Philips IntelliVue MP40, IntelliVue MP50 User manual

IntelliVue MP40/50
Service Guide
IntelliVue Patient Monitor
MP40/50
Patient Monitoring
Part Number M8000-9361D 4512 610 14411
*M8000-9361D*
1Table of Contents
1 Introduction 13
Who Should Use This Guide 13 How to Use This Guide 13 Abbreviations 13 Responsibility of the Manufacturer 14 Passwords 14 Warnings and Cautions 14
2 Theory of Operation 17
Monitor Theory of Operation 17
System Boundaries 17 Hardware Building Blocks 19
IntelliVue MP40 19 IntelliVue MP50 20 Optional Hardware 21 Compatible Devices 22 Power Supply 25 CPU Boards 25 System Interface and I/O Boards 26
Data Flow 27
Data Acquisition 27 Data Provider System Service 27 Persistent Data Storage System Service 28 Display and User Interface Service 28 Data Output 28 Monitor Applications 28 Internal LAN (Measurement Server Link) 28 Philips Clinical Network 29
How does the Support Tool Work with the Monitor 30 Monitor Software Block Diagram 30 Block Diagram Legend 32
3 Testing and Maintenance 37
Concepts 37 Test Reporting 37 Frequency 38 Tests When Performing... 39
Installation 39 Repair 39
Preventive Maintenance 39 Performance Verification 40 Upgrades 40
Preventive Maintenance Procedures 40
NBP PerformanceTests 40
NBP Accuracy Test 41 NBP Leakage Test 42 NBP Linearity Test 42 Valve Test 42
Microstream CO2 Performance Test 43
Barometric Pressure Check and Calibration 43 Leakage Check 44 Pump Check 45 Flow Rate Check and Calibration 45 Noise Check 45 CO
Gas Measurement Calibration Check 45
2
Calibration Verification 46 Reset Time Counters 46
CO2Pump / CO2 Scrubber Replacement 47
M3014A Capnography Extension Performance Tests 47
Mainstream CO2 Accuracy Check 48
Procedure 48
Sidestream CO2 Accuracy Check 48
Procedure 48
Sidestream CO2 Flow Check 49
Spirometry Performance Tests 49
Equipment Required 49 Flow Test 49 Leakage Test 50 Barometer Check 50
Te s t s 50
Visual Test 50 Power On Test 50 Temperature Performance Test 51 ECG/Resp Performance Test 51
ECG Performance 51 Respiration Performance 51
Invasive Pressure Performance Test 51 SpO2 Performance Test 52
Measurement Validation 52
Cardiac Output Performance Test 52
Service Tool Procedure, Version 1 52 Service Tool Procedure, Version 2 53
BIS Performance Test 53
PIC/DSC Test 53
Nurse Call Relay Performance Test 53
Phone Jack Type Connector Test (Traditional Nurse Call) 54 Multi-Port Nurse Call Connector Test (Flexible Nurse Call) 54
ECG Sync Performance Test 55 Docking Station Performance Test 56 VueLink Tests using VueLink Test Module 56
Tes t P r o c e d u r e 56
IIT Communication Test 57
IntelliVue 802.11 Bedside Adapter Communication Test 58
Safety Testing 58
Warnings, Cautions, and Safety Precautions 58 Safety Test Procedures 59
Battery Handling, Maintenance and Good Practices 62
About the Battery 62 Checking the Battery Status 62 Battery Status on the Main Screen 64 Battery Status Window 66
Viewing Individual Battery Status 66 Documenting Battery Status 67 Battery Implications 67
Conditioning a Battery 67
What is Battery Conditioning? 67 Why is Battery Conditioning Necessary? 67 When Should Battery Conditioning be Performed? 67 What Causes the Conditioning Message on the Monitor? 68
Conditioning Batteries 68
Battery Conditioning in the Monitor 68 Battery Conditioning with an External Charger 69
Touchscreen Calibration 69 Disabling/Enabling Touch Operation 71 Printer Test Report 71
4 Troubleshooting 73
Introduction 73 How To Use This Section 73 Who Should Perform Repairs 73 Replacement Level Supported 73 Software Revision Check 74 Software Compatibility Matrix 74 Obtaining Replacement Parts 75 Troubleshooting Guide 75
Checks for Obvious Problems 75 Checks Before Opening the Instrument 75
Checks with the Instrument switched Off 75 Checks with the Instrument Switched On, AC connected, without battery 75 Checks with the Instrument switched On, AC connected, with battery 76
Checks with the Instrument switched On, AC not connected, with battery 76 Initial Instrument Boot Phase 76
Troubleshooting Tables 78
