Who Should Use This Guide13
How to Use This Guide13
Abbreviations13
Responsibility of the Manufacturer14
Passwords14
Warnings and Cautions14
2 Theory of Operation15
Monitor Theory of Operation15
System Boundaries15
Hardware Building Blocks16
IntelliVue MP9017
Optional Hardware18
Compatible Devices18
Power Supply19
CPU Boards19
I/O Boards20
Data Flow20
Data Acquisition20
Data Provider System Service21
Persistent Data Storage System Service21
Display and User Interface Service21
Data Output21
Monitor Applications22
Internal LAN (Measurement Server Link)22
Philips Clinical Network23
How does the Support Tool Work with the Monitor23
Monitor Software Block Diagram24
Block Diagram Legend26
3 Testing and Maintenance31
Concepts31
Test Reporting31
Recommended Frequency32
Tests Recommended When Performing...33
Checks for Obvious Problems52
Checks Before Opening the Instrument52
Checks with the Instrument switched Off52
Checks with the Instrument Switched On, AC connected53
Initial Instrument Boot Phase53
Troubleshooting Tables54
How to use the Troubleshooting tables54
Boot Phase Failures55
External Display is blank58
External Touch Display not functioning58
General Monitor INOP Messages59
Remote Alarm Device60
Remote Extension Device60
Keyboard/Mouse not functioning61
Network related problems62
Wireless Network63
Multi-Measurement Server64
MSL-related problems64
Alarm Issues66
Alarm Lamps66
Alarm Tones67
Alarm Behavior67
Individual Parameter INOPS67
Flexible Module Server68
Printer69
Recorder70
MIB / RS23270
Flexible Nurse Call Relay71
Troubleshooting the ECG OUT72
Data Flow Marker In and ECG Wave72
Status Log73
Troubleshooting with the Support Tool74
Troubleshooting the Individual Measurements or Applications75
5 Repair and Disassembly77
Tools Required77
MP90 CMU Disassembly77
Removing I/O Boards77
Removing the Top Cover79
Removing the Plastic Feet and/or the Locking Cam81
Removing the optional Fans (Dual CPU Versions only)*82
Replacing the Second (Independent) Video Board
(Dual CPU Versions only)
83
5
Page 6
Removing the Second CPU/Main Board (Dual CPU Versions only)84
Accessing the Main CPU or Primary Video Board (Dual CPU Versions)86
Replacing the Primary Video Board88
Removing the Main Board90
Removing the Power Supply93
Removing the Speaker95
Removing the Power On/Off Switch95
Flexible Module Server (FMS) Disassembly97
Removing the Handle and the Measurement Server Mount97
Disassembly Procedures for the Measurement Server Extension105
Removing the Front Cover105
Removing the Extension Bottom Cover106
Removing the CO2 Scrubber107
Removing the Pump108
Refit Procedures for the Measurement Server Extension108
Refitting the CO2 Scrubber109
Refitting the Pump109
Refitting the Extension Bottom Cover109
Refitting the Front Cover110
General Reassembly/Refitting Comments110
Following Reassembly110
6 Parts111
MP90 Parts112
Exchange Parts112
Replacement Parts113
Flexible Module Server Parts114
Exchange and Replacement Parts114
Multi-Measurement Server Parts115
Measurement Server Extension Parts (M3015A and M3016A)117
Single-Width Plug-In Module122
Double-Width Plug-In Module122
Plug-in Module Replaceable Parts123
Plug-In Module Language Specific Front Housings, Table 1123
Plug-In Module Language Specific Front Housings, Table 2123
Plug-In Module Specific Bezels124
6
Page 7
BIS Module Replaceable Parts124
BIS Module Components125
tcpO2/tcpCO2 Module Accessories125
External Display Part Numbers126
Remote Input Devices Part Numbers127
Remote Alarm Device Part Numbers128
Remote Extension Device Part Numbers129
7 Installation Instructions131
Unpacking the Equipment131
Initial Inspection132
Mechanical Inspection132
Electrical Inspection132
Claims For Damage and Repackaging132
Claims for Damage132
Repackaging for Shipment or Storage132
Installing the Monitor (M8010A)132
Mounting Instructions132
Assembling Mounts132
Connections133
Installing Interface Boards134
Connection of MIB Devices135
Installing Remote Devices136
Mounting the Remote Display (M8031A)136
Connections136
Mounting the 17” Remote Display (M8033A)137
Flexible Module Server and/or Multi-Measurement Server137
Positioning the Measurement Server on a Clamp Mount137
Mounting the MMS Mount to the FMS (M8048A)137
Attaching the MMS to a Mount138
Detaching the Measurement Server from a Mount138
Mounting the Remote Extension Device to the FMS139
Mounting the BIS Engine to the FMS139
Mounting the FMS140
Connections141
MSL Cable Termination141
Setting Altitude and Line Frequency148
Configuring the Equipment Label148
Configuring the printer148
Setting the Display Resolution (M8033A only)148
8 Site Preparation149
Introduction149
Site Planning149
Roles & Responsibilities149
Site Preparation Responsibilities149
Procedures for Local Staff150
Procedures for Philips Personnel151
Monitor M8010A Site Requirements151
Space Requirements151
Environmental Requirements152
Te m p e r at u r e15 2
Humidity152
Altitude152
Electrical and Safety Requirements (Customer or Philips)152
Safety Requirements152
Electrical Requirements153
Remote Device Site Requirements153
Multi-Measurement Server M3001A or Flexible Module Server M8048A154
Space Requirements Multi-Measurement Server M3001A154
Space Requirements Flexible Module Server M8048A154
Environmental Requirements Multi-Measurement Server M3001A155
Environmental Requirements Flexible Module Server M8048A155
Cabling Options and Conduit Size Requirements155
Remote Displays - M8031A156
Space Requirements156
Environmental Requirements156
Electrical and Safety Requirements157
Remote Displays - M8033A157
Space Requirements157
Environmental Requirements157
Electrical and Safety Requirements157
Cabling Options and Conduit Size Requirements158
Remote Alarm Devices158
Space Requirements158
8
Page 9
Cabling Options and Conduit Size Requirements158
Remote Extension Device159
Space Requirements159
Cabling Options and Conduit Size Requirements159
Input Devices159
Local Printer160
Philips Medical LAN160
MIB Interface160
Flexible Nurse Call Relay Interface161
ECG Out Interface161
Explanation of Symbols Used179
Power Supply Requirements180
Grounding the System180
Equipotential Grounding181
Combining Equipment181
Checking and Calibrating the Anesthetic Gas Module181
Access Service Functions of the M1026A Anesthetic Gas Module181
When and how to check the Philips M1026A Anesthetic Gas Module183
Equipment required for checking183
Checks and adjustments184
Performance Leakage Check184
Performance Diagnostic Check185
Performance Flowrate Check185
Total Flowrate Check and Adjustment in Purge Mode185
Measurement Path Flowrate Check and Adjustment186
Total Flowrate Check in Normal Mode188
Zero Calibration188
Barometric Pressure Check and Calibration189
Span Calibration Check190
Other factors to maximize uptime or reduce cost of ownership:198
Troubleshooting the Anesthetic Gas Module198
Compatibility Criteria for the AGM and the IntelliVue Monitors198
Flow Charts for Communication and Measurement Type Problems198
Hardware Related Troubleshooting Strategy203
INOPs204
Calibration Checks206
Calibration Checks Troubleshooting Table207
10
Page 11
Diagnostic Checks208
Problem Solving Hierarchy209
Pneumatic System Diagnostic Checks210
O2 Assembly Diagnostic Checks210
Optical Path Disgnostic Checks213
IR Measurement Assembly Diagnostic Checks214
Agent ID Assmebly Diagnostic Checks215
Power Supply Diagnostic Checks216
Operating Temperature Diagnostic Checks217
Test Points, Connectors and Jumpers217
Tes t Poi n t s21 7
Connectors218
Jumpers218
Repairing the Anesthetic Gas Module220
Introduction220
The Top Cover222
Removal222
Replacement222
Lifting the IR Measurement Mounting Bracket224
Removal225
Replacement225
Infrared Measurement Assembly Head227
Transferring NVRAM Data to a Replacement Head227
Sample Cell231
Removal231
Replacement231
Solenoid Valve #1235
Removal235
Replacement235
Power Supply Unit237
Removal237
Replacement237
Main PC Board238
Removal238
Replacement239
O2 Sensor240
Removal240
Replacement241
Agent Identification Head243
Removal243
Replacement244
Pump245
Removal245
Replacement245
Fan 246
11
Page 12
Removal246
Replacement246
Solenoid Valve #2248
Removal248
Replacement248
Top C ov e r PC B o a rd249
Removal249
Replacement249
Watertrap Manifold and Protector251
Removal251
Replacement251
Power Fuses252
Removal252
Replacement252
Test and Inspection Matrix252
When to Perform Test Blocks256
Safety Test Appendix257
Parts List259
Calibration Equipment 264
12
Page 13
This Service Guide contains technical details for the IntelliVue MP90 Patient Monitor, the MultiMeasurement Server (MMS), the Flexible Module Server (FMS) and the Measurement Server
Extensions.
This guide provides a technical foundation to support effective troubleshooting and repair. It is not a
comprehensive, in-depth explanation of the product architecture or technical implementation. It offers
enough information on the functions and operations of the monitoring systems so that engineers who
repair them are better able to understand how they work.
It covers the physiological measurements that the products provide, the Measurement Server that
acquires those measurements, and the monitoring system that displays them.
Who Should Use This Guide
1
1Introduction
This guide is for biomedical engineers or technicians responsible for troubleshooting, repairing, and
maintaining Philips’ patient monitoring systems.
How to Use This Guide
This guide is divided into eight sections. Navigate through the table of contents at the left of the screen
to select the desired topic. Links to other relevant sections are also provided within the individual
topics. In addition, scrolling through the topics with the page up and page down keys is also possible.
Abbreviations
Abbreviations used throughout this guide are:
NameAbbreviation
IntelliVue MP90 Patient Monitorthe monitor
Flexible Module ServerFMS
Multi-Measurement ServerMMS
Measurement Server LinkMSL
Medical Information BusMIB
Anesthetic Gas ModuleAGM
13
Page 14
1 IntroductionResponsibility of the Manufacturer
Responsibility of the Manufacturer
Philips only considers itself responsible for any effects on safety, reliability and performance of the
equipment if:
• assembly operations, extensions, re-adjustments, modifications or repairs are carried out by persons
authorized by Philips, and
• the electrical installation of the relevant room complies with national standards, and
• the instrument is used in accordance with the instructions for use.
To ensure safety, use only those Philips parts and accessories specified for use with the monitor. If nonPhilips parts are used, Philips is not liable for any damage that these parts may cause to the equipment.
This document contains proprietary information which is protected by copyright. All Rights Reserved.
Reproduction, adaptation, or translation without prior written permission is prohibited, except as
allowed under the copyright laws.
Philips Medizin Systeme Böblingen GmbH
Hewlett-Packard Str. 2
71034 Böblingen, Germany
The information contained in this document is subject to change without notice.
Philips makes no warranty of any kind with regard to this material, including, but not limited to, the
implied warranties or merchantability and fitness for a particular purpose.
Philips shall not be liable for errors contained herein or for incidental or consequential damages in
connection with the furnishing, performance, or use of this material.
Passwords
In order to access different modes within the monitor a password may be required. The passwords are
listed below.
Monitoring Mode: No password required
Configuration Mode: 71034
Demo Mode: 14432
Service Mode: 1345
Consult the configuration guide before making any changes to the monitor configuration.
Warnings and Cautions
In this guide:
•A warning alerts you to a potential serious outcome, adverse event or safety hazard. Failure to
observe a warning may result in death or serious injury to the user or patient.
14
•A caution alerts you where special care is necessary for the safe and effective use of the product.
Failure to observe a caution may result in minor or moderate personal injury or damage to the
product or other property, and possibly in a remote risk of more serious injury.
Page 15
2Theory of Operation
Monitor Theory of Operation
The IntelliVue Patient Monitor:
• displays real-time data
• controls the attached measurement servers
• alarms in the case of patient or equipment problems
• offers limited data storage and retrieval (trending)
• interfaces to the Philips Clinical Network and other equipment
2
A monitor with just a single integrated measurement server can be connected to additional building
blocks to form a monitoring system with a large number of measurements, additional interface
capabilities and multiple slave displays. These elements cooperate as one single integrated real-time
measurement system.
System Boundaries
The following diagram discusses specific boundaries within the overall system with respect to their
openness and real-time requirements:
15
Page 16
2 Theory of OperationMonitor Theory of Operation
Philips Clinical Network
Measurement LAN
combines components of one patient monitor;
real time requirements across all interconnected
elements
Philips Clinical Network (wired LAN)
connects multiple patient monitors,
information centers, application servers; closed
system, only Philips qualified products (tested
and with regulatory approval) are connected,
Philips is responsible for guaranteed real-time
functionality and performance
Philips Clinical Network (wireless)
like Philips Clinical Network (wired) LAN,
however due to current wireless technologies
available it has reduced bandwidth, longer
latencies, reduced functionality
Hospital LAN, Internet
Standard Network, not under Philips control,
no guaranteed service, no real-time
requirements
Hardware Building Blocks
The following hardware building blocks make up the monitoring system:
16
Page 17
Monitor Theory of Operation2 Theory of Operation
IntelliVue MP90
The MP90 monitor:
• can be used with the standalone M8031A 15-inch or the M8033A 17-inch color LCD TFT display
with touchscreen operation.
