Olympus BX61 User Manual [ru]

Page 1
INSTRUCTIONS
BX61
MOTORIZED SYSTEM
MICROSCOPES
This instruction manual is for the Olympus Motorized System Microscopes Models BX61 and BX62. To ensure the safety, obtain optimum performance and to familiarize yourself fully with the use of this microscope, we recommend that you study this manual thoroughly before operating the micro­scope. Retain this instruction manual in an easily accessible place near the work desk for future reference.
A X 7 6 1 1
Page 2
Page 3
CONTENTS
Correct assembly and adjustments are indispensable to allow the microscope manifest its full performance. If you are going to assemble the microscope by yourself, please read chapter 7, “ASSEMBLY” (pages. 29 to 32).
BX61
IMPORTANT — Be sure to read this section for safe use of the equipment. —
1 NOMENCLATURE
2
TRANSMITTED LIGHT BRIGHTFIELD OBSERVATION PROCEDURE
3 USING THE CONTROLS
3-1 Base ................................................................................................................................................................................................................. 8-10
1 Voltage Indication 2 Transmitted/Reflected Light Switch 3 Light Preset Switch 4 Using the Filters
3-2 Focusing Block............................................................................................................................................................................... 11- 12
1 Replacing the Focus Adjustment Knob 2 F/C button 3 Stage UP/DOWN Buttons 4 Stage Escape Button
3-3 Stage ............................................................................................................................................................................................................ 13-15
1 Placing the Specimen 2 Adjusting the X/Y-Axis Knob Tension 3 Rotating the Stage 4 Adjusting the Stage Height
3-4 Observation Tube ....................................................................................................................................................................... 16-18
1 Adjusting the Interpupillar Distance 2 Adjusting the Diopter
3 Using the Eye Shades 4 Using the Eyepiece Micrometer Disk
5 Selecting the Light Path of the Trinocular Tube 6 Adjusting the Tilt
3-5 Condenser .......................................................................................................................................................................................... 19-20
1 Centering the Condenser 2 Compatibility of Objectives and Condensers
3-6 Immersion Objectives................................................................................................................................................................... 21
3-7 Objectives with Correction Collar ............................................................................................................................... 21
1-3
4-5
6-7
8-21
4 TROUBLESHOOTING GUIDE
5 SPECIFICATIONS
6 OPTICAL CHARACTERISTICS <<UIS2 (UIS) Series>>
ASSEMBLY
7
PROPER SELECTION OF THE POWER SUPPLY CORD...................................................................
See this section for the replacement of the light bulb. —
22-24
25-26
27-28
29-32
33-34
Page 4
This device complies with the requirements of directive 98/79/EC concerning in vitro diag­nostic medical devices. CE marking means the conformity to the directive.
NOTE: This equipment has been tested and found to comply with the limits for a Class A digital device,
pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
FCC WARNING: Changes or modifications not expressly approved by the party responsible for compliance
could void the user’s authority to operate the equipment.
Page 5
IMPORTANT
Each of these microscopes employs a UIS2/UIS (Universal Infinity System) optical design, and should be used only with modules designed for the BX2 series (which belong to the Olympus BX series microscopes) and objectives/eyepieces for the UIS series. For the applicable module, please consult Olympus or the catalogues. Less than optimum performance may result if inappropriate accessories are used.
Configuration of Instruction Manuals
Since these microscopes are expandable to a variety of systems, separate instruction manuals are provided so that the user has to read only the manuals according to the user’s own system.
Manual Name Main Contents
BX61 Transmitted light brightfield observation procedure
BX-UCB/U-HSTR2
BX2 Software for PC (CD-ROM) Methods of PC Control of Functions
BX-RFAA
BX-RLAA Reflected light brightfield/darkfield observation
U-UCD8A Motorized universal condenser
U-FWT/FWR/FWO Motorized filter wheels
Functions of the Control Box (incorporating the power supply) and Hand Switch
Reflected light fluorescence observation procedure
BX61
SAFETY PRECAUTIONS
1. After the equipment has been used in an observation of a specimen that is accompanied with a potential of infection, clean the parts coming in contact with the specimen to prevent infection.
· Moving this product is accompanied with the risk of dropping the
specimen. Be sure to remove the specimen before moving this product.
· In case the specimen is damaged by erroneous operation, promptly
take the infection prevention measures.
· The product becomes unstable if its height is increased by an accessory
mounted on it. In this case, take anti-toppling measures to prevent the specimen from being dropped when the product topples down.
2. Install the microscope on a sturdy, level table or bench so as not to block the air vents on the underside of the base. Do not place the microscope on a flexible surface, as this could result in blocking the air vents and cause overheating or a fire.
3. If a foreign object is caught during motorized focusing operation, there will be an error in the focusing block and the motorized focusing opera­tion will be suspended.
Recovery procedure
· If there is no error in motorized operation, the caught object can be removed by turning the focusing knob.
· If there is an error in motorized operation, the focusing knob becomes inoperable. Disassemble the relevant modules to remove the caught object. Replace the relevant modules afterward.
· Turn off the power and then on again. The system will restart unless there is a malfunction in the motor.
4. To activate temporary stop during focusing operation, turn the focusing knob (or dial) on the microscope frame (in either direction) or press any of the FOCUS control buttons ( , , F/C and ESC) (except during PC communication). When the main switch of the BX-UCB control box is set to “ I ” (ON), the focusing operation starts automatically as part of initialization (this opera­tion consists of temporary lowering and then returning to the original position of the stage). If any of the operations mentioned above is per­formed, an emergency stop will also occur. In this case, set the main switch to “ ” (OFF) and then “ I ” (ON) again.
1
Page 6
Fig. 1
²
@
5. When lowering the stage by pressing the stage DOWN button @, be careful not to have your hand caught between the bottom ² of the con­denser and the base.
6. The surfaces of the lamp housing on the rear of the microscope will become extremely hot during operation. When installing the microscope, make sure to allow ample free space (10 cm or more) around and in particular above the lamp housing.
7. When installing the microscope, route the power cord away from the lamp housing. Should the power cord come in contact with the hot lamp housing, the power cord could melt and cause electric shock.
8. To avoid potential shock hazards and burns when replacing the light bulb, set the main switch ³ to “ ” (OFF) then disconnect the power cord from the wall outlet in advance. Whenever you replace the bulb during use or right after use, allow the lamp housing | and bulb to cool before touching. (Fig. 2)
|
³
Fig. 2
Designated haiogen bulbs:
# The microscope also incorporate a fuse (this should be
replaced by the manufacturer or an authorized agent).
9. Always use the power cord provided by Olympus. If no power cord is provided, please select the proper power cord by referring to the section “PROPER SELECTION OF THE POWER SUPPLY CORD” at the end of this instruction manual. If the proper power cord is not used, product safety performance cannot be warranted.
10.
Always ensure that the grounding terminal of the microscope and that of the wall outlet are properly connected. If the equipment is not grounded, Olympus can no longer warrant the electrical safety performance of the equipment.
11.
Never insert metallic objects into the air vents of the microscope frame as this could result in electrical shock, personal injury and equipment damage. When using the U-FWR motorized filter wheel for reflected light observa-
12. tion, remove the push ring driver from the position where it is attached by magnetism in advance to prevent it from dropping. If the driver drops in a certain place, an electric shock and/or burns may result.
12V100WHAL (PHILIPS 7724) 12V50WHAL-L (LIFE JC)
Safety Symbols
The following symbols are found on the microscope. Study the meaning of the symbols and always use the equip­ment in the safest possible manner.
Symbol Explanation
2
Indicates that the surface becomes hot, and should not be touched with bare hands.
Before use, carefully read the instruction manual. Improper use could result in personal injury to the user and/or damage to the equipment.
l
Indicates that the main switch is ON.
Indicates that the main switch is OFF.
