Liquid Controls HML4-F1 User Manual

HML4-F1 Converter
LCMag Electromagnetic Flowmeters
Installation & Operation
Release number: 4F1_EN_LC_R1_3_9X
Liquid ControLs Group
An IDEX Fluid & Metering Business
IOM: IEM200-20
HML4-F1
2
INDEX
q
Introduction _______________________________________________ pag.3
q Safety informations ___________________________________ pag.3
§ Safety conventions __________________________________________ pag.4
q Technical characteristics ____________________________________ pag.5
§ Electrical characteristics ______________________________________ pag.5
§ Environmental use conditions _________________________________ pag.5
§ Ambient temperature ________________________________________ pag.5
§ Data plate _________________________________________________ pag.5
§ Overall dimensions __________________________________________ pag.6
q Electrical connections __________________________________ pag.7
§ Grounding instructions _______________________________________ pag.7
§ Power supply converter ______________________________________ pag.7
§ Input/output cables _________________________________________ pag.8
§ Sensor cables ______________________________________________ pag.9
q Inputs/outputs ______________________________________
pag.10
§ Optional expansion modules __________________________________ pag.11
§ Digital Input _______________________________________________ pag.12
q Flags and led interpretation ____________________________
pag.16
q Flow rate visualization ________________________________ pag.1
7
q Access codes ________________________________________
pag.17
q Access to the configuration menu________________________
pag.18
q Visualization pages ___________________________________
pag.19
q Quick start menu ____________________________________
pag.20
q Examples __________________________________________
pag.21
q Programming functions _______________________________ pag.2
3
q Functions description _________________________________
pag.25
q Batch _____________________________________________ pag.3
3
§ Enable batch _______________________________________________ pag.33
§ Programming batch _________________________________________ pag.33
§ Start/stop batch ____________________________________________ pag.34
§ Important notes ____________________________________________ pag.34
q Alarm messages _____________________________________ pag.3
5
§ Causes and action to be taken _________________________________ pag.35
§ Anomalies codes ____________________________________________ pag.35
q Conformity declaration ________________________________ pag.3
9
HML4-F1
3
INTRODUCTION
These operating instructions and description of device functions are provided as part of the scope of supply. They could be modified without prior notice. The improper use, possible tampering of the instrument or parts of it and substitutions of any components not original, renders the warranty automatically void. The converter is to be used only to be coupled directly with a sensor device to realize a measuring device, flow meter, that is to be used for measuring the flow of conductive liquids in closed pipes with conductivity greater than 5µS/cm.
The reproduction of this manual and any supplied software is strictly forbidd
en.
S
AFETY INFORMATIONS
Any
use other than described in this manual affects the protection provided by the manufacturer and compromises the safety of people and the entire measuring system and is, therefore, not permitted. The manufacturer is not liable for damaged caused by improper or non-designated use.
1) Transport the measuring device to the measuring point in the original packaging. Do not remove covers or caps until immediately before installation. In case of cartons packaging it is possible to place one above the other but no more than three cartons. In case of wooden packaging do not place one above the other.
2) Disposal of this product or parts of it must be carried out according to the local public or private waste collection service regulations.
3) The converter must only be installed, connected and maintained by qualified and authorized specialists (e.g. electrical technicians) in full compliance with the instructions in these Operating Instructions, the applicable norms, legal regulations and certificates (depending on the application).
4) The specialists must have read and understood these Operating Instructions and must follow the instructions it contains. The Operating Instructions provide detailed information about the converter. If you are unclear on anything in these Operating Instructions, you must call the service.
CONVERTER
SENSOR
HML4-F1
4
5) The converter should only be installed after have verified technical data provided in these operating instructions and on the data plate.
6) Specialists must take care during installation and use personal protective equipment as provided by any related security plan or risk assessment.
7) Never mount or wire the converter while it is connected to the power supply and avoid any liquid contact with the instrument’s internal components. To connect remove the terminals from the terminal block.
8) Before connecting the power supply check the functionality of the safety equipment.
9) Repairs may only be performed if a genuine spare parts kit is available and this repair work is expressly permitted.
10) For the cleaning of the device use only a damp cloth, and for the maintenance/repairs contact the service center.
Before starting up the equipment please verify the following:
q Power supply voltage must correspond to that specified on the data plate q Electric connections must be completed as described q Ground (earth) connections must be completed as specified
Verify periodically (every 3-4 months):
q The power supply cables integrity, wiring and other connected electrical parts q The converter housing integrity q The suitable tightness of the sealing elements q The mechanical fixing of the converter to the pipe or wall stand.
SAFETY CONVENTIONS
DANGER
W
arning indicates an action or procedure which, if not performed correctly, can result in injury or a safety hazard. Comply strictly with the instructions and proceed with care.
WARNING
Caution indicates an action or procedure which, if not performed correctly, can result in incorrect operation or destruction of the device. Comply strictly with the instructions.
NOTES
Note indicates an action or procedure which, if not performed correctly, can have an indirect effect on operation or trigger an unexpected response on the part of the device.
HML4-F1
5
TECHNICAL CHARACTERISTICS
ENVIRONMENTAL USE CONDITIONS
AMBIENT TEMPERATURE
Da
ta plate
On the data plate there is some technical information:
Conv. Mod.: converter Model Conv. s/n: converter Serial Number P.S.: power supply Power: maximum power consumption Sensor Mod.: serial number of sensor coupled Sens s/n: serial Number of the sensor DN: sensor nominal diameter PN: sensor nominal pressure IP: Protection grade Fittings: process connections Lining: sensor lining Max Temp: maximum sensor temperature Electrodes: sensor electrodes KA: sensor KA ITEM: free for user
Min. *
Max. 1
°C
-20
60
°F
-4
140
ELECTRICAL CHARACTERISTICS
Converter classification: class I, IP 67,
installation category (overvoltage) II, rated
pollution degree 2.
Power supply
versions
Power supply
voltage
Max
power
LLV
18-30V
10W
q Voltage variations must not exceed ±10% of the nominal one. q Input/output are insulated up to 500V q The output 4-20 mA (optional) is electrically connected
q
The converter can be installed inside or outside buildings
q Altitude: from –200 to 2000m (from -656 to 6560 feet)
q Humidity range: 0-100% (IP 67)
Consider the ambient temperatures more restrictive between converter and sensor.
*
= For discontinuous use, the installation of a heating resistance is necessary.
1
= For liquid temperature > 60°C , ambient temperature max 40°C
130°C Cip (cleaning in place) is allowed up to 30’ (unlimited time if converter switch off)
HML4-F1
6
65
53
107
89
177
OVERALL DIMENSIONS
Weight: 2kg
CABLE GLAND
TORQUES
To guarantee the housing’s IP degree the following torques are required
:
q cable glands: 1.8Nm (in reference conditions)
HML4-F1
7
-
+
GROUNDING INSTRUCTIONS
ELECTRICAL CONNECTIONS
ALWAYS ensure that the converter and the sensor are grounded (earthed) correctly. The grounding of the sensor and converter ensures that the equipment and liquid are equipotential.
