All rights reserved. It is prohibited to reproduce this documentation, or any part thereof, without
the prior written authorisation of KROHNE Messtechnik GmbH.
8.3.2 UFC 030 signal converter...................................................................................................... 67
9 Notes68
4
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
Page 5
UFM 530 HT
1.1 Software history
Release dateHardwareSoftwareDocumentation
1990-052134721100-200μP2: 90-05MA UFM 530 HT R01
1.2 Intended use
CAUTION!
Responsibility for the use of the measuring devices with regard to suitability, intended use and
corrosion resistance of the used materials against the measured fluid lies solely with the
operator.
INFORMATION!
The manufacturer is not liable for any damage resulting from improper use or use for other than
the intended purpose.
SAFETY INSTRUCTIONS 1
microcontroller: 01.13
DSP: 10.20
This product is designed for the measurement of liquids with high temperatures
up to 500°C / 932°F (Ex hazardous areas are limited to 440°C / 824°F).
1.3 Certifications
1.3.1 CE-certification
The device fulfils the statutory requirements of the following EC directives:
• Low voltage directive (73/23/EEC), product is designed in accordance with EN IEC 61010-1
first and second edition (safety requirements for electrical equipment for measurement,
control and laboratory use part 1).
• Pressure equipment directive (Module H of 97/23/EC, full quality assurance).
• ATEX directive (94/9/EC)
DANGER!
Local safety regulations shall be observed in combination with the measures special to this
product to avoid dangerous situations.
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
5
Page 6
1 SAFETY INSTRUCTIONS
1.4 Safety instructions from the manufacturer
1.4.1 Copyright and data protection
The contents of this document have been created with great care. Nevertheless, we provide no
guarantee that the contents are correct, complete or up-to-date.
The contents and works in this document are subject to copyright. Contributions from third
parties are identified as such. Reproduction, processing, dissemination and any type of use
beyond what is permitted under copyright requires written authorisation from the respective
author and/or the manufacturer.
The manufacturer tries always to observe the copyrights of others, and to draw on works created
in-house or works in the public domain.
The collection of personal data (such as names, street addresses or e-mail addresses) in the
manufacturer's documents is always on a voluntary basis whenever possible. Whenever
feasible, it is always possible to make use of the offerings and services without providing any
personal data.
UFM 530 HT
We draw your attention to the fact that data transmission over the Internet (e.g. when
communicating by e-mail) may involve gaps in security. It is not possible to protect such data
completely against access by third parties.
We hereby expressly prohibit the use of the contact data published as part of our duty to publish
an imprint for the purpose of sending us any advertising or informational materials that we have
not expressly requested.
1.4.2 Disclaimer
The manufacturer will not be liable for any damage of any kind by using its product, including,
but not limited to direct, indirect or incidental and consequential damages.
This disclaimer does not apply in case the manufacturer has acted on purpose or with gross
negligence. In the event any applicable law does not allow such limitations on implied warranties
or the exclusion of limitation of certain damages, you may, if such law applies to you, not be
subject to some or all of the above disclaimer, exclusions or limitations.
Any product purchased from the manufacturer is warranted in accordance with the relevant
product documentation and our Terms and Conditions of Sale.
The manufacturer reserves the right to alter the content of its documents, including this
disclaimer in any way, at any time, for any reason, without prior notification, and will not be liable
in any way for possible consequences of such changes.
6
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
Page 7
UFM 530 HT
1.4.3 Product liability and warranty
The operator shall bear responsibility for the suitability of the device for the specific purpose.
The manufacturer accepts no liability for the consequences of misuse by the operator. Improper
installation and operation of the devices (systems) will cause the warranty to be void. The
respective "Standard Terms and Conditions" which form the basis for the sales contract shall
also apply.
1.4.4 Information concerning the documentation
To prevent any injury to the user or damage to the device it is essential that you read the
information in this document and observe applicable national standards, safety requirements
and accident prevention regulations.
If this document is not in your native language and if you have any problems understanding the
text, we advise you to contact your local office for assistance. The manufacturer can not accept
responsibility for any damage or injury caused by misunderstanding of the information in this
document.
This document is provided to help you establish operating conditions, which will permit safe and
efficient use of this device. Special considerations and precautions are also described in the
document, which appear in the form of underneath icons.
SAFETY INSTRUCTIONS 1
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
7
Page 8
1 SAFETY INSTRUCTIONS
1.4.5 Warnings and symbols used
Safety warnings are indicated by the following symbols.
DANGER!
This information refers to the immediate danger when working with electricity.
DANGER!
This warning refers to the immediate danger of burns caused by heat or hot surfaces.
DANGER!
This warning refers to the immediate danger when using this device in a hazardous atmosphere.
DANGER!
These warnings must be observed without fail. Even partial disregard of this warning can lead to
serious health problems and even death. There is also the risk of seriously damaging the device
or parts of the operator's plant.
UFM 530 HT
WARNING!
Disregarding this safety warning, even if only in part, poses the risk of serious health problems.
There is also the risk of damaging the device or parts of the operator's plant.
CAUTION!
Disregarding these instructions can result in damage to the device or to parts of the operator's
plant.
INFORMATION!
These instructions contain important information for the handling of the device.
LEGAL NOTICE!
This note contains information on statutory directives and standards.
• HANDLING
HANDLING
HANDLINGHANDLING
This symbol designates all instructions for actions to be carried out by the operator in the
specified sequence.
iRESULT
RESULT
RESULTRESULT
This symbol refers to all important consequences of the previous actions.
1.5 Safety instructions for the operator
WARNING!
In general, devices from the manufacturer may only be installed, commissioned, operated and
maintained by properly trained and authorized personnel.
This document is provided to help you establish operating conditions, which will permit safe and
efficient use of this device.
8
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
Page 9
UFM 530 HT
2.1 Scope of delivery
INFORMATION!
Do a check of the packing list to make sure that you have all the elements given in the order.
INFORMATION!
Inspect the cartons carefully for damages or signs of rough handling. Report damage to the
carrier and to the local office of the manufacturer.
DEVICE DESCRIPTION 2
Figure 2-1: Standard scope of delivery, independent from the sensor version illustrated below
1 Signal converter UFC 030 in ordered version
2 Product documentation
3 Factory calibration report
4 Special wrench for opening the converter housing
5 Signal cable
INFORMATION!
Assembly materials and tools are not part of the delivery. Use the assembly materials and tools
in compliance with the applicable occupational health and safety directives.
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
9
Page 10
2 DEVICE DESCRIPTION
2.2 Device description
The UFM 530 HT is a bi-directional flowmeter (the arrow on the flowmeter indicates the positive
direction). It consists of a UFS 500 HT flow sensor and an UFC 030 signal converter.
The signal converter is installed separately from the high-temperature flow sensor.
The flow sensor is designed for process temperatures between -25...+500°C / -13...+932°F (Ex
hazardous areas are limited to -25...+440°C / -13...+824°F) and is manufactured from
316 stainless steel. Depending on the diameter it uses single beam or dual beam technology. All
coax cables are protected with metal tubes. On top of the flow sensor a connection box is
mounted.
INFORMATION!
Other diameters, pressure classes, materials or customized designs are available on request.
UFM 530 HT
Figure 2-2: Available sensor versions
1 Building construction "single beam" (DN25...40)
2 Building construction "single beam" (DN50...80)
3 Building construction "dual beam" (≥DN100)
2.3 Nameplates
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate.
10
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
Page 11
UFM 530 HT
2.4 Example nameplate
DEVICE DESCRIPTION 2
Figure 2-3: Example of nameplate
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
11
Page 12
3 INSTALLATION
3.1 General notes on installation
INFORMATION!
Inspect the cartons carefully for damages or signs of rough handling. Report damage to the
carrier and to the local office of the manufacturer.
INFORMATION!
Do a check of the packing list to make sure that you have all the elements given in the order.
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate.
3.2 Storage
• Store the device in a dry, dust-free location.
• Avoid continuous direct sunlight.
• Store the device in its original packaging.
• Storage temperature: -50...+70°C / -58...+158°F
UFM 530 HT
3.3 Transport
Figure 3-1: Correct lifting with the help of straps or crane hooks.
12
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
Page 13
UFM 530 HT
3.4 Inlet and outlet
INFORMATION!
