Keithley 2100 Datasheet

Page 1
2100
High precision 6½-digit DMM
for critical measurements at a 5½-digit price
11 measurement functions
cover most commonly measured parameters
Fully specified accuracies on
all functions for ISO-compliant results
TMC compliant USB 2.0
interface for use with SCPI test programs
Included PC software utilities
for graphing and data sharing in both Microsoft® Word and Excel
Rugged construction for
durability in bench/portable applications
Selectable front/rear inputs
facilitate bench or rack use
Includes all accessories, such
as startup software, USB cable, power cable, and safety test leads, for lowest total cost
CE compliant
The Model 2100 USB Digital Multimeter is the newest member of Keithley’s family of high performance DMMs. Its high accuracy (38ppm), 6½-digit resolution is ideal for critical measure­ments. The Model 2100 features 11 measurement functions and 8 math functions to easily accom­modate the most commonly measured param­eters. All accessories, such as USB cable, probes, and software, are included with the Model 2100. With its unique combination of high precision and low total cost of ownership, the Model 2100 is an unbeatable value for R&D engineers, test engineers, scientists, and students making basic precision measurements on the bench and in system applications.
High Precision, Low Cost
The Model 2100 provides stability, accuracy, and
speed at a very low cost. It has 0.0038% 1-year basic DC voltage accuracy on the 10V range and 0.013% 1-year basic resistance accuracy on the 10kW range. At 6.5 digits, the Model 2100 delivers 50 triggered rdgs/s via the USB remote interface. At the fast 4.5 digit setting, it reads over 2000 rdgs/s into its 2000 reading internal buffer.
The Model 2100 provides a wide number of measurement ranges and functions:
• DC voltage: 0.1V, 1V, 10V, 100V, and 1000V
• AC voltage: 0.1V, 1V, 10V, 100V, and 750V
• DC current: 10mA, 100mA, 1A, and 3A
• AC current: 1A and 3A
• Two- and four-wire resistance: 100W, 1kW, 10kW, 100kW, 1MW, 10MW, and 100M
• Frequency: From 3Hz to 300kHz
• Period measurement
• Diode measurement
• Programmable A-D converter and filter settings for signal to noise optimization
Additionally, eight mathematical operations can be performed on measurement readings: RATIO, %, Min/Max, NULL, Limits, mX+b, dB, and dBm testing. Microsoft allow remote storage and recall of the measured values from these applications. A graphing utility enables charting of measurements versus time for trending and noise observations.
TMC compliant USB remote interface enables control from a PC for consistent test/calibration proce­dure execution and easy re-use of existing SCPI programs, including Agilent Model 34401A command emulation.
Simple to Use
The Model 2100 can be setup quickly and is very easy to use. It has a high contrast front panel and keypad that are intuitive and user-friendly. An easy to read 5×7 dot matrix, vacuum fluorescent dis­play (VFD) offers three-color annunciators so users can easily distinguish each function symbol by its color.
Strength and Versatility
With its rugged construction and rubber bumpers, the Model 2100 has the durability to withstand bench, portable, or stacking applications. A sturdy carrying handle facilitates transportability.
®
Office, Word, and Excel add-in tools
W
Low cost 6½-digit DMM for portable, bench, and USB-based system applications
1.888.KEITHLEY
www.keithley.com
(U.S. only)
DIGITAL MULTIMETERS & SYSTEMS
A G R E A T E R M E A S U R E O F C O N F I D E N C E
Page 2
2100
Ordering Information
2100/120 6½-digit USB
Digital Multimeter
Extended warranty, service, and calibration contracts are available.
Accessories Supplied
Instruction manual on CD, Specifications, LabVIEWTM Driver, Keithley I/O Layer, USB Cable, Power Cable, Safety Test Leads, KI-Tool and KI­Link Add-in (Both Microsoft Word and Excel versions)
ACCESSORIES AVAILABLE
RACK MOUNT KITS
4299-3 Single Rack Mount Kit 4299-4 Dual Rack Mount Kit 8605 High Performance Modular Test Leads 8606 High Performance Modular Probe Kit
Applications
The Model 2100 USB Digital Multimeter is ideal for applications in electronic device, circuit, module, and product testing; low cost production testing of electrical and electronic components, sub-assem­blies, and end products; and student lab assignments. Typical applications include:
• Test Engineers: Manual and semi-automatic electrical functional test
• Development Engineers: Electrical/electronic circuit and product validation
• Service/Calibration Technicians: Electronic product repair and calibration
• Research Scientists: Electrical and physics experiments testing
• Engineering Students: Electronic device and circuits experiment testing
Specifications
DC CHARACTERISTICS: ACCURACY1 ±(% of reading + % of range)
Function Range Resolution Input Resistance 1 Year, 23°C ±5°C
100.0000 mV 0.1 µV >10 G
1.000000 V 1.0 µV >10 G
DC Voltage
Function Range Resolution Shunt Resistance 1 Year
DCI (DC Current)
10.00000 V 10 µV >10 G
100.0000 V 100 µV 10 M
1000.000 V 1 mV 10 M
10.00000 mA10.00000 mA mAmA 10 nA 5.1
100.0000 mA 100 nA 5.1
1.000000 A1.0000 00 A AA 1 µA 0.1
3.00000 A 10 µA 0.1
W
W
W
W
W
W
W
W
W
0.0055 + 0.0 040
0.0045 + 0.0 008
0.0038 + 0.0006
0.0050 + 0.0007
0.0055 + 0.0 010
0.055 + 0.025
0.055 + 0.006
0.120 + 0.015
0.150 + 0.025
Function Range Resolution Test Current 1 Year
100.0000
1.000000 k
2
Resista nce
Low cost 6½-digit DMM for portable, bench, and USB-based system applications
Diode Test
Continuity
DC NOTES:
1. Specif ications va lid after t wo hour war m-up. a. AD C set for contin uous trigger operation. b. Input bias current <30p A at 25°C. c. Input protection 100 0V all ra nges (2W input). d. Measureme nt rate set to 1 PLC.
