Leica HistoCore Arcadia H Instructions For Use Manual

HistoCore Arcadia H
Paraffin Embedding Station
Instructions for Use English
Order No.: 14 0393 81101 - Revision K
Always keep this Instructions for Use with the instrument. Read carefully before working with the instrument.
Version 1.8, Revision K - 10.2018
Note
The information, numerical data, notes and value judgments contained in this Instructions for Use represent the current state of scientific knowledge and state-of-the-art technology as we understand it following thorough investigation in this field.
We are under no obligation to update the present Instructions for Use periodically and on an ongoing basis according to the latest technical developments, nor to provide our customers with additional copies, updates etc. of this Instructions for Use.
To the extent permitted in accordance with the national legal system as applicable in each individual case, we shall not be held liable for erroneous statements, drawings, technical illustrations etc. contained in this Instructions for Use. In particular, no liability whatsoever is accepted for any financial loss or consequential damage caused by or related to compliance with statements or other information in this Instructions for Use.
Statements, drawings, illustrations and other information regarding the contents or technical details of the present Instructions for Use are not to be considered warranted characteristics of our products.
These are determined only by the contract provisions agreed between ourselves and our customers.
Leica Biosystems reserves the right to change technical specifications as well as manufacturing processes without prior notice. Only in this way is it possible to continuously improve the technology and manufacturing techniques used in our products.
This document is protected under copyright laws. All copyrights to this documentation are held by Leica Biosystems Nussloch GmbH.
Any reproduction of text and illustrations (or of any parts thereof) by means of print, photocopy, microfiche, web cam or other methods – including any electronic systems and media – requires express prior permission in writing by Leica Biosystems Nussloch GmbH.
For the instrument serial number and year of manufacture, please refer to the nameplate on the back of the instrument.
HistoCore Arcadia H
Leica Biosystems Nussloch GmbH Heidelberger Str. 17 - 19 D-69226 Nussloch Germany Tel.: +49 - (0) 6224 - 143 0 Fax: +49 - (0) 6224 - 143 268 Web: www.LeicaBiosystems.com
Assembly contracted to Leica Microsystems Ltd. Shanghai
Table of contents
1. Important Information ................................................................................................................................ 6
1.1 Naming conventions ....................................................................................................................................... 6
1.2 Symbols in the text and their meanings ....................................................................................................... 6
1.3 Instrument type .............................................................................................................................................. 10
1.4 Intended use of instrument .......................................................................................................................... 10
1.5 Qualification of personnel ............................................................................................................................ 10
2. Safety .......................................................................................................................................................... 11
2.1 Safety Notes ................................................................................................................................................... 11
2.2 Warnings ......................................................................................................................................................... 12
2.3 Integrated safety devices............................................................................................................................. 13
3. Instrument Components and Specifications ........................................................................................14
3.1 Overview – Instrument components .......................................................................................................... 14
3.2 Main features of the instrument.................................................................................................................. 15
3.3 Technical Data ............................................................................................................................................... 16
4. Setting up the instrument ........................................................................................................................ 18
4.1 Site requirement ............................................................................................................................................ 18
4.2 Standard delivery – packing list .................................................................................................................. 18
4.3 Unpacking and installation........................................................................................................................... 19
4.4 Necessary assembly work ........................................................................................................................... 20
4.5 Electrical connection .................................................................................................................................... 22
4.6 Moving the instrument .................................................................................................................................. 22
5. Operation .................................................................................................................................................... 23
5.1 Instrument parts/functions........................................................................................................................... 23
5.2 Switching the instrument on ........................................................................................................................ 28
5.3 Control panel functions................................................................................................................................. 29
5.4 Operation modes ............................................................................................................................................ 33
5.5 Instrument heater .......................................................................................................................................... 34
6. Maintenance and Cleaning .....................................................................................................................35
6.1 Cleaning the instrument................................................................................................................................ 35
6.2 Maintenance instructions ............................................................................................................................ 36
7. Troubleshooting ........................................................................................................................................ 37
7.1 Error message ................................................................................................................................................ 37
7.2 Possible faults ................................................................................................................................................ 40
7.3 Changing a fuse ............................................................................................................................................. 41
8. Optional Accessories ...............................................................................................................................43
8.1 Magnifier ......................................................................................................................................................... 43
8.2 Foot switch ...................................................................................................................................................... 43
8.3 Pre-filter cup ................................................................................................................................................... 45
8.4 Basket handle ................................................................................................................................................. 45
8.5 Ordering information ..................................................................................................................................... 45
Version 1.8, Revision K
Table of contents
9. Warranty and Service .............................................................................................................................. 47
10. Decontamination Confirmation .............................................................................................................. 48
HistoCore Arcadia H
Important Information
1

1. Important Information

1.1 Naming conventions

Note
The full name of the device is HistoCore Arcadia H Paraffin Embedding Station. The device is called HistoCore Arcadia H to ensure that the Instructions for Use are well legible.