To avoid measuring errors and to get proper measuring results an inlet section upstream and an
outlet section downstream is necessary. Regard the following illustration and table and regard
the information for different fluid products:
Figure 3-2: Recommended straight inlet and outlet
1 DN25...80: ≥ 50 DN
DN100...300: ≥ 15 DN
2 DN25...80: ≥ 10 DN
DN100...300: ≥ 5DN
INSTALLATION 3
INFORMATION!
Different fluid products
Different fluid products
Different fluid productsDifferent fluid products
To mix different fluid products, install the flow meter upstream of mixing point or at minimum
distance of 30 DN downstream of the mixing point.
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
13
Page 14
3 INSTALLATION
3.5 Mounting
3.5.1 Mounting position of the flow sensor
Installation of the flow sensor is allowed in horizontal, slightly ascending and vertical pipe
sections with up going flow direction (see next section). If installed in a horizontal or slightly
ascending pipeline, the connection box of the flow sensor has to be up or down.
UFM 530 HT
Figure 3-3: Allowed position of the flow sensor (up and down)
WARNING!
Do not unscrew the flanged transducer construction. This will cause direct contact with the high
temperature liquid running through the flow sensor.
3.5.2 Mounting location of the flow sensor
INFORMATION!
For proper flow measurement the measuring tube must be completely filled at all times. When
the sensors become non-wetted, a loss of signal message will be displayed. There is no damage
when this occurs.
Observe the following precautions to avoid measuring errors or malfunctioning of the flow meter
due to gas or air inclusions or an empty pipe.
Since gas will collect at the highest point of a pipe, installation of the flowmeter at that location
should always be avoided. In long horizontal pipes the flow meter has to be installed in a slightly
ascending pipe section. If not possible, ensure adequate velocity to prevent air, gas or vapour
from collecting in the upper part of the flow tube.
14
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
Page 15
UFM 530 HT
INSTALLATION 3
Figure 3-4: Avoid locations where gas can be present.
Also installation in a down going pipe should be avoided since a completely filled pipe may not be
guaranteed due to cascading effects. Additionally flow profile distortion is possible.
Figure 3-5: Avoid locations where gas can be present
Figure 3-6: Ensure you have a completely filled pipe.
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
15
Page 16
3 INSTALLATION
Figure 3-7: Air vent
1 Level difference > 5 m / 16 ft
2 Install an air vent.
UFM 530 HT
Figure 3-8: Install control valve downstream of the flow meter.
Figure 3-9: Install pump upstream of flowmeter.
16
Figure 3-10: Avoid installation in vibrating pipelines.
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
Page 17
UFM 530 HT
3.5.3 Insulation
WARNING!
Complete insulation of the UFS 500 HT flow sensor is prohibited. Insulation is allowed up to the
first flange of each transducer.
WARNING!
The connection box and the flanged transducers require adequate cooling by ambient air and
must be protected against heat radiation by surrounding equipment.
INSTALLATION 3
Figure 3-11: Approved insulations.
1 Building construction "single beam" for DN25...40
2 Building construction "single beam" for DN50...80
3 Building construction "dual beam" for ≥ DN100
3.5.4 Pipe flanges
INFORMATION!
Refer to dimensional drawings for flange spacing and in addition allowance for thickness of
gaskets.
Figure 3-12: Maximum deviation between flanges
1 L
max
2 L
min
CAUTION!
Max. permissible deviation of pipe flange faces:
- L
L
max
≤ 0.5 mm / 0.02"
min
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
17
Page 18
3 INSTALLATION
3.5.5 Pipes with cathodic protection
Pipes with electric corrosion protection are generally insulated inside and outside so that the
fluid has no conductive connection to the ground. The flow meter must be insulated from the
pipe. Observe the following instructions when installing the flow meter:
• The pipe flanges must be connected to each other using a copper cable (L), but must not be
connected to the flowmeter.
• The bolts for the flange connections and the gaskets must be insulated. Use sleeves and
washers that are made of insulating material (these must be provided by customer).
UFM 530 HT
Figure 3-13: Cathodic protection
1 Flanges (left one: of flow sensor, right one: of pipe)
2 Nut
3 Washer
4 Insulating sleeve
5 Bolt
6 Gasket
18
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
Page 19
UFM 530 HT
4.1 Safety instructions
DANGER!
All work on the electrical connections may only be carried out with the power disconnected. Take
note of the voltage data on the nameplate!
DANGER!
Observe the national regulations for electrical installations!
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex
documentation.
WARNING!
Observe without fail the local occupational health and safety regulations. Any work done on the
electrical components of the measuring device may only be carried out by properly trained
specialists.
ELECTRICAL CONNECTIONS 4
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate.
4.2 Power supply of the converter
Environmental conditions
The flowmeter is designed to operate safe under the following conditions. Observe them before
the connection to the mains supply voltage is established:
• Indoor and outdoor use, usable up to protection category IP67 according to IEC 60529
(Note: IP67 is only warranted when using suitable cabling with the cable glands and covers
mounted as specified).
• Maximum altitude: up to 2000 m above see level.
• Maximum relative humidity: up to 80%.
• Operation ambient temperature range: -40...+65°C.
• Storage temperature range: -40...+70°C.
CAUTION!
Never allow dirt to accumulate on the gasket of the rear (blind) cover. A dirty gasket has to be
cleaned, a damaged gasket must be replaced immediately.
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
19
Page 20
4 ELECTRICAL CONNECTIONS
Before the cables can be fastened to the power supply terminal, the rear (blind) cover has to be
removed.
Figure 4-1: Terminals for power supply
UFM 530 HT
Item number FunctionSpecification
1Neutral power supply.
2Life power supply.Mains voltage AC supply: 100 VAC < U < 240 VAC (-15%, +10%),
Protective conductor clamp terminal. Conductors up to
4mm2(11 AWG) need to be connected to this terminal.
20
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
Page 21
UFM 530 HT
4.3 Connection of signal cables
Connect the signal cable from the connection box of the UFS 500 HT sensor to the UFC 030
signal converter according to underneath drawings for the single beam and dual beam
construction.
ELECTRICAL CONNECTIONS 4
Figure 4-2: Connection of the sensor cables for dual beam (left) and
single beam (right) building construction (sensor side)
Figure 4-3: Connection of the sensor cables, converter side
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
21
Page 22
4 ELECTRICAL CONNECTIONS
4.4 Electrical connections of the signal inputs and outputs
CAUTION!
For wiring of the signal inputs and outputs it is advised to use unshielded twisted pairs.
CAUTION!
Please observe instrument polarity: current (I) is always flowing towards I, C, P, A1, A2 terminals
(current sink).
The signal inputs and outputs terminals are located in the converter terminal box. It is
accessible after removing the rear (blind) cover of the converter. There are versions for non Ex
and for Ex applications.
4.4.1 Non Ex versions
Non Ex standard version
UFM 530 HT
Figure 4-4: Terminals for standard instrument.
TerminalFunctionSpecification
1DC power supply from converter for active
2Combined current output (I) and digital
3Pulse / frequency output.I
4Analog input 2, for temperature or
5Analog input 1, for temperature
6Common ground-
wiring of inputs and outputs.
input (C). Current output (I) includes
®
-Communication.
HART
pressure measurement.
measurement.
22 VDC at full load, 24 VDC maximum,
I ≤ 100 mA.
Current output (I): I ≤ 22 mA,
≤ 680 Ω, U
R
load
Digital input (C): low = 0...5 VDC,
high = 15...32 VDC (will be switched off
when current output activated).
= 150 mA, U
max
maximal frequency = 2 kHz.
0(4)...20 mA, Ri= 58,2 Ω, fuse = 50 mA.
0(4)...20 mA, Ri= 58,2 Ω, fuse = 50 mA.
max
max
= 15 VDC.
CAUTION!
Never use the active and passive mode at the same terminal simultaneously.
®
If HART
-Communication is used, do not connect the pulse/frequency output P in active mode.
=32VDC/ 24VAC,
22
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
Page 23
UFM 530 HT
ELECTRICAL CONNECTIONS 4
INFORMATION!
The electrical input and output signals can be connected either in active or in passive mode. In
active mode DC supply voltage is provided from the terminal V+. In passive mode supply voltage
is provided from an external source.
Non Ex version with Profibus PA
Figure 4-5: Terminals for instrument with Profibus PA (non Ex).
TerminalFunctionSpecification
1Communication connection -For Fieldbus communication
2Communication connection +For Fieldbus communication
3Combined current output (I), digital
input (C) and Pulse / frequency output.
Current output (I) includes HART
Communication.
4Analog input 2, for temperature or
pressure measurement.
