Keithley 2015P, 2015 User Manual

Model 2015/2015PTHD Multimeter
User’s Manual
A GREATER MEASURE OF CONFIDENCE
WARRANTY
Keithley Instruments, Inc. warrants this product to be free from defects in material and workmanship for a period of 3 years from date of shipment.
Keithley Instruments, Inc. warrants the following items for 90 days from the date of shipment: probes, cables, rechargeable batteries, diskettes, and documentation.
During the warranty period, we will, at our option, either repair or replace any product that proves to be defective.
To exercise this warranty, write or call your local Keithley representative, or contact Keithley headquarters in Cleveland, Ohio. You will be given prompt assistance and return instructions. Send the product, transportation prepaid, to the indicated service facility. Repairs will be made and the product returned, transportation prepaid. Repaired or replaced products are warranted for the balance of the original warranty period, or at least 90 days.
LIMITATION OF WARRANTY
This warranty does not apply to defects resulting from product modification without Keithley’s express written consent, or misuse of any product or part. This warranty also does not apply to fuses, software, non-rechargeable batteries, damage from battery leakage, or problems arising from normal wear or failure to follow instructions.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR USE. THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND EXCLUSIVE REMEDIES.
NEITHER KEITHLEY INSTRUMENTS, INC. NOR ANY OF ITS EMPLOYEES SHALL BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES ARISING OUT OF THE USE OF ITS INSTRUMENTS AND SOFTWARE EVEN IF KEITHLEY INSTRUMENTS, INC., HAS BEEN ADVISED IN ADVANCE OF THE POSSIBILITY OF SUCH DAMAGES. SUCH EXCLUDED DAM­AGES SHALL INCLUDE, BUT ARE NOT LIMITED TO: COSTS OF REMOVAL AND INSTALLATION, LOSSES SUSTAINED AS THE RESULT OF INJURY TO ANY PERSON, OR DAMAGE TO PROPERTY.
Keithley Instruments, Inc.
Sales Offices: BELGIUM: Bergensesteenweg 709 • B-1600 Sint-Pieters-Leeuw • 02-363 00 40 • Fax: 02/363 00 64
CHINA: Yuan Chen Xin Building, Room 705 • 12 Yumin Road, Dewai, Madian • Beijing 100029 • 8610-6202-2886 • Fax: 8610-6202-2892 FINLAND: Tietäjäntie 2 • 02130 Espoo • Phone: 09-54 75 08 10 • Fax: 09-25 10 51 00 FRANCE: 3, allée des Garays • 91127 Palaiseau Cédex • 01-64 53 20 20 • Fax: 01-60 11 77 26 GERMANY: Landsberger Strasse 65 • 82110 Germering • 089/84 93 07-40 • Fax: 089/84 93 07-34 GREAT BRITAIN: Unit 2 Commerce Park, Brunel Road • Theale • Berkshire RG7 4AB • 0118 929 7500 • Fax: 0118 929 7519 INDIA: Flat 2B, Willocrissa • 14, Rest House Crescent • Bangalore 560 001 • 91-80-509-1320/21 • Fax: 91-80-509-1322 ITALY: Viale San Gimignano, 38 • 20146 Milano • 02-48 39 16 01 • Fax: 02-48 30 22 74 KOREA: FL., URI Building • 2-14 Yangjae-Dong • Seocho-Gu, Seoul 137-130 • 82-2-574-7778 • Fax: 82-2-574-7838 NETHERLANDS: Postbus 559 • 4200 AN Gorinchem • 0183-635333 • Fax: 0183-630821 SWEDEN: c/o Regus Business Centre • Frosundaviks Allé 15, 4tr • 169 70 Solna • 08-509 04 679 • Fax: 08-655 26 10 SWITZERLAND: Kriesbachstrasse 4 • 8600 Dübendorf • 01-821 94 44 • Fax: 01-820 30 81 TAIWAN: 1FL., 85 Po Ai Street • Hsinchu, Taiwan, R.O.C. • 886-3-572-9077• Fax: 886-3-572-9031
28775 Aurora Road • Cleveland, Ohio 44139 • 440-248-0400 • Fax: 440-248-6168
1-888-KEITHLEY (534-8453) • www.keithley.com
© Copyright 2001 Keithley Instruments, Inc.
Printed in the U.S.A.
11/01
Model 2015/2015P THD Multimeter
User’s Manual
When this manual refers to the Model 2015, it is also referring to the Model 2015P. Information that pertains exclusively to the Model 2015P will be clearly indicated.
©1998, Keithley Instruments, Inc.
All rights reserved.
Cleveland, Ohio, U.S.A.
Fifth Printing, December 2001
Document Number: 2015-900-01 Rev. E
Manual Print History
The print history shown below lists the printing dates of all Revisions and Addenda created for this manual. The Revision Level letter increases alphabetically as the manual undergoes sub­sequent updates. Addenda, which are released between Revisions, contain important change in­formation that the user should incorporate immediately into the manual. Addenda are numbered sequentially. When a new Revision is created, all Addenda associated with the previous Revision of the manual are incorporated into the new Revision of the manual. Each new Revision includes a revised copy of this print history page.
Revision A (Document Number 2015-900-01)................................................................. May 1998
Revision B (Document Number 2015-900-01) ............................................................. August 1998
Revision C (Document Number 2015-900-01) ................................................................. June 1999
Revision D (Document Number 2015-900-01)............................................................October 2000
Revision E (Document Number 2015-900-01) ........................................................December 2001
All Keithley product names are trademarks or registered trademarks of Keithley Instruments, Inc. Other brand names are trademarks or registered trademarks of their respective holders.
S
afety Precautions
The following safety precautions should be observed before using this product and any associated instrumentation. Although some instruments and accessories would normally be used with non-hazardous voltages, there are situations where hazardous conditions may be present.
This product is intended for use by qualified personnel who recognize shock hazards and are familiar with the safety precautions required to avoid possible injury. Read and follow all installation, operation, and maintenance information carefully before us­ing the product. Refer to the manual for complete product specifications.
If the product is used in a manner not specified, the protection provided by the product may be impaired.
The types of product users are:
Responsible body
ment is operated within its specifications and operating limits, and for ensuring that operators are adequately trained.
Operators
instrument. They must be protected from electric shock and contact with hazardous live circuits.
Maintenance personnel
voltage or replacing consumable materials. Maintenance procedures are described in the manual. The procedures explicitly state if the operator may perform them. Otherwise, they should be performed only by service personnel.
