Philips IntelliVue MP80, IntelliVue 90 User manual

Page 1
IntelliVue MP80/90 & D80
Service Guide
IntelliVue Patient Monitor
MP80/90 & D80
Patient Monitoring
Page 2
Part Number M8000-9351K 4535 641 12591
*M8000-9351K*
Page 3
Table of Contents
1 Introduction 9
Who Should Use This Guide 9 How to Use This Guide 9 Abbreviations 9 Responsibility of the Manufacturer 10 Passwords 11 Warnings and Cautions 11
2 Theory of Operation 13
Monitor Theory of Operation 13
System Boundaries 14 Hardware Building Blocks 15 Data Flow 19 How does the Support Tool Work with the Monitor 23 Monitor Software Block Diagram 24 Block Diagram Legend 25
3 Testing and Maintenance 33
Introduction 33 Terminology and Definitions 34 Recommended Frequency 35 When to Perform Tests 36 Testing Sequence 40 Visual Inspection 41
Before Each Use 41 After Each Service, Maintenance or Repair Event 41 Power On Test 41
Safety Tests 42
Warnings, Cautions, and Safety Precautions 43 Safety Test Procedures 44
Preventive Maintenance Procedures 84
Noninvasive Blood Pressure Measurement Calibration 84
Performance Assurance Tests 84
Basic Performance Assurance Test 84 Full Performance Assurance Test 85 ECG/Resp Performance Test 85 ECG Sync Performance Test 86 SpO2 Performance Test 86 NBP PerformanceTest 87 Invasive Pressure Performance Test 89
Page 4
Temperature Performance Test 90 M3014A Capnography Extension Performance Tests 90 Microstream CO2 Performance Test 93 Spirometry Performance Tests 98 Cardiac Output Performance Test 100 BIS Performance Test 101 Vuelink Performance Test 102 IntelliBridge Performance Test 103 EEG, SvO2 and tcGas Performance Tests 103 Nurse Call Relay Performance Test 103 Power Loss Alarm Buzzer Performance Test (only if Multi-Port Nurse Call Connector Board is installed) 105 IntelliVue 802.11 Bedside Adapter Communication Test 106
Reporting of Test Results 107
Carrying Out and Reporting Tests 108 Evaluation of Test Results 111
Other Regular Tests 112 Touchscreen Calibration 112 Disabling/Enabling Touch Operation 112 Printer Test Report 113 After Installation, Testing or Repair 113
4 Troubleshooting 115
Introduction 115 How To Use This Section 115 Who Should Perform Repairs 115 Replacement Level Supported 116 Hardware Revision Check 116 Hardware/Software Compatibility Matrix 117 Software Revision Check 118 Software Compatibility Matrix 119
Compatibilty with MMS 119 Compatibilty with FMS 120 Compatibility with Information Center 120 Number of Supported Parameter Modules 121
Obtaining Replacement Parts 123 Troubleshooting Guide 123
Checks for Obvious Problems 123 Checks Before Opening the Instrument 124 Troubleshooting Tables 126 Status Log 154 List of Error Codes 156 Troubleshooting with the Support Tool 156 Troubleshooting the Individual Measurements or Applications 157
Page 5
5 Repair and Disassembly 159
Tools Required 159 MP80/D80/MP90 CMU Disassembly 159
Removing I/O Boards 160 Removing the Top Cover 162 Removing the Plastic Feet and/or the Locking Cam 163 Removing the optional Fans (MP90 Dual CPU Versions only)* 163 Replacing the Second (Independent) Video Board(MP90 Dual CPU Versions only) 164 Removing the Second CPU/Main Board (MP90 Dual CPU Versions only) 165 Accessing the Main CPU or Primary Video Board (MP90 Dual CPU Versions) 167 Replacing the Primary Video Board 168 Removing the Main Board 170 Removing the Power Supply 172 Removing the Speaker (MP80/MP90 only) 174 Removing the Power On/Off Switch 174
Flexible Module Rack (FMS) Disassembly 175
Removing the Handle and the Measurement Server Mount 175
Plug-in Modules 180
Plug-In Module Disassembly 181
Multi-Measurement Module (MMS) Disassembly 184
Tools required 184 Removing the Front Cover 184 Removing the Mounting Pin 185 Removing the Top Cover 185 Removing the DC/DC Board 186 Removing the MSL Flex Assembly 186 Reassembling the MSL Flex Assembly 187 Removing the NBP pump 189 Refitting the new NBP Pump 189 Refitting the DC/DC board 191 Refitting the Cover 191 Refitting the Front Cover 192 Final Inspection 193
MMS Extensions - Exchanging the Top Cover, MSL Flex Cable and the Dual Link Bar 193
Exchange Procedures 194
Disassembly Procedures for the M3015A MMS Extension (HW Rev. A) 205
Removing the Front Cover 205 Refit Procedures for the MMS Extension 209
6 Parts 211
MP80/MP90/D80 Parts 212
Exchange Parts 212 Replacement Parts 213
Flexible Module Rack (FMS) Parts 216
Page 6
Exchange and Replacement Parts 216
Multi-Measurement Module (MMS) Parts 218
MMS Part Number Overview and Identification 218 MMS Firmware Overview 220 MMS Part Numbers - Front Bezel for M3001 #A01 & #A03 221 MMS Part Numbers - Front Bezel for M3001 #A02 221 MMS Part Numbers - Top Cover and MSL Assembly 222 MMS Exchange Part Numbers 223 MMS Part Numbers - Label Kits 225 MMS Part Numbers - NBP Assembly 225
MMS Extension Parts (M3012A, M3014A, M3015A and M3016A) 225
MMS Extension Part Numbers - Release Mechanisms 226 MMS Extension Part Numbers - Top Cover, Flex Cable and Link Bar 226 MMS Extension Part Numbers - Front Bezels 226 Exchange Parts List 228
IntelliVue X2 Part Numbers 229 Plug-in Modules Part Numbers 229
Part Number Table 230 Plug-In Modules Replaceable Parts 233 BIS Solution Replaceable Parts 238 BISx Solution Replacable Parts 239 tcpO2/tcpCO2 Module Accessories 240
IntelliVue 802.11 Bedside Adapter Part Numbers* 241 External Display Part Numbers 241 SpeedPoint Part Numbers 244 Remote Alarm Device Part Numbers 245 Remote Extension Device Part Numbers 245
7 Installation Instructions 247
Installation Checklist 247 Unpacking the Equipment 248 Initial Inspection 248
Mechanical Inspection 248 Electrical Inspection 248 Claims For Damage and Repackaging 249
Installing the M8008A/M8010A/M8016A CMU 249
Mounting Instructions 250
Connecting the Monitor to AC Mains 251
Connections 251 Installing Interface Boards 252 Connection of Devices via the MIB/RS232 Interface G.00.xx or higher 257 Connection of Devices via the MIB/RS232 Interface (Rev. D.00.58 to F.01.42) 258 Connection of Devices via the MIB/RS232 Interface (Rev. A.10.15 to C.00.90) 258 Connection of MIB Devices (Rev. below A.10.15) 259
Page 7
Connection of USB Devices 260
Setting Up Multiple Displays 264
Installation of Multiple Displays 264 Configuring Multiple Displays 266 Examples for Multiple Display Use Models 270
Installing Remote Devices 272
Mounting the 15” Remote Display (M8031A) 272 Mounting the 15” Remote Display (M8031B) 273 Mounting the 17” Remote Display (M8033A/B/C) 274 Hardware Settings 277 Flexible Module Rack and/or Multi-Measurement Module 277 Remote Alarm Devices 283 Remote Extension Device 284 PS/2 Keyboard/Mouse 286
Philips Clinical Network (Wired) 286 Philips Clinical Network (Wireless) 286 Nurse Call Relay 287
Connections 287
ECG Out Functionality 287
Connections 287
Configuration Tasks 288
Checking Country-Specific Default Settings 289 Setting Altitude, Line Frequency, ECG Cable Colors and Height & Weight Units 289 Setting Altitude and Line Frequency 290 Configuring the Equipment Label 290 Configuring the printer 290 Configuring IP Address, Subnet Mask and Default Gateway 290 Configuration Settings for CSCN Routed Bedside Monitors (RBM) 291 Configuring Routed Bedside Monitors Support 291 Display Settings 292
IntelliBridge EC10 293
Accessing the IntelliBridge EC10 Service Interface 293 Firmware Upgrade 294 Uploading and Removing Device Drivers 295 Generating and Uploading Clone Files 295 Viewing System Information 296
Handing Over the Monitor 297
8 Site Preparation 299
Introduction 299
Site Planning 299 Roles & Responsibilities 300
M8008A/M8010A/M8016A Site Requirements 302
Space Requirements 302 Environmental Requirements 303
Page 8
Electrical and Safety Requirements (Customer or Philips) 303
Remote Device Site Requirements 304
Connecting Non-Medical Devices 305 Multi-Measurement Module (MMS) M3001A, IntelliVue X2 M3002A or Flexible Module Rack (FMS) M8048A305 Remote Displays (M8031A) 310 Remote Displays (M8031B) 311 Remote Displays - M8033A 312 Remote Displays - M8033B 313 Remote Displays - M8033C 314 Remote Alarm Devices 316 Remote Extension Device 317 IntelliBridge 319 Local Printer 319
Philips Medical LAN 319 RS232/MIB/LAN Interface 320 Nurse Call Relay Interface 321 ECG Out Interface 321
9 Gas Analyzers 323 10 Philips 15210B Calibration Unit 325
Unpacking the Instrument 325
Initial Inspection 326 Instrument Identification 326 Specification 326 Operating Environment 327 Operating Information 327 Fitting the Gas Cylinders 327 Storage of Gas Cylinders 327 Disposal of Used Gas Cylinders 327
Routine Maintenance 327
Changing the Gas Cylinders 327 Care and Cleaning 328
Theory of Operation 328 Gas Flow Performance Check 329
Test Procedure 329
Disassembly 331 Parts List 332
11 IntelliVue Product Structure 335
Upgrade Options 337
12 Index 341
Page 9
1

1Introduction

This Service Guide contains technical details for the IntelliVue MP80/90 Patient Monitor, the Multi-Measurement Module (MMS), the IntelliVue X2, the Flexible Module Rack (FMS) and the Measurement Server Extensions.
This guide provides a technical foundation to support effective troubleshooting and repair. It is not a comprehensive, in-depth explanation of the product architecture or technical implementation. It offers enough information on the functions and operations of the monitoring systems so that engineers who repair them are better able to understand how they work.
It covers the physiological measurements that the products provide, the Measurement Server that acquires those measurements, and the monitoring system that displays them.

Who Should Use This Guide

This guide is for biomedical engineers or technicians responsible for troubleshooting, repairing, and maintaining Philips’ patient monitoring systems.

How to Use This Guide

This guide is divided into eight sections. Navigate through the tab le of conte nts at the left of the screen to select the desired topic. Links to other relevant sections are also provided within the individual topics. In addition, scrolling through the topics with the page up and page down keys is also possible.

