedPoint defective
Remote Extension Device defective
Remote Device I/O board in the wrong slot
Remote Device I/O board defective
SpeedPoint defective
SpeedPoint or other input device defective
Check cabling and connections
Check cabling to SpeedPoint in the Remote Extension Device
Replace SpeedPoint
Replace Remote Extension Device
Check I/O Matrix in Installation
Instructions
Replace I/O board
Replace Speed Point
Perform a visual and functional check of all the monitor input devices. Replace input devices if necessary.
Symptoms Possible Causes of Failure Failure Isolation and Remedy
Keyboard/Mouse attached directly to the monitor not functioning
Keyboard/Mouse attached to Remote Extension Box not functioning
Keyboard/Mouse not connected properly
Keyboard/Mouse defective
PS/2 I/O board in wrong slot
PS/2 I/O Board defective
Remote Extension Box is not connected to the monitor or Input Device is not connected to Remote Extension Box
Keyboard/Mouse defective
Remote Extension Device defective
Remote Device I/O board in wrong slot
Remote Device I/O board defective
Check cabling
Replace Keyboard/Mouse
Check I/O Matrix in “Theory of Operation”
replace I/O board
Check cabling and connections
Replace Keyboard/Mouse
Replace Remote Extension Device
Check I/O Matrix in Theory of
Operation
Replace I/O board
61
Page 62
4 Troubleshooting Troubleshooting Guide
Network related problems
Symptoms Possible Causes of Failure Failure Isolation and Remedy
Prompt Message “no central assigned to this bed” is issued
INOP “Unsupported LAN” is issued.
No connectivity to PIC, no prompt or error message on monitor
Other Bed Overview not available
“Other Bed” Alarms are not appearing
The monitor label is not set in the monitor (if the beds are “monitor labeled” in the Philips Information Center)
Problem with the Philips Information Center to Switch communication (if the beds are “port mapped” in the Philips Information center
Network failure Check if switches, Philips
Monitor connected to wrong network
IP address conflict after infrastructure re-installation
Hardware Defect Check LAN cable connection
Configuration problem
Configuration Problem Check configuration in PIC
This function is not available for wireless beds
Configuration problem Verify configuration in PIC, in
Set Monitor Label in Config Mode
Check PIC to Switch communication, Switch configuration and Firmware status
Information Center and Database Server are all running and connected to the network
Check if monitor has been connected for example to a different hospital network instead of the Philips Clinical Network
Reboot Database Server and Philips Information Center
Check NGN Connector board in Monitor
Check Switch
Check switch configuration and firmware revision
regarding other bed overview (care group assignment)
Verify configuration of switch (setting of multicast filters)
Switch to a wired configuration
Monitor (Config Mode) and check that the feature is not temporarily disabled by the user (Bed Info Window)
62
Page 63
Troubleshooting Guide 4 Troubleshooting
Wireless Network
Symptoms Possible Causes of Failure Failure Isolation and Remedy
No central monitoring possible (see also Network related problems)
No connectivity (coverage area consists of multiple access points and in some parts of the area there is no connectivity)
Frequent dropouts and network disconnects
Overview, Printing does not work
Monitor is out of range of the access point (in this case the yellow sync LED on the wireless adapter on the bottom of the device next to the power connector is not on steady)
Wireless Adapter has no power (LEDs on adapter are all off)
(Only after first install) Firmware revision in adapter is wrong
Wireless adapter defective (the red status LED on the adapter’s top panel is on)
Wrong configuration in wireless adapter or in access point
Configuration problem
Excessive interference by other radio equipment or by microwave ovens
System capacity exceeded in coverage area
Configuration problem
Some functions are not available on a wireless network
Move monitor back into coverage area
Verify size of coverage with the site survey tool
Check splitter cable and replace if necessary
Check network adapter board in monitor and replace if necessary
Check adapter itself. Replace if necessary
Update adapter firmware with wireless support tool
Replace wireless adapter
Check configuration with wireless support tool
Verify the channel, domain and security ID settings of the access points in the coverage area
Check statistics that can be read from the wireless adapter via RS232 or via logging application in the PIC. Remove interfering equipment.
Check configuration guidelines for number of monitors per access point.
Check access point configuration with wireless support tool, in particular the multicast filters
Connect to cabled network
63
Page 64
4 Troubleshooting Troubleshooting Guide
Multi-Measurement Server
Symptoms Possible Causes of Failure Failure Isolation and Remedy
Prompt message “Measurement Server Configuration not supported” is issued
INOP Message “MsmtSrv not Supp” is issued
