Connecting Gas Analyzer Accessories5
Using the Gas Analyzer Setup Menus7
Choosing Numerics for Display7
Humidity Correction for CO
Adjusting Wave Scales8
Changing the Apnea Alarm Delay9
Deriving Limit Alarms from awRR9
Alarms and Zero Calibration9
Automatic Alarm Suppression9
Agent Identification10
If Agent ID is Set to Manual10
If Agent ID is Set to Auto (M1019A IntelliVue G5 & M1026B AGM)10
Exchanging Agents (M1019A IntelliVue G5 & M1026B AGM)11
Agent ID During Emergence from Anesthesia
(M1019A IntelliVue G5 & M1026B AGM)
MAC Calculation11
Uncorrected MAC12
Ambient Pressure Corrected MAC (not available in the USA)12
Enhanced MAC Correction (not available in the USA)13
Removing Gas from the Circuit15
Returning the Gas Sample15
Removing the Gas Sample15
Entering Gas Analyzer Standby Mode16
Zero Calibration17
Automatic Zero Calibration17
M1013A IntelliVue G1 and M1019A IntelliVue G517
M1026B AGM17
Carrying Out Manual Zero Calibration17
Suppressing Zero Calibration17
Using the Gas Analyzer During a Cardiopulmonary Bypass18
Safety Information18
2
11
8
i
2 Maintenance and Troubleshooting21
Inspecting the Equipment and Accessories21
Inspecting the Cables and Cords21
Maintenance Task and Test Schedule21
Troubleshooting23
Disposing of the Gas Analyzer23
Disposing of Empty Calibration Gas Cylinders23
The M1013A IntelliVue G1, the M1019A IntelliVue G5 and the M1026B Anesthetic Gas Module
(AGM), hereafter referred to as the “gas analyzers”, measure patients’ anesthetic and respiratory
gases.The following table shows the main features and the patient monitor compatibility for the three
gas analyzers:
Automatic
Gas Analyzer
Agent ID
Compatibility
Number of Gases measured
M1013A IntelliVue G131n/aIntelliVue MP20/30/40/50
M1019A IntelliVue G5322 out of 5IntelliVue MP20/30/40/50/60/70/80/90
M1026B AGM311 out of 5IntelliVue MP40/50/60/70/80/90 and
The gas analyzers measure the Airway Respiration Rate (awRR) and provide end tidal (et) and inspired
(in) values for the following gases:
Respiratory GasesAnesthetic Agents
Carbon dioxide (CO2)Halothane
Nitrous oxide (N
Oxygen (O
The gas analyzers must only be used by qualified personnel.
O)Isoflurane
2
) Enflurane
2
Number of Agents
measured
Sevoflurane
Desflurane
Philips ACMS; Philips V24/26
1
1 Using the Gas AnalyzerUnderstanding the Gas Analyzer Display
Understanding the Gas Analyzer Display
The gas analyzers can send waves and numerics for all measured gases for display on the monitor
screen. This example shows the CO
configured to look different.
, O2, and N2O waves and numerics. Your display may be
2
CO2
O2
N2O
etCO2
imCO2
etN2O
inN2O
awRR
etO2
inO2
rpm
M1013A IntelliVue G1 & M1019A IntelliVue G5
Major Parts and Keys
Setup LED
Watertrap
Standby LED
Power on/off
Standby Key
Setup Key
The setup LED lights when the Setup Gas Analyzer menu is open, when the module is first
switched on (for 5 - 10 seconds), and if there is a problem with the communication between the gas
analyzer and the monitor. The standby LED lights up when the gas analyzer is in standby.
2
Quick Mount Release
Power LED
M1026B Major Parts and Keys1 Using the Gas Analyzer
The RJ-45 connector is the interface connector for the Philips IntelliVue patient monitors.
WARNING• The M1013A IntelliVue G1 may only be used with the Philips IntelliVue MP20/30/40/50 patient
monitors. Connections to other devices may result in a safety hazard.
• The M1019A G5 may only be used with the Philips IntelliVue MP20/30/40/50/60/70/80/90
patient monitors. Connections to other devices may result in a safety hazard.