• can also be used with other XGA and SXGA standalone off-the-shelf displays which comply with
medical standards such as IEC 60601-1 and IEC 60601-1-2.
• has the central processing unit in a separate module
• uses the Philips SpeedPoint as primary input device whereas the Philips Touchscreen and computer
devices such as mice, trackball, and keyboard can be added optionally
• supports the Flexible Module Server (FMS)
Building Blocks:
17
Page 18
2 Theory of OperationMonitor Theory of Operation
|| I/F To Local Printer
Power Supply
I/F
Boards
1st Main Board
Remote Device I/F to
Alarm Device I/F*
Alarm Device and Remote Input
MIB To AGM
1 Main Board
I/F
2nd
Main Board
Boards
FMS CPU
FMS M/B
MSL
OR
MMS CPU
Video I/F
Board
Video I/F
Board
To Ext. Display
I/F
To Ext. Display
Optional Hardware
The M8031A 15-inch color LCD TFT display or the M8033A 17-inch color LCD TFT display (both
with touchscreen operation) can be ordered optionally. Additional input devices such as mice, trackball
or keyboard can also be added. If the monitor is ordered with the wireless LAN option an external
wireless transmitter is required. For further details regarding the wireless network please refer to the
M3185A Philips Clinical Network documentation.
Compatible Devices
18
Figure 1 M8048A Flexible Module Server (FMS)
Page 19
Monitor Theory of Operation2 Theory of Operation
Figure 2 M3001A Multi-Measurement Server (MMS)
Power Supply
The AC/DC converter transforms the AC power coming from the power plug into 48 V/120W DC
source and isolates the monitoring system from the AC power mains.The 48V is distributed via power
bus and supplies power to all the components of the system: The 56 V DC power needed for the FMS,
MMS and measurement server extension is created by an isolating DC/DC converter. The CPU is
supplied with 3.3 V and 5 V DC power. The transformation is performed in two steps: The first DC/
DC converter is a power regulator which reduces the variations caused by load changes on the 48V
power bus. The second DC/DC converter converts the power to the needed voltage. Interface boards
require a power of 10V AC.
CPU Boards
The CPU boards have an MPC860 50 MHz processor that provides a number of on-chip,
configurable interfaces. An array of 12 fast UARTs with configurable protocol options are
implemented in an ASIC (along with other system functions such as independent watchdogs etc.),
providing interfacing capabilities to measurement modules and I/O boards. The serial interfaces can
easily be electrically isolated. The main board contains additional video hardware.
Flexible Module Server
CPU
Bank of I/Os
Modules
The CPUs provide two LAN interfaces to interconnect CPUs (via the Internal LAN) or to connect to
the Philips Clinical Network.
IntelliVue Patient MonitorMulti-Measurement Server
CPUVideo
Bank of I/Os
CPU
Bank of I/Os
Measurement
Interfaces
Acquisition
The CPU capabilities are identical. Different loading options are coded on serial EEPROMs to support
the automatic configuration of the operating system at boot time.
19
Page 20
2 Theory of OperationMonitor Theory of Operation
I/O Boards
Interfaces to the monitor are implemented via I/O boards. The location of these boards is restricted by
general rules. The I/O slot designations diagram and the I/O matrix which outline the I/O board
placement rules can be found in the Installation Instructions section.
The following is a list of Interface (I/O) boards which may be present in your monitor, depending on
your purchased configuration:
•MSL
•Video
• Philips Clinical Network (LAN wired or wireless)
The specifications for the above listed interfaces can be found in the technical data sheet for the
monitor and in the Specifications chapter of the Instructions for Use.
Data Flow
The following diagram shows how data is passed through the monitoring system. The individual stages
of data flow are explained below.
Display
and User
Interface
Data
Acquisition
Data
Provider
Applications
Service
Persistent
Data
Storage
Data
Output
Data Acquisition
Monitoring data (for example patient measurement data in the form of waves, numerics and alerts) is
acquired from a variety of sources:
20
Page 21
Monitor Theory of Operation2 Theory of Operation
•Measurement Servers
The Measurement Servers connected to the internal LAN convert patient signals to digital data and
apply measurement algorithms to analyze the signals.
• External measurement devices
Data can be also acquired from devices connected to interface boards of the monitor. Software
modules dedicated to such specific devices convert the data received from an external device to the
format used internally. This applies to parameter modules and the Anesthetic Gas Module.
• Server systems on the Philips Clinical Network
To enable networked applications such as the other bed overview, data can be acquired from server
systems attached to the Philips Clinical Network, for example a Philips Information Center.
Data Provider System Service
All data that is acquired from measurement servers or external measurement devices is temporarily
stored by a dedicated data provider system service. All monitor applications use this central service to
access the data in a consistent and synchronized way rather than talking to the interfaces directly.
This service makes the applications independent of the actual type of data acquisition device.
The amount of data stored in the data provider system service varies for the different data types. for
example several seconds of wave forms and the full set of current numerical values are temorarily stored
in RAM.
Persistent Data Storage System Service
Some applications require storage of data over longer periods of time. They can use the persistent data
storage system service. Dependent on the application requirements, this service can store data either in
battery backed-up (buffered) memory or in flash memory. The buffered memory will lose its contents
if the monitor is without power (not connected to mains) for an extended period of time. The flash
memory does not lose its contents.
The trend application for example stores vital signs data in a combination of flash memory and
buffered memory, while the system configuration information (profiles) is kept purely in flash
memory.
Display and User Interface Service
Applications can use high level commands to display monitoring data or status and command windows
on the LCD panel. These commands are interpreted by the display manager application. This
application controls the dedicated video hardware which includes video memory and a special ASIC.
User input is acquired from a variety of input devices, for example the SpeedPoint, the touchscreen or
other standard input devices (keyboard, mouse) which may be attached to I/O boards. The system
software makes sure that the user input is directed to the application which has the operating focus.
Data Output
The monitoring system is very flexible and customizable regarding its data output devices. Built-in
devices (for example LAN, speaker, video) provide the basic output capabilities.
These capabilities can be enhanced by adding additional I/O boards, as required in the specific enduser setup. The additional I/O boards typically provide data to externally attached devices, for example
to printers, RS232 based data collection devices, nurse call systems etc.
21
Page 22
2 Theory of OperationMonitor Theory of Operation
The monitor can identify I/O boards by means of a serial EEPROM device that stores type and version
information. The operating system detects the I/O boards and automatically connects them with the
associated (interface driver) application. For some multi-purpose cards it is necessary to configure the
card for a particular purpose first (for example the dual MIB/RS232 card can support external touch
display, data import, data export).
Monitor Applications
The monitor applications provide additional system functionality over the basic measurement and
monitoring capabilities. This includes for example trending, report generating, event storage or derived
measurements.
In general, the monitor applications use the data provider system service to access the measurement
data. Application interfaces to the other system services allow the application to visualize data, to store
data over extended periods of time or to output data to other devices.
Internal LAN (Measurement Server Link)
All components of the monitoring system (including measurement servers and CPUs in the monitor)
communicate using an IEEE802.3/ Ethernet LAN in the Measurement Server Link (MSL). This
network is used to distribute data between the components, for example:
• Digitized patient signals including wave data, numerical data and status information (typically from
the measurement server to a display unit)
• Control data representing user interactions (typically from the display unit to a measurement server)
• Shared data structures, for example representing patient demographical data and global
configuration items
The internal LAN allows plug and play configuration of the monitoring system. The system
automatically detects plugging or unplugging of measurement servers and configures the system
accordingly.
The components on the internal LAN are time-synchronized to keep signal data consistent in the
system. Dedicated hardware support for synchronization eliminates any latency of the network driver
software.
The integrated LAN provides deterministic bandwidth allocation/reservation mechanisms so that the
real-time characteristic of signal data and control data exchange is guaranteed. This applies to the data
flow from the measurement server to the monitor (for example measurement signal data) and the data
flow from the monitor to a measurement server (for example to feed data to a recorder module).
Integrated communication hubs in the monitor and the FMS allow flexible cabling options (star
topology, daisy chaining of servers).
22
Page 23
Monitor Theory of Operation2 Theory of Operation
MDSE Internal LAN
MDSE
Internal
LAN
Philips Clinical Network
The monitoring system may be connected to the Philips Clinical Network, for example to provide
central monitoring capabilities or other network services. This connection may be through a normal
wired connection or through a wireless connection.
The monitor supports the connection of an external off-the-shelf wireless adapter. This allows a simple
field upgrade as well as a technology upgrade in the future. Switching between wired and wireless
networks is automatically triggered by the plugging or unplugging of the network cable.
The Philips Clinical Network protocols function very similarly to the protocols used on the internal
LAN.
MDSE
Internal
LAN
After configuration, the monitoring system sends the digitized patient signals including wave data,
numerical data and status information onto the network. Control data representing user interactions
can be exchanged between the monitoring system and a central station bi-directionally.
Additional protocols are supported for networked applications, for example for the other bed overview
function, which allows viewing of monitoring data from other patients on the network.
For plug and play operation, the monitoring system uses the standard BootP protocol to automatically
acquire a network address.
How does the Support Tool Work with the Monitor
The support tool is a Windows application typically installed on the laptop of a customer engineer or a
biomedical engineer working in the customer’s own service department.
The purpose of the support tool is to upgrade, configure and diagnose all monitoring components
(modules, measurement servers, and monitors) in the system over the network.
23
Page 24
2 Theory of OperationMonitor Theory of Operation
The service protocol developed for this purpose uses a raw access to the devices without the need for IP
addresses etc. over a standard customer network installation, so that even defective devices can be
upgraded as long as the few kBytes of initial boot code are working. The boot code itself can also be
upgraded using the same protocol.
The tool allows access to internal service information and to serial numbers. It can be remotecontrolled, for example via a dial-up connection from a response center, provided the proper
infrastructure is in place.
For details see the Instructions for Use for the Support Tool.
Monitor Software Block Diagram
Figure 3 shows the functional block diagram for the monitoring system. A legend explaining terms and
diagram elements follows. The information below varies depending on the purchased monitor options.
Figure 3 IntelliVue Patient Monitoring System Functional Block Diagram
25
Page 26
2 Theory of OperationMonitor Theory of Operation
Block Diagram Legend
Functional BlockDescription
Services
Operating SystemThe Operating System (OS) provides a layer of isolation between
the specific hardware implementation and the application
software. The OS performs system checks and allocates resources
to ensure safe operation when the system is first started. This
includes internal self-tests on several hardware modules and
configuration checks for validity of configuration with the
operating software. During normal operation, the OS continues
to run checks on system integrity. If error conditions are detected
the OS will halt monitoring operations and inform the operator
about the error condition.
System ServicesThe System Services provide generic common system services.
In particular:
It uses a real-time clock component to track time. It synchronizes
to network time sources and verifies the accuracy of the system
time information. It is also responsible for managing persistent
user configuration data for all Measurement Servers, Flexible
Module Servers and IntelliVue Patient Monitoring System
software modules. User configuration data is stored in a nonvolatile read/write storage device
Applications
ReportsThe Reports Service retrieves current and stored physiological
data and status data to format reports for printing paper
documentation. The following reports are supported:
• Other reports (e.g. Loops, Review Applications, Drug
report)
The Reports service generates report data which can be printed
on a local or a central printer.
26
Page 27
Monitor Theory of Operation2 Theory of Operation
Functional BlockDescription
RecordThe Record Service retrieves current and stored physiological
data and status data to format a continuous strip recording. A
recording can be triggered manually by the operator or
automatically by an alarm condition. The Record Service uses the
services of the Recorder Interface to control an M1116B
Recorder in the FMS. The Record Service can also send data to a
central recorder.
AlarmThe Alarm Service contains logic that prioritizes alarm conditions
that are generated either by the Measurement Servers, Flexible
Module Server, or by IntelliVue Patient Monitoring System
software modules. Visual alarm messages are displayed at the top
of the IntelliVue Patient Monitoring System display and alarm
sounds are generated by a loudspeaker. Alarm conditions may be
generated when a physiological parameter exceeds preselected
alarm limits or when a physiological parameter or any other
software module reports an inoperative status (technical alarm,
for example, the ECG leads may have fallen off the patient). The
Alarm service manages the alarm inactivation states, for example
suspension of alarms, silencing of alarms, and alarm reminder.
Alarm signals may also be configured as latching (alarm signals
are issued until they are acknowledged by the operator, even
when the alarm condition is no longer true). The Alarm service
controls the visual alarm signals (alarm lamps).
TrendThe Trend service stores the sample values of physiological data
and status data with a resolution of 12 seconds, 1 minute or 5
minutes for a period of up to 48 hours. The data is kept in
battery buffered read/write storage and flash memory devices to
be preserved across power failures. The stored data is protected
via consistency checks and checksums. When a new patient is
admitted, the trend database erases all data of the previous
patient.
