2.3 Flanged type to DIN 2501 / ANSI B16.5 (SCH-40) 6
2.4 Temperature / Pressure Measurement for
External Density Compensation 6
3. Electrical connection 7-8
3.1 Installation location and cable diameter 7
3.2 Connection to power
3.3 Outputs
3.3.1 Abbreviations 7
3.3.2 Current (analog) output I 7
3.3.3 Connection between VFS1000 (I) F and VFC 091 F 8
3.3.4 Pulse Output P 8
4. Start-up 8
Part B Signal converter VFC 091
5 Operation of the signal converter 9
5.1 General 9
5.1.1 Starting up signal converter 9
5. I.2 Measurement mode 9
5.1.3 Programming or menu mode 9
5.1.4 Error handling 9
5.2 Operating and check elements 10
5.3 Program organization and programming chart 11
5.3.1 Menu levels 11
5.3.2 Programming chart 12
5.3.3 Description of keys 12
5.4 Programming and function of keys 14+15
5.5 Error Messages 16
5.5.1 Error Messages in Measurement mode 16
5.5.2 Error Messages in Programming mode 16
2
6. Description of programme functions 7-23
6.1 Numerical order description 7-20
6.2 Functional order description 20
6.2.1 Physical units 20
6.2.2 Numerical format 21
6.2.3 Display 21
6.2.4 Flow range and meter size 21
6.2.5 Primary information 22
6.2.6 Application information 22
6.2.7 Internal Electronic Totalizer 22
6.2.8 Current (analog) output I 22
6.2.9 Pulse. output P 22-23
6.2.10 Coding desired for entry into programming mode 23
6.2.11 Behaviour of outputs during programming 23
6.2.12 HART® Programmability 23
6.2.13 Functions through HHC 24
6.2.14 Generic Online Menu Tree 25
Part C Functional checks and Trouble shooting hints
7. Functional checks 26
7.1 Primary head functional checks 26
7.1. Vortex Sensor 26
7.2 Signal converter functional checks 26
7.2.1 Self diagnostics 26
7.2.2 Display check 26
7.2.3 Current output check 26
7.2.4 Pulse output check 26
7.2.5 Frequency measurement check 26
8. Trouble shooting hints 27
Part D VFM 1091 (I) Ex
9. Description of the system 28
9.1 Temperature class 28
9.2 Electrical connection 28
10. Field connection diagram 29
11. Nameplate & Certificates of VFM 1091(I) Ex 29-30
Part E Technical data 31-50
3
Part A system installation and start-up
1. Description
KRONHE MARSHALL vortex flowmeters operate on the Karman
vortex street principle to measure volumetric flow rate of gases /steam
and liquids. VFM 1091 computes normalized volumetric and mass flow
rates from operating & normalized density values supplied through
operator interface.
Items included with shipment for Compact flowmeter
• Compact Vortex flowmeter and operating instructions
Items included with shipment for field mounted
Flowmeter
• Field mounted sensor VFS 1091 F
• Pipe/Wall mounted separated signal converter VFC 091F
• Connecting Cable (10m) between VFS 1091 F & VFC 091F
• Installation and Operating Instructions
• Mounting bolts, washers, nuts
• Programming chart indicating factory setting.
Provided by customer
• All cables for electrical connections.
2. Installation in the pipeline
2.Difference between inside diameters of primary head and
pipeline
DΝ Meter size of primary head in millimeters or Inches
Φ Inside diameter of primary head in millimeters Max.
