Partlow MIC 8200 Operating Manual

Form 3032 Edition 4 ©July 1993 Updated Jan. 1994
MIC 8200
Installation, W iring, Operation Manual
Brand
PAGE 2
nformation in this installation, wiring, and operation
I
manual is provided with each instrument at the time of shipment. Extra copies are available at the price published on the front cover.
Copyright © July 1993, all rights reserved. No part of this publication may be reproduced, transmitted, transcribed or stored in aretrieval system, or translated into any language in any form by any means without the written permission of the factory.
This is the Fourth Edition of the manual. It was written and produced entirely on a desk-top-publishing system. Disk versions are available by written request to the factory ­Advertising and Publications Department.
NOTE
We are glad you decided to open this manual. It is written so that you can take full advantage of the features of your new dual display process controller.
It is strongly recommended that factory equipped applications incorporate a high or low limit protective device which will shut down the equipment at a preset process condition in order to preclude possible damage to property or products.
Table of Contents
SECTION 1 - GENERAL Page Number
1.1 Product Description 5
SECTION 2 - INSTALLATION & WIRING
2.1 Installation and Wiring 7
2.2 Input Connections 8
2.3 Output Connections 13
SECTION 3 - CONFIGURATION & OPERATION
3.1 Configuration and Operation 21
3.2 Operation Summary 22
3.3 Configuration Summary 23
3.4 Auto Tune Method 36
3.5 Manual Tuning Method 39
SECTION 4 - CONTROL CAPABILITY
4.1 Control Capability 40
4.2 Control Responses 40
4.3 Direct/Reverse Operation of Control Outputs 40
4.4 On-Off Control 41
4.5 Time Proportioning Control 41
4.6 Current Proportioning Control 41
4.7 Position Proportioning Control 41
4.8 Dual Output Control 43
4.9 Manual Operation of Proportional Outputs 44
4.10 Automatic Transfer Function 44
4.11 Setpoint Adjustments 45
PAGE 3
SECTION 5 - SERVICE
5.1 Service 48
5.2 Calibration 48
5.3 Test Mode 52
5.4 Troubleshooting and diagnostics 56
APPENDICES
A - Board Layout - Jumper Positioning
Figure A-1 Power Supply Board 64 Figure A-2 Processor Board 65
Figure A-3 Option Board 66, 67 B - Glossary of terms 68 C - Model Number Hardware Matrix Details 73 D - Specifications 74 E - Software Record/Reference Sheet 77 Warranty Inside back cover
PAGE 4
FIGURES & TABLES
Figure 1-1 Controller Display Illustration 5 Figure 2-1 Panel Opening Sizes and Installation 7 Figure 2-2 Noise Suppression 9 Figure 2-3 Noise Suppression 10 Figure 2-4 Wiring Label 12 Figure 2-5 AC Power 13 Figure 2-6 Thermocouple Input 13 Figure 2-7 RTD Input 14 Figure 2-8 Volt, mV, mADC Input 14 Figure 2-9 24 Volt Transmitter Power Supply 15 Figure 2-10 Remote Setpoint Input 16 Figure 2-11 Remote Setpoint Selection 17 Figure 2-12 Remote Digital Comm. 7 & 8 17 Figure 2-13 Remote Digital Comm. G & H 18 Figure 2-14 Relay Output 18 Figure 2-15 SSR Driver Output 19 Figure 2-16 mADC Output 20 Figure 2-17 Position Proportioning Output 20 Figure 3-1 Front Panel 21 Figure 4-1 Proportional Bandwidth effect on Output 42 Figure 4-2 Dual Proportional Outputs 43 Figure 4-3 Setpoint Ramp Rate Example 45 Figure 4-4 Re-transmission Example 46 Table 3-1 Enable Mode Configuration Procedures 24 Table 3-2 Program Mode Configuration Procedures 29 Table 3-3 Tune Mode Configuration Procedures 35 Table 5-1 Calibration Procedures 48 Table 5-2 Test Procedures and Description 53
FLOW CHARTS
Flow - Calibration 49 Flow - Enable Mode 25 Flow - Program Mode 26 Flow - Test 52 Flow - Tune Mode 34 Flow - Setpoint Select 44
Product Description 1.1
1.1.1 GENERAL
This instrument is a microprocessor based single loop controller capable of measuring, displaying and controlling temperature, pressure, flow, and level from a variety of inputs. Most heating outputs are easily tuned using the instrument’s Auto Tune function with several choices for control algorithms and control responses.
Control functions, alarm settings and other parameters are easily entered through the front keypad. All user's data can be protected from unauthorized changes with it’s Enable mode security system. Battery back-up protects against data loss during AC power outages.
The input is user configurable to directly connect to either thermocouple, RTD, mVDC, VDC or mADC inputs. Thermocouple and RTD linearization, as well as thermocouple cold junction compensation is performed automatically. The sensor input is isolated . The instrument can be specified to operate on either 115VAC or 230VAC power at 50/60Hz. It is housed in an extruded aluminum enclosure suitable for panel mounting and may be surface mounted using an optional adaptor. For installation in washdown areas, a watertight cover is available (see the instrument price list order matrix).
FIGURE 1-1
PAGE 5
OUT2
ALRM
°C
°F
U
RSP
PO1
PO2
PV
SP1
SP2
MAN
AUTO
MAN
AUTO TUNE
OUT1
SP1 SP2
1.1.2 DISPLAYS
Each instrument is provided with dual digital displays and status indicators as shown in Figure 1-1. The upper digital display is programmable to show the process variable or the deviation from setpoint value. The lower digital display will be the active setpoint value or the percent­age of the proportional output indicated by the indicator light. Status indication is as shown (Figure 1-1). Display resolution is programmable for 0 to 3 decimal places depending upon the input type selected.
PAGE 6
1.1.3 CONTROL
The instrument can be programmed for on-off, time proportioning, current proportioning, or position proportioning control implementations depending on the output(s) specified for the instrument in the model number. The Auto Tune function can be used for a heating output assigned to output 1 at the Setpoint 1 value. A second control output is an available option. Proportional control implementations are provided with fully programmable separate PID
parameters.
1.1.4 ALARM
Alarm indication is standard on all instruments. Alarm type may be set as PROCESS DIRECT or REVERSE (High or Low), DEVIATION DIRECT or REVERSE (Above or Below setpoint), or DEVIATION BAND TYPE (Closed or Open within the band). Alarm status is indicated by LED. An alarm output can be provided by assigning any output(s) SPST relay(s) or SSR Driver(s) to the alarm.
1.1.5 PROCESS VALUE RE-TRANSMISSION OUTPUT
If an instrument is specified with a mADC current output, this output may be programmed to operate as a process value re-transmission output (range scaled by user). If an output is used as a process value output, it is not available for use as a control output.
T
Installation and Wiring 2.1
Prior to proceeding with installation, verify the AC power input required by the instrument. AC power input is either 115 VAC or 230 VAC and is specified in the model number and on the wiring label affixed to the instrument housing. See Figure 2-4 (page 12) for a wiring label description.
230 VAC models may be converted to 115 VAC operation by the user by changing the position of jumpers soldered on the Power Supply Board, see Appendix A-1 (page 50) for details. (Note: 115VAC units cannot be field converted to 230VAC)
Electrical code requirements and safety standards should be observed and installation performed by qualified personnel.
The electronic components of the instrument may be removed from the housing during installation. To remove the components, loosen the locking screw located in the lower center of the instrument’s front panel. Pull the entire instrument straight out of the housing. During re-installation, the vertically mounted circuit boards should be properly aligned in the housing. Be sure that the instrument is installed in the original housing. This can be verified by matching the serial number on the unit to the serial number on the housing. (Serial numbers are located on the inside of the housing enclosure and on the label on the underside of the front panel). This will insure that each instrument is accurate to its published specifications. The ambient compensator on the rear of the housing enclosure is calibrated to the electronics of the instrument at the factory.
PAGE 7
Recommended panel opening sizes are illustrated below (Figure 2-1). After the opening is properly cut, insert the instrument housing into the panel opening. Insert the two panhead screws provided, through the holes in the mounting bracket into the holes in the rear of the instrument as shown in Figure 2-1. Firmly tighten the screws. Instruments are shipped standard for panel mounting. To surface mount, an adaptor is required and should be specified when ordering. For installation in wash-down areas, a watertight cover is available.
FIGURE 2-1 PANEL OPENING SIZES AND INSTALLATION
165.9 (6.53)
146.8 (5.78)
Side View
MOUNTING BRACKE
4.8 (.188) MAX PANEL THICKNESS
90.4 (3.560)
96.0 (3.78)
92 + or - 0.8 (3.622 + or - .031)
96.0 (3.78)
Top View
PANEL CUTOUT
92 + or - 0.8 (3.622 + or ­ .031)
DIMENSIONS ARE IN MM (IN)
90.4 (3.560)
PAGE 8
Preparation for Wiring 2.2
2.2.1 WIRING GUIDELINES
Electrical noise is a phenomenon typical of industrial environments. The following are guidelines that must be followed to minimize the effect of noise upon any instrumentation.
2.2.1.1 INSTALLATION CONSIDERATIONS
Listed below are some of the common sources of electrical noise in the industrial environ­ment:
• Ignition Transformers
• Arc Welders
• Mechanical contact relay(s)
• Solenoids Before using any instrument near the devices listed, the instructions below should be fol-
lowed:
1. If the instrument is to be mounted in the same panel as any of the listed devices, separate them by the largest distance possible. For maximum electrical noise reduction, the noise generating devices should be mounted in a separate enclosure.
2. If possible, eliminate mechanical contact relay(s) and replace with solid state relays. If a mechanical relay being powered by an instrument output device cannot be replaced, a solid state relay can be used to isolate the instrument.
3. A separate isolation transformer to feed only instrumentation should be considered. The transformer can isolate the instrument from noise found on the AC power input.
4. If the instrument is being installed on existing equipment, the wiring in the area should be checked to insure that good wiring practices have been followed.
2.2.1.2 AC POWER WIRING
Earth Ground The instrument includes noise suppression components that require an earth ground connec­tion to function. To verify that a good earth ground is being attached, make a resistance check from the instrument chassis to the nearest metal water pipe or proven earth ground. This reading should not exceed 100 ohms. Use a 12 gauge (or heavier) insulated stranded wire.
Neutral (For 115VAC) It is good practice to assure that the AC neutral is at or near ground potential. To verify this, a voltmeter check between neutral and ground should be done. On the AC range, the reading should not be more than 50 millivolts. If it is greater than this amount, the secondary of this AC transformer supplying the instrument should be checked by an electrician. A proper neutral will help ensure maximum performance from the instrument.
2.2.1.3 WIRE ISOLATION
Four voltage levels of input and output wiring may be used with the unit:
• Analog input or output (i.e. thermocouple, RTD, VDC, mVDC or mADC)
• SPST Relays
• SSR driver output
• AC power The only wires that should be run together are those of the same category. If they need to be
run parallel with any of the other lines, maintain a minimum 6 inch space between the wires. If wires must cross each other, do so at 90 degrees. This will minimize the contact with each other, do so at 90 degrees. This will minimize the contact with each other and reduces "cross talk". "Cross talk" is due to the EMF (Electro Magnetic Flux) emitted by a wire as current passes through it. This EMF can be picked up by other wires running the same bundle or conduit.
In applications where a High Voltage Transformer is used, (i.e. ignition systems) the second­ary of the transformer should be isolated from all other cables.
This instrument has been designed to operate in noisy environments, however, in some cases even with proper wiring it may be necessary to suppress the noise at its source.
2.2.1.4 USE OF SHIELDED CABLE
Shielded cable helps eliminate electrical noise being induced on the wires. All analog signals should be run with shielded cable. Connection lead length should be kept as short as possible, keeping the wires protected by the shielding. The shield should be grounded at one end only. The preferred grounding location is the sensor, transmitter, or transducer.
2.2.1.5 NOISE SUPPRESSION AT THE SOURCE
Usually when good wiring practices are followed, no further noise protection necessary. sometimes in severe electrical environments, the amount of noise is so great tht it has to be suppressed at the source. Many manufacturers of relays, contactors, etc., supply "surge suppressors" which mount on the noise source.
For these devices that do not have surge suppressors supplied, RC (resistance-capacitance) networks and/or MOC (,etal oxide varistors) may be added.
Inductive Coils - MOV's are recommended for transient suppression in inductive coils con­nected in parallel and as close as possible to the coil. See Figure 2-2. Aditional protection may be provided by adding an RC network across the MOV.
PAGE 9
FIGURE 2-2
0.5 mfd 1000V
Coil
220 ohms
115V 1/4W 230V 1W
Contacts - Arcing may occur across contacts when the contact opens and closes. This results in electrical noise as well as damage to the contacts. Connecting a RC network properly sized can eliminate this arc.
For circuits up to 3 amps, a combination of a 47 ohm resistor and 0.1 microfarad capacitor (1000 volts) is recommended. For circuits from 3 to 5 amps, connect 2 of these in parallel. See Figure 2-3, page 10.
PAGE 10
FIGURE 2-3
MOV
R
C
Inductive Load
2.2.2 SENSOR PLACEMENT (Thermocouple or RTD)
Two wire RTD's should be used only with lead lengths less then 10 feet. If the temperature probe is to be subjected to corrosive or abrasive conditions, it should be
protected by the appropriate thermowell. The probe should be positoned to reflect true process temperature:
In liquid media - the most agitated area. In air - the best circulated area.
THERMOCOUPLE LEAD RESISTANCE
Thermocouple lead length can affect instrument accuracy since the size (gauge) and the length of the wire affect lead resistance.
To determine the temperature error resulting from the lead length resistance, use the following equation:
Terr = TLe * L where; TLe = value from appropriate table below
L = length of leadwire in thousands of feet
TABLE 1 Temperature error in °C per 1000 feet of Leadwire
AWG Thermocouple Type: No. J K T R S E B N C 10 .34 .85 .38 1.02 1.06 .58 7.00 1.47 1.26 12 .54 1.34 .61 1.65 1.65 .91 11.00 2.34 2.03 14 .87 2.15 .97 2.67 2.65 1.46 17.50 3.72 3.19 16 1.37 3.38 1.54 4.15 4.18 2.30 27.75 5.91 5.05 18 2.22 5.50 2.50 6.76 6.82 3.73 44.25 9.40 8.13 20 3.57 8.62 3.92 10.80 10.88 5.89 70.50 14.94 12.91 24 8.78 21.91 9.91 27.16 27.29 14.83 178.25 37.80 32.64
TABLE 2 Temperature Error in °F per 1000 feet of Leadwire
AWG Thermocouple Type: No. J K T R S E B N C 10 .61 1.54 .69 1.84 1.91 1.04 12.60 2.65 2.27 12 .97 2.41 1.09 2.97 2.96 1.64 19.80 4.21 3.66 14 1.57 3.86 1.75 4.81 4.76 2.63 31.50 6.69 5.74 16 2.47 6.09 2.77 7.47 7.52 4.14 49.95 10.64 9.10 18 4.00 9.90 4.50 12.17 12.28 6.72 79.95 10.64 9.10 20 6.43 15.51 7.06 19.43 19.59 10.61 126.90 26.89 23.24 24 15.80 39.44 17.83 48.89 49.13 26.70 320.85 68.03 58.75
Example: An MIC is to be located in a control room 660 feet away from the process. Using 16 AWG, type J thermocouple, how much error is induced?