How to use the Troubleshooting tables 78 Boot Phase Failures 79 Integrated Display is blank 81 Integrated Touch Display not functioning 81 External Display is blank (Slave Display) 82 External Touch Display not functioning 83 General Monitor INOP Messages 83 Remote Alarm Device 84 Remote Extension Device 85 Navigation Point 85 Keyboard/Mouse not functioning 86 Battery related problems 86 Bedside Network Status Icons 88 Network related problems 88 Wireless Ethernet Adapter (Proxim) 89 IIT-related Problems 90 IntelliVue 802.11 Bedside Adapter Problems 91 Multi-Measurement Server 92 MSL-related problems 92 Alarm Issues 94 Alarm Lamps 94 Alarm Tones 94 Alarm Behavior 95 Individual Parameter INOPS 95 Integrated 4-Slot Rack 96 Printer 96 Recorder 97 MIB / RS232 98 Flexible Nurse Call Relay 99 Basic Nurse Call Relay 99 Troubleshooting the ECG OUT/Alarm LED 99 Docking Station Problems 100 Data Flow Marker In and ECG Wave 101
Status Log 102 List of Error Codes 103 Troubleshooting with the Support Tool 104 Troubleshooting the Individual Measurements or Applications 104
5 Repair and Disassembly 105
Tools Required 105 Removing the I/O Boards 106 Removing the Interface Board 108 Separating the Front and Back Half of the monitor 109
Exchanging the Backlight Tubes 111 Removing the Flex Panel Adapter 113 Removing the Touchscreen 116 Removing the Flat Panel 119 Removing the Backlight Inverter Board 122 Removing the Silicon Mat for the Power Switch and the LEDs 123 Removing the Navigation Point Assembly 124 Removing the ECG Out/Alarm LED Board 126 Removing the Handle 127 Removing the Main Board 129 Removing the MSL Board 132 Removing the Internal Module Rack 133 Removing the Power Supply Assembly 133 Reassembling the Power Supply Assembly 135 Removing the Loudspeaker 139 Changing the Serial Number Plate 141 Exchanging the battery door 142 Plug-in Modules and MMS Extensions 144
Plug-In Module Disassembly 144
tcpO2/tcpCO2 Calibration Chamber Kit 146 Recorder Module Paper 147
Disassembly Procedures for the Measurement Server Extension (M3015A) 148
Removing the Front Cover 148 Removing the Extension Bottom Cover 148 Removing the CO2 Scrubber 150 Removing the Pump 150
Refit Procedures for the Measurement Server Extension 151
Refitting the CO2 Scrubber 151 Refitting the Pump 151 Refitting the Extension Bottom Cover 152 Refitting the Front Cover 152 General Reassembly/Refitting Comments 152 Following Reassembly 152
Smart Battery Charger LG1480 (M8043A) 153
Cleaning the Air Filter Mats 153 Replacing the Fan 154
IntelliVue Instrument Telemetry (IIT) 156
Exchanging the PCA board or the Flex Cable 156 Exchanging the IIT Module 157
Docking Station 158
Exchanging the Main Board 159 Exchanging the Flex Cable 162
6 Parts 165
MP40/MP50 Parts 166 Multi-Measurement Server Parts 169
MMS Part Numbers - Front Bezel for M3001#A01 170 MMS Part Numbers - Front Bezel for M3001#A02 171 MMS Exchange Part Numbers 172
M3001A #A01 Philips FAST SpO2 MMS Exchange Numbers 173 M3001A #A02 Nellcor OxiMAX-compatible MMS Exchange Numbers 174
MMS Part Numbers - Label Kits 175 MMS Part Numbers - Release Mechanisms 175
Measurement Server Extension Parts (M3012A, M3014A, M3015A and M3016A)175
Exchange Parts List 181
Plug-in Modules Part Numbers 183
Part Number Table 183
Exchange Modules, Table 1 184 Exchange Modules, Table 2 185
Plug-In Modules Replaceable Parts 187
Single-Width Plug-In Module 187 Double-Width Plug-In Module 187 Plug-in Module Replaceable Parts 188 Plug-In Module Language Specific Front Housing Kits (incl. Silicone Buttons, Frames & Bezels),
Table 1 188 Plug-In Module Language Specific Front Housing Kits (incl. Silicone Buttons, Frames & Bezels),
Table 2 189 Plug-In Module Specific Bezels 190
BIS Solution Replaceable Parts 190
BIS Solution Components 191
BISx Solution Replacable Parts 191
BISx Solution Components 192
tcpO2/tcpCO2 Module Accessories 192
Smart Battery Charger Part Numbers 193 IntelliVue Instrument Telemetry Part Numbers 193 IntelliVue 802.11 Bedside Adapter Part Numbers* 194 Docking Station Part Numbers 194 External Display Part Numbers 195 Remote Alarm Device Part Numbers 197 Remote Extension Device Part Numbers 198
7 Installation Instructions 199
Unpacking the Equipment 199 Initial Inspection 200
Mechanical Inspection 200 Electrical Inspection 200 Claims For Damage and Repackaging 200
Claims for Damage 200 Repackaging for Shipment or Storage 200
