Part Number 4535 641 80621
Printed in Germany 10/09
*4535 641 80621*
1Table of Contents
1 Introduction7
Who Should Use This Guide7
How to Use This Guide7
Abbreviations7
Responsibility of the Manufacturer7
Passwords8
Warnings and Cautions8
2 Theory of Operation9
Monitor Theory of Operation9
3 Testing and Maintenance25
Introduction25
Terminology and Definitions25
Recommended Frequency26
When to Perform Tests27
Testing Sequence30
Visual Inspection30
Safety Tests31
System Test57
Preventive Maintenance Procedures68
Performance Assurance Tests69
Reporting of Test Results85
Other Regular Tests89
Touchscreen Calibration89
Disabling/Enabling Touch Operation90
Printer Test Report90
Battery Handling, Maintenance and Good Practices91
After Installation, Testing or Repair97
4 Troubleshooting99
Introduction99
How To Use This Section99
Who Should Perform Repairs99
Replacement Level Supported99
Software Revision Check100
Obtaining Replacement Parts101
Troubleshooting Guide101
3
5 Repair and Disassembly133
Who Should Perform Repairs133
Tools required134
Removing the Handle or Bedhanger134
Removing the Predictive Temperature Assembly135
Disassembling the Predictive Temperature Assembly137
Separating the Front and Back of the Monitor139
Removing the Recorder Slot Cover143
Removing the Internal Quick Mount Solution144
Removing the Short Range Radio (SRR) Interface144
Removing the IntelliVue 802.11 Bedside Adapter Antenna or IIT Antenna (not for MP5T)146
Removing the IntelliVue 802.11 Bedside Adapter (WLAN) (not for MP5T)147
Removing the IntelliVue Instrument Telemetry (IIT) Module (not for MP5T)150
Removing the IntelliVue 802.11 Bedside Adapter/IIT Holder (not for MP5T)151
Removing the MSL Board (not for MP5T)152
Removing the Backlight Inverter Board153
Removing the Power Supply155
Removing the System Interface Board156
Removing the Recorder Board157
Removing the Microstream CO2 Assembly159
Removing the Measurement Block165
Removing the NBP Pump169
Removing the Main Board171
Removing the Touch Assembly174
Removing the Loudspeaker176
Removing the Power Button and LED Assembly176
Removing the LCD Panel178
Exchanging the Backlight179
Modifying the Nurse Call Relay179
6 Parts183
MP5/MP5T Parts183
External Display Part Numbers (not for MP5T)196
7 Installation Instructions199
Out-Of-Hospital Transport - Standards Compliance199
Electromagnetic Emissions201
Electromagnetic Interference (SRR)201
Installation Checklist202
Unpacking and Checking the Shipment202
Installing the Predictive Temperature Probe204
Mounting the Monitor205
Connecting the Monitor to AC Mains210
Checking Out The Monitor210
Loading Paper212
Configuration Tasks213
4
Handing Over the Monitor216
Installing Remote Devices (not for MP5T)216
Philips Clinical Network (Wired) (not for MP5T)218
Philips Clinical Network (Wireless) (not for MP5T)218
Philips IntelliVue Information Center219
IntelliVue Instrument Telemetry (IIT)(not for MP5T)219
Short Range Radio220
Connecting the MP5 to a Host Monitor (not for MP5T)222
Nurse Call Relay (not for MP5T)223
ECG Out Functionality (ECG Sync) (not for MP5T)223
ECG Sync Pulse (not for MP5T)224
8 Site Preparation225
Introduction225
Monitor Site Requirements227
Remote Device Site Requirements229
Philips Medical LAN232
RS232/MIB Interface233
Telemetry Device (Patient Worn Device) cables233
Nurse Call Relay Interface234
ECG Out Interface234
9 Gas Analyzers (not for MP5T)235
10 MP5 Product Structure237
Upgrades242
11 Default Settings Appendix245
Country-Specific Default Settings245
Index255
5
6
This Service Guide contains technical details for the IntelliVue MP5/MP5T Patient Monitor
This guide provides a technical foundation to support effective troubleshooting and repair. It is not a
comprehensive, in-depth explanation of the product architecture or technical implementation. It offers
enough information on the functions and operations of the monitoring system so that engineers who
repair them are better able to understand how it works.
Who Should Use This Guide
This guide is for biomedical engineers or technicians responsible for installing, troubleshooting,
repairing, and maintaining Philips’ patient monitoring systems.
1
1Introduction
How to Use This Guide
This guide is divided into eight sections. Navigate through the table of contents at the left of the screen
to select the desired topic. Links to other relevant sections are also provided within the individual
topics. In addition, scrolling through the topics with the page up and page down keys is also possible.
Abbreviations
Abbreviations used throughout this guide are:
NameAbbreviation
IntelliVue MP5/MP5T Patient Monitorthe monitor
Medical Information BusMIB
Responsibility of the Manufacturer
Philips only considers itself responsible for any effects on safety, EMC, reliability and performance of
the equipment if:
•assembly operations, extensions, re-adjustments, modifications or repairs are carried out by
persons authorized by Philips, and
7
1Introduction
•the electrical installation of the relevant room complies with national standards, and
•the instrument is used in accordance with the instructions for use.
To ensure safety and EMC, use only those Philips parts and accessories specified for use with the
monitor. If non-Philips parts are used, Philips is not liable for any damage that these parts may cause to
the equipment.
This document contains proprietary information which is protected by copyright. All Rights Reserved.
Reproduction, adaptation, or translation without prior written permission is prohibited, except as
allowed under the copyright laws.
Philips Medizin Systeme Böblingen GmbH
Hewlett-Packard Str. 2
71034 Böblingen, Germany
The information contained in this document is subject to change without notice.
Philips makes no warranty of any kind with regard to this material, including, but not limited to, the
implied warranties or merchantability and fitness for a particular purpose.
Philips shall not be liable for errors contained herein or for incidental or consequential damages in
connection with the furnishing, performance, or use of this material.
Passwords
In order to access different modes within the monitor a password may be required. The passwords are
listed below.
Monitoring Mode: No password required
Configuration Mode: 71034
Demo Mode: 14432
Service Mode: 1345
Consult the configuration guide before making any changes to the monitor configuration.
Warnings and Cautions
In this guide:
•A warning alerts you to a potential serious outcome, adverse event or safety hazard. Failure to
observe a warning may result in death or serious injury to the user or patient.
•A caution alerts you where special care is necessary for the safe and effective use of the product.
Failure to observe a caution may result in minor or moderate personal injury or damage to the
product or other property, and possibly in a remote risk of more serious injury.
8
2Theory of Operation
Monitor Theory of Operation
The IntelliVue MP5/MP5T Patient Monitor:
•displays real-time data
•alarms in the case of patient or equipment problems
•offers limited data storage and retrieval (trending)
•interfaces to the Philips Clinical Network and other equipment (not MP5T)
The monitor can be configured with various different measurement and interface capabilities.
The following comparison table shows in detail the differences between MP5 and MP5T:
connects multiple patient monitors, information centers,
application servers; closed system, only Philips qualified products
(tested and with regulatory approval) are connected, Philips is
responsible for guaranteed real-time functionality and performance
Philips Clinical Network (wireless)
like Philips Clinical Network (wired) LAN, however due to current
wireless technologies available it has reduced bandwidth, longer
latencies, reduced functionality
Hospital LAN, Internet
Standard Network, not under Philips control, no guaranteed
service, no real-time requirements
11
2 Theory of Operation
Hardware Building Blocks
The following hardware building blocks make up the monitoring system. (Note that the MP5T does
not include all the hardware components shown below):
12
Figure 2 MP5 Hardware Building Blocks
IntelliVue MP5/MP5T
The MP5/MP5T monitor:
•integrates the display and processing unit into a single package
•uses a 8.4” TFT SVGA color display
•uses the Touchscreen as input device
•integrates the measurement block (Front End 1 (FE1) and Front End 2 (FE2)) with optional
parameter sets
Optional Hardware
•One slot is provided for one of three available system interface boards. An optional built-in
wireless network interface IntelliVue 802.11 Bedside Adapter or IntelliVue Instrument Telemetry)
is supported. For further details regarding the wireless network please refer to the M3185A Philips
Clinical Network documentation.
•optional recorder
•optional battery
•optional MSL board
•optional Short Range Radio (SRR) board
2 Theory of Operation
Power Distribution
Figure 3 Power Distribution Architecture
The AC/DC converter transforms the AC power (100-240 V AC range) coming from the power plug
into a 24 V / 50W DC source and isolates the monitoring system from the AC power mains.
The 24 V DC is distributed via the Interface Board to the optional battery charging circuit and to the
main- and recorder board.
13
2 Theory of Operation
If the interface board contains the optional battery circuits, the power is used to charge the battery and
supply the monitoring system. As soon as the AC power source is disconnected, the optional battery
starts and keeps the system powered (battery mode). A DC/DC converter on the interface board
converts the 8-12.6 V DC power supplied by the battery into 15 V DC power, which is distributed to
the monitoring system.
The main board and recorder board contain power supply circuits, which convert the 24 /15 V DC
into several voltages supplying the particular components of the monitoring system.
The realtime clock and the buffered RAM is supplied with cont. 3.6 V DC power, provided either by
the 24 / 15 V DC system power or by the battery power and converted to 3.6 V DC.
The CPU board has an MPC852/62 MHz processor in the patient monitor that provides a number of
on-chip, configurable interfaces. An array of fast UARTS with configurable protocol options are
implemented in an ASIC (along with other system functions such as independent watchdogs, video,
etc.), providing interfacing capabilities to integrated measurements and System Interface Boards. The
serial interfaces can easily be electrically isolated. The main board contains additional video hardware.
The CPUs provide a LAN interface to connect to the Philips Clinical Network (Ethernet).
NOTE
An MP5 in companion mode does not receive its power from the host monitor via the MSL. MP5 is
always powered by AC power or battery.
System Interfaces
The following is a list of Interface boards which may be present in your monitor, depending on your
purchased configuration:
Note that WLAN, IIT and MSL Interface require the full system interface board.
The MP5T is delivered only with the battery system interface board.
The specifications for the above listed interfaces can be found in the technical data sheet for the
monitor and in the Installation and Specifications chapter of the Instructions for Use.
#J02
2
#J40
2
14
Compatible Devices
Figure 4 IntelliVue G1/G5 Anesthetic Gas Module
Data Flow
The following diagram shows how data is passed through the monitoring system. The individual stages
of data flow are explained below.
2 Theory of Operation
Data Acquisition
Monitoring data (for example patient measurement data in the form of waves, numerics and alerts) is
acquired from a variety of sources:
•Measurement Block
The integrated measurements convert patient signals to digital data and apply measurement
algorithms to analyze the signals.
•External measurement devices
Data can be also acquired from devices connected to interface boards of the monitor. Software
modules dedicated to such specific devices convert the data received from an external device to
the format used internally. This applies to the IntelliVue G1/G5 Anesthetic Gas Module.
Figure 5 Data Flow
15
2 Theory of Operation
•Server systems on the Philips Clinical Network
To enable networked applications such as the other bed overview, data can be acquired from server
systems attached to the Philips Clinical Network, for example a Philips Information Center
Data Provider System Service
All data that is acquired from integrated measurements or external measurement devices is temporarily
stored by a dedicated data provider system service. All monitor applications use this central service to
access the data in a consistent and synchronized way rather than talking to the interfaces directly.
This service makes the applications independent of the actual type of data acquisition device.
The amount of data stored in the data provider system service varies for the different data types. For
example several seconds of wave forms and the full set of current numerical values are temporarily
stored in RAM.
Persistent Data Storage System Service
Some applications require storage of data over longer periods of time. They can use the persistent data
storage system service. Dependent on the application requirements, this service can store data either in
battery backed-up (buffered) memory or in flash memory. The buffered memory will lose its contents
if the monitor is without power (not connected to mains) for an extended period of time. The flash
memory does not lose its contents.
The trend application for example stores vital signs data in a combination of flash memory and
buffered memory, while the system configuration information (profiles) is kept purely in flash
memory.
Display and User Interface Service
Applications can use high level commands to display monitoring data or status and command windows
on the internal LCD panel. These commands are interpreted by the display manager application. This
application controls the dedicated video hardware which includes video memory and a special
hardware in the ASIC.
User input is acquired from the touchscreen. The system software makes sure that the user input is
directed to the application which has the operating focus.
Monitor Applications
The monitor applications provide additional system functionality over the basic measurement and
monitoring capabilities. This includes for example trending, report generating, event storage or derived
measurements.
In general, the monitor applications use the data provider system service to access the measurement
data. Application interfaces to the other system services allow the application to visualize data, to store
data over extended periods of time or to output data to other devices.
Internal LAN (Measurement Link)
The MP5/MP5T communicates as a Multi-Measurement Module (MMS) in companion mode when
connected to a host monitor using an IEEE802.3/Ethernet LAN in the Measurement Link (MSL).
This network is used to distribute data between the the MP5/MP5T and the host monitor, for
example:
16
2 Theory of Operation
•Digitized patient signals including wave data, numerical data and status information (typically from
the measurement server to a display unit)
•Control data representing user interactions (typically from the display unit to a measurement
server)
•Shared data structures, for example representing patient demographical data and global
configuration items
The internal LAN allows plug and play configuration of the monitoring system. The system
automatically detects plugging or unplugging of measurement servers on the host monitor and
configures the system accordingly.