Prompt message “Measurement Server not supported, unplug device, switch monitor off/on” and INOP “Bad Measurement Server are issued
MSL-related problems
An unsupported Measurement Server Extension has been connected
Measurement Server Extension is defective
Measurement Server defective
Wrong So f tware Re v ison Upgrade monitor and/or
Too many measurement servers connected
Unsupported type of measurement server (for example M3000A on a M800xA monitor) connected.
M3000A Measurement Server Revision A is plugged. This Measurement Server is not compatible with the IntelliVue patient monitors
Disconnect the measurement server extension
Replace Measurement Server Extension
Replace Measurement Server
measurement server to a matching software version
Disconnect unsupported measurement servers for proper operation
Disconnect the unsupported measurement server
Disconnect the measurement server and cycle power.
Symptoms Possible Causes of Failure Failure Isolation and Remedy
Measurement Server does not start up (no LEDs active), no INOP or prompt displayed
Measurement Server does not start but LEDs are normal
MSL Power High INOP is issued
Note: if this condition persists for longer than 15 minutes, the INOP MSL Power Off will appear (see below)
No Power
Communication lines in MSL cable or MSL connector broken
MSL connector board defective
Attached devices drawing too much power from the monitor. Too many FMS and MMS connected to the monitor
Check MSL cable and replace if necessary
Check MSL connector board and replace if necessary
Check MSL cable and MSL connectors
Check MSL connector board and replace if necessary
Reduce to a limit of 1 FMS and 1 MMS connected to the monitor
64
Page 65
Troubleshooting Guide 4 Troubleshooting
Symptoms Possible Causes of Failure Failure Isolation and Remedy
MSL Power Off INOP is issued
MSL Power Overload INOP is issued
INOP BAD SERVER LINK is issued
INOP Message SERVERLINK MALF is displayed, audible indicator: a beep every two seconds
Attached devices drawing too much power from the monitor. Too many FMS and MMS connected to the monitor
Short Circuit within MSL system Disconnect all MSL connections,
Unexpected data detected on MSL Check cable and power cycle the
An FMS or MMS with an incompatible software revision is connected to the monitor.
Communication between the components not functioning
The hardware for communicating with the Multi-Measurement Server is faulty.
Disconnect all FMS and MMS from the monitor
Cycle power to restore power to the MSL devices.
If the message disappears, reconnect FMS and MMS one at a time, waiting 15 minutes between each device to see if message reoccurs. If yes, the respective MMS or FMS is faulty. See Multi-Measurement Server or Flexible Module Server for troubleshooting tasks. If no, add front-end modules one at a time, waiting 15 minutes between each module to see if message reappears, Replace module if faulty.
Note: If an individual defective device is connected the MSL Power High o r MSL Power Overload INOPs will appear initially. The MSL Power Off INOP will not occur for at least 15 minutes.
2nd MSL interface, 1st MSL interface. Reconnect devices one at a time. If message persists, replace main board.
monitor
Connect FMS or MMS with compatible software revsion
Check software versions and model number of devices for compatibility
Check MSL cable, replace if necessary.
Check MSL I/O board. Replace if necessary
Check FMS or MMS connector board. Replace if necessary.
65
Page 66
4 Troubleshooting Troubleshooting Guide
Symptoms Possible Causes of Failure Failure Isolation and Remedy
Alarm Issues
Alarm Lamps
A measurement supported by a server does not come up on the monitor
Prompt message “Too many <label> modules connected” is issued
The ECG Out/ Marker In function does not funcion
Label conflict A parameter label from this
measurement is already in use in the monitor. Check the conflict window to select the measurement.
There are more modules of the type <label> connected than supported by the software
Hardware problem Check MSL cable
Remove the unsupported module or use the lebel manager application in the monitor to disable the module.
Check ECG Out Hardware in the monitor
Check the MSL connector in the measurement server
Symptoms Possible Causes of Failure Failure Isolation and Remedy
INOP Message Check Alarm Lamps is issued
Alarm occurs, but no LED lights up
Remote Alarm Device is not connected to the monitor
Remote Alarm Device defective
Remote Device I/O board in the wrong slot
Remote Device I/O board defective
Main board defective
Environmental Lighting too bright Place the alarm device in a darker
Remote Alarm Device is not connected to the monitor
Remote Alarm Device defective
Remote device I/O board in the wrong slot.
Remote device I/O board defective
Main Board defective
Check cabling and connections
Replace Remote Alarm Device
Check I/O Matrix in Installation
Instructions
Replace I/O board
Replace main board
environment
Check cabling and connections
Replace Remote Alarm Device
Check I/O matrix in Installation
Instructions
Replace I/O board
Replace Main board
66
