Kodak DirectView User manual

{TheoryGuide}{Production}{Carestream Health}{Restricted}
Publication No. TG5258-1
18JAN08
Restricted
THEORY GUIDE
for the
Service Codes: 5258, 5259
Important
Qualified service personnel must repair this equipment.
When performing the procedures outlined in this document, personnel must always
employ safe work practices and wear the appropriate personal protective equipment (e.g., safety eyewear) in accordance with Company Standard Operating Procedures.
H219_0001GC
© CARESTREAM HEALTH, INC.
THEORY GUIDE
18JAN08
TG5258-1
Page
2 of 113
PLEASE NOTE The information contained herein is based on the experience and knowledge relating to t he subject
matter gained by Carestream Health, Inc. prior to publication. No patent license is granted by this information.
Carestream Health, Inc. reserves the right to chang e this information withou t notice, and ma kes no
warranty, express or implied, with respect to this infor mation. Carestr eam Health shall not be liable for any loss or damage, including consequ ential o r special damages, resulting fr om any use of this information, even if loss or damage is caused by Carestream Health’s negligence or other fault.
This equipment includes parts and assemblies sensitive to damage from electrostatic discharge. Use caution to prevent damage during all service procedures.
Table of Contents
Description Page
Equipment Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Features and Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Main Subsystems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Radiography Theory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Comparison of Film/Screen and Computed Radiography (CR) . . . . . . . . . . . . 15
Overview of CR Technology . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Operations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Exposing the STORAGE PHOSPHOR SCREEN. . . . . . . . . . . . . . . . . . . . . . 19
Stimulating the PHOSPHOR. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Changing Light Energy to an Analog Signal . . . . . . . . . . . . . . . . . . . . . . . . 22
Changing Analog Signals to Digital Signals . . . . . . . . . . . . . . . . . . . . . . . . 23
Processing the Digital Image. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Sequence of Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Overview of Workflow Using the CLASSIC/ELITE CR SYSTEM . . . . . . . . . . . . 25
Before Loading the CASSETTE. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Loading the CASSETTE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Removing the PLATE from the CASSETTE. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Scanning the SCREEN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Erasing the SCREEN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Inserting the PLATE back into the CASSETTE SHELL . . . . . . . . . . . . . . . . . . . 31
Removing the CASSETTE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
STORAGE PHOSPHOR CASSETTE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Size and Resolution of SCREENS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Fast Scan / Slow Scan Directions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
THEORY GUIDE
18JAN08
TG5258-1
Page
3 of 113
Image Matrix Size . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Reading the BAR CODE LABEL of the CASSETTE . . . . . . . . . . . . . . . . . . . . . . 40
CASSETTE HANDLING. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
CASSETTE Load . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
EXTRACTION BAR MOTOR AY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
CLAMP MOTOR. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
PLATE HANDLING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Optical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
LASER . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
GALVO . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
COLLECTOR and PHOTOMULTIPLIER TUBE (PMT). . . . . . . . . . . . . . . . . . . . . . 61
Scan/Erase. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
PLATE SUPPORT AY - Version 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
PLATE SUPPORT AY - Version 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
LEAD SCREW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
EXTRACTION BAR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
REFERENCE SENSOR S5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
PLATE PRESENT SENSOR S6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
SLOW SCAN MOTOR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
ENCODER. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
ERASE AY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
LAMP CURRENT SENSORS CS1 - CS5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Imaging Sequence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Scanning the SCREEN - Slow Scan/Fast Scan. . . . . . . . . . . . . . . . . . . . . . . . . . 81
Obtaining the Image Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Processing the Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Processing the Image. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Logic and Control. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
Operator Input Compone nts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
MONITOR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
LOCAL USER INTERFACE (LUI) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
BOARDS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
Distribution of Images to the Network. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
Sequence of Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
18JAN08
TG5258-1
Page
4 of 113
THEORY GUIDE
Power Distribution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
POWER SUPPLY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
Power Distribution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
INTERLOCK . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
TRANSFORMER T1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Logs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
Error and Activity Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
Actuation Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
Glossary. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
18JAN08
TG5258-1
Page
5 of 113
THEORY GUIDE Equipment Description