The components on the internal LAN are time-synchronized to keep signal data consistent in the
system. Dedicated hardware support for synchronization eliminates any latency of the network driver
software.
The integrated LAN provides deterministic bandwidth allocation/reservation mechanisms so that the
real-time characteristic of signal data and control data exchange is guaranteed. This applies to the data
flow from the measurement server to the monitor (for example measurement signal data) and the data
flow from the monitor to a measurement server (for example to feed data to a recorder module).
Integrated communication hubs in the monitor allow flexible cabling options (star topology, daisy
chaining of servers).
NOTE
The MP5/MP5T does not support any MMS on the MSL.
Philips Clinical Network
The monitoring system may be connected to the Philips Clinical Network, for example to provide
central monitoring capabilities or other network services. This connection may be through a normal
wired connection or through a wireless connection.
The monitor supports the connection of an internal wireless adapter (#J35, #J45, #J47). Switching
between wired and wireless networks is automatically triggered by the plugging or unplugging of the
network cable.
After configuration, the monitoring system sends the digitized patient signals including wave data,
numerical data and status information onto the network. Control data representing user interactions
can be exchanged between the monitoring system and a central station bi-directionally.
Additional protocols are supported for networked applications, for example for the other bed
overview function, which allows viewing of monitoring data from other patients on the network.
For plug and play operation, the monitoring system uses the standard BootP protocol to automatically
acquire a network address.
How does the Support Tool Work with the Monitor
The support tool is a Windows application typically installed on the laptop of a customer engineer or a
biomedical engineer working in the customer’s own service department.
The purpose of the support tool is to upgrade, configure and diagnose all monitoring components in
the system over the network.
17
2 Theory of Operation
The service protocol developed for this purpose uses a raw access to the devices without the need for
IP addresses etc. over a standard customer network installation, so that even defective devices can be
upgraded as long as the few kBytes of initial boot code are working. The boot code itself can also be
upgraded using the same protocol.
The tool allows access to internal service information and to serial numbers. It can be remotecontrolled, for example via a dial-up connection from a response center, provided the proper
infrastructure is in place.
For details see the Instructions for Use for the Support Tool.
18
Monitor Software Block Diagram
The following shows the functional block diagram for the monitoring system. A legend explaining
terms and diagram elements follows. The information below varies depending on the purchased
monitor options.
2 Theory of Operation
Figure 6 IntelliVue Patient Monitoring System Functional Block Diagram
19
2 Theory of Operation
Block Diagram Legend
Functional BlockDescription
Services
Operating SystemThe Operating System (OS) provides a layer of isolation
System ServicesThe System Services provide generic common system
Applications
ReportsThe Reports Service retrieves current and stored physiological
between the specific hardware implementation and the
application software. The OS performs system checks and
allocates resources to ensure safe operation when the system
is first started. This includes internal self-tests on several
hardware modules and configuration checks for validity of
configuration with the operating software. During normal
operation, the OS continues to run checks on system integrity.
If error conditions are detected the OS will halt monitoring
operations and inform the operator about the error condition.
services.
In particular:
They use a real-time clock component to track time. They
synchronize to network time sources and verify the accuracy
of the system time information. They are also responsible for
managing persistent user configuration data for all
Measurement parameters and IntelliVue Patient Monitoring
System software modules. User configuration data is stored in
a non-volatile read/write storage device
data and status data to format reports for printing paper
documentation. Examples of supported reports:
20
•Vital Signs Report
•Graphical Trend Report
•Event Review Report
•Event Episode Report
•ECG Report (12 Lead/Multi-Lead)
•Test Report
The Reports service generates report data which can be
printed on a central printer.
RecordThe Record Service retrieves current and stored physiological
data and status data to format a continuous strip recording. A
recording can be triggered manually by the operator or
automatically by an alarm condition. The Record Service uses
the services of the Recorder Interface to control a recorder.
The Record Service can also send data to a central recorder.
2 Theory of Operation
Functional BlockDescription
AlarmThe Alarm Service contains logic that prioritizes alarm
conditions that are generated by IntelliVue Patient Monitoring
System software modules. Visual alarm signals (messages) are
displayed at the top of the IntelliVue Patient Monitoring
System display and alarm sounds are generated by a
loudspeaker. Alarm conditions may be generated when a
physiological parameter exceeds preselected alarm limits or
when a physiological parameter or any other software module
reports an inoperative status (technical alarm, for example, the
ECG leads may have fallen off the patient). The Alarm service
manages the alarm inactivation states, for example suspension
of alarms, silencing of alarms, and alarm reminder. Alarm
signals may also be configured as latching (alarm signals are
issued until they are acknowledged by the operator, even
when the alarm condition is no longer true). The Alarm
service controls the visual alarm signals (alarm lamps).
TrendThe Trend service stores the sample values of physiological
data and status data with a resolution of 12 seconds, 1 minute
or 5 minutes for a period of up to 48 hours. The data is kept
in battery buffered read/write storage and flash memory
devices to be preserved across power failures. The stored data
is protected via consistency checks and checksums. When a
new patient is admitted, the trend database erases all data of
the previous patient.
OxyCRGThe OxyCRG (Oxygen CardioRespiroGram) service derives a
high-resolution trend graph from the Beat-to-Beat Heart
Rate, SpO2, and Respiration physiological data. The OxyCRG
is specialized for neonatal applications, allowing the operator
to identify sudden drops in Heart Rate (Bradycardia) and
SpO2 (Desaturation), and supporting the operator in
visualizing Apnea situations.
ADTThe ADT (Admit/Discharge/Transmit) service maintains the
patient demographics information. The operator may admit a
new patient, discharge the old patient and enter or modify the
patient demographics.
calculations on physiological numerical values to derive
calculated parameters like Temperature Difference.
Interface Managers
21
2 Theory of Operation
Functional BlockDescription
MDSEThe MDSE (Medical Data Service Element) Interface
PrinterThe Printer Interface Manager provides a high level interface
Display & Operator InterfaceThe Display and Operator Interface Manager performs the
Manager is responsible for the exchange of real-time data
between the IntelliVue Patient Monitoring System display unit
and the Measurement parameters and other devices attached
to the network. MDSE establishes and maintains a data
communication link between the devices. It provides
configuration information about the remote device to
applications in the local device and it allows the exchange of
measurement data and status information between the
devices.
to a printer. It provides means to:
•establish a connection to the printer
•transfer data to the printer
•get status of the printer
•close connection to the printer
The Printer Interface Manager also supervises the connection
to the printer and whether the printer accepts data (for
example paper out). The Printer Interface Manager notifies
the operator in such cases.
following tasks:
Interfaces
•Screen presentation of real-time and stored physiological
measurement data, alarm condition data and status
information received from the MDSE interface manager,
the Alarm service or other IntelliVue Patient Monitoring
System modules
•Screen presentation of operating controls (control
windows)
•Processing of operating control commands received from
HIF Control interface. The module verifies and interprets
the received commands and forwards them to other
software modules of the IntelliVue Patient Monitoring
System display unit or measurement parameters.
•Sound generation (issues audible alarm signals and
generates audible information signals, for example QRS
and SpO2 tones, operator audible feedback)
22
2 Theory of Operation
Functional BlockDescription
LANThe LAN interface implements the physical layer of IEEE
802.3. The LAN interface performs Manchester encoding/
decoding, receive clock recovery, transmit pulse shaping,
jabber, link integrity testing, reverse polarity detection/
correction, electrical isolation, and ESD protection.
Electronically separated interfaces are used for
communication to the Measurement parameters and to the
network.
Display ControllerThe Display Controller Interface consists of a video
controller, video RAM and the controlling software. The
Display Controller interface processes the high level display
commands (character and graphic generation, wave drawing)
and translates them into pixels, which are written into the
video RAM where the video controller chip generates the
video synchronization signals and the pixel stream for the
Color LCD Display.
HIF ControlThe HIF (Human Interface Control) interface scans the
Human Interface devices for operator controls (Touch
Screen), formats the collected data and sends it to the display
and Operating Interface.
ECG-Out Marker-In
(not for MP5T)
The ECG Out/Marker In interface receives the ECG
waveform directly from the ECG/Resp Arrhythmia STSegment physiological algorithm running on the main CPU
and converts the digital ECG signal to an analog ECG signal.
In addition, the ECG Out hardware receives from a
connected device the marker information and forwards this
data to the ECG/Resp Arrhythmia ST-Segment physiological
algorithm. The converted analog signal is used to synchronize
a connected device to the patient’s ECG
Nurse Call
(not for MP5T)
MIB
(not for MP5T)
The Nurse Call board contains a phone jack type connector
with a single close-on-alarm relay.
The MIB interface allows full-duplex, short-haul
asynchronous binary communication between the monitor
and an arbitrary (medical/non-medical) device using an eightpin RJ45 modular connector. Communication protocols using
this interface can be configured.
ECG Sync Out
(not for MP5T)
A pulse signal is provided on the RS-232 interface to allow
synchronisation with other medical devices.
23
2 Theory of Operation
Functional BlockDescription
IIT
(not for MP5T)
The built-in IIT adapter allows operation of the MP5
monitors within IntelliVue Instrument Telemetry
infrastructure.
WLAN
(not for MP5T)
The bulit-in WLAN interface allows wireless operation of the
MP5 monitors with the IntelliVue 802.11 Bedside Adapter.
SRRThe built-in SRR interface allows wireless communication of
the MP5 and MP5T monitors with an IntelliVue Instrument
Telemetry Transceiver.
MSL
(not for MP5T)
All components of the monitoring system communicate using
an IEEE802.3/ Ethernet LAN in the Measurement Link
(MSL). This network is used to distribute data between the
components
24
3Testing and Maintenance
Introduction
This chapter provides a checklist of the testing and maintenance procedures to ensure the performance
and safety of the monitor.
These tests must be performed only by qualified personnel certified by the responsible organization.
Qualifications required are: training on the subject, knowledge, experience and acquaintance with the
relevant technologies, standards and local regulations. The personnel assessing safety must be able to
recognize possible consequences and risks arising from non-conforming equipment.
All recurring safety and performance assurance tests must be performed under equal environmental
conditions to be comparable.
Preventive Maintenance refers specifically to the series of tests required to make sure the measurement
results are accurate. The accuracy and performance procedures are designed to be completed as
specified in the following sections or when readings are in question.
3
For detailed instructions on the maintenance and cleaning of the monitor and its accessories, see Care
and Cleaning, Using Batteries and Maintenance and Troubleshooting in the monitor's Instructions for Use.
Terminology and Definitions
The following terms and definitions are used throughout this chapter and taken from the international
standards IEC 60601-1, IEC 60601-1-1 and IEC 62353.
•Medical System: a medical electrical system is a combination of at least one medical electrical
device and other electrical equipment, interconnected by functional connection or use of a
multiple portable socket-outlet.
•Patient Vicinity: any area in which intentional or unintentional contact can occur between the
patient and parts of the medical system or between the patient and other persons who have had
contact with parts of the medical system. The patient vicinity is defined anywhere within 1.5m (5
feet) of the perimeter of the patient's bed and 2.5m (8.2 feet) from the floor.
•Separation Device/Transformer: a component or arrangement of components with input parts
and output parts that, for safety reasons, prevent a transfer of unwanted voltage or current
between parts of a medical system.
•Multiple Portable Socket-Outlet: a combination of two or more socket-outlets intended to be
connected to or integrated with flexible cables or cords, which can easily be moved from one place
to another while connected to the power mains.
•Functional Connection: an electrical connection for transfer of signals and/or power.
25
3 Testing and Maintenance
•Tests: Safety or Performance Assurance test procedures which may consist of several steps.
Recommended Frequency
Perform the procedures as indicated in the suggested testing timetable. These timetable
recommendations do not supersede local requirements.
Table 1 Table 1: Suggested Testing Timetable
Test sFrequency
Preventive Maintenance
Other Regular Tests
Performance Assurance
Test s
Safety
Test s
Visual
Electrical
NBP Performance Once every two years, or more often if
specified by local laws.
Microstream CO2 CalibrationOnce a year or after 4000 hours of
continuous use and following any
instrument repairs or the replacement of
any instrument parts.
Visual InspectionBefore each use.
Power On Test
ECG/Resp PerformanceOnce every two years, or if you suspect the
ECG Out Sync Performance*
ECG Sync Pulse Performance*
SpO2 Performance
NBP Performance
Invasive Pressure Performance*
Temperature Accuracy*
Predictive Temperature Accuracy Check
Mainstream CO2 Accuracy Check*
Sidestream CO2 Accuracy Check and
Flow Check*
Microstream CO2 Performance Test*
Nurse Call Relay Performance*
Power Loss Alarm Buzzer Performance
Visual InspectionAfter each service event
Protective EarthOnce every two years and after repairs
Equipment Leakage Current
Patient Leakage Current
System TestOnce every two years
measurement is incorrect, except
Mainstream CO2 Accuracy Check,
Sidestream CO2 Accuracy Check and Flow
Check and Predictive Temperature
Accuracy Check - required once a year.
where the power supply has been removed
or replaced or the monitor has been
damaged by impact.