CONVERTER POWER SUPPLY
q
Before connecting the power supply, verify that the
mains voltage is
within the limits indicated on data
plate.
q ATTENTION:
the converters on dc power supply line
are not protected against the inversions of polarity.
q
For the connections use only approved conductors, with
fire-
proof properties, whose section varies from
0.25mm2 to 2.50mm
2
, based on distance/power;
additionally fix the
power supply wires with a additional
fastening system located close to the terminal.
q
The power supply line must be equipped with an external protection for overload current (fuse or
automatic line breaker).
q Provide in close proximity the converter
a circuit
breaker easily accessible for the operator and clearly
identified;
whose symbols must conform to the
electrical safety and local electrical requirements.
q
Ensure that the component complies with the requirements of the standard for
electrical safety distance.
q Check chemical compatibility of
materials used in the connection security systems in
order to minimize electrochemical corrosion.
q
The sensor, hardwired inputs and outputs are connected to the converter through a
terminal blocks located inside the converter.
q To locate terminal blocks loosen the 2 covers, front and
back. The terminal block is
the hardwire connection of the converter to external equipment, including the
sensor.
The following pages give informations on the terminal block numbering, and the respective connecting of the sensor cables, and inputs/outputs.
HML4-F1
8
+
-
OUTPUT 4-20
"A" RS485 "B" RS485
JACK FOR IF2X
PROFIBUS
PROFIBUS
IN / OUT CABLES
15
16
14
17
18
19
20
21
22
23
24
25
26
- INPUT 1
+ INPUT 1
E. OUTPUT 4- INPUT 3
C. OUTPUT 4+ INPUT 3
E. OUTPUT 2
C. OUTPUT 2
E. OUTPUT 3
C. OUTPUT 3
- INPUT 2
+ INPUT 2
E. OUTPUT 1
C. OUTPUT 1
MAIN POWER
+
-
NOTE: shielded cables are recommended for inputs and outputs wiring.
INPUT/OUTPUT CABLES
HML4-F1
9
CABLES FROM SENSOR
13
12
11
1
2
3
4
COMMON ELECTRODE 2 ELECTRODE 1
COIL 1 COIL 2
SENSOR CABLES
Dangers voltage:
- 60V max
- 250V max commutation coils circuit
HML4-F1
10
17 (Out 1) 21 (Out 2)
18 (Out 1) 22 (Out 2)
+
-
INPUTS/OUTPUTS
Output 1-2
20 (Out 3) 24 (Out 4)
19 (Out 3) 23 (Out 4)
47K
5V1-1/2 W
+
-
Output 3-4/Input 2-3
OUTPUTS technical characteristics
INPUTS technical characteristics
OUTPUTS technical characteristics
25
47K
5V1-1/2 W
26
+
-
Input 1
INPUTS technical characteristics
PROTECTIONS
The terminals are protected for the
accidental polarity inversion; the output is protected
from the extra voltage due to the inductive loads (connections of coils or relay).
q Max voltage 30V
q Max load: 100mA @24V , q Max frequency: 1250Hz.
q Max voltage 30V
q Min voltage for recognized logical state
enables: 18V
q Entry resistance of the circuit: 47kΩ
q Max voltage 30V
q Max load: 100mA @24V , q Max frequency: 1250Hz.
q Max voltage 30V
q Min voltage for recognized logical state
enables: 18V
q Entry resistance of the circuit: 47kΩ
HML4-F1
11
4-20mA OUT
mA
14
I
external supply
+
-
Output 4-20mA
14 15 16
A B
RS 485
or
PROFIBUS DP
+
-
MODULE ME35/36
OPTIONAL
FOR COMMUNICATION MODULE SEE THE SPECIFIC MANUALS
OPTIONAL EXPANSION MODULES
q Power supply 18-30V
q Max load resistive to the output: 800
with 24V of power supply
q Minimum load recommended 500 q Setting time: 5ms
ATTENTION: The power supply of communication module is the same of that of entire instruments, therefore is not insulated from it.
Terminal block M2
ATTENTION: The power supply of 4­20mA is the same of that of entire instruments, therefore is not insulated from it.
HML4-F1
12
DIGITAL INPUT
The functions refer to the inputs could be substantially divided in three groups:
1.
only assignable functions to the input 1
2.
functions that act directly on the inputs independently from the select input
3. only assignable functions to the input 2 and only to the input 3 which they
interact between them
Remember that the activation of any functions of batch automatically disable the operation o
f other.
The list of such functions is suitable in the table of the batch
function configuration section.
5/33 VDC (ON) 0/1.5 VDC (OFF)
25 (+)
26 (-)
10 K
Input electric wiring
ATTENTION: where not specified the time T must be to Tmin
Speed rate
Tmin
20 Hz 50 Hz
80 Hz
150 Hz
110 ms
45 ms 30 ms
15 ms
60 Hz 40 ms
HML4-F1
13
0-1.5 V
3-40 V
AUTOCALIB. OFF
T
RESET
0-1.5 V
3-40 V
Tmin = 100ms
BLOCK
0-1.5 V
3-40 V
T
0-1.5 V
3-40 V
INPUT OPERATION STAGE (GENERIC FUNCTIONS)
Auto-calibration
Reset totalizers
Block totalizers
Block totalizers
Totalizers active
Necessary conditions for enable the function
Necessary conditions for enable the function
Necessary conditions for enable the function
Tmin<T<1sec. = autocalibration
T > 1 sec. = Autozero
POS. 5.5 ENABLED
POS. 5.7-8-9 batch functions assign to input 1-2-3
(optional) DISABLED
(op
POS. 6.1-2-3-4 batch functions assign to output 1-2-
3-4 DISABLED
POS. 5.1-2 ENABLED at least one
POS. 5.4 ENABLED
POS. 9.5 (auto-batch) DISABLED POS. 9.7 (cons. mode) DISABLED
POS. 5.6 ENABLE
POS. 5.7 (batch on input 1) DISABLED POS. 5.8-9 batch functions assign to input 2-3
(optional) DISABLED
Necessary conditions for enable the function
Range change
THE FUNCTIONS POINT OUT ABOVE ARE ENABLED ONLY ON INPUT 1
Scale 2
Scale 1
HML4-F1
14
T
T
OPERATION STAGE ON INPUT 1 OR 2 OR 3 (BATCH FUNCTION)
Start batch from remote input
Start batch from consent (remote)
Start batch from remote input with auto-
batch enabled
POS. 5.7 ENABLE or POS. 5.8 or 5.9 on batch
POS. 6.1-4 on end batch
Start batch from remote input with
automatic selection of formula 00/03
N.B.: THE ACTIVATION OF BATCH FUNCTIONS ON INPUT 2
PREVENTS THE ACTIVATION OF BATCH FUNCTIONS ON INPUT 1
Necessary conditions for enable the function Necessary conditions for enable the function
POS. 5.7 ENABLE or POS. 5.8 or 5.9 on batch
POS. 6.1-4 on end batch POS. 9.7 (CONSENT MODE) ENABLED
POS. 5.7 ENABLE or POS. 5.8 or 5.9 on batch
POS. 6.1-4 on end batch POS. 9.5 (auto-batch) ENABLED
POS. 9.7 (consent mode) DISABLED
Necessary conditions for enable the function
T= 100ms ±50ms for select the formula 00
T= 200ms ±50ms for select the formula 01 T= 300ms ±50ms for select the formula 02 T= 400ms ±50ms for select the formula 03 In case of stop batch from remote input the time of input pulse must be > 50ms
POS. 5.7 ENABLE or POS. 5.8 or 5.9 on batch
POS. 6.1-2 on end batch POS. 9.6 (BM auto sel) ENABLED POS. 9.7 (consent mode) DISABLED POS. 5.8-9 functions assigned to BM