Installing the Monitor (M8003A or M8004A) 200
Mounting Instructions 201
Assembling Mounts 201
Connections 201 Installing Interface Boards 203 Connection of Devices via the MIB/RS232 Interface 205
Installing the Docking Station 205 Installing Remote Devices 206
Mounting the 15” Remote Display (M8031A) 206
Connections 206
Mounting the 15” Remote Display (M8031B) 206
Connections 207
Mounting the 17” Remote Display (M8033A/B/C) 207
Connections 207 Video Cable Wiring Schematics 209
Multi-Measurement Server 209
Attaching the MMS to a Mount 209 Detaching the Measurement Server from a Mount 210 Positioning the Measurement Server on a Clamp Mount 210 Mounting the BIS Engine to the Monitor 211 Mounting the Wireless Ethernet Adapter to the Monitor 212 Connections 215 MSL Cable Termination 215
PS/2 Keyboard/Mouse 217
Philips Clinical Network (Wired) 218 Philips Clinical Network (Wireless) 218 Nurse Call Relay 218
Connections 218
ECG Out Functionality 219
Connections 219
Configuration Tasks 219
Setting Altitude and Line Frequency 220 Configuring the Equipment Label 220 Configuring the printer 220
8 Site Preparation 221
Introduction 221
Site Planning 221 Roles & Responsibilities 221
Site Preparation Responsibilities 221 Procedures for Local Staff 222 Procedures for Philips Personnel 223
Monitor M8003A and M8004A Site Requirements 223
Space Requirements 223 Environmental Requirements 223
Te m p e r at u r e 2 2 3 Humidity 223 Altitude 224 Battery Storage 224
Electrical and Safety Requirements (Customer or Philips)224
Safety Requirements224 Electrical Requirements224
Remote Device Site Requirements224
Connecting Non-Medical Devices225 Multi-Measurement Server M3001A 226
Space Requirements Multi-Measurement Server M3001A226 Environmental Requirements Multi-Measurement Server M3001A226 Cabling Options and Conduit Size Requirements226 Mounting227
Remote Displays (M8031A)227
Space Requirements227 Environmental Requirements227 Electrical and Safety Requirements228
Remote Displays (M8031B)228
Space Requirements228 Environmental Requirements228 Electrical and Safety Requirements228
Remote Displays - M8033A229
Space Requirements229 Environmental Requirements229 Electrical and Safety Requirements229
Remote Displays - M8033B229
Space Requirements229 Environmental Requirements229 Electrical and Safety Requirements230
Remote Displays - M8033C230
Space Requirements230 Environmental Requirements230 Electrical and Safety Requirements230 Cabling Options and Conduit Size Requirements231 Touch Cable231
Remote Alarm Devices232
Space Requirements232 Mounting232 Cabling Options and Conduit Size Requirements232
Remote Extension Device232
Space Requirements232 Mounting233 Cabling Options and Conduit Size Requirements233 Input Devices233
Local Printer234
Philips Medical LAN235 RS232/MIB/LAN Interface235 Nurse Call Relay Interface236
10
ECG Out Interface236
9 Gas Analyzers 237
10 Philips 15210B Calibration Unit 239
Unpacking the Instrument239
Initial Inspection239
Claims for Damage240 Repacking for Shipment or Storage240
Instrument Identification240 Specification240 Operating Environment240 Operating Information240 Fitting the Gas Cylinders241 Storage of Gas Cylinders241 Disposal of Used Gas Cylinders241
Routine Maintenance241
Changing the Gas Cylinders241 Care and Cleaning241
Theory of Operation242 Gas Flow Performance Check242
Tes t P ro c e du re 242
Action if outside specification243
Disassembly244 Parts List246
11 IntelliVue Product Structure 247
Upgrade Options249
11
12
This Service Guide contains technical details for the IntelliVue MP40 and MP50 Patient Monitor, the measurement modules, the Multi-Measurement Server (MMS), and the Measurement Server Extensions.
This guide provides a technical foundation to support effective troubleshooting and repair. It is not a comprehensive, in-depth explanation of the product architecture or technical implementation. It offers enough information on the functions and operations of the monitoring systems so that engineers who repair them are better able to understand how they work.
It covers the physiological measurements that the products provide, the Measurement Server that acquires those measurements, and the monitoring system that displays them.