∆Φ allowable difference between inside
diameters of primary head and pipeline
SIZE φ∆φ SIZE φ∆φ
DN
(mm)
mm
(inch)
10S 8.9
(0.35)
10 12.6
(0.50)
15 14.9
(0.59)
20 20.9
(0.82)
25 28.5
(1.12)
40 43.1
(1.70)
50 54.5
(2.15)
80 82.5
(3.25)
100 107.1
(4.22)
150 159.3
(6.27)
200 206.5
(8.13) 1 (0.039)
mm
(inch)
0.4
(0.016)
0.4
(0.016)
0.4
(0.016)
0.4
(0.016)
0.4
(0.016)
0.4
(0.016)
0.6
(0.024)
0.6
(0.024)
0.6
(0.024)
0.8
(0.031)
DN
(inch)
3/8"S 8.9
mm
(inch)
(0.35)
3/8" 12.6
(0.50)
1/2" 14.9
(0.59)
3/4" 20.9
(0.82)
1" 26.7
(1.05)
1/5" 40.9
(1.61)
2" 52.6
(2.07)
3" 78
(3.07)
4" 102.4
(4.03)
6" 154.2
(6.07)
8" 202.7
(7.98) 1 (0.039)
mm
(inch)
(0.016)
(0.016)
(0.016)
(0.016)
(0.016)
(0.016)
(0.024)
(0.024)
(0.024)
(0.031)
Ensure that the bore of upstream and down-stream pipes are smooth
and without deposits or scaling of welding beads.
0.4
0.4
0.4
0.4
0.4
0.4
0.6
0.6
0.6
0.8
2.1 General
1. Flow direction and meter position -
• Flow must always be in the direction of the arrow, bluff side of
vortex-shedding body facing incoming flow i.e. upstream side.
• Vertical pipe run: upward flow direction
• Horizontal pipe run: see below
3. Straight, unimpeded inlet and outlet runs
D = Meter size (Nominal Dia. DN)
CONTROL VALVE
BALL/GATE VALVE
(FULLY OPEN)
0
2 X 90
BENDS
3-DIMENSIONAL
(SAME FOR 3 X 90
0
2 X 90
BENDS
0
90
BENDS
OR “T” PIEC
0
BENDS)
4
EXPANDER
REDUCER
Flexible pipes or bellows may be used before the upstream and after the
downstream in case the vibrations are too high. Pipe vibration limit is
0.2 g peak to peak up to 8-500 Hz crossover frequency. (which equals
0.75 mm)
FLOW STRAIGHTENER
SPACIAL FLOW STRAIGHTENER
With flow straightener the inlet pipe length may be halved e.g. for a
control valve the inlet length is 25D instead of 50D. The minimum inlet
pipe length including flow straightener must always be 12 D.
The small sizes (below 1" or DN 25) come with upstream &
downstream pipelines & flange connections to suit customer's flanges.
4. Pipe vibration
Caused, for example, by the action of pumps, valves, etc., will falsify
measurements particularly at low flow velocities. Support the pipeline
on both sides of the flowmeter; in the direction perpendicular to both
the pipeline & bluffbody axis.
SANDWICH VERSION
5. Pipeline along a wall
Where possible, the distance between the pipe centerline and wall
should be greater than 0.5 m (20”). If it is less, first connect all cables
to terminals in the connection compartment (power supply and outputs)
and run them via an intermediate connection box (see also Section 3)
before installing the flowmeter.
Do not lag (Thermally insulate) the meter body.
6. Orientation
• Turn the Display board through ±90 or 180 to obtain horizontal
positioning of the display.
• Turn the signal converter housing through ±900 or 1800 should that
be more suitable for the location of the installation.
7. State of Medium
• Ensure single-phase flow. Liquid droplets in gas / vapour, solid
particles in gas/liquid & gas bubbles in liquid are not permitted.
In Liquid application e.g. water; to prevent cavitation, minimum D/S
pressure is given by the relation:
Pds (bar_g) >= (2.9* DP)+ (1.3 * Ps) - 1.013
where DP= pr.drop of VFM in Bar from sizing
program.
Ps= Sat.pr.in Bar at op.temp
FLANGED VERSION
• In case of steam or compressed gas, a moisture separator may be used
compensation
50D upstream of the meter if the dryness fraction is less than 95%.