5Analog input 1, for temperature
measurement.
6Common ground
®
-
Current output (I): I ≤ 22 mA,
R
≤ 680 Ω, U
load
Digital input (C): low = 0...5 VDC,
high = 15...32 VDC (will be switched off
when current output activated).
Pulse output: I
U
= 32 VDC / 24 VAC, maximal
max
frequency = 2 kHz.
0(4)...20 mA, Ri= 58,2 Ω, fuse = 50 mA.
0(4)...20 mA, Ri= 58,2 Ω, fuse = 50 mA.
max
= 150 mA,
max
= 15 VDC.
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
23
Page 24
4 ELECTRICAL CONNECTIONS
Non Ex HiPower version
Figure 4-6: Terminals for non Ex HiPower instrument.
TerminalFunctionSpecification
UFM 530 HT
1DC power supply from converter for active
wiring of inputs and outputs.
2Combined current output (I) and digital
input (C). Current output (I) includes
®
HART
-Communication.
22 VDC at full load, 24 VDC maximum,
I ≤ 100 mA.
Current output (I): I ≤ 22 mA,
R
≤ 680 Ω, U
load
max
= 15 VDC.
Digital input (C): low = 0...5 VDC,
high = 15...32 VDC (will be switched off
when current output activated).
3Pulse / frequency output.I
= 150 mA, U
max
=32VDC/ 24VAC,
max
maximal frequency = 2 kHz.
4Analog input 2, for temperature or
0(4)...20 mA, Ri= 58,2 Ω, fuse = 50 mA.
pressure measurement.
5Analog input 1, for temperature
0(4)...20 mA, Ri= 58,2 Ω, fuse = 50 mA.
measurement.
6Common ground-
24
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
Page 25
UFM 530 HT
4.4.2 Ex versions
WARNING!
The electrical input and output signals must be connected in passive mode. The supply voltage
must be provided from an external source.
Ex standard version
Figure 4-7: Terminals for Ex standard instrument
ELECTRICAL CONNECTIONS 4
TerminalFunctionSpecification
1Neutral mains power supply100...240 VAC, 24 VAC or 24 VDC
2Live mains power supply100...240 VAC, 24 VAC or 24 VDC
3Combined current output (I) and digital
4Pulse / frequency outputI
5Analog input 2, for temperature or
6Analog input 1, for temperature
7Common ground
input (C). Current output (I) includes
®
-Communication.
HART
pressure measurement.
measurement.
Current output (I): I ≤ 22 mA,
≤ 680 Ω, U
R
load
Digital input (C): low = 0...5 VDC,
high = 15...32 VDC (will be switched off
when current output activated).
= 150 mA, U
max
maximal frequency = 2 kHz.
0(4)...20 mA, Ri= 58,2 Ω, fuse = 50 mA.
0(4)...20 mA, Ri= 58,2 Ω, fuse = 50 mA.
= 15 VDC.
max
=32VDC/ 24VAC,
max
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
25
Page 26
4 ELECTRICAL CONNECTIONS
Ex NAMUR version
Figure 4-8: Terminals for Ex instrument with NAMUR
TerminalFunctionSpecification
1Neutral mains power supply100...240 VAC, 24 VAC or 24 VDC
2Live mains power supply100...240 VAC, 24 VAC or 24 VDC
3Combined current output (I) and digital
input (C). Current output (I) includes
®
HART
-Communication.
4Pulse / frequency outputI
5Ground for pulse output
6Analog input 1, for temperature
7Common ground
measurement.
UFM 530 HT
Current output (I): I ≤ 22 mA,
R
≤ 680 Ω, U
load
Digital input (C): low = 0...5 VDC,
high = 15...32 VDC (will be switched off
when current output activated).
= 150 mA, U
max
maximal frequency = 2 kHz.
0(4)...20 mA, Ri= 58,2 Ω, fuse = 50 mA.
= 15 VDC.
max
=32VDC/ 24VAC,
max
INFORMATION!
The current output of the UFC 030 F-EEx can be set according to NAMUR NE43. The current
output will go either to 3.6 or 21.5 mA in case of failure indication.
26
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
Page 27
UFM 530 HT
ELECTRICAL CONNECTIONS 4
The following Ex-i Modis versions have two Modis modules, providing intrinsically safe input /
output circuits. Modis versions don't have analogue inputs A1 / A2.
Ex-i (Modis) version
Figure 4-9: Terminals for instrument with Ex-i (Modis)
TerminalFunctionSpecification
1Neutral mains power supply100...240 VAC, 24 VAC or 24 VDC
2Live mains power supply100...240 VAC, 24 VAC or 24 VDC
3Ground for pulse, frequency or status
output
4Pulse, frequency or status outputI
5Current outputCurrent output (I): I ≤ 22 mA,
6Ground for current output
= 150 mA, U
max
maximal frequency = 2 kHz.
R
≤ 680 Ω, U
load
=32VDC/ 24VAC,
max
= 15 VDC.
max
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
27
Page 28
4 ELECTRICAL CONNECTIONS
Ex-i (Modis) version with Profibus PA
Figure 4-10: Terminals for Ex-i (Modis) version with Profibus PA
TerminalFunctionSpecification
1Neutral mains power supply100...240 VAC, 24 VAC or 24 VDC
2Live mains power supply100...240 VAC, 24 VAC or 24 VDC
3Profibus communication -
4Profibus communication +
5Current outputCurrent output (I): I ≤ 22 mA,
6Ground for current output
R
load
≤ 680 Ω, U
max
UFM 530 HT
= 15 VDC.
28
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
Page 29
UFM 530 HT
4.5 Connection diagram examples
INFORMATION!
The connection diagrams on the next pages are valid for most versions. However, not all
versions have the same connection possibilities. Versions that don't have the V+ terminal (for
instance Ex versions) can only be connected in a passive way, e.g. by using an external power
supply.
4.5.1 Current output
Current output active
• V+: 22 VDC at full load, 24 VDC maximum, I ≤ 100 mA
≤ 680 Ω
• R
L
• I < 22 mA
• U
= 15 VDC
max
ELECTRICAL CONNECTIONS 4
Figure 4-11: Current output (active)
Current output passive
• U
• I ≥ 22 mA (for supply)
• For Ex-i modis versions: I = 4..20 mA
• For Ex-i modis versions: R
Figure 4-12: Current output (passive)
= 15...24 VDC
ext
(For Ex-i modis versions: U
= 8,1...30 VDC)
ext
≤ (U
L
- 8) / 0,022
ext
INFORMATION!
Ex-i modis versions do not have an "I/C" terminal. Use the "I" terminal instead.
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
29
Page 30
4 ELECTRICAL CONNECTIONS
4.5.2 Pulse output
Pulse output active
• V+: 22 VDC at full load, 24 VDC maximum, I ≤ 100 mA
• Ri ≥ 470 Ω
• frequency ≤ 2 kHz
Figure 4-13: Pulse output (active)
UFM 530 HT
Pulse output passive
• U
• I ≤ 150 mA
• Ri ≥ 470 Ω
• frequency ≤ 2 kHz
Figure 4-14: Pulse output (passive)
≤ 32 VDC or U ≤ 24 VAC
ext
(For Ex-i modis versions: U
= 6...30 VDC)
ext
(For Ex-i modis versions: I ≤ 110 mA)
INFORMATION!
Ex-i modis versions have different terminal codings.
30
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
Page 31
UFM 530 HT
Figure 4-15: Pulse output (passive) for Ex-i Modis
4.5.3 Digital input
INFORMATION!
The digital input will be switched off when the current output is activated.
Digital input active
• V+: 22 VDC at full load; 24 VDC maximum, I ≤ 100 mA
• Low = 0...5 VDC
• High = 15...32 VDC
ELECTRICAL CONNECTIONS 4
Figure 4-16: Digital input (active)
Digital input passive
• U
• I ≥ 1.5 mA
• Low = 0...5 VDC
• High = 15...32 VDC
= 15...30 VDC
ext
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
31
Page 32
4 ELECTRICAL CONNECTIONS
Figure 4-17: Digital input (passive)
4.5.4 Analog input
Analog input
• 0 (4)...20 mA
• Ri = 58.2 Ω
• Fuse 50 mA
UFM 530 HT
Figure 4-18: Analog input
32
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
Page 33
UFM 530 HT
5.1 Start-up procedure
DANGER!
Hazardous voltages are present within this product during normal operation. Do not operate it
with the covers removed!