Who Should Use This Guide

1

1Introduction

This guide is for biomedical engineers or technicians responsible for troubleshooting, repairing, and maintaining Philips’ patient monitoring systems.

How to Use This Guide

This guide is divided into eight sections. Navigate through the table of contents at the left of the screen to select the desired topic. Links to other relevant sections are also provided within the individual topics. In addition, scrolling through the topics with the page up and page down keys is also possible.

Abbreviations

Abbreviations used throughout this guide are:
Name Abbreviation
IntelliVue MP40/MP50 Patient Monitor the monitor
Multi-Measurement Server MMS
Measurement Server Link MSL
Medical Information Bus MIB
Anesthetic Gas Module AGM
13
1 Introduction Responsibility of the Manufacturer

Responsibility of the Manufacturer

Philips only considers itself responsible for any effects on safety, EMC, reliability and performance of the equipment if:
• assembly operations, extensions, re-adjustments, modifications or repairs are carried out by persons authorized by Philips, and
• the electrical installation of the relevant room complies with national standards, and
• the instrument is used in accordance with the instructions for use.
To ensure safety and EMC, use only those Philips parts and accessories specified for use with the monitor. If non-Philips parts are used, Philips is not liable for any damage that these parts may cause to the equipment.
This document contains proprietary information which is protected by copyright. All Rights Reserved. Reproduction, adaptation, or translation without prior written permission is prohibited, except as allowed under the copyright laws.
Philips Medizin Systeme Böblingen GmbH
Hewlett-Packard Str. 2
71034 Böblingen, Germany
The information contained in this document is subject to change without notice.
Philips makes no warranty of any kind with regard to this material, including, but not limited to, the implied warranties or merchantability and fitness for a particular purpose.
Philips shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.

Passwords

In order to access different modes within the monitor a password may be required. The passwords are listed below.
Monitoring Mode: No password required
Configuration Mode: 71034
Demo Mode: 14432
Service Mode: 1345
Consult the configuration guide before making any changes to the monitor configuration.

Warnings and Cautions

In this guide:
14
•A warning alerts you to a potential serious outcome, adverse event or safety hazard. Failure to observe a warning may result in death or serious injury to the user or patient.
•A caution alerts you where special care is necessary for the safe and effective use of the product. Failure to observe a caution may result in minor or moderate personal injury or damage to the product or other property, and possibly in a remote risk of more serious injury.
Warnings and Cautions 1 Introduction
NOTE
When an IntelliVue MP40/MP50, software revision B.0 with battery option installed is used together with an IntelliVue Infromation center D.01 or E.0 and the monitor issues battery-related INOPs, these INOPS are displayed as “UNKNOWN” on the IntelliVue Information Center. Upgrade the Information Center text catalog to E.01 if using an MP40/MP50 monitor with the Information Center.
15
1 Introduction Warnings and Cautions
16

2Theory of Operation

Monitor Theory of Operation

The IntelliVue MP40/MP50 Patient Monitor:
• displays real-time data
• controls the attached measurement server
• alarms in the case of patient or equipment problems
• offers limited data storage and retrieval (trending)
• interfaces to the Philips Clinical Network and other equipment
A monitor with just a single integrated measurement server can be connected to additional building blocks to form a monitoring system with a large number of measurements, additional interface capabilities and one slave display. These elements cooperate as one single integrated real-time measurement system.
2