26
* These tests do not apply for MP5T.
When to Perform Tests
This table tells you when to perform specific tests.The corresponding test procedures are described in
the following sections All tests listed below must be performed on the monitor.
Table 2 When to perform tests
3 Testing and Maintenance
Service Event
(When performing...
Installation
Installation of a monitor in combination with a
medical or non-medical device connected to the
same multiple socket outlet.
Installation of a standalone monitor with no
display connected to the video output
Installation of a monitor with a medical display
specified by Philips
Installation of a monitor with an off-the-shelf
display (non-compliant with IEC60601-1)
Installation of a monitor with IntelliVue G1/
G5, connected to separate mains sockets.
Installation of monitor with IntelliVue
Instrument Telemetry (IIT)
Installation of a monitor with IT equipment e.g.
PC connected via a functional connection e.g.
Centronics or USB.
Installation of monitor with IntelliVue 802.11
Bedside Adapter
Installation of a monitor with Short Range
Radio (SRR)
Installation of networked monitor (LAN)Perform Visual Inspection and Power On Test
Preventive Maintenance
Preventive Maintenance*Perform preventive maintenance tests and
Tests Required
...Complete these tests)
Perform Visual Inspection, Power On and System
Tests
Perform Visual Inspection and Power On Test
Perform Visual Inspection and Power On Test
Perform Visual Inspection, Power On and System
Test
Perform Visual Inspection and Power On Tests
Perform Visual Inspection, Power On and IIT
communication test
Perform Visual Inspection, Power On and System
Tests
Perform Visual Inspection, Power On and
IntelliVue 802.11 Bedside Adapter
Communication Test
Perform Visual Inspection, Power On and SRR
communication test
procedures:
•NBP calibration
•Microstream CO2 calibration
Other Regular Tests and Tasks
Visual InspectionPerform Visual Inspection
Power On TestPerform Power On test
Repairs
Repairs where the monitor has been damaged by
impact, liquid ingression, fire, short circuit or
electrical surge.
Perform Visual Inspection, Power On, all Safety
Tests and Full Performance Assurance Tests
27
3 Testing and Maintenance
Service Event
(When performing...
Repairs where the power supply, the mains
socket or an interface board is removed or
Tests Required
...Complete these tests)
Perform Visual Inspection, Power On, all Safety
Tests and Basic Performance Assurance Test
replaced or the protective earth ground
connection is disrupted.
Repairs where the main board has been replaced. Perform Visual Inspection, Power On, Basic
Performance Assurance Test and NBP Accuracy
Test and Calibration.
Repairs where the measurement block has been
removed or replaced
Perform Visual Inspection, Power On, all Safety
Tests and Basic Performance Assurance Test.
If a certain parameter seems suspicious, perform
Full Performance Assurance Test for this
parameter.
Repairs where the NBP pump has been replaced Perform Visual Inspection, Power On, all Safety
Tests, Basic Performance Assurance Test and NBP
Performance Test and Calibration
Repairs of IntelliVue Instrument Telemetry (IIT)
Module
Perform Visual Inspection, Power On Test Block
and IIT communication test
Repairs of IntelliVue 802.11 Bedside AdapterPerform Visual Inspection, Power On and
IntelliVue 802.11 Bedside Adapter
Communication Test
Repairs of Short Range Radio (SRR) InterfacePerform Visual Inspection, Power On and SRR
Communication Test
Repairs of the IntelliVue G1/G5Perform Basic Performance Assurance Test. For
further testing requirements, see IntelliVue G1/
G5 Service Guide
All other IntelliVue Monitoring System repairs
(except when power supply is removed)
Perform Visual Inspection, Power On Test and
Basic Performance Assurance Test
Performance Assurance
Basic Performance AssurancePerform basic performance assurance tests for the
respective monitoring system component.
Full Performance AssurancePerform all accuracy and performance test
procedures listed in the following sections. If a
particular measurement is in question, perform the
measurement performance test only.
Upgrades
Software UpgradesPerform Visual Inspection, Power On Test and
Basic Performance Assurance Test unless
otherwise specified in the Upgrade Installation
Notes shipped with the upgrade.
Hardware UpgradesPerform Visual Inspection, Power On Test and
Basic Performance Assurance Test unless
otherwise specified in the Upgrade Installation
Notes shipped with the upgrade.
28
3 Testing and Maintenance
Service Event
(When performing...
Hardware Upgrades where IntelliVue
Instrument Telemetry (IIT) is installed
Hardware Upgrades where IntelliVue 802.11
Bedside Adapter is installed
Hardware Upgrades where Short Range Radio
(SRR) is installed
Installation of Interfaces or Hardware Upgrades
where the power supply or parameter boards
need to be removed.
Combining or Exchanging System
Components (non-medical equipment
connected to an IntelliVue monitor or medical
system equipment operated on a multiple socket
outlet)
Tests Required
...Complete these tests)
Perform Visual Inspection, Power On Test, Basic
Performance Assurance Test and IIT
communication Test
Perform Visual Inspection, Power On Test, Basic
Performance Assurance Test and IntelliVue 802.11
Bedside Adapter Communication Test
Perform Visual Inspection, Power On Test, Basic
Performance Assurance Test and SRR
communication Test
Perform Visual Inspection, Power On Test, Basic
Performance Tests and all Safety Tests
Perform the System Test for the respective system
components
NOTE
It is the responsibility of the facility operator or their designee to obtain reference values for recurring
safety and system tests. These reference values are the results of the first test cycles after an installation.
You may also purchase this service from Philips.
29
3 Testing and Maintenance
Testing Sequence
Summary of the recommended sequence of testing:
NOTE
If any single test fails, testing must be discontinued immediately and the device under test must be
repaired or labeled as defective.
Visual Inspection
Before Each Use
Check all exterior housings for cracks and damage. Check the condition of all external cables,
especially for splits or cracks and signs of twisting. If serious damage is evident, the cable should be
replaced immediately. Check that all mountings are correctly installed and secure. Refer to the
instructions that accompany the relevant mounting solution.
After Each Service, Maintenance or Repair Event
Ensure all fuses accessible from the outside comply with the manufacturer’s specification.
Check:
•the integrity of mechanical parts, internally and externally.
•any damage or contamination, internally and externally
30
•that no loose parts or foreign bodies remain in the device after servicing or repair.
•the integrity of all relevant accessories.
Power On Test
1Connect the monitoring system to mains and switch it on. This includes connected displays and
gas analyzers.
2Make sure that all steps listed in the table Initial Instrument Boot Phase in the Troubleshooting section
are completed successfully and that an ECG wave appears on the screen.
The expected test result is pass: the monitor boots up and displays an ECG wave. The wave might be
a flat line if no simulator is attached.
Safety Tests
Safety tests are comprised of the following tests performed on the monitoring system:
•protective earth resistance
•equipment leakage current
•applied part leakage current
3 Testing and Maintenance
•system test (if required)
Safety test requirements are set according to international standards, their national deviations and
specific local requirements. The safety tests detailed in this Service Guide are derived from
international standards but may not be sufficient to meet local requirements. We recommend that you
file the results of safety tests. This may help to identify a problem early particularly if the test results
deteriorate over a period of time.
Each individual piece of equipment of the monitoring system which has its own connection to mains
or which can be connected or disconnected from mains without the use of a tool must be tested
individually. The monitoring system as a whole must be tested according to the “System Test” on
page 57 procedure.
Accessories of the monitoring system which can affect the safety of the equipment under test or the
results of the safety test must be included in the tests and documented.
Warnings, Cautions, and Safety Precautions
•These tests are well established procedures of detecting abnormalities that, if undetected, could
result in danger to either the patient or the operator.
•Disconnect the device under test from the patient before performing safety tests.
•Disconnect the device under test from mains before performing safety tests. If this is not possible,
ensure that the performance of these tests does not result in danger to the safety analyzer operator,
patients or other individuals.
•Test equipment (for example, a Safety Analyzer) is required to perform the safety tests. Please refer
to Annex C of IEC/EN 62353 for exact requirements for the measurement equipment and for
measurement circuits for protective earth resistance and leakage currents. Refer to the
documentation that accompanies the test equipment. Only certified technicians should perform
safety testing.
31
3 Testing and Maintenance
•The consistent use of a Safety Analyzeras a routine step in closing a repair or upgrade is emphasized
as a mandatory step to maintain user and patient safety. You can also use the Safety Analyzer as a
troubleshooting tool to detect abnormalities of line voltage and grounding plus total current loads.
•During safety testing, mains voltage and electrical currents are applied to the device under test.
Ensure that there are no open electrical conductive parts during the performance of these tests.
Avoid that users, patients or other individuals come into contact with touch voltage.
•For Europe and Asia/Pacific, the monitor complies with:
IEC60601-1:1988 + A1:1991 + A2:1995 = EN60601-1:1990 +A1:1993 + A2:1995
IEC60601-1-1:2000
For USA, the monitor complies with:
UL60601-1
For Canada, CAN/CSA C22.2#601.1-M90
•Local regulations supersede the testing requirements listed in this chapter.
•If a non-medical electrical device is connected to a medical electrical device, the resulting medical
electrical system must comply with IEC/EN 60601-1-1.
•Perform safety tests as described on the following pages.
Safety Test Procedures
Use the test procedures outlined here only for verifying safe installation or service of the product. The
setups used for these tests and the acceptable ranges of values are derived from local and international
standards but may not be equivalent. These tests are not a substitute for local safety testing where it is
required for an installation or a service event. If using an approved safety tester, perform the tests in
accordance with the information provided by the manufacturer of the tester and in accordance with
your local regulations, for example IEC/EN 60601-1, UL60601-1 (US), IEC/EN 62353, and IEC/EN
60601-1-1. The safety tester should print results as detailed in this chapter, together with other data.
Please refer to Annex C of IEC/EN 62353 for requirements for the measurement equipment and for
measurement circuits for protective earth resistance and leakage currents.
32
3 Testing and Maintenance
The following symbols are used in the diagrams illustrating the safety tests:
Resistance measuring deviceConnection to accessible
conductive parts
.........Optional connection
CAUTION
After each service, maintenance or repair event:
Ensure all fuses accessible from the outside comply with the manufacturer’s specification.
Check:
•the integrity of mechanical parts, internally and externally.
•any damage or contamination, internally and externally.
•that no loose parts or foreign bodies remain in the device after servicing or repair.
•the integrity of all relevant accessories.
Hints for Correct Performance of Safety Tests
•Perform a visual inspection on all detachable power cords used with the monitoring system and
include these in all safety test procedures.
•Connection lines such as data lines or functional earth conductors may appear to act like protective
earth connections. These may lead to incorrect measurements and need to be considered during
testing. If necessary, unplug these connections.
33
3 Testing and Maintenance
•Position all cables and cords in such a manner that they do not influence the safety tests.
•Measurement of insulation resistance is not required.
Guideline for Performance of Safety Tests
Connect the detachable power cord of the device under test to the safety analyzer's test mains port.
Connect the enclosure test lead of the safety analyzer to the enclosure of the device under test, e.g. to
the equipotential connector. For testing the applied part leakage current, connect all applied parts to
the safety analyzer using the appropriate patient lead or adapter cable. For the ECG parameter all ten
ECG-leads need to be connected to the safety analyzer. If necessary, use an adapter cable to connect
all ten ECG-leads. If necessary, repeat the safety test procedure until all available applied parts have
been tested. Refer to the documentation that accompanies the safety analyzer for further details on
how to set up and perform the test.
Protective Earth Resistance Test - Setup Example
Equipment Leakage Current Test - Setup Example
34
3 Testing and Maintenance
Applied Part Current Test - Setup Example
NOTE
The above graphics resemble the Metron QA-90 setup and are protected by copyright. Copyright
owned by Fluke (Metron).
Safety Test Adapter Cable - Schematics
The following graphics provide schematics of safety test (patient lead) adapter cables which can be
used for electrical safety testing. These schematics can also be used as a guideline for making your own
safety test adapter cables. Alternatively, other methods to make safety test adapter cables can be used,
e.g. using a modified accessory cable.
NOTE
You may not need all of the cables displayed below for electrical safety testing of your respective
monitor.
Use Clamp Adapter Cable and M1034-61650 BIS sensor simulator.
44
VueLink
3 Testing and Maintenance
4 = 220 Ohm
45
3 Testing and Maintenance
IntelliBridge
46
EEG
3 Testing and Maintenance
47
3 Testing and Maintenance
SvO2 (M1021A)
48
ScVO2 (M1011A)
3 Testing and Maintenance
49
3 Testing and Maintenance
tcpO2/tcpCO2
50
MP5 Predictive Temperature
3 Testing and Maintenance
51
3 Testing and Maintenance
MP5 TAAP
52
S(1): Protective Earth Resistance Test
Test to perform:
3 Testing and Maintenance
Measuring circuit for the measurement of Protective Earth Resistance in medical electrical
equipment that is disconnected from the supply mains.
This measures the impedance of the Protective Earth (PE) terminal to all exposed metal parts of the
Instrument under Test (IUT), which are for safety reasons connected to the Protective Earth (PE).