Abbreviations

Abbreviations used throughout this guide are:
Name Abbreviation
IntelliVue MP80/90 Patient Monitor the monitor Flexible Module Rack FMS
Page 10
1 Introduction Responsibility of the Manufacturer
Name Abbreviation
Multi-Measurement Module MMS Measurement Link MSL Medical Information Bus MIB Anesthetic Gas Module AGM

Responsibility of the Manufacturer

Philips only considers itself responsible for any effects on safety, EMC, reliability and performance of the equipment if:
- assembly operations, extensions, re-adjustments, modifications or repairs are carried out by
persons authorized by Philips, and
- the electrical installation of the relevant room complies with national standards, and
- the instrument is used in accordance with the instructions for use.
To ensure safety and EMC, use only those Philips parts and accessories specified for use with the monitor. If non-Philips parts are used, Philips is not liable for any damage that these parts may cause to the equipment.
This document contains proprietary information which is protected by copyright. All Rights Reserved. Reproduction, adaptation, or translation without prior written permission is prohibited, except as allowed under the copyright laws.
Philips Medizin Systeme Böblingen GmbH Hewlett-Packard Str. 2 71034 Böblingen, Germany The information contained in this document is subject to change without notice. Philips makes no warranty of any kind with regard to this material, including, but not limited to,
the implied warranties or merchantability and fitness for a particular purpose. Philips shall not be liable for errors contained herein or for incidental or consequential damages
in connection with the furnishing, performance, or use of this material.
Page 11
Passwords 1 Introduction

Passwords

In order to access different modes within the monitor a password may be required. The passwords are listed below.
Monitoring Mode: No password required Configuration Mode: 71034 Demo Mode: 14432 Service Mode: 1345 Consult the configuration guide before making any changes to the monitor configuration.

Warnings and Cautions

In this guide:
- A warning alerts you to a potential serious outcome, adverse event or safety hazard. Failure
to observe a warning may result in death or serious injury to the user or patient.
- A caution alerts you where special care is necessary for the safe and effecti ve use of the
product. Failure to observe a caution may result in minor or moderate personal injury or damage to the product or other property, and possibly in a remote risk of more serious injury.
Page 12
1 Introduction Warnings and Cautions
12 13
Page 13

2Theory of Operation

Monitor Theory of Operation

The IntelliVue MP80/MP90 Patient Monitor:
- displays real-time data
- controls the attached measurement servers
- alarms in the case of patient or equipment problems
- offers limited data storage and retrieval (trending)
2
- interfaces to the Philips Clinical Network and other equipment A monitor with just a single integrated measurement server can be connected to additional
building blocks to form a monitoring system with a large number of measurements, additional interface capabilities and multiple slave displays. These elements cooperate as one single integrated real-time measurement system.
Page 14
2 Theory of Operation Monitor Theory of Operation

System Boundaries

The following diagram discusses specific boundaries within the overall system with respect to their openness and real-time requirements:
Measurement LAN combines components of one patient monitor; real time requirements
across all interconnect ed elements Philips Clinical Network (wired LAN) connects multiple patient monitors, information centers, application
servers; closed system, only Philips qualified products (tested and with regulatory approval) are connected, Philips is responsible for guaranteed real-time functionality and performance
Philips Clinical Network (wireless) like Philips Clinical Network (wired) LAN, however due to current
wireless technologies available it has reduced bandwidth, longer latencies, reduced functionality
Hospital LAN, Internet Standard Network, not under Philips control, no guaranteed service,
no real-time requirements
Page 15
Monitor Theory of Operation 2 Theory of Operation

Hardware Building Blocks

The following hardware building blocks make up the monitoring system:
IntelliVue MP80/MP90
The MP80/MP90 monitor:
- can be used with the standalone M8031A 15-inch or the M8033A 17-inch color LCD TFT
display with touchscreen operation.
- can also be used with other XGA and SXGA standalone off-the-shelf displays which comply
with medical standards such as IEC 60601-1 and IEC 60601-1-2.
- has the central processing unit in a separate module
- uses the Philips SpeedPoint as primary input device whereas the Philips Touchscreen and
computer devices such as mice, trackball, and keyboard can be added o pti on ally
- supports the Flexible Module Rack (FMS)
Page 16
2 Theory of Operation Monitor Theory of Operation
Building Blocks:
Optional Hardware
The M8031A 15-inch color LCD TFT display or the M8033A 17-inch color LCD TFT display (both with touchscreen operation) can be ordered optionally. Additional input devices such as mice, trackball or keyboard can also be added. If the monitor is ordered with the wireless LAN option a wireless transmitter is required. For further details regarding the wireless network please refer to the M3185A Philips Clinical Network documentation.
Page 17
Monitor Theory of Operation 2 Theory of Operation
Compatible Devices
M3001A Multi-Measurement Module (MMS)
M3002A IntelliVue X2
M3012A, M3014A, M3015A, M3016A MMS Extensions
Page 18
2 Theory of Operation Monitor Theory of Operation
Power Supply
The AC/DC converter transforms the AC power coming from the power plug into 48 V/120W DC source and isolates the monitoring system from the AC power mains.The 48V is distributed via power bus and supplies power to all the components of the system: The 56 V DC power needed for the FMS, MMS and MMS Extension is created by an isolating DC/DC converter. The CPU is supplied with 3.3 V and 5 V DC power. The transformation is performed in two steps: The first DC/DC converter is a power regulator which reduces the variations caused by load changes on the 48V power bus. The second DC/DC converter converts the power to the needed voltage. Interface boards require a power of 10V AC.
CPU Boards
The CPU boards have an MPC860 50 MHz or MPC86x 100 MHz processor that provides a number of on-chip, configurable interfaces. An array of 12 fast UARTS with configurable protocol options are implemented in an ASIC (along with other system functions such as independent watchdogs etc.), providing interfacing capabilities to measurement modules and I/O boards. The serial interfaces can easily be electrically isolated. The main board contains additional video hardware.
The CPUs provide two LAN interfaces to interconnect CPUs (via the MSL) and to connect to the Philips Clinical Network.
The CPU capabilities are identical. Different loading options are coded on serial EEPROMs to support the automatic configuration of the operating system at boot time.
Page 19
Monitor Theory of Operation 2 Theory of Operation
I/O Boards
Interfaces to the monitor are implemented via I/O boards. The location of these boards is restricted by general rules. The I/O slot designations diagram and the I/O matrix which outline the I/O board placement rules can be found in the Installation Instructions section.
The following is a list of Interface (I/O) boards which may be present in your monitor, depending on your purchased configuration:
- MSL
- Video (analog)
- Philips Clinical Network (LAN wired or wireless)
- Basic Alarm Relay (Nurse Call)
I/O boards:
- PS/2
- MIB/RS232
- USB

Data Flow

- Flexible Nurse Call
- Parallel printer
- Remote devices (Remote Alarm Device, Remote Extension Device)
- IntelliVue 802.11 Bedside Adapter
The specifications for the above listed interfaces can be found in the technical data sheet for the monitor and in the Installation and Specifications chapter of the Instructions for Use.
The following diagram shows how data is passed through the monitoring system. The individual stages of data flow are explained below.
Page 20
2 Theory of Operation Monitor Theory of Operation
Data Acquisition
Monitoring data (for example patient measurement data in the form of waves, numerics and alerts) is acquired from a variety of sources:
- Measurement Servers
The Measurement Servers connected to the internal LAN convert patient signals to digital data and apply measurement algorithms to analyze the signals.
- External measurement devices
Data can be also acquired from devices connected to interface boards of the monitor. Software modules dedicated to such specific devices convert the data received from an external device to the format used internally. This applies to parameter modules and the Anesthetic Gas Module.
- Server systems on the Philips Clinical Network
To enable networked applications such as the other bed overview, data can be acquired from server systems attached to the Philips Clinical Network, for example a Philips Information Center
Data Provider System Service
All data that is acquired from measurement servers or external measurement devices is temporarily stored by a dedicated data provider system service. All monitor applications use this central service to access the data in a consistent and synchronized way rather than talking to the interfaces directly.
This service makes the applications independent of the actual type of data acquisition device. The amount of data stored in the data provider system service varies for the different data types.
For example several seconds of wave forms and the full set of current numerical values are temorarily stored in RAM.
Persistent Data Storage System Service
Some applications require storage of data over longer periods of time. They can use the persistent data storage system service. Dependent on the application requirements, this service can store data either in battery backed-up (buffered) memory or in flash memory. The buffered memory will lose its contents if the monitor is without power (not connected to mains) for an extended period of time. The flash memory does not lose its contents.
The trend application for example stores vital signs data in a combination of flash memory and buffered memory, while the system configuration information (profiles) is kept purely in flash memory.
Page 21
Monitor Theory of Operation 2 Theory of Operation
Display and User Interface Service
Applications can use high level commands to display monitoring data or status and command windows on the internal LCD panel. These commands are interpreted by the display manager application. This application controls the dedicated video hardware which includes video memory and a special ASIC.
User input is acquired from a variety of input devices, for example the SpeedPoint, the touchscreen or other standard input devices (keyboard, mouse) which may be attached to I/O boards. The system software makes sure that the user input is directed to the application which has the operating focus.
Data Output
The monitoring system is very flexible and customizable regarding its data output devices. Built-in devices (for example LAN, alarm lamps, speaker, video) provide the basic output capabilities.
These capabilities can be enhanced by adding additional I/O boards, as required in the specific end-user setup. The additional I/O boards typically provide data to externally attached devices, for example to printers, RS232 based data collection devices, nurse call systems etc.
The monitor can identify I/O boards by means of a serial EEPROM device that stores type and version information. The operating system detects the I/O boards and automatically connects them with the associated (interface driver) application. For some multi-purpose cards it is necessary to configure the card for a particular purpose first (for example the dual MIB/RS232 card can support external touch display , data import, data export).
Monitor Applications
The monitor applications provide additional system functionality over the basic measurement and monitoring capabilities. This includes for example trending, report generating, event storage or derived measurements.
In general, the monitor applications use the data provider system service to access the measurement data. Application interfaces to the other system services allow the application to visualize data, to store data over extended periods of time or to output data to other devices.
Page 22
2 Theory of Operation Monitor Theory of Operation
Internal LAN (Measurement Link)
All components of the monitoring system (including measurement servers and CPUs in the monitor) communicate using an IEEE802.3/ Ethernet LAN in the Measurement Link (MSL). This network is used to distribute data between the components, for example:
- Digitized patient signals including wave data, numerical data and status information
(typically from the measurement server to a display unit)
- Control data representing user interactions (typically from the display unit to a measurement
server)
- Shared data structures, for example representing patient demographical data and global
configuration items
The internal LAN allows plug and play configuration of the monitoring system. The system automatically detects plugging or unplugging of measurement servers and configures the system accordingly.
The components on the internal LAN are time-synchronized to keep signal data consistent in the system. Dedicated hardware support for synchronization eliminates any latency of the network driver software.
The integrated LAN provides deterministic bandwidth allocation/reservation mechanisms so that the real-time characteristic of signal data and control data exchange is guaranteed. This applies to the data flow from the measurement server to the monitor (for example measurement signal data) and the data flow from the monitor to a measurement server (for example to feed data to a recorder module).
Integrated communication hubs in the monitor and the FMS allow flexible cabling options (star topology, daisy chaining of servers).
Page 23
Monitor Theory of Operation 2 Theory of Operation
Philips Clinical Network
The monitoring system may be connected to the Philips Clinical Network, for example to provide central monitoring capabilities or other network services. This connection may be through a normal wired connection or through a wireless connection.
The monitor supports the connection of an external wireless adapter or an internal wireless adapter (#J35). Switching between wired and wireless networks is automatically triggered by the plugging or unplugging of the network cable.
The Philips Clinical Network protocols function very similarly to the protocols used on the internal LAN.
After configuration, the monitoring system sends the digitized patient signals including wave data, numerical data and status information onto the network. Control data representing user interactions can be exchanged between the monitoring system and a central station bi-directionally.
Additional protocols are supported for networked applications, for example for the other bed overview function, which allows viewing of monitoring data from other patients on the network.
For plug and play operation, the monitoring system uses the standard BootP protocol to automatically acquire a network address.