System Boundaries

The following diagram discusses specific boundaries within the overall system with respect to their openness and real-time requirements:
17
2 Theory of Operation Monitor Theory of Operation
Philips Clinical Network
Measurement LAN
combines components of one patient monitor; real time requirements across all interconnected elements
Philips Clinical Network (wired LAN)
connects multiple patient monitors, information centers, application servers; closed system, only Philips qualified products (tested and with regulatory approval) are connected, Philips is responsible for guaranteed real-time functionality and performance
Philips Clinical Network (wireless)
like Philips Clinical Network (wired) LAN, however due to current wireless technologies available it has reduced bandwidth, longer latencies, reduced functionality
Hospital LAN, Internet
Standard Network, not under Philips control, no guaranteed service, no real-time requirements
18
Monitor Theory of Operation 2 Theory of Operation

Hardware Building Blocks

The following hardware building blocks make up the monitoring system:
IntelliVue MP40
The MP40 monitor:
• integrates the display and processing unit into a single package
• uses a 12.1” TFT SVGA color display
• uses the Philips Navigation Point as primary input device; computer devices such as mice, trackball, and keyboard can be added optionally
• has an optional 4-slot rack
• supports the MMS and MMS extensions.
19
2 Theory of Operation Monitor Theory of Operation
Building Blocks:
IntelliVue MP50
The MP50 monitor:
• integrates the display and processing unit into a single package
Power Supply
Backlight Inv. Board
LCD
Assembly
Panel Adapter
Flex
Battery Board
Main Board
Rack Board
ECG Out
4 Modules
I/O
Board
System
Interface
Board
MSL
I/F
Dual MIB/ RS232 Interface Parallel Printer Interface Flexible Nurse Call Relay I/F Dual PS2 Interface Remote Device Interface
LAN (wireless) LAN (wired) Bas. Alarm Relays Slave Video out
Docking interface
MMS CPU
• uses a 12.1” TFT XGA color display
• uses the Philips Touchscreen and Philips Navigation Point as primary input devices. Computer devices such as mice, trackball, and keyboard can be added optionally.
• has an optional 4-slot module rack
• supports the MMS and MMS extensions
20
Monitor Theory of Operation 2 Theory of Operation
Building Blocks:
Assembly
Inv. Board
Optional Hardware
An integrated 4-Slot module rack and a battery board can be ordered optionally. One slot is provided for one of two available system interface boards. If the monitor is ordered with the wireless LAN option a wireless transmitter is required. For further details regarding the wireless network please refer to the M3185A Philips Clinical Network documentation.
Power Supply
LCD
Touch
Panel
Backlight
Touch
Controller
Panel Adapter
Battery Board
Flex
Rack Board
Main Board
ECG Out
4 Modules
I/O
Board
System
Interface
Board
MSL
I/F
Dual MIB/ RS232 Interface Parallel Printer Interface Flexible Nurse Call Relay I/F Dual PS2 Interface Remote Device Interface
LAN (wireless) LAN (wired) Bas. Alarm Relays Slave Video out
Docking interface
MMS CPU
Powersupply Heatsink
Figure 1 MP40/50 Rear
MMServer Mount
4-Slot Rack
Battery Compartment
Powerplug
21
2 Theory of Operation Monitor Theory of Operation
IIT Adapter
IntelliVue 802.11 Bedside Adapter
Figure 2 MP40/50 Rear with internal IntelliVue 802.11 Bedside Adapter (left) and
with external IIT Adapter (US only) (right)