Section 1: Equipment Description

Features and Functions

The Kodak DirectView CLASSIC/ELITE CR SYSTEM is a LASER SCANNER that reads a latent image made on a STORAGE PHOSPHOR SCREEN during an X-ray exam and provides a digital image. Physicians and radiologists can then view, improve, store and make a print of the image, and send the image across a computer network.
Feature Function
SCANNER for the CLASSIC or ELITE CR SYSTEM
BAR CODE READERS
Size: 43.8 x 60.3 cm (19.0 x 23.75 in.)
single CASSETTE load with integrated LOCAL USER INTERFACE
uses DirectView CR CASSETTES
ELITE CR SYSTEM provides maximum CASSETTE throughput
capability
CLASSIC CR SYSTEM provides throughput of approximately 25 - 30%
lower than the ELITE CR SYSTEM
EXTERNAL BAR CODE READER:
hand-held READERused to scan the BAR CODE LABEL on CASSETTES and other
bar codes used for entering data
INTERNAL BAR CODE READER:
automatically scans the BAR CODE LABEL on CASSETTES that
are loaded
provides information about the size, speed, and serial number of
the CASSETTE
18JAN08
TG5258-1
Page
6 of 113
THEORY GUIDE Equipment Description
Feature Function
WORKFLOW and IMAGE VIEWING CONSOLE (WAIV)
MONITOR - options:
17 in. FLAT PANEL DISPLAY without TOUCH SCREEN - requires use
of KEYBOARD
19 in. FLAT PANEL DISPLAY with TOUCH SCREEN
allows the operator to touch or click areas displayed on the screen to:
enter exam and patient informationview and improve images
allows the FE to do service diagnostics
EXTERNAL PC:
includes software for:
acquiring images from the CLASSIC/ELITE CR SYSTEMprocessing imagesproviding communication with external devices and the computer
network
Software Options Available
Furniture Options Available
available at all times to the FE
new EVP Plus Software
Administrative Analysis and Reporting
Total Quality Tool
Mammography Option - outside US and Canada
Software Refresh
FLOOR STAND
WALL STAND
either holds MONITOR, KEYBOARD, MOUSE, and BAR CODE
READER
FLOOR STAND also holds 10 CASSETTES
18JAN08
TG5258-1
Page
7 of 113
THEORY GUIDE Equipment Description
Feature Function
Kodak DirectView
REMOTE OPERATIONS PANEL (ROP)
A device that is installed on the wall in an area separate from the CLASSIC/ELITE CR SYSTEM, used for viewing images and entering data. The ROP includes:
computer running Microsoft Windows XP
TOUCH SCREEN MONITOR - SVGA device with a 1024 x 768 pixel
resolution
EXTERNAL BAR CODE READER - can read all formats identified for
the hand-held BAR CODE READER on the CLASSIC/ELITE CR SYSTEM
The ROP allows operators to:
enter patient, exam, and CASSETTE (PEC) data into a CLASSIC/
ELITE CR SYSTEM
check patient data
view scanned X-ray images
send images to other nodes on the network
The PEC data entered through an ROP and sent across the network is associated with the correct image.
18JAN08
TG5258-1
Page
8 of 113
THEORY GUIDE Equipment Description
Feature Function
Configurations STANDALONE - the CLASSIC/ELITE CR SYSTEM is not connected
for use with other CR SYSTEMS on a network:
CLASSIC/ELITE CR SYSTEM supports use with a maximum of 10
REMOTE DEVICES
network connects to WORKSTATIONS for viewing, or reading the
images, and to PRINTERS to obtain hardcopy output
CAPTURE LINK SYSTEM - uses a CAPTURE LINK SERVER to
support shared use of 2 - 5 CR SYSTEMS on a network
SIMPLE CAPTURE LINK - software option allows shared use of 2 CR