Measurements shall be performed using a measuring device capable to deliver a current of at least
200 mA into 500 mOhms with maximum 24V
This safety test is based on IEC/EN 60601-1, IEC/EN 62353, UL2601-1 Ed. 2/UL60601-1:2003 and
CSA 601.1-M90.
For measurement limits, refer to Safety (1) test, Test and Inspection Matrix.
Report the highest value (X1).
TestExpected test results
Protective Earth Resistance Test (with
X1 <= 300mOhms
mains cable)
NOTE
•If the protective earth resistance test fails, testing must be discontinued immediately and the device
under test must be repaired or labeled as defective.
•All values for current and voltage are the root mean square (r.m.s.) values, unless otherwise stated.
•Flex the power cord during the protective earth resistance test to evaluate its integrity. If it does
not pass the test, exchange the power cord.
53
3 Testing and Maintenance
S(2): Equipment Leakage Current Test - Normal Condition
Test to perform:
Measuring circuit for the measurement of Equipment Leakage Current - Direct method
according to IEC/EN 62353.
This test measures leakage current of exposed metal parts of the monitor and the functional earth
leakage current. It tests normal and reversed polarity. Perform the test with S1 closed (Normal
Condition).
There are no parts of the equipment that are not protectively earthed.
This safety test is based on IEC/EN 60601-1, IEC/EN 62353, UL2601-1 Ed. 2/UL60601-1:2003 and
CSA 601.1-M90.
For measurement limits, refer to Safety (2) test, Test and Inspection Matrix.
Report the highest value (X1).
TestExpected test results
Equipment Leakage Current Test
X1 <= 100μA
(Normal Condition - with mains cable)
NOTE
All values for current and voltage are the root mean square (r.m.s.) values, unless otherwise stated.
54
S(3): Equipment Leakage Current Test - Single Fault Condition
Test to perform:
Measuring circuit for the measurement of Equipment Leakage Current - Direct method
according to IEC/EN 62353.
3 Testing and Maintenance
This test measures leakage current of exposed metal parts of the monitor and the functional earth
leakage current. It tests normal and reversed polarity. Perform the test with S1 open (Single Fault
Condition).
There are no parts of the equipment that are not protectively earthed.
This safety test is based on IEC/EN 60601-1, IEC/EN 62353, UL2601-1 Ed. 2/UL60601-1:2003 and
CSA 601.1-M90.
For measurement limits, refer to Safety (3) test, Test and Inspection Matrix.
Report the highest value (X2).
TestExpected test results
Equipment Leakage Current Test
X2 <= 300μA
(Single Fault Condition - with mains
cable)
NOTE
All values for current and voltage are the root mean square (r.m.s.) values, unless otherwise stated.
55
3 Testing and Maintenance
S(4): Applied Part Leakage Current - Mains on Applied Part
NOTE
During measurement of the Applied Part Leakage Current it is possible that the measured current can
exceed the allowed limit (per IEC/EN 60601-1 or IEC/EN 62353).
This can occur when the safety tester is connected to the invasive blood pressure and temperature
connectors at the same time during the applied leakage current measurement.
The connectors for the invasive blood pressure and temperature are independently functioning
connectors.
Although there are individual connectors on the front end, internally those parameters use the same
electrical insulation interface and are hardwired to each other. This results in an electrical short of
those connectors during measurement if a test current is applied simultaneously. Therefore this should
be avoided.
Due to the combined insulation interface, it is sufficient to connect to only one parameter interface
(that is, Invasive Blood Pressure or Temperature) of the invasive blood pressure/temperature
measurement block. This avoids a short and the potential of exceeding the limit for the current.
Test to perform:
Measuring circuit for the measurement of Applied Part Leakage Current - Direct method
according to IEC/EN 62353.
56
This test measures applied part leakage current from applied part to earth caused by external main
voltage on the applied part. Each polarity combination possible shall be tested. This test is applicable
for ECG measurement inputs.
There are no parts of the equipment that are not protectively earthed.
3 Testing and Maintenance
This safety test is based on IEC/EN 60601-1, IEC/EN 62353, UL2601-1 Ed. 2/UL60601-1:2003 and
CSA 601.1-M90.
For measurement limits and test voltage, refer to test block Safety (4), Test and Inspection Matrix.
Report the highest value. (X1).
TestExpected test results
Applied Part Leakage Current Test
(Single Fault Condition - mains on
applied part)
NOTE
All values for current and voltage are the root mean square (r.m.s.) values, unless otherwise stated.
X1 <= 50μA
System Test
After mounting and setting up a system, perform system safety tests according to IEC/EN 60601-1-1.
What is a Medical Electrical System?
A medical electrical system is a combination of at least one medical electrical piece of equipment and
other electrical equipment, interconnected by functional connection or use of a multiple portable
socket-outlet.
•Devices forming a medical electrical system must comply with IEC/EN 60601-1-1.
•Any electrical device such as IT equipment that is connected to the medical electrical equipment
must comply with IEC/EN 60601-1-1 and be tested accordingly.
General Requirements for a System
After installation or subsequent modification, a system must comply with the requirements of the
system standard IEC/EN 60601-1-1. Compliance is checked by inspection, testing or analysis, as
specified in the IEC/EN 60601-1-1 or in this book.
Medical electrical equipment must comply with the requirements of the general standard IEC/EN
60601-1, its relevant particular standards and specific national deviations. Non-medical electrical
equipment shall comply with IEC safety standards that are relevant to that equipment.
Relevant standards for some non-medical electrical equipment may have limits for equipment leakage
currents higher than required by the standard IEC/EN 60601-1-1. These higher limits are acceptable
only outside the patient environment. It is essential to reduce equipment leakage currents to values
specified in IEC 60601-1 when non-medical electrical equipment is to be used within the patient
environment.
57
3 Testing and Maintenance
System Example
This illustration shows a system where both the medical electrical equipment and the non-medical
electrical equipment are situated at the patient’s bedside.
WARNING
•Do not use additional AC mains extension cords or multiple portable socket-outlets. If a multiple
portable socket-outlet is used, the resulting system must be compliant with IEC/EN 60601-1-1.
Do not place multiple socket-outlets on the floor. Do not exceed the maximum permitted load for
multiple socket-outlets used with the system. Do not plug additional multiple socket outlets or
extension cords into multiple socket outlets or extension cords used within the medical electrical
system.
•Do not connect any devices that are not supported as part of a system.
•Do not use a device in the patient vicinity if it does not comply with IEC/EN 60601-1. The whole
installation, including devices outside of the patient vicinity, must comply with IEC/EN 60601-1-
1. Any non-medical device placed and operated in the patient’s vicinity must be powered via a
separating transformer (compliant with IEC/EN 60601-1-1) that ensures mechanical fixing of the
power cords and covering of any unused power outlets.
System Installation Requirements
•Ensure that the the medical electrical system is installed in a way that the user achieves optimal use.
•Make sure the user is informed about the required cleaning, adjustment, sterilization and
disinfection procedures listed in the Instructions for Use.
•The medical electrical system must be installed in such a way that the user is able to carry out the
necessary cleaning, adjustment, sterilization and disinfection procedures listed in the Instructions
for Use.
58
•Ensure that the medical electrical system is installed in a way that an interruption and restoration
of power to any part of the medical electrical system does not result in a safety hazard.
3 Testing and Maintenance
•We recommend using fixed mains socket outlets to power the medical system or parts thereof.
Avoid using multiple portable socket-outlets.
•Any multiple portable socket outlets used must be compliant with IEC 60884-1 and IEC 60601-1-
1.
•Ensure that any part of the system connected to multiple portable socket-outlets is only removable
with a tool, i.e. the multiple portable socket-outlet provides a locking mechanism to prevent power
cords from being plugged or unplugged unintentionally. Otherwise, the multiple portable socketoutlet must be connected to a separation device. Multiple Socket Outlets used within the medical
electrical system must only be used for powering medical electrical equipment which is part of the
system.
•Ensure that any functional connections between parts of the medical electrical system are isolated
by a separation device according to IEC 60601-1-1 to limit increased equipment leakage currents
caused by current flow through the signal connections. This only works if the equipment leakage
current of the respective medical electrical system parts is not exceeded under normal conditions.
•Avoid increase of equipment leakage currents when non-medical electrical equipment within the
medical electrical system is used. This only works if the equipment leakage current of the
respective medical electrical system parts is not exceeded under normal conditions. Use additional
protective earth connection, separation device or additional non-conductive enclosures.
•Within the patient environment it is important to limit electrical potential differences between
different parts of a system. If necessary, use potential equalization equipment (equipotential cable)
or additional protective earth connections.
•Medical electrical equipment used in medical rooms must be connected to potential equalization
equipment (equipotential cable) to avoid electrical potential differences. Check your local
requirements for details.
59
3 Testing and Maintenance
Required Protective Measures at System Installation
For any IT equipment (IEC60950) operated in patient vicinity ensure that the equipment leakage
current does not exceed the limits described in IEC 60601-1. Use a separation device to ensure
compliance. After installation of IT equipment in patient vicinity, an enclosure leakage current test is
required.
60
Case 1: Medical Device Combined with Medical Device
If you combine a medical device with another medical device (incl. Philips specified displays) to form a
medical electrical system according to IEC60601-1-1, no additional protective measures are required.
The medical electrical devices may be located in or outside the patient vicinity in a medically used
room. This is valid as long as the medical devices are connected to separate mains outlets. No system
test is required.
3 Testing and Maintenance
61
3 Testing and Maintenance
If the combined medical devices are connected to the same multiple portable socket outlet an
enclosure leakage current test of the entire device combination on the multiple portable socket outlet is
required to ensure that the resulting protective earth leakage current and equipment leakage current
does not exceed the limits of IEC 60601-1-1. Avoid using multiple portable socket outlets. The
medical electrical devices may be located in or outside the patient vicinity in a medically used room. If
the limits are exceeded, additional protective measures are required, e.g. a separation device or the
connection of each device to separate mains.
62
Case 2: Medical Device Combined with a Non-Medical Device
If you combine a medical device with a non-medical device to form a medical electrical system
according to IEC60601-1-1, additional protective measures are required, e.g. usage of a separation
device. The medical electrical devices or the IT equipment may be located in or outside the patient
vicinity in a medically used room. After system installation incl. protective measures, a system test is
required to ensure that the resulting equipment leakage current and applied part leakage current does
not exceed the limits of IEC 60601-1-1.
3 Testing and Maintenance
For any IT equipment (IEC60950) operated in patient vicinity ensure that the equipment leakage
current does not exceed the limits described in IEC 60601-1. Use a separation device to ensure
compliance. After installation of IT equipment in patient vicinity, an enclosure leakage current test is
required.
63
3 Testing and Maintenance
64
If the combined devices forming the medical electrical system are connected to the same multiple
portable socket outlet, ensure that the resulting protective earth leakage current and equipment leakage
current do not exceed the limits of IEC 60601-1-1. The medical electrical devices or IT equipment may
be located in or outside the patient vicinity in a medically used room. Avoid using multiple portable
socket outlets. If the limits of IEC 60601-1-1 are exceeded, additional protective measures are
required, e.g. a separation device or the connection of each device to separate mains.
3 Testing and Maintenance
For any IT equipment (IEC60950) operated in patient vicinity ensure that the equipment leakage
current does not exceed the limits described in IEC 60601-1. Use a separation device to ensure
compliance. After installation of IT equipment in patient vicinity, an enclosure leakage current test is
required.
65
3 Testing and Maintenance
Case 3: Medical Device Combined with a Medical or Non-Medical Device with one
Device in a Non-Medically-Used Room
If you combine a medical device with a medical or non-medical device to form a medical electrical
system according to IEC60601-1-1 using a common protective earth connection and one of the
devices is located in a non-medically used room, additional protective measures are required, e.g. usage
of a separation device or additional protective earth connection. The medical electrical devices or IT
equipment may be located in or outside the patient vicinity. After system installation incl. protective
measures, a system test is required to ensure that the resulting equipment leakage current does not
exceed the limits of IEC 60601-1-1.
66
3 Testing and Maintenance
If you combine a medical device with a medical or non-medical device to form a medical electrical
system according to IEC60601-1-1 using two separate protective earth connections and one of the
devices is located in a non-medically used room creating a potential voltage difference, additional
protective measures are required, e.g. usage of a separation device or additional protective earth
connection. The medical electrical devices or IT equipment may be located in or outside the patient
vicinity. After system installation incl. protective measures, a system test is required to ensure that the
resulting equipment leakage current does not exceed the limits of IEC 60601-1-1.
67
3 Testing and Maintenance
System Test Procedure
If the medical electrical device has already been tested as a standalone device e.g. during factory safety
testing, an equipment leakage current test must only be performed once the device is connected to the
LAN network. If the medical electrical system has not been tested as a standalone device, the device
has to be tested as a standalone device (without connection to the system) and as part of the system
(with connection to the system).
Connect the detachable power cord of the device under test to the safety analyzer's test mains port.
Connect the enclosure test lead of the safety analyzer to the enclosure of the device under test, e.g. to
the equipotential connector. Refer to the documentation that accompanies the safety analyzer for
further details on how to set up the test.