NOTE The IntelliVue 802.11 Bedside Adapter and the IIT Adapter require a monitor with Software Rel. C.0
or higher.
Compatible Devices
22
Figure 3 M8045A Docking Station
Monitor Theory of Operation 2 Theory of Operation
Figure 4 M3001A Multi-Measurement Server (MMS)
Figure 5 M3012A, M3014A, M3015A, M3016A MMS Extensions
23
2 Theory of Operation Monitor Theory of Operation
Figure 6 Parameter Modules
List of supported modules:
• M1006B Invasive Blood Pressure Module
• M1029A Temperature Module
• M1012A Cardiac Output / Continuous Cardiac Output Module
• M1018A Transcutaneous Gas Module
• M1020B SpO
Module
2
• M1027A EEG Module
• M1034A BIS Module
• M1116B Thermal Array Recorder Module
• M1032A VueLink Device Interface Module
24
Monitor Theory of Operation 2 Theory of Operation
Power Supply
CPU Boards
Figure 7 Power Supply Architecture
The AC/DC converter transforms the AC power coming from the power plug into 14 V/80W DC source and isolates the monitoring system from the AC power mains.The 14V is distributed via power bus and supplies power to all the components of the system: The 48V DC power needed for the MMS and measurement server extension is created by an isolating DC/DC converter. The power needed for the backlights is converted to 12V DC by the backlight DC/DC converter. The CPU and the non­isolated I/O boards are supplied with 3.3 V and 5 V DC power. Isolated interface boards require a power of 10V AC. The remote HIF board and the LEDs are supplied with 12V DC power.
The CPU boards have an MPC852/50 MHz processor in the patient monitor and an MPC860/ 50MHz in the MMS that provides a number of on-chip, configurable interfaces. An array of fast UARTS with configurable protocol options are implemented in an ASIC (along with other system functions such as independent watchdogs etc.), providing interfacing capabilities to measurement modules and System Interface and I/O boards. The serial interfaces can easily be electrically isolated. The main board contains additional video hardware.
25
2 Theory of Operation Monitor Theory of Operation
IntelliVue Patient Monitor Multi-Measurement Server
CPU Video
The CPUs provide two LAN interfaces to interconnect CPUs (via the MSL) and to connect to the Philips Clinical Network.
The CPU capabilities are identical. Different loading options are coded on serial EEPROMs to support the automatic configuration of the operating system at boot time.
System Interface and I/O Boards
Interfaces to the monitor are implemented via I/O boards. The location of these boards is restricted by general rules. The I/O slot designations diagram and the I/O matrix which outline the I/O board placement rules can be found in the Installation Instructions section.
The following is a list of Interface (I/O) boards which may be present in your monitor, depending on your purchased configuration:
System Interface boards:
Bank of I/Os
Interfaces
CPU
Bank of I/Os
Measurement
Acquisition
•MSL
• Video for slave display
• Philips Clinical Network (LAN wired or wireless)
• Basic Alarm Relay (Nurse Call)
•Docking Interface
I/O boards:
•PS/2
• MIB/RS232
•Flexible Nurse Call
•Parallel printer
• Remote devices (Remote Alarm Device, Remote Extension Device)
• IntelliVue 802.11 Bedside Adapter
The specifications for the above listed interfaces can be found in the technical data sheet for the monitor and in the Installation and Specifications chapter of the Instructions for Use.
26
Monitor Theory of Operation 2 Theory of Operation