Page 67
Troubleshooting Guide 4 Troubleshooting
Alarm Tones
Symptoms Possible Causes of Failure Failure Isolation and Remedy
INOP Message SPEAKER MALFUNCTION is displayed
Alarm occurs but no alarm sound is issued
Alarm occurs on device connected to VueLink but no alarm sound is issued on the monitor
Alarm sound is issued on monitor but not on Remote Alarm Device
Speaker cable disconnected Reconnect speaker cable
Speaker defective
Sound amplifier on main board defective
Audible alarm indicators have been switched off
Volume set to 0
Speaker defective
Sound amplifier on main board defective
Configuration of VueLink is incorrect
Remote Alarm Device is not connected to the monitor.
Remote Alarm Device defective
Remote Device I/O board in the wrong slot
Remote Device I/O board defective
Replace speaker
Replace main board
Switch audible alarm indicators back on
Increase volume
Replace speaker
Replace main board
Check VueLink configuration
Check cabling and connections.
Replace Remote Alarm Device
Check I/O matrix in Installation
Instructions
Replace I/O board
Alarm Behavior
If your monitor did not alarm in the way in which the end user expected, please consult the Instructions for Use for possible setup issues or configuration settings which could affect alarm behavior.
Individual Parameter INOPS
If any of the following parameter INOP messages are issued try the respective parameter in another device. If the INOP message persists replace the parameter module, the MMS or other indicated device.
•CO
•ECG EQUIP MALF
•NBP EQUIP MALF
•P1 EQUIP MALF
• RESP EQUIP MALF
•SpO
•SpO
•SvO
•tcpO
EQUIP MALF
2
EQUIP MALF
2
TRANSDUC MALF
2
EQUIP MALF
2
(or tcpCO2) EQUIP MALF
2
67
Page 68
4 Troubleshooting Troubleshooting Guide
•T1 EQUIP MALF
• VueLnk EQU. MALFI
Flexible Module Server
Symptoms Cause of Failure Failure Isolation and Remedy
Prompt Message “Unrecognized Measurement Module in slot s” is issued
Prompt message “Measurement Module in slot n is currently ignored” is issued
Red Error LED stays on Unrecoverable hardware selftest
Red Error LED flashes Hardware selftest error If system comes up, check status
Flexible Module Server LEDs ok, Front End Measurement Module not recognized (no prompt or INOP)
An unsupported module has been plugged into the Flexible Module Server
Too many modules of the same kind have been plugged into the Flexible Module Server
error:
MSL cable defective
Flex connector from main board to MSL defective
CPU module defective
Measurement Module or Measurement Module Connector defective
No Front-End power because MSL voltage from the monitor is too high or too low
Mother board or connector on Mother Board defective
Unplug the unsupported module
Unplug module in slot n
Try to attach the MMS directly to the MSL cable. If the measurements show up on the screen, the fault is in the FMS
If the measurements do not show up when the MMS is connected directly to the MSL cable, then replace MSL cable
Replace connector
Replace CPU module
log. Otherwise see above
Replace Measurement Module
Try a new MSL cable. Replace if failure is rectified.
Replace mother board
If the voltage is in range, or there is obvious damage to a connector, replace mother board
68
Page 69
Troubleshooting Guide 4 Troubleshooting
Printer
Symptoms Cause of Failure Failure Isolation and Remedy
Prompt message “Print job could not be queued” is issued. No print device is found.
Status message “Print device Local 1 (Local 2) unavailable” is issued. Printer job is stalled.
Status message “Print device Remote 1 (Remote 2, Remote 3) unavailable” is issued. Printer job is stalled
Status message “Printing on device Remote 1... (Remote 2, Remote 3)” is issued but no report is printed
Printouts are not as expected
Printer is disabled in the Setup Printers menu
Paper size of printer does not match paper size of report
Printer not switched on
Printer paper tray empty
Cabling not connected correctly
I/O board defect
Print error on Philips Information Center
Network Connection to Philips Information Center not functioning
Print queue on Philips Information Center is full. Reasons for this may be:
- Printer is not switched on
- Printer paper tray is empty
Printer paper size is not correctly configured
Printer resolution is not correctly configured
Printer color support is configured to “On” although the printer does not support color
Printer not compatible
Enable the correct printer in the Setup Printers menu
Change paper size of the printer in the Setup Printers menu or change paper size of the report in the Setup Reports menu.
Switch on printer power
fill printer paper tray
Check cabling
replace I/O board
Print a test report on the Philips Information center. If this fails, refer to Philips Information Center documentation
Check that the network connection between the monitor and the Philips Information Center is working
Switch on printer power
Fill printer paper tray
Configure the paper size according to the inserted print media
Configure the printer resolution according to the printer capabilities
Configure the printer color support to “Off ”
Check specifications
69