• For gases, a filter or strainer may be used to remove solid particles.
This is specially important for meter sizes below 1".
2.2 Sandwich type to DIN 19205 / ANSI
• Meter sizes DN25, 40, 50, 80, 100 and 150 (1" to 6")
• check flange connection for leak-tightness after
flowmeter installation.
2.4 Temperature / Pressure measurement for External density
To measure the temperature and pressure of the medium for either mass
flow/normalized f1ow computation using external flow computer or
measure operating density using external pressure and temperature
measurement (e.g. to feed the density value in the VFM off-line mass
or normalized flow computation) suitable measuring points should
• Bolts, nuts and washers are supplied with meter, check flange
connection for leak-tightness after flowmeter installation.
Installation upstream of f1owmeter
Min. distance: 20*DN (DN = meter size)
Installation downstream of flowmeter
Min. distance: 5*DN (DN = meter size)
Allowance must be made for the pressure drop in the flowmeter as
correction value for operating conditions prevailing upstream of the
flowmeter.
3. Electrical Connection
3.1 Installation Location & Cable diameter
Location
• Do not expose the compact flowmeter to direct sunlight Install a
sunshade if necessary.
• Do not expose to intense vibration. If necessary support the pipeline
to the left right of the flowmeter.
• The rotating design of the housing makes it easier to connect the two
cables for power and outputs to the terminals in the rear terminal
box.
Cable diameter
• To conform to protection category requirements, observe
• the following recommendations :
• Cable diameter: 8 to 13 mm (0.31" to 0.51")
• Enlarge the inside diameter of the Screwed conduit entry by
removing the appropriate onion ring(s) from the seal, only if cables
have extremely tight fit.
• Fit blanking plug PG16 and apply sealant to unused cable entries.
• Do not kink cables at conduit entries
• Provide water drip point (U bend in cable).
Conduit Installation, general wiring considerations
When electrical codes require conduit, it must be installed in such a
manner that the meter connection compartment remains at all dry
times.
3.2 Connection to loop power supply
• In case of functional extra low voltages (24VDC) protective
separation in conformity with VDE 400 part 410, or equivalent
national standard, must be ensured.
• Hazardous area application will require the connection of
protective Earth.
• Connection to power, VFM 1091 K
• Current and frequency outputs are galvanically isolated when the
optional ground above is not used and can therefore be
simultaneously connected to a receiver instrument which is
grounded or separately connected to two receiver instruments.
• Ensure that 24V, 4-20 mA loop is grounded at one point only (i.e.
either at the input of the signal converter or at the input of the
indicator, or at the output of the Power Supply). Do not ground the
loop at more than one point.
• The current output is galvanically isolated from Pulse output P.
• Max. load at terminals 5/6
U – 12 V
max. load in Kohms
R =
B
B
20mA
<1200 ohms
Programming
via Fct. No....
-
-
1.3.X
1.3.2
1.1.2
1.1.2+1.3.2
Description
see Sect....
3.2+3.3.4
3.2
6.1+6.2.9
6.2.9
6.1
6.1+6.2.9
7
3.3.3 Connection between VFS1091 & VFC 091F
3.3.4 Pulse out put
• The pulse output is galvanically isolated from the 4-20mA current output and may be optionally grounded.
• All functions and operating data are programmable; see Section 5, 6.1 6.2.9
• Factory-set data and functions are listed in the enclosed report on settings. This can also be used to record any changes made to the operating
parameters.
•Passive frequency output, open collector for connection of active electronic counters EC or switchgear, input voltage 5 to 25V, load current max.
100 mA, min load resistance RL 250, selectable pulse width duty cycle or 250 mSec
4. Start-up
• Check that the system has been correctly installed as described in Sect. 1, 2 and 3.
• Before initial start-up check that the following details on the nameplate agree with the data specified in the report of settings for the signal
converter. If not, reprogramming will be necessary.