TestExpected test results
Equipment Leakage Current Test
(Normal Condition)
Equipment Leakage Current Test
(Single Fault Condition)
After the testing of the device as a standalone device and as part of the system, check that the resulting
values (without connection and with connection to the system) do not differ by more than +/- 10%
from each other.
If the devices in the medical electrical system are connected to a multiple portable socket outlet the
resulting protective earth leakage current needs to be determined. All system components must be
connected to the multiple portable socket outlet and be switched on during this measurement.
TestExpected test results
Protective Earth Leakage Current of
Multiple Socket Outlets
Refer to the documentation that accompanies the safety analyzer for further details on how to set up
the test.
Carry out the noninvasive blood pressure measurement performance tests at least every two years , or
as specified by local laws (whichever comes first).
Microstream CO2 Calibration
Carry out the Microstream CO2 calibration once a year or after 4000 hours of continuous use and
following any instrument repairs or the replacement of any instrument parts.
68
Performance Assurance Tests
Some of the following test procedures must be performed in service mode. To enter service mode
select
Operating Modes in the main menu. Then select Service Mode and enter the password.
3 Testing and Maintenance
If required, open the screen menu in the monitor info line at the top of the screen and select
access the service screen. This is required particularly for Anesthetic Gas Module testing procedures.
Basic Performance Assurance Test
This section describes the basic performance test procedure. Please refer to the section “When to
Perform Tests” on page 27 for detailed information on when which test procedure is required.
Procedure:
Power on the monitoring system and go into demo mode. Check that each parameter (incl. Gas
Analyzer) displays values.
Full Performance Assurance Test
The following sections describe the full performance testing procedures i.e. detailed testing of each
parameter with a patient simulator or specified tools. Please refer to the section When to perform
Tests for information on when which testing procedure is required.
ECG/Resp Performance Test
This test checks the performance of the ECG and respiration measurements.
Tools required: Patient simulator.
ECG Performance
1Connect the patient simulator to the ECG/Resp connector on the monitor, or connect the
M4841A/M4851A via TAAP cable or Short Range Radio to the monitor and connect the patient
simulator to the ECG/Resp connector.
Service to
2Configure the patient simulator as follows:
–ECG sinus rhythm.
–HR = 100 bpm or 120 bpm (depending on your patient simulator).
3Check the displayed ECG wave and HR value against the simulator configuration.
4The value should be 100bpm or 120 bpm+/- 2 bpm.
Respiration Performance (not available via TAAP or Short Range Radio)
1Change the Patient Simulator configuration to:
–Base impedance line 1500 Ohm.
–Delta impedance 0.5 Ohm.
–Respiration rate 40 rpm or 45 rpm.
69
3 Testing and Maintenance
The value should be 40 rpm +/- 2 rpm or 45 rpm +/- 2 rpm.
2
TestExpected test results
ECG Performance Test100bpm +/- 2bpm or
120bpm +/- 2bpm
Respiration Performance Test40 rpm +/- 2 rpm or
45 rpm +/- 2 rpm
ECG Out Sync Performance Test (not available via TAAP or SRR)
This test checks the performance of ECG synchronization between the monitor and a defibrillator. It
only needs to be performed when this feature is in use as a protocol at the customer site.
Tools required:
•Defibrillator with ECG Sync and Marker Output.
•Patient simulator.
1Connect the patient simulator to the ECG connector and the defibrillator to the ECG Sync
Output on the monitor.
2Set the patient simulator to the following configuration:
–HR = 100 bpm or 120 bpm (depending on your patient simulator).
–ECG sinus rhythm.
3Switch the defibrillator to simulation mode.
4Check that the marker pulse is displayed before the T-wave begins.
TestExpected test results
ECG Out Sync Performance TestMarker pulse is displayed before the
ECG Sync Pulse Performance Test
1Make sure the interface is properly configured. See configuration guide for details
2Provide a clean ECG signal (from patient or simulator) to the monitor.
3Connect the cable to the monitor.
4Marker pulses should show on the screen.
SpO2 Performance Test
T-wave begins
70
This test checks the performance of the SpO2 measurement.
Tools required: none
1Connect an adult SpO2 transducer to the SpO2 connector.
2Measure the SpO
value on your finger (this assumes that you are healthy).
2
The value should be between 95% and 100%.
3
TestExpected test results
SpO2 Performance Test95% and 100%
Measurement Validation
The SpO2 accuracy has been validated in human studies against arterial blood sample reference
measured with a CO-oximeter. In a controlled desaturation study, healthy adult volunteers with
saturation levels between 70% and 100% SaO2 were studied. The population characteristics for those
studies were:
•about 50% female and 50% male subjects
•age range: 18 to 45
•skin tone: from light to black
NOTE
A functional tester cannot be used to assess the accuracy of a pulse oximeter monitor. However, it can
be used to demonstrate that a particular pulse oximeter monitor reproduces a calibration curve that has
been independently demonstrated to fulfill a particular accuracy specification.
3 Testing and Maintenance
NBP PerformanceTest
This section describes NBP test procedures.The monitor must be in service mode and the screen
“Service A” must be selected to perform these tests. The NBP Performance Test consists of:
•NBP Accuracy Test
•NBP Leakage Test
•NBP Linearity Test
•Valve Test
NBP Accuracy Test
This test checks the performance of the non-invasive blood pressure measurement. Connect the
equipment as shown:
Tools required:
•Reference manometer (includes hand pump and valve), accuracy 0.2% of reading.
•Expansion chamber (volume 250 ml +/- 10%)
71
3 Testing and Maintenance
•Appropriate tubing.
In service mode, the systolic and diastolic readings indicate the noise of NBP channels 1 and 2
respectively. When static pressure is applied, the reading in NBP channel 1 should be below 50. The
value in parentheses indicates the actual pressure applied to the system.
1Connect the manometer and the pump with tubing to the NBP connector and to the expansion
chamber.
2In service mode, select the Setup NBP menu.
3Select Close Valves: On
4Raise the pressure to 280 mmHg with the manometer pump.
5Wait 10 seconds for the measurement to stabilize.
6Compare the manometer values with the displayed values.
7Document the value displayed by the monitor (x1).
8If the difference between the manometer and displayed values is greater than 3 mmHg, calibrate
the MMS. If not, proceed to the leakage test.
9To calibrate the MMS, select Close Valves off then Calibrate NBP and wait for the instrument to
pump up the expansion chamber.Wait a few seconds after pumping stops until
highlighted and then move the cursor to the value shown on the manometer. If one of the
following prompt messages appears during this step, check whether there is leakage in the setup:
–NBP unable to calibrate–cannot adjust pressure
–NBP unable to calibrate–unstable signal
EnterPrVal is
10 Press Confirm.
If the INOP NBP Equipment Malfunction message occurs in monitoring mode, go back to service
mode and repeat the calibration procedure.
NBP Leakage Test
The NBP leakage test checks the integrity of the system and of the valve. It is required once every two
years and when you repair the monitor or replace parts.
1If you have calibrated, repeat steps 2 to 6 from the accuracy test procedure so that you have 280
mmHg pressure on the expansion chamber.
2Watch the pressure value for 60 seconds.
3Calculate and document the leakage test value (x2).
x2 = P1 - P2
where P1 is the pressure at the beginning of the leakage test and P2 is the pressure displayed after
60 seconds.
The leakage test value should be less than 6 mmHg.
NBP Linearity Test
1Reduce the manometer pressure to 150 mmHg.
2Wait 10 seconds for the measurement to stabilize.
3After these 10 seconds, compare the manometer value with the displayed value.
72
4Document the value displayed by the monitor (x3)
Valve Test
3 Testing and Maintenance
If the difference is greater than 3 mmHg, calibrate the MMS (see steps 9 to 10 in the accuracy test
5
procedure).
1Raise the pressure again to 280 mmHg.
2Select Close valves: Off.
3Wait five seconds and then document the value displayed. The value should be less than 10 mmHg.
4Document the value displayed by the monitor (x4).
TestExpected test results
Accuracy testx1 = value displayed by monitor
Difference ≤ 3mmHg
Leakage testx2 = leakage test value
x2 < 6 mmHg
Linearity testx3 = value displayed by monitor
Difference ≤ 3mmHg
Valve Testx4 = value < 10 mmHg
Invasive Pressure Performance Test
This test checks the performance of the invasive pressure measurement.
Tools required: Patient simulator.
1Connect the patient simulator to the pressure connector.
2Set the patient simulator to 0 pressure.
3Make a zero calibration.
4Configure the patient simulator as P(static) = 200 mmHg.
5Wait for the display.
6The value should be 200 mmHg ± 5 mmHg. If the value is outside these tolerances, calibrate the
Invasive Pressure measurement. If the measurement was calibrated with a dedicated reusable
catheter, check the calibration together with this catheter.
This test checks the performance of the temperature measurement.
Tools required: Patient simulator (with 0.1°C or 0.2°F).
1Connect the patient simulator to the temperature connector.
2Configure the patient simulator to 40°C or 100°F.
73
3 Testing and Maintenance
The value should be 40°C ± 0.2°C or 100°F ± 0.4°F.
3
TestExpected test results
Temperature Performance Test40°C ± 0.2°C or 100°F ± 0.4°F
Predictive Temperature Accuracy Check
Tools required:
•Calibration Key (CalKey) - Part No. 453564033691
The Calkey can be used to verify correct function of the module (in continuous mode). It does not test
the probe. The CalKey contains a known resistance which is converted to a specific temperature by the
module.
The monitor can stay in monitoring mode for this procedure.
Procedure:
1Disconnect the probe and connect the CalKey.
2Remove the probe from the holder. The monitor software switches to continuous mode
automatically.
3Observe the displayed temperature. The value should read: 97.3°F ±0.2°F (36.3°C ±0.1°C).
Document whether the predictive temperature module passed or failed the accuracy check.
4Disconnect the CalKey, connect the probe and return it to the holder.
5The module test with the CalKey should be performed once a year.
TestExpected test results
Predictive Temperature Accuracy Test97.3°F ±0.2°F (36.3°C ±0.1°C)
The 9600 Plus Calibration Tester from Welch Allyn provides a convenient way of testing the entire
thermometer system, module and probe. It is not intended for use by clinical users. Nevertheless,
biomeds or Philips field personnel may use it for probe verification. Currently, the tester is orderable
from Welch Allyn only. Follow the instructions provided with the tester.
Mainstream CO2 Accuracy Check
Tools Required:
•three airway adapters
•Verification Gas M2506A
•Gas cylinder regulator M2505A
You also need a local barometric pressure rating received from a reliable local source (airport, regional
weather station or hospital weather station) which is located at the same altitude as the hospital.
74
Procedure:
1Attach the M2501A CO
Make sure that the sensor is disconnected from the patient circuit.
2Switch on the patient monitor.
3Enter the monitor’s Service Mode.
sensor to the patient monitor. Attach an airway adapter to the sensor.
2
3 Testing and Maintenance
Using the sensor status provided in the M2501A Serial protocol, wait for the M2501A sensor to
4
warm up to its operating temperature.
5The default setting for gas temperature is 22°C. If the gas temperature is significantly above or
below this value, correct the gas temperature setting.
6Zero the sensor on the airway adapter being used in this test. Ensure Zero Gas is set to Room Air
7Attach a regulated flowing gas mixture of 5% CO2, balance N2 to the airway adapter.
8Set the gas correction to off.
9Allow a few seconds for the gas mixture to stabilize and observe the CO2 value. The expected
value is 5% of the ambient pressure ±2mmHg
NOTE
Make sure that you follow the above steps correctly. If the sensor fails this check it must be exchanged.
The sensor cannot be calibrated.
Example for an expected test result:
The expected test result for an altitude of 0 m (sea level) at approximately 760 mmHg ambient
pressure is:
TestExpected test results (x1)Acceptance Range
Mainstream CO2 Accuracy
5% of 760 mmHg pressure ±2mmHg36 mmHg - 40 mmHg
Test
NOTE
The expected test results will differ depending on the conditions (i.e. altitude or ambient pressure).
Sidestream CO2 Accuracy Check
Tools Required:
•Cal gas flow regulator M2267A
•Cal tube 13907A
•Verification Gas M2506A
•Straight Sample Line M2776A
You also need a local barometric pressure rating received from a reliable local source (airport, regional
weather station or hospital weather station) which is located at the same altitude as the hospital.
Procedure:
1Attach the M2741A CO2 sensor to the patient monitor. Attach the sample line and the cal tube to
the sensor. Make sure that the sensor is disconnected from the patient circuit.
2Switch on the patient monitor.
3Enter the monitor’s Service Mode.
4Using the sensor status provided in the M2741A Serial protocol, wait for the M2741A sensor to
warm up to its operating temperature.
5Zero the sensor. Ensure Zero Gas is set to Room Air
6Attach a regulated flowing gas mixture of 5% CO2, balance N2 to the cal tube.
7Set the gas correction to off.
75
3 Testing and Maintenance
Allow a few seconds for the gas mixture to stabilize and observe the CO2 value. The expected
8
value is 5% of the ambient pressure ±2mmHg
NOTE
Make sure that you follow the above steps correctly. If the sensor fails this check it must be exchanged.