Page 70
4 Troubleshooting Troubleshooting Guide
Recorder
Symptom Possible Cau se Corrective Action
System thinks that door is open when it is not.
System thinks that the recorder is out of paper when it is not.
Recorder not communicating with System.
Recorder won't run. Recorder interface not
Poor print quality. Printhead dirty. Clean the Printhead.
Paper not feeding properly.
Module does not lock into FMS.
Defective door switch. Replace door switch. Exchange module.
Paper-out sensor dirty. Clean paper-out sensor.
Poor connection to the front-end FMS.
Only one recorder module may be used with each monitor.
System not configured properly.
Too many modules connected.
working correctly.
Printhead failure.
Paper roll off center. Center paper roll on roller guides.
Dirty roller.
Locking plates defective. Remove and exchange the locking plates.
Unplug the module. Plug it back in and try it again in a few seconds. (Watch for the LED to flash.)
Remove one of the recorder modules.
Check the configuration of the connected monitor.
Check and remove the extra modules.
Unplug the module. Plug it back in and try it again in a few seconds. (Watch for the LED to flash.)
Exchange the module.
Clean roller.
MIB / RS232
70
Symptoms Cause of Failure Failure Isolation and Remedy
AGM connected to an RS232 port not functioning
The MIB/RS232 port is not configured for AGM
The cable between AGM and monitor is not connected correctly or defective
The MIB/RS232 board is in a wrong slot (slot has been changed after software configuration or an additional board has been plugged in)
The MIB/RS232 board is defective
Check configuration of the MIB/ RS232 ports in configuration mode
Check cable connection, replace cable if necessary
Verify correct placement of the I/ O boards
Check board and replace if necessary
Page 71
Troubleshooting Guide 4 Troubleshooting
Symptoms Cause of Failure Failure Isolation and Remedy
External device not receiving data
Detailed Protocol Problem Consult the Data Export Protocol
The MIB/RS232 port is not configured for data export
The wrong data export protocol driver is configured in the monitor
The cable between the external device and the monitor is not connected correctly or defective
The external device does not support the version of the data export protocol used in the monitor
A terminal concentrator is used in between the device and the monitor and a protocol with dynamic speed negotiation is used
The MIB/RS232 board is in a wrong slot (slot has been changed after software configuration or an additional board has been plugged in)
The MIB/RS232 board is defective
Check configuration of the MIB/ RS232 ports in configuration mode
Check the export protocol required by the attached device and configure the monitor accordingly
Check cable and replace if necessary
Check if the device supports the version of the data export protocol. Upgrade device or monitor if necessary (if matching versions exist).
Some terminal concentrators do not support changing the transmission speed (baud rate) dynamically. Check if the connection works without the concentrator
Verify correct placement of the I/ O boards
Check board and replace if necessary
document.
Flexible Nurse Call Relay
Symptoms Cause of Failure Failure Isolation and Remedy
INOP message CHECK NURSE CALL RELAY is issued
Monitor alarmed, Nurse Call did not activate
Nurse Call Relay board defective Replace Nurse Call Relay I/O
board.
Incorrect configuration (Relay latency, Relay trigger)
Connection of cable to monitor or nurse call system not correct
Nurse Call Relay board is in the wrong slot.
The Nurse Call Relay bosrd is defective
Check monitor configuration (see configuration guide)
Check cable connections
Verify correct placement of the I/O boards
Replace Nurse Call Relay board
71
Page 72
4 Troubleshooting Troubleshooting Guide
Troubleshooting the ECG OUT
Symptoms Cause of Failure Failure Isolation and Remedy
No marker pulse is displayed on the Monitor or no ECG- OUT signal to the Defib
Disconnect the MMS and Defib cable.
Switch the Monitor off then on again. Observe the red LED in the ECG OUT section. (Note that the LED can only be observed if the housing bottom is removed).
If the red LED does not switch on for about 1 second at power on:
Cabling not connected
ECG OUT board defective Replace ECG OUT board
ECG OUT Board defective
Main Board defective Replace Main Board
ECG OUT board defective
Main Board defective Replace main board
Check cabling
If the red LED switches on and remains on for more than 20 seconds:
Replace ECG OUT Board
Connect Known good Defib, Defib cable MMS and MMS cable. Check Marker pulse and ECG OUT signal at defib again. If there is still no signal:
Replace ECG OUT board
Data Flow Marker In and ECG Wave
The following illustration of the data flow for Marker In and ECG Wave may assist in troubleshooting.
72
Page 73
Troubleshooting Guide 4 Troubleshooting
Defib Cable
Isolat.
ECG_OUT Board
CPU,Communication System and Video
MSL Connector
LAN
Display Assembly
ECG W a ve
Marker
DEFIB
Marker In Data Path
Conn.
RS485
Monitor System Board
ECG Wave Data Path