Terr = TLe * L
TLe = 2.47 (°F/1000 ft) from Table 2 Terr = 2.47 (°F/1000 ft) * 660 ft Terr = 1.6 °F
PAGE 11
RTD LEAD RESISTANCE
Rtd lead length can affect instrument accuracy, since the size (gauge) and length of the wire affect lead resistance.
To determine the temperatire error resulting from the lead length resistance, use the following equation:
Terr = TLe * L where; TLe = value from Table 3 if 3 wire RTD or Table 4 if 2 wire RTD
L = length of lead wire in thousands of feet.
TABLE 3 3 Wire RTD AWG No. Error °C Error °F
10 +/-0.04 +/-0.07 12 +/-0.07 +/-0.11 14 +/-0.10 +/-0.18 16 +/-0.16 +/-0.29 18 +/-0.26 +/-0.46 20 +/-0.41 +/-0.73 24 +/-0.65 +/-1.17
TABLE 4 2 Wire RTD AWG No. Error °C Error °F
10 +/-5.32 +/-9.31 12 +/-9.31 +/-14.6 14 +/-13.3 +/-23.9 16 +/-21.3 +/-38.6 18 +/-34.6 +/-61.2 20 +/-54.5 +/-97.1 24 +/-86.5 +/-155.6
(Continued on next page)
PAGE 12
A
A
C
+
C
-
A
A
B
C
B
C
(Continued from page 11) Example:
An application uses 2000 feet of 18 AWG copper lead wire for a 3 wire RTD sensor. What is the worst case error due to this leadwire length?
Terr = TLe * L
TLe = +/-.46 (°F/1000 ft) from Table 3 Terr = +/-.46 (°F/1000 ft) * 2000 ft Terr = +/- 0.92°F
FIGURE 2-4 WIRING LABEL
RELAY
RELAY
RELAY
115 230
VA
H
G
SERIAL
SERIAL
POS.PROP. WIPER
POS.PROP. HIGH
F
E
INPUT RATINGS:
D
C
115/230 VAC 50/60 HZ 15VA MAX RELAY OUTPUT RATINGS:
115VAC 5.0A RESISTIVE 230VAC 2.5A RESISTIVE 230VAC 1/8 HP 115/230VAC 250VA
MAXIMUM AMBIENT : 55°C
B
REMOTE
8
SETPT
OUT2
7
4-20mA
OUT1
6
4-20mA
RETURN
5
4
3
SIGNAL
2
1
SIGNAL
+
+
+
CJ
CJ
GROUND
MADE IN U.S.
Input Connections 2.3
E
w
e
E
w
In general, all wiring connections are made to the instrument after it is installed.
electrical shock. AC power wiring must not be connected to the source distribution panel until all wiring connection procedures are completed.
2.3.1 INPUT CONNECTIONS
FIGURE 2-5
AC Power Connect 115 VAC hot and neutral to terminals B and A respectively as illustrated below. Connect 230 VAC as described below. Connect Earth ground to the ground screw as shown.
Avoid
PAGE 13
115 VAC INSTRUMENT VOLTAG
Rear Vie
.5 AMP* FUSE
L1 L2
*Supplied by custome
B
A
GROUND
230 VAC INSTRUMENT VOLTAG
Rear Vie
.25 AMP* FUSE
L1
L2
B
A
GROUND
*Supplied by the custom
FIGURE 2-6
Thermocouple (T/C) Input Make thermocouple connections as illustrated below. Connect the positive leg of the thermo­couple to terminal 3, and the negative to terminal 1. For industrial environments with com­paratively high electrical noise levels, shielded thermocouples and extension wire are recom­mended. Be sure that the input conditioning jumpers are properly positioned for a thermo­couple input. See Appendix A-2 (page 65) and A-3 (page 66 and 67).
THERMOCOUPLE INPUT
Rear view
8
7
6
5
4
3
2
1
+
-
300 OHMS MAXIMUM LEAD
PAGE 14
T
w
T
w
FIGURE 2-7
RTD Input Make RTD connections as illustrated below. For a three wire RTD, connect the resistive leg of the RTD to terminal 3, and the common legs to terminal 1 and 5. For a two wire RTD, connect one wire to terminal 1 and the other wire to terminal 3 as shown below. A jumper wire supplied by the customer must be installed between terminals 1 and 5. Be sure that the input conditioning jumpers are properly positioned for an RTD input. See Appendix A-2 (page 65) and A-3 (page 66 and 67).
2 WIRE RTD INPU
Rear Vie
8
7
6
5
JUMPER*
4
3
2
1
100 OHM* PLATINUM
10 FEET LEAD MAXIMUM
3 WIRE RTD INPU
Rear Vie
8
7
6
5
4
3
2
1
*Supplied by the custome
100 OHM* PLATINUM
*Supplied by custome
FIGURE 2-8
Volt, mV, mADC Input Make volt, millivolt and milliamp connections as shown below. Terminal 3 is positive and terminal 1 is negative. Milliamp input requires a 250 ohm shunt resistor (supplied with the instrument) installed across the input terminals and by configuring the instrument for either 0 to 5 or 1 to 5 VDC input. If desired, milliamp DC input can be facilitated by installing an optional 2.5 ohm resistor across the input terminals and configuring the instrument for either 0 to 50 or 10 to 50 mVDC. Be sure that the input conditioning jumpers are properly positioned for the input type selected. See Appendix A-2 (page 65) and A-3 (page 66 and 67).
MILLIAMP DC INPUT
Rear View
8
7
6
5
4
3
2
1
Shielded Twisted Pair
+
-
MILLIAMP DC SOURCE 250 OHM SHUNT RESISTER REQUIRED
MILLIAMP DC INPUT
Rear View
8
7
6
5
4
3
2
1
Shielded Twisted Pair
+
-
MILLIAMP DC SOURCE
2.5 OHM SHUNT RESISTER REQUIRED
PAGE 15
MILLIVOLT DC INPUT
Rear View
8
7
6
5
4
+
3
2
1
-
Shielded Twisted Pair
MILLIVOLT DC SOURCE 50 MILLIVOLT DC MAXIMUM
VOLT DC INPUT
Rear View
8
7
6
5
4
+
3
2
1
-
Shielded Twisted Pair
VOLT DC SOURCE 5 VOLT DC MAXIMUM
FIGURE 2-9A
24 Volt Transmitter Power Supply (XP Option) Make connections as shown below. Terminal 3 is positive (+) and terminal 1 is negative (-). Be sure the input conditioning jumpers are properly positioned for the input type selected. See Figure A-2 Processor Board, page 65, and Figure A-3 Option Board, page 66 or 67. Note the 250 ohm shunt resistor is not required.
+3
2
-1
+
Two Wire Transmitter
-
FIGURE 2-9B
24 Volt Power Supply (XA Option) Make connections as shown below. Terminal G is positive (+) and terminal H is negative (-). Be sure the input conditioning jumpers are properly positioned. See Figure A-2 Processor Board, page 65 and Figure A-3 Option Board, page 66 or 67.
H -
24VDC
G +
PAGE 16
w
r
FIGURE 2-10
Remote Setpoint Input - VDC and mADC and Potentiometer Input connections are illustrated below. Terminal 8 is positive and terminal 5 is negative. The remote setpoint input can be configured for either 0 to 5VDC or 1 to 5 VDC input. Make sure that the voltage input matches the voltage configuration selected in the Program mode. For mA inputs, a 250 ohm shunt resistor must be installed between terminals 5 and 8. For remote setpoint using a potentiometer, JU1 on options board must be in MM/PP (see page 66 and 67).
CURRENT DC REMOTE SETPOINT
Rear Vie
8
7
6
5
4
3
2
1
Shielded Twisted Pai
+
-
MILLIAMP SETPOINT SIGNAL 250 OHM SHUNT RESISTER NEEDED
POTENTIOMETER
Rear View
VOLT DC REMOTE SETPOINT
Rear View
8
7
6
5
4
3
2
1
8
7
6
5
4
3
2
1
150 ohm to 10 K ohm
+
-
Shielded Twisted Pair
VOLT DC SETPOINT SIGNAL 5VDC MAXIMUM
FIGURE 2-11
Remote Setpoint Selection of one of two preset setpoint values (Optional) A programmable feature allows for the setpoint value to be toggled between two preselected values when a dry contact closure is sensed between terminals 8 and 5. For more information see section 3 (page 21).
Rear View
8
DRY CONTACT
7
6
5
(SWITCH, RELAY, ETC.) SUPPLIED BY CUSTOMER
PAGE 17
4
3
2
1
SHIELDED WIRING IS RECOMMENDED
FIGURE 2-12
Remote Digital Communications RS 485 Terminals 7 & 8 (Optional) If the communications network continues on to other units, connect the shields together, but not to the instrument. A terminating resistor should be installed at the terminals of the last instrument in the loop. The shield should be grounded at the computer or the convertor box, if used. See the Protocol Manual (Form 2878) for more details on the use of the digital communications option.
DIGITAL COMMUNICATIONS CONNECTIONS - TERMINALS 7 & 8
Terminals 7 & 8 are used for communications when the model number is 82XYX3X, 82XYX5X where X = any valid number and Y = 0, 1, or 2. No Second Output 4-20mA
Output 2 cannot be DC Current
8
7
6
5
4
3
2
1
FROM HOST COMPUTER
TO OTHER INSTRUMENTS
PAGE 18
w
FIGURE 2-13
Alternate Remote Digital Communications RS 485 Terminals G & H (Optional) If the communications network continues on to other units, connect the shields together, but not to the instrument. A terminating resistor should be installed at the terminals of the last instrument in the loop. The shield should be grounded at the computer or the convertor box , if used. See the Protocol Manual (Form 2878) for more details on the use of the digital communications option.
DIGITAL COMMUNICATIONS CONNECTIONS - TERMINALS G & H
Terminals G & H are used for communications when the model number is 82XY04X, 82XY06X where
From Host Computer
Output 3 Must Be 0
Rear Vie
H
G
X = any valid number and Y = 3, 4, or 5. Use when Second Output is 4-20mA.
INPUT POWER
To Other Instruments
F
E
D
C
B
A
GROUND
Output Connections 2.4
FIGURE 2-14
Relay Output Connections are made to relay A as illustrated below. Connect relay(s) B & C (if present) in the same manner. Relay contacts are rated at 5 amp Resistive load 115 VAC.
RELAY A
Rear View
LOAD
L2
LOAD
L2 L1
INPUT POWER
D
C
B
A
GROUND
L1
INPUT POWER
RELAY B
Rear View
H
G
F
E
D
C
B
A
GROUND
w
w
)
w
RELAY C
PAGE 19
Rear View
H
G
F
E
D
C
B
A
GROUND
INPUT POWER
LOAD
L2 L1
FIGURE 2-15
SSR Driver Output Connections are made to the solid state relay driver output located in the Relay A position as shown. The solid state relay driver is a 5 VDC current sink output type. Connect the solid state relay driver(s) in the Relay B and C position (if present) in the same manner.
INPUT POWER
SSR DRIVER (RELAY A)
SOLID STATE RELAY
Rear Vie
H
G
F
E
+
D
-
C
B
A
GROUND
INPUT POWER
SOLID STATE RELAY
SSR DRIVER (RELAY B)
Rear Vie
H
G
+
F
-
E
D
C
B
A
GROUND
SSR DRIVER (RELAY C
SOLID STATE RELAY
+
-
INPUT POWER
Rear Vie
H
G
F
E
D
C
B
A
GROUND
PAGE 20
A
B
r
w
FIGURE 2-16
mADC Output Connections are made for current outputs 1 or 2 as shown below. Connect the positive lead to terminal 6 for Output 1 or terminal 7 for Output 2 , the negative leads connect to terminal 5. Current outputs will operate up to 650 ohms maximum load. The current output(s) can be selected for either 4 - 20 mADC or 0 - 20 mADC. If dual current outputs are both used, connect the returns to terminal 5.
DC CURRENT OUTPUT 1
Rear View
8
7
6
5
4
3
2
1
Shielded Twisted Pair
+
-
LOAD 650 OHMS MAXIMUM
DC CURRENT OUTPUT 2
Rear View
8
7
6
5
4
3
2
1
Shielded
+
Twisted Pair
-
LOAD 650 OHMS MAXIMUM
FIGURE 2-17
Position Proportioning Output The relay and slidewire feedback connections are made as illustrated below. The relay assigned to Output 1 will be used to drive the motor in the open direction and the relay assigned to Output 2 will be used to drive the motor in the closed direction. The minimum slidewire feedback resistance is 135 ohms, the maximum resistance is 10K ohms.
L2
OPEN
L
CLOSE
H
Modulating Moto
Slidewire Feedback Resistance min. 135
POS.PROP.
8
WIPER
POS.PROP.
+
7
HIGH
F
RELAY
E
D
L1
RELAY
C
Rear Vie
5
RETURN
ohms max. 10K ohms
Configuration and Operation 3.1
3.1.1 POWER UP PROCEDURE
Verify all electrical connections have been properly made before applying power to the instrument.
If the instrument is being configured (set up) for the first time, it may be desirable to discon­nect the controller output connections. The instrument will go into the Control mode following the power up sequence and the output(s) may turn on. During power up, the seven digit model number will be displayed. Next, the EPROM tab number will be displayed, followed by the software revision level. Instrument self test 1 through 3 will take place as they are displayed. After completion of the tests Ctrl will be displayed for 3 seconds. At this time another mode of operation may be selected by pressing the SCROLL key.
3.1.2 CONFIGURATION PROCEDURE
Parameter selections and data entry are made via the front keypad. To ease configuration and operation, the user selectable features have been divided into several sections (modes). Data and parameter entries are made by stepping through each mode and making an appropriate response or entry to each step as necessary for the application.
PAGE 21
FIGURE 3-1
PV
SP1
SP2
MAN
AUTO
MAN
AUTO TUNE
OUT1
SP1 SP2
OUT2
ALRM
°C
°F
U
RSP PO1
PO2
PAGE 22
Operation Summary 3.2
3.2.1 KEYPAD OPERATION
AUTO/MANUAL KEY
This key is used to enter the Manual mode (Standby) of operation from the Control mode and visa versa.
AUTO TUNE KEY
This key is used to initiate the Auto Tuning of the Output 1 proportional output for heating applications. If Auto Tune is being performed, pressing this key will abort the Auto Tune function. The instrument will Auto Tune the process to control at the Setpoint 1 value.
SP1/SP2 KEY
This key is used to change the setpoint from one preselected value to the other preselected value.
SCROLL KEY
This key is used to:
1. Display enabled modes of operation
2. Display a mode parameter value
3. Advance display from a parameter value to the next parameter code
4. Exit some calibration/test functions
5. Used with other keys:
A. With UP key to view output percentages of proportional output(s)
B. With DOWN key
1. On power up to alter model number
2. Enter calibration /test functions
3. To view output percentage of proportional Output 2
UP KEY
This key is used to:
1. Increase displayed parameter value
2. Increase setpoint (press and hold)
3. With a parameter code displayed
A. Press once to exit mode
B. Press twice to enter Control mode
4. Used with other keys
A. In Control mode with SCROLL key to view output percentage of proportional
output 1.
B. With DOWN Key
1. On power up resets instrument
2. Lamp test (press and release)
3. Enter Enable Mode (press and hold)
DOWN KEY
This key is used to:
1. Decrease displayed parameter value
2. Decrease setpoint (press and hold)
3. Enter modes
4. While in a mode, will sequence the parameter codes
5. Used with other keys
A. With SCROLL key
1. On power up to alter model number
2. Enter calibration/test functions
3. To view the output percentage of proportional output 2
B. With UP key
1. On power up resets instrument
2. Lamp test (press and release)
3. Enter enable mode (press and hold)
3.2.2 CONFIGURATION DISPLAYS
During configuration, the upper display shows the parameter codes. The lower digital display shows the parameter value. During operation, the upper display is used to indicate process value or deviation from setpoint. The lower display can be used to indicate setpoint value or proportional output percentage.