The sensor cannot be calibrated
Example for an expected test result:
The expected test result for an altitude of 0 m (sea level) at approximately 760 mmHg ambient
pressure is:
TestExpected test results (x2)Acceptance Range
Sidestream CO2 Accuracy Test 5% of 760 mmHg pressure ±2mmHg36 mmHg - 40 mmHg
NOTE
The expected test results will differ depending on the conditions (i.e. altitude or ambient pressure).
Sidestream CO2 Flow Check
Check the flow rate in the Sidestream CO2 extension as follows:
1Connect the flowmeter to the sample line
2Check on the flowmeter the flow that the Sidestream CO
ml/min ± 10 ml/min. If the value is not within tolerance check your setup again and perform
another flow check. If it fails again, the sensor must be replaced. The sensor cannot be calibrated.
Example for an expected test result:
The expected test result for an altitude of 0 m (sea level) at approximately 760 mmHg ambient
pressure is:
The expected test results will differ depending on the conditions (i.e. altitude or ambient pressure).
Microstream CO2 Performance Test
Allow five seconds between individual service procedures to ensure stable equipment conditions.
When certain monitor procedures are running, service procedures are not possible and trying to start
them will result in a message
monitor completes the current operation, then restart the service procedure.
This test checks the performance of the Microstream CO2 measurement. The Microstream CO2
measurement can either be integrated into the IntelliVue MP5 monitor or, for other IntelliVue
monitors, into the M3015A MMS Extension. The Microstream CO2 performance test is required once
per year and when the instrument is repaired or when parts are replaced.
Service Operation Failed in the monitor’s status line. Wait until the
extension pump draws. It should be 50
2
min
76
This test uses calibration equipment that you can order (see the Parts section for the part number). The
procedure is summarized in the following steps. Refer to the documentation accompanying the
equipment for detailed instructions.
Tools Required:
•Standard tools, such as screwdriver, tweezers
•Electronic flowmeter, M1026-60144
•Gas calibration equipment:
3 Testing and Maintenance
•Cal 1 gas 15210-64010 (5% CO
•Cal 2 gas 15210-64020 (10% CO
)
2
)
2
•Cal gas flow regulator M2267A
•Cal tube 13907A
•Calibration Line M3015-47301
You also need a local barometric pressure rating received from a reliable local source (airport, regional
weather station or hospital weather station) which is located at the same altitude as the hospital.
The CO2 calibration for the Microstream extension consists of the following steps:
•Leakage check
•Barometric pressure check and calibration, if required.
•Pump check
•Flow check and calibration, if required
•Noise check
•CO2 Cal check and calibration, if required
•CO2 Cal verification using 2nd cal gas
Perform all checks in the same session.
Leakage Check
The leakage check consists of checking the tubing between:
•the pump outlet and the mCO
•the pump inlet and FilterLine inlet.
Check the user’s guide of the flowmeter for details on how to make a correct flow reading.
Part 1
1Go into service mode and select Setup CO2 menu.
2Connect a FilterLine to the Microstream CO
3Check the ambient pressure and the cell pressure shown in the monitor’s status line. The cell
4Connect the flowmeter outlet to the FilterLine inlet using a flexible connecting tube.
5Block the mCO
6If the value is outside the tolerance limits, there is a leakage between the pump outlet and the
outlet and
2
input to start the pump running.
2
pressure should be approximately 20 mmHg lower than ambient pressure.
outlet using your fingertip and observe the flowmeter display. The value on the
2
flowmeter (x1) should decrease to between 0 and 4 ml/min, accompanied by an audible increase in
pump noise. If the value is within the tolerance limits, continue with part 2 of the leakage check.
outlet.
mCO
2
77
3 Testing and Maintenance
Open the MMS Extension or MP5 and check the tubing connections at the pump outlet and the
7
extension gas outlet. If the connections are good, then there is a leakage in the tubing and you
must exchange the MMS Extension or the mCO
Part 2
1Disconnect the flowmeter from the Part 1 setup and connect the flowmeter inlet to the M3015A
gas outlet or the MP5 mCO
2Leave the Filterline connected to the M3015A inlet or the MP5 mCO
3Block the inlet of the FilterLine using your fingertip and observe the flowmeter display. The value
on the flowmeter (x2) should decrease to between 0 and 4 ml/min, accompanied by an audible
increase in pump noise. The cell pressure shown in the status line on the display should decrease to
between 300 and 500 mmHg. Do not block the inlet for longer than 25 seconds as this will lead to
an “Occlusion” INOP. If the value is within the tolerance limits, there are no leakages and the
leakage check is completed; proceed to the pump check.
4If the value is not within the tolerance limits, there is a leakage between the FilterLine inlet and the
pump inlet.
5Check the FilterLine connections and open the M3015A or MP5 to check the tubing connections
at the pump inlet and the M3015A or MP5 mCO
replacing the FilterLine and repeating the leakage check. If the situation remains, there is a leakage
in the tubing and the M3015A or the mCO
gas outlet.
2
Assembly of the MP5 respectively.
2
inlet..
2
gas inlet. If the connections are good, try
2
assembly of the MP5 must be exchanged.
2
Barometric Pressure Check and Calibration
Check the barometric pressure value in the M3015A MMS Extension or the MP5 as follows:
1Go into service mode and select Setup CO
2Connect a FilterLine to the Microstream CO
Extension or the MP5.
3The status line at the bottom of the screen displays “CO
yyy” where xxx is the ambient pressure and yyy is the measured cell pressure. Check whether the
ambient pressure value (x3) matches (within the acceptable tolerance of ±12mm Hg) the reference
value you have received. If so, proceed to the leakage check. If the value is not correct, calibrate as
follows.
CO
a.Select
b. Select the value in the table which matches the reference value received from a reliable local source
(airport, regional weather station or hospital weather station). (The values are displayed with a
resolution of 2 mmHg up to 500 mmHg and a resolution of 1 mmHg from 500 mmHg to 825
mmHg.) Note: the selected value must be within ±10% of the current measured ambient pressure,
otherwise an error message will occur at restarting the monitor.
c. Confirm the barometric pressure setting.
d. Check that the ambient pressure displayed in the status line at the bottom of the screen is the same
as the value which you selected from the list in step b.
then select Barom.Press to activate a table of values.
2
menu.
2
input. This activates the pump in the M3015A MMS
2
pressure reading (ambient/cell) xxx/
2
Pump Check
78
1Connect the flowmeter inlet to the mCO
2Connect the FilterLine to the mCO
inlet.
2
gas outlet.
2
Block the inlet of the FilterLine using your fingertip and observe the cell pressure on the monitor
3
display. The cell pressure (x4) should be more than 120 mmHg below the ambient pressure shown.
If the pressure difference is less than 120 mmHg, the pump is not strong enough and you should
replace it, irrespective of the Pump OpTime.
Flow Rate Check and Calibration
Check the flow rate in the M3015A MMS Extension or the MP5 as follows:
3 Testing and Maintenance
Noise Check
1Connect the flowmeter to the CO
2Check on the flowmeter the flow that the M3015A MMS Extension or MP5 mCO2 pump draws
(x5). It should be 50 ml/min ± 7.5 ml/min. If the value is within tolerance, proceed to the CO
FilterLine.
2
2
Gas calibration check. If the value is not within tolerance, calibrate as follows.
3Adjust the flow in the instrument by selecting Increase Flow or Decrease Flow until it is as close
as possible to 50 ml per minute as indicated on the flowmeter gauge.
4When you are satisfied that the flow is set as close as possible to 50 ml per minute, select Store
Flow
and confirm the setting. If you do not store the adjusted flow within 60 seconds of the
adjustment, the old flow setting is restored.
5If you cannot adjust the flow to within tolerance, replace the pump . If you still cannot make the
flow adjustment, this indicates a fault in the measurement extension, which must be replaced.
Note that the pump can only be replaced on M3015A with the old hardware Rev. A (i.e. Serial No.
DE020xxxxx
1With the monitor in service mode, select Setup CO
2Disconnect the flowmeter and connect the 5% calibration gas and flow regulator in its place.
3Open the valve to apply the 5% calibration gas and wait until the value is stable.
4Check the noise index (x6) displayed next to the CO
of noise on the CO
wave). If the value exceeds 3 mmHg, replace the measurement extension.
2
menu.
2
value on the display (this indicates the level
2
CO2 Gas Measurement Calibration Check
After switching the measurement extension on, wait at least 20 minutes before checking the
calibration. Check the calibration of the CO
1Check that the 5% calibration gas and flow regulator are connected.
2Calculate the expected measurement value in mmHg as follows:
0.05 x (ambient pressure) = value mmHg
for example 0.05 x 736 = 36.8 mmHg (with an ambient pressure of 736 mmHg)
3Open the valve on the flow regulator to allow 5% CO
value to stabilize.
4Check that the value on the instrument (measurement value on the main screen, x7) matches the
calculated mmHg value ± 2.6 mmHg. If the value is outside the tolerance, calibrate as described in
step in this procedure onwards.
5Disconnect the 5% calibration gas and connect the 10% calibration gas.
gas measurement as follows:
2
gas to flow into the extension. Allow the
2
79
3 Testing and Maintenance
Calculate the expected measurement value and tolerance in mmHg as follows:
6
0.1 x (ambient pressure) = value mmHg
±0.07 x (value mmHg) = tolerance
for example 0.1 x 737 mmHg = 73.7 mmHg (with an ambient pressure of 737 mmHg)
±0.07 x 73.7 mmHg = ±5.16 mmHg tolerance
7Open the valve on the flow regulator to allow 10% CO
value to stabilize.
8Check that the value on the instrument (x8) matches the calculated mmHg value within the
calculated tolerance. If so, the measurement extension is correctly calibrated. If the value is outside
the tolerance, calibrate as follows.
9If not already connected, connect the 5% calibration gas.
10 Select Cal. CO
11 Select the value for the calibration gas. (The default value is 5.0%.)
12 Open the valve on the calibration gas to allow CO
to stabilize before the start of the calibration. Leave the valve open until the instrument gives a
prompt that gas can be removed.
13 The extension calibrates and prompts when calibration is successful.
Calibration Verification
1Reopen the 5% gas valve and allow the value to stabilize.
2Check that the value displayed on the monitor is correct within the tolerance (see step above).
3Disconnect the 5% calibration gas and connect the 10% calibration gas.
4Open the valve on the flow regulator to allow 10% CO2 gas to flow into the extension. Allow the
value to stabilize.
gas to flow into the extension. Allow the
2
.
2
gas to flow into the extension. Allow the value
2
5Check that the value displayed on the monitor is correct within the tolerance (see step above).
If one or both values are not within tolerances, you must exchange the M3015A MMS Extension or
the MP5 mCO
Assembly.
2
TestExpected Test Results
Leakage Check parts
1 and 2
x1 = value of part 1 leakage check on flowmeter
(x1< 4.0 ml/min)
x2 = value of part 2 leakage check on flowmeter
(x2< 4.0 ml/min)
Barometric Pressure
Check
x3 = difference between the reference pressure and the
measured ambient pressure displayed on the monitor
(x3<12 mmHg)
Pump Checkx4 = difference in pressure between cell pressure and
ambient pressure displayed on the monitor during
occlusion (x4 >120 mmHg)
80
TestExpected Test Results
Flow Checkx5 = difference between measured value and 50.0 ml/min
Noise Checkx6 = noise index displayed on monitor (x6<3.0)
CO
Gas Calibration
2
Check
CO2 Cal Verification x8 = difference between measured CO2 value and
Reset Time Counters
NOTE
This procedure only applies to M3015A with the old hardware Rev. A (i.e. Serial No. DE020xxxxx
3 Testing and Maintenance
(x5<7.5 ml/min)
x7 = difference between measured CO2 value and
calculated value, based on 5% CO
cal. gas. (x7 < 2.6
2
mmHg)
calculated value, based on 10% CO
cal. gas.
2
(x8 < ± {0.07 x value calculated})
You must check the time counters on the Microstream CO
extension before calibrating the
2
instrument. As well, when parts are replaced, the appropriate counters must be reset to zero.
The counters for CO2 pump, IR Src and Last Cal are displayed in the status line. The values are
updated when entering the
Setup CO2 menu.
Observe the following guidelines:
•When calibrating the CO
Reset PumpOpTime and Reset IRSourceTime selections to make sure that they are within
extension, if no parts have been replaced, check the displayed values of
2
suggested guidelines for use (15, 000 hours of continuous use). If the counter time is greater than
15, 000 hours, replace the appropriate part. See Repair and Disassembly for details.
•When calibrating the CO
the
Reset PumpOpTime and Reset IRSourceTime selections. See Repair and Disassembly for details.
Resetting the PumpOpTime generates the INOP: “CO
extension, if parts have been replaced, reset the appropriate values using
2
OCCLUSION”. To clear this INOP you
2
must perform a flow check and store the flow in service mode (select
Nurse Call Relay Performance Test
The nurse call relay performance test can be performed either at the phone jack type connector (this
only tests one relay) or at the multi-port nurse call connector (to test all three relays).
Phone Jack Type Connector Test (Traditional Nurse Call)
This test checks the operation of the traditional Nurse Call Relay. The Nurse Call Relay test is
recommended for customer sites where the nurse call is in use. The Nurse Call relay functions as
follows:
Store Flow).