How does the Support Tool Work with the Monitor

The support tool is a Windows application typically installed on the laptop of a customer engineer or a biomedical engineer working in the customer’s own service department.
The purpose of the support tool is to upgrade, configure and diagnose all monitoring componen ts (modules, measurement servers, and monitors) in the system over the network.
The service protocol developed for this purpose uses a raw access to the devices without the need for IP addresses etc. over a standard customer network installation, so that even defective devices can be upgraded as long as the few kBytes of initial boot code are working. The boot code itself can also be upgraded using the same protocol.
The tool allows access to internal service information and to serial numbers. It can be remote­controlled, for example via a dial-up connection from a response center, provided the proper infrastructure is in place.
For details see the Instructions for Use for the Support Tool.
Page 24
2 Theory of Operation Monitor Theory of Operation

Monitor Software Block Diagram

shows the functional block diagram for the monitoring system. A legend explaining terms and diagram elements follows. The information below varies depending on the purchased monitor options.
IntelliVue Patient Monitoring System Functional Block Diagram
Page 25
Monitor Theory of Operation 2 Theory of Operation

Block Diagram Legend

Functional Block Description
Services Operating System The Operating System (OS) provides a layer of isolation
between the specific hardware implem entati on an d the application software. The OS performs system checks and allocates resources to ensure safe operation when the system is first started. This includes internal self-tests on several hardware modules and configuration checks for validity of configuration with the operating software. During normal operation, the OS continues to run checks on system integrity. If error conditions are detected the OS will halt monitoring operations and inform the operator about the error condition.
System Services The System Services provide generic common system
services. In particular: They use a real-time clock component to track time. They synchronize to network time sources and verify the accuracy of the system time information. They are also responsible for managing persistent user configuration data for all Measurement Servers, Flexible Module Racks and IntelliVue Patient Monitoring System software modules. User configuration data is stored in a non-volatile read/write storage device
Applications Application Server Client The Application Server Client provides the Citrix1 thin
client functionality.
Page 26
2 Theory of Operation Monitor Theory of Operation
Functional Block Description
Reports The Reports Service retrieves current and stored
physiological data and status data to format reports for printing paper documentation. The following reports are supported:
- Vital Signs Report
- Graphical Trend Report
- Event Review Report
- Event Episode Report
- ECG Report (12 Lead/Multi-Lead)
- Cardiac Output Report
- Calculations Report
(Hemodynamic/Oxygenation/Ventilation)
- Calculations Review Report
- Wedge Report
- Test Report
- Other reports (e.g. Loops, Review Applications, Drug
report)
The Reports service generates report data which can be printed on a local or a central printer.
Record The Record Service retrieves current and stored
physiological data and status data to format a continuous strip recording. A recording can be triggered manually by the operator or automatically by an alarm condition. The Record Service uses the services of the Recorder Interface to control an M1116B Recorder in the FMS. The Record Service can also send data to a central recorder.
Page 27
Monitor Theory of Operation 2 Theory of Operation
Functional Block Description
Alarm The Alarm Service contains logic that prioritizes alarm
conditions that are generated either by the Measurement Servers, Flexible Module Rack, or by IntelliVue Patient Monitoring System software modules. Visual alarm signals (messages) are displayed at the top of the IntelliVue Patient Monitoring System display and alarm sounds are generated by a loudspeaker. Alarm conditions may be generated when a physiological parameter exceeds preselected alarm limits or when a physiological parameter or any other software module reports an inoperative status (technical alarm, for example, the ECG leads may have fallen off the patient). The Alarm service manages the alarm inactivation states, for example suspension of alarms, silencing of alarms, and alarm reminder. Alarm signals may also be configured as latching (alarm signals are issued until they are acknowledged by the operator, even when the alarm condition is no longer true). The Alarm service controls the visual alarm signals (alarm lamps).
Trend The Trend service stores the sample values of
physiological data and status data with a resolution of 12 seconds, 1 minute or 5 minutes for a period of up to 48 hours. The data is kept in battery buffered read/write storage and flash memory devices to be preserved across power failures. The stored data is protected via consistency checks and checksums. When a new patient is admitted, the trend database erases all data of the previous patient.
HiRes The OxyCRG (Oxygen CardioRespiroGram) service
derives a high-resolution trend graph from the Beat-to-Beat Heart Rate, SpO2 or tcpO2, and Respiration physiological data. The OxyCRG is specialized for neonatal applications, allowing the operator to identify sudden drops in Heart Rate (Bradycardia) and SpO2 or tcpO2 (Desaturations), and supporting the operator in visualizing Apnea situations.
ADT The ADT (Admit/Discharge/Transmit) service maintains
the patient demographics information. The operator may admit a new patient, discharge the old patient and enter or modify the patient demographics. The ADT service also supports the transport of a patient (trend database) with the M3001A Multi-Measurement Module. The ADT service controls the deletion of old patient data, the upload of trend data from the M3001A and the switching back of all settings to user defaults. It also synchronizes patient information with a central station on the network.
Page 28
2 Theory of Operation Monitor Theory of Operation
Functional Block Description
Calc Param The Calc Param (Calculated Parameters) service accesses
current, stored and manually entered physiological data as input to calculation formulas. With these formulas, derived hemodynamic, oxygenation and ventilation variables are computed. The calculation results, including the input parameters, are stored for later review using the Trend service.
Drug Calc The Drug Calc application aids in calculating drug
dosages for patients.
PV Loops The PV Loops application compares graphic
representations of airway waves to help detect changes in
the patient airway condition. Interface Managers MDSE The MDSE (Medical Data Service Element) Interface
Manager is responsible for the exchange of real-time data
between the IntelliVue Patient Monitoring System display
unit and the Measurement Servers and Flexible Module
Rack as well as between the IntelliVue Patient
Monitoring System display unit and other devices
attached to the network. MDSE establishes and maintains
a data communication link between the devices. It
provides configuration information about the remote
device to applications in the local device and it allows the
exchange of measurement data and status information
between the devices. Printer The Printer Interface Manager provides a high level
interface to a printer. It provides means to:
- establish a connection to the printer
- transfer data to the printer
- get status of the printer
- close connection to the printer
The Printer Interface Manager also supervises the
connection to the printer and whether the printer accepts
data (for example paper out). The Printer Interface
Manager notifies the operator in such cases.
Page 29
Monitor Theory of Operation 2 Theory of Operation
Functional Block Description
Display & Operator Interface
The Display and Operator Interface Manager performs the
following tasks:
- Screen presentation of real-time and stored
physiological measurement data, alarm condition data and status information received from the MDSE interface manager, the Alarm service or other IntelliVue Patient Monitoring System modules
- Screen presentation of operating controls (control
windows)
- Processing of operating control commands received
from HIF Control interface. The module verifies and interprets the received commands and forwards them to other software modules of the IntelliVue Patient Monitoring System display unit, Measurement Servers or Flexible Module Rack
- Sound generation (issues audible alarm signals and
generates audible information signals, for example QRS and SpO2 tones, operator audible feedback)
Interfaces LAN The LAN interface implements the physical layer of IEEE
802.3. The LAN interface performs Manchester
encoding/decoding, receive clock recovery, transmit pulse
shaping, jabber, link integrity testing, reverse polarity
detection/correction, electrical isolation, and ESD
protection. Electronically separated interfaces are used for
communication to the Measurement Servers or Flexible
Module Rack and to the network. Centronics The Centronics interface implements the standard
signaling method for bi-directional parallel peripheral
devices according to IEEE 1284-I. The interface is used
as a parallel interface to a standard printer with electrical
isolation and ESD protection.
Page 30
2 Theory of Operation Monitor Theory of Operation
Functional Block Description
Display Controller The Display Controller Interface consists of a video
controller chip, video RAM and the controlling software.
The Display Controller interface processes the high level
display commands (character and graphic generation,
wave drawing) and translates them into pixels, which are
written into the video RAM where the video controller
chip generates the video synchronization signals and the
pixel stream for the Color LCD Display. HIF Control The HIF (Human Interface Control) interface scans the
Human Interface devices for operator controls (Touch
Screen, Speed Point, USB and PS/2 devices), formats the
collected data and sends it to the display and Operating
Interface. ECG-Out Marker-In The ECG Out/Marker In interface receives the ECG
waveform directly from the ECG/Resp Arrhythmia
ST-Segment physiological algorithm via an RS-422 serial
interface and converts the digital ECG signal to an analog
ECG signal. In addition, the ECG Out controller receives
from a connected device the marker information and
forwards this data to the ECG/Resp Arrhythmia
ST-Segment physiological algorithm. The converted
analog signal is used to synchronize a connected device to
the patient’s ECG RS-422 The serial link RS-422 interface communicates the ECG
signal to the ECG Output/Marker In of the IntelliVue
Patient Monitoring System display unit. The interface is a
serial, differential, full-duplex link. The interface is ESD
protected. PS/2 The PS/2 interface supports the serial protocol of standard
PS/2 devices (mouse). The PS/2 serial protocol is
interpreted by the HIF Control interface. Nurse Call The Nurse Call board contains 2 connectors. A phone
jack type connector and a multi-port connector. The
phone jack type connector has a single close-on-alarm
relay. The multi-port connector has three alarm relays
which are configurable to be open or closed on alarm. In
addition, this interface has an audible alert capability for
loss of AC power. MIB The MIB interface allows full-duplex, short-haul
asynchronous binary communication between the monitor
and an arbitrary (medical/non-medical) device using an
eight-pin RJ45 modular connector. Switching between
MIB and RS232 protocol is possible.
Page 31
Monitor Theory of Operation 2 Theory of Operation
Functional Block Description
USB Interface The USB interface allows connection of USB devices
(Mouse, Keyboard, Barcode Scanner, Printer) to the
monitor.
Page 32
2 Theory of Operation Monitor Theory of Operation
32 33
Page 33

3Testing and Maintenance

Introduction

This chapter provides a checklist of the testing and maintenance procedures to ensure the performance and safety of the monitor, the Multi-Measurement Module (MMS),the MMS Extensions and the Flexible Module Rack (FMS) associated modules.
These tests must be performed only by qualified personnel certified by the responsible organization. Qualifications required are: training on the subject, knowledge, experience and acquaintance with the relevant technologies, standards and local regulations. The personnel assessing safety must be able to recognize possible consequences and risks arising from non-conforming equipment.
3
All recurring safety and performance assurance tests must be performed under equal environmental conditions to be comparable.
Preventive Maintenance refers specifically to the series of tests required to make sure the measurement results are accurate. The accuracy and performance procedures are designed to be completed as specified in the following sections or when readings are in question.
For detailed instructions on the maintenance and cleaning of the monitor and its accessories, see
Care and Cleaning, Using Batteries and Maintenance and Troubleshooting in the monitor's Instructions for Use.
Page 34
3 Testing and Maintenance Terminology and Definitions

Terminology and Definitions

The following terms and definitions are used throughout this chapter and taken from the international standards IEC 60601-1, IEC 60601-1-1 and IEC 62353.
- Medical System: a medical electrical system is a combination of at least one medical
electrical device and other electrical equipment, interconnected by functional connection or use of a multiple portable socket-outlet.
- Patient Vicinity: any area in which intentional or unintentional contact can occur between
the patient and parts of the medical system or between the patient and other persons who have had contact with parts of the medical system. The patient vicinity is defined anywhere within 1.5m (5 feet) of the perimeter of the patient's bed and 2.5m (8.2 feet) from the floor.
- Separation Device/Transformer: a component or arrangement of components with input
parts and output parts that, for safety reasons, prevent a transfer of unwanted voltage or current between parts of a medical system.
- Multiple Portable Socket-Outlet: a combination of two or more socket-outlets intended to
be connected to or integrated with flexible cables or cords, which can easily be moved from one place to another while connected to the power mains.
- Functional Connection: an electrical connection for transfer of signals and/or power.
- Tests: Safety or Performance Assurance test procedures which may consist of several steps.
Page 35
Recommended Frequency 3 Testing and Maintenance

Recommended Frequency

Perform the procedures as indicated in the suggested testing timetable. These timetable recommendations do not supersede local requirements.
Tests Frequency
Preventive Maintenance*
Table 1: Suggested Testing Timetable
NBP Performance Once every two years, or more often if
specified by local laws.
Microstream CO2 Calibration Once a year or after 4000 hours of
continuous use and following any instrument repairs or the replacement of any instrument parts.
Other Regular Tests
Performance Assurance Tests
Visual Inspection Power On Test
ECG/Resp Performance ECG Sync Pulse Performance SpO2 Performance NBP Performance Invasive Pressure Performance Temperature Accuracy M3014A Capnography Extension
Performance Tests Microstream CO2 Performance Test Spirometry Accuracy Test C.O. Performance BIS Performance VueLink Performance IntelliBridge Performance Test
Before each use.
Once every two years, or if you suspect the measurement is incorrect, except Mainstream CO2 Accuracy Check, Sidestream CO2 Accuracy Check and Flow Check - required once a year.
Safety Tests
Visual
Electrical
Nurse Call Relay Performance Visual Inspection After each service event
Protective Earth Equipment Leakage Current Patient Leakage Current
Once every two years and after repairs where the power supply has been removed or replaced or the monitor has been damaged by impact.
System Test Once every two years
Page 36
3 Testing and Maintenance When to Perform Tests
*M3015A with the old hardware Rev. A (i.e. Serial No. DE020xxxxx) also require the CO2 pump/CO
scrubber replacement procedure. This is required every three years or after 15000
2
operating hours.
NOTE
The EEG, SvO2 and tcGas parameters do not require performance testing. See EEG, SvO2 and tcGas Performance Tests (on page
103) for details.