3.2.3 MODE SELECTION
If the instrument is in the Control mode, repeated depressions of the SCROLL key will cause the instrument to display the code corresponding to each mode that is enabled. To enter a mode, with the mode displayed, depress the DOWN key. Entry into any mode except the Control, Tune and Enable modes will cause the output(s) to turn off.
Configuration Summary 3.3
All configurable parameters are provided in Tables 3-1 thru 3-3 on the following pages. These tables illustrate the display sequence, parameter adjustment and factory setting for each step.
The instrument is provided with a “time-out” feature. If the instrument is in any mode, other than the Control mode, and no keypad activity takes place for 30 seconds, the mode will be exited automatically. The instrument will then display the code for the respective mode. If a mode code is displayed for five seconds with no key stroke activity the “time-out” will cause the instrument to return to the Control mode of operation.
PAGE 23
Control (CtrL)
Test (tESt)
Calibrate (CAL)
Program (Prog)
Tune (tunE)
Setpoint Select (SPS)
3.3.1 ENABLE MODE CONFIGURATION
The Enable Mode provides a means of enabling or disabling access to setpoint changes and each of the non-control modes. In the Enable mode, each mode except Control, will be displayed. Either “on” (enabled) or “oFF” (disabled) may be selected. See Table 3-1 (page
24) for the Enable mode procedure. For additional security the Enable mode may be locked out by using a hardware jumper, JU 2, located on the Processor board. See Appendix A-2 (page 65).
3.3.2 PROGRAM MODE CONFIGURATION
The Program mode is used to configure or re-configure the instrument. The input and output selections are made in the Program mode. All possible parameters are illustrated in Table 3-2 (page 29) for illustrative purposes. Only those parameters that are applicable to the hardware options chosen or to previous parameter selections will be displayed.
PAGE 24
3.3.3 TUNE MODE CONFIGURATION
The Tune mode is used to adjust the tuning parameters and the alarm setting needed for operation of the instrument. If Auto Tuning is used to determine the parameters for the heating output (Output 1), those parameters in the Tune mode (except Cycle time) need not be configured.
TABLE 3-1 ENABLE MODE CONFIGURATION PROCEDURE
To enter the Enable mode, depress and hold the UP and DOWN keys. All display lamps will light, after ten seconds the upper display will read EnAb. If EnAb does not appear, check the position of the Enable mode jumper, JU 2, located on the Processor board (See Appen­dix A-2, page 65). The jumper must be in the unlocked position for the Enable mode to function. Release the keys and the upper display will then change to EtSt. Depress the SCROLL key to review the state (on or off) of the mode (will appear on the lower display). Use the UP key to enable a mode that is off. Use the DOWN key to disable a mode that is on. When all selections have been made, to exit the Enable mode depress the UP key with a mode code displayed EtSt, ECAL, etc.
STE P DESCRIPTION DISPLAY AVAILABLE FACTORY YOUR
CODE SETTINGS SETTING SETTING
1 Test Mode EtSt on or oFF oFF 2 Calibration Mode ECAL on or oFF oFF 3 Program Mode EPro on or oFF on 4 Tune Mode Etun on or oFF on 5 Standby Mode ESby on or oFF on 6 Setpoint Select ESPS on or oFF oFF 7 Setpoint Changes ESPC on or oFF on 8 Auto Tune EAtn on or oFF on
If Standby is disabled and Auto Tune Abort is 0 or 1, then Standby is automatically turned on and cancels setting in the Enable mode.
ENABLE MODE FLOW CHART
PAGE 25
EnAb
EtSt
ECAL
EPro
Etun
Press UP and DOWN ARROWS for 10 seconds to enter this loop.
ON OFF
ON OFF
ON OFF
ON OFF
ESbY
ESPS
ESPC
EAtn
ON OFF
ON OFF
ON OFF
ON OFF
ON OFF
Key
Actual Display
On/Off Display ­Use arrow keys to turn on or off
Scroll Key
Numeric Display ­Use arrow keys to change value
Up Arrow Key
Down Arrow
PAGE 26
PROGRAM MODE FLOW CHART
Prog
A
inPS
iCor
out1
o1uL
o1LL
out2
o2uL
rLyA
rLyb
rLyC
diSP
dPoS
Euu
EuL
ON OFF
Key
Actual Display
On/Off Display ­Use arrow keys to turn on or off
Scroll Key
Numeric Display ­Use arrow keys to change value
Up Arrow Key
Down Arrow
o2LL
out3
HyCo
HyAo
A
B
PAGE 27
B
SPC
rSPu
rSPL
SPuL
SPLL
AtFr
C
Prnd
Colr
Co2r
Pout
Pou
PoL
PFF
dFF
FSCn
C
P1EC
P2EC
SPrr
D
PAGE 28
D
Com (Optional)
CCon
CbS
CAd
AduL
AdLL
ASuL
E
AAo
AtL
ASo
ON OFF
Key
Actual Display
On/Off Display ­Use arrow keys to turn on or off
Scroll Key
Numeric Display ­Use arrow keys to change value
Up Arrow Key
Down Arrow
ASLL
CrC
CAC
E
TABLE 3-2 PROGRAM MODE CONFIGURATION PROCEDURE
Press and release the SCROLL key until Prog is displayed. Use the DOWN key to enter the Program mode. Depress and release the SCROLL key to advance the display through the parameters and their values. The upper display will show the parameter codes. The lower display will show the parameter value selected. Use the UP and DOWN keys to adjust the parameter values. After adjusting a parameter, depress the SCROLL key to proceed to the next parameter. After all selections have been made, depress the UP key with a parameter code in the upper display and the lower display blank to exit the mode.
Note that parameter values are referred to in Degrees (°) and Engineering Units in the following tables. The input selection determines what the parameter values will be.
STEP DESCRIPTION DISPLAY AVAILABLE FACTORY YOUR
CODE SETTINGS SETTING SETTING
PAGE 29
1 Input Select inPS 0 = J °C Thermocouple
1 = J °F 2 = K °C 3 = K °F 4 = T °C 5 = T ° F 6 = R °C 7 = R °F 8 = S °C 9 = S °F 10 = E °C 11 = E °F 12 = B °C 13 = B °F 14 = N °C 15 = N °F 16 = C °C 17 = C °F 20 = RTD °C 21 = RTD °F 30 = 0 - 5VDC / 0 to 20mA 31 = 1 - 5VDC / 4 to 20mA 32 = 0 - 50mVDC 33 = 10 - 50mVDC
34 = 0 - 25mVDC 2 Input Correction iCor -300° to 300°/Units 3 Output 1 out1 1 = On-Off Direct (Cooling)
2 = On-Off Reverse (Heating)
3 = Time Proportioning -
Direct (Cooling)
4 = Time Proportioning -
Reverse (Heating)
5 = Current Proportioning -
Direct (Cooling)
6 = Current proportioning -
Reverse (Heating)
7 = Position Proportioning -
Open
1
0 2
4 Output 1 % o1uL 1 to 100%
Upper Limit (o1uL and o1LL will be displayed if out1 is not selected as 1 or 2)
5 Output 1 % o1LL 0 to 100%
Lower Limit
100
0
PAGE 30
STEP DESCRIPION DISPLAY AVAILABLE FACTORY YOUR
CODE SETTINGS SETTING SETTING
6 Output 2 out2 0 = None or Position
Proportioning Direct -Close 1 = On-Off Direct (Cooling) 2 = On-Off Reverse (Heating) 3 = Time Proportioning-
Direct (Cooling)
4 = Time Proportioning-Reverse
(Heating)
5 = Current Proportioning-
Direct (Cooling)
6 = Current Proportioning-
Reverse (Heating)
7 = Position Proportioning
Reverse -Close
7 Output 2 % o2uL 1 to 100%
Upper Limit (o2uL and o2LL will be displayed if out2 is selected as 3,4,5,6)
8 Output 2 % o2LL 0 to 100%
Lower Limit
9 Output 3 out3 0 = None
1 = Process Alarm-Direct 2 = Process Alarm-Reverse 3 = Deviation Alarm-Direct 4 = Deviation Alarm-Reverse 5 = Deviation Band Alarm-
Open within band
6 = Deviation Band Alarm-
Closed within band
0
100
0
0
10 Relay A rLyA 0 = Not assigned
Assignment 1 = Assigned to Output 1
2 = Assigned to Output 2 3 = Assigned to Output 3
11 Relay B rLyb Same selection as Relay A
Assignment (rLyb will be displayed if the relay is specified at the time of order)
12 Relay C rLyC Same selection as Relay A
Assignment (rLyC will be displayed if the relay is specified at the time of order)
13 Upper Display diSP 1 = Process Value (PV)
Select 2 = Deviation
14 Decimal Position dPoS 0 or 1 for T/C and RTD Input
0 to 3 for volt/mV Input
15 Engineering units Euu -9999 to 9999
Upper Value (Euu and EuL will be displayed if inPS= 30, 31, 32, 33, 34)
1
2
3
1
0
1000
STEP DESCRIPION DISPLAY AVAILABLE FACTORY YOUR
CODE SETTINGS SETTING SETTING
PAGE 31
16 Engineering units EuL -9999 to 9999
Lower Value
17 Hysteresis for HyCo 0 to 300°/Units
On/Off Control (width of hysteresis band) Output(s)
18 Hysteresis for HyAo 0 to 300 °/Units
Alarm Output (width of hysteresis band)
19 Setpoint SPC 0 to 4
Configuration 0 = Single Local Setpoint
**1 = 1 to 5VDC Remote
Setpoint and Single Local
Setpoint
**2 = 0 to 5VDC Remote
Setpoint and Single Local
Setpoint
3 = Dual Local Setpoint -
keypad selectable
**4 = Dual Local Setpoint -
Remote Dry Contact
Closure Selectable
**These features can be selected in the setpoint configuration but will not function unless the Remote Setpoint option is present, model #82XXX2X.
20 Remote Setpoint rSPu† -9999° to 9999°/Units
Upper Limit (rSPu and rSPL will be displayed if SPC is selected as 1 or 2 and model #82XXX2X has been selected)
0
3
3
0
1400*
21 Remote Setpoint rSPL† -9999° to 9999°/Units
Lower Limit
22 Setpoint SPuL# -9999° to 9999°/Units
Upper Limit
23 Setpoint SPLL# -9999° to 9999°/Units
Lower Limit
24 Automatic Transfer AtFr 0 = No automatic transfer
1 = Transfer when PV
goes below setpoint
2 = Transfer when PV
goes above setpoint
25 Process PFF 1 to 20 (# of scans averaged)
Filter Factor 1 = no filtering
26 Display dFF 1 to 20 (# of scans averaged)
Filter Factor 1 = No Filtering
27 Fast Scan FSCn 0 or 1
0 = Standard Scan -
1 per second
1 = Fast Scan -
3 per second
0*
1400*
0
0
1
1
0
† Sets scale of remote signal. # Both Local & Remote Limits
PAGE 32
STEP DESCRIPION DISPLAY AVAILABLE FACTORY YOUR
CODE SETTINGS SETTING SETTING
28 Process Rounding Prnd 1 to 100 degrees/units
1 = no rounding
29 Current Output 1 Co1r 0 = 0 to 20mADC
Range 1 = 4 to 20mADC
30 Current Output 2 Co2r 0 = 0 to 20mADC
Range 1 = 4 to 20mADC
31 Process Output Pout 0 = Not selected
1 = Assigned to
Current Output 1
2 = Assigned to
Current Output 2
32 Process Output Pou -9999 to 9999 degrees/units
Upper Value (Pou and PoL will not be seen if Pout=0)
33 Process Output PoL -9999 to 9999 degrees/units
Lower Value
34 Proportional P1EC 0 - 100%
Output 1 Action on Error Condition (P1EC will not be seen if out1=1,2)
35 Proportional P2EC 0 - 100%
Output 2 Action on Error Condition (P2EC will not be seen if out2=0,1,2,7)
1
1
1
0
2000
0
0
0
* Factory setting for Total Access
36 Setpoint SPrr 0 to 100°/Units per minute
Ramp Rate 0 = not used (Cannot be used in conjunction with Auto Tune)
0.0
Communication Parameters 37-39 are optional and will only be displayed on models 82XXX3X, 82XXX4X, 82XXX5X or 82XXX6X
37 Communications CCon 0 = Off
Configuration 1 = Monitor (Read Only)
2 = Full Communications
(Read & Write)
38 Communications CbS 1 = 300 bit rate
Bit Rate 2 = 600 bit rate
3 = 1200 bit rate 4 = 2400 bit rate 5 = 4800 bit rate 6 = 9600 bit rate
39 Communications CAd 0 to 99
Address
0, 4*
6
0, 1*
STEP DESCRIPION DISPLAY AVAILABLE FACTORY YOUR
CODE SETTINGS SETTING SETTING
PAGE 33
40 Auto Tune AduL 0° to 1000°
Deviation Upper Limit
40 Auto Tune AdLL 0 to 5000°
Deviation 0 = no lower limit Lower Limit
40 Auto Tune Setpoint ASuL -9999° to 9999°
Upper Limit
41 Auto Tune Setpoint ASLL -9999° to 9999°
Lower Limit
42 Control Response CrC 1.0 to 2.0
Criteria 1.0 = 1/4 Amplitude Decay
Response
2.0 = Damped Response
43 Control Algorithm CAC 1 = PID
2 = PI
3 = P 44 Auto Tune AAo 0 = Go into Standby at
Abort Option 0% output
1 = Go into Standby at
o1LL % output
2 = Go into Control with last
PID parameters 3 = Go into control with PID parameters of ArS1 = 0, rt1 = 0 and a) if dPoS = 0, Pb1 = 100 b) if dPoS = 1, Pb1 = 10
0
0
1400*
0
2.0
2
0
45 Auto Tune AtL 0 = No Limit
Time Limit 1 to 500 minutes
46 Auto Tune Select ASo 0 = On demand not selected
Option for On Demand 1 = On demand selected
0
0
* Whenever inPS is changed, the parameter is set to the upper limit of advertised span as indicated in the specifications section (Appendix D, page 74)
PAGE 34
TUNE MODE FLOW CHART
tunE
A
PAL
dAL
dbAL
Pb1
Pb2
rSt
rt2
Ct1
Ct2
SEnS
FoP
SoP
ON OFF
Key
Actual Display
On/Off Display ­Use arrow keys to turn on or off
Scroll Key
Numeric Display ­Use arrow keys to change value
Up Arrow Key
Down Arrow
ArS1
ArS2
rt1
A
TABLE 3-3 TUNE MODE CONFIGURATION PROCEDURE
Depress and release the SCROLL key until tunE is displayed. Use the DOWN key to enter the Tune mode. Depress and release the SCROLL key to sequence through the parameters and their values. The upper display will be the parameter code, the lower display will indicate the parameter value selected. Use the UP and DOWN keys to adjust the values. After adjusting a parameter, depress the SCROLL key to proceed to the next parameter. Use the DOWN key to advance to the next parameter code when a parameter code is showing in the upper display and the lower display is blank. After all selections have been made, depress the UP key with a parameter code showing in the upper display and the lower display blank, to exit the mode.