SELECT
Necessary conditions for enable the function
Start batch from remote input 1 reset p+
enabled on remote input 1
T BEETWEEN 1 E 4 = RESET TOTALIZER
T<1 = START E RESET TOTALIZER
Necessary conditions for enable the function
POS. 5.7 (batch on input 1) ENABLED
POS. 6.1-2 on AND BATCH POS. 5.2 (reset T2) ENABLED
0-1.5 V
3-40 V
INPUT
OUTPUT
Closing valve
Opening valve
BATCH
INPUT
OUTPUT
0-1.5 V
3-40 V
Opening valve
Opening valve
Closing valve
BATCH
3-40 V
0-1.5 V
5 Sec.
<5 Sec.
INPUT
OUTPUT
Closing valve quantity to batch memorized
Opening valve
Closing valve
AUTO-BATCH
BATCH
3-40 V
0-1.5 V
Formula selection
Start
INPUT OUTPUT
Opening valve
Closing valve
BATCH
INPUT OUTPUT
3-40 V
0-1.5 V
Start
Opening valve
Closing valve
BATCH
Start
HML4-F1
15
T
T T
OPERATION STAGE ON INPUT 1 and 2 or 1 and 3 (BATCH FUNCTION)
Start batch on remote input 1 stop from output
selection formula 00 o 01 from remote input 2
Block totalizer from remote input 1 start
batch from remote input 2
Block and reset totalizer from remote input 1
start batch from remote input 2 consent mode
to batch enable
Necessary conditions for enable the function
POS. 5.7 ENABLED
POS. 6.1 or 6.3 on END BATCH POS. 5.8 or/and 5.9 on BM SELECT
The block of totalizer always determine
the interruption of the batch. Exciting again the input 2 or 3 is been able to get 2 results:
1) T< 1Sec = restart interrupted batch
2) T between 1 e 4 Sec = reset interrupted batch. N.B.: will be necessary give a new impulse of start to the input 2 (T< 1Sec) for begin a new batch
Necessary conditions for enable the
function
POS. 5.4 (Count lock) ENABLE
POS. 6.1 or 6.3 on END BATCH POS. 5.8 or 5.9 on BATCH POS. 5.2 (T2) ENABLE
The block of the totalize always determines
the stop of batch. With the function of reset T2 enable on descent front of input 1 comes reset the totalize of the batch in course. therefore the presence of the consent or a new pulse on the input 2 or 3 they will determine the start of a new batch
Necessary conditions for enable the function
POS. 5.4 (Cout lock) ENABLE
POS. 5.8 or 5.9 on BATCH POS. 9.7 (consent mode) ENABLE POS. 5.2 (T2) ENABLE
3-40 V
0-1.5 V
Start
Opening valve
Closing valve
Opening valve
Closing valve
BATCH FORMULA 00
BATCH FORMULA 01
BATCH FORMULA 00
BATCH FORMULA 01
3-40 V
INPUT 2
OUTPUT
INPUT 1
3-40 V
0-1.5 V
Start
Start
Opening valve
Closing valve
Opening valve
Closing valve
NEW BATCH
RESTART INTERRUPTED
BATCH
OUTPUT
OUTPUT
INPUT 2
INPUT 2
INPUT 1
T<1 Sec.
T compreso tra 1 e 4 Sec.
3-40 V
0-1.5 V
Start
RESET
T<1 Sec.
Start
Opening valve
Closing valve
Opening valve
Closing valve
INTERRUPTED BATCH
0-1.5 V
3-40 V
BLOCK
0-1.5 V
3-40 V
Start
INPUT 2
Opening valve
Closing valve
Opening valve
Closing valve
0-1.5 V
3-40 V
BLOCK
RESET T2
INPUT 1
OUTPUT
INTERRUPTED BATCH
NEW BATCH
INTERRUPTED BATCH
HML4-F1
16
15
16
14
17
18
19
20
21
22
23
24 25
26
FLAGS INTERPRETATION
FLAG
DESCRIPTION
Alarm max/min activated
!
- Interruption coils circuit
- Signal error
- Empty pipe
C
Calibration running
S
Simulation
Pulse output saturation (reduce TIME PULSE)
FLAGS AND LED INTERPRETATION
FLAGS
LED
LED INTERPRETATION
PERMANENT LIGHT:
initialization
FLASHING
LIGHT ( 1 sec.): normal function
FLASHING
LIGHT (<1 SEC.): alarm on
The LED signals the real alarm status only if the display visualizes one of the visualization pages
HML4-F1
17
Functions in the converter main menu are enabled by the access codes. The information of this manual is related to all the functions available with the L2 level. All the functions available through higher level are protected and reserved to the service.
Description of the L2 access code
(menu “11 Internal data” pos. 11.1)
q with code L2 = 00000 (only with this code) you can disable the L2 code request q * with L2 customized
(freely chosen by the user) you can program all the functions
up to L2 security level, entering the code whenever you need enter the Main menu.
* ATTENTION:
take careful note of the customized code, since
there is no way for the
user to retrieve or reset it if lost.
Factory preset access codes
The converter is delivered with the
default L2 (level 2) access code:
11111
The instrument can show a 5 digit character display for flow rate units; this mean the maximum flow rate value that can be represented on the display is
99999
(no matter the
positioning of the decimal point). The minimum is
0.0025.
The representable measure unit depends from sensor flow rate and diameter; the permitted units are those, set the instrument full scale value, allow to be represented with a numerical field which the maximum value do not exceed
99999.
Example for DN 300:
- Full scale value: 3m/s
- PERMITTED measure unit (example): l/s (216.00); m3/h (777.60); m3/s (0.2160)
- NOT PERMITTED measure unit (example): l/h (777600)..
FLOW RATE VISUALIZATION
ACCESS CODES
HML4-F1
18
Functions can be accessed in two different ways:
q The Quick start menu makes direct access
possible to a range of principal
functions for setting the scale reading and display characteristics of the sensor.
Quick start menu can be enabled by function 8.4 (menu display)
q Through the Main menu
it is possible to access functions with the access code
level 2. These control the sensor’s monitoring, data processing, inpu
t/output as
well further options.
E
xamples in the next pages show how to change the Full scale by Quick Start menu
and
by the Main menu.
The converter configuration menu can be accessed ONLY using ISOCON interface (
cable and software needed).
Isocon is a Windows® software that allows to set all the converter functions and personalize the menu (IF2X is required), see suitable manual for details.
THE CONFIGURATION MENU
Factory pre-setting:
QUICK START MENU=ON
ACCESS TO THE CONFIGURATION MENU
ISOCON INTERFACE
HML4-F1
19
* The maximum digits by from the totalizer is 999999999 independently from the number of
selected decimal. Beyond this value the totalizers are reset.