When to Perform Tests

This table tells you when to perform specific tests.The corresponding test procedures are described in the following sections All tests listed below must be performed on the monitor itself, any attached MMS/X2 and FMS incl. parameter modules.
When to perform tests
Service Event
(When performing...
Installation Installation of a monitor in combination with
a medical or non-medical device connected to the same multiple socket outlet.
Installation of a monitor with no display connected to the video output
Installation of a monitor with a medical display specified by Philips
Installation of a monitor with an off-the-shelf display (non-compliant with IEC 60601-1)
Installation of a monitor with AGM or IntelliVue G1/G5, connected to separate mains sockets.
Installation of a monitor with a Vuelink connection to another medical device (compliant with IEC 60601-1), connected to separate mains sockets.
Tests Required
...Complete these tests)
Perform Visual Inspection, Power On and System Tests
Perform Visual Inspection and Power On Test
Perform Visual Inspection and Power On Test
Perform Visual Inspection, Power On and System Test
Perform Visual Inspection and Power On Tests
Perform Visual Inspection and Power On Tests
Installation of a monitor with an
Perform Visual Inspection and Power On Tests IntelliBridge connection to another medical device (compliant with IEC 60601-1), connected to separate mains sockets.
Installation of a monitor with IT equipment e.g. printer, PC connected via a functional
Perform Visual Inspection, Power On and
System Tests connection e.g. Centronics or USB.
Page 37
When to Perform Tests 3 Testing and Maintenance
Service Event
(When performing...
Installation of monitor with IntelliVue
802.11 Bedside Adapter
Tests Required
...Complete these tests)
Perform Visual Inspection, Power On and
IntelliVue 802.11 Bedside Adapter
Communication Test
Installation of networked monitor (LAN) Perform Visual Inspection and Power On Test
Preventive Maintenance
Preventive Maintenance* Perform preventive maintenance tests and
procedures:
NBP calibration
Microstream CO2 calibration Other Regular Tests and Tasks
Visual Inspection Perform Visual Inspection
Power On Test Perform Power On test
Repairs Repairs where the monitor, FMS, parameter
modules, MMS or X2 have been damaged by impact, liquid ingression, fire, short circuit or electrical surge.
Repairs where the power supply, the mains socket or an interface board is removed or replaced or the protective earth ground connection is disrupted.
Repairs of IntelliVue 802.11 Bedside Adapter
Repairs of the parameter modules, FMS, MMS or X2 (all service events where the parameter modules, FMS, MMS or X2 have been opened)
Repairs where the NBP pump of the MMS or X2 has been replaced
Perform Visual Inspection, Power On, all
Safety Tests and Full Performance Assurance
Tests
Perform Visual Inspection, Power On, all
Safety Tests and Basic Performance Assurance
Test
Perform Visual Inspection, Power On and
IntelliVue 802.11 Bedside Adapter
Communication Test
Perform Visual Inspection, Power On, all
Safety Tests and Basic Performance Assurance
Test.
If a certain parameter seems suspicious,
perform Full Performance Assurance Test for
this parameter.
Perform Visual Inspection, Power On, all
Safety Tests, Basic Performance Assurance
Test and NBP Performance Test and
Calibration
Page 38
3 Testing and Maintenance When to Perform Tests
Service Event
(When performing...
Tests Required
...Complete these tests)
Repairs of the AGM or IntelliVue G1/G5 Perform Basic Performance Assurance Test.
For further testing requirements, see AGM or
IntelliVue G1/G5 Service Guide Repairs where the parameter module, MMS
or X2 has been replaced. Repairs where the printer connected via
Centronics or USB I/O board has been
Perform Visual Inspection, Power On and
Basic Performance Assurance
Perform Visual Inspection, Power On, System
Test and Printer Test. replaced.
All other IntelliVue Monitoring System repairs (except when power supply is
Perform Visual Inspection, Power On Test and
Basic Performance Assurance Test removed)
Performance Assurance
Basic Performance Assurance Perform basic performance assurance tests for
the respective monitoring system component. Full Performance Assurance Perform all accuracy and performance test
procedures listed in the following sections. If a
particular measurement is in question, perform
the measurement performance test only. Upgrades
Software Upgrades Perform Visual Inspection, Power On Test and
Basic Performance Assurance Test unless
otherwise specified in the Upgrade Installation
Notes shipped with the upgrade. Hardware Upgrades Perform Visual Inspection, Power On Test and
Basic Performance Assurance Test unless
otherwise specified in the Upgrade Installation
Notes shipped with the upgrade. Hardware Upgrades where IntelliVue 802.11
Bedside Adapter is installed
Perform Visual Inspection, Power On Test,
Basic Performance Assurance Test and
IntelliVue 802.11 Bedside Adapter
Communication Test Installation of Interfaces or Hardware
Upgrades where the power supply or
Perform Visual Inspection, Power On Test,
Basic Performance Tests and all Safety Tests parameter boards need to be removed.
Page 39
When to Perform Tests 3 Testing and Maintenance
NOTE
Service Event
(When performing...
Combining or Exchanging System Components
Tests Required
...Complete these tests)
Perform the System Test for the respective
system components
*M3015A with the old hardware Rev. A (i.e. Serial No. DE020xxxxx) also require the pump and scrubber replacement procedures.
It is the responsibility of the facility operator or their designee to obtain reference values for recurring safety and system tests. These reference values are the results of the first test cycles after an installation. You may also purchase this service from Philips.
Page 40
3 Testing and Maintenance Testing Sequence

Testing Sequence

Summary of the recommended sequence of testing:
Start
Select the test
Visual Inspection
Safety Tests
Performance Tests
Reporting of Results
Evaluation of Results
See When to Perform Tests (on page
36)
See Visual Test (see " page
41).
Before Each Use" on
See Safety Test Procedures (on page
44).
See Performance Assurance Tests (on page
84).
See Reporting of Test Results (on page
107)
See Evaluation of Test Results (on page
111)
NOTE
Check and prepare for normal use
If any single test fails, testing must be discontinued immediately and the device under test must be repaired or labeled as defective.
Page 41
Visual Inspection 3 Testing and Maintenance

Visual Inspection

Before Each Use

Check all exterior housings for cracks and damage. Check the condition of all external cables, especially for splits or cracks and signs of twisting. If serious damage is evident, the cable should be replaced immediately. Check that all mountings are correctly installed and secure. Refer to the instructions that accompany the relevant mounting solution.

After Each Service, Maintenance or Repair Event

Ensure all fuses accessible from the outside comply with the manufacturer’s specification. Check:
- the integrity of mechanical parts, internally and external ly .
- any damage or contamination, internally and externally
- that no loose parts or foreign bodies remain in the device after servicing or repair.
- the integrity of all relevant accessories.

Power On Test

1. Connect the monitoring system to mains and switch it on. This includes connected displays,
2. Make sure that all steps listed in the table Initial Instrument Boot Phase in the
The expected test result is pass: the monitor boots up and displays an ECG wave. The wave might be a flat line if no simulator is attached.
MMS, MMS Extensions, X2, FMS and FMS associated modules, gas analyzers and Vuelink devices.
Troubleshooting section are completed successfully and that an ECG wave appears on the screen.
Page 42
3 Testing and Maintenance Safety Tests

Safety Tests

Safety tests are comprised of the following tests performed on the monitoring system:
- protective earth resistance
- equipment leakage current
- applied part leakage current
- system test (if required)
Safety test requirements are set according to international standards, their national deviations and specific local requirements. The safety tests detailed in this Service Guide are derived from international standards but may not be sufficient to meet local requirements. We recommend that you file the results of safety tests. This may help to identify a problem early particularly if the test results deteriorate over a period of time.
Each individual piece of equipment of the monitoring system which has its own connection to mains or which can be connected or disconnected from mains without the use of a tool must be tested individually. The monitoring system as a whole must be tested according to the System Test (on page
69) procedure.
Accessories of the monitoring system which can affect the safety of the equipment under test or the results of the safety test must be included in the tests and documented.
Page 43
Safety Tests 3 Testing and Maintenance

Warnings, Cautions, and Safety Precautions

- These tests are well established procedures of detecting abnormalities that, if undetected,
could result in danger to either the patient or the operator.
- Disconnect the device under test from the patient before performing safety tests.
- Disconnect the device under test from mains before performing safety tests. If this is not
possible, ensure that the performance of these tests does not result in danger to the safety analyzer operator, patients or other individuals.
- Test equipment (for example, a Safety Analyzer) is required to perform the safety tests.
Please refer to Annex C of IEC/EN 62353 for exact requirements for the measurement equipment and for measurement circuits for protective earth resistance and leakage currents. Refer to the documentation that accompanies the test equipment. Only certified technicians should perform safety testing.
- The consistent use of a Safety Analyzer as a routine step in closing a repair or upgrade is
emphasized as a mandatory step to maintain user and patient safety. You can also use the Safety Analyzer as a troubleshooting tool to detect abnormalities of line voltage and grounding plus total current loads.
- During safety testing, mains voltage and electrical currents are applied to the device under
test. Ensure that there are no open electrical conductive parts during the performance of these tests. Avoid that users, patients or other individuals come into contact with touch voltage.
- For Europe and Asia/Pacific, the monitor complies with:
IEC60601-1:1988 + A1:1991 + A2:1995 = EN60601-1:1990 +A1:1993 + A2:1995 IEC60601-1-1:2000 For USA, the monitor complies with: UL60601-1 For Canada, CAN/CSA C22.2#601.1-M90
- Local regulations supersede the testing requirements listed in this chapter.
- If a non-medical electrical device is connected to a medical electrical device, the resulting
medical electrical system must comply with IEC/EN 60601-1-1.
- Perform safety tests as described on the following pages.
Page 44
3 Testing and Maintenance Safety Tests