STEP DESCRIPION DISPLAY AVAILABLE FACTORY YOUR
CODE SETTINGS SETTING SETTING
PAGE 35
1 Process Alarm PAL -9999 to 9999 °/units
(PAL will be seen if out3=1 or 2)
2 Deviation Alarm dAL -3000 to 3000 °/units
(dAL will be seen if out3=3 or 4)
3 Deviation Band dbAL 1 to 3000°/units
Alarm (dbAL will be seen if out3=5 or 6)
4 1st Output Pb1 1 to 3000°/units
Proportional Band Width (Pb1 will not be seen if out1=1,2)
5 2nd Output Pb2 1 to 3000°/units
Proportional Band Width (Pb2 will not be
seen if out2=0,1,2,7) 6 Manual Reset rSt -1500 to 1500°/units 7 Automatic Reset ArS1 0.0 to 100.0 repeats
Output 1 (Integral) per minute 8 Automatic Reset ArS2* 0.0 to 100.0 repeats
Output 2 (Derivative) per mintue
0
0
1
100
100
0
0.0
0.0
9 Rate (Derivative) rt1 0.0 to 10.0 minutes
Output 1 10 Rate (Derivative) rt2* 0.0 to 10.0 minutes 11 Cycle Time Ct1 1 to 240 seconds
Output 1
(Ct1 will be seen if
out1=3,4,7) 12 Cycle Time Ct2 1 to 240 seconds
Output 2
(Ct2 will be seen
if out2=3 or 4) (Continued on next page)
0.0
0.0 30
30
PAGE 36
(Continued from page 35)
STEP DESCRIPION DISPLAY AVAILABLE FACTORY YOUR
CODE SETTINGS SETTING SETTING
13 Position Prop. SEnS 0.0 to 50.0%
Sensitivity (SEnS will be seen if out1=7 and out2=0 or 7)
14 First Output FoP -1000 to 1000°/units
Position
15 Second Output SoP -1000 to 1000°/units
Position (SoP will not be seen if out2=0
* ArS2 and rt2 are not used by the Control algorithm if both Output 1 and Output 2 are selected for the same proportional control (reverse or direct). The parameters are used when one output is selected for direct and the other is selected as reverse.
Note: The Program, Tune and Enable Mode selections can be conveniently recorded on the Software Reference Sheet located in Appendix E (page 76).
1.0
0
0
Auto Tune Method 3.4
The Auto Tune function will select the tuning parameters for a proportional control heating application assigned to Output 1. For the Auto Tune to properly calculate the Tune mode parameters, the Program and Tune mode parameters listed below must be correctly se­lected.
3.4.1 PROGRAM MODE PARAMETERS THAT AFFECT AUTO TUNING
1. Output 1 out1 must be set for proportioning reverse (heating) (4, 6, 7) output action.
2. Output 1 upper limit o1uL can be used to limit the maximum heating output percentage. This will affect the process response curve used to calculate the tuning parameters. If overshooting or Er56 occurs, reducing the maximum output percentage may be necessary.
3. Output 1 lower limit o1LL can be used to select a minimum output value. The instrument can be directed to output this minimum value if the Auto Tune aborts (fails) by use of the Auto Tune Abort AAo option.
4. Output 2 out2 can not be selected as time or current proportioning reverse (4 or 6). If out1 = 7, then out2 must be 7. Out2 may be used for direct cooling action.
5. Auto Tune can only be initiated when Setpoint Configuration SPC is 0, 1, 2, or 3 and SP1 is active. In other words, when SP2 or remote setpoint is active, Auto Tune can not be initiated and the AUTO TUNE key is ignored. If SPC is 4, Auto Tune can not be initiated.
6. The Auto Tune will not function if the Setpoint Ramp Rate is selected other than 0.0.
7. The Auto Tune Deviation upper limit AduL serves 2 functions: (which depend upon the Auto Tune Select option parameter selected, see step 46 on page 33).
A. If the Auto Tune Select option ASo = 0, then the process value (temperature)
must be less then the setpoint value minus the AduL value in order for the
Auto Tune to function. Auto Tune will not function if the PV > SP - AduL.
Example: if PV = 200, SP = 230 and AduL = 50, the Auto Tune will not
function (see Appendix B, page 68).
B. If ASo = 1 and the process value is greater than the setpoint value minus
AduL, the heating out1 control output will be turned off when the AUTO
TUNE key is pressed. When the process value drops below the setpoint
value minus the AduL value, the heating control output will be turned on so
the Auto Tune function can begin (see Appendix B, page 68).
Note: In order for AduL to have an effect on Auto Tune, the AduL value must
be greater than 20 degrees or 5 % of the setpoint value, whichever is greater,
initiating the Auto Tune function.
8. Auto Tune Deviation lower limit AdLL:
A. If AdLL = 0 when the Auto Tune key is pressed the Auto Tune process
response calculations will begin when the process value reaches the point
1/2 way between the setpoint value and the process value at the time when
the AUTO TUNE key was pressed. Example: If SP = 1200 and PV = 400,
then the response calculations will be considered when the PV > 800.
PAGE 37
B. If AdLL > 0, when the Auto Tune key is pressed, the Auto Tune process
response calculations will begin when the process value rises above the point
that is the result of subtracting 1/2 of the AdLL value from the setpoint value.
9. Auto Tune setpoint upper limit ASuL sets a maximum setpoint limit over which the auto tune will not initiate. Typically selected at application maximum setpoint value plus 10%.
10. Auto Tune setpoint lower limit ASLL sets a minimum setpoint limit under which the Auto Tune will not work. ASLL must be lower than ASuL (see Appendix B, page 68)
11. The Control Response Criteria CrC is used to select the desired type of control response for the process. Selecting 1.0 will provide good response to system upsets but may allow overshooting of the setpoint. Selecting a value of 2.0 may result in a slow response to system upsets but provide a stable process control. Selecting values between 1.0 and 2.0 will result in process control somewhere between the two extremes described. Actual process response will depend upon the application.
12. Control algorithm choice CAC allows selection of the type of control that best suits the process. For example, if the process acts a little unstable after Auto Tuning with PID selected, changing to the CAC PI and re-Auto Tuning may improve process stability.
13. Auto Tune abort option AAo is used to select what the controller will do if the Auto Tune function can not complete. Select the AAo parameter code that is best for your application.
14. Auto Tune time limit AtL selects a time limit that will cause the Auto Tune to abort if the process response calculations have not been completed. Start at 0, no time limit, if unfamiliar with the process reaction time needed.
15. The Auto Tune on demand ASo parameter, if selected as 0, will disable the Auto Tune function when the process variable is within the AduL value below setpoint. If
ASo is selected as 1, the Auto Tune will work when the process variable is within the AduL value below setpoint as described in number 6 previously (page 37).
PAGE 38
3.4.2 TUNE MODE
1. Manual Reset rSt should be set to 0 when performing the initial Auto Tune. This parameter may be adjusted later, if desired.
2. Cycle Time for Output 1 Ct1 may need to be adjusted when using time proportioning control. Typically the lowest cycle time settings result in the smoothest process control. However, low cycle time will reduce the life of mechanical relays. For motor modulation control, the cycle time setting must be the stroke time of the motor. Adjusting the cycle time affects the instrument operation. Shorter cycle time causes more accurate control and shorter life span of electro-mechanical components. Longer cycle time causes less control accuracy and longer life span of electro-mechancial components.
3. First Output Position deviation from setpoint FoP should be set to 0 when performing the initial Auto Tune. This may be adjusted later, if desired .
4. Second Output Position deviation from setpoint SoP, depending upon the application, may affect the process control response curve that is used by the Auto Tune calculations. Set SoP to 0 when performing Auto Tune.
3.4.3 AUTO TUNE OPERATION
1. Select the Program and Tune mode parameters as necessary for the application as described in this section.
2. Use the UP or DOWN key to select the setpoint 1 value for the application.
3. Press the AUTO TUNE key.
4. The lower display will show Atun to indicate that the Auto Tune function is operating. When the Auto Tune function is complete, Atun will not be displayed.
5. Observe the process response, if any error codes appear, consult the Trouble-shooting Section for the appropriate response (page 56).
6. If you wish to abort (stop) the Auto Tune, press the AUTO TUNE key once more. This will cause Er58 to be displayed and the controller will operate as selected by the AAo parameter.
7. For optimum control, some applications may require manual adjustments of the Tune mode parameters.
8. When the Auto Tune function has completed and the process control is satisfactory, you may wish to disable the Auto Tune function and the Tune mode to prevent inadvertant changes to the tuning parameters.
Manual Tuning Method 3.5
1. Cycle Time - Time Proportioning Outputs A. Adjusting the cycle time affects instrument operation
1. Shorter Cycle Time a. More accurate control b. Shorter life span of electro-mechanical components
2. Longer Cycle Time a. Less control accuracy b. Longer life span of electro-mechanical components
2. Proportional Bandwidth A. Proportional Bandwidth is the inverse of gain.
Increased Bandwidth = Decreased Gain
B. Increase the Proportional Bandwidth if:
1. The process overshoots excessively.
2. The process oscillates excessively.
C. Decrease the Proportional Bandwidth if:
1. The process responds slowly
2. The process fails to reach setpoint
3. Add Automatic Reset A. Increase the Automatic Reset in steps of .2 repeats per minute until the process becomes unstable, then decrease until stability is restored. B. Be sure to allow sufficient time for the process and the instrument to react.
PAGE 39
4. Rate Adjustment A. Rate can cause process instability. Typically add Rate as 1/10th of the automatic reset value. B. Decrease Rate if:
5. Manual Reset A. After making all other adjustments, use if an offset exists between the setpoint and the process variable. B. If the process is:
1. The process overshoots/undershoots
2. If the process oscillates excessively
1. Below setpoint use a positive Manual Reset value equal to the difference.
2. Above the setpoint use a negative Manual Reset value equal to the difference.
PAGE 40
Control Capability 4.1
A variety of user programmable control features and capabilities are available including:
• AutoTune • On-Off Control
• Time Proportioning Control • Current Proportioning
• Position Proportioning Control • Alarm Functions
• Dual Output Control • Auto/Manual Switching
• Automatic Transfer • Setpoint Adjustment
• Process Re-transmission The capabilities available in a specific unit are dependent upon the hardware options specified
when the instrument is ordered. Refer to Appendix C (page 58) for the decoding of the instrument model number. Current proportioning control cannot be implemented if a current output was not ordered. Position proportioning cannot be implemented if two relays (Outputs 1 and 2) and the option have not been ordered. The available output types and quantity of each are as follows:
Type of Output Quantity Available * SPST mechanical relay output Up to three * SSR Driver Up to three * mADC current output Up to two
The maximum number of SPST relay and/or SSR driver outputs available on a single instru­ment is three. Relay and SSR drivers may be assigned as either control or alarm outputs. The mADC current output(s) may be assigned control or process value retransmission output functions.
Control Responses 4.2
Each instrument may be configured to provide 3 mode proportional control. Proportional control is provided with Proportional Band, Integration, and Derivative responses. The PID parameters are defined as follows:
Out 1 Out 2 P (Proportional) Proportional Band Pb1 Pb2 I (Integration) Automatic Reset ArS1 ArS2 D (Derivative) Rate rt1 rt2
Manual Reset is provided for use in lieu of, or in conjunction with automatic reset. A cycle time adjustment parameter is provided for use with each time proportioning control output.
Direct/Reverse Operation of Outputs 4.3
Direct operation is typically used with cooling applications. On-Off direct output(s) will turn on when the process variable exceeds setpoint. Proportional direct output(s) will increase the percentage of output as the process value increases within the proportional band.
Reverse operation is typically used with heating applications. On-Off reverse output(s) will turn off when the process variable exceeds setpoint. Proportional reverse output(s) will decrease the percentage of output as the process value increases within the proportional band.
On-Off Control 4.4
On-Off control can be implemented with SPST relay or SSR driver output(s) . On-Off operation can be assigned to either or both Output 1 and 2. A hysteresis adjustment is provided for On-Off Outputs. This adjustment is in terms of degrees/engineering units and defines the bandwidth of the hysteresis. The hysteresis value straddles the setpoint. Relay chatter can be eliminated by proper adjustment of this parameter. When operating in On-Off control, the output(s) will turn on or off depending upon the setpoint, the process value, Tune mode selections, and the hysteresis adjustment.
Time Proportioning Control 4.5
Time Proportioning control can be implemented with a SPST relay or SSR driver. Time Proportioning control can be selected for either Output 1 and/or Output 2, depending on hardware configuration. Time Proportioning control is accomplished by cycling the output on and off during a prescribed period of time when the process variable is within the proportional band.
Ex: Calculated output % = 40%; Cycle time adjustment = 20 seconds Output on time = .4 x 20 = 8 seconds Output off time = .6 x 20 = 12 seconds
PAGE 41
When the unit is operating in the Control mode, the control algorithm determines the output % required to correct for any difference between the process value and the setpoint. The output calculation is affected by Tune mode parameter adjustments.
See Figure 4-1 (page 42) for proportional bandwidth effect on the output.
Current Proportioning Control 4.6
Current Proportioning control can be implemented on units provided with mADC current output(s). Current Proportioning control provides a 4 to 20mADC or 0 to 20mADC output in response to process value and setpoint. As with Time proportioning, the calculated output % for Current proportioning control is affected by the Tune mode parameter adjustments.
See Figure 4-1 (page 42) for proportional bandwidth effect on the output.
Position Proportioning Control 4.7
Position Proportioning Control can be implemented on those units provided with two SPST relay or two SSR driver outputs and the Position Proportioning (slidewire feedback) option.
Position Proportioning control permits the use of PID control when the final control element is a modulating device such as a motorized valve. Two outputs are required to control the valve. One output opens the valve, the second output closes the valve. The slidewire feedback is used to indicate the valve position to the instrument. The valve position will be dependent upon the process value, the setpoint and Tune mode parameters. (Continued on next page)
PAGE 42
(Continued from page 41)
A Position Proportioning sensitivity adjustment is provided, which specifies a deadband around the setpoint to prevent the valve from oscillating. The valve rotation time must be entered, for proper operation, into the Tune mode paramter Ct1.
See Figure 4-1 for proportional bandwidth effect on the output.
FIGURE 4-1
Proportional Bandwidth Effect On Output
100%
Output Action
100%
Output Action
0%
0%
PB=100
100 200
125 175
150
Setpoint
PB=50
150
Setpoint
Process Variable
The Proportional Bandwidth is the area where the output is a percentage of the full output. The size of the proportional band determines what change in the output will result from a change in the process variable. In the upper figure when the process variable is at 125 the output will be at 75% of full output. In the lower figure the proportional bandwidth is smaller. When the process variable is at 125 the output is now at 100%. The larger the proportional band the smaller the "gain" and vice versa.
Y
X
Dual Output Control 4.8
Dual output control can be performed when two outputs are specified. The outputs may be programmed for On-Off, Time Proportioning, or Current Proportioning, as applicable. To utilize the Auto Tune feature, Output 1 must be programmed for proportional reverse action.
The output action is dependent upon the setpoint, the process value, and Tune mode parame­ters. If two proportional outputs are selected, both output proportional bands will be biased so that 0 % of output is seen when the process value equals setpoint. The output(s) can be biased by the use of the Tune mode parameters FoP and SoP as shown below.
FIGURE 4-2
PAGE 43
100% Proportional Output 1
First Output Position = X
The first output is programmed as a proportional reverse output and the second as a propor­tional direct output. (See Glossary, page 68, for definitions of these terms). Dual propor­tioning outputs are provided with separate proportional band; auto reset, rate, and cycle time adjustments for each output.