Flow rate value
% full scale
Direct/reverse totalizer
Sampling rate
Three vertical bars
Flow direction +/-
See: anomalies codes and
flags interpretation
C=Calibration
S=Simulation
Flow speed
Scale (1=low); (2=high)
Flow rate value
% full scale
Measure units
Flow direction
Time scale
% full scale
Active scale
The three vertical bars represent the filters behaviors: first
bar, “F”, is the FAST filter, second bar, “S”, is the “SLOW”
filter and the third bar represents
the variation of the new
measure samples respect to the averaged value. It can give
an indication about the total noise of the measure.
*
VISUALIZATION PAGES
Push to change
visualization
HML4-F1
20
The set up functions (batching, regulation, flow and SMART measure) configure
automatically the instrument (filters) for the operation set modifying optimally the parameters related to the function chosen. To enable one of
four types operation press the key Enter while the function is
highlighted, to confirm the changes press the key Del/Canc or Esc. Fi
lters settings for different S.rate (sensor menu, not user modifiable):
Batching
Regulation
Flow measure
SMART measure
MFSLN
MFFLN
MFSLN
MFFLN
MFSLN
MFFLN
MFSLN
MFFLN
20Hz 0 0 5 0
10 0 50
5
50Hz 0 0 2 0 4 0
20
2
60Hz 0 0
1.7 0 3.3
0
16.7
1.67
80Hz 0 0
1.3 0 2.5
0
12.5
1.25
150Hz 0 0
0.7 0 1.3 0 6.7
0.67
See the functions description section for more details regarding filters.
QUICK START MENU
The user has immediate access to the Quick Start menu when the converter is powered up by pressing the key Enter. If access to the quick start menu does not occur, then it has been disabled through the function 8.4
See function menu section
Automatic parameters optimization (see below)
Access to all functions
HML4-F1
21
EXAMPLE: modifying the full scale value from 4dm³/s to 5dm³/s,
from the “Quick start menu”
Enter in the
Quick start menu
Access to the
function “Fs1”
Push
repeatedly
Change
the value
Confirm
the new value
Main page
HML4-F1
22
EXAMPLE: modifying the full scale value from 4dm³/s to 5dm³/s,
from the “Main Menu” (quick start menu enabled)
Enter in the
Quick start menu
Access to the
Main Menu
Access to the Scale menu
Access to the
function “Fs1”
Change the value
Confirm
the new value
X 5 TIMES
Push
repeatedly
Confirm the keycode
Main page
HML4-F1
23
1.1 Insert the Nominal Diameter of the sensor (0-3000mm)
1.2 Calibration data of sensor visualized on sensor label
1.3 Sensor model: enter the first two characters of the sensor serial number
1.4 Factory parameters
1.5 Enables the empty pipe detection feature
1.7* Value of empty pipe sensibility detection
1.8* Enables the automatic zero calibration system
1.6* Electrodes cleaning
2.1* Full scale value set for range N.1
2.2* Full scale value set for range N.2
2.3* Unit of measure and number of decimal totalizes
2.4* Pulse value on output 1
2.6* Duration of the pulse generated on output 1
2.7* Duration of the pulse generated on output 2
2.8 Specific gravity set in kg/dm³ (enable only if FS1 or FS2 are weigh/time)
2.5* Pulse value on output 2
3.3* Main frequency filter
3.5 Enable every hour an internal cycle of calibration. The measure is stopped for 8-15 sec.
3.6* Automatic change of scale
3.1 Low flow zero threshold: 0-25% of full scale value
3.2* Measure filter
3.4 Only for service purposes
4.1 Maximum value alarm set for direct flow rate
4.2 Maximum value alarm set for reverse flow rate
4.3 Minimum value alarm set for direct flow rate
4.4 Minimum value alarm set for reverse flow rate
4.5 Hysteresis threshold set for the minimum and maximum flow rate alarms
4.6* Current output value in case of failure
4.7* Batch safety timer
5.8* Functions assigned to input 2 (automatically disabled if OUT3 is enabled)
5.9* Functions assigned to input 3 (automatically disabled if OUT4 is enabled)
5.1* Total direct (positive) flow totalizer reset enable
5.2* Partial direct (positive) flow totalizer reset enable
5.3 Reset totalizer of pulse from digital input (see page 13)
5.4 Totalizer counting lock command (see page 13)
5.5* Autozero calibration external command
5.7 Batch start/stop external command (see batch functions)
5.6 Range change external command (see pos. 3.6)
Functions assigned on input 1
PROGRAMMING FUNCTIONS
(functions with access code < 3, those with symbol “*” see next section)
Functions in grey colour are visualized on display only with other active functions or with optional modules
HML4-F1
24
6.1* Output 1 functions
6.2* Output 2 functions
6.3* Output 2 functions
6.4* Output 4 functions
6.5* Choice of the function and the range of current output n.1
8.1 Choice of the language: EN= English, IT=italian, FR= French, SP= Spanish
8.2 Display totilizer mode
8.3 Updating frequency on the display: 1-2-5-10 Hz
8.4 Quick start menu visualization
8.5* Volume totalizer 1 reset
8.6* Volume totalizer 2 reset
9.1* Number of batch cycles to be done to define the value of compensation.
9.2* % limit of compensation threshold
9.3* Compensation value
9.4* Prebatch value
9.5* Auto-batch
9.7* Static consent of batch
9.6* Automatic selection of batch formula
7.1 Choice of the communication protocol for the IF2 device
7.2 Address value of the converter
7.3 Speed of the RS485 output (possible choices: 4800, 9600, 19200, 38400 bps)
7.4 Instrument answer delay
11.1 Level 2 access code enter
11.2 Load factory data pre-set
11.3 Load user data saved
11.4 Save user data
11.6 Ks coefficient (only for service purposes)
11.5 Ignore the calibration error during the switch on test
10.1* Enable the calibration of the converter
10.2* Converter autotest
10.3* Flow rate simulation enabling
10.4 Stand-by of converter to reduce the consumption during service operation
10.5 Firmware revision/version
Menu 9 visualized only IF batch is active
HML4-F1
25
MENU 1.SENSOR
(POS. 1.1) Nominal diameter of sensor
[ND= XXXX]
FUNCTIONS DESCRIPTION
(description of the functions with access code< 3)
The following pages give a description of the most important functions and how they can be changed or enabled/disabled by the user.
MENU 1 - SENSOR
(POS. 1.6) Electrodes cleaning [E. cleaning=MIN]
Selectable options: OFF, minimum, average and maximum. It is not recommended the use of this function (let OFF) when the liquid has a conductivity less than 100µS/cm.
(POS. 1.7) Empty pipe threshold [E.p. thr.=XXX]
This value represents the threshold of measure block to zero in empty pipe condition; the span function is 20-250. Since the sensibility of empty pipe detection could be changed by the liquid conductivity, ground connections, type of lining, in case of needed it is possible to adjust manually the preset threshold. It is recommended to check periodically its proper functioning. Value increasing means sensibility decreasing.
(POS. 1.8) Autozero calibration [Autozero cal.]
