9. Some Useful Information about Blood Pressure .............47
2
HEM-7080IT-E.book 3 ページ 2011年11月14日 月曜日 午後1時53分
Before using the unit
Introduction
Thank you for purchasing the OMRON M10-IT Intellisense Upper Arm
Blood Pressure Monitor.
The OMRON M10-IT Intellisense is a fully automatic blood pressure
monitor, operating on the oscillometric principle. It measures your
blood pressure and pulse rate simply and quickly. For comfortable
controlled inflation without the need for pressure pre-setting or reinflation the device uses its advanced “Intellisense” technology.
The monitor stores measurement results for two people and features
morning and evening averages. The auto mode allows for three
consecutive measurements with timed intervals for an average
reading.
Omron’s blood pressure management software included with this unit
allows you to view, manage and print data about blood pressure
measured by the OMRON M10-IT Intellisense. Included software is in
English only.
Please read this instruction manual thoroughly before using
the unit. For specific information about your own blood
pressure, CONSULT YOUR DOCTOR.
Introduction
EN
3
HEM-7080IT-E.book 4 ページ 2011年11月14日 月曜日 午後1時53分
ImportantSafety Information
Important Safety Information
Consult your doctor during pregnancy, arrhythmia and arteriosclerosis.
Use the unit only to store the designated users measurement results.
Please read this section carefully before using the unit.
Caution:
(General Usage)
• Always consult your doctor. You should never stop or change the
dosage of medication prescribed by your doctor.
• Do not leave the unit unattended with infants or persons who cannot
express their consent.
• Do not use the unit for any purpose other than measuring blood
pressure.
• Use only the approved arm cuff for this unit. Use of other arm cuffs
may result in incorrect measurement results.
• Do not disassemble the unit or arm cuff.
• Make sure that the air tube is not wrapped around other parts of your
body when taking measurements at night. This could result in injury
when the air pressure in the air tube is increased.
• Do not wrap the arm cuff tightly when taking measurements at night.
This could result in injury.
• Do not inflate the arm cuff over 299 mmHg.
• Do not allow objects to come into contact with the USB connector on
the right of the unit when taking measurements.
(Optional AC Adapter Usage)
• Use only the original AC adapter designed for this unit. Use of
unsupported adapters may damage and/or may be hazardous to the
unit. Plug the AC adapter into the appropriate voltage outlet. Do not
overload power outlets or use extension cords.
• Do not use the AC adapter if the unit or the power cord is damaged.
Turn off the power and unplug the power cord immediately.
4
HEM-7080IT-E.book 5 ページ 2011年11月14日 月曜日 午後1時53分
Important Safety Information
• Never plug in or unplug the power cord from the electric outlet with
wet hands.
(Battery Usage)
• If battery fluid should get in your eyes, immediately rinse with plenty
of clean water. Consult a doctor immediately.
• If battery fluid should get on your skin or clothing, immediately rinse
with plenty of clean water.
• Use only four “AA” alkaline batteries with this unit. Do not use other
types of batteries.
• Do not insert the batteries with their polarities incorrectly aligned.
• Replace old batteries with new ones immediately. Replace all four
batteries at the same time.
• Remove the batteries if the unit will not be used for three months or
more. If the batteries have been removed for more than 30 seconds,
the date and time setting will need to be reset.
• Do not use new and used batteries together.
• Read and follow the “Important information regarding Electro
Magnetic Compatibility (EMC)” in the Technical Data Section.
General Safety Precautions
• Do not forcibly bend the arm cuff or bend the air tube excessively.
• When removing the air tube, pull on the air plug at the connection with
the main unit not the tube itself.
• Do not apply strong shocks and vibrations to or drop the main unit
and arm cuff.
• Do not inflate the arm cuff when it is not wrapped around your arm.
• Do not wash the arm cuff or immerse it in water.
• Do not take measurements after bathing, drinking alcohol, smoking,
exercising or eating.
• Read and follow the “Correct Disposal of This Product” in the
Technical Data Section when disposing of the device and any used
accessories or optional parts.