•Standard Operation—Relay open.
•Alarm Condition—Relay closed.
Tools required: Ohmmeter.
1Plug a phono connector into the Nurse Call Relay connector.
81
3 Testing and Maintenance
Connect the ohmmeter.
2
3If no alarm occurs, the relay contacts are open. When an alarm occurs, the relay contacts close.
4The expected test result is: Alarm condition - Relay closed.
Some customers may have an Open-On-Alarm relay instead of a Closed-On-Alarm for their Nursecall
system.
The modified Nurse Call relay functions as follows:
•Standard Operation—Relay closed.
•Alarm Condition—Relay open.
Tools required: Ohmmeter.
1Plug a phono connector into the Nurse Call Relay connector.
2Connect the ohmmeter and verify the above conditions.
3If no alarm occurs, the relay contacts are closed. When an alarm occurs, the relay contacts open.
4The expected test result is: Alarm Condition - Relay open.
TestExpected test results
Nurse Call Relay Performance TestAlarm Condition—Relay open
Power Loss Alarm Buzzer Performance Test
1Switch on the monitor.
2Remove the battery and disconnect the monitor from AC power.
3The Power Loss Alarm Buzzer should beep for about one minute.
82
3 Testing and Maintenance
To switch off the alarm sound, either press the power button, connect the monitor to AC power or
4
insert a battery
TestExpected test results
Power Loss Alarm Buzzer Performance
Beep for one minute
Test
IntelliVue 802.11 Bedside Adapter Communication Test (not for MP5T)
1Make sure the LAN cable is disconnected from the rear of the monitor, then switch on the
monitor.
2Go into Service Mode and select Main Setup -> Network -> Setup WLAN. In the Setup WLAN
menu:
–set
–set
–set the
–set the
3Select Main Setup -> WLAN Diagnostic to access the service window.
4Proper installation of the IntelliVue 802.11 Bedside Adapter is assured by connecting to an access
point over the wireless link. Place the monitor with the IntelliVue 802.11 Bedside Adapter installed
in close proximity to the access point (e.g. if the access point is mounted on the ceiling, place the
monitor directly below). Wait until the
(for Rel. C.0 monitors)or Connected (for Rel D.0 or higher). Take the monitor approximately 5 m
away from the access point. There should be no walls or other obstacles between the monitor and
the access point. The following should apply:
–Observe the
Mode to either 802.11Ah, 802.11G, 802.11Bg (not recommended), Auto (not
recommended) or
None (this setting disables the wireless LAN functionality permanently), to
match your wireless infrastructure installation.
SSID to match your installation.
Country code to “1000”. Setting the country code to this value will automatically adjust
the regulatory domain to match the configuration of the infrastructure. Do not set the country
code to values other than “1000” unless otherwise instructed.
Security Mode to WPA(PSK) and enter the WPA password (string between 8 and 63
characters).
Conn.Status field in the service window shows Authenticatd
RSSI (Received Signal Strength Indicator) value for at least 5 - 10 seconds. The
RSSI value wil fluctuate but should stay above 30 in a 5 m distance from the access point used.
The wireless link should be active, i.e. the
Conn.Status field should be Authenticatd (for Rel.
C.0 monitors)or Connected (for Rel D.0 or higher), and the other fields should contain values. If
RSSI value is significantly lower, check the distance to the access point and the antenna
the
orientation at the monitor. The antenna orientation should be vertical, but the physical
placement of the monitor or other equipment within its vicinity as well as walls or other
obstacles may influence the antenna orientation required to receive the best RSSI value.
5If this test fails, retry in a different physical area with a different access point.
6Perform the Wireless Switch test blocks as described in the Philips IntelliVue 802.11 a/g
Infrastructure Installation and Configuration Guide.
TestExpected test results
IntelliVue 802.11 Bedside Adapter
RSSI value above 30
Performance Test
83
3 Testing and Maintenance
IIT Communication Test (not for MP5T)
1Make sure the LAN cable is disconnected from the rear of the monitor, then switch on the
monitor.
2Go into Configuration mode and, in the Network menu, set the RF Access Code in each profile to
match your installation.
3Go into Service Mode. Select Main Setup -> Instr. Telemetry to access the Instrument Telemetry
Service window.
4Proper installation of the IIT module is assured by connecting to an access point over the wireless
link. Place the monitor with the IIT module installed in close proximity to the access point (e.g. if
the access point is mounted on the ceiling, place the monitor directly below). Wait until the
Conn.Status field in the Instrument Telemetry Service window shows Active. Take the monitor
approximately 5 m away from the access point. There should be no walls or other obstacles between the monitor and
the access point. The following should apply:
–Observe the RSSI (Received Signal Strength Indicator) value for at least 5 - 10 seconds. The
RSSI value should be around -50 ±10 in a 5 m distance from the access point used and the IIT
link should be active, i.e. the
contain values. If the
and the antenna orientation at both the monitor and the access point (both should be vertical).
–Remove the antenna. The
the connection could be unreliable. The
Seeking. If the difference between the
significantly lower, check the antenna and the antenna connector for damage and verify that
the cable from the IIT adapter to the antenna connector plate is connected properly.
Conn.Status field should be Active and the other fields should
RSSI value is significantly lower, check the distance to the access point
RSSI value should be around -90 ±10. The IIT link may be active but
Conn. Status field may toggle between Inactive and
RSSI values measured with and without antenna is
5If this test fails, retry in a different physical area with a different access point.
Error Conditions:
–The field
MAC Instr. Tele should show a value unequal to 0000 0000 0000. If it does not,
there is a communication problem between the monitor and the IIT adapter.
–With an incorrect RF Access Code or an incorrect or defective antenna installation, the fields
IP Address,Server IP, Subnet Mask, and RSSI in the Instrument Telemetry Service window
will stay blank. The field
6Perform the Access Point Controller (APC) test blocks as described in the Philips IntelliVue
Conn. Status will slowly toggle between Inactive and Seeking.
Wireless Network Installation and Configuration Guide.
Short Range Radio (SRR) Performance Test
1Make sure that the short range radio interface is configured as follows: SRR On and appropriate
channel selected.
2Assign a telemetry transceiver to the IntelliVue Monitor according to the procedure described in
the Instructions for Use of the patient monitor.
3Check that the following conditions are fulfilled:
a. Place the telemetry transceiver close to the monitor.
b. The telemetry transceiver status is displayed on the monitor in the measurement selection
window.
c. Waves or numerics from the telemetry transceiver are displayed on the monitor. There a re no
dropouts or gaps in waves or numeric transmission.
84
d. The battery status of the telemetry transceiver is displayed in the measurement selection
window.
e. The Signal Quality Indicator shows at least
4Check that the data from the telemetry transceiver is transmitted to the monitor within a 1m radius
and that there are no dropouts or gaps in waves or numerics.
5Check whether the connection remains stable within a 5m radius from the monitor.
6Switch on all telemetry transceivers used on the site and check that there are no interferences
between the transceivers and their assigned monitors.
7Check and record the coverage area of the telemetry transceivers and inform the customer about
this coverage area.
Reporting of Test Results
Philips recommends all test results are documented in accordance with local laws. Authorized Philips
personnel report test result back to Philips to add to the product development database. While hospital
personnel (biomedical engineers or technicians) do not need to report results to Philips, Philips
recommends that they record and store the test results in accordance with local laws.
3 Testing and Maintenance
The following table lists what to record after completing the tests in this chapter. Record the results in
the empty column in Table 16.
The following is a guide as to what your documentation should include:
•Identification of the testing body (for example, which company or department carried out the
tests).
•Name of the person(s) who performed the tests and the concluding evaluation.
•Identification of the device(s) and accessories being tested (serial number, etc.).
•The actual tests (incl. visual inspections, performance tests, safety and system tests) and
measurements required
•Date of testing and of the concluding evaluation.
•A record of the actual values of the test results, and whether these values passed or failed the tests.
•Date and confirmation of the person who performed the tests and evaluation.
The device under test should be marked according to the test result: passed or failed.
85
3 Testing and Maintenance
Carrying Out and Reporting Tests
Test Report
Testing Organization:
Name of testing person:
Responsible Organization:
Device Under Test:ID-Number
Product Number:Serial No.:
Accessories:
Measurement Equipment (Manufacturer, Type, Serial
No.):
Functional Test (parameters tested):
(Check one of the following three options)
Test before putting into service (reference value)
Recurrent Test
Test after Repair
86
Test and Inspection Matrix
3 Testing and Maintenance
TestTest or Inspection
to be Performed
Visual
Inspection
Power OnPower on the unit.
Noninvasive
Blood Pressure
Performance
Tests
Temperature
Performance
Test
All other
performance
tests
Safety (1)Perform Safety Test
Safety (2)Perform Safety Test
Safety (3)Perform Safety Test
Perform Visual
Inspection
Does the self-test
complete
successfully
Perform the
Accuracy Test
Performance
Leakage Test
Performance
Linearity Test
Performance Valve
Test
Perform the
Temperature
Performance Test
Perform the
remaining
parameter
performance tests,
if applicable
(1): Protective
Earth Resistance
(2): Equipment
Leakage Current Normal Condition.
(3): Equipment
Leakage Current Single Fault
Condition (Open
Earth)
Expected Test ResultsRecord the Results (mandatory for
Philips Personnel only)
What to recordActual Results
Pass or FailV:P or V:F
If Yes, Power On test is passedPO:P or PO:F
X1 = value displayed by monitor
Difference <= 3mmHg
X2 = leakage test value
X2 < 6 mmHg
X3 = value displayed by monitor
Difference <= 3mmHg
X4 = value < 10 mmHgPN:P/X4 or
X1= 40°C ± 0.2°C or 100°F ±
0.4°F
See expected results in test
procedures
With mains cable:
Maximum impedance (X1):
<=300 mOhms
With mains cable:
Maximum leakage current
(X1):<= 100 μA
With mains cable:
Maximum leakage current
(X2):<= 300 μA
PN:P/X1 or
PN:F/X1
PN:P/X2 or
PN:F/X2
PN:P/X3 or
PN:F/X3
PN:F/X4
PT: P/X1 or
PT: F/X1
P: P or
P: F
S(1):P/X1 or
S(1):F/X1
S(2): P/X1 or
S(2): F/X1
S(3): P/X2 or
S(3): F/X2
87
3 Testing and Maintenance
TestTest or Inspection
Expected Test ResultsRecord the Results (mandatory for
to be Performed
Safety (4)Perform Safety Test
(4): Patient Leakage
Maximum leakage current (X1):
<=50 μA
Current - Single
Fault Condition,
mains on applied
part.
System
(Sys 1-2)
System
(Sys 3)
Perform the system
test according to
subclause 19.201 of
IEC/EN 60601-11, if applicable, after
forming a system
Perform the system
test according to
subclause 19.201 of
IEC/EN 60601-11, if applicable, after
forming a system
Equipment Leakage Current:
Sys1 <= 100 μA (Normal
Condition)
Sys2 <= 300μA (Single Fault
Condition
Protective Earth Leakage
Current if medical electrical
system components are
connected to the same Multiple
Portable Socket Outlet:
Sys3 <= 300 μA
Key: P = Pass, F = Fail, X or Sys = test value to be recorded
Philips Personnel only)
What to recordActual Results
S(4): P/X1 or
S(4): F/X1
Sys: PSys1/PSys2
or
Sys: FSys1/Fsys2
Sys: PSys3
or
Sys: FSys3
NOTE
All values for current and voltage are the root mean square (r.m.s.) values, unless otherwise stated.
Evaluation
Safety and Functional Test passed
Repair required at a later date, safety and functional test passed
Device must be taken out of operation until repair and passed tests
The evaluation of the test results must be performed by appropriately trained personnel with sufficient
product, safety testing and application knowledge.
If any test results are between 90% and 100% of the respective expected result, the previously
measured reference values must be taken into consideration for the assessment of the electrical safety
of the device under test. If no reference values are available, you should consider shorter intervals
between upcoming recurrent tests.
3 Testing and Maintenance
NOTE
If any single test fails, testing must be discontinued immediately and the device under test must be
repaired or labeled as defective. Be sure to inform the user about the test failure in writing.
Other Regular Tests
The care and cleaning requirements that apply to the monitor and its accessories are described in the
Instructions for Use. This section details periodic maintenance procedures recommended for the
monitor and its accessories.
Touchscreen Calibration
To access the touchscreen calibration screen:
1Enter service mode
2Select Main Setup
3Select Hardware
4Select Touch Calibration
89
3 Testing and Maintenance
Figure 7 Touchscreen Calibration Screen
Make sure you complete the calibration procedure without powering off the monitor mid-way. If the
monitor is powered off after the first point is touched, the touch panel will be deactivated until the
touch calibration is performed again.
If the touchscreen is accidentally mis-calibrated by selecting the wrong spot, you must use another
input device to re-enter calibration mode. If you have the support tool, you can select
Calibration to Default
calibration menu again via the touchscreen.
Please refer to the documentation shipped with your selected display for further details on touchscreen
calibration procedures.
and it will create a rough calibration which will allow you to access the
Disabling/Enabling Touch Operation
To temporarily disable touchscreen operation of the monitor, press and hold the Main Screen key. A
padlock symbol will appear on the key. Press and hold the
touchscreen operation.