1.2 Symbols in the text and their meanings

Symbol: Title of the symbol: Warning
Description: Warnings appear in a white box and are marked by a
warning triangle.
Symbol: Title of the symbol: Note
Description: Notes, i. e. important user information, appear in a
white box and are marked by an information symbol.
Symbol: Title of the symbol: Item number
"Fig. 7-1" Description: Item numbers for numbering illustrations. Numbers in
red refer to item numbers in illustrations.
Symbol: Title of the symbol: Function key
Save Description: Function keys, that have to be pressed on the display
screen displayed as bold, gray text.
Symbol: Title of the symbol: Warning, hot surface
Description: Instrument surfaces which become hot during
Symbol: Title of the symbol: Flammable
Description: Flammable reagents, solvents and cleaning agents
Symbol: Title of the symbol: Caution
Description: Indicates the need for the user to consult the
operation are marked with this symbol. Avoid direct contact to prevent risk of burning.
are labeled with this symbol.
Instructions for Use for important cautionary information such as warnings and precautions that cannot, for a variety of reasons, be presented on the medical device itself.
Symbol: Title of the symbol: ON (Power)
Description: The power supply is connected upon pushing the
power switch.
Version 1.8, Revision K
Important Information 1
Symbol: Title of the symbol: OFF (Power)
Description: The power supply is disconnected upon pushing the
power switch.
Symbol: Title of the symbol: Consult Instructions for Use
Description: Indicates the need for the user to consult the
Instructions for Use.
Symbol: Title of the symbol: Manufacturer
Description: Indicates the manufacturer of the medical product.
Symbol: Title of the symbol: Manufacturing date
Description: Indicates the date when the medical device was
manufactured.
Symbol: Title of the symbol: Alternating current
Symbol: Title of the symbol: PE terminal
Symbol: Title of the symbol: Article number
Description: Indicates the manufacturer's catalog number so that
the medical device can be identified.
Symbol: Title of the symbol: Serial number
Description: Indicates the manufacturer's serial number so that a
specific medical device can be identified.
Symbol: Title of the symbol: China ROHS
Description: Environmental protection symbol of the China RoHS
directive. The number in the symbol indicates the "Environment-friendly Use Period" of the product in years. The symbol is used if a substance restricted in China is used in excess of the maximum permitted limit.
Symbol: Title of the symbol: WEEE Symbol
Description: The WEEE symbol, indicating separate collection
for WEEE – Waste of electrical and electronic equipment, consists of the crossed-out wheeled bin (§ 7 ElektroG).
Symbol: Title of the symbol: CE Compliance
Description: The CE marking is the manufacturer's declaration
that the medical product meets the requirements of the applicable EC directives.
HistoCore Arcadia H
Important Information
1
Symbol: Title of the symbol: CSA Statement (Canada/USA)
Description: This product fulfills the requirements of the CAN/
CSA-C22.2 No. 61010.
Symbol: Title of the symbol: Fragile, handle with care
Description: Indicates a medical device that can be broken or
damaged if not handled carefully.
Symbol: Title of the symbol: Store dry
Description: Indicates a medical device that needs to be
protected from moisture.
Symbol: Title of the symbol: This way up
Description: Indicates correct upright position of the transport
package.
Symbol: Title of the symbol: Stack limit
Description: It allows maximum 2 stacks layers.
Symbol: Title of the symbol: Temperature limit for transport
Description: Indicates the temperature limits for transport to
which the medical device can be safely exposed.
Symbol: Title of the symbol: Temperature limit for storage
Description: Indicates the temperature limits for storage to which
the medical device can be safely exposed.
Version 1.8, Revision K
Important Information 1
Symbol: Title of the symbol: Humidity limitation for transport and storage
Description: Indicates the range of humidity for transport and
storage to which the medical device can be safely exposed.
Symbol: Title of the symbol: Shockdot Impact Indicator
Description: In the Shockwatch system, a shock dot shows
shocks or impacts that are above a specified intensity through red coloration. Exceeding a defined acceleration (g value) causes the indicator tube to change color.
Symbol: Title of the symbol: Tilt indicator
Description: Indicator to monitor whether the shipment has been
transported and stored in upright position according to your requirements. With a pitch of 60° or more, the blue quartz sand flows into the arrow-shaped indicator window and sticks there permanently. Improper handling of the shipment is immediately detectable and can be proven definitively.
Symbol: Title of the symbol: Recycling
Description: Indicates, that the item can be recycled where
correct facilities exist.
Symbol: Title of the symbol: Regulatory Compliance Mark (RCM)
Description: The Regulatory Compliance Mark (RCM) indicates a
device’s compliance with applicable ACMA technical standards of New Zealand and Australia – that is, for telecommunications, radio communications, EMC and EME.
HistoCore Arcadia H
Important Information
1