Service personnel
trained service personnel may perform installation and service procedures.
Keithley products are designed for use with electrical signals that are rated Installation Category I and Installation Category II, as described in the International Electrotechnical Commission (IEC) Standard IEC 60664. Most measurement, control, and data I/O signals are Installation Category I and must not be directly connected to mains voltage or to voltage sources with high tran­sient over-voltages. Installation Category II connections require protection for high transient over-voltages often associated with local AC mains connections. Assume all measurement, control, and data I/O connections are for connection to Category I sourc­es unless otherwise marked or described in the Manual.
Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable connector jacks or test fixtures. The American National Standards Institute (ANSI) states that a shock hazard exists when voltage levels greater than 30V RMS,
42.4V peak, or 60VDC are present.
circuit before measuring.
Operators of this product must be protected from electric shock at all times. The responsible body must ensure that operators are prevented access and/or insulated from every connection point. In some cases, connections must be exposed to potential human contact. Product operators in these circumstances must be trained to protect themselves from the risk of electric shock. If the circuit is capable of operating at or above 1000 volts,
Do not connect switching cards directly to unlimited power circuits. They are intended to be used with impedance limited sourc­es. NEVER connect switching cards directly to AC mains. When connecting sources to switching cards, install protective de­vices to limit fault current and voltage to the card.
Before operating an instrument, make sure the line cord is connected to a properly grounded power receptacle. Inspect the con­necting cables, test leads, and jumpers for possible wear, cracks, or breaks before each use.
When installing equipment where access to the main power cord is restricted, such as rack mounting, a separate main input pow­er disconnect device must be provided, in close proximity to the equipment and within easy reach of the operator.
For maximum safety, do not touch the product, test cables, or any other instruments while power is applied to the circuit under test. ALWAYS remove power from the entire test system and discharge any capacitors before: connecting or disconnecting ca­bles or jumpers, installing or removing switching cards, or making internal changes, such as installing or removing jumpers.
is the individual or group responsible for the use and maintenance of equipment, for ensuring that the equip-
use the product for its intended function. They must be trained in electrical safety procedures and proper use of the
perform routine procedures on the product to keep it operating properly, for example, setting the line
are trained to work on live circuits, and perform safe installations and repairs of products. Only properly
A good safety practice is to expect that hazardous voltage is present in any unknown
no conductive part of the circuit may be exposed.
Do not touch any object that could provide a current path to the common side of the circuit under test or power line (earth) ground. Al­ways make measurements with dry hands while standing on a dry, insulated surface capable of withstanding the voltage being measured.
The instrument and accessories must be used in accordance with its specifications and operating instructions or the safety of the equipment may be impaired.
Do not exceed the maximum signal levels of the instruments and accessories, as defined in the specifications and operating in­formation, and as shown on the instrument or test fixture panels, or switching card.
When fuses are used in a product, replace with same type and rating for continued protection against fire hazard.
Chassis connections must only be used as shield connections for measuring circuits, NOT as safety earth ground connections.
If you are using a test fixture, keep the lid closed while power is applied to the device under test. Safe operation requires the use of a lid interlock.
If a screw is present, connect it to safety earth ground using the wire recommended in the user documentation.
!
The symbol on an instrument indicates that the user should refer to the operating instructions located in the manual.
The symbol on an instrument shows that it can source or measure 1000 volts or more, including the combined effect of normal and common mode voltages. Use standard safety precautions to avoid personal contact with these voltages.
The
WARNING
information very carefully before performing the indicated procedure.
The
CAUTION
ranty.
Instrumentation and accessories shall not be connected to humans.
Before performing any maintenance, disconnect the line cord and all test cables.
To maintain protection from electric shock and fire, replacement components in mains circuits, including the power transformer, test leads, and input jacks, must be purchased from Keithley Instruments. Standard fuses, with applicable national safety ap­provals, may be used if the rating and type are the same. Other components that are not safety related may be purchased from other suppliers as long as they are equivalent to the original component. (Note that selected parts should be purchased only through Keithley Instruments to maintain accuracy and functionality of the product.) If you are unsure about the applicability of a replacement component, call a Keithley Instruments office for information.
To clean an instrument, use a damp cloth or mild, water based cleaner. Clean the exterior of the instrument only. Do not apply cleaner directly to the instrument or allow liquids to enter or spill on the instrument. Products that consist of a circuit board with no case or chassis (e.g., data acquisition board for installation into a computer) should never require cleaning if handled accord­ing to instructions. If the board becomes contaminated and operation is affected, the board should be returned to the factory for proper cleaning/servicing.
heading in a manual explains dangers that might result in personal injury or death. Always read the associated
heading in a manual explains hazards that could damage the instrument. Such damage may invalidate the war-
11/01

