Nihon Kohden YG211T Owner's Manual

OPERATOR’S MANUAL
Airway Adapter
First Edition: 01 Oct 2008
Attention, consult operator’s manual Expiration date
Year and month of manufacture
Lot number
Single use only
Second Edition: 20 May 2009
General
The YG-211T airway adapter is used with a TG-221T CO2 sensor
of the TG-970P CO2 sensor kit to measure the partial pressure of
the expired CO2 of the patient.
Connects to a respiration device
Connects to a mask or an endotracheal tube
YG-211T0604-904708A
The CE mark is a protected conformity mark of the European Community. The products herewith comply with the requirements of the Medical Device Directive 93/42/EEC.
CAUTION
Connect the airway adapter to the respiration circuit
so that the UP mark on the airway adapter faces
upward. Otherwise, water droplets may get into the
airway adapter and affect the measurement accuracy.
Transparent film made from special anti-fogging material
Dead space volume: 4 mL (for patients more than approximately
7 kg)
This airway adapter can be used with Nihon Kohden monitors
which measure the partial pressure of the expired CO2. Refer to
the operator’s manual for your Nihon Kohden monitor to check
if the YG-211T airway adapter can be used with your Nihon
Kohden monitor. Read the operation manuals for the monitor,
CO2 sensor and CO2 sensor kit together with this manual.
UP mark
WARNING
When using the airway adapter on patients with low
ventilatory amount, due to the airway adapter’s dead
space volume (4 mL), the CO2 sensor might have
difficulty in detecting no breath. Perform ventilation
taking into consideration the 4 mL dead space volume.
CAUTION
The airway adapter is non-sterilized and disposable.
Use only for a single patient and single use. Failure to
follow this instruction causes cross infection.
CAUTION
Failure to follow the instructions below degrades the
anti-fogging ability of the transparent film and results
in incorrect measurement.
• Replace the airway adapter with a new one every 24
hours.
• Replace the airway adapter with a new one if blood,
sputum or mucus adhere to the transparent film.
• Do not damage the transparent film. Do not let dust
or detergent contact the transparent film. Do not
touch, wipe or clean the transparent film with fingers
or cleaners.
CAUTION
United States law restricts this product to sale by or on
the order of a physician.
NOTE
• Open the package just before use.
• Do not leave the airway adapter in a place with high
temperature such as on a dashboard of a car. This
may deform the airway adapter and cause inaccurate
measurement.
• Loose or damaged connections may compromise
ventilation or result in inaccurate measurements of
respiratory gases. Securely connect all components and
check the connections regularly.
Replacing the Airway Adapter
NIHON KOHDEN CORPORATION
1-31-4 Nishiochiai, Shinjuku-ku, Tokyo 161-8560, Japan Phone +81 (3) 5996-8036 Fax +81 (3) 5996-8100
NIHON KOHDEN AMERICA, INC.
90 Icon Street, Foothill Ranch, CA 92610, USA Phone +1 (949) 580-1555 Fax +1 (949) 580-1550
NIHON KOHDEN EUROPE GmbH
Raiffeisenstrasse 10, 61191 Rosbach, Germany Phone +49 6003 827-0 Fax +49 6003 827-599
NIHON KOHDEN TRADING (SHANGHAI) CO., LTD
5F Hui Jin Business Mansion No. 259 Wuzhong Road, Shanghai 201103, China Phone +86 (21) 6270-0909 Fax +86 (21) 6270-9700
Hold the CO2 sensor as shown and insert the sensor into the
airway adapter until it clicks.
Refer to the operator’s manual of the CO2 sensor kit together
with this manual.
When removing the airway adapter, hold the CO2 sensor by the
ridges and pull it straight up.
NOTE
Do not pull the CO2 sensor cable. This may break the cable.
Options
Description Model Q’ty Code No.
CO2 sensor kit TG-970P 1 P909
CO2 adapter JG-970P 1 K987
CO2 sensor TG-221T 1 P924
Cleaning and Disinfecting the Airway Adapter
This airway adapter is disposable. You cannot clean or disinfect
it. Immediately replace the airway adapter with a new one when
it becomes dirty.
Expiration Date of the Airway Adapter
This airway adapter is disposable. The expiration date is 36
months from the date of manufacture.
Disposing of the Airway Adapter
Follow your local laws for disposing of infectious medical waste.
Specifications
Operating Environment
Temperature: 10 to 40°C (50 to 104°F)
Humidity: 30 to 85% (noncondensing)
Atmospheric pressure: 700 to 1060 hPa
Storage Environment
Temperature: –20 to +65°C (–4 to +149°F)
Humidity: 10 to 95%
Atmospheric pressure: 700 to 1060 hPa
• The Japanese manual is published according to
guidelines of the Health, Labor and Welfare Ministry in
Japan. Content of the English and Japanese manuals
may differ.
• The entire contents of this manual are copyrighted by
Nihon Kohden. All rights are reserved.