Equipotential Grounding Terminal
RJ-45
Gas Outlet
Make sure that the anesthetic gas outlet at the rear of the module is connected to the gas scavenging
system or the gas return line.
See your gas analyzer’s service guide for further information on connecting devices.
M1026B Major Parts and Keys
irway Gases LED
irway Gases Key
Power On/
Off switch
Power LED
Wat er tr ap
The setup airway gases LED lights when the Setup Gas Analyzer menu is open, when the module
is first switched on (for 5 - 10 seconds), and if there is a problem with the communication between the
M1026B and the monitor.
3
1 Using the Gas AnalyzerM1026B Major Parts and Keys
M1026B Rear Panel
Gas Outlet
Make sure all devices connected to the RS232 connectors are isolated. Make sure that the anesthetic
gas outlet at the rear of the module is connected to the gas scavenging system or the gas return line.
See the M1026B Anesthetic Gas Module Service Guide for further information on connecting devices.
Watertrap M1657B
RS232
Connector
The watertrap prevents water and other fluids from passing into the
gas analyzer and causing contamination and/or internal occlusions. It
has a water reservoir in which fluids are collected, two water
separation filters, and two shut-off fuses as a backup mechanism for
the water separation filters.
The watertrap is for multi-patient use. It must be exchanged at least
every two weeks or when watertrap is full.
Fuses
Power
Inlet
Equipotential
Grounding
4
Understanding the Gas Measurement1 Using the Gas Analyzer
Understanding the Gas Measurement
M1013A IntelliVue G1 & M1019A IntelliVue G5
The M1013A IntelliVue G1 and the M1019A IntelliVue G5 use a technique called Non-Dispersive
Infrared Gas Concentration Measurement (NDIR) to measure the concentration of certain gases.
The gases which can be measured by the M1013A IntelliVue G1 and the M1019A IntelliVue G5
absorb infrared (IR) light. Each gas has its own absorption characteristic. The gas is transported into a
sample cell, and an optical IR filter selects a specific band of IR light to pass through the gas. For
multiple gas measurement, such as in the M1013A IntelliVue G1 or the M1019A IntelliVue G5, there
are multiple IR filters. The higher the concentration of gas in a given volume the more IR light is
absorbed. This means that higher concentrations of IR absorbing gas cause a lower transmission of IR
light. The amount of IR light transmitted after it has been passed through an IR absorbing gas is
measured. From the amount of IR light measured, the concentration of gas present can be calculated.
This calculation provides the gas measurement value. Oxygen is measured by an additional sensor in
the M1013A IntelliVue G1 and the M1019A IntelliVue G5 using its paramagnetic properties. The gas
is transported into a sample cell. The higher the oxygen concentration, the higher the measured effect.
The oxygen concentration can be calculated from the amplitude of the effect.
NOTE The presence of organic cleaning solutions or gases containing freon may impact the accuracy of the
infrared gas measurement.
M1026B AGM
The M1026B Anesthetic Gas Module uses a technique called Dispersive Infrared (DIR) to measure the
concentration of certain gases. The gases measured (except oxygen) by the M1026B Anesthetic Gas
Module absorb infrared (IR) light. Each gas has its own absorption characteristic. The gas is
transported into a sample cell. A diffraction grating is used to scan the relevant wavelength range of the
IR light that passes through the sample cell. The higher the concentration of gas the more IR light is
absorbed, and from the amount of IR light measured, the concentration of gas present can be
calculated.
Individual gases have an individual spectral fingerprint. A mathematical algorithm is used to analyze
the spectrum and to identify the anesthetic agents in the gas. Oxygen is measured by an additional
sensor in the M1026B Anesthetic Gas Module using its paramagnetic properties. The gas is
transported into a sample cell. The higher the oxygen concentration, the higher the measured effect.
The oxygen concentration can be calculated from the amplitude of the effect.
NOTE The presence of organic cleaning solutions or gases containing freon may impact the accuracy of the
infrared gas measurement.
Connecting Gas Analyzer Accessories
The gas analyzer accessories and part numbers are listed in the accessories section.