When powered, "start-up" is shown at the display for a short time. After that the normal
measuring mode starts.
INFORMATION!
The flowmeter is programmed at the factory according to your order. No changes are necessary.
START-UP 5
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
33
Page 34
6 OPERATION
6.1 Signal converter: front panel and operating keys
The front panel and its operating keys are accessible after removing the front cover of the
electronics section (use the special wrench for it). If it is not allowed to open the housing, for
instance in hazardous areas, you can program the converter with the magnetic pin.
CAUTION!
When removing the cover, do not damage the screw thread and the gasket, never allow dirt to
accumulate and make sure that they are well greased using Teflon. A damaged gasket must be
replaced immediately!
UFM 530 HT
1 Magnetic sensors to program the signal converter by means of a hand-held bar magnet (optional) without having to
remove the front cover (left sensor is equivalent to the "→" ke y, r igh t se nso r to "↑" key and sensor at the top to "^" key.
2 Compass needles.
3 Measured value ("E" stands for "Exponent").
4 Unit of measured value.
5 Indication of actual displayed value.
6 Operating keys (→, ^, ↑) for programming the signal converter.
Function of operating keys depending on selected mode
Key / symbolMeasuring modeMenu modeData level
→Go to the parameter setting
mode. If access CODE 1 is
activated, CODE 1 must be
entered first.
^Go to the error/totalizer
reset mode (via CODE 2).
↑Cycle through measured
values.
Go to the next, lower menu
level.
Return to the previous
(higher) menu level or
leave the menu mode.
Cycle through menu
options within actual menu
level.
Go to the next character or
change line (only when two
lines are displayed).
Accept entered value.
Raise active digit.
34
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
Page 35
UFM 530 HT
6.2 Available versions
INFORMATION!
The flowmeter is programmed at the factory according to your order. No changes are necessary.
All standard UFC 030 signal converters can be programmed in menu 3.02.01.
1. Standard version
1. Standard version
1. Standard version1. Standard version
2. CORR T: on-site interchangeable to "standard".
2. CORR T: on-site interchangeable to "standard".
2. CORR T: on-site interchangeable to "standard".2. CORR T: on-site interchangeable to "standard".
Temperature correction via analogue input 1 (temp range -50... +150°C) ; with this version the
outputs are non-Ex-I.
3. CORR T+P: on-site interchangeable to "standard".
3. CORR T+P: on-site interchangeable to "standard".
3. CORR T+P: on-site interchangeable to "standard".3. CORR T+P: on-site interchangeable to "standard".
Temperature correction via analogue input 1, pressure correction via analogue input 2 (press
range 0...100 bar) ; with this version the outputs are Ex-I.
Corrected volume calculation can be done based on temperature correction or temperature and pressure correction.
The volume correction factor is based on API 2540; chapter 11.1 for the temperature correction
and chapter 11.2 for the pressure correction.
For liquid oil products the volume correction factors can be estimated as 0.1% per degree
Celsius and 0.01% per bar as a rule of thumb. Therefore for normal industrial processes these
corrections are hardly ever applied.
OPERATION 6
4. HEAT: on-site interchangeable to "standard".
4. HEAT: on-site interchangeable to "standard".
4. HEAT: on-site interchangeable to "standard".4. HEAT: on-site interchangeable to "standard".
The UFC 030 signal converter is programmed for energy measurement, based on temperature
measurements connected via analogue inputs A1 and A2 (temperature range: -50...+150°C);
with this version the outputs are non-Ex-I. The following settings are needed:
No. of menuDisplay textSetting/description and functions
3.02.02INP1 4mAminimum value of temperature sensor 1 on inlet
3.02.03INP1 20mAmaximum value of temperature sensor 1 on inlet
3.02.04INP1 4mAminimum value of temperature sensor 2 on outlet
3.02.05INP1 20mAmaximum value of temperature sensor 2 on outlet
3.02.12full scalexxx (unit yyy)
3.03.02function
totalizer
3.03.05unit for total
energy
3.05.01function pulse
output
3.05.10pulse/unitselect unit
select direction
select unit
CORR FLOW
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
35
Page 36
6 OPERATION
5. BATCH: on-site interchangeable to "standard".
5. BATCH: on-site interchangeable to "standard".
5. BATCH: on-site interchangeable to "standard".5. BATCH: on-site interchangeable to "standard".
The following settings are needed:
No. of menuDisplay textSetting/description and functions
UFM 530 HT
3.04.01FUNCTION
3.05.01FUNCTION
3.06.01FUNCTION
3.02.13BATCH VOLxxx(unit yyy)
6. MODIS: setting can not be changed.
6. MODIS: setting can not be changed.
6. MODIS: setting can not be changed.6. MODIS: setting can not be changed.
(current
output)
(pulse output)
(digital output)
OFF
BATCH
BATCH
The analogue inputs 1 and 2 are not available in this version.
The Ex-I 1 version has a current output (incl. HART
The Ex-I 2 version has a current output (incl. HART
INFORMATION!
Depending on the programming of function 3.03.07 the additional or selectable indications can
be manually selected by pressing ↑ key, or they are alternating with the display of the measured
value(s). The corrected volume flow indication or volume flow totalizer indication is marked with
the letter "C" at the left of the display's second (middle) line. The batch totalizer indication is
marked with the letter "B".
6.3 Signal converter: menu structure
INFORMATION!
Since the UFC 030 converter can be equipped with different options, the availability of certain
options depends on the function of the converter.
®
) and a pulse output.
®
) and a Profibus PA output.
The menu structure consists of five user accessible parts. They are described on the following
pages.
Error / totalizer menu
No. of menuDisplay textSetting/description and functions
0.00.01VIEW ERRView error messages list.
0.00.02RST ERRReset error messages:
1) NO RESET: keep error messages list.
2) RESET: reset error messages list.
0.00.03RST TOTALReset display totalizer(s). Note: all totalizer values are reset!
Only availabe when function 3.07.08 is set to YES. Available options:
1) RESET ALL: reset all totalizer values.
2) NO RESET: keep totalizer values.
36
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
Page 37
UFM 530 HT
OPERATION 6
INFORMATION!
The functions in this menu are a subset of menu 3.00.00. They are selected in this menu as most
commonly used functions for a quick installation. Note: parameters set in these functions are
automatically set in both menus!
Operation menu
No. of menuDisplay textSetting/description and functions
1.01.00, FLOW
1.01.01FULL SCALEFull-scale value for 100% volume flow rate.
1.01.02ZERO VALUEZero value.
1.01.03ZERO CALZero calibration.
1.01.04MASTER TCMaster time constant.
1.01.05LF CUTOFFLow-flow cut-off.
1.01.06CUTOFF ONCut-off active.
1.01.07CUTOFF OFFCut-off de-active.
1.02.00, DISPLAY
1.02.01DISP FLOWDisplay of flow.
1.02.02FUNCT TOTFunction of totalizer.
1.02.03TOTAL VOLDisplay of totalizer.
1.03.00, PULSE OUTP (Pulse output)
1.03.01PULSE RATEPulse frequency value for 100% scale.
1.03.02PULSE/UNITPulse value per volume flow unit.
1.03.03PULSE/UNITPulse value per energy unit.
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
37
Page 38
6 OPERATION
Test functions menu
No. of menuDisplay textSetting/description and functions
This menu is for testing the display, the in- and outputs and for information on hard- and software
numbers.
2.01.00, DISPLAY
2.01.01DISPLAYTest display, lights all pixels (end with ^ key).
2.02.00, OUTPUTS
2.02.01CURRENTTest current output (0, 4, 12, 20 and 22 mA).
2.02.02PULSETest pulse/frequency output (1, 10, 100, 1000 and 2000 Hz).
2.03.00, INPUTS
2.03.01AN INP 1Test analog input 1.
2.03.02AN INP 2Test analog input 2.
2.03.03DIG INPUTTest digital input. Measure level at digital input.
2.03.04SENSORSensor status: good, open, short
UFM 530 HT
Use ↑ key to go to next value. Displayed value directly present at
current output. End with ^ key.
Use ↑ key to advance. Displayed value directly present at pulse
output. End with ^ key.
Measure current at analog input 1.
End with ^ key.
Measure current at analog input 2.
End with ^ key.
End with ^ key.
2.04.00, DEV INFO (Device information)
2.04.01MANUFACTDisplay manufacturer.
2.04.02MODEL NODisplay model number.
2.04.03SERIAL NODisplay serial number.