Safety Test Procedures

Use the test procedures outlined here only for verifying safe installation or service of the product. The setups used for these tests and the acceptable ranges of values are derived from local and international standards but may not be equivalent. These tests are not a substitute for local safety testing where it is required for an installation or a service event. If using an approved safety tester, perform the tests in accordance with the information provided by the manufacturer of the tester and in accordance with your local regulations, for example IEC/EN 60601-1, UL60601-1 (US), IEC/EN 62353, and IEC/EN 60601-1-1. The safety tester should print results as detailed in this chapter, tog ether with other data.
Please refer to Annex C of IEC/EN 62353 for requirements for the measurement equipment and for measurement circuits for protective earth resistance and leakage currents.
The following symbols are used in the diagrams illustrating the safety tests:
CAUTION
Supply mains
Protective earth
L, N Supply mains terminals PE Protective earth terminal
Mains part
F-type applied part
Applied part
Measuring device
Resistance measuring device
......... Optional connection
Connection to accessible conductive parts
After each service, maintenance or repair event: Ensure all fuses accessible from the outside comply with the manufacturer’s specification. Check:
- the integrity of mechanical parts, internally and external ly .
- any damage or contamination, internally and externally.
- that no loose parts or foreign bodies remain in the device after servicing or repair.
- the integrity of all relevant accessories.
Page 45
Safety Tests 3 Testing and Maintenance
Hints for Correct Performance of Safety Tests
- Perform a visual inspection on all detachable power cords used with the monitoring system
and include these in all safety test procedures.
- Connection lines such as data lines or functional earth conductors may appear to act like
protective earth connections. These may lead to incorrect measurements and need to be considered during testing. If necessary, unplug these connections.
- Position all cables and cords in such a manner that they do not influence the safety tests.
- Measurement of insulation resistance is not required.
Page 46
3 Testing and Maintenance Safety Tests
Guideline for Performance of Safety Tests
Connect the detachable power cord of the device under test to the safety analyzer's test mains port. Connect the enclosure test lead of the safety analyzer to the enclosure of the device under test, e.g. to the equipotential connector. For testing the applied part leakage current, connect all applied parts to the safety analyzer using the appropriate patient lead or adapter cable. For the ECG parameter all ten ECG-leads need to be connected to the safety analyzer. If necessary, use an adapter cable to connect all ten ECG-leads. If necessary, repeat the safety test procedure until all available applied parts have been tested. Refer to the documentation that accompanies the safety analyzer for further details on how to set up and perform the test.
Protective Earth Resistance Test - Setup Example
Equipment Leakage Current Test - Setup Example
Page 47
Safety Tests 3 Testing and Maintenance
Applied Part Current Test - Setup Example
NOTE
The above graphics resemble the Metron QA-90 setup and are protected by copyright. Copyright owned by Fluke (Metron).
Page 48
3 Testing and Maintenance Safety Tests
Safety Test Adapter Cable - Schematics
The following graphics provide schematics of safety test (patient lead) adapter cables which can be used for electrical safety testing. These schematics can also be used as a guideline for making your own safety test adapter cables. Alternatively, other methods to make safety test adapter cables can be used, e.g. using a modified accessory cable.
NOTE
You may not need all of the cables displayed below for electrical safety testing of your respective monitor.
ECG:
Page 49
Safety Tests 3 Testing and Maintenance
SpO2 (MP2/X2, MP5, M3001A & M1020B #A01, #A02, #A03):
Page 50
3 Testing and Maintenance Safety Tests
SpO2 (M1020A):
Page 51
Safety Tests 3 Testing and Maintenance
Invasive Pressure:
M1006B #C01:
Page 52
3 Testing and Maintenance Safety Tests
Temperature:
Page 53
Safety Tests 3 Testing and Maintenance
CO2 (MP5, M3014A):
Page 54
3 Testing and Maintenance Safety Tests
CO2 (M1016A, M3016A):
4 = all resistors 120 KOhm
Page 55
Safety Tests 3 Testing and Maintenance
Cardiac Output:
Page 56
3 Testing and Maintenance Safety Tests
BIS: Use Clamp Adapter Cable and M1034-61650 BIS sensor simulator.
Page 57
Safety Tests 3 Testing and Maintenance
VueLink:
4 = 220 Ohm
Page 58
3 Testing and Maintenance Safety Tests
IntelliBridge:
Page 59
Safety Tests 3 Testing and Maintenance
EEG:
Page 60
3 Testing and Maintenance Safety Tests
SvO2 (M1021A):
Page 61
Safety Tests 3 Testing and Maintenance
ScvO2 (M1011A):
Page 62
3 Testing and Maintenance Safety Tests
tcpO2/tcpCO2:
Page 63
Safety Tests 3 Testing and Maintenance
MP5 predicitive Temperature:
Page 64
3 Testing and Maintenance Safety Tests
MP5 TAAP:
Page 65
Safety Tests 3 Testing and Maintenance
S(1): Protective Earth Resistance Test
Test to perform:
Measuring circuit for the measurement of Protective Earth Resistance in medical electrical equipment that is disconnected from the supply mains.
NOTE
This measures the impedance of the Protective Earth (PE) terminal to all exposed metal parts of the Instrument under Test (IUT), which are for safety reasons connected to the Protective Earth (PE).
Measurements shall be performed using a measuring device capable to deliver a current of at least 200 mA into 500 mOhms with maximum 24V
This safety test is based on IEC/EN 60601-1, IEC/EN 62353, UL2601-1 Ed. 2/UL60601-1:2003 and CSA 601.1-M90.
For measurement limits, refer to Safety (1) test, Test and Inspection Matrix. Report the highest value (X1).
Test Expected test results
Protective Earth Resistance Test (with
X1 <= 300mOhms
mains cable)
- If the protective earth resistance test fails, testing must be discontinued immediately and the
device under test must be repaired or labeled as defective.
- All values for current and voltage are the root mean square (r.m.s.) values, unless otherwise
stated.
- Flex the power cord during the protective earth resistance test to evaluate its integrity. If it
does not pass the test, exchange the power cord.
Page 66
3 Testing and Maintenance Safety Tests
S(2): Equipment Leakage Current Test - Normal Condition
Test to perform:
Measuring circuit for the measurement of Equipment Leakage Current - Direct method according to IEC/EN 62353.
NOTE
This test measures leakage current of exposed metal parts of the monitor and the functional earth leakage current. It tests normal and reversed polarity. Perform the test with S1 closed (Normal Condition).
There are no parts of the equipment that are not protectively earthed. This safety test is based on IEC/EN 60601-1, IEC/EN 62353, UL2601-1 Ed. 2/UL60601-1:2003
and CSA 601.1-M90. For measurement limits, refer to Safety (2) test, Test and Inspection Matrix. Report the highest value (X1).
Test Expected test results
Equipment Leakage Current Test
X1 <= 100μA
(Normal Condition - with mains cable)
All values for current and voltage are the root mean square (r.m.s.) values, unless otherwise stated.
Page 67
Safety Tests 3 Testing and Maintenance
S(3): Equipment Leakage Current Test - Single Fault Condition
Test to perform:
Measuring circuit for the measurement of Equipment Leakage Current - Direct method according to IEC/EN 62353.
This test measures leakage current of exposed metal parts of the monitor and the functional earth leakage current. It tests normal and reversed polarity. Perform the test with S1 open (Single Fault Condition).
NOTE
There are no parts of the equipment that are not protectively earthed. This safety test is based on IEC/EN 60601-1, IEC/EN 62353, UL2601-1 Ed. 2/UL60601-1:2003
and CSA 601.1-M90. For measurement limits, refer to Safety (3) test, Test and Inspection Matrix. Report the highest value (X2).
Test Expected test results
Equipment Leakage Current Test
X2 <= 300μA
(Single Fault Condition - with mains cable)
All values for current and voltage are the root mean square (r.m.s.) values, unless otherwise stated.
Page 68
3 Testing and Maintenance Safety Tests
S(4): Applied Part Leakage Current - Mains on Applied Part
NOTE
During measurement of the Applied Part Leakage Current it is possible that the measured current can exceed the allowed limit (per IEC/EN 60601-1 or IEC/EN 62353).
This can occur when the safety tester is connected to the invasive blood pressure and temperature connectors at the same time during the applied leakage current measurement.
The connectors for the invasive blood pressure and temperature are independently functioning connectors.
Although there are individual connectors on the front end, internally those parameters use the same electrical insulation interface and are hardwired to each other. This results in an electrical short of those connectors during measurement if a test current is applied simultaneously . Therefore this should be avoided.
Due to the combined insulation interface, it is sufficient to connect to only one parameter interface (that is, Invasive Blood Pressure or Temperature) of the invas ive bl ood pressure/temperature measurement block. This avoids a short and the potential of exceeding the limit for the current.
Test to perform:
Page 69
Safety Tests 3 Testing and Maintenance
Measuring circuit for the measurement of Applied Part Leakage Current - Direct method according to IEC/EN 62353.
This test measures applied part leakage current from applied part to earth caused by external main voltage on the applied part. Each polarity combination possible shall be tested. This test is applicable for ECG measurement inputs.
There are no parts of the equipment that are not protectively earthed. This safety test is based on IEC/EN 60601-1, IEC/EN 62353, UL2601-1 Ed. 2/UL60601-1:2003
and CSA 601.1-M90. For measurement limits and test voltage, refer to Safety (4) test, Test and Inspection Matrix. Report the highest value. (X1).
Test Expected test results
Applied Part Leakage Current Test
S4 <= 50μA
(Single Fault Condition - mains on applied part)
NOTE
All values for current and voltage are the root mean square (r.m.s.) values, unless otherwise stated.
System Test
After mounting and setting up a system, perform system safety tests according to IEC/EN 60601-1-1.
What is a Medical Electrical System?
A medical electrical system is a combination of at least one medical electrical piece of equipment and other electrical equipment, interconnected by functional connection or use of a multiple portable socket-outlet.
- Devices forming a medical electrical system must comply with IEC/EN 60601-1-1.
- Any electrical device such as IT equipment that is connected to the medical electrical
equipment must comply with IEC/EN 60601-1-1 and be tested accordingly.
Page 70
3 Testing and Maintenance Safety Tests
General Requirements for a System
After installation or subsequent modification, a system must comply with the requirements of the system standard IEC/EN 60601-1-1. Compliance is checked by inspection, testing or analysis, as specified in the IEC/EN 60601-1-1 or in this book.
Medical electrical equipment must comply with the requirements of the general standard IEC/EN 60601-1, its relevant particular standards and specific national deviations. Non-medical electrical equipment shall comply with IEC safety standards that are relevant to that equipment.
Relevant standards for some non-medical electrical equipment may have limits for equipment leakage currents higher than required by the standard IEC/EN 60601-1-1. These higher limits are acceptable only outside the patient environment. It is essential to reduce equipment leakage currents to values specified in IEC 60601-1 when non-medical electrical equipment is to be used within the patient environment.
System Example
This illustration shows a system where both the medical electrical equipment and the non-medical electrical equipment are situated at the patient’s bedside.
Page 71
Safety Tests 3 Testing and Maintenance
WARNING
- Do not use additional AC mains extension cords or multiple portable socket-outlets. If a
multiple portable socket-outlet is used, the resulting system must be compliant with IEC/EN 60601-1-1. Do not place multiple socket-outlets on the floor. Do not exceed the maximum permitted load for multiple socket-outlets used with the system. Do not plug additional multiple socket outlets or extension cords into multiple socket outlets or extension cords used within the medical electrical system.
- Do not connect any devices that are not supported as part of a system.
- Do not use a device in the patient vicinity if it does not comply with IEC/EN 60601-1. The
whole installation, including devices outside of the patient vicinity, must comply with IEC/EN 60601-1-1. Any non-medical device placed and operated in the patient’s vicinity must be powered via a separating transformer (compliant with IEC/EN 60601-1-1) that ensures mechanical fixing of the power cords and covering of any unused power outlets.
Page 72
3 Testing and Maintenance Safety Tests
System Installation Requirements
- Ensure that the the medical electrical system is installed in a way that the user achieves
optimal use.
- Make sure the user is informed about the required cleaning, adjustment, sterilization and
disinfection procedures listed in the Instructions for Use.
- The medical electrical system must be installed in such a way that the user is able to carry
out the necessary cleaning, adjustment, sterilization and disinfection procedures listed in the Instructions for Use.
- Ensure that the medical electrical system is installed in a way that an interruption and
restoration of power to any part of the medical electrical system does not result in a safety hazard.
- We recommend using fixed mains socket outlets to power the medical system or parts
thereof. Avoid using multiple portable socket-outlets.
- Any multiple portable socket outlets used must be compliant with IEC 60884-1 and IEC
60601-1-1.
- Ensure that any part of the system connected to multiple portable socket-outlets is only
removable with a tool, i.e. the multiple portable socket-outlet provides a locking mechanism to prevent power cords from being plugged or unplugged unintentionally. Otherwise, the multiple portable socket-outlet must be connected to a separation device. Multiple Socket Outlets used within the medical electrical system must only be used for powering medical electrical equipment which is part of the system.
- Ensure that any functional connections between parts of the medical electrical system are
isolated by a separation device according to IEC 60601-1-1 to limit increased equipment leakage currents caused by current flow through the signal connections. This only works if the equipment leakage current of the respective medical electrical system parts is not exceeded under normal conditions.
- Avoid increase of equipment leakage currents when non-medical electrical equipment within
the medical electrical system is used. This only works if the equipment leakage current of the respective medical electrical system parts is not exceeded under normal conditions. Use additional protective earth connection, separation device or additional non-conductive enclosures.
- Within the patient environment it is important to limit electrical potential differences
between different parts of a system. If necessary, use potential equalization equipment (equipotential cable) or additional protective earth connections.
- Medical electrical equipment used in medical rooms must be connected to potential
equalization equipment (equipotential cable) to avoid electrical potential differences. Check your local requirements for details.
Page 73
Safety Tests 3 Testing and Maintenance
Required Protective Measures at System Installation
For any IT equipment (IEC60950) operated in patient vicinity ensure that the equipment leakage current does not exceed the limits described in IEC 60601-1. Use a separation device to ensure compliance. After installation of IT equipment in patient vicinity, an enclosure leakage current test is required.
Page 74
3 Testing and Maintenance Safety Tests
Case 1: Medical Device Combined with Medical Device
If you combine a medical device with another medical device (incl. Philips specified displays) to form a medical electrical system according to IEC60601-1-1, no additional protective measures are required. The medical electrical devices may be located in or outside the patient vicinity in a medically used room. This is valid as long as the medical devices are connected to separate mains outlets. No system test is required.
Page 75
Safety Tests 3 Testing and Maintenance
If the combined medical devices are connected to the same multiple portable socket outlet an enclosure leakage current test of the entire device combination on the multiple portable socket outlet is required to ensure that the resulting protect ive earth leakage current and equipment leakage current does not exceed the limits of IEC 60601-1-1. Avoid using multiple portable socket outlets. The medical electrical devices may be located in or outside the patient vicinity in a medically used room. If the limits are exceeded, additional protective measures are required, e.g. a separation device or the connection of each device to separate mains.
Page 76
3 Testing and Maintenance Safety Tests
Case 2: Medical Device Combined with a Non-Medical Device
If you combine a medical device with a non-medical device to form a medical electrical system according to IEC60601-1-1, additional protective measures are required, e.g. usage of a separation device. The medical electrical devices or the IT equipment may be located in or outside the patient vicinity in a medically used room. After system installation incl. protective measures, a system test is required to ensure that the resulting equipment leakage current and applied part leakage current does not exceed the limits of IEC 60601-1-1.
Page 77
Safety Tests 3 Testing and Maintenance
For any IT equipment (IEC60950) operated in patient vicinity ensure that the equipment leakage current does not exceed the limits described in IEC 60601-1. Use a separation device to ensure compliance. After installation of IT equipment in patient vicinity, an enclosure leakage current test is required.
Page 78
3 Testing and Maintenance Safety Tests
If the combined devices forming the medical electrical system are connected to the same multiple portable socket outlet, ensure that the resulting protective earth leakage current and equipment leakage current do not exceed the limits of IEC 60601-1-1. The medical electrical devices or IT equipment may be located in or outside the patient vicinity in a medically used room. Avoid using multiple portable socket outlets. If the limits of IEC 60601-1-1 are exceeded, additional protective measures are required, e.g. a separation device or the connection of each device to separate mains.
Page 79
Safety Tests 3 Testing and Maintenance
For any IT equipment (IEC60950) operated in patient vicinity ensure that the equipment leakage current does not exceed the limits described in IEC 60601-1. Use a separation device to ensure compliance. After installation of IT equipment in patient vicinity, an enclosure leakage current test is required.
Page 80
3 Testing and Maintenance Safety Tests
Case 3: Medical Device Combined with a Medical or Non-Medical Device with one Device in a Non-Medically-Used Room
If you combine a medical device with a medical or non-medical device to form a medical electrical system according to IEC60601-1-1 using a common protective earth connection and one of the devices is located in a non-medically used room, additional protective measures are required, e.g. usage of a separation device or additional protective eart h connec tion. The medical electrical devices or IT equipment may be located in or outside the patient vicinity. After system installation incl. protective measures, a system test is required to ensure that the resulting equipment leakage current does not exceed the limits of IEC 60601-1-1.
Page 81
Safety Tests 3 Testing and Maintenance
Page 82
3 Testing and Maintenance Safety Tests
If you combine a medical device with a medical or non-medical device to form a medical electrical system according to IEC60601-1-1 using two separate protective earth connections and one of the devices is located in a non-medically used room creating a potential voltage difference, additional protective measures are required, e.g. usage of a separation device or additional protective earth connection. The medical electrical devices or IT equipment may be located in or outside the patient vicinity. After system installation incl. protective measures, a system test is required to ensure that the resulting equipment leakage current does not exceed the limits of IEC 60601-1-1.
Page 83
Safety Tests 3 Testing and Maintenance
System Test Procedure
If the medical electrical device has already been tested as a standalone device e.g. during factory safety testing, an equipment leakage current test must only be performed once the device is connected to the LAN network. If the medical electrical system has not been tested as a standalone device, the device has to be tested as a standalone device (without connection to the system) and as part of the system (with connection to the system).
Connect the detachable power cord of the device under test to the safety analyzer's test mains port. Connect the enclosure test lead of the safety analyzer to the enclosure of the device under test, e.g. to the equipotential connector. Refer to the documentation that accompanies the safety analyzer for further details on how to set up the test.
Test Expected test results
Equipment Leakage Current Test (Normal Condition)
Equipment Leakage Current Test (Single Fault Condition)
Sys1 <= 100μA
Sys2 <= 300μA
Page 84
3 Testing and Maintenance Preventive Maintenance Procedures
After the testing of the device as a standalone device and as part of the system, check that the resulting values (without connection and with connection to the system) do not differ by more than +/- 10% from each other.
If the devices in the medical electrical system are connected to a multiple portable socket outlet the resulting protective earth leakage current needs to be determined. All system components must be connected to the multiple portable socket outlet and be switched on during this measurement.
Test Expected test results
Protective Earth Leakage Current of
Sys3 <= 300μA
Multiple Socket Outlets Refer to the documentation that accompanies the safety analy z er for further detai ls on how to s et
up the test.