Warnings
Warning engraving/stickers are placed at parts where special precaution is required when handling and using the microscope. Always heed the warnings.
Warning engraving position
Lamp housing (Warning against high temperature)
Page 7
1 Getting Ready
Fig. 3
BX61
1. A microscope is a precision instrument. Handle it with care and avoid subjecting it to sudden or severe impact.
2. Do not use the microscope where it is subjected to direct sunlight, high temperature and humidity, dust or vibrations. (For the operating condi­tions, refer to chapter 5, “SPECIFICATIONS”.)
3. When moving the microscope, remove the modules that may drop including the specimen from the microscope in advance, hold it by the hand hook on the rear and the base as shown on the left (Fig. 3) and carry carefully (Weight: approx. 16 kg).
#Before transporting the microscope, be sure to remove the speci-
men, eyepieces and color glass filter so that they will not drop. Also remove other modules attached to the microscope because they increase the system weight.
# Damage to the microscope will occur if you grasp it by the stage,
focus adjustment knob or binocular section of the observation tube.
4. The BX61 series can be used with up to two intermediate attachments including a U-CA magnification changer, U-EPA2 eyepoint adjuster, etc. For restrictions when using two intermediate attachments, make sure to read the instruction manuals provided with the respective intermediate attachments.
2 Maintenance and Storage
1. To clean the lenses and other glass components, simply blow dirty away using a commercially available blower and wipe gently using a piece of cleaning paper (or clean gauze). If a lens is stained with fingerprints or oil smudges, wipe it gauze slightly moistened with commercially available absolute alcohol.
!Since the absolute alcohol is highly flammable, it must be handled carefully.
Be sure to keep it away from open flames or potential sources of electrical sparks –– for example, electrical equipment that is being switched on or off, which could cause ignition of a fire. Also remember to always use it only in a well-ventilated room.
2. Do not attempt to use organic solvents to clean the microscope components other than the glass components. To clean them, use a lint-free, soft cloth slightly moistened with a diluted neutral detergent.
3. Do not disassemble any part of the microscope as this could result in malfunction or reduced performance.
4. When not using the microscope, set the main switch to “ ” (OFF), wait until the lamp housing has cooled down, and cover the microscope with a dust cover before storage.
5. When disposing of the microscope, check the regulations and rules of your local government and be sure to observe them.
3 Caution
If the microscope is used in a manner not specified by this manual, the safety of the user may be imperiled. In addition, the equipment may also be damaged. Always use the equipment as outlined in this instruction manual.
The warning, caution and other notes in this manual use the following symbols. !: Indicates that failure to follow the instructions in the warning could result in bodily harm to the
user and/or damage to equipment (including objects in the vicinity of the equipment). # : Indicates that failure to follow the instructions could result in damage to equipment. } : Indicates commentary (for ease of operation and maintenance).
Intended use
4
This instrument has been designed to be used to observe magnified images of specimens in routine and research applications. Do not use this instrument for any purpose other than its intended use.
3
Page 8
1
}If you have not yet assembled the microscope, read chapter 7, “ASSEMBLY” (pages 29 to 32).
Transmitted light specification models
Interpupillary distance adjustment scale (Page 16)
NOMENCLATURE
Light path selector knob (Page 17)
Condenser height adjustment screw (Page 19)
Allen screwdriver
(accommodation position)
Diopter adjustment ring (Page 16)
Specimen holder (Page 13)
Aperture iris diaphragm ring
(Page 20)
Filter mount (Page 9)
Field iris diaphragm ring
(Page 19)
Brightness adjustment buttons
(Page 8)
Lamp voltage indicator
Transmitted/reflected light switch (Page 8)
PHOTO indicator
(Page 8)
Y-axis knob (Page 14)
X-axis knob (Page 14)
Optional filter knob
4
Filter knobs (Page 9)
ND6, ND25 and LBD from the front to the rear
Page 9
Base
Focus adjustment knob (Page 11)
BX61
Stage escape button
(Page 12)
F/C button (Page 11)
Reflected light specification models
Motorized reflected light illuminator*
· Motorized fluorescence illuminator (BX-RFAA)
· Motorized reflected light illuminator (BX-RLAA)
UCB connector
Light Preset switch (Page 8)
Lamp ON-OFF switch
* For the reflected illuminators and reflected lamp housings, please also
refer to their instruction manuals.
· 100 W mercury burner housing
· 100 W mercury apo-burner housing
· 75 W xenon lamp housing
· 100 W halogen lamp housing
· 50 W metal hydride lamp housing
Stage UP button (Page 12)
Stage DOWN button (Page 12)
Reflected lamp housing*
(U-LH100HG)
(U-LH100HGAPO)
(U-LH75XEAPO)
(U-LH100-3)
(U-LH50MH)
5
Page 10
TRANSMITTED LIGHT BRIGHTFIELD OBSERVATION PROCEDURE
Set the main switch to “ I ” (ON).
Select the transmitted light and adjust the brightness.
Engage the LBD filter in light path. ƒLBD filter knob
Select the light path (when a trinocular tube is used).
Place the specimen on the stage.
Engage the 10X objective in the light path.
Bring the specimen in focus.
(Controls Used)
@Main switch “ I ” (ON) ²Lamp ON-OFF switch ON
³Transmitted/reflected light switch (P. 8) |Brightness adjustment buttons (P. 8)
…Light path selector knob (P. 17)
†Specimen holder (P. 13) ‡X/Y-axis knobs (P. 14)
ŠRevolving nosepiece button
‰Stage UP/DOWN button, focus
adjustment knob
(Page)
(P. 12)
Adjust the interpupillary distance. Adjust the diopter. Adjust the light axis.
Adjust the aperture iris and field iris diaphragms.
Engage the objective to be used in the light path and bring the specimen in focus.
Engage the required filters.
Adjust the brightness.
Start observation.
‹Binocular tube (P. 16) ŒDiopter adjustment ring (P. 16) ™Condenser height adjustment screw (P. 19) šCondenser centering screws (P. 19)
›Aperture iris diaphragm ring (P. 20) œField iris diaphragm ring (P. 19)
ŠRevolving nosepiece button ‰Stage UP/DOWN button, focus
adjustment knob
ŸFilters (P. 9)
|Brightness adjustment buttons (P. 8)
(P. 12)
6
Page 11
Œ
Use the Allen screwdriver.
BX61
š
Use the Allen screwdriver.
œ
|
Ÿ
ƒ
³
@
Control Box* BX-UCB
* Set the Z-board or AF-board. For the setup
²
Hand Switch U-HSTR2
} Make a photocopy of the observation procedure pages and post it near your microscope.
Š
and installation of the Z-board, refer to page 30. For the information regarding the AF-board, refer to the instruction manual of the AF unit.
7
Page 12
USING THE CONTROLS
3-1 Base
ƒ
²
@
³
Fig. 4
Fig. 5
|
1 Voltage Indication
1. Press the Brightness adjustment button @ to increase the voltage and make illumination brighter. Pressing the button ² makes the illumination darker.
2. The numerals on the left of the lamp voltage LEDs ³ indicate the approximate voltage.
#The LED may sometimes go out (temporarily) when the motor is
driving. Even in this case, there is no fluctuation in illumination voltage.
2 Transmitted/Reflected Light Switch
The illumination can be switched between the transmitted light and re­flected light by pressing the transmitted/reflected light switch |.
: Reflected light illumination : Transmitted light illumination
The indicator of the selected illumination lights.
3 Light Preset Switch
}The Light Preset switch sets the light intensity voltage to a voltage suit-
able for color photography* (the factory default is 9 V) regardless of the current setting of the light intensity adjustment buttons.
* Achieved by engaging the built-in LBD filter.
Setting the Desired Brightness (Figs. 4 & 5)
1. Press the Light Preset switch ƒ The PHOTO indicator … lights up.
2. Press one of the light intensity adjustment buttons @ ² to set the de­sired brightness.
3. Press the Light Preset switch again. The indictor turns off, the brightness returns to the original brightness but the setting made above is stored in memory.