HiResThe OxyCRG (Oxygen CardioRespiroGram) service derives a
high-resolution trend graph from the Beat-to-Beat Heart Rate,
SpO
or tcpO2, and Respiration physiological data. The
2
OxyCRG is specialized for neonatal applications, allowing the
opeartor to identify sudden drops in Heart Rate (Bradycardia)
and SpO
or tcpO2 (Desaturations), and supporting the operator
2
in visualizing Apnea situations.
ADTThe ADT (Admit/Discharge/Transmit) service maintains the
patient demographics information. The operator may admit a
new patient, discharge the old patient and enter or modify the
patient demographics. The ADT service also supports the
transport of a patient (trend database) with the M3001A MultiMeasurement Server. The ADT service controls the deletion of
old patient data, the upload of trend data from the M3001A and
the switching back of all settings to user defaults. It also
synchronizes patient information with a central station on the
network.
27
Page 28
2 Theory of OperationMonitor Theory of Operation
Functional BlockDescription
Calc ParamThe Calc Param (Calculated Parameters) service accesses current,
stored and manually entered physiological data as input to
calculation formulas. With these formulas, derived
hemodynamic, oxygenation and ventilation variables are
computed. The calculation results, including the input
parameters, are stored for later review using the Trend service.
Interface Managers
MDSEThe MDSE (Medical Data Service Element) Interface Manager is
responsible for the exchange of real-time data between the
IntelliVue Patient Monitoring System display unit and the
Measurement Servers and Flexible Module Server as well as
between the IntelliVue Patient Monitoring System display unit
and other devices attached to the network. MDSE establishes and
maintains a data communication link between the devices. It
provides configuration information about the remote device to
applications in the local device and it allows the exchange of
measurement data and status information between the devices.
PrinterThe Printer Interface Manager provides a high level interface to a
printer. It provides means to:
• establish a connection to the printer
• transfer data to the printer
• get status of the printer
• close connection to the printer
The Printer Interface Manager also supervises the connection to
the printer and whether the printer accepts data (for example
paper out). The Printer Interface Manager notifies the operator
in such cases.
28
Page 29
Monitor Theory of Operation2 Theory of Operation
Functional BlockDescription
Display & Operator InterfaceThe Display and Operator Interface Manager performs the
following tasks:
• Screen presentation of real-time and stored physiological
measurement data, alarm condition data and status
information received from the MDSE interface manager,
the Alarm service or other IntelliVue Patient Monitoring
System modules
• Screen presentation of operating controls (control
windows)
• Processing of operating control commands received from
HIF Control interface. The module verifies and interprets
the received commands and forwards them to other
software modules of the IntelliVue Patient Monitoring
System display unit, Measurement Servers or Flexible
Module Server
• Sound generation (issues audible alarm signals and
generates audible information signals, for example QRS
and SpO
Interfaces
LANThe LAN interface implements the physical layer of IEEE 802.3.
The LAN interface performs Manchester encoding/decoding,
receive clock recovery, transmit pulse shaping, jabber, link
integrity testing, reverse polarity detection/correction, electrical
isolation, and ESD protection. Electronically separated interfaces
are used for communication to the Measurement Servers or
Flexible Module Server and to the network.
CentronicsThe Centronics interface implements the standard signaling
method for bi-directional parallel peripheral devices according to
IEEE 1284-I. The interface is used as a parallel interface to a
standard printer with electrical isolation and ESD protection.
Display ControllerThe Display Controller Interface consists of a video controller
chip, video RAM and the controlling software. The Display
Controller interface processes the high level display commands
(character and graphic generation, wave drawing) and translates
them into pixels, which are written into the video RAM where
the video controller chip generates the video synchronization
signals and the pixel stream for the external display.
HIF ControlThe HIF (Human Interface Control) interface scans the Human
Interface devices for operator controls (Touch Screen, Trim
Knob, and PS/2 devices), formats the collected data and sends it
to the display and Operating Interface.
tones, operator audible feedback)
2
29
Page 30
2 Theory of OperationMonitor Theory of Operation
Functional BlockDescription
ECG-Out Marker-InThe ECG Out/Marker In interface receives the ECG waveform
directly from the ECG/Resp Arrhythmia ST-Segment
physiological algorithm via an RS-422 serial interface and
converts the digital ECG signal to an analog ECG signal. In
addition, the ECG Out controller receives from a connected
device the marker information and forwards this data to the
ECG/Resp Arrhythmia ST-Segment physiological algorithm.
The converted analog signal is used to synchronize a connected
device to the patient’s ECG
RS-422The serial link RS-422 interface communicates the ECG signal
to the ECG Output/Marker In of the IntelliVue Patient
Monitoring System display unit. The interface is a serial,
differential, full-duplex link. The interface is ESD protected.
PS/2The PS/2 interface supports the serial protocol of standard PS/2
devices (mouse). The PS/2 serial protocol is interpreted by the
HIF Control interface.
Remote Device InterfaceThe remote device interface is an interface to the remote alarm
device and the remote extension device. The interface
communicates with the remote devices via a differential serial
link.
Nurse CallThe Nurse Call board contains 2 connectors. A phone jack type
connector and a multi-port connector. The phone jack type
connector has a single close-on-alarm relay. The multi-port
connector has three alarm relays which are configurable to be
open or closed on alarm. In addition, this interface has an audible
alert capability for loss of AC power.
binary communication between the monitor and an arbitrary
(medical/non-medical) device using an eight-pin RJ45 modular
connector. Switching between MIB and RS232 protocol is
possible.
30
Page 31
3Testing and Maintenance
Concepts
This chapter provides a checklist of the testing and maintenance procedures for the monitor, the MMS
the Measurement Server Extensions and the FMS associated modules.
Preventive Maintenance refers specifically to the series of tests required to make sure the Instrument
measurement results are accurate. The measurements requiring these reported tests are NBP and
sidestream CO
readings are in question or as specified.
. The accuracy and performance procedures are designed to be completed when
2
Test Reporting
Authorized Philips personnel report test results back to Philips to add to the product development
database. Hospital personnel, however, do not need to report results. This table shows you what to
record on the service record after completing the tests in this chapter.
3
TestWhat to record
VisualV:P or V:F
Power OnPO:P or PO:F
P NIBPPN:P/X1/X2/X3/X4 or
PN:F/X1/X2/X3/X4
PCO
2
SafetyS(1):P/x1/x2 or
Where P = Pass, F = Fail and X/x are the measured values as defined in the tests described in this
chapter.
PCO2:P/X1/X2/X3/X4/X5/X6/X7/X8 or
PCO2:F/X1/X2/X3/X4/X5/X6/X7/X8
S(1):F/x1/x2
S(2): P/x1 or
S(2): F/x1
S(3): P/x1 or
S(3): F/x1
31
Page 32
3 Testing and MaintenanceRecommended Frequency
Recommended Frequency
The testing checklist appears in the next section of this chapter. Perform the procedures as indicated in
the suggested testing timetable. These timetable recommendations do not supersede local
requirements.
Suggested Testing TimetableFrequency
Preventive Maintenance Tests
•NBP Calibration
• Sidestream CO
•CO
pump / CO2 scrubber replacement
2
Calibration
2
Performance and Safety Tests
• Temperature Accuracy
• ECG/Resp Performance
• Invasive Pressure Performance
•SpO
•Mainstream CO
Per formance
2
Per formance
2
•EEG Performance
• C.O. Performance
•BIS Performance
•SvO2 Performance
Required
• Once a year, or as specified by local laws.
• Once a year or after 4,000 hours continuous
use and following any instrument repairs or
the replacement of any instrument parts.
• Once every three years or after 15 000
operating hours
Once every two years, or if you suspect the
measurement is incorrect
32
•tcGas Performance
• VueLink Performance
• Nurse Call Relay Performance*
• ECG Sync Performance*
*Only when in use as part of hospital protocols
Safety Checks (in accordance with IEC 60601-
1)
• System Enclosure Leakage Current
•Protective Earth
• Patient Leakage Current
Once every two years and after repairs where
the power supply is replaced or the monitor
has been damaged by impact.
Page 33
Tests Recommended When Performing...3 Testing and Maintenance
Tests Recommended When Performing...
Installation
Service Event
(When performing...
Installation of monitor with a non-medical device
connected.
Repair
Service Event
(When performing...
Repairs of M3015APerform Power On and M3015A tests
Repairs where the monitor has been damaged by
impact
Repairs where the power supply is replacedPerform Safety (2) Test Block
All other IntelliVue Monitoring System repairsPerform Power On Test Block
Preventive Maintenance
Perform preventive maintenance tests:
Tes t Bl o c k s R e q u i r e d
...Complete these tests)
Perform Visual, Power On and Safety (1) Test
Blocks
Tes t Bl o c k s R e q u i r e d
...Complete these tests)
Perform Power On and Safety (2) and (3) Test
Blocks
•NBP calibration
• Sidestream CO
• Pump and scrubber replacement.
calibration
2
Performance Verifications
Perform all safety, accuracy and performance test procedures listed in the following sections. If a
particular measurement is in question, perform the measurement performance test only.
33
Page 34
3 Testing and MaintenanceTests
Upgrades
Service Event
(When performing...
Hardware and software upgradesPerform Power On Test Block unless otherwise
Tests
Some of the following testprocedures must be performed in service mode. To enter service mode select
Operating Modes in the main menu. Then select Service Mode and enter the password.
If required, open the screen menu in the monitor info line at the top of the screen and select
Service to access the service screen. This is required particularly for Anesthetic Gas Module testing
procedures.
Visual Test
Inspect the system for obvious signs of damage. Also check all external leads and accessories.
The expected test result is pass: the system has no obvious signs of damage.
Power On Test
Tes t Bl o c k s R e q u i r e d
...Complete these tests)
specified in the Upgrade Installation Notes shipped
with the upgrade.
NOTE An external display must be connected to the MP90 computer module unit.
1Switch on the monitor and connect the MMS.
2Observe whether the system boots up successfully and if an ECG wave appears on the screen.
The expected test result is pass: the monitor boots up and displays an ECG wave. The wave might be a
flat line if no simulator is attached.
NBP Tests
This section describes NBP test procedures.The monitor must be in service mode to perform these
tests.
34
Page 35
Tests3 Testing and Maintenance
NBP Accuracy Test
This test checks the performance of the non-invasive blood pressure measurement. Connect the
equipment as shown:
Expansion Chamber
Tubing
Manometer
To NBP Input
Tools required:
• Reference manometer (includes hand pump and valve), accuracy 0.2% of reading.
• Expansion chamber (volume 250 ml +/- 10%)
• Appropriate tubing.
In service mode, the systolic and diastolic readings indicate the noise of NBP channels 1 and 2
respectively. When static pressure is applied, the reading in NBP channel 1 should be below 50. The
value in parentheses indicates the actual pressure applied to the system.
1Connect the manometer and the pump with tubing to the NBP connector on the MMS and to the
expansion chamber.
2In service mode, select the Setup NBP menu.
3Select Close Valves: On
4Raise the pressure to 280 mmHg with the manometer pump.
5Wait 10 seconds for the measurement to stabilize.
6Compare the manometer values with the displayed values.
7Document the value displayed by the monitor (x1).
8If the difference between the manometer and displayed values is greater than 3 mmHg, calibrate
the MMS. If not, proceed to the leakage test.
9To calibrate the MMS, select Close Valves off then Cali brate NBP and wait for the
instrument to pump up the expansion chamber.Wait a few seconds after pumping stops until
EnterPrVal is highlighted and then move the cursor to the value shown on the manometer. If
one of the following prompt messages appears during this step, check whether there is leakage in
the setup:
– NBP unable to calibrate – cannot adjust pressure
– NBP unable to calibrate – unstable signal
10 Press Confirm.
If the INOP NBP Equipment Malfunction message occurs in monitoring mode, go back to service
mode and repeat the calibration procedure.
35
Page 36
3 Testing and MaintenanceTests
NBP Leakage Test
The NBP leakage test checks the integrity of the system and of the valve. It is required once per year
and when you repair the monitor or replace parts.
1If you have calibrated, repeat steps 2 to 6 from the accuracy test procedure so that you have 280
mmHg pressure on the expansion chamber.
2Watch the pressure value for 60 seconds.
3Calculate and document the leakage test value (x2).
x2 = P1 - P2
where P1 is the pressure at the beginning of the leakage test and P2 is the pressure displayed after
60 seconds.
The leakage test value should be less than 6 mmHg.
NBP Linearity Test
1Reduce the manometer pressure to 150 mmHg.
2Wait 10 seconds for the measurement to stabilize.
3After these 10 seconds, compare the manometer value with the displayed value.
4Document the value displayed by the monitor (x3)
Valve Test
5If the difference is greater than 3 mmHg, calibrate the MMS (see steps 9 to 10 in the accuracy test
procedure).
1Raise the pressure again to 280 mmHg.
2Select Close valves: Off.
3Wait five seconds and then document the value displayed. The value should be less than 10
mmHg.
4Document the value displayed by the monitor (x4).