Preventive Maintenance Procedures

Noninvasive Blood Pressure Measurement Calibration

Carry out the noninvasive blood pressure measurement performance tests at least every two years , or as specified by local laws (whichever comes first).

Performance Assurance Tests

Some of the following test procedures must be performed in service mode. To enter service mode select Operating Modes in the main menu. Then select Service Mode and enter the
password. If required, open the screen menu in the monitor info line at the top of the screen and select
Service to access the service screen. This is required particularly for Anesthetic Gas Module testing procedures.

Basic Performance Assurance Test

This section describes the basic performance test procedure. Please refer to the section When to Perform Tests (on page
Procedure:
36) for detailed information on when which test procedure is required.
Power on the monitoring system and go into demo mode. Check that each connected parameter (integrated, module, MMS, Gas Analyzer, Vuelink connected device) displays values.
Page 85
Performance Assurance Tests 3 Testing and Maintenance

Full Performance Assurance Test

The following sections describe the full performance testing procedures i.e. detailed testing of each parameter with a patient simulator or specified tools. Please refer to the section When to perform Tests (on page
36) for information on when which testing procedure is required.

ECG/Resp Performance Test

This test checks the performance of the ECG and respiration measurements. Tools required: Patient simulator.
ECG Performance
1. Connect the patient simulator to the ECG/Resp connector.
2. Configure the patient simulator as follows:
- ECG sinus rhythm.
- HR = 100 bpm or 120 bpm (depending on your patient simulator).
3. Check the displayed ECG wave and HR value against the simulator configuration.
4. The value should be 100bpm or 120 bpm+/- 2 bpm.
Respiration Performance
1. Change the Patient Simulator configuration to:
- Base impedance line 1500 Ohm.
- Delta impedance 0.5 Ohm.
- Respiration rate 40 rpm or 45 rpm.
2. The value should be 40 rpm +/- 2 rpm or 45 rpm +/- 2 rpm.
Test Expected test results
ECG Performance Test 100bpm +/- 2bpm or
Respiration Performance Test 40 rpm +/- 2 rpm or
120bpm +/- 2bpm
45 rpm +/- 2 rpm
Page 86
3 Testing and Maintenance Performance Assurance Tests

ECG Sync Performance Test

This test checks the performance of ECG synchronization between the monitor and a defibrillator. It only needs to be performed when this feature is in use as a protocol at the customer site.
Tools required:
- Defibrillator with ECG Sync and Marker Output.
- Patient simulator.
1. Connect the patient simulator to the ECG connector and the defibrillator to the ECG Sync
Output on the monitor.
2. Set the patient simulator to the following configuration:
- HR = 100 bpm or 120 bpm (depending on your patient simulator).
- ECG sinus rhythm.
3. Switch the defibrillator to simulation mode.
4. Check that the marker pulse is displayed before the T-wave begins.
Test Expected test results
ECG Sync Performance Test Marker pulse is displayed before

SpO2 Performance Test

This test checks the performance of the SpO2 measurement. Tools required: none
1. Connect an adult SpO2 transducer to the SpO2 connector.
2. Measure the SpO
3. The value should be between 95% and 100%.
Test Expected test results
SpO2 Performance Test 95% and 100%
the T-wave begins
value on your finger (this assumes that you are healthy).
2
Page 87
Performance Assurance Tests 3 Testing and Maintenance
Measurement Validation
The SpO2 accuracy has been validated in human studies against arterial blood sample reference measured with a CO-oximeter. In a controlled desaturation study, healthy adult volunteers with saturation levels between 70% and 100% SaO2 were studied. The population characteristics for those studies were:
- about 50% female and 50% male subjects
- age range: 18 to 45
- skin tone: from light to black
NOTE
A functional tester cannot be used to assess the accuracy of a pulse oximeter monitor. However, it can be used to demonstrate that a particular pulse oximeter monitor reproduces a calibration curve that has been independently demonstrated to fulfill a particular accuracy specification.