4. Hereafter, pressing the Light Preset switch sets the brightness to the in­tensity value set in step 2 above.
#Be sure to set the light intensity voltage back to 9 V when color
photography is to be performed.
}When the Light Preset switch ƒ is pressed, a note sound (short beep) is
generated at the 9 V position.
(Fig. 4)
(Fig. 4)
(Fig. 5)
8
4 Using the Filters
}You can place a filter in the light path with one of the following four
methods.
· Push in the knob of a built-in filter to engage it in the light path. (Page 9)
· Place a filter on the filter mount on the base and engage in the light path. (Page 9)
· Insert a filter in the U-FC filter cassette, attach it on the filter mount and engage it in the light path by sliding the filter lever. (Page 9)
· Insert a filter in the U-FWT filter wheel, attach it on the filter mount and engage it in the light path by motorized operation (using the PC). (Refer to the separate instruction manual.)
(Figs. 6 - 11)
Page 13
@
Fig. 6
²
³
|
Using Built-In Filters (Fig. 6)
Each filter can be engaged into the light path by pressing one of the filter knobs @ to | in. Pressing the button again disengages the filter from the light path.
Filter Type/Purpose
@
ND6 (Neutral Density filter for light adjustment, transmittance 6%)
²
ND25 (Neutral Density filter for light adjustment, transmittance 25%)
³ LBD (for color balancing, daylight filter)
| OP (optional) filter set*
* Ask your Olympus representative to mount the filters.
BX61
ƒ
Fig. 7
Fig. 8
Mounting a Single Filter (Fig. 7)
A filter with a diameter of 45 mm ƒ can be placed on the filter mount on the base. If you need to place multiple filters, please purchase the U-FC filter cassette.
# Even when a filter cassette is used, a filter with thickness of no more
than 3 mm can still be placed on the filter mount.
Using the Filter Cassette (Figs. 8 - 11)
Loading Filters into Filter Cassette
}The filter cassette accommodates filters with a diameter of 45 mm and
thickness of 2.7 mm or less.
}The filter cassette has two filter levers on the right side and one on the left
side.
1. Move all filter levers to the OUT position except for the one belonging to the slot into which the filter is to be inserted.
2. Slide the lever … to the IN position. Make sure that it clicks securely into place. (Fig. 8)
3. Holding the lever in the position shown in Fig. 8, put the filter into the cassette by inserting it in the direction indicated by the arrow.
4. Insert the other two filters in the same manner.
9
Page 14
Fig. 9
Fig. 10
Š
Mounting the Filter Cassette
1. Fully loosen the filter cassette clamping screw †. (Fig. 9)
2. Align the key ‡ on the bottom surface of the filter cassette with the positioning slot Š on the filter mount, then snap the filter cassette into place from above.
3. Rotate the filter cassette to align its sides with the base. (Fig. 10)
4. Align the clamping screw † with the positioning hole ‰ on the filter mount, then tighten the screw to fasten the filter cassette.
#When the filter cassette is installed, the condenser may hit the filter
cassette when the stage is lowered. Therefore, exercise caution when lowering the stage with the filter cassette installed.
Fig. 11
Using the Filter Cassette (Fig. 11)
Usable Filters Applications
45ND-6, 45ND-25 Neutral density filter
45G-530, 45G-533, 45IF550 Green
45Y-48 Yellow B&W contrast filter
45O-560 Orange
Up to three of the above filters can be loaded in the filter cassette. Mov­ing the levers ‹ on the left and right sides of the cassette to the IN position brings the corresponding filter into the light path.
10
Page 15
3-2 Focusing Block
²
@
³
Fig. 12
BX61
#If the substage is not installed in the correct position, it
may deviate from the coverage range of the damper spring of the objective and damage the specimen.
Ensure that the top surface @ of the substage is flush with the bottom surface of the notch ² on the mount dovetail. If they are not flush, loosen the clamping screw ³ and lower the substage until it is stopped.
@
³
²
Fig. 13
@
|
1 Replacing the Focus Adjustment Knob
# The focus adjustment knob is installed on the right side of the micro-
scope when it left the factory.
}The focus adjustment knob is designed detachable to prevent interfer-
ence with hand during manipulation of the focus adjustment knob or X/ Y-axis knobs. Usually attach the knob on the opposite side to the stage feed knobs.
1. Loosen the clamping screw @ and remove the focus adjustment knob ².
2. Remove the seal from the focus adjustment knob screw hole on the other side and attach the knob by reversing the removal procedure.
3. Attach a provided seal on the screw hole | of the focus adjustment dial ³, from which the focus adjustment knob has been removed.
}The focus adjustment dial ³ can be operated with your fingertip or finger
surface at the same time as manipulating the X/Y-axis knobs.
2 F/C button
}This button switches the movement of the stage using the focusing
knob @ or dial between F (Fine) and C (Coarse). (For the purpose of safety, this button is set automatically to F at the moment the main switch of the BX-UCB control box is set to “ I ” (ON).)
· Each press of the F/C button ² switches F and C alternately.
(Fig. 13)
(Fig. 14)
²
Fig. 14
Stage Fine movement: 0.1 mm per turn Stage Coarse movement: 1 mm per turn
11
Page 16
²
@
Fig. 15
3 Stage UP/DOWN Buttons
!When lowering the stage, be careful not to have your hand caught
between the bottom of the condenser and the base.
· Press the stage UP button @ to raise the stage and press the stage DOWN button ² to lower the stage.
· The stroke is 25 mm. It becomes 14 mm when the U-FWT filter wheel is used.
· The upper limit of stage movement has been set so that the stage stops within the coverage range of the damper spring of objective when a slide specimen (thickness 1.2 mm) is used.
#When observing a metallurgic specimen with thickness above 1.2
mm, take care so that the stage does not hit the objective.
(Fig. 15)
@
Fig. 16
4 Stage Escape Button
When replacing the specimen, press the stage escape button @. The stage will lower by 5 mm (in 1 sec.). Pressing the button again returns the stage to the original height.
(Fig. 16)
12
Page 17
BX61
3-3 Stage
}The following description deals with biological specimens. When observing a metallurgic specimen, it is more convenient
to replace the special stage or slide holder with the stage plate. (For details, refer to the instruction manual for the BX-RLAA.)
²
Fig. 17
Fig. 18
@
1 Placing the Specimen
# The dimensions of the slide glass should be 26 x 76 mm with thick-
ness of 0.9 to 1.2 mm, and the cover glass should have thickness of
0.17 mm.
# When observing very large specimens, remove the slide holder and
place the specimen directly on the stage.
Microscopy with Double-Slide Holder (Fig. 17)
1. Press the stage DOWN button @ or stage escape button to lower the stage.
2. Open the specimen holder clamping lever ² and place the specimen slides on the stage by sliding each slide from the front.
3. After placing the slides as far as they will go, gently release the clamping lever.
Microscopy with Single-Slide Holder (Fig. 18)
The specimen slide can easily be placed by sliding it into the slide holder from the front.
Examining a whole specimen
Use one of the thin slide holders listed below that cause less interference with the objectives*.
· U-HRD-4 · U-HLD-4 · U-HLS-4 * The applicable objectives are ones are ones with 40X or less
magnification (except Apo series).
³
Using an Oil Immersion Objective
Adsorption of immersion oil can cause the specimen to drift. In such cases, it is recommended to use the optional BH2-SCB-3 specimen clip ³ for oil immersion objectives. (Fig. 19)
Using an Oil Immersion Condenser
It may be effective to use the optional U-SVRO (right-hand knob) or U­SVLO (left-hand knob) slotted stage which prevents close contact be­tween the specimen and stage.