SYSTEMS on a network without using a CAPTURE LINK SERVER
In a SIMPLE or CAPTURE LINK SYSTEM:
CR SYSTEMS configured can be Kodak DirectView CR 825/850/
950/975 SYSTEMS or CLASSIC/ELITE CR SYSTEMS
up to 20 remote devices can be configured for useworkflow is distributed by allowing patient data, CASSETTE ID
information, CASSETTE scanning, and image review functions to be shared between CR SYSTEMS and remote devices
CR SYSTEMS and remote devices can only share information and
function together within the same SIMPLE or CAPTURE LINK SYSTEM
CR SYSTEMS network connect to WORKSTATIONS for viewing or
reading the images and/or to PRINTERS to obtain hardcopy output
Remote devices include:
REMOTE OPERATION PANELS (ROP)
Customer provided PC using REMOTE ACCESS SOFTWARE (RAS)
allows the PC to be used:
as a REMOTE PATIENT DATA ENTRY STATION (RPDES)for performing REMOTE KEY OPERATOR functions.
18JAN08
TG5258-1
Page
9 of 113
THEORY GUIDE Equipment Description
Feature Function
Network Communications
All CLASSIC/ELITE CR SYSTEMS and ROP devices:
connect to the 10 Base-T, 100 Base-T, or 1000 Base-T Ethernet
network of the site
can send information to all networked DICOM digital imaging
equipment that is qualified with the CLASSIC/ELITE CR SYSTEM
use CATEGORY 5 CABLES to connect to the network
use a gateway device qualified by Carestream Health to enable access
to the HIS/RIS system. The customer must provide this device
On-site Service CASTERS allow the CLASSIC/ELITE CR SYSTEM to be moved for
service without leveling
DATA PLATES and MODIFICATION LABELS are located for easy
access and viewing
PLUGS and CONNECTORS are identified
data in the Error and Activity logs can be sorted by field for
troubleshooting, for example, by date and error code number
FEs can view internal diagnostics, including error codes, component
tests, and tests of the SENSORS from the MONITOR
18JAN08
TG5258-1
Page
10 of 113
THEORY GUIDE Equipment Description
Feature Function
Remote Service remote access options:
dedicated MODEM connected to the CLASSIC/ELITE CR SYSTEMMODEM SERVER provided by the customer
one point of access to the service functions of all CLASSIC/ELITE CR
SYSTEMS on the customer network from the remote service access connection
access to all service functions, except running the SCAN/ERASE
subsystem
remote service:
installing softwaresetting up the configuration for the CLASSIC/ELITE CR SYSTEMretrieving and clearing Error and Activity Logsretrieving Image Processing Library (IPL) diagnostic images
Note
FEs providing remote service cannot view the information about the patient on images.
THEORY GUIDE Equipment Description
18JAN08
TG5258-1
Page
11 of 113
The following table describes the specifications for the number of CASSETTES per hour:
CLASSIC CR SYSTEM ELITE CR SYSTEM
Size
High Speed and
Reduced Border
Scan Mode
Standard Speed
High Speed and Reduced Border
Scan Mode
Standard Speed
24 x 18 GP 77 77 100 100 24 x 30 GP 58 58 76 76 24 x 18 HR 77 77 101 101 24 x 30 HR 58 58 76 76 24 x 18 EHR-M 60 60 79 79 24 x 18 EHR-M2 60 60 79 79 24 x 30 EHR-M 45 45 58 58 24 x 30 EHR-M2 45 45 58 58 30 x 15 GP 92 92 122 122 35 x 35 GP 77 53 102 71 35 x 43 GP 69 46 90 61 35 x 35 GP+ 53 53 70 70 35 x 43 GP+ 46 46 61 61 35 x 43 LONG-
66 66 87 87 LENGTH CASSETTE
35 x 84 GP LLI 68 68 88 88
Tolerance is ± 5
18JAN08
TG5258-1
Page
12 of 113
THEORY GUIDE Equipment Description