This function activates the zero calibration automatic system. To perform this operation, it is necessary for the sensor to be full of liquid, and the liquid be perfectly still. Even minimal movement of the liquid may affect the result, and consequently, the accuracy of the system. Once you are sure the above conditions have been fulfilled (and the percentage flow rate value is stable) press the key and automatically activates the zeroing procedure, after which it is necessary to check that the percentage value showed goes to zero, repeat the operation otherwise. When the value is stable at zero, then press the key Enter.
MENU 2 - SCALES
(POS. 2.1-2) Full scale n° 1-2 [FS1-2=
dm³/S
X.XXXX]
The full scale is used to indicate to the maximum meter’s flow rate; a volume per time is required.
The full scale should be chosen carefully as it’s parameters are used for several other parameters. There are four fields to fill in order to set this parameter, from left to right: 1) volume unit of measure, 2) type of unit, 3) time unit of measure and 4) numeric value. The selection is made by positioning the cursor on the field to modify. To change the type of unit of measure (metric, British
or American, mass or volume) the cursor has to be positioned on the symbol “/” (field N. 2). When
the nominal diameter is set to zero it is possible to modify only the numeric field, since the unit of measure stays at m/sec. The following tables show the units of measure available and the conversion factor by comparison with 1 dm3 and 1 kg. The converter accepts any kind of co
mbination of units of measure satisfying both the following conditions:
1. Numeric field value
99999
2. 1/25 fs
max
numeric field value fs
max
.
Menu visualized on the converter (from 1 to 11)
Synthetic description of the function
Identification of the function (not visualized on display)
Converter request
HML4-F1
26
Where fs
max
is the maximum full scale value corresponding to the sensor, equal to a 10 m/sec liquid speed. The units of measure are shown as appear on the display. The British and American units are diversified by using capital and small characters. Available units of mass and volume:
When a mass unit of measure is set, the specific gravity function is automatically enabled by the system. The units of measure of time may be chosen among the values: s=second, m=minute,
h=hour, d=day.
(POS. 2.3) Measure unit and number decimal totalizers [MU.tot:dm³X.XXX]
Setting the unit of measure and number of decimals for visualized the totalizers or the volumes to batch. For set the unit of measure, position the cursor on field of the actual unit of measure; to set the type of unit, position the cursor on the blank space between the unit of measure and the numeric value; to set the number of decimal totalizers position the cursor on numeric field and choose one of the possible combinations: 1000-01.00-001.0-00001.
(POS. 2.4-5) Pulse value output 1-2 and unit of measure of tot. [Imp1-2=
dm³
X.XXXXX]
Setting of the pulse volume corresponding to output 1-2 and of the totalizers measure units. There are three fields to fill in to set this parameter, from left to right: 1) measure unit, 2) unit type and
3) numeric value. The selection is performed by positioning the cursor on the field to be modified. To change the unit type (metric, British or American, mass or volume) just position the cursor on the blank space between the measure unit and the numeric value. When the nominal diameter is set to zero it is possible to modify only the numeric field since the measure unit stays at meter (m) or feet (ft). The possible measure units are those above described
(POS. 2.6-7) Pulse duration on output 1-2 [Tpul1-2=msXXXX.XX]
Setting of the duration of the pulse generated on output 1-2. Its value is expressed in milliseconds and has to be between 0.4 and 9999.99. When the high frequency output is present, then the minimum value can go down to 0.04 milliseconds.
ATTENTION
: since the instrument cannot detect which type of device it is connected to, it is up to the user to verify the set pulse duration is compatible with the external device processing such pulses. If, for example, an electro-mechanical pulse counter is connected, then two kind of problems may occur: if the pulse is too long than the coil may burn or, if it is too short, the counter may not be able to count and eventually even cause the damaging of the output itself.
cm3
Cubic centimetre
ml
Millilitre
l
Liter
dm3
Cubic decimeter
dal
Decalitre
hl
Hectolitre
m3
Cubic metre
in3
Cubic inch
Gal
American gallon
GAL
British gallon
ft3
Cubic foot
Bbl
Standard barrel
BBL
Oil barrel
yd3
Cubic yard
kgl
KAmerican gallon
KGL
KBritish gallon
G
Gram
Kg
Kilogram
T
Ton
Oz
Ounce
Lb
Pound
Ton
short tons
HML4-F1
27
MENU 3 – MEASURE
(POS. 3.2) Measure filter [M. filter=s XXX.X]
This section of the manual is extremely important because the right settings of filters allow to get a proper meter reaction according to the flow rate behaviour.
WORKING PRINCIPLE
There are two filters:
MFSLN (Measure Filter Slow LeNght): available via ETP and through the display, in menu
3. Measure, function “M. filter”
The M. filter function effects the converter response time while measuring (it is like LOW PASS filter); the value needs to be chosen carefully as it effects the meter behaviours. A smaller value with a fluctuating flow will lead to unstable flow readings as the integration filter will be measuring over very short periods. A higher value corresponds to a more stable but slower measure, a smaller value gives a faster measure but is more unstable.
MFFLN (Measure Filter Fast LeNght): is available ONLY via ETP
The MFFLN must be set by the user in accordance with the MFSLN (M. filter) to meet the required meter behaviour. How the system functions: If the period of flow rate variation is over the MFFLN time (i.e. a slow flow variation), the outputs (display/4-20mA/Alarms) are refreshed according to MFSLN (M. filter); as soon as the flow rate variation is within MFFLN time, the output will be set immediately at the new value.
The combination of MFSLN & MFFLN is useful as it allows the converter to react quickly to large flow rate variations, regardless of the MFSLN (M. filter) set, therefore delaying (dampening) smaller fluctuations in flow, resulting in very stable measures. The value of those parameters has to be within the range from 0 (filter disabled) to the maximum value defined on below table:
Sampling rate
M Filter (MFSLN)
F Filter (MFFLN)
20
50.0
7.50
50
20.0
3.00
60
16.7
2.50
80
12.5
1.88
150
6.7
1
HML4-F1
28
10
20
30
40
50
60
0
0 2 4 6 8 10 12 14 16 18 20 22
MFSLN: 0,0s MFFLN: 0,0s
FLOW RATE MEASURE
FILTERS
>>>>
>>>>
10
20
30
40
50
60
0
0 2 4 6 8 10 12 14 16 18 20 22
10
20
30
40
50
60
0
0 2 4 6 8 10 12 14 16 18 20 22
1,0sec.
1,0sec.
1,0sec.
1,0sec.
1,0sec.
5,0sec.
IN
OU T
MFSLN: 5,0s MFFLN: 1,0s
FLOW RATE MEASURE
FILTERS
>>>>
>>>>
IN
OU T
MFSLN: 5,0s MFFLN: 0,0s
FLOW RATE MEASURE
FILTERS
>>>>
>>>>
IN
OU T
FLOW RATE (%)
TIME (Sec)
TIME (Sec)
FLOW RATE (%)
TIME (Sec)
FLOW RATE (%)
The following diagrams show the response of the instrument for a flow rate variation from 0 to 100% of full scale set using different settings of filters.
Both filters to zero
(batching setup)
: the measure quickly follows the flow rate
behaviour
Both filters are used (Smart mode): prompt reaction to quick flow variation (based on the MFFLN value); smoothing the measure result to slow flow variations (based
on MFSLN value)
Only M. filter (MFSLN)
is u
sed (regulation
and flow measure
setup)
: smoothing the measure result (based on MFSLN value) in reaction to quick and slow flow variation
(MFFLN does no
t
work).