Table of Contents

1 General Information
Introduction ................................................................................ 1-2
Feature overview ........................................................................ 1-2
Warranty information ................................................................. 1-3
Manual addenda ......................................................................... 1-3
Safety symbols and terms .......................................................... 1-3
Specifications ............................................................................. 1-4
Inspection ................................................................................... 1-4
Options and accessories ............................................................. 1-5
General purpose probes ....................................................... 1-5
Low thermal probes ............................................................ 1-5
Cables and adapters ............................................................. 1-6
Rack mount kits .................................................................. 1-6
Carrying case ...................................................................... 1-6
2 Basic Measurements
Introduction ................................................................................ 2-2
Front panel summary ................................................................. 2-3
Rear panel summary ................................................................... 2-6
Power-up .................................................................................... 2-8
Line power connection ........................................................ 2-8
Setting line voltage and replacing fuse ............................... 2-9
Power-up sequence ........................................................... 2-10
High energy circuit safety precautions ............................. 2-11
Power-on defaults .............................................................. 2-12
GPIB primary address ....................................................... 2-16
Warm-up time ................................................................... 2-16
Display ..................................................................................... 2-16
Status and error messages ................................................. 2-16
Measuring voltage .................................................................... 2-17
Connections ....................................................................... 2-17
Crest factor ........................................................................ 2-17
Low level considerations ................................................... 2-18
Measuring current .................................................................... 2-21
Connections ....................................................................... 2-21
AMPS fuse replacement ................................................... 2-22
Measuring resistance ................................................................ 2-23
Connections ....................................................................... 2-23
Shielding ........................................................................... 2-24
Measuring frequency and period .............................................. 2-25
Trigger level ...................................................................... 2-25
Gate time ........................................................................... 2-25
Connections ....................................................................... 2-26
Measuring temperature ............................................................. 2-27
Connections ....................................................................... 2-27
Configuration ..................................................................... 2-28
Math .......................................................................................... 2-29
MX + B .............................................................................. 2-29
Percent ............................................................................... 2-30
dBm calculation ................................................................. 2-31
dB calculation .................................................................... 2-32
Measuring continuity ................................................................ 2-33
Connections ....................................................................... 2-33
Threshold resistance level ................................................. 2-33
Testing diodes ........................................................................... 2-34
Connections ....................................................................... 2-34
Range ................................................................................. 2-34
Measuring distortion ................................................................. 2-35
Configuration ..................................................................... 2-36
Connections ....................................................................... 2-41
Measurement examples ..................................................... 2-43
Analyzing frequency spectrum (Model 2015P only) ............... 2-48
Peak commands ................................................................. 2-48
Operation overview ........................................................... 2-49
Typical peak analysis ......................................................... 2-49
Frequency list analysis ...................................................... 2-50
Peak analysis programming example ................................ 2-50
Delta programming example ............................................. 2-52
Frequency list programming example ............................... 2-54
Measurement Queries ............................................................... 2-55
:FETCh? ............................................................................ 2-55
:READ? ............................................................................. 2-55
:MEAS:[function]? ............................................................ 2-56
[:SENSe[1]]:DATA:FRESh? .............................................. 2-56
[:SENSe[1]]:DATA:LATest? .............................................. 2-57
3 Measurement Options
Introduction ................................................................................ 3-2
Measurement configuration ........................................................ 3-3
Range ................................................................................... 3-3
Filter .................................................................................... 3-4
Relative ................................................................................ 3-5
Digits ................................................................................... 3-5
Rate ...................................................................................... 3-6
Trigger operations ...................................................................... 3-8
Trigger model ...................................................................... 3-8
Reading hold (autosettle) .................................................. 3-10
External triggering ............................................................ 3-11
Buffer operations ...................................................................... 3-17
Storing readings ................................................................ 3-17
Recalling readings ............................................................. 3-18
Buffer statistics ................................................................. 3-19
Limit operations ....................................................................... 3-20
Setting limit values ........................................................... 3-20
Enabling limits .................................................................. 3-21
Scan operations ........................................................................ 3-22
Scanning overview ............................................................ 3-22
Front panel scanner controls ............................................. 3-22
Stepping and scanning trigger model additions ................ 3-22
Using SHIFT-CONFIG to configure stepping
and scanning ................................................................ 3-24
Scanning examples ............................................................ 3-25
System operations .................................................................... 3-27
Calibration ......................................................................... 3-27
4 Remote Operation
Introduction................................................................................. 4-2
Selecting an interface .......................................................... 4-2
RS-232 ................................................................................ 4-3
GPIB bus ............................................................................. 4-3
RS-232 operation ....................................................................... 4-4
Sending and receiving data ................................................. 4-4
Selecting baud rate .............................................................. 4-4
Selecting signal handshaking (flow control) ....................... 4-5
Setting terminator ................................................................ 4-5
RS-232 connections ............................................................ 4-6
Error messages .................................................................... 4-6
GPIB bus operation and reference ............................................. 4-7
Introduction ......................................................................... 4-7
GPIB bus standards ............................................................. 4-7
GPIB bus connections ......................................................... 4-8
Selecting the primary address ........................................... 4-10
QuickBASIC 4.5 programming ........................................ 4-10
General Bus Commands ................................................... 4-12
Front panel GPIB operation .............................................. 4-16
Status structure ......................................................................... 4-17
Condition registers ............................................................ 4-18
Event registers ................................................................... 4-18
Enable registers .................................................................. 4-19
Queues ............................................................................... 4-22
Status Byte and Service Request (SRQ) ........................... 4-23
Trigger model (GPIB operation) .............................................. 4-27
Idle and initiate .................................................................. 4-28
Trigger model operation .................................................... 4-28
Programming syntax ................................................................. 4-30
Command words ............................................................... 4-30
Query commands ............................................................... 4-32
Case sensitivity .................................................................. 4-32
Long-form and short-form versions .................................. 4-32
Short-form rules ................................................................ 4-33
Program messages ............................................................. 4-34
Response messages ........................................................... 4-36
Message exchange protocol ............................................... 4-36
Common Commands ................................................................ 4-37
5 SCPI Command Reference
SCPI Signal oriented measurement commands .......................... 5-3
SCPI command subsystems reference tables ............................. 5-7
Calculate subsystem ................................................................. 5-23
DISPlay subsystem ................................................................... 5-29
:FORMat subsystem ................................................................. 5-31
OUTPut subsystem ................................................................... 5-36
ROUTe subsystem .................................................................... 5-38
[SENSe[1]] subsystem .............................................................. 5-39
STATus subsystem .................................................................... 5-59
:SYSTem subsystem ................................................................. 5-67
:TRACe subsystem ................................................................... 5-72
Trigger subsystem ..................................................................... 5-74
:UNIT subsystem ...................................................................... 5-77
A Specifications
B Status and Error Messages
C Example Programs
Program examples ..................................................................... C-2
Changing function and range ............................................. C-2
One-shot triggering ............................................................ C-4
Generating SRQ on buffer full ........................................... C-5
Storing readings in buffer .................................................. C-6
Taking readings using the :READ? command ................... C-8
Controlling the Model 2015 via the RS-232 COM2 port .. C-8
D IEEE-488 Bus Overview
Introduction ............................................................................... D-2
Bus description .......................................................................... D-4
Bus lines .................................................................................... D-6
Data lines ........................................................................... D-6
Bus management lines ....................................................... D-6
Handshake lines ................................................................. D-7
Bus commands .......................................................................... D-8
Uniline commands ............................................................. D-9
Universal multiline commands .......................................... D-9
Addressed multiline commands ....................................... D-10
Address commands .......................................................... D-10
Unaddress commands ...................................................... D-10
Common commands ........................................................ D-11
SCPI commands ............................................................... D-11
Command codes ............................................................... D-11
Typical command sequences ............................................ D-13
IEEE command groups .................................................... D-14
Interface function codes .......................................................... D-15
E IEEE-488 and SCPI Conformance Information
Introduction ................................................................................ E-2