Reverse Acting Output
Setpoint
Control
-
Direct Acting Output
Second
+
Output Position = Y
100% Proportional Output 2
Process Value
PAGE 44
Manual Operation of Proportional Outputs 4.9
To enter the Manual mode, press and release the AUTO/MANUAL key. If the Standby mode is on in the Enable mode the instrument will enter the Manual mode. The Manual mode status LED will light to indicate that the Manual mode is in use. Shifting from the Control to the Manual mode is bumpless. The proportional output(s) will stay at the last value(s) calculated by the control algorithm. The upper display will show the current process value. If Output 1 is a proportional output, the lower display will show the Output 1 percentage of output value and the PO1 status lamp will light. If PO1 is not a proportional output, the lower display will show the Output 2 percentage of output and the PO2 status lamp will light. If dual proportional outputs have been selected, press the SCROLL key to toggle the lower display between the PO1 and the PO2 values. To change the percentage of output value, press the SCROLL key to display the percentage output value that you desire to adjust. Use the UP or DOWN key to change the percentage value as desired.
To exit the Manual mode of operation press the AUTO/MANUAL key once more. The Manual mode status LED will go out. The Auto Transfer to the Control mode function can be selected in the Program mode to shift the instrument from Manual to Control mode automatically when the process variable reaches setpoint.
The proportional control output value(s) may change rapidly when returning to the Control mode. The output change will depend upon the Tune mode selections and the process value deviation from setpoint at the time of transfer.
ON OFF
Key
Actual Display
On/Off Display ­Use arrow keys to turn on or off
Scroll Key
Numeric Display ­Use arrow keys to change value
Up Arrow Key
Automatic Transfer Function 4.10
Automatic transfer provides automatic shifting from the Manual mode to the Control mode of operation when the process value reaches setpoint. This feature is selectable in the Program mode.
SETPOINT SELECT FLOW CHART
SPS
LOC
rSP
CtrL
Down Arrow
Setpoint Adjustments 4.11
g
Local Single Setpoint Local single setpoint adjustment, if selected in the Program mode, is accomplished by using the keypad. Press the UP key to increase the setpoint value. Press the DOWN key to decrease the setpoint value. Holding the key pressed will cause the value to change slowly at first then increasingly faster. The range of setpoint values can be limited by selecting the desired setpoint upper limit SPuL and the setpoint lower limit SPLL values in the Program mode. The setpoint value can be protected from inadvertent changes by disabling the Setpoint Change, ESPC, in the Enable mode.
Local Dual Setpoint Local dual setpoint adjustment, if selected in the Program mode, is accomplished by using the keypad. Press the SP1/SP2 key to select either SP1 or SP2. Press the UP key to increase the setpoint value displayed. Press the DOWN key to decrease the setpoint value displayed. Press the SP1/SP2 key to display the alternate setpoint value. Use the UP or DOWN key(s) as necessary to adjust the alternate setpoint value. The range of setpoint values can be restricted by selecting the setpoint upper limit SPuL and the setpoint lower limit SPLL values in the Program mode. Press the SP1/SP2 key to toggle the setpoint value from SP1 to SP2 and visa versa. The Auto Tune will function at the Setpoint 1 value. If the second setpoint is active when the AUTO TUNE key is pressed, the key will be ignored.
Ramp Rate A selectable Ramp Rate function can be used to limit the rate at which the setpoint used by the control algorithm will change. This feature will also establish a soft startup. Upon power up, the instrument will take the initial process value as the setpoint. A setpoint ramp rate will be calculated to increase the setpoint from the initial process value to the setpoint selected. The setpoint ramp rate feature cannot be used with the Auto Tune function.
PAGE 45
Sudden changes in the setpoint value entered via the keypad can be inhibited from affecting the control outputs by use of this feature. The internal setpoint used to control the process will ramp to the setpoint value entered at the rate of change selected.
Note: The displayed SP is not the same as the ramp SP.
FIGURE 4-3
Setpoint Ramp
205 204
Setpoint in
rees
De
203 202
201
200
05
Time in Minutes
10
PAGE 46
Remote Setpoint (Optional) The instrument setpoint can be adjusted by supplying a signal to the remote setpoint terminals as indicated in the installation section. Local or Remote setpoint operation is selected by pressing and releasing the SCROLL key until the upper display reads setpoint select SPS. Press the DOWN key to enter the Setpoint Select mode. The lower display will change to show the current setpoint mode, either local loc or remote rSP. To change the setpoint mode press the SCROLL key. To exit the setpoint mode press the UP key. To prevent unwanted setpoint mode changes, the Setpoint Select mode can be disabled in the Enable mode, ESPS. When remote setpoint is active, the AUTO TUNE key is ignored.
Remote Selection of Dual Local Setpoint (Optional) To use this feature, a remote dry contact closure needs to be connected to the instrument between terminals 8 and 5 as shown in Section 2.2, page 11. In the Program mode, set the setpoint configuration value to 3 - Local Dual Setpoint. Exit the Program mode and follow the instructions for the Local Dual Setpoint to adjust the two setpoint values that are desired. Return to the Program mode and change the setpoint configuration parameter value from 3 to 4 - Remote Selection of Dual Local Setpoint. In this configuration, the AUTO TUNE key is ignored.
When a dry contact closure is sensed between terminals 8 and 5, the setpoint value will be the SP1 value. If no contact closure is sensed, the controller will be using the SP2 value.
The setpoint values can be adjusted by using the Digital Communcations Option. Refer to the Protocol Manual (Form 2878) for more details about this option.
Process Re-transmission Output If the instrument is provided with a current output not used for process control, this output may be assigned to provide a linear re-transmission of the process value. This output can be used to provide a process signal to remotely installed recorders, panel meters and data loggers. The process output is scaled for the application by using the Program mode parameters, process value upper Pou and process value lower PoL. The current output resolution is 200 steps, so for the best re-transmission accuracy the span between Pou and PoL should be as small as possible.
The example illustrated in Figure 4-4 shows a process re-transmission application for 0 to 200 degrees F.
FIGURE 4-4
PROCESS OUTPUT / RETRANSMISSION VALUES EXAMPLE
100 %
OUTPUT
0%
0
F 200
INPUT
F
4.1.2 OUTPUT ACTION ON ERROR CONDITION
If the instrument displays a sensor problem code Hi, Lo or Snsr or any of the error codes 1-36, the On/Off Output(s) Control and Alarm will go off. The Proportional Control Outputs will go to a user selectable output % (P1EC, P2DC in the Program mode). The Process Re-transmission proportional output will go to 0%. Proportional control can be adjusted in the event of an error condition in the Manual mode. On/Off relays can be activated individually in the Test mode (Test 6).
PAGE 47
PAGE 48
Service 5.1
This section contains Calibration , Test and Trouble-shooting procedures that can be per­formed by the user. Instruments are calibrated to all input types at the factory prior to shipment. Re-calibration should not be necessary under normal operating conditions.
Calibration 5.2
Caution: Do not attempt any of these calibrations without the proper test equipment with specifications equal to or better than those listed.
Press and release the SCROLL key to sequence the upper display until CAL appears. If CAL does not appear refer to Section 3 for instructions on how to enable the Calibration mode. When CAL appears on the upper display, press the DOWN key. The display will read CAL
1. CAL 1 can be initiated at this time or press the SCROLL key to advance the display to the other calibrations available. The lower display will remain blank in the Calibration mode.
TABLE 5-1 CALIBRATION PROCEDURES
Calibration Procedure Description
CAL 1 Re-initialization of Program and Tune Mode values. CAL 2 Main Calibration used by all inputs. This is the only calibration required
for voltage and millivolt inputs.
CAL 3 Cold Junction Compensation calibration used to correct for component
variation in CJC circuit.
CAL 4 Cold Junction utility. The temperature of the cold junction is displayed.
No adjustment is made with this procedure.
CAL 5 RTD input calibration used to correct for component differences in the
RTD input circuit. CAL 6 CJC turn on/off CAL 7 Factory Use Only
CALIBRATION FLOW CHART
PAGE 49
CAL
CAL1
CAL2
CAL3
CAL4
CAL5
Prog
ON OFF
Key
Actual Display
On/Off Display ­Use arrow keys to turn on or off
CAL6
CAL7
5.2.1 CAL 1 PARAMETER INITIALIZATION
This procedure is performed to erase the information that was entered in the Program and Tune modes . All parameters will be reset to default values. Prior to beginning this procedure record the Program and Tune mode parameters so that they can be re-entered. No special test equipment is required.
With CAL1 displayed, depress and hold the DOWN key, then press the SCROLL key. The display will momentarily go blank. Release the keys. CAL1 will reappear on the display. This calibration can be done again or another may be selected.
Scroll Key
Numeric Display ­Use arrow keys to change value
Up Arrow Key
Down Arrow
PAGE 50
5.2.2 CAL2 MAIN CALIBRATION
This procedure determines and saves calibration values which correct for component varia­tions relating to the input measuring function of the instrument. CAL2 is the only calibration required for the volt and millivolt inputs. Additional calibration procedures are required for thermocouple and RTD inputs.
A 50.00 ± .01mVDC source is required for calibrating. In addition make sure that JU1 on the Processor board is in the “non volt” position. See Appendix A-2 (page 65).
With CAL2 displayed, press and hold the DOWN key, then press the SCROLL key. Release both keys and the instrument will display hLd1. Short the input terminals 1 and 3 or apply
0.00 ± mV to the input. Depress the DOWN key; dELy will appear for up to ten seconds, then SCAn will appear for up to ten seconds. A calibration reference number, which should be 0, ± 50, should then appear. With a number within tolerance display, connect a 50.00 ± .01mV source to the input terminals. Press the DOWN key and deLy will be displayed for ten seconds and the SCAn for ten seconds. Then CAL2 will reappear. If there is a problem, the appropriate error code will be displayed. Restore JU1 to the position necessary for the input type.
If the calibration reference number falls outside the 0, ± 50 tolerance, depress the SCROLL key and CAL2 will be displayed. Depress the DOWN key and perform the calibration once more. Repeat the calibration until the number falls within the tolerance limits. If the calibra­tion number remains outside these limits, check the connections to the test equipment and try the calibration again. If the number still does not approach the tolerance limits contact an Applications Service Engineer at the factory or a local representative.
Error Recovery - see 5.4 (page 56) for details. However, be sure that the millivolt source is securely connected, functioning properly and the polarity is correct. Press the DOWN key to bring the instrument back to dELy and try the calibration again. The calibration can be exited at anytime. hLd1 or the reference number is displayed by pressing the SCROLL key.
CAL2 QUICK CALIBRATION This routine will allow the operator to execute a rough calibration on their unit via the keypad with no other equipment or disturbance to established wiring. It is intended to provide a partial recovery from a calibration corruption where the necessary equipment indicated in CAL 2-5 may not be available. Is should be noted that this is not intended as a substitution to the main calibration procedure described earlier and may considerably deter from the accuracy of the instrument.
With CAL2 displayed, press and hold the DOWN ARROW key, then press the SCROLL key. Release both keys and the instrument will display hLd1. Press and hold the UP ARROW key, then press the SCROLL key. The display will momentarily blank and then CAL1 will be displayed. Release both keys and depress the UP ARROW key. CAL will be displayed.
5.2.3 CAL 3 COLD JUNCTION COMPENSATION
This procedure determines and saves calibration values which correct component variations relating to the cold junction compensation. This calibration must be preceeded by CAL2 the main calibration, to properly calibrate the instrument. These two calibrations are the only ones needed for proper operation with a thermocouple input.
For test equipment: one type J thermocouple and one mercury thermometer accurate to ± .25 degrees C or equivalent is required. Allow 30 minutes of warm up time, with the thermocouple connected, before proceeding with calibration.
With CAL3 displayed, depress and hold the DOWN key. Then press the SCROLL key and the unit will display hoLd. Release both keys. Connect the J thermocouple to the input terminals and place thermometer at the back of the unit. Press the DOWN key and deLy will
be displayed for ten seconds, then SCAn for ten seconds. The instrument will compute and display the cold junction temperature to the nearest tenth of a degree C. Compare reading with thermometer and use the UP and DOWN keys to correct the reading, if necessary. To end the procedure press the SCROLL key and CAL3 will be displayed again.
The instrument may stay in SCAn. To establish a reasonable starting point, with SCAn displayed, press the SCROLL key. CAL3 should be displayed. With CAL3 displayed, while pressing the DOWN key, press the SCROLL key. The instrument will display hoLd. Press the UP key. The instrument will begin the calibration procedure with a default value and proceed to deLy. Complete calibration as described above.
Error Recovery - see 5.4 ( page 56) for details on specific errors. The calibration can be exited at any time. hoLd is displayed by pressing the SCROLL key.
5.2.4 CAL 4 COLD JUNCTION TEMPERATURE UTILITY
This procedure displays the temperature sensed by the cold junction compensator (CJC). No special test equipment is required. With CAL4 displayed, press and hold the DOWN key then press the SCROLL key and
release both keys. SCAn will be displayed for ten seconds while the instrument computes the CJC temperature. The result will then be displayed to a tenth of a degree C. The input terminals must be shorted with a jumper wire. Remember, the temperature displayed is that of the CJC terminals not the ambient temperature. To exit, press the SCROLL key and CAL4 will be displayed.
PAGE 51
5.2.5 CAL 5 RTD INPUT
This procedure determines and saves calibration values which correct for component varia­tions relating to RTD inputs. This calibration must be preceded by CAL2 to properly calibrate the unit.
Test equipment needed will include a Decade Box (Resistance Substitution ) with .01% resolution or equivalent. Make sure the jumpers JU1 (Processor Board), JU2 and JU3 (Options boards) are in the proper positions for RTD input. See Appendix A-2 (page 65) and A-3 (page 66 and 67).
With CAL5 displayed press and hold the DOWN key, then press the SCROLL key and release both keys. hLd1 will then be displayed. Connect the Decade Box at 100 ohm setting across the input terminals 1 and 3 and Jumper terminals 1 and 5. Press the DOWN key and dELy will be displayed for up to ten seconds, then SCAn for ten seconds. When hLd2 is displayed, connect 277 ohms to the input and press the DOWN key. Again dELy will display for up to ten seconds, followed by SCAn for ten more seconds. CAL5 will be displayed after the calibration is completed.
Error Recovery - See section 5.4 (page 56) for details about specific errors. The Calibration mode can be exited any time the unit displays hLd1 or hLd2 by pressing the
SCROLL key.
5.2.6 CAL 6 COLD JUNCTION ON/OFF
With CAL 6 displayed, while pressing the DOWN ARROW key, press the SCROLL key. The instrument will display C6 and the number of the mode in effect. Mode 0 is the normal operating mode. The cold junction compensation is on. Mode 1 is the cold junction compen­sation disabled (off). Pressing the UP ARROW or DOWN ARROW will change the mode selection. The Mode 1 functions to facilitate input testing with a non-temperature compen­sated millivolt source used to simulate thermocouple input.
PAGE 52
Test Mode 5.3
The Test mode can be entered, if enabled, by pressing and releasing the SCROLL until tESt is displayed in the upper display. Press the DOWN key and tSt1 will be displayed. This test can be initiated at this time or press the SCROLL key to advance to the desired test. Test 1, 2 and 3 are performed as a block so the display will advance from tSt1 to tSt4. During the Test mode, with the exception of Test 5, the lower display will be blank.
All available test procedures are listed in TABLE 5-2 (page 53). Test 1, 2, and 3 are per­formed on start up, periodically during Control, and on entry into the Test mode. Test 4 is executed on entry into and periodically during the Control mode. These tests can be used as trouble-shooting aids.