EN
5
F
E
G
B
A
C
D
H
J
I
HEM-7080IT-E.book 6 ページ 2011年11月14日 月曜日 午後1時53分
1. Overview
1. Overview
Main Unit
A. Display
B. SET button
C. Morning/Evening Average
(/) buttons
D. MEMORY button
E. USB connector jack (USB 1.1
compatible)
F. O/I START button
G. User ID selection switch
(A & B)
6
H. Battery compartment
I. Air jack
J. AC adapter jack (for AC
adapter)
K
ML
HEM-7080IT-E.book 7 ページ 2011年11月14日 月曜日 午後1時53分
Arm Cuff
K. Arm cuff
(Cuff: arm circumference 22-42 cm)
L. Air Plug
M. Air Tube
1. Overview
EN
7
N
V
Y
Q
O
S
U
P
R
X
Z
AB
AD
W
AA
AC
T
HEM-7080IT-E.book 8 ページ 2011年11月14日 月曜日 午後1時53分
1. Overview
Display
N. Systolic blood pressure
O. Diastolic blood pressure
P. Heartbeat symbol
1. (Flashes during measurement)
2. (If flashing after measurement
completed indicates blood
pressure out of recommended
range)
Q. Movement symbol
(Displayed if you move your
body during the measurement.)
R. Irregular heartbeat symbol
S. Deflation symbol
T. Auto mode symbol
(Displayed when Auto Mode is
set.)
U. Battery low symbol
V. User ID symbol (A or B)
(Indicates user A or user B as
selected by the user to take a
measurement or when using
the memory function.)
W. Pulse display
X. Early morning hypertension
symbol
(Displayed if the morning
average reading for a week is
above 135 for the systolic
blood pressure value and/or 85
for the diastolic blood pressure
value.)
Y. Morning average symbol
(Displayed when viewing
morning averages using the
memory function.)
Z. Evening average symbol
(Displayed when viewing
evening averages using the
memory function.)
AA.Average value symbol
(Displayed when viewing value
for the last three
measurements)
AB.Memory symbol
(Displayed when viewing
values stored in memory)
AC.Week display
AD.Date/Time display
direction of the arrow while
pressing the ribbed part of the
cover.
3.Install or replace four “AA” size
batteries so that the + (positive)
and - (negative) polarities
match the polarities indicated
on the battery compartment.
4.Put the battery cover back in place.
Slide the battery cover as indicated, until it clicks into place.
Note: The measurement values continue to be stored in memory even
after the batteries are replaced.
10
HEM-7080IT-E.book 11 ページ 2011年11月14日 月曜日 午後1時53分
2. Preparation
Battery Life & Replacement
If the battery low symbol () appears on the display, replace all four
batteries at the same time.
- When the battery low symbol () starts to blink, you will still be
able to use the unit for a short while. You should replace the
batteries with new ones ahead of time.
- When the symbol () remains lit, the batteries are exhausted. You
should replace the batteries with new ones at once.Turn the unit off
before replacing the batteries.
• Remove the batteries if the unit will not be used for three months or
more.
• If the batteries have been removed for more than 30 seconds, the
Date/Time setting will need to be reset. See Chapter 2.2 for details.
• Dispose of batteries according to applicable local regulations.
Four new
“AA” alkaline batteries will last for approximately
1400 measurements, when used to take six measurements a day.
Since the supplied batteries are for monitoring use only, they may have
a shorter life and not last for 1400 measurements.
EN
11
HEM-7080IT-E.book 12 ページ 2011年11月14日 月曜日 午後1時53分
2. Preparation
2.2Setting the Date and Time
Your blood pressure monitor automatically stores up to 84 individual
measurement values with the date and time.
To make use of the memory and average values functions:
• Set the unit to the correct date and time before taking a measurement
for the first time.
• If the batteries have been removed for more than 30 seconds, the
Date/Time setting will need to be reset.
1.When the O/I START button is
pressed to turn the unit on for
the first time after inserting
batteries, the year digits
(2007) will flash on the
display.
2.Press the MEMORY button to
advance the digits one at a
time.
Notes:
• The range for the year
setting is 2007 to 2030. If the
year reaches 2030, it will
return to 2007.
• If you hold down the
MEMORY button, the digits
will advance rapidly.
3.Press the SET button to confirm the setting when the desired
number appears on the display.
The year is set and the month digits flash on the display.
12
HEM-7080IT-E.book 13 ページ 2011年11月14日 月曜日 午後1時53分
4.Repeat steps 2 and 3 to set
the month.
The month is set and the day
digits flash on the display.
5.Repeat steps 2 and 3 to set
the day.
The day is set and the hour
digits flash on the display.
2. Preparation
6.Repeat steps 2 and 3 to set
the hour.
The hour is set and the
minutes digits flash on the
display.
EN
13
HEM-7080IT-E.book 14 ページ 2011年11月14日 月曜日 午後1時53分
2. Preparation
7.Repeat steps 2 and 3 to set
the minutes.
The minutes settings is set.
The unit automatically turns
itself off after the minute
setting has been set.
To adjust the date and time,
press the SET button while
the unit is in standby mode.