List of Illustrations

2 Basic Measurements
Figure 2-1 Model 2015 front panel .......................................................... 2-3
Figure 2-2 Model 2015 rear panel ........................................................... 2-6
Figure 2-3 Power module ........................................................................ 2-8
Figure 2-4 DC and AC voltage measurements ...................................... 2-18
Figure 2-5 DC and AC current measurements ....................................... 2-21
Figure 2-6 Two- and four- wire resistance measurements ..................... 2-24
Figure 2-7 Frequency and period measurements ................................... 2-26
Figure 2-8 Thermocouple temperature measurements .......................... 2-27
Figure 2-9 Continuity measurements .................................................... 2-33
Figure 2-10 Diode testing ........................................................................ 2-34
Figure 2-11 Distortion measurements ..................................................... 2-42
Figure 2-12 DUT connections when using the internal sine source
to stimulate a device ....................................................... 2-47
Figure 2-13 Peak analysis ........................................................................ 2-49
3 Measurement Options
Figure 3-1 Moving average and repeating filters ..................................... 3-4
Figure 3-2 Front panel triggering without stepping/scanning ................. 3-8
Figure 3-3 Rear panel pinout ................................................................. 3-11
Figure 3-4 Trigger link input pulse specifications (EXT TRIG) ........... 3-12
Figure 3-5 Trigger link output pulse specifications (VMC) .................. 3-12
Figure 3-6 DUT test system .................................................................. 3-13
Figure 3-7 Trigger link connections ...................................................... 3-13
Figure 3-8 Operation model for triggering example ............................. 3-14
Figure 3-9 DIN to BNC trigger cable .................................................... 3-16
Figure 3-10 Buffer locations .................................................................... 3-18
Figure 3-11 Using limit test to sort 100W, 10% resistors ........................ 3-21
Figure 3-12 Front panel triggering with stepping .................................... 3-23
Figure 3-13 Front panel triggering with scanning ................................... 3-23
Figure 3-14 External scanning example with Model 7001 ...................... 3-26
4 Remote Operation
Figure 4-1 RS-232 interface connector ................................................... 4-6
Figure 4-2 IEEE-488 connector ............................................................... 4-8
Figure 4-3 IEEE-488 connections ........................................................... 4-8
Figure 4-4 IEEE-488 connector location ................................................. 4-9
Figure 4-5 Model 2015 status register structure .................................... 4-17
Figure 4-6 Standard event status ............................................................ 4-20
Figure 4-7 Operation event status .......................................................... 4-20
Figure 4-8 Measurement event status .................................................... 4-21
Figure 4-9 Questionable event status ..................................................... 4-21
Figure 4-10 Status byte and service request (SRQ) ................................. 4-23
Figure 4-11 Trigger model (remote operation) ........................................ 4-27
Figure 4-12 Device action (trigger model) .............................................. 4-29
Figure 4-13 Standard event enable register .............................................. 4-39
Figure 4-14 Standard event status register ............................................... 4-41
Figure 4-15 Service request enable register ............................................. 4-47
Figure 4-16 Status byte register ............................................................... 4-49
5 SCPI Command Reference
Figure 5-1 ASCII data format ................................................................ 5-32
Figure 5-2 IEEE754 single precision data format (32 data bits) ........... 5-32
Figure 5-3 IEEE754 double precision data format (64 data bits) .......... 5-33
Figure 5-4 Measurement event register .................................................. 5-60
Figure 5-5 Questionable event register .................................................. 5-61
Figure 5-6 Operation event register ....................................................... 5-61
Figure 5-7 Measurement event enable register ...................................... 5-63
Figure 5-8 Questionable event enable register ....................................... 5-63
Figure 5-9 Operation event enable register ............................................ 5-63
Figure 5-10 Key-press codes .................................................................... 5-71
D IEEE-488 Bus Overview
Figure D-1 IEEE-488 bus configuration .................................................. D-5
Figure D-2 IEEE-488 handshake sequence ............................................. D-7
Figure D-3 Command codes .................................................................. D-12