1.3 Instrument type

All information provided in these Instructions for Use applies only to the instrument type indicated on the cover page. A nameplate is attached to the back of the instrument and a serial number label is on the side of the instrument.

1.4 Intended use of instrument

The HistoCore Arcadia H is a modern paraffin embedding station with a microprocessor control system.
The HistoCore Arcadia H is designed for embedding histological tissue specimens in molten paraffin for use in pathology laboratories.
It is only for the following tasks:
• Melt solid paraffin for sample embedding and maintain the molten paraffin at the required temperature.
• Dispense paraffin into embedding molds where the specimens are placed.
• Heat and maintain the temperatures of embedding cassettes with specimens and molds as well as the required forceps.
Warning
Any other use of the instrument will be considered as improper use!

1.5 Qualification of personnel

• The HistoCore Arcadia H may be operated by trained laboratory personnel only.
• All laboratory personnel designated to operate this instrument must read these Instructions for Use carefully and must be familiar with all technical features of the instrument before attempting to operate it.
10
Version 1.8, Revision K
Safety 2

2. Safety

2.1 Safety Notes

Warning
The safety and caution notes in this chapter must be observed at all times. Be sure to read these notes even if you are already familiar with the operation and use of other Leica Biosystems products.
These Instructions for Use include important instructions and information related to the operating safety and maintenance of the instrument.
The Instructions for Use are an important part of the product, and must be read carefully prior to startup and use and must always be kept near the instrument.
This instrument has been built and tested in accordance with the safety requirements for electrical equipment for measurement, control, and laboratory use.
To maintain this condition and ensure safe operation, the user must observe all notes and warnings contained in these Instructions for Use.
Note
These Instructions for Use must be appropriately supplemented as required by the existing regulations on accident prevention and environmental safety in the operator‘s country.
Warning
• The protective devices on the instrument and its accessories must not be removed or modified. Only service personnel qualified by Leica Biosystems may repair the instrument and access the instrument's internal components.
• Only use the provided power cable. This power cable must be replaced if the power plug does not fit your countries socket. Please contact Leica Biosystems Service.
• The maximum load of the paraffin tank lid is 1 kg. Do not exceed 1 kg, otherwise damage could occur to the paraffin tank lid.
Residual risks
• The instrument has been designed and constructed with the latest state-of-the-art technology and according to recognized standards and regulations with regard to safety technology. Operating or handling the instrument incorrectly can place the user or other personnel at risk of injury or can cause damage to the instrument or other property. The instrument may be used only as intended and only if all of its safety features are in proper working condition. Malfunctions that impede safety must be remedied immediately.
Note
For current information about applicable guidelines, please refer to the CE declaration of conformity and on our Internet site at:
http://www.LeicaBiosystems.com
HistoCore Arcadia H
11
Safety
2