Data Flow

The following diagram shows how data is passed through the monitoring system. The individual stages of data flow are explained below.
Display
and User
Interface
Data
Acquisition
Data
Provider Service
Applications
Data Acquisition
Monitoring data (for example patient measurement data in the form of waves, numerics and alerts) is acquired from a variety of sources:
•Measurement Server
The Measurement Server connected to the internal LAN converts patient signals to digital data and applies measurement algorithms to analyze the signals.
• External measurement devices
Data can be also acquired from devices connected to interface boards of the monitor. Software modules dedicated to such specific devices convert the data received from an external device to the format used internally. This applies to parameter modules and the Anesthetic Gas Module.
• Server systems on the Philips Clinical Network
To enable networked applications such as the other bed overview, data can be acquired from server systems attached to the Philips Clinical Network, for example a Philips Information Center
Persistent
Data
Storage
Data
Output
Data Provider System Service
All data that is acquired from measurement servers or external measurement devices is temporarily stored by a dedicated data provider system service. All monitor applications use this central service to access the data in a consistent and synchronized way rather than talking to the interfaces directly.
This service makes the applications independent of the actual type of data acquisition device.
The amount of data stored in the data provider system service varies for the different data types. For example several seconds of wave forms and the full set of current numerical values are temorarily stored in RAM.
27
2 Theory of Operation Monitor Theory of Operation
Persistent Data Storage System Service
Some applications require storage of data over longer periods of time. They can use the persistent data storage system service. Dependent on the application requirements, this service can store data either in battery backed-up (buffered) memory or in flash memory. The buffered memory will lose its contents if the monitor is without power (not connected to mains) for an extended period of time. The flash memory does not lose its contents.
The trend application for example stores vital signs data in a combination of flash memory and buffered memory, while the system configuration information (profiles) is kept purely in flash memory.
Display and User Interface Service
Applications can use high level commands to display monitoring data or status and command windows on the internal LCD panel. These commands are interpreted by the display manager application. This application controls the dedicated video hardware which includes video memory and a special ASIC.
User input is acquired from a variety of input devices, for example the Navigation Point, the touchscreen or other standard input devices (keyboard, mouse) which may be attached to I/O boards. The system software makes sure that the user input is directed to the application which has the operating focus.
Data Output
The monitoring system is very flexible and customizable regarding its data output devices. Built-in devices (for example LAN, alarm lamps, speaker, video) provide the basic output capabilities.
These capabilities can be enhanced by adding additional I/O boards, as required in the specific end­user setup. The additional I/O boards typically provide data to externally attached devices, for example to printers, RS232 based data collection devices, nurse call systems etc.
The monitor can identify I/O boards by means of a serial EEPROM device that stores type and version information. The operating system detects the I/O boards and automatically connects them with the associated (interface driver) application. For some multi-purpose cards it is necessary to configure the card for a particular purpose first (for example the dual MIB/RS232 card can support external touch display (only slave display), data import, data export).
Monitor Applications
The monitor applications provide additional system functionality over the basic measurement and monitoring capabilities. This includes for example trending, report generating, event storage or derived measurements.
In general, the monitor applications use the data provider system service to access the measurement data. Application interfaces to the other system services allow the application to visualize data, to store data over extended periods of time or to output data to other devices.
Internal LAN (Measurement Server Link)
All components of the monitoring system (including measurement servers and CPUs in the monitor) communicate using an IEEE802.3/ Ethernet LAN in the Measurement Server Link (MSL). This network is used to distribute data between the components, for example:
28
• Digitized patient signals including wave data, numerical data and status information (typically from the measurement server to a display unit)
Monitor Theory of Operation 2 Theory of Operation
• Control data representing user interactions (typically from the display unit to a measurement server)
• Shared data structures, for example representing patient demographical data and global configuration items
The internal LAN allows plug and play configuration of the monitoring system. The system automatically detects plugging or unplugging of measurement servers and configures the system accordingly.
The components on the internal LAN are time-synchronized to keep signal data consistent in the system. Dedicated hardware support for synchronization eliminates any latency of the network driver software.
The integrated LAN provides deterministic bandwidth allocation/reservation mechanisms so that the real-time characteristic of signal data and control data exchange is guaranteed. This applies to the data flow from the measurement server to the monitor (for example measurement signal data) and the data flow from the monitor to a measurement server (for example to feed data to a recorder module).
Integrated communication hubs in the monitor allow flexible cabling options (star topology, daisy chaining of servers).
Philips Clinical Network
The monitoring system may be connected to the Philips Clinical Network, for example to provide central monitoring capabilities or other network services. This connection may be through a normal wired connection or through a wireless connection.
The monitor supports the connection of an external wireless adapter or an internal wireless adapter (#J35). Switching between wired and wireless networks is automatically triggered by the plugging or unplugging of the network cable.
The Philips Clinical Network protocols function very similarly to the protocols used on the internal LAN.
After configuration, the monitoring system sends the digitized patient signals including wave data, numerical data and status information onto the network. Control data representing user interactions can be exchanged between the monitoring system and a central station bi-directionally.
MDSE Internal LAN
Additional protocols are supported for networked applications, for example for the other bed overview function, which allows viewing of monitoring data from other patients on the network.
29
2 Theory of Operation Monitor Theory of Operation
For plug and play operation, the monitoring system uses the standard BootP protocol to automatically acquire a network address.

How does the Support Tool Work with the Monitor

The support tool is a Windows application typically installed on the laptop of a customer engineer or a biomedical engineer working in the customer’s own service department.
The purpose of the support tool is to upgrade, configure and diagnose all monitoring components (modules, measurement servers, and monitors) in the system over the network.
The service protocol developed for this purpose uses a raw access to the devices without the need for IP addresses etc. over a standard customer network installation, so that even defective devices can be upgraded as long as the few kBytes of initial boot code are working. The boot code itself can also be upgraded using the same protocol.
The tool allows access to internal service information and to serial numbers. It can be remote­controlled, for example via a dial-up connection from a response center, provided the proper infrastructure is in place.
For details see the Instructions for Use for the Support Tool.

Monitor Software Block Diagram

Figure 8 shows the functional block diagram for the monitoring system. A legend explaining terms and diagram elements follows. The information below varies depending on the purchased monitor options.
30
Loading...
+ 222 hidden pages