Fig. 19
13
Page 18
2 Adjusting the X/Y-Axis Knob Tension
(Fig. 20)
²
|
³
@
Fig. 20
1. Hold the X-axis knob @ and slide up the Y-axis knob ² up to expose the adjustment knobs.
2. Turning the X-axis adjustment knob ³ or Y-axis adjustment knob | clock­wise (in the direction of the arrow) increases the rotation tension and counterclockwise decreases it.
#If the tension is adjusted too tight, a creaking sound may be heard
during stage travel, and the stage stopping accuracy may be imper­iled.
CAUTION
[Treatment] X-axis direction: Hold the specimen holder and move the
Rubber Stage Knob Caps (Optional)
}When the X/Y-axis knobs are fitted with the rubber caps, the knobs can
be adjusted without slipping and fine adjustment is possible by holding the knobs with a very light force. The knob rubber caps also reduce fatigue after long hours of operation. The U-SHGT thick type (thickness 5 mm) and U-SHG thin type (thickness 2 mm) knob rubber caps are available.
To attach the knob rubber caps:
First fit the larger rubber knob cap to the Y-axis (upper) knob from below it, then fit the smaller rubber knob cap to the X-axis (lower) knob from below it.
After long hours of use, the stage guide may be devi­ated and the stage travel range may be decreased. However, this is not malfunction and can be corrected easily as described below.
stage guide to the left and right so that it hits the stoppers.
Y-axis direction: Hold the upper stage and move it to the front
and rear so that it hits the stoppers.
14
@
Fig. 21
3 Rotating the Stage
1. Slightly loosen the stage clamping knob @.
2. The stage can be rotated both clockwise and counterclockwise while holding the stage clamping knob.
#A click may be heard and felt during rotation. However, this is due to
the construction of the substage and does not indicate a malfunc­tion.
}The angle of rotation varies depending on the positions of the feed knobs.
Angle of Rotation
Clockwise Counterclockwise
Right-hand knobs 230° 20°
Left-hand knobs 20° 230°
(Fig. 21)
Page 19
BX61
²
Fig. 22
|
Fig. 23
@
ƒ
³
4 Adjusting the Stage Height
}By lowering the position of the substage, the microscope will accommo-
date specimens with maximum height of 35 mm. This is useful when observing metallurgical specimens and other thick objects.
1. Press the stage DOWN button @ to lower the stage to the lower limit, then remove the stage from the microscope.
2. Using an Allen screwdriver, loosen the substage bracket clamping screw ² and remove the substage.
3. Press the stage UP button ³ and raise the focusing block ƒ to where the stopper screw | on the arm becomes visible.
4. Using the Allen screwdriver, loosen and remove the upper stopper screw |.
5. Reattach substage bracket and stage.
}Retain the removed stopper screw | and take care not to lose it.
(Figs. 22 & 23)
15
Page 20
3-4 Observation Tube
²
Fig. 24
Fig. 25
@
1 Adjusting the Interpupillar Distance
While looking through the eyepieces, adjust for binocular vision until the left and right fields of view coincide completely. The index dot · indicates the interpupillary distance.
}Note your interpupillary distance so that it can be quickly duplicated.
2 Adjusting the Diopter
1. Looking through the eyepiece without the diopter adjustment ring, rotate the coarse and fine focus adjustment knob to bring the specimen into focus.
2. Looking through the eyepiece with the diopter adjustment ring @, turn it to focus on the specimen. (Fig. 25)
(Figs 25 & 26)
(Fig. 24)
Fig. 26
Using a Finder Eyepiece
1. Looking through the right eyepiece with your right eye, turn the top of the eyepiece ² until a clearly defined double crossline can be seen in the field of view. (Figs. 25 & 26)
2. Looking through the right eyepiece, rotate the coarse and fine focus ad­justment knob to bring the specimen and double crossline into simulta­neous focus.
3. Looking through the left eyepiece with your left eye, turn the diopter ad­justment ring @ to focus on the specimen.
Using an Super Widefield Observation Tube
The operation is similar to the above. However, as the super widefield tube does not have the diopter adjustment ring on the left sleeve, focus on the specimen by turning the top of the left sleeve.
3 Using the Eye Shades
When Wearing Eyeglasses
Use with the eye shades in the normal, folded-down position. This will prevent the eyeglasses from being scratched.
When Not Wearing Eyeglasses
Extend the folded eye shades in the direction of the arrow to prevent extraneous light from entering between the eyepieces and eyes.
(Fig. 27)
16
Fig. 27
Page 21
BX61
Fig. 28
Fig. 29
@
@
²
4 Using the Eyepiece Micrometer Disk
When the WHN10X-H (or WHN10X) eyepieces are used, an eyepiece micrometer disk can be inserted in one of them. When the eyepiece does not have a diopter adjustment mechanism, however, it is hard to focus on the micrometer disk if the operator has poor eyesight. Should that be the case, adjust the focus with eyeglasses on. Use 24 mm dia. x 1.5 mm thick micrometer disks. Following Fig. 28, turn the built-in micrometer-mounting frame ² counterclockwise to remove it from the eyepiece and place a micrometer disk. Screw the micrometer mounting frame back into the eyepiece sleeve.
Selecting the Light Path of the
5
Trinocular Tube
Slide the light path selector knob @ to select the desired light path.
Trinocular tube
U-TR30-2
U-SWTR-3
U-TR30NIR
*
100% for binocu­lar eyepieces
Light path selector knob position
Pushed in
Middle position
20% for binocular eyepieces, 80% for TV/photography
50% for binocular eyepieces, 50% for TV/photography
100% for TV/ photography
(Fig. 28)
(Fig. 29)
Pulled out
* With the infrared trinocular tube, infrared observation up to 1000 nm is
possible. For details, consult your Olympus representative.
17
Page 22
Fig. 30
6 Adjusting the Tilt (with the U-TBI3)
}Adjust the height and tilt of the observation tube to obtain the most
comfortable viewing position. Holding the binocular section with both hands, raise or lower it to the desired position.
# Never attempt to force the binocular section past the upper or lower
stop position. Applying excessive force could destroy the limiting mechanism.
}The U-TBI3 can be used in combination with only one intermediate at-
tachment. However, no intermediate attachment can be combined when the BX-RFAA or BX-RLAA reflected light illuminator is used.
}If you need photography using the U-TBI3, also use the U-TRU trinocular
intermediate tube.
With the U-ETBI/U-TTBI (Fig. 31)
The U-ETBI and U-TTBI are ergonomic observation tubes with normal field, capable of the tilting adjustment as well as the adjustment of the eyepiece position toward the front and rear (by 45 mm). The U-ETBI is the erect image model and the U-TTBI is the inverted image model, and both models are of the same size.
#The intermediate attachments that can be combined with the U-
TTBI are limited. For details, please contact Olympus.
(Fig. 30)
Fig. 31
18
Page 23
3-5 Condenser
BX61
@
²
Fig. 32
³
1 Centering the Condenser
1. Fit the Allen screwdriver @ into the condenser height adjustment screw and turn the tool counterclockwise to raise the condenser to its upper limit.
2. Focus on the specimen using the 10X objective.
# When using a swing-out condenser, move the top lens into the light
path.
3. Rotate the field iris diaphragm ring ² in the direction of the arrow so that the diaphragm image comes inside the field of view.
4. Turn the Allen screwdriver @ to focus on the diaphragm image.
5. Fit the Allen screwdrivers into the two condenser centering screws ³ and turn them to move the iris diaphragm image to the center of the field of view.
6. Gradually open the field iris diaphragm. The condenser is properly cen­tered if the iris image is centered and inscribed in the field of view.
7. During actual use, open the field diaphragm slightly until its image cir­cumscribes the field of view.
(Figs. 32 & 33)
Fig. 33
Effects of Field Iris Diaphragm (Fig. 33)
The field iris diaphragm restricts the diameter of the beam of light enter­ing the objective and thus excludes extraneous light, improving image contrast. The diameter of the field iris should e adjusted for objective power to the extent that it just circumscribes the field of view. (See “Com­patibility of Objectives and Condensers” on the next page.)