TEST MODE FLOW CHART
tESt
tSt1
ON OFF
Key
Actual Display
On/Off Display ­Use arrow keys to turn on or off
Scroll Key
Numeric Display ­Use arrow keys to change value
tSt4
tSt5
tSt6
tSt7
tSt8
tSt9
Up Arrow Key
Down Arrow
tStA
TABLE 5-2 TEST PROCEDURES AND DESCRIPTION
TEST DESCRIPTION Test 1 Microprocessor internal RAM test.; used to verify that the processor RAM is
functioning correctly.
Test 2 External RAM test; used to test the instrument’s RAM for proper
function. Test 3 EPROM checksum test; used to check program for correct data. Test 4 External RAM checksum test; displays the number of times Error 16
and 17 have occurred. Test 5 Verifies that all keys are functional and all LED displays are working. Test 6 Used to verify that all relays and/or solid state relay driver outputs are working. Test 7 Used to check the operation of Output 1, mA current output. Test 8 Used to check the operation of Output 2, mA current output. Test 9 Auxiliary input test; used to test position proportioning (slidewire
feedback or remote setpoint voltage levels).
PAGE 53
Test A Communications hardware test; tests the send and receive functions.
5.3.1 TEST 1 - INTERNAL RAM TEST
Checks the Random Access Memory in the microprocessor. No special test equipment is required for this test. With Test 1 displayed tSt1 press and hold the DOWN key then press the SCROLL key. tSt1 will be displayed momentarily while the test is in progress. Upon successful completion the instrument will initiate Test 2 automatically.
5.3.2 TEST 2 - EXTERNAL RAM TEST
Checks the operation of the RAM external to the microprocessor. No special test equipment is required. After completion of Test1, tSt2 will be displayed momentarily while the test is in progress. Upon successful completion of Test 2, Test 3 will be initiated.
5.5.3 TEST 3 PROGRAM - EPROM TEST
This is a checksum test to verify data integrity of the stored program. No special test equip­ment is required for this test. After completion of Test 2, tSt3 will be displayed momentarily while the test is in progress. Upon successful completion the instrument will display tSt1.
5.3.4 TEST 4 - EXTERNAL RAM CHECKSUM TEST
This is a checksum test to verify the integrity of data stored in RAM and indicate the number of times the instrument has had an Error 16 or 17. No special test equipment is required for this test. With tSt4 displayed, press and hold the DOWN key then press the SCROLL key. The display will go blank momentarily, then briefly display two numbers and then tSt4 will be displayed. These numbers indicate the number of times Error 16 and 17 have occurred respectively. Test 4 can be executed again, or another test may be selected. Test 4 occurs when the instrument enters the Control mode and periodically during Control mode operation.
PAGE 54
5.3.5 TEST 5 - KEYPAD/DISPLAY TEST
This test allows the operator to verify that the keys work and that all display elements can be lighted. No special test equipment is required for this test. With tSt5 displayed press and hold the DOWN key then press the SCROLL key. The display will go blank. Release both keys, then press each key to be tested.
KEY DISPLAY SCROLL SCrL UP KEY uAro DOWN KEY dAro UP AND DOWN KEYS ALL LED’s AND SEGMENTS LIGHTED, both displays AUTO/MANUAL Auto AUTOTUNE Atun SP1/SP2 SP
To exit Test 5, press the SCROLL and UP key simultaneously. tSt5 will be displayed.
5.3.6 TEST 6 RELAY/SSR DRIVER OUTPUT TEST
Verifies that the Relay/SSR Driver output(s) are working. A volt/ohm meter will be useful to verify the output operation. With tSt6 displayed press and hold the DOWN key then press the SCROLL key. oFF will be displayed. For SPST relay outputs, connect the volt/ohm meter, set to ohms, across the relay outputs. For SSR driver outputs, connect the volt/ohm meter across the output terminals in the volt/DC mode. Depress the DOWN key repeatedly to advance through the following sequence:
DISPLAY RELAY ON
rLYA A Only rLYb B Only rLYC C Only oFF None
The relays should be checked for continuity when on and high impedance when off. SSR drivers will output 5 VDC when on and 0 VDC when off. This sequence may be repeated by using the DOWN key. To exit press the SCROLL key and tSt6 will be displayed. The existence of relay SSR outputs is dependent upon the hardware configuration.
5.3.7 TEST 7 - CURRENT OUTPUT 1 TEST
This test allows the user to verify that current Output 1 is functioning properly and will allow the adjustment of the current output value for testing of associated equipment. A volt meter with an appropriate shunt resistor or milliamp meter will be needed to execute this test. With tSt7 displayed depress and hold the DOWN key, then press the SCROLL key. Connect the DVM or milliamp meter across the output terminals 5 and 6. The display will indicate 4 milliamps output. Use the UP and DOWN keys to vary the output in 1mA steps. The current output reading should be + /- 0.5mA at any output value. To exit the test, press the SCROLL key and “tSt7” will be displayed. The existence of the mADC current output is dependent upon the hardware configuration as indicated by the model number.
5.3.8 TEST 8 - CURRENT OUTPUT 2 TEST
This test is the same as Test 7 except for Output 2. Check the output at terminals 7 and 5.
5.3.9 TEST 9 - AUXILIARY INPUT TEST
This test allows the operator to verify that the auxiliary inputs used for position proportioning (slidewire) feedback or remote setpoint is functioning properly. A variable voltage source, 5 VDC will be required to execute this test. With tSt9 displayed, press and hold the DOWN key then press the SCROLL key. The Auxiliary input voltage will be displayed to the nearest hundredth of a volt. Connect the +5V source across the Auxiliary input terminals (terminals 8 and 5) and adjust the voltage. Verify that the voltage displayed changes accordingly. The displayed voltage should be typically 0 - 5VDC +/-
0.3 volts. To terminate the test, press the SCROLL key. The display will show tSt9. The existence of the auxiliary input tested in Test 9 depends upon the hardware
configuration as indicated by the model number.
5.3.10 TEST A - COMMUNCATIONS HARDWARE TEST (Communications Option only)
This test allows the operator to verify that the communications hardware is functioning properly. With tStA displayed, press and hold the DOWN key then press the SCROLL key. The display will indicate SEnd. Each time the DOWN key is depressed, the unit will toggle between SEnd and rEC (receive). With the desired function selected, depress the SCROLL key.
In the SEnd (send or transmit) mode, the instrument will repeat the following sequence. First, the transmitter will go logic 1 for one second. Next, the transmitter will change the logic level to 0 for one second. Then, the transmitter will be disabled for one second. In the rEC mode, the transmitter will be disabled. In either mode, the instrument will monitor the line logic level. The display will be rEC0 when a logic 0 is on the line . The display will be rEC1 when logic 1 is on the line. In the SEnd mode, the unit will display rEC when the transmitter is disabled.
PAGE 55
To perform an internal test to verify the operation of the hardware, place the instrument in the Send mode. Verify that the display cycles through rEC1, rEC0, and rEC. To verify that the transmitter functions properly, two LED’s, each with a current limiting resistor, can be connected to the communications terminals, with their polarities connected opposite of each other. The following three states will be produced: one LED on, then the other LED on, then both off. Alternately, a load resistor can be placed on the terminals, the voltage generated across the load resistor is as follows: > +3 VDC then > -3VDC and then 0 VDC. The terminals used depends on the hardware ordered, either 8 and 7 or G and H.
Another test method, would be to connect one or more instruments in the Receive mode to an instrument in the Send mode. The instruments in the Receive mode should have their display alternating in sync with the instrument that is in the Send mode. When the sending unit displays rEC, the receiving units should display rEC1.
To terminate the test, press the SCROLL key for one second. Upon exit, tStA will be displayed.
PAGE 56
Trouble-shooting and Diagnostics 5.4
This section consists of two columns. The first column is a list of some possible instrument conditions. The second column is a list of steps that should improve the condition. The steps should be performed in order until the condition improves or all the steps have been tried. If the instrument condition has not improved, contact the nearest representative or the factory for assistance.
Trouble-shooting should be performed by qualified personnel using the proper equipment and following all safety precautions. Whenever possible, the trouble-shooting should be accom­plished with the electrical power disconnected. The instrument contains static sensitive components so care should be taken to observe anti-static procedures.
Condition Correction Steps
Display is blank (dark) 1. Verify that the correct instrument power, as
indicated on the wiring label on the housing, is supplied to terminals A & B. If the voltage is not correct, check the power source.
2. Turn off the instrument power. Wait about 5 seconds, then turn the power on again.
3. Turn off the instrument power, loosen the front panel screw, and remove the instrument from the housing. Inspect the instrument for poor connections.
a. The white ribbon cable that connects the Processor board (Appendix A-2, page 65) to the Power Supply Board (Appendix A­ 1, page 64) must be properly aligned and seated. b. The Front Display board pins should be properly aligned and seated in the sockets on the Processor board (Appendix A-2, page 65) and the Power Supply board (Appendix A-1, page 64). c. The Display Driver (U-1), located on the Display board, must be free of corrosion and firmly seated in the socket. Reinsert the instrument in the housing, tighten the panel screw, and turn on the power.
4. Turn off the instrument power. Press and hold the UP and DOWN keys. Turn on the power. Hold the keys depressed for about 10 seconds. If the display lights the model number, Program and Tune mode parameters will need to be re-entered (page 29 & 35 or the Software Ref. Sheet, page 77, if already filled out).
Model Number Displayed 1. Turn off the instrument power, wait 5 seconds then is incorrect reapply the power. Verify that the number dis-
played during the power up sequence is the same
Note: To re-initialize, follow steps 2 and 3.
as indicated on the label affixed to the lower front of the display bezel.
2. Turn off the power to the instrument. Press and hold the UP and DOWN keys and turn on the power. Keep the keys depressed until the model number resets to 8200. Release the keys and turn off the power.
3. To enter the correct model number press and hold the SCROLL and DOWN keys and turn on the instrument power, 8200 should be displayed. Wait about 5 seconds and release the keys. The display should remain 8200. Use the UP/DOWN keys as necessary to change the displayed number to match the first 4 digits of the model number. After adjusting the first 4 digits to the proper values, press the SCROLL key and the display will change to 000-. Use the UP/DOWN keys to set the last 3 digits of the model number to the correct values. Press the SCROLL key and the power up sequence will complete. The Program and Tune mode parameters will need to be re-entered (page 29 & 35 or the Software Ref. Sheet, page 77, if already filled out).
Relay/SSR Driver Output(s) 1. Verify that the Program and Tune mode Malfunction parameters are correctly set (pages 29 & 35 or the
Software Ref. Sheet, page 77, if already filled out).
2. Turn off the power to the instrument. Wait about 5 seconds and turn the power on again. Confirm that the model number displayed during the power up sequence indicates that the output(s) is/are present in the instrument. This number should match the number on the label affixed to the lower front of the display bezel. If model # is incorrect, follow steps for "Model # displayed is incorrect".
3. Turn off the power to the instrument. Loosen the front panel screw and remove the unit from the housing. Inspect the Power Supply board (Appendix A-1, page 64) for the presence of the output device(s). Relay A is located at K1, Relay B at K2, and Relay C at K3. A relay output will appear to be a cube. The SSR Driver will appear as a resistor and a jumper wire. The output will not work if the hardware is not present.
4. Check the output operation by performing Test 6 as described in the Test section (page 54). If the output(s) function(s) in the Test Mode re-examine the Program and Tune Mode Parameters settings (page 29 & 35, or the Software Ref. Sheet, page 77, if already filled out).
5. If the output appears not to turn off remove the power to the instrument. Loosen the front panel screw and take the unit out of the housing. Clip the resistor located on the Power Supply board (Appendix A-1, page 64) for the output(s) that seem to stay on. A .01 microfarrad, 1 KV should be connected from the terminal listed below, for the output where the resistor indicated was removed, to the AC ground.
Relay A R12 Terminal C Relay B R13 Terminal E
Relay C R14 Terminal G Return the instrument to the case and tighten the front panel screw. Turn the power on to the instrument and check the operation of the output(s).
PAGE 57
PAGE 58
mADC Output(s) 1. Verify that the Program and Tune mode Malfunction parameters are correctly set (page 29 & 35 or the
Software Ref. Sheet, page 77, if already filled out).
2. Turn off the power to the instrument . Wait about 5 seconds and turn the power on again. Confirm that the model number displayed during the power up sequence indicates that the output is present in the instrument. The number should match the model number on the label located on the lower front of the display bezel. If model # is incorrect, follow steps for "Model # displayed is incorrect" (page 56).
3. Turn off the power to the instrument. Loosen the front panel screw and remove the unit from the housing. Inspect the Option board (Appendix A-3, page 66 and 67) for the presence of the Current Output Driver IC. Current 1 output is U-1 and Current 2 output is U-5. The current output cannot function without the hardware being present . Return the instrument to the housing and tighten the front panel screw.
4. Refer to the Test section (page 52) and carry out the procedure for the output(s) that is/are not working. Test 7 operates current Output 1 and Test 8 for current Output 2. If the current output operates properly in the Test mode re-check the Program and Tune mode parameters (page 29 & 35 or the Software Ref. Sheet, page 77, if already filled out).
Error Code Displayed - The display of error codes except Er 40 - 60 will cause on/off outputs to turn off and proportional outputs to be the value selected in the Program Mode parameters P1EC and/or P2EC
SnSr 1. Inspect the sensor for proper operation and Sensor Break or out of range connection to the instrument. Acceptable sensor
ranges for the instrument are listed in the Specifications section of Appendix D (page 73).
2. Verify that the Program Mode input selection matches the sensor input connected.
3. Check that the input conditioning jumpers on the Processor board (Appendix A-2, page 65) and the Option Board (Appendix A -3, page 66 and 67) are in the proper position for the sensor input.
4. Perform the calibration procedure(s), as described in the Calibration section (page 48) , for the sensor input type.
rSEr 1. Check that the Remote Setpoint signal is present Remote Setpoint Error and of the polarity between terminals 8 (+) and
5 (-).
2. Perform the Auxiliary Input Test, Test 9 as described in the Test section (page 55), the voltage indicated during the test should be the same as measured in the preceeding step.
3. Verify that the Remote Setpoint input voltage range selected in the Program Mode (page 29) is the same as the voltage that is present at the Remote Setpoint input terminals.
FbEr 1. Inspect the Slidewire Feedback connections at Slidewire Feedback Error terminals 8, 7, and 5. Be sure that the
connections are the same as shown in the position proportioning illustration (page 20).
2. Measure the resistance of the Slidewire segment. The minimum resistance must be 135 ohms, the maximum 10 K ohms.
3. Perform the Auxiliary Input Test. Test 9 as described in the Test section, (page 55) the voltage indicated should be between 0 and 5 VDC.
4. Turn off the power to the instrument. Loosen the front panel screw and take the instrument out of the housing. Verify that the jumper JU-1 on the Option Board (Appendix A-3, page 66 and 67) is in the Motor Modulation position.
Hi - Input more than 10% 1. Perform the steps listed for the SnSr error Over Span condition (page 58).
Lo - Input more than 10% 1. Perform the steps listed for the SnSr error Under Span condition (page 58).
o - display overrange 1. If this error code is displayed as a Program or Tune (the “broken 6” appears mode parameter value , perform the Cal 1 on the left side of the display) procedure as described in the Calibration
section (page 48).
2. If this error code appears as part of the model number during the power up sequence, follow the steps listed for the "Model number incorrect" condition (page 56).
