The OMNIS Liquid Adapter (6.01600.010) is a quick-change coupling
for chemical bottles with the following sealing mechanisms:
■Bottle cap, single-use (mounted on purchased chemical bottles)
■Bottle cap, multi-use (6.01600.100)
It takes only a single hand movement to move the OMNIS Liquid Adapter
from one bottle to the next. Cable connections and tubing connections
remain in place.
The bottle cap, single-use is equipped with an RFID tag on which the data
of the chemicals in the bottle is stored. The OMNIS Liquid Adapter has an
RFID reader that reads out this data and transfers it to the software.
1 Overview
1.2 Overview
Figure 1
Cable
1
RFID reader
3
Tubing adapter
5
part of 6.01600.xxx
Liquid Adapter - Parts
2
4
Status display
Aspiration tube
part of 6.01600.xxx
OMNIS Liquid Adapter
■■■■■■■■
1
Page 8
2.1 General safety instructions
2 Safety
TODO
Work on this Safety chapter is not yet finished
2.1 General safety instructions
To use the OMNIS Liquid Adapter safely, you must be aware of the fundamentals of safe working practices in laboratories. You must be familiar
with the safety apparatus in your laboratory and know how to use it.
Observe the following safety instructions for safe operation of the OMNIS
Liquid Adapter.
CAUTION
■■■■■■■■■■■■■■■■■■■■■■
Problems with leak-tightness
The leak-tightness of the OMNIS Liquid Adapter can no longer be guaranteed if the sealing lip or the sealing surface of
the aspiration tube is soiled or damaged.
The sealing lip and the sealing surface of the aspiration tube
must always be kept clean and intact.
■Check the status of the aspiration tube every time a bot-
tle is replaced.
CAUTION
Damage to the electronics
The OMNIS Liquid Adapter is not resistant to splash water.
Water can seep into the interior during cleaning and cause
damage to the electronics.
■Do not clean the OMNIS Liquid Adapter under running
water.
■Do not use a wash bottle to clean the OMNIS Liquid
Adapter.
■■■■■■■■
2
OMNIS Liquid Adapter
Page 9
■■■■■■■■■■■■■■■■■■■■■■
3 Additional information
TODO
The definitive Internet address of the Metrohm Information
portal has not yet been established.
Additional information concerning the OMNIS Liquid Adapter can be
found:
■in the Help section of the software
■in the Metrohm Information portal on the Internet Path Information
portal
3 Additional information
OMNIS Liquid Adapter
■■■■■■■■
3
Page 10
4.1 Instrument
4 Technical specifications
4.1 Instrument
High-frequency
part
Frequency13.56 MHz
TechnologyISO/IEC 14443 A, MIFARE, inductive
DistanceUp to 50 mm
Connector
Plug connectionRound plug, size 00, 4-pin, locked
Power supply
Supply voltage10 V - 27 V DC voltage
Power consumption
Approx. 4 Wp
■■■■■■■■■■■■■■■■■■■■■■
Status displayMulti-colored, can be configured via software
4.2 Ambient conditions
Nominal function
range
Temperature+5 - +45°C
HumidityMax. 85%
Storage and transport
Temperature-40 - +70°C
■■■■■■■■
4
OMNIS Liquid Adapter
Page 11
■■■■■■■■■■■■■■■■■■■■■■
5 Declaration of conformity
5 Declaration of conformity
5.1 Safety specifications
This instrument fulfills the following electrical safety requirements:
CE marking in accordance with the EU directives:
■2014/35/EC (Low Voltage Directive, LVD)
■2014/30/EC (EMC Directive, EMC)
■2014/53/EC (Radio Systems (RED))
Design and testingAccording to EN/IEC 61010-1, protection class III, EN/IEC 60950-1, EN/
IEC 60529, degree of protection IP20.
Safety instructionsThis document contains safety instructions which have to be followed
by the user in order to ensure safe operation of the instrument.