19
Page 24
Aperture iris diaphragm image
Objective pupil
Fig. 34
70-80%
30-20%
²
Aperture Iris Diaphragm (Figs. 34 & 35)
· The aperture iris diaphragm determines the numerical aperture of the illumination system. It has an effect of adjusting image resolution and contrast. Stopping down the aperture iris diaphragm increases the depth of focus.
· Since the contrast of microscope specimens is ordinarily low, setting the condenser aperture iris diaphragm to between 70 and 80% of the N.A. ² of the objective in use is usually recommended. Adjust the ratio by removing the eyepiece and looking into the eyepiece sleeve while adjusting the aperture iris diaphragm ring @ until the image shown in Fig. 34 is seen.
}Using the numerical aperture scale:
You can use the condenser numerical aperture scale by adjusting the aperture iris diaphragm ring @. (Fig. 35)
@
Fig. 35
N.A. scale
2 Compatibility of Objectives and Condensers
Objective Magnification
1.25X Applicable to FN 22
2X
4X
10-60X
100X
* When using the U-SC3 swing-out condenser together with a 1.25X to 4X objective, fully open the condenser aperture iris
diaphragm and use the field iris diaphragm in the base as aperture diaphragm. With the 1.25X to 2X objectives, the surroundings of the field of view may become dark.
}To obtain better illumination in photomicrography using a 1.25X to 4X objective, use of the U-ULC-2 for ultralow magnifi-
cation is recommended.
Achromat/aplanat
U-AAC
Applicable
(FN 26.5)
Applicable (FN 26.5) by moving top lens out of the light path.*
Engage top lens in light path.
Condenser
Swing-out
U-SC3
(FN 26.5)
Motorized universal
U-UCD8A
Applicable when U-FWCO1.25X is combined.
Applicable (FN 26.5) by moving top lens out of the light path.
Engage top lens in light path.
(FN 26.5)
20
Page 25
3-6 Immersion Objectives
# Be sure to use the provided Olympus Immersion oil.
BX61
Fig. 36
1 Using Immersion Objectives
1. Focus on the specimen using all objectives, starting from the lowest­power objective to higher-power objectives.
2. Press the stage escape button to lower the stage, then place a drop of provided immersion oil onto the specimen at the area to be observed.
3. Turn the revolving nosepiece to engage the immersion objective.
4. Press the stage escape button to raise the stage, then turn the focus adjustment knob to focus on the specimen.
# Since air bubbles in the oil will affect the image quality, make sure
that the oil is free of bubbles.
a. To check for bubbles, remove the eyepiece and fully open the field and
aperture iris diaphragms, then look at the exit pupil of the objective inside the observation tube. (The pupil should appear round and bright.)
b. To remove bubbles, turn the revolving nosepiece toward the shorter ob-
jectives and repeatedly defocus and refocus the oil immersion objective.
}If the condenser engraving shows a numerical aperture (NA) of 1.0 or
higher, the number applies only when oil is applied between the slide glass and the top surface of the condenser. When oil is not present, the NA is about 0.9.
5. After use, remove oil from the objective front lens by wiping with gauze slightly moistened with absolute alcohol.
!Caution in use of immersion oil
If immersion oil enters your eyes or contacts with your skin, immediately take the following treatment.
Eyes: Rinse with fresh water (for 15 minutes or more).
Skin: Rise with water and soap. If the appearance of the eyes or skin is altered or pain persists, immedi­ately see your doctor.
(Fig. 36)
3-7 Objectives with Correction Collar
}If the cover glass thickness is not 0.17 mm, the objectives cannot mani-
fest their performances. If a correction collar equipped objective is used in this case, the difference in thickness can be compensated for by ad­justing the collar.
Adjustment Procedure
· If the cover glass thickness is known, set the correction collar @ to that
@
Fig. 37
value. (Fig. 37)
· If the cover glass thickness is unknown, adjust the correction collar @ and focus adjustment knob alternately until the positioning with the highest resolution is obtained.
# Be careful not to touch the correction collar @ when turning the
revolving nosepiece.
21
Page 26
TROUBLESHOOTING GUIDE
Under certain conditions, performance of the unit may be adversely affected by factors other than defects. If problems occur, please review the following list and take remedial action as needed. If you cannot solve the problem after checking the entire list, please contact your local Olympus representative for assistance.
Problem
1. Optical System
a) Bulb does not light. Power cord of BX-UCB is unplugged.
Main switch of BX-UCB is not on. Set main switch to “ I ” (ON).
Lamp ON-OFF switch on the base of BX61 is not on.
Bulb is burned out. Replace bulb with a new specified
Transmitted/reflected light switch is set to reflected light ( ).
b) Bulb operates, but field of view re-
mains dark.
c) Field of view is obscured or not
evenly illuminated
Aperture and field iris diaphragms are not opened wide enough.
Condenser is lowered too much. Adjust the condenser height position. 19
Light path selector knob is set to the position.
Light path selector knob is in an posi­tioned halfway.
Revolving nosepiece is not correctly en­gaged.
Condenser is not attached properly. Re-attach it. 32
Revolving nosepiece is not attached properly.
An objective that falls outside of the condenser’s illumination range is used.
Condenser is not properly centered. Center the condenser. 19
Cause Remedy Page
Insert power plug firmly into power outlet.
Set the lamp ON-OFF switch to ON.
bulb.
Set the switch to (transmitted light illumination).
Adjust them to proper sizes.
Move the knob to the or posi­tion.
Set the knob according to the obser­vation method.
Make sure that the revolving nose­piece clicks properly into place.
Push the side dovetail all the way un­til it is stopped.
Use a condenser to match the pur­pose.
31
8
19/20
17
17
20
22
d) Dirt or dust is visible in the field of
view.
e) Visibility is poor.
· Image is not sharp.
· Contrast is poor.
· Details are indistinct.
· Image glares.
Field iris diaphragm is stopped down too far.
Bulb is not mounted correctly. Push the pins of halogen bulb all the
Dirt/dust on the eyepiece Clean thoroughly.
Dirt or the to surface of the condenser
Dirt/dust on the specimen
A non-UIS2/UIS objective is used. Use only UIS2/UIS series objectives
Condenser is lowered too far. Adjust the condenser height position. 19
Aperture iris diaphragm is stopped down too far.
Revolving nosepiece is not mounted properly.
Open the field iris diaphragm until it circumscribes the field.
way until the stop position.
with this microscope.
Open aperture iris diaphragm.
Push the slide dovetail all the way until it is stopped.
19
31
27/28
20
3
Page 27
BX61
Problem
e) Visibility is poor.
· Image is not sharp.
· Contrast is poor.
· Details are indistinct.
· Image glares.
f ) One side of image is blurred. Revolving nosepiece is not correctly
g) Image appears to waver. Revolving nosepiece is not corrected
Correction collar on correction collar equipped objective is not properly ad­justed.
Front lens of objective is dirty. Clean objective. 3
Immersion oil is not being used with an oil immersion objective.
Immersion oil contains bubbles. Remove the bubbles. 21
Recommended immersion oil is not used.
Dirt/dust on specimen. Clean it.
Dirt/dust on condenser
Inappropriate object side or cover glass thickness.
mounted.
Stage is not correctly mounted. Re-attach it. 15
Objective is not correctly engaged in light path.
Specimen is not correctly mounted on stage.
mounted.
Objective is not correctly engaged in light path.
Condenser is not properly centered. Center the condenser. 19
Cause Remedy Page
While focusing, turn the correction col­lar to find the best position.
Use immersion oil.
Use the provided immersion oil.
Replace with glass of recommended thickness.
Push slide dovetail all the way until it is stopped.