NBP PerformanceTest

This section describes NBP test procedures.The monitor must be in service mode and the screen “Service A” must be selected to perform these tests. The NBP Performance Test consists of:
- NBP Accuracy Test
- NBP Leakage Test
- NBP Linearity Test
- Valve Test
NBP Accuracy Test
This test checks the performance of the non-invasive blood pressure measurement. Connect the equipment as shown:
Page 88
3 Testing and Maintenance Performance Assurance Tests
Tools required:
- Reference manometer (includes hand pump and valve), accuracy 0.2% of reading.
- Expansion chamber (volume 250 ml +/- 10%)
- Appropriate tubing.
In service mode, the systolic and diastolic readings indicate the noise of NBP channels 1 and 2 respectively. When static pressure is applied, the reading in NBP channel 1 should be below 50. The value in parentheses indicates the actual pressure applied to the system.
1. Connect the manometer and the pump with tubing to the NBP connector on the MMS and to
the expansion chamber.
2. In service mode, select the Setup NBP menu.
3. Select Close Valves: On
4. Raise the pressure to 280 mmHg with the manometer pump.
5. Wait 10 seconds for the measurement to stabilize.
6. Compare the manometer values with the displayed values.
7. Document the value displayed by the monitor (x1).
8. If the difference between the manometer and displayed values is greater than 3 mmHg,
9. To calibrate the MMS, select Close Valves off then Calibrate NBP and wait for the
10. Press Confirm. If the INOP NBP Equipment Malfunction message occurs in monitoring mode, go back to
service mode and repeat the calibration procedure.
NBP Leakage Test
The NBP leakage test checks the integrity of the system and of the valve. It is required once every two years and when you repair the monitor or replace parts.
1. If you have calibrated, repeat steps 2 to 6 from the accuracy test procedure so that you have
calibrate the MMS. If not, proceed to the leakage test.
instrument to pump up the expansion chamber.Wait a few seconds after pumping stops until EnterPrVal is highlighted and then move the cursor to the value shown on the manometer. If one of the following prompt messages appears during this step, check whether there is leakage in the setup:
- NBP unable to calibrate–cannot adjust pressure
- NBP unable to calibrate–unstable signal
280 mmHg pressure on the expansion chamber.
2. Watch the pressure value for 60 seconds.
3. Calculate and document the leakage test value (x2).
x2 = P1 - P2 where P1 is the pressure at the beginning of the leakage test and P2 is the pressure displayed after 60 seconds. The leakage test value should be less than 6 mmHg.
Page 89
Performance Assurance Tests 3 Testing and Maintenance
NBP Linearity Test
1. Reduce the manometer pressure to 150 mmHg.
2. Wait 10 seconds for the measurement to stabilize.
3. After these 10 seconds, compare the manometer value with the displayed value.
4. Document the value displayed by the monitor (x3)
5. If the difference is greater than 3 mmHg, calibrate the MMS (see steps 9 to 10 in the
accuracy test procedure).
Valve Test
1. Raise the pressure again to 280 mmHg.
2. Select Close valves: Off.
3. Wait five seconds and then document the value displayed. The value should be less than 10
mmHg.
4. Document the value displayed by the monitor (x4).
Expected Test Results for NBP Accuracy Test, Leakage Test, Linearity Test & Valve Test
Test Expected test results
Accuracy test x1 = value displayed by monitor
Difference 3mmHg
Leakage test x2 = leakage test value
x2 < 6 mmHg
Linearity test x3 = value displayed by monitor
Difference 3mmHg
Valve Test x4 = value < 10 mmHg

Invasive Pressure Performance Test

This test checks the performance of the invasive pressure measurement. Tools required: Patient simulator.
1. Connect the patient simulator to the pressure connector.
2. Set the patient simulator to 0 pressure.
3. Make a zero calibration.
4. Configure the patient simulator as P(static) = 200 mmHg.
5. Wait for the display.
Page 90
3 Testing and Maintenance Performance Assurance Tests
6. The value should be 200 mmHg ± 5 mmHg. If the value is outside these tolerances, calibrate
the Invasive Pressure measurement. If the measurement was calibrated with a dedicated reusable catheter, check the calibration together with this catheter.
Table 4:
Test Expected test results
Invasive Pressure Performance Test 200 mmHg ± 5 mmHg

Temperature Performance Test

This test checks the performance of the temperature measurement. Tools required: Patient simulator (with 0.1°C or 0.2°F).
1. Connect the patient simulator to the temperature connector.
2. Configure the patient simulator to 40°C or 100°F.
3. The value should be 40°C ± 0.2°C or 100°F ± 0.4°F.
Table 2:
Test Expected test results
Temperature Performance Test 40°C ± 0.2°C or 100°F ± 0.4°F

M3014A Capnography Extension Performance Tests

The procedures below describe the mainstream and sidestream CO2 performance tests for the M3014A Capnography Extension.
Mainstream CO2 Accuracy Check
Tools Required:
- three airway adapters
- Verification Gas M2506A
- Gas cylinder regulator M2505A
You also need a local barometric pressure rating received from a reliable local source (airport, regional weather station or hospital weather station) which is located at the same altitude as the hospital.
Procedure:
1. Attach the M2501A CO
sensor to the patient monitor. Attach an airway adapter to the
2
sensor. Make sure that the sensor is disconnected from the patient circuit.
2. Switch on the patient monitor.
3. Enter the monitor’s Service Mode.
Page 91
Performance Assurance Tests 3 Testing and Maintenance
4. Using the sensor status provided in the M2501A Serial protocol, wait for the M2501A
sensor to warm up to its operating temperature.
5. The default setting for gas temperature is 22°C. If the gas temperature is significantly above
or below this value, correct the gas temperature setting.
6. Zero the sensor on the airway adapter being used in this test. Ensure Zero Gas is set to Room
Air
7. Attach a regulated flowing gas mixture of 5% CO2, balance N2 to the airway adapter.
8. Set the gas correction to off.
9. Allow a few seconds for the gas mixture to stabilize and observe the CO2 value. The
expected value is 5% of the ambient pressure ±2mmHg
NOTE
Make sure that you follow the above steps correctly. If the sensor fails this check it must be exchanged. The sensor cannot be calibrated.
Example for an expected test result: The expected test result for an altitude of 0 m (sea level) at approximately 760 mmHg ambient pressure is:
Table 6:
Test Expected test results (x1) Acceptance Range
Mainstream CO2 Accuracy Test
NOTE
The expected test results will differ depending on the conditions (i.e. altitude or ambient pressure).
Sidestream CO2 Accuracy Check
Tools Required:
- Cal gas flow regulator M2267A
- Cal tube 13907A
- Verification Gas M2506A
- Straight Sample Line M2776A
You also need a local barometric pressure rating received from a reliable local source (airport, regional weather station or hospital weather station) which is located at the same altitude as the hospital.
Procedure:
5% of 760 mmHg pressure ±2mmHg
36 mmHg ­40 mmHg
1. Attach the M2741A CO2 sensor to the patient monitor. Attach the sample line and the cal
tube to the sensor. Make sure that the sensor is disconnected from the patient circuit.
2. Switch on the patient monitor.
3. Enter the monitor’s Service Mode.
Page 92
3 Testing and Maintenance Performance Assurance Tests
4. Using the sensor status provided in the M2741A Serial protocol, wait for the M2741A
sensor to warm up to its operating temperature.
5. Zero the sensor. Ensure Zero Gas is set to Room Air
6. Attach a regulated flowing gas mixture of 5% CO2, balance N2 to the cal tube.
7. Set the gas correction to off.
8. Allow a few seconds for the gas mixture to stabilize and observe the CO2 value. The
expected value is 5% of the ambient pressure ±2mmHg
NOTE
Make sure that you follow the above steps correctly. If the sensor fails this check it must be exchanged. The sensor cannot be calibrated
Example for an expected test result: The expected test result for an altitude of 0 m (sea level) at approximately 760 mmHg ambient pressure is:
Test Expected test results (x2) Acceptance Range
Sidestream CO2 Accuracy Test
5% of 760 mmHg pressure ±2mmHg
36 mmHg ­40 mmHg
NOTE
The expected test results will differ depending on the conditions (i.e. altitude or ambient pressure).
Sidestream CO2 Flow Check
Check the flow rate in the Sidestream CO2 extension as follows:
1. Connect the flowmeter to the sample line
2. Check on the flowmeter the flow that the Sidestream CO2 extension pump draws. It should
be 50 ml/min ± 10 ml/min. If the value is not within tolerance check your setup again and perform another flow check. If it fails again, the sensor must be replaced. The sensor cannot be calibrated.
Page 93
Performance Assurance Tests 3 Testing and Maintenance