2.04.04UP2 HW NODisplay μP2 hardware number.
2.04.05UP2 SW NODisplay μP2 software number.
2.04.06FRNT HW NODisplay front end hardware number.
2.04.07DSP HW NODisplay D.S.P. hardware number.
2.04.08DSP SW NODisplay D.S.P. software number.
2.04.09TIME COUNTDisplay time counter.
38
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
Page 39
UFM 530 HT
OPERATION 6
Installation menu
No. of menuDisplay textSetting/description and functions
3.01.00, FLOW (Volume flow parameters)
3.01.01FULL SCALEFull-scale value for 100% volume and flow rate units (see
3.01.02ZERO VALUEAlthough zero calibrated at the factory the flow sensor might still give
3.01.03ZERO CALZero calibration (see function 1.01.03). Note: Carry out only at "zero"
3.01.04MASTER TCThe master time constant apply for the display and the current and
3.01.05LF CUTOFFLow-flow cut-off for display and outputs.
3.01.06CUTOFF ONCut off "active" value.
3.01.07CUTOFF OFFCut off "de-active" value. Note: value "off" must be greater than value
3.01.08METER SIZEThe nominal diameter of the measuring tube. See the flow sensor
3.01.09GK VALUEAt the factory, each flow sensor is calibrated and supplied with a
3.01.10FLOW DIRThe forward flow direction is indicated with an arrow on the flow
function 1.01.01). The selection of units may be limited to SI units
only.
Available units: m3/s, m3/min, m3/hr, L/s, L/min, L/hr,
US.Gal/s, US.Gal/min, US.Gal/hr, bbls/hr, bbls/day or free user
configurable unit, set by using function 3.07.05...3.07.07).
an offset flow reading, at "zero" flow in the pipeline. Available
settings:
1) FIXED: factory zero setting.
2) MEASURED: value measured with function 3.01.03 (ZERO CAL).
flow and with completely filled measuring tube!
Duration approximately 15s with display
indicating "BUSY":
1) STORE NO: preserve old zero value.
2) STORE YES: store new zero value.
pulse outputs. It does not apply for totalisation and for the current
output in F/R setting. If required, a different time constant value can
be set for the pulse/frequency output under Function 3.05.06.
Range: 0.02...99.99 s.
1) NO: fixed tripping points (ON = 0.1%, OFF = 0.2%).
2) YES: see function 3.01.06 and 3.01.07.
Range: 1 through 19% of Q
"on"!
Range: 2 through 20% of Q
nameplate. Selection of size from meter size table:
25...3000 mm / 1...120".
calibration constant (GK). This constant can be found on the flow
sensor nameplate and must equal the value on the flow sensor
nameplate.
Range: 0.02...20.
sensor. Available settings:
1) POSITIVE: If the actual flow is in the direction of the arrow then the
flow is in the positive direction and the converter will have a positive
flow reading.
2) NEGATIVE: reversed reading of the converter, useful when the
process flow direction is changed so the flow sensor will not need to
be reversed.
100%
100%
.
.
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
39
Page 40
6 OPERATION
No. of menuDisplay textSetting/description and functions
3.01.11MIN VOSMinimum velocity of sound (VOS). Value used for I0% or P0% when
3.01.12MAX VOSMaximum velocity of sound.Value used for I
3.02.00, VERSION
3.02.01FUNCTIONFunction of converter. This is factory pre-set and can only be changed
3.02.02INP1 4 mA4 mA reference for analog input 1, 4 mA temperature reference.
3.02.03INP1 20 mA20 mA reference for analog input 1, 20 mA temperature reference.
3.02.04INP2 4 mA4 mA reference for analog input 2, 4 mA temperature reference.
3.02.05INP2 20 mA20 mA reference for analog input 2, 20 mA temperature reference.
3.02.06INP2 4 mA4 mA reference for analog input 2, 4 mA pressure reference.
3.02.07INP2 20 mA20 mA reference for analog input 2, 20 mA pressure reference.
3.02.08K0Product constant K0.
UFM 530 HT
function "VOS" selected in function 3.04.01 or 3.05.01.
Units: m/s or feet/s,
range: 0...4999 m/s (0...16.400 feet/s).
or P
function "VOS" selected in function 3.04.01 or 3.05.01. Note:
maximum value must be greater than minimum value!
Units: m/s or feet/s,
range: 1...4999 m/s (0...16.400 feet/s).
to standard from any setting:
1) STANDARD.
2) CORR T: temperature correction of the measured flow via analog
input 1.
3) CORR T + P: temperature and pressure correction of the measured
flow via analog input 1 (pressure) and analog input 2 (temperature).
4) HEAT: reserved for measurement of heat power and heat energy
totalisation.
5) BATCH: batch volume.
6) MODIS: Cannot be altered.
Unit: °Cor°F, range: -50...+150°C / -58...302°F.
Unit: °Cor°F, range: -50...+150°C / -58...302°F.
Unit: °Cor°F, range: -50...+150°C / -58...302°F.
Unit: °Cor°F, range: -50...+150°C / -58...302°F.
Unit: bar(a) or psi(a), range: 0...100 bar(a).
Unit: bar(a) or psi(a), range: 0...100 bar(a).
100%
100%
when
40
Range: 10-9...109.
3.02.09K1Product constant K1.
Range: 10-9...109.
3.02.10K2Product constant K2.
Range: 10-9...109.
3.02.11DENSITY 15Product density at T = 15°C/ 59°F.
Range: 500...2000 kg/m3.
3.02.12FULL SCALEHeat measurement. Full scale setting and unit for heat power.
Available units: GJ/s, GJ/hr, MJ/s, MJ/hr, GCal/s, GCal/hr, MCal/s,
MCal/hr.
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
Page 41
UFM 530 HT
OPERATION 6
No. of menuDisplay textSetting/description and functions
3.02.13BATCH VOLBatch volume total size and units.
Availabe units: m3, Liter, US-Gallon, Barrel or user definable unit.
Range: 0.025...100000 m3.
3.03.00, DISPLAY.
3.03.01DISP FLOWDisplay of flow. Available options:
1) RATE: full-scale units (flow is shown with the unit as set in function
3.01.01).
2) PERCENT: flow is shown as a percentage (0...100%) of the full
scale value.
3) NO DISPLAY: no flow is shown.
3.03.02FUNCT TOTFunction of totalizer. Two totalizers (counters) are available. The
3.03.03DISP TOTALDisplay of totalizer. A totalizer can be selected here for displaying:
3.03.04TOTAL VOLUnit for volume totalizer. Note: the maximal value of the totalizer is
totalizer values are incremented and stored once a second. The
following settings are available:
1) ACT FLOW: actual flow units, used for counting the total volume in
the totalizer (for each direction one totalizer is available, the sum of
the two can also be displayed).
2) CORR FLOW: the corrected flow is used for counting the total
volume in the totalizers.
3) POS BOTH: both the actual and the corrected flow are used for
counting the total volume in the totalizers. Note: both are only
counted in the positive direction!
1) TOTAL OFF: both totalizers are switched off, counting stops.
2) FORWARD: forward volume units.
3) REVERSE: reverse volume units.
4) BOTH: both totalizers, alternating.
5) SUM: sum of both totalizers.
6) BOTH + SUM: both totalizers und their sums show alternating.
7) NO DISPLAY: totalizers are not displayed but totalizers count.
3
99999999 x 10 m
and will roll over to 0 at overrun!
Available units: X10 m3, US-Gallon, m3, barrel, liter.
3.03.05TOTAL ENERUnit for the heat energy totalizer. Note: the maximal value of the
totalizer is 99999999 x 10 GJ and will roll over to 0 at overrun!
Available units: X10 GJ, GJ, MJ, GCal., MCal.
3.03.06VOSUnit for velocity of sound. Available options:
1) NO DISPLAY: no display of the velocity of sound.
2) m/s, feet/s.
3.03.07CYCL DISPIf more than one measured value are to be displayed (e.g. flow rate
and totalizer), each value can be selected manually by pressing the
↑ key, or the values can be alternately displayed each five seconds by
turning the cyclic display function on.
3.03.08ERROR MSGDisplay error messages
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
41
Page 42
6 OPERATION
No. of menuDisplay textSetting/description and functions
UFM 530 HT
3.03.10AN INPUTEnabling or disabling the display of the analog inputs. Note: this
3.03.11SIGN LEVEL This function enables or disables the display of the signal level from
function is only available for converter function CORR T and
CORR T+P (see function 3.02.01)!