Make sure that revolving nosepiece clicks into place correctly.
Place specimen correctly on to of stage and secure it with slide holder.
Push slide dovetail all the way until it is stopped.
Make sure that revolving nosepiece clicks into place correctly.
21
21
21
3
13
13
h) Field of view becomes only slightly
brighter when the voltage is raised.
2. Electrical System
a) Bulb intermittently lights and goes
out.
b) Bulb burns out almost immediately. Wrong type of bulb is being used. Use correct bulb type. 31
c) All voltage indicator LEDs light and
voltage cannot be varied by press­ing light intensity adjustment but­ton.
d) Brightness does not change when
you press light intensity adjustment button.
3. Focusing Block
a) Image cannot be focused. When adjusting stage height, you forgot
b) Stage rises too high and destroys
specimen.
Condenser is not properly centered. Center the condenser. 19
Condenser is lowered too far. Adjust the condenser height position. 19
Bulb is nearly burned out. Replace bulb. 31
A connector is not properly connected.
Bulb is not installed. Install bulb. 31
Bulb is burned out. Replace bulb. 31
Lamp housing is not connected. Connect lamp housing correctly.
Bulb is burned out. Replace bulb.
to reattach upper stopper screw.
Substage is not installed properly. Re-attach substage.
Check all connectors.
Re-attach upper stopper screw.
31
15
11
23
Page 28
Problem
c) Coarse adjustment will not go all
the way down.
d) Objective makes contact with speci-
men before focus is obtained.
4. Observation Tube
Cause Remedy Page
Condenser holder is too low. Raise condenser holder.
Specimen is mounted upside down.
Mount specimen correctly.
32
a) Field of view of one eye does not
match that of the other.
5. Stage
a) Image shifts when you touch stage. Stage is not properly mounted. Clamp stage.
b) Specimen stops midway on the X-
axis traverse.
c) X-axis and/or Y-axis knobs are too
tight or too loose.
d) Stroke is decreased. Stage guide is deviated. Correct deviation as instructed on
Interpupillary distance is incorrect. Adjust interpupillary distance. 16
Incorrect diopter adjustment. Adjust diopter. 16
Different eyepieces are used on left and right.
Your view is not accustomed to micro­scope observation.
Specimen is not correctly positioned. Place specimen correctly.
X-axis and/or Y-axis adjustment knobs are set at a position where tension is too high or too low.
Change on eyepiece to match the other so that both sides are the same type.
Upon looking into eyepieces, try look­ing at overall field before concentrat­ing on specimen range. You may also find it helpful to look up and into dis­tance for a moment before looking back into microscope.
Adjust tension.
page 14.
13
14
14
24
Page 29
SPECIFICATIONS
Item Specification
1. Optical system UIS2/UIS (Universal Infinity System) optical system (featuring infinity correction)
2. Illumination
3. Focusing mechanism Drive system: Motorized focusing using a stepping motor and ball screw.
Built-in transmitted Koehler illumination 12 V , 100 W long-life halogen bulb (pre-centered) 12 V 100 W HAL-L (PHILIPS 7724) or 12 V , 50 W long-life halogen bulb (pre-centered) 12 V 50 W HAL-L (LIFE JC) (Average life time: Approx. 2,000 hr. when used as directed) Light intensity voltage range: 2 V or less to 12.0 V (continuous) With a light intensity preset switch (voltage variation range: 2 V or less to 12.0 V ) Rated voltage of external power supply (BX-UCB): 100-120/220-240 V , 50/60 Hz, 3.5 A /1.5 A Power consumption (BX-UCB):50–300 W (variable with number of connected modules)
(Ex.) Approx. 200 W with 100 W halogen bulb and motorized focusing unit
Stage height movement by cross roller guides. Finest adjustment scale: 1 μm (fine movement sensitivity 1 μm) Resolution: 0.01 μm Highest stage speed: 3 mm/sec. Stroke per rotation: 0.1 mm (fine), 1 mm (coarse) Full stroke range: 25 mm to 14 mm (when the U-FWT filter wheel is used)
BX61
4. Revolving nosepiece
5. Observation tube Type U-BI30-2 U-TBI3
6. Stage Type U-SVRB-4 U-SVLB-4
Manual revolving nosepieces:
U-5RE-2, U-D6RE, U-D7RE, U-P6RE
Motorized revolving nosepieces:
U-D6REM, U-D5BDREM, U-D6REMC, U-D6BDREMC, U-D5BDREMC, U-P5REMC, U-P5BDREMC
U-TR30-2
Widefield bin­ocular tube
Field No. 22 26.5
Tube inclination
Interpupillary distance ad­justment
Light path se­lector
30°
Common-axis knobs on bottom right. Common-axis knobs on bottom left.
Widefield tilting binocular
5°-35°
continuous
50 mm to 76 mm
None
Rectangular ceramic coated. wire driven stage
Widefield trin­ocular tube
U-TR30NIR
Super widefield trinocular
30°
3 steps: @ Bi 100%
² Bi 20%, Photo 80% Bi 50%, Photo 50% for U-TR30NIR only
³ TV, Photo 100%
U-SWTR-3
Size 156 mm (H) x 191 mm (W)
Movement mechanism
Specimen holders (single slide holder)
Specimen holders (double slide holder)
X/Y-axis knobs with adjustable tension. Movement range: 52 mm in Y-axis direction, 76 mm in X-axis direction.
U-HLST-4 (thick slide holder)
U-HLS-4 (thin slide holder)
U-HLDT-4 (left opening, thick slide holder)
U-HLD-4 (left opening, thin slide holder)
U-HRDT-4 (right opening, thick slide holder)
U-HRD-4 (right opening, thin slide holder)
25
Page 30
Item Specification
7. Condenser Type
N.A. 1.40 0.9 - 0.1 1.4 - 0.2
Aperture iris diaphragm
Objective range
U-AAC
Achromat/aplanat
10X - 100X (for FN 26.5 super widefield)
U-SC3
Swing-out
With numerical aperture scale
1.25X (for FN 22 widefield), 2X - 100X (for FN 26.5 super widefield)
U-UCD8A
Motorized, achromat/
aplanat, swing-out
2X - 100X (for FN 26.5 super widefield)
8. Operating environment
· Indoor use.
· Altitude: Max. 2000 meters
· Ambient temperature: 5° to 40°C (41° to 104° F)
· Maximum relative humidity: 80% for temperatures up to 31°C (88°F), decreasing linearly through 70% at 34°C (93°F), 60% at 37°C (99°F), to 50% relative humidity at 40°C (104°F).
· Supply voltage fluctuations; Not to exceed ±10% of the normal voltage.
· Pollution degree: 2 (in accordance with IEC60664)
· Installation/Overvoltage category: II (in accordance with IEC60664)
26
Page 31
BX61
OPTICAL CHARACTERISTICS «UIS2 (UIS) Series»
UIS series objectives not listed here can also be combined with this microscope.
The following table shows the optical characteristics of com­binations of eyepieces and objectives. The figure on the right shows the performance data engraved on the objectives.
NOTE
Refer to the latest catalogue or consult your local Olympus representative for the updated information on the eyepieces and objectives that can be combined with this microscope.