Notes on Adjusting the Date and Time Setting
If you need to adjust the date and time for some reason, or if the date
and time has been reset after replacing the batteries, adjust the date
and time between 10:00 and 18:59. This will avoid problems with
morning and evening weekly averages stored in memory.
To adjust the date or time at a later stage, start the unit in the off
position. Follow the steps below.
1.Press and hold the SET button.
The auto mode symbol and the current setting (on or
oFF) will appear on the display for the user ID selected.
If the year is not flashing, press the SET button once or twice
until the year is flashing.
The year flashes on the display.
14
HEM-7080IT-E.book 15 ページ 2011年11月14日 月曜日 午後1時53分
2. Preparation
2.Refer to Chapter 2.2.2 to start setting the date and time.
With this manual setting of date and time the unit needs to be
manually shut off. It will not automatically turn off after the
minutes have been set.
Notes:
• Please ensure that the date and time have been set correctly so that
the weekly averages will be displayed correctly. If the Date/Time
settings have been accidentally set incorrectly, press the SET button
to circle through the settings in order to adjust to their correct values.
• In the case where the time displays incorrectly after a long period of
non-use of the unit, reset the Date and Time (refer to Chapter 2.2 for
instructions on setting the clock values). Erase the recorded values
when reusing the unit (refer to Chapter 3.5 for directions on erasing
the recorded values).
15
EN
Correct
Posture
Remove tight fitting clothing from
your upper arm, and any thick
clothing such as a sweater.
Do not place the cuff over thick
clothes and do not roll up your sleeve
if it is too tight.
Sit upright
with your
back straight.
Place your arm on a table so
that the cuff will be at the same
level as your heart.
The gap between the
chair and the top of the
table should be between
25cm to 30cm.
HEM-7080IT-E.book 16 ページ 2011年11月14日 月曜日 午後1時53分
Operating instructions
3. Using the Unit
3. Using the Unit
3.1How to Sit Correctly When Taking a Measurement
You can take a measurement on either your left or right arm.
Notes:
• Measurements should be taken in a quiet place and you should be in
a relaxed, seated position. Make sure that the room is not too hot or
cold.
• Avoid eating, smoking, or exercising for at least 30 minutes before
taking a measurement.
• Do not move or talk during measurement.
16
Incorrect Posture
These situations could lead to higher blood pressure values due to
strain or the arm cuff being lower than the heart.
If the arm cuff is at a lower position than your heart use cushions
etc., to adjust the height of your arm.
• Arched back (leaning
forwards)
• Sitting cross-legged
• Sitting on a sofa or at a low
table so that you tend to lean
forward
HEM-7080IT-E.book 17 ページ 2011年11月14日 月曜日 午後1時53分
3. Using the Unit
Notes:
• Correct posture during measurement is necessary to get accurate
results.
• You should also try to measure your blood pressure at the same time
each day. (Within 1 hour after waking up is recommended.)
EN
17
Air plug
Air jack
HEM-7080IT-E.book 18 ページ 2011年11月14日 月曜日 午後1時53分
3. Using the Unit
3.2Applying the Arm Cuff
You can wrap the cuff either on your right or left arm.
• Remove tight-fitting clothing from your
upper arm.
• Do not place the cuff over thick clothes and
do not roll up your sleeve if it is too tight.
1.Insert the air plug into the air jack (on the left side of the
device).
The cuff must be fully deflated when it is inserted into the air
jack.
2.Sit in a chair with your feet flat on the floor and place your arm
on a table so that the cuff will be at the same level as your
heart.
18
Grip
Fabric fastener
Blue strip
Bottom edge 1 to 2 cm
Air tube
HEM-7080IT-E.book 19 ページ 2011年11月14日 月曜日 午後1時53分
3. Using the Unit
3.Hold the grip on the cuff securely
with your hand.
4.Turn the palm of your hand upward.
5.Apply the cuff to your upper arm so that the blue strip is centred
on the middle of your inner arm and points down the inside of
the arm. The air tube should run down the inside of your
forearm and be in line with your middle finger.
The bottom of the cuff should be approximately 1 to 2 cm
above the elbow.
EN
19
HEM-7080IT-E.book 20 ページ 2011年11月14日 月曜日 午後1時53分
3. Using the Unit
6.When the cuff is positioned correctly,
close the fabric fastener FIRMLY.
Notes:
• Make certain the cuff fits snugly around your arm.
• The cuff should make good contact with your skin. You should
be able to fit your index finger between the cuff and your arm
easily, so you can pull the cuff off and on.
• Make sure that there are no kinks in the air tubing.
Taking measurements on the right arm
Please note the following points when
applying the cuff to your right arm. Apply the
cuff so that the air tube is at the side of your
elbow.