Main Subsystems

To Network
MONITOR
EXTERNAL
BARCODE
READER
EXTERNAL
PC
AC Power
90-264 VAC
H219_7500DC
ERASE LAMPS
VOLTAGE
SELECTION
JUMPERS
MINI - MCB
BOARD
A1
INTERNAL BARCODE
READER
ISOLATION
TRANSFORMER
RS-232
POWER SUPPLY
PS1
SLOW
SCAN
ENCODER
SLOW
SCAN
MOTOR
LASER DIODE
DRIVER
BOARD
A17
LASER
IEB
BOARD
A2
LOCAL
USER
INTERFACE
GALVO
2 PMTS
PMT/DAS
BOARD
A5
COLLECTOR
Hospital Network
18JAN08
TG5258-1
Page
13 of 113
THEORY GUIDE Equipment Description
Subsystem Description See:
CASSETTE includes:
STORAGE PHOSPHOR
SCREEN that captures and stores the X-ray
STORAGE PHOSPHOR CASSETTE
image for processing
CASSETTE SHELL that holds the PLATE
available in 5 sizes and 3 resolutions (GP, HR,
and EHR)
CASSETTE HANDLING
allows the operator to load the CASSETTE into
the CLASSIC/ELITE CR SYSTEM
removes the PLATE from the CASSETTE SHELL
after scanning, installs the PLATE in the
CASSETTE SHELL
allows the CASSETTE to be removed from the
CLASSIC/ELITE CR SYSTEM
Optical controls and moves the laser beam to the
SCREEN
captures the blue light emitted from the SCREEN
Scan/Erase moves the PLATE at a uniform speed:
through the scanning areato the erase position
removes the residual image on the SCREEN by
exposing it to maximum light
inserts the PLATE into the CASSETTE SHELL
again
CASSETTE HANDLING
Optical
Scan/Erase
Imaging assembles the data from the SCREEN and
changes it to digital format
processes the image
Imaging Sequence
18JAN08
TG5258-1
Page
14 of 113
THEORY GUIDE Equipment Description
Subsystem Description See:
Logic and Control processes commands from the operator
controls the operation of all subsystems
Logic and Control
sends processed images to the network for
distribution
Pow e r D i s t r ib u ti o n provides power for all subsystems
has an INTERLOCK SWITCH that actuates when
the FRONT COVER is removed
Error and Activity Logs
records logs of errors in the system
records user actions
Power Distribution
Logs
18JAN08
TG5258-1
Page
15 of 113
THEORY GUIDE Radiography Theory

Section 2: Radiography Theory

Comparison of Film/Screen and Computed Radiography (CR)

FILM/
SCREEN
STORAGE
PHOSPHOR
SCREEN
H194_5012HC
X-RAY latent image TUBE
AERIAL IMAGE
X-RAY TUBE
AERIAL IMAGE
(On Film)
(Storage Phosphor)
FILM PROCESSING
CONVERSIONS
visible image
(On Film)
visible imagelatent image
(CRT)
FINAL VISIBLE
IMAGE
(Film or Viewer)
ENHANCED
PROCESSING
X-rays are used in medical imaging to make an image of given body parts on a surface, which can be read by a Radiologist or other medical personnel. The available systems for capturing these images are:
Screen/film - captures a projection image on an X-ray film
Computed Radiography (CR) - captures a digital image
THEORY GUIDE Radiography Theory
18JAN08
TG5258-1
Page
16 of 113
The following phases are necessary to capture and process projection radiographs for both screen/film systems and CR systems:
Phase of Image Capture Description
Phase 1 ­Making the aerial image
In both screen/film and CR SYSTEMS:
an X-ray TUBE emits X-rays in the direction of an IMAGE
RECEPTOR
when the X-rays reach the body of the patient, some are
absorbed by the patient and some are not. The result is an “aerial” image with varying degrees of X-ray power (varying numbers of X-ray PHOTONS)
Phase 2 ­Capturing the latent image
When the IMAGE RECEPTOR is exposed to the X-rays in the
aerial image, a latent image is captured on the RECEPTOR:
screen/film systems - image is captured on sensitized
radiographic film
CR SYSTEMS - image is captured on a STORAGE
PHOSPHOR SCREEN. The X-ray PHOTONS that reach the SCREEN charge the PHOSPHOR, making a latent image on the screen
Phase 3 ­Capturing, changing, and storing the visible image
The latent image must be changed to a visible image, which
can be read by the Radiologist, moved from one place to
another, and stored for use at another time:
screen/film systems - radiographic film is processed through
chemicals and the latent image is fixed onto the film
CR SYSTEMS - the latent image on the STORAGE
PHOSPHOR SCREEN is scanned by a laser beam, which stimulates the charged PHOSPHOR on the SCREEN. Blue light is emitted from the stimulated PHOSPHOR, assembled, and changed into analog electrical signals. The analog image is then changed into digital signals and processed. The digital image is stored and displayed by a computer system and can be routed to other computers and PRINTERS through a network
THEORY GUIDE Radiography Theory
18JAN08
TG5258-1
Page
17 of 113
The following table compares the analog and digital health image capture systems.
Analog Screen/Film Systems Digital CR Systems
Uses “Rare Earth” SCREENS ­GADOLINIUM OXYSULFIDE or
Uses a BARIUM FLOUROHALIDE STORAGE PHOSPHOR SCREEN.
LANTHANUM OXYBROMIDE. Speed range from 100 - 1000. Phosphor SCREEN types. Use:
General Purpose (GP) for most general
radiography exams
High Resolution (HR) for general
radiography extremity exams
Enhanced High Resolution (EHR) for
mammography exams
Film processing parameters are important
No film processing is necessary. in determining the quality of the image, for example: chemical temperature and timing.
It is hard to obtain the same print quality when copies are necessary because of
The user can print a copy of the digital image
at any time with a consistent level of quality. variations in GENERATORS, PROCESSORS, positions, procedures, and conditions for developing the film.
Overexposing or underexposing an image normally makes it necessary to expose the patient to ionizing radiation again.
Image quality is changed by conditions in the environment, for example temperature or humidity.
The image cannot be viewed in more than one place at a time.
Exposure factors do not normally make it
necessary to expose the patient to ionizing
radiation again.
Image quality is not changed by conditions in
the environment.
CR images can be viewed at more than one
place at the same time, in the same building or
in remote nodes on the network.
18JAN08
TG5258-1
Page
18 of 113
THEORY GUIDE Radiography Theory
Analog Screen/Film Systems Digital CR Systems
Recording medium - film
Output medium - film
Storing medium - film
Recording medium - STORAGE
PHOSPHOR SCREEN
Output medium - film, paper, digital display
Storing medium - digital
Image density and contrast are controlled by kilovolts peak (kvP), milliampere seconds (mA.s), and film type.
Viewing quality can only be improved by increasing the brightness of the LAMP that illuminates the film.
The quality of films that are not exposed correctly cannot be improved.
Density and contrast are controlled by image
processing parameters. kvP, and mA.s continue
to be important image control factors for details
and noise in the digital image.
Digital images can be improved by processing
on a computer to change the density, contrast,
sharpness, and other factors.
Images that were not exposed correctly can be
improved. For example, software parameters
can improve image density and contrast.
18JAN08
TG5258-1
Page
19 of 113
THEORY GUIDE Radiography Theory