HML4-F1
29
(POS. 3.3) Measure filter AC enable [AC Line fil.=ON/OFF]
This function allow to filter electrical noise on the liquid coming from the main frequency. (POS. 3.6) Automatic scale change enable [Autorange=ON/OFF] Enables the automatic change of scale. The meter may have two different working ranges in order to suit to the variable process conditions. In order to get the best results out of this function it is important range N.2 is bigger than N.1. When the flow rate increases and reaches the 100% of the full scale 1, then the meter automatically switches to scale 2. When the flow rate decreases again reaching a value on scale 2 equal to the 90% of full scale N.1, then the active scale is 1 again. Allowed values for this parameter: ON/OFF. N.B.: the autorange does not allow using the manual change of range (pos. 5.6)
MENU 4 - ALARMS
(POS. 4.6) Current output value in case of failure [mA v.fault =% XXX]
Setting of the value the 4-20mA current output has to be in one of the following cases: empty pipe; coils interrupted; ADC error The allowed range is from 0 to 120% of the 0-20 mA scale, 120% corresponds to 24mA and does not depend on the selected range (0-20/4-20mA). The NAMUR NE43 recommendation asks for a alarms signalling value for the current output lower than 3.6mA (<18%) or bigger than 21mA (>105%). It would then be preferable to set the value of this function at the 10%, so that the current value in case of the a.m. cases would be 2mA, allowing the following diagnostics:
1. current < 2 mA - 5%: line interrupted, power supply failure or faulty converter;
2. 2mA -5%
current 2mA + 5%: hardware alarm;
3. 4mA current 20mA: normal working range;
4. 20mA < current 22mA: out of range, measure above 100% f.s.
N.B.: To set this parameter to zero corresponds to disable the alarm (POS. 4.7) Batch safety timer [Timeout=s XX.X]
With this function is possible to set a maximum time for the batch; over this time will be generate an alarm. This function is useful when you need control one or both of the followings condition:
q batch valve open and flow rate is zero q batch valve closed and flow rate different to zero
When this alarm is activated, they are aborted the batch operation and the power supply of the valve. The values of function are from 0 to 25.5 seconds and is active only if one or more of the batch functions are enable.
MENU 5 - INPUTS
(POS. 5.1-2) Modify/reset totalizer enable [T1/2 reset=ON/OFF]
When one of this function is enabled, the related totalizer may be reset through the on/off input. (POS. 5.5) Autozero calibration external command enable [Calibration=ON/OFF] When this function is active, applying a voltage on the on/off input terminals the meter performs a autozero calibration cycle. ATTENTION: if the voltage pulse is less 1 sec., the meter performs a calibration cycle for compensate possible thermal drifts. If the voltage pulse is more 1 sec, the meter performs a zero calibration of measure. This function enables/disables the automatic zero calibration system. To perform the sensor it is absolutely necessary the sensor is full of liquid and that the liquid is perfectly staying still. Even very small movement of the liquid may affect the result of this function, and, consequently, the accuracy of the system. (POS. 5.8-9) Functions assigned to input 2-3 [INP.2-3=XXXXXX] Choice of the function to associate the input 2 The functions are listed in the table below.
FUNCTIONS FOR INPUTS 2-3
OFF:
DISABLE
BATCH
: START/STOP BATCH
BM SELECT:
STATIC SELECTION OF FORMULA
VALV.
OPEN: OPEN VALVE COMMAND
STAND
-BY: STAND-BY COMMAND (see function 10.4)
HML4-F1
30
MENU 6 - OUTPUTS
(POS. 6.1-4) Function corresponding to on/off output 1-2-3-4 [OUT1=XXXXXX]
Choice of the function corresponding to digital Output 1. The functions are listed in the table below:
(POS. 6.5) Function and the range of current output n.1 [OUT.mA1=X÷XX+] Choice of function and range of current output N.1. There are three fields to modify:
q Scale zero: 4 or 0 mA; Full scale: 20 or 22 mA q Field: + = positive, - = negative, blank = both, -0+ = central zero scale
The values corresponding to the scale points are shown in the following chart:
In hardware alarm conditions “HW ALARM” (interrupted coils, empty pipe, measure error) the current value is programmed by the function “mA v. fault” (pos. 4.6) and it is expressed as
percentage of a fixed current range, where: 0% = 0mA e 110% = 22mA.
CURRENT VALUES IN mA ASSOCIATE TO THE % VALUE OF FULL SCALE
POSSIBLE FIELD
REVERSE FLOW VALUE
ZERO
DIRECT FLOW VALUE
-110%
-100%
0%
+100%
+110%
OutmA = 0 ÷ 20 +
0 0 0
20
20
OutmA = 0 ÷ 22 +
0 0 0
20
22
OutmA = 4 ÷ 20 +
4 4 4
20
20
* OutmA = 4 ÷ 22 +
4 4 4
20
22
OutmA = 0 ÷ 20 -
20
20 0 0
0
OutmA = 0 ÷ 22 -
22
20 0 0
0
OutmA = 4 ÷ 20 -
20
20 4 4
4
OutmA = 4 ÷ 22 -
22
20 4 4
4
OutmA = 0 ÷ 20
20
20 0 20
20
OutmA = 0 ÷ 22
22
20 0 20
22
OutmA = 4 ÷ 20
20
20 4 20
20
OutmA = 4 ÷ 22
22
20 4 20
22
OutmA = 0 ÷ 20 –0+
0 0 10
20
20
OutmA = 0 ÷ 22 –0+
0 1 11
21
22
** OutmA = 4 ÷ 20 –0+
4 4 12
20
20
OutmA = 4 ÷ 22 –0+
4
4.8
12.8
20.8
22
22 mA
20 mA
-110% F.S. -100% F.S. +100% F.S. +110% F.S. zero
4 mA
I (mA)
I (mA) 20 mA
-110% F.S. -100% F.S. +100% F.S. +110% F.S. zero
12 mA
4 mA
FUNCTIONS FOR OUTPUTS 1, 2, 3, 4
OFF: DISABLED PLS+: PULSE FOR POSITIVE FLOW RATE (ONLY OUTPUT 1-2) PLS-: PULSE FOR NEGATIVE FLOW RATE (ONLY OUTPUT 1-2) PLS: PULSE FOR POSITIVE AND NEGATIVE FLOW RATE (ONLY OUTPUT 1-2) SIGN: FLOW DIRECTION OUTPUT (ENERGIZED = -) RANGE: RANGE INDICATION OUTPUT (ENERGIZED = SCALE 2) MAX AL+: MAX DIRECT FLOW RATE OUTPUT(ENERGIZED = AL. OFF) MAX AL-: MAX REVERSE FLOW RATE OUTPUT(ENERGIZED = AL. OFF) MAX AL: MAX DIRECT/REVERSE FLOW RATE OUTPUT(ENERGIZED = AL. OFF) MIN AL+: MIN DIRECT FLOW RATE OUTPUT(ENERGIZED = AL. OFF) MIN AL-: MIN REVERSE FLOW RATE OUTPUT(ENERGIZED = AL. OFF) MIN AL: MIN DIRECT/REVERSE FLOW RATE OUTPUT(ENERGIZED = AL. OFF) MAX+MIN±: MAX AND MIN FLOW RATE ALARM OUTPUT (ENERGIZED = AL. OFF) P.EMPTY: EMPTY PIPE ALARM OUTPUT (ENERGIZED = FULL PIPE) OVERFLOW.: OUT OF RANGE ALARM OUTPUT (ENERGIZED = FLOW RATE OK) HARDW AL.: CUMULATIVE ALARM OUTPUT interrupt coils, empty pipe, measure error (ENERGIZED = NO ALARMS EXT. COMM.: ONLY AVAILABLE WITH DATA LOGGER MODULE BATCH AL: BATCH ALARM BATCH SYN.: AT THE AND OF BATCH THE OUTPUT CHANGE STATUS END BATCH.: END BATCH OUTPUT (ENERGIZED =BATCH IN PROGRESS) PREBATCH.: PREBATCH OUTPUT (ENERGIZED = PREBATCH IN PROGRESS)
* Example 1: out 4÷22 +
** Example 2: out 4÷20 –0+
HML4-F1
31
MENU 8 - DISPLAY
(POS. 8.5-6) Totalizer 1-2 reset [T1/2 reset]
These functions are activated pressing the key Enter during the function visualization. When "EXECUTE?" is required, press Esc/Del/Canc to proceed; or any other key to delete the operation.