The display of the values represented by the current signal coming
from the temperature and pressure transducers. Available settings:
1) NO
2) YES
the sensors. For each measuring path this level is displayed as a gain
value 0...80 dBV at the input amplifier. Available options:
1) NO
2) YES
42
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
Page 43
UFM 530 HT
OPERATION 6
3.04.00, CURR OUTP (Current output)
3.04.01FUNCTIONThe current output can be programmed for the following functions:
1) OFF: switched off, current output steady at current value for 0 %
scale.
2) ACT FLOW: proportional with the actual volume flow.
3) CORR FLOW: proportional with the corrected volume flow. Note:
this function ist only available if the converter function is set to
CORR T or CORR T+P (see function 3.02.01)!
4) F/R IND: forward/reverse indication of actual flow.
100 pct mA value for forward flow, 0 pct mA value for reverse flow.
5) VOS: proportional with the velocity of sound (range is defined in
function 3.01.11 and 3.01.12).
6) GAIN: sensor signal gain, range is 0...100 dBV.
7) AN INP 1: proportional with the signal on analog input 1. Note: this
function is only available if the converter function is set to CORR T or
CORR T+P!
8) AN INP 2: proportional with the signal on analog input 2. Note: this
function is only available if the converter function is set to CORR T or
CORR T+P!
3.04.02DIRECTIONDirection of current output, only available when ACT FLOW or
3.04.03RANGERange of current output (Note: maximum reading is 22 mA!):
3.04.040pctSetting of mA for 0% of the range.
3.04.05100 pctSetting of mA for 100% of the range. Note: value must be at least
3.04.06LIMITLimitation of current output. While default is 22 mA it can be switched
3.04.07ERR INDICOnly available for NAMUR devices. Options:
CORR FLOW is selected in function 3.04.01. In this case there are the
following options:
1) FORWARD: the current output will only be active when the flow is
in the forward flow direction as defined in function 3.01.10.
2) BOTH: the current output will be active when the flow is in the
forward ord reverse flow direction, indicating both in the same range.
3) F/R SPEC: forward and reverse flow measurement indicated in
different range. Use this function to indicate the reverse flow in the
range from 0 mA through 0 pct mA. Note: when the flow goes from
the forward direction to the negative direction, the current output will
pass the "0 pct" mA value down to 0 mA, where it stops!
1) OTHER: user defined span.
2) 0...20/22 mA: 0...100 pct / limit.
3) 4...20/22 mA: 0...100 pct / limit.
Range: 0...16 mA (default: 4 mA).
4 mA greater than current value for 0% scale!
Range: 4...20 mA (default: 20 mA).
to 20 mA when safety system reserve higher currents as fault codes.
Range: 20...22 mA.
1) I
=3.6mA.
ERR
2) I
= 21.5 mA.
ERR
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
43
Page 44
6 OPERATION
3.05.00, PULSE OUTP (Pulse output)
3.05.01FUNCTIONFunction of pulse output with the following options:
3.05.02DIRECTIONDirection of pulse output, only availabe when ACT FLOW or
3.05.03DIG OUTPUTFunction of digital output, only available when DIG OUTPUT is
3.05.04TRIP PNT 1First trip point.
3.05.05TRIP PNT 2Second trip point.
3.05.06TIME CONSTTime constant of pulse output (Note: the time constant setting only
UFM 530 HT
1) OFF: switched off, contact closed.
2) ACT FLOW: proportional with the actual volume flow.
3) CORR FLOW: proportional with the corrected volume flow.
4) F/R IND: forward/reverse flow indication. Contact closed for
forward flow, contact open for reverse flow .
5) VOS: proportinal with the velocity of sound (range defined in
function 3.01.11 and 3.01.12).
6) DIG OUTPUT: digital output.
7) BATCH OUTP: batch output indication, only available for the batchversion. Contact closes at start of the batch and opens when the
batch is reached.
8) GAIN: gain of sensor amplifier, proportional with the signal level.
Range: 0...100 dBV.
9) AN INP 1: proportional with the signal on analog input 1, only
available for the converter functions CORR T and CORR T+P.
11) AN INP 2: proportional with the signal on analog input 2, only
available for the converter functions CORR T and CORR T+P.
CORR FLOW is selected in function 3.05.01.
1) FORWARD: the pulse output will only be active when the flow is in
the forward flow direction as defined in function 3.01.10.
2) BOTH: the pulse output will be active in both flow directions
(forward and reverse), indicating both in the same range.
selected in function 3.05.01. The pulse output now acts as a digital
output and can be programmed for the following functions:
6) TRIP POINT: status output trips if actual flow (Q) goes over a set
limit (an hysteresis is build in). Set points can be set using function
3.05.04 and 3.05.05.
Range: 0...120% of Q
Range: 0...120% of Q
applies to actual flow and corrected flow!):
1) 25 ms (lowest value).
2) MASTER TC: time constant has the value set in function 3.01.04.
100%
100%
.
.
44
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
Page 45
UFM 530 HT
OPERATION 6
3.05.07OUTPUTUnit of pulse output:
1) PULSE RATE (frequency): pulses per unit time, setting by entering
a frequency at 100% volume flow rate.
2) PULSE/UNIT (totalizer pulse output): pulses per unit volume ,
setting by entering a value for the number of pulses for each volume
or energy unit. Each pulse has a fixed volume (i.e. 1 pulse per
0.1 liter). This is the best method of remote totalizing, as pulses
simply need to be counted.
3.05.08PULSE RATEPulse rate (frequency) value for 100% scale. If function 3.05.07 is set
3.05.09PULSE/UNITPulse value per volume unit for totalisation. Note: Check that the
3.05.10PULSE/UNITPulse value per heat energy unit for totalisation. If function 3.05.07 is
3.05.11PULS WIDTH For frequencies ≤10 Hz the following pulse widths can be applied:
to PULSE RATE, the following units can be chosen using this function:
pulse/s, pulse/hr, pulse/min.
Range: 10 pulses/h...2000 pulses/s, default setting: 1000 Hz.
maximum flow span will not cause the number of pulses generated
per second to exceed the maximum of 2000 Hz!
If function 3.05.07 is set to PULSE/UNIT, the unit and number of
pulses per unit can be set for the flow measurement using this
function. Available options: pulse/m
3
, pulse/l , pulse/US-Gallon,
pulse/bbl, free user definable unit.
Default setting: 1, maximum number of pulses per unit: 7870000.
set to PULSE/UNIT, the unit and number of pulses per unit can be set
for the heat power measurement using this function. Available
options: pulse/MJ, pulse/GCal, pulse/MCal, pulse/GJ.
Default setting: 1, maximum number of pulses per unit: 1000000.
1) 25 ms for P
2) 50 ms for P
3) 100 ms for P
4) 200 ms for P
5) 500 ms for P
100%
100%
100%
100%
100%
<10Hz.
<10Hz.
<5Hz.
<2.5Hz.
<1Hz.
For frequencies 10...1000 Hz: 50% duty cycle, 1000...2000 Hz: 70/30%
duty cycle (Note: pulse width may vary 5 ms, pulse period may vary
25 ms!)
3.06.00, DIG INPUT (Digital input)
3.06.01FUNCTIONFunction of digital input. The digital input terminal is the same as the
current output terminal. Therefore, when a digital input function is
selected, the function of the current output needs to be set to OFF
and the current output range has to be set to 0...20 mA. The following
options are available for the digital input:
1) OFF: switched off, no function.
2) RST TOTAL: reset display totalizer(s), independent of the
programming of function 3.07.08.
3) RST ERROR: reset error messages.
4) FORCE ZERO: force display and outputs to their "zero" values.
5) BATCH: start (input high) or stop (input low) a batch, only available
for the batch-version.
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
45
Page 46
6 OPERATION
3.07.00, USER DATA
3.07.01LANGUAGELanguage for display texts:
3.07.02ENTRY CODEEntry code for setting mode:
3.07.03CODE 1Press any 9-keystroke combination and then press the same
3.07.04LOCATIONTag name setting. Free settable tag for identification, maximum
3.07.05UNIT TEXTText for user-defined unit. Definition: volume/time.
3.07.06UNIT VOL
UFM 530 HT
1) GB/USA: English.
2) D: German.
3) F: French.
1) NO: entry with → key only.
2) YES: entry with → key and code 1, factory set on 9 x key.
combination again. Each keystroke is acknowledged by "?" in the
display. If both combinations are equal, "CODE OK" appears and the
new code can be stored. Otherwise "WRONG CODE" appears and the
desired code has to be entered again.