TestExpected test results
Accuracy testx1 = value displayed by monitor
Difference ≤ 3mmHg
Leakage testx2 = leakage test value
x2 < 6 mmHg
Linearity testx3 = value displayed by monitor
Difference ≤ 3mmHg
36
Valve Testx4 = value < 10 mmHg
Page 37
Tests3 Testing and Maintenance
Sidestream CO2 Performance Test
Allow five seconds between individual service procedures to ensure stable equipment conditions. When
certain monitor procedures are running, service procedures are not possible and trying to start them
will result in a message Service Operation Failed in the monitor’s status line. Wait until
the monitor completes the current operation, then restart the service procedure.
This test checks the performance of the CO
measurement for the sidestream extension. The CO2
2
performance test is required once per year and when the instrument is repaired or when parts are
replaced.
This test uses calibration equipment that you can order (see the Parts section for the part number). The
procedure is summarized in the following steps. Refer to the documentation accompanying the
equipment for detailed instructions.
Tools Required:
• Standard tools, such as screwdriver, tweezers
• Electronic flowmeter, M1026-60144.
• Gas calibration equipment:
• Cal 1 gas 15210-64010 (5% CO
• Cal 2 gas 15210-64020 (10% CO
)
2
)
2
• Cal gas flow regulator M2267A
• Cal tube 13907A
You also need a local barometric pressure rating received from a reliable local source (airport, regional
weather station or hospital weather station) which is located at the same altitude as the hospital.
The CO
calibration for the sidestream extension consists of the following steps:
2
• Barometric pressure check and calibration, if required.
• Leakage check
•Pump check
• Flow check and calibration, if required.
•Noise check
•CO
•CO
Cal check and calibration, if required.
2
Cal verification using 2nd cal gas
2
Perform all checks in the same session.
Barometric Pressure Check and Calibration
Check the barometric pressure value in the sidestream CO2 extension as follows:
1Go into service mode and select Setup CO
2Connect a FilterLine to the sidestream CO
Extension.
3The status line at the bottom of the screen displays “CO2 pressure reading (ambient/cell) xxx/yyy”
where xxx is the ambient pressure and yyy is the measured cell pressure. Check whether the
ambient pressure value (x1) matches (within the acceptable tolerance of ±12mm Hg) the reference
menu.
2
input. This activates the pump in the sidestream CO2
2
37
Page 38
3 Testing and MaintenanceTests
value you have received. If so, proceed to the leakage check. If the value is not correct, calibrate as
follows.
Leakage Check
The leakage check consists of checking the tubing between:
• the pump outlet and the measurement server extension outlet and
• the pump inlet and FilterLine inlet.
Check the user’s guide of the flowmeter for details on how to make a correct flow reading.
Part 1
1Go into service mode and select Setup CO
2Connect a FilterLine to the sidestream CO
3Check the ambient pressure and the cell pressure shown in the monitor’s status line. The cell
a. Select CO
then select Barom.Press to activate a table of values.
2
b. Select the value in the table which matches the reference value received from a reliable local
source (airport, regional weather station or hospital weather station). (The values are displayed
with a resolution of 2 mmHg up to 500 mmHg and a resolution of 1 mmHg from 500 mmHg
to 825 mmHg.) Note: the selected value must be within ±10% of the current measured ambient
pressure, otherwise an error message will occur at restarting the monitor.
c. Confirm the barometric pressure setting.
d. Check that the ambient pressure displayed in the status line at the bottom of the screen is the
same as the value which you selected from the list in step b.
menu.
2
input to start the pump running.
2
pressure should be approximately 20 mmHg lower than ambient pressure.
4Connect the flowmeter outlet to the FilterLine inlet using a flexible connecting tube.
5Block the measurement server extension outlet using your fingertip and observe the flowmeter
display. The value on the flowmeter (x2) should decrease to between 0 and 4 ml/min, accompanied
by an audible increase in pump noise. If the value is within the tolerance limits, continue with part
2 of the leakage check.
6If the value is outside the tolerance limits, there is a leakage between the pump outlet and the
measurement server extension gas outlet.
7Open the measurement server extension and check the tubing connections at the pump outlet and
the extension gas outlet. If the connections are good, then there is a leakage in the tubing and you
must exchange the measurement server extension.
Part 2
1Disconnect the flowmeter from the Part 1 setup and connect the flowmeter inlet to the M3015A
gas outlet.
2Leave the Filterline connected to the M3015A inlet.
3Block the inlet of the FilterLine using your fingertip and observe the flowmeter display. The value
on the flowmeter (x3) should decrease to between 0 and 4 ml/min, accompanied by an audible
increase in pump noise. Do not block the inlet for longer than 25 seconds as this will lead to an
“Occlusion” INOP. If the value is within the tolerance limits, there are no leakages and the leakage
check is completed; proceed to the pump check.
38
Page 39
Tests3 Testing and Maintenance
If the value is not within the tolerance limits, there is a leakage between the FilterLine inlet and the
4
pump inlet.
5Check the FilterLine connections and open the M3015A to check the tubing connections at the
pump inlet and the M3015A gas inlet. If the connections are good, try replacing the FilterLine and
repeating the leakage check. If the situation remains, there is a leakage in the tubing and the
M3015A must be exchanged.
Pump Check
1Connect the flowmeter inlet to the M3015A gas outlet.
2Connect the FilterLine to the M3015A inlet.
3Block the inlet of the FilterLine using your fingertip and observe the cell pressure on the M3046A
display. The cell pressure (x4) should be more than 120 mmHg below the ambient pressure shown.
If the pressure difference is less than 120 mmHg, the pump is not strong enough and you should
replace it, irrespective of the Pump OpTime.
Flow Rate Check and Calibration
Check the flow rate in the sidestream CO2 extension as follows:
Noise Check
1Connect the flowmeter to the CO
2Check on the flowmeter the flow that the sidestream CO
50 ml/min ± 7.5 ml/min. If the value is within tolerance, proceed to the CO
FilterLine.
2
extension pump draws (x5). It should be
2
Gas calibration
2
check. If the value is not within tolerance, calibrate as follows.
3Adjust the flow in the instrument by selecting Increase Flow or Decrease Flow until it
is as close as possible to 50 ml per minute as indicated on the flowmeter gauge.
4When you are satisfied that the flow is set as close as possible to 50 ml per minute, select Store
Flow and confirm the setting. If you do not store the adjusted flow within 60 seconds of the
adjustment, the old flow setting is restored.
5If you cannot adjust the flow to within tolerance, replace the pump. If you still cannot make the
flow adjustment, this indicates a fault in the measurement extension, which must be replaced.
1With the monitor in service mode, select Setup CO
2Disconnect the flowmeter and connect the 5% calibration gas and flow regulator in its place.
3Open the valve to apply the 5% calibration gas and wait until the value is stable.
4Check the noise index (x6) displayed next to the CO
of noise on the CO
wave). If the value exceeds 3 mmHg, replace the measurement extension.
2
menu.
2
value on the display (this indicates the level
2
CO2 Gas Measurement Calibration Check
After switching the measurement extension on, wait at least 20 minutes before checking the
calibration. Check the calibration of the CO
1Check that the 5% calibration gas and flow regulator are connected.
gas measurement as follows:
2
39
Page 40
3 Testing and MaintenanceTests
Calculate the expected measurement value in mmHg as follows:
2
0.05 x (ambient pressure) = value mmHg
for example 0.05 x 736 = 36.8 mmHg (with an ambient pressure of 736 mmHg)
3Open the valve on the flow regulator to allow 5% CO
gas to flow into the extension. Allow the
2
value to stabilize.
4Check that the value on the instrument (measurement value on the main screen, x7)) matches the
calculated mmHg value ± 2.6 mmHg. If the value is outside the tolerance, calibrate as described in
step 9 in this procedure onwards.
5Disconnect the 5% calibration gas and connect the 10% calibration gas.
6Calculate the expected measurement value and tolerance in mmHg as follows:
0.1 x (ambient pressure) = value mmHg
±0.07 x (value mmHg) = tolerance
for example 0.1 x 737 mmHg = 73.7 mmHg (with an ambient pressure of 737 mmHg)
±0.07 x 73.7 mmHg = ±5.16 mmHg tolerance
7Open the valve on the flow regulator to allow 10% CO
gas to flow into the extension. Allow the
2
value to stabilize.
8Check that the value on the instrument (x8) matches the calculated mmHg value within the
calculated tolerance. If so, the measurement extension is correctly calibrated. If the value is outside
the tolerance, calibrate as follows.
9If not already connected, connect the 5% calibration gas.
10 Select Cal. CO
11 Select the value for the calibration gas. (The default value is 5.0%.)
.
2
12 Open the valve on the calibration gas to allow CO
to stabilize before the start of the calibration. Leave the valve open until the instrument gives a
prompt that gas can be removed.
13 The extension calibrates and prompts when calibration is successful.
Calibration Verification
1Reopen the 5% gas valve and allow the value to stabilize.
2Check that the value displayed on the monitor is correct within the tolerance (see step 2 above).
3Disconnect the 5% calibration gas and connect the 10% calibration gas.
4Open the valve on the flow regulator to allow 10% CO
value to stabilize.
5Check that the value displayed on the monitor is correct within the tolerance (see step 6 above).
If one or both values are not within tolerances, you must exchange the measurement server extension.
Reset Time Counters
You must check the time counters on the sidestream CO2 extension before calibrating the instrument.
As well, when parts are replaced, the appropriate counters must be reset to zero.
gas to flow into the extension. Allow the value
2
gas to flow into the extension. Allow the
2
40
Page 41
Tests3 Testing and Maintenance
The counters for CO2 pump, IR Src and Last Cal are displayed in the status line. The values are
updated when entering the Setup CO
menu.
2
Observe the following guidelines:
• When calibrating the CO
extension, if no parts have been replaced, check the displayed values of
2
Reset PumpOpTi me and Reset IRSourceTime selections to make sure that they are
within suggested guidelines for use (15, 000 hours of continuous use). If the counter time is greater
than 15, 000 hours, replace the appropriate part. See Repair and Disassembly for details.
• When calibrating the CO
extension, if parts have been replaced, reset the appropriate values using
2
the Reset PumpOpTime and Reset IRSourceTime selections. See Repair and Disassembly
for details.
Resetting the PumpOpTime generates the INOP: “CO
OCCLUSION”. To clear this INOP you
2
must perform a flow check and store the flow in service mode (select Store Flow).
Table 1 Documenting CO
Test Results
2
TestExpected Test Results
Barometric Pressure
Check
x1 = difference between the reference pressure and the
measured ambient pressure displayed on the monitor
(x1<12 mmHg)
Leakage Check parts
1 and 2
x2 = value of part 1 leakage check on flowmeter
(x2< 4.0 ml/min)
x3 = value of part 2 leakage check on flowmeter
(x3< 4.0 ml/min)
Pump Checkx4 = difference in pressure between cell pressure and ambient
Flow Checkx5 = difference between measured value and 50.0 ml/min
Noise Checkx6 = noise index displayed on monitor (x6<3.0)
CO
Gas
2
Calibration Check
CO
2 Cal Verification
Temperature Accuracy
This test checks the performance of the temperature measurement.
Tools required: Patient simulator (with 0.1
1Connect the patient simulator to the temperature connector on the MMS or measurement server
extension.
2Configure the patient simulator to 40
pressure displayed on the monitor during occlusion (x4 >120
mmHg)
(x5<7.5 ml/min)
x7 = difference between measured CO2 value and calculated
value, based on 5% CO
cal. gas. (x7 < 2.6 mmHg)
2
x8 = difference between measured CO2 value and calculated
value, based on 10% CO
cal. gas.
2
(x8 < ± {0.07 x value calculated})
o
C or 0.2oF).
o
C or 100 oF.
41
Page 42
3 Testing and MaintenanceTests
The value should be 40 oC ± 0.2 oC or 100 oF ± 0.4 oF.
3
ECG/Resp Performance Test
This test checks the performance of the ECG and respiration measurements.
Tools required: Patient simulator.
ECG Performance
1Connect the patient simulator to the ECG/Resp connector on the measurement server.
2Configure the patient simulator as follows:
– ECG sinus rhythm.
– HR = 100 bpm.
3Check the displayed ECG wave and HR value against the simulator configuration.
4The value should be 100bpm +/- 2bpm.
Respiration Performance
1Change the Patient Simulator configuration to:
– Base impedance line 1500 Ohm.
– Delta impedance 0.5 Ohm.
– Respiration rate 40 rpm.
2The value should be 40 rpm +/- 2 rpm.
Invasive Pressure Performance Test
This test checks the performance of the invasive pressure measurement.
Tools required: Patient simulator.
1Connect the patient simulator to the pressure connector on the MMS or the measurement server
extension.
2Set the patient simulator to 0 pressure.
3Make a zero calibration.
4Configure the patient simulator as P(static) = 200 mmHg.
5Wait for t he d isp lay.
6The value should be 200 mmHg ± 5 mmHg. If the value is outside these tolerances, calibrate the
MMS or measurement server extension. If the MMS was calibrated with a dedicated reusable
catheter, check the calibration together with this catheter.
SpO2 Performance Test
This test checks the performance of the SpO2 measurement.
Tools required: none
42
1Connect an adult SpO
2Measure the SpO
value on your finger (this assumes that you are healthy).
2
transducer to the SpO2 connector on the MMS.