2. Specif ications for 4 W ohms mode. For 2 W ohms, use ze ro null or subs tract lead r esistance fr om displayed r eading. a. Ma ximum lead re sistance 10% of range per lead for 100W and 1kW range s; add 1kW per lea d for all ot her ranges.
10.00000 k
100.0000 k
1.000000 M
10.00000 M
100.0000 MW 100
1.0000 V 10 µV 1 mA1 m A mAmA 0.040 + 0.020
1000.00
MEASUREMENT NOISE REJECTION DC (60Hz/50Hz)
Rate Digits CMRR1 NMRR2
10PLC 140 dB 60 dB
1PLC 140 dB 60 dB
1. For 1kW unba lance in LO le ad.
2. For line frequency ± 0 .1%.
W
100 µ
W
1 m
W
10 m
W
100 m
W
1
W
10
W
10 m
W
1 mA 0.015 + 0.005
W
1 mA 0.015 + 0.002
W
100 µA 0.013 + 0.002
W
10 µA10 µA µAµA 0.015 + 0.002
W
5 µA5 µA µAµA 0.017 + 0.002
W
500 n A 0.045 + 0.002
W
500 nA��10 M500 nA��10 M nA��10 MnA��10 M
W
1 mA 0.024 + 0.030
W
1.00 + 0.020
TEMPERATURE (RTD)
Range Resolution
–100°C to +100°C 0.001°C ±0.1°C –200°C to +630°C 0.001°C ±0.2°C
RTD T YPE: 100W platinum (PT100), D100, F100, PT385, or
PT3916.
MAXIMUM LEAD RESISTANCE (each lead): 12W (to achieve
rated accuracy).
SENSOR CURRENT: 1mA (pulsed).
1. Excludi ng probe erro rs. 23°C ±5°C.
4-Wire Accuracy1,
1 Year
1.888.KEITHLEY
DIGITAL MULTIMETERS & SYSTEMS
www.keithley.com
(U.S. only)
A G R E A T E R M E A S U R E O F C O N F I D E N C E
Page 3
2100
AC CHARACTERISTICS: ACCURACY1 ±(% of reading + % of range)
Function Range
Frequency and Period
Function Range Resolution
ACV (AC TRMS Voltage)
ACI (AC TRMS Current)
100 mV to 750 V
100.0000 mV 0.1 µV
1.000000 V to
2
750.000 V
1.000000 A 1 µA
3.000000 A 10 µA
2
1.0 µV to
1 mV
Frequency
(Hz)
3–5 0.10 5–40 0.05
40–300k 0.01
Frequency
(Hz)
3 – 5 1.15 + 0.05 5 – 10 0.45 + 0.05
10 – 20k 0.08 + 0.05 20k – 50k 0.15 + 0.06 50k – 100k 0.70 + 0.09
100k – 300k 4.25 + 0.60
3 – 5 1.10 + 0.04 5 – 10 0.4 + 0.04
10 – 20k 0.08 + 0.04 20k – 50k 0.14 + 0.06 50k – 100k 0.70 + 0.08
100k – 300k 4.35 + 0.50
3 – 5 1.10 + 0.05 5 – 10 0.40 + 0.05
10 – 5k 0.15 + 0.05
3 – 5 1.25 + 0.07 5 – 10 0.45 + 0.07
10 – 5k 0.20 + 0.07
1 Year
±(% of reading)
23°C ±5°C
1 Year
23°C ±5°C
GENERAL
AC CMRR: 70dB (for 1kW unbalance LO lead). POWER SUPPLY: 120V/220V/240V. POWER LINE FREQUENCY: 50/60Hz auto detected. POWER CONSUMPTION: 25VA max. DIGITAL I/O INTER FACE: USB-compatible Type B connection. ENVIRONMENT: For indoor use only. OPERATING TEMPERATURE: 5° to 40°C. OPERATING HUMIDITY: Maximum relative humidity 80% for temperature up to 31°C,
decreasing linearly to 50% relative humidity at 40°C.