10 characters.
Characters assignable to each place: A...Z, blank character, 0...9.
Factory setting: KROHNE.
Characters assignable to each place: A...Z, blank character, 0...9.
Note: fraction bar "/" in 5th place is unalterable!
Factory setting: XXXX/YYY.
User-defined unit volume. Quantity of user-defined volume per m3.
Range: 10-5...107, factory setting: 1.
3.07.07UNIT TIMEUser-defined unit time. Amount of user-defined time in seconds.
Range: 10-5...107, factory setting: 1.
3.07.08RST ENABLEEnable the reset of the display totalizer(s). Note: this function doesn't
3.07.09ERR LIMITError limit in % of measured value for plausibility filter on the sensor
3.07.10CNT DECRCounter decrement setting for plausibility filter.
3.07.11CNT LIMITCounter limit for plausibility filter. When "0" is set, the plausibility
affect the ability to reset the display totalizer(s) using the digital
input! Available options:
1) NO: totalizer reset disabled.
2) YES: totalizer reset enabled.
paths.
filter will become inactive.
Range: 0...1000, factory setting: 0.
46
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
Page 47
UFM 530 HT
OPERATION 6
3.09.00, COMMUNICATION
3.09.01PROTOCOLCommunication protocol:
1) OFF: no communication.
2) HART: HART® (standard available).
3) PROFIB PA: PROFIBUS PA (optional available).
3.09.02HART ADDR
3.09.03PP/FF ADDRPROFIBUS PA address.
INFORMATION!
In the next part, multiple error messages are shown. Details about the specific functions are
described earlier in the "Installation menu".
Parameter error menu
HART® address.
Range: 00...16.
Range: 000...126.
No. of menuDisplay textSetting/description and functions
4.01.00, FLOW VELOC
Volume flow velocity (v) value incorrect. The flow speed is calculated from the full scale volume flow and
the meter size. Note: ensure condition 0.5 m/s ≤ v ≤ 20 m/s (1.64 feet/s ≤ 65.62 feet/s) is met!
4.01.01FULL SCALEFull-scale value for 100% volume flow rate.
4.01.02METER SIZEMeter size.
4.02.00, CURR OUTP (Current output)
Current output range incorrect. Setting for 100% is compared with setting for 0%. Note: ensure condition
100 pct-0 pct ≥ 4mA is met!
4.02.01RANGERange of current output.
4.02.020pctCurrent value for 0% scale.
4.02.03100 pctCurrent value for 100% scale.
4.03.00, LF CUTOFF (Low-flow cut-off)
Low-flow cut-off range incorrect. If low flow cut-off is set to "on", the value for CUTOFF-OFF is compared
with the value of CUTOFF-ON on. Note: ensure condition CUTOFF-OFF – CUTOFF-ON ≥ 1% is met!
4.03.01LF CUTOFFLow-flow cut-off.
4.03.02CUTOFF ONCutoff "on" value.
4.03.03CUTOFF OFFCutoff "off" value.
4.04.00, ENERGY
Full scale value for heat energy rate (E) incorrect. The fullscale value is compared with the maximum
value that can be measured and should meet the condition E
maximum value that can be measured is at maximum flow and 200°C temperature difference!
4.04.01HEAT FSFull-scale value for 100% heat energy rate.
max<Efullscale<Emax
/1000. Note: the
4.05.00, PULSE/VOS (Pulse/Velocity of sound)
Unit of pulse output for velocity of sound function incorrect. Note: ensure that "PULSE RATE" is selected
for "VOS"!
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
47
Page 48
6 OPERATION
No. of menuDisplay textSetting/description and functions
4.05.01PULS FUNCTFunction of pulse output.
4.05.02PULSE OUTPUnit of pulse output.
4.06.00, VOS (Velocity of sound)
Velocity of sound range incorrect. Note: ensure condition MAX VOS - MIN VOS ≥ 1 m/s (3.28 feet/s) is met!
4.06.01MIN VOSMinimum velocity of sound.
4.06.02MAX VOSMaximum velocity of sound.
4.07.00, PULSE OUTP (Pulse output)
Pulse output frequency value (f) incorrect. The maximum frequency is calculated from the PULSE/UNIT
setting and the maximum value of the measured value. Note: ensure condition
1pulse/hr≤ f ≤ 2000 pulse/s is met!
4.07.01PULSE/UNITPulse value for volume flow rate unit.
4.07.02PULSE/UNITPulse value for heat power rate unit.
4.08.00, PULS WIDTH
Pulse output pulse width incorrect. Note: ensure condition pulse width ≤ 0.5 x pulse period time is met!
4.08.01PULS WIDTHPulse width for frequencies ≤ 10 Hz.
UFM 530 HT
4.09.00, HART
Current output range for HART® incorrect. If HART® is activated, the minimum possible current should
be 4 mA. Note: ensure condition CURR 0 pct ≥ 4mA is met!
4.09.01CURR RANGERange of current output.
4.09.02CURR = PCTCurrent value for 0% scale.
4.10.00, INP/OUTP (Input/Output)
The digital input (C) and current output (I) are not allowed to be switched on simultaneously. If the
Profibus option is activated, only one of the following input/output functions can be used: digital input (C),
current output (I), pulse output (P). The current output is deactivated by setting the function of current
output to "off" and setting the range of current output to 0...20 mA.
4.10.01INP FUNCTFunction of digital input.
4.10.02CURR FUNCTFunction of current output.
4.10.03CURR RANGERange of current output.
4.10.04PULS FUNCTRange of pulse output.
4.13.00, EPROM
EPROM checksum error. Resetting device is necessary.
48
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
Page 49
UFM 530 HT
6.4 Important menu functions in detail
INFORMATION!
Since the UFC 030 converter can be equipped with different options, the availability of certain
options depends on the function of the converter.
In this chapter the different functions of the menu structure are described in more depth.
6.4.1 Error/totalizer (Menu 0.00.00)
This menu is accessible from the measuring mode by pressing the ^ key to enter "CODE 2" and
afterwards pressing the key ↑ and and then →.
INFORMATION!
Only when function 3.03.08 is set to YES, errors occurring during process flow measurement are
represented with flashing display lines and/or a compass field!
Depending on the programming of function 3.03.07, the error messages alternate with the
display of the measured value(s) every five seconds or they can be manually selected by pressing
the ↑ key.
OPERATION 6
1 Compass needle, indicating measuring path error(s): depending on the version of the sensor triangle a appears when
there is no measured value from path 2 and triangle b from path 1; triangle c appears when there is too much noise
on the measuring path(s) for proper measurement.
2 Flashing bar, indicating "new" errors, i.e. not yet acknowledged.
3 Flashing line with description of error message(s).
4 Flashing line with number of errors that have occurred.
The following list gives an alphabetical overview of error messages that can occur during
process flow measurement and what to do:
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
49
Page 50
6 OPERATION
UFM 530 HT
Error
Description of error messageWhat to do
message
ADC AN INPAnalog input internal error (A1or A2).Switch off the flowmeter and switch it on
COMMUNICCommunication device internal error.Reset the error and wait for one minute. If
All occurred error messages are stored in an error messages list and can be viewed using
function 0.00.01. The messages are kept in this list until the cause of the error(s) has been
removed and the error messages have been reset using function 0.00.02.
Errors that have been reset, but whose cause has not been removed, are kept in the list but are
displayed without bar. This allows identification of previously acknowledged and
unacknowledged errors.
6.4.3 Full-scale value for 100% volume and flow rate units (menu 3.01.01)
The measuring range depends on the diameter (DN) of the pipe and the volume flow velocity:
v
min
Q
(m3/h)0.9 x DN
min
Q
(m3/h)
max
Q
(US.Gal/min)
min
Q
(US.Gal/min)-
max
= 0.5 m/s / 1.64 ft/sv
2
-
3.9 x DN
2
= 20 m/s / 65.62 ft/s
max
-
31.25 x DN
-
138 x DN
2
2
6.4.4 Low-flow cut-off, cut-off "on"/"off" value (menu 3.01.05...3.01.07)
Due to the extreme low flow sensitivity of the UFM 3030, it will detect the slightest movement of
fluid, even at zero flow. To avoid these measurements causing outputs and totalizer changes, the
low-flow cut-off can be used to force reading to zero. These are set as a percentage of full scale,
as configured in function 1.01.01 or 3.01.01:
• When the flow rate decreases below the "on" value, the display and outputs are set to their
"zero" values.