2
Page 43
Tests3 Testing and Maintenance
The value should be between 95% and 100%.
3
Cardiac Output Performance Test
These tests check the performance of the cardiac output measurement.
1Connect the patient simulator to the C.O. module using the patient cable.
The nurse call relay performance test can be performed either at the phone jack type connector (this
only tests one relay) or at the multi-port nurse call connector (to test all three relays).
Phone Jack Type Connector Test
This test checks the operation of the Nurse Call Relay. The Nurse Call Relay test is recommended for
customer sites where the nurse call is in use. The Nurse Call relay functions as follows:
• Standard Operation—Relay open.
• Alarm Condition—Relay closed.
Tools required: Ohmmeter.
1Plug a phono connector into the Nurse Call Relay connector.
2Connect the ohmmeter.
3If no alarm occurs, the relay contacts are open. When an alarm occurs, the relay contacts close.
Multi-Port Nurse Call Connector Test
This test checks the operation of the Flexible Nurse Call Relay. The Nurse Call Relay test is
recommended for customer sites where the nurse call is in use. The following diagram and table show
the pins and relay identifiers of the connector:
44
Page 45
Tests3 Testing and Maintenance
PinCable Color Coding Relay
1black R2-closure
2brownR2-middle
3redR2-opener
4orangeR3-closure
5yellowR3-middle
6greenR3-opener
7bluen/a
8purplen/a
9grayn/a
10whiten/a
11pinkR1-closure
12light greenR1-middle
13black/whiteR1-opener
14brown/whiten/a
15red/whiten/a
16orange/whiten/a
17blue/whiteR_failure_closure
18purple/whiteR_failure_middle
19green/whiteR_failure_opener
20red/blackn/a
The Nurse Call relay functions as follows:
• During standard operation R1,R2,R3 _opener are closed; R1,R2,R3_closure are open.
• During alarm condition R1,R2,R3_opener are open; R1,R2,R3_closure are closed.
Tools required: Ohmmeter.
1Plug an M8087-61001 cable into the Nurse Call Relay connector.
2Connect the ohmmeter and measure the pins as indicated in the diagram and table.
3The relay contacts should behave as described above. The behavior may vary depending on
configuration choices. See the Configuration Guide for details on Alarm Relay settings.
ECG Sync Performance Test
This test checks the performance of ECG synchronization between the monitor and a defibrillator. It
only needs to be performed when this feature is in use as a protocol at the customer site.
Tools required:
• Defibrillator with ECG Sync and Marker Output.
•Patient simulator.
1Connect the patient simulator to the ECG connector on the Measurement server and the
defibrillator to the ECG Sync Output on the monitoring.
45
Page 46
3 Testing and MaintenanceTests
Set the patient simulator to the following configuration:
2
– HR = 100 bpm.
– ECG sinus rhythm.
3Switch the defibrillator to simulation mode.
4Check that the marker pulse is displayed before the T-wave begins.
VueLink Tests using VueLink Test Module
Use the VueLink plug-in test module ( M1186-60510) to test M1032A VueLink modules.
Test Procedure
You must preselect the test module to ON in Configuration Mode. Therefore, the test module must be
one of the devices made available for selection during configuration of the VueLink module.
Carry out the test itself in monitoring mode. For information concerning the configuration of
VueLink modules see the M1032A VueLink Module Handbook.
1Plug the VueLink module into the FMS.
2Press the Setup key on the front of the VueLink module.
3In the Setup Vu eL in k menu select Device, then select Test Module.
4Select Confirm to store the selection and wait for the message “Switched to new device”.
5Plug in the test module.
6Connect the modules by plugging one end of the cable (part number M1032-61661) into the
connector on the front of the VueLink Module, and the other end into the connector on the front
of the Test Module.
7Select the wave segment on the screen, where you want the waves to appear. In the wave menu,
select Change Wave, then select WAVE.
8Select the VueLink SmartKey, then select the TEST Plug-In pop-up key
The test module acts in the same way as an external device would, and sends signals to the VueLink
module in both analog and digital form. The computer module checks these signals for validity, and
then displays “passed” or “failed” on the screen.
The wave segment displays two waveforms, a triangular one and a rectangular one. These are displayed
alternately and for a period of ten seconds each. The expected curve type is indicated below the wave.
There are two pairs of gridlines that indicate the permitted range for the max/min values of these
waves. If all the data received by the test module is correct, the waves will lie within the specified
ranges. If either limit of either wave falls outside the respective gridlines, then the module being tested
is faulty regardless of the passed/failed messages.
When the test is complete:
46
1Disconnect the cable that joins the test module to the VueLink module.
2In configuration mode, ensure that the test module is not selected, and the preselected devices are
the same ones as before the test. Also, verify the settings for these devices.
3Return to monitoring mode.
4Press the Setup key on the front of the VueLink module and select the required device by selecting
Device in the Setup VueLink menu.
Page 47
Tests3 Testing and Maintenance
NOTE
It is important to ensure that the preselected device drivers are configured exactly the same as they were
before the test, including their default settings.
Safety Testing
You are recommended to file the results of safety tests. This may help to identify a problem early
particularly if the test results deteriorate over a period of time.
Warnings, Cautions, and Safety Precautions
• These tests are a proven means of detecting abnormalities that, if undetected, could prove dangerous
to either the patient or the operator.
• You can perform all tests using commercially available Safety Analyzer test equipment. You can
perform basic measurements with widely available multifunction instruments such as the HP 3469A
multimeter or equivalent.
• The consistent use of a Safety Analyzer as a routine step in closing a repair or upgrade is emphasized
as a mandatory step to maintain approval agency status. You can also use the Safety Analyzer as a
troubleshooting tool to detect abnormalities of line voltage and grounding plus total current loads.
• For Europe and Asia/Pacific according to:
IEC60601-1:1988 + A1:1991 + A2:1995 = EN60601-1:1990 +A1:1991 + A2:1995
For USA according to:
UL2601-1
• Additional tests may be required according to local regulations.
•Normally, a Safety Analyzer is used to perform these procedures. Popular testers include the
DEMPSEY 232D, or for use in Europe, testers like the Rigel, Metron or Gerb. Follow the
instructions of the Instrument manufacturer. If the Dempsey is used for an extended length of time,
it could be damaged by the high amp current draw of the system.
• Any device with mains connection that is connected to the medical device must comply with
IEC60601-1 if within patient vicinity and be separately tested at the same intervals as the monitor
Safety Test Procedures
Use the test procedures outlined here only for verifying safe installation or service ofthe product. The
setups used for these tests and the acceptable ranges of values are derived from local and international
standards but may not be equivalent. These tests are not a substitute for local safety testing where it is
required for an installation or a service event. If using the Metron Safety tester, perform the tests in
accordance with your local regulations, for example in Europe use IEC60601-1/IEC60601-1-1 and in
the US use UL2601-1. The Metron Report should print results with the names listed below, together
with other data.
NOTE For any system with external displays: Disconnect the display from the medical device and perform
S(1) and S(2) on each device with a mains cable. If both pass the tests reconnect the display and
procede with normal use.
47
Page 48
3 Testing and MaintenanceTests
S(1) Part 1: System Enclosure Leakage Current - NC (normal condition)
Medical electrical system
Instrument under test
L (N)
Signal parts
in- and/or
output
Signal parts
(**)
in- and/or
output
Other
Instrument
N (L)
S1
S4
(*)
S2
PE
MD
(*) Not present in Class 2.
(**) Can be multiple different connections to different equipment at same time.
This measures leakage current of exposed metal parts of Instrument under Test (IUT) and between
parts of the system within the patient environment; normal and reversed polarity using S2.
Safety test according IEC 60601-1 / UL2601-1
S(1) Part 2: System Enclosure Leakage current - Single Fault (open earth)
48
Medical electrical system
Instrument under test
L (N)
S2
PE
(**) Can be multiple different connections to different equipment at same time.
This measures leakage current of exposed metal parts of Instrument under Test (IUT) with Protective
Earth (PE) open circuit (S4 = open) and between parts of the system within the patient environment;
normal and reversed polarity using S2.
S(2) Protective Earth Continuity
Instrument under test
L (N)
Ri
Ohm
6V
50 Hz 25 A or 1.5 Ir
N (L)
PE
(*)
Insulating pad
(*) If equipotential connection present : measure also
with yellow/green E.P. conductor connected.
This measures impedance of Protective Earth (PE) terminal to all exposed metal parts of Instrument
under Test (IUT), which are for safety reasons connected to the Protective Earth (PE). Test current 25
Amp applied for 5 to 10 seconds.
49
Page 50
3 Testing and MaintenanceTouchscreen Calibration
S(3) Patient Leakage current - Single Fault Condition (S.F.C.) mains on applied part
Measures patient leakage current from applied Part to earth caused by external main voltage on applied
Part with switch S5 open and closed. Each polarity combination possible is tested using S2 and S6.
This test is applicable for every measurement input .
Touchscreen Calibration
Please refer to the documentation shipped with your selected display for details on touchscreen
calibration procedures.
Instrument under test
PE
Insulating pad
S5
Applied part
Signal part
in- and/or
output
R
MD
50
Page 51
4Troubleshooting
Introduction
This section explains how to troubleshoot the monitor if problems arise. Links to tables that list
possible monitor difficulties are supplied, along with probable causes, and recommended actions to
correct the difficulty.
How To Use This Section
Use this section in conjunction with the sections Testing and Maintenance and Parts. To remove and
replace a part you suspect is defective, follow the instructions in the section Repair and Disassembly.
The Theory of Operation section offers information on how the monitor functions.
4
Who Should Perform Repairs
Only qualified service personnel should open the monitor housing, remove and replace components, or
make adjustments. If your medical facility does not have qualified service personnel, contact Philips’
Response Center or your local Philips representative.
WARNINGHigh Voltage - Voltages dangerous to life are present in the instrument when it is connected to the
mains power supply. Do not perform any disassembly procedures (other than server and extension
removal) with power applied to the instrument. Failure to adhere to this warning could cause serious
injury or death.
Replacement Level Supported
The replacement level supported for this product is to the printed circuit board (PCB) and major
subassembly level. Once you isolate a suspected PCB, follow the procedures in the Repair and Disassembly section, to replace the PCB with a known good PCB. Check to see if the symptom
disappears and that the monitor passes all performance tests. If the symptom persists, swap back the
replacement PCB with the suspected malfunctioning PCB (the original PCB that was installed when
you started troubleshooting) and continue troubleshooting as directed in this section.
51
Page 52
4 TroubleshootingSoftware Revision Check
Software Revision Check
Some troubleshooting tasks may require that you identify the Software Revision of your monitor. You
can find the software revision along with other information, such as the system serial number, in the
monitor revision screen.To access the monitor revision screen:
1Enter the Main Setup menu and select Revision
2Select the pop-up key Software
NOTE The system serial number can also be found on the right side and on the bottom of the MP90 CMU.
Obtaining Replacement Parts
See Parts section for details on part replacements.
Troubleshooting Guide
Problems with the monitor are separated into the categories indicated in the following sections and
tables. Check for obvious problems first. If further troubleshooting instructions are required refer to
the Troubleshooting Tables.
Taking the recommended actions discussed in this section will correct the majority of problems you
may encounter. However, problems not covered here can be resolved by calling Philips Response
Center or your local representative.
Checks for Obvious Problems
When first troubleshooting the instrument, check for obvious problems by answering basic questions
such as the following:
1Is the power switch turned on?
2Is the AC power cord connected to the instrument and plugged into an AC outlet?
3Are the MMS and, if present, the measurement server extension inserted correctly?
4Are the cables connected properly to the FMS?
5Are the parameter modules plugged into the FMS correctly?
Checks Before Opening the Instrument
You can isolate many problems by observing indicators on the instrument before it is necessary to open
the instrument.
NOTE It takes several seconds for the AC Power LED to switch on / off after the mains power cord has been
connected / disconnected.
Checks with the Instrument switched Off
• AC connected:
– AC Power LED is on (green).
• No AC connected:
52
Page 53
Troubleshooting Guide4 Troubleshooting
–All LEDs are off.
Checks with the Instrument Switched On, AC connected
When the monitor is first switched on, all the LEDs on the power supply and the remote alarm device
light up momentarily. The location of the power supply LEDs is shown in the following photograph:
∼AC Power LED
(Green)
Error LED
(Red)
Power On LED
(Green)
Initial Instrument Boot Phase
The following tables describe the regular initial boot phase of the monitor and its components. If the
boot phase does not proceed as described below go to Boot Phase Failures for Troubleshooting
information.
Monitor Boot Phase:
For these steps it is assumed that the Monitor is powered correctly and the +5V System Board supply
voltage is okay. This is indicated by the green Power On LED.
Time (sec.)
after Power
On
0When the Power On/Off button is pressed, the green Power On LED and the red error LED
1The alarm LEDs on the remote alarm device are switched on with low intensity. Colors: Left
3Red Error LED is switched off.
4Boot Screen with the Philips Logo appears on the display. Test Sound is issued.
5All Alarm LEDs are switched off.
8Fixed screen elements (for example smart keys, alarm fields) appear on the screen.
15-30First measurement information appears on the screen,user input devices (for example
Event
switch on immediately.