Overview of CR Technology

Operations

The following operations are necessary to capture the latent image in the STORAGE PHOSPHOR SCREEN and change it to a digital image that can be viewed on a computer screen and sent to a PRINTER.
Exposing the STORAGE PHOSPHOR SCREEN
Stimulating the PHOSPHOR
Changing Light Energy to an Analog Signal
Changing Analog Signals to Digital Signals
Processing the Digital Image

Exposing the STORAGE PHOSPHOR SCREEN

X-RAY TUBE
H194_5033BC
aerial
image
STORAGE PHOSPHOR SCREEN
Charged storage phosphors proportional to X-ray energy
absorbed by screen.
latent
image
Lighter values indicate that more x-rays were absorbed by the SCREEN - bone tissue
Mid-range values indicate that fewer x-rays were absorbed by the SCREEN - soft tissue
Darker values indicate that most x-rays were absorbed by the SCREEN - did not pass through the body
THEORY GUIDE Radiography Theory
18JAN08
TG5258-1
Page
20 of 113
When a STORAGE PHOSPHOR SCREEN is exposed to X-rays:
special PHOSPHOR on the SCREEN absorbs the radiation in degrees of intensity
determined by the body part and the type of SCREEN:
soft body tissues absorb a small quantity of radiation - these areas are indicated in the
X-ray image by mid-range values
bone tissues absorb most of the radiation - these areas are indicated in the X-ray
image by light values
X-rays that do not hit any obstructions are indicated in the X-ray image by dark valuesHigh Resolution SCREENS absorb less energy than General Purpose SCREENS
SCREEN has a latent image in the areas that were exposed to the radiation. The quantity
of stored energy or charge on the SCREEN is proportional to the quantity of X-ray energy absorbed by the SCREEN.
Characteristics of the
STORAGE PHOSPHOR
SCREEN
Description
X-ray absorption About 50% of the X-ray energy is released in the form of
fluorescence when the SCREEN is exposed. The X-ray energy remaining makes the latent image on the SCREEN.
Photostimulable luminescence
When the charged PHOSPHOR on the SCREEN is stimulated by light, the PHOSPHOR releases or discharges blue light proportional to the energy the PHOSPHOR has stored.
Fading The latent image fades with time, but it is possible to read data from
the SCREEN for a number of days after scanning.
Residual image After a SCREEN is erased by exposing it to light, it keeps some
charge from the latent image. This charge does not make the SCREEN less effective when it is used again.
18JAN08
TG5258-1
Page
21 of 113
THEORY GUIDE Radiography Theory
Characteristics of the
STORAGE PHOSPHOR
SCREEN
Description
Signal accumulation Signals can accumulate on SCREENS that are not used for more
than 24 hours. Erasing these SCREENS decreases the residual image to the optimum range for using the SCREEN again. Failure to erase these signals can result in artifacts.
Long life The photostimulable luminescent quality of the SCREEN does not
decrease with time. The life of a SCREEN can be decreased by damage to the material.