List of Tables

2 Basic Measurements
Table 2-1 Fuse ratings ............................................................................ 2-9
Table 2-2 Factory defaults .................................................................... 2-13
Table 2-3 Crest factor limitations ......................................................... 2-17
Table 2-4 PEAK commands ................................................................. 2-48
Table 2-5 Peak analysis programming example .................................. 2-51
Table 2-6 Delta programming example ............................................... 2-53
Table 2-7 Frequency list programming example ................................. 2-54
3 Measurement Options
Table 3-1 Rate settings for the measurement functions ......................... 3-7
Table 3-2 Auto delay settings ................................................................ 3-9
4 Remote Operation
Table 4-1 RS-232 connector pinout ....................................................... 4-6
Table 4-2 General bus commands and associated statements .............. 4-12
Table 4-3 IEEE-488.2 common commands and queries ...................... 4-37
5 SCPI Command Reference
Table 5-1 Signal oriented measurement command summary ................ 5-3
Table 5-2 CALCulate command summary ............................................ 5-8
Table 5-3 DISPlay command summary ................................................. 5-9
Table 5-4 FORMat command summary ............................................... 5-10
Table 5-5 OUTPut command summary ............................................... 5-10
Table 5-6 ROUTe command summary ................................................. 5-11
Table 5-7 SENSe command summary ................................................. 5-11
Table 5-8 STATus command summary ................................................ 5-19
Table 5-9 SYSTem command summary .............................................. 5-20
Table 5-10 TRACe command summary ................................................ 5-20
Table 5-11 Trigger command summary ................................................. 5-21
Table 5-12 UNIT command summary ................................................... 5-22
B Status and Error Messages
Table B-1 Status and error messages ...................................................... B-2
D IEEE-488 Bus Overview
Table D-1 IEEE-488 bus command summary ....................................... D-8
Table D-2 Hexadecimal and decimal command codes ........................ D-11
Table D-3 Typical addressed command sequence ............................... D-13
Table D-4 Typical addressed command sequence ................................ D-13
Table D-5 IEEE command groups ....................................................... D-14
Table D-6 Model 2015 interface function codes .................................. D-15
E IEEE-488 and SCPI Conformance Information
Table E-1 IEEE-488 documentation requirements ................................ E-2
Table E-2 Coupled commands ............................................................... E-3
1
General
Information
1-2 General Information

Introduction

NOTES
mation is organized as follows:
The Models 2015 and 2015P are almost identical in function. The exception is that the Model 2015P has the capability to analyze the frequency spectrum of a signal, while the Model 2015 does not.
Over the bus, the *IDN? command can be used to identify the model number of the unit. The response message will be either Model 2015 or Model 2015P (see “Com­mon commands” in Section 4).
When this manual refers to the Model 2015, it is also referring to the Model 2015P. Information that pertains exclusively to the Model 2015P will be clearly indicated.
This section contains general information about the Model 2015 THD Multimeter. The infor-
Feature overview
Warranty information
Manual addenda
Safety symbols and terms
Specifications
Inspection
Options and accessories
If you have any questions after reviewing this information, please contact your local Keithley representative or call one of our Applications Engineers at 1-800-348-3735 (U.S. and Canada only). Worldwide phone numbers are listed at the front of this manual.

Feature overview

The Model 2015 is a 6½-digit high-performance digital multimeter. It has 0.002% 90-day basic DC voltage accuracy and 0.008% 90-day basic resistance accuracy. At 6 timeter delivers 50 triggered readings/sec over the IEEE-488 bus. At 4 2000 readings/sec into its internal buffer. The Model 2015 has broad measurement ranges:
THD, THD+n, and SINAD from 20Hz to 50kHz with 0.0001% (0.00001 dB) resolution.
Frequency spectrum analysis from 20Hz to 50kHz (Model 2015P only).
DC voltage from 0.1
AC (RMS) voltage from 0.1
DC current from 10nA to 3A.
AC (RMS) current from 1
Two and four-wire resistance from 100µ
Frequency from 3Hz to 500kHz.
Thermocouple temperature from -200°C to +1372°C.
µ
V to 1000V.
µ
µ
A to 3A.
V to 750V, 1000V peak.
to 120MΩ.
½
-digits, the mul-
½
-digits, it can read up to
Some additional capabilities of the Model 2015 include:
Full range of functions — In addition to those listed above, the Model 2015 functions include period, dB, dBm, continuity, diode testing, mX+b, and percent.
Optional scanning — For external scanning, the Model 2015 is compatible with Kei­thley's Model 7001 and 7002 switch matrices and cards.
Programming language and remote interfaces — The Model 2015 has the SCPI pro­gramming language and two remote interface ports (IEEE-488/GPIB and RS-232C).
Reading and setup storage — Up to 1024 readings and two setups (user and factory defaults) can be stored and recalled.
Closed-cover calibration — The instrument can be calibrated either from the front panel or remote interface.

Warranty information

Warranty information is located at the front of this instruction manual. Should your Model 2015 require warranty service, contact the Keithley representative or authorized repair facility in your area for further information. When returning the instrument for repair, be sure to fill out and include the service form at the back of this manual to provide the repair facility with the necessary information.
General Information 1-3

Manual addenda

Any improvements or changes concerning the instrument or manual will be explained in an addendum included with the manual. Be sure to note these changes and incorporate them into the manual.

Safety symbols and terms

The following symbols and terms may be found on the instrument or used in this manual.
!
The symbol on the instrument indicates that the user should refer to the operating instructions located in the manual.
The
symbol
Use standard safety precautions to avoid personal contact with these voltages.
The
WARNING
injury or death. Always read the associated information very carefully before performing the indicated procedure.
The
CAUTION
instrument. Such damage may invalidate the warranty.
on the instrument shows that high voltage may be present on the terminal(s).
heading used in this manual explains dangers that might result in personal
heading used in this manual explains hazards that could damage the
1-4 General Information
Specifications
Full Model 2015 specifications are included in Appendix A.