STORAGE TEMPERATURE: –25° to 65°C. OPERATING ALTIT UDE: Up to 2000m above sea level. BENCH DIMENSIONS (with handles and feet): 112mm high × 256mm wide × 375mm112mm high × 256mm wide × 375mm
deep (4.4 in. × 10.1 in. × 14.75 in.).
WEIGHT: 4.1kg (9 lbs.). SAFETY: Conforms to European Union Directive 73/23/ECC, EN61010-1. EMC: Conforms to European Union Directive 89/336/EEC, EN61326-1. WARR ANTY: One year.
Model 2100 specifications
AC NOTES:
1. Specif ications va lid for two hou r warm-up at 6½ digits. a. Slow AC filter (3Hz b andwidth).
b. Pure si ne wave input gr eater than 5% o f range.
2. 750VAC range is limited t o 100kHz.
Model 2100 rear panel
1.888.KEITHLEY
www.keithley.com
(U.S. only)
DIGITAL MULTIMETERS & SYSTEMS
A G R E A T E R M E A S U R E O F C O N F I D E N C E
Page 4
2100
Startup Software, PC utilities Included
KI-Tool application provides charting and graphing capabilities with­out programming to simplify setup, checkout, and basic measurement applications requiring graphical data representation. Scale, offset, and level can be adjusted to fine tune images for visual evaluation of sig­nal and noise elements over time. Also includes tabular data and SCPI command prompt windows for maximum flexibility. Data sets can also be saved to disk files.
Microsoft Excel Add-In utility is also included and provides quick data import into standard Microsoft Excel spreadsheet, including selectable graphing, instrument settings, and number of data points collected. Data can then be analyzed through standard or options Microsoft Excel functions, including graphical, statistical, and trend charting. A version supporting Microsoft Word is also included for direct data import into reports.
Low cost 6½-digit DMM for portable, bench, and USB-based system applications
KEITHLEY INSTRU MEN TS, INC . 28775 AU ROR A ROAD CLEVEL AND, OH IO 44139 -1891 44 0-24 8-0 400 Fax: 440-248- 6168 1-888- KEI THL EY www.k eithley.com
BE LGIUM
Sint- Pieters- Leeuw Ph : 02-363 00 40 Fax: 02-363 0 0 6 4 www. kei thley.nl
ITALY
Mi lan o Ph: 02-553842.1 Fax: 02-55384228 www. kei thley.it
CH IN A
Be iji ng Ph : 8610-82255010 Fax: 8610- 82255018 www. kei thley.co m.c n
JA PAN
Tok yo Ph: 81-3-5733-7555 Fax: 81-3-5733-7556 www. kei thley.jp
SW ED EN
So lna Ph: 08-50 90 46 00 Fax: 08-655 26 10 www. kei thley.co m
All Kei thley trademarks and trade name s are the prop erty of Keithley Instrume nts, Inc .
All other trademarks and trade names are the proper ty of their respec tive companies.
A G R E A T E R M E A S U R E O F C O N F I D E N C E
Specifications are subject to change withou t notice.
FI NL AND
Es poo Ph : 09-88171661 Fax: 09- 8817166 2 www. kei thley.co m
KOREA
Se oul Ph: 82-2-574-7778 Fax: 82-2-574-7838 www. kei thley.co .kr
SW IT ZERLAND
Zü ric h Ph: 044 -821 94 44 Fax: 41-44-820 30 81 www. kei thley.ch
FR ANC E
Sa int -Aubin Ph : 01- 64 53 20 20 Fax: 01-60 -11-77-26 www. kei thley.fr
MA LAYSI A
Kuala Lumpur Ph : 60-3 -4 041-0899 Fax: 60- 3-4042-0 899 www. kei thley.co m
TAIWA N
Hs inc hu Ph: 886-3-572-9077 Fax: 886-3-572-9031 www. kei thley.co m.t w
GE RM AN Y
Ge rme ri ng Ph : 089- 84 93 07-4 0 Fax: 089 -84 93 07-3 4 www. kei thley.de
NE TH ERLANDS
Go rin chem Ph: 0183-63 53 33 Fax: 0183-63 08 21 www. kei thley.nl
UN IT ED KINGD OM
Th eal e Ph : 0118- 929 75 0 0 Fax: 0118-929 75 19 www. kei thley.co .uk
IN DI A
Ba nga lor e Ph: 91-80-2677-1071 Fax: 91-80-2677-1076 www. kei thley.co m
SI NG APORE
Singa por e Ph: 65-6747-9077 Fax: 65-6747-2991 www. kei thley.co m.sg
DIGITAL MULTIMETERS & SYSTEMS
© Copyright 2007 Keithley Instruments, Inc. Printed in the U.S.A . No. 2826 July 07
Loading...