Optical
Objective
UIS2 Series
character
PlanN-P Plan Achromat for polarized light (FN22)
AchN-P Achromat for polarized light (FN22)
PlanN Plan Achromat (FN22)
UPlanFLN Plan Semi Apochromat (FN26.5)
UPlanSApo Plan Apochromat (FN26.5)
PlanApoN Plan Apochromat (FN26.5)
MPlanFLN Plan
Semi Apochromat (FN26.5)
MPlanApoN No Cover Plan Apochromat (FN26.5)
Magnifi-
cation
4X 0.1 18.5
10X 0.25 6.0 20X 0.4 3.0 0.17 0.84 200X 9.3 1.1 40X 0.65 0.45 0.17 0.52 400X 2.0 0.55
100XO 1.25 0.13
2X 0.06 5.8 4X 0.1 18.5
10X 0.25 10.6 20X 0.4 1.2 0.17 0.84 200X 9.27 1.1 40X 0.65 0.6 0.17 0.52 400X 3.04 0.55
50XOI 0.5-0.9 0.2 0.17 0.37 500X 1.7 0.44 100XO 1.25 0.15
4X 0.13 17.0
10X2 0.3 10.0
20X 0.5 2.1 0.17 0.67 200X 7.0 1.1 40X 0.75 0.51 0.17 0.45 400X 2.52 0.55
40XO 1.3 0.2 0.17 0.26 400X 1.27 0.55
60X 0.9 0.2 0.17 0.37 600X 1.5 0.37
60XOI 0.65-1.25 0.12 0.17 0.27 600X 0.98 0.37 100XO2 1.30 0.2 0.17 0.26 1000X 0.66 0.22 100XOI2 0.6-1.30 0.2 0.17 0.26 1000X 0.66 0.22
4X 0.16 13.0
10X2 0.4 3.1 0.17 0.84 100X 15.9 2.2
20X 0.75 0.6 0.17 0.45 200X 4.29 1.1
20XO 0.85 0.17 40X2 0.95 0.18 0.17 0.35 400X 1.9 0.55 60XW 1.2 0.28 0.17 0.28 600X 1.03 0.37 60XO 1.35 0.15 0.17 0.25 600X 0.89 0.37 100XO 1.4 0.13 0.17 0.24 1000X 0.59 0.22
1.25X 0.04 5.0 2X 0.08 6.2
60XO 1.42 0.15 0.17 0.24 600X 0.83 0.37
40X 0.75 0.63 0 0.45 400X 1.7 0.55
100XO 1.40 0.10 0 0.24 1000X 0.59 0.22
N.A.
W.D.
(mm)
Cover
glass thick-
ness
(mm)
— —
— —
— —
— —
Magnification
Mechanical tube length
Color band
Reso­lution
(μm)
3.40 40X 180.0 5.5
1.30 100X 28.0 2.2
0.27 1000X 0.69 0.22
5.59 20X 560.1 11.0
3.36 40X 175.0 5.5
1.34 100X 28.0 2.2
0.27 1000X 0.69 0.22
2.58 40X 127.2 5.5
1.12 100X 22.4 2.2
2.50 40X 99.6 5.5
0.39 200X 3.5 1.1
8.39 12.5X 1326.8 17.6
4.19 20X 398.3 11.0
Total mag.
Eyepiece
WHN10X (FN22)
Depth of
focus
(μm)
Objective type
Number of aperture
Oil: Oil immersion Iris: Iris diaphragm
Field number (FN)
Cover glass thickness
Field of
view
Remark
Oil immersion/Iris Oil immersion
Oil immersion Correction collar Oil immersion/Iris Oil immersion Oil immersion/Iris
Oil immersion Correction collar Water immersion Oil immersion Oil immersion
Oil immersion
Oil immersion
27
Page 32
Objective
UIS Series
Optical character
PlanFl Plan Semi Apochromat (FN26.5)
UApo Apochromat (FN22)
Cover
Magnifi-
cation
100X 0.95 0.20 0.14-0.20 0.35 1000X 1.0 0.22
20X3/340 0.75 0.55 0.17 0.45 200X 4.3 1.1 20XW3/340 40X3/340 0.90 0.20 0.11-0.23 0.37 400X 2.0 0.55 40XW3/340 40XOI3/340
N.A.
0.70 0.40 0.17 0.48 200X 4.08 1.1
1.15 0.26 0.13-0.25 0.29 400X 1.29 0.55
0.65-1.35 0.10 0.17 0.25 400X 1.2 0.55
W.D.
(mm)
glass thick-
ness
(mm)
Reso­lution
(μm)
Total mag.
Eyepiece
WHN10X (FN22)
Depth of
focus
(μm)
Field of
Remark
view
Correction collar
Correction collar Correction collar Oil immersion/Iris
28
Page 33
BX61
ASSEMBLY
7-1 Assembly Diagram
The diagram below shows the sequence of assembly of the various modules. The numbers indicate the order of assembly. The BX61TRF microscope frame is assembled in the same procedures. The module model numbers shown in the following diagram are merely the typical examples. For the modules with which the model numbers are not given, please consult your Olympus representative or the catalogues.
# When assembling the microscope, make sure that all parts are free of dust and dirt, and avoid scratching any parts
or touching glass surfaces.
Assembly steps enclosed in will be detailed on the subsequent pages.
}Most assembly operations are possible by using the Allen screwdriver ( ) provided with the microscope. How-
ever, the assembly of the reflected light illuminator requires the use of the provided Allen wrench ( ) for clamping the internal screws (to ensure the performance, we recommend that you have your Olympus representative assemble or remove this module).
Eyepiece WHN10X (FN 22) 35WHN10X (FN 22) SWH10X-H (FN 26.5) 35SWH10X (FN 26.5)
Observation tube U-BI30-2 (FN 22) U-TR30-2 (FN 22) U-TR30NIR (FN 22) U-TBI3 (FN 22) U-SWTR-3 (FN 26.5) U-ETBI (FN 22) U-TTBI (FN 22)
Reflected light illuminator BX-RFAA BX-RLAA
Reflected lamp housing
Slide holder U-HLDT-4 U-HRDT-4 U-HLST-4
Stage clamping knob
Stage U-SVRB-4 U-SVLB-4 U-SRG
Condenser U-SC3 U-AAC U-UCD8A
Revolving nosepiece U-D6RE U-D6REM U-D6REMC, etc.
UIS2/UIS series objective
Specimen holder clamping knobs
Condenser clamping screw
Intermediate attachment U-EPA2 U-DO3 U-TRU, etc.
Tube clamping screw
Revolving nosepiece clamping screw
Microscope frame (Transmitted light specification) BX61TRF
Microscope frame (Transmitted/reflected light specification) BX61TRF
100 W halogen lamp housing U-LH100-3
Lamp socket clamping screw
}For the electrical connections, refer to the instruction manual of the BX-UCB control box.
For the setup and installation of the Z-board, see page 30.
100 W halogen bulb 12V100WHAL-L 12V50WHAL-L
29
Page 34
7-2 Detailed Assembly Procedures
# For use in focusing control, only either the Z-board or AF control board can be installed on the BX-UCB.
Installing different boards on it may result in malfunction. When using the AF control board instead of the Z-board, set the on-board DIP switches on the AF control board to the same settings as the following settings that are originally intended for the Z-board.
Setup and Installation of the Z-Board
}The on-board DIP switches on the Z-board have been set up for the
BX61 or BX62 (all of switches S1, S2 and S3 are set to OFF).
²
S1 S2 S3
Fig. 38
@
Fig. 39
³
Changing the Settings of On-Board DIP Switches (Fig. 38)
#Leave all other DIP switches than those mentioned below in the OFF
positions.
}Switch only the DIP switches mentioned below to the ON positions.
S2 No. 4
· This replaces the functions of the F/C switching and stage escape but­tons on the left side of the BX61/BX62 with the stage DOWN and UP buttons.
S3 No. 1
· This turns off the automatic initialization and sets the remote mode which is the default.
S3 No. 4
· When the U-FWT filter wheel is not used, this changes the focus adjust­ment stroke from 15 mm to 25 mm.
}When the U-LWUCD long-WD universal condenser is used, set this to
the OFF position.
Mounting the Z-Board (Fig. 39)
}Be sure to set the main switch on the BX-UCB control box @ to “ ”
(OFF) before proceeding to mounting the Z-board.