PAGE 59
Er 1 - Microprocessor RAM 1. Turn off the power to the instrument. Failure 2. Loosen the front panel screw and remove the
instrument from the housing. Inspect that the microprocessor (U1) is properly seated in the socket located on the Processor board (Appendix A-2, page
65). Return the instrument to the housing and tighten the front panel screw. Turn on the power.
Er 2 - External RAM Failure 1. Turn off the power to the instrument. Wait 5
seconds, and turn the power on.
Er 3 - EPROM Checksum 1. Perform the steps listed for Er 1 except that the Failure EPROM (U2) on the Processor board should be
inspected.
Er 4 - RTD Mismatch Error 1. Check the connections to the instrument for the
RTD Input Calibration CAL5 as described in the Calibration section (page 51). Repeat the RTD Input Calibration.
Er 5 - No Zero Crossings 1. Turn off the power to the instrument. Wait 5 Detected seconds and turn the power on.
2. Turn off the instrument power. Loosen the front panel screw and remove the instrument from the housing. Inspect the white ribbon cable that connects the Processor board to the Power Supply board. Be sure that the cable is properly aligned and seated in socket on the Power Supply board. Return the instrument to the housing and tighten the front panel screw. Turn on the power to the instrument.
3. Connect the instrument to another AC power source.
PAGE 60
Er 6 - AC line below 45 HZ 1. Turn off the power to the instrument. Wait 5
seconds and turn the power on.
2. Turn off the instrument power. Loosen the front panel screw and remove the instrument from the housing. Inspect the white ribbon cable that connects the Processor board to the Power Supply board. Be sure that the cable is properly aligned and seated in the socket on the Power Supply board. Return the instrument to the housing and tighten the front panel screw. Turn on the power to the instrument.
3. Connect the instrument to another AC power source.
Er 7 - AC line over 65 HZ 1. Turn off the power to the instrument. Wait 5
seconds and turn the power on.
2. Turn off the instrument power. Loosen the front panel screw and remove the instrument from the housing. Inspect the white ribbon cable that connects the Processor board to the Power Supply board. Be sure that the cable is properly aligned and seated in the socket on the Power Supply board.
3. Connect the instrument to another AC power source.
Er 8 - Cal 2 Volt Input Error 1. Check that 50 mVDC is properly connected to the
instrument and is within the tolerance limits as indicated in the CAL 2 procedure of the Calibration section (page 50).
2. Loosen the front panel screw and remove the instrument from the housing. Inspect the Processor board (Appendix A-2, page 65) to insure that the input conditioning jumper JU 1 is in the non-volt position.
3. Perform the CAL 2 procedure as described in the Calibration section (page 50).
Er 9 - ADC Reference Number 1. Turn off power to the instrument, wait 5 seconds, Error then turn the power on.
Er10 - ADC Reference Voltage 1. Turn off power to the instrument, wait 5 seconds, Error then turn the power on.
Er 11 - Cold Junction 1. Be sure the Cold Junction Sensor is firmly Compensation Error attached to terminals 2 and 4.
2. Perform the CAL 3 procedure as described in the Calibration section (page 50).
Er 12 - CAL 2 Voltage Error 1. Check that 50 mVDC is properly connected to the
instrument and is within the tolerance limits as indicated in the CAL 2 procedure of the Calibration section (page 50).
2. Loosen the front panel screw and remove the instrument from the housing. Inspect the Processor board (Appendix A-2, page 65) to insure that the input conditioning jumper JU1 is in the non-volt position.
3. Perform the CAL 2 procedure as described in the Calibration section (page 50).
Er 13 - RTD CAL 5 Input Error 1. Check that the resistance device is of the correct
value and properly connected to the instrument and is within the tolerance limits as indicated in the CAL 5 procedure of the Calibration section (page
51).
2. Loosen the front panel screw and remove the instrument from the housing. Inspect the Processor board (Appendix A-2, page 51) to insure that the input conditioning jumper JU1 is in the non-volt position and that the Option board jumpers JU2 and JU3 are in the RTD position.
3. Perform the CAL 5 procedure as described in the Calibration section (page 51).
Er 14 - Cold Junction 1. Be sure the Cold Junction Sensor is firmly Compensation Error attached to terminals 2 and 4.
2. Perform the CAL 3 procedure as described in the Calibration section (page 50).
Er 15 - Ground Reference 1. Turn off power to the instrument, wait 5 seconds, Tolerance Error then turn the power on.
Er 16 - Program/Tune Mode 1. Record all Program and Tune mode Parameters. Checksum Error Perform the CAL 1 procedure as described in the
Calibration section (page 49). Re-enter the Program and Tune mode Parameters (page 29 & 35 or the Software Ref. Sheet, page 77, if already filled out).
PAGE 61
Er 17 - Calibration Checksum 1. Perform the calibration procedures that are needed Error for the input sensor that will be used.
Er 20 - Setpoint Validation 1. Use the UP or DOWN key to change the setpoint Error value.
Er 36 - Incorrect Crystal 1. Turn off the power to the instrument, wait 5 seconds, For Digital Comm. then turn the power on.
Er 37 - Incorrect Micro. 1. Turn off the power to the instrument, wait 5 seconds, For Digital Comm. then turn the power on.
Er 40 - Process Value not 1. Auto Tune will not function unless the process 20 Degrees value is at least 20 degrees below the setpoint below Setpoint Value value when the Auto Tune Select ASo, is set for 0.
Changing the ASo to 1 will allow the Auto Tune to function when the process value is within 20 degrees of setpoint.
Er 41 - Process Value not 1. Auto Tune will not function unless the process 5 % of the Setpoint Value value is at least 5 % of the setpoint value below Below the Setpoint Value the setpoint when Auto Tune Select is set for 0.
Changing the ASo to 1 will allow the Auto Tune to function when the process value is within 5 % of the setpoint value below setpoint.
Er 42 - Process Value not 1. The Auto Tune will not function unless the process Below the AduL exceeds the Auto Tune Deviation from setpoint value from Setpoint value, selected in the Program Mode, if ASo is
selected as 0. Changing the ASo value to 1 will allow the Auto Tune to function when the process value is within the AduL range of the setpoint value.
PAGE 62
Er 43 - Setpoint above 1. Auto Tune will not function if the Setpoint value is ASuL value greater than the Auto Tune Setpoint Upper Limit
ASuL selected in the Program mode. Increase the ASuL value to be greater than the desired setpoint.
Er 44 - Setpoint value 1. Auto Tune will not function if the Setpoint value is ASLL value less than the Auto Tune Setpoint Value ASLL
selected in the Program mode. Decrease the ASLL value to be lower than the desired setpoint.
Er 45 - Incorrect Output 1. Auto Tune will not function unless the Output Selection(s) Configuration, in the Program mode is correct.
Output 1 must be selected as 4, 6 or 7 depending on the instrument hardware and the application. Output 2 cannot be 2, 4 or 6. (If Output 1 = 7, then Output 2 must be 7).
Er 46 - Setpoint Ramp Rate 1. Auto Tune will not function if the Setpoint Ramp Feature in Use Rate feature has been enabled. For the Auto
Tune to function the Setpoint Ramp Rate feature must be selected as 0.
Er 47 - Not in Control or 1. Auto Tune will not function unless the instrument is Standby Mode in the Control or Standby modes. Be sure that the
instrument is in either of these modes and re­ attempt the Auto Tune.
Er 48 - Auto Tune is not 1. Auto Tune will not function if Auto Tune mode is Enabled not Enabled. Enable the Auto Tune mode and
re-attempt the Auto Tune. Refer to the Enable mode section of the manual (page 24) for instructions on how to Enable the Auto Tune feature.
Er 49 - Output(s) Not 1. Auto Tune will not function unless the Output Selected Properly Configuration, in the Program mode is correct.
Output 1 must be selected as 4, 6 or 7 depending on the instrument hardware and the application. Output 2 cannot be 2, 4 or 6. (If Output 1 = 7, then Output 2 must be 7).
Er 53 - Process Value Not 1. The Auto Tune feature must sense an increasing Increasing process response to calculate the Tune mode
parameter values. Check the control device for proper operation and re-attempt the Auto Tune (page 36).
Er 54 - Process Value Not 1. The Auto Tune feature must sample a decreasing Decreasing process response during part of the Auto Tune
function to calculate the Tune mode parameter values. Check the control device for proper operation and re-attempt the Auto Tune (page 36).
Er 55 - Auto Tune Time Out 1. The Auto Tune was unable to complete the
calculations within the time allowed in the Auto Tune Time Limit AtL parameter in the Program mode. Increase the time limit value and re-try the Auto Tune (page 36).
Er 56 - Process Overshot 1. The process value exceeded the setpoint value too the Setpoint quickly for the Auto Tune calculations to complete.
a) Lower the process value further before
re-attempting Auto Tune (page 36).
b) If not used, select out2 =3 or 5 then
re-attempt Auto Tune (page 36).
c) Reduce the o1uL percentage of output
value in steps of 10% and re-attempt Auto
Tune until Er56 doesn't appear (page 36).
Er 57 - Failed Noise Test 1. The process response samples are checked to
detect erroneous values that might be caused by noise on the sensor input. If noise is detected the Auto Tune will abort. Inspect the instrument and the sensor for proper installation. Re-attempt the Auto Tune (page 36).
Er 58 - Cancelled by Operator 1. If the AUTO TUNE key is pressed while the Auto
Tune feature is active the Auto Tune will abort. Re-attempt the Auto Tune (page 36).
Er 59 - Error Occurs During 1. If a non-Auto Tune Error condition occurs while the Auto Tune Auto Tune is active, the Auto Tune will abort.
Clear the error condition and re-try the Auto Tune (page 36).
PAGE 63
Er 60 - Power Interrupt 1. If the instrument power goes off while the Auto During Auto Tune Tune is active it will cause the Auto Tune to abort.
Re-attempt the Auto Tune (page 36).
Momentary Er 70 - 1. Tried to communicate while unit was in a non- Controller unable to respond control mode. within 250 milliseconds
Momentary Er 71 - 1. The unit received a request before proper amount Byte received before the of time has elapsed since last request. response was transmitted
Momentary Er 72 - 1. Data received not valid, possible corruption on Incorrect Block Check the commlink. Possible noise. character was received
Momentary Er 73 - 1. Improper parity selection on the transmitting Byte received with incorrect terminal. parity 2. Incorrect baud rate.
3. Noise.
PAGE 64
Appendix A Board Layout - Jumper Positioning
FIGURE A-1 - Power Supply Board
TOP
K3
R14
R13
RELAY C
K2
RELAY B
115 VAC JUMPER POSITION
JU6 JU5
FRONT OF UNIT
R12
RELAY A
K1
COMPONENT SIDE
230 VAC UNITS MAY BE FIELD CONVERTED TO 115 VAC BY MOVING JUMPERS AS SHOWN ABOVE . (115 VAC UNITS CANNOT BE FIELD CONVERTED TO 230 VAC!!)
T1
230 VAC JUMPER POSITION
JU6 JU5
T1
FIGURE A-2 - Processor Board
PAGE 65
Revision L, M, and M2
FRONT OF UNIT
JU1
MICRO U1
BATTERY
T/C,mV,RTD,CAL2
VOLT
TOP
JU2
Y1
REV
EPROM U2
RAM U3
COMPONENT SIDE
JU2
ENABLE MODE LOCKED
ENABLE MODE UNLOCKED
Note: Locked and unlocked positions differ from Rev J and below and M3 and above.
Located on solder side
JU12
JU1
JU12
IF NOT OPTION BOARD
FRONT OF UNIT
JU1
Revision J and below AND M3 and above
TOP
MICRO U1
EPROM U2
BATTERY
T/C,mV,RTD,CAL2
VOLT
RAM U3
COMPONENT SIDE
JU2
JU2
Y1
REV
ENABLE MODE UNLOCKED
ENABLE MODE LOCKED
Located on solder side
JU12
Rev. M2 and above only
JU1
JU12
IF NOT OPTION BOARD
PAGE 66
FIGURE A-3 - Option Board, Revision E and above
2nd 4-20
Position
Prop.
Com
RS-485
JU1
JU3
JU2
T/C, mV, V
RTD
JU2
J15
J16
JU3
JU1
JU13
T/C, mV, V
RTD
REV
JU14
Com &
2nd 4-20
Pos. Prop. &
Alt. Com
RSP &
Com
RRH &
Com
Alt.
Com
XPS
JU13
XPS
J15 - AC Input XPS cable from transformer J16 - XPS to Relay C
XPS
JU14
XPSNo XPS No XPS
FIGURE A-3 - Option Board, Revision D and below
REV
FRONT OF UNIT
For 2nd 4-20mA, U5 is populated
PAGE 67
TOP
JU1
U5
U1
For 1st 4-20mA, U1 is populated
JU1
2ND 4-20 mADC
MOTOR MODULATION/ POSITION PROPORTIONING POTENTIOMETER REMOTE SETPOINT
DIGITAL COMMUNICATIONS 422/485
COMPONENT SIDE
JU2
JU3
JU2
RTD
T/C, mV, VOLT
JU3
(NON-RTD)
RTD
T/C , mV , VOLT (NON-RTD)
PAGE 68
Appendix B Glossary of Terms
Automatic Reset (Integration)
Automatic reset is a Tune mode parameter that will bias the proportional output(s) to compen­sate for process load variations. This parameter is adjustable from 0.0 to 100.0 repeats per minute. Factory default is 0.0. The display codes are ArS1 for proportional Output 1 and/or ArS2 for proportional Output 2. The Auto Tune feature will select the ArS1 setting for a heating output.
Automatic Transfer
Automatic transfer is a feature selected in the Program mode that will allow the instrument to shift from the Manual to the Control mode of operation automatically when the process value reaches setpoint.
Auto Tune
Auto Tune automatically determines the Tune mode parameters for a proportional heating output assigned to Output 1. The Ziegler - Nichols method is used to determine the Tune mode parameters.
Auto Tune Deviation Lower Limit
If AdLL = 0, when the AUTO TUNE key is pressed, the process response calculations will occur during the time the process variable rises to the point 1/2 of the way between the setpoint value and the process value when the key was pressed.
If AdLL > 0, when the AUTO TUNE key is pressed the process response calculations will begin when the process value rises above the point that is the result of subtracting AdLL from setpoint.
Auto Tune Deviation Upper Limit If ASo = 0, the Auto Tune function will not operate unless the process value is < the SP- AduL value. This can be useful to prevent unwanted retuning of the process when the
process value (PV) is > the Setpoint value minus the AduL value. If ASo = 1, Auto Tune will function if the PV is within the AduL value from setpoint. However,
if the PV is > SP-AduL, the heating output will be turned off until the PV < SP-AduL. At this point the heating output will be turned on so the control repsonse can be calculated.
In order for AduL to have an effect on Auto Tuning, the AduL value should be greater than 20 degrees or 5 % of the setpoint value, whichever is greater, to initiate the Auto Tune function.
SP
Auto Tune will not initiate if PV is above this line, when the AUTO TUNE key is depressed
SP - AduL
Auto Tune will function if the PV is within this area (below SP-AduL), when the AUTO TUNE key is depressed
ASuL and ASLL are Program mode parameters that can be used to establish upper and
L
L
e
e
e
lower setpoint values outside of which the Auto Tune feature will not function. The Auto Tune feature will not function if the process value is greater than ASuL or below ASLL.
No Auto Tun
ASu
Auto Tun
PAGE 69
ASL
No Auto Tun
Auto Tune Time Limit
This feature is used to automatically abort the Auto Tune function if the control response calculations have not been completed within the time allotted.