LED:cyan; Middle LED:red; Alarm Suspend LED (right): red
Boot Screen with the Philips Logo disappears
Mouse, Touch, Speed Point) are functional
53
Page 54
4 TroubleshootingTroubleshooting Guide
Flexible Module Server Boot Phase
For these steps it is assumed that the Flexible Module Server is connected via MSL-cable to the
monitor
Time (sec. )
after
Monitor
Power On
0Red Error LED switches on immediately
1Green “Ready” LED switches on
3Red Error LED is switched off
5Module Power is switched on
5-8Module Status LEDs blink once or twice (Module dependent)
Event
Troubleshooting Tables
The following tables list troubleshooting activities sorted according to symptoms. Click on the links
below to view a particular table.
How to use the Troubleshooting tables
The possible causes of failure and the remedies listed in the troubleshooting tables should be checked
and performed in the order they appear in the tables. Always move on to the next symptom until the
problem is solved.
Boot Phase Failures
External Display is blank
External Touch Display not functioning
Remote Alarm Device
Remote Extension Device
Keyboard/Mouse not functioning
Network related problems
Wireless Network
Multi-Measurement Server
MSL-related problems
Alarm Lamps
Alarm Tones
Individual Parameter INOPS
Flexible Module Server
Printer
MIB / RS232
Flexible Nurse Call Relay
54
Page 55
Troubleshooting Guide4 Troubleshooting
Boot Phase Failures
SymptomsPossible Causes of FailureFailure Isolation and Remedy
AC LED does
not light up
Green Power On
LED and Red
Error LED
remain off after
pressing power
on button on
power supply
AC Connection not okCheck that the AC-Mains are powered and
the power cord is ok and connected
LED defective
Power Switch board not connected to the
main board
Power supply defective
Secondary main board defective
Primary Video Board defective
Backplane Bottom defective
Primary Main Board defective
Remote Devices
Power Switch Micro Controller hung
Power switch board not connected to the
main board
Power Switch Board defective
I/O Board defective
Primary Video Board defective
Primary Main Board defective
Try to switch on the monitor. If it operates
normally , the LED is defective => exchange
Power Switch board.
Check if power switch board is connected
correctly to the Main Board
Remove power supply and check if output
voltage is within the specifications (47V 49V). Measure on multi-colored wired
connection between red and black wires
Exchange power supply if defective
Disconnect cable from the backplane top to
backplane bottom.
If failure persists continue with ‘Primary
Video Board defective”
Otherwise: Reconnect cable, remove
secondary video board and check again.
Exchange main board and check again.
Replace backplane top.
Remove primary Video Board and check
again
Remove Backplane Bottom and check again
Exchange primary Main Board
Disconnect all connections to the remote
devices and try to switch on the monitor
again
Unplug AC Mains and replug after 10
seconds. Try to switch on the monitor again.
Check if power switch board is connected
correctly to the primary main board.
Exchange Power Switch BOard and try to
switch the monitor on again.
Remove all I/O boards and try to switch the
monitor on again
Remove primary video board and try again
Exchange primary main board. Add boards
in reverse order and try again with each
board.
55
Page 56
4 TroubleshootingTroubleshooting Guide
SymptomsPossible Causes of FailureFailure Isolation and Remedy
Green Power On
LED and Red
Error LED
remain off after
pressing power
on button on
remote extension
device or remote
alarm device.
Green On/
Standby LED or
Red Error LED
remain off after
pressing Power
on button:
Red Error LED
stays on
continuously
Red Error LED
blinks
(indicating cyclic
reboots)
Cable to remote device
Remote device
Power switch board not connected to the
main board
Remote device interface board
Power Switch Board not connected to the
main board
Power Switch board defective
Main board defective
External connected device defectivedisconnect all external cables (except AC)
I/O Board defective
Backplane top defective or secondary main
board defective (opt. #E30)
Primary Video Board defective
Backplane Bottom defective
Primary Main board defective
Hardware Failure
Software Fault
Hardware Failure
Check if cable to remote device is connected
correctly to the remote device interface board
and the remote device.
Use another remote device on this port of the
remote device interface board and try to
switch on the monitor again.
Check if power switch board cable is
connected to the primary and secondary
main board (opt.#E30)
Check if monitor can be switched on by the
power switch on the power supply. If this is
possible replace the remote device interface
board. Otherwise proceed as described in
“Green Power On LED and Red Error LED
remain off after pressing power on button on
power supply” above.
check if power switch board is connected
correctly to the main board
exchange Power switch board
exchange main board
and switch the monitor on again
Remove all I/O boards and switch the
monitor on again.
Disconnect cable from the backplane top to
the backplane bottom.
Remove primary video board and switch on
again
Remove Backplane Bottom and check again.
Exchange primary Main board
connect Support Tool directly to monitor
with crossover cable and start “search for
defective devices”
If no device is detected, proceed as described
above in section “Red error LED stays on
continuously”
If the Support Tool can detect the device and
it indicates the Operating Mode is ‘Boot’,
download and store the status log. Reload
software and re-clone the monitor. If this
fixes the problem e-mail the status log to
your local response center
If this does not rectify the problem follow
instructions under “Red Error LED stays on
continuously”.
56
Page 57
Troubleshooting Guide4 Troubleshooting
SymptomsPossible Causes of FailureFailure Isolation and Remedy
No Test Sound
issued
Speaker defective
Main board defective
check for INOPs and follow instructions
exchange speaker
exchange main board
57
Page 58
4 TroubleshootingTroubleshooting Guide
External Display is blank
SymptomsPossible Causes of FailureFailure Isolation and Remedy
External Display is blank Video cable to external display
not connected
External display has no power
External display is defective
Video board defective
Main board defective
External Touch Display not functioning
SymptomsPossible Causes of FailureFailure Isolation and Remedy
Touch Screen not
functioning
Touch position invalidTouch not calibratedPerform touch calibration:
External Touch cable not connected Check cable connection from
External Touch driver configuration
MIB Board defective
External touch defective
Main board defective
Check video cable connection to external
display
Check electricity supply of external
display
Check external display and video cable
on another monitor or PC
Replace video baord
Replace main board
external touch to MIB board
Check RS232/MIB
configuration:
1. Enter Main Setup menu
2. Select Monitor
3. Select Hardware
4. Reconfigure RS232/MIB
drivers
5. if problem persists, proceed to
the next step
Replace MIB board
Replace external touch
Replace Main board
1. Enter Main Setup menu
2. Select Monitor
3. Select Hardware
4. Select Touch Driver
5. Select Calibrate
58
Page 59
Troubleshooting Guide4 Troubleshooting
General Monitor INOP Messages
INOP MessagePossible Causes of FailureFailure Isolation and Remedy
CHECKINTERNVOLTAGE Problem with the voltages (5V, 12V) in
the monitor
CHECK MENU LABELS
This message indicates that
the monitor has performed a
cold start and attempted to
reset the menu labels to
defaults. If the defaults are
present, the recovery was
successful. Default settings
should be recorder as part of
your documented
configuration setup. If the
menu labels do not match the
configuration file settings or
the INOP persists, perform
the troubleshooting tasks
listed in this table.
CHECK MONITOR TEMP The temperature inside the monitor is
CHECK SETTINGS
This message indicates that
the monitor has performed a
cold start and attempted to
reset the menu labels to
defaults. If the defaults are
present, the recovery was
successful. Default settings
should be recorded as part of
your documented
configuration setup. If the
menu labels do not match the
configuration file settings or
the INOP persists, perform
the troubleshooting tasks
listed in this table.
Memory space in which the menu labels
are stored has been corrupted
Main Board defective
too high
Monitor ventilation obstructed
Main Board defective
Memory space in which the settings are
stored has been corrupted
Main board defective
Remove all I/O boards and put them
back in one at a time to isolate any
defective board. If this does not
resolve the problem, replace the main
board
Reclone configuration file. This will
reload the memory space
Replace Main Board
Check the environment for possible
causes
Clean the monitor ventilation
internally and then cool monitor
down for 8 hours
replace Main Board
Reclone configuration file. This will
reload the memory space.
Replace Main board
59
Page 60
4 TroubleshootingTroubleshooting Guide
INOP MessagePossible Causes of FailureFailure Isolation and Remedy
SETTINGS
MALFUNCTION
INTERNAL COMM.MALF. Problem with the I2C Bus
Problem during cloning process.Reclone configuration file
Memory space in which the settings are
stored has been corrupted
Main board defective
communication in the monitor
Video board defective
Main board defective
Remote Alarm Device
SymptomsPossible Causes of FailureFailure Isolation and Remedy
Alarm LEDs illuminate,
but no alarm sound is
issued
Alarm occurs on screen, but
no LED or alarm sound on
the alarm device
Alarm sound is isued, but
no LEDs light up
wrong I/O slotcheck I/O matrix in the Theory
speaker defective
remote device I/F defective
cabling not connectedcheck cabling
cabling defetive
I/O board defective
Remote Alarm Device defective
LED failureReplace Alarm Device
Reclone configuration file. This will
reload the memory space.
Replace Main board
Disconnect the external display and
try another one
Replace Video board
Replace Main board
of Operation section of this
manual
replace remote alarm device
replace I/O board
replace cable
replace I/O board
replace Remote Alarm Device
Remote Extension Device
SymptomsPossible Causes of FailureFailure Isolation and Remedy
Remote input device (for
example mouse/keyboard)
attached to the Remote
Extension Device does not
function
Buttons on the Remote
Extension Device do not
function but input device
attached is functioning
60
See table Keyboard/Mouse not
functioning
Remote Extension Device defective
See tables Keyboard/Mouse not
functioning
replace Remote Extension Device
Page 61
Troubleshooting Guide4 Troubleshooting
SymptomsPossible Causes of FailureFailure Isolation and Remedy
Speed Point attached to
Remote Extension Device
not functioning
Speed Point Knob
Rotation, Joystick Control
or Selection control not
functioning
INOP Message CHECK
INPUT DEVICES is
issued
Keyboard/Mouse not functioning
Remote Extension Device is not
connected to the monitor.
SpeedPoint not connected properly
SpeedPoint defective
Remote Extension Device defective
Remote Device I/O board in the
wrong slot
Remote Device I/O board defective
SpeedPoint defective
SpeedPoint or other input device
defective
Check cabling and connections
Check cabling to SpeedPoint in
the Remote Extension Device
Replace SpeedPoint
Replace Remote Extension
Device
Check I/O Matrix in Installation
Instructions
Replace I/O board
Replace Speed Point
Perform a visual and functional
check of all the monitor input
devices. Replace input devices if
necessary.
SymptomsPossible Causes of FailureFailure Isolation and Remedy
Keyboard/Mouse attached
directly to the monitor not
functioning
Keyboard/Mouse attached
to Remote Extension Box
not functioning
Keyboard/Mouse not connected
properly
Keyboard/Mouse defective
PS/2 I/O board in wrong slot
PS/2 I/O Board defective
Remote Extension Box is not
connected to the monitor or Input
Device is not connected to Remote
Extension Box
Keyboard/Mouse defective
Remote Extension Device defective
Remote Device I/O board in wrong
slot
Remote Device I/O board defective
Check cabling
Replace Keyboard/Mouse
Check I/O Matrix in “Theory of
Operation”
replace I/O board
Check cabling and connections
Replace Keyboard/Mouse
Replace Remote Extension
Device
Check I/O Matrix in Theory of
Operation
Replace I/O board
61
Page 62
4 TroubleshootingTroubleshooting Guide
Network related problems
SymptomsPossible Causes of FailureFailure Isolation and Remedy
Prompt Message “no
central assigned to this bed”
is issued
INOP “Unsupported
LAN” is issued.
No connectivity to PIC, no
prompt or error message on
monitor
Other Bed Overview not
available
“Other Bed” Alarms are not
appearing
The monitor label is not set in the
monitor (if the beds are “monitor
labeled” in the Philips Information
Center)
Problem with the Philips
Information Center to Switch
communication (if the beds are
“port mapped” in the Philips
Information center
Network failureCheck if switches, Philips
Monitor connected to wrong
network
IP address conflict after
infrastructure re-installation
Hardware DefectCheck LAN cable connection
Configuration problem
Configuration ProblemCheck configuration in PIC
This function is not available for
wireless beds
Configuration problemVerify configuration in PIC, in
Set Monitor Label in Config
Mode
Check PIC to Switch
communication, Switch
configuration and Firmware
status
Information Center and Database
Server are all running and
connected to the network
Check if monitor has been
connected for example to a
different hospital network instead
of the Philips Clinical Network
Reboot Database Server and
Philips Information Center
Check NGN Connector board in
Monitor
Check Switch
Check switch configuration and
firmware revision
regarding other bed overview
(care group assignment)
Verify configuration of switch
(setting of multicast filters)
Switch to a wired configuration
Monitor (Config Mode) and
check that the feature is not
temporarily disabled by the user
(Bed Info Window)
62
Page 63
Troubleshooting Guide4 Troubleshooting
Wireless Network
SymptomsPossible Causes of FailureFailure Isolation and Remedy
No central monitoring
possible (see also Network
related problems)
No connectivity (coverage
area consists of multiple
access points and in some
parts of the area there is no
connectivity)
Frequent dropouts and
network disconnects
Overview, Printing does
not work
Monitor is out of range of the access
point (in this case the yellow sync
LED on the wireless adapter on the
bottom of the device next to the
power connector is not on steady)
Wireless Adapter has no power
(LEDs on adapter are all off)
(Only after first install) Firmware
revision in adapter is wrong
Wireless adapter defective (the red
status LED on the adapter’s top
panel is on)
Wrong configuration in wireless
adapter or in access point
Configuration problem
Excessive interference by other radio
equipment or by microwave ovens
System capacity exceeded in
coverage area
Configuration problem
Some functions are not available on
a wireless network
Move monitor back into coverage
area
Verify size of coverage with the
site survey tool
Check splitter cable and replace if
necessary
Check network adapter board in
monitor and replace if necessary
Check adapter itself. Replace if
necessary
Update adapter firmware with
wireless support tool
Replace wireless adapter
Check configuration with wireless
support tool
Verify the channel, domain and
security ID settings of the access
points in the coverage area
Check statistics that can be read
from the wireless adapter via
RS232 or via logging application
in the PIC. Remove interfering
equipment.