1Insert the M1657B watertrap into the watertrap socket by gently pushing
it up and in. Make sure that the watertrap snap lock clicks into place
2Connect the gas sample tubing to the Luer connector of the watertrap.
5
1 Using the Gas AnalyzerConnecting Gas Analyzer Accessories
3Connect the other end of the gas sample tubing to the patient via the airway adapter.
2
3
4
1
5
6
7
1 M1657B Watertrap
2 Watertrap connector prongs
3 Luer connector (gas sample inlet)
4Snap lock
5 M1658A Gas Sample Tubing
6Gas sample port
7Airway Adapter, either
13902A Elbow Airway Adapter
or M1612A Straight Airway Adapter
8 Connects to the patient
9 Connects to the Anesthesia Machine
9
8
WARNINGMake sure that you do not accidentally connect the luer connector of the gas sample line to an infusion
link or any other links in the patient vicinity.
6
Using the Gas Analyzer Setup Menus1 Using the Gas Analyzer
CAUTION Airway Adapter: Use a Philips Airway Adapter listed in the Accessories section of this manual and
position it so that the part connecting to the gas sample tube is pointing upwards. This prevents
condensed water from passing into the gas sample tube and causing an occlusion. Philips airway
adapters have a built-in port extending from the adapter wall, which reduces the risk of a blockage
occurring.
Watertrap: To minimize the risk of internal contamination, never leave the gas analyzer running
without a watertrap attached (except during a watertrap exchange).
Gas Sample Tube: Do not use the gas sample tube if it is kinked, as it may cause an occlusion or
leakage.
Room Ventilation Make sure that the room in which the gas analyzer is used is well-ventilated with
fresh air. Gases or fumes that mix with and contaminate the room air may degrade measurement
accuracy. Use either a Gas Exhaust Return Filter/Gas Exhaust return Tubing to return gas samples to
the breathing circuit or connect a scavenging system to the gas exhaust port and remove the gas sample.
Note that Gas Exhaust return tubing may not be available for use in all geographies.
Do not use the gas analyzer in a hyperbaric chamber with oxygen enrichment. Also, the ambient air
must be free of CO
enrichment.
2
WARNINGEnsure that the connections are tight. Any leak in the system can result in erroneous readings due to
ambient air mixing with patient gases.
Using the Gas Analyzer Setup Menus
Many gas analyzer settings can be changed just like other measurement settings. These are described in
the chapter on Basic Operation in the Instructions for Use of your patient monitor, only gas analyzerspecific settings are described here.
To change settings for individual gases, enter the setup menu for the individual gas:
♦select the measurement numeric on the monitor screen, or
♦select the required gas label in the Setup <Gas Analyzer> menu.
To change Gas Analyzer settings, enter the
♦select one of the gas analyzer numerics on the monitor screen and then select the menu item
Setup <Gas Analyzer>, or press the Setup hardkey or Airway Gases hardkey on the gas
analyzer.
Choosing Numerics for Display
For each gas the gas analyzer measures, you can choose which numerics are displayed with the
waveform on the screen:
Setup <Gas Analyzer> menu:
et displays the endtidal numerics,
–
in displays the inspiratory numerics,
–
et+in displays both endtidal and inspiratory numerics.
–
Off switches off measurement of that particular gas.
–
7
1 Using the Gas AnalyzerUsing the Gas Analyzer Setup Menus
– MAC displays the minimum alveolar concentration of an anesthetic agent at which patients do not
respond with movement to a painful stimulus.
– MACawk (MAC awake) displays the minimum alveolar concentration of an anesthetic agent at
which patients respond to verbal command.
No waveforms or numerics will be shown for gases set to
To change the displayed numeric, in the
measured to call up a pop-up list of numerics available and then select the numeric you want to display.
As the inspired minimum is measured for CO
Humidity Correction for CO2
The gas analyzer is configured to correct the CO2 measurement for either Body Temperature Pressure
Saturated (BTPS), to account for humidity in the patient’s breath, or Ambient Temperature Pressure
Dry (ATPD).
♦In the Setup CO2 menu, see the menu item Humidity Corr. to see which correction
applies. It is either
Please refer to the Measurement Specifications in the Installation and Specifications chapter of this
manual for details on the humidity correction.