Stimulating the PHOSPHOR

It is necessary to stimulate the PHOSPHOR in the SCREEN to read the latent image. The following components of the CR SYSTEMS provide this function:
light source:exposes the SCREEN with high-intensity light that stimulates the ELECTRONS and
causes the ELECTRONS to be luminescent
laser beam moves from one side of the SCREEN to the other to expose the image
GALVO MIRROR:moves the laser beam across the SCREEN and then back to the start position. At the
same time, the SCREEN moves perpendicular to the scanning direction of the laser beam
is continually monitored and adjusted to check that the scanning operation is correct
and has a continual speed
scanning optics:focuses and shapes the laser beam, keeping the speed and angle of the beam the
same when it moves across the SCREEN
angle of a laser beam determines the size, shape, and speed of the beam. An
example is the beam of a FLASHLIGHT moving across a flat surface from one edge to the center and to the other edge
THEORY GUIDE Radiography Theory
18JAN08
TG5258-1
Page
22 of 113

Changing Light Energy to an Analog Signal

The following components of CR SYSTEMS change the light energy in the exposed SCREEN to an analog signal:
LIGHT COLLECTOR:provides the collection of the blue light emitted when the SCREEN is stimulated by the
laser beam
CLASSIC/ELITE CR SYSTEM uses an INTEGRATING CAVITY with MIRRORS to
provide the collection of the blue light
BLUE FILTER:does not allow any red light reflected from the SCREEN to reach the LIGHT
DETECTORS
allows only the blue light to reach the LIGHT DETECTORS
LIGHT DETECTORS:are normally PHOTOMULTIPLIER TUBES (PMT)receive light that enters the COLLECTORchange the light PHOTONS into ELECTRONS when the PHOTONS enter through a
PHOTOCATHODE. When the ELECTRONS move through the LIGHT DETECTORS, the ELECTRONS increase in number - “gain”
when more than one LIGHT DETECTOR is used in a system, the system adds and
changes the signals into one output. The output from the added PMTs can include frequencies that are outside of the limits of the system - “noise”. An ANALOG FILTER limits this noise
THEORY GUIDE Radiography Theory
18JAN08
TG5258-1
Page
23 of 113

Changing Analog Signals to Digital Signals

SAMPLING
Y
image
sample
grid
X
analog image digital image
(continual values) (discrete values)
pixel code value
(0 - 4095)
image matrix
H194_5014HC
Analog signals are changed to digital signals by sampling the blue light from the STORAGE PHOSPHOR SCREEN and moving it through an ANALOG-TO-DIGITAL CONVERTER to make a digital value for the brightness of each sample.
Sampling is similar to making a photograph of the signal at a given time. The sample has both a horizontal and a vertical value. The size of the sample is defined in the system software for both the horizontal and vertical directions.
The horizontal value indicates a point in time in the motion of the laser beam across the
SCREEN.
The vertical value indicates a line on the screen at a right angle to the scanning direction.
THEORY GUIDE Radiography Theory
18JAN08
TG5258-1
Page
24 of 113
If you locate both the horizontal and the vertical points of the sample on an imaginary matrix, similar to the one in the graphic, the result indicates one pixel in the digital image.
Continual analog input values are changed to output values. In this process, the replacement of small ranges of analog input values with one digital output value occurs. The digital output value indicates one pixel of information on the MONITOR.
The output is a linear digital signal. The CLASSIC/ELITE CR SYSTEM emits a 16-bit digital signal with a total signal range of 65,536 levels. Because it is not possible for the human eye to see this range of separate values, the CLASSIC/ELITE CR SYSTEM changes the 16-bit linear image data to 12-bit log data. This 12-bit log provides data from 0 - 4095 values. These values are used in the CLASSIC/ELITE CR SYSTEM.