Inspection

The Model 2015 was carefully inspected electrically and mechanically before shipment. After unpacking all items from the shipping carton, check for any obvious signs of physical damage that may have occurred during transit. (Note: There may be a protective film over the display lens, which can be removed.) Report any damage to the shipping agent immediately. Save the original packing carton for possible future reshipment. The following items are included with every Model 2015 order:
Model 2015 THD Multimeter with line cord.
Safety test leads (Model 1751).
Accessories as ordered.
Certificate of calibration.
Model 2015/2015P User's Manual (P/N 2015-900-00).
Model 2015/2015P Service Manual (P/N 2015-902-00).
If an additional manual is required, order the appropriate manual package. The manual pack­ages include a manual and any pertinent addenda.

Options and accessories

The following options and accessories are available from Keithley for use with the
Model 2015.
General purpose probes
General Information 1-5
Model 1754 Universal Test Lead Kit:
lugs, two banana plugs, two hooks, and two alligator clips.
Model 8605 High Performance Modular Test Leads:
test probes and leads. The test leads are terminated with a banana plug with retractable sheath on each end.
Model 8606 High Performance Probe Tip Kit:
clips, and two spring hook test probes. (The spade lugs and alligator clips are rated at 30V RMS,
42.4V peak; the test probes are rated at 1000V.) These components are for use with high perfor­mance test leads terminated with banana plugs, such as the Model 8605.
The following test leads and probes are rated at 30V RMS, 42.4V peak:
Models 5805 and 5805-12 Kelvin Probes:
with banana plug termination. Designed for instruments that measure 4-terminal resistance. The Model 5805 is 0.9m long; the Model 5805-12 is 3.6m long.
Model 5806 Kelvin Clip Lead Set:
plug termination. Designed for instruments that measure 4-terminal resistance. A set of eight replacement rubber bands is available as Keithley P/N GA-22.
Model 8604 SMD Probe Set:
face mount device “grabber” clip on one end and a banana plug with a retractable sheath on the other end.
Low thermal probes
Consists of one set of test leads (0.9m), two spade
Consists of two high voltage (1000V)
Consists of two spade lugs, two alligator
Consists of two spring-loaded Kelvin test probes
Includes two Kelvin clip test leads (0.9m) with banana
Consists of two test leads (0.9m), each terminated with a sur-
Model 8610 Low Thermal Shorting Plug:
1-inch square circuit board, interconnected to provide a short circuit among all plugs.
Model 8611 Low Thermal Patch Leads:
banana plug with a retractable sheath at each end. These leads minimize the thermally-induced offsets that can be created by test leads.
Model 8612 Low Thermal Spade Leads:
with a spade lug on one end and a banana plug with a retractable sheath on the other end. These leads minimize the thermally-induced offsets that can be created by test leads.
Consists of four banana plugs mounted to a
Consists of two test leads (0.9m), each with a
Consists of two test leads (0.9m), each terminated
1-6 General Information
Cables and adapters
Models 7007-1 and 7007-2 Shielded GPIB Cables:
bus using shielded cables and connectors to reduce electromagnetic interference (EMI). The Model 7007-1 is 1m long; the Model 7007-2 is 2m long.
Models 8501-1 and 8501-2 Trigger Link Cables:
ments with Trigger Link connectors (e.g., Model 7001 Switch System). The Model 8501-1 is 1m long; the Model 8501-2 is 2m long.
Model 8502 Trigger Link Adapter:
of the Model 2015 to instruments that use the standard BNC trigger connectors.
Model 8504 DIN to BNC Trigger Cable:
(Voltmeter Complete) and two (External Trigger) of the Model 2015 to instruments that use BNC trigger connectors. The Model 8504 is 1m long.
Rack mount kits
Model 4288-1 Single Fixed Rack Mount Kit:
19-inch rack.
Model 4288-2 Side-by-Side Rack Mount Kit:
486, 487, 2000, 2001, 2002, 2010, 2015, 6517, 7001) side-by-side in a standard 19-inch rack.
Model 4288-3 Side-by-Side Rack Mount Kit:
side-by-side in a standard 19-inch rack.
Model 4288-4 Side-by-Side Rack Mount Kit:
ment (Models 195A, 196, 220, 224, 230, 263, 595, 614, 617, 705, 740, 775, etc.) side-by-side in a standard 19-inch rack.
Connect the Model 2015 to the GPIB
Connect the Model 2015 to other instru-
Allows you to connect any of the six Trigger Link lines
Allows you to connect Trigger Link lines one
Mounts a single Model 2015 in a standard
Mounts two instruments (Models 182, 428,
Mounts a Model 2015 and a Model 199
Mounts a Model 2015 and a 5.25-inch instru-
Carrying case
Model 1050 Padded Carrying Case:
shoulder strap.
A carrying case for a Model 2015. Includes handles and
2
Basic
Measurements
2-2 Basic Measurements

Introduction

This section summarizes front panel operation of the Model 2015. It is organized as follows:
Front panel summary —
connections.
Rear panel summary —
Power-up —
the warm-up time, and default conditions.
Display —
instrument.
Measuring voltage —
level voltage considerations.
Measuring current —
fuse replacement.
Measuring resistance —
shielding considerations.
Measuring frequency and period —
nections.
Measuring temperature —
surements.
Math —
readings.
Measuring continuity —
Testing diodes —
Measuring distortion
culations used.
Analyzing frequency spectrum (Model 2015P only)
Model 2015P to analyze the frequency spectrum of a signal over the bus.
Includes an illustration and summarizes keys, display, and
Includes an illustration and summarizes connections.
Describes connecting the instrument to line power, the power-up sequence,
Discusses the display format and messages that may appear while using the
Covers DC and AC voltage measurement connections and low
Covers DC and AC current measurement connections and current
Details two and four-wire measurement connections and
Covers frequency and period measurement con-
Describes the use of thermocouples for temperature mea-
Covers the mX+b, percent, dBm, and dB math functions performed on single
Explains setting up and measuring continuity of a circuit.
Describes testing general-purpose and zener diodes.
— Explains how to measure total harmonic distortion and the cal-
— Explains how to use the