MENU 9 - BATCH Visualized only with batch active (output on batch and/or pos. 5.8 enable or pos. 5.9 on batch)
(POS. 9.1) Number of batch samples [N.SAMPLES=XXX]
Number of batch cycles to be done to define the value of compensation. This function allows to automatically determine the average value for automatic compensation of system delay (pos. 9.3). Set to ZERO this function for manually introduction of the compensation value. (POS. 9.2) % limit of compensation [HYST.=%XXX] This value defines the percentage of maximum difference between the compensation value set (pos. 9.3) and the average compensation value defined with the function 9.1. Over this threshold the new compensation value will be automatically set (if number of batch samples is not zero) (POS. 9.3) Compensation value [V.COM.=XX.XXX] This value, expressed in the same selected volume unit of measure, is the result of the difference between the batch value set and the quantity of product really supplied due to the system delays: closing valves, stop pumps, stop motors, etc. Attention: if you need to set manually the value of compensation, preset to ZERO the Number of batch samples (pos. 9.1) (POS. 9.4) Prebatch value [V.PRE.=XX.XXX] set the volume of liquid at which you want to enable the pre-batch. When the pre-batch volume “V Pre” is reached the output (if enabled) is de-activated. This value is constant for all quantities to be batched and must be set in current volume unit of measure. The pre-batch function is useful when you need fast and accurate fillings.
(POS. 9.5) Enable/disable auto-batch function [AUTO BATCH=ON/OFF]
Applying a voltage on the on/off input terminals for more than 5 second the valve controlled by the meter stands open while the voltage is applied on the input. When the product has reached the desired volume/level, removing the voltage from the input, the meter closes the valve and memorizes the supplied product volume in the current memory batch (see BATCH FUNTIONS); the value obtained with this procedure will be the volume supplied in every following batch. In order to modify this value, repeat the operations above. This procedure set the safety timer at a value 1.25 times greater than the time used to reach the batched quantity; after that the counter will be reset.
(POS. 9.6) Automatic selection of batch formula [BM AUTO SEL=ON/OFF]
The function allows the automatic selection of the first 4 formulas depending on the duration of the pulse of the batch start (see section Input operation stage). This function is active only if the function cons. mode (pos. 9.7) has not enabled. Besides, activating this function, the automatic compensation of the batch volume is also excluded (the value of the parameter "N.samples" (pos.
9.1) will be automatically set to zero). However the manual compensation is possible introducing the opportune value on the parameter "V.com" (pos. 9.3)
(POS. 9.7) Static consent of batch [CONS. MODE=ON/OFF]
The function enable the dosing start and stop using a static signal, instead of an impulsive, applied to the input (see Digital input section), this signal will have to stand applied all through the batch. The function automatically disables functions BM AUTO SEL (pos. 9.6) and AUTO BATCH (pos. 9.5).
HML4-F1
32
MENU 10 - DIAGNOSTIC
(POS. 10.1) Meter calibration [CALIBRATION]
Enable the calibration of the meter. The activation of this function happens pressing the key Enter during the function visualization. Will be visualized the following question: "EXECUTE?" press the key Esc/Del/Canc to proceed. Press any other key to delete the operation (POS. 10.2) Autotest function enable [SELF TEST] Meter autotest function. This function stops the normal functions of the meter and performs a complete test cycle on the measure input circuits and on the excitation generator. To activate this function, after select it, push key Enter, at the question: “EXECUTE?” push the key Esc/Del/Canc. For start autotest, or any other key for delete operation. The result of the test is shown on the display. At the end of operation will have visualized one of visualization page. This function is automatically performed when switching on the device. (POS. 10.3) Flow rate simulation [SIMULATION=ON/OFF] Flow rate simulation enabling. With this function it is possible to generate an internal signal that simulates the flow rate, allowing the outputs and all the connected instruments test. After enabling it, a ‘S’ appears in the top left of the screen and the flow rate simulation can be: q set: by pushing the key Enter from one of visualization pages, to set the required flow rate
(Fl.rate=) and the same key to confirm the value;
q finished: by pushing the key Enter from visualization pages and then by the key Esc/Del/Canc.
HML4-F1
33
BATCH FUNCTION
For each formula you can
associate:
q Product quantity q Product name q
Maximum time for batch (safety time for
each formula)
After activating the batch function from visualization page, proceed as in the
aside example.
VISUALIZATION PAGE WITH BATCH FUNCTION ENABLE
PROGRAMMING BATCH
Programming formula n°
product Name
Batch in progress
Visualizations:
1) batch off: n° batch effected
2) batch on: decrease safety timer
3) programming: programming safety timer indefinite if timer=0 timer disabled
Programming batch quantity
Visualization batched product
Input key code
Choose the formula number for associate
quantity batch (between 00 and 15)
Input quantity product for each batch
Input product name for batch (max 8 characters.)
Input maximum time for batch
If timer = 0, safety timer disabled max time set = 6000 sec. N.B. if to one output is assigned the function of batch alarm, to the attainment of the maximum time, beyond to the interruption of the batch,
will have activated also the alarm output
From the visualization pages
ENABLE BATCH
Enable one of the following functions to enable and program the batch on the converter:
q POS. 5.7-8-9: START/STOP batch from input q POS. 6.1-2-3-4: assign one of the functions to one of two output
Some examples of operation of such functions are visualized from page 14
HML4-F1
34
START STOP BATCH
START: it is possible activate the start of batch in two different way:
1. from remote input: assigning the
functions of start/stop batch to the input 1 (POS.
5.7) or input 2-3 (POS. 5.8-9) and using the input/s like visualized from page 14.