Meter size Fct3.1.1, Sect. 6.1, 6.2.5.
K-Factor Fct 3.1.2, Sect. 6.1, 6.2.5.
• The flowmeter is ready for service 15 minutes (waiting time) after switching on the power source. Increase flow velocity slowly and steadily.
• Avoid abrupt changes in pressure in the pipeline.
• If the process medium is steam, condensate may form initially and cause faulty measurements when the system is started up for the first time.
• When powered, the signal converter normally operates in the measurement mode. The power-on sequence to measurement mode is as follows:
“TEST” is displayed for approx. 3 seconds followed by
“VFM 1091 ” the instrument type followed by
“Ver x.xx” the software version of the instrument.
Then instrument operates in measurement mode where it displays the parameter being measured Or FATAL ERROR if there are one or more
critical errors detected (For description on errors refer Sect. 5.5.)
8
Part B Signal Converter VFC 091
5. Opening of the signal converter
5.1 General
5.1.1 Starting up signal converter
When power is switched ON to signal converter it displays TEST, VFM, 1091 & Ver.x.x.x and then goes to measurement mode. In this initial
sequence VFM 1091 (I) carries out self-diagnostics to check its own functiona1elements and loads the configuration data from non-volatile
memory. If any error(s) are detected in power on diagnostics converter displays
to carry normal measurements. The first measured parameter displayed is the one being displayed when power supply was removed last time.
5.1.2 Measurement Mode
In measurement mode, the parameters that converter measures/computes are shown on display in the appropriate units. (See Sect. 5.2 for display
details). As per the configuration, display can be either in non- cyclic/cyclic mode. In non cyclic mode of display, use key to see the next
parameter on display. In cyclic mode display shows all parameters one after another, wherein each parameter is displayed for about 6 seconds.
While in the measurement mode, the flow and the current values can be monitored using the HART
values, the communicator also displays the minimum and maximum values of flow values, which are programmed into the device.
5.1.3 Programming or menu mode
All the configuration/settings/test functions are grouped in the form of menu having a tree structure (see Sect. 5.3.1 for details) and are accessible
in the programming mode. Operator can view or alter the present settings, data values by the use of functions available in this mode. VFC091
implements all the universal commands and relevant common practice commands of HART
parameters using HART
All changes made in programming mode are stored in non-volatile memory and have appropriate effect on the operation of signal converter. While
being programmed (i.e. while in the menu), the instrument will stop making further measurements and the current O/P will be frozen to the last
value, pulse O/P will also stop. The changes made using the HART
data to the device. After data is received, VFC091 updates the configuration n with thenewly received data. During the updation period, the
measurement is interrupted. (The updation time is less than half second).
5.1.4 Error handling
®
Communicator. These are mainly flow-related parameters.
®
communicator are stored in the communicator till you request to "Send" the
FATAL. ERROR since instrument has critical error(s) and is not able
®
Communicator. Other than flow and current
®
this enables you to program some of the main
Converter can detect errors during power-on diagnostics as well as when in normal measurement mode. Errors are put into two main categories
viz. fatal errors and non-fatal errors. Fatal errors cause measurement to stop since they are serious in nature. Non-fatal errors do not affect
functionality of the converter.
If one or more errors are present, display (in measurement mode) starts blinking. If programmed so, error information is shown on display interleaved
between display of two parameters.
9
5.2 Operating & check elements
Caution
To avoid damage to electronics, be certain that the area around the
meter is dry before removing electronics compartment cover.
The operating elements are accessible after removing the cover of
the electronics section using the special wrench supplied.
Caution
Do not damage screw thread, never allow dirt to accumulate,
and make sure it is well greased at all times.
1
Display, 1st line
2
Display, 2nd line
3
Display, 3rd line
q v Volumetric flow rate
q n Normalised volumetric flow rate
q m Massflow rate
S Totalizer value
u Velocity of medium
F Vortex frequency
4
Keys for Programming the Signal Converter, refer
To Sect.5.4 for function of keys.