Status Log

Marker merged to ECG Wave here
Server Application
MEASUREMENT
SERVER
Software
Many events that occur during start-up or regular monitoring are logged in the Status Log. The Status Log can be printed and cleared. Not all entries in the Status Log are errors.
Monitor H 1720 20050 1 4 Apr 02 16:37 C 1721 21050 1 4 Apr 02 15:37
73
Page 74
4 Troubleshooting Troubleshooting Guide
The window title is either Monitor or MeasServ, dependent on which system component’s status log is currently displayed.
The Status Log window shows logged events which caused a reboot of the system component (monitor or measurement server).
The first column in the log identifies the event class (“C”: caused a cold start, “H”: caused a hot start, “N”: no retstart, for information only). Column 3 and 4 identify the event source and event code. Column 4 counts the number of occurrences of the event. The last column shows the time and date of the last occurrence of the event.
The following pop-up keys overlay the SmartKeys:
Clear
StatLog
Clear StatLog
This key clears the currently displayed Status Log
Revision
This key switches to the Revision Screen of the currently displayed system component
M8010
This key switches to the Monitor Revision Window
M8048
This key switches to the Flexible Module Server (FMS) Revision Window
M3001
This key switches to the Multi Measurement Server (MMS) Revision Window
If an event occurs repeatedly, contact your Philips Service Representative.
NOTE It is possible, using the support tool, to download the status log and send it to your Philips Service
Representative as a file (for example via e-mail).
Revision M8010 M8048 M3001

Troubleshooting with the Support Tool

Using the support tool you can:
• access the full status log which can be saved as a file
•reload software
• identify defective devices
• reset touch screen calibration
For details on how to perform these tasks see the Support Tool User Manual.
74
Page 75
Troubleshooting Guide 4 Troubleshooting

Troubleshooting the Individual Measurements or Applications

For problems isolated to an individual parameter or application such as event review, please consult the Instructions for Use and configuration information.
If the instructions for use did not resolve an individual parameter problem, then another module or measurement server should be tried.
If you are getting questionable readings for individual measurements you may want to do the Performance Verification tests in the Testing and Maintenance section.
The performance of the individual applications (event review, arrhythmia, trending) are affected by the configuration of the monitor. When contacting Philips support you may be asked about the configuration of the monitor to aid in troubleshooting.
75
Page 76
4 Troubleshooting Troubleshooting Guide
76
Page 77

5Repair and Disassembly

The following section describes the disassembly and reassembly procedures for the monitor and its components

Tools Required

• Torx screwdriver (size 10)
• Torx screwdriver (size 20)
• ESD mat and wrist strap
• 2 small flat blade screwdrivers

MP90 CMU Disassembly

5
NOTE All screws for the MP90 are the same size Torx. We recommend that you keep them separated and
counted as you disassemble and reassemble the device.