Front panel summary

The front panel of the Model 2015 is shown in Figure 2-1. This figure includes important
abbreviated information that should be reviewed before operating the instrument.
Basic Measurements 2-3
Figure 2-1
Model 2015 front panel
SENSE
INPUT
4 WIRE
5
1 3
SHIFT
LOCAL
POWER
SOURCE
1 Function keys
SCAN
CH1REM
STEP CH2 CH3 CH4 CH5 CH6 CH7 CH8 CH9 CH10
TALK LSTN SRQ SHIFT
TIMER
HOLD TRIG FAST MED SLOW AUTO ERR
MATH
THD
DCV
ACV
HOLD
TRIG
EX TRIG
SAVE SETUP
MEAS
THD
dBm
DCI
LIMITS ON/OFFDELAY
STORE
RECALL
CONFIG HALT
STEP SCAN
ACI
dB
REL FILT
CONT
Ω2Ω
TEST
GPIB
DIGITS RATE
(shifted and unshifted)
RS232
CAL
RELFILTER
BUFFER
4
PERIOD TCOUPL
FREQ
MATH REAR
4W
STAT
2015 THD MULTIMETER
TEMP
RANGE
AUTO
EXIT ENTER
RANGE
4
350V PEAK
F
INPUTS
FRONT/REAR
78 2
HI
1000V
!
PEAK
LO
500V PEAK
R
3A 250V
AMPS
Select measurement function (DC and AC voltage, DC and AC current, 2-wire and 4-wire resistance, frequency, period, temperature with thermocouples), math function (mX+b, %, dBm, dB), THD (total harmonic distortion) or special function (continuity, diode test).
2 Operation keys
EXTRIG Selects external triggers (front panel, bus, trigger link) as the trigger
source. TRIG Triggers a measurement from the front panel. STORE Enables reading storage. RECALL Displays stored readings and buffer statistics (maximum, minimum, aver-
age, standard deviation). Use ▲ and ▼ to scroll through buffer; use
and to toggle between reading number and reading. FILTER Displays digital filter status for present function and toggles filter on/off. REL Enables/disables relative reading on present function (not applicable for
distortion measurements).
and Moves through selections within functions and operations. If scanner
card installed, manually scans channels. THD SOURCE Selects and configures the internal function generator. THD MEAS Configures distortion measurements. STEP Steps through channels; sends a trigger after each channel. SCAN Scans through channels; sends a trigger after last channel. DIGITS Changes number of digits of resolution. RATE Changes reading rate: fast, medium, slow. EXIT Cancels selection, moves back to measurement display. ENTER Accepts selection, moves to next choice or back to measurement display. SHIFT Used to access shifted keys. LOCAL Cancels GPIB remote mode.
6
2-4 Basic Measurements
3 Shifted operation keys
DELAY Sets user delay between trigger and measurement. HOLD Holds reading when the selected number of samples is within the selected
LIMITS Sets upper and lower limit values for readings. ON/OFF Enables/disables limits; selects beeper operation for limit testing. TEST Selects built-in tests, diagnostics, display test. CAL Accesses calibration. SAVE Saves present configuration for power-on user default. SETUP Restores factory or user default configuration. CONFIG Selects minimum/maximum channels, timer, and reading count for step/scan. HALT Turns off step/scan. GPIB Enables/disables GPIB interface; selects address and language. RS232 Enables/disables RS-232 interface; selects baud rate, flow control, terminator.
4 Range keys
▲ ▼
AUTO Enables/disables autorange. SHIFT-AUTO displays the most recent error
5 Annunciators
* (asterisk) Reading being stored.
(diode) Instrument is in diode testing function.
)
)
)
(speaker) Beeper on for continuity or limits testing. (more) Indicates additional selections are available. 4W 4-wire resistance reading displayed. AUTO Autoranging enabled. BUFFER Recalling stored readings. ERR Questionable reading; invalid cal step. FAST Fast reading rate. FILT Digital filter enabled. HOLD Instrument is in hold mode. LSTN Instrument addressed to listen over GPIB. MATH Math function (mX+b, %, dB, dBm) enabled. MED Medium reading rate. REAR Reading acquired from rear inputs. REL Relative reading displayed. REM Instrument is in GPIB remote mode. SCAN Instrument is in scan mode. SHIFT Accessing shifted keys. SLOW Slow reading rate. SRQ Service request over GPIB. STAT Displaying buffer statistics. STEP Instrument is in step mode. TALK Instrument addressed to talk over GPIB. TIMER Timed scans in use. TRIG Indicates external trigger (front panel, bus, trigger link) selected.
tolerance.
Moves to higher range; increments digit; moves to next selection. Moves to lower range; decrements digit; moves to previous selection.
message.
Basic Measurements 2-5
6 Input connections
INPUT HI and LO Used for making DC volts, AC volts, 2-wire resistance measurements. AMPS Used in conjunction with INPUT LO to make DC current and AC cur-
SENSE Ω4 WIRE Used with INPUT HI and LO to make 4-wire resistance measure­HI and LO ments.
rent measurements. Also holds current input fuse (3A, 250V, fast blow, 5×20mm).
7 INPUTS
Selects input connections on front or rear panel.
8 Handle
Pull out and rotate to desired position.
2-6 Basic Measurements