10
F
F
F
F
F
F
F
F
F
5.3 Program organization & programming chart
5.3.1 Menu levels
The program for the signal converter consists of 5 levels. The 1
st
line of display will identify the menu levels during programming
Actual Display
Fct 1.0
OPERATION
Fct
2.0
TEST
Fct 3.0
INSTALL.
Fct 1.1.0
BASIS PARAM
Fct 1.2.0
DISPLAY
Fct 1.3.0
PULS O/P.
Fct 3.1.0
BASIS PARAM
ct 3.2.0
USER DATA
ct3.3.0
APPLICAT
ct 1.1.1-1.1.4
ct 1.2.1-1.2.6
ct 1.3.1-1.3.3
ct 2.1-2.3
ct 3.1.1-3.1.2
ct 3.2.1-3.2.3
ct 3.3.1-3.3.4
A
C
C
E
S
S
T
O
S
E
T
T
I
N
G
S
O
R
D
A
T
A
P
O
S
S
I
B
L
E
A
T
T
H
I
S
L
E
V
E
L
O
N
L
Y
Measuring
mode level
Main
menu
Submenu
level
Function
level
Data
level
level
11
12
13
5.4 Program organization & function keys
Function of keys in measurement mode:
After power-on, the signal converter enters the normal measuring mode. Display shows the actual measured value of the parameter, units and
arrow markers to identify parameter. A steady (non-blinking) display indicates that there are no errors in measuring mode. Use key to get the
next parameter on display. If display is in cyclic mode (Fct 1.2.6 CYCLE DISP as YES) the next parameter is displayed after every 6 seconds
and key has no function.
Use the key to go to the programming mode If Fct.3.2.1 ENTRY. CODE.1 is YES then converter will ask for the Code 1 password. Password
is the following sequence of 9 key operations Press key to go back to normal display.
Use the key results in prompt for Code 2 password. Code 2 password is predefined and reserved for Khrone Marshall service person.
Operator shouldnot use key at normal display. If used by accident then give any arbitrary key sequence to code 2 prompt until display gets
back to normal.
Using keys in measuring mode
KEY FUNCTION
Without
Followed by 9 keys Coding, dependent on programming Fct.3.2.1. Go to programming mode.
With
Displaying next parameter measured or next error message
(Fct.1.2.5 as YES for error message)
if in non cyclic mode (Fct.1.2.6 as NO) In cyclic mode the key has no function
Reserved for use by Khrone Marshall
Functions of keys in programming mode:
Programming and other functions are grouped in a menu having tree structure. To navigate in the menu tree use keys as follows.
To go into the branch i.e., from main menu level to submenu level or submenu level to function level. If you go were already at function
level then that function will be executed.
Selects other options/branches at the same level.
Takes you one level back. If you were already at main menu level then this key takes you back to measurement mode.
Programming functions involve one or more of the following types of data.
1. Selecting an option from option list – Initially present option is displayed in second line of displayed and display blinks to indicate that there
are other choices. Use the key to scroll through all options one by one. Using will cause that option to be selected and also completes
selection process.
2. Entering a numeric value-You can enter +ve or –ve number in floating point with exponent notation.
-9
Rang : 0.0001 x 10
to 9.9999 x 10 9
Format: d.dddd E-d
e.g 1.2345 E-3, 1.2345E6
Method of entry: Initially present value is displayed in the first line display. A flashing digit will change by using key. Select next digit
position in sequence and terminates entry of the number.
Note:
a) Digits cycle through 0-9 values. A sign fields cycles as “ “ and “ – “ for exponent negative
b) When you enter a value beyond its limit then you get message d.ddddEd (MIN. VALUE) or d.ddddEd (MAX VALUE). Press after
you have noted limit and then correct value to be in valid limits.
c) If you don’t want to change value press at the beginning itself !