• When the flow increases above the "off" value, measurements are resumed.
INFORMATION!
The "off" value must be larger than the "on" value by at least 1%.
With function 3.01.05 set to NO, factory settings are used for the "on" and "off" values.
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
51
Page 52
6 OPERATION
6.4.5 Direction of current output (menu 3.04.02)
INFORMATION!
This function is only available when ACT FLOW or CORR FLOW is selected in function 3.04.01.
When VOS or GAIN is set, only the forward characteristic applies.
A plausibility filter can be set for the sensor measuring paths. There are three settings:
Error limit (menu 3.07.09)
Every measured value outside the error limit is not processed and will increase an internal
plausibility counter by 1, until a counter limit has been reached. The corresponding measuring
path will then be made inactive.
Range: 1 through 99%, default setting: 20%.
INFORMATION!
Whenever one or more measuring paths are made inactive, the display compass needle will
indicate that, combined with the display of error messages.
Error counter decrease step value (menu 3.07.10)
Every measured value within the error limit will decrease the internal plausibility counter by the
number programmed in this function. The higher the number, the faster an inactive measuring
path will become active again.
OPERATION 6
Range: 1 through 99, default setting: 4.
Error counter limit (menu 3.07.11
The counter limit is set using menu 3.07.11. Default setting is 0, i.e. the function is switched off.
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
53
Page 54
7 SERVICE
7.1 Electronics exchange
7.1.1 Exchange of the electronics unit
DANGER!
All work on the electrical connections may only be carried out with the power disconnected. Take
note of the voltage data on the nameplate!
WARNING!
Observe without fail the local occupational health and safety regulations. Any work done on the
electrical components of the measuring device may only be carried out by properly trained
specialists.
UFM 530 HT
Figure 7-1: Switching off power
1 Remove the back cover using the wrench.
2 Disconnect all cables.
Figure 7-2: Removing display print
1 Remove the front cover with the wrench.
2 Loosen the screws of the display.
54
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
Page 55
UFM 530 HT
SERVICE 7
Figure 7-3: Disconnecting display cable
1 Pull the display towards you
2 Pull out the connector from the electronics.
Figure 7-4: Removing electronics unit
1 Loosen the three screws as indicated.
2 Take out the electronics unit.
The reassembling takes place in the reverse order. Please regard the following items:
• Sensor specific data must be entered in the converter.
• All application specific settings must be re-entered after replacement of the electronic unit.
The supplied report on settings contains the standard factory setting.
• It is recommended to check the zero point and eventually perform a new zero measurement.
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
55
Page 56
7 SERVICE
7.1.2 Replacing the mains fuse
DANGER!
For continued fire protection or protection against other damage replace the old fuse only with a
fuse specified in this section.
Fuse specification
Dimensions5 x 20 mm (miniature cartridge fuse)
Rated current230 VAC power supply: 800 mA
CharacteristicTime delay (T)
Rated voltage250 V
Breaking capacity1500 A (ceramic body fuse, high breaking capacity)
StandardIEC 60127-2
ApprovalsUL and OR CSA, VDE, SEMKO, BSI
The fuse should not blow normally unless a problem has developed in the instrument. Try to
determine and correct the cause of the blown fuse before you replace it. The main fuse can only
be reached by removing the complete electronics from the housing as explained in the previous
section. Then proceed as follows:
UFM 530 HT
24 VACDC power supply: 1.25 A
56
Figure 7-5: Removing fuse
1 Find the location of the fuse at the power supply board (in this example the 230 VAC version is shown).
2 Pull out the fuse and replace it.
INFORMATION!
The reassembling takes place in the reverse order.
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
Page 57
UFM 530 HT
7.2 Maintenance
Contact your local representative for maintenance or repair.
DANGER!
The following instructions are for use by qualified personnel only. To avoid electric shock, do not
perform any servicing other than described in the operating instructions of this manual unless
you are qualified to do so.
7.2.1 Admonitions for replacement of flow sensor in separate systems
DANGER!
Always switch off power source before starting work!
Specific calibration data for each flow sensor are determined during factory calibration. Flow
sensor specific data must be set in the UFC 030 ultrasonic flow converter when a flow sensor is
replaced.
• Enter value of flow sensor primary constant GK in menu 3.01.09 (it is specified on the
nameplate).
• If the new flow sensor has a different meter size, you have to program this menu 3.1.8.
• Also set the full-scale range for Q
• Make a zero check after setting of new data.
in menu 3.1.1.
100%
SERVICE 7
7.3 Cleaning
CAUTION!
Scaling at the innerside of the pipe effects the flow measurement. Clean the pipe in such case.
INFORMATION!
Each time a housing cover is opened, the thread should be cleaned and greased. Use only resinfree and acid-free grease.
Ensure that the housing gasket is properly fitted, clean and undamaged.
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
57
Page 58
7 SERVICE
7.4 Spare parts availability
The manufacturer adheres to the basic principle that functionally adequate spare parts for each
device or each important accessory part will be kept available for a period of 3 years after
delivery of the last production run for the device.
This regulation only applies to spare parts which are subject to wear and tear under normal
operating conditions.
7.5 Availability of services
The manufacturer offers a range of services to support the customer after expiration of the
warranty. These include repair, maintenance, technical support and training.
INFORMATION!
For more precise information, please contact your local sales office.
7.6 Returning the device to the manufacturer
UFM 530 HT
7.6.1 General information
This device has been carefully manufactured and tested. If installed and operated in accordance
with these operating instructions, it will rarely present any problems.
CAUTION!
Should you nevertheless need to return a device for inspection or repair, please pay strict
attention to the following points:
•
Due to statutory regulations on environmental protection and safeguarding the health and
safety of our personnel, manufacturer may only handle, test and repair returned devices that
have been in contact with products without risk to personnel and environment.
•
This means that the manufacturer can only service this device if it is accompanied by the
following certificate (see next section) confirming that the device is safe to handle.
CAUTION!
If the device has been operated with toxic, caustic, flammable or water-endangering products,
you are kindly requested:
•
to check and ensure, if necessary by rinsing or neutralising, that all cavities are free from
such dangerous substances,
•
to enclose a certificate with the device confirming that is safe to handle and stating the
product used.
58
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
Page 59
UFM 530 HT
7.6.2 Form (for copying) to accompany a returned device
Company:Address:
Department:Name:
Tel. no.:Fax no.:
Manufacturer's order no. or serial no.:
The device has been operated with the following medium:
SERVICE 7
This medium is:water-hazardous
toxic
caustic
flammable
We checked that all cavities in the device are free from such
substances.
We have flushed out and neutralized all cavities in the
device.
We hereby confirm that there is no risk to persons or the environment through any residual media
contained in the device when it is returned.
Date:Signature:
Stamp:
7.7 Disposal
CAUTION!
Disposal must be carried out in accordance with legislation applicable in your country.
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
59
Page 60
8 TECHNICAL DATA
8.1 Measuring principle
• Like canoes crossing a river, acoustic signals are transmitted and received along a diagonal
measuring path.
• A sound wave going downstream with the flow travels faster than a sound wave going
upstream against the flow.
• The difference in transit time is directly proportional to the mean flow velocity of the medium.
UFM 530 HT
Figure 8-1: Measuring principle
1 Flow velocity
2 Transducer A
3 Acoustic path
4 Transducer B
5 Angle (between flow vector and acoustic path vector)
6 Diameter
60
www.krohne.com12/2013 - 4002393201- HB UFM 530 HT R02 en
Page 61
UFM 530 HT
8.2 Technical data
INFORMATION!
•
The following data is provided for general applications. If you require data that is more
relevant to your specific application, please contact us or your local sales office.
•
Additional information (certificates, special tools, software,...) and complete product
documentation can be downloaded free of charge from the website (Download Center).
Measuring system
Measuring principleUltrasonic transit time.
Application range(Non) conductive fluids.
Measured value
Measured value
Measured valueMeasured value
Primary measured valueTransit time.
Secondary measured valueVolume flow rate, totalised volume, velocity of sound (VoS), signal
Design
Measuring range0.5...20 m/s / 1.7...66 ft/s
Signal converter
Signal converter
Signal converterSignal converter
Wall-mounted housing (W) remote version
Measuring sensor
Measuring sensor
Measuring sensorMeasuring sensor
UFS 500 HT is available in the following pipe diameters and building constructions: