Manual Inflation Blood Pressure Monitor
Model M1 Eco
• Instruction Manual
РУКОВОДСТВО ПО ЭКСПЛУАТАЦИИ
1649511-0F
12K2607
EN
RU
HEM-4011C-EN-RU.book 2 ページ 2012年1月6日 金曜日 午後2時50分
Contents
Before using the unit
Introduction ...........................................................................3
Important Safety Information ...............................................4
1.Overview ............................................................................. 6
2.Preparation .........................................................................9
2.1 Installing/Replacing the Batteries..................................9
2.2 Setting the Date and Time ..........................................11
2.3 Adjusting the Settings Modes ..................................... 14
Operating instructions
3.Using the Unit................................................................... 15
3.1 How to Sit Correctly When Taking a Measurement....15
3.2 Applying the Arm Cuff .................................................17
3.3 Taking a Reading........................................................20
3.4 Using the Memory Function ........................................ 25
4.Quick Reference Guide.................................................... 28
Care and maintenance
5.Handling Errors and Problems .......................................29
5.1 Error Messages...........................................................29
5.2 Troubleshooting ..........................................................31
6.Maintenance and Storage................................................33
7.Optional Parts...................................................................36
8.Technical Data..................................................................37
9.Some Useful Information about Blood Pressure ..........39
2
HEM-4011C-EN-RU.book 3 ページ 2012年1月6日 金曜日 午後2時50分
Introduction
Thank you for purchasing the OMRON M1 Eco Upper Arm Blood Pressure Monitor.
The OMRON M1 Eco is a manual inflation blood pressure monitor, operating on the oscillometric principle. It measures your blood pressure and pulse rate simply and quickly.
The unit also stores up to 42 measurements in memory and calculates an average reading based on the three most recent measurements taken within 10 minutes of the last reading. (If there are only two readings in memory for that period, the average will be based on two readings. If there is only one reading in memory for that period, that reading will be displayed as the average value.)
Please read this instruction manual thoroughly before using the unit. For specific information about your own blood pressure, CONSULT YOUR DOCTOR.
EN
3
HEM-4011C-EN-RU.book 4 ページ 2012年1月6日 金曜日 午後2時50分
Important Safety Information
Consult your doctor during pregnancy, arrhythmia and arteriosclerosis. You should never change the dosage of medication prescribed by your doctor. Use the M1 Eco only for yourself.
Warning:
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
(General Usage)
• Always consult your doctor. Self-diagnosis of measurement results and self-treatment are dangerous.
• People with severe blood flow problems, or blood disorders, should consult a doctor before using the unit. Cuff inflation can cause internal bleeding.
(Battery Usage)
• If battery fluid should get in your eyes, immediately rinse with plenty of clean water. Consult a doctor immediately.
Caution:
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or patient or damage to the equipment or other property.
(General Usage)
• Do not leave the unit unattended with infants or persons who cannot express their consent.
• Do not use the unit for any purpose other than measuring blood pressure.
• Do not disassemble the unit or arm cuff.
• Do not inflate the arm cuff over 299 mmHg.
• Do not use a mobile phone, or other devices that emit electromagnetic fields, near the unit. This may result in incorrect operation of the unit.
• Do not operate unit in a moving vehicle (car, airplane).
4
HEM-4011C-EN-RU.book 5 ページ 2012年1月6日 金曜日 午後2時50分
Important Safety Information
(Battery Usage)
• If battery fluid should get on your skin or clothing, immediately rinse with plenty of clean water.
• Use only four “AA” alkaline batteries with this unit. Do not use other types of batteries.
• Do not insert the batteries with their polarities incorrectly aligned.
• Replace old batteries with new ones immediately. Replace all four batteries at the same time.
• Remove the batteries if the unit will not be used for three months or more.
• When the batteries are replaced, you may need to reset the date and time. If the year is flashing on the display screen, refer to “2.2 Setting the Date and Time”.
• Do not use new and used batteries together.
General Safety Precautions
• Do not inflate the arm cuff when it is not wrapped around your arm.
• Do not apply strong shocks and vibrations to or drop the unit.
• Do not take measurements after bathing, drinking alcohol, smoking, exercising or eating.
• Do not wash the arm cuff or immerse it in water.
• Read and follow the “Important information regarding Electro Magnetic Compatibility (EMC)” in the Technical Data Section.
• Read and follow the “Correct Disposal of This Product” in the Technical Data Section when disposing of the device and any used accessories or optional parts.
Save these instructions for future reference.
EN
5
B
A
C C
E
D
F
M1 Eco
HEM-4011C-EN-RU.book 6 ページ 2012年1月6日 金曜日 午後2時50分
1.Overview
1. Overview
Main unit
A. Display B. SET button C. Air connector D. MEMORY button E. I/O button
6
F. Battery compartment
HG J
K
I
HEM-4011C-EN-RU.book 7 ページ 2012年1月6日 金曜日 午後2時50分
Arm Cuff
G. Arm cuff
(Medium cuff: arm circumference 22-32 cm)
H. Air Tube
If air starts to leak from the arm cuff, please replace with a new one. (Refer to Chapter 7 “Optional Parts”)
Inflation Bulb
1.Overview
EN
I. Air Release Button J. Air Tube K. Air Inflation Bulb
7
L
R
T
W
M
O
P
N
S
U
V
Q
YX
HEM-4011C-EN-RU.book 8 ページ 2012年1月6日 金曜日 午後2時50分
1.Overview
Display
L. Systolic blood pressure M. Diastolic blood pressure N. Memory symbol
(Displayed when viewing values stored in memory)
O. Average value symbol
(Displayed when viewing average value for the last three
measurements) P. Buzzer symbol Q. Battery low symbol
Package contents
X. Storage case Y. Four “AAA” alkaline (LR03)
batteries
R. Heartbeat symbol
1. (Flashes during measurement)
2. (If flashing after measurement is complete, indicates blood pressure out of recommended range)
S. Pulse T. Deflation symbol U. Date/Time V. Irregular heartbeat symbol W. Reinflation symbol
• Instruction manual
• Guarantee card
• Blood pressure pass
• Blood Pressure Monitor unit
8
HEM-4011C-EN-RU.book 9 ページ 2012年1月6日 金曜日 午後2時50分
2. Preparation
2.1 Installing/Replacing the Batteries
1. Turn the main unit upside down.
2. Slide the battery cover in the
direction of the arrow while pressing the ribbed part of the cover.
3. Install or replace four AAA
size batteries so that the + (positive) and - (negative) polarities match the polarities indicated on the battery compartment.
4. Put the battery cover back in
place.
Slide the battery cover as indicated, until it clicks into place.
2.Preparation
EN
Note: The measurement values continue to be stored in
memory even after the batteries are replaced.
9
HEM-4011C-EN-RU.book 10 ページ 2012年1月6日 金曜日 午後2時50分
2.Preparation
Battery Life & Replacement
If the battery low symbol ( ) appears on the display, replace all four batteries at the same time.
- When the battery low symbol ( ) starts to blink, you will still be able to use the unit for a short while. You should replace the batteries with new ones ahead of time.
- When the symbol ( ) remains lit, the batteries are exhausted. You should replace the batteries with new ones at once.Turn the unit off before replacing the batteries.
• Remove the batteries if the unit will not be used for three months or more.
• If the batteries are removed for 30 seconds, the Date/Time setting will need to be reset. See “2.2 Setting the Date and Time” for details.
• Dispose of batteries according to applicable local regulations.
Four new
AAA alkaline batteries will last for approximately
1500 measurements, when used to take two measurements a day.
Since the supplied batteries are for monitoring use only, they may have a shorter life and not last for 1500 measurements.
10
HEM-4011C-EN-RU.book 11 ページ 2012年1月6日 金曜日 午後2時50分
2.2 Setting the Date and Time
Your blood pressure monitor automatically stores up to 42 measurements values, including date and time of measurement, in its memory and calculates an average value based on the three most recent measurements taken within 10 minutes of the last reading.
To make use of the memory and average values functions:
• Set the unit to the correct date and time before taking a measurement for the first time.
• If the batteries have been removed for more than 30 seconds, the date and time setting will need to be reset.
1. When the I/O button is
pressed to turn the unit on for the first time after inserting batteries, the year digits (2005) will flash on the display.
2. Press the MEMORY button to
advance the digits one at a time.
2.Preparation
EN
Notes:
• The range for the year setting is 2005 to 2030. If the year reaches 2030, it will return to 2005.
• If you hold down the MEMORY button, the digits will advance rapidly.
11
HEM-4011C-EN-RU.book 12 ページ 2012年1月6日 金曜日 午後2時50分
2.Preparation
3. Press the SET button to confirm the setting when the desired
number appears on the display.
The year is set and the month digits flash on the display.
4. Repeat steps 2 and 3 to set
the month.
The month is set and the day digits flash on the display.
5. Repeat steps 2 and 3 to set
the day.
The day is set and the hour digits flash on the display.
6. Repeat steps 2 and 3 to set
the hour.
The hour is set and the minutes digits flash on the display.
12
HEM-4011C-EN-RU.book 13 ページ 2012年1月6日 金曜日 午後2時50分
7. Repeat steps 2 and 3 to set
the minutes.
The minutes settings is set.
The unit automatically switches to standby mode after the minute setting has been set.
To adjust the date and time, press the SET button twice while the monitor is in standby mode.
Notes:
• After the buzzer setup is completed, the time/date setting will flash on the display.
• After each setup is completed, please push the I/O button to turn off the monitor.
2.Preparation
13
EN
HEM-4011C-EN-RU.book 14 ページ 2012年1月6日 金曜日 午後2時50分
2.Preparation
2.3 Adjusting the Settings Modes
You can modify the options for the various settings of your unit. This is done by pressing the SET button to select a setting, then pressing the MEMORY button to select the options for that setting. After selecting a setting, press the I/O button to confirm the setting and turn the power off.
Buzzer Mode
When Buzzer mode is on, the unit emits sounds, during measurement (to indicate that pulse waves are detected), and when measurement is complete.
1. Press the SET button once
while the unit is in standby mode.
The buzzer symbol appears on the display.
2. Press the MEMORY button to select
on or off.
The setting alternates between “on” and “off” with each press of the MEMORY button.
The default setting is “on”.
3. When you have selected the desired setting, press the I/O
button to turn the unit off.
Notes:
• If you want to adjust the date or time, press the SET button after setting the buzzer setting, follow steps for setting date and time (Refer to 2.2) press the I/O button to turn the unit off.
• If you forget to turn the unit off, it will automatically shut itself off after five minutes.
14
Correct Posture
Remove tight fitting clothing from your upper arm, and any thick clothing such as a sweater.
Do not place the cuff over thick clothes and do not roll up your sleeve if it is too tight.
Sit upright with your back straight.
Place your arm on a table so that the cuff will be at the same level as your heart.
The gap between the chair and the top of the table should be between 25cm to 30cm.
HEM-4011C-EN-RU.book 15 ページ 2012年1月6日 金曜日 午後2時50分
3. Using the Unit
3.Using the Unit
3.1
How to Sit Correctly When Taking a Measurement
You can take a measurement on either your left or right arm.
Note:
• Measurements should be taken in a quiet place and you should be in a relaxed, seated position. Make sure that the room is not too hot or cold.
• Do not take measurement after bathing, drinking alcohol, or exercising.
• Do not move or talk during measurement.
EN
15
Incorrect Posture
These situations tend to lead to higher blood pressure values due to strain or the arm cuff being lower than the heart.
If the arm cuff is at a lower position than your heart use cushions etc., to adjust the height of your arm.
Arched back (leaning
forwards)
Sitting cross-legged
Sitting on a sofa or at a low
table so that you lean forward
HEM-4011C-EN-RU.book 16 ページ 2012年1月6日 金曜日 午後2時50分
3.Using the Unit
Notes:
• Correct posture during measurement is necessary to get accurate results.
• You should also try to measure your blood pressure at the same time each day. (Within 1 hour after waking up is recommended.)
16
M1 Eco
HEM-4011C-EN-RU.book 17 ページ 2012年1月6日 金曜日 午後2時50分
3.2 Applying the Arm Cuff
Notes:
• Be sure to wrap the arm cuff correctly so that you get accurate results.
• Measurements can be taken in light clothing. However, please remove thick clothes, such as sweaters, before taking a reading.
• You can take a measurement on either your left or right arm. The blood pressure can differ between the right arm and the left arm and therefore also the measured blood pressure values can be different. Omron recommends to always use the same arm for measurement. If the values between the two arms differ substantially, please check with your doctor which arm to use for your measurement.
1. Connect the air tube of the arm cuff and the inflation bulb to
their respective air connectors.
3.Using the Unit
EN
17
HEM-4011C-EN-RU.book 18 ページ 2012年1月6日 金曜日 午後2時50分
3.Using the Unit
2. Put your left arm through the cuff
loop.
Note: If the cuff is not assembled,
pass the end of the cuff furthest from the tubing through the metal D-ring to form a loop. The smooth cloth should be on the inside of the cuff loop.
18
1) The air tube should run down the inside of your forearm and be in line with your middle finger.
2) The bottom of the cuff should be approximately 1 to 2 cm above your elbow.
3) Apply the cuff to your upper arm so that the coloured marker (blue arrow under tube) is centered on the middle of your inner arm and points down the inside of the arm.
1 to 2 cm
HEM-4011C-EN-RU.book 19 ページ 2012年1月6日 金曜日 午後2時50分
3. Position the arm correctly.
Taking measurements on the right arm
Apply the cuff so that the air tube is at the side of your elbow.
• Be careful not to rest your arm on the air tube, or otherwise restrict the flow of air to the cuff.
• The cuff should be 1 to 2 cm above the elbow.
4. When the cuff is positioned correctly, close the fabric fastener
FIRMLY.
3.Using the Unit
EN
19
HEM-4011C-EN-RU.book 20 ページ 2012年1月6日 金曜日 午後2時50分
3.Using the Unit
3.3 Taking a Reading
1. Preparation
1) Press the air release button to release any air in the arm cuff.
2) Press the I/O button to turn the unit on.
3) All items in the display will be displayed briefly.
The battery symbol will also be displayed for test purposes. This does not mean the batteries are empty.
4) The current date is displayed and the deflation symbol will flash.
5) Finally, the heartbeat symbol will appear and the current time
and 0 is displayed.
Note: If the deflation symbol does not disappear soon, press
the air release button to release any air in the arm cuff.
2. Pump the inflation bulb to inflate the arm cuff.
1) Inflate the cuff until it is 30 to 40
mmHg above your expected systolic blood pressure value.
ex) If your expected blood pressure is around 140mmHg, inflate the arm cuff to between 170 and 180 mmHg. Inflate the cuff rapidly so that the pressure is reached in about five seconds.
2) When the desired pressure has been achieved, release the
inflation bulb. Remain still and do not talk.
20
HEM-4011C-EN-RU.book 21 ページ 2012年1月6日 金曜日 午後2時50分
• The current time continues to be displayed while the cuff is being inflated.
Note: Do not inflate the arm cuff more than necessary.
3. Measurement starts.
Measurement starts automatically after you stop inflating the arm cuff. Decreasing numbers appear on the display and the heartbeat symbol flashes.
Remain still and do not talk during measurement.
During measurement, the buzzer (if set to “on”) will beep in rhythm with your heartbeat.
Note: To cancel a measurement, press the I/O button to turn
off the unit and press the air release button to release the air in the arm cuff.
The deflation symbol flashes on the display. Blood pressure and pulse rate values are displayed.
3.Using the Unit
4. Measurement ends.
Press the air release button to release the air in the arm cuff until the deflation symbol is no longer displayed.
EN
21
HEM-4011C-EN-RU.book 22 ページ 2012年1月6日 金曜日 午後2時50分
3.Using the Unit
5. Check the measurement results.
The unit automatically stores blood pressure and pulse rate into its memory, including date and time of measurement. Refer to “3.4 Using the Memory Function”.
The buzzer will sound (if switched on) when measurement is completed.
Note: Wait 2-3 minutes before taking another blood pressure
measurement. Waiting between readings allows the arteries to return to the condition prior to taking the blood pressure measurement.
Warning:
Self-diagnosis of measured results and treatment are dangerous. Please follow the instructions of your doctor.
Common Causes of Inaccurate Results
Arm cuff is too loose. Movement or talking
during measurement.
Systolic blood pressure
Diastolic blood pressure
Pulse
Leaning forwards.
Note: If the arm cuff is at a lower position than
your heart use cushions etc., to adjust the height of your arm.
6. Undo the fastener and remove the arm cuff.
7. Press the I/O button to turn the unit off.
Note: If you forget to turn the unit off, it will automatically shut
itself off after five minutes.
22
HEM-4011C-EN-RU.book 23 ページ 2012年1月6日 金曜日 午後2時50分
Important:
• If your systolic or diastolic pressure is outside the standard range, the heartbeat symbol will blink when the measurement result is displayed.
Recent research suggests that the following values can be used as a guide to high blood pressure for measurements taken at home.
Systolic Blood Pressure Above 135 mmHg
Diastolic Blood Pressure Above 85 mmHg
This criteria is for home blood pressure measurement. For professional office blood pressure measurement criteria, please refer to Chapter 9 “Some Useful Information about Blood Pressure”
• Your blood pressure monitor includes an irregular heartbeat feature. Irregular heartbeats can influence the results of the measurement. The irregular heartbeat algorithm automatically determines if the measurement is usable or needs to be repeated. If the measurement results are affected by irregular heartbeats but the result is valid, the result is shown together with the irregular heartbeat symbol ( ). If the irregular heartbeats cause the measurement to be invalid, no result is shown. If the irregular heartbeat symbol ( ) is shown after your have taken a measurement, repeat the measurement. If the irregular heartbeat symbol ( ) is shown frequently, please make your doctor aware of it.
3.Using the Unit
EN
23
Normal Heartbeat
HEM-4011C-EN-RU.book 24 ページ 2012年1月6日 金曜日 午後2時50分
3.Using the Unit
What is Irregular Heartbeat?
An irregular heartbeat is a heartbeat rhythm that varies by more than 25% from the average heartbeat rhythm detected while the unit is measuring the systolic and diastolic blood pressure.
If such an irregular rhythm is detected more than twice
Blood pressure
Blood pressure
Pulse
Irregular Heartbeat
Pulse
during measurement, the irregular heartbeat symbol ( ) appears on the symbol when the measurement results are displayed.
What is Arrhythmia?
A heartbeat is stimulated by electrical signals that cause the heart to contract.
Arrhythmia is a condition where the heartbeat rhythm is abnormal due to flaws in the bio-electrical system that drives the heartbeat. Typical symptoms are skipped heartbeats, premature contraction, an abnormally rapid (tachycardia) or slow (bradycardia) pulse. This can be caused by heart disease, aging, physical predisposition, stress, lack of sleep, fatigue etc. Arrhythmia can only be diagnosed by a doctor through a special examination.
Whether the appearance of the irregular heartbeat symbol ( ) in the results indicates arrhythmia or not can only be
determined by an examination and diagnosis by your doctor.
Warning:
If the irregular heartbeat symbol ( ) is shown frequently, please make your doctor aware of it. Conducting self­diagnosis and treatment based on measurement results is dangerous. Be sure to follow the instructions of your doctor.
Short
Long
24
HEM-4011C-EN-RU.book 25 ページ 2012年1月6日 金曜日 午後2時50分
3.4 Using the Memory Function
This unit has a memory capable of storing 42 sets of readings. Every time you complete the measurement, the unit automatically stores the blood pressure and pulse rate.
Note: When 42 sets of readings are stored in memory, the
oldest set will be deleted to store a new set.
1. Press the O/I button to turn the
power on.
2. When the heartbeat symbol appears
on the display, press the MEMORY button.
If there are three readings stored in memory that were taken within 10 minutes of the last reading, an average value for those reading is displayed. (If there are only two readings in memory for that period, the average will be based on two readings. If there is only one reading in memory for that period, that reading will be displayed as the average value.)
3.Using the Unit
EN
Note: If there are no measurements results stored in
memory, the screen to the right is displayed.
25
HEM-4011C-EN-RU.book 26 ページ 2012年1月6日 金曜日 午後2時50分
3.Using the Unit
3. Press the MEMORY button again,
while the average values screen is displayed.
The result from the most recent measurement is displayed.
The date and time of each measurement is displayed alternately with the measurement values.
Note: If your systolic or diastolic pressure is
outside the standard range, the heartbeat symbol will blink when the measurement result is displayed. Refer to 3.3, Important.
4. Press the MEMORY button
repeatedly to cycle through the previous measurement results.
Keep the button pressed down to cycle rapidly through the previous results.
If the irregular heartbeat symbol was displayed at the time of a measurement, this is displayed when that result is displayed.
5. Press the I/O button to turn the unit off.
If you forget to turn the unit off, it will automatically shut itself off after five minutes.