Wet for BTPS or Dry for ATPD.
Adjusting Wave Scales
1In the Wave menu or the Setup menu for the gas, select Scale.
2Choose a suitable scale range from the pop-up list.
Off, and no alarms will be generated.
Setup <Gas Label> menu, select the label of the gas
(imCO2), the numeric label is im instead of in.
2
8
Changing the Apnea Alarm Delay1 Using the Gas Analyzer
Changing the Apnea Alarm Delay
The apnea alarm delay time determines the time limit after which the monitor alarms if the patient
stops breathing.
1In the Setup CO2 menu, select awRR.
2In the SetupawRR menu, select Apnea Time.
3Choose the apnea alarm delay time.
WARNINGThe safety and effectiveness of the respiration measurement method in the detection of apnea,
particularly the apnea of prematurity and apnea of infancy, has not been established.
Deriving Limit Alarms from awRR
1In the Setup CO2 menu or in the Setup <Gas Analyzer> menu, select awRR.
2In the Setup awRR menu, select Alarms.
3Select On to derive alarms from the airway respiration signal or Off to disable them.
Alarms and Zero Calibration
When a zero calibration is in progress, the physiological alarm detection is suspended. When the
calibration is finished, the gas analyzer resumes alarm detection. If an alarm condition is present after
the zero calibration, the alarm will be activated within the specified alarm delay time.
WARNINGIf an apnea occurs during a zero calibration, the time delay between the start of apnea and the
activation of the apnea alarm could be up to 24 seconds plus the configured apnea delay time. After
startup or after continuous operation of the M1013A IntelliVue G1 or the M1019A IntelliVue G5 of
4 months or more the time delay could be up to 93 seconds plus the configured apnea time for a single
time.
Automatic Alarm Suppression
Your monitor can be set to suppress alarms until it detects that a patient has been connected to the gas
analyzer (when breathing is detected). This feature is called
or
Off in the monitor’s Configuration Mode.
No Al til Breath and can be set to On
9
1 Using the Gas AnalyzerAgent Identification
Agent Identification
The following table shows the Agent Identification possibilities of the different gas analyzers:
Gas AnalyzerManual Agent
Identification
M1013A IntelliVue G1yesnon/a
M1019A IntelliVue G5noyes2
M1026B AGMyesyes1
NOTE Only the M1026B AGM allows switching the agent identification mode between Agent Id:
Manual and Agent Id: Auto. Nevertheless, agent identification behavior for the M1013A
IntelliVue G1 and the M1019A IntelliVue G5 is described below in the Manual Agent ID (M1013A)
and the Automatic Agent ID (M1019A) sections.
Setting the agent identification mode to Agent Id: Manual lets you choose the anesthetic agent
manually. If you choose the setting Agent Id: Auto, the gas analyzer automatically identifies the
predominant anesthetic agent(s) in the breathing circuit.
♦To change the agent identification mode, in the SetupAGT menu, select Agent Id: to toggle
between the settings
Auto and Manual.
Automatic Agent
Identification
Number of agents
identified
If Agent ID is Set to Manual
To change the agent monitored, when Agent Id is set to Manual:
♦In the Setup <Agent label> menu, select Agent to call up a pop-up list of available agents
and select the agent you want to monitor. For example, Setup HAL.
M1026B only: If the manually selected agent does not match the agent detected, the INOP CHECK
AGENT appears.
WARNINGMake sure to select the correct agent for monitoring. Selecting the wrong agent will cause erroneous
readings.
If Agent ID is Set to Auto (M1019A IntelliVue G5 & M1026B AGM)
As soon as the M1019A IntelliVue G5 or the M1026B AGM has detected the agent(s), a waveform
and numerics for this agent appears on the monitor screen, if they are configured to be displayed.
During the process of identification, the generic label
G5) is shown as a placeholder.
For an anesthetic agent to be detected by automatic agent identification, its concentration must exceed
the identification threshold. The presence of other substances in the patient’s breathing circuit such as
methanol or acetone can influence the agent identification and lead to incorrect values and incorrect
identification.
If the anesthetic agent administered to the patient changes, a mixture of both gases is detected by the
M1019A IntelliVue G5 or the M1026B AGM during the transition. The time needed to complete the
exchange depends on the type of anesthesia (low flow or high flow), and the characteristics of the
agents administered (pharmacokinetics). During the exchange, you will see the INOP message
MIXTURE
AGM only: If you are using automatic agent identification, when one of the agents decreases below its
threshold and the other agent predominates, the monitor will recognize the exchange.
M1013A IntelliVue G1 and M1026B AGM only: If you are using manual agent identification, you
must change the agent in the
and (with AGM only) -?- next to the affected numerics.
Agent Setup menu to match the administered agent.
AGT
Agent ID During Emergence from Anesthesia
(M1019A IntelliVue G5 & M1026B AGM)
If automatic agent identification is selected during emergence from anesthesia and the agent
concentration falls below the identification threshold, the agent will no longer be identified. The agent
label will remain on the display and the numeric will show
IntelliVue G5) until the monitor detects that a patient is no longer connected. After this, the generic
label
AGT (AGT1 / AGT2 for M1019A IntelliVue G5) will be shown.
0.00 % (numeric unavailable with
♦To display the correct agent and value, change to manual identification and select the agent
manually.(AGM only)
MAC Calculation
The MAC (Minimum Alveolar Concentration) value of an anesthetic gas or agent denotes the
concentration at which 50% of a population of anesthetized patients do not respond with movement
to a painful stimulus (e.g. a standardized incision through the skin). The MAC awake represents the
concentration at which 50% of a population of anesthetized patients responds to verbal command.
The Philips IntelliVue patient monitors offer three configurable methods of MAC calculation:
•Uncorrected MAC
•Ambient Pressure Corrected MAC
• Enhanced MAC Correction
The preferred method must be set in configuration mode of your patient monitor. The total MAC and
MACawk values can then be selected for display on your monitor. The sections below describe how
these values are calculated with the different methods.
To switch the MAC and/or the MACawk parameter on, set MAC and /or MACawk to ON in the gas
analyzer setup menu.
NOTE • The MACawk value can only be displayed if MAC Correction is configured to “Enhanced”.
• MAC Calculation is only available in IntelliVue patient monitors with software revision C.0 or
higher.
• Ambient Pressure Corrected MAC and Enhanced MAC Correction are not available in the USA.
11
1 Using the Gas AnalyzerMAC Calculation
Uncorrected MAC
If the MAC Correction is configured to “Off” the uncorrected MAC is calculated. The MAC value is
not corrected for ambient pressure, age, temperature or any other individual factors influencing the
effect of volatile anesthetic agents.
In order to calculate the MAC value the standard 1MAC concentrations of anesthetic agents and
nitrous oxide are required. The following table lists these concentrations (according to the EN ISO
21647:2004 standard). The values are based on the assumptions that the patient is 40 years old (except
for Desflurane where 25 years are assumed), the body temperature is 37° and the atmospheric pressure
is 760 mmHg (1 atm):
For each volatile anesthetic agent detectable by the gas analyzer the MAC value for the specific agent
(MAC(AA)) is calculated as follows:
etConc(AA)
MAC(AA)
-----------------------------
=
1MAC(AA)
where AA = Anesthetic Agent and etConc = end-tidal concentration
In the same way, the MAC value for nitrous oxide (MAC(N2O) is derived from the measured value of
the nitrous oxide end-tidal concentration (etConc(N2O)):
etCONC(N2O)
MAC(N2O)
Finally, the total MAC value of nitrous oxide and the selected anesthetic agent is calculated as follows:
------------------------------------
=
1MAC(N2O)
MACMAC(N2O)MAC(AA)+=
NOTE Gas components (N2O and/or anesthetic agent) which are switched off, are not included in the total
MAC computation.
Ambient Pressure Corrected MAC (not available in the USA)
If the MAC Correction is configured to “Amb. Pressure”, the MAC is corrected to reflect the effect of
a different partial pressure at another altitude.
12
Loading...
+ 40 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.