Olympus WM-P1 User manual

INSTRUCTIONS
MOBILE WORKSTATION
WM-DP1
Contents
Important Information — Please Read Before Use. . . . . . . . . . . . . . . 2
Intended use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Signal words . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Chapter 1 Standard Set and Accessories . . . . . . . . . . . . . . . . . . . 3
1.1 Optional accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Chapter 2 Warnings and Cautions . . . . . . . . . . . . . . . . . . . . . . . . 5
Chapter 3 Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Chapter 4 Preparation for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
4.1 General information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
4.2 Shelf positions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
4.3 Remote switch. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
4.4 Equipment installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
4.5 Cable hanger and blanking caps. . . . . . . . . . . . . . . . . . . . . . 16
4.6 Water container holder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.7 Rear panels. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.8 Fitting the MAJ-181 restrictor . . . . . . . . . . . . . . . . . . . . . . . . 18
Chapter 5 Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Chapter 6 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
6.1 Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
6.2 Manoeuvring and Transportation . . . . . . . . . . . . . . . . . . . . . 24
Chapter 7 Care, Storage and Maintenance . . . . . . . . . . . . . . . . . . 26
7.1 Care. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
7.2 Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
7.3 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Chapter 8 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
8.1 WM-DP1 mobile workstation. . . . . . . . . . . . . . . . . . . . . . . . . 28
8.2 Separation transformer (WM-T1) . . . . . . . . . . . . . . . . . . . . . 28
Chapter 9 Spares . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Chapter 10 Technical Description . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Chapter 11 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Chapter 12 End of Life . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
1
WM-DP1 MOBILE WORKSTATION
Important Information — Please Read Before Use
Important Information — Please Read Before Use
Intended use
The Olympus WM-DP1 mobile workstation is intended for use in medical facilities under the direction of a trained physician, and has been designed to be used with a range of Olympus equipment to facilitate GI endoscopy, endoscopic ultrasound, respiratory and surgical endoscopic procedures. Do not use the equipment for any purpose other than its intended application.
The safety and performance of an endoscopic system depends not only on the endoscope but also on the ancillary equipment used with it. Ensure any ancillary equipment is compatible with the endoscope and other equipment used.
Instruction manual
Please read this manual carefully before attempting to use the Olympus WM-DP1 mobile workstation as it contains important information on the proper care, handling and use of the equipment. In addition, read the manuals of any other units (light source, endoscope, etc.) that form part of the system.
If you have any questions or comments regarding the information in this manual, please contact Olympus. These instructions should be retained for reference during the life of the product.
Signal words
The following signal words are used throughout this manual:
WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices or potential equipment damage.
NOTE
Indicates additional helpful information.
MOBILE WORKSTATION WM-DP1
2
Standard Set and Accessories
Chapter 1 Standard Set and Accessories
Please refer to the standard set below and check that all items are present. Contact Olympus if any parts are damaged or missing. Workstations that are fitted with antistatic castors can be identified by the following label affixed to the rear of each shelf.
1
4
5
2
3
6
9
8
INSTRUCTIONS
10
7
Contents Qty
1 WM-DP1 mobile workstation (includes separation transformer) 1 2 Rear panel - narrow 1 3 Rear panel - wide 1 4 Blister pack 1
Blanking cap (cable port, 2pcs) Blanking cap (monitor port, 2pcs) Blanking cap (scope pole, 2pcs) L-wrench ball driver 6mm A/F, 1 piece Cable tie, 6 pcs
5 Blister pack 1
Blanking cap (cable port, 2pcs)
Blanking cap (monitor port, 1 piece) 6 Instruction manual 1 7 Cable hanger 2 8 Universal IEC cable 2.5m 4
Universal IEC cable 1.5m 4 9 IEC adaptor cable (USA only) 2
10 MAJ-181 restrictor assembly 2
3
MOBILE WORKSTATION WM-DP1
1.1 Optional accessories
The following items are available as optional accessories. Each accessory is supplied complete with the necessary fixings, tools and instructions to enable assembly to the mobile workstation.