2. from keyboard: pressing of the key N.B.:
the start of batch from keyboard is always on the descent front (release of the
key) and is not available with the function of batch consent (POS. 9.7)
STOP: the stop of batch can be due to three events:
1. keyboard or remote input (manual stop): pressing of the key
2. end of batch:
in this case the stop of batch will have activated from a output signal to
the attainment of the batch quantity
3. maximum time of batch:
if a maximum batch time has been set and this is exceeded,
the batch in progress is stopped independently from the batched quantity
Notes:
q
during the batch the symbol of the active batch
and the name of the
formula are visualized on video.
q When the batch outputs are enable, pushing the key , the outputs will remai
n
energized till the key is released. On the display, in place of the CT and ST totalises the
following messages will appear:
!! VALVE !!
!! OPENED !!
IMPORTANT NOTES
Cross of relation between functions assigned to the input and automatic settin
gs of IN/OUT:
THEN
IF
POS 9.5
POS 9.6
POS 5.4 POS 9.1
POS
3.6-
5.6
POS 5.5
POS 5.7
POS 5.8
POS
5.9
AUTO BATCH
BM AUTO SEL
COUNT LOCK
N. SAMPLES
AUTO RANGE
CHANGE OR FROM
INPUT
CALIBRATION
INPUT 1
(BATCH)
INPUT 2
INPUT 3
POS 9.5
AUTO BATCH
*DISABLE
POS 9.6 BM AUTO SEL ** DISABLE
POS 9.7
CONS. MODE
DISABLE
DISABLE
* DISABLE
POS 5.7 INPUT 1 ON BATCH
DISABLE
DISABLE
POS 5.8
INPUT 2 ON BATCH
DISABLE
DISABLE
DISABLE
POS 5.9
INPUT 3 ON BATCH
DISABLE
DISABLE DISABLE
DISABLE
POS 6.1-2-3-4
OUTPUT 1-2-3-
4 ON BATCH
FUNCTIONS
DISABLE
DISABLE
DISABLE
POS 6.3-4
OUTPUT 3-4 ON ANY
FUNCTION
DISABLE
DISABLE
POS 5.8-9
INPUT 2-3 ON BM SELECT
DISABLE
** DISABLE
* IF IMPUT 1 USED
** SET VALUE TO ZERO
To
optimize the performances of the meter used as a batch instrument (batch time < 5s), it is recommended to set it as prompt as possible according to the plant requirements, choosing the batching setup for the automatic parameter optimization of the quick start menu. It is also reccomended to disactivate the function 3.3, AC line filter.
HML4-F1
35
CODES
ANOMALIE DESCRIPTIONS
ACTION TO TAKE
0001
problem with watch-dog circuit
ADDRESS TO SERVICE
0002
wrong configuration work data in eeprom
0004
wrong configuration safety data in eeprom
0008
defective eeprom
0010
defective keyboard (one or more key are pushed during the test)
0020
Power supply voltage (+3.3) is out of range
0040
Power supply voltage (+13) is too low (<10V)
0080
Power supply voltage (+13) it’s too high (>14V)
0200
timeout calibration input (input circuit is broken)
0400 Input stage gaining is out of range
Check the status of the cables connecting the sensor to the converter, the grounding connections of the devices or the possible presence of strong and anomalous noise sources
0800 Interruption on the coils circuit
Check the status of the cables connecting the sensor to the converter
0C00 Cumulative alarm 0800 + 0400
see single code
ALARMS
Messages ANOMALIES ACTION TO TAKE
NO ALARMS
All works regularly
-----
MAX ALARM
The flow rate is higher than the maximum threshold set
Check the maximum flow rate threshold set and the process conditions
MIN ALARM
The flow rate is lower than the minimum threshold set
Check the minimum flow rate threshold set and the process conditions
FLOW RATE >FS
The flow rate is higher than the full scale value set on the instrument
Check the full scale value set on the instrument and the process conditions
PULSE/FREQ>FS
The pulse generation output of the device is saturated and cannot generate the sufficient number of impulses
Set a bigger unit of volume or, if the connected counting device allows it, reduce the pulse duration value
EMPTY PIPE
The measuring pipe is empty or the detection system has not been properly calibrated
Check whether the pipe is empty or perform again the empty pipe calibration procedure
BATCH ALARM
Batch interrupted for the followings
condition:
q Timer batch expired before the end of the
batch
q Batch valve open and flow rate to zero for a
time longer to the safety timer set
q
batch valve closed and flow rate different from zero for a time longer to the safety timer set
Verify:
q presetting q system condition
INPUT NOISY
The measure is strongly effected by external noise or the cable connected the converter to the sensor is broken
Check the status of the cables connecting the sensor, the grounding connections of the devices or the possible presence of noise sources
EXCITATION FAIL
The coils or the cable connecting the sensor are interrupted
Check the connecting cables to the sensor
CURR. LOOP
OPEN
The 0/4...20mA output on board or the optional one are not correctly closed on a valid load
Verify the load is applied to the output (max 1000 ohm).
To disable the alarm,set the “mA VAL.FAULT” value (
menu alarm ) to 0.
P.SUPPLY FAIL
Power supply different from that suitable in the label.
Verify that the power supply is that suitable on label
Causes and actions to be taken
Anomalies codes
CONFORMITY DECLARATION
it declares under the own responsibility that the product:
Converter model:
HML 110
Sensors model:
HMS 501 – HMS 600 – HMS 1000 – HMS 2410 – HMS 2500 – HMS 377
0
HMS 3800 HMS 5000 to which this declaration refers, is in compliance with the following
Harmonized
European Norms:
§ EN 61010-1:2010
§ EN 61326-1:2006
and therefore answering to essential requirement of CE directives:
§ 2006/95/CE (Low voltage directive – LVD)
§ 2004/108/CE (Electromagnetic compatibility Directive – EMC)
04/03/2011
Liqui d c onT roL S
105 Albrecht Drive Lake Bluff, IL 60044 (847) 295-1050
Liqui d c onT roL S e uro pe/ SAm pi
Via Amerigo Vespucci 1 55011 Altopascio (Lucca), Italy +39 0583 24751
ideX FLu id And me Ter ing pv T. LTd.
Survey No. 256, Alindra Savli GIDC, Manjusar Dist. Vadodara 391 770 Gujarat, India +91 265 2631855
Liqui d c onT roL S S pon SLe r
105 Albrecht Drive Lake Bluff, IL 60044 (847) 295-1050
TopTe ch SySTe mS
1124 Florida Central Parkway Longwood, FL 32750 (407) 332-1774
Nateus Business Park Nieuwe Weg 1-Haven 1053 B-2070 Zwijndrecht (Antwerp), Belguim +32 (0)3 250 60 60
FAu re her mAn
Route de Bonnetable B.P. 20154 72406 La Ferté-Bernard Cedex, France +33 (0)2 43 60 28 60
6961 Brookhollow West Drive Houston, TX 77040 (713) 623-0808
corke n
3805 Northwest 36th St. Oklahoma City, OK 73112 (405) 946-5576
105 Albrecht Drive Lake Bluff, IL 60044-2242
1.800.458.5262 • 847.295.1050 Fax: 847.295.1057 www.lcmeter.com
© 2012 Liquid Controls
Pub. No. 502062
(6/12)
Loading...