Removing I/O Boards

1 Pull off the cable management cover at the rear of the Computer Module Unit (CMU).
77
Page 78
5 Repair and Disassembly MP90 CMU Disassembly
2 Pull off the Power Cord Cover at the front of the CMU and make sure the power cord is
unplugged.
CAUTION Make sure to unplug the AC power cord before removing the I/O boards. Failure to remove the power
cord could cause damage to the CMU as it is still under power.
78
Page 79
MP90 CMU Disassembly 5 Repair and Disassembly
Unscrew the four screws at the rear of the CMU and remove the metal I/O board cover.
3
4 Use the board removal tool located inside the cable management cover to remove the I/O boards.
Make sure to insert the end with the hook underneath the I/O board.
Board Removal
Hook

Removing the Top Cover

1 Flip over the CMU so the bottom faces upwards. With single CPU units it may be helpful to
reinsert the I/O board cover without screws before flipping over the CMU in order to have a more stable platform.
Tool
79
Page 80
5 Repair and Disassembly MP90 CMU Disassembly
Remove the four screws next to the plastic feet.
2
3 Flip the CMU back over, remove the I/O board cover if still present, and slide off the top cover
towards the rear.
Reassembly Note: Insert the I/O board cover before flipping over the CMU to aid with reinserting screws.
80
Page 81
MP90 CMU Disassembly 5 Repair and Disassembly

Removing the Plastic Feet and/or the Locking Cam

Some mounting options may require the removal of the plastic feet or the locking cam.
1 Place the CMU upside down and remove the screws securing each of the feet and the locking cam.
Plastic
Plastic Feet
Feet
Locking Cam
CAUTION When you remove the feet and/or the locking cam you may find there are standoffs to increase the
height under the CMU. You must refit these standoffs with the plastic feet and the locking cam or use shorter screws to refit the feet and the locking cam alone. Screws cannot go more than 5 mm inside the CMU (measured from the exterior of the metal chassis). Inserting screws which are too long into the CMU will damage the main board.
81
Page 82
5 Repair and Disassembly MP90 CMU Disassembly

Removing the optional Fans (Dual CPU Versions only)*

1 Remove the top cover.
2 Disconnect the fans and pull them out from the sheet metal.
Remove Latch
from metal
Fan Connector
* The MP90 dual CPU version may have fans added during installation. The fans are not mandatory and do not ship with the product.
82
Page 83
MP90 CMU Disassembly 5 Repair and Disassembly

Replacing the Second (Independent) Video Board (Dual CPU Versions only)

1 Remove the 5 screws and take out the video board, unplugging it from its connector.
Lift board here to unplug connector
Connector
83
Page 84
5 Repair and Disassembly MP90 CMU Disassembly
NOTE
If not already separated, you may need to break apart the two pieces of the repaired/exchange board.

Removing the Second CPU/Main Board (Dual CPU Versions only)

1 Remove the second independent videoboard as described in Replacing the Second (Independent)
Video Board (Dual CPU Versions only).
2 Remove the top three I/O boards.
84
Page 85
MP90 CMU Disassembly 5 Repair and Disassembly
Remove the white fan connector and the screw securing the main CPU interconnection board.
3
4 Pull off the main CPU innterconnection board in the direction shown.
Interconnection Board Connector
NOTE If you wih to remove the interconnection board entirely, you must also disconnect the interconnection
board connector.
85
Page 86
5 Repair and Disassembly MP90 CMU Disassembly
Remove the 10 screws and pull the mainboard straight up to remove it. Please note that there are
5
sensitve components near the screws on the bottom of the main board.