Description Ref No. GI SP EUS
14” CRT Platform WM-NP1 MAJ-1620 
20” CRT Platform WM-NP1 MAJ-1621 
Cable retaining hook MAJ-195 
cylinder bracket kit large MAJ-188
CO
2
cylinder bracket kit small MAJ-1614
CO
2
Drawer pack narrow deep MAJ-194 
Drawer pack narrow standard MAJ-193 
Drawer pack wide deep MAJ-1612 
Drawer pack wide standard MAJ-1611 
Standard Set and Accessories
Recommended use
ECS cable kit 10m MAJ-186
ECS cable kit 15m MAJ-1616
Equipment support rail MAJ-190 
EU-C60/2000 shelf kit MAJ-1627
EUS Arm Mount Kit WM-P1 MAJ-1626
IV pole kit MAJ-187
Keyboard arm side mounted MAJ-180  –
Keyboard tray sliding MAJ-179 
EU-M30 Keyboard Tray MAJ-1625
LCD handle MAJ-1610 
LCD monitor arm MAJ-181 
Nurse’s control panel arm (long) MAJ-1628 
Nurse’s control panel arm (short) MAJ-1624
Scope pole kit MAJ-183 
Shelf kit narrow MAJ-198 
Shelf kit wide MAJ-199 
Universal display arm CRT MAJ-1613
Universal stowage container MAJ-185 
Videoscope cable holder MAJ-1630 
compatible – incompatible
CAUTION
Do not attempt to install any other accessories on the WM-DP1.
MOBILE WORKSTATION WM-DP1
4
Warnings and Cautions
Chapter 2 Warnings and Cautions
WARNING
• The WM-DP1 mobile workstation should only be used in a medical facility under the direction of a trained physician.
• Explosion hazard - never install or use the WM-DP1 mobile workstation within the zone of risk of flammable gases. If the workstation is fitted with antistatic castors (see Chapter 8), none of the medical electrical equipment with which the workstation is intended to be used is ‘AP’ or ‘APG’ rated, therefore never install or use the WM-P1 series mobile workstation within the zone of risk of flammable gases.
• To prevent patient shock, the combination of equipment used on the WM-DP1 mobile workstation should never be applied directly to the heart.
• If the fuse holder cap on the transformer is not fitted correctly, there is a possibility that the fuse holder will overheat. When replacing the fuse, ensure the fuse holder cap (Figure 4.3) is tightened using a 6mm to 8mm flat bladed screwdriver. The recommended tightening torque is
0.5Nm.
• Should the equipment become soiled with blood or other potentially infectious materials, first wipe off all gross debris and then decontaminate the equipment using a surface disinfectant, otherwise, blood, mucus and other potentially infectious material from the patient could pose an infection risk. Confirm that the equipment is completely dry before use.
• Use personal protective equipment to guard against dangerous chemicals and potentially infectious material, otherwise, blood, mucus and other potentially infectious material from the patient could pose an infection risk. During cleaning, wear appropriate protective equipment, such as eye wear, face mask, moisture-resistant clothing and chemical­resistant gloves that fit properly and are long enough so that your skin is not exposed.
• To prevent user or patient cross-contamination, do not hang contaminated endoscopes on the scope hanger.
• The WM-DP1 should be used in accordance with the following. Otherwise, the workstation will be ‘top heavy’ and may lose stability and topple, causing equipment damage and/or personal injury.
- Two MAJ-181 restrictors are supplied which must be fitted to MAJ-181 LCD monitor arm(s) if used (refer to section 4.8).
- LCD monitor arms and monitors should be installed after all other equipment is installed on an appropriate place on the workstation.
- LCD monitor arms should not be installed or used to the rear of the workstation, see Figure 2.1.
- Confirm that the workstation does not lose stability when the monitor arm is extended. If the workstation looks unstable, relocate heavy equipment from the upper shelf to the bottom shelf.
5
MOBILE WORKSTATION WM-DP1
Warnings and Cautions
MAJ-181
(restrictor fitted)
MAJ-1624
or
MAJ-1628
Do not install or operate
LCD monitor arms in this zone
• Thoroughly review the manuals of all other equipment which will be used. The manuals contain essential information on using this equipment safely and effectively.
CAUTION
• Do not exceed the load capacity detailed in Chapter 8.
• Do not lean or stand on any part of the mobile workstation or accessory as failure of the product may result.
• When assembling equipment onto the mobile workstation, the installation should meet the requirements of EN IEC 60601-1-1:2001, the safety standard for medical electrical systems.
• When configuring a system, the user should ensure that there is sufficient spare equipment available that in the event of a fault in any part of the system, the endoscopy procedure may be completed or terminated without endangering the patient.
• Before commissioning an installation including a separation transformer, ensure its rating will not be exceeded by the equipment connected to it.
• When installing the mobile workstation in the procedure or operating room, ensure that the mains cable has sufficient slack that it cannot be pulled taught as the mobile workstation is moved around the working area. If necessary, relocate the mains connector to a nearer supply socket.
Figure 2.1
• Section 6.2 Manoeuvring and Transportation gives detailed instructions and precautions which must be read and understood before attempting to move the mobile workstation.
MOBILE WORKSTATION WM-DP1
• Spray type medical agents, such as lubricants, anaesthetic or alcohol, should be used away from the mobile workstation.
6
Warnings and Cautions
• Keep open containers of liquid away from electrical equipment used on the mobile workstation to prevent operator electrical shock and equipment damage. Do not use the system if liquid has entered the equipment.