14
Using keys in programming mode
Key Main menu level Submenu Data level
Function level
Option/Units Numerical values / strings
Select main menu Select submenu or function select next Change flashing digit or
Proposal Character Proposal
Enter displayed Enter displayed submenu Shift flashing digit or
Main menu (execute) function character position
Quit programming Return to main menu Select displayed Enter the displayed numerical
Mode and go back or submenu proposal and return value or alphanumerical
To measurement to function level if no string. Then return to
Mode further data entry is function level if no further
required by the required by that function
function
Newly entered data will be saved in non-volatile memory and accepted by measuring program only after termination of programming mode as
described below.
Pressing the key at main menu level quits the programming mode. Before exiting programming mode, device updates the configuration
according to the changes done in the menu.
Programming using HART® based communicator
While in the measuring mode the communicator displays the measured values of flow, the computed current output value and Qmax, Qmin
values programmed into the device. Using the communicator, you can see / modify many of the device parameters. For details of using HART®
Communicator please refer the HART
reference. Please refer 6.2.12 for details of HART® programmability.
®
Communicator manual. The Generic Online Menu of the communicator is given in 6.2.14 for your quick
15
5.5 Error messages
5.5.1 Error messages in Measurement mode
Error message Type Description Corrective action required
INTL ERR. nn*Internal error in Switch off the power and try again. If the problem converter operation persists contact KHRONE MARSHALL service.
NO SIGNAL NNo signal from No flow through the primary or Vortex sensor problem. the vortex sensor If sensor problem, contact KHRONE MARSHALL service.
LOW FREQ NVortex frequency Check Flow rate>q min else call KHRONE MARSHALL
Service.
too low
HIGH FREQ NVortex frequency Check Flow rate<q max else call KHRONE MARSHALL
Service.
too high
LOW FLOW NFlow rate lower than minimum flow Converter will continue to display actual flow rate. However
rate q min accuracy of measurement may suffer.
HIGH FLOW NActual flow rate Corrective action depends on application process. If flow higher than q max rate exceeds too much further it may damage entire instrument
physically!
INV. CONFIG FConfiguration date Check entire configuration again. If error persists-call in non-volatile memory KHRONE MARSHALL service.
is not valid.
UC FAIL F Micro-Controller failure Contact KHRONE MARSHALL Service.
LCDC FAIL F LCD Controller IC (s) has Contact KHRONE MARSHALL Service.
failed (This error is
applicable to users having
display option)
KEY FAIL FFailure of Keys Contact KHRONE MARSHALL Service. (applicable to instrument
which have display option)
NOVRAM FAIL FNon Volatile Memory Contact KHRONE MARSHALL Service.
has failed
LINE INTER NSupply interrupted Press keys to reset this error.
· This text is displayed for Fatal Errors. The nature of error is beyond the scope of user (failure of IC or other hardware inside converter electronics
etc.). All that can be done is to switch OFF power and then try again. If error message continues, call KHRONE MARSHALL service. internal
error nn = internal error number.
· Type N indicates non-fatal errors whereas F indicates fatal errors.
· Measurement stops if any Fatal error is encountered. This means that device makes flow rate equal to zero,
· Current output to minimum value 4m A & pulse output OFF.
· When errors are displayed during the measuring mode, 'n Err" (n = number) will appear in the 1st line. n gives the number of momentarily
occurring errors that are displayed alternately with the actual measured value.
· Error messages disappear when their cause disappears
5.5.2 Error Messages in Programming mode
Error in programming mode can only occur while entering any numerical value. When you enter numerical value outside possible limits you get
message " n.nnn E n" in 1st line and MIN. VALUE or MAX VALVE or in the 2nd line ("MIN VALUE" if entered value is less than lower limit and
“MAX VALVE “ if higher limit is crossed). Note the message indicating permissible limit then press the key to continue.
16
Loading...
+ 35 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.