Main Screen key again to re-enable
Printer Test Report
To verify your printer configuration you may want to print a test report.
To print a test report select
Your test report should look like this:
Main Setup -> Reports -> Setup Printers -> Print Test Rep.
Reset Touch
90
3 Testing and Maintenance
Battery Handling, Maintenance and Good
Practices
This section provides some information on how to handle and maintain the battery in order to get the
best usage from them. Additionally, some good working practices are also given regarding the correct
disposal of the batteries. This section only applies if a system interface board with battery functionality
is installed in the monitor.
NOTE
If your monitor is connected to an IntelliVue Patient Monitors Information Center, you should make
sure that the IIC uses the text catalog revision E.0 or later, otherwise battery INOPs may not display
correctly on the IIC. Consult your IIC documentation for instructions on upgrading the text catalog.
About the Battery
The rechargeable Lithium-Ion battery used in the monitor is regarded as a Smart battery because it has
built-in circuitry. (This circuitry communicates battery-status information to the Monitor.)
To get the most out of the batteries, observe the following guidelines:
•Condition the battery only upon maintenance request prompt on display.
•If a battery shows damage or signs of leakage, replace it immediately. Do not use a faulty battery in the Monitor.
•Capabilities of integrated battery charger: 12.6V, 2.5 Amps max.
Actual current / voltage: depends on smart battery request and monitor configuration
The approximate charging time is 4 hours with the monitor switched off and up to 12 hours or
more during monitor operation, depending on the monitor configuration.
NOTE
In certain situations, where many measurements are in use plus the recorder, the load on the monitor
may be so high that the batteries will not charge. In this case you must use the M8043A Smart Battery
Charger to charge the battery.
91
3 Testing and Maintenance
•Battery Disposal—The battery should be disposed of in an environmentally-responsible manner.
Consult the hospital administrator or your local Philips representative for local arrangements.
Do not dispose of the battery in normal waste containers.
•Battery Storage — Batteries should not remain inside the monitor if they are not used for a
longer period of time. Batteries should be max. 50% charged for storage.
NOTE
The battery will discharge over time if it is stored inside the monitor without AC power connection.
The reported values “remaining capacity” and “runtime” will become less accurate when the battery is
stored inside the monitor without AC Power connection for a longer period of time (i.e. several
weeks).
Checking the Battery Status
When the Monitor is connected to the AC power supply, the battery charges automatically. The battery
can be charged remotely from the Monitor by using the battery charger. Use only the M8043A Smart
battery charger.
Battery status (level of charge) is indicated in several ways:
•LED on the front panel of the Monitor.
•Battery gauge.
•Display of battery time below gauge.
•Battery status window.
•INOP messages.
The AC Power LED is only on when the power cord is connected and AC power is available to the
Monitor. In this case, the battery can be either charging or fully charged.
The battery LED can be green, yellow, or red depending on the following conditions:
Battery LED ColorsIf the monitor is connected to
AC power, this means
Green
Yell ow
Red, flashing
Red, flashes
intermittently
battery full (≥90%)
battery charging
(battery power < 90%)
battery or charger
malfunction1,2
If the monitor is running on
battery power, this means
≤ 10 minutes power remaining
battery or charger
malfunction1,2
92
1 indicated by malfunction symbol and INOP
2 for further details see Troubleshooting section
Battery Status on the Main Screen
Battery status information can be configured to display permanently on all Screens. It shows the status
of the battery and the battery power and battery time remaining. The battery time is only displayed
when the monitor is not running on AC power. Note that the battery status information may take a
few minutes after the monitor is switched on to stabilize and show correct values.
Battery power gauge:
This shows the remaining battery power. It is divided into sections, each representing 20% of the total
power. If three and a half sections are shaded, as in this example, this indicates that 70% battery power
remains. If no battery is detected, the battery gauge is greyed out.
Battery malfunction symbols:
If a problem is detected with the battery, these symbols are displayed. They may be accompanied by an
INOP message or by a battery status message in the monitor information line (if battery window is
open) providing more details.
Battery Status Symbols
3 Testing and Maintenance
Battery requires maintenance
Battery is empty
Battery not charging as the temperature is above or below the specified range
Charging stopped to protect the battery
Battery Malfunction Symbols
Incompatible Battery
Battery Malfunction
Battery temperature too high
Battery has no power left
Explanations of Battery Status and Malfunction Symbols:
Battery requires maintenance: The battery requires conditioning. Refer to “Conditioning Batteries” for
details.
Battery is empty: The capacity of the battery is ≤10 min. Recharge the battery as soon as possible.
93
3 Testing and Maintenance
Temperature outside specified range: The charging of the battery is stopped if the temperature is below 15°C
or above 50°C in order to protect the battery. Charging is resumed as soon as the temperature is within
this range.
Incompatible Battery: The inserted battery is checked for certain battery internal parameters. If these are
not correct, the incompatible battery symbol is displayed. Please use only M4605A batteries with the
MP5 monitor. Note that the incompatible battery symbol may also appear if there is a communication
problem between the battery and the battery board.
Battery Malfunction:Communication between the battery and the battery board could not be established
within about 4 minutes or battery internal data indicates malfunction. Please see the
“Troubleshooting” section for remedies.
Battery Temperature too high: This symbol is displayed if the battery temperature goes above 65°C. In
addition the INOP message CHECK BATT TEMP is displayed. If the battery temperature increases
further above 70°C the batteries will switch off for safety reasons. Allow the battery to cool down to
avoid the monitor switching off.
Battery has no power left: If the monitor is not running on AC power: battery will switch off power
delivery at any moment - in this case recharge the battery immediately - or, if the monitor is running on
AC power, the battery is in deep discharge and requires pre-charging to restore communication. To
avoid this condition charge batteries to 50% for storage. Note that the battery malfunction INOP will
eventually be issued if the pre-charging does not restore battery communication within about 4
minutes.
Battery Status Window
♦To access the Battery Status window and its associated pop-up keys, select the battery status
information on the Screen, or select
Capacity, Remaining tells you how much power is left in the battery.
Capacity, Full Charge tells you how much power the battery can hold when fully charged.
Time To Empty tells you approximately how long you can continue to use the monitor with this
battery. Note that this time fluctuates depending on the system load (how many measurements and
recordings you carry out), the age of the battery, and the remaining capacity of the battery. The time
indication appears after AC has been unplugged for about 30 seconds (after finishing calculation of the
Time to Empty)
Time To Full is shown in place of Time To Empty if the monitor is connected to AC power, and tells
you how much time is left until the battery is charged to 90%. Please allow indication to stabilize for 3
to 5 minutes after beginning the charging cycle. If the battery is charged over 90%
is displayed until they are charged to 100%. Then
Main Setup -> Battery.
Battery Full (>90%)
Batt Fully Charged is displayed.
94
Viewing Battery Details
♦To view detailed information for the battery, select the pop-up key Batt.
Documenting Battery Status
To print all battery information in the Battery Status window,
1Select the battery status information on the Screen or select Main Setup -> Battery to open the
Battery Status window
2Select the Record Status pop-up key to print the information on a recorder
or
Select the
Print Status pop-up key to print the information on a connected printer.
3 Testing and Maintenance
Conditioning a Battery
What is Battery Conditioning?
Battery conditioning recalibrates the battery to ensure that it has accurate information on the actual
battery capacity.
Why is Battery Conditioning Necessary?
The capacity of a battery decreases gradually over the lifetime of a battery. Each time a battery is
charged its capacity decreases slightly. Therefore, the operating time of a monitor running on batteries
also decreases with each charge cycle.
Battery conditioning ensures that the value stored in the battery for its full capacity takes account of
this decrease, so that the remaining battery charge can be calculated accurately, and the low battery
warning given at the right time.
When Should Battery Conditioning be Performed?
Battery conditioning should be performed when indicated by the Battery Status.
NOTE
When the battery status signals a conditioning request, the displayed Time to Full or Time to Empty
may not be reliable.
What Causes the Conditioning Message on the Monitor?
In addition to the value for the full capacity, the battery also stores a value for the Max Error. The Max
Error tracks the maximum possible deviation of the estimated charge of a battery from the actual
charge.
95
3 Testing and Maintenance
If a battery is charged or discharged partially, or if it is charged while the monitor is being used, the
accuracy of the “reference points” for the fully discharged and fully charged states decreases, causing
an increase in the value for the Max Error (see diagram, below).
When the Max Error rises over a certain limit, a message is displayed prompting the user to condition
the battery, as described in “Conditioning Batteries” on page 96.
You can reset the value for the Max Error before the battery needs conditioning, by performing the
steps described in”Conditioning Batteries”. The minimum value of the Max Error after conditioning is
2%.
Conditioning Batteries
Battery conditioning can either be performed in the monitor or with an external battery charger.
Philips recommends using the M8043A Smart Battery Charger to condition batteries.
Battery Conditioning in the Monitor
CAUTION
Do not use a monitor being used to monitor patients to condition batteries. The monitor switches off
automatically when the battery is empty.
You should condition a battery when its “battery requires maintenance” symbol shows on the Screen.
If conditioning is not performed immediately the monitor will still function according to specifications.
However, the displayed time to empty and time to full will show increasing inaccuracy. Do not
interrupt the charge or discharge cycle during conditioning. To condition a battery,
1Insert the battery into a monitor connected to mains power.
96
Charge the battery until it is completely full. Switch the monitor off to decrease the charging time
2
When the battery LED turns green i.e. the battery is >90% charged, switch on the monitor and
open the
3Disconnect the monitor from AC power, and let the monitor run until the battery is empty and the
monitor switches itself off.
4Reconnect the monitor to AC power and charge the battery until it is full for use or charge to 50%
for storage.
Battery Status window. Check that the Battery fully charged message is displayed.
Battery Conditioning with an External Charger
You can use the M8043A Smart Battery Charger for external battery conditioning. For details please
see the IfU for the Smart Battery Charger. Use only the M8043A Smart battery charger.
After Installation, Testing or Repair
Before handing the patient monitor over to the end-user, make sure it is configured appropriately and
that it is in monitoring mode. Ensure that the user receives the current revision of the monitor
documentation.
3 Testing and Maintenance
97
3 Testing and Maintenance
98
4Troubleshooting
Introduction
This section explains how to troubleshoot the monitor if problems arise. Links to tables that list
possible monitor difficulties are supplied, along with probable causes, and recommended actions to
correct the difficulty.
How To Use This Section
Use this section in conjunction with the sections Testing and Maintenance and Parts. To remove and
replace a part you suspect is defective, follow the instructions in the section Repair and Disassembly. The
Theory of Operation section offers information on how the monitor functions.
4
Who Should Perform Repairs
Only qualified service personnel (biomedical engineers or technicians) should open the monitor
housing, remove and replace components, or make adjustments. If your medical facility does not have
qualified service personnel, contact Philips’ Response Center or your local Philips representative.
WARNING
High Voltage - Voltages dangerous to life are present in the instrument when it is connected to the
mains power supply. Do not perform any disassembly procedures (other than server removal) with
power applied to the instrument. Failure to adhere to this warning could cause serious injury or death.
Replacement Level Supported
The replacement level supported for this product is to the printed circuit board (PCB) and major
subassembly level. Once you isolate a suspected PCB, follow the procedures in the Repair and Disassembly section, to replace the PCB with a known good PCB. Check to see if the symptom
disappears and that the monitor passes all performance tests. If the symptom persists, swap back the
replacement PCB with the suspected malfunctioning PCB (the original PCB that was installed when
you started troubleshooting) and continue troubleshooting as directed in this section.
99
4 Troubleshooting
Software Revision Check
Some troubleshooting tasks may require that you identify the Software Revision of your monitor. You
can find the software revision along with other information, such as the system serial number, in the
monitor revision screen.To access the monitor revision screen:
1Enter the Main Setup menu and select Revision
2Select Product
3Select Software Revision
4Select the pop-up key for the device you want to check (e.g. M8105A)
NOTE
The part numbers listed in the monitor revision screen do not necessarily reflect the part numbers
required for ordering parts. Please refer to the Parts section for the ordering numbers.
NOTE
The system serial number can also be found on the back of the monitor.
Compatibilty of MP5 in Companion Mode with IntelliVue Patient
Monitors
The following table shows the compatibility between IntelliVue Patient Monitor and MP5 software
revisions when MP5 is used in Companion Mode as an MMS.
Host
Monitor
Software
F. 0Ye sYe s
G.0YesYes
MP5 Software
F. 0G.0
Compatibility with Information Center
The following tables show the compatibility between the MP2/X2/MP5 and the Information Center
software revisions. The first table shows the compatibility if MP2/X2/MP5 are used as pure monitor
or measurement module. The second table shows the compatibility if the MP2/X2/MP5 are used in
companion mode i.e. as monitor and measurement module.
MP5/
MP2/X2
Software
E.0NoNoNoYesYesYesYesYesYes
F. 0N oNoNoNoYesYesYe sYesYes
G.0NoNoNoNoYesYesYesYesYes
Information Center Software
D.01E.0E.01F. 0G.0H.0J.0K.0L.0
100
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.