1. Loosen the 6 clamping knobs on the 2 option slot covers on the rear of the BX-UCB and remove the covers and knobs.
2. Insert the Z-board ² into the BX-UCB by aligning the orientation of the connector inside the BX-UCB and that of the Z-board. Push in the board along the board rails all the way so that the connector is plugged in securely.
3. Clamp the Z-board ² using two of the clamping knobs removed above. Also attach one of the removed covers ³ using the other four clamping knobs removed above.
}Store the other, removed cover in a safe place.
30
Page 35
BX61
@
²
Fig. 40
Fig. 41
ƒ
³
1 Installing the Halogen Bulb
}Use only the designated bulb 12V100WHAL-L (PHILIPS 7724) or
12V50WHAL-L (LIFE JC).
1. Fully loosen the lamp housing clamping screw @ on top of the lamp housing cover with the provided Allen screwdriver.
2. Lift the lamp housing cover ² upward to remove it.
3. Turn the lamp socket ³ by 90° in the direction indicated by the arrow.
4. Holding the bulb ƒ with gloves or a piece of gauze, depress the bulb clamping lever | and insert the bulb pins … fully into the sections † as far as they will go. Gently release the bulb clamping lever to the original position to secure the bulb.
To prevent reduced bulb life or cracking, do not touch the bulb with bare hands. If fingerprints are accidentally left on the bulb, wipe the bulb with a soft cloth.
5. Side the lamp housing cover onto the housing base from the above. Tighten the clamping screw @ while pressing downward on the cover. (Fig. 40)
Caution for Bulb Replacement During Use or Right After Use
The bulb, lamp socket and areas near these will be extremely hot during and right after use. Set the main switch to “ ” (OFF), disconnect the power cord from the wall outlet, then allow the old bulb and lamp housing to cool before replacing the bulb with a new of the designated type.
(Figs. 40 - 42)
|
Fig. 42
Fig. 43
@
²
6 Attaching the Eyepiece
Fit the eyepiece into each sleeve as far as it goes.
# When using the U-BI30-2 binocular tube, eyepiece with built-in mi-
crometer disk cannot be attached.
# When using a finder eyepiece or an eyepiece with focusing mirror,
attach it to the right-hand eyepiece sleeve. When doing so, make sure that the eyepiece positioning pin @ fits into the notch ² on the eyepiece sleeve.
# The super-widefield trinocular observation tube is equipped with a
positioning notch on both eyepiece sleeves. Make sure that the positioning pins on both eyepieces fits into the respective notches.
(Fig. 43)
31
Page 36
Fig. 44
³
²
@
9 Attaching the Condenser
}This operation necessitates the operation of the motorized focusing. Con-
nect the cables and set the main switch to “ I ” (ON) before proceeding to the following.
1. Press the stage UP button @ to raise the stage to its upper limit.
2. Fit the Allen screwdriver into the condenser height adjustment screw ² and turn the tool counterclockwise to lower the condenser to its lower limit.
3. Fully loosen the condenser clamping screw ³ using the Allen screw­driver.
}When attaching the U-SC3 swing-out condenser, align the positioning
pin at the back of the condenser with the groove in the condenser sleeve.
#When attaching a condenser with swing-out top lens, swing out the
top lens out of the way before inserting the condenser.
4. Position the condenser so that the aperture scale or optical device indi­cation plate is located in front. Gently slide the condenser along the dovetail all the way until it is stopped.
5. Tighten the condenser clamping screw ³ and gently raise the condenser holder to its highest position.
(Fig. 44)
32
Page 37
BX61
PROPER SELECTION OF THE POWER SUPPLY CORD
If no power supply cord is provided, please select the proper power supply cord for the equipment by referring to “ Specifications ” and “ Certified Cord ” below:
CAUTION:
In case you use a non-approved power supply cord for Olympus products, Olympus can no longer warrant the electrical safety of the equipment.
Specifications
Voltage Rating Current Rating Temperature Rating Length Fittings Configuration
125V AC (for 100-120V AC area) or, 250V AC (for 220-240V AC area) 6A minimum 60°C minimum
3.05 m maximum Grounding type attachment plug cap. Opposite terminates in molded-on IEC con­figuration appliance coupling.
Table 1 Certified Cord
A power supply cord should be certified by one of the agencies listed in Table 1 , or comprised of cordage marked with an agency marking per Table 1 or marked per Table 2. The fittings are to be marked with at least one of agencies listed in Table 1. In case you are unable to buy locally in your country the power supply cord which is approved by one of the agencies mentioned in Table 1, please use replacements approved by any other equivalent and authorized agencies in your country.
Country Agency
Argentina
Australia
IRAM
SAA
Certification
Mark
Country Agency
Italy
Japan
IMQ
JET, JQA, TÜV, UL-APEX / MITI
Certification
Mark
Austria
Belgium
Canada
Denmark
Finland
France
Germany
Ireland
ÖVE
CEBEC
CSA
DEMKO
FEI
UTE
VDE
NSAI
Netherlands
Norway
Spain
Sweden
Switzerland
United Kingdom
U.S.A.
KEMA
NEMKO
AEE
SEMKO
SEV
ASTA BSI
UL
33
Page 38
Table 2 HAR Flexible Cord
APPROVAL ORGANIZATIONS AND CORDAGE HARMONIZATION MARKING METHODS
Approval Organization
Comite Electrotechnique Belge (CEBEC)
Verband Deutscher Elektrotechniker (VDE) e.V. Prüfstelle
Union Technique de l´Electricite´ (UTE)
Instituto Italiano del Marchio di Qualita´ (IMQ)
British Approvals Service for Electric Cables (BASEC)
N.V. KEMA
SEMKO AB Svenska Elektriska Materielkontrollanstalter
Österreichischer Verband für Elektrotechnik (ÖVE)
Danmarks Elektriske Materialkontroll (DEMKO)
Printed or Embossed Harmoniza­tion Marking (May be located on jacket or insulation of internal wir­ing)
CEBEC <HAR>
<VDE> <HAR>
USE <HAR>
IEMMEQU <HAR>
BASEC <HAR>
KEMA-KEUR <HAR>
SEMKO <HAR>
<ÖVE> <HAR>
<DEMKO> <HAR>
Alternative Marking Utilizing Black-Red-Yellow Thread (Length of color section in mm)
Black Red Yellow
10 30 10
30 10 10
30 10 30
10 30 50
10 10 3 0
10 30 30
10 10 5 0
30 10 50
30 10 30
National Standards Authority of Ireland (NSAI)
Norges Elektriske Materiellkontroll (NEMKO)
Asociacion Electrotecnica Y Electronica Espanola (AEE)
Hellenic Organization for Standardization (ELOT)
Instituto Portages da Qualidade (IPQ)
Schweizerischer Elektro Technischer Verein (SEV)
Elektriska Inspektoratet
Underwriters Laboratories Inc. (UL) SV, SVT, SJ or SJT, 3 X 18AWG Canadian Standards Association (CSA) SV, SVT, SJ or SJT, 3 X 18AWG
<NSAI> <HAR>
NEMKO <HAR>
<UNED> <HAR>
ELOT <HAR>
np <HAR>
SEV <HAR>
SETI <HAR>
30 30 50
10 10 70
30 10 70
30 30 70
10 10 9 0
10 30 90
10 30 90
34
Page 39
Page 40
EC REP
Shinjuku Monolith, 3-1, Nishi Shinjuku 2-chome, Shinjuku-ku, Tokyo, Japan
Wendenstraße 14-18, 20097 Hamburg, Germany
3500 Corporate Parkway, P.O. Box 610, Center Valley, PA 18034-0610, U.S.A.
One Corporate Drive, Orangeburg, NY 10962, U.S.A.
491B River Valley Road, #12-01/04 Valley Point Office Tower, Singapore 248373
31 Gilby Road, Mount Waverley, VIC., 3149, Australia
Blue Lagoon Drive, Suite 290 Miami, FL 33126, U.S.A.
5301
01/10
Loading...