Balanceless Transfer
This feature prevents changes in proportional output when changing from the Manual to Control mode of operation. When transferring from the manual mode to the control mode, the proportional outputs will be "Balanceless" regardless of whether the unit is inside or outside the proporitonal band. This only holds true if the Auto Reset (ArSt) value is greater than 0.
Bumpless Transfer
This feature prevents changes in proportional outputs when changing from the Control to the Manual mode of operation.
Choice of Control Algorithm
This Program mode parameter is used to select the control algorithm that will be used for Auto Tune calculations.
Control selections are PID, PI, or P only . PI is the factory default.
Control Algorithm
A pre-programmed series of instructions that are used by the instrument when determining the status of the output(s).
Control Response Criteria
This Program mode parameter selects the type of control response desired for the application when using the Auto Tune feature.
The control response selections are from 1.0 for 1/4 wave decay response to 2.0 which is a damped response.
Setpoint
CRC - Control Response Criteria available settings
1.0 = 1/4 Wave Decay Response
2.0 = Damped Response
NOTE: Actual damped response may vary depending on the control system and the application.
PAGE 70
e
e
Cycle Time
This Tune mode parameter is used to select the on/off cycle time for time proportioning outputs (Ct1 for Output 1 and/or Ct2 for Output 2). (See page 41, Section 4.5)
When using the Position Proportioning option, Ct1 must be selected for the stroke time of the motor.
Deviation Band Alarm (Output 3)
This feature can be used to provide an indication that the process value has deviated outside of a selectable deviation tolerance band value that strattles the setpoint. If in the Program mode out3 is selected as 5, an assigned output will be off as the process value is within the deviation band about the setpoint. The output will turn on if the process value falls below or goes above the deviation band about the setpoint. The amount of the deviation band is selected in the Tune mode parameter, dbAL.
On
Output Action
Deviation Band
Off
Process Valu
If out3 is selected as 6, an assigned output will be on as long as the process value is within the deviation band about the setpoint. The output will turn off if the process value falls below or goes above the deviation band about the setpoint. The deviation band value is selected in the Tune mode, dbAL.
SP
On
Output Action
Deviation Band
Off
Process Valu
Display Filter Factor
This Program mode parameter is used to dampen the process value displayed. The selec­tions range from 1 through 20, the value represents the number of process scans that will be averaged for the display value. Factory default is 1, no filtering.
SP
Engineering Units Upper and Engineering Units Lower
These Program mode parameters are used with volt, millivolt, and milliamp inputs. The Engineering Units Upper Euu should be selected as the value to be displayed when the input is at maximum. Th Engineering Units Lower EuL should be selected as the value to be displayed when the input is at minimum.
First Output Position
This parameter is adjustable from -1000 to 1000 units and represents a shift or offset of the on-off actuation points or proportional band for the first output relative to the normal position. For example, a negative value could be used to offset an expected overshoot. First OutputPosition also shifts the proportional band with respect to the process value range output of which integral action is inhibited. Factory default is 0. Display code is FoP.
Hysteresis for Alarm Outputs
This Program mode parameter is used to create a deadband for the alarm output(s). For example, a process direct alarm is selected for 200 °/units with a HyAo set for 20°/units. The output assigned will turn on when the process value exceeds 210°/units. The output will turn off when the process value goes below 190°/units.
Be aware that this parameter will also effect the output action if used as a Deviation or Deviation Band Alarm. (i.e.the Deviation Band will be increased by the amount of HyAo selected)
This parameter is adjustable from 0 to 300 °/units. The factory default is 3. The display code is HyAo.
PAGE 71
Hysteresis for Control Outputs
This Program mode parameter is used to create a deadband for On/Off control outputs. This parameter is adjustable from 0 to 300 degrees/units. Factory default is 3. This feature can be used to reduce the cycling of the on/off outputs. The display code is HyCo.
Input Correction
This parameter is used to adjust the process variable value to compensate for sensor errors. This Program mode parameter is selectable from -300 to + 300 degrees/units. The factory default is 0. The display code is iCor.
Manual Reset
This parameter is adjustable from -1500 to 1500 units representing a manual shift of propor­tional band(s) relative to the normal position. Manual reset is intended to be used when automatice reset is not used to allow compensation for deviations from setpoint which remain after the process has stabilized. Factory default is 0. Increasing the value increases the process variable, i.e. if the process variable stabilized too low, increase the manual set. Integral action, and conversely reset-windup inhibit apply over the same process value range regardless of the manual reset value. The display code is rSt.
Process Filter Factor
This Program mode parameter is used to dampen the process value used to calculate output action. The process value is averaged to dampen the control outputs. This parameter is adjustable from 1 to 20 . Factory default is 1. The display code is PFF.
Process Retransmission Output
This parameter allows for a linear milliamp proportional output relative to the process value. The current output may be scaled over a range selectable by the user. This output can be used to supply the process variable signal to remote chart recorders, panel meters, and data logger instruments.
PAGE 72
Process Rounding
This Tune mode parameter is used to determine the step size of the process value that will be seen on the display. This feature can be used to reduce display fluctuation. This parameter is adjusted from 1 to 100 degrees/units. The factory default is 1,no rounding (e.g. Process rounding = 2, Process Value Display - 4, -2, 0, 2, 4, etc.).
Process Variable
The process variable refers to the condition of the process being measured (sensed). The instrument will accept process inputs other than temperature (pressure, level, flow, etc.).
For the Auto Tune feature to work, the process value must be temperature from a thermocouple or an RTD.
Proportional Band
This Tune mode parameter selects the span of the proportional output range. This parameter is adjustable from 1 to 3000 degrees/units. Factory default is 100. If Output 1 is selected as a proportional output, a display code of Pb1 will be seen. If Output 2 is selected as a propor­tional output, the display code will be Pb2. The Pb1 value will be automatically selected if the Auto Tune function is used.
Rate (Derivative)
This Tune mode parameter is adjustable from 0.0 to 10.0 minutes. The value selected represents how much sooner a PID instrument will recover from a process upset than a PI only instrument will recover to the same process upset. If Output 1 is selected as a propor­tional output, rt1 will be displayed. If Output 2 is selected as a proportional output then rt2 will be displayed. The rt1 value will be automatically selected if the Auto Tune function is used.
Second Output Position
This Tune mode parameter is used to shift (deviate) the Output 2 operating point from setpoint. This parameter is adjustable from -1000 to + 1000 degrees/units. Factory default is
0. The display code is SoP.
Setpoint Ramp Rate
This Program mode parameter provides a rate of change control of the instrument setpoint value. This parameter is used to inhibit sudden upsets in the instrument control caused by large setpoint changes. This feature also creates a soft start when the instrument power is turned on. The instrument will read the process value at the time the power was turned on as the setpoint value. A rate of change ramp will change the internal setpoint to the setpoint selected.
Appendix C - Order Matrix
PAGE 73
8
Output 1
1 Relay 2 SSR Driver 3 4-20mA & Relay 4 4-20mA & SSR Driver
Output 2
0 None 1 Relay 2 SSR Driver 3 4-20mA 4 4-20mA & Relay 5 4-20mA & SSR Driver
Alarm/ Output 3
0 None 1 Relay 2 SSR Driver
Remote
0 None 1 Position Prop. * 2 Remote Setpoint 3 RS 422/485 Std. Com ** 4 RS 422/485 Std. Com *** 5 RS 422/485 Total Access Com** 6 RS 422/485 Total Access Com with Alternate Conn.***
2
Voltage
1 115VAC Input & Relays 2 230VAC Input & Relays 3 115VAC Input, 230VAC Relays
Option Suffix
None (blank) AB Extended Features Software XP 24VDC Transmitter Power Supply XA 24VDC Power Supply
* Limited to Model 8211X1X or 8222X1X ** Limited to Model 82X0X3X, 82X1X3X or 82X2X3X and 82X0X5X, 82X1X5X or 82X2X5X. Output 2 cannot be 3,4,5. *** Limited to Model 82X304X,82X404X, 82X504X and 82X306X, 82X406X, or 82X506X. The Alarm/ Output 3 must be 0.
PAGE 74
Appendix D - Specifications Input Specifications
THERMOCOUPLE
TYPE RANGE TYPE RANGE
J 0 TO 760C E 0 TO 750C
0 TO 1400F 0 TO 1400F
K 0 TO 1370C B 200 TO 1800C
0 TO 2500F 400 TO 3300F
T -200 TO 400C N 0 TO 1300C
-330 TO 750F 0 TO 2370F
R 200 TO 1650C C 200 TO 2300C
400 TO 3000F 390 TO 4170F
S 200 TO 1650C
400 TO 3000F
RTD VOLTS MILLIVOLTS
100 ohm 0 to 5 VDC 0 to 25 mVDC (.00385 OHM/OHM/C) 1 to 5 VDC 0 to 50 mVDC
-140 to 400C 10 to 50 mVDC
-220 to 750F
MILLIAMPS REMOTE
SETPOINT
* 0 to 20 mADC 0 to 5 VDC
1 to 5 VDC * 4 to 20 mADC is accommodated via the 1-5 VDC input selection with the addition of a shunt resistor.
SENSOR FAULT DETECTION
Displays Hi or Lo process input for thermocouple or RTD inputs (10% above or below range) and sensor break, SnSr. On/Off outputs go off, proportional outputs go to user selectable output %. Sensor fault detection is not functional for 0 to 5 VDC.
Output Specifications
CONTROL OUTPUT 1 AND 2
Relay Output SPST
115 VAC: 5.0 A Resistive; 1/8HP or 250 VA 230 VAC: 2.5 A Resistive; 1/8HP or 250 VA
SSR Driver Open collector output
Short circuit protected at 100 mA maximum Provides 4 VDC at 20 mA or 3 VDC at 40 mA
Current Output 0-20mADC or 4-20 mADC into 650 ohms maximum.
ALARM OUTPUT
Relay Output SPST
115 VAC: 5.0 A Resistive; 1/8HP or 250 VA 230 VAC: 2.5 A Resistive; 1/8HP or 250 VA
SSR Driver Open collector output
Short circuit protected at 100 mA maximum Provides 4 VDC at 20 mA or 3 VDC at 40 mA
PAGE 75
Display Specifications
Upper Digital Display Four (4) 7 segment LED’s each; .36 inches high Lower Digital Display Four (4) 7 segment LED's each; 36 inches high
Status Indicators Individual LED indicators for Remote Setpoint, Setpoint 1,
Setpoint 2, Process Value,Out 1, Out 2, Manual, Alarm, Degrees F, Degrees C, or Engineering Units, minus sign for negative values (one for each display), Percentage of Output 1, and Percentage of Output 2.
Alarm Adjustment Specifications
Process Alarm -9999 to 9999 units Deviation Alarm -3000 to 3000 units Deviation Band Alarm 1 to 3000 units
Control Adjustments Specifications
On/Off Hysteresis 0 to 300 units Proportional Band 1 to 3000 units Manual Reset -1500 to 1500 units Auto Reset 0.0 to 100.0 repeats/minute Rate 0.0 to 10.0 minutes Cycle Time 1 to 240 seconds Position Proportioning Sensitivity 0.0 to 50.0 % First Output Position -1000 to 1000 units Second Output Position -1000 to 1000 units
PAGE 76
Performance Specifications
Measurement Error Limit • Type J,K,T,E,N, & C thermocouples and RTD
+ or - 0.25% of reading plus 1 degree at 25°C
• Type R,S, & B thermocouple + or - 0.25% of span at 25°C
• mVDC, mADC and VDC + or - 0.25% of scaled span plus
1 least significant digit at 25 degrees C Ambient Temp. Error 0.01% of span per degree C deviation from 25 degrees C Scan Rate 1 scan per second, 3 scan per second selectable Display Resolution 0 to 3 decimal places (depending upon input type selected) Auto Reset Windup Inhibit Auto reset is disabled when the process is outside of the
proportional band Cold Junction Self compensation for ambient temperature. All calibration
Compensation values are stored in memory Noise Rejection Normal mode, 85dB minimum at 60 Hz or greater.
Common mode, 90dB minimum + /- 24VAC maximum for RTD
input, 115 VAC maximum for other inputs Line Voltage 115/230 VAC + /- 10% 50/60 Hz Power Consumption 15VA maximum Operating Temperature 0 to 55 degrees C
32 to 131 degrees F Storage Temperature -40 to 65 degrees C
-40 to 149 degrees F Humidity 0 to 90% RH, noncondensing Dimensions 1/4 DIN front panel (96mm X 96mm) 5.8 inches deep Weight 3 pounds maximum Vibration 0.5 to 100 Hz at 0.5g Agency Approvals UL and CSA Warranty 3 years, details on the inside back cover.
Appendix E Software Record/Reference Sheet
Model Number
PAGE 77
Program Mode
InPs Icor out1 o1uL o1LL out2 o2uL o2LL out3 rLyA rLyb rLyC diSP dPoS Euu EuL HyCo HyAo SPC rSPu rSPL SPuL SPLL AtFr PFF dFF FACn Prnd Co1r Co2r Pout Pou PoL P1EC P2EC
Program Mode Continued
SPrr CCon CbS
Comm
CAd AduL AdLL ASuL ASLL CrC CAC AAo AtL ASo
Tune Mode
PAL dAL dbAL Pb1 Pb2 rSt ArS1 ArS2 rt1 rt2 Ct1 Ct2 SEnS FoP SoP
Enable Mode
ENAB ON OFF EtSt ECAL EPro Etun ESby ESPS ESPC EAtn
Warranty and Return Statement
These products are sold by the factory under the warranties set forth in the following paragraphs. Such warranties are extended only with respect to a purchase of these products, as new merchandise, directly from the factory or from a factory distributor, representative or reseller, and are extended only to the first buyer thereof who purchases them other than for the purpose of resale.
Warranty
These products are warranted to be free from functional defects in materials and work­manship at the time the products leave the factory and to conform at that time to the specifications set forth in the relevant instruction manual or manuals, sheet or sheets, for such products for a period of three years.
THERE ARE NO EXPRESSED OR IMPLIED WARRANTIES WHICH EXTEND BEYOND THE WARRANTIES HEREIN AND ABOVE SET FORTH. THE FACTORY MAKES NO WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE WITH RESPECT TO THE PRODUCTS.
Limitations
The factory shall not be liable for any incidental damages, consequential damages, special damages, or any other damages, costs or expenses excepting only the cost or expense of repair or replacement as described above.
Products must be installed and maintained in accordance with the instructions. Users are responsible for the suitability of the products to their application. There is no warranty against damage resulting from corrosion, misapplication, improper specifications or other operating condition beyond our control. Claims against carriers for damage in transit must be filed by the buyer.
This warranty is void if the purchaser uses non-factory approved replacement parts and supplies or if the purchaser attempts to repair the product themselves or through a third party without factory authorization.
Returns
The factory’s sole and exclusive obligation and buyer’s sole and exclusive remedy under the above warranty is limited to repairing or replacing (at factory’s option), free of charge, the products which are reported in writing to the factory at its main office indicated below.
The factory is to be advised of return requests during normal business hours and such returns are to include a statement of the observed deficiency. The buyer shall pre-pay shipping charges for products returned and the factory or its representative shall pay for the return of the products to the buyer.
Approved returns should be sent to: 2 CAMPION ROAD
NEW HARTFORD, NY 13413 USA
PAGE 80
THE PARTLOW-WEST COMPANY
2 CAMPION ROAD • NEW HARTFORD, NY 13413 USA
1-800-866-6659 • 315-797-2222 • FAX 315-797-0403
Loading...