Notes:
• Be careful not to rest your arm on the air
tube, or otherwise restrict the flow of air to
the cuff.
• Apply the arm cuff so that no part of the
cuff is positioned over the elbow joint. The
cuff should be 1 to 2 cm above the elbow.
• If used 2 times per day, the cuff will last for 3 years.
• The blood pressure can differ between the right arm and the left arm,
and therefore also the measured blood pressure values can be
different. Omron recommends to always use the same arm for
measurement. If the values between the two arms differ substantially,
please check with your physician which arm to use for your
measurement.
20
Press for at least 3 seconds
A and B display
goes out
HEM-7080IT-E.book 21 ページ 2011年11月14日 月曜日 午後1時53分
3. Using the Unit
3.3Taking a Reading
The unit is designed to take measurements and store the
measurement values in the memory for two people using user ID A and
user ID B.
The unit can also be used to take a single measurement for other
persons using the guest mode. The user ID symbol is not displayed
when using the guest mode. Measurements taken using the guest
mode are not stored in the memory.
Using the Guest Mode
1.Press and hold the O/I START button until the A and B user
symbols disappear from the display.
2.Release the O/I START button.
The cuff starts to inflate automatically.
Note: To stop the inflation press and release the O/I START
button. The unit will stop inflating, start deflation and turn
off.
3.Inflation stops and the measurement is started. When the
measurement is completed, your blood pressure and pulse rate
appear on the display.
EN
21
HEM-7080IT-E.book 22 ページ 2011年11月14日 月曜日 午後1時53分
3. Using the Unit
Selecting the User ID
Select your user ID. Always use the same user ID when taking a
measurement. The unit stores the measurement values in the memory
by the user ID selected. These measurement values are used to
calculate the morning averages and evening averages.
Slide the user ID selection switch up or down to select user A or B.
Using the User ID
When you select a user ID the unit can be set to take one
measurement using the single mode or three consecutive
measurements using the auto mode when you press the O/I START
button.
Note: The default value is set to the single mode to take one
measurement.
The auto mode takes three consecutive measurements. The unit will
inflate the cuff and take three measurements separated by a short
interval of time between each measurement.
22
HEM-7080IT-E.book 23 ページ 2011年11月14日 月曜日 午後1時53分
Using the Single Mode
1.Select your user ID (A or B).
2.Press the O/I START button.
All symbols appear on the display. The user ID symbol you
selected flashes on the display.
The cuff starts to inflate automatically. As the cuff inflates, the
unit automatically determines your ideal inflation level. This unit
detects the pulse during inflation. Do not move your arm and
remain still until the entire measurement process is completed.
3. Using the Unit
Note: To stop the inflation or measurement, press and release
the O/I START button. The unit will stop inflating, start
deflating and will turn off.
23
EN
HEM-7080IT-E.book 24 ページ 2011年11月14日 月曜日 午後1時53分
3. Using the Unit
3.Inflation stops automatically and the measurement is started.
As the cuff deflates, decreasing numbers appear on the
display. The heartbeat symbol () flashes at every heartbeat.
4.When the measurement is complete, the arm
cuff completely deflates. Your blood pressure
and pulse rate are displayed.
5.Press the O/I START button to turn
the unit off.
24
Note: If you forget to turn the
monitor off, it will
automatically turn off after
five minutes.
Notes:
• Self-diagnosis of measured results and treatment are
dangerous. Please follow the instructions of your doctor.
• Wait 2-3 minutes before taking another blood pressure
measurement. Waiting between readings allows the arteries
to return to the condition prior to taking the blood pressure
measurement.
HEM-7080IT-E.book 25 ページ 2011年11月14日 月曜日 午後1時53分
Selecting the Auto Mode
Auto Mode may be selected for either A or B user.
1.Make sure the unit is powered off.
2.Select your user ID (A or B).
3.Press and hold the SET button.
The auto mode symbol () and the oFF setting appears on
the display.
3. Using the Unit
4.Press the MEMORY button ()
to select on.
Note: Press the MEMORY
button to alternate
between on and oFF.
EN
25
HEM-7080IT-E.book 26 ページ 2011年11月14日 月曜日 午後1時53分
3. Using the Unit
5.Press the SET button to confirm the change.
The number of seconds between each measurement is
displayed.
Note: The default value between each measurement is
60 seconds.
6.Press the MEMORY button () to change the interval of time
between each measurement.
The interval can be set to 15 seconds, 30 seconds, 60 seconds
or 120 seconds. Continue to press the MEMORY button until
the desired interval appears on the display.
