VARIFLUX electromagnetic flowmeters are precision measuring instruments designed for the
linear flow measurement of process liquids.
The process liquids must be electrically conductive: ≥ 5 µS/cm ≥ 20 µS/cm for demineralized cold water
The full-scale rangeQ
DN 2.5 - 150 / 1/10” - 6” Q
This is equivalent to a flow velocity of 0.3 - 12 m/s, or 1 - 40 ft/s.
can be set as a function of the meter size:
100%
VARIFLUX electromagnetic flowmeters are designed solely for measuring the volumetric flowrate
of electrically conductive, liquid process products.
Special codes and regulations apply to their use in hazardous locations, and these are referred to
in the special “EEx” installation and operating instructions (supplied only with hazardous-duty
equipment).
Please note: Flowmeters for use in hazardous locations (EEx) have different dimensions and
changed constructions.
Responsibility as to suitability and intended use of these electromagnetic flowmeters rests solely
with the operator.
Improper installation and operation of the flowmeters (systems) may lead to loss of warranty.
In addition, the ”General conditions of sale” forming the basis of the purchase contract are
applicable.
If VARIFLUX flowmeters need to be returned to Krohne, please note the information given on the
last-but-one page of this manual. Krohne regret that they cannot repair or check your
flowmeter(s) unless accompanied by the completed form sheet.
System description
= 0.01 - 763 m3/hr = 0.02 - 3361 US Gal/min
100%
Product liability and warranty
2 Installation instructions VARIFLUX 6000
Standards and approvals
Please refer to the installation and operating instructions for the signal converter.
The VARIFLUX flowmeters are approved to 3A and to EHDEG for the
adapters to DIN 11850, ISO 2037 and DIN 11864-2a.
Items included with supply
VARIFLUX 6000 flow sensor
• flow sensor in the size as ordered • compact flowmeter in the size as ordered
• certificate of calibration data • certificate of calibration data
Weld on to … Screwed to … Clamp joint to … Flanged to
11850
ISO 2037 DIN 11851
(dairy screw)
VARIFLUX 6010 K, VARIFLUX 6020 K and
VARIFLUX 6080 K compact flowmeters
• installation and operating instructions for the
signal converter
SMS
1145
ISO 2852 TRI-
CLOVER
DIN
32676
11864-2a
DIN
Installation instructions VARIFLUX 6000 3
1 Important information for installation: PLEASE NOTE !
• Handling
Do not lift flowmeter by the signal converter
housing or the terminal box.
Do not set flowmeter down on signal converter
housing or terminal box.
PLEASE NOTE: the temperature limits for storage and transport, see below.
• Use only solventless detergents to clean the signal converter housing (polycarbonate).
• Temperatures
Refer to Section 11 “Limits” for operating pressure and vacuum load based on flange
standards and type of tube liner.
Ambient temperature Process temperature
Compact systems -25 to +60 °C (-13 to +140 °F) -20 to + 60 °C (-4 to +140 °F)
-25 to +40 °C (-13 to +104 °F) -20 to +140 °C (-4 to +284 °F)
VARIFLUX 6000 -25 to +60 °C (-13 to +140 °F) -20 to + 60 °C (-4 to +140 °F)
-25 to +40 °C (-13 to +104 °F) -20 to +140 °C (-4 to +284 °F)
• Location and position as required,
but electrode axis
X – • – • – • – X
must be approximately horizontal in a horizontal pipe run.
Y terminal box or converter housing
• Measuring tube must be completely filled at all times.
• Direction of flow is arbitrary. Arrow on flowmeter can normally be ignored. For exceptions,
refer to Section “Factory settings” in the installation and operating instructions for the signal
converter.
• Stud bolts and nuts: to fit, make sure there is sufficient room next to the pipe flanges.
• Vibration: support the pipeline on both sides of the compact flowmeter.
Level of vibration in conformity with IEC 068-2-34: 20-500 Hz, random / 2g runs / 30 minutes /
x, y, z directions.
• Do not expose to direct sunlight, fit a sunshade if necessary, not included with flowmeter,
to be provided by customer.
4 Installation instructions VARIFLUX 6000
• Strong electromagnetic fields, avoid in vicinity of flowmeter
• Straight inlet run minimum of 5 × DN and outlet run minimum of 2 × DN,
(DN = meter size), measured from the electrode axis.
• Vortex and corkscrew flow: increase length of inlet and outlet runs or install flow
conditioners.
• Mixing different process liquids: install flowmeter upstream of mixing point or at an
adequate distance downstream (minimum of 30 × DN), otherwise display may be unsteady.
• Insulated pipeline: do not insulate flowmeter.
• Zero setting not necessary. To check, it should be possible to set “zero” flow velocity in the
completely filled measuring tube. Shutoff valves should therefore be provided either
downstream of the flowmeter or upstream and downstream of the flowmeter.
2 Suggestions for installation
To avoid measuring errors due to gas/air inclusion or to pipe running empty, please observe the
following:
Highest point of pipe run Downpipe over 5 m (16 ft) length
(Air bubbles collect in measuring
tube - faulty measurements!) flowmeter.
Install air valve ✇ downstream of
Preferred
locations
Downpipe
“Zero” flow
velocity.
Line drained.
Faulty
measurements!
Long pipeline
Always install control and shutoff valves
downstream of flowmeter.
Horizontal pipe run
Install in slightly inclining pipe section. If not
possible, assure adequate velocity to prevent
air, gas or vapour from collecting in upper
part of flow tube.
Pumps
Open feed or discharge Never install flowmeter on pump suction side.
Install meter in low section of pipe.
open discharge
Installation instructions VARIFLUX 6000 5