• To minimise risk of injury, it is recommended that heavy items are placed on the mobile workstation by two people working together. This is particularly important when placing a monitor on the optional monitor platform (if fitted). Ensure electrical items used on the mobile workstation are positioned centrally on the shelves.
• Never place any items on top of a monitor (if fitted) as these may topple when the mobile workstation is moved.
• Where fitted, the electrical resistance of the conductive antistatic castors should be tested at regular intervals in accordance with ISO 2878 to assess antistatic effectiveness.
• Where antistatic castors are fitted, avoid contact of the castor wheels with oil and avoid a build up of floor cleaning wax and polish as these will impair the antistatic effectiveness of the castors.
• Store and use the WM-DP1 mobile workstation within the environmental conditions described in Chapter 8; failure to do so may lead to equipment malfunction or failure.
• There are no operator-serviceable components on the WM-DP1 mobile workstation. Refer servicing to qualified service personnel.
• To minimise the risk of injury and/or damage, do not fit accessories other than those described in Section 1.1 to the WM-DP1 mobile workstation.
7
MOBILE WORKSTATION WM-DP1
Chapter 3 Symbols
The following symbols are used on the separation transformer
Refer to instructions Potential equalisation terminal
Input Output
Symbols
Not exceeding
Standby position Fuse (output)
Output ON position Remote switch
Non-inherently short circuit proof separation transformer
Unit complies with relevant European directives
(MAJ-175 only)
View of the separation transformer rear panel
Ta=40°C
Class B
Maximum ambient temperature
Insulation temperature rating of transformer according to IEC 60085:2004
Disposal conditions - see End of Life (Chapter 12) (MAJ-175 only)
NOTE
MAJ-175 (220-240V) is shown. Other versions are similar.
MOBILE WORKSTATION WM-DP1
8
Preparation for Use
Chapter 4 Preparation for Use
NOTE
• Multiple fixing positions are provided enabling the shelves to be raised or lowered to accommodate varying equipment configurations (see 4.2).
• The mobile workstation is designed for use with Olympus OEV LCD monitors and other monitors that meet the VESA standard. Contact Olympus regarding compatibility of other monitors.
4.1 General information
Installation of endoscopic video systems requires an understanding of EN IEC 60601-1-1:2001, the international collateral standard for the safety of medical electrical systems.
This standard requires systems to meet the leakage current requirements of the general medical electrical safety standard, EN IEC 60601-1, which allows a maximum earth leakage current of
0.5mA in normal condition.
A separation transformer is required under certain circumstances, as combined earth leakage currents may result in excessive enclosure leakage current under single fault conditions, such as the interruption of a common protective earth path.
In order to meet EN IEC 60601-1-1:2001, a separation transformer will be required in a system under one of the following conditions:
(i) Ancillary equipment that does not meet the leakage current requirements of
EN IEC 60601-1 is used (or could be used) within the ‘patient environment’ (as defined in EN IEC 60601-1-1:2001, that is within a 1.5m radius of, or 2.5m above, the patient bed). The separation transformer must be used to supply the ancillary equipment, but may also be used to supply medical electrical equipment, as long as the maximum transformer rating is not exceeded.
(ii) The total earth leakage current from a combination of medical equipment is above
0.5mA. The separation transformer must be used to supply sufficient of the medical electrical equipment to reduce the total system earth leakage current to below the above limit, as long as the maximum transformer rating is not exceeded. High intensity light sources use arc lamps, which require a high start-up current. It is not recommended for these to be supplied from the isolation transformer, otherwise its maximum rating may be exceeded. The transformer may also be used to supply ancillary equipment forming part of the system as long as the maximum transformer rating is not exceeded.
9
MOBILE WORKSTATION WM-DP1
Preparation for Use
Some system installations, for instance those using two mobile workstations, may require two separation transformers to be fitted.
NOTE
• The total impedance of the protective earth path for a system may be up
Ω
to 0.4
or higher if the conditions of clause 18 g) of EN IEC 60601-1 are
satisfied.
• The power outlets should only be used for powering equipment which forms part of the endoscopy imaging system.
• All equipment forming part of the system should be powered from the transformer, unless appropriate isolation is employed as described in Annex BBB of EN IEC 60601-1-1:2001. This particularly applies where ancillary equipment as described in (i) forming part of the system is remotely situated.
CAUTION
• Failure to observe either of the above mentioned points could result in the isolation of the entire system being compromised, defeating the protection provided by the transformer in a single fault condition.
Additional fixings are provided (Figure 4.1) to facilitate shelf adjustment or the fitting of the optional drawer pack(s).
MOBILE WORKSTATION WM-DP1
10
Loading...
+ 25 hidden pages