Processing the Digital Image

Digital imaging allows users to improve diagnostic images by processing the images. After the digital image is made, the digital data is processed using parameters set up in the software. In the CLASSIC/ELITE CR SYSTEM, this processing occurs in the EXTERNAL PC.
Examples of image processing used for digital images:
segmentation
tone scaling
edge enhancement
brightness - level
contrast - window
18JAN08
TG5258-1
Page
25 of 113
THEORY GUIDE Sequence of Operation

Section 3: Sequence of Operation

Overview of Workflow Using the CLASSIC/ELITE CR SYSTEM

1 The Radiology Department receives an exam request. 2 The Radiology Technologist (RT) assembles the patient information. Examples of patient
information are patient name, ID, and accession number.
Note
On the CLASSIC/ELITE CR SYSTEM, you can receive patient information through a DICOM­MODALITY WORKLIST PROVIDER.
3 The operator can select network nodes to send the image data. 4 The operator uses a CR CASSETTE to do the exam, capturing the latent image on the
STORAGE PHOSPHOR SCREEN.
5 Using the CLASSIC/ELITE CR SYSTEM or the ROP, the operator enters the CASSETTE
ID Information by scanning the CASSETTE BAR CODE or entering it manually.
6 The operator inserts the exposed CASSETTE into the CLASSIC/ELITE CR SYSTEM.
The system scans the SCREEN, capturing the latent image on the SCREEN and changing it to a digital image. After scanning, the SCREEN is automatically erased and inserted into the CASSETTE SHELL.
7 The CLASSIC/ELITE CR SYSTEM processes the image. If the system is in:
“Pass-Through Mode” - the image is automatically sent to all network nodes
“QA Mode” - the operator can process the image and then send it to other network
nodes
8 If necessary, the image can be processed and sent to network nodes again.
THEORY GUIDE Sequence of Operation
18JAN08
TG5258-1
Page
26 of 113

Before Loading the CASSETTE

1 After initializing, the CLASSIC/ELITE CR SYSTEM is ready to receive a CASSETTE for
scanning. The LOCAL USER INTERFACE (LUI) displays:
Status ICON: green
name of the system in white text
2 The RT uses a CR CASSETTE to capture the latent image of the body part on the
SCREEN.
Status Summary: Ready to Receive a CASSETTE
CLAMP BAR is in the open position
PIVOTING PLUSH is in the open position
LIGHT SEAL BAR is in the open position
EXTRACTION BAR is at the home position
HOOKS are in the down position

Loading the CASSETTE

1 The RT loads the CASSETTE into the INPUT SLOT until the CASSETTE reaches the
CASSETTE ENTRY SENSOR S3.
2 The CASSETTE ENTRY SENSOR S3 detects the CASSETTE, and the SLOW SCAN
MOTOR moves up to lock the TOP CAP of the EXTRACTION BAR.
3 When the CASSETTE is inserted, the INTERNAL BAR CODE READER reads the size,
speed, and serial number of the CASSETTE, then:
emits a sound
sends information to the IEB BOARD A2:“CASSETTE Detected” messagesize of the CASSETTE
THEORY GUIDE Sequence of Operation
18JAN08
TG5258-1
Page
27 of 113
4 When the CASSETTE reaches the CASSETTE EXTRACTION SENSOR S4, the CLAMP
MOTOR closes the CLAMP BAR until the CLAMP CLOSED SENSOR S2 is blocked.
Note
If the INTERNAL BAR CODE READER did not read the bar code correctly, the operator must enter the data manually at the LUI. The CLAMP MOTOR will not close the CLAMP until the operator enters the size of the CASSETTE.
5 The MINI-MCB BOARD A1 sends “CASSETTE Detected” message to the IEB BOARD
A2.
6 The IEB BOARD A2 sends a “Scan Request” message to the EXTERNAL PC. 7 The EXTERNAL PC:
checks that it has the quantity of memory necessary to receive an image
makes a raw image file to receive the image
sends a “Scan Request Reply” message to the IEB BOARD A2 with a value of “OK”
8 The IEB BOARD A2 receives the information about the size and speed of the
CASSETTE.
Status Summary: CASSETTE Loaded
SLOW SCAN is in the up position
TOP CAP on the EXTRACTION BAR is locked
CLAMP BAR is closed
PLATE remains inside the CASSETTE
HOOKS on the EXTRACTION BAR are not extended
bar code was read
18JAN08
TG5258-1
Page
28 of 113
THEORY GUIDE Sequence of Operation