Check configuration guidelines
for number of monitors per access
point.
Check access point configuration
with wireless support tool, in
particular the multicast filters
Connect to cabled network
63
Page 64
4 TroubleshootingTroubleshooting Guide
Multi-Measurement Server
SymptomsPossible Causes of FailureFailure Isolation and Remedy
Prompt message
“Measurement Server
Configuration not
supported” is issued
INOP Message “MsmtSrv
not Supp” is issued
Prompt message
“Measurement Server not
supported, unplug device,
switch monitor off/on” and
INOP “Bad Measurement
Server are issued
MSL-related problems
An unsupported Measurement
Server Extension has been connected
Measurement Server Extension is
defective
Measurement Server defective
Wrong So f tware Re v isonUpgrade monitor and/or
Too many measurement servers
connected
Unsupported type of measurement
server (for example M3000A on a
M800xA monitor) connected.
M3000A Measurement Server
Revision A is plugged. This
Measurement Server is not
compatible with the IntelliVue
patient monitors
Disconnect the measurement
server extension
Replace Measurement Server
Extension
Replace Measurement Server
measurement server to a
matching software version
Disconnect unsupported
measurement servers for proper
operation
Disconnect the unsupported
measurement server
Disconnect the measurement
server and cycle power.
SymptomsPossible Causes of FailureFailure Isolation and Remedy
Measurement Server does
not start up (no LEDs
active), no INOP or
prompt displayed
Measurement Server does
not start but LEDs are
normal
MSL Power High INOP is
issued
Note: if this condition
persists for longer than 15
minutes, the INOP MSL
Power Off will appear (see
below)
No Power
Communication lines in MSL cable
or MSL connector broken
MSL connector board defective
Attached devices drawing too much
power from the monitor. Too many
FMS and MMS connected to the
monitor
Check MSL cable and replace if
necessary
Check MSL connector board and
replace if necessary
Check MSL cable and MSL
connectors
Check MSL connector board and
replace if necessary
Reduce to a limit of 1 FMS and 1
MMS connected to the monitor
64
Page 65
Troubleshooting Guide4 Troubleshooting
SymptomsPossible Causes of FailureFailure Isolation and Remedy
MSL Power Off INOP is
issued
MSL Power Overload
INOP is issued
INOP BAD SERVER
LINK is issued
INOP Message
SERVERLINK MALF is
displayed, audible
indicator: a beep every two
seconds
Attached devices drawing too much
power from the monitor. Too many
FMS and MMS connected to the
monitor
Short Circuit within MSL systemDisconnect all MSL connections,
Unexpected data detected on MSLCheck cable and power cycle the
An FMS or MMS with an
incompatible software revision is
connected to the monitor.
Communication between the
components not functioning
The hardware for communicating
with the Multi-Measurement Server
is faulty.
Disconnect all FMS and MMS
from the monitor
Cycle power to restore power to
the MSL devices.
If the message disappears,
reconnect FMS and MMS one at
a time, waiting 15 minutes
between each device to see if
message reoccurs. If yes, the
respective MMS or FMS is faulty.
See Multi-Measurement Server or
Flexible Module Server for
troubleshooting tasks. If no, add
front-end modules one at a time,
waiting 15 minutes between each
module to see if message
reappears, Replace module if
faulty.
Note: If an individual defective
device is connected the MSL
Power High o r MSL Power
Overload INOPs will appear
initially. The MSL Power Off
INOP will not occur for at least
15 minutes.
2nd MSL interface, 1st MSL
interface. Reconnect devices one
at a time. If message persists,
replace main board.
monitor
Connect FMS or MMS with
compatible software revsion
Check software versions and
model number of devices for
compatibility
Check MSL cable, replace if
necessary.
Check MSL I/O board. Replace if
necessary
Check FMS or MMS connector
board. Replace if necessary.
65
Page 66
4 TroubleshootingTroubleshooting Guide
SymptomsPossible Causes of FailureFailure Isolation and Remedy
Alarm Issues
Alarm Lamps
A measurement supported
by a server does not come
up on the monitor
Prompt message “Too
many <label> modules
connected” is issued
The ECG Out/ Marker In
function does not funcion
Label conflictA parameter label from this
measurement is already in use in
the monitor. Check the conflict
window to select the
measurement.
There are more modules of the type
<label> connected than supported
by the software
Hardware problemCheck MSL cable
Remove the unsupported module
or use the lebel manager
application in the monitor to
disable the module.
Check ECG Out Hardware in the
monitor
Check the MSL connector in the
measurement server
SymptomsPossible Causes of FailureFailure Isolation and Remedy
INOP Message Check
Alarm Lamps is issued
Alarm occurs, but no LED
lights up
Remote Alarm Device is not
connected to the monitor
Remote Alarm Device defective
Remote Device I/O board in the
wrong slot
Remote Device I/O board defective
Main board defective
Environmental Lighting too brightPlace the alarm device in a darker
Remote Alarm Device is not
connected to the monitor
Remote Alarm Device defective
Remote device I/O board in the
wrong slot.
Remote device I/O board defective
Main Board defective
Check cabling and connections
Replace Remote Alarm Device
Check I/O Matrix in Installation
Instructions
Replace I/O board
Replace main board
environment
Check cabling and connections
Replace Remote Alarm Device
Check I/O matrix in Installation
Instructions
Replace I/O board
Replace Main board
66
Page 67
Troubleshooting Guide4 Troubleshooting
Alarm Tones
SymptomsPossible Causes of FailureFailure Isolation and Remedy
INOP Message SPEAKER
MALFUNCTION is
displayed
Alarm occurs but no alarm
sound is issued
Alarm occurs on device
connected to VueLink but
no alarm sound is issued on
the monitor
Alarm sound is issued on
monitor but not on Remote
Alarm Device
Speaker cable disconnectedReconnect speaker cable
Speaker defective
Sound amplifier on main board
defective
Audible alarm indicators have been
switched off
Volume set to 0
Speaker defective
Sound amplifier on main board
defective
Configuration of VueLink is
incorrect
Remote Alarm Device is not
connected to the monitor.
Remote Alarm Device defective
Remote Device I/O board in the
wrong slot
Remote Device I/O board defective
Replace speaker
Replace main board
Switch audible alarm indicators
back on
Increase volume
Replace speaker
Replace main board
Check VueLink configuration
Check cabling and connections.
Replace Remote Alarm Device
Check I/O matrix in Installation
Instructions
Replace I/O board
Alarm Behavior
If your monitor did not alarm in the way in which the end user expected, please consult the
Instructions for Use for possible setup issues or configuration settings which could affect alarm
behavior.
Individual Parameter INOPS
If any of the following parameter INOP messages are issued try the respective parameter in another
device. If the INOP message persists replace the parameter module, the MMS or other indicated
device.
•CO
•ECG EQUIP MALF
•NBP EQUIP MALF
•P1 EQUIP MALF
• RESP EQUIP MALF
•SpO
•SpO
•SvO
•tcpO
EQUIP MALF
2
EQUIP MALF
2
TRANSDUC MALF
2
EQUIP MALF
2
(or tcpCO2) EQUIP MALF
2
67
Page 68
4 TroubleshootingTroubleshooting Guide
•T1 EQUIP MALF
• VueLnk EQU. MALFI
Flexible Module Server
SymptomsCause of FailureFailure Isolation and Remedy
Prompt Message
“Unrecognized
Measurement Module in
slot s” is issued
Prompt message
“Measurement Module in
slot n is currently ignored”
is issued
Red Error LED stays onUnrecoverable hardware selftest
Red Error LED flashesHardware selftest errorIf system comes up, check status
Flexible Module Server
LEDs ok, Front End
Measurement Module not
recognized (no prompt or
INOP)
An unsupported module has been
plugged into the Flexible Module
Server
Too many modules of the same kind
have been plugged into the Flexible
Module Server
error:
MSL cable defective
Flex connector from main board to
MSL defective
CPU module defective
Measurement Module or
Measurement Module Connector
defective
No Front-End power because MSL
voltage from the monitor is too high
or too low
Mother board or connector on
Mother Board defective
Unplug the unsupported module
Unplug module in slot n
Try to attach the MMS directly to
the MSL cable. If the
measurements show up on the
screen, the fault is in the FMS
If the measurements do not show
up when the MMS is connected
directly to the MSL cable, then
replace MSL cable
Replace connector
Replace CPU module
log. Otherwise see above
Replace Measurement Module
Try a new MSL cable. Replace if
failure is rectified.
Replace mother board
If the voltage is in range, or there
is obvious damage to a connector,
replace mother board
68
Page 69
Troubleshooting Guide4 Troubleshooting
Printer
SymptomsCause of FailureFailure Isolation and Remedy
Prompt message “Print job
could not be queued” is
issued. No print device is
found.
Status message “Print
device Local 1 (Local 2)
unavailable” is issued.
Printer job is stalled.
Status message “Print
device Remote 1 (Remote
2, Remote 3) unavailable”
is issued. Printer job is
stalled
Status message “Printing on
device Remote 1... (Remote
2, Remote 3)” is issued but
no report is printed
Printouts are not as
expected
Printer is disabled in the Setup
Printers menu
Paper size of printer does not match
paper size of report
Printer not switched on
Printer paper tray empty
Cabling not connected correctly
I/O board defect
Print error on Philips Information
Center
Network Connection to Philips
Information Center not functioning
Print queue on Philips Information
Center is full. Reasons for this may
be:
- Printer is not switched on
- Printer paper tray is empty
Printer paper size is not correctly
configured
Printer resolution is not correctly
configured
Printer color support is configured
to “On” although the printer does
not support color
Printer not compatible
Enable the correct printer in the
Setup Printers menu
Change paper size of the printer
in the Setup Printers menu or
change paper size of the report in
the Setup Reports menu.
Switch on printer power
fill printer paper tray
Check cabling
replace I/O board
Print a test report on the Philips
Information center. If this fails,
refer to Philips Information
Center documentation
Check that the network
connection between the monitor
and the Philips Information
Center is working
Switch on printer power
Fill printer paper tray
Configure the paper size
according to the inserted print
media
Configure the printer resolution
according to the printer
capabilities
Configure the printer color
support to “Off ”
Check specifications
69
Page 70
4 TroubleshootingTroubleshooting Guide
Recorder
SymptomPossible Cau se Corrective Action
System thinks that door is
open when it is not.
System thinks that the
recorder is out of paper
when it is not.
Recorder not
communicating with
System.
Recorder won't run.Recorder interface not
Poor print quality.Printhead dirty.Clean the Printhead.
Paper not feeding
properly.
Module does not lock
into FMS.
Defective door switch.Replace door switch. Exchange module.
Paper-out sensor dirty.Clean paper-out sensor.
Poor connection to the
front-end FMS.
Only one recorder module
may be used with each
monitor.
System not configured
properly.
Too many modules
connected.
working correctly.
Printhead failure.
Paper roll off center.Center paper roll on roller guides.
Dirty roller.
Locking plates defective.Remove and exchange the locking plates.
Unplug the module. Plug it back in and
try it again in a few seconds. (Watch for
the LED to flash.)
Remove one of the recorder modules.
Check the configuration of the connected
monitor.
Check and remove the extra modules.
Unplug the module. Plug it back in and
try it again in a few seconds. (Watch for
the LED to flash.)
Exchange the module.
Clean roller.
MIB / RS232
70
SymptomsCause of FailureFailure Isolation and Remedy
AGM connected to an
RS232 port not
functioning
The MIB/RS232 port is not
configured for AGM
The cable between AGM and
monitor is not connected correctly
or defective
The MIB/RS232 board is in a
wrong slot (slot has been changed
after software configuration or an
additional board has been plugged
in)
The MIB/RS232 board is defective
Check configuration of the MIB/
RS232 ports in configuration
mode
Check cable connection, replace
cable if necessary
Verify correct placement of the I/
O boards
Check board and replace if
necessary
Page 71
Troubleshooting Guide4 Troubleshooting
SymptomsCause of FailureFailure Isolation and Remedy
External device not
receiving data
Detailed Protocol ProblemConsult the Data Export Protocol
The MIB/RS232 port is not
configured for data export
The wrong data export protocol
driver is configured in the monitor
The cable between the external
device and the monitor is not
connected correctly or defective
The external device does not support
the version of the data export
protocol used in the monitor
A terminal concentrator is used in
between the device and the monitor
and a protocol with dynamic speed
negotiation is used
The MIB/RS232 board is in a
wrong slot (slot has been changed
after software configuration or an
additional board has been plugged
in)
The MIB/RS232 board is defective
Check configuration of the MIB/
RS232 ports in configuration
mode
Check the export protocol
required by the attached device
and configure the monitor
accordingly
Check cable and replace if
necessary
Check if the device supports the
version of the data export
protocol. Upgrade device or
monitor if necessary (if matching
versions exist).
Some terminal concentrators do
not support changing the
transmission speed (baud rate)
dynamically. Check if the
connection works without the
concentrator
Verify correct placement of the I/
O boards
Check board and replace if
necessary
document.
Flexible Nurse Call Relay
SymptomsCause of FailureFailure Isolation and Remedy
Connection of cable to monitor or
nurse call system not correct
Nurse Call Relay board is in the
wrong slot.
The Nurse Call Relay bosrd is
defective
Check monitor configuration (see
configuration guide)
Check cable connections
Verify correct placement of the
I/O boards
Replace Nurse Call Relay board
71
Page 72
4 TroubleshootingTroubleshooting Guide
Troubleshooting the ECG OUT
SymptomsCause of FailureFailure Isolation and Remedy
No marker pulse is
displayed on the Monitor
or no ECG- OUT signal to
the Defib
Disconnect the MMS and Defib
cable.
Switch the Monitor off then on again.
Observe the red LED in the ECG
OUT section. (Note that the LED can
only be observed if the housing
bottom is removed).
If the red LED does not switch on for
about 1 second at power on:
Cabling not connected
ECG OUT board defectiveReplace ECG OUT board
ECG OUT Board defective
Main Board defectiveReplace Main Board
ECG OUT board defective
Main Board defectiveReplace main board
Check cabling
If the red LED switches on and
remains on for more than 20 seconds:
Replace ECG OUT Board
Connect Known good Defib, Defib
cable MMS and MMS cable. Check
Marker pulse and ECG OUT signal at
defib again. If there is still no signal:
Replace ECG OUT board
Data Flow Marker In and ECG Wave
The following illustration of the data flow for Marker In and ECG Wave may assist in troubleshooting.
72
Page 73
Troubleshooting Guide4 Troubleshooting
Defib
Cable
Isolat.
ECG_OUT
Board
CPU,Communication
System and Video
MSL
Connector
LAN
Display
Assembly
ECG W a ve
Marker
DEFIB
Marker In
Data Path
Conn.
RS485
Monitor System Board
ECG Wave
Data Path
Status Log
Marker merged to
ECG Wave here
Server
Application
MEASUREMENT
SERVER
Software
Many events that occur during start-up or regular monitoring are logged in the Status Log. The Status
Log can be printed and cleared. Not all entries in the Status Log are errors.
The window title is either Monitor or MeasServ, dependent on which system component’s status
log is currently displayed.
The Status Log window shows logged events which caused a reboot of the system component (monitor
or measurement server).
The first column in the log identifies the event class (“C”: caused a cold start, “H”: caused a hot start,
“N”: no retstart, for information only). Column 3 and 4 identify the event source and event code.
Column 4 counts the number of occurrences of the event. The last column shows the time and date of
the last occurrence of the event.
The following pop-up keys overlay the SmartKeys:
Clear
StatLog
Clear StatLog
This key clears the currently displayed Status Log
Revision
This key switches to the Revision Screen of the currently displayed system component
M8010
This key switches to the Monitor Revision Window
M8048
This key switches to the Flexible Module Server (FMS) Revision Window
M3001
This key switches to the Multi Measurement Server (MMS) Revision Window
If an event occurs repeatedly, contact your Philips Service Representative.
NOTE It is possible, using the support tool, to download the status log and send it to your Philips Service
Representative as a file (for example via e-mail).
RevisionM8010M8048M3001
Troubleshooting with the Support Tool
Using the support tool you can:
• access the full status log which can be saved as a file
•reload software
• identify defective devices
• reset touch screen calibration
For details on how to perform these tasks see the Support Tool User Manual.
74
Page 75
Troubleshooting Guide4 Troubleshooting
Troubleshooting the Individual Measurements or Applications
For problems isolated to an individual parameter or application such as event review, please consult the
Instructions for Use and configuration information.
If the instructions for use did not resolve an individual parameter problem, then another module or
measurement server should be tried.
If you are getting questionable readings for individual measurements you may want to do the
Performance Verification tests in the Testing and Maintenance section.
The performance of the individual applications (event review, arrhythmia, trending) are affected by the
configuration of the monitor. When contacting Philips support you may be asked about the
configuration of the monitor to aid in troubleshooting.
75
Page 76
4 TroubleshootingTroubleshooting Guide
76
Page 77
5Repair and Disassembly
The following section describes the disassembly and reassembly procedures for the monitor and its
components
Tools Required
• Torx screwdriver (size 10)
• Torx screwdriver (size 20)
• ESD mat and wrist strap
• 2 small flat blade screwdrivers
MP90 CMU Disassembly
5
NOTE All screws for the MP90 are the same size Torx. We recommend that you keep them separated and
counted as you disassemble and reassemble the device.
Removing I/O Boards
1Pull off the cable management cover at the rear of the Computer Module Unit (CMU).
77
Page 78
5 Repair and DisassemblyMP90 CMU Disassembly
2Pull off the Power Cord Cover at the front of the CMU and make sure the power cord is
unplugged.
CAUTION Make sure to unplug the AC power cord before removing the I/O boards. Failure to remove the power
cord could cause damage to the CMU as it is still under power.
78
Page 79
MP90 CMU Disassembly5 Repair and Disassembly
Unscrew the four screws at the rear of the CMU and remove the metal I/O board cover.
3
4Use the board removal tool located inside the cable management cover to remove the I/O boards.
Make sure to insert the end with the hook underneath the I/O board.
Board Removal
Hook
Removing the Top Cover
1Flip over the CMU so the bottom faces upwards. With single CPU units it may be helpful to
reinsert the I/O board cover without screws before flipping over the CMU in order to have a more
stable platform.
Tool
79
Page 80
5 Repair and DisassemblyMP90 CMU Disassembly
Remove the four screws next to the plastic feet.
2
3Flip the CMU back over, remove the I/O board cover if still present, and slide off the top cover
towards the rear.
Reassembly Note: Insert the I/O board cover before flipping over the CMU to aid with reinserting
screws.
80
Page 81
MP90 CMU Disassembly5 Repair and Disassembly
Removing the Plastic Feet and/or the Locking Cam
Some mounting options may require the removal of the plastic feet or the locking cam.
1Place the CMU upside down and remove the screws securing each of the feet and the locking cam.
Plastic
Plastic
Feet
Feet
Locking Cam
CAUTION When you remove the feet and/or the locking cam you may find there are standoffs to increase the
height under the CMU. You must refit these standoffs with the plastic feet and the locking cam or use
shorter screws to refit the feet and the locking cam alone. Screws cannot go more than 5 mm inside the
CMU (measured from the exterior of the metal chassis). Inserting screws which are too long into the
CMU will damage the main board.
81
Page 82
5 Repair and DisassemblyMP90 CMU Disassembly
Removing the optional Fans (Dual CPU Versions only)*
1Remove the top cover.
2Disconnect the fans and pull them out from the sheet metal.
Remove Latch
from metal
Fan Connector
* The MP90 dual CPU version may have fans added during installation. The fans are not mandatory
and do not ship with the product.
82
Page 83
MP90 CMU Disassembly5 Repair and Disassembly
Replacing the Second (Independent) Video Board
(Dual CPU Versions only)
1Remove the 5 screws and take out the video board, unplugging it from its connector.
Lift board here to unplug connector
Connector
83
Page 84
5 Repair and DisassemblyMP90 CMU Disassembly
NOTE
If not already separated, you may need to break apart the two pieces of the repaired/exchange board.
Removing the Second CPU/Main Board (Dual CPU Versions only)
1Remove the second independent videoboard as described in Replacing the Second (Independent)
Video Board (Dual CPU Versions only).
2Remove the top three I/O boards.
84
Page 85
MP90 CMU Disassembly5 Repair and Disassembly
Remove the white fan connector and the screw securing the main CPU interconnection board.
3
4Pull off the main CPU innterconnection board in the direction shown.
Interconnection Board Connector
NOTE If you wih to remove the interconnection board entirely, you must also disconnect the interconnection
board connector.
85
Page 86
5 Repair and DisassemblyMP90 CMU Disassembly
Remove the 10 screws and pull the mainboard straight up to remove it. Please note that there are
5
sensitve components near the screws on the bottom of the main board.
Accessing the Main CPU or Primary Video Board (Dual CPU
Versions)
NOTE If the second independent video board or the second CPU board do not require replacement, you do
not need to remove them to access the first CPU and primary video boards.
1Unplug the interconnector board connector, the fan connector and the power switch cable
connector.
Interconnector Board Connec tor
Fan Connector
Power Switch Cable
Connector
86
Page 87
MP90 CMU Disassembly5 Repair and Disassembly
Remove the six screws and take off the upper sheet metal plate. (The top three I/O boards do NOT
2
have to be removed).
87
Page 88
5 Repair and DisassemblyMP90 CMU Disassembly
Replacing the Primary Video Board
1Remove the five screws securing the primary video board.
2Lift up the video board at the back to unplug the connector. Then slide the board backwards to
remove it.
Connector
88
Page 89
MP90 CMU Disassembly5 Repair and Disassembly
Reassembly Note: The flex cable on the primary video board must go up.
Reassembly Note: When replacing the primary video board, the board must slide under the sheet
metal.
89
Page 90
5 Repair and DisassemblyMP90 CMU Disassembly
NOTE
If not already separated, you may need to break apart the two pieces of the repaired/exchange board.
Removing the Main Board
1Remove the video board as described in Replacing the Primary Video Board.
2Remove all I/O boards
90
Page 91
MP90 CMU Disassembly5 Repair and Disassembly
Remove power supply and power switch connectors and the speaker connector.
3
Speaker
Power
Power
Switch
Supply
4Disconnect the I/O board backplane connectors.
91
Page 92
5 Repair and DisassemblyMP90 CMU Disassembly
Remove the two screws securing the backplane (on the metal, NOT the board) and take out the
5
backplane.
92
Page 93
MP90 CMU Disassembly5 Repair and Disassembly
Remove the CPU interconnection board.
6
7Remove the remaining ten screws and pull the main board straight up to remove it. Please note
that there are sensitive components near the screws on the bottom of the main board.
Removing the Power Supply
1Remove the second CPU sheet metal assembly, if present. (See Accessing the Main CPU or Primary
Video Board (Dual CPU Versions)
93
Page 94
5 Repair and DisassemblyMP90 CMU Disassembly
Pull out the power supply and power switch connectors.
2
3Flip over the CMU, support it with the cable management cover and take off the four screws.
94
Cable Manageme nt Cover
Page 95
MP90 CMU Disassembly5 Repair and Disassembly
Pull the CMU base off of the power supply.
4
Reassembly Note: When reassembling, ensure that all cables are led through the dedicated holes and
not pinched underneath the metal.
Removing the Speaker
1Remove the power supply.
2Disconnect the speaker cable, unscrew the three black screws using a flat blade screwdriver, and
then turn and pull out the speaker.
Screws
Removing the Power On/Off Switch
1Remove the power supply.
95
Page 96
5 Repair and DisassemblyMP90 CMU Disassembly
Remove the screw securing the power switch and remove the switch.
2
96
Page 97
Flexible Module Server (FMS) Disassembly5 Repair and Disassembly
Flexible Module Server (FMS) Disassembly
Removing the Handle and the Measurement Server Mount
Please note that any combination of handles and mounts is possible.
1Remove the two screws on the bottom with a T20 screwdriver.
2Slide the handle up and pull it out.
3Remove the MMS if connected. Slide the MMS mount up and remove it by pulling directly
perpendicular to the FMS.
97
Page 98
5 Repair and DisassemblyFlexible Module Server (FMS) Disassembly
NOTE
There is a connector located on the MMS mount. If you tilt the MMS mount as you remove it, you
may damage the connector.
4Remove the connector housings on each side of the FMS by compressing the cover slightly using
two screwdrivers.
98
Page 99
Flexible Module Server (FMS) Disassembly5 Repair and Disassembly
Remove the two white pins on each side with a small screwdriver.
5
6Tak e of f th e r e ar h ou s in g .
Rear
Front
7Remove the four screws on the CPU board, pull it gently off the mother board, unplugging the
connector at the same time.
99
Page 100
5 Repair and DisassemblyFlexible Module Server (FMS) Disassembly
Remove the four remaining screws on the mainboard.
8
9Pull off side connector brackets by pulling them gently away from the housing on each side and
lifting carefully.
10 Lift up and pull on the tabs to remove the connector holders.
Reassembly Note: The connector holders are side specific.
11 Unsnap the lightpipe.
100
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.