2.2 Warnings

The safety devices installed in this instrument by the manufacturer only constitute the basis for accident prevention. Operating the instrument safely is, above all, the responsibility of the owner, as well as the designated personnel who operate, service or repair the instrument.
To ensure trouble-free operation of the instrument, make sure to comply with the following instructions and warnings.
Warnings – Safety notes on the instrument itself
Warning
• Safety notes on the instrument itself, which are marked with a warning triangle, indicate that the correct operating instructions (as defined in these Instructions for Use) must be followed when operating or replacing the item marked. Nonobservance can cause accidents, injuries and/or damage to the instrument/accessories.
• Certain surfaces of the instrument are hot during operation under normal conditions. They are marked with this warning sign. Touching these surfaces without suitable safety measures can cause burns.
Safety instructions – Transport and installation
Warning
• After unpacking the instrument it may only be transported in an upright position.
• Place the instrument on a laboratory table and ensure it is level.
• The instrument must not be exposed to direct sunlight (window)!
• Only plug the instrument into a grounded power socket. If an extension cable has to be used, make sure that it has a protective grounding conductor.
• Connect the instrument with the right power socket voltage 100~120 V or 220~240 V according to the instrument type.
• The installation location must be well-ventilated; there should be no ignition sources there of any kind.
• The instrument may not be operated in hazardous locations.
• Extreme temperature fluctuations between storage facility and setup site as well as high humidity may cause condensation to form. In this case, wait at least two hours before switching on.
12
Version 1.8, Revision K
Safety 2
Safety instructions – Working with the instrument
Warning
• Paraffin is flammable and should therefore be handled with due care.
• Do not use sharp tools to remove solidified paraffin from the work areas, as this may destroy the coating on the surface. Use the plastic spatula supplied with the instrument.
• During operation, the paraffin tank, embedding mold tray, cassette tray, work area as well as the forceps holder are hot.
• Risk of burning!
• Do not move the instrument during operation.
• Do not store any combustible and flammable substances near the instrument. There is a fire hazard if work with an exposed flame (e.g. Bunsen burner) is carried out in the direct vicinity of the instrument (solvent vapors). Therefore, keep all ignition sources at least 2 meters away from the instrument!
• Please allow 30 minutes before touching the instrument after shutting down.
• Failure to observe the instructions specified by the manufacturer may result in damage to the protection provided by the instrument.
Hazards – Servicing and cleaning
Warning
• Switch off the instrument each time before servicing and pull out the power plug.
• When using cleaners, please comply with the safety instructions of the manufacturer and the laboratory safety regulations.
• Before changing defective fuses, the instrument has to be disconnected from the power supply. Fuses in the fuse holder on the back panel can be replaced by the user.
• Liquid must not get into the instrument while operating or cleaning it.

2.3 Integrated safety devices

The instrument is equipped with the following safety features and devices:
Fuses in the heating elements
All of the instrument's AC heating elements are equipped with overheating fuses, which trip if an AC heating element overheats and switches the element off.
Warning
• Fuses will only automatically reset when the instrument is disconnected from AC power and the heating element temperature drops below 50 °C.
• Note the only way the user can disconnect the instrument from a power source is to remove the power plug.
HistoCore Arcadia H
13
Instrument Components and Specifications
3

3. Instrument Components and Specifications

3.1 Overview – Instrument components

Instrument front view
1
2 12
13
3
4
5
6
7
8
Fig. 1
1 Power switch 8 Cold spot
11
10
9
14
2 Control panel 9 Paraffin collection trays
3 Dispenser 10 Lid for right tray
4 Forceps holder 11 Right tray
5 Left tray 12 Work area illumination
6 Lid for left tray 13 Paraffin tank
7 Working surface
Version 1.8, Revision K
Instrument rear view
1
2
Instrument Components and Specifications 3
3
5
Fig. 2
1 Port for foot switch 4 AC fuses
2 Back panel 5 Foot
3 Port for power inlet

3.2 Main features of the instrument

• Paraffin tank with a capacity of 4 litres.
• 5.7 inch LCD display and integrated capacitive touch keys.
• The paraffin flow is activated by means of a height-adjustable, pivotable clip – activated either manually by pushing or via a foot switch (optional).
• Controllable flow rate.
• Removable paraffin collection trays.
• Spacious, easy-to-clean, heated work area, with integrated cold spot, also for extra large cassettes ("Super Cassettes") with paraffin drain system.
• Trays for cassettes and/or molds with foldable lid, removable and interchangeable.
• Removable, heated forceps holder for 6 forceps, accessible from both sides.
• Optimum illumination of the working surface by LED lamp, controlled by the key on LCD control panel.
• Temperature range of cassette and embedding mold tray, work area and paraffin tank adjustable from 50 °C (122 °F) to 75 °C (167 °F).
• The beginning and end of the work time and work days can be programmed.
• Provide the error message for operation condition monitoring.
• Provide the enhance heating function for quicker paraffin melting.
4