Removing the PLATE from the CASSETTE

1 The IEB BOARD A2 sends a message to the MINI-MCB BOARD A1 to load the PLATE. 2 The SLOW SCAN MOTOR moves down to unlock the TOP CAP, and the HOOKS
MOTOR moves to the “HOOKS in Lane” position.
3 The SLOW SCAN MOTOR moves the HOOKS up into the CASSETTE. 4 The HOOKS MOTOR moves the HOOKS to the “Unlatch” position. 5 The PLATE is “Unlatched” from the CASSETTE, and the SLOW SCAN MOTOR moves
the PLATE to the “Star t of Scan” position.
6 The MINI-MCB BOARD A1 sends a message to the IEB BOARD A2 that the PLATE is
loaded.
Note
If any MOTOR does not move correctly, the MINI-MCB BOARD A1 sends an error message to the IEB BOARD A2.
Status Summary: PLATE Fastened
CLAMP BARS are holding the CASSETTE
PLATE is fastened to the EXTRACTION BAR
EXTRACTION BAR is at the “Start of Scan” position with the fastened PLATE
PIVOTING PLUSH has made a light-tight environment around the CASSETTE
18JAN08
TG5258-1
Page
29 of 113
THEORY GUIDE Sequence of Operation

Scanning the SCREEN

1 The IEB BOARD A2:
actuates the GALVO
sends a signal to the MINI-MCB BOARD A1 to start the scan, which star ts the SLOW
SCAN MOTOR
sends a signal to the EXTERNAL PC that the scan is star ting
2 The EXTERNAL PC displays a PROGRESS BAR on the Scan Status screen. This is a
graphic display only and not a real-time indication of the status of the scanning operation. The LUI also displays a PROGRESS BAR for the scan status.
3 The SLOW SCAN MOTOR rotates, moving the PLATE at a continual speed through the
field of scan in the slow scan direction.
4 The IEB BOARD A2 controls the motion of the laser beam across the SCREEN in the
fast scan direction. The SCREEN is scanned one pixel at a time, one line at a time. See
Scanning the SCREEN - Slow Scan/Fast Scan.
Note
The fast scan motion is an almost horizontal trace across the SCREEN, from the back of
the SCREEN toward the front. When it reaches the end of a line, it does a fast retrace to start another line. During the scanning, the SCREEN is moving down at a controlled speed to make each fast scan trace one pixel line higher up on the SCREEN than the line before. The result is that the fast scan is in a slightly diagonal trace across the SCREEN.
The slow scan runs for a determined number of lines in the vertical direction. A set
number of samplings occur for each line. The number is determined by the size of the SCREEN. Both the number of lines and the number of samplings are set up in the calibration for that size of SCREEN.
5 When the end of the scan is reached, the MINI-MCB BOARD A1 sends a status
message to the IEB BOARD A2.
6 The IEB BOARD A2 de-energizes the PMTs, GALVO, and the LASER. 7 The IEB BOARD A2 sends a “Scan End” message to the EXTERNAL PC. The
PROGRESS BAR displays until the image is transferred to the EXTERNAL PC.
18JAN08
TG5258-1
Page
30 of 113
THEORY GUIDE Sequence of Operation

Erasing the SCREEN

1 The IEB BOARD A2 sends an “Erase Plate” command to the MINI-MCB BOARD A1,
which sends a signal to the SLOW SCAN to start the erasing operation.
2 The SLOW SCAN MOTOR actuates and moves the PLATE into the erase position,
determined by the counts of the SLOW SCAN ENCODER.
3 The MINI-MCB BOARD A1 energizes the ERASE LAMPS to illuminate for 2 - 16 seconds
to remove the image from the SCREEN.
Note
The length of time the ERASE LAMPS illuminate is determined by the highest pixel code value of the image that was scanned. If one pair of LAMPS is not operating, the time increases by a factor of 2. If more than one pair of LAMPS is not operating, a message displays on the MONITOR.
4 When the SCREEN is erased, the MINI-MCB BOARD A1 sends the “Erase Done” status
to the IEB BOARD A2.
Loading...
+ 83 hidden pages