7.Press the SET button to confirm the
setting.
The year flashes on the display.
8.Press the O/I START button to turn
the unit off.
26
HEM-7080IT-E.book 27 ページ 2011年11月14日 月曜日 午後1時53分
Using the Auto Mode
1.Select your user ID (A or B).
2.Press the O/I START button.
All symbols appear on the display. The user ID Symbol you
selected flashes on the display. The cuff starts to inflate
automatically.
Note: To stop the inflation or measurement, press the O/I
START button. The unit will stop inflating, start deflating
and will turn off.
3.Inflation stops automatically and the measurement is started.
As the cuff deflates, decreasing numbers appear on the
display. The heartbeat symbol () flashes at every heartbeat.
When the measurement is complete, the arm cuff completely
deflates.
Note: Do not move your arm and remain still until the entire
measurement process is completed.
3. Using the Unit
4.Check the progress of the consecutive measurements.
The first measurement is complete, waiting to start the next
measurement. The interval of time set between each
measurement will appear on the display.
EN
27
HEM-7080IT-E.book 28 ページ 2011年11月14日 月曜日 午後1時53分
3. Using the Unit
The second measurement starts automatically.
The second measurement is complete, waiting to start the next
measurement.
The third measurement starts automatically.
The consecutive measurements are complete. The average for
the three measurements appears on the display.
The three individual measurement results are not displayed
while Auto Mode is taking the measurements. However, the
results for each individual measurement can be displayed after
all the measurements have been completed.
Note: During the Auto Mode the unit will repeat the
measurement in case of an error. After 3 errors the unit
will stop taking further measurements. If only two
measurements are successful the average of these two
measurements will be displayed. If only one
measurement is successful, this result will be displayed.
5.Press the SET button to display the blood pressure and pulse
rate for the first measurement. Press the SET button again to
display the next measurement.
6.Press the O/I START button to turn the unit off.
28
HEM-7080IT-E.book 29 ページ 2011年11月14日 月曜日 午後1時53分
Important:
• If your systolic or diastolic pressure is
outside the standard range, the heartbeat
symbol will blink when the measurement
result is displayed.
Recent research suggests that the following
values can be used as a guide to high blood
pressure for measurements taken at home.
Systolic Blood PressureAbove 135 mmHg
Diastolic Blood PressureAbove 85 mmHg
This criteria is for home blood pressure measurement.
For professional office blood pressure measurement criteria,
please refer to Chapter 9.
• Your blood pressure monitor includes an
irregular heartbeat feature. Irregular
heartbeats can influence the results of the
measurement. The irregular heartbeat
algorithm automatically determines if the
measurement is usable or needs to be
repeated. If the measurement results are
affected by irregular heartbeats but the result
is valid, the result is shown together with the irregular
heartbeat icon. If the irregular heartbeats cause the
measurement to be invalid, no result is shown. If the irregular
heartbeat symbol ( ) is shown after you have taken a
measurement, repeat the measurement. If the irregular
heartbeat icon is shown frequently, please make your doctor
aware of it.
3. Using the Unit
EN
29
Normal Heartbeat
Irregular Heartbeat
Pulse
Blood pressure
Blood pressure
Short
Long
Pulse
HEM-7080IT-E.book 30 ページ 2011年11月14日 月曜日 午後1時53分
3. Using the Unit
What is Irregular Heartbeat?
An irregular heartbeat is a
heartbeat rhythm that varies
by more than 25% from the
average heartbeat rhythm
detected while the unit is
measuring the systolic and
diastolic blood pressure.
If such an irregular rhythm is
detected more than twice
during measurement, the irregular heartbeat symbol ( )
appears on the symbol when the measurement results are
displayed.
What is Arrhythmia?
A heartbeat is stimulated by electrical signals that cause the
heart to contract.
Arrhythmia is a condition where the heartbeat rhythm is
abnormal due to flaws in the bio-electrical system that drives
the heartbeat. Typical symptoms are skipped heartbeats,
premature contraction, an abnormally rapid (tachycardia) or
slow (bradycardia) pulse. This can be caused by heart disease,
aging, physical predisposition, stress, lack of sleep, fatigue etc.
Arrhythmia can only be diagnosed by a doctor through a
special examination.
Whether the appearance of the irregular heartbeat symbol
() in the results indicates arrhythmia or not can only be
determined by an examination and diagnosis by your doctor.
If the irregular heartbeat symbol ( ) is shown frequently,
please make your doctor aware of it. Conducting self-diagnosis
and treatment based on measurement results is dangerous. Be
sure to follow the instructions of your doctor.
30
Loading...
+ 320 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.