Microstream CO2 Performance Test

Allow five seconds between individual service procedures to ensure stable equipment conditions. When certain monitor procedures are running, service procedures are not possible and trying to start them will result in a message Service Operation Failed in the monitor’s status line. Wait until the monitor completes the current operation, then restart the service procedure.
This test checks the performance of the Microstream CO2 measurement. The Microstream CO2 measurement can either be integrated into the IntelliVue MP5 monitor or, for other IntelliVue monitors, into the M3015A MMS Extension. The Microstream CO2 performance test is required once per year and when the instrument is repaired or when parts are replaced.
This test uses calibration equipment that you can order (see the Parts section for the part number). The procedure is summarized in the following steps. Refer to the documentation accompanying the equipment for detailed instructions.
Tools Required:
- Standard tools, such as screwdriver, tweezers
- Electronic flowmeter, M1026-60144
- Gas calibration equipment:
- Cal 1 gas 15210-64010 (5% CO
- Cal 2 gas 15210-64020 (10% CO
)
2
)
2
- Cal gas flow regulator M2267A
- Cal tube 13907A
- Calibration Line M3015-47301
You also need a local barometric pressure rating received from a reliable local source (airport, regional weather station or hospital weather station) which is located at the same altitude as the hospital.
The CO2 calibration for the Microstream extension consists of the following steps:
- Leakage check
- Barometric pressure check and calibration, if required.
- Pump check
- Flow check and calibration, if required
- Noise check
- CO2 Cal check and calibration, if required
- CO2 Cal verification using 2nd cal gas
Perform all checks in the same session.
Page 94
3 Testing and Maintenance Performance Assurance Tests
Leakage Check
The leakage check consists of checking the tubing between:
- the pump outlet and the mCO
outlet and
2
- the pump inlet and FilterLine inlet. Check the user’s guide of the flowmeter for details on how to make a correct flow reading.
Part 1
1. Go into service mode and select Setup CO2 menu.
2. Connect a FilterLine to the Microstream CO
input to start the pump running.
2
3. Check the ambient pressure and the cell pressure shown in the monitor’s status line. The cell
pressure should be approximately 20 mmHg lower than ambient pressure.
4. Connect the flowmeter outlet to the FilterLine inlet using a flexible connecting tube.
5. Block the mCO
outlet using your fingertip and observe the flowmeter display. The value on
2
the flowmeter (x1) should decrease to between 0 and 4 ml/min, accompanied by an audible increase in pump noise. If the value is within the tolerance limits, continue with part 2 of the leakage check.
6. If the value is outside the tolerance limits, there is a leakage between the pump outlet and the
mCO
outlet.
2
7. Open the MMS Extension or MP5 and check the tubing connections at the pump outlet and
the extension gas outlet. If the connections are good, then there is a leakage in the tubing and you must exchange the MMS Extension or the mCO
Assembly of the MP5 respectively.
2
Part 2
1. Disconnect the flowmeter from the Part 1 setup and connect the flowmeter inlet to the
M3015A gas outlet or the MP5 mCO
2. Leave the Filterline connected to the M3015A inlet or the MP5 mCO
gas outlet.
2
inlet..
2
3. Block the inlet of the FilterLine using your fingertip and observe the flowmeter display. The
value on the flowmeter (x2) should decrease to between 0 and 4 ml/min, accompanied by an audible increase in pump noise. The cell pressure shown in the status line on the display should decrease to between 300 and 500 mmHg. Do not block the inlet for longer than 25 seconds as this will lead to an “Occlusion” INOP. If the value is within the tolerance limits, there are no leakages and the leakage check is completed; proceed to the pump check.
4. If the value is not within the tolerance limits, there is a leakage between the FilterLine inlet
and the pump inlet.
5. Check the FilterLine connections and open the M3015A or MP5 to check the tubing
connections at the pump inlet and the M3015A or MP5 mCO
gas inlet. If the connections
2
are good, try replacing the FilterLine and repeating the leakage check. If the situation remains, there is a leakage in the tubing and the M3015A or the mCO must be exchanged.
assembly of the MP5
2
Page 95
Performance Assurance Tests 3 Testing and Maintenance
Barometric Pressure Check and Calibration
Check the barometric pressure value in the M3015A MMS Extension or the MP5 as follows:
Pump Check
1. Go into service mode and select Setup CO
2. Connect a FilterLine to the Microstream CO
menu.
2
input. This activates the pump in the M3015A
2
MMS Extension or the MP5.
3. The status line at the bottom of the screen displays “CO
pressure reading (ambient/cell)
2
xxx/yyy” where xxx is the ambient pressure and yyy is the measured cell pressure. Check whether the ambient pressure value (x3) matches (within the acceptable tolerance of ±12mm Hg) the reference value you have received. If so, proceed to the leakage check. If the value is not correct, calibrate as follows.
a. Select CO
then select Barom.Press to activate a table of values.
2
b. Select the value in the table which matches the reference value received from a reliable local
source (airport, regional weather station or hospital weather station). (The values are displayed with a resolution of 2 mmHg up to 500 mmHg and a resolution of 1 mmHg from 500 mmHg to 825 mmHg.) Note: the selected value must be within ±10% of the current measured ambient pressure, otherwise an error message will occur at restarting the monitor.
c. Confirm the barometric pressure setting. d. Check that the ambient pressure displayed in the status line at the bottom of the screen is the
same as the value which you selected from the list in step b.
1. Connect the flowmeter inlet to the mCO
gas outlet.
2
2. Connect the FilterLine to the mCO
3. Block the inlet of the FilterLine using your fingertip and observe the cell pressure on the
monitor display. The cell pressure (x4) should be more than 120 mmHg below the ambient pressure shown. If the pressure difference is less than 120 mmHg, the pump is not strong enough and you should replace it, irrespective of the Pump OpTime.
Flow Rate Check and Calibration
Check the flow rate in the M3015A MMS Extension or the MP5 as follows:
1. Connect the flowmeter to the CO
2. Check on the flowmeter the flow that the M3015A MMS Extension or MP5 mCO2 pum p
draws (x5). It should be 50 ml/min ± 7.5 ml/min. If the value is within tolerance, proceed to the CO
Gas calibration check. If the value is not within tolerance, calibrate as follows.
2
3. Adjust the flow in the instrument by selecting Increase Flow or Decrease Flow until
it is as close as possible to 50 ml per minute as indicated on the flowmeter gauge.
4. When you are satisfied that the flow is set as close as possible to 50 ml per minute, select
Store Flow and confirm the setting. If you do not store the adjusted flow within 60 seconds of the adjustment, the old flow setting is restored.
inlet.
2
FilterLine.
2
Page 96
3 Testing and Maintenance Performance Assurance Tests
5. If you cannot adjust the flow to within tolerance, replace the pump . If you still cannot make
the flow adjustment, this indicates a fault in the measurement extension, which must be replaced.
Note that the pump can only be replaced on M3015A with the old hardware Rev. A (i.e. Serial No. DE020xxxxx
Noise Check
1. With the monitor in service mode, select Setup CO
2. Disconnect the flowmeter and connect the 5% calibration gas and flow regulator in its place.
3. Open the valve to apply the 5% calibration gas and wait until the value is stable.
4. Check the noise index (x6) displayed next to the CO
level of noise on the CO extension.
wave). If the value exceeds 3 mmHg, replace the measurement
2
CO2 Gas Measurement Calibration Check
After switching the measurement extension on, wait at least 20 minutes before checking the calibration. Check the calibration of the CO
1. Check that the 5% calibration gas and flow regulator are connected.
2. Calculate the expected measurement value in mmHg as follows:
0.05 x (ambient pressure) = value mmHg for example 0.05 x 736 = 36.8 mmHg (with an ambient pressure of 736 mmHg)
3. Open the valve on the flow regul ator to allow 5% CO
the value to stabilize.
4. Check that the value on the instrument (measurement value on the main screen, x7) matches
the calculated mmHg value ± 2.6 mmHg. If the value is outside the tolerance, calibrate as described in step in this procedure onwards.
menu.
2
value on the display (this indicates the
2
gas measurement as follows:
2
gas to flow into the extension. Allow
2
5. Disconnect the 5% calibration gas and connect the 10% calibration gas.
6. Calculate the expected measurement value and tolerance in mmHg as follows:
0.1 x (ambient pressure) = value mmHg ±0.07 x (value mmHg) = tolerance
for example 0.1 x 737 mmHg = 73.7 mmHg (with an ambient pressure of 737 mmHg) ±0.07 x 73.7 mmHg = ±5.16 mmHg tolerance
7. Open the valve on the flow regulator to allow 10% CO
gas to flow into the extension.
2
Allow the value to stabilize.
8. Check that the value on the instrument (x8) matches the calculated mmHg value within the
calculated tolerance. If so, the measurement extension is correctly calibrated. If the value is outside the tolerance, calibrate as follows.
9. If not already connected, connect the 5% calibration gas.
10. Select Cal. CO
.
2
11. Select the value for the calibration gas. (The default value is 5.0%.)
Page 97
Performance Assurance Tests 3 Testing and Maintenance
12. Open the valve on the calibration gas to allow CO2 gas to flow into the extension. Allow the
value to stabilize before the start of the calibration. Leave the valve open until the instrument gives a prompt that gas can be removed.
13. The extension calibrates and prompts when calibration is successful.
Calibration Verification
1. Reopen the 5% gas valve and allow the value to stabilize.
2. Check that the value displayed on the monitor is correct within the tolerance (see step
above).
3. Disconnect the 5% calibration gas and connect the 10% calibration gas.
4. Open the valve on the flow regulator to allow 10% CO2 gas to flow into the extension.
Allow the value to stabilize.
5. Check that the value displayed on the monitor is correct within the tolerance (see step
above).
If one or both values are not within tolerances, you must exchange the M3015A MMS Extension or the MP5 mCO
Assembly.
2
Test Expected Test Results
Leakage Check parts 1 and 2
x1 = value of part 1 leakage check on flowmeter (x1< 4.0 ml/min)
x2 = value of part 2 leakage check on flowmeter (x2< 4.0 ml/min)
Barometric Pressure Check
x3 = difference between the reference pressure and the measured ambient pressure displayed on the monitor
(x3<12 mmHg)
Pump Check x4 = difference in pressure between cell pressure and
ambient pressure displayed on the monitor during occlusion (x4 >120 mmHg)
Flow Check x5 = difference between measured value and 50.0
ml/min (x5<7.5 ml/min)
Noise Check x6 = noise index displayed on monitor (x6<3.0) CO2 Gas
Calibration Check
x7 = difference between measured CO2 value and calculated value, based on 5% CO
cal. gas. (x7 < 2.6
2
mmHg)
CO2 Cal Verification
x8 = difference between measured CO2 value and calculated value, based on 10% CO
cal. gas.
2
(x8 < ± {0.07 x value calculated})
Page 98
3 Testing and Maintenance Performance Assurance Tests
Reset Time Counters
NOTE
This procedure only applies to M3015A with the old hardware Rev. A (i.e. Serial No. DE020xxxxx
You must check the time counters on the Microstream CO instrument. As well, when parts are replaced, the appropriate counters must be reset to zero.
The counters for CO
pump, IR Src and Last Cal are displayed in the status line. The values are
2
updated when entering the Setup CO2 menu.
Observe the following guidelines:
- When calibrating the CO
extension, if no parts have been replaced, check the displayed
2
values of Reset PumpOpTime and Reset IRSourceTime selections to make sure that they are within suggested guidelines for use (15, 000 hours of continuous use). If the counter time is greater than 15, 000 hours, replace the appropriate part. See Repair and Disassembly for details.
- When calibrating the CO
extension, if parts have been replaced, reset the appropriate values
2
using the Reset PumpOpTime and Reset IRSourceTime selections. See Repair and Disassembly for details.
Resetting the PumpOpTime generate s the INOP: “CO must perform a flow check and store the flow in service mode (select Store Flow).
CO2 Pump / CO2 Scrubber Replacement
NOTE
This procedure only applies to M3015A with the old hardware Rev. A (i.e. Serial No. DE020xxxxx
extension before calibrating the
2
OCCLUSION”. To clear this INOP you
2
Refer to the Repair and Disassembly section for the replacement procedures.

Spirometry Performance Tests

These tests verify the performance accuracy of the M1014A Spirometry module.
Equipment Required
- Leak test kit (Part number: M1014-64100)
- calibrated barometer
- M2785A Pediatric/Adult Flow Sensor
- 500ml calibration syringe, Hans Rudolph model 5550 or equivalent
Flow Test
1. Connect the M1014A Spirometry Module to the host monitor and go into service mode.
2. Connect the flow sensor to the module.
Page 99
Performance Assurance Tests 3 Testing and Maintenance
3. Connect the 500ml calibration syringe to the flow sensor. Make sure the syringe is set to the
“empty” position.
4. Press the Setup key on the module and select Show all Values in the Setup
Spirometry menu.
5. Pump the calibration syringe back and forth with a steady motion at a rate of 20 cycles and
verify that the readings for TVexp and TVin are 500 ± 25 ml.
If the readings are not within the specified range, try another flow sensor. Ensure that the syringe is calibrated correctly and that the procedure is performed exactly as described above. If the test fails again, replace the module.
Test Expected test results
Flow Test
TVexp and TVin are 500 ± 25 ml
Leakage Test
1. Connect the M1014A Spirometry Module to the host monitor and go into service mode.
2. Connect the leak test adapter to the module.
3. Press the Setup key on the module and then select Show all Values in the Setup
4. Press the Purge key on the module and start a purge cycle. At the end of the purge cycle, the
5. Verify that the pressure difference between Ppeak and Paw remains less than 10 cmH2O
If the readings are not within the specified range or if an INOP (e.g. SPIRO PURGE FAILED) is issued, check the leak test adapter for any leaks. Disconn ect the adapter from the module and start the test procedure from the beginning. If the test fails again, replace the module.
Test Expected test results
Leakage Test Paw and Ppeak >100 cmH2O
Barometer Check
1. Connect the M1014A Spirometry Module to the host monitor and go into service mode.
2. Attach any airway adapter to the module.
Spirometry menu.
values for Paw and Ppeak should both be above 100 cmH2O.
after 30 seconds.
3. Press the Setup key on the module and then select Show all Values in the Setup
Spirometry menu.
4. Check that the barometric reading (PB) is within ± 5 mmHg of a reference barometer.
5. If the readings are not within the specified range, check the accuracy of the barometric
pressure reference again. If the test fails again, replace the module.
Page 100
3 Testing and Maintenance Performance Assurance Tests
Test Expected test results
Barometer Check
PB is within ± 5 mmHg of a reference barometer
NOTE
The built-in barometer cannot be recalibrated.

Cardiac Output Performance Test

These tests check the performance of the cardiac output measurement.
1. Connect the patient simulator to the C.O. module using the patie nt cab le.
2. Configure the patient simulator as follows:
Injection temperature: 2 °C Computation Const: 0.542 (Edward's Catheter) Flow: 5 l/min
3. Check displayed value against the simulator configuratio n.
4. Expected test result: C.O. = 5 +/– 1 l/min.
Test Expected test results
Cardiac Output Performance Test C.O. = 5 +/– 1 l/min
Service Tool Procedure, Version 1
This procedure applies for Service Tool M1012-61601 in combination with C.O. modules without option C10 and M3012A MMS extensions with option C05.
1. In monitoring mode, connect the C.O. interface cable to the module.
2. Connect one side of the service tool to the injectate receptacle of C.O. interface cable and
the other side to catheter cable receptacle.
3. Enter the C.O. Procedure window and check the results. The expected test result is:
o
Tblood = 37.0
C +/- 0.1oC
Test Expected test results
Cardiac Output Service Tool Procedure Version 1
Tblood = 37.0oC +/- 0.1oC
Loading...