Accessing the Main CPU or Primary Video Board (Dual CPU Versions)

NOTE If the second independent video board or the second CPU board do not require replacement, you do
not need to remove them to access the first CPU and primary video boards.
1 Unplug the interconnector board connector, the fan connector and the power switch cable
connector.
Interconnector Board Connec tor
Fan Connector
Power Switch Cable
Connector
86
Page 87
MP90 CMU Disassembly 5 Repair and Disassembly
Remove the six screws and take off the upper sheet metal plate. (The top three I/O boards do NOT
2
have to be removed).
87
Page 88
5 Repair and Disassembly MP90 CMU Disassembly

Replacing the Primary Video Board

1 Remove the five screws securing the primary video board.
2 Lift up the video board at the back to unplug the connector. Then slide the board backwards to
remove it.
Connector
88
Page 89
MP90 CMU Disassembly 5 Repair and Disassembly
Reassembly Note: The flex cable on the primary video board must go up.
Reassembly Note: When replacing the primary video board, the board must slide under the sheet
metal.
89
Page 90
5 Repair and Disassembly MP90 CMU Disassembly
NOTE
If not already separated, you may need to break apart the two pieces of the repaired/exchange board.

Removing the Main Board

1 Remove the video board as described in Replacing the Primary Video Board.
2 Remove all I/O boards
90
Page 91
MP90 CMU Disassembly 5 Repair and Disassembly
Remove power supply and power switch connectors and the speaker connector.
3
Speaker
Power
Power
Switch
Supply
4 Disconnect the I/O board backplane connectors.
91
Page 92
5 Repair and Disassembly MP90 CMU Disassembly
Remove the two screws securing the backplane (on the metal, NOT the board) and take out the
5
backplane.
92
Page 93
MP90 CMU Disassembly 5 Repair and Disassembly
Remove the CPU interconnection board.
6
7 Remove the remaining ten screws and pull the main board straight up to remove it. Please note
that there are sensitive components near the screws on the bottom of the main board.

Removing the Power Supply

1 Remove the second CPU sheet metal assembly, if present. (See Accessing the Main CPU or Primary
Video Board (Dual CPU Versions)
93
Page 94
5 Repair and Disassembly MP90 CMU Disassembly
Pull out the power supply and power switch connectors.
2
3 Flip over the CMU, support it with the cable management cover and take off the four screws.
94
Cable Manageme nt Cover
Page 95
MP90 CMU Disassembly 5 Repair and Disassembly
Pull the CMU base off of the power supply.
4
Reassembly Note: When reassembling, ensure that all cables are led through the dedicated holes and
not pinched underneath the metal.

Removing the Speaker

1 Remove the power supply.
2 Disconnect the speaker cable, unscrew the three black screws using a flat blade screwdriver, and
then turn and pull out the speaker.
Screws

Removing the Power On/Off Switch

1 Remove the power supply.
95
Page 96
5 Repair and Disassembly MP90 CMU Disassembly
Remove the screw securing the power switch and remove the switch.
2
96
Page 97
Flexible Module Server (FMS) Disassembly 5 Repair and Disassembly

Flexible Module Server (FMS) Disassembly

Removing the Handle and the Measurement Server Mount

Please note that any combination of handles and mounts is possible.
1 Remove the two screws on the bottom with a T20 screwdriver.
2 Slide the handle up and pull it out.
3 Remove the MMS if connected. Slide the MMS mount up and remove it by pulling directly
perpendicular to the FMS.
97
Page 98
5 Repair and Disassembly Flexible Module Server (FMS) Disassembly
NOTE
There is a connector located on the MMS mount. If you tilt the MMS mount as you remove it, you may damage the connector.
4 Remove the connector housings on each side of the FMS by compressing the cover slightly using
two screwdrivers.
98
Page 99
Flexible Module Server (FMS) Disassembly 5 Repair and Disassembly
Remove the two white pins on each side with a small screwdriver.
5
6 Tak e of f th e r e ar h ou s in g .
Rear
Front
7 Remove the four screws on the CPU board, pull it gently off the mother board, unplugging the
connector at the same time.
99
Page 100
5 Repair and Disassembly Flexible Module Server (FMS) Disassembly
Remove the four remaining screws on the mainboard.
8
9 Pull off side connector brackets by pulling them gently away from the housing on each side and
lifting carefully.
10 Lift up and pull on the tabs to remove the connector holders.
Reassembly Note: The connector holders are side specific.
11 Unsnap the lightpipe.
100
Loading...