Rear panel summary

The rear panel of the Model 2015 is shown in Figure 2-2. This figure includes important
abbreviated information that should be reviewed before operating the instrument.
Figure 2-2
Model 2015 rear panel
23 4
1
HI
!
LO
42V PEAK
INPUT
1000V PEAK
!
INV/PULSE
SOURCE OUTPUT
500V PEAK
350V PEAK
SOURCE OUTPUT
SENSE
4W
7
6
FUSE
500 mAT
(SB)
250 mAT
(SB)
TRIGGER
LINK
!
3 5
1
VMC
4 6
2
EXT TRIG
!
LINE
LINE RATING
100 VAC
50, 60Hz
120 VAC
40VA MAX
220 VAC 240 VAC
8
7
5
2
1
#2
EXTERNAL TRIGGER INPUT
Trigger Reading
>10µsec
TTL HI
TTL LO
MADE IN
U.S.A.
RS232
(CHANGE IEEE ADDRESS
FROM FRONT PANEL)
6
34
#1
VOLT METER COMPLETE OUTPUT
Reading
Complete
>10µsec
TTL HI
TTL LO
IEEE-488
120
5
Basic Measurements 2-7
1 Input connections
INPUT HI and LO Used for making DC volts, AC volts, 2-wire resistance measurements
SENSE
4 WIRE Used with INPUT HI and LO to make 4-wire resistance measurements
and for connecting scanner card.
HI and LO and also for connecting scanner card.
2 TRIGGER LINK
One 8-pin micro-DIN connector for sending and receiving trigger pulses among other instru­ments. Use a trigger link cable or adapter, such as Models 8501-1, 8501-2, 8502, 8504.
3 RS-232
Connector for RS-232 operation. Use a straight-through (not null modem) DB-9 cable.
4 IEEE-488
Connector for IEEE-488 (GPIB) operation. Use a shielded cable, such as Models 7007-1 and 7007-2.
5 Power module
Contains the AC line receptacle, power line fuse, and line voltage setting. The Model 2015 can be configured for line voltages of 100V/120V/220V/240VAC at line frequencies of 45Hz to 66Hz.
6. INV/PULSE SOURCE OUTPUT
A BNC connector that provides an inverted sine wave or pulsed output for exciting devices under test during distortion measurement or for general purpose use. This con­nector also may provide a square wave pulse for triggering or synchronizing other sys­tems to the SOURCE OUTPUT sine wave.
7 SOURCE OUTPUT
A BNC connector that provides a sine wave output for exciting devices under test during dis­tortion measurement or for general purpose use.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
2-8 Basic Measurements

Power-up

Line power connection
Follow the procedure below to connect the Model 2015 to line power and turn on the
instrument.
1. Check to see that the line voltage selected on the rear panel (see Figure 2-3) is correct for the operating voltage in your area. If not, refer to the next procedure, “Setting line voltage and replacing fuse.”
CAUTION Operating the instrument on an incorrect line voltage may cause damage to
2. Before plugging in the power cord, make sure that the front panel power switch is in the off (0) position.
3. Connect the female end of the supplied power cord to the AC receptacle on the rear panel. Connect the other end of the power cord to a grounded AC outlet.
WARNING The power cord supplied with the Model 2015 contains a separate ground
the instrument, possibly voiding the warranty.
wire for use with grounded outlets. When proper connections are made, instrument chassis is connected to power line ground through the ground wire in the power cord. Failure to use a grounded outlet may result in per­sonal injury or death due to electric shock.
4. Turn on the instrument by pressing the front panel power switch to the on (1) position.
Figure 2-3
Power module
350V PEAK
SOURCE OUTPUT
Model 2015
HI
1000V
TRIGGER
PEAK
!
500V PEAK
LO
SENSE
INPUT
4W
42V PEAK
INV/PULSE
SOURCE OUTPUT
3 5
1
4 6
2
!
!
FUSE LINE
100 VAC
500 mAT
120 VAC
(SB)
220 VAC
250 mAT
240 VAC
(SB)
LINK
VMC
EXT TRIG
LINE RATING
50, 60Hz
40VA MAX
MADE IN
U.S.A.
RS232
!
IEEE-488
(CHANGE IEEE ADDRESS
FROM FRONT PANEL)
120
Fuse
Spring
120
Line Voltage Selector
Window
Fuse Holder Assembly
Setting line voltage and replacing fuse
A rear panel fuse located next to the AC receptacle protects the power line input of the instru­ment. If the line voltage setting needs to be changed or the line fuse needs to be replaced, per­form the following steps.
WARNING Make sure the instrument is disconnected from the AC line and other equip-
ment before changing the line voltage setting or replacing the line fuse.
1. Place the tip of a flat-blade screwdriver into the power module by the fuse holder assem­bly (see Figure 2-3). Gently push in and to the left. Release pressure on the assembly and its internal spring will push it out of the power module.
2. Remove the fuse and replace it with the type listed in Table 2-1.
CAUTION For continued protection against fire or instrument damage, only replace
fuse with the type and rating listed. If the instrument repeatedly blows fus­es, locate and correct the cause of the trouble before replacing the fuse. See the optional Model 2015 Service Manual for troubleshooting information.
3. If configuring the instrument for a different line voltage, remove the line voltage selector from the assembly and rotate it to the proper position. When the selector is installed into the fuse holder assembly, the correct line voltage appears inverted in the window.
4. Install the fuse holder assembly into the power module by pushing it in until it locks in place.
Basic Measurements 2-9
Table 2-1
Fuse ratings
Line voltage Fuse rating Keithley P/N
100/120V 220/240V
0.5A, 250V, slo-blo, 5 × 20 mm
0.25A, 250V, slo-blo, 5 × 20 mm
FU-71 FU96-4
Loading...
+ 254 hidden pages